Page 1: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

Stratos®Pro A4... MSPH/MSPHInstruction Manual

www.knick.deLatest Product Information:

Page 2: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

2

WarrantyWarrantyDefects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender).Sensors, fittings, and accessories: 1 year.

Subject to change without notice.

Return of Products Under WarrantyPlease contact our Service Team before returning a defective device. Ship the cleaned device to the address you have been given. If the device has been in contact with process fluids, it must be decontaminated/disinfected before shipment. In that case, please attach a corresponding certificate, for the health and safety of our service personnel.

DisposalPlease observe the applicable local or national regulations concerning the disposal of “waste electrical and electronic equipment”.

Page 3: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

3

Documents Supplied

CD-ROMComplete documentation:• Instruction manuals• Safety instructions • Short instructions

Safety InformationIn official EU languages and others.• ATEX / IECEX / FM / CSA• EC Declarations of Conformity

Short InstructionsIn German, English, French, Russian,Spanish, Portuguese, Swedish, and Dutch. More languages on CD-ROM and on our website: www.knick.de• Installation and commissioning• Operation• Menu structure• Calibration• Error messages and recommended actions

Specific Test Report

Stratos® Pro A211 / A411

Other languages: www.knick.de

Kurzübersicht .............................. 15

QuickStart .................................... 27

Быстрый старт ........................... 39

Inicio rápido ................................ 51

Início rápido ................................ 63

Guida rapida ................................ 75

Short Instructions .........................3

Snabbstart.................................... 87

QuickStart .................................... 99

Stratos® Pro SeriesSafety Instructions

www.knick.de

Page 4: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

4

ContentsDocuments Supplied ..................................................................... 3Introduction .................................................................................... 7

Intended Use ......................................................................................... 7

Safety Information ......................................................................... 8Overview of Stratos Pro A4... MSPH/MSPH ............................... 9Assembly ........................................................................................10

Package Contents ..............................................................................10Mounting Plan, Dimensions ..........................................................11Pipe Mounting, Protective Hood ..................................................12Panel Mounting ..................................................................................13

Installation .....................................................................................14Installation Instructions ...................................................................14Rating Plates / Terminal Assignments ........................................14Power Supply, Signal Lines .............................................................15Connecting Sensors A and B .........................................................16Memosens Cable ................................................................................17Memosens Cable CA/MS-... ............................................................18

Connecting a Memosens Sensor ...............................................19User Interface, Keypad ................................................................20Display ............................................................................................21

Signal Colors (Display Backlighting) ...........................................21

Measuring Mode...........................................................................22Selecting the Mode / Entering Values ......................................23Operating Modes ..........................................................................24

Menu Structure of Modes and Functions .................................25HOLD Mode .........................................................................................26Alarm ......................................................................................................27

Overview of Configuration .........................................................28

Page 5: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

5

ContentsSetup and Channel Selection on the Device ..........................29

Channel Selection and Display Assignment ............................29

Calculations (CALC) ......................................................................30Configuration ................................................................................31

Configuration (Original for Copy) ................................................35Sensor A ................................................................................................38Sensor B .................................................................................................44Measuring Mode ................................................................................50Current Output 1 ................................................................................52Current Output 2 ................................................................................58Alarm Settings .....................................................................................60Limit Function .....................................................................................62Pulse Length / Pulse Frequency Controller ..............................71Controller .............................................................................................72WASH Contact .....................................................................................76Time and Date .....................................................................................78Tag Number .................................................................................................78

Calibration .....................................................................................80Selecting a Calibration Mode ........................................................81Zero Adjustment (ISFET) .................................................................82Automatic Calibration (Calimatic) ................................................84Manual Calibration with Buffer Entry .........................................86Data Entry of Premeasured Sensors............................................88Product Calibration (pH) .................................................................90

Measurement ................................................................................93Diagnostics ....................................................................................95Service ..........................................................................................100Operating States .........................................................................104Product Line and Accessories ..................................................105Specifications ..............................................................................106

Page 6: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

6

Error Handling .............................................................................113Error Messages ............................................................................114Sensoface .....................................................................................117FDA 21 CFR Part 11 ....................................................................120

Electronic Signature – Passcodes ...............................................120Audit Trail ............................................................................................120

Index .............................................................................................121Trademarks .........................................................................................131

Passcodes .....................................................................................132

Contents

Page 7: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

7

IntroductionIntended UseStratos Pro A4... MSPH/MSPH is a two-channel process unit for pH measurement using 2 Memosens pH sensors. The two measured values can be used for various differential calculations (pH, mV, temperature).The sturdy molded enclosure can be fixed into a control panel or mounted on a wall or at a post. The protective hood, which is avail-able as accessory, provides additional protection against direct weather exposure and mechanical damage.

Plain-text messages in a large, backlit display allow intuitive operation. The colored display backlighting signals alarm messages (red) or HOLD mode (orange).The “Sensocheck“ automatic monitoring of sensor and cables and the “Sensoface“ function for clear indication of the sensor condition provide excellent diagnostics. The internal logbook (TAN SW-A002) can handle up to 100 entries – up to 200 with AuditTrail (TAN SW-A003). Password protection for granting access rights during operation can be configured. The floating, digital control input “HOLD“ allows remote controlled switching to HOLD mode.The “CONTROL“ input allows evaluation of external monitoring equip-ment, such as a flow monitoring system.For that purpose, you can also monitor the current input (optional) for keeping a desired value.The device provides two current outputs (for transmission of mea-sured value and temperature, for example) and four floating relay contacts. Current is provided through a universal power supply 24 ... 230 V AC/DC, AC: 45 ... 65 Hz.

Page 8: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

8

Safety InformationSafety informationBe sure to read and observe the following instructions!The device has been manufactured using state of the art technology and it complies with applicable safety regulations.When operating the device, certain conditions may nevertheless lead to danger for the operator or damage to the device.

See also separate document:• “Safety Instructions“.

(EC Declaration of Conformity, FM, CSA, ATEX (if applicable) Certificates)

CAUTION!

Please note:Before commissioning, you must prove that the device may be connected with other equipment.

Commissioning must only be performed by trained personnel autho-rized by the operating company! Whenever it is likely that protection has been impaired, the device shall be made inoperative and secured against unintended operation. The protection is likely to be impaired if, for example:• the device shows visible damage• the device fails to perform the intended measurements• after prolonged storage at temperatures above 70 °C• after severe transport stresses

Before recommissioning the device, a professional routine test must be performed. This test must be carried out at the manufacturer's factory.

Terminals:Screw terminal, suitable for single wires / flexible leads up to 2.5 mm2 (AWG 14).Recommended torque for the terminal screws: 0.5 ... 0.6 Nm.

Page 9: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

9

OverviewOverview of Stratos Pro A4... MSPH/MSPH

8

10

9

5

6

Output 1

input

Current

Output 2 7

14

15

16

17

18

19

20

21

22

R1

R2

Alarm

R3

Wash

Power

11

12

13

HOLD

input

input

Control

1

2

3

4

Sensor APower supply

RS 485 A

RS 485 B

GND/Shield

A

B

C

D

Sensor B“MK-MS” module

Power supply

RS 485 A

RS 485 B

GND

EShield

Page 10: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

10

AssemblyPackage ContentsCheck the shipment for transport damage and completeness!The package should contain:• Front unit, rear unit, bag containing small parts• Specific test report• Documentation (cf Pg 3)• CD-ROM

Fig.: Assembling the enclosure

1) Jumper (3 x)2) Washer (1 x), for conduit

mounting: Place washer between enclosure and nut

3) Cable tie (3 x)4) Hinge pin (1 x), insertable

from either side5) Enclosure screw (4 x)

6) Sealing insert (1 x)7) Rubber reducer (1 x)8) Cable gland (3 x)9) Filler plug (3 x)10) Hexagon nut (5 x)11) Sealing plug (2 x), for sealing

in case of wall mounting

11

10

9

8

7 6 5 4

1

2

3

Page 11: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

11

AssemblyMounting Plan, Dimensions

Fig.: Mounting plan (All dimensions in mm!)

1) Cable gland (3 x)2) Knockouts for cable gland or

½" conduit, 21.5 mm dia. (2 knockouts) Conduits not included!

3) Knockout for pipe mounting (4 x)

4) Knockout for wall mounting (2 x)

148

148

42 4242

14

11741

21431

2

34 80

746.

2

4

3

Page 12: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

12

AssemblyPipe Mounting, Protective Hood

Fig.: ZU 0737 protective hood for wall and pipe mounting (All dimensions in mm!)

Fig.: ZU 0274 pipe-mount kit (All dimensions in mm!)

147199

91

185

1

2

3

4

1) Hose clamp with worm gear drive to DIN 3017 (2 x)

2) Pipe-mount plate (1 x)3) For vertical or horizontal posts

or pipes4) Self-tapping screw (4 x)

ø40...ø60

Page 13: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

13

AssemblyPanel Mounting

Fig.: ZU 0738 panel-mount kit (All dimensions in mm!)

1) Circumferential sealing (1 x)

2) Screw (4 x)3) Position of control panel4) Span piece (4 x)5) Threaded sleeve (4 x)

Cutout 138 x 138 mm (DIN 43700)

<30 76 31

1...221

2345

Page 14: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

14

InstallationInstallation Instructions• Installation of the device must be carried out by trained experts in

accordance with this instruction manual and as per applicable local and national codes.

• Observe the technical specifications and input ratings during installation!

• Be sure not to notch the conductor when stripping the insulation!• The supplied current must be galvanically isolated.

If not, connect an isolator module.• All parameters must be set by a system administrator prior to

commissioning!

Terminals: suitable for single wires / flexible leads up to 2.5 mm2 (AWG 14)

Rating Plates / Terminal Assignments

Fig.: Stratos Pro A4...N rating plate at bottom of front

Fig.: Terminal assignments of Stratos Pro A4...

Page 15: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

15

Areas for placing the screwdriver to pull out the terminals

Fig.: Terminals, device opened, back of front unit

1

Power Supply, Signal LinesConnect the power supply for Stratos Pro A4... MSPH/MSPH to terminals 21 and 22(24 ... 230 V AC, 45  ... 65 Hz / 24 ... 80 V DC)

22

Assignments

Page 16: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

16

Areas for placing the screwdriver to pull out the terminals

1 22

Fig.: Terminals, device opened, back of front unit

Sensor B terminals

Connect the sensor lines with the respective sensor terminals.

Connecting Sensors A and B

Sensor A terminals

Page 17: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

17

Memosens Cable

Power supply +

RS485 (A)

RS485 (B)

Power supply – (GND)

Shield

Brown

Green

Yellow

White

Transparent

Page 18: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

18

Memosens Cable CA/MS-...

Intended UseConnecting cable for contactless inductive digital transmission of measured signals (Memosens®).The connecting cable consists of an inductive connector for digital Memosens® sensors (bayonet lock). It allows connecting the ferrule-terminated wires from the sensor loop of the transmitter. Contactless inductive digital transmission of signals and energy eliminates the influence of humidity, electromagnetic fields and corrosion.

SpecificationsMaterial PEEKCable diameter 6.3 mmCable 2x2, twisted wire pairsLength up to 100 mProcess temperature –20 °C ... 135 °CIngress protection IP 68

Model CodeCable type Cable length Order numberMemosens® cable 3 m CA/MS-003NAA

5 m CA/MS-005NAA10 m CA/MS-010NAA20 m CA/MS-020NAA

Memosens® cable, Ex* 3 m CA/MS-003XAA5 m CA/MS-005XAA10 m CA/MS-010XAA20 m CA/MS-020XAAOther lengths available on request.

*) Ex-certified, ATEX II IG Ex ia IIC T3/T4/T6; The Type-Examination Certificate is enclosed with each Ex sensor.

Page 19: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

19

Connecting a Memosens Sensor

1 2 3 4

HE

RS 4

85 (A

)

RS 4

85 (B

)

GN

D/s

hiel

d

A B C D

HE

RS 4

85 (A

)

RS 4

85 (B

)

GN

D

E

shie

ld

Mem

osen

s

Mem

osen

s

Example:

Measuring task: pH/ORP, temperatureSensors (example): SE532/1-NMSN (Knick)Cable CA/MS-003-NAA

Yello

w

Whi

te

Shie

ld,

tran

spar

ent

Brow

n

Gre

en

Sens

or(s

)Ca

ble

Dev

ice

Yello

w

Whi

te

Shie

ld,

tran

spar

ent

Brow

n

Gre

en

Channel ATerminal row 1

Channel BModule terminals

Page 20: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

20

User Interface, Keypad

1 IrDA transmitter/receiver2 Display3 Keypad4 Rating plate

(enclosure bottom)

Key Functionmeas • Return to last menu level

• Directly to measuring mode (press > 2 s)info • Retrieve information

• Show error messagesenter • Measuring mode: Display output current

• Configuration: Confirm entries, next configuration step

• Calibration: Continue program flow

Arrow keysup / down

• Measuring mode: Call menu• Menu: Increase/decrease a numeral• Menu: Selection

Arrow keysleft / right

• Measuring mode: Call menu• Previous/next menu group• Number entry: Move between digits

1

3

4

2

MEMO SENS

Page 21: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

21

Display

MEMO SENS

1 Temperature2 Sensocheck3 Interval/response time4 Sensor data5 Digital sensor devaluated6 Limit values7 Alarm8 Service9 Sensor A/B10 Calibration11 Digital sensor 12 Waiting time running

13 Info available14 Hold mode active15 Main display16 Secondary display17 Proceed using enter18 Digital sensor19 Diagnostics20 Configuration mode21 Calibration mode22 Measuring mode23 Sensoface24 Unit symbols

1 2 3 4 5 6 7 8 9 10 11

171819202122

121314

15

16

2324

Signal Colors (Display Backlighting)

Red AlarmOrange HOLD mode (Calibration, Configuration, Service)Turquoise DiagnosticsGreen InfoPurple Sensoface message

Page 22: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

22

Measuring ModeAfter the operating voltage has been connected, the analyzer automatically goes to “Measuring“ mode. To call the measuring mode from another operating mode (e.g. Diagnostics, Service): Hold meas key depressed (> 2 s).

Process variable

Temperature

Sensoface indicator(sensor status)

Time

Mode indicator (measuring)

Hold meas key depressed to call the measuring mode(pressing once more switches the display)

Display sensor A / sensor B

enter key

In measuring mode the display indicates:• Measured values as selected (OUT 1 / OUT 2) and time (24/12 h

AM/PM) as well as temperature in °C or °F (formats selected during configuration)

By pressing the meas key in measuring mode you can view the follow-ing displays (for approx. 60 sec):• Measured values channel A and measured values channel B• Measured value and tag (point of measurement designation –

entered during configuration)• Time and date• Controller (if configured), upper display: controller output Y,

lower display: setpoint (can be edited) Pressing the enter key shows the output currents. They are displayed as long as enter is held depressed, then the measured-value display will return after 3 sec.

You must configure the device for the respective measurement task!

Page 23: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

23

Selecting the Mode / Entering Values

To enter a value:5) Select numeral: left / right arrow key6) Change numeral: up / down arrow key7) Confirm entry by pressing enter

To select the operating mode:1) Hold meas key depressed (> 2 s) (measuring mode)2) Press any arrow key: the selection menu appears3) Select operating mode using left / right arrow key4) Press enter to confirm the selected mode

Selection menu

Selected mode (blinks)

1 2

3

4

5 6

7

Page 24: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

24

DiagnosticsDisplay of calibration data, display of sensor data, performing a device self-test, viewing the logbook entries, display of hardware/software versions of the individual components. The logbook can store 100 events (00...99). They can be displayed directly on the device. The logbook can be extended to 200 entries using a TAN (Option).

HOLDManual activation of HOLD mode, e.g. for replacing a digital sensor. The signal outputs adopt a defined state.

CalibrationEvery sensor has typical characteristic values, which change in the course of the operating time. Calibration is required to supply a cor-rect measured value. The device checks which value the sensor delivers when measuring in a known solution. When there is a deviation, the de-vice can be “adjusted“. In that case, the device displays the “actual“ value and internally corrects the measurement error of the sensor. Calibration must be repeated at regular intervals. The time between the calibration cycles depends on the load on the sensor. During calibration the device is in HOLD mode. During calibration the device remains in the HOLD mode until it is stopped by the operator.

ConfigurationThe analyzer must be configured for the respective measurement task. In the “Configuration“ mode you select the connected sensor, the measuring range to be transmitted, and the conditions for warning and alarm messages. During configuration the device is in HOLD mode. Configuration mode is automatically exited 20 minutes after the last keystroke. The device returns to measuring mode.

ServiceMaintenance functions (current source, relay test, controller test), IrDA operation, passcode assignment, reset to factory settings, enabling of options (TAN).

Operating Modes

Page 25: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

25

Menu Structure of Modes and Functions

Measuring mode

TAG display

Manual activation of HOLD mode, e.g. for sensor replacement.The signal outputs behave as configured (e.g. last measured value, 21 mA)

Calibrating sensor A

Adjusting the pH

Product calibration

Zero adjustment (for ISFET only)

Calibrating sensor B (see sensor A for details)

Configuration

Display of measured values for validation (simulators)

Current source, output 1

Current source, output 2

Relay test

Controller: manual specification of controller output

Activating the IrDA interface

Specifying access codes for operating modes

Reset to factory setting

Enabling an option via TAN

Pressing any arrow key opens the selection menu. Select the menu group using the left/right arrow keys. Pressing enter opens a menu item. Press meas to return.

(Access via code, factory setting:5555)

meas Controller parameter display(if configured)

meas meas

after 60 s after 60 s

Display of calibration and sensor data Sensor A

Display of calibration and sensor data Sensor B

Self test: RAM, ROM, EEPROM, module

100 events with date and time

Display of measured values (mV_pH, mV_ORP, RTD, resistances of glass electrode, reference electrode)

Display of software version, model designation, serial number

CLK displaymeas

Page 26: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

26

HOLD ModeThe HOLD mode is a safety state during configuration and calibration. Output current is frozen (Last) or set to a fixed value (Fix). Alarm and limit contacts are inactive. The HOLD mode is indicated by orange display backlighting.

Terminating the HOLD mode The HOLD mode is ended by switching to measuring mode (hold meas key depressed). The display reads “Good Bye“, after that, the HOLD mode is exited. When the calibration mode is exited, a confirmation prompt ensures that the installation is ready for operation (e.g.: sensor reinstalled, located in process).

HOLD mode, display icon:

HOLD active

21

4HOLD active

Output current[mA]

Output signal for HOLDFIX setting = 21.0 mA

Output signal for HOLDLAST setting

Output signal during HOLD:

Output signal response• Last: The output current is frozen at its last value. Recommended

for short configuration procedures. The process should not change decisively during configuration. Changes are not noticed with this setting!

• Fix: The output current is set to a value that is noticeably different from the process value to signal the control system that the device is being worked at.

Page 27: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

27

Alarm

Manual activation of HOLDThe HOLD mode can be activated manually from the HOLD menu. This allows checking or replacing a sensor, for example, without provoking unintended reactions of outputs or contacts.Press meas key to return to selection menu.

AlarmWhen an error has occurred, Err xx is displayed immediately.Only after expiry of a user-defined delay time will the alarm be registered and entered in the logbook.During an alarm the display blinks, the display backlighting turns red.

Error messages can also be signaled by a 22 mA output current (see Configuration).The alarm contact is activated by alarm and power failure, see also “Configuration / Alarm Settings“.2 sec after the failure event is corrected, the alarm status will be deleted.

External activation of HOLD (SW-A005)The HOLD mode can be activated from outside by sending a signal to the Hold input (e.g. from the process control system).

11

12

Process control system

Power supply12...24 V AC/DC

Stratos Pro A4...

input

HOLD

HOLD inactive 0...2 V AC/DCHOLD active 10...30 V AC/DC

Page 28: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

28

Overview of ConfigurationThe configuration steps are assigned to different menu groups. Using and you can jump between the individual menu groups.Each menu group contains menu items for setting the parameters. Pressing enter opens a menu item. The values are edited using and . Pressing enter confirms/saves the settings. Return to measurement: Hold meas key depressed (> 2 s).

Select menu group Menu group Code Display Select menu

item

Sensor A parameters S_A:

Menu item 1

...

Menu item ...

Sensor B parameters S_B:

Measuring mode MES:

Current output 1 OT1:

Current output 2 OT2:

Alarm mode ALA:

Relay outputs REL:

Cleaning WSH:

Setting the clock CLK:

Tag number TAG:

enter

enter

enter

enter

Page 29: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

29

Setup and Channel Selection on the DeviceSensors A and B – Typical Arrangement

DIFF pH (A–B)DIFF mV (A–B)DIFF TMP (A–B)

Analyzer Stratos Pro A4... MSPH/MSPH

Display OUT 2

Display OUT 1DISPLAY:

Channel Selection and Display Assignment

Outlet:Sensor B with fitting

Inlet:Sensor A with fitting

Connectionvia Memosens cable, max. length 100 m

Page 30: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

30

Calculations (CALC)

CONF Calculation Formula Display-C1- Difference pH pH A – pH B dPH-C2- mV difference mV A – mV B dmV-C3- Temp difference TMP A – TMP B d°C (d°F)

Page 31: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

31

Configuration

Configuration Choices DefaultSensor AS_A: SENSOR ON / OFF ON

CALMODE AUTOMANDAT

AUTO

AUTO BUFFER SET -00- ... -09-, -U1- -00-

Please note: Pressing info displays nominal values and type of buffer set

U1(For specifiable buffer set,see Appendix: “Buffer Tables“)

EDIT BUFFER 1 (NO, YES) NO

Enter values for buffer 1

EDIT BUFFER 2 (NO, YES) NO

Enter values for buffer 2

CAL TIMER OFF, FIX, ADAPT OFF

ON CAL-CYCLE 0...9999 h 0168 h

CIP COUNT ON/OFF OFF

SIP COUNT ON/OFF OFFSensor BS_B: SENSOR ON / OFF OFF

CALMODE AUTOMANDAT

AUTO

AUTO BUFFER SET -00- ... -09-, -U1- -00-

Please note: Pressing info displays nominal values and type of buffer set

U1(For specifiable buffer set,see Appendix: “Buffer Tables“)

EDIT BUFFER 1 (NO, YES) NO

Enter values for buffer 1

EDIT BUFFER 2 (NO, YES) NO

Enter values for buffer 2

CAL TIMER OFF, FIX, ADAPT OFF

ON CAL-CYCLE 0...9999 h 0168 h

CIP COUNT ON/OFF OFF

SIP COUNT ON/OFF OFF

Page 32: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

32

Configuration

Output 1 (OUT1) / Output OT2 (OUT2) identical except for CHANNEL PH B

OT1: RANGE 0–20 mA, 4–20 mA 4-20 mA

CHANNEL PH A / PH B / mV A / mV B / TMP A / TMP B /CALC

PH A

PH BEGIN 4 mA (0 mA) –2.00...16 PH 00.00 PH

END 20 mA –2.00...16 PH 14.00 PH

mV BEGIN 4 mA (0 mA) -1999...1999 mV

END 20 mA -1999...1999 mV

TMP °C

BEGIN 4 mA (0 mA) -20...300 °C Select °C/°F

END 20 mA -20...300 °C at

TMP °F

BEGIN 4 mA (0 mA) -4...572 °F "MEAS_MODE“

END 20 mA -4...572 °F

dPH BEGIN 4 mA (0 mA) -19.99 ... 19.99 PH

END 20 mA

dmV BEGIN 4 mA (0 mA) -1999 ... 1999 mV

END 20 mA

d°C BEGIN 4 mA (0 mA) -199.9 ... 300.0 °C

END 20 mA

d°F BEGIN 4 mA (0 mA) -199.9 ... 500.0 °F

END 20 mA

FILTERTIME 0...120 SEC 0000 SEC

22mA-FAIL ON/OFF OFF

HOLD MODE LAST/FIX LAST

FIX HOLD-FIX (0) 4...22 mA 021.0 mA

Configuration Choices DefaultMEAS_MODEMES: TEMP UNIT °C / °F °C

CALCULATION ON/OFF OFF

ON (Selected in text line) -C1- Difference PH -C2- Difference mV-C3- Difference TMP

-C1-Difference PH

Page 33: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

33

Configuration Choices Default

Configuration

Alarm (ALARM)ALA: DELAYTIME 0...600 SEC 0010 SEC

SENSOCHECK ON/OFF OFF

CONTROL IN 1) ON/OFF OFF

LIMIT-IN 1) 2) ON/OFF OFF

ON FUNCTION Lo LEVL / Hi LEVL Lo LEVL

LEVEL 0 ... 22.00 mA 12 mA

HYSTERESIS 0 ... 10.00 mA 01.00 mA

Relay 1/2 (RL1/RL2)REL: (Selected in text line) LIMITS, CONTROLLER LIMITS

RL1: CHANNEL PH A / PH B / mV A / mV B / TMP A / TMP B /CALC

PH A

FUNCTION Lo LEVL, Hi LEVL Lo LEVL

CONTACT N/O, N/C N/O

LEVEL –2.00...16 PH(-1999...1999 mV)(-20...200 °C)

00.00 PH

HYSTERESIS 0...10.00 PH 00.50 PH

DELAYTIME 0...9999 SEC 0010 SEC

RL2: CHANNEL PH A / PH B / mV A / mV B / TMP A / TMP B /CALC

PH A

FUNCTION Lo LEVL, Hi LEVL Hi LEVL

CONTACT N/O, N/C N/O

LEVEL –2.00...16 PH(-1999...1999 mV)(-20...200 °C)

14.00 PH

HYSTERESIS 0...10.00 PH 00.50 PH

DELAYTIME 0...9999 SEC 0010 SEC

Page 34: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

34

Configuration

1) These menu items are displayed only when the respective hardware is installed and enabled.

2) LIMIT-IN can be used to measure and m onitor the current input. For monitoring, you can enter a setpoint which generates an alarm.

REL: CTR CHANNEL PH A / PH B / mV A / mV B / TMP A / TMP B

PH A

TYPE PLC / PFC PLC

PLC PULSE LEN 0...0600 SEC 0010 SEC

PFC PULSE FREQ 0...0180 P/M 0060 P/M

SETPOINT –2.00...16 PH –1500...1500 mV

07.00 PH

DEAD BAND 0...10.00 PH 01.00 PH

P-GAIN 10...9999% 0100%

I-TIME 0...9999 SEC 0000 SEC

D-TIME 0...9999 SEC 0000 SEC

HOLD MODE Y LAST/Y OFF Y LAST

Cleaning contact (WASH)WSH: WASH CYCLE 0.0...999.9 h 000.0 h

WASH TIME 0...9999 SEC 0060 SEC

CONTACT N/O, N/C N/O

Real-time clock (CLOCK)CLK: FORMAT 24 h / 12 h

24 h TIME hh/mm 00..23:00...59

12 h TIME hh/mm 00...11 AM/PM:00...59

DAY/MONTH 01...31/01...12

YEAR 2000...2099

Tag number (TAG)TAG: (Input in text line) ___

Configuration Choices DefaultRelay 1/2 (RL1/RL2) -continued-

Page 35: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

35

Configuration (Original for Copy)Please note:Fill in your configuration data on the following pages or use them as original for copy.

ParameterS_A: Sensor On/Off

S_A: Calibration mode

S_A: Select buffer set (-U1-, see Appendix)

S_A: Calibration timer

S_A: Calibration cycle

S_A: CIP counter

S_A: SIP counter

S_B: Sensor On/Off

S_B: Calibration mode

S_B: Select buffer set (-U1-, see Appendix)

S_B: Calibration timer

S_B: Calibration cycle

S_B: CIP counter

S_B: SIP counter

MES: Temperature unit

MES: Calculation

OT1: Current range

OT1: Channel (process variable)

OT1: Current start

OT1: Current end

OT1: Filter time

OT1: 22 mA error current

OT1: HOLD mode

OT1: HOLD-FIX current

Page 36: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

36

(Original for Copy) ConfigurationParameterOT2: Current range

OT2: Channel (process variable)

OT2: Current start

OT2: Current end

OT2: Filter time

OT2: 22 mA error current

OT2: HOLD mode

OT2: HOLD-FIX current

ALA: Delay

ALA: Sensocheck on/off

ALA: "Conrol" input

ALA: Current input

- Function (effective direction)

- Level

- Hysteresis

REL: Usage

RL1: Channel (process variable)

RL1: Function

RL1: Contact response

RL1: Setpoint

RL1: Hysteresis

RL1: Delay

RL2: Channel (process variable)

RL2: Function

RL2: Contact response

RL2: Setpoint

Page 37: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

37

Configuration (Original for Copy)ParameterRL2: Hysteresis

RL2: Delay

CTR: Channel (process variable)

CTR: Controller type

CTR: Pulse length

CTR: Pulse frequency

CTR: Setpoint

CTR: Neutral zone

CTR: P gain

CTR: I time

CTR: D time

CTR: HOLD mode

WSH: Wash cycle

WSH: Wash duration

WSH: Contact response

CLK: Time format

CLK: Time hh/mm

CLK: Day/month

CLK: Year

TAG: Tag number

Page 38: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

38

Configuration

1

Sensor ASelecting sensor A, calibration mode

1 Press any arrow key.2 Select CONF using keys, press enter.3 Select "Sensor A“ using ,

press enter, then select "MEAS_MODE = ON“. Confirm by pressing enter.

4 All items of this menu group are indicated by the “S_A:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

MEMO SENS

MEMO SENS

2

3

4

Select sensor A

Measuring channel on/off

Calibration mode

(AUTO: Buffer set)

Calibration timer

Calibration cycle

Cleaning cycles

Sterilization cycles

4enter

enter

5

enter

meas

Page 39: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

39

ConfigurationMenu item Action ChoicesSelect sensor

MEMO SENS

Select sensor using keys.

Press enter to confirm.

SENSOR_ASENSOR_B

Measuring channel

MEMO SENS

Press enter to select "On".

Switch to "OFF" if you do not want to use this chan-nel for measuring or if a sensor is not connected.

ON / OFF

Calibration mode

MEMO SENS

Select CALMODE using keys:AUTO: Calibration with Calimatic buffer set recognitionMAN: Manual entry of buffer solutionsDAT: Direct entry of adjustment dataPress enter to confirm.

AUTOMANDAT

(AUTO: Buffer set) Select buffer set using keys (see buffer tables for nominal values).

Press enter to confirm.

-00-...-09-, -U1-

Pressing the info key dis-plays the manufacturer and nominal values in the lower line.

4

Page 40: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

40

Configuration

1

Sensor ASelecting cal timer, cal cycle

1 Press any arrow key.2 Select CONF using keys, press enter.3 Select "Sensor A“ using ,

press enter, then select "MEAS_MODE = ON“. Confirm by pressing enter.

4 All items of this menu group are indicated by the “S_A:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

MEMO SENS

2

3

4

Select sensor A

Measuring channel on/off

Calibration mode

(AUTO: Buffer set)

Calibration timer

Calibration cycle

Cleaning cycles

Sterilization cycles

4enter

enter

5

enter

meas

Page 41: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

41

ConfigurationMenu item Action ChoicesCalibration timer Adjust CALTIMER using

keys:OFF: No timerADAPT: Maximum cal cycle (adjust in the next step)FIX: Fixed cal cycle (adjust in the next step)Press enter to confirm.

OFF/ADAPT/FIX

With ADAPT, the calibra-tion cycle is automatically reduced depending on the sensor load (high temperatures and pH val-ues) and the sensor wear.

Calibration cycle Only with FIX/ADAPT:Modify digit using keys, select next digit using keys.Press enter to confirm.

0...9999 h

Note for the calibration timer:When Sensocheck has been activated in the Configuration > Alarm menu, the expiration of the calibration interval is indicated by Senso-face:

Display Status

+

+

Over 80% of the calibration interval has already past.

The calibration interval has been exceeded.

The time remaining until the next due calibration can be seen in the diagnostics menu (see Diagnostics chapter).

4

Page 42: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

42

ConfigurationSensor AAdjusting CIP cleaning cycles, SIP sterilization cycles

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select "Sensor A“ using ,

press enter, then select "MEAS_MODE = ON“. Confirm by pressing enter.

4 All items of this menu group are indicated by the “S_A:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

MEMO SENS

2

3

4

Select sensor A

Measuring channel on/off

Calibration mode

(AUTO: Buffer set)

Calibration timer

Calibration cycle

Cleaning cycles

Sterilization cycles

4enter

enter

5

enter

meas

Page 43: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

43

ConfigurationMenu item Action ChoicesCIP / SIP

Cleaning cyclesOn / Off

Select ON or OFF using keys.Activates/deactivates log-ging in extended logbookPress enter to confirm.

ON/OFF

Sterilization cyclesOn / Off

Select ON or OFF using keys.Activates/deactivates log-ging in extended logbookPress enter to confirm.

ON/OFF

The cleaning and sterilization cycles are counted to measure the load on the sensor. Suitable for biochemical applications (process temp approx. 0...50 °C, CIP temperature > 55 °C, SIP temperature > 115 °C).

4

Page 44: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

44

Configuration

1

Sensor BSelecting sensor A, calibration mode

1 Press any arrow key.2 Select CONF using keys, press enter.3 Select "Sensor B“ using ,

press enter, then select "MEAS_MODE = ON“. Confirm by pressing enter.

4 All items of this menu group are indicated by the “S_B:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

MEMO SENS

MEMO SENS

2

3

4

Select sensor B

Measuring channel on/off

Calibration mode

(AUTO: Buffer set)

Calibration timer

Calibration cycle

Cleaning cycles

Sterilization cycles

4enter

enter

5

enter

meas

Page 45: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

45

ConfigurationMenu item Action ChoicesSelect sensor

MEMO SENS

Select sensor using keys.

Press enter to confirm.

SENSOR_A SENSOR_B

Measuring channel

MEMO SENS

Press enter to select "On".

Switch to "OFF" if you do not want to use this chan-nel for measuring or if a sensor is not connected.

ON / OFF

Calibration mode

MEMO SENS

Select CALMODE using keys:AUTO: Calibration with Calimatic buffer set recognitionMAN: Manual entry of buffer solutionsDAT: Direct entry of adjustment dataPress enter to confirm.

AUTOMANDAT

(AUTO: Buffer set) Select buffer set using keys (see buffer tables for nominal values).

Press enter to confirm.

-00-...-09-, -U1-

Pressing the info key dis-plays the manufacturer and nominal values in the lower line.

4

Page 46: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

46

Configuration

1

Sensor BSelecting cal timer, cal cycle

1 Press any arrow key.2 Select CONF using keys, press enter.3 Select "Sensor B“ using ,

press enter, then select "MEAS_MODE = ON“. Confirm by pressing enter.

4 All items of this menu group are indicated by the “S_B:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

MEMO SENS

2

3

4

Select sensor B

Measuring channel on/off

Calibration mode

(AUTO: Buffer set)

Calibration timer

Calibration cycle

Cleaning cycles

Sterilization cycles

4enter

enter

5

enter

meas

Page 47: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

47

ConfigurationMenu item Action ChoicesCalibration timer Adjust CALTIMER using

keys:OFF: No timerADAPT: Maximum cal cycle (adjust in the next step)FIX: Fixed cal cycle (adjust in the next step)Press enter to confirm.

OFF/ADAPT/FIX

With ADAPT, the calibra-tion cycle is automatically reduced depending on the sensor load (high temperatures and pH values) and for digital sensors also depending on the sensor wear

Calibration cycle Only with FIX/ADAPT:Modify digit using keys, select next digit using keys.Press enter to confirm.

0...9999 h

Note for the calibration timer:When Sensocheck has been activated in the Configuration / Alarm menu, the expiration of the calibration interval is indicated by Sensoface:

Display Status

+

+

Over 80% of the calibration interval has already past.

The calibration interval has been exceeded.

The time remaining until the next due calibration can be seen in the diagnostics menu (see Diagnostics chapter).

4

Page 48: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

48

ConfigurationSensor BAdjusting CIP cleaning cycles, SIP sterilization cycles

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select "Sensor B“ using ,

press enter, then select "MEAS_MODE = ON“. Confirm by pressing enter.

4 All items of this menu group are indicated by the “S_B:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

MEMO SENS

2

3

4

Select sensor B

Measuring channel on/off

Calibration mode

(AUTO: Buffer set)

Calibration timer

Calibration cycle

Cleaning cycles

Sterilization cycles

4enter

enter

5

enter

meas

Page 49: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

49

ConfigurationMenu item Action ChoicesCIP / SIP

Cleaning cyclesOn / Off

Select ON or OFF using keys.Activates/deactivates log-ging in extended logbookPress enter to confirm.

ON/OFF

Sterilization cyclesOn / Off

Select ON or OFF using keys.Activates/deactivates log-ging in extended logbookPress enter to confirm.

ON/OFF

The cleaning and sterilization cycles are counted to measure the load on the sensor. Suitable for biochemical applications (process temp approx. 0...50 °C, CIP temperature > 55 °C, SIP temperature > 115 °C).

4

Page 50: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

50

Configuration

Temperature unitOffsetting channels A/B

Measuring ModeSpecifying the temperature unit and calculation method

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select MEAS_MODE menu using keys,

press enter. 4 All items of this menu group are indicated by

the “MES:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

MEMO SENS

2

3

4enter

enter

5

enter

4

meas

Page 51: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

51

ConfigurationMenu item Action ChoicesTemperature unit Select °C or °F using

keys.

Press enter to confirm.

°C / °F

Calculations Select CALCULATION=ON using arrow keys if you want to use the value calculat-ed from the process variables of both channels.Press enter to confirm.

OFF/ON

pH difference Use arrow keys to select calculation method.Press enter to confirm.

C1 Difference PHC2 Difference mV C3 Difference TMP

mV difference

Temp difference

4

Page 52: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

52

Configuration

Current rangeProcess variableCurrent startCurrent endTime averaging filterOutput current during error messageOutput current during HOLDOutput current for HOLD FIX

Current Output 1Output current range. Current start. Current end.

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select OUT1 menu using keys, press

enter. 4 All items of this menu group are indicated by

the “OT1:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

4enter

enter

5

enter

4

meas

Page 53: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

53

ConfigurationMenu item Action ChoicesCurrent range Select 4-20 mA or

0-20 mA range using keys.

Press enter to confirm.

4-20 mA / 0-20 mA

Process variable Select using keys:PH: pH valueORP: Redox potentialTMP: TemperaturePress enter to confirm.

PH A/PH B/mV A/mV B/TMP A/TMP B/CALC:C1: DIFF PH/C2: DIFF mVC3: DIFF °C (DIFF °F)

Current start Modify digit using keys, select next digit using keys.

Press enter to confirm.

–2...16 pH (PH) –1999...1999 mV (ORP) –20...300 °C / –4...572 °F (TMP)-19.99 ... 19.99 (dPH)-1999 ... 1999 mV (d mV)-199.9 ... 300.0 °C (d°C)-199.9 ... 500.0 °F (d°F)

Current end Enter value using keys.

Press enter to confirm.

–2...16 pH (PH) –1999...1999 mV (ORP) –20...300 °C / –4...572 °F (TMP)-19.99 ... 19.99 (dPH)-1999 ... 1999 mV (d mV)-199.9 ... 300.0 °C (d°C)-199.9 ... 500.0 °F (d°F)

Assignment of measured values: Current start and current end

Output current

[pH]

20 4

14

020 4

7

5

[mA]

Example 1: Range pH 0...14

[mA]

Output current

75

4

Example 2: Range pH 5...7 Advantage: higher resolution in range of interest[pH]

Page 54: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

54

ConfigurationCurrent Output 1Adjusting time interval of output filter

Current rangeProcess variableCurrent startCurrent endTime averaging filterOutput current during error messageOutput current during HOLDOutput current for HOLD FIX

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select OUT1 menu using keys, press

enter. 4 All items of this menu group are indicated by

the “OT1:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

4enter

enter

5

enter

4

meas

Page 55: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

55

ConfigurationMenu item Action ChoicesTime averaging filter Enter value using

keys.

Press enter to confirm.

0...120 SEC (0000 SEC)

Time averaging filterTo smoothen the current output, a low-pass filter with adjustable filter time constant can be switched on. When there is a jump at the input (100 %), the output level is at 63 % after the time interval has been reached. The time interval can be set from 0 to 120 sec. If the time interval is set to 0 sec, the current output directly follows the input.Please note: The filter only acts on the current output, not on the display, the limit values, or the controller!During HOLD the filter is not applied. This prevents a jump at the output.

Time interval 0...120 s

4

pH

0/4-20 mA

Time interval 0...120 s

Temp

Display / Controller

Page 56: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

56

ConfigurationCurrent Output 1Output current during Error and HOLD

Current rangeProcess variableCurrent startCurrent endTime averaging filterOutput current during error messageOutput current during HOLDOutput current for HOLD FIX

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select OUT1 menu using keys, press

enter. 4 All items of this menu group are indicated by

the “OT1:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

4enter

enter

5

enter

4

meas

Page 57: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

57

ConfigurationMenu item Action ChoicesOutput current during error message

Select ON or OFF using keys.Press enter to confirm.

ON/OFF

Output current during HOLD

LAST: During HOLD the last measured value is maintained at the output.FIX: During HOLD a value (to be entered) is main-tained at the output.Select using Press enter to confirm.

LAST/FIX

Output current for HOLD FIX

Only with FIX selected:Enter current which is to flow at the output during HOLDEnter value using keys.

Press enter to confirm.

00.00...22.00 mA(21.00 mA)

HOLD active

21

4HOLD active

Output current[mA]

Output signal for HOLDFIX setting = 21.0 mA

Output signal for HOLDLAST setting

Output signal during HOLD:

4

Page 58: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

58

ConfigurationCurrent Output 2Output current range. Current start. Current end.

Current rangeProcess variableCurrent startCurrent endTime averaging filterOutput current during error messageOutput current during HOLDOutput current for HOLD FIX

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select OUT2 menu using keys, press

enter. 4 All items of this menu group are indicated by

the “OT2:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

4enter

enter

5

enter

4

meas

Page 59: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

59

ConfigurationMenu item Action ChoicesCurrent range Select 4-20 mA or

0-20 mA range using keys.

Press enter to confirm.

4-20 mA / 0-20 mA

Process variable Select using keys:PH: pH valueORP: Redox potentialTMP: TemperaturePress enter to confirm.

PH A/PH B/mV A/mV B/TMP A/TMP B/CALC:C1: dPH/C2: dmVC3: d°C (d°F)

.

.

.

All the following adjustments are made as for current output 1 (see there)!

4

Page 60: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

60

ConfigurationAlarm SettingsDelay. Sensocheck.

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select ALARM menu using keys, press

enter. 4 All items of this menu group are indicated by

the “ALA:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

enter

enter

5

enter

4

measDelaySensocheckCONTROL inputCurrent input (I-INPUT)

4

Error messages can be signaled by a 22 mA output current (see Error Messages and Configuration of Output 1/Output 2). The alarm contact opens. The alarm delay time delays the color change of the display backlighting to red, the 22 mA signal (if configured), and the alarm contact switching.

Page 61: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

61

ConfigurationMenu item Action ChoicesDelay Enter value using

keys.Press enter to confirm.

0...600 SEC010 SEC)

Sensocheck Select Sensocheck (continuous monitoring of sensor).Select ON or OFF using keys.Press enter to confirm.

ON/OFF

CONTROL input Activate CONTROL input (switches alarm contact)Select ON or OFF using keys.Press enter to confirm.

ON/OFF

Current input Activate current input (switches alarm contact)Select ON or OFF using keys.Press enter to confirm.

ON/OFF

Function Signal active with: Hi LEVL / Lo LEVL

LEVEL Threshold value 0 ... 22.00 mA(12.00 mA)

HYSTERESIS Hysteresis 0 ... 10.00 mA(01.00 mA)

4

Page 62: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

62

ConfigurationLimit FunctionRelay 1

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select REL1/REL2 menu using keys,

press enter. 4 All items of this menu group are indicated by

the “RL1:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

enter

5

enter

4

meas

Use of relaysSelect process variableLimit 1 switching characteristics (function)Limit 1 contact type Limit 1 setpointLimit 1 hysteresisLimit 1 delay

enter

4

Page 63: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

63

ConfigurationMenu item Action ChoicesUse of relays Select in the text line

using keys:• Limit function (LIMITS)• Controller

(CONTROLLER)

Press enter to confirm.

LIMITS / CONTROLLER

Please note: Selecting CONTROLLER leads to Controller menu group CTR.

Select process variable

Select desired process variable using keys.

Press enter to confirm.

PH A/PH B/mV A/mV B/TMP A/TMP B/CALC:C1: dPH/C2: dmVC3: d°C (d°F)

Limit 1 function Select desired function using arrow keys.LoLevel: active if value falls below / HiLevel: active if value exceeds setpointPress enter to confirm.

Lo LEVL / Hi LEVL

Limit 1 contact response

N/O: normally open contactN/C: normally closed contactSelect using keys.Press enter to confirm.

N/O / N/C

Limit 1 setpoint Enter setpoint using keys.

Press enter to confirm.

–2.00...16.00 PH(00.00 PH) /-1999...1999 mV /-20...200 °C

4

Page 64: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

64

Configuration

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select REL1/REL2 menu using keys,

press enter. 4 All items of this menu group are indicated by

the “RL1:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

enter

5

enter

4

meas

Use of relaysSelect process variableLimit 1 switching characteristics (function)Limit 1 contact type Limit 1 setpointLimit 1 hysteresisLimit 1 delay

enter

4

Limit FunctionRelay 1

Page 65: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

65

ConfigurationMenu item Action ChoicesLimit 1 hysteresis Select hysteresis using

keys.

Press enter to confirm.

0...10.00 PH(00.50 PH)

Limit 1 delay The contact is activated with delay (deactivated without delay)Adjust delay using keys.

Press enter to confirm.

0...9999 SEC (0010 SEC)

Limit Lo

Hysteresis +Setpoint

Contact0

1

Signal

Limit Hi

Hysteresis -

Signal

Setpoint

Contact0

1

4

Page 66: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

66

Limit FunctionRelay 2

Configuration

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select REL1/REL2 menu using keys,

press enter. 4 All items of this menu group are indicated by

the “RL2:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

enter

5

enter

4

meas

Use of relaysSelect process variableLimit 2 switching characteristics (function)Limit 2 contact typeLimit 2 setpointLimit 2 hysteresisLimit 2 delay

enter

4

Page 67: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

67

ConfigurationMenu item Action ChoicesSelect process variable(CHANNEL)

Select desired process variable using keys.Press enter to confirm.

PH A/PH B/mV A/mV B/TMP A/TMP B/CALC:C1: dPH/C2: dmVC3: d°C (d°F)

Limit 2 function(FUNCTION)

Select desired function using arrow keys.Press enter to confirm.

Lo LEVL / Hi LEVL

Limit 2 contact type (CONTACT)

N/O: normally open contactN/C: normally closed contactSelect: Press enter to confirm.

N/O / N/C

Limit 2 setpoint (LEVEL)

Enter setpoint using keys.Press enter to confirm.

–2.00...16.00 PH(14.00 PH) /-1999...1999 mV /-20...200 °C

Limit 2 hysteresis(HYSTERESIS)

Select hysteresis using keys.Press enter to confirm.

0...10.00 PH(00.50 PH)

Limit 2 delay(DELAYTIME)

The contact is activated with delay (deactivated without delay)Adjust delay using keys.Press enter to confirm.

0...9999 SEC (0010 SEC)

Limit Lo

Hysteresis +Setpoint

Contact0

1

Signal

Limit Hi

Hysteresis -

Signal

Setpoint

Contact0

1

4

Page 68: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

68

Page 69: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

69

Controller Functions

Setpoint

Controller output Yp [%]

+100 %

-100 %

Deviation Xw [pH]

Neutral zone Yp=0

Relay 1

Relay 2

Typical Applications

P controller Application for integrating control systems (e.g. closed tank, batch processes).

PI controllerApplication for non-integrating control systems (e.g. drains).

PID controller The additional derivative action compensates for measurement peaks.

Controller Characteristic

Page 70: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

70

Controller Functions

Proportional action YP

Setpoint - Meas. valueConstant

YP = *

Controller output Y =

I action

with:YP Proportional actionTN Reset time [s]TD Rate time [s]KC Controller gain [%]Constant 5 (for pH) 500 mV (for ORP)

D action

dYPdt

1TR

P action

Controller equations

YP + YPdt + TD

KC

Neutral zoneTolerated deviation from desired value. With the setting “1 pH”, for example, a deviation of ± 0.5 pH from the desired value is tolerated.

100 %

50 %

Xw

Process variables: pH/ (mV),underneath: temp [K]

Cont

rolle

r ou

tput

Y

KC = 500 % KC = 200 %

KC = 100 %

KC = 50 %

1 / (100) 2 / (200) 3 / (300) 4 / (400)

Proportional action (Gradient KC [%])

Deviation 5 / (500)

[10] [20] [30] [40] [50]

Page 71: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

71

Controller FunctionsPulse Length / Pulse Frequency ControllerPulse length controller (PLC)The pulse length controller is used to operate a valve as an actuator. It switches the contact on for a time that depends on the controller output. The period is constant. A minimum ON time of 0.5 sec is maintained even if the controller output takes corresponding values.

Output signal (relay contact) of pulse length controller

ON time (Y = 80 %)

ON time (Y = 20 %)

Pulse length

1

0

1

0

Pulse frequency controller (PFC)The pulse frequency controller is used to operate a frequency- controlled actuator (metering pump). It varies the frequency with which the contacts are switched on. The maximum pulse frequency [pulses/min] can be defined. It depends on the actuator.The contact ON time is constant. It is automatically calculated from the user-defined maximum pulse frequency:

Output signal (relay contact) of pulse frequency controller

ON time

Pulse frequency (Y = 20%)

Pulse frequency (Y = 80%)

1

0

1

0

Page 72: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

72

ConfigurationController (For description, see Controller Functions)Process variable. Controller type. Setpoint.

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select REL1/REL2 menu using keys,

press enter. 4 All items of this menu group are indicated by

the “CTR:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

enter

5

enter

4

meas

Use of relaysSelect process variableController typePulse length Pulse frequencySetpointNeutral zone Controller: P actionController: I actionController: D actionBehavior during HOLD

enter

4

Page 73: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

73

ConfigurationMenu item Action ChoicesSelect process variable

Select desired process variable using keys.

Press enter to confirm.

PH A/PH B/mV A/mV B/TMP A/TMP B

Controller type Pulse length controller (PLC)or pulse frequency controller (PFC)Select using keys.Press enter to confirm.

PLC/PFC

Pulse length Only with PLC: Pulse lengthAdjust using keys.

Press enter to confirm.

0...0600 SEC0010 SEC)

Pulse frequency Only with PFC: Pulse frequencyAdjust using keys.

Press enter to confirm.

0...0180 P/M(0060 P/M)

(pulses per minute)

Setpoint Adjust setpoint using keys.

Press enter to confirm.

–2.00...16.00 PH (07.00 PH) /-1500...1500 mV

4

Page 74: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

74

Controller (For description, see Controller Functions)Neutral zone. P, I, D actions. Behavior during HOLD

1 1 Press any arrow key.2 Select CONF using keys, press enter.3 Select REL1/REL2 menu using keys,

press enter. 4 All items of this menu group are indicated by

the “CTR:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

enter

5

enter

4

meas

Use of relaysSelect process variableController typePulse length Pulse frequencySetpointNeutral zone Controller: P actionController: I actionController: D actionBehavior during HOLD

enter

4

Configuration

Page 75: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

75

ConfigurationMenu item Action ChoicesNeutral zone Adjust neutral zone using

keys.

Press enter to confirm.

0...10.00 PH (01.00 PH) /0...2000 mV

Controller: P action Adjust P action using keys.

Press enter to confirm.

10...9999% (0100%)

Controller: I action Adjust I action using keys.

Press enter to confirm.

0...9999 SEC (0000 SEC)

Controller: D action Adjust D action using keys.

Press enter to confirm.

0...9999 SEC (0000 SEC)

Behavior during HOLD

Select response using keys.

Press enter to confirm.

Y LAST / Y OFF

4

Page 76: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

76

ConfigurationWASH ContactControlling a rinsing probe or signaling the parameter set

1 1 Press any arrow key.2 Select CONF using keys, press enter. 3 Select WASH menu using keys, press

enter. 4 All items of this menu group are indicated by

the “WSH:” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

4

enter

meas

5

enter

FunctionCleaning intervalCleaning durationContact type

enter

4

Page 77: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

77

ConfigurationMenu item Action ChoicesFunction Select WASH contact

function using keys.

Press enter to confirm.

WASH / PARSET A/B

WASH: Controlling a rinsing probe

With PARSET A/B selected, the contact signals:“Parameter set A“ (open contact)“Parameter set B“ (closed contact)

Cleaning interval Only with WASH:Adjust value using keys.

Press enter to confirm.

0.0...999.9 h (000.0 h)

Cleaning duration Only with WASH:Adjust value using keys.

Press enter to confirm.

0...9999 SEC (0060 SEC)

Contact type Only with WASH:N/O: normally open contactN/C: normally closed contactSelect using keys.Press enter to confirm.

N/O / N/C

4

Cycle interval

Cleaning Relax time (approx. 20 s)

HOLD

Page 78: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

78

Configuration

1 Press any arrow key.2 Select CONF using keys, press enter.3 Select CLOCK or TAG using keys, press

enter. 4 All items of this menu group are indicated by

the “CLK:” or “TAG” code. Press enter to select menu, edit using arrow keys (see next page). Confirm (and proceed) using enter.

5 End: Press meas key until the [meas] mode indicator is displayed.

2

3

4

enter

enter

Time formatTimeDay and monthYear

Tag number

enter

4

5

meas

Time and DateTag Number

1

Page 79: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

79

Time and DateIn measuring mode the time is shown in the lower display. In addition, the logbook entries (cf Diagnostics) are provided with a time stamp.

Please note:• After prolonged power outage (> 5 days) the time display is

replaced by dashes and cannot be used for processing. Enter the correct time.

• There is no automatic switchover from winter to summer time! Be sure to manually adjust the time!

Configuration

Menu item Action ChoicesTag number Select character using

keys,select next digit using keys.

Press enter to confirm.

A...Z, 0...9, – + < > ? / @

The first 10 characters are seen in the display with-out scrolling.

Tag Number (“TAG“)You can enter a designation for the point of measurement (tag number) in the lower display line. Up to 32 digits are possible.Pressing meas (repeatedly) in the measuring mode indicates the tag number.Being part of the device configuration, the “TAG“ can be read out via IrDA. A standardized tag number helps, for example, to correctly re-install a device after repair.

4

Page 80: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

80

CalibrationPlease note:• All calibration procedures must be performed by trained personnel.

Incorrectly set parameters may go unnoticed, but change the measuring properties.

• The response time of the sensor and temperature probe is considerably reduced when the sensor is first moved about in the buffer solution and then held still.

• The device can only operate properly when the buffer solutions used correspond to the configured set. Other buffer solutions, even those with the same nominal values, may demonstrate a different temperature response. This leads to measurement errors.

When using an ISFET Memosens sensor, the nominal zero point must be adjusted each time a new sensor is connected. This is important if you want to obtain reliable Sensoface messages. The Sensoface messages issued during all further calibrations are based on this basic calibration.

Page 81: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

81

Calibration is used to adapt the device to the individual sensor characteristics, namely asymmetry potential and slope. Access to calibration can be protected with a passcode (SERVICE menu). First, open the calibration menu and select the sensor (SENSOR_A / SENSOR_B) and then the calibration mode:

SENSOR_A Calibrating sensor ACAL_PH Adjusting the pHP_CAL Product calibration

(calibration with sampling)ISFET-ZERO ISFET operating point

(displayed only when ISFET Memosens sensor is connected)

Selecting a Calibration Mode

To preset CAL_PH (CONF menu / configuration):1) Hold meas key depressed (> 2 s) (measuring mode)2) Press any arrow key: the selection menu appears3) Select CONF mode using left / right arrow key4) Select “SENSOR“ – “CALMODE“: AUTO, MAN, or DAT.

Press enter to confirm.

Selection menu

Selected mode (blinks)

1 2

3

4

Page 82: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

82

Zero Adjustment (ISFET)This adjustment allows the use of ISFET Memosens sensors. The func-tion is available when an ISFET Memosens sensor is connected. The adjustment is made using a zero buffer (pH 7.00). Permitted range for buffer value: pH 6.5 ... 7.5 Temperature-corrected input. Maximum zero offset: ±750 mV

Display Action RemarkSelect Calibration.Press enter to proceed.

Ready for calibration. Hourglass blinks.

Display (3 sec)Now the device is in HOLD mode.

Immerse sensor in a pH 7.00 buffer. Enter the temperature-corrected pH value in the range 6.50 to 7.50 using the arrow keys (see buffer table).Press enter to confirm.

If the zero offset of the sensor is too large (> ± 750 mV), a CAL ERR error mes-sage is generated. In that case the sensor cannot be calibrated. At 500 mV Sensoface becomes sad.

Stability check.The measured value [mV] is displayed.The “hourglass” icon is blinking.

Please note: Stability check can be stopped (by pressing enter). However, this re-duces calibration accuracy.

Zero adjustment

Page 83: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

83

Display Action RemarkAt the end of the ad-justment procedure the zero offset [mV] of the sensor is displayed (based on 25 °C).Sensoface is active.Press enter to proceed.

This is not the final calibration value of the sensor! Asymmetry poten-tial and slope must be determined with a complete 2-point calibration.

Use the arrow keys to select:• Repeat

(repeat calibration) or • Measuring. Press enter to confirm.Place sensor in process.End zero calibration with enter.

After end of calibra-tion, the outputs re-main in HOLD mode for a short time.

Note for Zero AdjustmentAfter having adjusted the zero offset, be sure to calibrate the sensor following one of the procedures as described on the next pages.

Zero Adjustment (ISFET)

Page 84: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

84

Adjustment (Calimatic)

Automatic Calibration (Calimatic)The AUTO calibration mode must have been preset during configura-tion. Make sure that the buffer solutions used correspond to the con-figured buffer set. Other buffer solutions, even those with the same nominal values, may demonstrate a different temperature response. This leads to measurement errors.

Display Action RemarkSelect Calibration.Press enter to proceed.

Ready for calibration. Hourglass blinks. Select calibration method: CAL_PH Press enter to proceed.

Display (3 sec)Now the device is in HOLD mode.

Remove the sensor, clean it, and immerse it in the first buffer solu-tion (it does not matter which solution is taken first). Press enter to start.Buffer recognition.While the “hourglass” icon is blinking, the sensor remains in the first buffer solution.

To reduce the sensor response time, first move it about in the buffer solution and then hold it still.

Buffer recognition terminated, the nomi-nal buffer value is displayed.

Page 85: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

85

Display Action RemarkAt the end of the stabil-ity check, the value will be saved and the asym-metry potential will be displayed.Calibration with the first buffer is terminated. Remove the sensor from the first buffer solution and rinse it thoroughly.Use the arrow keys to select:• END (1-point cal)• CAL2 (2-point cal)• REPEATPress enter to proceed.

Please note:Stability check can be stopped after 10 sec (by pressing enter). However, this reduces calibration accuracy. Display for 1-point cal:

Sensoface is active. End by pressing enter

2-point calibration:Immerse the sensor in the second buffer solution. Press enter to start.

The calibration process runs as for the first buffer.

Retract sensor out of second buffer, rinse off, re-install.Press enter to proceed.

The slope and asym-metry potential of the sensor (based on 25 °C) are displayed.

Use the arrow keys to select:• MEAS (end)• REPEATPress enter to proceed.End: HOLD is de-activated with delay.

When 2-point cal is ended:

Automatic Calibration (Calimatic)

Page 86: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

86

Manual Calibration with Buffer EntryThe MAN calibration mode must have been preset during configura-tion. For calibration with manual buffer specification, you must enter the pH value of the buffer solution used in the device for the proper temperature. Any desired buffer solution can be used for calibration.

Display Action RemarkSelect Calibration.Press enter to proceed.

Ready for calibration.Hourglass blinks.

Display (3 sec)Now the device is in HOLD mode.

Remove the sensor, clean it, and immerse it in the first buffer solution. Press enter to start.Enter the pH value of your buffer solution for the proper temperature. While the “hourglass” icon is blinking, the sensor remains in the first buffer solution.

To reduce the sensor response time, first move it about in the buffer solution and then hold it still.

Page 87: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

87

Display Action RemarkAt the end of the sta-bility check, the value will be saved and the asymmetry potential will be displayed.Calibration with the first buffer is terminated. Remove the sensor from the first buffer solution and rinse it thoroughly.Use the arrow keys to select:• END (1-point cal)• CAL2 (2-point cal)• REPEATPress enter to proceed.

Please note:Stability check can be stopped after 10 sec (by pressing enter). However, this reduces calibration accuracy. Display for 1-point cal:

Sensoface is active. End by pressing enter

2-point calibration:Immerse sensor in second buffer solution.Enter pH value. Press enter to start.

The calibration process runs as for the first buffer.

Rinse off and re-install the sensor.Press enter to proceed.

Display of slope and new asymmetry potential (based on 25 °C).

Use the arrow keys to select:• MEAS (end)• REPEATPress enter to proceed.End: HOLD is deacti-vated with delay.

When 2-point cal is ended:

Manual Calibration with Buffer Entry

Page 88: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

88

Data Entry of Premeasured SensorsThe DAT calibration mode must have been preset during configuration. You can directly enter the values for slope and asymmetry potential of a sensor. The values must be known.

Display Action RemarkSelect Calibration.Press enter to proceed.

“Data Input“Ready for calibration. Hourglass blinks.

Display (3 sec)Now the device is in HOLD mode.

Enter asymmetry potential [mV].Press enter to proceed.

Enter slope [%].

The device displays the new slope and asymmetry potential (at 25 °C).Sensoface is active.Use the arrow keys to select:• MEAS (end)• REPEATPress enter to proceed.

End: HOLD is deactivated with delay.

Page 89: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

89

Converting Slope to mV

Converting slope [%] to slope [mV/pH] at 25 °C

% mV/pH78 46.280 47.482 48.584 49.786 50.988 52.190 53.392 54.594 55.696 56.898 58.0100 59.2102 60.4

Converting asymmetry potential to sensor zero point

= Sensor zero

= Asymmetry potential

= Slope

ZERO = 7 - ZERO

VAS

S

VAS

[mV]

S [mV / pH]

Page 90: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

90

Product calibration

Product Calibration (pH)Calibration by sampling (one-point calibration). During product calibration the sensor remains in the process. The measurement process is only interrupted briefly.Procedure: 1) The sample is measured in the lab or directly on the site using a

portable meter. To ensure an exact calibration, the sample tempera-ture must correspond to the measured process temperature. During sampling the device saves the currently measured value and then returns to measuring mode. The “calibration“ mode indicator blinks.

2) In the second step you enter the measured sample value in the device. From the difference between the stored measured value and entered sample value, the device calculates the new asymmetry potential.

If the sample is invalid, you can take over the value stored during sampling. In that case the old calibration values are stored. Afterwards, you can start a new product calibration.

Display Action RemarkSelect product calibration: P_CAL.Press enter to proceed.

Ready for calibration. Hourglass blinks.

Display (3 sec)Now the device is in HOLD mode.

Take sample and save value. Press enter to proceed.

Now the sample can be measured.

Page 91: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

91

Display Action RemarkThe device returns to measuring mode.

From the blinking CAL mode indicator you see that product calibration has not been terminated.

Product calibration step 2 

Display (3 sec)Now the device is in HOLD mode.

The stored value is displayed (blinking) and can be overwritten with the measured sample value.Press enter to proceed.Display of new asymmetry potential (based on 25°C).Sensoface is active.To end calibration:Select MEAS, then enter

Repeat calibration: Select REPEAT, then enter

Calibration is terminated.

After end of calibration, the outputs remain in HOLD mode for a short time.

Product Calibration (pH)

Page 92: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

92

Page 93: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

93

MeasurementDisplay Remark

From the configuration or calibration menus, you can switch the device to measuring mode by pressing the meas key.In the measuring mode the main display shows the configured process variable, the secondary display shows the time and the second configured process variable. The [meas] mode indicator lights.Please note:• After prolonged power outage (> 5 days)

the time display is replaced by dashes and cannot be used for processing. Enter the correct time.

Pressing the enter key briefly shows the output currents. By pressing the meas key you can step through the following displays. When no key has been pressed for 60 sec, the device returns to the standard display.

or AM/PM and °F:

Further displays (each by pressing meas).

1) pH and temperature channel A2) pH and temperature channel B

(without figure)3) Display of time and date

(without figure)4) Display of tag number (“TAG“)

Page 94: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

94

MeasurementDisplay RemarkWith activated controller you can also step through the following displays by pressing the meas key. When no key has been pressed for 60 sec, the device returns to the standard display.

Main display: Controller output YSecondary display: SetpointDepending on configuration setting:pH, mV, or temperature.

Direct adjustment of setpointThe first digit of the setpoint (lower display) blinks, you can enter the value using the cursor keys. Press enter to save the new value for configuration. 60 sec after the last keystroke, the display switches back to measure-ment.This direct entry is only possible when the configuration passcode has been switched off (0000). If configuration is password-protected, the setpoint is only displayed but cannot be edited.

Page 95: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

95

DiagnosticsIn the Diagnostics mode you can access the following menus without interrupting the measurement:

Viewing sensor A data Viewing sensor B data Starting a device self-test Viewing the logbook entries Displaying currently measured values

Displaying device type, software version, serial number

Access to diagnostics can be protected with a passcode (SERVICE menu).

Please note:HOLD is not active during Diagnostics mode!

Action Key RemarkActivate diagnostics

Press any arrow key to call the selection menu.(Display color changes to turquoise.)Select DIAG using keys, confirm by pressing enter.

Select diagnostics option

Use keys to select from:SENSOR A SENSOR B SELFTEST LOGBOOK MONITOR VERSIONSee next pages for further proceeding.

End meas End by pressing meas.

Diagnostics (DIAG)

Page 96: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

96

DiagnosticsDisplay Menu item

Displaying the calibration dataSelect SENSOR A or SENSOR B, then select CALDATA usingkeys, press enter to confirm.Use thekeys to select the desired parameter from the bottom line of the display (LAST_CAL ISFET-ZERO ZERO SLOPE NEXT_CAL).The selected parameter is shown in the main display.

Press meas to return to measurement.

Displaying the sensor dataManufacturer, type, serial number and last calibration date.In each case Sensoface is active.

Display data using keys,Press enter or meas to return.

Page 97: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

97

DiagnosticsDisplay Menu item

Device self-test(To abort, you can press meas.)1) Display test: Display of all segments with

changing background colors white/green/red. Proceed by pressing enter

2) RAM test: Hourglass blinks, then display of --PASS-- or --FAIL-- Proceed by pressing enter

3) EEPROM test: Hourglass blinks, then display of --PASS-- or --FAIL-- Proceed by pressing enter

4) FLASH test: Hourglass blinks, then display of --PASS-- or --FAIL-- Proceed by pressing enter

5) Module test: Hourglass blinks, then display of --PASS-- or --FAIL-- Press enter or meas to return to measuring mode.

Page 98: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

98

DiagnosticsDisplay Menu item

Displaying the logbook entriesSelect LOGBOOK using, press enter to confirm.

Using the keys, you can scroll backwards and forwards through the logbook (entries -00-...-99-), -00- being the last entry.

If the display is set to date/time, you can search for a particular date using the keys.Pressto view the corresponding message text.

If the display is set to the message text, you can search for a particular message using the keys.Pressto display the date and time.

Press meas to return to measurement.

Extended logbook / Audit Trail (via TAN)Using the keys, you can scroll backwards and forwards through the extended logbook (entries -000-...-199-), -000- being the last entry.Display: CFRAudit Trail also records function activations (CAL CONFIG SERVICE), some Sensoface messages (cal timer, wear), and opening of the enclosure.

Page 99: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

99

Diagnostics

Display Menu item

Display example:

Displaying the currently measured values (sensor monitor)Select MONITOR using, press enter to confirm. Use thekeys to select the desired parameter from the bottom line of the display: mV_PH(A) mV_PH(B) R_GLASS(A) R_GLASS(B) OPERATION TIME(A) OPERATION TIME(B) SENSOR WEAR(A) SENSOR WEAR(B) LIFETIME(A) LIFETIME(B) I-INPUT. The selected parameter is shown in the main display.Press meas to return to measurement.

Display of mV_pH(for validation, sensor can be immersed in a calibra-tion solution, for example, or the device is checked by using a simulator)

Display of remaining dynamic lifetime

Display of sensor operating time

Display of sensor wear

VersionDisplay of device type, software/hardware version, and serial number for all device components.Use the keys to switch between software and hardware version. Press enter to proceed to next device component.

Page 100: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

100

ServiceIn the Service mode you can access the following menus:MONITOR Displaying currently measured values OUT1 Testing current output 1OUT2 Testing current output 2RELAY Testing the function of the 4 relaysCONTROL Testing the controller functionIRDA Activating and communicating via the IrDA interfaceCODES Assigning and editing passcodesDEFAULT Resetting the device to factory settingsOPTION Enabling options via TAN.

Please note:HOLD is active during Service mode!

Action Key/Display RemarkActivate Service

Press any arrow key to call the selection menu.Select SERVICE using keys, confirm with enter

Passcode Enter passcode “5555“ for service mode using the keys.

Press enter to confirm.

Display In service mode the following icons are displayed:• HOLD triangle• Service (wrench)

End meas End with meas.

Service (SERVICE) A

Page 101: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

101

ServiceMenu item Remark

Displaying the currently measured values (sensor monitor) with HOLD mode activated:Select MONITOR using, press enter to confirm.Select variable in the bottom text line using.

The selected parameter is shown in the main display.As the device is in HOLD mode, you can perform validations using simulators without influencing the signal outputs.

Press meas to return to the service menu.Return to measurement: Press meas once more.

Specifying the current at outputs 1 and 2:Select OUT1 or OUT2 using thekeys, press enter to confirm.Enter a valid current value for the respective output using keys.Press enter to confirm.For checking purposes, the actual output current is shown in the bottom right corner of the display.End by pressing enter or meas.

Relay test (manual test of contacts):Select RELAIS using, press enter to confirm.Now the status of the 4 relays is “frozen“. The 4 digits of the main display represent the respective states (from left to right: REL1, REL2, ALARM, WASH). Select one of the 4 relays using the keys, close (1) or open (0) using the keys. The selected position blinks. The lower display line shows the relay function, e.g. “WASH” or “ALARM”.End by pressing enter. The relays will be re-set corresponding to the measured value.

Press meas to return to measurement.REL1

REL2

ALA

RMW

ASH

Page 102: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

102

ServiceMenu item Remark

Controller characteristic

The arrows indicate which relay (valve) is active:

Relay 2 active (Meas. value > setpoint)

Relay 2 active (Meas. value < setpoint)

Controller test (manual specification of controller output):This function is used to start up control loops or check the actuators.For bumpless changeover to automatic operation (exiting this function), configure an I-action component (reset time).

The lower display shows the currently adjusted controller output Yp.

Specify new value for controller output Yp:Enter sign and value in the main display using the keys. Press enter to confirm.

The new value is taken into the lower display.

Press enter or meas to return to the service menu.Return to measurement: Press meas once more.

Controller output –100...0% Relay 2 active

Controller output 0...+100% Relay 1 active

IrDA communication:Select IRDA using, press enter to confirm.

When IrDA communication is active, the device remains in the HOLD mode for reasons of safety. Further operation is performed via IrDA.

End communication by pressing meas.

Exception: Firmware update (must not be interrupted!)

Setpoint

Controller output Yp [%]

+100 %

-100 %

Deviation Xw [mS/cm]

Neutral zone Yp=0

Relay 1

Relay 2

Page 103: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

103

ServiceMenu item Remark

Assigning passcodes:In the “SERVICE - CODES“ menu you can assign pass-codes to DIAG, HOLD, CAL, CONF, and SERVICE modes (Service preset to 5555).When you have lost the Service passcode, you have to request an “Ambulance TAN“ from the manufac-turer specifying the serial number of your device. To enter the “Ambulance TAN“, call the Service func-tion and enter passcode 7321. After correct input of the ambulance TAN the device signals “PASS“ for 4 sec and resets the Service passcode to 5555.

Reset to factory settings:In the “SERVICE - DEFAULT“ menu you can reset the device to factory settings.Caution!After a reset to factory setting the device must be reconfigured completely, including the sensor parameters!

Option request:Communicate the serial number and hardware/ software version of your device to the manufacturer. These data can be viewed in the Diagnostics/Version menu.The “transaction number“ (TAN) you will then receive is only valid for the device with the corresponding serial number.Release of options:Options come with a “transaction number“ (TAN). This TAN must be entered and confirmed using enter to release the option.

Page 104: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

104

Operating States

Operating status

OU

T 1

OU

T 2

REL1

/2

(Lim

it)

REL1

/2

(Con

trol

)

ALA

RM

cont

act

WA

SH

cont

act

Tim

e ou

t

Measuring -

DIAG 60 s

CAL_PH No

P_CAL No

CONF 20 min

SERVICE 20 min

SERVICE OUT 1 20 min

SERVICE OUT 2 20 min

SERVICE RELAIS 20 min

SERVICE CONTROL

20 min

Cleaning function

No

HOLD No

Explanation: as configured (Last/Fix or Last/Off )

active

manual

Betriebszustände A

Page 105: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

105

Product Line and AccessoriesOrder Code Stratos Pro A 4...

Channel 1 Channel 2 TANExample A 4 1 1 N - PH / 0

4-wire / 20...254 V AC/DC A 4 B,C,E

CommunicationWithout (HART retrofittable via TAN) 0 AHART 1

Version numberVersion 1

ApprovalsGeneral Safety NATEX / IECEX / FM / CSA Zone 2 / Cl 1 Div 2 B

Meas. channel 1 / Meas. channel 2Memosens pH (ORP) Digital MSPH 0Memosens pH (ORP) / pH (ORP) Digital MSPH MSPHMemosens pH (ORP) / Oxy Digital MSPH MSOXYMemosens COND Digital MSCOND 0Memosens COND / COND Digital MSCOND MSCONDMemosens Oxy Digital MSOxy 0Dual COND (2x2-electrode, analog) Module CC 0pH / ORP value (ISM digital: TAN) Module PH 0 FCond, 2-/4-electrode Module COND 0Conductivity, electrodeless Module CONDI 0Oxygen (ISM digital/Traces: TAN) Module OXY 0 D, FCarbon dioxide (ISM digital: TAN) Module CO2 0 F

TAN optionsHART SW-A001 (A)Logbook SW-A002 (B)Extended logbook (Audit Trail) SW-A003 (C)Trace oxygen measurement SW-A004 (D)Current input + 2 digital inputs SW-A005 (E)ISM digital SW-A006 (F)

Mounting accessoriesPipe-mount kit ZU 0274Protective hood ZU 0737Panel-mount kit ZU 0738

Page 106: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

106

SpecificationspH input Input for Memosens sensors (RS485)

Display range pH value –2.00 ... 16.00

pH sensor standardization* pH calibration

Operating modes BUF Calibration with Calimatic automatic buffer recognition

MAN Manual calibration with entry of individual buffer values

DAT Data entry of pre-measured electrodes

Product calibration

Calimatic buffer sets * -00- Knick 2.00/4.01/7.00/9.21

-01- Mettler-Toledo 2.00/4.01/7.00/9.21

-02- Merck/Riedel 2.00/4.00/7.00/9.00/12.00

-03- Ciba (94) 2.06/4.00/7.00/10.00

-04- NIST technical 1.68/4.00/7.00/10.01/12.46

-05- NIST standard 1.679/4.006/6.865/9.180

-06- HACH 4.00/7.00/10.01

-07- WTW techn. buffers 2.00/4.01/7.00/10.00

-08- Hamilton 4.01/7.00/10.01

-09- Reagecon 2.00/4.00/7.00/9.00/12.00

-U1- Specifiable buffer set with 2 buffer solutions

Zero adjustment ± 750 mV (for ISFET)

Max. calibration range Asymmetry potential ±60 mV

Slope 80 ... 103 % (47.5 ... 61 mV/pH)

(possibly restricting notes from Sensoface)

Adaptive cal timer v

(Pat. DE 101 41 408)Interval 0000 ... 9999 h

Calculations (CALC)

-C1- Difference pH A – pH B [dpH]

-C2- Difference mV A – mV B [dmV]

-C3- Difference Temp A – Temp B [d°C / d°F]

Sensocheck Automatic monitoring of glass electrode (can be disabled)

Delay Approx. 30 s

Sensoface Provides information on the sensor condition, evaluation of zero/slope, response time, calibration interval, wear, Sensocheck

Page 107: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

107

Specifications

I input Current input 0/4 ... 20 mA / 50 Ω for flow monitoring

Setpoint As desired within range

Characteristic Linear

Measurement error 1,3) < 1% current value + 0.1 mA

HOLD input Galvanically separated (OPTO coupler)

Function Switches device to HOLD mode

Switching voltage 0 ... 2 V (AC/DC) HOLD inactive

10 ... 30 V (AC/DC) HOLD active

CONTROL input Galvanically separated (OPTO coupler)

Function Input for external monitoring equipment, e.g. flow, can be disabled

Message via 22mA signal and alarm contact

Switching voltage 0 ... 2 V (AC/DC) Flow too low

10 ... 30 V (AC/DC) Flow okay

Output 1 0/4 ... 20 mA, max. 10 V, floating (galv. connected to output 2)

Process variable * pH, mV, temperature, or difference (Calc)

Characteristic Linear

Overrange * 22 mA in the case of error messages

Output filter * PT1 filter, time constant 0 ... 120 s

Measurement error 1,2,3) < 0.25 % current value + 0.025 mA

Start/end of scale* Configurable within the measuring ranges for pH, mV, °C, °F

Output 2 0/4 ... 20 mA, max. 10 V, floating (galv. connected to output 1)

Process variable * pH, mV, temperature, or difference (Calc)

Characteristic Linear

Overrange * 22 mA in the case of error messages

Output filter * PT1 filter, time constant 0 ... 120 s

Measurement error 1,2,3) < 0.25 % current value + 0.025 mA

Start/end of scale * Configurable within the measuring ranges for pH, mV, °C, °F

Page 108: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

108

Specifications

Alarm contact Relay contact, floating

Contact ratings AC < 250 V / < 3 A / < 750 VA

DC < 30 V / < 3 A / < 90 W

Contact response N/C (fail-safe type)

Response delay * 0000 ... 0600 s

Wash contact Relay contact, floating

For controlling a cleaning system

Contact ratings AC < 250 V / < 3 A / < 750 VA

DC < 30 V / < 3 A / < 90 W

Contact response * N/C or N/O

Interval * 000.0 ... 999.9 h (000.0 h = cleaning function switched off)

Cleaning duration* 0000 ... 1999 s

Limit values REL1/REL2 REL1/REL2 contacts, floating, but inter-connected

Contact ratings AC < 250 V / < 3 A / < 750 VA

DC < 30 V / < 3 A / < 90 W

Contact response * N/C or N/O

Response delay * 0000 ... 9999 s

Setpoints* As desired within range

Hysteresis * User-defined

PID process controller Output via REL1/REL2 relay contacts (see limit values)

Setpoint specification* Within selected range

Neutral zone * pH 0 ... 5 / 0 ... 500 mV / 0 ... 50 K

Proportional action* Controller gain Kp: 0010 ... 9999 %

Integral action* Reset time Tr: 0000 ... 9999 s (0000 s = no integral action)

Derivative action* Rate time Td: 0000 ... 9999 s (0000 s = no derivative action)

Controller type * Pulse length controller or pulse frequency controller

Page 109: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

109

Specifications

Pulse period * 0001 ... 0600 s, min. ON time 0.5 s (pulse length controller)

Max. pulse frequency* 0001 ... 0180 min-1 (pulse frequency controller)

Real-time clock Different time and date formats selectable

Power reserve > 5 days

Display LC display, 7-segment with icons

Main display Character height approx. 22 mm, unit symbols approx. 14 mm

Secondary display Character height approx. 10 mm

Text line 14 characters, 14 segments

Sensoface 3 status indicators (friendly, neutral, sad face)

Mode Indicators meas, cal, conf, diag

Further icons for configuration and messages

Alarm indication Red backlighting in case of alarm

Keypad Keys: meas, info, 4 cursor keys, enter

HART communication HART version 6 Digital communication by FSK modulation of output current 1

Device identification, measured values, status and messages, parameter setting, calibration, records

Conditions Output current ≥ 3.8 mA and load resistance ≥ 250 Ω

IrDA interface Infrared interface for transmission of records and logbook, parameter setting, calibration, firmware update

FDA 21 CFR Part 11 Access control by editable passcodesLogbook entry and flag via HART in the case of configuration changesMessage and logbook entry when enclosure is opened

Diagnostics functions

Calibration data Calibration date, zero, slope, response time

Device self-test Displaytest, automatic memory test (RAM, FLASH, EEPROM), module test

Logbook 100 events with date and time

Extended logbook (TAN) AuditTrail: 200 events with date and time

Page 110: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

110

Specifications

Service functions

Sensor monitor Display of direct sensor signals (mV/temperature ...)

Current source Current specifiable for output 1 and 2 (00.00 ... 22.00 mA)

Relay test Manual control of the four switching contacts

Manual controller Controller output entered directly (start of control process)

IrDA Activating the IrDA function

Passcodes Assigning passcodes for menu access

Factory setting Resetting all parameters to factory settingException: Calibration data

TAN Enabling optionally available additional functions

Data retention Parameters, calibration data, logbook > 10 years (EEPROM)

EMC EN 61326-1 (General Requirements)

Emitted interference Class B (residential area)

Immunity to interference Industry EN 61326-2-3 (Particular Requirements for Transmitters)

Explosion protectionStratos Pro A4...B MSPHMSPH

USA: FM Class I Div 2 / Zone 2 (pending)

Canada: cCSAus Cl I Div 2 / Zone 2

Power supply 24 (–15%) ... 230 (+10%) V AC/DC4);

approx. 12 VA, 4 W AC: 45 ... 65 Hz

Overvoltage category II, protection class II

Nominal operating conditions

Ambient temperature –20 ... +55 °C

Transport/Storage temperature

–20 ... +70 °C

Relative humidity 10 ... 95% not condensing

Power supply 24 (–15%) ... 230 (+10%) V AC/DC (DC ≤ 80V)

Frequency for AC 45 ... 65 Hz

Page 111: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

111

Specifications

Enclosure Molded enclosure made of PBT/PC, glass reinforced

Mounting Wall, pipe/post, or panel mounting

Color Gray, RAL 7001

Ingress protection IP 67

Flammability UL 94 V-0

Dimensions 148 mm x 148 mm

Control panel cutout 138 mm x 138 mm to DIN 43 700

Weight 1.2 kg (1.6 kg incl. accessories and packaging)

Cable glands 3 knockouts for M20 x 1.5 cable glands

2 knockouts for NPT ½” or rigid metallic conduit

Connections Terminals, conductor cross section max. 2.5 mm2

* User-defined

1) Acc. to EN 60746, at nominal operating conditions

2) ± 1 count

3) Plus sensor error

4) DC ≤ 80 V

5) At room temperature

Page 112: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

112

Specifications

Page 113: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

113

Error HandlingAlarm condition:• The display backlighting turns red• The alarm icon is displayed• The complete measured-value display blinks• “ERR xxx“ is displayed in the lower menu linePress the [info] key to view a short error text:• The error text appears in the lower menu line• The main display reads “InFo“.

Parameter errors:Configuration data such as current range, limit values, etc are checked during the input.If they are out of range,• “ERR xxx“ is displayed for 3 sec,• the display backlighting flashes red,• the respective maximum or minimum value is shown,• input must be repeated.

If a faulty parameter arrives through the interface (IrDA, HART),• an error message will be displayed: “ERR 100...199“• the faulty parameter can be localized by pressing the [info] key

Calibration errors:If errors occur during calibration,• an error message will be displayed• calibration will be restarted

Sensoface:If the Sensoface becomes sad,• the display backlighting will turn purple• the cause can be seen by pressing the info key• the calibration data can be seen in the Diagnostics menu

Page 114: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

114

Error codes

Error Messages

ErrorInfo text(is displayed in case of fault when the Info key is pressed)

Problem Possible causes

ERR 99 DEVICE FAILURE Error in factory settingsEEPROM or RAM defectiveThis error message only occurs in the case of a total defect. The device must be repaired and recalibrated at the factory.

ERR 98 CONFIGURATION ERROR Error in configuration or calibration dataConfiguration or calibration data defectiveReset device to factory settings (SERVICE/DEFAULT), then calibrate

ERR 95 SYSTEM ERROR System errorRestart required. If error still persists, send in the device for repair.

ERR 01 NO SENSOR pH sensorSensor defectiveSensor not connectedBreak in sensor cableBoth channels switched off

ERR 02 WRONG SENSOR Wrong sensor type e.g. oxygen sensor

Page 115: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

115

ErrorInfo text(is displayed in case of fault when the Info key is pressed)

Problem Possible causes

ERR 11 PH RANGE pH display range violation< -2 or > 16

ERR 12 MV RANGE mV rangeERR 13 TEMPERATURE RANGE Temperature range

violationERR 15 SENSOCHECK GLASS-EL Sensocheck glassERR 59 INVALID CALCULATION Calculation errorERR 60 OUTPUT LOAD Load errorERR 61 OUTPUT 1 TOO LOW Output current 1

< 3.8 mAERR 62 OUTPUT 1 TOO HIGH Output current 1

< 20.5 mAERR 63 OUTPUT 2 TOO LOW Output current 2

< 3.8 mAERR 64 OUTPUT 2 TOO HIGH Output current 2

< 20.5 mAERR 69 TEMP. OUTSIDE TABLE Temperature

value outside tableERR 100 INVALID SPAN OUT1 Configuration error

Output range 1

ERR 101 INVALID SPAN OUT2 Configuration errorOutput range 2

Error Messages

Page 116: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

116

ErrorInfo text(is displayed in case of fault when the Info key is pressed)

Problem Possible causes

ERR 102 FAILURE BUFFERSET -U1- Configuration errorSpecifiable buffer set U1

ERR 104 INVALID PARAMETER CONTROLLER

Configuration errorController

ERR 105 INVALID SPAN I-INPUT Configuration errorCurrent input

ERR 106...255

VOID PARAMETER Invalid parameter

Error Messages

Page 117: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

117

Sensoface(Sensocheck must have been activated during configuration.)

The smiley in the display (Sensoface) alerts to sensor problems (defec-tive sensor, sensor wear, defective cable, maintenance request). The permitted calibration ranges and the conditions for a friendly, neutral, or sad Sensoface are summarized in the following table. Additional icons refer to the error cause.

SensocheckContinuously monitors the sensor and its wiring. Critical values make the Sensoface “sad” and the corresponding icon blinks:

The Sensocheck message is also output as error message Err 15 (glass electrode). The alarm contact is active, the display backlighting turns red, output current 1 is set to 22 mA (when configured correspond-ingly). Sensocheck can be switched off during configuration (then Sensoface is also disabled). Exception: After a calibration a smiley is always displayed for confirmation.

Please note:The worsening of a Sensoface criterion leads to the devaluation of the Sensoface indicator (Smiley becomes “sad”). An improvement of the Sensoface indicator can only take place after calibration or removal of the sensor defect.

Page 118: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

118

Sensoface

Display Problem StatusAsymmetry potential and slope

Asymmetry potential (zero) and slope of the sensor are still okay.The sensor should be replaced soon.

Asymmetry potential and slope of the sensor have reached values which no lon-ger ensure proper calibration. Replace sensor.

Calibration timer

Over 80% of the calibration interval has already past.

The calibration interval has been exceeded.

Sensor defect Check the sensor and its connections (see also Err 15, Error Messages).

Response time

Sensor response time has increased.The sensor should be replaced soon.To achieve an improvement, clean or wet the sensor.

Sensor response time significantly increased ( > 72 s, calibration aborted after 120 s)Replace sensor.

Page 119: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

119

Sensoface

Display Problem StatusSensor wear High temperatures and pH

values have caused a wear of over 80%. The sensor should be replaced soon.

Wear is at 100%. Replace sensor.

Page 120: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

120

FDA 21 CFR Part 11Conformity with FDA 21 CFR Part 11In their directive “Title 21 Code of Federal Regulations, 21 CFR Part 11, Electronic Records; Electronic Signatures“ the US American health agency FDA (Food and Drug Administration) regulates the production and processing of electronic documents for pharmaceutical devel-opment and production. This results in requirements for measuring devices used for corresponding applications. The following features ensure that the measuring devices of this Series meet the demands of FDA 21 CFR Part 11:

Electronic Signature – PasscodesAccess to the device functions is regulated and limited by individually adjustable codes – “Passcodes“ (see SERVICE). This prevents unauthor-ized modification of device settings or manipulation of the measure-ment results. Appropriate use of these passcodes makes them suitable as electronic signature.

Audit TrailEvery (manual) change of device settings can be automatically docu-mented. Each change is tagged with a “Configuration Change Flag“, which can be interrogated and documented using HART communica-tion. Altered device settings or parameters can also be retrieved and documented using HART communication.

Extended logbookAudit Trail also records function activations (CAL, CONFIG, SERVICE), some Sensoface messages (cal timer, wear), and opening of the enclosure.

Page 121: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

121

AAccess codes 120, 132Accessories 105Alarm 27Ambulance TAN 103Approvals for application in hazardous locations 110Assembly 10Asymmetry potential 89Audit Trail 98, 120Automatic calibration (Calimatic) 84

BBacklighting 21

CCalculation method 50Calibration 24, 80

Automatic calibration (Calimatic) 84Calibration errors 113Calibration timer 41, 47, 118Data entry of premeasured sensors 88ISFET sensors 80Manual calibration with buffer entry 86Product calibration (pH) 90Zero adjustment 83

Calibration cycle 40, 46Calibration data 96Calibration errors 113Calibration mode 38, 44, 81Calimatic 84CD-ROM 3Channel selection and display assignment 29CIP 43, 49Configuration 24

Alarm 60Cleaning cycles 42, 48

Index

Page 122: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

122

Controller 72Current output 1 52Current output 2 58Individual configuration data 35Limit function 62Output current during Error and HOLD 56Overview 28, 31Sensor 38, 40Sterilization cycles 42, 48Tag number 78Time and date 78Time averaging filter 54WASH contact 76

Connection length for sensors 29Connections 15, 19Controller

Configuration 72Controller equations 70Controller functions 69Controller test 102Manual specification of controller output 102

Controller characteristic 69Converting slope to mV 89

DData entry of premeasured sensors 88Date 79

Display 93Device self-test 97Diagnostics 24, 95

Calibration data 96Device self-test 97Logbook 98Sensor data 96Sensor monitor 99Version 99

Index

Page 123: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

123

IndexDigital sensors

Sensor type selection 39, 45Dimensions 11Display 21Display backlighting 21Disposal 2Documentation 3

EEEPROM test 97Electronic Signature 120Enclosure 11Enclosure components 10Error handling 113Error messages 114Explosion protection 110Extended logbook 98, 120

FFDA 21 CFR Part 11 120FLASH test 97

HHOLD 24, 26

Controller behavior during HOLD 74End 26External activation of HOLD 27Manual activation of HOLD 27Output signal during HOLD 26, 57Output signal response 26

Hysteresis 65, 67

IInstallation 14Intended use 7IrDA communication 102

Page 124: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

124

IndexKKeypad 20

LLimit 1 62Limit 2 66Logbook 98

MManual calibration with buffer entry 86Measured values, display 99Measurement 22, 93, 94Measuring mode 50Memosens

Wiring example 19Memosens cable 17, 18Menu structure 25Mounting plan 11

OOffsetting channels A/B 50Operating modes 24Operating mode, selection 23Operating states 104Options 103, 105

TAN 103Order code 105Output current, fixed value 101Output current range 52, 58Output filter 54Output signal during HOLD 26, 57Overview 9

Page 125: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

125

PPackage contents 3, 10Panel mounting 13Parameter error 113Parameter set A/B

Individual configuration data 35Passcodes 120, 132

Setting 103PFC 71Pipe mounting 12PLC 71Point of measurement, arrangement 29Power supply 15, 16, 110Presetting pH calibration 81Process variable 51, 53, 59Product calibration 90Product line 105Protective hood 12Pulse frequency controller (PFC) 71

Configuration 73Pulse length controller (PLC) 71

Configuration 73

RRAM test 97Rating plates 14Registered trademarks 131Relay 1 62Relay 2 66Relay test 101Release of options 103Reset to factory settings 103Return of products under warranty 2

Index

Page 126: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

126

SSafety information 7, 8Safety instructions 3Selection menu 23Sensocheck 60, 117

Configuration 61Sensoface 113, 117Sensor A 38Sensor B 44Sensor connection 16Sensor data, display 96Sensor defect 118Sensor monitor 99, 101Sensor wear 119Service 24, 100

Controller test 102Factory setting 103IrDA communication 102Passcodes 103Relay test 101Releasing options 103Sensor monitor 101Specifying current outputs 101

Service passcode lost 103Setpoint adjustment (controller) 94Signal colors 21Signal lines 15SIP 43, 49Specifications 106Start-up 8

Index

Page 127: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

127

TTAG 79Tag number (“TAG“) 79TAN options 103, 105

Entering a TAN 103Temperature unit 50Terminal assignments 14Terminals 14Time 79

Display 93Time averaging filter 55Trademarks 131

UUser interface 20

WWarranty 2WASH contact

Configuration 76Wiring 15

ZZero adjustment (ISFET) 82

Index

Page 128: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

128

Page 129: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

129

Page 130: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

130

Page 131: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

131

Trademarks

The following names are registered trademarks. For practical reasons they are shown without trademark symbol in this manual.Stratos®Sensocheck®Sensoface®Calimatic®GainCheck®InPro® is a registered trademark of Mettler-Toledo.Memosens® is a registered trademark of Endress+Hauser Conducta GmbH and Knick Elektronische Meßgeräte GmbH & Co. KG.HART® is a registered trademark of the HART Communication Foundation.

Page 132: Stratos Pro A4 MSPH/MSPH - knick-international.com... MSPH/MSPH Instruction Manual Latest Product Information: 2 Warranty Warranty Defects occurring within 3 years from delivery date

TA-212.016-MS-KNE01 20100110 Software version: 1.x

KnickElektronische Messgeräte GmbH & Co. KGP.O. Box 37 04 15D-14134 Berlin

Phone: +49 (0)30 - 801 91 - 0Fax: +49 (0)30 - 801 91 - 200Internet: http://[email protected]

PasscodesIn the SERVICE – CODES menu you can assign passcodes to protect the access to certain functions.

Operating Mode Passcode

Service (SERVICE) 5555

Diagnostics (DIAG)

HOLD mode

Calibration (CAL)

Configuration (CONF)