Page 1: Shirley trush analyzer

Welcome to my presentation

Page 2: Shirley trush analyzer

Topic of the presentation:

Shirley Trash Analyzer

Page 3: Shirley trush analyzer

Presented By…

141-23-3756 Habibur Rahman

Page 4: Shirley trush analyzer

Introduction

To determine the percentage of trash, lint and dust content within

a sample of raw cotton, Blow room lap, carding sliver etc. and to

open and clean fibres for further testing.

Page 5: Shirley trush analyzer

Features

50/100 gram samples at the opening rate of 16+ grams lint per minute.

Compact, aerodynamic shape, silent operation and nonpolluting nature boosts laboratory environment.

Digital monitor based user friendly operating system.

Solid-state control electrical system to enhance feasibility an operator safety.

Page 6: Shirley trush analyzer

Safety

While designing this instrument, great care has been taken to minimize the risk of operator injury. However it user disregrated the sefety precautions, then there is a risk of damage.

Attention Power Supply Fuse Rotating Parts Hazard Hot

Page 7: Shirley trush analyzer

Technical SpecificationCapacity / sample size : 50 or 100 gram per test

Measuring principal : Buoyancy separation principal

Opening rate : 16 grams per minute

Micro dust filtering stage : Fine-150 micron; Course-300 micron

Licker-in speed : 1200(±2%) RPM

Feed roller speed : 3.2 ± 0.2 RPM

Balance Accuracy : ±0.002 grams

Control System : Microcomputer based solid state electrical system control

Power supply : Three phase 220v AC @50Hz

Power consumption : ~2000 watts

Applicable standard : ASTM D2812 & ITME

Physical Dimensions : Size(WDH): 620*900*550mm; Weight:180 kg

Page 8: Shirley trush analyzer
Page 9: Shirley trush analyzer

Parts of the trash analyzer

Front door

Start button

Stop button

Feeding tray

Control Unit

Emergency Stop

Page 10: Shirley trush analyzer

Atmospheric & sample condition

Standard practice for conditioning textile for testing. The standard atmosphere having a relative humidity of 65±2% and at a temperature of 27±2ºC must be maintained. For international testing, ISO 139 may be used.

Page 11: Shirley trush analyzer

Installation

Instrument should keep on the flat floor. It is essential that the instrument should be perpendicularly to spirit level. Care should be taken not to expose the instrument to any unnecessary conditions such as vibration, moisture, dust or heat.

Pre-requisition and installation:

Suitable working table

Precision electronic balance.

Stabilized three phase 415v AC @ 50 Hz ported at switched terminal.

Suitable qualified operator and test samples.

Page 12: Shirley trush analyzer

Sample preparation

The test cotton of 50/100 gram mass will be using to determine the trash.

During handling the sample, care must be taken for opening of the cotton and loss of heavy material such as send, metal particles should not be inserted in to the feeding parts.

Before weighing, the sample use must be kept at the standard atmosphere for a minimum of 4 hours.

Use the cotton tray for the weighted cotton at the time of 50 or 100 gram testing.

While opening the cotton use the feeding tray & cotton tray and ensure the trash particle should be inserted in the feeding tray along with cotton.

Page 13: Shirley trush analyzer

ThanksTHANKS

TO ALL