Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes, irises and equators. Now we are investigating STF EP process using old MHI-#0 cavity (with simple beam pipe). They seems to be Niobium-Oxide. However, the problem is not yet solved, not yet fixed. 07.17.2009 H. Hayano, KEK ???

Upload: jacqueline-golly

Post on 15-Jan-2016

214 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

Stain appeared after EP acid renewal

Field emission at low gradient happened on MHI-06, MHI-07, in VT.

Kyoto-camera found brown stain trace in beam pipes, irises and equators.

Now we are investigating STF EP process using old MHI-#0 cavity (with simple beam pipe).

They seems to be Niobium-Oxide.

However, the problem is not yet solved, not yet fixed.

07.17.2009

H. Hayano, KEK

???

Page 2: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

BL#6 1-cell equator, t = 306 deg. Downstream : Outside weld area

BL#6 #9-BP, t = 241 deg. -1

BL#6 #9-BP, t = 241 deg. -2

Brown spot and traces;

MHI-06: spot or traces on BP-#1 、 #1、 #2、 #3、 #4、 #8、 #9、 #9-BP (red indicate bad)

MHI-07: BP-#1、 #1、 #2、 #3、 #4、 #5、 #6、 #7、 #8、 #9、 #9-BP

Kyoto-camera picture examples using new LED illumination

Page 3: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

BL#6 9-cell equator, t = 034 deg. Upstream : Outside weld area

BL#6 3-cell equator, t = 291 deg.

BL#6 3-4 iris, t = 055 deg.

BL#6 8-9 iris, t = 053 deg.

BL#6 9 equator, t = 037 deg.

BL#6 9-cell equator, t = 184 deg. Upstream : Outside weld area

Page 4: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

OLD_MHI-#0 cavity EP test 2009/07/07

EP condition : 4th EP after acid renewal

removal : 20 um

cavity surface area : 9000 cm2

acid soaking area : 5400 cm2

target average current : 270 A

acid flow : 15 litter / min

cavity rotation : 1 rpm

cavity acid temp at start :  20.9℃

acid tank temp :  20.1℃

room temp :  31℃

Typical EP parameter

Page 5: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

2009/5/27, BL#6, 1st EP after acid renew

2009/6/10, BL#7, 2nd EP after acid renew

2009/7/01, BL#8, 3rd EP after acid renew

2009/4/08, BL#5, EP before acid renewal

4div

4div

4div

2div

Page 6: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

2009/3/25, ERL-2cell #1, EP before acid renew

2div

2008/12/10, BL#6, EP before acid renew

2div

2009/7/07, OLD_MHI-#0 4th EP after acid renew

4div

2009/7/07, OLD_MHI-#05th EP after acid renew

4div

Page 7: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

After EP-20um, 5min rotate, HPR#1 equator, t = 248 deg.

After EP-20um, 5min rotate, HPR#1 equator, t = 304 deg.

After EP-20um, 30min rotate, HPR#1 equator, t = 248 deg.

After EP-20um, 30min rotate, HPR#1 equator, t = 304 deg.

after 4th EP after 5th EP

after 4th EP after 5th EP

Page 8: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

After EP-20um, 5min rotate, HPR#1 equator, t = 349 deg.

After EP-20um, 30min rotate, HPR#1 equator, t = 349 deg.

stains went away by EP 20um.

after 4th EP after 5th EP

Page 9: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

After EP-20um, 30min rotate, HPR#1 equator, t = 342 deg.

After EP-20um, 30min rotate, HPR#1 equator, t = 328 deg.

After EP-20um, 30min rotate, HPR#1 equator, t = 000 deg.

new stains comes up

after 5th EP

after 5th EP

after 5th EP

Page 10: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

stain in beam pipe edge

scrape out

scraped sample

sample was glued by silver paste on Si substrate.Then, apply XPS analysis.

results;Nb 11 at%O 48 at%C 27 at%Si 14 at% (substrate)

No special material,just Niobium-oxide, maybe.

MHI#0 cavity

Page 11: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

AES001 had also similar stains, observed as received.( Jlab EP )

Page 12: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

AES001 #2-3 iris 212°

1mm

iris

back side of EL

Other spots that have no wall gradientsOther spots that have no wall gradients

Appendixback side of EL

to Equatorof #3 cell

to Equatorof #2 cell

arc scar?The result of SI measurements shows that this spot is flat.This spots looks like a sign of Field Emission or Arcing. Many spots like this were observed.

Page 13: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

AppendixAES001 #4-5 iris 233°

1mm

irisback side of EL back side of EL

bright belt

to Equatorof #5 cell

All the Iris regions are yellowish. This spot is locally bright. It may be caused by EP?

to Equatorof #4 cell

Page 14: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

AppendixAES001 #4-5 iris 136°

1mm

iris

to Equatorof #5 cellDropped

coffee smear?

back side of EL back side of EL

to Equatorof #4 cell

Page 15: Stain appeared after EP acid renewal Field emission at low gradient happened on MHI-06, MHI-07, in VT. Kyoto-camera found brown stain trace in beam pipes,

AppendixAES001 #7-8 iris 279°

1mmiris

to Equatorof #8 cell

back side of EL back side of EL

to Equatorof #7 cell

Dropped coffee smear?

scar?