Slit lamp

By

praveen.p

D.optom,B.optm

Page 2: Slit lamp

Slit lamp are used in the examination of the anterior segment of the eye .

Where they allow the practitioner to perform a careful examination of cornea ,conjunctiva and lids.

In modern instrument the microscope and light source are coupled so that the light shines on the same part of the eye that the microscope is focused .

Page 3: Slit lamp

This coupling facilitates examination of the eye .

The instrument is called a slit lamp because in the normal mode of operation .

The light source produces a thin vertical slit of light at the eye .

When this slit of light shines along a different axis to that of the microscope the clinician see a section of living eye.

Page 4: Slit lamp

Main parts of slit lamp

1. Silt lamp microscopes2. Illumination systems .3. Mechanical coupling .

Page 5: Slit lamp
Page 6: Slit lamp

Attachments of slit lamp

Fundus examination Gonioscopy TonometryLaser photo coagulation Measuring the visual acuity in

patients with hazy media .

Page 7: Slit lamp

Slit lamp microscopes

The magnification of slit lamp microscope is usually with in the range of *6 to * 40 .

At higher magnifications the small and often uncontrollable movement of the patient’s eye make the image jump around so much that an examination cannot be performed satisfactorily .

This range of magnifications and the large working distance mean that compound rather than simple microscope have to be used .

Page 8: Slit lamp

In most basic form composed of microscopes are composed of two optical elements .

An objective and an eyepiece .[in the fig shown as two simple positive lenses.

In reality the eye piece and occasionally the objective are composed of a series of lenses to reduce aberrations .

Page 9: Slit lamp

Range of magnifications

1. The use of different objectives.2. The use of different eyepieces. 3. The Littmann- Galilean telescope

principle .4. With a zoom system .

Page 10: Slit lamp

The use of different objectives

This is one of the oldest and possibly still the most frequently used technique for obtaining different magnifications .

The different objectives are usually placed on a turret type of arrangement that allows them to be fairy rapidly changed during an examination .

Page 11: Slit lamp

The system is usually limited to two sets of objective lenses due to the confinements of space in and around the objective area of slit-lamp.

Page 12: Slit lamp

The use of different eye pieces

This technique is usually used in the cheaper slit lamps it is the only means of changing the magnification .

It is not a very convenient technique as it requires the practitioner to pull out the current eye pieces and replace them

Normally they are kept in a drawer attached to the instrument table .

Page 13: Slit lamp

More than two pairs of eye pieces to be provide with the slit lamp .

The technique alone provides a very limited range of magnifications.

Page 14: Slit lamp

The littmann Galilean telescope principle

The Galilean magnification changer developed by littmann [1950]

It is completely separate optics that sit nearly b\w the objective and eye piece lenses and does not require either of them to change .

It provides a larger range of magnifications than the other technique .

Page 15: Slit lamp

Typically five via a turrent arrangement which is completely enclosed within the microscope’s body .

It is called Galilean system because it utilizes Galilean telescopes to alter the magnification .

The Galilean telescopes have two optical components a positive and a negative lens .

Page 16: Slit lamp

The parallel light rays both enters and leaves the system and under goes some degree of magnification which is dependent upon the power and separation of two lens .

Galilean telescope fits with in the standard slit lamp microscope along with a rely lens in the manner .

By reversing the order of the lenses in the telescope a different magnification can be achieve with out altering any optical elements.

Page 17: Slit lamp

Zoom systems

Recently expensive slit-lamps have been produced with a zoom systems that allow a continuously variable degree of magnification .

The Nikon instrument contains the zoom system with in the objective of the microscope and offers a range of magnifications from *7 to *35.

Page 18: Slit lamp

Illumination systems

The objective of the slit-lamp illumination system is to produce a bright light evenly illuminated ,finely focused ,adjustable slit of light at the eye .

Almost all slit –lamp manufactures have adopted the koller illumination system .

Page 19: Slit lamp

Koller illumination system

The filament of the bulb is imaged by the condenser lenses at or close to projector lens .

The projector lens forms an image of the slit at the eye .

The diameter of the projection lens is usually fairly small in size.

This has two advantage :- first is keep the aberrations of the lens down ,which results in a better quality image .

Page 20: Slit lamp

Second :- it increases the depth of focus of the slit and thereby produces a better optical section of the eye .

While the optics of the condenser are not as critical as that of the other elements of the slit – lamp care must be taken in their design so as not to introduce too much chromatic aberration which will tend to cause fingers at the slit image .

Page 21: Slit lamp

To reduce the aberration is normally achieved by using two or more lenses in the condenser system .

The light source used in slit –lamp illumination systems is a tungsten filament bulb .

For safety reasons ,they are run on relatively low voltages

Slit-lamp can now be obtained with halogen filled bulbs.

Page 22: Slit lamp
Page 23: Slit lamp

The diag slit-lamp

It use a optic fiber to convey the light from a high –power halogen bulb .

It is housed in a special fan cooled box to the slit lamp .

Page 24: Slit lamp

Advantage

Decrease the weight of slit- lamp Allow it to be hand held by the

practitioner .

Page 25: Slit lamp

Mechanical coupling of microscope and illumination system

Microscope and illumination system slit-lamp are linked together around a common axis or rotation that co insides with their focal plans.

Microscope focused where the illumination system is focused .

Without changing the focus the microscope and the illumination system can rotated around their axis .

Page 26: Slit lamp

The movement towards or away from the eye and from side to side is usually achieved via a joy stick .

Up and down movement is obtained via some screw device .

Page 27: Slit lamp