Page 1: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 1 GWDAW 10, Brownsville, December 2005

Status of GEO 600Stefan Hild, AEI Hannover

for the GEO-team

Page 2: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 2 GWDAW 10, Brownsville, December 2005

Overview

•Some examples from commissioning work

- increasing circulating power

- radiation pressure compensation

- reduction of feedback noise

• Calibration

• Data quality

• Detector characterization, stationarity

• Sensitivity progress

Page 3: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 3 GWDAW 10, Brownsville, December 2005

GEO 600 layout

3 kW

0.09

9 W

4 W

5 kW

• Increased to full input power • Still not the full power build-up IFO due to unexplained power losses

Page 4: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 4 GWDAW 10, Brownsville, December 2005

Radiation pressure compensation

High power inside the modecleaners in combination with low weight suspended mirrors causes strong radiation pressure effects

Right after lock acquisition mirrors are pushed by RP Saturation of the actuator (Laser PZT) Solution: Apply a bias force to the mirrors for acquisition and reduce this force in lock corresponding to the power build-up.

Page 5: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 5 GWDAW 10, Brownsville, December 2005

Optical transfer function

Signal Recycling provides an adjustable optical response

Lock acquistion at 2.5 kHz, OPERATION NOW at 350 Hz.

Page 6: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 6 GWDAW 10, Brownsville, December 2005

Noise projections / instrumental vetoes

• Useful tool for commissioning• Can be used for instrumental vetoes

Page 7: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 7 GWDAW 10, Brownsville, December 2005

Noise hunting example 1:Reduction of feedback noise in SR-loop

FB noise in Signal Recycling loop:

Shot noise from camera causes FB noise in detection band

Solution: Using a digital loop with strong filtering above unity gain.

Improved sensitivity below 500 Hz

Page 8: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 8 GWDAW 10, Brownsville, December 2005

Noise hunting example 2:Reduction of feedback noise in MI-loop

Noise mainly caused in HVA for ESD

Due to dynamic range constraints use whitening-dewhitenig

Passive dewhitening done in HV path(0-1kV)

Improved sensitivity from 50 to 600 Hz

Page 9: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 9 GWDAW 10, Brownsville, December 2005

Noise hunting example 3:Attacking Phase noise

• Indication that Phase noise is nearly limiting

• Implementation of a new RF-setup for Michelson modulation (higher signal levels, higher quality components)

Increased robustness of the detector, but no sensitivity improvement.

Page 10: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 10 GWDAW 10, Brownsville, December 2005

Calibration of GEO600

• 2 main outputs (each containing GW signal)• both calibrated to strain using time-domain method• making an optimal combination

Calibration quality can be judged by chi^2 value (data quality flag)

Page 11: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 11 GWDAW 10, Brownsville, December 2005

Null-Stream Veto

Page 12: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 12 GWDAW 10, Brownsville, December 2005

Independent check of calibration using:

Wavelength: 1035 nm Max. power: 1.4 W

Photon pressure calibrator

Good agreement with ESD.

Page 13: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 13 GWDAW 10, Brownsville, December 2005

Sensitivity improvement of GEO

Peak sensitivity = 4e-22 @ 550 Hz

Page 14: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 14 GWDAW 10, Brownsville, December 2005

Inspiral range

Optimal orientation

Detector characterization information available on daily basis: www.geo600.uni-hannover.de/georeports/

Page 15: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 15 GWDAW 10, Brownsville, December 2005

Noise stationarity / BLRMS of h(t)

Page 16: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 16 GWDAW 10, Brownsville, December 2005

Future plans

• No hardware changes foreseen

• Continue intense noise hunting

• Overnight and weekend runs

Best guess: start around Christmas

Provide a well-calibrated detector output

• Join S5

When additional improvements will take too long or are too risky to implement

Page 17: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 17 GWDAW 10, Brownsville, December 2005

E n d

Page 18: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 18 GWDAW 10, Brownsville, December 2005

Sensitivity improvement of GEO

Peak sensitivity = 4e-22 @ 550 Hz

Page 19: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 19 GWDAW 10, Brownsville, December 2005

Instrumental vetos

Page 20: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 20 GWDAW 10, Brownsville, December 2005

CHI^2 data quality flag

HACRmon

h(t) @ 1kHz

chi2

Example: MU3-Glitches

(Problem was fixed during S4)

Page 21: Stefan Hild 1GWDAW 10, Brownsville, December 2005 Status of GEO 600 Stefan Hild, AEI Hannover for the GEO-team

Stefan Hild 21 GWDAW 10, Brownsville, December 2005

S4 duty cycle (chi^2 cutoff dependent)