spectra optia service manual - english, en...spectra optia apheresis system service manual v fan 3...
TRANSCRIPT
-
Spectra Optia®APHERESIS SYSTEM
SERVICE MANUAL
-
Spectra Optia®Apheresis SystemService Manual
Part No. 777401-0422011-01
-
©2011CaridianBCT, Inc.
CaridianBCT, Inc.10811 W. Collins AvenueLakewood, Colorado 80215USAUSA Phone: +1.877.339.4228Phone: +1.303.231.4357USA Fax: +1.866.715.6768Fax: +1.303.542.5215
CaridianBCT Europe N.V.Ikaroslaan 411930 ZaventemBelgiumPhone: +32.2.715.05.90Fax: +32.2.721.07.70
www.caridianbct.com
-
Contents
Preface ................................................................................................................. 15
1: Operational DescriptionOperational Description ..................................................................................................................................... 1-2
System Components ........................................................................................................................................... 1-3
Modes of Operation and Procedure States .......................................................................................................... 1-5
User Interface ..................................................................................................................................................... 1-7
Exchange Disposable Tubing Set ...................................................................................................................... 1-11
Collection Disposable Tubing Set .................................................................................................................... 1-15
Therapeutic Plasma Exchange .......................................................................................................................... 1-19
Mononuclear Cell Collection ........................................................................................................................... 1-22
Red Blood Cell Exchange Procedure Types ...................................................................................................... 1-24
RBCX Functional Description ......................................................................................................................... 1-26
2: System DescriptionSystem Description ............................................................................................................................................. 2-2
E-Box and Computer Systems ............................................................................................................................ 2-3
Control CCA ............................................................................................................................................. 2-5
Control Computer CCA ............................................................................................................................ 2-7
Control Ethernet CCA .............................................................................................................................. 2-8
Other Control Stack CCAs ........................................................................................................................ 2-9
Safety CCA .............................................................................................................................................. 2-11
Safety Computer CCA ............................................................................................................................. 2-12
Other Safety Stack CCA .......................................................................................................................... 2-14
AIM System CCAs .................................................................................................................................. 2-14
Motor Driver CCA .................................................................................................................................. 2-18
Display CCA ........................................................................................................................................... 2-26
Top Cap and Motherboard CCAs ........................................................................................................... 2-28
Spectra Optia Apheresis System Service Manual i
-
Power System .......................................................................................................................................... 2-29
Cooling Fans ........................................................................................................................................... 2-32
Display System ................................................................................................................................................. 2-33
Display Assembly ..................................................................................................................................... 2-35
Pump System ................................................................................................................................................... 2-38
Pump Assembly ....................................................................................................................................... 2-41
Valve System .................................................................................................................................................... 2-43
Valve Assembly ........................................................................................................................................ 2-45
Sensor System ................................................................................................................................................... 2-48
Pressure Sensors ....................................................................................................................................... 2-48
Reservoir Level Sensors ............................................................................................................................ 2-52
Reservoir Management ............................................................................................................................ 2-55
Fluid Sensors ........................................................................................................................................... 2-56
Return Line Air Detector ......................................................................................................................... 2-58
Set ID Detector ....................................................................................................................................... 2-60
RBC Detector .......................................................................................................................................... 2-63
Fluid Leak Detector ................................................................................................................................. 2-66
Linear Actuator System .................................................................................................................................... 2-68
AIM System ..................................................................................................................................................... 2-71
Camera System ........................................................................................................................................ 2-71
Lighting System ....................................................................................................................................... 2-73
Centrifuge System ............................................................................................................................................ 2-81
Centrifuge Drive Assembly ...................................................................................................................... 2-85
Filler Assembly ......................................................................................................................................... 2-87
Centrifuge Motor Controller ................................................................................................................... 2-89
Mechanical System ........................................................................................................................................... 2-92
Dual IV Pole ............................................................................................................................................ 2-92
Wheel and Brake System ......................................................................................................................... 2-95
3: Software DescriptionSoftware Description .......................................................................................................................................... 3-2
Base Layer Software States .................................................................................................................................. 3-4
Power Control State ................................................................................................................................... 3-4
Startup State .............................................................................................................................................. 3-7
T1 Pump Power Test State ...................................................................................................................... 3-12
T1 Device Test State ............................................................................................................................... 3-14
T1 APC Test State ................................................................................................................................... 3-16
T1 Centrifuge Power Test State ............................................................................................................... 3-18
Load Disposable State .............................................................................................................................. 3-21
RBC Detector Calibration State .............................................................................................................. 3-26
ii Spectra Optia Apheresis System Service Manual
-
Disconnect Test State .............................................................................................................................. 3-28
Unload Disposable State .......................................................................................................................... 3-30
Set Layer Software States .................................................................................................................................. 3-32
Collection States ...................................................................................................................................... 3-32
Exchange States ....................................................................................................................................... 3-54
Protocol Layer Software States .......................................................................................................................... 3-77
MNC Run State ...................................................................................................................................... 3-77
RBCX Run State ..................................................................................................................................... 3-82
TPE Run State ......................................................................................................................................... 3-88
4: TroubleshootingNon-Alarm Troubleshooting .............................................................................................................................. 4-2
Touchscreen Troubleshooting ................................................................................................................... 4-2
General AIM Troubleshooting ................................................................................................................... 4-3
Troubleshooting Failures During Start Up ................................................................................................ 4-5
Datalog Files ...................................................................................................................................................... 4-7
Alarm Troubleshooting ...................................................................................................................................... 4-8
Spectra Optia Alarms ....................................................................................................................................... 4-11
AC fluid detector detected fluid too soon. ................................................................................................ 4-12
AC fluid detector did not detect fluid. ..................................................................................................... 4-14
AC infusion data was invalid. ................................................................................................................... 4-16
AC infusion data was invalid. ................................................................................................................... 4-17
AC occlusion test failed. ........................................................................................................................... 4-18
AC pump did not stop when commanded. .............................................................................................. 4-20
AC pump malfunctioned. ........................................................................................................................ 4-22
AC pump software error occurred. ........................................................................................................... 4-24
AC pump speed was too high. ................................................................................................................. 4-25
AC pump turned faster than inlet pump. ................................................................................................. 4-26
AC volume limit was exceeded. ................................................................................................................ 4-28
AC was not detected. ............................................................................................................................... 4-29
AIM subsystem failed. .............................................................................................................................. 4-31
AIM system's interface measurement data was not updated. .................................................................... 4-33
AIM system could not establish target interface. ...................................................................................... 4-35
AIM system detected RBC interface near top of channel. ........................................................................ 4-37
AIM system did not complete calibration test. ......................................................................................... 4-40
AIM system did not complete channel imaging test. ................................................................................ 4-42
AIM system did not complete illumination test. ...................................................................................... 4-44
AIM system did not complete interface measurement. ............................................................................. 4-46
AIM system did not detect interface. ....................................................................................................... 4-48
AIM system did not respond to system command. ................................................................................... 4-51
AIM system failed calibration test. ........................................................................................................... 4-53
Spectra Optia Apheresis System Service Manual iii
-
AIM system failed channel imaging test. .................................................................................................. 4-55
AIM system failed illumination test. ........................................................................................................ 4-58
AIM system failed to measure interface. ................................................................................................... 4-61
Air may not have been removed from return line. .................................................................................... 4-64
Air removal is complete. ........................................................................................................................... 4-66
Air was detected in return line. ................................................................................................................ 4-67
Analog sensor data was not available. ....................................................................................................... 4-69
Bolus volume caused fluid balance to exceed limits. ................................................................................. 4-70
Calculation of pump rpm was not consistent. .......................................................................................... 4-72
Canceling custom prime recommendation. .............................................................................................. 4-74
Cassette mechanism that raises and lowers cassette malfunctioned. .......................................................... 4-75
Cassette sensor malfunctioned. ................................................................................................................ 4-77
Cassette software error occurred. .............................................................................................................. 4-79
Cassette tray was not correctly positioned. ............................................................................................... 4-80
Cassette was not detected in all positions. ................................................................................................ 4-82
Cassette was not lowered. ......................................................................................................................... 4-84
Centrifuge control malfunctioned. ........................................................................................................... 4-85
Centrifuge did not stop during alarm. ...................................................................................................... 4-87
Centrifuge did not stop when commanded. ............................................................................................. 4-88
Centrifuge door lock malfunctioned. ....................................................................................................... 4-90
Centrifuge door lock power circuit failed. ................................................................................................ 4-91
Centrifuge door lock sensor malfunctioned. ............................................................................................. 4-92
Centrifuge door lock software error occurred. .......................................................................................... 4-94
Centrifuge door lock test malfunctioned. ................................................................................................. 4-95
Centrifuge door position sensor malfunctioned. ....................................................................................... 4-97
Centrifuge door was not closed. ............................................................................................................... 4-99
Centrifuge door was opened or unlocked when centrifuge was spinning. ............................................... 4-100
Centrifuge exceeded maximum allowed speed. ....................................................................................... 4-102
Centrifuge motion detected was improper. ............................................................................................ 4-104
Centrifuge power could not be disabled. ................................................................................................ 4-106
Centrifuge power could not be enabled. ................................................................................................. 4-107
Centrifuge power did not turn off. ......................................................................................................... 4-108
Centrifuge pressure exceeded limit. ........................................................................................................ 4-110
Centrifuge pressure sensor failed calibration test. ................................................................................... 4-113
Centrifuge pressure sensor failed post-load test. ..................................................................................... 4-115
Centrifuge pressure sensor failed zero differential test. ............................................................................ 4-117
Centrifuge pressure sensor failed zero limit test. ..................................................................................... 4-119
Centrifuge pressure sensor failed zero range test. .................................................................................... 4-121
Centrifuge pressure sensor malfunctioned. ............................................................................................. 4-123
Centrifuge software error occurred. ........................................................................................................ 4-125
iv Spectra Optia Apheresis System Service Manual
-
Centrifuge speed sensors malfunctioned. ................................................................................................ 4-126
Centrifuge speed was not sufficient. ....................................................................................................... 4-128
Centrifuge speed was too high while procedure was paused. .................................................................. 4-130
Centrifuge was commanded to spin during alarm. ................................................................................. 4-132
Centrifuge was spinning in reverse. ........................................................................................................ 4-133
Chamber took too long to fill. ............................................................................................................... 4-135
Change replacement fluid to RBC for exchange. .................................................................................... 4-137
Change to patient data caused fluid balance to exceed limits. ................................................................. 4-138
Check connection from return line to blood warmer tubing. ................................................................. 4-141
Clamp inlet and return lines. ................................................................................................................. 4-142
Clamp inlet line. .................................................................................................................................... 4-143
Clamp return line. ................................................................................................................................. 4-144
Clamp return line. ................................................................................................................................. 4-145
Clear saline from drip chamber. ............................................................................................................. 4-146
Close inlet saline line. Clamp return line. .............................................................................................. 4-147
Close lines to prepare for tubing set test. ................................................................................................ 4-148
Close return saline line. .......................................................................................................................... 4-149
Close return saline line. .......................................................................................................................... 4-150
Close saline lines and clamp extra remove line. ...................................................................................... 4-151
Collection bag reached capacity. ............................................................................................................ 4-152
Collect recirculation flow path was not primed. ..................................................................................... 4-153
Confirming selection to begin rinseback. ............................................................................................... 4-155
Confirming selection to disable AIM system. ......................................................................................... 4-156
Confirming selection to disable RBC detector. ...................................................................................... 4-157
Confirming selection to disconnect patient and raise cassette. ................................................................ 4-158
Confirming selection to disconnect patient and raise cassette. ................................................................ 4-159
Confirm return line did not contain air. ................................................................................................ 4-160
Confirm saline lines were closed. ........................................................................................................... 4-161
Control flow was not consistent. ............................................................................................................ 4-162
Could not confirm saline lines were closed. ............................................................................................ 4-163
Current run targets could not be attained with run value entered. ......................................................... 4-164
Current run targets could not be attained with value entered. ................................................................ 4-165
Current run targets could not be attained with value entered. System paused. ....................................... 4-166
Custom prime volume was not sufficient. .............................................................................................. 4-167
Door-closed sensor was not consistent. .................................................................................................. 4-168
Door lock sensor was not consistent. ...................................................................................................... 4-170
Fan 1 failed. ........................................................................................................................................... 4-172
Fan 1 malfunctioned. ............................................................................................................................. 4-174
Fan 2 failed. ........................................................................................................................................... 4-176
Fan 2 malfunctioned. ............................................................................................................................. 4-178
Spectra Optia Apheresis System Service Manual v
-
Fan 3 failed. ........................................................................................................................................... 4-180
Fan 3 malfunctioned. ............................................................................................................................. 4-182
Fluid balance change caused fluid balance to exceed limits. .................................................................... 4-184
Fluid balance no longer exceeded limits. ................................................................................................ 4-186
Fluid leak detector failed self-test. .......................................................................................................... 4-188
Hall sensor and encoder in AC pump malfunctioned. ............................................................................ 4-189
Hall sensor and encoder in inlet pump malfunctioned. .......................................................................... 4-191
Hall sensor and encoder in plasma pump malfunctioned. ...................................................................... 4-193
Hall sensor and encoder in replace pump malfunctioned. ...................................................................... 4-195
Hall sensor and encoder in return pump malfunctioned. ....................................................................... 4-197
High-level reservoir sensor detected excess fluid. .................................................................................... 4-199
High-level reservoir sensor detected fluid too soon. ................................................................................ 4-202
High-level reservoir sensor did not detect fluid. ..................................................................................... 4-204
High-level reservoir sensor failed fluid check. ......................................................................................... 4-206
High-level reservoir sensor failed timing test. ......................................................................................... 4-208
High-level reservoir sensor malfunctioned. ............................................................................................. 4-210
Hypervolemia condition was possible. ................................................................................................... 4-212
Hypervolemia limit was exceeded. ......................................................................................................... 4-213
Hypovolemia condition was possible. .................................................................................................... 4-214
Hypovolemia limit was exceeded. .......................................................................................................... 4-216
Identification of tubing set did not match selected procedure. ............................................................... 4-217
Improper state transition occurred. ........................................................................................................ 4-219
Inlet:AC ratio data was invalid. .............................................................................................................. 4-220
Inlet:AC ratio data was invalid. .............................................................................................................. 4-221
Inlet:AC ratio was above maximum. ...................................................................................................... 4-222
Inlet:AC ratio was below minimum. ...................................................................................................... 4-223
Inlet line prime failed. ............................................................................................................................ 4-224
Inlet line was not clamped. .................................................................................................................... 4-225
Inlet pressure sensor failed calibration test. ............................................................................................. 4-226
Inlet pressure sensor failed post-load test. ............................................................................................... 4-228
Inlet pressure sensor failed zero differential test. ..................................................................................... 4-230
Inlet pressure sensor failed zero limit test. .............................................................................................. 4-232
Inlet pressure sensor failed zero range test. ............................................................................................. 4-234
Inlet pressure sensor malfunctioned. ...................................................................................................... 4-236
Inlet pressure was too low. ..................................................................................................................... 4-238
Inlet pump did not stop when commanded. .......................................................................................... 4-240
Inlet pump malfunctioned. .................................................................................................................... 4-242
Inlet pump software error occurred. ....................................................................................................... 4-244
Inlet saline line was obstructed. .............................................................................................................. 4-245
Inlet saline line was obstructed during rinseback. ................................................................................... 4-247
vi Spectra Optia Apheresis System Service Manual
-
Interface still took too long to establish. ................................................................................................. 4-248
Interface took too long to establish. ....................................................................................................... 4-251
Internal timing was not consistent. ........................................................................................................ 4-254
Leak was detected in centrifuge. ............................................................................................................. 4-256
LED sensor malfunctioned. ................................................................................................................... 4-259
Low-level reservoir sensor continued to detect excess fluid. .................................................................... 4-260
Low-level reservoir sensor detected air. ................................................................................................... 4-263
Low-level reservoir sensor detected excess fluid. ..................................................................................... 4-265
Low-level reservoir sensor detected fluid too soon. ................................................................................. 4-268
Low-level reservoir sensor did not detect fluid. ....................................................................................... 4-270
Low-level reservoir sensor failed timing test. .......................................................................................... 4-272
Low-level reservoir sensor malfunctioned. .............................................................................................. 4-274
Low-level reservoir sensor malfunctioned. .............................................................................................. 4-276
Low-level reservoir sensor malfunctioned. .............................................................................................. 4-278
Low-level reservoir sensor persisted in detecting excess fluid. ................................................................. 4-279
Minus twelve volt power supply was out of range. ................................................................................. 4-282
NVRAM safety test failed. ..................................................................................................................... 4-283
NVRAM test failed. ............................................................................................................................... 4-284
Open and close centrifuge door. ............................................................................................................ 4-285
Open inlet saline line. ............................................................................................................................ 4-286
Open return saline line. ......................................................................................................................... 4-287
Open saline lines. ................................................................................................................................... 4-288
Patient's fluid balance may be 10% higher than reported. ...................................................................... 4-289
Patient's fluid balance may be 10% lower than reported. ....................................................................... 4-291
Patient's fluid balance may be 15% higher than reported. ...................................................................... 4-293
Patient's fluid balance may be 15% lower than reported. ....................................................................... 4-295
Patient's fluid balance may be 5% higher than reported. ........................................................................ 4-297
Patient's fluid balance may be 5% lower than reported. ......................................................................... 4-299
Patient fluid balance may be lower than reported. .................................................................................. 4-301
Patient height was invalid. ..................................................................................................................... 4-303
Patient height was invalid. ..................................................................................................................... 4-304
Patient hematocrit data was invalid. ....................................................................................................... 4-305
Patient hematocrit data was invalid. ....................................................................................................... 4-306
Patient sex was invalid. .......................................................................................................................... 4-307
Patient sex was invalid. .......................................................................................................................... 4-308
Patient TBV was invalid. ....................................................................................................................... 4-309
Patient TBV was invalid. ....................................................................................................................... 4-310
Patient weight was invalid. ..................................................................................................................... 4-311
Patient weight was invalid. ..................................................................................................................... 4-312
Pause button was touched. ..................................................................................................................... 4-313
Spectra Optia Apheresis System Service Manual vii
-
Plasma bag reached capacity. .................................................................................................................. 4-314
Plasma pump did not stop when commanded. ....................................................................................... 4-315
Plasma pump malfunctioned. ................................................................................................................ 4-317
Plasma pump software error occurred. ................................................................................................... 4-319
Plasma recirculation flow path was not primed. ..................................................................................... 4-320
Plasma remove flow path was not primed. ............................................................................................. 4-322
Plasma valve malfunctioned. .................................................................................................................. 4-324
Plasma valve sensor malfunctioned. ........................................................................................................ 4-325
Plasma valve software error occurred. ..................................................................................................... 4-326
Plasma valve was not correctly positioned. ............................................................................................. 4-327
Predicted fluid balance exceeded limits. ................................................................................................. 4-328
Predicted procedure time exceeded 8 hours. ........................................................................................... 4-330
Prediction software malfunctioned. ........................................................................................................ 4-331
Pressure in collect line was too high. ...................................................................................................... 4-333
Pressure sensor 4 failed calibration test. .................................................................................................. 4-335
Pressure sensor 4 failed post-load test. .................................................................................................... 4-337
Pressure sensor 4 failed zero differential test. .......................................................................................... 4-339
Pressure sensor 4 failed zero limit test. ................................................................................................... 4-341
Pressure sensor 4 failed zero range test. .................................................................................................. 4-343
Prime saline has been diverted. Close return saline line. ......................................................................... 4-345
Prime saline was not diverted. ................................................................................................................ 4-346
Procedure configuration data was corrupt. ............................................................................................. 4-347
Procedure data error detected. ................................................................................................................ 4-348
Procedure data error detected. ................................................................................................................ 4-349
Procedure time exceeded 8 hours. .......................................................................................................... 4-350
Pump 1 failed rotation test. .................................................................................................................... 4-351
Pump 2 failed rotation test. .................................................................................................................... 4-353
Pump 3 failed rotation test. .................................................................................................................... 4-355
Pump 4 failed rotation test. .................................................................................................................... 4-357
Pump 5 failed rotation test. .................................................................................................................... 4-359
Pump did not function correctly. ........................................................................................................... 4-361
Pump motion detected was improper. ................................................................................................... 4-363
Pump power could not be disabled. ....................................................................................................... 4-364
Pump power could not be enabled. ........................................................................................................ 4-365
Pumps and centrifuge did not stop during alarm. .................................................................................. 4-366
Pumps continued to turn during alarm. ................................................................................................. 4-367
Pumps did not stop during alarm. .......................................................................................................... 4-368
Pumps have been paused for 10 minutes. ............................................................................................... 4-369
Pumps have been paused for 3 minutes. ................................................................................................. 4-370
Pumps were commanded to turn during alarm. ..................................................................................... 4-371
viii Spectra Optia Apheresis System Service Manual
-
Pumps were commanded to turn instead of stop. ................................................................................... 4-372
Pumps were paused for 3 minutes. ......................................................................................................... 4-373
Pumps were turned during alarm. .......................................................................................................... 4-375
RBC detector calibration error occurred. ............................................................................................... 4-376
RBC detector did not respond. .............................................................................................................. 4-378
RBC detector did not transmit data. ...................................................................................................... 4-380
RBC detector software error occurred. ................................................................................................... 4-382
RBC recirculation flow path was not primed. ........................................................................................ 4-383
RBC remove flow path was not primed. ................................................................................................ 4-385
RBC valve malfunctioned. ..................................................................................................................... 4-387
RBC valve malfunctioned. ..................................................................................................................... 4-388
RBC valve sensor malfunctioned. ........................................................................................................... 4-389
RBC valve software error occurred. ........................................................................................................ 4-390
RBC valve was not correctly positioned. ................................................................................................ 4-391
Read-type communication error from RBC detector occurred. .............................................................. 4-392
Recommending custom prime. .............................................................................................................. 4-394
Red blood cells were detected in collect line too soon. ........................................................................... 4-395
Red blood cells were detected in plasma line from centrifuge. ................................................................ 4-398
Remove bag reached capacity. ................................................................................................................ 4-401
Remove valve malfunctioned. ................................................................................................................ 4-402
Remove valve sensor malfunctioned. ...................................................................................................... 4-403
Remove valve software error occurred. ................................................................................................... 4-404
Remove valve was not correctly positioned. ............................................................................................ 4-405
Replacement fluid detector detected fluid too soon. ............................................................................... 4-406
Replacement fluid was not detected. ...................................................................................................... 4-408
Replace pump did not stop when commanded. ..................................................................................... 4-410
Replace pump malfunctioned. ............................................................................................................... 4-412
Replace pump software error occurred. .................................................................................................. 4-414
Reservoir did not contain enough fluid. ................................................................................................. 4-415
Reservoir sensors malfunctioned. ........................................................................................................... 4-416
Reservoir was full during air removal. .................................................................................................... 4-417
Return line air detector detected fluid too soon. ..................................................................................... 4-419
Return line air detector failed fluid check. .............................................................................................. 4-421
Return line prime failed. ........................................................................................................................ 4-423
Return line was not clamped during air removal. ................................................................................... 4-425
Return line was not primed. ................................................................................................................... 4-427
Return pressure sensor failed calibration test. ......................................................................................... 4-429
Return pressure sensor failed post-load test. ........................................................................................... 4-431
Return pressure sensor failed zero differential test. ................................................................................. 4-433
Return pressure sensor failed zero limit test. ........................................................................................... 4-435
Spectra Optia Apheresis System Service Manual ix
-
Return pressure sensor failed zero range test. .......................................................................................... 4-437
Return pressure was too high. ................................................................................................................ 4-439
Return pressure was too high when system was diverting saline. ............................................................. 4-441
Return pump did not stop. .................................................................................................................... 4-442
Return pump did not stop when commanded. ....................................................................................... 4-443
Return pump malfunctioned. ................................................................................................................ 4-445
Return pump malfunctioned. ................................................................................................................ 4-447
Return pump software error occurred. ................................................................................................... 4-449
Return pump speed was too high. .......................................................................................................... 4-450
Return pump turned too fast. ................................................................................................................ 4-452
Return saline line failed integrity test. .................................................................................................... 4-454
Return saline line was not closed during air removal. ............................................................................. 4-456
Return saline line was obstructed. .......................................................................................................... 4-458
Return saline line was obstructed during air removal. ............................................................................ 4-460
Run target will be attained before next collection phase. ........................................................................ 4-462
Safety system could not confirm position of collect valve. ...................................................................... 4-463
Safety system could not confirm valve position. ..................................................................................... 4-464
Safety system could not read backup configuration file. ......................................................................... 4-466
Safety system could not read default configuration file. .......................................................................... 4-467
Safety system could not read primary configuration file. ........................................................................ 4-468
Safety system detected air in return line. ................................................................................................ 4-469
Safety system detected air in return line. ................................................................................................ 4-470
Safety system detected excess AC infusion. ............................................................................................. 4-471
Safety system detected internal timing inconsistency. ............................................................................. 4-473
Safety system detected software communication error. ........................................................................... 4-475
Safety system detected that patient hematocrit was lower than minimum. ............................................. 4-477
Safety system detected too much fluid in collection bag. ........................................................................ 4-478
Safety system detected too much fluid in plasma bag. ............................................................................ 4-479
Safety system detected too much fluid in remove bag. ............................................................................ 4-480
Safety system detected valve open longer than expected. ........................................................................ 4-481
Safety system did not respond to power request. .................................................................................... 4-483
Safety system did not respond to request to turn off power. ................................................................... 4-484
Safety system did not respond to request to turn on power. ................................................................... 4-485
Safety system failed power test. .............................................................................................................. 4-486
Safety system failed start process. ........................................................................................................... 4-488
Safety system lost internal communication capability. ............................................................................ 4-490
Safety system stopped air removal. ......................................................................................................... 4-492
Saline lines were not closed. ................................................................................................................... 4-493
Saline line was obstructed. ..................................................................................................................... 4-494
Sixty-four volt power supply was out of range. ....................................................................................... 4-496
x Spectra Optia Apheresis System Service Manual
-
Software consistency error occurred. ...................................................................................................... 4-497
Software control data was not updated. .................................................................................................. 4-498
Software CRC failed. ............................................................................................................................. 4-500
Software memory for AIM system was not sufficient. ............................................................................. 4-501
Software memory for control system was not sufficient. ......................................................................... 4-502
Software memory for safety system was not sufficient. ........................................................................... 4-503
Software module was not consistent. ...................................................................................................... 4-504
Stop button was touched. ...................................................................................................................... 4-505
System cannot verify that patient was disconnected. .............................................................................. 4-506
System could not load correct software. ................................................................................................. 4-507
System could not read backup configuration file. ................................................................................... 4-508
System could not read default configuration file. ................................................................................... 4-509
System could not read primary configuration file. .................................................................................. 4-510
System did not correctly monitor AC infusion. ...................................................................................... 4-511
System malfunctioned when it raised or lowered cassette. ...................................................................... 4-513
System paused: Centrifuge door was not closed. .................................................................................... 4-514
System was reset. .................................................................................................................................... 4-516
System was reset. .................................................................................................................................... 4-518
System was reset. .................................................................................................................................... 4-520
TBV restricted volume available to confirm volume in reservoir. ........................................................... 4-521
Test of negative inlet pressure failed. ...................................................................................................... 4-522
Test of negative return pressure failed. ................................................................................................... 4-524
Timing discrepancy in low-level reservoir sensor occurred. .................................................................... 4-526
Too much AC was detected during prime. ............................................................................................. 4-528
Tubing set did not match selected procedure. ........................................................................................ 4-529
Tubing set failed dwell test for negative inlet pressure. ........................................................................... 4-531
Tubing set failed dwell test for negative return pressure. ........................................................................ 4-533
Tubing set failed dwell test for positive inlet pressure. ............................................................................ 4-535
Tubing set failed dwell test for positive return pressure. ......................................................................... 4-537
Tubing set failed test for negative inlet pressure. .................................................................................... 4-539
Tubing set failed test for positive inlet pressure. ..................................................................................... 4-541
Tubing set failed test for positive inlet pressure. ..................................................................................... 4-543
Tubing set failed test for positive return pressure. .................................................................................. 4-545
Tubing set failed test of negative return pressure. ................................................................................... 4-547
Tubing set identification error occurred. ................................................................................................ 4-549
Tubing set may have contained fluid. .................................................................................................... 4-552
Twelve volt power supply was out of range. ........................................................................................... 4-554
Twenty-four volt power supply was out of range. ................................................................................... 4-555
Unclamp return line. ............................................................................................................................. 4-556
Unclamp return line. ............................................................................................................................. 4-557
Spectra Optia Apheresis System Service Manual xi
-
Valve position error occurred. ................................................................................................................ 4-558
Vent bag contained fluid during prime. ................................................................................................. 4-560
Vent bag contained fluid during run. ..................................................................................................... 4-562
Verification test that patient was disconnected from inlet line failed. ..................................................... 4-564
Verification test that patient was disconnected from return line failed. ................................................... 4-565
Voltage of fluid leak detector was above normal range. .......................................................................... 4-566
Voltage of fluid leak detector was below normal range. .......................................................................... 4-567
Volume entered for custom prime was too low. ..................................................................................... 4-569
Volume in collection bag may be lower than reported. ........................................................................... 4-570
Write-type communication error to data logging subsystem occurred. ................................................... 4-572
Write-type communication error to RBC detector occurred. ................................................................. 4-573
5: Maintenance and CalibrationMaintenance and Calibration ............................................................................................................................. 5-2
Removing the Side Panels .......................................................................................................................... 5-2
Raising the Pump Panel ............................................................................................................................. 5-5
Removing the Linear Actuator ................................................................................................................... 5-8
Calibrating the Centrifuge Motor Controller ........................................................................................... 5-10
Making a Test Cassette ............................................................................................................................ 5-12
Positioning the Linear Actuator Sensors ................................................................................................... 5-14
Printer and FTP Connectivity .......................................................................................................................... 5-17
Connecting a Printer ............................................................................................................................... 5-18
Connecting a Computer .......................................................................................................................... 5-20
6: Spectra Optia Spare PartsLiterature Group ................................................................................................................................................ 6-2
Documentation Group ....................................................................................................................................... 6-2
Display Head Group .......................................................................................................................................... 6-3
Pump Panel Group ............................................................................................................................................. 6-4
Centrifuge Basin Group ..................................................................................................................................... 6-6
E-box Group ...................................................................................................................................................... 6-7
Seal Safe Group .................................................................................................................................................. 6-9
Exterior Parts Group ........................................................................................................................................ 6-10
Tools Group ..................................................................................................................................................... 6-11
Other Parts Group ........................................................................................................................................... 6-11
Software Group ................................................................................................................................................ 6-12
Software Multipack Group ............................................................................................................................... 6-14
xii Spectra Optia Apheresis System Service Manual
-
7: SpecificationsSpecifications ...................................................................................................................................................... 7-2
Glossary
Index
Spectra Optia Apheresis System Service Manual xiii
-
xiv Spectra Optia Apheresis System Service Manual
-
PrefaceThe Spectra Optia Therapeutic Apheresis System Service Manual provides the information neededto service and troubleshoot the system.
Spectra Optia Apheresis System Service Manual 15
-
Who Should Read This ManualThis manual is intended for CaridianBCT service technicians and employees, trained and qualifiedcustomer technical staff, and CaridianBCT service partners.
How to Use This ManualThis manual is divided into sections that can be read and used separately.
Title Description
OperationalDescription
Describes the operating principles of the machine. Includes information about systemcomponents, operator controls, disposables, and procedures.
System Description Describes the functional and electronic principles of the machine
Software Description Describes the software that the machine uses to operate.
Troubleshooting Describes non-alarm and alarm troubleshooting. Provides alarm text and alarminformation.
Maintenance andCalibration
Describes maintenance and service procedures for the machine.
Spare Parts Describes the spare parts available for the machine.
Preface
16 Spectra Optia Apheresis System Service Manual
-
1Operational Description
Spectra Optia Apheresis System Service Manual 1-1
-
Operational DescriptionThe Spectra Optia Apheresis System is a transportable, automatic blood component separator thatuses centrifugation and optical detection (automated interface management or AIM) to performexchange and collection procedures.
The apheresis system operator uses the Spectra Optia system’s touch screen to communicate with thesystem. The operator follows the instructions on the screens to enter patient and procedure data,load and prime the disposable tubing set, and perform and troubleshoot procedures. During theprocedure, the system displays detailed procedural information, enabling the operator to adjustvalues to achieve a specific outcome or troubleshoot a condition. After the procedure, the systemprovides a summary of procedure data for the patient record. The operator may copy the data fromthe screen, or print the data if the system is connected to a printer or a computer.
Operational Description
1-2 Spectra Optia Apheresis System Service Manual
-
System ComponentsThe Spectra Optia system components include the video display, front panel, centrifuge chamber,and electronics compartment.
Figure 1-1: Spectra Optia system, front and back views
Table 1-1: Spectra Optia System Components
Component Function
1 Monitor with touch screen Allows you to communicate with the system.
2 Dual IV pole Contains hooks for hanging bags and containers. Adjusts up anddown for transport.
3 Front panel Holds the pumps, valves, sensors and detectors.
4 Seal Safe system Seals the lines of the tubing set.
5 Cassette tray Holds the tubing set cassette in place.
System Components
Spectra Optia Apheresis System Service Manual 1-3
-
Table 1-1: Spectra Optia System Components (continued)
Component Function
6 Power switch Allows you to turn power to the system on and off.
7 View port Allows you to look through the centrifuge door to see the interfacein the channel.
8 Centrifuge door Allows access to the centrifuge chamber.
9 Wheel pedal Allows you to adjust the direction of the wheels for moving thesystem, or lock the wheels in place so that the system does not roll.
10 Wheels Used to transport the system.
11 Handle wheels (2) Facilitates transport of the system in a horizontal position.
12 System serial number Unique number that identifies the system.
13 IV pole release button Allows you to lower the IV pole by pressing the button.
14 Power cord Connects the system to a power source.
15 Power cord holder Secures the power cord during transport.
16 Ethernet connection Allows you to connect the system to a printer or a computer to printa procedure report.
17 Blood warmer bracket (notshown)
Holds the blood warmer, if used.
Operational Description
1-4 Spectra Optia Apheresis System Service Manual
-
Modes of Operation and Procedure StatesThe Spectra Optia system has three modes of operation: Automatic mode, Semi-Automatic andService mode. In Automatic mode, the AIM system controls the interface. While in Semi-Automaticmode, AIM system control of the interface is disabled. There are nine procedure states that performmany individual steps to complete a procedure.
Automatic ModeDuring Automatic mode, the system uses configured parameters and entered data to set and controlthe pump flow rates, centrifuge speed, packing factor, run targets, and establish the interface. Foroptimal separation in the channel, the target packing factor is 20. The system then uses feedback itreceives from the AIM system to control the interface, so it can collect a specific blood componentand return the rest to the patient.
Semi-Automatic ModeDuring Semi-Automatic mode, the operator monitors the interface and adjusts it, as needed. Thesystem displays the interface on the screen, but does not control it. When certain alarm conditionsoccur, you may be required to disable the AIM system before you continue the procedure in Semi-Automatic mode. Under those conditions, the interface position does not appear on the screen andyou must use the viewport to see the interface.
Service ModeService mode allows technicians to connect to the Spectra Optia system using service software fortroubleshooting and maintenance.
Procedure StatesThe procedure states listed in the following table describe system activity from the perspective of theoperator. For specific information on the software states, see the Software Description section of thismanual.
Table 1-2: Procedure States
State Description
Load Operator selects the procedure and loads theappropriate tubing set.
Tubing set test System verifies that the correct tubing set is loadedproperly, and confirms the function of the pumps,valves, and sensors.
Prime System primes the tubing set to remove air from theset, and performs volumetric tests.
Modes of Operation and Procedure States
Spectra Optia Apheresis System Service Manual 1-5
-
Table 1-2: Procedure States (continued)
State Description
Run Patient is connected, and the system performsapheresis, according to the selected procedure.
Rinseback System returns the blood remaining in the tubing set tothe patient after the run is complete to reduce thepatient’s RBC loss. Operator may choose to perform orskip rinseback.
Disconnect test System verifies that the lines to the patient are clampedbefore it raises the cassette.
Unload System raises the cassette and unloads the pumps.
Procedure summary System displays the final procedure values.
Next procedure System prepares for the next procedure.
Operational Description
1-6 Spectra Optia Apheresis System Service Manual
-
User InterfaceThe Spectra Optia system’s user interface allows you to perform, adjust, and monitor procedures.
Navigating the Screen
Figure 1-2: Example screen
Table 1-3: User Interface elements
Section Description
1 Menu bar Contains menu buttons that you touch to show or hide menu tabs.
2 Menu tabs Appear according to the menu button that you touch. Tabs provideaccess to screens that allow you to view information or modify theprocedure.
User Interface
Spectra Optia Apheresis System Service Manual 1-7
-
Table 1-3: User Interface elements (continued)
Section Description
3 Message bar Displays information about what is currently occurring, andoccasionally displays instructions.
4 Activity area Displays tasks to complete, buttons that allow you to
• Navigate to another screen
• Change procedure conditions
• Change values or information about the procedure.
5 Action bar Contains buttons that you touch to perform the action stated on thebutton, or to navigate to a different screen.
Screen ColorsThe Spectra Optia system uses a color scheme to indicate the different states and conditions of theprocedure, allowing you to quickly determine the progress of the procedure.
Table 1-4: Corresponding color schemes for operating states and conditions
State or Condition Color
Load, prepare procedure, tubing set test, prime (patientnot connected)
Run, rinseback, disconnect test (patient connected)
Unload, procedure summary, next procedure (patientdisconnected)
Caution
Screen ButtonsTable 1-5: Active and Inactive Buttons
Active button Active buttons are colored andappear raised. Touch an activebutton to enter, change, or savedata, or to proceed to the nextactivity or action.
Inactive button Inactive buttons are gray. Theyrepresent an action or informationthat is not available.
Operational Description
1-8 Spectra Optia Apheresis System Service Manual
-
The buttons and icons described in the following table appear on the monitor or the screens, andeither allow you to operate and adjust the Spectra Optia system during an apheresis procedure, orcommunicate important information.
Table 1-6: Button and Icons
Graphic Description Location Function
Go back button Bottom right side of thescreen
Returns you to theprevious screen. Availableonly when it is safe toreturn to the previousscreen. You must confirmany data you changedbefore touching the goback button or the datawill not be saved.
Mute button Bottom left side of thescreen
Allows you to mute thesound during an alarmcondition. The sound isrestored if another alarmoccurs.
Pause button Bottom right corner of themonitor
Pauses the pumps.
Stop button Bottom left corner of themonitor
Stops the centrifuge andthe pumps.
Active alarm button Bottom left side of thescreen Indicates an activealarm.
Takes you to the activealarm screen, or to a listof active alarms if morethan one has occurred.
Clock button Run and the exchangestatus screens
Allows you to view clocktime or elapsed time forthe procedure.
Up/down arrow button Main run screen Allows you to adjust theposition of the interfaceduring Semi- Automaticmode.
User Interface
Spectra Optia Apheresis System Service Manual 1-9
-
Table 1-6: Button and Icons (continued)
Graphic Description Location Function
Left/right arrow button Strobe screen Adjusts the timing of thestrobe. Allow you to viewdifferent areas of thechannel.
Up/down scroll button Active alarm, alarmhistory, and reportscreens
Allows you to viewadditional information onthe screen.
Left/right scroll button Alarm action screens Allows you to scrollthrough the differentalarm action screens, soyou do not have to returnto the main alarm screento troubleshoot an alarm.
Check mark icon Buttons on the prepareprocedure screens
Indicates completion of arequired task, or confirmsa selection.
Do not connect patienticon
Bottom right side of thescreens
Safety feature thatreminds you that thepatient should not beconnected during thecurrent operating state.
Operator override icon Total Blood Volumebutton on the data entryscreen
Indicates that the operatorentered the patient’s totalblood volume whennecessary, or instead ofusing the volume that thesystem calculated.
Semi-Automatic modeicon
Bottom right side of therun screens
Indicates that the systemis operating in Semi-Automatic mode.
Operational Description
1-10 Spectra Optia Apheresis System Service Manual
-
Exchange Disposable Tubing SetThe Exchange disposable tubing set is used for therapeutic exchange procedures.
Table 1-7: Exchange Set Catalog and Set ID Numbers
Catalog Number Set ID
10200 • 0020 (4 L remove bag)
• 0120 (6 L remove bag)
10220 0122
Note: The 10200 Exchange set can only be used with TPE procedures. The 10220 Exchange set can be usedwith both TPE and RBCX procedures.
Figure 1-3: The Exchange set (10200)
Exchange Disposable Tubing Set
Spectra Optia Apheresis System Service Manual 1-11
-
Figure 1-4: The Exchange set (10220)
Table 1-8: Exchange Set Components
Component Function
1 Inlet line Carries anticoagulated blood from the patient to the channel. Theinlet luer (male) connects to the patient inlet site. The inlet lineclamp (red) is used to clamp the inlet line.
2 Inlet line manifold Consists of an injection site and connections for the inlet line,anticoagulant line, and inlet saline line.
3 Inlet saline clamp (red) Used to close the inlet saline line or control the flow of saline to thepatient.
4 Flow restrictor Allows AC to be pumped through the tubing but prevents the freeflow of AC.
5 Return line Carries blood components from the reservoir to the patient. Theluer (male) connects to the patient return access.
Operational Description
1-12 Spectra Optia Apheresis System Service Manual
-
Table 1-8: Exchange Set Components (continued)
Component Function
6 Return line manifold Consists of an injection site and connections for the return line andreturn saline line.
7 Return saline clamp (blue) Used to close the inlet saline line or control the flow of saline to thepatient.
8 Anticoagulant (AC) line (orange) Carries AC to the inlet manifold. The spike (orange) is used to spikethe AC container.
9 Saline line (green) Line that carries saline to the inlet manifold and return manifold fordelivery to the patient. The spike (green) is used to spike the salinecontainer.
10 Inlet line trap Prevents large, solid cellular matter from entering the channel.
11 Inlet pressure sensor Diaphragm that attaches to the inlet pressure sensor on the frontpanel to monitor the pressure to the patient.
12 Centrifuge pressure sensor Diaphragm that attaches to the centrifuge pressure sensor on thefront panel to monitor centrifuge pressure.
13 Reservoir Contains the fluid that is returned to the patient. Includes a 200-micron filter.
14 Return pressure sensor Diaphragm that attaches to the return pressure sensor on the frontpanel to monitor return pressure to the patient.
15 Centrifuge loop Four-lumen tubing that carries fluids in and out of the channel. Thesleeves reinforce the tubing at flex points. The collars fix the twoends of the loop to the centrifuge. The bearings are the contactpoints between the centrifuge arm and the loop.
16 Channel Used during centrifugation to separate cellular components fromplasma.
17 Connector Holds the lines where the separated components exit the channel.
18 Vent bag Holds air displaced from the system.
19 Remove bag Holds removed blood components.
Exchange Disposable Tubing Set
Spectra Optia Apheresis System Service Manual 1-13
-
Table 1-8: Exchange Set Components (continued)
Component Function
20 Remove line Carries the removed blood components to the remove bag.
21 Extra remove line Provides a connection (female luer) for an extra remove bag. Theclamp (yellow) is used to clamp the extra remove line.
22 Replace line Carries the replacement fluid from the containers to the reservoir.The spikes (white) are used to spike the replacement fluidcontainer. The clamps (white) are used to clamp the replacementline. The luer (male/female) allows connection of a filter or bloodwarmer tubing.
Operational Description
1-14 Spectra Optia Apheresis System Service Manual
-
Collection Disposable Tubing Set
Figure 1-5: The Collection set
Catalog number: 10110. Set ID number: 0011.
Table 1-9: Collection Set Components
Component Function
1 Inlet line Inlet line: Used to carry anticoagulated blood from the patient to thechannel. Also includes:
• Luer connector (male): Connects the inlet line to the needle line.
• Inlet line clamp (red): Used to clamp the inlet line.
• Clamp (white): Used to clamp the needle line.
2 Diversion bag Used to capture a skin plug after performing a peripheralvenipuncture, or to collect a blood sample. Also includes:
• Clamp (yellow): Used to clamp the line to the diversion bag.
Collection Disposable Tubing Set
Spectra Optia Apheresis System Service Manual 1-15
-
Table 1-9: Collection Set Components (continued)
Component Function
3 Inlet line manifold Provides a connection point for the inlet line, the AC line, and theinlet saline line. Includes an injection site.
4 Inlet saline line clamp (red) Used to close the inlet saline line or to control the flow of saline tothe patient.
5 AC check valve Allows the system to pump AC through the tubing, but prevents thefree flow of AC.
6 Return line Used to carry fluid from the reservoir to the patient. Also includes:
• Luer connector (male): Connects the return line to the patientreturn access.
• Return line clamp (blue): Used to clamp the return line.
7 Return line manifold Provides a connection point for the return line and the return salineline. Includes an injection site.
8 Return saline line clamp (blue) Used to close the return saline line or to control the flow of saline tothe patient.
9 Anticoagulant (AC) line (orange) Used to carry AC from the AC container to the inlet line manifold.Also includes:
• Spike (orange): Used to spike the AC container.
• Sterile barrier filter (0.2 microns): Prevents bacteria fromentering the system to maintain a functionally closed system.
10 Saline line (green) Used to carry saline from the saline container to the inlet linemanifold and the return line manifold. Also includes:
• Spike (green): Used to spike the saline container.
• Sterile barrier filter (0.2 microns): Prevents bacteria fromentering the system to maintain a functionally closed system.
11 Inlet line trap Prevents larger solid cellular matter from entering the channel.
12 Inlet pressure sensordiaphragm
Attaches to the inlet pressure sensor on the front panel to monitorpressure in the inlet line.
13 Centrifuge pressure sensordiaphragm
Attaches to the centrifuge pressure sensor on the front panel tomonitor pressure in the