Shower study for 1-5 GeV energy electron beam with ASIC
SKIROC readout
Elmaddin Guliyev
LLR – Ecole Polytechnique, CNRS/IN2P3
Palaiseau, France
26.09.2012- Analysis meeting
![Page 2: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/2.jpg)
e- beam
Detector Interface (DIF)
Very front-electronics (FVE8)ASIC+Si wafer
to DAQ (LDA)
Tungsten plate
e- beam energy: 1-5 GeVRate: 1 KHz – 5 Hz
6 layer of detector=6*(Si wafer+4ASIC+DIF)
![Page 3: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/3.jpg)
1 GeV e- hit with 6 layer of detector: between the layers and in front of 1st layer W plate placed.
All the run was same: 3600 sec.
![Page 4: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/4.jpg)
Example of MIP for one chip, channel,column
Pedestal
1st MIP
2nd MIP
![Page 5: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/5.jpg)
MIP position variety:
![Page 6: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/6.jpg)
MIP position variety:
![Page 7: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/7.jpg)
Shower profile for 6 layer/detector for 1 GeV electron / X0
![Page 8: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/8.jpg)
Beam alignment or leakage in configuration????
![Page 9: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/9.jpg)
Hit per event for different radiation length
![Page 10: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/10.jpg)
X profile GAUS fit
Use to estimate:
Shower size-sigma from gausBeam alignment-mean from gaus
![Page 11: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/11.jpg)
Deviation ==> beam alignment
Mean value of Gaus fit of shower on X profile
![Page 12: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/12.jpg)
Shower size for different radiation length: Size of shower determined from sigma of Gaus fit of shower profile.
Show
er size (sigma gaus)
![Page 13: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/13.jpg)
e-
Shower study in radial direction for different radiation length.
![Page 14: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/14.jpg)
Efficiency for the layers (detectors)
1 GeV electron beam
X0 and 6X0 configuration
![Page 15: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/15.jpg)
Hit for one event in same BCID
![Page 16: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/16.jpg)
Summary:
Hit distribution for different radiation length analysed.The size of shower estimated for different radiation length.
It seems beam not aligned proper or we had a shower leakage due to particle loose, which gives the discrepancy in end of X0 and beginning of3X0 configuration (same for 3X0 and 6X0).
Need to verify it with same geometry with simulation!
![Page 17: shower study for Si-W prototype ECAL with ASIC readout](https://reader038.vdocument.in/reader038/viewer/2022103002/55b60502bb61ebfa438b45c0/html5/thumbnails/17.jpg)
Hold calibration
Optimized value: 100-110 uDelay