Side-by-side ambient comparison of Capture vs
Standard Vaporizer
Weiwei Hu, Pedro Campuzano-Jost, Doug Day, Jose L. Jimenez, Phil Croteau, and John T. Jayne
University of Colorado-Boulder & Aerodyne Research
7-Sep-2013: AMS Users Meeting, Prague(Analysis still preliminary)
1
![Page 2: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/2.jpg)
Motivation: Field Comparison of OA for 3 AMSs + 1 ACSM (all standard vaporizer)
2
![Page 3: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/3.jpg)
So just compare the AMS w SMPS!
3
![Page 4: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/4.jpg)
Experimental Setup
HR-AMS 2STD
VAPORIZER,ePToF
HR-AMS 1Capture
VAPORIZER
CU Roof
10 lpm
1.2 lpm
0.6 lpm
0.6 lpm
5 min V5 min W
5 min V/PToF5 min V/ePToF
Both AMS were AN, AS and size calibrated before the comparison
Capture Vaporizer was being run at 4 W, PToF sample time ~ 1 min/5 min cycle
Dry (< 15% RH)
CE = 0.5 for Std vaporizer (per Middlebrook et al., AS&T 2012)
John Jaynes’s presentation forePToF results
4
![Page 5: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/5.jpg)
Time Series of Main Species
Using measured IE, NH4 RIE and SO4 RIE on each vaporizer
Fresh Plumes Period, CE=1 was applied to Org in standard Vaporizer
5
CaptureStandard
![Page 6: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/6.jpg)
Species scatter plots
6
Cap
ture
Standard
![Page 7: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/7.jpg)
7
Tge
Monodisperse NH4NO3 in PToF
Capture vaporizer has delayed signals (e.g 30 s for 400 nm), and broader peaks
Capture
Standard
![Page 8: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/8.jpg)
Monodisperse (NH4)2SO4 in PToF
The size distribution of pure (NH4)2SO4 in the capture vaporizer was broadened due to the longer vaporization time. 8
Capture
Standard
![Page 9: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/9.jpg)
Ambient size distribution: Std vs captureShifting the standard vapPToF by 500 us makes the high volatity species (AN) collapse on top of each other. So while delayed and slightly smeared, PToFseems useful!
Warning: Slightly unfair comparison 1% chopper (capture vap) vs ePToF (standard vap)9
CaptureStandard
![Page 10: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/10.jpg)
Sulfate Fragmentation
• Only SO2+ & SO+
• SO2+/SO+ ratio slightly larger 10
CaptureStandard
![Page 11: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/11.jpg)
Nitrate Fragmentation
While detecting RONO2 with the capture vaporizer is not impossible, at ambient concentrations it might be quite challenging… 11
CaptureStandard
![Page 12: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/12.jpg)
HR average spectra (Normalized to 1)
More small fragments (e.g m/z 28, 40, 44) in capture vaporizer, and less higher m/z 12
Capture
Standard
Capture - Standard
![Page 13: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/13.jpg)
13
HR Spectra of Fresh “Cooking” Plumes
Capture
Standard
Capture - Standard
![Page 14: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/14.jpg)
f44 comparison and Sally’s Triangle
• f44 is probably still an O/C proxy but some new complications, may need replacement for f43
• m/z 40 and 55 can be enhanced, so possibly we can use those instead 14
f 44
Cap
ture
f44 Standard
Standard
Capture
![Page 15: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/15.jpg)
Nominally higher O/C and lower H/C w/ capture vap.
15
Still shows the same evolution Need to re-calibrate the elemental analysis
Capture
Standard
![Page 16: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/16.jpg)
CO+/CO2+ & H2O+/CO2
+
16
Cap
ture
Stan
dard
V
Cap
ture
Stan
dard
W
Frag
Tab
le
Stan
dard
V
Stan
dard
W
Frag
Tab
le
![Page 17: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/17.jpg)
Biomass Burning Tracer f60
17
Capture
Standard
Non-BB background
Non-BB background
![Page 18: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/18.jpg)
PMF Comparison
18
CaptureStandard
![Page 19: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/19.jpg)
Spectra of OA PMF factors
19
CaptureStandard
![Page 20: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/20.jpg)
Time series of PMF OA factors
20
CaptureStandard
![Page 21: Side-by-side ambient comparison of Capture vs Standard ...cires.colorado.edu/jimenez...Std_Vaporizer_Comp_Hu.pdf•Capture vaporizer works, CE ~1 ! •Need to recalibrate RIEs, O/C…](https://reader034.vdocument.in/reader034/viewer/2022051921/600eeab4ccfb9417fc31ae95/html5/thumbnails/21.jpg)
Conclusions• Capture vaporizer works, CE ~1 !
• Need to recalibrate RIEs, O/C…• Slower evaporation, PToF ok for ambient• MS shifted to smaller fragments
• Still see same trends in VK diagram, Sally’s triangle
• Much higher signal at CO2+, H2O+
• Seem to lose f60, organic nitrates• Other tracers may still be usable
• Additional testing for SOAS, to be analyzed21