WarrantyInformation...............................................................................2ProductRegistration................................................................................2ChemicalCompatibility............................................................................2SafetyInformation...................................................................................2Dimensions..............................................................................................2Specifications..........................................................................................3Installation...............................................................................................3Wiring......................................................................................................6BatteryInstallationandReplacement......................................................6Operation.................................................................................................7StandardMenuSettings..........................................................................7SecurityCodes........................................................................................8TotalizerOperation..................................................................................9AutomaticCalibration............................................................................11FlowK-Factor........................................................................................11TotalK-Factor........................................................................................12Timebase...............................................................................................13DecimalPointPlacement......................................................................13Sensitivity..............................................................................................14Speed....................................................................................................15FlowandTotalizerK-FactorSelection...................................................15K-FactorCharts.....................................................................................16Troubleshooting.....................................................................................18OrderingInformation..............................................................................20
Operating Instructions
Description
Table of Contents
TheSignet8150FlowTotalizerisabattery-poweredinstrumentcapableofprovidinguninterruptedflowandtotalvolumeinformationfor4yearsandbeyond.Theuniquefeaturesofthe8150include:
• Easysetupanddisplayselectionwith4-buttonkeypad.• Userselectablesecurityaccesstopreventunwantedprogrammingchanges.• Displaysflowratefrom0.001to9999engineeringunitswithanauto-ranging
decimalpoint.• Threetotalizers,onepermanentandtwothatareindependentlyresettable.• Displayselapsedtimebetweenoperatingperiods.• Integralmountandpanelmountoptions.• 3.6Vlithiumbatterieslast4yearsnominalinmostapplications.• Non-volatilememorystoresallprogrammingandtotalizervaluesevenwhen
batteriesareremoved.
Signet 8150 Flow Totalizer
3-8150.090-1 Rev.K9/14 English
*3-8150.090-1*
2 Signet 8150 Flow Totalizer
Caution / Warning / DangerIndicatesapotentialhazard.Failuretofollowallwarningsmayleadtoequipmentdamage,injury,ordeath
Electrostatic Discharge (ESD) / Electrocution DangerAlertsusertoriskofpotentialdamagetoproductbyESD,and/orriskofpotentialofinjuryordeathviaelectrocution.
Hand tighten only!
Hand Tighten OnlyOvertighteningmaypermanentlydamageproductthreadsandleadtofailureoftheretainingnut.
Do Not Use ToolsUseoftool(s)maydamageproductbeyondrepairandpotentiallyvoidproductwarranty.
Dimensions
RefertoyourlocalGeorgFischerSalesofficeforthemostcurrentwarrantystatement.
Allwarrantyandnon-warrantyrepairsbeingreturnedmustincludeafullycompletedServiceFormandgoodsmustbereturnedtoyourlocalGFSalesofficeordistributor.ProductreturnedwithoutaServiceFormmaynotbewarrantyreplacedorrepaired.
Signetproductswithlimitedshelf-life(e.g.pH,ORP,chlorineelectrodes,calibrationsolutions;e.g.pHbuffers,turbiditystandardsorothersolutions)arewarrantedoutofboxbutnotwarrantedagainstanydamage,duetoprocessorapplicationfailures(e.g.hightemperature,chemicalpoisoning,dry-out)ormishandling(e.g.brokenglass,damagedmembrane,freezingand/orextremetemperatures).
ThankyouforpurchasingtheSignetlineofGeorgFischermeasurementproducts.Ifyouwouldliketoregisteryourproduct(s),youcannowregisteronlineinoneofthefollowingways: • Visitourwebsitewww.gfsignet.com.
UnderService and SupportclickonProduct Registration Form
• Ifthisisapdfmanual(digitalcopy),click here
Chemical Compatibility
Warranty Information
Product Registration
Safety Information
Integral Mount
96 mm(3.8 in.)
96 mm(3.8 in.)
FRONT VIEW
106 mm(4.2 in.)
82 mm(3.2 in.)
Universal Mount
92 mm(3.6 in.)
97 mm(3.8 in.)
56 mm(2.2 in.)
41 mm (1.6 in.)
OptionalRearCover
Panel MountSIDE VIEW
102 mm(4.0 in.)
42 mm (1.7 in.)
60 mm (2.3 in.)
96 mm(3.8 in.)
Theretainingnutsofpaddlewheelsensorsarenotdesignedforprolongedcontactwithaggressivesubstances.Strongacids,causticsubstancesandsolventsortheirvapormayleadtofailureoftheretainingnut,ejectionofthesensorandlossoftheprocessfluidwithpossiblyseriousconsequences,suchasdamagetoequipmentandseriouspersonalinjury.Retainingnutsthatmayhavebeenincontactwithsuchsubstances,e.g.,duetoleakageorspilling,mustbereplaced.
Integral Mount
96 mm(3.8 in.)
96 mm(3.8 in.)
FRONT VIEW
106 mm(4.2 in.)
82 mm(3.2 in.)
Universal Mount
92 mm(3.6 in.)
97 mm(3.8 in.)
56 mm(2.2 in.)
41 mm (1.6 in.)
OptionalRearCover
Panel MountSIDE VIEW
102 mm(4.0 in.)
42 mm (1.7 in.)
60 mm (2.3 in.)
96 mm(3.8 in.)
3Signet 8150 Flow Totalizer
SpecificationsGeneralCompatibility...................... Signet515,525FlowSensorsInputFrequencyRange..... 1to400HzAccuracy............................ ±0.5%ofreading
EnclosureDimensions........................ 1/4DIN(96mm×96mm×50mm) (3.8in.×3.8in.×2.0in.)Casematerial.................... PBTresinKeypadmaterial................. Sealed4-keysiliconrubber
Display: LCD type4-digitupperline................ Flowrate8-digitlowerline................. Threetotalizeroptions: PermanentTotalizer....... forlifeofinstrument Totalizer1...................... resettablefromkeypadorremote 30m(100ft) Totalizer2...................... ResettablewithsecuritycodeonlyDisplayContrast................ Automatic
ElectricalBattery............................... Two3.6Vlithiumthionylchloride, AA-size(UseSAFTLS14500 LithiumBatteriesorEquivalentOnly)Sensorpoweroutput......... +3.6VDC@20µABatterylife.......................... 4yearsnominal@50°C(122°F)LowBatteryindication....... BatterysymbolonLCDdisplayEnvironmentalOperatingTemperature...... -10°Cto65°C(14°Fto149°F)StorageTemperature......... -40°Cto100°C(-40°Fto212°F)RelativeHumidity.............. 0to95%,non-condensingRating................................ NEMA4X/IP65(frontpanel)
Shipping Weight.............. 0.5kg(1.1lbs.)
Quality StandardsCE,CUL,ULManufacturedunderISO9001forQuality,ISO14001forEnvironmentalManagementandOHSAS18001foroccupationalhealthandsafety.
ChinaRoHS(Gotowww.gfsignet.comfordetails)
Declaration of Conformity according to FCC Part 15ThisdevicecomplieswithPart15oftheFCCrules.Operationissubjecttothefollowingtwoconditions:(1)Thisdevicemaynotcauseharmfulinterference,and,(2)Thisdevicemustacceptanyinterferencereceived,includinginterferencethatmaycauseundesiredoperation.
Panel cutout
92 x 92 mm(3.62 x 3.62 in.)
Output -
Output +
System Pwr
Loop -
System Pwr
Loop +
2
1
4
3
Sensr Gnd
(SHIELD)
Sensr IN
(RED)
Sensr V+
(BLACK)
7
6
5
quick-clips
gasketpanel
terminalsmountingbracket
latch
Panel MountInstallation Detail
The8150-1PpanelmountTotalizerisastandard¼DINpackage.Usea92×92mmpunchtooltomakethepanelcutout.Minimumspacingof25mm(1in.)betweenpanelunitsisrecommended.
Panel Mount Installation Detail
Installation
PanelCutout
PanelMountInstallationDetail
4 Signet 8150 Flow Totalizer
InstallationIntegral Installation with Field Mount Totalizer
3-8150-P03-8150-P13-8150-T03-8150-V0
(3-8510-XX)
(3-8051-X)
(3-8150-1)
FlowTotalizer
IntegralMounting Kit
IntegralFlow Sensor
SignetFittings
Integral Mount
• LubricateO-ringswithanon-petroleumbased,viscouslubricant(grease)compatiblewiththesystem.
• Usinganalternating/twistingmotion,lowerthesensorintothefitting,makingsuretheinstallationarrowsontheblackcaparepointinginthedirectionofflow.
• Engageonethreadofthesensorcapthenturnthesensoruntilthealignmenttabisseatedinthefittingnotch.
Hand tighten the sensor cap.DO NOT use any tools on the sensorcap or the cap threads and/or fi ttingfl ange threads will be damaged,
1. Insertthewiresfromthesensorthroughtheyellowconduitbaseandlockingring.
2. Insertthelockingringintotheconduitbase,aligningitsothesquaretabisclosetotheconduitports.
3. Insertthemountingscrewintotheconduitbasesotheheadofthescrewpressesdownonthelockingringwhentightened.
4. Connectsensorwirestotheterminalconnectionsontheintegraltotalizerorterminalcover.
5. Removetheplasticpulltabsprotectingthebatteries,thiswillcausethetotalizerunittopowerup.
6. Forremoteassembly,connectoutputwirestoterminalcoveroutput.
7. Routetheoutputwiresthroughtheconduitportintheconduitbase.Usealiquid-tightconnectororconduitconnectortopreventmoisturefromenteringtheassembly.
8. Placetotalizerorcoverontoconduitbaseandtwisttolockinplace.
Conduit Base Assembly Detail
The Mounting Screw for the Locking Ring goes into the Conduit Base (not the Locking Ring) -- the screw head rests on the lip of the Locking Ring.
Do not use this hole.
Sensor wires
Locking Ring
Conduit Base
Conduit Ports
Mounting Screw
ConduitBase Kit
3-8150 Flow Totalizer
Square Tab
sensor bale
sensor cap
tab
blackconduit
cap
notch
Hand tighten only!
Plastic Sensor Installation Tips
5Signet 8150 Flow Totalizer
UniversalConduit Base(3-8050)
Terminal Cover(4-3111)
Liquid-tight Connector(3-9000.392-1)
FlowTotalizer(3-8150-1)
IntegralFlow Sensor(3-8510)
SignetFittings
IntegralMounting Kit(3-8051-X)
Remote Field Mount on Wall
Thepartsidentifiedinbold typearerequiredfortheinstallation.Otherpartsareshownforreferenceonly.Use2-conductorshieldedcablenomorethan30m(100ft)long.
Remote Field mount on Pipe
FlowTotalizer
(3-8150-1)
SignetFittings
Standard MountPaddlewheel(P51530-X)
Universal Mounting Kit(3-8050)
NOTE: Pipe strapsnot included
Remote Installation with Panel Mount Totalizer
TotalizerPanel Mount(3-8150-1P)
SignetFittings
Standard MountPaddlewheel(P51530-X)
Installation
6 Signet 8150 Flow Totalizer
1 2 3 4 5 6
+
-
+
-
RED#1 #2
Totalizer 1Remote reset
Connect shieldto Ground wire
Splice BLK and SHLDtogether outside of terminal
BLACK
SHIELD
515 525
Battery Installation and ReplacementWiringThewiringisidenticalforthepanelmountandthefieldmountversionsoftheTotalizer.• Onlyonewireshouldbeinsertedintoaterminal.Splice
doublewiresoutsidetheterminal.• External Reset for Total #1:Usenomorethan
30m(100ft)of2-conductortwisted-paircableconnectedtoadrycontact(forexample,anordinarydoor-bellbuttonorrelaycontact).
• OnlyTotalizer#1canberesetbytheexternalconnection.• Total#1willnotbedisplayedunlessitisthestandard
totalizerselection.
Instructions1. Remove10mm(3/8in.)ofinsulationfrom
sensorcableconductors.2. Pressdownonorangelevertoopenterminal.3. Insertwireintoterminaluntilithitsbottom.4. Releasethelevertosecurewire.
10 mm
Input WiringUsethiswiringschemeforSignetmodels515and525.
Terminal Function 1 SignalGround 2 Sensorsignal 3 OpenCollectorSignal 4 DCPowertosensor 5 Ground 6 Ext.Reset
Two3.6 V lithium thionyl chloride batteries,AA-size(7400-0011)areinstalledintheTotalizer.
NOTE: The8150willnotoperatewithstandard1.5Valkalinebatteries.USE3.6VSAFTLS14500LITHIUMBATTERIESOREQUIVALENTONLY!
CAUTION When replacing batteries,
remove and replace one battery at a time
Remove and replace this battery first.
1 2 3 4 5 6
Note: Lithium Batteries Dispose of properly!
Replace with 3.6 V Lithium battery
+ +
- -
1. Removepulltabsfromthebatteriestopowerupthe8150.2. Note:Observe polarity!Bothbatteriesshouldfacethe
samedirection.3. Whenthe“lowbattery”indicatorappearsonthedisplay,
bothbatteriesshouldbereplacedwithin90days.4. Removeandreplacebattery#1first,thenremoveand
replacebattery#2.Thisensuresthatallsettingsandtotalizervaluesaresaved.
5. Ifthelowbatterysymbolreappearsformorethan10secondsafterinstallingnewbatteries,onebatteryisreversed,orbattery#2wasinstalledbeforebattery#1.
6. Securethebatteriesbyfasteningthehook-and-loopstraps.
DISPOSE OF EXPENDED BATTERIES PROPERLY!Lithiumbatteriescontainhazardouschemicals.Disposeofbatteriesinaccordancewithlocalregulation.
Shipping Notice:Ifthebatterypulltabshavebeenremoved,removethebatteriesfromthetotalizerpriortoshipping.
7Signet 8150 Flow Totalizer
Totalizersareshippedfromthefactorywiththesestandardsettings:
No Flow and ELAPSED TIME DisplayIftheflowstops,theFlowRatedisplaysthenumberofhourssinceflowwaslastdetected.ThisdisplaywillalternatewiththenormalTotalizedValueandstandard"noFLo"displayeveryfiveseconds.AnymovementoftherotorinthepipewillresettheELAPSEDTIMEdisplay.
The8150-1displayshowstheflowrateinlargenumeralsandatotalizervalueinsmallernumerals.Anyoneofthreedifferenttotalizerscanbeselectedasthestandarddisplay(Seepage9fordetailedinformationonthetotalizers).
Pressthe▲or▼keystoscrollthroughallthreetotalizervaluesduringnormaloperation.
Illustrated:Noflowfor3hours
Operation Standard Menu Settings
ENTER
5 sec.
Function Factoryset Description
AUTO CALIBRATION Nosetting Seepage11fordetailed
information.
FLOW K-FACTOR 60
Numberofsensorpulsespervolumetricunit;Refertosensormanual.
TOTAL K-FACTOR 1
Setthenumberofvolumetricunitspertotalizercount;seepage12.
TIMEBASE Minutes
Selectflowrateinseconds,minutes,hoursordays.Page13.
DECIMAL XXX.XSetthemaximumdecimalresolution.Page13.
SPEED 30s
Zeroto120secondsaveragingstabilizesreadingsinerraticflowconditions.Page15.
SENSITIVITY 6
MomentarilyoverridesSPEEDwhenflowratechangessignificantly.Page14.
SECURITY CODE 0-0-0-0
Setaprivatecodetopreventtampering.Page8.
DEFAULT TOTALIZER Permanent
Selectfromthreetotalizeroptions.Page9.
TOTALIZER #2 RESET
ResetTotalizer#2afterenteringthesecuritycode.Page9.
8 Signet 8150 Flow Totalizer
Changing the Security Code
Thesecuritycodepreventsunauthorizedtamperingwithcalibrationandoperationalsettingsinthe8150.Thefactorystandardcodeis0-0-0-0.Changethecodetoany4-digitnumberbyfollowingthesesteps:
Example: Change the security code from the factory standard 0-0-0-0 to custom setting 1-0-0-1
NOTE:Record and store your security code in a safe place!
12 s
2 3
4 5
6 7
8
3x
3x
Enter Security code
1. HoldtheENTERkeyfor2seconds,thenenterthecurrentSECURITYCODE.Ifworkingwithanewunit,presstheENTERkeyagain.Thedisplayshowsthefirstmenuitem(AutoCal)
2. Pressthe▲keythreetimestoscrolltoSecCode.3. Press►toeditthecode.
Theleadingdigitonthedisplaywillflash.4. Pressthe▲keyonetimetoscrolltheflashingdigitto1.5. Pressthe►keythreetimestoadvancetheflashingelement
tothelastdigit.6. Pressthe▲keyonetimetoscrolltheflashingdigitto1.7. PresstheENTERkeytocompletetheeditprocess.8. Pressthe▲and▼keystogethertoexitthe
Calibratemenu,storethenewsettingsandreturntonormaloperation.
Using the Security Code• Anumericalcode(0-0-0-0to9-9-9-9)mustbeentered
beforeanyofthemenuselectionscanbemodified.• Thecodeissetat0000fromthefactory.Tochangethe
code,seeabove.• Tousethefactorysetting,completestep1andthenstep8
above.
Example: Enter security code of 1001:
1. HoldtheENTERkeyfor2seconds.Thedisplayshowsfactorystandardaccesscodeof0000,withthefirstzeroflashing.
2. Pressthe▲keyonetimetoscrolltheflashingzeroto1.3. Pressthe►keythreetimestoadvancetheflashing
charactertothelastplacevalue.4. Pressthe▲keyonetimetoscrolltheflashingzeroto1.5. PresstheENTERkey.Thedisplaynowshowsthefirstitem
intheEDITMENU.
ENTER
ENTER
Signet Total Flow
Signet Total Flow
Signet Total Flow
Signet Total Flow
Signet Total Flow1 2
4
5
3
3x
2 s
9Signet 8150 Flow Totalizer
Totalizer Setup and OperationDuringnormaloperationthe8150displaystheflowrateandoneselectedtotalizervalue.Anyoneofthethreetotalizerscanbesetasthestandarddisplay:Theothertwototalizerscanbeviewedbypressingthekeypad.Thedisplaywillautomaticallyreturntothestandardselectionafterfiveminutes.
• ThePERMANENT TotalizerisidentifieddirectlyontheLCD.• Thistotalizerrecordsallinputfromthetimeofmanufacture.• Thepermanenttotalizercannotbereset.
Application:Thepermanenttotalizershouldbeselectedasthestandardifthesystemismonitoredandthetotalrecordedregularly.
• Total 1 (tot1)canberesetfromthekeypadorfromtheexternalRESET(seewiring,page6)withoutthesecuritycode.
• Total 1 isidentifiedbyaflashingdisplayeverysixseconds.
Application:UseTotal1tomeasurewaterusageforarecurringperiod,asforadailydischargevolume.
• Total 2 (tot2)canberesetonlybyenteringthesecuritycodeinthecalibrationmenu.
• Total 2 isidentifiedbyaflashingdisplayeverysixseconds.
Application:UseTotal2forextendedmeasurementperiods,asforamonthlydischargevolume.
Define the Standard TotalizerAnyofthethreetotalizerfunctionscanbesetasthestandarddisplay,orselectSCANtodisplayallthreetotalizersinsequence.ThePERMANENTtotalizeristhefactorystandardselection.
Example: Change the standard Totalizer from PERMANENT to Totalizer #1
1. PressENTERkeyfor2seconds.(Displayshowssecuritykeysymboland0-0-0-0.SetsecuritycodeandpressENTERkey.)
2. Press▲keytwotimes.Displayshows"deftot"andthe"PERMANENT"label.)
3. Press►key.(PERMANENTlabelbeginstoflash.4. Press▲keyonetime.Displaychanges
toflashing"tot1".5. PressENTERkeytocompletetheeditprocess.6. Press▲and▼keystogethertostorenewvalue
inthememory.7. Thedisplaywillshow"Storing"forafewseconds,
thenreturntonormaloperation.
NOTE:Thenewtotalizerselectionwillappearaftera5minutedelay.
2x
5 m
Enter security code2 s
1 2
3 4 5
6
10 Signet 8150 Flow Totalizer
Resetting Totalizer 11. Press▲keytoscrollto"total 1" display.2. Pressandholdthe►keyuntilthedisplayshows
"rst tot1".Thetotalizerwillflashfor8secondsandthenitwillautomaticallyresetto00000000.• PresstheENTERkeywhilethedisplayisflashingtoresetimmediately.
• Whilethetotalvalueisflashing,youcancanceltheresetbypressing▲and▼keystogether.
• Totalizer#1willbedisplayedfor5minutesafterthereset,thenthestandardtotalizerselectionwillreturn.
• Pressthe▲or▼keytoscrollbacktothestandarddisplayimmediately.
External Reset:Seepage6:WiringforinformationonresettingTotalizer#1remotelyfromupto30metersdistance.
NOTE:WhenTotal#1isresetfromanexternalswitch,thedisplaywillnotshowtotalizer#1unlessithasbeensetasthestandardtotalizer.
Resetting Totalizer 2 Totalizer#2canberesetONLYbyenteringthesecuritycode.1. PressENTERkeyfor2seconds.(Displayshowssecurity
keysymboland0-0-0-0)2. Setthesecuritycodeintheflashingdisplayandthenpress
theENTERkey.3. Press▲keyonetime.(Displayshows"tot2reset")4. Press►key.Thetotalizervaluewillbeginflashing.
Thetotalizerwillautomaticallyresetto00000000in8seconds.Whilethedisplayisflashing,youcancanceltheresetbypressing▲and▼keystogether.
5. Press▲and▼keystogethertoreturntonormaloperation.
Saving Totalizer ValuesToconservebatterylife,totalizervaluesarestoredinthememoryevery12hours.Ifbothbatteriesareremovedfromtheunit,thetotalizersretainthelastsavedvalues,sotheunitmayloseseveralhoursofdata.Topreventthisloss,enterthesecuritycode,thenenteranymenuitemandstorethesetting.
Wheneverthe8150storesasetting,italsostoresallcurrenttotalizerdata:
1. Enterthesecuritycode.2. Pressthe▲keytoscrolltothelastiteminthemenu
(DEFAULTTOTALIZER)3. Pressthe►keytoentertheeditmode
(flashingdisplaymode)4. PresstheENTERkeytoretainthecurrentsettings.5. Pressthe▲and▼keyssimultaneouslytoinitiatethe
"Storing"function.Thebatteriescannowberemovedandreplacedwithoutlosinganytotals.
1 2
8 s 5 m
2 s
1 2 3
4
5
8 s
2 sEnter security code
Totalizer Setup and Operation
Pressthe▲and▼keyssimultaneouslyfromthemainmenutoreturntonormaloperation.
Aslongasanyelementisflashing,youcanabortthechangeandreturntotheoriginalvaluebypressing▲and▼keyssimultaneously.
Exit Without Changing?* Finished Editing?**
11Signet 8150 Flow Totalizer
TheAutoCALfeatureallowstheTotalizertobeadjustedtomatchtheflowratetoanyexternalreference.• Flowinthepipeshouldbeasstableaspossibleforbest
results.• Iftheflowratedisplayiserratic,settheSPEED
(page15)to120secondsduringtheAutoCALprocedure.• Thetimebaseonthereferencemetermustbethesameas
the8150Totalizer.
Example: The Totalizer flow rate shows 60 GPM, while an external reference indicates a true flow rate of 70 GPM. Change the flow rate from 60 GPM to 70 GPM using AutoCAL.
1. HoldtheENTERkeyfor2seconds,thenentertheSECURITYCODE.AutoCAListhefirstiteminthemenu.
2. Pressthe►keytoselecttheAutoCALfunction.ThedisplaywillshowAutoCALandthecurrentflowratewillbeflashing.
3. Pressthe►keyagaintochangetheflowrate.Thedisplayshows"SetFlo"andthefirstdigitoftheflowratewillbeginflashing.
4. Pressthe►keytoadvancetheflashingelementtothe"6".
5. Pressthe▲keyonetimetochangethe"6"to"7".6. PresstheENTERkeytocompletetheautomatic
calibrationprocess.ThedisplayshowsanewK-Factorwiththefirstdigitflashing.ThisK-Factorisbasedonthechangeinflowrate.
7. PresstheENTERkeyagaintoacceptthenewvalue.NOTE:Ifthedisplayshows"ERR SetFlo" theprocedurewasunsuccessfulbecausethecalculatedK-Factorislessthan0.001orgreaterthan999999.VerifytheflowrateandstarttheAutoCALprocedurefromstep1.
8. Press▲and▼keystogethertostorethenewvalueinthememory.Thedisplaywillshow"storing"forafewseconds,thenreturntonormaloperation.
AutoCAL Calibration
Enter security code2 s
1 2
3 4 5
6 7
8
TheK-Factoristhenumberofpulsesgeneratedbytheflowsensorforeachmeasureofwaterthatmovespastthesensor.YourflowsensormanualcontainsK-FactordataintermsofU.S.gallonsandliters.LocatetheK-Factorthatmatchesyourpipesizematerial.Ifnecessary,youcanconverttheK-Factorintootherunitsofmeasure.TheminimumK-Factorvalueis0.001,maximumis999999.
Example: Change the Flow K-Factor from 060.000 to 095.000
1. HoldtheENTERkeyfor2seconds,thenentertheSECURITYCODE.(Factorydefaultis0-0-0-0)Thedisplayshowsthefirstiteminthemenu(AutoCAL).
2. Pressthe▼keytoscrolltotheFlowK-Factor(thedisplayshowsthecurrentK-Factorsetting).
3. Pressthe►keytoselecttheFlowK-Factorforediting.(ThefirstelementoftheK-Factorwillbeginflashing.)
4. Pressthe►key1timetoadvancetheflashingelementtothe"6".
5. Pressthe▲keythreetimestochangethe"6"to"9".6. Pressthe►keytoadvancetheflashingelementto
the"0."7. Pressthe▼keyfivetimestochangethe"0"to"5".8. PresstheENTERkeytoreturntothe
CALIBRATEmenu.9. Press▲and▼keystogethertostorethenewvalue
andreturntonormaloperation.
Flow K-Factor
2 s
1 2
3 4 5
6 7 8
9
Enter security code
**
****
12 Signet 8150 Flow Totalizer
Total K-Factor
TheTOTALK-FactorisamultipleoftheFLOWK-Factor.UseittoprogramtheincrementalcountsizeoftheTotalizer.
Example:IftheflowRATEregistersinlitersperminute,thetotalizermaybesetto1(factorystandard),soitcountsin1-literincrements,oritmaybesetto1000soitcountsin1kiloliter(1m3)increments.ByconvertingtheFlowK-Factor,thetotalizercanalsobesettocountinotherengineeringunits.Seepage15foradditionalinformationaboutTotalK-Factoradjustments.
Example: Change the totalizer from 1-kiloliter increments to count in 10-kiloliter increments.
1. HoldtheENTERkeyfor2seconds,thenentertheSECURITYCODE.Thedisplayshowsthefirstiteminthemenu,AutoCAL.
2. Pressthe▼keytwotimestoscrolltotheTotalK-Factor.3. Pressthe►keytoselecttheTotalK-Factorforediting.
ThefirstelementoftheTotalK-Factorwillbeginflashing.4. Pressthe►keysixtimestoadvancetheflashingelementto
thedecimalpoint.5. Pressthe▼keyonetimetomovethedecimalpointone
positiontotheright.6. PresstheENTERkeytoreturntothemenu.7. Press▲and▼keystogethertoexitthemenuandreturn
tonormaloperation.Thedisplayshows"Storing"forafewseconds,thenreturnstothenormaloperatingdisplay.
ENTER
6x
2x2 s
1
3 4 5
6 7
2
Sicherheitscode eingeben
Adjusting the Flow K-FactorIftheTotalizeryieldsaconsistenterror,makecorrectionsbyeitherusingtheAutoCALfunction(page11)orbymanuallyadjustingtheFlowK-Factorbythepercentageoferror.AsmallerK-Factorincreasestheflowrate,whilelargerK-Factorsreducetheflowrate.
Example:• TheFlowK-Factorissetat480.19pulsespergallon.• Thetotalizerregisters10gallonswhentheactualvolumeisknowntobe11gallons.• Theerroris1gallondividedby10gallons,or-10%.(Thetotalizeriscounting10%low,andtheflowrateisreading10%slow.)• ReducetheFlowK-Factorby10%:480.19-10%=432.17.• ChangetheFlowK-Factorto432pulsespergallon.
Theresult:Thetotalizermustcount10%fewerpulsesfromtheflowsensortoregisteronegallon,soboththetotalizerandtheflowratewillincreaseby10%.
Flow K-Factor
13Signet 8150 Flow Totalizer
Selectthetimebasefortheflowrate.Theavailableselectionsareseconds,minutes,hoursordays.
Example: Change the Timebase from MINUTES (factory standard) to DAYS
1. HoldtheENTERkeyfor2seconds,thenentertheSECURITYCODE.ThedisplayshowsthefirstitemintheCALIBRATEmenu,AutoCAL.
2. Pressthe▼keythreetimestoscrolltotheTimebase.3. Pressthe►keytoselecttheTimebaseforediting.4. Pressthe▲keytwotimetoscrollfromMINtoDAY.5. PresstheENTERkeytoreturntothemenu.6. Press▲and▼keystogethertoexitthemenuandreturn
tonormaloperation.Thedisplayshows"Storing"forafewseconds,thenreturnstothenormaloperatingdisplay.
Enter security code2 s
1
3 4 5
6
2
2x
3x
*
**
Timebase
Decimal Point for Flow Display
Selectthemaximumdecimalresolutionfortheflowratedisplay.Theavailableselectionsarehundredths(xx.xx),tenths(xxx.x)orwholenumbersonly(xxxx.).Thedecimalwillauto-rangedowntothissetting.• Ifthedecimalissettowholenumbers,theflowrate
displaywillnotauto-range.• Ifthedecimalissettotenths,theflowratedisplaywill
showtenthsupto999.9,thentheauto-rangewillswitchtowholenumbers(1000to9999.)
• Ifthedecimalissettohundredths,theflowratedisplaywillshowhundredthsupto99.99,thententhsfrom100.1to999.9,thenwholenumbersto9999.
Example: Change the maximum decimal display from hundredths to tenths:
1. HoldtheENTERkeyfor2seconds,thenentertheSECURITYCODE.Thedisplayshowsthefirstiteminthemenu,AutoCAL.
2. Pressthe▼keyfourtimestoscrolltotheDecimalsetting.Thedisplayshowsfourdashesandthecurrentdecimalsetting.
3. Pressthe►keytoselectthedecimalforediting.Thedecimalpointwillbegintoflash.
4. Pressthe▼keyonetimetomovetheflashingdecimalfromhundredthstotenths.
5. PresstheENTERkeytoreturntothemenu.Thedecimalwillstopflashing.
6. Press▲and▼keystogethertoexitthemenuandreturntonormaloperation.
7. Thedisplayshows"Storing"forafewseconds,thenreturnstothenormaloperatingdisplay.
Available Display Selections
Your Flow rate
will read on display as:
hundredths(XX.XX) 10.55 10.55
tenths(XXX.X) 10.55 10.6wholenumbers(XXXX.) 10.55 11
2 s
1
3 4 5
6
2
4xEnter Security code
**
*
14 Signet 8150 Flow Totalizer
Sensitivity
TheSENSITIVITYsettingdetermineshowthe8150respondstosuddensurgesintheflowrate.It"overrides"theSPEEDfunctionjustlongenoughtoallowanactualchangeinflowratetobedisplayed,thenresumestheaveraging.Theresultisasmoothflowdisplayandaquickresponsetolargeshiftsintheflowrate.
No AVERAGING, no SENSITIVITY
AVERAGING and SENSITIVITY
AVERAGING only
10 s 2 s 20 s 30 s 40 s 50 s 60 s 70 s
No SPEED, no SENSITIVITY
SPEED and SENSITIVITY
SPEED only
No SPEED, no SENSITIVITY
SPEED and SENSITIVITY
SPEED only
WithSPEEDaveragingsetto0(zero)andwithSENSITIVITYsettozero,theflowratemaybeveryunstable.Thislinerepresentstheactualoutputoftheflowsensorasitrespondstounstableflowconditionsinthepipe.
WithSPEEDsetto60secondsandSENSITIVITYstillsettozerotheflowrateisstabilized,butasharpchangeinflowrateisnotrepresentedfor60secondsorlonger.(dottedgreenline).
WithSPEEDat60secondsandSENSITIVITYsetto6,theflowrateisstabilized,whilethesuddenshiftinflowisreflectedveryquickly(dottedblueline).
NOTE:TheSENSITIVITYfunctionisineffectiveiftheSPEEDfunctionissettozero(seconds).
No SPEED, no SENSITIVITY
SPEED and SENSITIVITY
SPEED only
ENTER
3x
Enter Security code2 s
1
3 4 5
6
2
Example: Change the SENSITIVITY from 5 to 8
1. HoldtheENTERkeyfor2seconds,thenentertheSECURITYCODE.Thedisplayshowsthefirstiteminthemenu,AutoCAL.
2. Pressthe▲keyoncetoscrolltoSENSITIVITY.ThedisplayshowsDISPSENSandthecurrentsensitivitysetting.
3. Pressthe►keytoselecttheSENSITIVITYforediting.ThecurrentSENSITIVITYsettingbeginsflashing.
4. Pressthe▲keythreetimestoscrollfrom5to8.5. PresstheENTERkeytoreturntothemenu.6. Press▲and▼keystogethertoexitthemenuandreturn
tonormaloperation.Thedisplayshows"Storing"forafewseconds,thenreturnstothenormaloperatingdisplay.
15Signet 8150 Flow Totalizer
SpeedSPEEDaveragingservestosmoothoutfluctuationsintheflowratethatmaybecausedbyinadequatestraightpiperunsafterpumps,valves,andelbowsinthepipe.Theselectionsare0,7,15,30,60and120seconds.Thefactorystandardsettingis30seconds.
• Usefaster(0s-30s)averagingforwell-established,stableflowconditions.
• Useslower(60s-120s)averagingiftheflowconditionsareunstable.
Note: WhiletheSPEEDsettinghelpstosmoothoutthefluctuationscausedbypipingconditions,italsocausesadelayinshowingactualchangesinflowrate.TheSENSITIVITYfunction(section15)isdesignedtohelpoffsetthiseffect.
Example: Change the SPEED setting from 60 seconds to 30 seconds.
1. HoldtheENTERkeyfor2seconds,thenentertheSECURITYCODE.ThedisplayshowsthefirstitemintheCALIBRATEmenu,AutoCAL.
2. Pressthe▼keyfivetimestoscrolltoDISPSPEED.ThedisplayshowsDISPSPEED,the"sec"annunciator,andthecurrentspeedsetting.
3. Pressthe►keytoselecttheDisplayspeedforediting.Thecurrentspeedselectionbeginsflashing.
4. Pressthe▼keyonetimetoscrollfrom60secondsto30seconds.
5. PresstheENTERkeytoreturntothemenu.6. Press▲and▼keystogethertoexitthemenuandreturn
tonormaloperation.Thedisplayshows"Storing"forafewseconds,thenreturnstothenormaloperatingdisplay.
2 s
1
3 4 5
6
2
5xEnter Security code
*
**
Pages16-18provideK-FactorsfortheSignet515and525flowsensors.UsethistabletoconverttheK-Factortootherunitsofmeasure,andtosettheTotalK-Factor.
NOTE:ThemaximumK-Factoris999999.TheminimumK-Factoris0.001.
Flow and Total K-Factor Selection If you want the
FLOW RATE to read in:
and you want the TOTALIZER
to count in:
Set the Flow K-Factor to:
and set the Total
K-Factor to:
Liters Liters K(Liter) 1Liters Kiloliters K(Liter) 1000Liters cubicmeters K(Liter) 1000
cubicmeters cubicmeters K(Liter)x1000 1cubicmeters Megaliters K(Liter)x1000 1000Kiloliters Kiloliters K(Liter)x1000 1Kiloliters Megaliters K(Liter)x1000 1000Megaliters Megaliters K(Liter)x1000
0001
U.S.gallons U.S.gallons K(gal) 1U.S.gallons U.S.gallons
x1000K(gal) 1000
U.S.gallons cubicfeet K(gal) 7.4805U.S.gallons acre-inches K(gal) 27154U.S.gallons Acre-feet K(gal) 325848U.S.gallons Kiloliters K(gal) 264.2Acre-Inches Acre-Inches K(gal)x27154 1
Acre-Inches Acre-feet K(gal)x27154 12Acrefeet Acre-feet K(gal)x325848 1Acrefeet Acre-Inches K(gal)x325848 0.083cubicfeet cubicfeet K(gal)x7.4805 1
16 Signet 8150 Flow Totalizer
K-Factor Charts515 Paddlewheel Flow SensorThefollowingcalibrationdataisreprintedfromtheinstructionmanualfortheSignet515Flowsensorforyourconvenience.
PIPESIZE(IN.)
FITTING515/8510-XX
U.S. GAL LITERS
SCH 80 PVC TEES FOR SCH 80 PVC PIPE
1/2 MPV8T005 520.12 137.42
3/4 MPV8T007 297.52 78.61
1 MPV8T010 172.07 45.46
1-1/4 MPV8T012 91.54 24.19
1-1/2 MPV8T015 62.22 16.44
2 MPV8T020 36.32 9.60
2-1/2 PV8T025 21.833 5.7683
3 PV8T030 13.541 3.5775
4 PV8T040 7.6258 2.0147
SCH 80 CPVC TEES FOR SCH 80 CPVC PIPE
1/2 MCPV8T005 520.12 137.42
3/4 MCPV8T007 297.52 78.61
1 MCPV8T010 172.07 45.46
1-1/4 MCPV8T012 91.54 24.19
1-1/2 MCPV8T015 62.22 16.44
2 MCPV8T020 36.32 9.60
SCH 80 PVC SADDLES FOR SCH 80 PVC PIPE
2 PV8S020 32.480 8.5812
2-1/2 PV8S025 21.833 5.7683
3 PV8S030 13.541 3.5775
4 PV8S040 7.6258 2.0147
6 PV8S060 4.1623 1.0997
8 PV8S080 2.3705 0.6263
10 PV8S100 1.5300 0.4042
12 PV8S120 1.0600 0.2801
SCH 80 PVC SADDLE ON SCH 40 PVC PIPE
2 PV8S020 27.350 7.2259
2-1/2 PV8S025 18.874 4.9866
3 PV8S030 12.638 3.3389
4 PV8S040 6.7282 1.7776
6 PV8S060 3.7297 0.9854
8 PV8S080 2.1527 0.5688
10 PV8S100 1.3500 0.3567
12 PV8S120 0.9600 0.2536
PIPESIZE(IN.)
FITTING515/8510-XX
U.S. GAL LITERS
CARBON STEEL TEES ON SCH 40 PIPE
1/2 CS4T005 370.20 97.808
3/4 CS4T007 212.06 56.027
1 CS4T010 141.14 37.289
1-1/4 CS4T012 60.655 16.025
1-1/2 CS4T015 45.350 11.982
2 CS4T020 26.767 7.0717
STAINLESS STEEL TEES ON SCH 40 PIPE
1/2 CR4T005 358.96 94.838
3/4 CR4T007 202.61 53.530
1 CR4T010 127.14 33.590
1-1/4 CR4T012 61.910 16.357
1-1/2 CR4T015 40.410 10.676
2 CR4T020 22.300 5.8917
GALVANIZED IRON TEES ON SCH 40 PIPE
1 IR4T010 104.54 27.619
1-1/4 IR4T012 62.979 16.639
11/2 IR4T015 46.688 12.335
2 IR4T020 29.459 7.7832
BRONZE TEES ON SCH 40 PIPE
1 BR4T010 104.54 27.619
1-1/4 BR4T012 62.979 16.639
1-1/2 BR4T015 46.688 12.335
2 BR4T020 29.459 7.7832
COPPER TEE FITTINGS ON COPPER PIPE SCH K
1/2 CUKT005 443.21 117.10
3/4 CUKT007 212.16 56.052
1 CUKT010 127.18 33.600
1-1/4 CUKT012 88.218 23.307
1-1/2 CUKT015 56.962 15.049
2 CUKT020 29.370 7.7595
COPPER TEE FITTINGS ON COPPER PIPE SCH L
1/2 CUKT005 414.41 109.49
3/4 CUKT007 191.09 50.485
1 CUKT010 119.84 31.662
1-1/4 CUKT012 85.451 22.576
1-1/2 CUKT015 55.160 14.573
2 CUKT020 28.605 7.5575
PIPESIZE(IN.)
FITTING515/8510-XX
U.S. GAL LITERS
STAINLESS STEEL WELDOLETS ON SCH 40 PIPE
2-1/2 CR4W025 18.800 4.9670
3 CR4W030 12.170 3.2153
4 CR4W040 6.9600 1.8388
5 CR4W050 5.2600 1.3897
6 CR4W060 3.6900 0.9749
8 CR4W080 2.1300 0.5627
10 CR4W100 1.3500 0.3567
12 CR4W120 0.9600 0.2536
CARBON STEEL WELDOLETS ON SCH 40 PIPE
2-1/2 CS4W025 18.800 4.9670
3 CS4W030 12.170 3.2153
4 CS4W040 6.9600 1.8388
5 CS4W050 5.2600 1.3897
6 CS4W060 3.6900 0.9749
8 CS4W080 2.1300 0.5627
10 CS4W100 1.3500 0.3567
12 CS4W120 0.9600 0.2536
COPPER/BRONZE BRAZOLETS ON SCH 40 PIPE
2-1/2 BR4B025 18.800 4.9670
3 BR4B030 12.170 3.2153
4 BR4B040 6.9600 1.8388
5 BR4B050 5.2600 1.3897
6 BR4B060 3.6900 0.9749
8 BR4B080 2.1300 0.5627
10 BR4B100 1.3500 0.3567
12 BR4B120 0.9600 0.2536
SCH 80 IRON SADDLES ON SCH 80 PIPE
2 IR8S020 32.360 8.5495
2-1/2 IR8S025 22.220 5.8705
3 IR8S030 13.420 3.5456
4 IR8S040 7.6600 2.0238
5 IR8S050 5.8600 1.5482
6 IR8S060 4.0900 1.0806
8 IR8S080 2.3300 0.6156
10 IR8S100 1.5300 0.4042
12 IR8S120 1.0600 0.2801
SCH 80 IRON SADDLE ON SCH 40 PIPE
2 IR8S020 26.820 7.0859
2-1/2 IR8S025 18.800 4.9670
3 IR8S030 11.990 3.1678
4 IR8S040 6.8500 1.8098
5 IR8S050 5.3300 1.4082
6 IR8S060 3.7600 0.9934
8 IR8S080 2.1300 0.5627
10 IR8S100 1.3500 0.3567
12 IR8S120 0.9600 0.2536
17Signet 8150 Flow Totalizer
ThefollowingdataisreprintedfromtheSignet525MetalexFlowsensormanualforyourconvenience.
525 Metalex Flow Sensor515 Paddlewheel Flow Sensor (continued)
PIPESIZE FITTING
515/8510-XX
U.S. GAL LITERS
POLYPROPYLENE FITTINGS (DIN/ISO AND BS AND ANSI)
DN15 PPMT005 481.55 127.23
DN20 PPMT007 277.09 73.207
DN25 PPMT010 141.18 37.300
DN32 PPMT012 83.540 22.071
DN40 PPMT015 51.265 13.544
DN50 PPMT020 29.596 7.8193
DN65 PPMT025 20.658 5.4579
DN80 PPMT030 13.330 3.5218
DN100 PPMT040 8.7077 2.3006
DN125 PPMT050 5.0667 1.3386
DN150 PPMT060 3.6892 0.9747
DN200 PPMT080 2.0398 0.5389
PVDF FITTINGS (DIN/ISO AND BS AND ANSI)
DN15 SFMT005 420.87 111.19
DN20 SFMT007 228.15 60.277
DN25 SFMT010 136.70 36.116
DN32 SFMT012 79.294 20.950
DN40 SFMT015 43.490 11.490
DN50 SFMT020 25.908 6.8450
DN65 SFMT025 18.067 4.7732
DN80 SFMT030 12.357 3.2648
DN100 SFMT040 8.0599 2.1294
DN125 SFMT050 4.4312 1.1707
DN150 SFMT060 3.2271 0.8526
DN200 SFMT080 2.0360 0.5379
PVC FITTINGS (DIN/ISO) - EUROPE ONLY
DN15 PVMT005 486.18 128.45
DN20 PVMT007 242.85 64.160
DN25 PVMT010 148.64 39.270
DN32 PVMT012 85.125 22.490
DN40 PVMT015 51.855 13.700
DN50 PVMT020 29.750 7.8600
DN65 PVMT025 17.487 4.6200
DN80 PVMT030 12.491 3.3000
DN100 PVMT040 8.1377 2.1500
DN150 PVMT060 4.0878 1.0800
DN200 PVMT080 2.0439 0.5400
SCH 40S STAINLESS STEEL PIPE PER ANSI B36.19
K-FACTOR
PIPESIZE
PULSES/U.S. GAL
PULSES/LITER
1/2IN. 873.03 230.66
3/4IN. 515.41 136.17
1IN. 266.17 70.322
11/4IN. 148.84 39.324
11/2IN. 107.98 28.528
2IN. 64.808 17.122
21/2IN. 44.685 11.806
3IN. 28.579 7.5506
4IN. 16.302 4.3070
5IN. 10.237 2.7046
6IN. 7.0057 1.8509
8IN. 3.9641 1.0473
10IN. 2.4690 0.6523
12IN. 1.6894 0.4463
K-Factor Charts
18 Signet 8150 Flow Totalizer
Troubleshooting
Display Condition Probable Cause Suggested Solutions
Batteriesaredeadormissing. Replacebothbatteries.
Pulltabshavenotbeenremoved. Removeplasticpulltabsprotectingthebatteries.
Theflowrateisgreaterthan“9999”.
1. Reducetheflowrate.2. ChangetheTimebasetoasmallervalue. (Example:Changefrom“Day”to“Hour”)3. Changetheflowunitstoalargermeasure. (Example:Changefrom“Liters”to“Gallons”) NOTE:IftheFlowK-Factorischanged,be
suretomakeacorrespondingchangetotheTOTALK-Factor.
8150isnotreceivingasignalfromtheflowsensor.
1. Thereisnoflowinthepipe.2. Flowsensorisnotturningduetoblockage
ordamage.3. Sensorwiringislooseorincorrect.
InAutoCal,thecalculatedK-Factorisoutsidetherangeofthe8150(lessthan0.001orgreaterthan99999).
Press►keytostartAutoCALprocedureagain.Makesurethattheflowrateenteredisaccurate.
Theflowratedisplayiserraticandnon-linear
Usuallycausedbyinadequatestraightpiperunupstreamofsensor.
1. Correctpipinglayouttoprovidemorestraightpipeupstreamofsensor.
2. SettheSPEEDtohighersettingtoaverageoutthefluctuationscausedbypipingconditions(seeSpeed,section14).
Bothbatteriesaretoodepletedtosafelystoresettings. Replacebattery#1,thenreplacebattery#2.
19Signet 8150 Flow Totalizer
Notes
Georg Fischer Signet LLC, 3401 Aero Jet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057For Worldwide Sales and Service, visit our website: www.gfsignet.com • Or call (in the U.S.): (800) 854-4090For the most up-to-date information, please refer to our website at www.gfsignet.com
3-8150.090RevK09/14English ©GeorgFischerSignetLLC2014
Mfr. Part No. Code Description3-8150-1 159000929 FlowTotalizer3-8150-1P 159000930 FlowTotalizer,PanelMount3-8150-P0 159000931 FlowTotalizersystem,Integral,for0.5to4in.pipes,PPhousing,BlkPVDFrotor,Titaniumpin3-8150-P1 159000932 FlowTotalizersystem,Integral,for0.5to8in.pipes,PPhousing,BlkPVDFrotor,Titaniumpin3-8150-T0 159001011 FlowTotalizersystem,Integral,for0.5to4in.pipes,naturalPVDFhousing,rotorandpin
Parts and AccessoriesMounting3-8050 159000184 Universalmountingkit3-0000.596 159000641 Heavydutywallmountbracket(forpanelmountonly)3-5000.598 198840225 Surfacemountbracket3-8050.395 159000186 Splashproofrearcoverforpanelmounttotalizer3-9000.392 159000368 Liquidtightconnectorkit(includes3connectors)3-9000.392-1 159000839 Liquidtightconnector,NPT(1connector)3-9000.392-2 159000841 Liquidtightconnector,PG13.5(1connector)3-8050.390-1 159001702 RetainingNutReplacementKit,NPT,Valox®
3-8050.390-3 159310116 RetainingNutReplacementKit,NPT,PP3-8050.390-4 159310117 RetainingNutReplacementKit,NPT,PVDF3-8050.392 159000640 Model200retrofitadaptor7400-0011 159000935 Lithiumbattery,3.6V,AAsizeNote:UseSaftLS14500LithiumBatteriesorequivalentONLY.5523-0222 159000392 Cable,twoconductorshielded,22AWG
Replacement parts for integral mount units3-8051 159000187 FlowSensorIntegralMountingKit,NPT,Valox®
3-8051-1 159001755 FlowSensorIntegralMountingKit,NPT,PP3-8051-2 159001756 FlowSensorIntegralMountingKit,NPT,PVDF3-8510-P0 198864504 Sensorfor0.5to4in.pipes,Polypropylenehousing,titaniumpin,BlackPVDFrotor3-8510-P1 198864505 Sensorfor5to8in.pipes,Polypropylenebody3-8510-T0 159000622 Sensorfor0.5to4in.pipes,naturalPVDFhousing,rotorandpin3-8510-V0 198864506 Sensorfor0.5to4in.pipes,PVDFbody
Ordering Information