U.S. Food & Drug Administration 10903 New Hampshire Avenue Doc ID# 04017.02.02 Silver Spring, MD 20993 www.fda.gov Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building, Yuquan road, Nanshan Shenzhen 518051, Guangdong CHINA Re: K173058 Trade/Device Name: P10 Series Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: November 15, 2017 Received: November 20, 2017 Dear Toki Wu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

Upload: others

Post on 07-Mar-2021

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

U.S. Food & Drug Administration 10903 New Hampshire Avenue D o c I D # 0 4 0 1 7 . 0 2 . 0 2 Silver Spring, MD 20993 www.fda.gov

Sonoscape Medical Corp.℅ Toki WuRegulatory Affairs Manager4/f, 5/f, 8/f, 9/f & 10/fYizhe building, Yuquan road, NanshanShenzhen 518051, GuangdongCHINA

Re: K173058Trade/Device Name: P10 Series Digital Color Doppler Ultrasound SystemRegulation Number: 21 CFR 892.1550Regulation Name: Ultrasonic pulsed doppler imaging systemRegulatory Class: IIProduct Code: IYN, IYO, ITXDated: November 15, 2017Received: November 20, 2017

Dear Toki Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

Page 2: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

Page 2 - Toki Wu K173058

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email ([email protected]) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Robert Ochs, Ph.D.DirectorDivision of Radiological HealthOffice of In Vitro Diagnosticsand Radiological Health

Center for Devices and Radiological Health

Enclosure

Page 3: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120Expiration Date: January 31, 2017See PRA Statement below.

510(k) Number (if known)

Device NameP10 Series Digital Color Doppler Ultrasound System

Indications for Use (Describe)The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human ServicesFood and Drug AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) [email protected]

“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”

Page 4: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 1 of38

Diagnostic Ultrasound Indications for Use Form

System: SonoScape P10 Exp P10 P10Pro P11 M10 P9 Diagnostic Ultrasound Pulsed Echo System Diagnostic Ultrasound Pulsed Doppler Imaging System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

Fetal P P P P P Note 1 Notes 2,4,5 Abdominal P P P P P Note 1 Notes 2,4,5 Intra-operative Specify Intra-operative Neuro Laparoscopic P P P P P Note 1 Notes 2,4 Pediatric P P P P P Note 1 Notes 2,4

Small Organ (specify) P P P P P Note 1 Notes 2,4,5,6,7

Neonatal Cephalic P P P P P P Note 1 Notes 2,3,4 Adult Cephalic P P P P P P Note 1 Notes 2,3,4 Trans-rectal P P P P P Note 1 Notes 2,4,5 Trans-vaginal P P P P P Note 1 Notes 2,4,5 Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) P P P P P Note 1 Notes 2,4

Musculo-skeletal (Superficial) P P P P P Note 1 Notes 2,4

Intravascular Other (Ob/GYN) P P P P P Note 1 Notes 2,4,5 Other (Urology) P P P P P Note 1 Notes 2,4,5

Cardiac

Cardiac Adult P P P P P P Note 1 Notes 2,3,4 Cardiac Pediatric P P P P P P Note 1 Notes 2,3,4 Intravascular(Cardiac) Trans-esoph.(Cardiac) P P P P P P Note 1 Notes 2,3,4 Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel P P P P P Note 1 Notes 2,4 Cerebral vascular P

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 5: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 2 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: C322 Curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

Fetal P P P P P Note 1 Notes 2,4 Abdominal P P P P P Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) P P P P P Note 1 Notes 2,4 Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 6: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 3 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 3C-A Curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

Fetal P P P P P Note 1 Notes 2,4 Abdominal P P P P P Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) P P P P P Note 1 Notes 2,4 Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 7: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 4 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: L741 Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric

Small Organ (specify) N N N N N Note 1 Notes 2,4,6,7

Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) N N N N N Note 1 Notes 2,4

Musculo-skeletal (Superficial) N N N N N Note 1 Notes 2,4

Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel N N N N N Note 1 Notes 2,4 Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 8: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 5 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: C1-6 Curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

Fetal P P P P P Note 1 Notes 2,4 Abdominal P P P P P Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) P P P P P Note 1 Notes 2,4 Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 9: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 6 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: VC6-2 Curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

Fetal P P P P P Note 1 Notes 2,4,5Abdominal P P P P P Note 1 Notes 2,4,5Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) P P P P P Note 1 Notes 2,4,5Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 10: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 7 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: C613 Micro-curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal P P P P P Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric P P P P P Note 1 Notes 2,4 Small Organ (specify) Neonatal Cephalic P P P P P P Note 1 Notes 2,3,4Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric P P P P P P Note 1 Notes 2,3,4Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 11: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 8 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 2P1 Phased Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Application Mode of Operation

General (TRACK 1

ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric N N N N N Note 1 Notes 2,4 Small Organ (specify) Neonatal Cephalic N N N N N N Note 1 Notes 2,3,4Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric N N N N N N Note 1 Notes 2,3,4Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 12: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 9 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: S1-5 Phased Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal P P P P P Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic P P P P P P Note 1 Notes 2,3,4 Adult Cephalic P P P P P P Note 1 Notes 2,3,4 Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult P P P P P P Note 1 Notes 2,3,4 Cardiac Pediatric P P P P P P Note 1 Notes 2,3,4 Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 13: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 10 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 7P-B Phased Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric N N N N N Note 1 Notes 2 Small Organ (specify) Neonatal Cephalic N N N N N N Note 1 Notes 2 Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric N N N N N N Note 1 Notes 2 Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 14: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 11 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: L746 Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) N N N N N Note 1 Notes 2,6,7Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) N N N N N Note 1 Notes 2

Musculo-skeletal (Superficial) N N N N N Note 1 Notes 2

Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel N N N N N Note 1 Notes 2 Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 15: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 12 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: L742 Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric

Small Organ (specify) P P P P P Note 1 Notes 2,4,6,7

Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) P P P P P Note 1 Notes 2,4

Musculo-skeletal (Superficial) P P P P P Note 1 Notes 2,4

Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel P P P P P Note 1 Notes 2,4Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 16: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 13 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 10I2 Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) P P P P P Note 1 Notes 2,4,6Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) P P P P P Note 1 Notes 2,4

Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel P P P P P Note 1 Notes 2,4Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 17: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 14 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 9L-A Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric

Small Organ (specify) P P P P P Note 1 Notes 2,4,6,7

Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) P P P P P Note 1 Notes 2,4

Musculo-skeletal (Superficial) P P P P P Note 1 Notes 2,4

Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel N N N N N Note 1 Notes 2,4Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 18: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 15 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 12L-B Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric

Small Organ (specify) P P P P P Note 1 Notes 2,4,6,7

Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) P P P P P Note 1 Notes 2,4

Musculo-skeletal (Superficial) P P P P P Note 1 Notes 2,4

Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel P P P P P Note 1 Notes 2,4Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 19: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 16 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 6V1 Micro-curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal P P P P P Note 1 Notes 2,4Trans-vaginal P P P P P Note 1 Notes 2,4Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology) P P P P P Note 1 Notes 2,4

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 20: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 17 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 6V3 Micro-curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal P P P P P Note 1 Notes 2,4Trans-vaginal P P P P P Note 1 Notes 2,4Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology) P P P P P Note 1 Notes 2,4

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 21: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 18 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 6V7 Micro-curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal P P P P P Note 1 Notes 2,4Trans-vaginal P P P P P Note 1 Notes 2,4Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology) P P P P P Note 1 Notes 2,4

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 22: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 19 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: EC9-5 Micro-curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal P P P P P Note 1 Notes 2,4Trans-vaginal P P P P P Note 1 Notes 2,4Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology) P P P P P Note 1 Notes 2,4

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 23: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 20 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: BCC9-5 Micro-curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal P P P P P Note 1 Notes 2,4Trans-vaginal P P P P P Note 1 Notes 2,4Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology) P P P P P Note 1 Notes 2,4

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 24: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 21 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: BCL10-5 Biplane Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal P P P P P Note 1 Notes 2,4Trans-vaginal P P P P P Note 1 Notes 2,4Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology) P P P P P Note 1 Notes 2,4

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 25: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 22 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: MPTEE Phased Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac)

Trans-esoph.(Cardiac) P P P P P P Note 1 Notes 2,3,4

Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 26: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 23 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: MPTEE mini Phased Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac)

Trans-esoph.(Cardiac) P P P P P P Note 1 Notes 2,3,4

Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 27: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 24 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: LAP7 Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Application Mode of Operation

General (TRACK 1

ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic P P P P P Note 1 Notes 2,4Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 28: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 25 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: PWD2.0 Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic PTrans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular P

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 29: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 26 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 13L-A Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric

Small Organ (specify) P P P P P Note 1 Notes 2,4,6,7

Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) P P P P P Note 1 Notes 2,4

Musculo-skeletal (Superficial) P P P P P Note 1 Notes 2,4

Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel P P P P P Note 1 Notes 2,4Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 30: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 27 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 18L-A Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric

Small Organ (specify) P P P P P Note 1 Notes 2,4,6,7

Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) P P P P P Note 1 Notes 2,4

Musculo-skeletal (Superficial) P P P P P Note 1 Notes 2,4

Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel P P P P P Note 1 Notes 2,4Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 31: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 28 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 10L-I Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) P P P P P Note 1 Notes 2,4,6Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) P P P P P Note 1 Notes 2,4

Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel P P P P P Note 1 Notes 2,4Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 32: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 29 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 6V3A Micro-curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal P P P P P Note 1 Notes 2,4Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 33: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 30 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 12LT-A Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) P P P P P Note 1 Notes 2,4,6Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) P P P P P Note 1 Notes 2,4

Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel P P P P P Note 1 Notes 2,4Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 34: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 31 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 12LI-A Linear Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) P P P P P Note 1 Notes 2,4,6Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) P P P P P Note 1 Notes 2,4

Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel P P P P P Note 1 Notes 2,4Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 35: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 32 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 6CI-A Curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

Fetal P P P P P Note 1 Notes 2,4 Abdominal P P P P P Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) P P P P P Note 1 Notes 2,4 Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 36: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 33 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: 6CT-A Curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

Fetal P P P P P Note 1 Notes 2,4 Abdominal P P P P P Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) P P P P P Note 1 Notes 2,4 Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 37: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 34 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: CWD2.0 Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult PCardiac Pediatric PIntravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 38: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 35 of38

Diagnostic Ultrasound Indications for Use FormTransducer: 3P-A Phased Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal N N N N N Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic N N N N N N Note 1 Notes 2,3,4 Adult Cephalic N N N N N N Note 1 Notes 2,3,4 Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult N N N N N N Note 1 Notes 2,3,4 Cardiac Pediatric N N N N N N Note 1 Notes 2,3,4 Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 39: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 36 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: C542 Micro-curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

FetalAbdominal N N N N N Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric N N N N N Note 1 Notes 2,4 Small Organ (specify) Neonatal Cephalic N N N N N N Note 1 Notes 2,3,4Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric N N N N N N Note 1 Notes 2,3,4Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 40: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 37 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: C361 Curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

Fetal N N N N N Note 1 Notes 2,4 Abdominal N N N N N Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) N N N N N Note 1 Notes 2,4 Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 41: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

Indications for Use Page 38 of38

Diagnostic Ultrasound Indications for Use Form

Transducer: C351 Curved Array Diagnostic Ultrasound Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Operation General

(TRACK 1 ONLY)

Specific (TRACKS 1 & 3) B M PWD CWD

Color Doppler

Power (Amplitude)

Doppler

Other* Combined

Other* Specify

Ophthalmic Ophthalmic

FetalImaging& Other

Fetal N N N N N Note 1 Notes 2,4 Abdominal N N N N N Note 1 Notes 2,4 Intra-operative Specify Intra-operative Neuro Laparoscopic Pediatric Small Organ (specify) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph.(non-Card) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Ob/GYN) N N N N N Note 1 Notes 2,4 Other (Urology)

Cardiac

Cardiac Adult Cardiac Pediatric Intravascular(Cardiac) Trans-esoph.(Cardiac) Intra-cardiac Other (specify)

Peripheral Vessel

Peripheral vessel Cerebral vascular

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M; B/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D Note 5: 4D Note 6: Small Organ: breast, thyroid, testes Note 7: Elastography

Page 42: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

510(k) Summary Page 1 of 8

510(k) Summary

1. Submitter [21 CFR807.92 (a) (1)]

Submitter: SONOSCAPE MEDICAL CORP.

Address: 4/f, 5/f, 8/f, 9/f & 10/f, Yizhe Building, Yuquan Road,

Nanshan, Shenzhen 518051, Guangdong, China

Contact Person: Toki Wu

Tel: +86 755 26722890

Fax: +86 755 26722850

Email: [email protected]

Date Prepared Sep 25, 2017

2. Device [21 CFR807.92 (a) (2)]

Trade Name: P10 Series Digital Color Doppler Ultrasound System

Common Name: Diagnostic Ultrasound System and Transducers

Classification Regulatory:

FR Number Product Code

Ultrasonic Pulsed Doppler Imaging System 892.1550 90-IYN

Ultrasonic Pulsed Echo Imaging System 892.1560 90-IYO

Diagnostic Ultrasound Transducer 892.1570 90-ITX

Classification Panel: Radiology

Device Class: II

3. Predicate Device(s) [21 CFR 807.92(a) (3)]

The identified predicate device within this submission is as follows:

SonoScape P50 Digital Color Doppler Ultrasound System K170999

4. Device Description [21 CFR 807.92(a) (4)]

This SonoScape P10 series Digital Color Doppler Ultrasound System is an integrated

preprogrammed color ultrasound imaging system, capable of producing high detail

resolution intended for clinical diagnostic imaging applications.

The basic principle is that system transmits ultrasonic energy into patient body and

implements post processing of received echoes to generate onscreen display of

Page 43: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

510(k) Summary Page 2 of 8

anatomic structures and fluid flow within the body.

This system is a Track 3 device that employs a wide array of probes that include linear

array, convex array and phased array with a frequency range of 1.0 MHz to 25.0 MHz.

This system consists of a mobile console with touch screen and keyboard control panel,

power supply module, color LCD monitor and optional probes.

This system is a mobile, general purpose, software controlled, color diagnostic

ultrasound system. Its basic function is to acquire ultrasound data and to display the

image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler,

Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination

of these modes, Elastography, 3D/4D.

5. Intended Use [21 CFR 807.92(a) (5)]

The device is a general-purpose ultrasonic imaging instrument intended for use by a

qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast,

testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral

Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac

(pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

6. Comparison with the Predicate device [21 CFR 807.92(a) (6)]

P10 Series Digital Color Doppler Ultrasound System is comparable with and

substantially equivalent to the predicate device:

SonoScape P50 Digital Color Doppler Ultrasound System K170999

Intended Use Comparison:

Compared with the predicate device SonoScape P50 (K170999), the Subject Device

P10 Series has the same intended use.

Technical Characteristics Comparison:

The basic and key technical features of the Subject Device P10 Series are the same as

the predicate device SonoScape P50 (K170999), including Design, Operation Controls,

Display Modes, Operation Modes, Measurement Items, Cine Loop, Operating and

Storage Condition.

The detailed technical features can be found in General Device Descriptions of this

submission.

Probes Comparison:

Subject device P10 Series has the similar probes as the predicate device SonoScape

P50 (K170999).

Page 44: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

510(k) Summary Page 3 of 8

Table 1 Probes Comparison

ID Comparison Items

Predicate device SonoScape P50 Exp/P50

Pro/P50S/P50/P47/P40 Exp/P40

Pro/P40S/P40/P37/P30 Exp/P30 Pro/P30S/P30/P27

Subject device

SonoScape P10 Exp,P10,P10

Pro,P11,M10,P9

Remark

10 Probes

C322 Micro-curved Array 3C-A Curved Array C1-6 Curved Array C1-6A Curved Array 6CT-A Curved Array 6CI-A Curved Array C2-9 Curved Array

C322 Micro-curved Array C351 Micro-curved Array C361 Micro-curved Array 3C-A Curved Array C1-6 Curved Array 6CT-A Curved Array 6CI-A Curved Array

SE Analysis1

VC6-2 Curved Array C613 Micro-curved Array VC2-9A Curved Array VC2-9 Curved Array

VC6-2 Curved Array C613 Micro-curved Array C542 Micro-curved Array

8P1 Phased Array 4P-A Phased Array S1-5 Phased Array 7P-A Phased Array

2P1 Phased Array S1-5 Phased Array 7P-B Phased Array 3P-A Phased Array

SE Analysis2

L742 Linear Array 10I2 Linear Array 9L-A Linear Array 12L-A Linear Array VL12-5 Linear Array 12L-B Linear Array 13L-A Linear Array 18L-A Linear Array 10L-I Linear Array ML3-18 Linear Array 12LT-A Linear Array 12LI-A Linear Array

L742 Linear Array L741 Linear Array L746 Linear Array 10I2 Linear Array 9L-A Linear Array 12L-B Linear Array 13L-A Linear Array 18L-A Linear Array 10L-I Linear Array 12LT-A Linear Array 12LI-A Linear Array

SE Analysis3

Page 45: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

510(k) Summary Page 4 of 8

6V1 Micro-curved Array 6V3 Micro-curved Array 6V3A Micro-curved Array 6V7 Micro-curved Array EC9-5 Micro-curved Array BCC9-5 Micro-curved Array BCL10-5 Biplane (Micro-curved + Linear Array) VE9-5 Micro-curved Array 12C-ER Micro-curved Array C3-10V Micro-curved Array VE3-10 Micro-curved Array

6V1 Micro-curved Array 6V3 Micro-curved Array 6V3A Micro-curved Array 6V7 Micro-curved Array EC9-5 Micro-curved Array BCC9-5 Micro-curved Array BCL10-5 Biplane (Micro-curved + Linear Array)

PWD2.0 Doppler CWD5.0 CW CWD8.0 CW CWD2.0 CW MPTEE Phased Array MPTEE mini Phased Array LAP7 Linear Array

PWD2.0 Doppler CWD2.0 CW MPTEE Phased Array MPTEE mini Phased Array LAP7 Linear Array

SE Analysis 1:

Compared with the predicate device, there are three new Curved Array probes as

followed:

(C542), which is similar with the probe (C613) cleared with predicate device SonoScape

P50 (K170999). (C351, C361), which is similar with the probe (C613) cleared with

predicate device SonoScape P50 (K170999).

SE Analysis 2:

Compared with the predicate device, there are three new Phase probes as followed:

(3P-A, 7P-B), which is similar with the probe (4P-A) cleared with predicate device

SonoScape P50 (K170999). (2P1), which is similar with the probe (8P1) cleared with

predicate device SonoScape P50 (K170999).

SE Analysis 3:

Compared with the predicate device, there are two new Linear Array probes as followed:

(L741, L746), which is similar with the probe (L742) cleared with predicate device

SonoScape P50 (K170999).

Biopsy Bracket Comparison:

Subject device P10 Series has the similar Biopsy Brackets as the predicate device

SonoScape P50 (K170999).

Page 46: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

510(k) Summary Page 5 of 8

Table 2 Biopsy Bracket Comparison

ID Comparison Items

Subject Device

SonoScape P10 Exp,P10,P10

Pro,P11,M10,P9

Predicate Device SonoScape P50

Exp/P50 Pro/P50S/P50/P47/P40

Exp/P40 Pro/P40S/P40/P37/P30

Exp/P30 Pro/P30S/P30/P27

(K170999)

Remark

11 Biopsy Bracket

NGBC613 NGB3C-A NGBC322 NGBC1-6 NGBVC6-2 NGBC542 NGBC351 NGBC361 NGB6CT-A NGB6CI-A

NGBC613 NGB3C-A NGBC322 NGBC1-6 NGBC1-6A NGBC2-9 NGBVC6-2 NGBVC2-9 NGBVC2-9A

SE Analysis 4

NGBL742-2 NGBL741 NGB9L-A NGB12L-B NGB13L-A NGB18L-A NGB10I2 NGB10L-I NGBL746 NGB 12LT-A NGB 12LI-A

NGBL742-2 NGB12L-A NGB9L-A NGB12L-B NGB13L-A NGB18L-A NGB10L-I NGBML3-18 NGBVL12-5

NGB4P-A NGBS1-5 NGB7P-B NGB2P1

NGB4P-A NGBS1-5 NGB7P-A

NGB6V1 NGB6V3 NGB 6V3A NGB6V7 NGBBCL10-5 NGBVE9-5 NGBBCC9-5 NGBEC9-5

NGB6V1 NGB6V3-2 NGB6V7 NGBBCL10-5 NGBVE9-5 NGBVE3-10 NGBC3-10V NGBBCC9-5

Page 47: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

510(k) Summary Page 6 of 8

ID Comparison Items

Subject Device

SonoScape P10 Exp,P10,P10

Pro,P11,M10,P9

Predicate Device SonoScape P50

Exp/P50 Pro/P50S/P50/P47/P40

Exp/P40 Pro/P40S/P40/P37/P30

Exp/P30 Pro/P30S/P30/P27

(K170999)

Remark

NGBEC9-5

SE Analysis 4:

Compared with predicate device, there are seven new biopsy brackets NGB2P1, NGBL741,

NGB7P-B, NGBC542, NGBL746, NGBC351, NGBC361, the seven biopsy brackets are not

cleared with predicate device (P50 series), but they have been cleared with other device.

The detail information of all the seven biopsy brackets are described in table 6 as followed.

Table 3 Technical specifications of new Biopsy brackets

Biopsy

brackets

Model

Available

probe

model

Patient

Contacting

Material

Type Application

K

number

Device

NGBL741 L741 Stainless

steel

(SUS 304)

Reusable Body

surface

K150045 S30

NGB7P-B 7P-B Stainless

steel

(SUS 304)

Reusable Body

surface

K150045 S30

NGBC542 C542 Stainless

steel

(SUS 304)

Reusable Body

surface

K150045 S30

NGBL746 L746 Stainless

steel

(SUS 304)

Reusable Body

surface

K123616 A6

NGBC351 C351 Stainless

steel

(SUS 304)

Reusable Body

surface

K123616 A6

NGB2P1 2P1 Stainless

steel

(SUS 304)

Reusable Body

surface

K150045 S30

NGBC361 C361 Stainless Reusable Body K102642 SSI-8000

Page 48: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

510(k) Summary Page 7 of 8

Biopsy

brackets

Model

Available

probe

model

Patient

Contacting

Material

Type Application

K

number

Device

steel

(SUS 304)

surface

7. Non-Clinical Tests [21 CFR 807.92(b) (1)]

The P10 Series Digital Color Doppler Ultrasound System has been evaluated for

electrical, mechanical, thermal and electromagnetic compatibility safety,

biocompatibility and acoustic output.

Laboratory tests (including Phantom tests) were conducted to verify that the P10 system

met all design specifications and the P10 system conformed to applicable medical

device standards.

The P10 Series system has been designed and manufactured to meet the following

standards:

IEC 60601-1:2005+A1:2012, Medical Electrical Equipment- Part 1: General

requirements for basic safety and essential performance [08/20/2012];

IEC 60601-1-2:2007, Medical Electrical Equipment -Part 1-2: General requirements for

basic safety and essential performance- Collateral standard: Electromagnetic

compatibility-Requirements and tests [03/30/2007];

IEC 60601-2-37:2007, Medical Electrical Equipment-Part 2-37: Particular requirements

for basic safety and essential performance of ultrasonic medical diagnostic and

monitoring equipment [08/09/2007];

ISO 10993-5:2009, Biological Evaluation of Medical Devices, Part 5-Tests for in vitro

cytotoxicity [06/01/2009];

ISO 10993-10:2010, Biological Evaluation of Medical Devices- Part 10: Tests for

irritation and skin sensitization [08/01/2010];

AIUM/NEMA UD 2:2004 (R2009), Acoustic output measurement standard for diagnostic

ultrasound equipment [08/21/2009]; and

AIUM/NEMA UD 3:2004 (R2009), Standard for real-time display of thermal and

mechanical acoustic output indices on diagnostic ultrasound equipment [08/13/2010].

8. Clinical Test [21 CFR 807.92(b) (2)]

No clinical testing was required.

Page 49: Sonoscape Medical Corp. Regulatory Affairs Manager Yizhe … · 2017. 11. 15. · Sonoscape Medical Corp. ℅ Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe building,

SONOSCAPE MEDICAL CORP. 510(k) Submission

510(k) Summary Page 8 of 8

9. Substantially Equivalent Conclusions [21 CFR 807.92(b) (3)]

In accordance with the 21 CFR Part 807 and based on the information provided in this

premarket notification, SONOSCAPE MEDICAL CORP. concludes that P10 Series

Digital Color Doppler Ultrasound System is substantially equivalent to the predicate

device with regard to safety and effectiveness.