    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E

    A G R E A T E R M E A S U R E O F C O N F I D E N C

    W W W . K E I T H L E Y . C O M

    http://www.keithley.com/http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    2/33

    2

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    INTRODUCTIONMany electronic test systems use relay switching to connect multipledevices to sources and measurement instruments. In some cases,multiple sources and measuring instruments are connected to a singledevice. Switching allows automating the testing of multiple devices,thereby reducing error and costs.

    Designing the switching for an automated test system demandsan understanding of the signals to be switched and the tests to be

    performed. Test requirements can change frequently, so automatedtest systems must provide the exibility needed to handle a variety ofsignals. Even simple test systems often have diverse and conictingswitching requirements. Given the versatility that test systems mustoffer, designing the switching function may be one of the most complexand challenging parts of the overall system design.

    As a signal travels from its source to its destination, it may encountervarious forms of interference or sources of error. Each time the signalpasses through a connecting cable or switch point, the signal may bedegraded. When calculating the overall system accuracy, the engineermust include not only the effects of the switch itself but all the switchinghardware in the system.

    The quality of a switch system depends in large part on its ability to

    preserve the characteristics of the test signals routed through it. Forexample, when the test signal is a low voltage, the switching systemmust minimize errors such as offset voltage and IR drops. Leakagecurrent may be a problem for high resistance and low current switchingapplications. Depending on the type of test signal involved, specicswitching techniques must be used to maintain signal integrity throughthe switch system. This e-handbook describes switching techniquesfor a variety of test signals.

    VOLTAGE SWITCHINGMany different applications involve switching a voltmeter or voltage sourceto multiple devices, including testing batteries, electrochemical cells,circuit assemblies, thermocouples, etc.

    The types of switch cards and the techniques used in these applicationswill depend on the magnitude and impedance of the voltages switched.

    The approximate level for low voltage switching is in the millivolt rangeor less, mid-range levels are from 1V to 200V, and voltages greater than200V demand the use of high voltage switching methods.

    Switching a Voltmeter to Multiple Sources in SeriesFigure 1 illustrates switching a voltmeter to a series string of 30 batteriesor voltage sources (VS). To avoidshort-circuiting one or more ofthese sources, it is necessaryto open a given channelbeforeclosing a second one (break-befor e-mak e operation). Toguard against short-circuiting,add fuses in series with eachvoltage source to prevent

    damage to the switch card. Besure not to exceed the common-mode rating of the switch card.In this example, each battery is12V and the total voltage acrossthe string is 360V. A channel-to-channel voltage rating and acommon-mode voltage rating ofat least 500V is desirable.

    Switching a Voltage Source to Multiple LoadsFigure 2 shows a single voltage source connected to multiple loads, suchas lamps. If two or more loads are connected to the source, the voltageat each load may be less than expected due to current ow through thecommon impedances (R), such as the test leads and trace resistance.As additional loads are connected, the total current will increase, therebyincreasing the voltage drop across the common impedances (R).

    Ch. 1 VS1

    Ch. 2 VS2

    Ch. 30 VS30

    V

    Figure 1. Switching a voltmeter to multiplesources in series

    Want to Explore FFEATURED RESOURCESn Switching in Multipoint Testing

    n Testing Devices withHigh Voltage andHigh Current

    ADDITIONAL RESOURCESn Optimizing a Switch System fo

    Mixed Signal Testing

    n Tips and Techniques for DesigEffective, Efcient Switch Syst

    http://www.keithley.com/http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/data?asset=4631http://www.keithley.com/data?asset=4631http://www.keithley.com/data?asset=3594http://www.keithley.com/data?asset=3594http://www.keithley.com/data?asset=3594http://www.keithley.com/data?asset=3594http://www.keithley.com/data?asset=55828http://www.keithley.com/data?asset=55828http://www.keithley.com/data?asset=55828https://event.on24.com/eventRegistration/EventLobbyServlet?target=registration.jsp&eventid=334696&sessionid=1&key=29041162A4AF068A3FF941FE2FBF43ED&sourcepage=registerhttps://event.on24.com/eventRegistration/EventLobbyServlet?target=registration.jsp&eventid=334696&sessionid=1&key=29041162A4AF068A3FF941FE2FBF43ED&sourcepage=registerhttps://event.on24.com/eventRegistration/EventLobbyServlet?target=registration.jsp&eventid=334696&sessionid=1&key=29041162A4AF068A3FF941FE2FBF43ED&sourcepage=registerhttp://www.keithley.com/data?asset=3594http://www.keithley.com/data?asset=4631https://event.on24.com/eventRegistration/EventLobbyServlet?target=registration.jsp&eventid=334696&sessionid=1&key=29041162A4AF068A3FF941FE2FBF43ED&sourcepage=registerhttps://event.on24.com/eventRegistration/EventLobbyServlet?target=registration.jsp&eventid=334696&sessionid=1&key=29041162A4AF068A3FF941FE2FBF43ED&sourcepage=registerhttp://www.keithley.com/data?asset=55828http://www.keithley.com/data?asset=55828http://www.keithley.com/data?asset=3594http://www.keithley.com/data?asset=3594http://www.keithley.com/data?asset=3594http://www.keithley.com/data?asset=4631http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    3/33

    3

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    Switch ResistanceWhen sw i tch ing avo l tage sou rce tomultiple devices, it maybecome necessary tocompensate for voltagedrops due to switchresistance. In particular,

    if the devices have lowresistance, the currentf lowing through theswitches may cause asignicant voltage drop.To prevent this problem,remote sensing can beused to correct for any voltage drops in switches and wiring. With remotesensing, external sense connections are made across the load. Therefore,the subsequent programmed output voltage will be the actual voltageacross the load.

    For example, Figure 3a shows a 5V source being switched to an integratedcircuit (IC). The contact resistance for each switch is 1. If the currentdrawn from the source is 500mA, the voltage drop across each switch

    will be 500mV, and the voltage at the integrated circuit will be reducedby a total of 1V. Operation of the IC will likely be unsatisfactory. Figure 3bshows a 5V source with remote sense. In this case, sense leads are alsoconnected to the load. This will ensure that the actual voltage across theload will be 5V, and the IC will operate as intended. Note that the voltage atthe source output terminals is 6V.

    Do not usehot-switching with remote sensing since the voltage from thesource may become excessive during switching.

    R Ch. 1

    Ch. 2

    Ch. 3

    Ch. n

    Figure 2. Switching a voltage source tomultiple loads

    HI500mA

    IC

  • 7/30/2019 Switching e Handbk Neu

    4/33

    4

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    LOW VOLTAGE SWITCHINGWhen switching low voltage signals (millivolts or less), special techniquesmust be used to avoid unwanted voltage errors. However, the requiredaccuracy will determine if these techniques are necessary. Theseunwanted voltage errors may be due to thermoelectric offset voltage in theswitch card and connecting cabling, switch lm contamination, magneticinterference, and ground loops. For more information on low voltagemeasurements, refer to Keithleys Low Level Measurements Handbook.

    Thermoelectric Offset VoltageThe contact potentialor thermoelectric offset voltage is the keyspecification of a switch card designed for low voltage switching.Thermoelectric voltage is the voltage generated by thermal differencesat the junction of dissimilar metals, such as between the nickel-ironreed relays and the copper conductor to which they are connected.The temperature gradient is typically caused by the power dissipated bythe energized coil. The offset voltage adds directly to the signal voltageand can be modeled as an unwanted voltage source in series with theintended signal. The offset voltage will cause an error in the appliedstimulus to a device under test or the value measured by the voltmeter.

    As shown in Figure 4a, the offset voltage (E) of a single pole relay isadded into the circuit. As a result, the measured voltage (VM) will be the

    sum of the source voltage (Vs) and the offset voltage (E). To minimizethis offset voltage, a low voltage switch card uses a two-pole relay, asshown in Figure 4b. Here, the offset voltage (ET1) in the circuit HI is largelycanceled by the offset voltage (ET2) in circuit LO. The contact potentialof some low voltage cards is specied with the card used as a two-poleswitch to take advantage of this cancellation. It may be specied as percontact pair. However, total cancellation cannot be achieved becausetemperature differences will cause ET1 and ET2 to be slightly different.

    The drift due to the thermoelectric voltage of a switch card depends onseveral factors, including the type of relays used (reed, solid-state, orelectromechanical). The drift also depends on the coil drive technique(latching ornon-latching). Finally, the material used for the contact plating(for example, nickel alloy or gold) also affects the thermoelectric voltage.

    The power diss ipated in the coil of the reed relay may cause thetemperature to increase for several minutes after it is energized, so itis important to make low voltage measurements within a few seconds

    after contact closure. If many measurements are taken over severalminutes after closure, a steadily increasing thermoelectric voltage will beadded to the reading. Thermal time constants may range from secondsto hours. Even though solid-state relays have no coil dissipation, heat

    generated by internal IR drops can still produce thermoelectric drift.Latching relays use a pulse of current to actuate them, so they havevery low thermoelectric drift.

    VS

    +

    VM

    +

    E

    VS = VM

    Figure 4a. Potential in loop is added to measurement

    VS

    +

    VM

    +

    ET1+

    + ET2

    If ET1 = ET2

    then VS = VM

    Figure 4b. Canceling of potentials in loop

    http://www.keithley.com/http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    5/33

    5

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    The connections to the switch card itself represent another source ofthermally generated voltages. Wherever possible, make connectionsto the card with untinned copper wire, and keep all leads at the sametemperature.

    The thermoelect ric offset voltage due to the switch card and theinterconnecting cable may be compensated for by using a short-circuitedchannel to establish a zero reference. Connect a clean copper wireconnected between the HI and LO terminals of an unused channel. Close

    the channel and measure the offset voltage. Then open this channel.This value can be subtracted from subsequent readings made on otherchannels. This approach is not ideal because the offset will change overtime due to self-heating and changes in the ambient temperature.

    When switching low voltages while making low resistance measurements,the thermoelectric offset voltages may be canceled by using offsetcompensation. This technique requires making two voltage measurementswith two different values of current. To determine the resistance, thedifference between the two resulting voltages is divided by the differenceof the two test currents:

    Refer to the section on Low Resistance Switchingfor more informationon this method.

    Switch Film ContaminationOver time, a contaminating film can form on the surface of a relaycontact. This lm can increase the relay contact resistance, which canmake the switched voltages erratic when measuring or sourcing lowvoltage. Voltages greater than 100mV are usually sufcient to clear thiscontamination. Usingscanner cards with solid state switches is one wayto avoid this problem.

    Magnetic InterferenceMagnetic interference can be a problem in low voltage circuits. A high rateof change in magnetic ux, such as that produced by a switching powersupply or by switching a high current signal on and off, can induce a pulseof many microvolts in an adjacent circuit. This can easily cause signicanterror in a low voltage circuit. This type of interference can be minimized byseparating thenoise source and the sensitive circuit as much as possible,by magneticshielding, and by reducing the enclosed area of the noisesource and signal conductors. Twist the HI and LO wires of each channeltogether to minimize the enclosed area.

    Ground LoopsGround loops can easily occur in a complex test system. If a smallpotential difference exists between two ground points, some groundcurrents may ow through a sensitive part of the system. This may occuronly when certain switches are closed, so it can be very difficult todiagnose. When possible, try to maintain a single system ground point.When this is not possible, isolation techniques using optical coupling orbalanced transformers may help by increasing the effective resistancebetween the two points, thereby reducing the common ground current toa negligible level.

    R=V1 V2

    I1 I2

    http://www.keithley.com/http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    6/33

    6

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    HIGH VOLTAGE SWITCHINGSome applications, such as testinginsulation resistance of cables andprinted circuit boards and high-pot testing, may require switching highvoltages. To avoid switch card damage, be particularly careful whenswitching voltages of ~200V or higher.

    Choose a card rated for the desired voltage and power levels. Coldswitching, if feasible, will extend the relay life and make it possible toincrease the allowable current. Be sure to use appropriately rated cables

    when switching high voltages.

    Reactive loads can cause excessive current and voltage transients,so current surge limiting for capacitive loads and voltage clampingfor inductive loads are required to prevent damage to the relays andexternal circuitry.

    The surge current from a capacitive load is i = C dV/dt and must be limitedto less than the rated current to protect the relays. Figure 5 shows aseries resistance (R) used to limit the charging current. The resistor mustbe able to withstand the applied voltage; otherwise, the high voltage mayarc across the resistor, damaging the device under test and the switchcard. All components must be rated for peak voltage and current.

    When determining the current limit for a reactive load, consider themaximum load in VA. For example, if the maximum load is 10VA and

    500V is switched, then the current must be limited to 20mA. The seriesresistance is then calculated as:

    Inductive reaction voltage (L(di/dt)) must be less than the scanner cardsmaximum voltage rating. Figure 6 shows two typical clamping circuits,one using a diode for clamping DC voltages and the other using back-to-back zener diodes for clamping AC voltages.

    R= 25k 500V

    20mA

    R

    IN

    High VoltageSwitch

    OUTCapacitiveLoad

    Figure 5. Limiting capacitive reaction current

    IN

    High VoltageSwitch

    OUT InductiveLoad

    +

    Figure 6. Limiting inductance reaction voltage

    http://www.keithley.com/http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    7/33

    7

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    HIGH IMPEDANCE VOLTAGE SWITCHINGHigh impedance voltage switching may be necessary in applicationssuch as monitoring electrochemical cells and measuring semiconductorresistivity. Switching and measuring voltage sources with high internalimpedance are subject to a number of errors, including offset currents,stray leakage paths, and electrostatic interference. Shunt capacitancemay increase the settling time. This section discusses these error sourcesand provides an example application.

    When choosing a card to switch high impedance voltage, make sure thecard has a low offset current. Any offset current owing through a highimpedance device will cause an unwanted voltage to appear across thedevice. This offset voltage will be added to the voltage measurement.

    High impedance circuitry is susceptible to electrostatic interference, souse shielding to avoid noise pickup. The device under test, as well as theconnecting cables, should be well shielded.

    Leakage paths can cause error by reducing the measured voltage. Suchleakage paths may be present in the test instrument, switching cards,cables, and fixtures. To minimize errors due to such leakage paths,choose a switch card with high isolation resistance and use guardingwherever possible, including in the test xturing and cabling. Also, selectinsulating materials with the highest possible insulation resistance.

    Response time is another concern when switching high impedancevoltage signals. Excessive response time may be caused by shuntcapacitance, both in the switch itself and in the associated cables. Insome cases, the shunt capacitance can be largely neutralized by the useof a driven guard, which will keep the shield of the cable at nearly thesame potential as the center conductor (or high impedance lead) of thecable. Figure 7a shows a high impedance voltage connected through aswitch to an electrometer voltmeter. Notice the slow response to a stepfunction. To guard the signal, make a connection between the guardoutput (unity gain or preamp output) of the electrometer and the shieldof the switch card, as shown in Figure 7b. Some electrometers, suchas Keithleys Model 6517B and Model 6514, can make this connectioninternally by enabling the internal guard connection. Enabling the guardeffectively reduces the cable and switch capacitance, thereby improving

    the electrometers response time.

    Switch cards appropriate for high impedance voltage switching includethe Models 7158 and 6522. A card with triax connections is necessary ifthe guard voltage could exceed 30VDC. This precaution is necessary toensure safety.

    High ImpedanceVoltage Source

    ElectrometerVoltmeter

    HI

    LO

    Volts

    Time

    Figure 7a. Switching a high impedance voltage source to an electrometer

    High ImpedanceVoltage Source

    ElectrometerVoltmeter

    HI

    LO

    Volts

    Time

    Guard

    Figure 7b. Using a driven guard to neutralize shunt capacitance

    Want to Explore FFEATURED RESOURCESn Troubleshooting Low Voltage

    Measurement Problems

    n Common Challenges inVoltage Switching

    ADDITIONAL RESOURCESn Gearing Up for Parametric Tes

    High Voltage Future

    n Low Level Measurements Han

    http://www.keithley.com/http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/support/data?asset=54358http://www.keithley.com/support/data?asset=54358http://www.keithley.com/support/data?asset=54358http://www.keithley.com/data?asset=54612http://www.keithley.com/data?asset=54612http://www.keithley.com/data?asset=54612http://www.keithley.com/data?asset=55801http://www.keithley.com/data?asset=55801http://www.keithley.com/data?asset=55801http://www.keithley.com/promo/wb/259http://www.keithley.com/promo/wb/259http://www.keithley.com/data?asset=54612http://www.keithley.com/support/data?asset=54358http://www.keithley.com/promo/wb/259http://www.keithley.com/data?asset=55801http://www.keithley.com/data?asset=55801http://www.keithley.com/data?asset=54612http://www.keithley.com/data?asset=54612http://www.keithley.com/support/data?asset=54358http://www.keithley.com/support/data?asset=54358http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    8/33

    8

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    CURRENT SWITCHINGApplications for current switching include testing of power supplies,insulation resistance, capacitor leakage, resistivity of materials, batteries,and semiconductors. This section discusses current switching in general.Special techniques are required when switching high current or lowcurrent. Current switching applications may require connecting multiplecurrent signals to a single meter, or may require connecting a singlesource to multiple loads.

    Each case has differentswitch configurations andconcerns.

    Connecting MultipleCurrent Signals to aSingle AmmeterMost current measuringa pp l i c a t i o n s r e qu i r ethat all current paths becontinuous, even when aparticular current signalis not connected to theammeter. To accomplishthis, switch cards designed

    for current switching oftenuse SPDT or Form Crelays. As shown in Figure8, when a channel is in an open state (i.e., not connected to the output),signal HI is connected to signal LO through the normally c losed (NC)contact. In this case, only signal HI is switched. Signal LO is hardwiredto the ammeter and to all other signal LOs. This is normally the casewhen testing components with relatively low currents (

  • 7/30/2019 Switching e Handbk Neu

    9/33

    9

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    Connecting a Current Source to Multiple LoadsUnlike switching current signals to an ammeter, switching a currentsource to multiple loads usually does not require maintaining the currentpath at all times.

    For this case, a card with Form A switches is suitable, as shown in Figure12. To connect the current source to Load #1, close relay Ch. 1. Ensurethat Ch. 1 is opened before Ch. 2 is closed if the loads are capable ofstoring energy, such as a battery or a large capacitor. Two high energy

    loads connected in parallel may damage the relays.Often, cards designed for voltage switching can be used for currentswitching for currents from 1A to 1A. For current less than 1A, choosea card that has high isolation resistance (across open contacts, from highto low, and between channels) and low offset current.

    Cold switching is always preferable to ensure maximum relay life. Coldswitching involves turning the source current off (placing it on standby)before opening or closing a relay.

    HI

    LO

    Ch. 1FromCurrentSource #1

    HI

    LO

    FromCurrentSource #2

    AHILO

    Model 7053High Current Card

    Ch. 2

    Jumper

    Jumper

    HI

    LO

    Ch. 1

    FromCurrentSource #1

    HI

    LO

    FromCurrentSource #2

    VHILO

    Ch. 2

    RS

    RS

    HI

    LO

    Ch. 1FromCurrentSource

    Load #1

    Ch. 2 Load #2

    Figure 10. Using the Model 7053 High Current Card to switch current

    Figure 11. Using shunt resistors to measure current

    Figure 12. Connecting a current source to multiple loads

    http://www.keithley.com/http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    10/33

    10

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    HIGH CURRENT SWITCHINGWhen designing a switching circuit for high current (>1A), pay particularattention to the maximum current, maximum voltage, and VAspecications of the switch card. Also, its important to choose a switchcard with low contact resistance to avoid excessive heating that cancause contact failure by welding the contacts together. Contact heating iscaused by I2R power dissipation.

    High current switching can be used for either switching a power supply

    to multiple loads or for switching an ammeter to multiple sources. Figure13 is an example of switching a power supply to multiple loads using amultiplexerscanner card. In this example, the power supply will output1A to each of four loads. This doesnt present a problem when only onechannel is closed at a time. However, when all four channels are closed,the power supply will output 4A through the common path. Unfortunately,even though the maximum current of a particular channel is specied at1A, the common path on the switch card may not be able to tolerate 4A.This is not usually specied for a switch card, but the limitation is usuallya function of the trace width and connector ratings. One way to avoid thisproblem is to use a switch card with independent (isolated) relays andconnect with wires rated to carry the total current.

    When currents that exceed the commercially available card ratings mustbe switched, then a general-purpose switch card can be used to controlexternal high current relays or contactors. The user must supply the coilcurrent for the external relays.

    Under no circumstances should unlimited power (direct from the powerline) ever be connected directly to a switch card.

    When switching high VA loads (power line to motors, pumps, etc.), solid-state relays (SSR) are often used. Industry standard SSR modules areavailable from many sources. These SSR modules are plugged into aboard, such as the Keithley Model PB-24SM, that can be controlled fromTTL-level digital outputs. These digital outputs can be from a board thatplugs into a PC or from a scanner mainframe. Some SSR modules canswitch up to 1kVA to high power loads.

    Concerns about switching transients with reactive loads also apply to highcurrent switching. Cold switching, where contacts do not make or breakcurrent or voltage, is recommended for currents in excess of 100mA toavoid radiated interference and to extend relay life.

    Ch. 1 Load #11A

    Ch. 2 Load #21A

    Ch. 3 Load #31A

    Ch. 4 Load #41A

    CommonPath

    4A

    PowerSupply

    Figure 13. Switching a power supply to multiple loads

    Table 1. High Current Switches

    Card Model Current Rating Compatible Mainframes

    7053 5A (10 channels) 7001, 7002

    7700 3A (two channels) 2700, 2701, 2750

    7702 3A (two channels) 2700, 2701, 2750

    7705 2A (40 SPST Isolated) 2700, 2701, 2750

    3740 3A (28 SPST Isolated)7A (4 SPST Isolated)

    3706A

    3721 2A switched, 3A carry (40 channels)3A (2 channels)

    3706A

    3730 616 Matrix (1A switched, 2A carry) 3706A

    7174A 2A carry 707A/708A

    http://www.keithley.com/http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    11/33

    11

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    LOW CURRENT SWITCHINGWhen switching currents of 1A or less, special techniques must be usedto minimize interferences, such as offset currents, leakage currents,electrostatic interference, triboelectric currents, and electrochemicalcurrents. These types of interferences may be due to the scanner carditself, the connecting cables, or the test fixturing. Allowing sufficientsettling time before making a measurement is also crucial when switchinglow current.

    Offset CurrentsThe offset current specication of the switch card should be as low aspossible for switching low current. Offset current is a spurious currentgenerated by a switching card even though no signals are applied.It is usually caused by galvanic sources on the card. Offset current isespecially signicant when measuring low currents if the magnitude ofthe offset is comparable to the low current being measured. Scannercards designed to minimize offset current are commercially available. Forexample, the Model 7158 Low Current Scanner Card has

  • 7/30/2019 Switching e Handbk Neu

    12/33

    12

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    In Figure 14b, a guard ring is added to the ionization chamber. This guardcircuit splits the leakage resistance into two parts. The coax connectionsofFigure 14a have been replaced with triax connections. The voltageacross RL1 is the picoammeter voltage burden, normally less than 1mV,so the resulting leakage current will be quite small. The full bias voltageappears across RL2 (the leakage resistance between the inside and outershields of a triax cable). However, the resulting leakage current does notow through the meter, so it is not added to the measurement.

    Guarding may also be necessary to prevent leakage current in low currenttest xturing.

    Electrostatic InterferenceShielding is required because high impedance circuitry is susceptible topickup of spurious radiated noise. Relay contacts should be shieldedfrom the coil to minimize induced noise from the relay power supply. TheDUTs and interconnect cabling should also be shielded to prevent noisepickup. Any good conductor can be used as a shield. All shields shouldbe connected to circuit LO.

    Triboelectric Currents

    Triboelectr ic currents are generated by charges created by frict ionbetween a conductor and an insulator, such as between the conductorand the insulation of a coax cable. This noise source can be reducedby using special low noise cable that has a conductive coating (such asgraphite) and securing the interconnect cabling to minimize movement.

    Electrochemical CurrentsElectrochemical currents are generated by galvanic battery action causedby contamination and humidity. Thorough cleansing of joints and surfacesto remove electrolytic residue, including PC etchants, body salts, andprocessing chemicals, will minimize the effect of these parasitic batteries.

    Settling TimeWhen a relay opens or closes, there is a charge transfer (on the order ofpicocoulombs), which causes a current pulse in the circuit. This charge

    transfer is due to the mechanical release or closure of the contacts,the contact-to-coil capacitance, and the stray capacitance betweensignal and relay drive lines. After a relay is closed, its important to allowsufcient settling time before taking a measurement. This time can beas long as several seconds. If a step voltage is applied to the circuit, atransient current is generated. This current will gradually decay to a steadyvalue. The time needed to reach the steady value, or the settling time,can be used to determine the proper delay time for the measurement.See Keithleys Low Level Measurements Handbookfor more detaileddiscussions of generated currents and guarding.

    RL

    pA

    HI

    LOI L

    IM

    CoaxConnectionIon Chamber IX

    RL2

    RL1 pA

    HI

    LO

    I L

    Guard

    IM

    TriaxConnectionIon Chamber

    GuardRing

    IX

    IM = IX + I L

    IM = IX

    Note: RLrepresentsthe leakageresistance of theion chamberinsulator aswell as thecable leakage.

    Cross-sectionalview

    Cross-sectionalview

    a.

    b.

    Figure 14. Guarding to reduce leakage currents when switching

    an ion chamber to a picoammeter

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W K

    http://www.keithley.com/http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    13/33

    13

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    LOW CURRENT MATRIX SWITCHINGSome low current applications require switching through amatrix. Thisis often the case when switching several Source-Measure Units (SMUs)to multipin devices or wafer level semiconductor measurements. Allthe special techniques used to minimize interferences for low currentmultiplexing also apply to low current matrix switching. However, offsetcurrent and leakage current are the types of interference most frequentlyencountered. In general, low current matrix switching can becomecomplex and is best explained by illustration.

    Figure 15 shows three SMUs from the Model 4200-SCS SemiconductorCharacterization System connected through a two-pole matrix (Model7174A Low Current Matrix Card) to a multipin device. Notice that theFORCE terminal of each SMU is connected to the HI terminal of theswitch and the Guard of the SMU is connected to the Guard terminal ofthe switch. In this example, guarding becomes important to avoid leakagecurrents because the FORCE terminal of the SMU is at the test potential.With the guard terminal at the same potential, the leakage current throughthe switch is minimized. Guarding will also speed up the response time.The ground terminal (GNDU) of the Model 4200-SCS is connected to asinglerow of the matrix.

    Figure 16 shows two independent SMUs with the Output LO terminalsconnected to separate rows. Note that each pin of the device under test,

    the FET, is connected to a single column. The guard terminals of thematrix card are not connected to the device. To connect the SMUs tothe transistor, close crosspointRow 1/Column 2 to connect SMU1 to thedrain, Row 3/Column 1 to connect SMU2 to the gate, and Row 2/Column3 and Row 4/Column 3 to connect both SMU Output LO terminals to thesource terminal of the FET.

    Remote sensing may be necessary if the current through a given pathis high enough to cause a significant voltage drop. Remote sensingcompensates for test lead and switch voltage drops and ensures that theprogrammed output voltage of the SMU is delivered to the load. Remotesensing allows making accurate load voltage measurements. Figure 17shows SMU1 connected to four rows to enable remote sensing, therebyallowing accurate measurement of the collector-emitter voltage of the BJT.SMU2 is used to supply the relatively small base current and requires only

    two rows. To connect SMU1 between the emitter and the collector, closecrosspoints Row 1/Column 4, Row 2/Column 5, Row 3/Column 2, andRow 4/Column 1. Note that remote sensing requires the use of four rows

    and four columns. To connect SMU2 between the base and the emitter,close crosspoints Row 5/Column 3 and Row 6/Column 1. Remote senseis not needed to source the base current because the lead resistancedoes not affect the current.

    HI

    Guard Row 4Force

    HI

    Guard Row 3

    Force

    Guard

    HI

    Guard Row 2Force

    Guard

    Column 4Column 3Column 2

    HI

    Guard Row 1ForceGuard

    Column 1

    GNDU

    SMU3

    SMU2

    SMU1

    Multi-Pin Device Under Test

    Model 4200-SCSSemiconductorCharacterizationSystem

    Model 7174A Low Current Matrix Card

    Figure 15. Using the Model 4200-SCS to test a multipin device

    HI

    Guard Row 4Output LO

    HI

    Guard Row 3Output HI

    Guard

    HI

    Guard Row 2Output LO

    Column 4Column 3Column 2

    HI

    Guard Row 1Output HI

    Guard

    Column 1

    SMU2

    SMU1

    Drain

    Gate

    Source

    Figure 16. Using two SMUs to test a FET

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W K

    http://www.keithley.com/http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    14/33

    14

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    When connecting two SMUs in a matrix, the Output LO terminals of bothSMUs can be tied together and connected to the device using one row.However, the SMU in remote sense may be outputting high current (>1mA)and the resulting voltage drop may interfere with the second SMU if it isused to measure small voltages (

  • 7/30/2019 Switching e Handbk Neu

    15/33

    15

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    RESISTANCE SWITCHINGSwitching an ohmmeter to a device under test is common to a varietyof applications, including measuring the insulation resistance ofmaterials, continuity testing of cables and connectors, contact resistancemeasurements, and measuring the resistance of components such asresistors, thermistors, and potentiometers.

    Resistance measurements may range from less than 1 to greater than1012, so the switching techniques required may vary signicantly based

    on the magnitude of the resistance. Low resistance measurements aregenerally those less than 100, mid-range are those from 100 to 107,and high resistance measurements are those greater than 107. As withany measurement, the exact point at which low resistance techniquesshould be used depends upon the desired measurement accuracy. Forexample, a 1ppm measurement of a 1k resistor implies an uncertainty ofless than 1m, which can only be achieved by using a four-wire technique.

    For scanning resistances in the mid-range (100 to 10M), either single-ended (one-pole) or differential (two-pole) methods can be used. Thesingle-ended method is shown in Figure 18a. Note that all the resistorsunder test have a common terminal that is connected to meter low.

    Figure 18b illustrates the differential method. A two-pole relay is used toconnect both terminals of the unknown resistor to the meter.

    The application usually dictates whether the single-ended or differentialmethod should be used for switching. The advantage of the single-endedmethod is that it requires only half as many switches as the differentialmethod. The only advantage of the differential method is that the offsetvoltages of the two switches tend to cancel each other. However, giventhat these are microvolt-level voltages, the differential method is notusually an issue for mid-range resistance switching.

    HI

    LO

    Ohmmeter

    Figure 18a. Single-ended resistance switching

    HI

    LO

    Ohmmeter

    Figure 18b. Differential method for switching resistances

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    http://www.keithley.com/http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    16/33

    16

    |

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    LOW RESISTANCE SWITCHINGAppli cations such as contact resistance measurements and cablecontinuity testing typically involve switching low resistances. Lowresistance (

  • 7/30/2019 Switching e Handbk Neu

    17/33

    17

    |

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    resistor (R), as shown in Figure 20a, the offsets are not likely to canceland may actually combine to cause an even greater error. Even a properlyconnected two-pole switch, like that shown in Figure 20b, will have somedifferential offset voltage that will not cancel out. The error caused by suchoffset voltage can be virtually eliminated by using offset compensation.

    Figure 21 shows an example of the preferred connection scheme whentesting metal-oxide varistors. The Model 2410-C SourceMeter instrumentis ideal for testing MOVs because of its ability to source up to 1100V at21mA. One Model 7154 High Voltage Card connects the source lines tothe devices and another Model 7154 connects the sense lines. The two-pole relays minimize lead resistance and voltage drop in the switch. This isparticularly important when measuring relatively small voltages (millivolts).

    HI Sense

    LO Sense

    Ch. 1R

    Ch. 2

    HI Source

    LO Source

    Ohmmeter

    Figure 20a. Offset voltage of Ch. 1 and Ch. 2 will be added to sense voltage

    HI Sense

    LO Sense

    Ch. 1

    R

    Ch. 2

    HI Source

    LO Source

    Ohmmeter

    Figure 20b. Offset voltages of Ch. 1 will tend to cancel

    Ch. 1

    Ch. 2

    Ch. 3

    Ch. 4

    OutputIn/Out HI

    In/Out LO

    TriggerLink

    Model 2410-CSourceMeter

    Ch. 1

    Ch. 2

    Ch. 3

    Ch. 4

    OutputSense HI

    Sense LO

    #1

    #2

    #3

    #4

    Model 7001 with 7154Slot #1

    Model 7154Slot #2

    Figure 21. Switching multiple varistors

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    18/33

    18

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    On the other hand, the example of Figure 22 minimizes lead resistanceonly. With the HI leads and LO leads on the same two-pole relays, offsetvoltages in the Model 3721 multiplexer switches will not cancel eachother because very little current ows in the high impedance sense leads.Depending on the needs of the application, the offset voltages can beaddressed with the of fset-compensated ohms feature of the Series 2600ASystem SourceMeter instrument.

    In some applications, the resistances to be tested may have a commonterminal. In this case, only one two-pole switch per resistor is necessary,as shown in Figure 23. This approach has the advantage that only half asmany switches are required, so it is more economical. The disadvantage isthat the offset voltage of the single pole in the sense circuit is added to themeasurement. Switching only a single sense lead will not cancel the offsetvoltage as happens when both sense leads are switched through the sametwo-pole relay. However, offset compensation will generally correct for thiserror source at the cost of increased measurement time. Some ohmmetersprovide this feature, including two Keithley product congurations, theModel 2700 DMM/Switch Mainframe with the Model 7701 32-ChannelDifferential MultiplexerCard and the Model 3706A System Switch/ DMMwith the Model 3721 Dual 120 Multiplexer Card. This feature is known ascommon side ohms. When operating in common side ohms mode, eachinstrument has the ability to recognize the appropriate card and convertthe two-wire connection into a four-wire measurement.

    Other IssuesSome low resistance applications may require sourcing a relatively hightest current and measuring very low voltages. This may require using ahigh current switch card for sourcing and a low voltage switch card formeasuring.

    When switching low impedances, shielding needs are minimal. Shieldingis generally needed only in electrically noisy environments.

    Series 2600System SourceMeter

    Instrument

    Ch. 1#1

    Ch. 2#2

    Ch. 3

    #3

    Ch. 4#4

    Ch. 21

    Ch. 22

    Ch. 23

    Ch. 24

    HI Source

    HI Sense

    LO Source

    LO Sense

    Output Output

    4-ElementDiode Array3721 Card #1 3721 Card #2

    Figure 22. Switching multiple diodes

    HI Source

    HI Sense

    Ch. 1 R1

    LO Source

    LO SenseOhmmeter

    Ch. 2R2

    Ch. 3R3

    Figure 23. Four-wire resistance switching with common terminal

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    19/33

    19

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    HIGH RESISTANCE SWITCHINGApplicati ons such as measuring capacitor leakage, multi-conductor

    cable insulation resistance, and pin-to-pin leakage of connectors require

    measuring high resistances through switches. Special techniques are

    required for switching resistances greater than 10M. Different high

    resistance switching techniques will be needed, depending on whether

    the source voltage/measure current or the source current/measure

    voltage method of determining resistance is used. However, techniques

    such as shielding and guarding apply to both methods.

    Source Voltage/Measure Current MethodThis technique for measuring high resistance involves sourcing a known

    voltage, measuring the resulting current, and calculating the resistance.

    As shown in the schematic in Figure 24, one side of each resistor is

    connected through a set of switches to the picoammeter (pA), while the

    other end is connected through a second set of switches to a DC voltage

    source. To measure the resistance of R1, close switches Ch. 1 and Ch. 4.

    If the measured current is less than 100nA, the switches used to connect

    the resistors to the picoammeter must have low offset current and high

    isolation resistance. In many cases, a Form C switch is preferred. The

    Form C switch keeps one end of the device under test at guard potential

    (approximately 0V) when the switch is de-energized. This prevents leakage

    current across the relay that would degrade the measurements. The Form

    C switch also allows the device under test to be charged for a controlled

    time interval before measuring.

    General-purpose switches are usually sufcient for switching the voltage

    source unless high voltages (>100V) are involved, in which case, high

    voltage switch cards are necessary. If high voltage is used, current

    limiting resistors are needed to avoid damage to the current switches in

    case the device under test breaks down. The current limiting resistor (R)is placed in series with the device under test. The value of this resistor is

    chosen such that, in the event of device failure, the short circuit current

    will not exceed the maximum current specication of the relays on either

    set of switch cards. The voltage rating of these resistors must be at least

    equal to the test voltage.

    In some cases, the device under test may have a common terminal, as

    shown in Figure 25. In this case, the voltage source will be applied to

    all the devices simultaneously. As a result, the excitation voltage time

    is different for each resistance to be measured. This will cause errors

    for some time-dependent applications, such as measuring insulation

    resistance or capacitor leakage. In these cases, the measured resistance

    is a function of the excitation time.

    HI

    LO

    Ch. 1

    Ch. 2

    Ch. 3

    R1

    R2

    R3

    R

    R

    R

    pAR = Current Limiting Resistor

    Ch. 4

    Ch. 5

    Ch. 6

    CurrentSwitches

    VoltageSource

    HI

    LO

    R1

    R2

    R3

    R

    R

    R

    pA

    R = Current Limiting Resistor

    Ch. 1

    Ch. 2

    Ch. 3

    CurrentSwitches

    VoltageSource

    Figure 24. Switching high resistance

    Figure 25. Switching high resistance with a common terminal

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    20/33

    20

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    When the measured current is large compared to the relay current offsetspecication, then Form A switches can provide a more economicalsolution, as shown in Figure 26. In this example, the leakage currentof all open relays will contribute to the measured current. Unfortunately,this leakage current cannot be guarded. Also, if there is a leakage pathbetween devices (as when measuring the insulation resistance betweentraces on a circuit board or multi-conductor cable), this will also degrademeasurement accuracy. In this diagram, each of the resistors has oneend connected together to the voltage source. In this case, voltage

    source switching is unnecessary, because the device under test cannotbe pre-charged.

    Special techniques must be used to minimize interferences such asoffset currents, leakage currents, electrostatic interference, triboelectriccurrents, and electrochemical currents. These sources of errors may bedue to the switch card or the connecting test xturing.

    Source Current/Measure Voltage MethodIn some cases, high resistance is measured by sourcing a current andmeasuring the resulting voltage. Although this is not the preferred methodfor measuring high resistance, it is necessary for some applications,such as van der Pauw resistivity measurements of semiconductors.High impedances are being switched, so some of the same switchingtechniques used when switching low current and high impedance

    voltages are equally applicable here. To avoid errors, choose a card withlow offset current and high isolation resistance. Shielding is necessary toprevent noise due to electrostatic interference.

    Figure 27 shows a simplied van der Pauw resistivity measurement witha current source and a differential voltmeter. By closing crosspoints Row1/Column 1 and Row 2/Column 2, the current source is connected tosample terminals 1 and 2. Closing crosspoints Row 3/Column 3 and Row4/Column 4 will connect the voltmeter between terminals 3 and 4. Formore detailed information on van der Pauw resistivity measurements,consult Keithleys Low Level Measurements Handbook.

    HI

    LO

    R1

    R2

    R3

    pA

    Ch. 1

    Ch. 2

    Ch. 3

    VoltageSource

    TestSample

    DifferentialVoltmeter

    2

    3

    4

    Col. 1 Col. 2 Col. 3 Col. 4

    Row 1

    Row 2

    Row 3

    Row 4

    +

    1

    Figure 26. Switching high resistances with common terminal andForm A switches

    Figure 27. Van der Pauw resistivity measur ements through a switch matrix

    Want to Explore FFEATURED RESOURCESn Accurate Low-Resistance

    Measurements Start withIdentifying Sources ofError

    n Hall EffectMeasurementsin Materials

    Characterization

    ADDITIONAL RESOURCESn Solutions for Production Testin

    n Production Testing of Thermist2400 Digital SourceMeter

    http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/support/data?asset=53682http://www.keithley.com/support/data?asset=53682http://www.keithley.com/support/data?asset=53682http://www.keithley.com/support/data?asset=53682http://www.keithley.com/support/data?asset=53682http://www.keithley.com/data?asset=55773http://www.keithley.com/data?asset=55773http://www.keithley.com/data?asset=55773http://www.keithley.com/data?asset=55773http://www.keithley.com/data?asset=55773http://www.keithley.com/data?asset=3668http://www.keithley.com/data?asset=3668http://www.keithley.com/data?asset=4100http://www.keithley.com/data?asset=4100http://www.keithley.com/data?asset=4100http://www.keithley.com/data?asset=55773http://www.keithley.com/support/data?asset=53682http://www.keithley.com/data?asset=4100http://www.keithley.com/data?asset=4100http://www.keithley.com/data?asset=3668http://www.keithley.com/data?asset=55773http://www.keithley.com/data?asset=55773http://www.keithley.com/data?asset=55773http://www.keithley.com/data?asset=55773http://www.keithley.com/support/data?asset=53682http://www.keithley.com/support/data?asset=53682http://www.keithley.com/support/data?asset=53682http://www.keithley.com/support/data?asset=53682http://gw1.vtrenz.net/?K5CHYUBW35
  • 7/30/2019 Switching e Handbk Neu

    21/33

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    22/33

    22

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    RF AND MICROWAVE SWITCHINGGiven the explosive growth of the communications industry, a tremendousamount of testing is being performed on the various components thatmake up dif ferent communications systems. These components rangefrom active components such as Radio Frequency Integrated Circuits(RFICs) and Microwave Monolithic Integrated Circuits (MMICs) tocomplete communication systems. While the testing requirements andprocedures for these components differ widely, all are tested at very highfrequencies, typically in the gigahertz range. The main components in atest system may include DC bias, DC measurement, RF power meter,network analyzer, RF sources, and other instruments. Automating the testprocess and improving test efciency demands integrating RF/microwaveand low frequency switching systems into the test system.

    Microwave Switch TypesAvailable microwave switch congurations include a simple single-poledouble throw (SPDT) switch, multi-position switch, matrix, and cascade.

    An SPDT switch has one input port, which can be connected to one oftwo output ports. A multi-position switch connects one input port to one ofseveral output ports. Keithleys System 46 can accommodate eight SPDTcoaxial microwave relays and four multi-pole coaxial microwave relays.

    A matrix switch can connect any input to any output. Two types of matricesare used in microwave switching: blocking and non-blocking. A blockingmatrix connects any one input to any one output. Other inputs and outputscannot be connected at the same time. Anon-blocking matrix allowsmultiple paths to be connected simultaneously through the matrix.

    The cascade switch conguration is an alternate form of multi-positionswitch. The cascade switch connects one input to one of many outputsusing multiple relays. The path length (and therefore, the phase delay)varies, depending on the number of relays that the signal must go through.

    RF Switch Card SpecicationsThe use of a switch will inevitably degrade the perfo rmance of themeasurement system, so it is important to consider several criticalparameters that may affect system performance signicantly. Duringthe design phase, the costs and benefits are often weighed againsteach other to achieve an optimal solution. When choosing an RF switchsystem, some of the critical electrical specications to review includecrosstalk(path isolation),insertion loss , voltage standing wave ratio(VSWR), andbandwidth.

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    23/33

    23

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    RF Switch Design ConsiderationsWhen designing an RF switch system, additional factors that may affectswitch system performance includeimpedance matching, termination,power transmission, signal lters,phase distortion, and cabling.

    Impedance Matching: Given that the switch is positioned between themeasurement instruments and the DUT, matching the impedance levels ofall elements in the system is critical. For optimal signal transfer, the outputimpedance of the source should be equal to the characteristic impedanceof the switch, the cables, and the DUT. In RF testing, the most commonly

    used impedance levels are 50 and 75. Whichever impedance level isrequired, proper matching will ensure the overall system integrity.

    The i nput VSWR and signal path VSWR determine the l imitati on onthe accuracy of the measurement:

    Mismatch Uncertainty (dB) = 20 log (1 sig pathinstr)

    where

    If both the signal path output and the instrument input have good

    VSWRs of 1.3:1 at a frequency, then the uncer tainty due to mismatchalone is 0.15dB.

    Termination:At high frequencies, all signals must be properly terminatedor the electromagnetic wave will be reected from the terminating point.This, in turn, will cause an increase in VSWR. An unterminated switchincreases VSWR in its off condition, while a terminated switch will try toprovide a 50 match on or off. The VSWR increase may even damage thesource if the reected power is large enough. All paths through a systemmust be terminated with their characteristic impedance.

    Power Transmission:Another important consideration is the systems

    ability to transfer the RF power from instrument to DUT. Due to insertionloss, the signal may require amplication. In other applications, it maybe necessary to reduce the signal power to the DUT. An amplifier orattenuator may be needed to ensure that the required level of power istransmitted through the switch.

    Signal Filter: Signal lters can be useful in a number of circumstances,such as when spurious noise is inadvertently added to the signal as ittravels through the switch. They can also be helpful if the original signalfrequency does not t in the DUT testing frequency. In these cases, lterscan be added to the switch to modify the signal frequency bandwidth,or spurious signals at unwanted frequencies can be eliminated from thesignal to the DUT.

    Phase Distortion:As the size of a test system expands, signals from the

    same source may travel to the DUT via paths of different lengths, resultingin phase distortion. This specication is often referred to as propagationdelay. For a given conducting medium, the delay is proportional to thelength of the signal path. Different signal path lengths will cause the signalphase to shift. This phase shift may cause erroneous measurementresults. To minimize phase distortion, keep the path lengths the same.

    Taking all of these design considerations into account when conguringan RF/microwave switch system can be simplified by ordering apackage solution, such as the System 46 RF/Microwave Switch System.It can be congured with up to 32 channels for controlling microwaveswitches. It also tracks contact closures for proactive maintenance ofrelays, and performance parameters, such as VSWR or insertion loss,for trend analysis.

    =VSWR 1

    VSWR + 1

    Want to Explore FFEATURED RESOURCESn Conguring an Optimal

    RF/Microwave SwitchSystem

    n Broadband SignalRouting SolutionsBrochure

    ADDITIONAL RESOURCESn Keithleys Switching Systems

    Control Solutions for DC, RF, a

    n RF/Microwave Switching SysteAchieving the Performance Yo

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/data?asset=5130http://www.keithley.com/data?asset=5130http://www.keithley.com/data?asset=5130http://www.keithley.com/data?asset=5130http://www.keithley.com/data?asset=4371http://www.keithley.com/data?asset=4371http://www.keithley.com/data?asset=4371http://www.keithley.com/data?asset=4371http://www.keithley.com/data?asset=6190http://www.keithley.com/data?asset=6190http://www.keithley.com/data?asset=6190http://www.keithley.com/data?asset=55589http://www.keithley.com/data?asset=55589http://www.keithley.com/data?asset=55589http://www.keithley.com/data?asset=4371http://www.keithley.com/data?asset=5130http://www.keithley.com/data?asset=55589http://www.keithley.com/data?asset=55589http://www.keithley.com/data?asset=6190http://www.keithley.com/data?asset=6190http://www.keithley.com/data?asset=4371http://www.keithley.com/data?asset=4371http://www.keithley.com/data?asset=4371http://www.keithley.com/data?asset=5130http://www.keithley.com/data?asset=5130http://www.keithley.com/data?asset=5130http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    24/33

    24

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    DIGITAL SWITCHINGHigh speed digital signals exhibit RF behavior in real-world devices, whichcreates a need for RF or microwave components when routing thesesignals in test systems. Digital (and optical) engineers must be aware ofRF parameters in product development and production testing.

    Digital Design ConsiderationsThe rst consideration when switching digital signals is providing adequatebandwidth to preserve the frequency harmonics of the digital pulse and

    their power spectral distribution. The switching solution should haveas a minimum the bandwidth of the DUT, and preferably equal to thebandwidth of any higher frequency source and measurement equipment.

    Because the 3dB specications of switch cards are based on sinewavesignals, an approximation is needed for square waves. If the frequency ofthe square wave is known, the required bandwidth is determined by thehighest harmonic that needs to pass without distortion, whether that isthe third, fth, seventh, or higher harmonic.

    If the frequency of the square wave is unknown, the following equationapproximates the minimum bandwidth for digital signals:

    Digital switching has some other considerations in addition to those listedin the section for RF Switch Design Considerations:

    Arcing: The RFI (radio frequency interference) generated by high voltageswitching may disrupt high-speed logic circuits.

    Hot switching:When digital signals are being switched, they maychange state with any length of power interruption. Using hot switchingprevents digital state changes.

    Make-before-break:Use Form A isolated switches instead of Form Crelays to allow for make-before-break operation.

    Bandwidth (Hz) = 0.35risetime (s)

    Want to Explore FFEATURED RESOURCESn Product Testing Where RF

    Meets Broadband Convergence of RF, Optical,and Digital Test Environments

    n S46 MicrowaveSwitch SystemsJust what you

    need Philosophy

    ADDITIONAL RESOURCESn Selector Guide and Data Sheet

    and Probes

    n Series 700 Switch Cards and A

    n Series 7000 Switch Card Acce

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/data?asset=9670http://www.keithley.com/data?asset=9670http://www.keithley.com/data?asset=9670http://www.keithley.com/data?asset=9670http://www.keithley.com/data?asset=9670http://www.keithley.com/data?asset=5960http://www.keithley.com/data?asset=5960http://www.keithley.com/data?asset=5960http://www.keithley.com/data?asset=5960http://www.keithley.com/data?asset=5960http://www.keithley.com/data?asset=4371http://www.keithley.com/support/data?asset=50660http://www.keithley.com/support/data?asset=50660http://www.keithley.com/support/data?asset=50660http://www.keithley.com/data?asset=9483http://www.keithley.com/data?asset=9483http://www.keithley.com/data?asset=550http://www.keithley.com/data?asset=550http://www.keithley.com/data?asset=5960http://www.keithley.com/data?asset=550http://www.keithley.com/data?asset=9483http://www.keithley.com/support/data?asset=50660http://www.keithley.com/support/data?asset=50660http://www.keithley.com/data?asset=4371http://www.keithley.com/data?asset=5960http://www.keithley.com/data?asset=5960http://www.keithley.com/data?asset=5960http://www.keithley.com/data?asset=5960http://www.keithley.com/data?asset=9670http://www.keithley.com/data?asset=9670http://www.keithley.com/data?asset=9670http://www.keithley.com/data?asset=9670http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    25/33

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W

    25

    A G R E A T E R M E A S U R E O F C O N F I D E N

    ACTUATION TIME.The time between application of the nominal relaycoil voltage and the nal closure of the relay contacts after the contactbounce interval.

    AUTOMATIC CJC (COLD JUNCTION COMPENSATION).A system formulti-channel thermocouple measurements where the temperature ateach thermocouple/switch card junction is sensed and compensatedfor automatically.

    BANDWIDTH.The range of frequencies that can be switched,

    conducted, or amplied within certain limits. Under given load condi-tions, bandwidth is dened by the 3dB (half-power) points.

    BANK.A group of relays with a common connection for scanning ormultiplex applications.

    BLOCKING.A multiplexer of matrix arrangement of relays where onlyone signal path is active at any given time. Common in RF/microwavetesting to maintain the characteristic impedance of the test system.

    BREAK-BEFORE-MAKE.Disconnecting the present circuit beforeconnecting a new circuit. Also known as Break/Make.

    CARRY CURRENT.The maximum continuous current of closed relaycontacts. Most relays are rated higher for carry current than switchedcurrent. (Heat is generated by I2R losses for carry current and I2Rlosses plus arcing for switched current.)

    CHANNEL. One of several signal paths on a switching card. For scanner(or multiplex) cards, the channel is used as a switched input in measuringcircuits or as a switched output in sourcing circuits. For switch cards,each channels signal paths are independent of other channels.

    CHANNEL CROSSTALK. Coupling of a signal from one channel toanother or to the output by stray capacitance, inductive coupling,or radiation. Crosstalk is expressed in decibels at a specied loadimpedance and at a specic frequency.

    CHANNEL ISOLATION. On a switching card, the isolation from signalHI and LO of one channel to signal HI and LO of any other channel (orthe output on switch or scanner cards). Specied as resistance andcapacitance, except for RF cards (decibels and frequency range).

    COAXIAL CABLE.A two-conductor cable that has a center conductorsurrounded by a shield conductor, the two being coaxial and separatedby an insulator.

    COIL RESISTANCE.A nominal value of the resistance of the coilwinding of a relay.

    COLD JUNCTION.The junction in a thermocouple circuit that is at aknown temperature. Also known as Reference Junction.

    COLD SWITCHING. Closing the relay contacts before applying voltageand current and removing voltage and current before opening thecontacts. (Contacts do not make or break current.)

    COLUMN.As viewed on the schematic of a matrix card/module, thevertical signal lines that are connected via relays to the horizontal rows.

    COMMON MODE. Between two signal lines and a third line (e.g., fromsignal HI and LO to chassis ground or guard).

    COMMON MODE ISOLATION. On a switching card/module, theisolation from signal HI and LO to guard (or shield) for a 3 pole circuit,or from signal HI and LO to chassis ground for a two-pole circuit.Specied as resistance and capacitance.

    COMMON MODE REJECTION RATIO.The ability of an instrument toreject interference from a common mode voltage at its input terminals

    with respect to ground. Usually expressed in decibels at a frequency.COMMON MODE VOLTAGE.A voltage between input low and chassisground of an instrument.

    CONTACT BOUNCE.The intermittent and undesired opening of relaycontacts during closure.

    CONTACT LIFE.The maximum number of expected closures beforefailure. Life is dependent on the switched voltage, current, and power.Failure usually occurs when the contact resistance exceeds the end oflife value.

    CONTACT POTENTIAL.A voltage produced between contact terminalsdue to the temperature gradient across the relay contacts, and thereed-to-terminal junctions of dissimilar metals. (The temperature gra-dient is typically caused by the power dissipated by the energized relaycoil.)

    CONTACT RATING.The voltage, current, and power capacities of relaycontacts under specied environmental conditions. See Carry Currentand Switched Current.

    CONTACT RESISTANCE. For a relay, the resistance in ohms acrossclosed contacts. For a Keithley switching card/module, also includesthe tape resistance and connector terminal resistance.

    CONTACTS. The surfaces of current carrying elements where electriccircuits are opened or closed.

    CROSSPOINT.The intersecting point of a column and row in a relaymatrix. Specied as (column, row) or (row, column).

    CROSSTALK.SeeChannel Crosstalk.

    CURRENT SURGE LIMITING.The circuitry necessary to protect relaycontacts from excessive transient current.

    DIFFERENTIAL INPUT ISOLATION.On a switching card, the isolationfrom signal HI to LO. Specied as resistance and capacitance.

    DIFFERENTIAL MULTIPLEXER.A type of switching card/module whereone input is routed to one of several outputs (e.g., sourcing) or oneof several inputs is routed to one output (e.g., measuring), and eachchannel uses a two-pole relay conguration (one pole for signal HI andone pole for signal LO).

    DIGITAL I/O.A TTL-level input/output port that is programmable by themainframe.

    DRY CIRCUIT SWITCHING. Switching below specied levels of voltage(e.g., 20mV) and current to minimize any physical and electricalchanges in the contact junction.

    DRY REED RELAY.A glass-enclosed, hermetically sealed, magneticallyactuated contact. No mercury or other wetting material is used.

    DUT.An abbreviation for Device Under Test.

    ELECTROCHEMICAL EFFECTS.The property that generates currentbetween circuit board conductors due to contamination of the card

    surface. Minimized with proper handling and cleaning.

    ELECTROMECHANICAL RELAY.A relay that uses an electromagnet tomove an armature and make or break the relay contacts.

    EMI.Abbreviation for Electromagnetic Interference. A term thatdenes unwanted electromagnetic radiation from a device that couldinterfere with desired signals in test or communication equipment, forexample. RFI (Radio Frequency Interference) and EMI are often usedinterchangeably.

    FLOATING.The condition where a common mode voltage existsbetween earth ground and the instrument or circuit of interest. (Low ofcircuit is not at earth potential.) The condition where signal LO in thesystem is electrically isolated from earth ground.

    FORM A.A single-pole, normally open contact conguration. Alsocalled SPST-NO. A two-pole conguration is called 2 Form A.

    FORM B.A single-pole, normally closed contact conguration. Alsocalled SPST-NC. A two-pole conguration is called 2 Form B.

    FORM C.A single-pole, double-throw contact conguration. Also

    known as a Transfer Switch. A two-pole conguration is called 2 FormC or DPDT.

    GUARDING.A technique that reduces leakage errors and decreasesresponse time. Consists of a guard conductor driven by a low-impedance source surrounding the lead of a high-impedance signal.The guard voltage is kept at or near the potential of the signal.

    HOT SWITCHING. Opening and closing relay contacts while voltageand current are applied. (Contacts make or break current.) Degradescontact life. Typically used for digital signal switching to prevent statechanges.

    IMPEDANCE MATCHING.To optimize power transfer and minimizemeasurement uncertainty in RF/microwave systems, the ohmic valuesof source, switch, and measure components are equalized, commonlyto 50W. Differences in impedances cause reection of the signal.

    INDEPENDENT SWITCH CARD/MODULE.A type of card where eachchannel is electrically isolated from all other channels. Also calledIsolated Switch Card/Module.

    INPUT ISOLATION. On a switching card, the isolation from signal HI

    to LO (or guard) for a two-pole circuit. Specied as resistance andcapacitance.

    INSERTION LOSS.The attenuation of signals being rswitching card. Specied as a decibel value over a freBecomes more important with low signal levels or hig

    INSULATION RESISTANCE.The ohmic resistance ofdegrades quickly as humidity increases.

    I/O.Abbreviation for input/output, which refers to theinformation to an external device (output), and the rection from an external device (input).

    ISOLATED SWITCH.A type of card where each chanisolated from all other channels. Also called IndependModule.

    ISOLATION. On a switching card/module, the impedany stated terminals. Specied as resistance and capmicrowave switching, the ratio of the power level betwchannels, which is expressed in decibels over a frequInput Isolation.

    ISOTHERMAL BLOCK.The thermally conductive blothe temperature of all thermocouple connections.

    LATCHING RELAY.A relay that maintains its contactposition assumed without the need to keep the coil e

    LEAKAGE CURRENT. Error current that ows througresistance when a voltage is applied.

    LOW NOISE CABLE.A cable that is coated with a cobetween the braid and inner insulator (coax and triaxinner shield (triax). This reduces triboelectric currents due to vibration, movement, or temperature uctuatio

    MAINFRAME.A switching instrument that operates auser commands to connect signals among sourcing instruments and devices under test. Signal switching cards that are inserted into the mainframe rack. A mareferred to as a Scanner, Multiplexer, Matrix, or Progr

    MAKE-BEFORE-BREAK.Connecting a new circuit bdisconnecting the present circuit. Also known as Mak

    MATRIX.Connecting multiple inputs to multiple outp

    MATRIX CARD/MODULE.A type of card with a switcthat has columns and rows of relay crosspoints. Withsince any one point can be connected to any other, yneously have one input with multiple outputs, multiploutput, or multiple inputs with multiple outputs.

    MERCURY WETTED RELAY.A reed relay in which thwetted by a lm of mercury. Usually has a required pooperation; some types are position insensitive.

    MODULE. SeeSwitch Cards/Modules.

    MULTIPLEX.Connecting one instrument to multiple dor multiple instruments to one device under test. Also

    SWITCHING GLOSSARY

    The following terms are dened to clarify or expand on how they relate specically to the material in this handbook.

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    26/33

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W

    26

    A G R E A T E R M E A S U R E O F C O N F I D E N

    MULTIPLEX CARD.See Scanner Card.

    NOISE.An undesirable electrical signal from an external source (suchas an AC power line, motors, generators, transformers, uorescentlights, CRT displays, computers, radio transmitters, and others) that isimposed on a desired signal.

    NON-BLOCKING.A multiplexer or matrix arrangement of relays whereit is possible to switch any signal to any DUT at any time. It has highexibility and high cost.

    NON-LATCHING RELAY.A relay that maintains its closed contactposition when the coil is energized and its open contact position when

    the coil is not energized.NORMALLY CLOSED CONTACTS.A contact pair that is closed whenthe relay is not energized.

    NORMALLY OPEN CONTACTS.A contact pair that is open when therelay is not energized.

    NORMAL MODE. Between two signal lines (e.g., from signal HI to signalLO).

    NORMAL MODE REJECTION RATIO.The ability of an instrument toreject an AC interference (usually of line frequency) across its inputterminals. Usually expressed in decibels at a frequency.

    NORMAL MODE VOLTAGE.A voltage applied between the input highand input low terminals of an instrument.

    OFFSET CURRENT.A current that comes from a switching card eventhough no signals are applied. It comes mostly from the nite coil-to-contact impedance. It is also generated by triboelectric, piezoelectric,and electrochemical effects present on the card.

    PATH ISOLATION. On a matrix switching card, the isolation from signal

    HI and LO of one path to signal HI and LO of any other path. Speciedas resistance and capacitance.

    PATH RESISTANCE. On a matrix switching card, the resistance perconductor of a closed path, including the contact resistance, wiringresistance, and connector terminal resistance.

    PHASE DISTORTION.The shifting of RF/microwave or digitalwaveforms due to signal paths of different lengths and differentpropagation delays through test systems. May cause errors in digitaltesting.

    PIEZOELECTRIC CURRENTS.The current caused by mechanicalstress to certain insulating materials. To minimize, the stress is removedfrom the insulators, and materials with low piezoelectric effect are used.

    POLE.A set of mating contacts in a relay or switch: normally open,normally closed, or both.

    PROPAGATION DELAY.The specied amount of time for a signal to berouted through a previously closed channel of a switching card. Thedelay must be considered, for example, when the switched signal isused to synchronize other signals.

    REED RELAY.A relay using one or more glass-enclosed, hermeticallysealed, magnetically actuated contact members. Some types are dry toimprove isolation or make them position independent. Some types aremercury wetted to improve switched current rating and make contactresistance more predictable. Also see Electromechanical RelayandSolid State Relay.

    REFERENCE CHANNEL. On a thermocouple scanner card, the channelthat measures the temperature of the isothermal block.

    REFERENCE OUTPUT.The output signal representing the temperatureof the reference channel, usually the isothermal block, on somethermocouple scanner cards/modules.

    RELAY.An electrically controlled mechanical device that opens andcloses an electrical contact. A relay provides isolation of control signalsfrom switched signals.

    RELAY DRIVE.The total current available from the power supply of themainframe to energize the relay coils on switching cards. This current isderated at elevated temperatures.

    RELAY DRIVE CURRENT.The amount of current necessary to energizea relay. Usually specied per channel.

    RELAY SETUP.A conguration of open and closed relays.

    RELEASE TIME.The time between the removal of the coil voltage andthe stabilized opening of the contacts.

    RETURN LOSS.A measure of the power reected back towardsthe source due to an impedance mismatch among the source,transmission line, and the load. Expressed in dB.

    ROW. As viewed on the schematic of a matrix card/module, thehorizontal signal lines that are connected via relays to the verticalcolumns.

    SCAN. Sequential connecting (usually break-before-make) of oneinstrument to multiple devices under test or multiple instruments to onedevice under test. Also see Multiplex.

    SCANNER.See Mainframe.

    SCANNER CARD.A type of switching card where one input is routedto one of several outputs (e.g., sourcing), or one of several inputsis routed to one output (e.g., measuring). The actual switching canbe nonsequential (multiplex) or sequential (scan). Also known as aMultiplex Card.

    SETTLE TIME.The time required for establishing relay connections andstabilizing user circuits.

    SHIELDING.A metal enclosure for the circuit being measured or ametal sleeving surrounding wire conductors (coax or triax cable) toreduce electrostatic interference. The shield is usually connected to theLO terminal of the sensitive instrument..

    SINGLE-ENDED MULTIPLEXER.A type of switching card/modulewhere one input is routed to one of several outputs (e.g., sourcing) orone of several inputs is routed to one output (e.g., measuring). Eachchannel uses a single-pole relay for signal HI and a single commonterminal is connected to signal LO for all channels.

    SOLID STATE RELAY.A relay that switches electric circuits by useof semiconductor elements without moving parts or conventionalcontacts.

    SMA.A type of miniature coaxial connector used in situations requiringshielded cable for signal connections.

    SMB.A type of miniature coaxial connector used in situations requiringshielded cable for signal connections.

    SWITCH/MEASURE SYSTEM.An instrument that integrates a switchmainframe and a digital meter into a single chassis. Signal switching isperformed on cards/modules that are inserted into the chassis.

    SWITCHED CURRENT. The maximum current level that can be reliablyhandled while opening and closing contacts. Also see Carry Current.

    SWITCHING CARD/MODULES.The general classication of relay cards.May be subdivided into: independent switch cards (1 input to 1 output),scanner or multiplex cards (1:N, N:1) and matrix cards (M:N).

    T/C COLD JUNCTION. See Cold Junction.

    TEST FIXTURE.An enclosure, sometimes shielded, used for mountingsingle or multiple devices for testing purposes.

    THERMAL OFFSET VOLTAGE. SeeContact Potential.

    TRIAXIAL CABLE.A three-conductor cable that has a centerconductor surrounded by an inner shield conductor that is in turnsurrounded by an outer shiel d conductor.

    TRIBOELECTRIC CURRENT.The current caused by friction betweena conductor and insulator when a cable exes due to vibrations,movement, or temperature uctuation. Also see Low Noise Cable.

    TRIGGER.An external stimulus that initiates one or more instrumentfunctions. Trigger stimuli include: the front panel, an external triggerpulse, and IEEE-488 bus X, Talk, and GET triggers.

    TSP. TEST SCRIPT PROCESSOR.An onboard processor, used ina growing number of Keithley instruments and switch mainframes,designed for executing test scripts within the instrument. UsingTSP test scripts instead of a PC for instrument control avoidscommunication delays between the PC controller and instrument,allowing improved test throughput. Test scripts can contain math anddecision-making rules that further reduce the interaction between ahost PC and the instrument.

    TSP-LINK.The TSP-Link master/slave connection offers easy systemexpansion between Keithleys Series 3700 mainframes. TSP-Link canalso be used to connect to other TSP-Link enabled instruments, suchas Series 2600A System SourceMeter instruments. All instrumentation

    connected via TSP-Link can be controlled by the master unit, just as ifthey were all housed in the same chassis.

    VOLTAGE CLAMPING.The circuitry necessary to protfrom excessive voltage caused by switching current in

    VSWR.Abbreviation for Voltage Standing Wave Ratiosignal reection along a transmission line. Expressedhighest voltage to the lowest voltage found along the

    For further reading on switching terminology, see treferences:

    ANSI/EIA RS-473-1981.Denitions and TerminRelays for Electronic Equipment.American Na

    Institute, 1981.

    ANSI/IEEE Std. 100-1992.IEEE Standard DictiElectrical and Electronics Terms. 5th edition.Electrical and Electronics Engineers, 1992.

    Engineers Relay Handbook. 5th edition. RelayIndustry Association (formerly National AssociaManufacturers), 1996.

    SWITCHING GLOSSARY

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W

    A G U I D E T O S W I T C H C O N S I D E R AT I O N S | B Y S I G N A L T Y P E W W W . K

    http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://gw1.vtrenz.net/?K5CHYUBW35http://www.keithley.com/http://forum.keithley.com/http://www.keithley.com/
  • 7/30/2019 Switching e Handbk Neu

    27/33

    27

    Q U E S T I O N S ? | Q U O T E ? | C O N T A C T U S N O W A G R E A T E R M E A S U R E O F C O N F I D E N

    SWITCH CARD AND SWITCH MODULE SELECTOR GUIDES

    Switching Cards for Series 3700 Mainframes**Additional Series 3700 cards are currently in development. For a current list of cards and specifications, v isit www.keithley.com.

    3720 3721 3722 3723 3724 3730 3731 3732 3740 3750

    No. of Channels 60 (Dual 130) 40 (dual 120) 96 (dual 148)60 (dual 130) or120 single pole

    (dual 160)60 (dual 130) 616 616

    448 crosspoints(Quad 428)

    32

    40 digital I/O, 4counter/