Sl No.
Items Requirement Imperial Corporation Ltd. Khan & Deen Traders IRMECO GmbH, Germany
01Integrity Test types
The below tests must be performed automatically-
1. Forward flow (Diffusive flow) test2. Bubble point test3. Combined forward flow (Diffusive flow) test and Bubble point test4. Water Intrusion test.
Sartocheck Junior BP Plus:
Does not meet the requirement
Palltronic Flowstar (FFS02):
Meets the requirement
Palltronic AquaWIT System (AW02) The equipment can perform the required tests but in semi-automatic way except one which does not meet the requirement. Sartocheck 4:
Meets the requirement
Palltronic Flowstar XC (FFXC):
Meets the requirement
02Measuring Range
1. Forward flow test:0.1 – 200 cc/min
2. Water intrusion test:0.01 – 50 ml/min
3. Bubble point test:700 – 6000 mbar/min
Sartocheck Junior BP Plus:
1. 0- 999 ml/min2. N/A3. Meets the requirement
Palltronic Flowstar (FFS02):
Meets the requirement
Palltronic AquaWIT System (AW02)
1. Meets the requirement
2. 0.05 – 50 ml/min which does not meet the requirement.
3. meets the requirement
Sartocheck 4:
1. no information2. no information3. no information
Palltronic Flowstar XC (FFXC):
Meets the requirement
03Resolution of test
1. Forward flow test:0.1 ml/min
2. Water intrusion test:0.01 ml/min
3. Bubble point test:50 mbar
Sartocheck Junior BP Plus:
No information
Palltronic Flowstar (FFS02):
Meets the requirement
Palltronic AquaWIT System (AW02)
Meets the requirement
Sartocheck 4:
1. no information2. no information3. no information
Palltronic Flowstar XC (FFXC):
Meets the requirement
Page 1 of 3
![Page 2: Technical comprsion (int. tester)](https://reader036.vdocument.in/reader036/viewer/2022082319/543d66e1afaf9f7b0c8b47cb/html5/thumbnails/2.jpg)
Sl No.
Items Requirement Imperial Corporation Ltd. Khan & Deen Traders IRMECO GmbH, Germany
04Test accuracy
1. Forward flow test:+/-3% or +/-0.05%, whichever is greater
2. Water intrusion test: +/-3% or +/- 0.02 ml/min, whichever is greater
Sartocheck Junior BP Plus:
No information
Palltronic Flowstar (FFS02):
No information
Palltronic AquaWIT System (AW02) Meets the requirement
Sartocheck 4: Does not meet the requirement
Palltronic Flowstar XC (FFXC):
Meets the requirement
05 Calibration
1. Forward flow test: Within 3% of measurement
2. Water intrusion test: Within 0.33% of full scale
Sartocheck Junior BP Plus:
No information
Palltronic Flowstar (FFS02):1. Meets the requirement2. does not meet the requirement
Palltronic AquaWIT System (AW02)
Meets the requirementSartocheck 4: 1. no information2. no information
Palltronic Flowstar XC (FFXC):
Meets the requirement
06Pneumatic specification
1. Maximum inlet pressure must be at least 8000 mbar
2. Test pressure range must be 50 – 6000 mbar
Sartocheck Junior BP Plus: 1. Meets the requirement2. Test pressure range: 100-6000 mbar which partially match the requirement.
Palltronic Flowstar (FFS02):
Meets the requirement
Palltronic AquaWIT System (AW02)
Meets the requirement
Sartocheck 4: 1. Meets the requirement2. Test pressure range: 100-8000 mbar which partially match the requirement.
Palltronic Flowstar XC (FFXC):
Meets the requirement
07Pneumatic connection
1. Compressed air inlet: Staubli Nipple type connection2. Compressed air outlet: Staubli coupling type connection3. Vent: Hose connection (dimension must match with that of hose)
Sartocheck Junior BP Plus:
No information
Palltronic Flowstar (FFS02):
Meets the requirement
Palltronic AquaWIT System (AW02)
Meets the requirementSartocheck 4:
No information
Palltronic Flowstar XC (FFXC):
Meets the requirement
08 Printer
Must have a built-in thermal type printer to print test result automatically
Sartocheck Junior BP Plus: Printer is ink-ribbon type and does not meet the requirement.
Palltronic Flowstar (FFS02):
Meets the requirement
Palltronic AquaWIT System (AW02)
No information
Sartocheck 4: Printer is ink-ribbon type and does not meet the requirement.
Palltronic Flowstar XC (FFXC):
Meets the requirement
Page 2 of 3
![Page 3: Technical comprsion (int. tester)](https://reader036.vdocument.in/reader036/viewer/2022082319/543d66e1afaf9f7b0c8b47cb/html5/thumbnails/3.jpg)
Sl No.
Items Requirement Imperial Corporation Ltd. Khan & Deen Traders IRMECO GmbH, Germany
09Operating temperature
5 - 50ºC Sartocheck Junior BP Plus:
No information
Palltronic Flowstar (FFS02):
Meets the requirement
Palltronic AquaWIT System (AW02)
Meets the requirementSartocheck 4: 15 - 35ºC which does not meet the requirement
Palltronic Flowstar XC (FFXC):
Meets the requirement
10 Others
1. The equipment must have option for automatic self test, flow check and air clean.
2. Must have display to show the test progress as bar chart or line graphs.
3. Both stabilization and test time must be automatically controlled.
4. The equipment must be supplied with inlet and outlet hoses.
Sartocheck Junior BP Plus:
1. no information2. no information3. no information4. offered
Palltronic Flowstar (FFS02):
1. No information2. Meets the requirement3. Meets the requirement4. No information
Palltronic AquaWIT System (AW02)
1. Meets the requirement2. Meets the requirement3. No information4. No information
Sartocheck 4:
1. Have cleaning option only.2. Touch screen color TFT display3. no information4. Offered
Palltronic Flowstar XC (FFXC):
Meets the requirement
11 Documentation
The equipment must be supplied with the following documentations-
1. Calibration and test certificate.
2. Material certificate of hose
3. Instruction manual
4. IQ, OQ documents.
Sartocheck Junior BP Plus:
1. offered2. not offered3. offered4. not offered
Palltronic Flowstar (FFS02):
1. not offered2. not offered3. not offered4. Offered
Palltronic AquaWIT System (AW02)
1. No information2. No information3. No information4. Offered
Sartocheck 4:
1. offered2. not offered3. offfered4. Offered
Palltronic Flowstar XC (FFXC):
1. Must be supplied2. Must be supplied3. Must be supplied4. Offered
Comments:
From the above technical evaluation Palltronic Flowstar XC (FFXC) covers the most kinds of integrity tests which meet our requirement. This model is also being used in SVPO and Cephalosporin plant and from our previous experience it is user friendly. Moreover, the calibration option of this model is convenient. So, we recommend Palltronic Flowstar XC (FFXC).