smart dynamic strain recorder dc-204r/dc-204ra v2 speed sampling 200khz strain dc voltage...

6
High speed sampling 200kHz Strain DC voltage Temperature SMART DYNAMIC STRAIN RECORDER Standard With analog output DC-204R DC-204R a Automatic data storing at power interruption onboard UPS Large data recording 2GB capacity TML Pam E-442B TML

Upload: halien

Post on 26-Mar-2018

213 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: smart dynamic strain recorder DC-204R/DC-204Ra v2 speed sampling 200kHz Strain DC voltage Temperature SMART DYNAMIC STRAIN RECORDER Standard With analog output DC-204R DC-204Ra Automatic

High speed sampling200kHz

StrainDC voltageTemperature

SMART DYNAMIC STRAIN RECORDERStandard     With analog output

DC-204R   DC-204Ra

Automatic datastoring at power interruptiononboardUPS

Large datarecording2GBcapacity

TML Pam E-442BTML

kikaku_mina
siro
Page 2: smart dynamic strain recorder DC-204R/DC-204Ra v2 speed sampling 200kHz Strain DC voltage Temperature SMART DYNAMIC STRAIN RECORDER Standard With analog output DC-204R DC-204Ra Automatic

The DC-204R/DC-204Ra is a compact flash recording type 4-channel dynamic strain recorder and measures strain, DC voltage and thermocouples. The DC-204Ra equips an analog output of ±5V to enable, waveform output to external recorders, display devices and automobile ECU. As compared with the previous model DC-104R/DC-104Ra, the frequency response is upgraded to 10kHz and sampling speed to 200kHz at the fastest. At the same time of measurement, measured data are automatically stored on a compact flash card up to 2GB. Using the controller DC-7204 as bundled software, the maximum 8 units (32 channels) including the DC-104 series can be used together. In addition to numerical monitor and waveform display, dynamically variable amount can be displayed in analog form and in real time. The battery-operating all-in-one model is optimum for measurement on a high speed moving body. AC operation needs a power adaptor as an option.

The recorder is compatible with a compact flash card adapter of 32MB to 2GB. The formatted flash cards are available from TML.

Measured data can be stored in not only CSV format but BIN format conforming to commercial software DADiSP for data analysis of dynamic phenomena. The bundled software CONTROLLER DC-7204 can get the data stored in CSV and BIN format.

The DC-204 series has a frequency response of DC to 10kHz. The maximum sampling speed is 200kHz with one channel. Reading of the stored data from the CF card during measurement and monitoring the measuring waveform data in real time are possible. (N.B. The sampling speed is restricted.)

The DC-204 series bundled software CONTROLLER DC-7204 can monitor dynamically variable amount in analog form and in real time with bar graph and circular graph as well as numeric values and waveform.

DADiSP is available as a separate software package from DSP Development Corporation in the USA.

It is compatible with a flash memory card of 32MB to 2GB in memory capacity. The formatted CF cards are available from TML. TML does not guarantee operation of commercial CF cards.

In case of accidental power degradation during measurement, the built-in UPS circuit stops measurement and gets measured data automatically and safely stored on the CF card. Even if power supply is suddenly interrupted, the power switch is designed to turn off after recording the measured data on the CF card.

Using the supplied analog output cable (CR-3610), the DC-204Ra can output 4-channel live waveform to external recorders and display devices.

It is as small as 15.7x8.4cm like postcard and 4.2cm in height, suitable for vehicle onboard measurement. The 4-channel unit can be connected in parallel up to 8 units (total 32 channels).

* 1 channel use

The highest sampling speed is 5µ sec.* with one channel and the measured data are recorded on a specified CF memory card at the same speed. The relationship of recording time (measuring time ) with sampling speed is tabulated below. The measuring time depends on the data size.

Sampling frequency Sampling speed 1 channel 2 channels 4 channels

200kHz 5μs 1.3 hours —   —   100kHz 10μs 2.7 hours 1.3 hours —   50kHz 20μs 5.5 hours 2.7 hours 1.3 hours 10kHz 100μs 27.4 hours 13.5 hours 6.7 hours 5kHz 200μs 55.4 hours 27.4 hours 13.5 hours 1kHz 1ms 11.4 days 5.6 days 67.5 hours 500Hz 2ms 23.1 days 11.4 days 5.6 days 50Hz 20ms 231 days 114 days 56.3 days 20Hz 50ms 572 days 281 days 140 days 10Hz 100ms 1155 days 572 days 281 days

In case of a 8M data size continuously recorded on a 2GB CF card

Page 3: smart dynamic strain recorder DC-204R/DC-204Ra v2 speed sampling 200kHz Strain DC voltage Temperature SMART DYNAMIC STRAIN RECORDER Standard With analog output DC-204R DC-204Ra Automatic

Windows PCUSB cableCR-6182(Standard accessory)

ControllerDC-7204

USB Hub (commercial product)

Synchronous cable CR-6181 (Option)

Possible connection up to 8 units of DC-104or DC-204 series(Mixing available)

Measurement with 1 unit

Measurement with 2 or more units

Analog output

DC-204Ra(DC-104Ra)Analog output toexternal devices

Analog output cable CR-3610

Recorders

Oscillographs

Digital displaydevices

Automobile ECU, etc.

Measurement Software

Controller DC-7204 (Standard accessory)

● Main menu

● Setting up Measurement Conditions

● Monitoring

● Data Processing

Main menu at thetime of start

Click the MeasureSetting to open dia-log for measurementconditions.

Set measuring capa-city, unit, coefficientetc.

Click the MonitoringSetting.Dynamically variableamount during mea-surement can beviewed in numeric,waveform, bar-graph,cirular-graph, etc.

Measured data arewritten directly on CFcard.Data format is BINor CSV.

The measured datasaved on CF card are read. Processingsuch as printout,display of maximum,minimum and meanvalues and cutoput is done.

Circular-graph monitor Bar-graph monitor

Related ProductsMeasurement Software Visual LOG● DC-7630 Dynamic strain recorder measurement and Data Processing Software

● DFA-7610 FFT Analysis and Processing Software

In addition to operation extended channel function, on-line or off-line automatic measurement can be made. An function of extract-ing /omitting measured data is provided. Graphic display in real time during sampling is possible.

Various analysis processing for data file gotten by the supplied software DC-7204 or optional software DC-7630 is done.

● Carring Case CA-10A

This is an aluminum carrying case to store the recorder, strain connectors, cables, etc.

Size: 400(W) x 260(D) x 100(H) mm

Page 4: smart dynamic strain recorder DC-204R/DC-204Ra v2 speed sampling 200kHz Strain DC voltage Temperature SMART DYNAMIC STRAIN RECORDER Standard With analog output DC-204R DC-204Ra Automatic

DC Power Supply DC10~16VThe supplied cable CR-1310 is used.

Battery driving in combination with the battery pack BA-104 (option) is possible.

Battery driving

AC Power Supply

A stain gauge bridge configuration connector for the DC-104/DC-204 series recorders

SB-120DD-1R 120 ohms 3-wire quarter bridge 120/350 ohms half bridgeSB-350DD-1R 350 ohms 3-wire quarter bridge 120/350 ohms half bridge SB-120DD-4R 120/350 ohms full bridge

Available is a remote sense cable (Option) to compensate for sensitivity degradation due to cable extension.

Used for connecting strain gauge based transducers (for force, displacement, pressure, acceleration, etc.) with a NDIS plug at the cable.

Used when inputting voltage signal. The measuring range of voltage signal is ±20V. The signal is attenu-ated to 1/1000 for input.

It enables temperature measurements using thermo-couples type T or K. No external power supplies is needed. The adapter is connected through the supplied sensor input conversion cable CR-6180.

Strain gauges

Strain gauge basedtransducers

DC voltage

Thermocouple(temperature)

Strain Connectors

Sensor input conversion cable CR-6180 (Standard accessory)

Attenuatioin cable CR-4010

Thermocouple adapter TA-01KT

In case of connecting 2 or more units of the DC-104R/DC-204R series, the use of the synchronous cable CR-6181 (option) makes simultaneous sampling possible. In this case, power supply is necessary for all the connected recorders.

Synchronous cable CR-6181

External Start/Stop and External Trigger

External control signals for start, stop and trigger are input to the OONTROL IN connector.

By connecting the supplied output cable CR-3610 to the OUTPUT connector, a voltage output of 0 ~ Å}5V is obtained. Voltage signals for 4 channels from the one connector cable can be output to external recorders, etc.

Using the supplied USB cable CR-6182, the DC-104R/DC-204R recorders can be hooked up to a computer.

Page 5: smart dynamic strain recorder DC-204R/DC-204Ra v2 speed sampling 200kHz Strain DC voltage Temperature SMART DYNAMIC STRAIN RECORDER Standard With analog output DC-204R DC-204Ra Automatic

Compact flash memory card

Status LEDSTART/STOP

TML specified CF card is used. The memory capacity is up to 2GB.

Each operational status is indicated.MES /MeasuringUSB /Communicating with USB cableSYNC /Synchronizing with 2 or more recorders.PWR /Switching on

Gets started or stopped.

BAL/HLDBy depressing the key 3 seconds or more, Balancing functiions for the model itself.

USB InterfaceThe USB cable (CR-6182) is plugged in. Measurement setup and data are transferred to a PC.

Power supplyOperates on DC power supply. The battery and recorder are connected with the power cable (CR-1310). For AC operation, use the optional AC adapter set CR-1860.

ControlFor synchronization of 2 or more record-ers. The optional synchronous cable (CR-6181) is needed.

Power supply

Input channelStrain gauges and strain gauge based transducers are plugged in through the supplied sensor input conversion cable (CR-6180). The voltage input attenuation cable CR-4010 (option) and thermo-couple adapter TA-01KT (option) can be also connected.

Analog output 【DC-204Ra only】The DC-204Ra has an analog output. By connecting the supplied analog output cable (CR-3610), a live wave form is output to an external recording device.

Page 6: smart dynamic strain recorder DC-204R/DC-204Ra v2 speed sampling 200kHz Strain DC voltage Temperature SMART DYNAMIC STRAIN RECORDER Standard With analog output DC-204R DC-204Ra Automatic

Measurement Measuring points 4 Inputs Strain DC voltage with optional cable CR-4010 Strain measurement

Gauge resistance 120Ω、350Ω(Full bridge) Bridge excitation DC2V、0.5V、ON/OFF control Measuring range 1000、2000、5000、10000、20000×10-6 strain in

5 ranges Maximum range ±80000×10-6 strain Accuracy ±0.2%FS Voltage measurement 【using CR-4010 option】 Measuring range 1、2、5、10、20V in 5 ranges Maximum range 20V Accuracy 0.3%FS Lowpass filter 10、30、100、300Hz、1k、3k、10kHz(Bessel type)Frequency response DC~ 10kHz (-3dB±1dB)

Temperature coefficient Stability on zero 1×10-6 strain/ºC with maximum sensitivity Stability on span 0.01%FS/ºC with maximum sensitivity Balancing method Electronics Balancing range 10000×10-6 strain Balancing accuracy 0.06%FS Balancing speed Approx. 1 sec. per channel A/D converter 16-bit successive approximation Resolution 1000×10-6 strain range : 1 or 0.1×10-6 strain 2000×10-6 strain range : 1 or 0.1×10-6 strain

5000×10-6 strain range : 1×10-6 strain 10000×10-6 strain range : 1×10-6 strain 20000×10-6 strain range : 1×10-6 strain Sampling method Simultaneous sampling Sampling speed 5µs in 1 channel mode (200kHz) 10µs in 2 channel mode (100kHz) 20µs in 4 channel mode ( 50kHz)Function Start/Stop Panel key operation, External signal Computer operation available Manual trigger External signal, Computer operation available Balance/Open check Panel key operation, Computer operation available Synchronous measurement Synchronous sampling available with 8 units 32

channels Combination use with DC-104R/DC-104Ra

available Displaly unit LED status Power、Synchronization、USB、MeasurementSetupProcedure By USB communication with bundled software

Channel 1、2、4 channel mode Trigger Trigger mode SINGLE、CONTINUE、FREERUN Trigger level (0%~ 100%) with 0.1% division for full scale

Data save Saving media Compact FLASH memory card 32M~2GB supplied from TML N.B.: High speed sampling use card is 128M~2GB

supplied from TML. Format Conforms to DADiSP/2000 Saving speed 5µs�1、10µs�2、20µs、50µs、100µs、200µs、500µs、

1ms、2ms、5ms、10ms、20ms、50ms、 (Sampling intervals) 100ms、200ms、500ms、1s �1:in 1-ch use only �2:in 1-、or 2-ch use only Other with up to 4 channels Data size 1k、2k、4k、8k、16k、32k、64k、128k、256k、512k

1M、2M、3M、4M、6M、8M、12M、16M、24M、32M、48M、60M、120M�3、240M�4[Data]

�3:in 1-ch use only �4:in 1-、or 2-ch use only Other with up to 4 channels For high speed sampling, the size is restricted to 64k

data over in 1-ch use, 32k data over in 2-ch use, and 16k data over in 4-ch use.

Pre-area Changeable by 10% division for 0~100% while 256k data in 1-ch is at maximum.

File numbers Maximum 255Analog output 【DC-204Ra only available】 Voltage output ±5V(5kΩ load) or ±1mA Output accuracy ±0.3%FS Output balancing ±5mV or less(at 5000×10-6 strain) accuracy Stability on zero ±2mV/ºC(at maximum sensitivity) Calibration output ±1V Lowpass filter 10、30、100、300Hz、1k、3k、10kHz (amplitude flat) S/N ratio 46dBp-p or over (with filter of 1kHz or less, 1000

×10-6 strain)General Specifications Power supply 10~ 16V dc DC-204R 0.4A MAX. DC-204Ra 0.4A MAX. Environment 0ºC~+50ºC 85%RH or less(no condensation) Vibration tolerance 49m/s2(5~ 50Hz)in 3 directions Dimensions 84(W)×42(H)×157(D)mm Weight 500g■ Product for CE marking is available on request.

1…………………………………… launam noitarepO1…………………………… 0131-RC elbac rewoP CD4……………………………… 0816-RC elbac rosneS1………………………………… 2816-RC elbac BSU

Output cable CR-3610【DC-204Ra only】 1………… Compact Flash Memory card High speed type (512MB) 1……………………… Measuring software Controller DC-7204 (CD-ROM) 1………………………………………

Standard accessory

Specifications are subject to change without prior notice.