slid e 2 introduction digital mammography what is ffdm qc ... · aver ag egla nd ula rd ose to a 4....
TRANSCRIPT
1
Slide 1
DigitalMammography
Detector QC
Eric A. Berns, Ph.D.Eric A. Berns, Ph.D.
[email protected] meric.bern [email protected]
Denver Health Medical CenterDenver Health Medical CenterUniversity of Colorad o at Denver and HSCUniv ersit y of Colorado at Denver and HSC
Denver, CODenver , CO
Slide 2
�� What is FFDM QC and why is itWhat is FFDM QC and why is it
important?important?
�� What to know before you startWhat to know before you start
�� Overview and compare QC testsOverview and compare QC tests
�� Key take home pointsKey take home points
IntroductionIntroduction
Slide 3
AccreditedFFDM units
CertifiedFacilities withFFDM units
Total AccreditedMammo Units
Total CertifiedMammoFacilities
+2,193
+1,375
+45
+5
Difference
5,119
3,366
13,450
8,837
July 1, 2008
2,926
1,991
13,405
8,832
July 1, 2007
Certified Statistics – Past Year
IntroductionIntroduction 8,331 SFM Units
Slide 4
FFDM vs. Date
0%
2%
4%
6%
8%
10%
12%
Ju l-98 Dec-99 Apr-01 Sep-02 Jan-04 May-05 Oct-06
% Fac w Dig
% Units FFDM
GE 2000DGE 2000DApp: Jan 2000App: Jan 2000
FischerFischer SenoscanSenoscanApp: SeptApp: Sept ‘‘0101
LoradLorad CCDCCDApp: MarApp: Mar ‘‘0202
LoradLorad SeleniaSeleniaApp: OctApp: Oct ‘‘0202
SiemensSiemens NovationNovationApp: AugApp: Aug ‘‘0404
GE DSGE DSApp: Feb 2004App: Feb 2004
GE EssentialGE EssentialApp: AprApp: Apr ‘‘0606
FujiFuji FCRmFCRmApp: JulApp: Jul ‘‘0606
2
Slide 5
�� MQSAMQSA
–– Mammography Quality Standards ActMammography Quality Standards Act
�� ACRACR
–– American College of RadiologyAmerican College of Radiology
IntroductionIntroductionSlide 6
ScreenScreen--FilmFilm
Slide 7
ScreenScreen--FilmFilm
Slide 8
ScreenScreen--FilmFilm
3
Slide 9
ScreenScreen--FilmFilm
Slide 10
ScreenScreen--FilmFilm
Slide 11
ScreenScreen--FilmFilm
Slide 12
FFDMFFDM
�� In FFDM, the manufacturer designs and mandates theirIn FFDM, the manufacturer designs and mandates theirown QC programown QC program
�� In FFDM, you must the manufacturersIn FFDM, you must the manufacturers’’ QC programQC program
4
Slide 13
RWSRWS
Slide 14
Vs.Vs.
Slide 15
Quality Control: Why?Quality Control: Why?
�� Reduce exposure to patients andReduce exposure to patients and
personnelpersonnel
�� Consistent image qualityConsistent image quality
�� Detect and correct for potential problems,Detect and correct for potential problems,
before they impact image qualitybefore they impact image quality
Slide 16
What is Quality Control?What is Quality Control?
�� Determination of what isDetermination of what is ““NormalNormal””
�� Detection of what isDetection of what is ““AbnormalAbnormal””
�� Understanding of how to return toUnderstanding of how to return to ““NormalNormal””
fromfrom ““AbnormalAbnormal””
�� In particular, in FFDM, how do you knowIn particular, in FFDM, how do you know
what you are seeing is what it is supposed towhat you are seeing is what it is supposed to
be?be?
5
Slide 17
�� Golden Rules for QCGolden Rules for QC–– Must use manufacturerMust use manufacturer’’s QC proceduress QC procedures
•• Mandate action limitsMandate action limits
–– ManufacturersManufacturers’’ QC may refer to Monitor & PrinterQC may refer to Monitor & Printer
ManufacturersManufacturers’’ QCQC
–– Multimodality Workstations have own separate QCMultimodality Workstations have own separate QC
–– Printers may have their own QCPrinters may have their own QC
–– Most failures result in stopping clinical imagingMost failures result in stopping clinical imaging
until failure can be correcteduntil failure can be corrected
Before You Begin QCBefore You Begin QCSlide 18
�� Obtain proper training & CE credits (8 hours)Obtain proper training & CE credits (8 hours)
–– HandsHands--on training onon training on actualactual unit:unit:
•• MechanicsMechanics
•• SoftwareSoftware
•• ArtifactsArtifacts
•• Learn vendor specific tests and tricksLearn vendor specific tests and tricks
Before You BeginBefore You Begin
Slide 19
�� Must have proper continuing experience:Must have proper continuing experience:
–– 2 facilities2 facilities
–– 6 units6 units
–– Within last 2 yearsWithin last 2 years
Before You BeginBefore You BeginSlide 20
Before You BeginBefore You Begin
��Note:Note:
––For new unitFor new unit: Must use most current: Must use most current
versionversion
––For renewal unitFor renewal unit: Can use older: Can use older
version (version used when testedversion (version used when tested
previously)previously)
6
Slide 21
�� ACR AccreditationACR Accreditation -- www.acr.orgwww.acr.org–– Physics formsPhysics forms
•• http://acr.org/accreditation/mammography/mammo_qc_forms.aspxhttp://acr.org/accreditation/mammography/mammo_qc_forms.aspx
–– GEGE SenographeSenographe 2000D, DS, and Essential2000D, DS, and Essential–– FischerFischer SenoscanSenoscan–– LoradLorad SeleniaSelenia–– SiemensSiemens NovationNovation–– FujiFuji FCRmFCRm
•• NoteNote: if using unit for both screen: if using unit for both screen--film & CR, you must accredit forfilm & CR, you must accredit forbothboth
�� FDA AccreditationFDA Accreditation–– www.fda.gov/cdrh/mammographywww.fda.gov/cdrh/mammography//–– Other vendors when approvedOther vendors when approved
Before You BeginBefore You BeginSlide 22
Before You BeginBefore You Begin
PersonnelEquipmentTestingTech QC FormsEquipment EvaluationAnn ual Survey Form s
Slide 23
Before You BeginBefore You Begin
GE Fischer Fuji
Lorad Siemen s Film
Slide 24
Summary FormsSummary Forms
7
Slide 25
Before You BeginBefore You BeginSlide 26
Site Report DateSurvey Date
X-Ray Unit Manufacturer GE Medical Systems ModelDate of Installation Room ID
Medical Physicist Signatu re
GE QC manual version(s) at facility (check all that apply):? 2277390-100 Rev 1, 2000 ? 2277390-100 Rev 3, 2001 ? 2277390-100 Rev 3+ Addendum 2354312-100, 2003? 2371472-100 Rev 0, 2003 ? Mobile 2371698-100Rev 0, 2003 ? SenoAdvantage 2391082-100 Rev 1, 2003
1. Flat Field (as described in radiologic technologist tests)2. Image Quality (Phantom) (as described in radiologic technologist tests)
CNR (required)Change in CNR =0.2 (NA for Equipment Evaluations)
Fibers Specks MassesPhantom IQ Test on AWSPhantom IQ Test on RWS – Left* (* NA for units with SenoPhantom IQ Test on RWS – Right* Advantage workstations)Phantom IQ Test on Printer
3. MTF Measurement (as described in radiologic technologist tests)4. AOP Mode and Signal-to-Noise (SNR) (as described in radiologic technologist tests)5. Collimation Assessment
Deviation between X-ray field and light field is less than 2% of SIDX-ray field does not extend beyond any side of the IR by more than 2% of SIDChest wall edge of compression paddle doesn't extend beyond IR by more than 1% of SID
6. Evaluation of Focal Spot Performa nceMeasured performance within acceptable limits for large focal spotMeasured performance within acceptable limits for small focal spot
7. Breast Entrance Exposure, Average Glandular Dose and Reproduci bil ityAverage glandular dose for average breast is below 3 mGy (300 mrad)Average glandular dose to a 4.2-cm-thick breast on your unit is mradExposure reproducibility (CV) for R and mAs must be less than 0.05
8. Artifact Evaluation and Flat Field Uniform ity9. Viewing Condition Check and Setting10. Monit or Calibration11. Image Quality – SMPTE Pattern12. Analysis of Review Work Station (RWS) or Seno Advantage Screen Uniformit y
(required for Equipment Evaluations and as necessary)13. kVp Accuracy and Reproducibility
Measured average kVp within ±5% of indicated kVpkVp coefficient of variation =0.02
14. Beam Quality Assessment (Half-Value Layer Measurement)15. Radiation Output
Radiation output is =800 mR/s mR/s16. Mammographic Unit Assembly Evaluati on
MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARYFull -Field Digital – GE Medical Systems
Medical Physicist's QC Tests
PASS/FAIL/NA
Senograp he 2000D
Site Report DateSurvey Date
X-Ray Unit Manufacturer Model
Date of Installation Room IDMedical Physicist Signature
Fischer QCmanual version ? P-55943-OMIssue 1, Rev. 000 (2002) ? P-55943-OM Issue 1, Rev. 1 (April 2003)? P-55943-OMIssue 1, Rev. 3 (July 2003) ? P-55943-OM Issue 1, Rev. 7 (March 2004)
1. X-Ray Field Size Al ignment and Chest Wall Missed Tissue ChecksX-ray field size aligns with field area indicated onbreast support within =2% SIDChest wall edge of the digital image to the ruler reference mark of breast support is =8.5 mm
2. Compression Paddl e Al ignme ntCompression paddle edges not visible within field of viewChest wall edge of compression paddle extends beyondimage by =1% SID
3. kVp Accuracy TestMeasured average kVp within ±5% of indicated kVpkVp coefficient of variation =0.02
4. Linearity, Reproducibi lit y and AccuracyLinearity <0.08Reproducibility <0.035 for each technique
5. Half-Value Layer and OutputHVL =0.33 mmAl at 30 kVp, 100 mA
6. Dosimetry – Average Glandul ar Dose and OutputAverage glandular dose for average breast is below 3 mGy (300 mrad)Average glandular dose to a4.2-cm-thick breast on your unit is mrad
7. Phantom Image Acquis ition TestNo obvious artifacts(ADU values required for both QCand Equipment Evaluations; P/F results are NA for Equipment Evaluations)Background StDev within +50/-0 ADUcounts of baselineBackground mean within ±100 ADUcounts of baselineADU level difference within ±300 ADUcounts of baseline
8. Image QualityLargest 4 fibers, 3 speck groups and 3 masses Fibers Specks MassesPhantom IQ Test on Review Work Station
9. System Resolution/Scan Speed Uniformit yStandard imaging mode @7 lp/mm: =5 transitions, =10% modulationHigh resolution imaging mode @ 11.1 lp/mm: =5 transitions, =5% modulation
10. Flat Field TestFlat field testDeviations betweencorner ROIs and center within ±20%
11. Geometric Distorti on and Resolut ion Uniformity12. Automatic Decompres sion Control13. System Artifacts14. Image Display Monitor(s) Check (review daily log)15. Image Viewing Room Illumi nance Test (=50 lux)
at facility (check one) :
MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARYFull-Field Digita l - Fisch er
Medical Physicist's QC Tests
PASS/FAIL/NA
Fischer SenoScan
Site Report DateSurvey Date
X-Ray Unit Manufacture r ModelDate of Installation Room IDMedical Physicist Signature
Lorad QC manual version at facil ity (check one) : ? June 11, 2002 ? 9-500-0285, Rev. 002 (2003)
? 9-500-0285, Rev. 003 (2003) ? 9-500-0285, Rev. 004 (2004)
1. Mammographic Unit Assembly EvaluationAutodecompression can be overidden to maintain compression (& status maintained)Manual emergency compression release can be activated in the event of power failure
2. Collimation Assess mentDeviation between X-ray field and light field is less than 2%of SIDX-ray field does not extend beyond any side of the IRby more than 2% of SIDChest wall edge of compression paddle doesn't extend beyond IRby more than 1%of SID
3. Artifact Evaluati onArtifacts were not apparent or not significant
4. kVp Accura cy and Reproduci bil ityMeasured average kVp within ±5% of indicated kVpkVp coefficient of variation =0.02
5. Beam Quality Assessm ent - HVL MeasurementHalf-value layer is within acceptable lower and upper limits at all kVp values tested
6. Evaluation of System Resolut ionMeasured performance within acceptable limits
7. Breast Entrance Exposure and Averag e Glandular DoseAverage glandulardose for average breast is below 3 mGy (300 mrad)Average glandulardose to a 4.2-cm-thick breast on your unit is mrad
8. Radiation Output RateRadiation output rate is greater than 800 mR/sec mR/sec
9. Phantom Image Quality Evaluat ion5 largest fibers, 4 largest speck groups and 4 largest masses are visiblePhantom image quality scores: Fibers Background Optical Density
Specks Disk Optical DensityMasses Disk Contrast
Hard copy background density must be =1.20 (with operating level =1.40)Hard copy density difference over acrylic disk must be =0.35 (with operating level =0.40)
10. Signal-To-Nois e Ratio and Contrast-To-Noi se Ratio Measuremen tSignal-To-Noise Ratio should be equal or greater to 40 SNRContrast-To-Noise Ratio should not vary by more than ±15% CNR
11. Viewbox Lumin ance and Room Illumin anceMammographic viewbox is capable of a luminance of at least 3000 cd/sq m (nit)Room illuminance (viewbox surface as seen by observer) is 50 lux or lessRoom illuminance (monitor surface) is 20 lux or less
12. Softcopy Workstat ion QCWhite level performanceBlack level performanceQuality level performanceUniformity performance
Selenia
MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARYFull-F ield Digital – Lorad
Lorad
Medical Physicist's QC Tests
PASS/FAIL
••ACR AccreditationACR Accreditation –– Equipment Evaluation FormsEquipment Evaluation Forms
Before You BeginBefore You Begin
Slide 27
Before You BeginBefore You BeginSlide 28
�� ScreenScreen--FilmFilm MammoMammo
–– Ensure locks, detents,Ensure locks, detents, angulationangulation indicators, andindicators, and
mechanical support devices operate properlymechanical support devices operate properly
�� FFDMFFDM
–– SameSame
MammoMammo Unit EvaluationUnit Evaluation
8
Slide 29
�� ScreenScreen--FilmFilm MammoMammo–– Assure XAssure X--ray field aligns with the light fieldray field aligns with the light field
–– Collimator allows full coverage of receptorCollimator allows full coverage of receptor
–– Chest wall edge of compression paddle aligns with chestChest wall edge of compression paddle aligns with chest
wall of receptorwall of receptor
�� FFDMFFDM
–– SameSame
–– MultipleMultiple targetarge/filters and/or multiple paddle positions/filters and/or multiple paddle positions
–– Chest Wall Missed TissueChest Wall Missed Tissue
–– Compression Plate Overlap on the Chest Wall SideCompression Plate Overlap on the Chest Wall Side
CollimationCollimationSlide 30
CollimationCollimation
XXCompression PlateCompression PlateOverlap on ChestOverlap on ChestWallWall
XXXXChest Wall & MissedChest Wall & MissedTissueTissue
XXXXXXMultiple paddleMultiple paddlepositionspositions
XXXXXXMultiple target/filterMultiple target/filtercombocombo’’ss
XXXXXXXXXXSame as ScreenSame as Screen--FilmFilm
SiemensSiemensNovationNovation
LoradLoradSeleniaSelenia
FischerFischerSenoscanSenoscan
GEGE2000D,2000D,
DS,DS,EssentialEssential
FujiFuji FCRmFCRm
Slide 31
�� ScreenScreen--FilmFilm MammoMammo
–– LineLine--pair test toolpair test tool
•• Focal spotFocal spot
�� FFDMFFDM
–– LineLine--pair test toolpair test tool
•• Focal spotFocal spot
•• Detector & imaging chainDetector & imaging chain
Limiting Spatial ResolutionLimiting Spatial ResolutionSlide 32
Spatial ResolutionSpatial Resolution
XXXXMTFMTF
XXXXXXMeasure using LPMeasure using LPphantom on top ofphantom on top ofacrylicacrylic
XXSame as ScreenSame as Screen--FilmFilm(Focal Spot(Focal Spot EvalEval))
SiemensSiemensNovationNovation
LoradLoradSeleniaSelenia
FischerFischerSenoscanSenoscan
GEGE2000D,2000D,
DS,DS,EssentialEssential
FujiFuji FCRmFCRm
9
Slide 33
�� Focal Spot EvaluationFocal Spot Evaluation
••Mo/MoMo/Mo
••Rh/RhRh/Rh
••Large & Small SpotsLarge & Small Spots
(1.5(1.5 magmag))
GE 2000D, DS, & EssentialGE 2000D, DS, & EssentialSlide 34
�� MTF MeasurementMTF Measurement –– MonthlyMonthly –– Raw ImageRaw Image
MTF (% ) = (Std. Dev.)x 222/ (Mean Dark ROIMT F (%) = (Std. Dev.)x 222/ (Mean Dark ROI –– Mean Light ROI)Mean Light ROI)
MTF (%) @M TF (%) @ 22 lplp/mm > 58% MTF (%) @/mm > 58% MTF (%) @ 44 lplp/mm > 25%/mm > 25%
GE 2000DGE 2000D
Slide 35
GE 2000DGE 2000D
�� 2000D Addendum: Sub2000D Addendum: Sub--system MTFsystem MTF–– Optional replacement for MTF & Focal spotOptional replacement for MTF & Focal spot
–– No need for filmNo need for film
•• Use specific bar pattern & 4.5 cm acrylic blockUse specific bar pattern & 4.5 cm acrylic block
•• Use ROI tools to calculate subUse ROI tools to calculate sub--system MTFsystem MTF
•• Repeat at 1.8Repeat at 1.8 magmag
Slide 36
�� MTF and CNR MeasurementMTF and CNR Measurement -- WeeklyWeekly–– Test consistency of CNR and ensureTest consistency of CNR and ensure
contrast is adequatecontrast is adequate•• IQST Test toolIQST Test tool
�� Action Limit:Action Limit:–– MTF @ 2MTF @ 2 lplp/mm > 58%, MTF @ 4/mm > 58%, MTF @ 4 lplp/mm > 25%/mm > 25%
–– Change in CNR must not exceed 0.2Change in CNR must not exceed 0.2
GE DS & EssentialGE DS & Essential
10
Slide 37
�� Evaluation of System ResolutionEvaluation of System Resolution–– 55--1515 lplp/mm Test Pattern/mm Test Pattern
�� The system limiting spatial resolution must be:The system limiting spatial resolution must be:> 7> 7 lplp/mm/mm
LoradLorad & Siemens& SiemensSlide 38
�� System ResolutionSystem Resolution
–– Up to 10Up to 10 lplp/mm Test Pattern/mm Test Pattern
–– 4 cm acrylic4 cm acrylic
–– Parallel & Perpendicular to tube with 3Parallel & Perpendicular to tube with 3--55
degree angledegree angle
�� The system limiting spatial resolution must be:The system limiting spatial resolution must be:
> 8> 8 lplp/mm/mm ++ 22 lplp/mm (6 to 10/mm (6 to 10 lplp/mm)/mm)
FujiFuji FCRmFCRm
Slide 39
�� ScreenScreen--FilmFilm MammoMammo
–– Measure optical densitiesMeasure optical densities
•• Different thicknesses using clinical modesDifferent thicknesses using clinical modes
•• Different density settings (Different density settings (--2,2, --1, 0, +1, +2, etc.)1, 0, +1, +2, etc.)
�� FFDMFFDM
–– Measure resultant techniquesMeasure resultant techniques
–– Measure signal, exposure, & SNR valuesMeasure signal, exposure, & SNR values
AEC PerformanceAEC PerformanceSlide 40
AECAEC
XXXXXXXXImage Mode TrackingImage Mode Tracking
XXACR ReproducibilityACR Reproducibility
XXXXXXSNRSNR
XXCNRCNR
XXXXXXDensity ControlDensity ControlFunctionFunction
Same as ScreenSame as Screen--FilmFilm
SiemensSiemensNovationNovation
LoradLoradSeleniaSelenia
FischerFischerSenoscanSenoscan
GEGE2000D,2000D,
DS,DS,EssentialEssential
FujiFuji FCRmFCRm
11
Slide 41
�� AOP Mode and SNR CheckAOP Mode and SNR Check–– Variable thicknesses of acrylicVariable thicknesses of acrylic–– Std, AutoStd, Auto–– Evaluate :Evaluate:
•• Correct techni ques?Correct techni ques ?
•• Adequate SNR?Adequate SNR?
2.5 cm2.5 cmAcrylicAcrylic
4.0 cm4.0 cmAcryli cAcryli c
6.0 cm Acryl ic6.0 cm Acrylic
AcrylicThickness
Target-Filter
SelectedkVp
SelectedmAs
2.5cm Mo-Mo 27 kVp 20-60
4.0 cm Mo-Rh 28 kVp 35-90
6.0cm Rh-Rh 32 kVp 35-90
EachEach ““ rawraw ”” image must have a measured SNR ofimage must have a measured SNR ofat least 50at least 50
GE 2000DGE 2000DSlide 42
GE DS & EssentialGE DS & Essential
�� DS specific QC testsDS specific QC tests–– AOP Mode and SNR CheckAOP Mode and SNR Check
•• Use builtUse built --in softw are for image acquis itionin softw are for image acqu is ition
•• AOP: STD/AutoAOP: STD/Auto
31314545--9595Rh/RhRh/Rh6060
29294040--9090Rh/RhRh/Rh5050
26262020--6060Mo/MoMo/Mo2525
kVkVmAsmAsTrack/FilterTrack/Filter
Exposure ParametersExposure ParametersFor AOP STD modeFor AOP STD mode olyoly
Acrylic ThicknessAcrylic Thickness(mm)(mm)
SNR must be > 50SNR must be > 50
Slide 43
LoradLorad SeleniaSelenia
�� Automatic Exposure ControlAutomatic Exposure Control
–– AUTOAUTO--FILTERFILTER –– 2, 4, 6, 8 cm2, 4, 6, 8 cm
–– 4 cm (4 cm (--5 thru +5, or5 thru +5, or --3 to +4)3 to +4)
–– Use ROI to measure mean pixel valueUse ROI to measure mean pixel value
inside virtual AEC detectorinside virtual AEC detector
�� Performance CriteriaPerformance Criteria
–– Pixel value should not vary by morePixel value should not vary by more
than 10% of meanthan 10% of mean
–– Exposure compensation steps shallExposure compensation steps shall
meet requirements in pixel value tablemeet requirements in pixel value table
Slide 44
�� AEC Image Stability and Reproducibility and SNRAEC Image Stability and Reproducibility and SNR
–– ACR PhantomACR Phantom
–– Mo/Mo, 28 kV,Mo/Mo, 28 kV, ““HH””, sensor 1, sensor 1
–– RecordRecord mAsmAs, SNR, Mean, Entrance Exposure (5 times), SNR, Mean, Entrance Exposure (5 times)
�� Action Limits:Action Limits:
–– CoefCoef. of. of VarVar forfor mASmAS and X < 5%and X < 5%
–– SNR and Mean shall not vary bySNR and Mean shall not vary by ++ 15% of Mean15% of Mean
SiemensSiemens NovationNovation
12
Slide 45
�� AEC Thickness TrackingAEC Thickness Tracking
–– 2, 4, and 6 cm2, 4, and 6 cm
–– ““ HH”” modemode
–– Max deviat ion = (Max difference / mean value ) * 100Max deviat ion = (Max differenc e / mean value ) * 100
SiemensSiemens NovationNovation DRDRSlide 46
�� Density Control FunctionDensity Control Function
–– 4 cm acrylic, ACR phantom technique4 cm acrylic, ACR phantom technique
–– Repeat atRepeat at --2,2, --1, 0, +1, +2 etc. density1, 0, +1, +2 etc. density
–– RecordRecord mAsmAs
–– mAsmAs change should be 5 to 15% per stepchange should be 5 to 15% per step
FujiFuji FCRmFCRm
Slide 47
�� Reproducibility & Image Mode TrackingReproducibility & Image Mode Tracking–– 4 cm acrylic with clinical technique4 cm acrylic with clinical technique
–– Position ion chamber in beamPosition ion chamber in beam
–– RecordRecord mAsmAs and Xand X
–– Repeat 3 timesRepeat 3 times
–– Repeat in each mode (small, large,Repeat in each mode (small, large, magmag, no grid), no grid)
–– Action LimitsAction Limits::
•• Coefficient of variation for Exposure orCoefficient of variation for Exposure or mAsmAs must not exceedmust not exceed
0.050.05
•• No significant difference in exposure between small and largeNo significant difference in exposure between small and large
buckybucky when using similar gridswhen using similar grids
FujiFuji FCRmFCRmSlide 48
�� CNR Per Object ThicknessCNR Per Object Thickness
–– CNR using clinical technique for 2 cmCNR using clinical technique for 2 cm
–– Repeat for 4 and 6 cmRepeat for 4 and 6 cm
–– Action LimitsAction Limits::
•• CNR of 2 cm relative to 4 cm must beCNR of 2 cm relative to 4 cm must be >> 100%100%
•• CNR of 6 cm relative to 4 cm must beCNR of 6 cm relative to 4 cm must be >> 75%75%
FujiFuji FCRmFCRm
13
Slide 49
�� ScreenScreen--FilmFilm MammoMammo
–– Measure if ODMeasure if OD’’s are consistents are consistent
–– Check forCheck for artifactsartifacts
�� FFDMFFDM
–– FlatFlat--field uniformityfield uniformity
–– Detector calibrationDetector calibration
–– Pixel correction testPixel correction test
–– CR Reader Scanner PerformanceCR Reader Scanner Performance
Uniformity of Screen Speed &Uniformity of Screen Speed &Detector PerformanceDetector Performance
Slide 50
Detector PerformanceDetector Performance
XXPixel CorrectionPixel Correction
XXGeometric DistortionGeometric Distortion
XXDynamic RangeDynamic Range
XXCR Reader ScannerCR Reader ScannerPerformancePerformance
XXXXGhostingGhosting
XXInterInter--PlatePlateConsistencyConsistency
XXPrimary ErasurePrimary Erasure
XXXXXXDetector CalibrationDetector Calibration
XXXXFlatFlat--Field UniformityField Uniformity
Same as ScreenSame as Screen--FilmFilm(Screen Uniformity)(Screen Uniformity)
SiemensSiemensNovationNovation
LoradLoradSeleniaSelenia
FischerFischerSenoscanSenoscan
GEGE2000D,2000D,
DS,DS,EssentialEssential
FujiFuji FCRmFCRm
Slide 51
�� Dynamic RangeDynamic Range
–– To confirm the dynamic range of reader and plateTo confirm the dynamic range of reader and plate
–– Use 2 & 4 cm acrylicUse 2 & 4 cm acrylic
–– Technique for 6 cmTechnique for 6 cm
–– View image to determine if 3 regions are visibleView image to determine if 3 regions are visible
and discernableand discernable
FujiFuji FCRmFCRmSlide 52
�� CR Reader Scanner PerformanceCR Reader Scanner Performance
–– Establish that IP reader & optics do not exhibit scan orEstablish that IP reader & optics do not exhibit scan or
print jitterprint jitter
–– NonNon--grid exposuregrid exposure
–– Position rulers inPosition rulers in ““TT”” formationformation
–– Expose & processExpose & process
–– Action Limit:Action Limit: Image must have smooth borders freeImage must have smooth borders free
from jagged edges or defectsfrom jagged edges or defects
FujiFuji FCRmFCRm
14
Slide 53
�� Primary Erasure (Additive & Multiplicative)Primary Erasure (Additive & Multiplicative)–– To assess the erasure performance of reader andTo assess the erasure performance of reader and
platesplates
–– Additive:Additive: Shoot phantom, process, wait 1 minute,Shoot phantom, process, wait 1 minute,
reprocess, change S value to 10X original,reprocess, change S value to 10X original,
inspect for visibility of phantom imageinspect for visibility of phantom image
–– Multiplicative:Multiplicative: Shoot & process phantom, shoot 4Shoot & process phantom, shoot 4
cm acrylic on same cassette, process and inspectcm acrylic on same cassette, process and inspect
for visibility of phantom imagefor visibility of phantom image
FujiFuji FCRmFCRmSlide 54
�� InterInter--Plate ConsistencyPlate Consistency
–– To confirm xTo confirm x--ray absorption & SNR consistencyray absorption & SNR consistency
–– Expose plates using clinical technique with 4cmExpose plates using clinical technique with 4cm
acrylicacrylic
–– RecordRecord mASmAS (must be within(must be within ++10%)10%)
–– Calculate SNR (must be withinCalculate SNR (must be within ++15%)15%)
FujiFuji FCRmFCRm
Slide 55
�� FlatFlat--field uniformityfield uniformity
–– Test to ensure detector performance acceptableTest to ensure detector performance acceptable
•• Measures detector uniformity (signal & noise)Measures detector uniformity (signal & noise)
•• Measures bad pixelsMeasures bad pixels
–– System automatically calculates pass/failSystem automatically calculates pass/fail
QAPQAP
GE 2000D, DS, & EssentialGE 2000D, DS, & EssentialSlide 56
�� Detector CalibrationDetector Calibration –– WeeklyWeekly
–– 4 cm acrylic block4 cm acrylic block –– no paddleno paddle
–– Built in Calibration softwareBuilt in Calibration software
–– Follow onFollow on--screen instructionsscreen instructions
•• Review image for artifactsReview image for artifacts
•• Repeat until you have 8 imagesRepeat until you have 8 images
•• End calibrationEnd calibration
•• Calibration is performed automaticallyCalibration is performed automatically
LoradLorad SeleniaSelenia
15
Slide 57
�� Detector UniformityDetector Uniformity
–– Make exposure: 28 kV, 50Make exposure: 28 kV, 50 mASmAS ,,
Mo/MoMo/Mo
–– Select ROI 128 x 128 pixelsSelect ROI 128 x 128 pixels
–– Record measurem ents in cornersRecor d measur ements in corners
and centerand center
–– Pixel value insi de each of 5Pixel value insi de each of 5 ROIROI’’ss
shall not diff er by more than 10%shall not diff er by more than 10%
of meanof mean
SiemensSiemens NovationNovation DRDRSlide 58
�� Detector Calibration TestDetector Calibration Test
–– Objective: to determine if systemObjective: to determine if system
was correctly gain calibratedwas correctly gain calibrated
–– Must be done weeklyMust be done weekly
(technologist)(technologist)
–– Follow onFollow on--screen instructionsscreen instructions
–– Acquire 8 imagesAcquire 8 images
–– Images should be free of artifactsImages should be free of artifacts
SiemensSiemens NovationNovation
Slide 59
�� Pixel Correction TestPixel Correction Test
–– Calibration modeCalibration mode
–– Create new pixel mapCreate new pixel map
–– Follow onFollow on--screen instructionsscreen instructions
–– Acquire imageAcquire image
–– Evaluate image is free fromEvaluate image is free from
collimator or debris artifactscollimator or debris artifacts
–– System compiles new pixel mapSystem compiles new pixel map
–– System display resultsSystem display results –– mustmust
meet requirementsmeet requirements
SiemensSiemens NovationNovation DRDRSlide 60
�� ScreenScreen--FilmFilm MammoMammo
–– Evaluate cassettesEvaluate cassettes
�� FFDMFFDM
–– Evaluate detectorEvaluate detector
–– CR: Imaging PlatesCR: Imaging Plates
ArtifactsArtifacts
16
Slide 61
ArtifactsArtifacts
XXXXXXXXXXExpose Using AcrylicExpose Using Acrylic
Same as ScreenSame as Screen--FilmFilm
SiemensSiemensNovationNovation
LoradLoradSeleniaSelenia
FischerFischerSenoscanSenoscan
GEGE2000D,2000D,
DS,DS,EssentialEssential
FujiFuji FCRmFCRm
Slide 62
ArtifactsArtifacts
Slide 63
Artifact evaluationArti fact evaluati on
Mo/MoMo/Mo Mo/Mo/RhRh Rh/RhRh/Rh
ArtifactsArtifactsSlide 64
�� ScreenScreen--FilmFilm MammoMammo
–– ACR Phantom ScoresACR Phantom Scores
–– Optical Density & ContrastOptical Density & Contrast
�� FFDMFFDM
–– ACR Phantom ScoresACR Phantom Scores
•• Pass/fail requirements differ by vendorPass/fail requirements differ by vendor
–– SignalSignal--toto--Noise Ratio (SNR)Noise Ratio (SNR)
–– ContrastContrast--toto--Noise Ratio (CNR)Noise Ratio (CNR)
Image QualityImage Quality
17
Slide 65
Image QualityImage Quality
XXXXXXXXXXCNRCNR
PartialPartialXXXXClinical TechniqueClinical Technique
XXXXManual TechniquesManual Techniques
Same as ScreenSame as Screen--FilmFilm
SiemensSiemensNovationNovation
LoradLoradSeleniaSelenia
FischerFischerSenoscanSenoscan
GEGE2000D,2000D,
DS,DS,EssentialEssential
FujiFuji FCRmFCRm
Slide 66
ACR Phantom ImagingACR Phantom Imaging
4433MassesMasses
4433SpecksSpecks
5544FibersFibers
LoradLorad &&SiemensSiemens
GE & Fischer &GE & Fischer &FujiFuji
Image QualityImage Quality
Slide 67
�� ACR Phantom ImagingACR Phantom Imaging
–– Manual technique (Mo/Mo, 26Manual technique (Mo/Mo, 26
kVpkVp, 125, 125 mASmAS ))
–– Score theScore the process edproc essed imageimage
–– Acquisition workstationAcquisition work stati on
–– Each monitor of the RWSEach monitor of the RWS
–– Laser imagerLaser imager
GE 2000DGE 2000DSlide 68
�� ContrastContrast--toto--Noise Test (CNR)Noise Test (CNR)
–– To examine consistency of CNRTo examine consistency of CNR
ratio measured over timeratio measured over time
–– Use theUse the rawraw imageimage
–– ++ 20% of baseline20% of baseline
Backg roun d ROIBackg roun d ROI Mass ROIMass ROI
CNR = (CNR = (MeanMeanbackgroundbackground -- MeanMeanmassmass)/SD)/SDbackgroundbackground
GE 2000DGE 2000D
18
Slide 69
GE DS & EssentialGE DS & Essential
�� Phantom Image QualityPhantom Image Quality
–– Manual technique:Manual technique: Rh/RhRh/Rh , 29, 29 kVpkVp, 56, 56 mAsmAs
�� MTF and CNR MeasurementMTF and CNR Measurement
–– Use IQST test toolUse IQST test tool
–– Use builtUse built --in software for image acq uisitionin software for image acq uisition
–– Manual technique:Manual technique: Rh/RhRh/Rh 3030 kVpkVp, 56, 56 mAsmAs
–– Results are autom atically displaye d (pass/fail )Result s are autom atically displayed (pass/fai l)
–– Same action li mits as 2000DSame acti on limits as 2000D
Slide 70
�� Phantom Image Qual ityPhantom Image Quali ty–– Select clinical exposu re mode (i.e. AUTOSelect clinic al exposu re mode (i.e. AUTO--FILTER)FILTER)
–– Print filmPrint film
–– Measure backgrou nd OD and density diff erenceMeasure backgroun d OD and density diff erence ––
Plot on tech worksheet sPlot on tech work sheets
•• Backgroun d must beBackgr oun d must be >> 1.20 OD1.20 OD ++ 0.200.20
•• DD mus t beDD mus t be >> 0.400.40 ++ 0.050.05
–– Score on each Sof t Copy Workstatio nScore on each Soft Copy Workstation
•• 5 fibers5 fibers
•• 4 speck groups4 speck gro ups
•• 4 masses4 masses
LoradLorad SeleniaSelenia
Slide 71
�� Phantom Image Qualit yPhantom Image Quality
–– Position phantom 1 cmPosition phantom 1 cm
over chest wall edg eover chest wall edge
–– Select: 28 kV, AECSelect: 28 kV, AEC--Auto,Auto,
Mo/MoMo/Mo
–– Score on RWS or FilmScore on RWS or Film
–– FiberFiber >> 55
–– Speck sSpecks >> 44
–– MassesMasses >> 44
SiemensSiemens NovationNovation DRDRSlide 72
�� SNR and CNR MeasurementsSNR and CNR Measurements
–– SNRSNR at leastat least >> 4040
–– CNRCNR should stay withinshould stay within ±±15% of baseline15% of baseline
•• Obtained during acceptance testingObtained during acceptance testing
LoradLorad & Siemens& Siemens
19
Slide 73
�� ContrastContrast--toto--Noise Test (CNR)Noise Test (CNR)
–– To examine consistency of CNR ratio measured over timeTo examine consistency of CNR ratio measured over time
–– Use 4 cm acrylic & 0.2 mm AlUse 4 cm acrylic & 0.2 mm Al
–– Manual technique (Mo/Mo, 26Manual technique (Mo/Mo, 26 kVpkVp, 125, 125 mAsmAs))
–– Calculate CNR using softwareCalculate CNR using software
–– ++ 20% of baseline20% of baseline
FujiFuji FCRmFCRmSlide 74
�� ScreenScreen--FilmFilm MammoMammo
–– Assure that actualAssure that actual kVpkVp isis
•• Accurate (Accurate (++ 5% of indicated)5% of indicated)
•• Reproducible (Reproducible (CoeffCoeff. Var.. Var. << 2%)2%)
�� FFDMFFDM
–– SameSame
kVpkVp
Slide 75
�� ScreenScreen--FilmFilm MammoMammo
–– Assure HVL is adequate to minimize patientAssure HVL is adequate to minimize patient
dosedose
–– HVL shall meetHVL shall meet FDAFDA’’ss Performance StandardsPerformance Standards
�� FFDMFFDM
–– SameSame
HVLHVL
Note the lead sheetNote the lead sheet
Slide 76
�� ScreenScreen--FilmFilm MammoMammo
–– Dose for single CC view of ACR phantom shallDose for single CC view of ACR phantom shall
not exceed 3.0not exceed 3.0 mGymGy per exposureper exposure
�� FFDMFFDM
–– SameSame
DoseDose
20
Slide 77
�� ScreenScreen--FilmFilm MammoMammo
–– System shall be capable of producing aSystem shall be capable of producing a
minimum output of 7.0minimum output of 7.0 mGymGy/sec @ 28/sec @ 28 kVpkVp
Mo/Mo over 3.0 second periodMo/Mo over 3.0 second period
�� FFDMFFDM
–– SameSame
–– S value confirmationS value confirmation
–– Exposure LinearityExposure Linearity
Radiation OutputRadiation OutputSlide 78
�� ScreenScreen--FilmFilm MammoMammo
–– Luminance (Luminance (>> 3000 cd/m3000 cd/m22))
–– RoomRoom illuminanceilluminance ((<< 5050 luxlux))
–– ViewboxViewbox cleaningcleaning
�� FFDMFFDM
–– SameSame
Viewing & Viewing ConditionsViewing & Viewing Conditions
Slide 79
�� ScreenScreen--FilmFilm MammoMammo
–– Assess siteAssess site’’s QC programs QC program
–– Check that QC tests are properly performedCheck that QC tests are properly performed
–– Independent checks of tech QC testsIndependent checks of tech QC tests
�� FFDMFFDM
–– SameSame
Assessing Tech QCAssessing Tech QCSlide 80
�� ScreenScreen--FilmFilm MammoMammo
–– Measure optical densitiesMeasure optical densities
•• Density difference, midDensity difference, mid--density,density, base+fogbase+fog
•• Measures consistencyMeasures consistency
�� FFDMFFDM
–– ManufacturersManufacturers’’ recommendationsrecommendations
–– Some refer to printer manufacturersSome refer to printer manufacturers’’ recommendationsrecommendations
–– Typically identical to ACR SFM ManualTypically identical to ACR SFM Manual
Film ProcessingFilm Processing
21
Slide 81
Film Processor QCFilm Processor QC
XXXXFollowFollow FFDMsFFDMs’’ QCQC
XXXXXXIf Not, Use TheirsIf Not, Use Theirs
XXXXXXFollow PrinterFollow PrinterManufacturers QCManufacturers QC
SiemensSiemensNovationNovation
LoradLoradSeleniaSelenia
FischerFischerSenoscanSenoscan
GEGE2000D,2000D,
DS,DS,EssentialEssential
FujiFuji FCRmFCRm
Slide 82
�� MidMid--DensityDensity
Film ProcessingFilm Processing
Northwestern Memorial Hospital - Proce ssor ID: 2 - MAR 2006Mid-Density
1.30
1.40
1.50
1.60
1.70
1.80
1.90
2.00
2.10
2.20
2.30
1-Mar-0
6
2-M
ar-0
6
3-M
ar-06
6-Mar-0
6
7-M
ar-0
6
8-Mar-0
6
9-M
ar-0
6
10-Mar-0
6
13-M
ar-0
6
14-Mar-0
6
serv
i ce3/1
4
15-M
ar-0
6
16-M
ar-06
17-M
ar-0
6
20-Mar-0
6
21-M
ar-0
6
22-Mar-0
6
23-M
ar-06
24-M
ar-0
6
27-M
ar-06
28-M
ar-0
6
29-Mar-0
6
30-M
ar-06
31-M
ar-0
6
Date
Mid-D ensi ty Upper Contr ol Limit = 2.06
Mid-D ensi ty Daily Value
Mid-D ensi ty Daily Value Oper ating Lev el = 1.91
Mid-D ensi ty Lower Control Limit = 1.76
MD Step=12
Nor thwester n Memorial Hospital - Proces sor ID: Kodak 8900 - Oct 2006Mid-Density
0.70
0.80
0.90
1.00
1.10
1.20
1.30
1.40
1.50
1.60
1.70
Oct
ober 2, 200
6
Oct
ober 3, 200
6
Oct
ober4,
2006
Oct
ober5,
2006
Oct
ober6,
2006
Oct
ober9,
2006
October 10,
2006
October 11,
2006
October 12,
2006
October 13,
2006
October 16,
2006
October 17,
2006
October 18,
2006
October 19,
2006
October 20,
2006
October
23,20
06
October
24,20
06
October
25,20
06
October
26,20
06
October
27, 2
006
October
30, 2
006
Date
Mid-Density Upper Control Limit = 1.48
Mid-Density Daily Valu e
Mid-Density Daily Valu e Operating Level = 1.33
Mid-Density Lower Control Limi t = 1.18
Wet Processing Laser Printer
Slide 83
RWS QCRWS QC
XXXXUse FFDMUse FFDMmanufacturersmanufacturers’’ss QCQC
XXXXXXIf Not, UseIf Not, Use FFDMFFDM’’ss
XXXXXXFollow monitorFollow monitormanufacturers QCmanufacturers QC
SiemensSiemensNovationNovation
LoradLoradSeleniaSelenia
FischerFischerSenoscanSenoscan
GEGE2000D,2000D,
DS,DS,EssentialEssential
FujiFuji FCRmFCRm
Slide 84
�� AGFA IMPAX MA3000AGFA IMPAX MA3000–– Clean monitor surf aceClean monitor surfa ce–– Measure display white & bla ckMeasure display white & black–– Measure quality levelMeasure quality leve l–– Measure uniformi tyMeasure uniformity–– Calibrate displaysCalibrate displays–– View SMPTE patternView SMPTE pattern–– Viewing conditionsViewing conditions
�� GE Seno AdvantageGE Seno Ad vantage–– Viewing con ditions check and sett ingViewing conditions check and setting–– Monitor calibr ationMonitor calibrati on–– Image qualityImage quali ty –– SMPTE patter nSMPTE pattern–– Anal ysis of scr een uniformi tyAnalysis of scr een unifo rm ity
�� SiemensSiemens Mammo ReportMammoReport PlusPlus–– Geometr ic distortio nGeometric distorti on–– ReflectionReflec tion–– Luminance ResponseLuminance Respons e–– Luminance Uniformit yLuminance Uniformi ty–– ResolutionResolution–– NoiseNoise–– Overall EvaluationOverall Evaluation
22
Slide 85
�� BarcoBarco–– II--Guard CheckGuard Check–– Quality LevelQuality Level–– Display WhiteDisplay White–– Calibr ation Setti ngs CheckCalibration Setti ngs Check
�� EizoEizo–– Pattern CheckPattern Check–– Lumina nce CheckLuminance Check–– Graysca leGrays cale CheckCheck–– Unifor mity CheckUniformity Check
�� PlanarPlanar–– Monitor CleaningMonitor Cleanin g–– Viewing Cond itions CheckViewing Conditi ons Check–– Monitor Calibration CheckMonitor Calibrati on Check–– Image QualityImage Quality –– SMPTESMPTE–– Visual Insp ecti on for DisplayVisual Inspection for Display
DefectsDefec ts
Slide 86
�� Tungsten w/ Silver or RhodiumTungsten w/ Silver or Rhodium
–– CalibrationCalibration: FDA requires all air: FDA requires all air kermakerma measuringmeasuring
instruments be calibrated to NISTinstruments be calibrated to NIST--traceabletraceable
standard for at least one targetstandard for at least one target--filter combination infilter combination in
thethe mammomammo xx--ray rangeray range……,,
–– Do not require such a calibration for W/Ag or W/Do not require such a calibration for W/Ag or W/RhRh
–– May require corrections for solid state instrumentsMay require corrections for solid state instruments
New FDA Hot TopicsNew FDA Hot Topics
Slide 87
�� Tungsten w/ Silver or RhodiumTungsten w/ Silver or Rhodium
–– kVpkVp MeasurementsMeasurements::
•• Scenario 1Scenario 1: If using: If using kVpkVp meter certified for W/Ag ormeter certified for W/Ag or
W/W/RhRh and aand a mammomammo calibrated ion chamber, then okcalibrated ion chamber, then ok
•• Scenario 2Scenario 2: If not, you need to include a correction: If not, you need to include a correction
factor infactor in kVpkVp measurements obtained frommeasurements obtained from kVpkVp metermeter
manufacturer, or, by working with service engineermanufacturer, or, by working with service engineer
and comparing measurementsand comparing measurements
•• Scenario 3Scenario 3: If not W certified and using a separate: If not W certified and using a separate
solidsolid--state probe, then follow scenario 2.state probe, then follow scenario 2.
New FDA Hot TopicsNew FDA Hot TopicsSlide 88
�� Tungsten w/ Silver or RhodiumTungsten w/ Silver or Rhodium
–– HVL MeasurementsHVL Measurements::
•• Scenario 1Scenario 1: If using integrated solid: If using integrated solid--state instrumentstate instrument
that is manufactured for W/Ag or W/that is manufactured for W/Ag or W/RhRh, then ok., then ok.
•• Scenario 2:Scenario 2: If using integrated solidIf using integrated solid--state instrumentstate instrument
that isthat is notnot manufactured for W/Ag or W/manufactured for W/Ag or W/RhRh, then, then
obtain necessary correction factors.obtain necessary correction factors.
New FDA Hot TopicsNew FDA Hot Topics
23
Slide 89
�� Use of Foam Cushions on FFDMUse of Foam Cushions on FFDM
––FDA recommends using pads for QCFDA recommends using pads for QC
if used clinicallyif used clinically
––Required if manufacturerRequired if manufacturer’’s QCs QC
manuals specifies the pad be presentmanuals specifies the pad be present
during QCduring QC
New FDA Hot TopicsNew FDA Hot TopicsSlide 90
�� QC testing for PrintersQC testing for Printers
–– Follow FFDM manufacturerFollow FFDM manufacturer’’s QC Manuals QC Manual
–– Then, follow printerThen, follow printer’’s QC manuals QC manual
–– To determine the appropriate testingTo determine the appropriate testing
New FDA Hot TopicsNew FDA Hot Topics
Slide 91
�� The current ACR QC Manual was written forThe current ACR QC Manual was written for
screenscreen--filmfilm–– Most does not apply to digitalMost does not apply to digital
�� Subcommittee on QA is writing new manualSubcommittee on QA is writing new manual–– ChairChair -- EricEric BernsBerns, Ph.D., Ph.D.
–– MembersMembers –– Radiologists, technologists, medical physicists, MITARadiologists, technologists, medical physicists, MITA
repsreps
�� Standardize QC tests, performance criteria andStandardize QC tests, performance criteria and
frequencies across all systemsfrequencies across all systems
�� Clinical section specific to digital issuesClinical section specific to digital issues
ACR FFDM QC Manual ProjectACR FFDM QC Manual ProjectSlide 92
�� QC testsQC tests–– Will apply to all manufacturers and modelsWill apply to all manufacturers and models
–– Fewer tests and written to beFewer tests and written to be ““tech friendlytech friendly””
–– New phantom to be more applicable to digital (but usable withNew phantom to be more applicable to digital (but usable with
screenscreen--film)film)
–– When ready, draft will be sent to manufacturers for their inputWhen ready, draft will be sent to manufacturers for their input beforebefore
it is finalit is final
�� When final, ACR will apply for FDA alternativeWhen final, ACR will apply for FDA alternative
standardstandard
�� Hope to have draft completed by end of 2008Hope to have draft completed by end of 2008
ACR FFDM QC Manual ProjectACR FFDM QC Manual Project
24
Slide 93
�� Archive QC images periodically on a CDArchive QC images periodically on a CD
�� Perform flatPerform flat--field imaging before phantomfield imaging before phantom
imaging testsimaging tests
�� Perform testing in the morningPerform testing in the morning
�� Have a system for monitoring theHave a system for monitoring the ““AutoAuto
Print/Auto PushPrint/Auto Push”” functionsfunctions
SuggestionsSuggestionsSlide 94
�� Repeat testRepeat test–– CleanClean plexiglassplexiglass & move slightly& move slightly
�� Questions to ask yourself:Questions to ask yourself:–– Are you using the correct image? Raw vs.Are you using the correct image? Raw vs.
Processed?Processed?
–– Are theAre the ROIROI’’ss in the correct location?in the correct location?
�� ReRe--boot systemboot system
�� Call serviceCall service–– ReRe--calibration solves most image quality problemscalibration solves most image quality problems
Troubleshooting QCTroubleshooting QC
Slide 95
�� Obtain relevant handsObtain relevant hands--on trainingon training
�� Must perform manufacturer specific QC testsMust perform manufacturer specific QC tests
�� ArtifactsArtifacts –– most problems can be seen heremost problems can be seen here
�� ReRe--booting and/or rebooting and/or re--calibrating fixes mostcalibrating fixes most
problemsproblems
�� Laser PrinterLaser Printer
–– DDmaxmax at least 3.5 ODat least 3.5 OD
–– MidMid--density about 1.5 ODdensity about 1.5 OD
Key Take Home PointsKey Take Home PointsSlide 96
�� Review workstation monitorsReview workstation monitors –– look at thelook at the
clinical images!clinical images!–– Do they match?Do they match?
–– Appropriate dark and light levelsAppropriate dark and light levels
�� Be aware of separate QC for printers &Be aware of separate QC for printers &
monitorsmonitors
�� Know, and be involved with, your site QCKnow, and be involved with, your site QC
programprogram
Key Take Home PointsKey Take Home Points
25
Slide 97
SAM QuestionsSAM Quest ions
Digital Mammography QCDigital Mammography QC
EricEric BernsBerns, PhD, PhD
Slide 98
In digital mammography, whoIn digital mammography, whomandates the pass /fail criteria for sitemandates the pass/fail crit eria for siteQC?QC?
20%
20%
20%
20%
20%
10
1.1. The American College of RadiologyThe American College of Radiology2.2. The FDAThe FDA3.3. The FFDM unit manufacturerThe FFDM unit manufacturer4.4. NEMANEMA5.5. MQSAMQSA
Slide 99
Answer: 3Answer: 3 -- The FFDM uni tThe FFDM uni tmanufac turermanufacturer
ReferencesReferences
�� MQSA Regulations 900.12(e)(6)MQSA Regulations 900.12(e)(6)�� http://www.fda.gov/CDRH/MAMMOGRAhttp://www.fda.gov/CDRH/MAMMOGRA
PHY/frmamcom2.html#s90012PHY/frmamcom2.html#s90012
Slide 100
Answe r 3: The FFDM unitAnswer 3: The FFDM unitmanufacturermanufa cturer
ExplanationExplanation
““the quality assurance program shall bethe quality assurance program shall besubstantially the same as the qualitysubstantially the same as the qualityassurance program recommended by theassurance program recommended by theimage receptor manufacturer, except thatimage receptor manufacturer, except thatthe minimum allowable dose for screenthe minimum allowable dose for screen--film systems in this sectionfilm systems in this section””
26
Slide 101
How do you accredit a Fuji CRHow do you accredit a Fuji CRMammograph y system which usesMammograp hy system which usesboth CR and screenboth CR and screen --film on the samefilm on the samexx--ray system?ray system?
25%
25%
25%
25%
10
1.1. As 1 mammography unit?As 1 mammography unit?
2.2. As 2 mammography units?As 2 mammography units?3.3. You cannot use CR and film on theYou cannot use CR and film on the
same unit.same unit.4.4. CR is exempt from accreditationCR is exempt from accreditation
Slide 102
Answer: 2Answer: 2 –– As 2 mammograph yAs 2 mamm ographyunitsuni ts
ReferencesReferences
�� http://www.acr.org/accreditation/mammography/http://www.acr.org/accreditation/mammography/mammo_faq/mammo_faq_mamac.aspx#8.0.1mammo_faq/mammo_faq_mamac.aspx#8.0.1
Slide 103
Answe r: 2Answer: 2 -- As 2 mammographyAs 2 mammographyunitsunits
ExplanationExplanation
As of November 15, 2006 facilities usingAs of November 15, 2006 facilities usingboth screenboth screen--film and CR on the samefilm and CR on the samemammography units must accredit thesemammography units must accredit these2 systems as 2 separate units.2 systems as 2 separate units.
Slide 104
What are the min imum passi ng ACRWhat are the mini mum passin g ACRphantom scores for thephantom scores for the Lora dLorad SeleniaSelenia forforweekly QC?weekly QC?
20%
20%
20%
20%
20%
10
1.1. 5 Fibers, 4 Speck Groups, 3 Masses5 Fibers, 4 Speck Groups, 3 Masses2.2. 4 Fibers, 3 Speck Groups, 3 Masses4 Fibers, 3 Speck Groups, 3 Masses3.3. 5 Fibers, 4 Speck Groups, 4 Masses5 Fibers, 4 Speck Groups, 4 Masses4.4. 4 Fibers, 4 Speck Groups, 3 Masses4 Fibers, 4 Speck Groups, 3 Masses5.5. 4 Fibers, 4 Speck Groups, 4 Masses4 Fibers, 4 Speck Groups, 4 Masses
27
Slide 105
Answer 3Ans wer 3:: 5 Fibers, 4 Speck Grou ps, 45 Fibers, 4 Speck Group s, 4MassesMasses
ReferencesReferences
�� HologicHologic.. LoradLorad SeleniaSelenia Quality ControlQuality ControlManual MANManual MAN--00093, Revision 003,00093, Revision 003,Bedford (MA):Bedford (MA): HologicHologic, 2005, 2005
Slide 106
Answer 3Answe r 3:: 5 Fibers, 4 Speck Grou ps, 45 Fibers , 4 Speck Groups, 4MassesMasses
ExplanationExplanation
TheThe LoradLorad SeleniaSelenia QC manual states that the ACRQC manual states that the ACRPhantom minimum passing scores are asPhantom minimum passing scores are asfollows:follows:
�� 5 Fibers5 Fibers�� 4 Speck Groups4 Speck Groups�� 4 Masses4 Masses
Slide 107
To meet the FDA requirement for continuingTo meet the FDA requiremen t for continuingexperienc e, how many mammo grap hyexperience , how many mammog raphyfacilities and mammography uni t surveysfacilities and mammo grap hy unit surveysmust be performed wi th in the previous 24must be per for med with in the previous 24months?months?
20%
20%
20%
20%
20%
10
1.1. 6 Facilities and 2 mammography units6 Facilities and 2 mammography units2.2. 4 Facilities and 12 mammography units4 Facilities and 12 mammography units3.3. 2 Facilities and 6 mammography units2 Facilities and 6 mammography units4.4. 1 Facilities and 6 mammography units1 Facilities and 6 mammography units5.5. 2 Facilities and 12 mammography units2 Facilities and 12 mammography units
Slide 108
Answer 3Answe r 3:: 2 Facilities and 6 Mammography2 Facilities and 6 MammographyUni tsUni ts
ReferencesReferences
�� http://www.fda.gov/CDRH/MAMMOGRAPHY/robohelp/http://www.fda.gov/CDRH/MAMMOGRAPHY/robohelp/START.HTMSTART.HTM
�� 900.12(a)(3)(iii)(B): Continuing experience. Following900.12(a)(3)(iii)(B): Continuing experience. Followingthe second anniversary date of the end of the calendarthe second anniversary date of the end of the calendarquarter in which the requirements of paragraphsquarter in which the requirements of paragraphs(a)(3)(i) and (a)(3)(ii) of this section were completed or(a)(3)(i) and (a)(3)(ii) of this section were completed orof April 28, 1999, whichever is later, the medicalof April 28, 1999, whichever is later, the medicalphysicist shall have surveyed at least twophysicist shall have surveyed at least twomammography facilities and a total of at least sixmammography facilities and a total of at least sixmammography units during the 24 months immediatelymammography units during the 24 months immediatelypreceding the date of the facilitypreceding the date of the facility’’s annual MQSAs annual MQSAinspection or the last day of the calendar quarterinspection or the last day of the calendar quarterpreceding the inspection or any date in between thepreceding the inspection or any date in between thetwo. The facility shall choose one of these dates totwo. The facility shall choose one of these dates todetermine the 24determine the 24--month period. No more than onemonth period. No more than onesurvey of a specific facility within a 10survey of a specific facility within a 10--month period ormonth period ora specific unit within a period of 60 days can bea specific unit within a period of 60 days can becounted towards this requirement.counted towards this requirement.