G O D D A R D S P A C E F L I G H T C E N T E R
Status and improvements of J1 OMPS pre-launch calibration
Matt Kowalewski, USRA
14 May 2014
![Page 2: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/2.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
Outline
• Instrument design changes – Wavelength coverage – QVD
•Calibration test phase summary •Calibration issues
– Diffuser stability – G/I and R recalibration summary
• Summary
2
![Page 3: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/3.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
JPSS OMPS instrument design changes
•No Limb sensor • J1 Nadir instrument overview
– 110deg FOV – Nadir Profiler: 250-310nm – Total Column: 305-380, 417nm* – Enhanced spatial resolution with new timing patterns*
•Nadir Profiler: 250km to TBD •TC Mapper: 50km to 15km
– 2 quasi-volume diffusers* – TC slit redesigned to reduce “puckering”* – Optical mounts redesigned to improve boresight
stability* *Differences wrt NPP OMPS
3
![Page 4: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/4.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
417 nm
J1 OMPS TC wavelength coverage
• J1 Total Column (TC) modified optical alignment permits wavelengths up to ~420nm to be measured.
4
![Page 5: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/5.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: -54deg Left: 417nm Right:372nm
5
![Page 6: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/6.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: -48deg Left: 417nm Right:372nm
6
![Page 7: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/7.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: -43deg Left: 417nm Right:372nm
7
![Page 8: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/8.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: -37deg Left: 417nm Right:372nm
8
![Page 9: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/9.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: -32deg Left: 417nm Right:372nm
9
![Page 10: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/10.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: -27deg Left: 417nm Right:372nm
10
![Page 11: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/11.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: -21deg Left: 417nm Right:372nm
11
![Page 12: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/12.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: -16deg Left: 417nm Right:372nm
12
![Page 13: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/13.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: -10deg Left: 417nm Right:372nm
13
![Page 14: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/14.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: -5deg Left: 417nm Right:372nm
14
![Page 15: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/15.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: 0deg Left: 417nm Right:372nm
15
![Page 16: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/16.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: +5deg Left: 417nm Right:372nm
16
![Page 17: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/17.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: +10deg Left: 417nm Right:372nm
17
![Page 18: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/18.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: +16deg Left: 417nm Right:372nm
18
![Page 19: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/19.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: +21deg Left: 417nm Right:372nm
19
![Page 20: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/20.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: +27deg Left: 417nm Right:372nm
20
![Page 21: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/21.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: +32deg Left: 417nm Right:372nm
21
![Page 22: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/22.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: +37deg Left: 417nm Right:372nm
22
![Page 23: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/23.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: +43deg Left: 417nm Right:372nm
23
![Page 24: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/24.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: +48deg Left: 417nm Right:372nm
24
![Page 25: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/25.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
OGTC Az: +54deg Left: 417nm Right:372nm
25
![Page 26: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/26.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
417nm Line Fits
26
![Page 27: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/27.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
372nm Line Fits
27
![Page 28: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/28.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
J1 OMPS QVD
•NPP OMPS utilized ground aluminum diffusers. •New diffuser (QVD) design implemented in order to
minimize spectral features in solar calibrations. – Reduces wavelength dependent albedo calibration
uncertainty. – Reduces time required for ground characterization.
•Design
28
Flight 1 Diffusers/Wipers
Ground Fused Silica
Aluminum & Chrome
Epoxy c/o
BATC
![Page 29: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/29.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
J1 OMPS QVD – Diffuser Features
• Diffuser features significantly reduced in J1 QVD. • Colored lines are individual rows. • Solid black is the macro-pixel average.
NPP J1
![Page 30: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/30.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
Outline
• Instrument design changes – Wavelength coverage – QVD
•Calibration test phase summary •Calibration issues
– Diffuser stability – G/I and R recalibration summary
• Summary
30
![Page 31: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/31.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
Calibration test phase summary
reference: IN0092-TST2-054
![Page 32: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/32.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
Calibration test phase highlights
•No significant changes to performance requirements. •QVD
– Smaller goniometry step size: was 0.5deg; is 1deg – Reflectivity changes and conditioning necessitated
goniometry, irradiance, and radiance calibration checks after ISS TVAC.
•Wavelength coverage – Band pass measurements at 417nm – Stray light PSF measurements at 417nm (TBC)
•Air to vacuum albedo check – Verify instrument albedo calibration consistent in air
and vacuum conditions. – Performed during ISS TVAC testing.
32
![Page 33: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/33.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
Outline
• Instrument design changes – Wavelength coverage – QVD
•Calibration test phase summary •Calibration issues
– Diffuser stability – G/I and R recalibration summary
• Summary
33
![Page 34: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/34.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
J1 OMPS QVD
• QVDs experienced optical degradation over course of ground testing. – Failure Review Board found
that epoxy exposure to UV light caused change in its optical characteristics.
– Aluminum coating on back side of diffuser did not fully cover roughened back surface, thus allowing UV light to interact with epoxy.
34
• BATC performed “conditioning” of diffusers in order to stabilize reflectivity. • Verification tests performed after conditioning.
– Goniometry and absolute irradiance calibration – Absolute radiance calibration
c/o BATC
![Page 35: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/35.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
J1 OMPS QVD – Verification Tests
• Irradiance and Goniometry – Repeated goniometry at 3 diffuser positions to verify
that QVD characterization from 2012 still valid. • 2012 and 2014 goniometry matched to within about
0.5% for repeated diffuser positions. •Correction methodology developed using 2014 data.
– Repeated absolute irradiance calibration at all positions for most accurate albedo calibration.
•Radiance Check and ReTest – Subset of radiance calibration performed to verify
2012 characterization still valid. – Differences ~2% seen, prompting repeat of full
radiance calibration.
35
![Page 36: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/36.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
J1 OMPS QVD – TC Goniometry Comparison
36
• Plots at right compare azimuthal dependence of Total Column flight diffuser gonimetry before (black) and after (green) “conditioning”.
• Differences are relatively small (~0.25%).
• Effect of this error would be to cause season dependence in derived ozone.
![Page 37: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/37.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
J1 OMPS QVD Stability
37
• Final UV soak of both flight diffusers performed to ensure stability.
• Plots at right show comparison between pre- and post-soak absolute irradiance calibration measurements.
• Multiple light sources used (colors). • Time separation between measurements
approximately 8 hours xenon arc exposure.
• Results demonstrate stability in both diffusers
to within measurement uncertainty (~0.75). NP
NP TC
TC
Conclusion: J1 OMPS calibration stability and accuracy meets science requirements.
![Page 38: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/38.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
Outline
• Instrument design changes – Wavelength coverage – QVD
•Calibration test phase summary •Calibration issues
– Diffuser stability – G/I and R recalibration summary
• Summary
38
![Page 39: Status and improvements of J1 OMPS pre-launch …...G O D D A R D S P A C E F L I G H T C E N T E R J1 OMPS QVD – Verification Tests •Irradiance and Goniometry – Repeated goniometry](https://reader034.vdocument.in/reader034/viewer/2022050323/5f7c448b21d15658a67b979a/html5/thumbnails/39.jpg)
G O D D A R D S P A C E F L I G H T C E N T E R
Summary
•QVD implementation yields improvements in the albedo uncertainty budget.
• Extended wavelength coverage potentially enhances science return and no significant stray light effects.
•No major differences in Acceptance Test Program.
39
Radiance Irradiance Albedo – Wvl Independent
Albedo – Wvl Dependent
NP TC NP TC NP TC NP TC NPP Goniometry 0 0 0.38 0.41 0.38 0.41 0.15 0.36
J1 Goniometry 0 0 0.21 0.21 0.21 0.21 0.1 0.11
NPP OMPS 3.383 3.067 3.499 3.194 1.653 1.717 0.426 0.497
J1 OMPS (Est) 2.36 1.81 2.57 2.04 1.62 1.71 0.29 0.31
Requirement 8 8 7 7 2 2 0.5 0.5