1 SOP for use of the SX19 Stopped-Flow Instrument The main components of the stopped flow system are: Always book the system before use. Starting up and preparing the SX19 for use: 1. Turn on the compressed air supply. An 8 bar gas supply power the pneumatic rams and stop valve of the sample-handling unit. Open the large lever value on the wall and check the pressure is 8 bar. 2. Turn on the water bath. Turn on the water bath on the floor under the instrument. Check there is sufficient water covering the heater coil, and top up with mQ water as necessary (there is a large plastic container by the sink in 0.14 for this purpose). (Let Katherine or Joe know if there is air in the sample handling unit, which the water bath fills and controls.) 3. Turn on the lamp. Power on the lamp power supply unit using the rocker switch at the rear. The green light on the “arc lamp ignite” switch should illuminate. Wait for at least 30 seconds before pressing the green button firmly to ignite the lamp. Ensure the shutter between the lamp housing and the monochromator is in the open position. You can check the lamp is on since light comes out of the slit in the lamp housing:

Upload: phamhanh

Post on 30-Jul-2019

216 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: SOP for use of the SX19 Stopped-Flow Instrument · SOP for use of the SX19 Stopped-Flow Instrument ... In any event it is best to ensure that mQ water/buffer is in the flow circuit

1

SOPforuseoftheSX19Stopped-FlowInstrument

Themaincomponentsofthestoppedflowsystemare:

Alwaysbookthesystembeforeuse.

StartingupandpreparingtheSX19foruse:

1. Turnonthecompressedairsupply.

An8bargassupplypowerthepneumaticramsandstopvalveofthesample-handlingunit.Openthelargelevervalueonthewallandcheckthepressureis8bar.

2. Turnonthewaterbath.

Turnonthewaterbathonthefloorundertheinstrument.Checkthereissufficientwatercoveringtheheatercoil,andtopupwithmQwaterasnecessary(thereisalargeplasticcontainerbythesinkin0.14forthispurpose).(LetKatherineorJoeknowifthereisairinthesamplehandlingunit,whichthewaterbathfillsandcontrols.)

3. Turnonthelamp.

Poweronthelamppowersupplyunitusingtherockerswitchattherear.Thegreenlightonthe“arclampignite”switchshouldilluminate.Waitforatleast30secondsbeforepressingthegreenbuttonfirmlytoignitethelamp.Ensuretheshutterbetweenthelamphousingandthemonochromatorisintheopenposition.

Youcancheckthelampisonsincelightcomesoutoftheslitinthelamphousing:

Page 2: SOP for use of the SX19 Stopped-Flow Instrument · SOP for use of the SX19 Stopped-Flow Instrument ... In any event it is best to ensure that mQ water/buffer is in the flow circuit

2

Ifthelampfailstoignite,thenwaitfortwominutesbeforepressingthegreenbuttonagain.

4. Turnontheelectronicsunit.

Powerontheelectronicsunitviatheblackrockerswitchonthefrontpanel.Therearethreelightemittingdiodes,onegreenandtwored.Thegreendiodewillflashforashorttimeaftertheunitispoweredon,butshouldthenremainonconstantly.Wheneverthesoftwareisrunningthereddiodeswillflashcontinuously.Itisrecommendedthatiftheinstrumentistobeuseddailytheelectronicsunitisleftpoweredon.

5. PoweronthecomputerandloadthePro-Datasoftware.

Poweronthecomputerandlogon.Thedefaultaccountis“lab”whichhasthepasswordbcssxxxx(seethestickeronthesideofthePC).LoadPro-DataSXusingthedesktopicon:

6. LoadthePro-DataViewersoftwareandsetthefilesavelocationandseednames.

StartthePro-DataViewerprogramusingtheViewericononthePro-DataViewertoolbar:

NavigateinthePro-DataViewerfilebrowsertoyourpreferredsavedirectoryandrightclickonitselectingSetWorkingDirectoryHere.Newdatafileswillnowautomaticallybesavedinthisdirectory:

NowopentheFilesNamesdialogboxbyclickingontheicononthetoolbar:

Enteranameforthenewdatafilesandbaselinefile(whererelevant).Subsequentfilesofeachtypewillbenumberedstartingfromthevaluessetinthewindow:

Page 3: SOP for use of the SX19 Stopped-Flow Instrument · SOP for use of the SX19 Stopped-Flow Instrument ... In any event it is best to ensure that mQ water/buffer is in the flow circuit

3

Preparingthesamplehandlingunit:

1. Configuretheflowlines.

Thefollowingdescriptionappliestoasingle-mixingstoppedflowexperimentwitha1:1mixingratio.(Seethemanualfordetailsofhowtoperformmorecomplicatedexperiments.)Thesingle-mixingexperimentalwaysutilisestheleft-handdriveramandthereagentsarealwaysloadedintheF(left)andC(right)drivesyringes:

Checkthattheinstrumentisintheconfigurationabovebeforeproceeding.(Thedrivevolumeispre-settoapproximately120μl.Donotadjustthisunderanycircumstances.)

2. Flushthesystemwithwater.

Thestopped-flowunitshouldhavebeenleftfromprevioususewithmQwater/ethanolintheflowcircuitandcell.InanyeventitisbesttoensurethatmQwater/bufferisintheflowcircuitandcellpriortoloadingsample.Todothis:

a)Setthebeigedrivevalvestothe“load”position:

Page 4: SOP for use of the SX19 Stopped-Flow Instrument · SOP for use of the SX19 Stopped-Flow Instrument ... In any event it is best to ensure that mQ water/buffer is in the flow circuit

4

b)Emptyanycontentsofthedrivesyringesintowastereservoirsyringesbypushingthedrivesyringepistonupwardsintoreservoirsyringes.Alwaysbecarefulwhenhandlingthedrivesyringepistonsastheycanbreak.

c)ReplacethereservoirsyringeswithonescontainingcleanmQwater/buffer.

d)Loadeachdrivesyringewithabout1mlofwater/bufferbypushingdownoneachreservoirsyringepistonwhileatthesametimepullingonthedrivesyringepiston.Takecarenottointroduceairbubblesintotheflow-circuit:oneofthemostcommoncausesofartifactsinstopped-flowdataisthepresenceofairbubbles.Repeattheupanddownmotionafewtimestodevelopyourtechniqueforeliminatingairbubbles*.

*Ingeneral,tominimisetheintroductionofbubblesandtofacilitatetheirremoval,withthedrivevalvesinthe“load”position,shuntliquidsbackandforthgentlythroughthedrivevalveswhilstkeepingtheliquidcompressedfrombothendsoftheflow(byapplyingdownwardpressureonthepistononthereservoirsyringeandupwardpressureonthedrivesyringepiston).Whenloadingdrivesyringesavoidsimplypullingthesyringesdown,tosuckliquidfromthereservoirs,asthiswillencouragecavitationandbubbleformation.Fillthembybothpushingliquidinfromthereservoirsabove,whilepullingthedrivesyringepistonsdown,tokeepapositivepressureinthevalveandsyringecavity.

e)Withthedrivevalvesstillinthe“load”position,raisethedriveramuntilitisincontactwithbothdrivesyringepistons.

f)Nowsetthebeigedrivevalvestothe“drive”position:

g)SelectDriveontheSamplehandlingunitpanel(usingtheDrivefunctionratherthantheAcquirefunctionensuresthattherewillbenodatacollectionandgenerationofadatafile*).

*ItishowevermoreefficienttorunwashesusingAcquire:use<10datapointsand<1stimeframe,tickthepanellabelled“repeats”andenter“10”asthenumberofrepeats.Thefilesgeneratedcaneasilybedeletedafterwards.

h)Doatleast5repeatdrivestoensurethattheflowcircuitandcellcontentshavebeencompletelydisplacedbythecleanwater.

Page 5: SOP for use of the SX19 Stopped-Flow Instrument · SOP for use of the SX19 Stopped-Flow Instrument ... In any event it is best to ensure that mQ water/buffer is in the flow circuit

5

Beforeexecutingadrivealwayscheckthatthedrivevalvesareinthe“drive”position(notthe“load”position)andthedriveramisincontactwiththesyringeplungers.Failuretodosomayresultindamagetotheflowcircuit.

Stopped-FlowAbsorbanceKinetics

1. SelectAbsorbancemode.

OntheSignalpanelofthePro-DatamainwindowselectAbsorbancefromthedropdownlist.EnsurethatAutoHVisticked:

2. Selectkineticsmode,settimebaseandpoints.

EnsurethatKineticsmodeisselectedontheSequencerpanel:

SelectExternalfromthedropdownlistintheTriggerpaneltoactivatestopped-flowoperation:

OntheTimebasepanelchooseanappropriateacquisitiontimebasedontheexpectedtimescaleofthereaction.ThePointsshouldnormallybesettoavaluebetween400and1000datapoints.

3. Selectmonochromatorwavelengthandslitwidth.

OntheMonochromatorpanelentertheselectedwavelength:

Usingthedialsonthefrontofthemonochromatorsettheslitwidthstogivetherequiredabsorbancesignal.

Page 6: SOP for use of the SX19 Stopped-Flow Instrument · SOP for use of the SX19 Stopped-Flow Instrument ... In any event it is best to ensure that mQ water/buffer is in the flow circuit

6

4. Zerothedetector.

Filltheopticalcellwithasuitablereferencematerial(e.g.waterorothersolvent).ClicktheReferencebuttonintheSignalpaneltozerotheabsorbancedetector.

5. Loadthesamples.

Samplesmaynowbeloadedintothedrivesyringes.Atleast1.3mLisnecessaryforacompleteseriesofapprox.25drives.

Donottopush/pullthedrivesyringepistonswhiletheyareinthe“drive”positionafterloading.Equalizingthepistonlevelsshouldonlybedoneinthe“load”position,otherwisethesamplefromonesyringe(i.e.enzyme)iseasilypushedintotheothersyringe(i.e.substrate).Thereactionwillthenoccurprematurelyinthedrivesyringesandyouwillhavetodiscardthecontaminatedmaterial,andthenwashandloadthesyringesagainusingfreshsamples.

6. Acquiredata.

Ensurethatthedrivevalvesareinthe“drive”positionandclicktheAcquirebuttontoperformadriveandcollectthedata.Thefirstfewdriveswillconsistofbufferandthendilutedsample.Bythefourthorfifthdrivecleansampleshouldbemixedwitheachdriveandresultsshouldbeconsistent.

Absorbancemeasurementswillnowbeinunitsofabsoluteabsorbancesincewaterwasusedwhenmeasuringthereference.Itispossibletomeasurethereferencewithsomethingotherthanwaterinthecell,inwhichcasethemeasurementswillnowberelativeabsorbance.Thismaybeconvenientwhenmeasuringsmallabsorbancechangesonsignificantbackgrounds.Collectingareferencewhenonlyonereagent(oftencolourless)ispushedintothecellisonetypicaloptionasisreferencecollectiontothechemicalendpointofthechemicalreactionunderinvestigation.

Whenyouhavefinishedyourexperiments:

1. QuitthesoftwareandsaveyourdatatoacleanUSBdrive.2. Poweroffthecomputer.3. Poweroffthelamp.4. Powerofftheelectronicsunit.5. Turnoffthewaterbath.

Cleaningtheobservationcellandflowcircuitafteruse:

1. Flushthedrivesyringes.

Withthebeigevalvesinthe“load”position,pushupthedrivesyringestoemptythemasfaraspossible(takingcareastheyarefragile).ReplacethereservoirsyringeswithonescontainingmQwaterandflushthedrivesyringeswithwater.Replacethereservoirsyringeswithonescontaining2%Hellmanex(providednexttotheinstrument;asupplyofneatHellmanexisunderthefumehoodandmQwaterisintheplasticcontainerbythesinkin0.14ifthissolutionrunsout).Flushthedrivesyringeswithatapprox.5mLofthiscleaningsolution.

2. Washthecellandstopsyringe.

Thecleaningsolutioncanbepushedthroughthecell,stopsyringeandassociatedcircuitryusingrepeateddrivesasdescribedonpages4-5(putthevalvesinthe“load”positionandthedriveramincontactwithbothdrivesyringepistons,thenputthevalvesinthe“drive”positionandperformrepeatedsingleDrives,orAcquirewithrepeats).However,thisisunlikelytofullyreplacetheliquidinthestopsyringe.Toflushthestopsyringe,cellandassociatedcircuitrymorecompletelyitisnecessarytoperformmanualdrives.Todothis,putthedrive

Page 7: SOP for use of the SX19 Stopped-Flow Instrument · SOP for use of the SX19 Stopped-Flow Instrument ... In any event it is best to ensure that mQ water/buffer is in the flow circuit

7

valvesinthe“load”positionandmovethedriveramsoitisincontactwiththesyringeplungers,thenputthevalvesinthe“drive”position.

Next,turnoffthepressureusingthelargevalvehandleonthewall,andthenreleaseresidualpressurebydepressingthein-lineredbuttonbehindthesample-handlingunit:

Identifythestopvalve,whichwillbeinthe“inject”position(facingforwards):

Turnitbyhand180°clockwise(“waste”position;itwillonlyturnifthereisnopressureleftinthesystem)andemptythestopsyringemanuallybypushingitupwardsbeforereturningthestopvalvetoitsnormalposition.Whiledoingthis,takegreatcarenotto(a)movethestopassembly,whichisonlyweaklyfixedtothesamplehandlingunit,or(b)touchthecoppertriggerplate.FreshmQwater/buffercanthenbeintroducedintothedrivesyringes(“load”position)andflushedthroughtheflowcircuitmanually(byraisingthedriveramupwardsagainstthedrivesyringepistonsinthe“drive”position).Thestopsyringewillfill,andcanbeemptiedagainbyturningthestopvalveto“waste”.Thisprocedurecanberepeated(e.g.x5)untiltheuserisconfidentthatthecircuitiscleanandalsofreefromairbubbles.

Washthestopsyringeusingthemanualdriveproceduredescribedabove.Repeatapprox.5timesuntilyouareconfidentthatthecircuitisclean.Leaveitfilledwith80:20water:ethanolwhennotinuse.

3. Finalwash.

Finishbyreplacingthereservoirsyringeswithonescontainingcleanpurifiedwaterand20%ethanol.Flushthedrivesyringeswiththissolutionandleavethemfilledwith80:20water:ethanolwhennotinuse.

4. Turnofftheairsupply.

Turnoffthecompressedairsupplyonthewall,andreleasetheresidualpressurebydepressingtheredin-linebuttonbehindthesample-handlingunit.

5. Tidyup!

Pleasekeeptheareaaroundtheinstrumentcleanandclear.Re-fillemptybottlesandwipeupanyspillsimmediately.

KatherineStottandJoeMaman,January2019