stryker air pump - rehabmart.com · educating and training staff on the safe operation of the pump....

58
2016/06 AO-SM60SG-01 Rev 2.0 Stryker Air Pump 2861 Service Manual

Upload: others

Post on 02-Aug-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0

Stryker Air™

Pump

2861

Service Manual

Page 2: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 2 of 58

Table of Contents

Page

Symbols and Definitions 3

Warning / Caution / Note Definitions 4

Introduction 5

Cleaning Instructions 8

Quick Reference Parts List 9

Membrane Replacement 10

Pump Replacement 14

Front Cover Replacement 19

Control Board Replacement 22

Circuit Breaker Replacement 25

Power Inlet Replacement 30

Bumper Replacement 35

Quick-Disconnect Replacement 39

Cable Replacement 43

Krytox® Application 48

Case Closure 49

Functionality Test 50

EMC Declarations 53

Warranty 56

Page 3: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 3 of 58

Symbols and Definitions

SYMBOLS

Warning/Caution - Consult accompanying documentation

Caution

Refer to Instruction Manual

Electrical Safety Mark

Consult Instructions for Use

Type BF Applied Part; Applied Part is the Support Surface

Protective Earth Terminal

IPX4 Protection from Powerful Jets of Water Rating

Equipment Emits Electromagnetic Energy

Manufacturer

Model Number

Serial Number

Power (ON/STANDBY)

Lock

Alarm Indicator

Operating Mode (IsoFlexLAL or SPRPlus)

SPRPlus Pressure Adjustment (Millimeters of Mercury = mmHg)

Warning: dangerous voltage

Page 4: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 4 of 58

Warnings and Cautions

WARNING / CAUTION

The words WARNING and CAUTION carry special meanings and should be carefully reviewed.

WARNING

Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may

also describe potential serious adverse reactions and safety hazards.

CAUTION

Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or

moderate injury to the user or patient or damage to the equipment or other property. This includes

special care necessary for the safe and effective use of the device and the care necessary to avoid

damage to a device that may occur as a result of use or misuse.

Use Extreme

Caution: 120V live

voltage

Page 5: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 5 of 58

Introduction

This manual is designed to assist with the operation and maintenance of the Stryker AirTM

Pump

(“Pump”) that is used to power both the Stryker SPRPlusTM

Single Patient Use Low Air Loss Overlay

System (“SPRPlus”) and IsoFlexLALTM

Support Surface. The SPRPlus is only for use in the United

States.

Carefully read this entire manual before using or beginning maintenance on the Pump. To ensure safe

operation of this equipment, it is recommended that methods and procedures are established for

educating and training staff on the safe operation of the Pump.

INTENDED USE OF THE PRODUCT

The Pump is an accessory used with the SPRPlus and IsoFlexLAL Support Surface to assist in the

prevention and treatment of pressure ulcers. This therapy is recommended to be used in combination

with clinical evaluation of risk factors and skin assessments made by a health care professional.

Patients may be excluded due to some medical conditions such as unstabilized spinal cord injuries or

other conditions as determined by the facilities professional staff.

This Pump is intended to be used:

− With the Stryker SPRPlus or IsoFlexLAL Support Surface only.

− With the hosing and connector assembly provided (Part Number 2861-001-001).

− In acute care, general hospital care, or other locations as prescribed by a physician.

This product is not intended to be used in a home health care environment.

PRODUCT DESCRIPTION

The Pump provides Low Air Loss Therapy and when used in conjunction with SPRPlus or IsoFlexLAL

Support Surface assists in the prevention and treatment of pressure ulcers and includes the following

features:

- The Pump is a versatile accessory that may be used with either the SPRPlus or the IsoFlexLAL

Support Surface.

- Easily attaches to the Footboard of the Bed Frame.

- Incorporates an alarm for detection of hose disconnection from the Pump

- Designed with a “User-Friendly” Control Panel.

OPERATING PRINCIPLE

The Stryker Air Pump is designed to work with either the IsoFlexLAL (Low Air Loss) mattress or the

SPRPlus mattress overlay. When used with the IsoFlexLAL mattress, the Pump provides a continuous flow

of air to the mattress for the Low Air Loss feature of the mattress. Air flows inside the mattress to the

underside of the cover to remove moisture. The mattress, by itself, provides all the necessary support for

the patient. The Pump does not provide support of the patient.

When the Pump is used with the SPRPlus mattress overlay, the Pump provides air for patient support and

Low Air Loss. The patient is supported by air pressure in the mattress overlay. The Pump adjusts its air

output to obtain the pressure value selected by the user. Small pin holes in the top side of the mattress

overlay “leak” air to achieve the Low Air Loss therapy.

Page 6: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 6 of 58

Introduction

SPECIFICATIONS

Pump

Dimensions Height: 8.5 in / 25.6 cm

Width: 8 in / 20.3 cm

Depth: 5 in/ 12.7 cm

Weight 5.5 lbs/ 2.5kgs

Input Voltage AC 110 - 127 Volts

Input Frequency 50 - 60 Hz

Current Consumption 0.25 Amps

Power Consumption < 60 VA

Circuit Protection Dual Circuit Breakers, 240V, 1.0A

Mode of Operation Continuous

Protection Against Electrical Shock Class I, Type BF Applied Part

Air Output − 12.5 Liters / Minute @ 30 mmHg

− 25 Liters / Minute @ 20 mmHg

Pressure Settings (mmHg) 18 to 30 mmHg in 3 mmHg Increments

Power Cord 3 ft/ 0.91m (For Bed Frame Outlet) 15 ft / 4.6m; (For Wall Outlet)

Air Hose 56 in / 142 cm

Air Hose Connection 3/8 Inch Flow Quick Coupling

Packaging 1 Piece per Box

Latex Content User accessible parts of the Pump are not made with natural rubber latex.

Alarms:

Sound Pressure Level 63.1 dB(A)

Hose Disconnection Alarm

(Low Priority Alarm)

− Indicated after 15 minutes for hose disconnection from Pump

(SPRPlus Mode)

Hose Disconnection / High Pressure Alarm (Medium Priority Alarm)

− Indicated after 30 minutes for hose disconnection from Pump

(SPRPlus Mode)

− Indicated after 30 seconds for excessive Pump pressure

System Alarm (Medium Priority Alarm)

− If system error is detected

− Stuck Button

Operating Conditions:

Ambient Temperature 40 to 90 o

F / 5 to 32 o

C

Relative Humidity 30 to 75 % Non-Condensing

Atmospheric Pressure 700 to 1060 hPa

Storage and Shipping Conditions:

Ambient Temperature -40 to 158 oF / -40 to 70

oC

Relative Humidity 10 to 95 %, Non-Condensing

Atmospheric Pressure 500 to 1060 hPa

Protection Against Harmful Ingress of Liquids:

Liquid Ingress Protection IPX4

Product Compliance:

Medical Equipment IEC 60601-1 (3

rd Edition)

UL 60601-1 CAN/CSA C22.2 NO. 601.1 (3

rd Edition)

Collateral Standards Electromagnetic Compatibility IEC 60601-1-2 (See Pages 53-55)

Usability IEC 60601-1-6

Stryker reserves the right to change specifications without notice.

Page 7: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 7 of 58

Introduction

CONTACT INFORMATION

Contact Stryker Customer Service or Technical Support at: (800) 327-0770 or (269) 324-6500.

Stryker Medical 3800 E. Centre Avenue Portage, MI 49002

USA

Please have the serial number of your Stryker product available when calling Stryker Customer

Service or Technical Support. Include the serial number in all written communication.

SERIAL NUMBER LOCATION

The serial number is located on the back of the Pump, on the rear label.

SERIAL NUMBER FORMAT (8 DIGITS):

Serial Number Example: 15J00234

1 5 J 0 0 2 3 4

Manufacture Date (YY/M): 2015 September Sequential Number (NNNNN): 00234

Month Legend (M)

January A

February B

March C

April D

May E

June F

July G

August H

September J

October K

November L

December M

Year Legend (YY)

2015 15

2016 16

2017 17

2018 18

2019 19

Sequential # Legend (NNNNN)

00001 - 99999

Y M N N Y N N N

Page 8: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 8 of 58

Cleaning Instructions

CLEANING / DISINFECTION The exterior of the Pump and hosing assembly should be wiped down between patient uses with a cloth

dampened with disinfectant.

WARNING

− DO NOT spray disinfectant directly on the electrical Pump, or immerse the Pump in any type of

liquid. This could result in a severe electrical hazard. − All disinfection should be done using a “hospital-grade” disinfectant registered with the

Environmental Protection Agency (EPA). − When disinfecting is required, check manufacturer’s instructions before use, and use disinfectant

in accordance with the manufacturer’s instructions.

Suggested Disinfectants

1. Quaternary Cleaners 2. Phenolic Cleaners 3. Chlorinated Bleach Solution (5.25% bleach diluted 1 part bleach to 10 parts water) 4. 70% Isopropyl Alcohol 5. Accelerated Hydrogen Peroxide (AHP)

CAUTION

− Disinfect the Pump, Power Cord, and Hosing Assembly between patient installations and when

servicing, utilizing standard hospital protocol and disinfectants. Failure to disinfect may risk cross-contamination and infection.

− DO NOT autoclave the Pump OR the Hosing Assembly.

− Unplug Pump from its source prior to cleaning. − Do not use harsh cleansers, solvents, or detergents on the Pump. Equipment damage could

occur.

Page 9: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 9 of 58

Quick Reference Replacement Parts List

Part Number Item Description

2861-001-009 COVER, FRONT

2861-001-013 FEMALE COUPLING, 3/8” QUICK DISCONNECT

2861-001-004 HANGING BRACKET

2861-001-006 HANGING BRACKET, PIN

2861-001-005 HANGING BRACKET, RING

2861-001-012 KIT, BUMPER

2861-001-014 KIT, CABLE ASSEMBLY

2861-001-011 KIT, CIRCUIT BREAKER

2861-001-015 KIT, CONTROL BOARD

2861-001-038 KIT, CONTROL MEMBRANE

2861-001-019 KIT, HANGING BRACKET REPLACEMENT

2861-001-018 KIT, INTERNAL PUMP ASSEMBLY

2861-001-008 KIT, POWER INLET

2861-001-016 KRYTOX GPL 205 SEALANT

2861-001-020 MODIFIED NUTDRIVER

2861-001-024 POWER CORDS, PLUG TYPE B – 3 FEET (1M) AND 15 FEET (4.5M)

2861-001-001 STRYKER AIR HOSE ASSEMBLY

2861-001-039 STRYKER AIR MANUAL

2861-001-037 STRYKER AIR SERVICE MANUAL

Page 10: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 10 of 58

Membrane Replacement

Tools Required:

Membrane Kit (2861-001-038)

Small Slotted Screwdriver

#2 Phillips Screwdriver

Acetone

ESD System (Static Strap)

Removal:

1. Unplug the power cord from the power inlet and remove the air hose from the

controller, if attached, and set aside.

2. Set the controller on a work surface.

3. With the controller sitting upright and facing away from you, use a #2 Phillips

screwdriver to remove the six screws and O-rings securing the front cover to the

back cover, see Figure 1.

Figure 1

4. Hold the front and back cases together and tip the controller backwards to remove

the screws and O-rings and set them aside.

5. With the controller upright and facing away from you, turn it counterclockwise 90o

and slightly separate the front and back cases. See Figure 2.

Page 11: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 11 of 58

Membrane Replacement

Figure 2

6. Once separated, use a small slotted screwdriver to unlock both sides of the

membrane ribbon cable connector at J10 and remove the ribbon cable from the

connector. See Figure 3.

Figure 3

7. Carefully peel the membrane ribbon cable up and away from the tape on the inside

of the front cover.

8. With the front case facing toward you, use a small slotted screwdriver to pry up the

membrane from the left side and peel off the membrane. See Figure 4.

Page 12: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 12 of 58

Membrane Replacement

Figure 4

9. Pull the membrane ribbon cable through the slot in the front cover.

10. Remove the two pieces of tape from the inside of the front case.

Installation:

1. Using acetone, clean and wipe down the recessed membrane area of the front case.

2. Slide the membrane ribbon cable through the slot in the front cover.

3. Remove the paper backing from the membrane and align it with the recessed area

on the front cover.

4. Apply the membrane to the recessed area and press down to seal.

5. Make a small loop at the top of the membrane ribbon cable and line up the

membrane ribbon cable with the adhesive tape and press down to seal.

See Figure 5.

Figure 5

6. Peel the plastic backing off of the 3M VHB tape that is adhered to the membrane tail.

7. Apply the adhesive membrane tail to the inside of the front case.

Inside of front case

Loop

Page 13: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 13 of 58

Membrane Replacement

Krytox® Application:

Refer to page 48 for Krytox® application instructions.

Case Closure:

Refer to page 49 for case closure instructions.

Functionality Test:

Refer to page 50 for functionality test procedures.

Page 14: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 14 of 58

Pump Replacement

Tools Required:

Pump Replacement Kit (2861-001-018)

Small Slotted Screwdriver

#2 Phillips Screwdriver

Needle Nose Pliers

Wire Cutters

ESD System (Static Strap)

Removal:

1. Unplug the power cord from the power inlet and remove the air hose from the

controller, if attached, and set aside.

2. Set the controller on a work surface.

3. With the controller sitting upright and facing away from you, use a #2 Phillips

screwdriver to remove the six screws and O-rings securing the front cover to the

back cover. See Figure 6.

Figure 6

4. Hold the front and back cases together and tip the controller backwards to remove

the screws and O-rings and set them aside.

5. With the controller upright and facing away from you, turn it counterclockwise 90o

and slightly separate the front and back cases. See Figure 7.

Page 15: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 15 of 58

Pump Replacement

Figure 7

6. Once separated, use a small slotted screwdriver to unlock both sides of the

membrane ribbon cable connector at J10 and remove the ribbon cable from the

connector. See Figure 8.

Figure 8

7. Using wire cutters, carefully clip the four (4) zip ties that hold the wires together.

8. Unplug the two (2) connectors from the control board connector at J8 and J9.

9. Disconnect the large and small silicone hoses from the muffler. Keep the hoses

attached to the board and air outlet fitting. See Figure 9.

Page 16: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 16 of 58

Pump Replacement

Figure 9

10. Slide the PCB out of the key in the back case and put aside.

11. Lay the back case down, face up, on a work surface.

12. Using needle nose pliers unhook and remove the two (2) pump isolators from the

bottom pump bracket.

13. Slide the two (2) pump isolators out of the black hooks on the top of the pump. See

Figure 10.

Figure 10

14. Lift the pump assembly up and remove it from the back case.

Pressure

transducer

hose

Air outlet

hose

Disconnected

pump isolator

Hook at top of

pump

Page 17: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 17 of 58

Pump Replacement

Installation:

1. Place the back case on a work surface, facing up.

2. Put the pump assembly onto the back metal bracket.

3. Using needle nose pliers, pull the bottom two (2) pump isolators through the holes in

the bottom pump bracket. See Figure 11.

Figure 11

4. Attach the top two (2) pump isolators to the black hooks on the top of the pump.

5. Attach the large and small silicone hoses to the muffler. See Figure 9 for correct

hose orientation.

6. Reinsert the PCB into the key in the back case.

7. Insert the two (2) connectors to the control board connector at J8 and J9.

8. Zip tie the wires. See Figure 12 for placement reference.

Bottom

pump

bracket

Bottom

pump

isolator

Page 18: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 18 of 58

Pump Replacement

Important! Wires must be constrained in accordance with the placement specified below for safety reasons. A) Pump wires: blue, white and black

B) AC inlet module wires: blue, brown and green/yellow

C) Pump and circuit breaker wires: blue, white and black (pump) with blue (right

circuit breaker) and brown (left circuit breaker)

D) Circuit breaker wires: blue (left and right) with brown (right side of left circuit

breaker)

Figure 12

Krytox® Application:

Refer to page 48 for Krytox® application instructions.

Case Closure:

Refer to page 49 for case closure instructions.

Functionality Test:

Refer to page 50 for functionality test procedures.

A

B

C

D

Page 19: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 19 of 58

Front Cover Replacement

Tools Required:

Front Cover Replacement Kit (2861-001-009)

Small Slotted Screwdriver

#2 Phillips Screwdriver

ESD System (Static Strap)

Removal:

1. Unplug the power cord from the power inlet and remove the air hose from the

controller, if attached, and set aside.

2. Set the controller on a work surface.

3. With the controller sitting upright and facing away from you, use a #2 Phillips

screwdriver to remove the six screws and O-rings securing the front cover to the

back cover. See Figure 13.

Figure 13

4. Hold the front and back cases together and tip the controller backwards to remove

the screws and O-rings and set them aside.

5. With the controller upright and facing away from you, turn it counterclockwise 90o

and slightly separate the front and back cases. See Figure 14.

Page 20: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 20 of 58

Front Cover Replacement

Figure 14

6. Once separated, use a small slotted screwdriver to unlock both sides of the

membrane ribbon cable connector at J10 and remove the ribbon cable from the

connector. See Figure 15.

Figure 15

7. Discard the entire front cover.

Page 21: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 21 of 58

Front Cover Replacement

Installation:

1. With the front case upright, line up the grooves with those of the back case.

2. With the controller upright and facing away from you, turn it counterclockwise 90o.

3. Insert the membrane ribbon cable at J10 and push down the sides of the connector

to lock it in place. See Figure 16.

Figure 16

Krytox® Application:

Refer to page 48 for Krytox® application instructions.

Case Closure:

Refer to page 49 for case closure instructions.

Functionality Test:

Refer to page 50 for functionality test procedures.

Page 22: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 22 of 58

Control Board Replacement

Tools Required:

PCB (2861-001-015)

Small Slotted Screwdriver

#2 Phillips Screwdriver

Needle Nose Pliers

ESD System (Static Strap)

Removal:

1. Unplug the power cord from the power inlet and remove the air hose from the

controller, if attached, and set aside.

2. Set the controller on a work surface.

3. With the controller sitting upright and facing away from you, use a #2 Phillips

screwdriver to remove the six screws and O-rings securing the front cover to the

back cover. See Figure 17.

Figure 17

4. Hold the front and back cases together and tip the controller backwards to remove

the screws and O-rings and set them aside.

5. With the controller upright and facing away from you, turn it counterclockwise 90o

and slightly separate the front and back cases. See Figure 18.

Page 23: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 23 of 58

Control Board Replacement

Figure 18

6. Once separated, use a small slotted screwdriver to unlock both sides of the

membrane ribbon cable connector at J10 and remove the ribbon cable from the

connector. See Figure 19.

Figure 19

7. Unplug the two (2) connectors from the control board connector at J8 and J9.

8. Partially remove the PCB from the key in the back case.

9. Remove the small silicone hose from the pressure transducer on the board at PT1.

See Figure 20 for reference.

Page 24: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 24 of 58

Control Board Replacement

Figure 20

10. Slide the PCB completely out of the key in the back case and dispose of according to

applicable local and federal regulations.

Installation:

1. Remove the sticker from the speaker on the bottom of the PCB.

2. Place the back case on a work surface, facing up.

3. Attach the small silicone tube from the muffler to PT1 on the board.

4. Insert the PCB into the key in the back case.

5. Insert the two (2) connectors to the control board connector at J8 and J9.

Krytox® Application:

Refer to page 48 for Krytox® application instructions.

Case Closure:

Refer to page 49 for case closure instructions.

Functionality Test:

Refer to page 50 for functionality test procedures.

Pressure

transducer

hose

Air outlet

hose

Page 25: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 25 of 58

Circuit Breaker Replacement

Tools Required:

Circuit Breaker (2861-001-011)

Small Slotted Screwdriver

#2 Phillips Screwdriver

Wire Cutters

Needle Nose Pliers

ESD System (Static Strap)

Removal:

1. Unplug the power cord from the power inlet and remove the air hose from the

controller, if attached, and set aside.

2. Set the controller on a work surface.

3. With the controller sitting upright and facing away from you, use a #2 Phillips

screwdriver to remove the six screws and O-rings securing the front cover to the

back cover. See Figure 21.

Figure 21

4. Hold the front and back cases together and tip the controller backwards to remove

the screws and O-rings and set them aside.

5. With the controller upright and facing away from you, turn it counterclockwise 90o

and slightly separate the front and back cases. See Figure 22.

Page 26: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 26 of 58

Circuit Breaker Replacement

Figure 22

6. Once separated, use a small slotted screwdriver to unlock both sides of the

membrane ribbon cable connector at J10 and remove the ribbon cable from the

connector. See Figure 23.

Figure 23

7. Using wire cutters, carefully clip the four (4) zip ties that hold the wires together.

8. Unplug the two (2) connectors from the control board connector at J8 and J9.

9. Partially remove the PCB from the key in the back case.

10. Remove the small silicone hose from the pressure transducer on the board at PT1.

11. Remove the large silicone hose from the muffler. See Figure 24 for reference.

Page 27: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 27 of 58

Circuit Breaker Replacement

Figure 24

12. Slide the PCB completely out of the key in the back case and set aside.

13. Using needle nose pliers, pull straight up on the fastons connected to the circuit

breaker.

14. Turn the back case on its side.

15. Holding the circuit breaker from the inside, unscrew the circuit breaker cover on the

outside of the back case.

16. Remove the circuit breaker and dispose of according to local and federal regulations.

17. Put the circuit breaker cover aside.

Installation:

1. Remove the nut from the circuit breaker and discard.

2. With the back case still on its side, insert the new circuit breaker.

3. Holding the circuit breaker from the inside, reinstall the circuit breaker cover to the

back of the case.

4. Place the back case on a work surface, facing up.

5. Using needle nose pliers, reconnect the fastons to the circuit breaker. See

Schematic 1 below for placement.

Pressure

transducer

hose

Air outlet

hose

Page 28: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 28 of 58

Circuit Breaker Replacement

Schematic 1

6. Insert the PCB into the key in the back case.

7. Attach the large silicone tube from the air outlet to the muffler.

8. Attach the small silicone tube from the muffler to PT1 on the board.

9. Insert the two (2) connectors to the control board connector at J8 and J9.

10. Zip tie the wires. See Figure 25 for placement reference.

Important! Wires must be constrained in accordance with the placement specified below for safety reasons. A) Pump wires: blue, white and black

B) AC inlet module wires: blue, brown and green/yellow

C) Pump and circuit breaker wires: blue, white and black (pump) with blue (right

circuit breaker) and brown (left circuit breaker)

D) Circuit breaker wires: blue (left and right) with brown (right side of left circuit

breaker)

View: looking down into the back case

Page 29: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 29 of 58

Circuit Breaker Replacement

Figure 25

Krytox® Application:

Refer to page 48 for Krytox® application instructions.

Case Closure:

Refer to page 49 for case closure instructions.

Functionality Test:

Refer to page 50 for functionality test procedures.

A

B

C

D

Page 30: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 30 of 58

Power Inlet Replacement

Tools Required:

Power Inlet (2861-001-008)

#2 Phillips Screwdriver

Torx Screwdriver T10

Torque Screwdriver

Wire Cutters

Needle Nose Pliers

ESD System (Static Strap)

Removal:

1. Unplug the power cord from the power inlet and remove the air hose from the

controller, if attached, and set aside.

2. Set the controller on a work surface.

3. With the controller sitting upright and facing away from you, use a #2 Phillips

screwdriver to remove the six screws and O-rings securing the front cover to the

back cover. See Figure 26.

Figure 26

4. Hold the front and back cases together and tip the controller backwards to remove

the screws and O-rings and set them aside.

5. With the controller upright and facing away from you, turn it counterclockwise 90o

and slightly separate the front and back cases. See Figure 27.

Page 31: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 31 of 58

Power Inlet Replacement

Figure 27

6. Once separated, use a small slotted screwdriver to unlock both sides of the

membrane ribbon cable connector at J10 and remove the ribbon cable from the

connector. See Figure 28.

Figure 28

7. Using wire cutters, carefully clip the four (4) zip ties that hold the wires together.

8. Unplug the two (2) connectors from the control board connector at J8 and J9.

9. Partially remove the PCB from the key in the back case.

10. Remove the small silicone hose from the pressure transducer on the board at PT1.

11. Slide the PCB completely out of the key in the back case and set aside.

12. Using needle nose pliers, pull straight up on the three (3) fastons connected to the

power inlet.

Page 32: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 32 of 58

Power Inlet Replacement

13. Lay the back case with the back facing up toward you.

14. With a Torx T10 screwdriver, remove the two (2) screws holding the power inlet in

place and set the screws aside.

15. Remove the power inlet and discard.

16. Remove the two (2) rubber gaskets and discard.

Installation:

1. Insert the two (2) new rubber gaskets over the new power inlet.

2. With the back case still lying flat, insert the power inlet with gaskets into the

appropriate slot in the back case.

Important! Ensure the power inlet is inserted in the correct orientation as shown in Figure 29

below.

Figure 29

3. Using a torque screwdriver, use the previously removed two (2) screws and torque

them down to 4 in-lbf being careful not to over torque them.

4. Place the back case on a work surface, facing up.

Page 33: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 33 of 58

Power Inlet Replacement

5. Using needle nose pliers, reconnect the fastons to the circuit breaker. See

Schematic 2 below for placement.

Schematic 2

6. Insert the PCB into the key in the back case.

7. Attach the small silicone tube from the muffler to PT1 on the board.

8. Insert the two (2) connectors to the control board connector at J8 and J9.

9. Zip tie the wires. See Figure 30 for placement reference.

Important! Wires must be constrained in accordance with the placement specified below for safety reasons. A) Pump wires: blue, white and black

B) AC inlet module wires: blue, brown and green/yellow

C) Pump and circuit breaker wires: blue, white and black (pump) with blue (right

circuit breaker) and brown (left circuit breaker)

D) Circuit breaker wires: blue (left and right) with brown (right side of left circuit

breaker)

View: looking down into the back case

Page 34: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 34 of 58

Power Inlet Replacement

Figure 30

Krytox® Application:

Refer to page 48 for Krytox® application instructions.

Case Closure:

Refer to page 49 for case closure instructions.

Functionality Test:

Refer to page 50 for functionality test procedures.

A

B

C

D

Page 35: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 35 of 58

Bumper Replacement

Tools Required:

Bumper (2861-001-012)

#2 Phillips Screwdriver

Wire Cutters

Modified nutdriver bit (2861-001-020)

Torque Screwdriver

Loctite 242 thread locker

ESD System (Static Strap)

Removal:

1. Unplug the power cord from the power inlet and remove the air hose from the

controller, if attached, and set aside.

2. Set the controller on a work surface.

3. With the controller sitting upright and facing away from you, use a #2 Phillips

screwdriver to remove the six screws and O-rings securing the front cover to the

back cover. See Figure 31.

Figure 31

4. Hold the front and back cases together and tip the controller backwards to remove

the screws and O-rings and set them aside.

5. With the controller upright and facing away from you, turn it counterclockwise 90o

and slightly separate the front and back cases. See Figure 32.

Page 36: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 36 of 58

Bumper Replacement

Figure 32

6. Once separated, use a small slotted screwdriver to unlock both sides of the

membrane ribbon cable connector at J10 and remove the ribbon cable from the

connector. See Figure 33.

Figure 33

7. Using wire cutters, carefully clip the four (4) zip ties that hold the wires together.

8. Unplug the two (2) connectors from the control board connector at J8 and J9.

9. Partially remove the PCB from the key in the back case.

10. Remove the small silicone hose from the pressure transducer on the board at PT1.

11. Slide the PCB completely out of the key in the back case and set aside.

12. Using the torque screwdriver and modified nutdriver bit remove the nut from the

inside of the back case and set aside.

Page 37: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 37 of 58

Bumper Replacement

See Figure 34 for reference.

Figure 34

13. Remove the fender washer and set aside.

14. From the other side of the back case, unscrew the bumper, remove the bumper and

neoprene washer and discard.

Installation:

1. Apply Loctite 242 thread locker to the new bumper screw threads.

2. Insert the neoprene washer and bumper into the hole in the back case.

3. From the inside of the back case, install fender washer and nut over the threads of

the bumper and torque to 8.5 in-lbf being careful not to over torque.

4. Place the back case on a work surface, facing up.

5. Insert the PCB into the key in the back case.

6. Attach the small silicone tube from the muffler to PT1 on the board.

7. Insert the two (2) connectors to the control board connector at J8 and J9.

8. Zip tie the wires. See Figure 35 for placement reference.

Important! Wires must be constrained in accordance with the placement specified below for safety reasons. A) Pump wires: blue, white and black

B) AC inlet module wires: blue, brown and green/yellow

C) Pump and circuit breaker wires: blue, white and black (pump) with blue (right

circuit breaker) and brown (left circuit breaker)

D) Circuit breaker wires: blue (left and right) with brown (right side of left circuit

breaker)

Bumper Bumper

Page 38: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 38 of 58

Bumper Replacement

Figure 35

Krytox® Application:

Refer to page 48 for Krytox® application instructions.

Case Closure:

Refer to page 49 for case closure instructions.

Functionality Test:

Refer to page 50 for functionality test procedures.

A

B

C

D

Page 39: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 39 of 58

Quick Disconnect Replacement

Tools Required:

Quick Disconnect Kit (2861-001-013)

#2 Phillips Screwdriver

Wrench, 21 mm.

Small Slotted Screwdriver

Torque Screwdriver

ESD System (Static Strap)

Removal:

1. Unplug the power cord from the power inlet and remove the air hose from the

controller, if attached, and set aside.

2. Set the controller on a work surface.

3. With the controller sitting upright and facing away from you, use a #2 Phillips

screwdriver to remove the six screws and O-rings securing the front cover to the

back cover. See Figure 36.

Figure 36

4. Hold the front and back cases together and tip the controller backwards to remove

the screws and O-rings and set them aside.

5. With the controller upright and facing away from you, turn it counterclockwise 90o

and slightly separate the front and back cases. See Figure 37.

Page 40: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 40 of 58

Quick Disconnect Replacement

Figure 37

6. Once separated, use a small slotted screwdriver to unlock both sides of the

membrane ribbon cable connector at J10 and remove the ribbon cable from the

connector. See Figure 38.

Figure 38

7. Disconnect hose from the back of the Quick-Disconnect. See Figure 39.

8. Remove Retaining Washer using 21 mm Wrench.

9. Remove Quick-Disconnect.

Page 41: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 41 of 58

Quick Disconnect Replacement

Figure 39

Installation:

1. Insert replacement Quick-Disconnect as shown in Figure 40.

Figure 40

2. Ensure that Quick-Disconnect release tab is oriented as shown in Figure 41.

Page 42: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 42 of 58

Quick Disconnect Replacement

Figure 41

3. Replace washer and tighten securely while holding the external end of the Quick-

Disconnect to keep release tab in the correct orientation.

4. Re-connect hose to the back of Quick-Disconnect.

Krytox® Application:

Refer to page 48 for Krytox® application instructions.

Case Closure:

Refer to page 49 for case closure instructions.

Functionality Test:

Refer to page 50 for functionality test procedures.

Page 43: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 43 of 58

Cable Replacement

Tools Required:

Cable Kit (2861-001-014)

#2 Phillips Screwdriver

Wire Cutters

Needle Nose Pliers

Nut Driver (11/32”)

Torque Screwdriver

ESD System (Static Strap)

Removal:

1. Unplug the power cord from the power inlet and remove the air hose from the

controller, if attached, and set aside.

2. Set the controller on a work surface.

3. With the controller sitting upright and facing away from you, use a #2 Phillips

screwdriver to remove the six screws and O-rings securing the front cover to the

back cover. See Figure 36.

Figure 42

4. Hold the front and back cases together and tip the controller backwards to remove

the screws and O-rings and set them aside.

5. With the controller upright and facing away from you, turn it counterclockwise 90o

and slightly separate the front and back cases. See Figure 37.

Page 44: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 44 of 58

Cable Replacement

Figure 43

6. Once separated, use a small slotted screwdriver to unlock both sides of the

membrane ribbon cable connector at J10 and remove the ribbon cable from the

connector. See Figure 38.

Figure 44

7. Using wire cutters, carefully clip the four (4) zip ties that hold the wires together.

8. Unplug the two (2) connectors from the control board connector at J8 and J9.

9. Partially remove the PCB from the key in the back case.

10. Remove the small silicone hose from the pressure transducer on the board at PT1.

11. Slide the PCB completely out of the key in the back case and set aside.

12. Using needle nose pliers, pull straight up on the fastons connected to the circuit

breakers (brown and blue wires) and dispose of wires.

Page 45: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 45 of 58

Cable Replacement

13. Using needle nose pliers, pull straight up on the fastons connected to the power inlet

(brown, blue and green/yellow) and dispose of wires.

14. Using a #2 Phillips screwdriver, remove the screw in the pump shelf and put aside.

Remove the green/yellow wire. See Figure 39 below for reference.

Figure 45

15. Using a 11/32” nutdriver, remove the nut on the ground lug and put aside. Remove the green/yellow wire and dispose of the wire. See Figure 40 below for reference.

Figure 46

Page 46: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 46 of 58

Cable Replacement

Installation:

1. Place the back case on a work surface, facing up.

2. Using the previously removed screw, install the green/yellow wire to the pump shelf.

3. Using the previously removed nut, install the green/yellow wire to the ground lug and

torque to 18 in-lbf being careful not to over torque.

4. Attach the green/yellow wire faston to the middle tab of the power inlet.

5. Attach the brown and blue wires to the circuit breakers and power inlet according to

Schematic 3 below.

Schematic 3

6. Insert the PCB into the key in the back case.

7. Attach the small silicone tube from the muffler to PT1 on the board.

8. Insert the two (2) connectors to the control board connector at J8 and J9.

9. Zip tie the wires. See Figure 41 for placement reference.

Page 47: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 47 of 58

Cable Replacement

Important! Wires must be constrained in accordance with the placement specified below for safety reasons. A) Pump wires: blue, white and black

B) AC inlet module wires: blue, brown and green/yellow

C) Pump and circuit breaker wires: blue, white and black (pump) with blue (right

circuit breaker) and brown (left circuit breaker)

D) Circuit breaker wires: blue (left and right) with brown (right side of left circuit

breaker)

Figure 47

Krytox® Application:

Refer to page 48 for Krytox® application instructions.

Case Closure:

Refer to page 49 for case closure instructions.

Functionality Test:

Refer to page 50 for functionality test procedures.

.

A

B

C

D

Page 48: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 48 of 58

Krytox® Application

Tools Required:

DuPont™ Krytox® Grease (2861-001-016)

Gloves

1. Apply Krytox® to the groove of the back case ensuring that there are no spaces. See

Figure 42 and Figure 43 for reference.

2. Remove any excess Krytox® from the inside of the back case.

Figure 48

Figure 49

Back case

Acceptable

Krytox®

application: no

gaps or voids

Back case

Unacceptable

Krytox®

application: gaps

and/or voids

Page 49: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 49 of 58

Case Closure

Tools Required:

#2 Phillips Screwdriver

Torque Screwdriver with #2 Phillips bit

Simple Green® or comparable degreaser

1. Line up the front and back cases with the controller facing toward you.

2. With the right side slightly ajar, insert the membrane ribbon cable into the connector

on the PCB board at J10.

3. Press down both sides of the connector to lock the membrane ribbon cable in place.

4. Slide the PCB into the key of the front cover and press the front and back case

together.

5. Place the controller face down with the back case facing up toward you.

6. Reinstall the 6 screws with O-rings, using a #2 Phillips screwdriver, previously

removed to secure the front case to the back case. DO NOT TIGHTEN.

7. Using a torque screwdriver, torque down the 6 screws to 8.5 in-lbf being careful not

to over torque them.

8. Wipe down the seam of the controller with Simple Green®, or comparable degreaser,

to remove any excess Krytox®.

Page 50: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 50 of 58

Functionality Test

Follow final test procedure to verify proper functionality prior to returning it to service.

Equipment List

Pressure Meter (Extech HD700 Differential Pressure Manometer or equivalent)

Stopwatch

Power Strip with Switch

SPR Mattress

IsoFlex Mattress

3/8 x 3/8 x 1/8 ‘T’ Fitting

6”of PVC tubing (5/16” ID x 1/2” OD) x 2

1’ of PVC tubing (1/8” ID x 1/4” OD)

Test Procedure

Step 1 Hanger Inspection

Inspect hangers to ensure each can be locked in all three height adjustment positions.

Acceptance Criteria: Locking pin engages on all three positions of both hangers.

Step 2 System Initialization

Connect power strip to appropriate voltage AC source for controller being tested.

Switch power strip off.

Connect unit to power strip

Switch power strip on.

Acceptance Criteria: All LEDs illuminate for 1 second and unit powers up (unit will power up in previous operating mode).

Step 3 Calibration

Acceptance Criteria: Alarm does not sound after unit power-up.

Step 4 ON/OFF Switch

Press the Power Button.

Acceptance Criteria: Blue LED is illuminated (under power button).

Press the Power Button.

Acceptance Criteria: The button is tactile and the white LED is illuminated (under power button).

Page 51: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 51 of 58

Functionality Test

Step 5 Button Test

Press the Mode button repeatedly.

Acceptance Criteria: Button is tactile and toggles between IsoFlex and SPR mode.

Press the Lock button twice.

Acceptance Criteria: Button is tactile and Lock toggles ON and OFF

Set to SPR mode.

Press the mmHg button, and note that pressure indicator on the membrane panel switches to the next value with each repeated press (18, 21, 24, 27, 30).

Acceptance Criteria: Button is tactile and increases pressure setting to the maximum then sequences to the lowest value.

Step 6 SPR+ Pressure Test

Connect the ‘T’ fitting to the three PVC hoses, the pressure meter, controller air output and CPR hose as seen in Figure 44 below.

Figure 50

Connect the CPR hose to the SPR mattress.

Set controller pressure level to 30 mmHg.

Measure pressure.

Acceptance Criteria: Pressure reading is in the range of 28 to 32 mmHg.

Step 7 IsoFlex Flow Test

Connect the ‘T’ fitting to the three PVC hoses, the pressure meter, controller air output and CPR hose as seen in Figure 28 above.

Connect the CPR hose to the IsoFlex mattress.

Toggle mode to IsoFlex.

Measure pressure.

Acceptance Criteria: Pressure reading is greater than 20 mmHg.

‘T’ Fitting

CPR Hose

Page 52: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 52 of 58

Functionality Test/Preventive Maintenance Check List

Step 8 High Pressure Regulation Test

Manually block air outlet to simulate a kinked hose (High Pressure)

Measure pressure after it stabilizes.

Acceptance Criteria: Pressure regulates to 48 to 52 mmHg.

Step 9 High Pressure Alarm Test

Wait for alarm (approximately 30 seconds).

Acceptance Criteria: System alarms.

Allow air to flow normally.

Acceptance Criteria: Alarm is disabled (approximately 5 seconds).

Step 10 Power Failure Test

Set unit to SPR mode, with 30 mmHg selected.

Cycle power on power strip to simulate a power failure.

Acceptance Criteria: Systems recalls mode and firmness setting after power returns.

Step 11 Electrical Safety Testing

Electrical safety test unit according to the table below:

TEST ACCEPTANCE CRITERIA

GROUND IMPEDENCE < 0.10 Ω

CURRENT LEAKAGE < 0.50 mA

INSULATION WITHSTAND > 1.488kVac

FUNCTIONAL 0.05A - 0.40A

Page 53: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 53 of 58

Product Compliance Declarations

GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

The Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Pump should

assure that it is used in such an environment.

Emissions test Compliance Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1 The Pump uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class A The Pump is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Harmonic emissions IEC 61000-3-2

Class A

Voltage fluctuations / flicker emissions IEC 61000-3-3

Complies

GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

The Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Pump should

assure that it is used in such an environment.

Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD) IEC 61000-4-2

±6 kV contact ±8 kV air

±6 kV contact ±8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.

Electrical fast transient/burst IEC 61000-4-4

±2 kV for power supply lines ±1 kV for input/output lines

±2 kV for power supply lines Not Applicable

Mains power quality should be that of a typical commercial or hospital environment.

Surge IEC 61000-4-5 ±1 kV line(s) to line(s) ±2 kV line(s) to earth

±1 kV line(s) to line(s) ±2 kV line(s) to earth

Mains power quality should be that of a typical commercial or hospital environment.

Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11

<5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec

<5 % UT (>95 % dip in UT) for 0,5 cycle 40 % UT (60 % dip in UT) for 5 cycles 70 % UT (30 % dip in UT) for 25 cycles <5 % UT (>95 % dip in UT) for 5 sec

Mains power quality should be that of a typical commercial or hospital environment. If the user of the Pump requires continued operation during power mains interruptions, it is recommended that the Pump be powered from an uninterruptible power supply or a battery.

Power frequency (50/60 Hz) magnetic field IEC 61000-4-8

3 A / m 3 A / m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE! UT is the a.c. mains voltage prior to application of the test level.

Page 54: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 54 of 58

Product Compliance Declarations

GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY - NON

LIFE SUPPORTING

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

The Pump is intended for use in the electromagnetic environment specified below. The customer or the user

of the Pump should assure that it is used in such an environment.

Immunity test IEC 60601 test

level Compliance

level Electromagnetic environment – guidance

Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3

3 Vrms 150 kHz to 80 MHz 3 V/m 80 MHz to 2,5 GHz

3 Vrms 3 V/m

Portable and mobile RF communications equipment should be used no closer to any part of the Stryker Pump , including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance

Pd 2.1

Pd 2.1 80 MHz to 800 MHz

Pd 3.2 800 MHz to 2,5 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,

a should be less than the compliance

level in each frequency range. b

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and

land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Pump is used exceeds the applicable RF compliance level above, the Pump should be observed to

verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the Pump.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Page 55: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 55 of 58

Product Compliance Declarations

GUIDANCE AND MANUFACTURER’S DECLARATION – RECOMMENDED SEPARATION

DISTANCES

Recommended Separation Distances Between Portable And Mobile RF Communications Equipment And The Pump

The Pump is intended for use in an electromagnetic environment in which radiated RF disturbances are

controlled. The customer or the user of the Pump can help prevent electromagnetic interference by

maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Pump as recommended below, according to the maximum output power of the

communications equipment.

Rated maximum output power of transmitter

W

Separation distance according to frequency of transmitter M

150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz

Pd 2.1 Pd 2.1 Pd 3.2

0,01 0.12 0.12 0.23

0,1 0.38 0.38 0.73

1 1.2 1.2 2.3

10 3.8 3.8 7.3

100 12 12 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Page 56: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 56 of 58

Warranty

LIMITED WARRANTY

The Stryker Air™ Pump has a warranty of THREE (3) YEARS under normal use, conditions, and with appropriate periodic maintenance as described in this manual. The Stryker Air™ Pump has an expected service life of FIVE (5) YEARS.

This statement constitutes Stryker’s entire warranty with respect to the aforesaid equipment. Stryker makes no other warranty or representation, either expressed or implied, except as set forth herein. There is no warranty of merchantability and there are no warranties of fitness for any particular purpose. In no event shall Stryker be liable here under for incidental or consequential damages arising from or in any manner related to sales or use of any such equipment.

CONDITIONS AND LIMITATIONS

This statement constitutes Stryker’s entire warranty with respect to the aforesaid equipment. Stryker makes no other warranty or representation, either expressed or implied, except as set forth herein. There is no warranty of merchantability and there are no warranties of fitness for any particular purpose. This warranty does not extend to, nor cover:

− Normal wear and tear; or

− Damage or product failure due to causes beyond Stryker’s control such as, but not limited to abuse, theft, fire, flood, wind, lightning, freezing, clogging of support surface pores due to tobacco smoke, unusual atmosphere conditions, material degradation due to exposure to moisture; or

− Damage to support surface or support surface handles through the use of the support surface for patient transfer or transport.

Normal use is defined as normal hospital or facility usage. Damages arising from abnormal use such as those caused by needle punctures, burns, chemicals, negligent use or improper care or improper cleaning or staining resulting from it are exempt from warranty coverage.

TO OBTAIN PARTS AND SERVICE

Stryker products are supported by a nationwide network of dedicated Stryker Field Service Representatives. These representatives are factory trained, available locally, and carry a substantial spare parts inventory to minimize repair time. Simply call your local representative or call Stryker Customer Service USA at (800) 327-0770 or (269) 324-6500.

RETURN AUTHORIZATION

Merchandise cannot be returned without approval from the Stryker Customer Service Department. An authorization number will be provided which must be printed on the returned merchandise. Stryker reserves the right to charge shipping and restocking fees on returned items. Special, modified, or discontinued items not subject to return.

DAMAGED MERCHANDISE

ICC Regulations require that claims for damaged merchandise must be made with the carrier within fifteen (15) days of receipt of merchandise. Do not accept damaged shipments unless such damage is noted on the delivery receipt at the time of receipt. Upon prompt notification, Stryker will file a freight claim with the appropriate carrier for damages incurred. Claim will be limited in amount to the actual replacement cost. In the event that this information is not received by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on the delivery receipt at the time of receipt, the customer will be responsible for payment of the original invoice in full. Claims for any short shipment must be made within thirty (30) days of invoice.

Page 57: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 57 of 58

Notes

Page 58: Stryker Air Pump - Rehabmart.com · educating and training staff on the safe operation of the Pump. INTENDED USE OF THE PRODUCT ... Air Hose Connection 3/8 Inch Flow Quick Coupling

2016/06 AO-SM60SG-01 Rev 2.0 Page 58 of 58