Strain Gage Calibration Factors for Constant Room Temperature Conditions (Or equivalently, measurement of the room temperature (Or equivalently, measurement of the room temperature Gage Resistance , Gage Factor and Transverse Sensitivity Coefficient )

Upload: others

Post on 28-Feb-2021

14 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Strain Gage Calibration Factors for Constant Room Temperature

Conditions

(Or equivalently, measurement of the room temperature (Or equivalently, measurement of the room temperature Gage Resistance, Gage Factor and Transverse

Sensitivity Coefficient)

Page 2: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

As previously discussed, photolithography is used to produce a large number of strain gages from a single parent metal foil…called a strain gage “lot”

Page 3: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

The lot number is provided by the manufacturer when a strain gage is purchased:

Typical data sheet provided with strain gages

Lot Number

Page 4: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Gage Resistance, Gage Factor and Transverse Sensitivity• The gage resistance, gage factor (Sg) and transverse sensitivity coefficient (Kt) are strain gage calibration constants measured by the gage manufacturer and provided to the user (average and standard deviation of the lot provided):

Gage Resistance

Typical data sheet provided with strain gages

Gage Factor, Sg

Transverse Sensitivity Coefficient, Kt

Page 5: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Gage Resistance, Gage Factor and Transverse Sensitivity

• The gage resistance, gage factor (Sg) and transverse sensitivity coefficient (Kt) are strain gage calibration constants measured by the gage manufacturer and provided to the user (average and standard deviation of the lot provided):

• The gage factor Sg relates the change in gage resistance due to axial strains• The transverse sensitivity coefficient Kt relates the change in gage resistance

to transverse strains to transverse strains

• BothSg and Kt are measured in accordance with ASTM E251-92

Page 6: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Measurement of the Gage Factor• Several gages from a lot of gages are bonded to a constant stress cantilever

beam described in the E251 standard and made from a standard calibration material (normally 1018 steel).

• Specimen design and loading frame allows both tensile and compressive stress/strains to be easily applied

Photo courtesy Becky Showalter, M-M GroupSketch of constant stress beam (from ASTM E251-92)

Page 7: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Measurement of the Gage Factor (cont’d)

• Initial resistance is measured (R)

• Stress is increased until an axial strain ε = ±1000 µε is induced at gage site(i.e., separate gages mounted on tensile and compressive sides of cantilever beam)

• Corresponding change in resistance is measured (∆R)

• Gage factor calculated:

• Measurement repeated for several gages; average and tolerance provided to user

( )ε

RR

Sg

∆=

Page 8: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Measurement of the Gage Factor (cont’d)

• Note: The strain sensitivity of the gage alloy, SA, is often confused with the gage factor , Sg:

( )ε

RR

Sg

∆=( )

ε

∆R

RSA =

(….confusion is understandable!!)

• Suppose a constantan foil gage is used. Would you expect the gage factor for the constantan foil gage to differ from the strain sensitivity of the constantan wire?

ε

Page 9: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Measurement of the Gage Factor (cont’d)

• During measurement of SA the constantan wire is subjected to a uniaxial stress and allowed to contract “naturally” (i.e., according to the Poisson ratio of constantan, νconst)

• During measurement of Sg the constantan foil is bonded to a calibration material, and the assembly is subjected to a uniaxial stress; the constantan material, and the assembly is subjected to a uniaxial stress; the constantan foil gage does not contract “naturally”, but rather is forced to contract according to the Poisson ratio of the calibration material, νo (Poisson ratio of 1018 steel is 0.285)

• So, the state of strain during measurement of SA is not necessarily equivalent to the state of strain during measurement of Sg ….

• Would you expect the gage factor for the constantan foil gage to differ from the strain sensitivity of the constantan wire?

Yes (….unless νo = νconst ….)

Page 10: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Significance of the Transverse Sensitivity Coefficient

• Given: a strain gage subjected to in-plane strains referenced to the a-tcoordinate system: εa, εt, γat

• Define three types of strain gage strain sensitivities

Sa = axial strain sensitivity = (∆R/R)/εa

St = transverse strain sensitivity = (∆R/R)/εt

Ss = shear strain sensitivity = (∆R/R)/γat

Page 11: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Significance of the Transverse Sensitivity Coefficient

• Assuming gage alloys exhibits linear strain sensitivities, principle of superposition applies:

• Experimental measurements show:

(6.4) atsttaa SSSR

R γεε ++=∆

Ss = shear strain sensitivity = 0

• Define the transverse sensitivity coefficient: Kt = St/Sa

Eq (6.4) becomes:

(6.5) )( ttaa KSR

R εε +=∆

Page 12: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Significance of the Transverse Sensitivity Coefficient

• During measurement of the Sg:

εt = -νoεa

• According to Eq (6.5) then:

R∆

• Therefore:

)1( atoa KSR

R εν−=∆

( ))1( toa

ag KSR

RS ν

ε−=

∆=

Page 13: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Measurement of the Transverse Sensitivity Coefficient• Kt is measured using a special

fixture that induces an (in-plane) uniaxial strain field (described in ASTM E251 standard)

• Separate gages oriented to measure- Axial strain: Sa = (∆R/R)/εa

- Transverse strain: St = (∆R/R)/εt

- Kt = St/Sa

Photo courtesy Becky Showalter, M-M GroupSketch of transverse sensitivity test rig (from ASTM E251-92)

Page 14: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Measurement of the Transverse Sensitivity Coefficient• Transverse sensitivity is an undesirable effect…gage manufacturers have

successfully minimized (but not eliminated) this phenomenon

Page 15: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Correcting for Transverse Sensitivity EffectsBiaxial (Tee) Rosettes

• Denote measured strains as εmx and εmy

• Strains corrected for transverse sensitivity effects are:

[ ]to KK εενε −

−=

)1(

where: Kt = transverse sensitivity coefficientνo = Poisson ratio of the calibration

material used by gage manufacturer (usually νo = 0.285)

[ ]

[ ]mxtmyt

toy

mytmxt

tox

KK

K

KK

K

εενε

εενε

−−

−=

−−

−=

)1(

)1(

)1(

)1(

2

2

Page 16: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Correcting for Transverse Sensitivity Effects3-Element Rectangular Rosettes

• Denote measured strains as εmx, εm45, and εmy

• Strains corrected for transverse sensitivity effects are:

[ ]mytmxto

x KK

K εενε −−

−=

)1(

)1(2

where: Kt = transverse sensitivity coefficientνo = Poisson ratio of the calibration

material used by gage manufacturer (usually νo = 0.285)

[ ]

[ ]

[ ]mxtmyt

toy

mmymxtmt

to

mytmxt

x

KK

K

KK

K

K

εενε

εεεενε

−−

−=

−+−−

−=

)1(

)1(

)()1(

)1(

)1(

2

4545245

2

Page 17: Strain Gage Calibration Factors for Constant Room ...courses.washington.edu/mengr556/StraingageCalibration.pdf• The gage resistance, gage factor ( Sg) and transverse sensitivity

Correcting for Transverse Sensitivity Effects3-Element Delta Rosettes

• Denote measured strains as εmx, εm60, and εm120

• Strains corrected for transverse sensitivity effects are:

( )

+−

+−

−=

3

2

31

)1(

)1(120602 mm

tmx

t

t

tox

KK

K

K εεενε

where: Kt = transverse sensitivity coefficientνo = Poisson ratio of the calibration

material used by gage manufacturer (usually νo = 0.285)

+−

+−

−=

+−

+−

−=

)(3

2

31

)1(

)1(

)(3

2

31

)1(

)1(

33)1(

601202120

12060260

2

mmxt

mt

t

to

mmxt

mt

t

to

t

KK

K

K

KK

K

K

K

εεενε

εεενε