technical developments of in vivo proton magnetic ...€¦ · editing, two-dimensional...
TRANSCRIPT
Technical Developments of in vivo Proton Magnetic Resonance Spectroscopy
by
Karl Landheer
A thesis submitted in conformity with the requirements for the degree of Doctor of Philosophy
Graduate Department of Medical Biophysics University of Toronto
Copyright © 2017 by Karl Landheer
ii
iii
Abstract
Technical Developments of in vivo Proton Magnetic Resonance
Spectroscopy
Karl Landheer
Doctor of Philosophy
Graduate Department of Medical Biophysics
University of Toronto
2017
Proton magnetic resonance spectroscopy (MRS) is a powerful non-invasive technique to probe
the biochemistry of the brain and body. Because MRS signals are collected from metabolites in
concentrations on the order of a few mM in biological tissues, (rather than from water at
approximately 40 M concentration, as in magnetic resonance imaging) low signal-to-noise ratio
(SNR) poses a major challenge. In the face of this challenge, there is an ongoing need for
improved techniques to enhance the clinical applicability of MRS. One area of interest, for
example, involves using MRS data as a biomarker for the assessment of early response in
radiation therapy.
This thesis focuses on the development of three MRS techniques, with the ultimate goal
of this work being used in early radiation detection response protocols. First, a technique is
developed to extend single voxel MRS to MRS of a small number of voxels (eg. two) without the
need for spatial frequency encoding, using customized selective radiofrequency (RF) excitation
iv
and the localized sensitivity of multichannel receiver coils. The SNR efficiency of this technique
is demonstrated in phantoms, healthy adult volunteers and patients with brain cancer. Second,
a novel inversion and saturation recovery sequence is developed in conjunction with a spectral
editing module to estimate the longitudinal relaxation time (T1) of lactate in vivo at 3 T. Lactate
is of significant interest due to its role in cellular metabolism, and particularly its elevation in
tumors. The resulting T1 estimate enables further optimization of MRS protocols and assists in
estimating the absolute concentration of lactate from MRS data. Third, a diffusion-weighted
two-dimensional spectroscopy sequence is developed, subsequently referred to as DW-JPRESS.
This sequence, as well as an optimized processing pipeline, is demonstrated to provide
estimates of the apparent diffusion coefficient (ADC) of brain metabolites beyond those
typically accessible at 3 T, namely glutamate, myo-inositol and scyllo-inositol. The DW-JPRESS
data provide unique information concerning the local diffusion environment of each measured
metabolite, with the potential to characterize microstructural changes from different brain
pathologies including cancer. Collectively, the technical development undertaken in this thesis
promises to enhance future clinical applications of MRS, such as its use in distinguishing
between neoplastic and nonneoplastic lesions, or for assessing tumor recurrence.
v
Acknowledgments
First of all I would like to thank my supervisor, Dr. Simon Graham, for this thesis would not be
possible without his constant guidance, support, and commitment. I would also like to thank
my supervisory committee members, Dr. Charles Cunningham and Dr. Kullervo Hynynen, for
their encouragement and enthusiasm throughout my project, as well as Chuck’s happiness to
discuss anything related to RF pulses.
I would like to thank Jeff Stainsby and Dr. Albert Chen for their help navigating the tricky
world of EPIC. I thank Dr. Arjun Sahgal for being the most enthusiastic clinical collaborator I
could hope for, as well as his record-breaking email response times. Thanks to Fred Tam for his
technical assistance and Rafal Janik for always being willing to explain why what I’m doing is the
wrong way of doing things. Thanks to Justin Lau for his help in understanding the quantum side
of this research. Thanks to all the volunteers, in particular James Mester, Lech Skórski and Philip
Chen who spent several hundred hours lying still inside the magnet without any complaints.
Thanks to Dr. Diana Sima for her help with AQSES, and Dr. Martin Wilson for his unending and
insightful support on the TARQUIN help forums. Thanks to Rolf Schulte and Ben Geraghty for
their help in implementing ProFit.
Thanks to all my friends for their much needed comic relief and Sarah for her love and
support. Additionally I acknowledge the pivotal roles my brother, Alex, and my grandparents,
Ann, Percy David, and Cath have played in my life – without them I would not be where I am
today.
Finally I am deeply grateful for the support of my parents Karen and Dolf, who always
nurtured my curiosity and pushed me to do my best.
vi
Table of Contents
Contents
List of Tables ................................................................................................................................... ix
List of Figures ................................................................................................................................... x
List of Acronyms ............................................................................................................................. xii
Chapter 1 ......................................................................................................................................... 1
1.1 Introduction ........................................................................................................................ 1
1.2 Basic Physics of Nuclear Magnetic Resonance Spectroscopy ............................................. 1
1.2.1 Quantum Mechanics of a Spin ½ Particle in a Magnetic Field................................ 1
1.2.2 Product Operator Formalism .................................................................................. 7
1.2.3 Relaxation ............................................................................................................. 13
1.2.4 Spectral Editing ..................................................................................................... 15
1.2.5 Two-dimensional Spectroscopy ............................................................................ 16
1.2.6 Relationship of the Semi-Classical Formalism to the Quantum Mechanical Formalism ............................................................................................................. 17
1.3 In vivo MRS of the brain .................................................................................................... 20
1.3.1 In vivo MRS acquisition ......................................................................................... 20
1.3.2 Biochemistry of the Brain ..................................................................................... 23
1.3.3 Gliomas ................................................................................................................. 27
1.3.4 Other brain lesions ................................................................................................ 30
1.3.5 Spatial Localization of in vivo MRS ........................................................................ 31
1.4 Parallel Imaging ................................................................................................................. 35
1.5 Absolute Quantitative MRS .............................................................................................. 37
1.6 Diffusion-weighted MRS ................................................................................................... 38
vii
1.7 Hypotheses and Thesis Outline......................................................................................... 41
Chapter 2 ....................................................................................................................................... 43
2.1 Introduction ...................................................................................................................... 43
2.2 Materials and Methods ..................................................................................................... 44
2.3 Results ............................................................................................................................... 51
2.4 Discussion.......................................................................................................................... 59
2.5 Conclusions ....................................................................................................................... 61
Chapter 3 ....................................................................................................................................... 62
3.1 Introduction ...................................................................................................................... 62
3.2 Theory ............................................................................................................................... 64
3.3 Methods ............................................................................................................................ 65
3.4 Results ............................................................................................................................... 72
3.5 Discussion.......................................................................................................................... 78
3.6 Conclusions ....................................................................................................................... 83
Chapter 4 ....................................................................................................................................... 84
4.1 Introduction ...................................................................................................................... 84
4.2 Methods ............................................................................................................................ 86
4.3 Results ............................................................................................................................... 94
4.4 Discussion........................................................................................................................ 100
4.5 Conclusions ..................................................................................................................... 106
Chapter 5 ..................................................................................................................................... 107
5.1 Summary ......................................................................................................................... 107
5.2 Future Directions for CSSMRS ......................................................................................... 109
5.3 Future Directions for DW-JPRESS ................................................................................... 111
5.4 Early Radiation Treatment Response ............................................................................. 114
viii
5.5 Final Remarks .................................................................................................................. 115
Bibliography ................................................................................................................................ 116
ix
List of Tables
Table 2.1: Summary of brain tumor patients for Chapter 2 ......................................................... 50
Table 2.2: Quantified NAA values from PRESS and CSSMRS ........................................................ 57
Table 2.3: Quantified Cho values from PRESS and CSSMRS ......................................................... 57
Table 2.4: Quantified Cr values from PRESS and CSSMRS ............................................................ 58
Table 2.5: Quantified Lac values from PRESS and CSSMRS .......................................................... 58
Table 3.1: Summary of radiofrequency (RF) pulses in the prototype pulse sequence ................ 68
Table 3.2: The TI, TR and total scan time values for all experiments in Chapter 3. ..................... 71
Table 3.3: T1 values measured within the white matter in two healthy volunteers .................... 75
Table 3.4: Estimated T1 values from six patients with high grade glioma .................................... 77
Table 3.5: Estimated absolute concentration of metabolites from six high grade glioma patients
....................................................................................................................................................... 78
Table 4.1: ADC estimates obtained from the “BRAINO” phantom............................................... 96
Table 4.2: ADCs estimated from 6 subjects for 2D and 1D pipelines. .......................................... 98
x
List of Figures
Figure 1.1: Graphical representation of the effect of an RF pulse on spin states .......................... 6
Figure 1.2: The conversion of to with an RF pulse .................................................................. 8
Figure 1.3: Effects of chemical shift and J-coupling on the measured absorption spectrum for
lactate ........................................................................................................................................... 12
Figure 1.4: 2D JPRESS signal as a function of time (horizontal axis) obtained with maximum-echo
sampling ........................................................................................................................................ 18
Figure 1.5: Logarithmically-compressed contour plot of the JPRESS absorption spectrum from
the “Braino” MRS phantom .......................................................................................................... 19
Figure 1.6: Spectrum obtained from parietal brain tissue of a healthy volunteer using PRESS
with TE = 30 ms echo time at 3 T .................................................................................................. 26
Figure 1.7: Pulse sequence diagram for PRESS ............................................................................. 33
Figure 1.8: Spatial localization obtained from PRESS ................................................................... 34
Figure 1.9: The effect of the order of summing and phasing the individual excitations in DW-
MRS ............................................................................................................................................... 40
Figure 2.1: a) Pulse diagram for CSSMRS. b) An anatomical T1-weighted image of a patient with
nominal voxel locations overlaid .................................................................................................. 46
Figure 2.2: Spectra from both a healthy volunteer (a and b) and a brain cancer patient (c and d)
measured with both CSSMRS and PRESS ...................................................................................... 48
Figure 2.3: Spectra from a healthy volunteer at 30-milisecond echo time, obtained by using
both CSSMRS and PRESS ............................................................................................................... 53
Figure 2.4: The unapodized spectrum obtained from CSSMRS from a patient 1along with the fit
....................................................................................................................................................... 53
xi
Figure 2.5: Measured and simulated differences between the CSSMRS and PRESS measurement
for six different voxel separations ................................................................................................ 54
Figure 2.6: Simulated metabolite quantification values for seven different voxel separations .. 56
Figure 3.1: Spectroscopic pulse sequence for measuring T1 relaxation ....................................... 66
Figure 3.2: T1 – weighted anatomical image with voxel placement ............................................. 72
Figure 3.3: Inversion recovery results for “Braino” phantom for the four major metabolites
observed at TE = 144 ms (creatine, lactate, NAA, choline).......................................................... 74
Figure 3.4: Singlet and doublet spectra for two different inversion times from high grade glioma
patient ........................................................................................................................................... 76
Figure 4.1: DW-JPRESS pulse sequence for the a) initial echo time and b) intermediate kth echo
time ............................................................................................................................................... 88
Figure 4.2: Flow chart of the processing steps used to estimate ADCs from the raw DW-JPRESS
data ............................................................................................................................................... 90
Figure 4.3: a) Axial prescription and b) coronal prescription of the DW-JPRESS voxel ................ 95
Figure 4.4: Water-suppressed JPRESS spectra obtained from healthy volunteer for two different
b-values, along with fit and residual obtained from ProFit .......................................................... 97
Figure 4.5: Plot of ADCs estimated from the 2D pipeline versus those estimated from the 1D
pipeline ......................................................................................................................................... 99
Figure 5.1: Four voxel profile overlaid on an anatomical axial MR image ................................. 110
xii
List of Acronyms
Acronym Definition
ADC Apparent Diffusion Coefficient
Ala Alanine
Asc Ascorbic acid (Vitamin C)
Asp Aspartate
AQSES An automated quantitation of short echo time
MRS spectra
BASING Band Selective Inversion With Gradient
Dephasing
CHESS Chemical Shift Selective
Cho Choline
Cr Creatine
CSSMRS Constrained Source Space Magnetic
Resonance Spectroscopy
DW-JPRESS Diffusion-Weighted J-Resolved Spectroscopy
DW-MRS Diffusion-weighted Magnetic Resonance
Spectroscopy
FGRE Fast Gradient Echo
FOV Field of View
xiii
FSPGR Fast Spoiled Gradient Echo
GABA -amintobutyric acid
Gd-DTPA Gadopentetic acid
g-factor Geometry factor
GRAPPA Generalized Autocalibrating Partially Parallel
Acquisitions
Gln Glutamine
Glu Glutamate
Gly Glycine
Glx Glutamine + Glutamate
GPC Glycerophosphorylcholine
GSH Glutathione
HVSD Hankel Singular Value Decomposition
IR Inversion Recovery
JPRESS J-Resolved Spectroscopy
Lac Lactate
MFIR Modified Fast Inversion-Recovery
MRI Magnetic Resonance Imaging
MRS Magnetic Resonance Spectroscopy
xiv
NAA N-Acetylaspartic acid
NAAG N-acetylaspartylglutamate
NMR Nuclear Magnetic Resonance
PCh Phosphorylcholine
PE Phosphorylethanolamine
PRESS Point-Resolved Spectroscopy
ProFit Two-dimensional Prior-Knowledge Fitting
RF Radiofrequency pulse
Scy Scyllo-inositol
SENSE Sensitivity Encoding
SLR Shinnar-Le Roux
SNR Signal-to-Noise Ratio
STEAM Stimulated Echo Acquisition Mode
TARQUIN Totally Automatic Robust Quantitation in NMR
TE Echo Time
TI Inversion Time
tNAA N-acetylaspartate + N-acetylaspartylglutamate
TR Repetition Time
Tau Taurine
1
Chapter 1
1.1 Introduction
This thesis develops several technical advances in the field of in vivo magnetic resonance
spectroscopy (MRS), focusing on applications to the brain. The following introduction provides
an overview of the necessary physics required to motivate the research, and to understand the
research methodology and analysis described in subsequent chapters. The specific introductory
topics of interest are the basic physics of nuclear magnetic resonance spectroscopy, spectral
editing, two-dimensional spectroscopy, in vivo MRS, magnetic resonance parallel imaging,
absolute quantitative spectroscopy and the measurement of the diffusion of brain metabolites.
1.2 Basic Physics of Nuclear Magnetic Resonance Spectroscopy
The basic physics underlying MRS is the logical starting point to understand the work described
in this thesis. This foundation is provided by physics underlying nuclear magnetic resonance
(NMR) for spin ½ particles, which fully applies to all the chemical species of specific interest
within the thesis context. Although the common “semi-classical” formalism used in magnetic
resonance imaging (MRI) is sufficient to describe the spin dynamics of uncoupled spin systems,
many of the chemical species of interest exhibit coupling phenomena. Thus, the full quantum
formalism must be summarized. A section of the introduction is also devoted to translating
between the semi-classical and quantum formalism, as the semi-classical formalism is used for
simplicity within this work when the effects of coupling can be neglected.
1.2.1 Quantum Mechanics of a Spin ½ Particle in a Magnetic Field
Elementary particles contain an intrinsic angular momentum referred to as spin. It is taken here
as an empirical fact that protons have an intrinsic angular momentum of ½ℏ, where ℏ is
Planck’s reduced constant (
6.626 x 10-34 m2 kg/s). The factor ℏ is frequently omitted due to
convenience and is implicitly assumed, thus a proton is hereafter referred to as a spin ½
particle, or in MR jargon as a “spin”.
2
Every NMR experiment measures how a group of spins evolve with time. A spin ½
particle in the presence of an applied magnetic field (by convention chosen to be along the z
direction, also referred to as the longitudinal direction) can occupy one of two stationary
Zeeman eigenstates of energy: one aligned with the field, referred to as the “spin up” state, and
one anti-aligned with the magnetic field, referred to as “spin down” state. These two
eigenstates are conveniently expressed using either the “bra-ket” notation or the vector
notation:
(
)
(1.1)
(
)
(1.2)
where is aligned with the field and is anti-aligned. These two spin states are referred
to as the Zeeman eigenbasis because any spin state can be expressed as a linear combination of
these two. The second value in the bra-ket notation indicates the projection of the spin angular
momentum onto the z-axis, as chosen by convention.
The matrix representations of the three angular momentum operators in the Zeeman
eigenbasis are
ℏ
(
)
(1.3)
ℏ
(
)
(1.4)
ℏ
(
)
(1.5)
It can easily be shown that the application of the z angular momentum operator to the spin
state or yields the eigenvalues ℏ/2 and -ℏ/2, respectively. By the Heisenberg
uncertainty principle, a spin cannot be in simultaneous eigenstates of all three operators, since
the operators do not commute. This is evident as the states or are not eigenvectors
3
of or , which provide the components of angular momentum in the “transverse plane”
orthogonal to the longitudinal direction.
The Hamiltonian for a single spin in the presence of an applied magnetic field is given by1
ℏ (
)
(1.6)
where is the gyromagnetic ratio (267.5 x 106 rad/s/T for protons) and is the magnitude of
the applied magnetic field. Application of this Hamiltonian to the two eigenstates yields the two
energy eigenvalues:
ℏ
(1.7)
ℏ
(1.8)
Thus, the spin-down state is at the higher energy state and the spin-up state is at the lower
energy state. The difference between the two energy levels is ℏ ℏ and spins can
undergo transitions between the two states when this amount of energy is applied in the form
of a radiofrequency (RF) pulse, as explained further below.
Although there are only two eigenstates for a spin ½ particle, a very small number of
spins in a typical NMR experiment are actually in either of the eigenstates due to various time-
dependent processes, such as molecular motion. The vast majority of spins are in a
superposition of spin-up and spin-down states, which can be written using the superposition
principle as
(1.9)
where and are complex numbers called position coefficients. These coefficients must be
normalized such that:
| |
(1.10)
4
The spin state obeys the time-dependent Schrödinger equation1:
( ) ( )
(1.11)
where is the spin Hamiltonian operator, the operator representation of the total energy of
the system. Schrödinger’s time-dependent equation is a first-order ordinary differential
equation with the solution,
( ) ( ) ( ) (1.12)
Equation 1.12 indicates that the spin state at a later time ( ) is completely determined by
knowledge of the initial spin state at time and the Hamiltonian of the system.
Typically the dynamics of spins are expressed in the “rotating reference frame” which
simplifies the solution of the Schrödinger equation in the presence of an applied RF pulse. This
is done by viewing the experiment from a frame that rotates about the -z axis at a chosen
reference angular frequency, . It can then be shown1 that the solution to the Schrödinger
equation in the rotating frame is given by
( ) ( ) ( ) (1.13)
where is the Hamiltonian in the rotating frame and is the state of the spin in the
rotating frame, given by
| ( )| (1.14)
and is the rotation matrix about the z axis in the rotating frame. The Hamiltonian in the
rotating frame, , can be related to the Hamiltonian in the laboratory frame by1
( ) ( ) (1.15)
The convenience of the rotating frame is not immediately clear from Equations 1.13 – 1.15.
However, if a circularly-polarized RF pulse is applied at the chosen reference angular frequency
, using an electromagnetic coil of appropriate geometry such that magnet field
5
components of the pulse rotate in the transverse plane, then the Hamiltonian in the laboratory
frame is given by1
{ ( ) ( )}
(1.16)
where is the amplitude of the applied RF pulse (given in units of Tesla), and is a chosen
phase factor that controls the orientation of the effect of the RF pulse. It can be shown by
applying Equation 1.16 to Equation 1.15 that the effect of the Hamiltonian of the RF pulse can
be written in the rotating frame as
( )
( ){ ( ) ( )}
(1.17)
where is the amplitude of the magnetic field of the applied RF pulse. It can then be shown by
applying Equation 1.17 to Equation 1.13 that the state after application of the RF pulse is
related to the state before application of the RF pulse by
( ) ( ) ( ) (1.18)
where ( ) is the rotation operator with matrix representation
( ) ( ( ) ( ) ( )
( ) ( ) ( ))
(1.19)
Equation 1.19 shows that the effect of an RF pulse is the rotation of the state of the particle in
the rotating frame by the “flip angle” . For all work presented here the phase factor is
arbitrarily set to be zero, thus the effects of the RF pulses are to rotate the spin state around
the x axis. If the phase factor was changed to then the effect of RF pulses would be to
rotate the spin state about the y axis instead. Modern NMR instrumentation provides full
control over the phase of the applied RF pulses, but for all the pulse sequences presented here
there is no such need, thus it is neglected for simplicity. For an amplitude-modulated pulse
( ) is given by
6
∫ ( )
(1.20)
where is the RF pulse duration. A graphical representation of the effects of Equation 1.19 is
shown in Figure 1.1, depicting how common RF pulses act on spins initially in state . A
rotation about the x axis by radians places spins in the state , which is an eigenstate
of the angular momentum operator . A further /2 rotation places spins in the state .
Figure 1.1: Graphical representation of the effect of Equation 1.19 (an RF pulse causes rotation
of spin states, represented by dark arrows, about the x axis in the rotating frame when the
phase equals zero). The initial, spin-up state is ; is an eigenstate of the
momentum operator ; is the spin-down state. x is the rotation operator about the x
axis with brackets indicating the flip angle by which the angular momentum operator is
rotated, according to Equation 1.20.
7
Alternatively, spins in the state can be placed into state by a rotation about the x
axis.
1.2.2 Product Operator Formalism
In a typical in vivo MRS experiment the chemical species are present in concentrations of a few
millimoles, which means the signal from roughly 1020 spins is measured. Two “tricks” have been
developed to circumvent the need to develop an eigenstate that spans all 1020 particles. These
two tricks are referred to as the density matrix and product operator formalism2. The density
matrix formalism is complete and is able to describe the result of any general NMR experiment.
The product operator formalism is only applicable in the “weak-coupling” regime, where the
Hamiltonian is dominated by the Zeeman term. In this scenario, all terms within the
Hamiltonian commute. This approximation is sufficient for this thesis as most biological
molecules are weakly-coupled. The benefit of the product operator formalism is that an implicit
expectation value is taken over many spins. This gives a clear physical meaning to the angular
momentum operators and much of the subsequent quantum mechanics can be ignored,
including the density matrix.
In the following, the terms and refer to the longitudinal, x and y
magnetization components of spin k, respectively. Strictly, spin k is not a single spin but the
average over all magnetically equivalent spins. For example, spin k could represent all the
hydrogens in water or the methyl protons in lactate. The measured signal from spin k in an MRS
experiment is given by (i.e., only magnetization in the transverse plane is measured).
The relevant Hamiltonians can be summarized by a few simple rules explained below. This
formalism offers the ability to predict the outcome of complicated multi-pulse experiments for
weakly-coupled spins with the successive application of simple rules. The effect of RF pulses on
the magnetization terms is identical for the individual spin states, as all individual spin states
experience the same rotation, as shown in Figure 1.2.
8
Figure 1.2: The conversion of to produced by an RF pulse. The net magnetization in these
directions is represented by the grey arrows. Notice that the effect of the RF pulses, ( )
and ( ) is identical for each different spin.
The product operator analog of Equation 1.13 is given by
( ) ( ) ( ) ( ) (1.21)
After application of an RF pulse, the spins are said to be in “free precession”. For a typical in
vivo proton MRS experiment involving N magnetically different types of spins in an isotropic
liquid, this condition is governed by two relevant terms in the Hamiltonian:
∑
∑
(1.22)
where is the spin angular momentum vector for the kth spin; is the scalar J-coupling value
between two different spins, a measured (field-independent) constant; and is the relative
Larmor frequency given by
(1.23)
where , which is dependent on the magnetic field experienced by the particular spin
and is a reference frequency which is under experimental control, typically tuned to water
9
due to its high signal. The first term in Equation 1.22 is referred to as the “chemical shift term”,
resulting from shielding of the applied magnetic field by the electron cloud surrounding the
proton nucleus. This value is by convention measured in reference to the molecule
tetramethylsilane (TMS) and is usually stated in parts per million (ppm). The conversion
between the angular frequency, measured in rad/s, and the chemical shift measured in ppm
is given by
(1.24)
where is the precessional frequency of the protons in TMS. Measuring chemical shift in
ppm is convenient because the magnetic field dependence of the frequency of spins is
removed.
The second term in Equation 1.22 characterizes the “J-coupling” interaction that arises
between neighboring nuclear spins from an indirect coupling that is mediated by the
surrounding electron cloud. Briefly, each spin has its own associated magnetic field which
slightly alters the electromagnetic characteristics of the surrounding electron cloud. This
altered electron cloud then affects the local magnetic field of the coupled spin, slightly changing
its precessional frequency. This mediation through the electron cloud is why J-coupling is also
referred to as indirect coupling. (As an aside, each spin can affect the local magnetic field of a
neighboring spin and is referred to as “direct dipole-dipole coupling”, or simply “dipole-dipole
coupling”. In an isotropic liquid it can be shown that the dipole-dipole effect averages to zero
and therefore does not affect the observed spectrum beyond influencing the rate of
“relaxation” processes1. More will be said about relaxation processes below.)
Using the product operator formalism by applying the Hamiltonian given in Equation
1.22 to Equation 1.21 the freely precessing magnetization of one spin weakly-coupled to
another (in the absence of relaxation), referred to as an “AX system”, is given by1
10
→( ( )( ( ) ( ) ))
( ( )( ( ) ( ) ))
(1.25)
→( ( )( ( ) ( ) ))
( ( )( ( ) ( ) ))
(1.26)
→
(1.27)
with similar transformations for and . Equations 1.25 and 1.26 are generated from
several other assumptions beyond weak spin coupling. The RF pulse duration is considered to
be short in comparison to both longitudinal and transverse relaxation processes. This is a good
approximation for the RF pulses used in this thesis, which have duration of 3-30 ms. For
substantially longer RF pulses, the effects of relaxation must be included for quantitative
measurements. Furthermore, for simplicity when applying the rules of the product operator
formalism, relaxation is often suppressed at the outset then added post-hoc by following the
appropriate echo pathway3 which generated the acquired signal. As can be seen from the time-
evolution of the transverse operators and (Equations 1.25 and 1.26), the terms ( )
and ( ) are to the chemical shift evolution, and terms ( ) and ( ) are due
to J-coupling evolution. The chemical shift evolution is simply an accrual of phase
(interconversion of and ), whereas J-coupling results in an inter-conversion from in-phase
magnetization ( ) to anti-phase magnetization ( , ). The term does not
evolve under chemical shift or J-coupling. It should also be mentioned that the process of spin
echo formation discussed above must be qualified slightly when J-coupled species are
considered. By allowing magnetization to evolve in the transverse plane for time and
then applying a pulse, the resulting magnetization after another evolution time is still
modulated by J-coupling effects whereas chemical shift effects are completely refocused. This is
what allows for J-resolved spectroscopy, as discussed in further detail in Section 1.2.4.
11
The most common in vivo MRS experiments use either point resolved spectroscopy
(PRESS)4 or stimulated echo acquisition mode (STEAM)5 sequences. Both sequences use
identical spatial localization techniques to obtain MRS data from a single coarse volume
element (voxel), but PRESS involves spin echo formation (a pulse followed by two pulses)
whereas STEAM uses a stimulated echo involving three pulses. The PRESS sequence offers
a factor of two improvement in the amplitude of the signal, but STEAM uses shorter RF pulses
(since, for a given max B1 amplitude, a pulse will be of shorter duration as evident from
Equation 1.20) and thus the echo time can be reduced, making STEAM better suited for
chemical species whose signal decays rapidly. Additionally pulses deposit a quarter of the
energy when compared to pulses and have more flexibility in the choice of bandwidth and
slice profile. Due to their inherent similarities STEAM, with suitable modifications, could be
used interchangeably for all the work presented here.
After the MRS signal is acquired in the time domain by PRESS or STEAM sequences, the
spectral information is usually interpreted in the frequency domain after Fourier
transformation of the signal. In the frequency domain, different chemical species are present at
their characteristic chemical shifts in the form of Lorentzian functions. The Lorentzian nature is
due to the monoexponential decay of the transverse magnetization, and deviation from this
monoexponential decay will result in other lineshapes. By taking the Fourier transform of a
measured time-dependent MRS signal with a relative frequency and transverse relaxation
value T2, the following spectral lineshape is obtained:
{ ( ) (
)}
( )
( )
( )
( )
( )
(1.28)
The real portion of Equation 1.28 is a Lorentzian lineshape, and is referred to as the
“absorption” portion of the spectrum, whereas the imaginary portion is referred to as the
“dispersion” portion. Of the two portions, the absorption lineshape has much narrower
linewidth and is conventionally displayed as a consequence. All spectra in this thesis consist of
absorption lineshapes. The measured signal is multiplied by an arbitrary phase factor to display
12
Figure 1.3: Effects of chemical shift and J-coupling on the measured absorption spectrum for
lactate. A molecule of lactate has two chemical groups which give measurable proton MRS
signal: a methine [CH] and a methyl [CH3] group. The methine proton gives rise to a quartet (left
side of the bottom spectrum) because it is coupled to three protons, whereas the methyl group
is a doublet (right side of the bottom spectrum) because it is coupled to a single proton. The
area under the doublet is three times as large as that under the quartet, due to the factor of
three times as many protons from the methyl group than the methane group. The chemical
shift and J-coupling value are not shown to scale, for display purposes. The hydroxyl group does
not contribute to measurable signal at room temperature due to its very fast T2 relaxation.
13
the real portion as the Lorentzian lineshape, which is referred to as “zero-order phasing” the
spectrum.
Considering now the MRS data from multiple weakly-coupled species, the resulting
absorption spectrum will consist of a set of Lorentzians separated by the relative chemical shift,
, of each species, with the Lorentzians split further by the J-coupling effects that occur
between each species. A peak which has no J-coupling is referred to as a “singlet”, and a peak
which is split into two peaks (due to its coupling with a single neighboring spin) is referred to as
a “doublet”. A peak which is split into three peaks (due to coupling with two neighboring spins
from the same nuclei) is referred to as a triplet, etc. The relative amplitudes of the split lines are
given by the binomial distribution. Splitting is additive: for example, if a spin is coupled to two
different nuclei each with a single spin, the resulting splitting would produce a doublet of
doublets (a quartet) with four spectral lines of identical amplitude. As another example, Figure
1.3 shows the effects of chemical shift and J-coupling on the spectrum from lactate.
1.2.3 Relaxation
Both spin-lattice (T1) and spin-spin (T2) relaxation are caused by fluctuations in the magnetic
field experienced by the spins due to thermal molecular motion. The T1 relaxation effect
characterizes the restoration (or “recovery”) of magnetization toward the equilibrium state of
alignment with the applied main magnetic field. This “longitudinal relaxation” occurs because
the spin state aligned with the field minimizes potential energy and after each thermal collision,
a small amount of energy is lost such that the magnetization will very slightly preferentially
align with the field. After very many collisions, the original thermal equilibrium magnetization is
restored. In practice, T1 is an empirically measured parameter on the order of 1000 ms for most
of the biologically relevant molecules considered in this thesis. The other relaxation parameter,
T2 characterizes how each molecular collision affects the phase of the individual spins. Initially
after RF excitation, a net polarization exists indicating that the spins are freely precessing “in
phase” in the transverse plane. After a molecular collision between spins, however, the phase
of the spins is altered. After many collisions, the phase across all spins approaches a uniform
distribution and the transverse magnetization decays (relaxes) to zero. For the biological
chemicals described here, T2 is typically on the order of 100 ms, although the value varies
14
significantly across chemical species. The T1 and T2 values for a group of spins are affected by
numerous physical factors such as the temperature, magnetic field strength, and the
surrounding molecular environment. Some chemical agents, such as the commonly used
contrast agent gadolinium, can drastically shorten the T1 of surrounding molecules due to its
unpaired electrons which cause magnetic field variations that induce relaxation.
The two most common NMR methods to measure longitudinal relaxation are the
Inversion Recovery6 and Saturation Recovery7 sequences of RF pulses. The Inversion Recovery
sequence initiates with an RF pulse of flip angle, referred to as an “inversion pulse” for its
ability to make the equilibrium magnetization point in the opposite direction. The time
between the inversion pulse and a subsequent RF excitation pulse, used to measure or
“read out” the NMR signal, is referred to as the inversion time (TI). The signal is measured at a
variety of TI values and then fit to a mono-exponential model to estimate the T1 value. In a
typical Inversion Recovery experiment, the magnetization is allowed to relax very nearly to
equilibrium between successive measurements at different TI values. Thus, the repetition time
(TR) is usually set to a value much greater than the T1 value of interest. In a Saturation Recovery
experiment, the signal is measured at a variety of different TR values, where the signal is not
allowed to relax to equilibrium in between successive RF excitations, and fit to a mono-
exponential model to estimate T2.
The T2 relaxation is typically measured via a spin-echo pulse sequence7. In this
experiment, the magnetization is excited by a RF excitation pulse and then allowed to relax
(de-phase) in the transverse plane for a duration . A pulse, referred to as a “refocusing
pulse” is then applied to flip the orientation of spins in the transverse plane. As the spins retain
the same precession characteristics after refocusing as they had before (see below for further
clarification), the magnetization re-phases at a time afterwards when a “spin-echo” is
said to have formed. To estimate T2 values, this spin echo sequence can be repeated for many
different values and the measured spin signal amplitudes can then be fit to a
monoexponential model as a function of TE. Alternatively, it is also possible to sample the T2
decay curve by creating multiple spin echoes with a single RF excitation pulse followed by
an appropriately-spaced train of refocusing pulses. These methods are in no way exhaustive,
15
as virtually any pulse sequence which causes magnetization to evolve in the transverse plane
and longitudinal plane will be sensitive to both T1 and T2, and in principle can be used to
perform relaxation measurements. However, the pulse sequences described have the merit
that they enable measurements of a single type of relaxation in a relatively straightforward
manner.
1.2.4 Spectral Editing
One challenge of in vivo MRS is that many of the chemical species of interest have very similar
chemical shift values. Thus, the overlap of spectral lines causes difficulties in accurately
separating the signals and quantifying relative concentrations. One method to overcome this
pitfall is referred to as “spectral editing”, which has been applied to a wide array of different
chemical species. Spectral editing falls into two categories: J-difference editing and multiple
quantum coherence editing. The former is used within this thesis and is subsequently discussed
below, whereas the latter is rarely used for in vivo MRS due to time constraints as well as
hardware limitations. The J-difference editing category exploits the differences in the J-coupling
between metabolites with comparable chemical shift. Initially, this spectral editing technique
used different echo times to produce different evolutions through J-coupling8. For example, it
can be shown that for an AX system (two individual weakly coupled spins) a spin-echo
experiment can be represented by
→ ( )
(1.29)
If the pulse sequence is repeated with twice the evolution time, then
→ ( )
(1.30)
The subtraction of the measured signal from these two experiments yields the magnetization
( ). For spins that are uncoupled, the resulting measured magnetization is (
) for both evolution times, thus subtraction eliminates the signal component from
uncoupled spins. This neglects transverse relaxation effects, however, as the measurement
16
with the longer echo time will have reduced signal. To circumvent this problem, improved
spectral editing techniques have been developed which use frequency-selective RF pulses to
refocus J-coupling effects9. These spectral editing techniques selectively edit one spin thereby
modifying the measured magnetization of the associated coupled spin. Typically two cycles are
repeated, one with the editing pulses off, and one with the editing pulses on. The
measurements are then subtracted to yield the J-coupled spin of interest. The results from
these selective-editing sequences are similar to the original spectral editing method, but are
insensitive to T2 relaxation effects because both cycles have equal echo time. This spectral
editing technique is used in Chapter 3 for the separation of the lactate methyl group and
contaminating lipid signals.
1.2.5 Two-dimensional Spectroscopy
An alternative to spectral editing pulses is two-dimensional (2D) MRS. Similar to spectral editing
techniques, 2D spectroscopy techniques exploit J-coupling to distinguish between overlapping
metabolites10. Used widely in the field of in vitro NMR, 2D MRS offers significantly more
information over 1D spectroscopy, such as molecular connectivities through correlation
spectroscopy (COSY11) and molecular distances through Nuclear Overhauser Effect
spectroscopy (NOESY12). Due to hardware constraints and scanning time, however, only the
most basic of 2D spectroscopy techniques have been applied in vivo13. The 2D spectroscopic
technique referred to as J-resolved spectroscopy (JPRESS) is used within this work due to its
comparatively shorter 2D spectroscopic acquisition time compared to other alternatives13
together with 2D spectral fitting software referred to as ProFit14,15.
The JPRESS experiment can be understood by considering how the magnetization of J-
coupled species evolves with time. From Equation 1.25 and Equation 1.26, J-coupling results in
an interconversion of in-phase and anti-phase magnetization. The resultant magnetization for a
spin-echo experiment with echo time with a reference frequency tuned to the Larmor
frequency of the spin is given by
→ ( ) ( )
(1.31)
17
Thus, the measured signal is modulated by the cosine term ( ) due to J-coupling.
The second term in Equation 1.31, the anti-phase magnetization, is not directly measured but
can be detected by converting it to in-phase magnetization through RF pulses or time-
evolution. By measuring the signal with sequentially increasing echo time, , metabolites that
overlap in chemical shift can be discerned by their J-coupling value. In principle, in vivo JPRESS
uses two pulses for spatial localization as necessitated by PRESS4 but the result of Equation
1.31 still holds. Typically, in vivo JPRESS spectra are acquired with a
maximum echo sampling scheme (data collection begins immediately after the last crusher
gradient, instead of the usual 1D approach of acquiring beginning at the peak of the echo) to
increase available SNR as well as to improve sensitivity by shifting the tails of the absorption
lineshapes off the horizontal axis16. Figure 1.5 displays an example 2D JPRESS signal obtained
over time from a “phantom” test object. The corresponding spectral data after Fourier
Transformation are shown in Figure 1.5. The directly measured frequency dimension is referred
to as (chemical shift) and the indirectly measured frequency dimension (from the sequential
increase in echo time) is referred to as (J-modulation). The benefits of acquiring with a
maximum echo sampling scheme can be readily observed from Figure 1.4 and Figure 1.5, as the
ripples in the time-domain data ensure that the tails of the lineshapes are now oriented at 45o.
If data sampling commenced at the spin echo maximum, then the tails would lie on the F1 = 0
axis, contaminating the measurement of smaller resonances along this line. The tails are
inherently lengthy due to the Lorentzian nature of the absorption spectrum, caused by taking
the Fourier transform of the exponential envelope of the time-dependent signal recorded
during JPRESS.
1.2.6 Relationship of the Semi-Classical Formalism to the Quantum Mechanical Formalism
In the classical formalism, the magnetization is governed by the classical equivalent of the
Schrödinger equation, the Bloch equations:
( ) ( ( ) ( ))
( )
(1.32)
18
( ) ( ( ) ( ))
( )
(1.33)
( ) ( ( ) ( ))
( )
(1.34)
where is the equilibrium magnetization value, and the magnetic field ( )
( ) ( ) . The terms ( ) and ( ) represent the magnetic field of
the RF pulse, and the terms , and are the longitudinal magnetic field gradients in the
Figure 1.4: 2D JPRESS signal as a function of time (horizontal axis) obtained with maximum-
echo sampling. The data were acquired with a total of 100 different echo times (“echo steps”,
vertical axis), each with an acquisition time of 409.6 ms. The Fourier transform of these data is
shown in Figure 1.5.
19
Figure 1.5: Logarithmically-compressed contour plot of the JPRESS absorption spectrum from
the “Braino” MRS phantom. The x axis is the chemical shift frequency dimension (measured in
ppm), and the y axis is the J-modulation frequency (measured in Hz because it is field-
independent).
three orthogonal directions. These gradient terms are critical for in vivo MRS, as they are used
to together with RF pulses to select regions of interest and to de-phase the signal from other
regions4,5, as explained in more detail in Section 1.3.5.
The classical formalism involves simpler equations than the quantum mechanical
formalism and thus is preferential when small effects such as J-coupling can be neglected. For
proton magnetic resonance imaging (MRI) which measures signal almost entirely from the
uncoupled protons of water and fat, or when the applied RF pulse dominates the Hamiltonian
(and thus J-coupling effects can be neglected) the classical formalism is the obvious method of
choice. Solving the Bloch equations reveals that the magnetization behaves identically to the
quantum mechanical formalism in the absence of coupling, namely RF pulses rotate the
magnetization, transverse magnetization (the measured portion of the magnetization) accrues
phase with time and relaxation of the magnetization vector components is identical to
relaxation of the angular momentum operator analogs. It is evident from solving Equations
20
1.32 – 1.34 that the transverse magnetization decays exponentially whereas the longitudinal
magnetization recovers exponentially.
In the semi-classical formalism the detected signal from a coil, ( ), is given by the
equation
( ) ( ( ) ( ( ) ( ) )
(1.35)
where ( ) is the vector field produced by a unit current of one of the fields at the position, .
This is a statement of Faraday’s law combined with the principal of reciprocity. In particular it is
the free-precession behavior (accrual of phase with time) of the magnetization that induces the
signal. For the quantum mechanical picture a similar expression can be developed using the
angular momentum operators instead. In a modern clinical scanner it is typical to receive the
signal using an array of 8 to 64 different channels. Data from these channels are then combined
to give a single measured signal from the sample.
1.3 In vivo MRS of the brain
1.3.1 In vivo MRS acquisition
The previous section described the physics required for in vivo MRS. In this section, applications
of MRS will be investigated focusing on the brain and brain tumors, due to the relevance with
the rest of the thesis. Because of its biological abundance and strong NMR signal, the proton,
also referred to as 1-Hydrogen (1H), is the nucleus of choice for MRS although 31P, and 13C are
also used in research settings17. To reiterate, the chemical environment slightly alters the
magnetic field experienced by the nucleus18 in the form of chemical shift and J-coupling effects,
as shown by the Hamiltonian in Equation 1.22. The measured signal in space and time is then a
superposition of precessing magnetization at several different frequencies for each chemical
species, also referred to as “compounds” or “metabolites”. Spectroscopy can be used to assess
non-invasively the in vivo relative concentrations of these different compounds, providing
additional biological information beyond anatomical MRI. Because these metabolites are found
at concentrations much smaller than water (on the order of a few mM) it is important to
21
remove the water signal so that dynamic range issues do not occur. The most common method
of water removal for in vivo MRS involves a chemical shift selective (CHESS) module19, which
selectively excites water and then dephases it with a large gradient.
The typical processing steps in almost every in vivo MRS experiment are:
Phasing of the data acquired from each coil element prior to summation by the inverse
of the coil sensitivity phase. The coil sensitivity is typically estimated from a water-
unsuppressed measurement, and the lack of this step can drastically reduce SNR if there
is a significant variation in phase across coils.
Water removal via Hankel singular value decomposition. Despite the use of a water
suppression module, there typically remains some residual water signal which can be
several times larger than any of the metabolites. The long tail of the residual water
lineshape can contaminate MRS results, and thus requires this additional processing
step.
Remove eddy currents (as explained below), either using a water-unsuppressed
acquisition20 or analytical methods21.
Zero filling, typically by a factor of two (appending zeros to the end of the measured
signal). This increases the spectral resolution by a factor of two after the measured
signals are Fourier transformed.
Fourier transformation of the signal to obtain the spectrum.
Zero order phasing (applying a single phase factor to the entire spectrum) to separate
the absorption (real) and dispersion (imaginary) line shapes.
First order phasing (applying a linear phase variation across the spectrum) to correct for
how the digital sampling comb used to measure the signal does not coincide precisely
with the spin echo maximum.
22
Reduction of spectral noise by mathematical convolution (optional). Convolution is
typically performed with a Gaussian kernel of 1 or 2 Hz full-width at half-maximum at
the cost of broader linewidths.
There are a variety of commonly encountered artefacts which can cause erroneous
quantification in both 1D and 2D in vivo MRS. Below is a summary of some of the most relevant
artefacts typically encountered, however a more comprehensive examination of all
encountered artefacts is available in the literature22.
Motion artefacts arise due to the movement of the subject either in-between or during
scans. Some degree of motion is unavoidable in in vivo experiments but attempts should
be taken to minimize excessive motion as it results in increased linewidths, improper
voxel localization and decreased water suppression efficacy as well as a reduction in the
peak area due to phase cancellation23. Some degree of motion can be corrected by
aligning peaks prior to averaging the spectra, however motion which happens during a
single acquisition cannot be fully corrected in this manner, as it will lead to signal
cancellation due to gradient dephasing, as explained further in Section 1.6.
Unbalanced crusher gradients can be caused by an improperly tuned final crusher
gradient or due to eddy currents. Unbalanced gradients manifest as a shift of the peak
of the echo, resulting in an alteration of the Lorentzian lineshape of the spectrum. This
can be corrected for post-acquisition by first order phasing, however the underlying
cause may also result in other artefacts.
Chemical shift artefact is caused by the frequency-selective RF localization pulses
exciting slightly different locations for different metabolite peaks. For singlets, this
results only in a positional shift of the voxel, however for coupled metabolites such as
lactate, it results in a positional shift as well as changes in phase and amplitude at the
edges of the voxel (referred to as anomalous J-modulation24). This can be corrected by
using larger bandwidth pulses or by inner volume saturation25, where voxel boundaries
are defined by the edges of saturation pulses instead of through the intersection of
three frequency-selective RF pulses.
23
Spurious echoes are the result of the refocusing of water or fat signals from outside the
voxel. When shimming to minimize the magnetic field homogeneity across a small voxel
a large set of linear shim gradients may be applied, which can shift the water resonance
of regions outside the voxel so that they are no longer captured within the water
suppression band. Because the water concentration is approximately 10,000 times
greater than the metabolites, even if a small fraction of the water signal is unspoiled it
can be comparable to the metabolite amplitudes. Spurious echoes can be effectively
eliminated by customizing the spoiling power and directions for the particular scan26.
Eddy currents are small electrical currents in the sample which are induced by changing
magnetic fields (such as the switching of gradients), inducing their own local magnetic
field fluctuations. As mentioned above, they can cause unbalanced gradient crushers
and they can also induce large phase distortions over the acquisition of the signal. Phase
distortions induced by eddy currents can be corrected, most commonly by using a
water-unsuppressed signal to estimate the phase distortion and deconvolving this
component from the spectra.
1.3.2 Biochemistry of the Brain
The exact roles of all the metabolites measurable by MRS are unknown. However, MRS has
played a critical role in improving understanding of the physiology of many brain metabolites.
This has been achieved due to the high specificity, non-invasiveness and minimal attributable
risk of MRS, permitting measurement of healthy volunteers as well as patients. A summary of
the most commonly measured metabolites is given below, although this is non-exhaustive as
virtually any molecular compound with hydrogen groups with sufficiently long T2 present in
concentrations of ~1mM or greater can in principle be measured by MRS. A more
comprehensive list of 35 metabolites that can be detected with 1H MRS, as well as the chemical
shifts, J-coupling patterns of each metabolite and full in vitro spectra can be obtained from
Govindaraju et al.27
The five brain metabolites most commonly measured by MRS are: N-Acetylaspartic acid
(NAA), myo-inositol (mI), choline (Cho), creatine (Cr) and lactate (Lac). Of these metabolites,
24
NAA exhibits the highest spectral amplitude in a healthy adult brain. This compound is
produced in the mitochondria of neurons and transported into the neural cytoplasm. The NAA
molecule is the second most abundant amino acid in the brain27 (the first being glutamic acid)
and is characterized in MRS by strong resonance at 2.0 ppm. Although its exact function is not
fully understood, NAA is believed to act as an osmolyte as well as a precursor to N-
acetylaspartylglutamate (NAAG), which is believed to be involved in glutamatergic
neurotransmission28. The NAA signal is treated as a biomarker of neuronal density, with a
reduction from normal levels reflecting either neuronal loss or dysfunction27.
Myo-inositol has several prominent resonances at ~3.5 ppm, and is primarily found
within glial cells,29 which are support cells within the brain. Myo-inositol is the most abundant
isomer of inositol and is an osmolyte and precursor to membrane phospho-inositides,
phospholipids and myelin sheet structures30, and plays a key role in signal transduction31. The
exact roles of mI are unknown, although it appears to be a storage form of glucose32. Myo-
inositol has been found to be increased in amyotrophic lateral sclerosis (ALS), with the ratio of
NAA/mI being decreased by 22% (P = 0.001) compared to healthy tissue33.
Creatine has two prominent singlets at 3.0 and 3.9 ppm, and is associated with energy
metabolism through its role in phosphocreatine phosphorylating adenodiphosphate into
adenotriphosphate in times of high energy demand27. Creatine is often used as a control for the
changes in other metabolites measured in MRS, as it does not vary significantly with age34 and a
variety of diseases. However, some diseases can affect Cr levels, such as brain tumors35, which
can confound ratios36.
Choline (Cho) has a strong peak at 3.2 ppm and is the precursor of the active form
phosphorylcholine37 (PCh), which plays a role in cell membrane turnover, and in the production
of acetyl-choline38, an excitatory neurotransmitter that plays a role in memory and learning39.
Typically, glycerophosphorylcholine (GPC) and PCh cannot be distinguished by in vivo MRS, and
the total spectral content is simply referred to as Cho. Choline has been found to have a
significant increase (P < 0.01) in both low and high grade brain tumors as compared to healthy
controls35.
25
Lactate is the end product of anaerobic glycolysis and is increased when normal
metabolism is disrupted40. The most prominent resonance of lactate is from the methyl group
(CH3) J-coupled to the methine group (CH). This J-coupling causes a splitting of the three
protons into a doublet located at 1.3 ppm. The CH group produces a quartet at 4.1 ppm which
is not typically quantified, due to its proximity to the very large resonance from water in
comparison. The CH2 chain of lipids also resonates at approximately 1.3 ppm, thus lactate is
typically measured at an echo time of 1/2J (144 ms) when the doublet is in phase but with
distinctive negative amplitude in relation to lipids. To quantify lactate accurately in the
presence of lipids, spectral editing techniques or 2D spectroscopy must be used. Due to the
very small concentration of lactate in the healthy brain, coupled with the overlap of lipids, this
metabolite is detected at elevated levels only under numerous pathological conditions such as
stroke41, bipolar disorder42, cancer35, among others.
All five of these metabolites are disrupted by a variety of disorders such as brain
tumors43, stroke,41 Alzheimer’s disease44 and Schizophrenia45. Indeed, most brain pathologies
that influence brain biochemistry are likely to result in measurable MRS changes, and the
implications of these changes are an aspect of ongoing research. The MRS focus on these five
metabolites is not due to their greater biological importance than other chemical species within
the brain but due to relative ease of measurement.
Figure 1.6 displays the MRS result obtained from a voxel within parietal white matter of
a healthy volunteer. The four main metabolites are labelled, as well as glutamine/glutamate
(Glx). Lactate is not present due to its very small concentration in a healthy brain.
26
Figure 1.6: Spectrum obtained from parietal brain tissue of a healthy volunteer using PRESS
with TE = 30 ms echo time at 3 T. Metabolites Cr, glutamate/glutamine (Glx), mI, Cr, Cho, NAA
and lipids are labelled.
Beyond these five metabolites (NAA, Cr, Cho, Lac, mI) the remaining metabolites are
more reliably measured at 3 T by employing special spectral techniques, such as editing
(Section 1.2.4) or 2D MRS (Section 1.2.5), due to their relatively low metabolite concentration
and strong spectral overlap with other resonances. Some of these lesser metabolites are
studied in Chapter 4, and are briefly summarized below. Glutamate (Glu) and gamma-
Aminobutyric acid (GABA) are of particular interest due to their roles as the main excitatory and
inhibitory neurotransmitters, respectively46, as well as a variety of other functions32. Glutamate
and GABA typically reside in the synaptic vesicles29,46 within neurons but are released into the
synaptic cleft where they bind to postsynaptic receptors. Alanine (Ala) is a non-essential amino
acid used in protein synthesis and in the transfer of ammonia from astrocytes to neurons in the
glutamate/glutamine cycle47. Ascorbic acid (Asc), more commonly known as Vitamin C, is found
in measurable quantities within the brain and is a vital antioxidant. It also plays a role in
catecholamine synthesis, collagen production and regulation of the protein HIF-1α48. Aspartate
(Asp) is a neurotransmitter that plays a role in the termination of the signals from
27
neurotransmitters at the excitatory synapse49. Glutamine is a precursor of several amino acids
such as glutamate, aspartate and GABA. Glutamine also plays a role in the synthesis of the
messenger molecule nitric oxide (NO) by controlling the supply of the precursor of NO,
arginine50. Glycine (Gly) plays a significant role as one of the other major inhibitory
neurotransmitters51. Glutathione (GSH) is a tripeptide that plays a critical role in disposal of
peroxides by brain cells and in the protection from reactive oxygen species due to the high rate
of oxidative metabolism within the brain52. Glutathione is also an anti-oxidant that is essential
in maintaining healthy red blood cell structure27. Phosphorylethanolamine (PE) is the main
precursor of ethanolamine, which is an essential structural component of cell membranes and
is involved in regulatory roles such as cell division, activation, cell signaling, autophagy and
phagocytosis53. Scyllo-inositol is the second most abundant naturally occurring isomer of
inositol and acts as a main precursor to mI31. Taurine (Tau) is one of the most abundant amino
acids in the brain and has been shown to activate glycine receptors as well as an activator of
extrasynaptic GABAA receptors54.
1.3.3 Gliomas
Gliomas are characterized by uncontrolled growth of a variety of different cell types within the
brain. There are many different types of glioma neoplasms, each with their own respective
biology. The incidence rate within the United States for primary brain tumors is 18.1 per
100,000 person-years, with a 2-, 5-, 10- and 20-year observed survival rates of 62%, 54%, 45%
and 30%, respectively55.
Despite their underlying biological differences the clinical presentation, diagnostic
approach and initial treatment plan are similar across different glioma types56. Initial symptoms
are categorized into two different types: generalized or focal. Generalized symptoms are
usually the results of increased intracranial pressure (due to tumor growth) and include
headaches as well as nausea, vomiting and sixth-nerve palsy in severe cases. The location of the
tumor impacts the presentation focal symptoms which can include hemiparesis (partial
paralysis on one side of the body) or aphasia (speech deficit). Seizures also occur in 15 to 95
percent of patients and may be either focal or generalized56. Neural stem cells, progenitor
cells or de-differentiated mature neural cells can all mutate into gliomas57. Depending on the
28
cellularity, cytonuclear atypia, tumor differentiation, mitotic activity, microvascular
proliferation and degree of necrosis, gliomas can be further subdivided into four grades
established by the World Health Organization (WHO). Grades 1 and 2 are diffuse infiltrating
low-grade gliomas, Grade 3 are anaplastic gliomas and Grade 4 are glioblastomas, with
increasing grade indicating increasing aggressiveness57. Grades 1-3 are considered low-grade,
whereas Grade 4 is considered high grade. Over time Grade 2 and Grade 3 will progress to high-
grade gliomas, which are referred to as secondary glioblastomas. Glioblastomas are the most
deadly and common form of glioma.
Magnetic Resonance Imaging is the most common diagnostic test for patients
presenting with symptoms of brain cancer56. A typical MRI protocol for this application,
depending on the healthcare centre and time permitting, consists of a localizer (T2-weighted
fast spin echo), T2 fluid-attenuated inversion recovery sequence followed by a pre- and post-
contrast (gadolinium) T1-weighted spin echo sequence57, such as the clinical portion of the
protocol used in Chapter 3. Furthermore other magnetic resonance techniques may provide
complimentary information and may also be applied, such as tumor-cell density from diffusion
imaging or proliferation rate of tumor cells as well as necrosis (from the presence of lactate and
lipids) from MRS57. Additionally other medical imaging such as positron emission tomography
(PET) is also sometimes supplemented, especially in patients with presumed low-grade
gliomas56, which are characterized by glucose hypometabolism58.
Although relatively independent of glioma type, as mentioned above, the exact
treatment plan depends on the grade of the glioma. Treatment of direct symptoms involves
steroids to relieve edema, anticonvulsants in patients with seizures and antianxiety or
antidepressants for help with the psychological effects. For gliomas treatment includes surgical
resection of the tumor, as well as radiotherapy (60 Gy to the tumor) with or without the use of
chemotherapy agents such as temozolomide57. For grade 3 gliomas typical treatment includes
maximal possible surgery and radiotherapy (60 Gy). Low grade gliomas account for only about
25 % of diffuse gliomas and present as a non-contrast enhancing lesion on MRI. Typically
disease progression involves slow growth followed by a malignant transformation to a
glioblastoma that is the cause of death around 5 – 15 years after onset57. Retrospective studies
29
have shown that patients with an early and greater extent of resection have postponed
transformation to a secondary glioblastoma and improved survival59. Radiation is also a
standard treatment (50 – 54 Gy), but the optimal timing and delivery is an open topic of
research57, as the tumors are relatively benign for a long period of time.
It is well known that the rate of anaerobic glycolysis is markedly elevated in neoplastic
cells even in the presence of sufficient oxygen, and this phenomenon is referred to as the
Warburg effect60. Because of this, MRS has been used to differentiate between neoplastic and
nonneoplastic lesions due to the differences in metabolism43, and to differentiate between low
and high grade gliomas61. Extensive literature has investigated these prospects with MRS meta-
analysis62 showing the use of MRS to supplement anatomical MRI in the diagnosing of brain
tumors, exhibit sensitivity and specificity of 80 % and 78 %, respectively. Another meta-analysis
observed no observed statistical significance in the accuracy of assessing tumor recurrence
between MRS and PET63. This is an important result as PET is recognized to be sensitive to
radiolabeled compounds at much lower concentrations (~nM) than MRS, which detects
metabolites at ~mM concentration at 3 T.
Compared to healthy tissue, neoplastic lesions have decreased NAA concentration due
to neuronal breakdown, increased Cho due to increased cellular turnover, increased levels of
mI due to increased number of glial cells, a large increase in lactate due to anaerobic glycolysis,
and increased lipids due to necrosis and membrane breakdown43. In addition, MRS provides a
measurement of 2-hydroxyglutarate64 which is only present in glioblastomas that have the
genetic mutations IDH1 or IDH265.Thus, MRS potentially provides useful information as a
diagnostic and prognostic biomarker in this context, as IDH1 and IDH2 confer improved
prognosis when compared to wild-type IDH65.
At present, it is very difficult to evaluate the efficacy of radiation therapy as outcomes
depend on the underlying pathology. Typical treatment responses are observable by solid
tumor changes on anatomical MR images after 6 to 8 weeks66. There are also short term and
long term side effects of radiation therapy. Short term side effects are generally mild and
include vomiting, tinnitus (ringing of the ears), alopecia (loss of hair) and skin changes, resolving
30
after the cessation of treatment. Long term side effects are more detrimental and include brain
necrosis, demyelination, calcifications, hearing loss and, due to the large radiation dose
involved, the potential for new tumors67. It is desirable to monitor response to treatment as
early as possible; in order to change treatment course and, if necessary, in non-responding
cases, avoid unnecessary harmful and expensive treatments. There are numerous techniques
which aim to detect treatment response before anatomical MRI changes. Three of the most
promising are diffusion-weighted MRI66, which measures microstructural changes, and chemical
exchange saturation transfer imaging68 and MRS69, which both measure changes in metabolism.
The application of MRS to early radiation response is currently limited by inherently low SNR
per unit time of the available methods. The aim of this thesis was to develop techniques which
can combat the issue of low SNR, improving clinical MRS, as well as develop new techniques
which could be used for early radiation treatment response. Within Section 5.4 a new study is
proposed which aims to utilize the techniques developed here to provide earlier detection of
treatment response, thereby potentially saving patients from harmful and unnecessary
radiation exposure.
1.3.4 Other brain lesions
Differentiating between tumor recurrence and radiation-induced necrosis is an ongoing issue in
neuro-oncology. Their appearance from diagnostic imaging and clinical symptoms are typically
similar, but their treatment course and outcome is different. Even specialized sequences are
not sufficient in some cases, as typically radiation necrosis is marked by elevated ADC on
diffusion-weighted imaging and low choline from MRS as compared to tumor recurrence but
both techniques yield results with substantial overlap, yielding in some cases false positive or
negatives70. The development of novel non-invasive diagnostic techniques to differentiate
between these two types of lesions is an active field of ongoing research which could have large
ramifications for those undergoing radiation treatment.
Although MRS has been shown to be useful in differentiating neoplastic and
nonneoplastic lesions another ongoing challenge is the differentiation of primary gliomas with
brain metastases, which may require different courses of treatment. Both typically exhibit
decreased NAA/Cr, increased Cho/Cr and a notable increase in lactate and lipids in the 1.3 ppm
31
region71 or an increase in ADC in diffusion-weighted images, as well as similarities in both T1
and T2-weighted anatomical MRI scans. Despite their similarities on currently available
diagnostic tests the prognosis and treatment is often quite dissimilar, thus it is clear that
improved diagnostic techniques could have large ramifications on the treatment of both
patients with metastases and gliomas.
1.3.5 Spatial Localization of in vivo MRS
There are two different categories of MRS: single voxel spectroscopy (SVS), which measures
only one coarse voxel; and magnetic resonance spectroscopic imaging (MRSI), which measures
many spectra simultaneously over a coarse Cartesian grid of voxels. The SVS category will be
explained more thoroughly as it is used preferentially in the thesis. To localize a coarse voxel,
SVS applies three orthogonal gradients during three frequency-selective RF pulses. The effect of
applying a gradient during an RF pulse is most easily understood from the small tip angle
approximation of the Bloch equations. Under this approximation72 and ignoring relaxation
(especially the change in longitudinal magnetization from RF pulses) it can be shown that the
resulting transverse magnetization profile is73
( ) ( )∫ ( ) ( )
(1.36)
This equation shows that the transverse magnetization as a function of z can be expressed as
the Fourier transform of the applied RF pulse waveform when in the presence of a gradient
applied along the z axis. To excite a “slice” of magnetization, mathematically expressed as a rect
function, the appropriate RF pulse waveform is a sinc function. This type of excitation is
desirable in many MRI applications as it simplifies spatial encoding of magnetization, defining
the “through-plane” resolution and requiring subsequent encoding in the remaining two spatial
dimensions within-plane. Equation 1.36 is used in Chapter 2, where two bands of
magnetization (instead of one) are excited by amplitude-modulating the excitation pulse by a
cosine function.
More generally, the relationship between the magnetization profile and the applied RF pulse is
obtainable by solving the Bloch equations using the Shinnar-Le Roux (SLR) transform method,
32
which is applicable in the small-tip as well as large-tip regimes74,75. Other specialized solutions
are possible, such as adiabatic pulses76,77, as used in Chapter 4, and RF pulse design continues
to be an important area of research. Irrespective of the precise nature of the RF pulse, spatial
localization is achieved in MRS through the subsequent application of RF pulses played out in
the presence of three orthogonal gradients, followed by large gradients referred to as
“crushers” which de-phase magnetization and eliminate any echo pathway outside the voxel78.
The area under the crusher gradients is carefully balanced to ensure that the echo pathway of
magnetization within the voxel, which experiences all three RF pulses, remains unaffected.
Figure 1.7 is the pulse sequence diagram for a typical PRESS experiment. In general, this pulse
sequence is repeated many times and the acquired signal from each repetition is averaged to
improve SNR. Figure 1.8 is a diagram depicting how spatial localization is achieved in most in
vivo SVS experiments. A magnetic field gradient, or “gradient”, is depicted as a function of time
and is a linear spatially-varying field term added to the main magnetic field
( ) ( ) (1.37)
where ( ) ( ) is the gradient vector, which is typically depicted by three
separate lines in a pulse sequence diagram, such as in Figure 1.8.
The most commonly used MRSI technique maps out spatial frequency using phase-
encoding gradients (referred to as “k-space”) to obtain spectroscopic data throughout the field
of view (FOV)79,80. Ignoring relaxation, the measured signal for a single coil can be expressed as
( ) ∫ ( ( ) ( )) ( )
(1.38)
where is the k-space vector, ( ) mapped out by the magnetic field gradient function,
i.e.
( ) ∫ ( )
(1.39)
33
( ) ∫ ( )
(1.40)
( ) ∫ ( )
(1.41)
Figure 1.7: Pulse sequence diagram for PRESS. Three orthogonal slices of magnetization are
excited by pulses , and . The crusher scheme dephases all magnetization outside
of the intersection of the three orthogonal slices, allowing the acquisition of a spin echo signal
from a single localized region, as shown in Figure 1.8. The data acquisition (DAQ) begins at the
spin echo maximum. The echo time is when the spin echo is fully rephased, and is equal to
twice the spacing between the two refocusing pulses. In principle the excitation pulse does not
34
need to be 90°, nor the refocusing pulses need not be 180°, however for a single shot this
produces the maximum signal.
Figure 1.8: Spatial localization obtained from the pulse sequence shown in Figure 1.7. Three
orthogonal bands of magnetization are excited and crushers are used to de-phase excited
magnetization outside the voxel. Spatial localization is identical for the STEAM pulse sequence,
except the RF pulses are each and a different set of crushers is used to select the
appropriate echo pathway.
This four-dimensional space can then be mapped out by applying a different combination of
gradients in the three orthogonal directions for each successive excitation (referred to as
“phase-encoding”). The result of taking the 4D Fourier transform of this data is a 4D space
where three of the dimensions are the position in each of the three orthogonal directions (as
prescribed by the phase-encoding gradients) and the fourth dimension is the associated
spectrum at each location. Due to time constraints typically only a two dimensional image is
encoded with k-space and the third spatial direction (usually the axial direction) is localized
using a slice-select RF pulse. Both MRSI and SVS are widely applied in humans81–83 to measure
35
metabolic changes, aid in diagnosis of various cancers and to monitor therapeutic response.
Typical MRSI scan times are long, however, and spatial resolution suffers from an inherently
broad sinc point spread function. The SVS method has improved SNR per unit time compared to
MRSI and when a single localized region is of interest, SVS is usually the preferable option. For
this reason, SVS is used exclusively throughout the thesis.
1.4 Parallel Imaging
Multi-channel receiver coils are an important feature of modern MR systems. Each individual
coil has its own spatial sensitivity for measuring magnetization at a particular position in space.
Each coil can be used to produce its own separate image and the individual images can then be
combined to produce a single image84, with an increase in overall SNR when compared to
imaging with a single large coil. The SNR increase is achieved because each individual coil is
designed with spatial sensitivity for a small fraction of the intended imaging volume, and thus is
only sensitive to noise sources arising from this fraction. In contrast, single channel coils are
designed for sensitivity to the entire imaging volume and all the associated noise sources.
The SNR benefits provided by multi-channel coils can also be used to speed up MRI. A
technique referred to as “parallel imaging” uses the signals from each coil creatively so that the
sampling distance between k-space lines can be increased when traversing k-space. This
decreases the acquisition time (fewer k-space lines are required), but also decreases the FOV
because the Nyquist sampling criterion is no longer satisfied over the entire imaging volume (ie.
high spatial frequency information will now masquerade as low frequency spatial information).
Aliased images with a characteristic “overlapping artifact” are formed if an inverse 2D Fourier
transform is used on undersampled raw k-space data. Several techniques have been developed
to disentangle the aliased images based on multi-channel coil information; the two most
common are generalized autocalibrating partially parallel acquisitions (GRAPPA)85 computed in
k-space, and sensitivity encoding (SENSE)86 computed in image space. These techniques use the
inherent spatially limited sensitivity of the receiver coils to estimate the data for the absent k-
space lines to reconstruct the unaliased image.
36
This thesis adopts SENSE reconstruction, as summarized below. The most common
SENSE image reconstruction is the weak reconstruction with SNR optimization. The unfolding
matrix, , used to disentangle the aliased images, is expressed as
( ) (1.42)
where the superscript indicates Hermitian conjugate, is the coil sensitivity matrix (coil
sensitivity for each voxel), and is the receiver noise covariance matrix (the covariance
between the noise from each of the coils). The disentangled signals, , are then obtained by
(1.43)
where is the measured signal. The SENSE reconstruction results in increased noise because
the coils are not perfectly uncoupled (the coils exhibit correlations between both MR signals
and noise) and because less k-space data are used for spatial encoding, such that
√
(1.44)
where is the SNR of an image reconstructed using SENSE, is the SNR of an
image from fully sampled k-space, g is the geometry factor or “g-factor” of the multi-channel
coils, and is the reduction factor by which k-space measurement is reduced. Using the
definition of the g-factor, the noise amplification due to the condition of the coil sensitivity
matrix86 is
( ) √( ) ( )
(1.45)
where n is the index for the nth reconstructed voxel.
The parallel imaging formalism described above was initially developed for imaging, however it
has since been used to speed up MRSI87,88. Previously, a novel technique for functional MR
imaging (fMRI) of brain activity was also developed, which used SENSE to select a few coarse
voxels89,90, instead of the more traditional method of acquiring an entire brain image. The
37
technique used cosine modulation of the RF excitation pulses (thus exciting two slabs of
magnetization instead of one), and coil sensitivity to reconstruct the signal from the
simultaneously excited voxels. This technique, however, could alternatively be used to speed up
MRS acquisitions instead of fMRI acquisitions. This idea is pursued in Chapter 2 of the thesis.
1.5 Absolute Quantitative MRS
Beyond qualitative analyses of MR spectra, quantitative MRS seeks to extract metabolite values
that are proportional to concentration by fitting the acquired spectra to a basis set containing
the quantum-mechanically simulated signal from all the different metabolites within the
spectrum. Several such packages exist such as LCModel91, jMRUI92, AQSES93, TARQUIN94 and
ProFit14,15. Often the estimated concentration values are expressed as ratios to Cr, as it is
considered the most stable metabolite. It has been previously shown, however, that implicitly
assuming Cr to be stable can confound the quantitative analysis based on ratios36, as Cr levels
can also be strongly affected by disease. Furthermore, the ratios depend on the concentration,
T1 and T2 of both the metabolite of interest and creatine. Thus, changes in the ratio could be
due to a variety of factors, and in some cases the ratio may not change when the underlying
values do. For this reason, it is desirable to estimate the physical concentration of these
metabolites (not just ratios), which is referred to as quantitative MRS.
It can be shown that for a particular chemical species, the magnetization measured with PRESS
is given by
(
)
[ (
) (
) (
)]
(1.46)
where is the equilibrium magnetization for the particular chemical species and TR is the
repetition time. Thus to obtain estimates of , which is proportional to physical
concentration, the measured signal must be multiplied by a correction factor:
38
(
)
[ (
) (
) ( )]
(1.47)
In quantitative MRS the values obtained from the fitting software are corrected for relaxation
and then scaled to represent actual physical concentrations. The two most common ways to
scale to absolute concentration are using an internal35,64 or external95,96 reference. When an
internal reference is used, the measured signal from a certain chemical species (typically water)
is set to be a concentration that is previously known from the scientific literature. When an
external reference is used, the known concentration is typically taken from a phantom that is
measured using MRS before or after in vivo MRS measurements are made. In both cases, the
measured reference signal must also be corrected by Equation 1.47 using the appropriate
relaxation values prior to scaling and obtaining quantitative MRS data.
1.6 Diffusion-weighted MRS
By applying a series of refocused gradients, it is possible to make proton MR signals attenuate
in a manner that is sensitive to diffusion7,97,98. The attenuation is a result of spins diffusing in-
between successive gradients, resulting in a slight phase accrual upon refocusing. This phase
accrual integrated over many spins results in a reduction in the measured signal dependent on
the amplitude of the amplitude ( ) and duration of the applied gradients ( ), the time between
refocusing gradients ( ) as well as the rate of diffusion of the spins. The diffusion
characteristics of spins are highly sensitive to tissue geometry and morphometry99.
Water is the molecule of interest in diffusion-weighted MRI (DW-MRI). The basis of
diffusion-weighted MRS (DW-MRS) is identical to DW-MRI, except that the molecules of
interest are now metabolites (typically NAA, Cr and Cho), and not water. The SNR is thus
substantially decreased for DW-MRS in comparison to DW-MRI, and this has meant that SVS
approaches have been adopted. However, DW-MRS has recently been extended to multiple
voxels using the MRSI technique100 (as explained in Section 1.3.5). The most basic of DW-MRS
experiments involves measuring the spectra twice: once using large applied diffusion-sensitizing
gradients (DSGs) and once using the identical pulse sequence with negligible DSG amplitude.
39
The apparent diffusion coefficient, ADC, of the metabolite can then be estimated through the
equation98
( )
( )
(1.48)
where is the value estimated for a particular metabolite by fitting the measured signal
obtained with negligible DSGs, and is the corresponding value with the DSG amplitude set to
. The spin accrues a phase , in the presence of gradients according to
∫ ( ) ( )
(1.49)
where is the position vector of the spin as a function of time. By Taylor series expanding the
position vector and ignoring the terms above linear motion it can be shown that the phase
accrued due to a bipolar gradient is
( ) (1.50)
where is the angle between the applied gradient direction and the motion and v is the
average velocity of the spin. Using Equation 1.50 and typical values for diffusion-sensitizing
gradients (as given in Chapter 4) a velocity of approximately 1 mm/s in the same direction as
the applied gradient results in a full phase accrual, which is approximately the macroscopic
motion of the head during in vivo experiments. It is therefore paramount to re-phase the
individual excitations prior to averaging due to the interaction with motion and the large DSGs
resulting in large phase variations from one TR to the next101. Figure 1.9 is a diagrammatic
explanation of the need for re-phasing MRS data prior to summing over successive excitations
in a healthy volunteer.
40
Figure 1.9: The effect of the order of summing and phasing the individual excitations in DW-
MRS from in vivo healthy volunteer data with 64 individual excitations. The vertical scales for
each row are constant. The measured amplitude of the metabolites is drastically reduced when
the excitations are summed prior to phasing.
In the presence of non-linear motion, Equation 1.50 is no longer valid and DW-MRS
signals suffer further attenuation artifact. Because of this, cardiac gating is usually necessary in
DW-MRS to limit the effects of non-linear motion from cardiac pulsatility101 in the brain. Failure
to correct for motion by re-phasing and cardiac gating will result in substantially reduced values
when measuring MRS signals at high-b values, producing ADC estimates that are artificially
high.
Diffusion-weighted MRS offers unique intracellular information as the metabolites
exhibit much more restricted diffusion than water, with correspondingly lower ADC values101–
109. The DW-MRS experiment is usually limited to investigating NAA, Cr and Cho, largely due to
the relatively high SNR and ease of measuring these metabolites in relation to the others that
have weaker single strength. Thus, there is a need to improve on the current capabilities of DW-
MRS, so that signals from other metabolites can be measured reliably.
41
1.7 Hypotheses and Thesis Outline
This thesis focuses on the technical development and application of three novel in vivo MRS
pulse sequences. Ultimately these sequences may be applied in a variety of different
applications, although their motivation was ultimately towards using MRS as a predictor for
early treatment response to radiation therapy, as discussed further in Section 5.4.
Chapter 2 tests the hypothesis that by using tailored RF excitation and SENSE parallel
imaging methods, it is possible to obtain high quality MRS data from two voxels simultaneously
without the need for a full k-space encoding procedure, as is typically done in MRSI. This new
approach should offer all the benefits of SVS over MRSI, such as shorter scan times and lack of
point spread function effects on spatial resolution, while allowing the simultaneous
measurement of both voxels. The method is demonstrated in phantoms, healthy controls and
patients with brain cancer. For this chapter I implemented the developed pulse sequence,
performed all experiments on phantoms, healthy volunteers and patients, and analyzed all data
using a combination of custom scripts and a quantitative MRS software package.
Chapter 3 develops a novel inversion recovery sequence which is then combined with
spectral editing to measure the longitudinal relaxation time of lactate in a cohort of glioma
patients. This technique enables T1 to be estimated with improved precision compared to the
use of standard inversion recovery for a fixed experiment time. This novel sequence is then
used to test the hypothesis that lactate has a significantly different T1 relaxation value than the
contaminating lipids in patients with brain cancer. The lactate T1 value is then used to obtain
estimates of absolute metabolite concentration and to optimize the TR value in MRS
experiments involving this metabolite. For this chapter I developed the concept for the pulse
sequence as well as identified the gap in the literature, in addition to writing the pulse
sequence, performing all experiments and analyzing all data using a combination of custom
scripts and a quantitative MRS software package.
Chapter 4 develops a novel technique that combines DW-MRS and JPRESS. This
sequence is then used to investigate what metabolites beyond NAA, Cr and Cho can have their
diffusion coefficients reliably estimated at 3 Tesla in healthy volunteers. For this chapter I
42
developed the concept for the pulse sequence, in addition to writing the pulse sequence,
performing all experiments and analyzing all data using a combination of custom scripts and a
quantitative MRS software package, with the help of Rofl Schulte and Ben Geraghty to
implement the software package.
Lastly, Chapter 5 provides the overall conclusions of the thesis and discusses work that
could be done in the future to extend the research, including technical improvements and
potential applications of the sequences developed here.
43
Chapter 2
Constrained Source Space MR Spectroscopy:
Multiple Voxels, No Gradient Readout
A paper published in American Journal of Neuroradiology, 2015, pp 1-8 by Karl Landheer, Arjun
Sahgal, Sunit Das and Simon J. Graham.
2.1 Introduction
There are two major categories of magnetic resonance spectroscopy (MRS) pulse sequences on
current clinical MRI systems: single voxel spectroscopy (SVS), which measures one voxel; and
magnetic resonance spectroscopic imaging (MRSI), which measures many spectra
simultaneously over a Cartesian grid of voxels. Both SVS and MRSI are widely applied in humans
to detect certain molecular constituents of normal and abnormal tissues, especially those
associated with cellular metabolism, and to monitor therapeutic response81–83. Each MRS
category has its application niche, as SVS and MRSI exploit different spatial and temporal
resolution trade-offs. SVS is attractive when anatomical MRI provides precise indication of
where spectral information should be collected. When pathology is more diffuse, widely
distributed, or not detectable on anatomical MRI, MRSI is the technique of choice to generate
spectra from many voxels using multiple repetitions for k-space encoding79,80. To reduce
spectroscopic scan times various "parallel imaging" approaches have been applied to reduce
the amount of k-space data acquired. These techniques exploit the spatial sensitivity of
individual elements in multi-channel receiver coils87,88,110,111 and can substantially reducing scan
times.
The spatial limitations of SVS are well recognized; it is usually the case that SVS spectra
are required at more than one location, either to compare spectra from diseased and normal
44
tissue, or in the case of multi-focal disease. This naturally leads to execution of SVS pulse
sequences successively for each voxel location. There have been some attempts to modify
spectroscopy acquisition to extend the volume of coverage of SVS, such as line scan echo planar
spectroscopic imaging112 which provides spectra from a column of voxels. However, for clinical
applications, standard SVS methods, notably point resolved spectroscopy (PRESS)4 and
stimulated echo acquisition mode(STEAM)113 remain entrenched.
Previously, a technique was developed that uses RF localization and sensitivity encoding
(SENSE)86 for fast functional magnetic resonance imaging90. It is reasonable that this approach,
appropriately modified for MRS applications, should be investigated more to determine
whether it usefully augments existing SVS capabilities. In the present work, referred to as
constrained source space magnetic resonance spectroscopy (CSSMRS), a prototype pulse
sequence is developed and analyzed for its ability to acquire and separate spectra from two
voxels simultaneously with no k-space encoding. The efficacy of spectral separation is
investigated for a variety of distances between the two voxels in a healthy volunteer.
Additionally, numerical simulations are preformed to assess the validity of certain assumptions
made in the reconstruction and to predict CSSMRS performance in cases where lengthy
experimentation is impractical. Lastly, two-voxel CSSMRS data are reported in relation to
conventional SVS data acquired successively at each voxel location for patients with a variety of
different brain cancers ranging from low grade to high grade.
2.2 Materials and Methods
All experimental data were collected using a GE 750MR 3.0T MRI system (General Electric
Healthcare, Waukesha WI) with a standard 8-channel head coil receiver. To achieve CSSMRS
for proof-of-principle demonstrations, a standard PRESS sequence was modified to excite two
voxels arbitrarily in space, instead of one (Figure 2.1a). Illustrative voxel locations are shown
overlaid on the anatomical image of a patient with brain cancer in Figure 2.1b (see Patient 6 in
Table 2.1 below). The two user inputs were the voxel size, chosen throughout as (20 mm)3; and
the x, y, z coordinates of each voxel location. In this approach, two arbitrarily positioned voxels
were excited via cosine modulation of the first RF pulse, which resulted in the excitation of two
45
parallel slices, followed by the standard spin echo formation process thereafter. Arbitrarily
localization is obtained by modifying the offset frequencies of the RF pulses and changing the
rotation array between logical and physical gradients. The three RF pulses were Shinnar-Le
Roux pulses75 with durations of 3600, 5200 and 5200 ms and bandwidths of 2366.67, 1384.62
and 1384.62 Hz for the first, second and third pulse, respectively.
The additional pulse sequence parameters for this initial work included a repetition time
(TR) and echo time (TE) of 1500 ms and 288 ms (unless otherwise stated), respectively; a flip
angle of 63° (approximately the Ernst angle); a readout bandwidth of 2500 Hz; and 1024 points
data acquisition (total acquisition time of 409.6 ms). The value TE = 288 ms was chosen because
it has been shown to have high MRS reproducibility114, an important clinical factor compared to
the other common TE values of 30 and 144 ms, despite the associated reduction in signal-to-
noise ratio (SNR). Water suppression was implemented using chemical shift selective
saturation19. Prior to all data acquisitions, 1st and 2nd order shimming was applied
encompassing most of the brain to decrease spectral linewidths. The typical linewidth of the
water peak was about 8 Hz. The total number of excitations was 128, with a total scan time of
3.2 minutes. The reconstruction of CSSMRS requires a calibration scan to measure the coil
sensitivity, which is explained below.
Regarding spatial reconstruction of CSSMRS data to separate spectra from the two
voxels, the governing equation can be expressed in matrix form as86,90
( ) ( ) ( ) (2.1)
where the sensitivity matrix relates how the magnetization signals ( ) from each voxel result
in the acquired signals ( ) from each element in the receiver coil, and ( ) represents coil
element-dependent noise. The sensitivity matrix is generated by assuming that the spatial
sensitivity of each coil element varies slowly over the extent of each voxel. For each of the
coils and slices,
46
∑ ∑ ( )
(2.2)
where is the number of pixels within the specified region inside the voxel, and are
the respective minimum and maximum row pixel limits on the voxel and and are the
respective minimum and maximum column pixel limits on the voxel. Equation 2.1 can be
solved by the SENSE formalism using weak reconstruction with SNR optimization86:
Figure 2.1: a) Pulse diagram for CSSMRS. The first RF pulse has a flip angle of α (where α < 90°)
and is cosine-modulated, such that the subsequent spin echo after the third RF pulse excites
two coarse voxels. Shaded gradients are crusher gradients. The slice-select rephasing lobe for
the y gradient is added directly to the first crusher. The gradient echo readout in the dotted box
is optional for voxel localization verification. See text for further details. b) An anatomical T1-
weighted image of patient 6 with the nominal voxel locations overlaid, and a brain tumor
evident in the left middle temporal gyrus. The two spectra for this patient are displayed in the
bottom row of Figure 2.2.
a)
b)
47
( ) ( ) ( ) (2.3)
where is the estimated magnetization signal for each voxel, denotes the pth repetition
and represents the noise covariance matrix between the coils. For example,
∑(
( ) )( ( ) )
(2.4)
where the complex noise samples can be taken from the last datum of each acquisition, and the
covariance is calculated over the total number of excitations,
The matrices can be estimated by various approaches 115–117 although previous CSS
work has shown that a simple procedure is sufficient for proof-of-concept implementation90.
Two sets of fast gradient echo (FGRE) images were acquired with the same pulse sequence
parameters (TE/TR = 1.3/34 ms, flip angle = 5°, field of view = 30 cm, 64 by 64 acquisition
matrix, 5 mm slice thickness): one set with the body coil, and one set with the multi-channel
head coil receiver. These images were then spectrally interpolated to produce 256 by 256
images with an isotropic in-plane resolution of 1.17 mm. For each of the coils and slices, the
coil sensitivity map at each in-plane coordinate, ( ), was estimated by dividing each
of the individual head coil images by the analogous body coil image, and then thresholding
using an “object indicator” to set the coil sensitivity to zero in regions where noise dominates
object signal.
The CSSMRS reconstruction was performed using specially-written scripts in MATLAB
(the Mathworks, Inc., Natick, MA). The two separated signals were first zero-filled by a factor of
two, then transformed to the spectral domain by fast Fourier transformation. The spectra were
then phase-corrected including zero and first order correction terms using an automated
algorithm based on minimizing entropy.118 The spectra were then shifted in frequency to place
the peak for N-Acetylaspartate (NAA) at 2.04 ppm; normalized by their L2-norm; and subjected
to Hankel Lanczos singular value decomposition93 for removal of residual spectral content
arising from water. Spectral components were then quantified automatically using the
48
Figure 2.2: Spectra from both a healthy volunteer (a and b) and a brain cancer patient (c and d)
measured with both CSSMRS and PRESS. Spectra from Patient 6 (See Table 2.1 for list of all
patients) are shown because this patient exhibited the median g-factor, typifying CSSMRS
reconstruction quality. Errors represent the standard deviation over 128 excitations. a.u. =
arbitrary units.
freeware “SPID” which utilizes a separable nonlinear least-squares fitting algorithm known as
automated quantitation of short echo time MRS spectra (AQSES).93 The AQSES algorithm
provides Cramer-Rao lower bound estimates of the standard deviation of each quantified
spectral component. The basis set used was simulated using Java Magnetic Resonance User
Interface (jMRUI) and the input scan parameters. The values obtained from the quantification
algorithm for NAA, Cho and Cr were then scaled by attenuation factors to account for
transverse and longitudinal relaxation effects using relaxation constants obtained in a normal
brain119. The values for lactate were not adjusted for attenuation according to common
practice.
Bloch equation simulations confirmed that cosine modulation had negligible effect on
the integrity of the spatial profile. A water/fat phantom was used to measure the bleed
49
between voxels. One voxel was placed inside a stationary fat container and another voxel was
placed inside the surrounding water bath. Typical scan parameters were used except an echo
time of 40 ms (for increased SNR) and a total number of excitations of 32. This scan was
repeated for centre-to-centre distances of 40.2 to 70.2 mm. The bleed was defined to be the
amplitude of the contaminating absorption spectrum divided by the amplitude of the main
absorption spectrum in the other voxel multiplied by 100%. Two validation experiments were
subsequently conducted on healthy volunteers and patients with brain cancer, to assess
CSSMRS capabilities in practical scenarios. All volunteers participated with free and informed
consent and with the approval of the hospital research ethics board.
Experiment One was performed to investigate how CSSMRS results are affected by voxel
placement in relation to coil sensitivity profiles. Because CSSMRS involves SENSE
reconstruction, overall performance depends on the condition number of the reconstruction
matrix, as quantified by the "g-factor"86:
( ) √( ) ( )
(2.5)
where the integer is used to denote the different voxels that are reconstructed (ie. [ ]
in this case). To assess CSSMRS results for various g-factors, one voxel was placed in a fixed
central location in the brain, and the other was placed to achieve centre-to-centre separations
between voxels varying from 20 mm (ie. adjacent voxels) to 70 mm in the radial direction
toward the head coil. SVS PRESS data were acquired in each successive location for
comparison. These CSSMRS and PRESS data were collected for one healthy young male adult
(23 years old). Equation 2.2 was then used to calculate the sensitivity matrix from the
measured coil sensitivities at each individual voxel location, which, along with the noise
covariance matrix (Equation 2.4) can be used to calculate the g-factor using Equation 2.5.
Experiment Two was performed to investigate how well CSSMRS distinguishes spectra
from cancerous and normal tissue over a representative range of clinical presentations. Six
patients with brain cancer were recruited from the Sunnybrook Odette Cancer Centre during
the course of their treatment (see Table 2.1 for tumor characteristics). Patients were included if
50
they presented with a tumor volume approximately the same size or larger than the prescribed
voxel. Tumor location was verified using a high resolution fast spoiled gradient echo with an
anatomical inversion recovery preparation (FSPGR IR prep, acquisition parameters given
below). For all patients, one voxel was placed at the centre of the tumor and the other was
placed on the contralateral side in the analogous neuroanatomical region within normal-
appearing brain tissue. PRESS data were also acquired successively in these two locations for
comparison purposes.
In both experiments, PRESS was performed with the identical acquisition parameters
used in CSSMRS and with the same spectral analysis pipeline. The total examination time for
comparing CSSMRS and PRESS data from two voxels was approximately 20 minutes, which
included scout images, anatomical MRI (FSPGR IR, 256 by 256 pixels, pixel size = 0.86 mm by
0.86 mm, TR/TE = 8.2/3.2 ms, flip angle = 8°), two FGRE scans (for measuring coil sensitivity as
described above), higher order shim, CSSMRS and PRESS acquisitions.
Table 2.1: Summary of brain tumor patients studied in Experiment Two.
Patient Age (years)
Sex Disease Radiation Treatment Status
Tumor Location Tumor Size (vs Voxel Size)
1 36 F grade II oligodendroglioma
none right cingulate gyrus
Larger
2 84 M grade IV glioblastoma
currently undergoing focused radiation
left middle temporal gyrus
Comparable
3 79 M grade IV glioblastoma
currently undergoing focused radiation
left superior temporal gyrus
Larger
4 61 F brain metastases from breast cancer
60 days since completion of focused radiation
left middle temporal gyrus
Smaller
5 79 M brain metastases from colon cancer
70 days since completion of focused radiation
right superior temporal gyrus
Smaller
6 61 M grade IV glioblastoma
currently undergoing focused radiation
left middle temporal gyrus
Larger
51
A simple numerical simulation was also written in MATLAB for additional insight into the
results of Experiments One and Two. The simulation assessed the impact on spatial
reconstruction of the important assumption underlying Equation 2.1, namely that coil
sensitivity could be reasonably approximated as a constant over each voxel. Given good
agreement between experimental results and simulations for Experiment One (see Results) the
simulation also was used to predict CSSMRS performance under conditions that were not
possible to measure experimentally during Experiment Two, due the inherent time restrictions
for collecting MRS data in patients.
The simulation used measured coil sensitivity data and PRESS data from two voxels as
initial inputs. In the context of the simulation, the PRESS data (obtained according to
experimental parameters given above, averaged over 128 excitations) were considered to
represent a situation in which signal components were uniformly concentrated over each voxel
volume. Simulated signals were then generated for each coil element while accounting for non-
uniform coil sensitivity, by performing the appropriate spatial integral. Complex Gaussian noise
was added to each simulated signal to approximate the levels observed experimentally for each
coil. These simulated coil signals were then used for spatial reconstruction of two voxel signals
according to the Equations 2.1 – 2.4 above, for subsequent comparison with the PRESS data
that were originally input. Spatial reconstruction, spectral processing and analysis were
conducted identically to the procedures outlined above for experimental data.
2.3 Results
For centre-to-centre spacings of 40.2, 50.2, 60.2, 70.2 mm the observed bleed of water into the
fat voxel was 2.0%, 1.4%, 1.5%, 1.7% and 1.3%, respectively, and the observed bleed of fat into
the water voxel was 3.7%, 4.7%, 4.0%, 3.0%, 0.7%, respectively.
For visual comparison, Figure 2.2 displays four representative spectra obtained by
CSSMRS (solid black line) and PRESS (dashed grey line), respectively. As commonly performed
for display purposes, all spectra were apodized by a Gaussian filter with 2 Hz full-width-at-half-
maximum. The spectra shown in Figure 2.2a and Figure 2.2b are qualitatively similar and were
obtained from a healthy volunteer with both voxels placed inside the prefrontal cortex. The
52
spectra shown in Figure 2.2c and Figure 2.2d were obtained from Patient 6 (see Table 2.1) and
are substantially different for the two voxels, with the spectra in Figure 2.2c obtained from
tumor tissue inside the left middle temporal gyrus and those in Figure 2.2d obtained from
contralateral homologous tissue, as shown in Figure 2.1b. Spectra from Patient 6 were chosen
for display in Figure 2.2 because CSSMRS results were obtained in this case with the median g-
factor observed over the patient cohort. Figure 2.3 displays spectra obtained from both
CSSMRS and PRESS for the minimum achievable echo time of this pulse sequence (30 ms).
Figure 2.4 shows the tumor spectrum obtained from CSSMRS for patient 1 and the fit obtained
from AQSES.
The results of Experiment One and related numerical simulations are shown in Figure
2.5, which plots the difference between quantified spectral components measured by CSSMRS
and PRESS for six different voxel separations (one voxel held fixed, one moved radially) and the
three main metabolites observed in Figure 2.2a and Figure 2.2b: NAA, creatine (Cr) and choline
(Cho). The difference values (CSSMRS minus PRESS) reported are specifically for the voxel that
was maintained in a fixed position. For both the experimental and simulated results, the
difference between CSSMRS and PRESS remained constant within error over all voxel
separations. Furthermore, the difference values for experimental and simulation results also
agreed within error, with the only exception being a slight bias in NAA quantification when
voxels were placed adjacent to one another (20 mm separation distance).
Given the good level of agreement between experiment and simulation observed in
Figure 2.5, numerical simulations were then extended to assess CSSMRS reconstruction quality
as a function of voxel separation with spectra that were substantially different in the two
voxels. Figure 2.6 plots the difference between quantified spectral components measured by
CSSMRS and PRESS in a manner analogous to that shown in Figure 2.5, however in this case the
inputs to the simulation were provided from Patient 6 with the difference values relating to
quantification of the tumor spectral components: NAA, Cho, Cr, and lactate (Lac). For
additional context, the difference values obtained experimentally for Patient 6 are also
indicated as single data points in Figure 2.6. Similar to Figure 2.5, Figure 2.6 shows difference
values of zero within error for all voxel separations and metabolites except choline for the first
53
Figure 2.3: Spectra from a healthy volunteer at 30-milisecond echo time, obtained by using
both CSSMRS and PRESS. The labeled metabolites are myo-inositol (mI), Cho, Cr, Glx, and NAA.
Figure 2.4: The unapodized spectrum obtained from CSSMRS from patient 1 (highest g-factor)
along with the automated quantitation of short echo time MR spectroscopy spectra (AQSES) fit.
54
two separations and NAA for adjacent voxels, indicating that good CSSMRS reconstruction
quality is maintained even when the two voxels are located in close proximity to one another.
Additionally the reconstruction tends to improve as the distance between the voxels increases
for all metabolites. The simulation results and experimental results also agree within error for
the single experimental data point.
Figure 2.5: Measured and simulated differences between the CSSMRS and PRESS measurement
for six different voxel separations for the three main metabolites within a healthy adult brain:
N-acetylaspartic acid (NAA), choline (Cho) and creatine (Cr). The signal from the CSSMRS coarse
voxel that was kept in fixed position was reconstructed and compared to the PRESS
measurement obtained from the same location. The black line and gray lines are the measured
and simulated values, respectively. The g-factors are also displayed for reference at the top x-
axis, although there is a non-linear relationship between g-factor and voxel displacement. Error
bars denote Cramer-Rao bounds. a.u. = arbitrary units.
Summarizing the results of Experiment Two, CSSMRS and PRESS results are quantified in Tables
2.2 – 2.5 for Patients 1-6 across tumor and normal tissue voxels for NAA, Cr, Cho, and Lac,
55
including the differences in spectral quantification. The CSSMRS g-factors for Patients 1-6 were
1.67, 1.01, 1.21, 1.23, 1.13 and 1.18, respectively; this indicates that there should be a SNR per
square root of unit time benefit for CSSMRS over PRESS in all cases except for Patient 1.
Overall, large decreases in NAA and increases in Lac and Cho were observed for tumor voxels in
relation to normal tissue voxels for CSSMRS and PRESS for most patients, consistent with
previous papers43. Tables 2.2 – 2.5 also show a large variability in the tumor spectra across
patients. A Mann-Whitney U test on the pooled values from all metabolites obtained from
CSSMRS versus PRESS yielded a p-value of 0.90, indicating no significant difference. It should be
mentioned that the bleed values estimated in a water-fat phantom may not be representative
of those obtained in vivo, due to the differences in linewidths in a phantom versus a human, in
addition to the bleed being estimated from water and fat and not from metabolites. There is no
evidence of significant voxel bleed in the in-vivo experiments, as no systematic increase in Lac
was observed in normal tissue CSSMRS voxels (Table 2.5), except that a large Lac value was
obtained from CSSMRS and PRESS spectra in healthy tissue for Patient 4. Voxel placement was
close to the scalp in this particular patient, which produced contaminating lipid signals that
were subsequently misinterpreted as Lac by the AQSES software. Thus, this specific result
should be discounted. In addition there is a significant increase observed for this patient in
both NAA and Cr in the tumor voxel from CSS. This is likely due to motion which exacerbated
bleed effects, as this particular patient had difficulty remaining still.
56
Figure 2.6: Simulated metabolite quantification values for seven different voxel separations for
the four main metabolites within the tumor spectra for Patient 6: NAA, Cho, Cr, and lactate
(Lac). The quantified values were from the stationary voxel placed within the tumor, and are
plotted in gray. The black data points located at 78 cm in each plot are the experimental results
for this patient, corresponding to the first difference column values listed in Tables 2.2 – 2.5 for
Patient 6. The estimated g-factors are also displayed at the top x-axis for reference, although
there is a non-linear relationship between g-factor and voxel displacement. a.u. = arbitrary
units.
57
Table 2.2: Quantified NAA values from PRESS and CSSMRS for both voxels in arbitrary units
(a.u.). The standard deviations are Cramer-Rao bounds.
Patient CSSMRS tumor voxel
PRESS tumor voxel
Difference CSSMRS healthy
voxel
PRESS healthy
voxel
Difference
1
2
3
4
5
6
Table 2.3: Quantified Cho values from PRESS and CSSMRS for both voxels in arbitrary units
(a.u.). The standard deviations are Cramer-Rao bounds.
Patient CSSMRS tumor voxel
PRESS tumor voxel
Difference CSSMRS healthy
voxel
PRESS healthy
voxel
Difference
1
2
3
5
6
58
Table 2.4: Quantified Cr values from PRESS and CSSMRS for both voxels in arbitrary units (a.u.).
The standard deviations are Cramer-Rao bounds.
Patient CSSMRS tumor voxel
PRESS tumor voxel
Difference CSSMRS healthy
voxel
PRESS healthy
voxel
Difference
1
2
3
4
5
6
Table 2.5: Quantified Lac values from PRESS and CSSMRS for both voxels in arbitrary units
(a.u.). The standard deviations are Cramer-Rao bounds.
Patient CSSMRS tumor voxel
PRESS Tumor voxel
Difference CSSMRS healthy
voxel
PRESS healthy
voxel
Difference
1
2
3
4
5
6
* Lipid contamination from scalp mislabelled as Lac in healthy voxel.
59
2.4 Discussion
This work has introduced a prototype pulse sequence for CSSMRS, a novel spectroscopy
technique that measures spectra from multiple voxels simultaneously without the need for k-
space encoding. Instead, spatial encoding is achieved by multi-voxel RF selective excitation,
signal readouts from a multi-channel receiver coil, and SENSE14 reconstruction to separate the
signals from each voxel. The CSSMRS method is important from the perspective of SNR per
square root of acquisition time, potentially providing efficiency in comparison to the standard
clinical practice of performing successive SVS acquisitions at different voxel locations.
Careful experiments and simulations were undertaken to investigate the capabilities of
CSSMRS for simultaneous measurement of two voxels. In particular, considerable attention was
paid to whether CSSMRS provides adequate spatial localization in relation to the standard SVS
PRESS method. In a water/fat experiment it was shown that the bleed was on the order of 2-
5%, which is acceptable for spectroscopic applications. Experiments One and Two, conducted
in healthy volunteers and a diverse group of six brain tumor patients with four different types
of cancer, showed overall that CSSMRS and successive PRESS spectra agreed within
experimental error. Furthermore, CSSMRS spatial reconstruction was shown to be robust over
a range of voxel prescriptions (with one voxel held fixed and the voxel separation varied), by
both experiment and numerical simulation. Experiment and simulation were in agreement for
a healthy volunteer, indicating excellent reconstruction even when the two voxels were placed
adjacent to one another. The only additional feature of note in this regard was a slight,
systematic discrepancy between the simulated and measured NAA values observed in Figure
2.5 for all voxel separations. This feature is likely due to the relatively simplistic nature of the
simulations, which did not account for various experimental factors. However, given that the
overall level of agreement between experiment and simulation was very good, these factors
evidently have a small influence. The simulation therefore helps to support the assumption
made in CSSMRS reconstruction that coil sensitivity variations can be neglected within the
voxels.
60
The agreement between these experiments and the simulation provided rationale for
using the simulation further to predict CSSMRS capabilities in a brain cancer patient. As
expected, slightly larger variations were observed as a function of voxel separation in this case,
likely due to the larger spectral differences between the two voxels. However, with the
exception of choline quantification for very closely spaced voxels (20 and 30 mm centre-to-
centre) and NAA with adjacent voxels, all CSSMRS results were predicted to be consistent with
PRESS results within error.
Given that CCSMRS has been demonstrated to provide robust, high-quality results,
discussion can turn productively to the potential efficiency of this pulse sequence in terms of
SNR per square root of acquisition time. In the two-voxel implementation investigated in the
present work, spectra were obtained in half the time compared to successive application of
PRESS. The quality of CSSMRS results is potentially affected by noise amplification in the SENSE
reconstruction, however, as parameterized by the g-factor. Therefore, the appropriate context
for using CSSMRS advantageously over PRESS is when the g-factor is less than √ . This
corresponds to a minimum centre-to-centre separation in voxels of about 55 mm near the
centre of the 8-channel head receiver coil used in this work. All but Patient 1 (who had a tumor
approaching the midline) had a g-factor below this threshold.
It is also interesting to note that CSSMRS is compatible with another approach that
avoids using k-space for spatially encoding spectral information. In principal, if the flip angles
assigned to each voxel can be modulated appropriately, then simple algebraic combinations of
the successive spectroscopic readouts can be used to localize each voxel without SENSE
reconstruction, as achieved in Hadamard Spectroscopic Imaging (HSI)24. The HSI approach is
independent of g-factor and also provides improvements in SNR per square root of time, but
has traditionally required excellent RF fidelity and is sensitive to how spatial RF nonuniformity
and patient motion influence algebraic combination and the subsequent leakage of signals
between voxels. In addition, the algebraic combination of multiple recordings reduces the
minimum temporal resolution that is achievable with HSI, whereas CSSMRS provides spectral
separation in as little as a single TR value. CSSMRS and HSI are not mutually exclusive,
61
however, and it is possible that a robust, hybrid technique can be developed in the future for
further scan time reductions.
Irrespective of developing such a hybrid technique, the present method has potential
applications in any in vivo spectroscopy experiment in which there are two regions are of
interest and the lengthy acquisition times of MRSI are impractical. CSSMRS could also be
beneficial in a research setting where sophisticated 2-dimensional MRS experiments have
inherently long acquisition times, such as JPRESS120. Another promising application of CSSMRS is
in functional spectroscopy where real-time changes in metabolic information could be
measured from multiple points within the brain simultaneously with high temporal resolution.
Further development and applications of CSSMRS will be explored in the future.
2.5 Conclusions
CSSMRS has been developed to extract signals from two localized regions
simultaneously and reliably. Utility was demonstrated in a clinical setting, although the
technique has promising applications in the research setting as well.
62
Chapter 3
A Rapid Inversion Technique for Measurement of
Longitudinal Relaxation Times of Brain Metabolites:
Application to Lactate in High Grade Gliomas at 3 T
A paper published in NMR in Biomedicine, 2016, vol. 29, pp 1381-1390 by Karl Landheer, Arjun
Sahgal, Sten Myrehaug, Albert P. Chen, Charles H. Cunningham and Simon J. Graham.
3.1 Introduction
Proton magnetic resonance spectroscopy (MRS) is a powerful non-invasive technique used to
measure biomarker activity within the brain and body. This technique has been used
extensively to investigate the biochemical profiles of brain tumors121. Typically, gliomas exhibit
a decrease in N-acetylaspartic acid (NAA) due to the degradation of neurons, and an increase in
choline (Cho) due to elevated cell density and membrane turnover in neoplasms. Especially
when necrosis is present, there is also often an accompanying increase in lipids. Lastly, a
significant increase in lactate is common, widely attributed to the increase in anaerobic
glycolysis40. Lactate is of particular interest due to its role in metabolism and its negative
correlation with survival time43.
For MRS results to be interpreted in detail, the magnetic resonance properties of each
spectral component must be well understood. One important property is the relaxation
time, which describes the timescale for longitudinal recovery of magnetization after resonant
excitation. The value depends on the static magnetic field strength and indirectly reflects
molecular dynamics within tissue microstructure. From an experimental standpoint, the
value is important for determining the optimal repetition time (TR) between spectral
acquisitions so that signal-to-noise ratio (SNR) is maximized (eg. according to the efficiency
63
metric √ ). Together with knowledge of the relaxation time, the value also
enables correction of spectral components to estimate absolute concentrations.
The values for the major spectral components of brain tumors have been well
studied at 3 T122, the preferable field strength for clinical MRS in terms of SNR and spectral
resolution. Lactate is the notable exception, however. Despite the importance of lactate as a
biomarker of tumor aggression, measurements of lactate are challenging because of three
inter-related factors. First, lactate is not typically detectable by MRS in normal brain,
necessitating that dedicated efforts must be made to measure values in patients prior to or
during their cancer treatment. Second, the most common methods for measurements are
time-consuming (see below). Third, the measurements of the lactate methyl [CH3] doublet
(coupled to a methine [CH] proton) are confounded by spectral overlap from lipids. To our
knowledge, the only pertinent human MRS data were acquired at 1.5 T without accounting for
the overlap123.
The present work was conceived to address these challenges and fill the gap in the
existing literature through dedicated study of brain cancer patients. Regarding the method for
measurement, the two most common choices are progressive saturation recovery124 and
inversion recovery (IR)125. When naively applied, both are inherently slow and suboptimal for
measuring low concentrations of in vivo brain metabolites in a time efficient manner in
patients, because they conventionally require use of a TR value that is multiples of the T1 value
of interest. More rapid methods such as the Look-Locker approach126 are difficult to combine
with the required spatial localization and spectral editing schemes. Alternatively, the modified
fast inversion-recovery (MFIR) method enables T1 measurements without full longitudinal
recovery127. This approach has been modified, characterized and validated appropriately for the
present spectroscopic application, allowing flexible choice of TR and a simple fitting approach
to estimate T1 values. In addition, the lactate and lipid signals are separated using the
radiofrequency (RF) band selective inversion with gradient dephasing (BASING)128 lactate-
editing sequence. The BASING sequence also provides lactate refocusing129, offsetting the
“anomalous” J-modulation24 and substantial signal reductions at odd multiples of 1/J that are
commonly observed using point resolved spectroscopy (PRESS)4. The anomalous J-modulation
64
of lactate is the result of the excited magnetization of the quartet and the doublet being slightly
shifted in space relative to each other. This shift results in voxel boundaries where both coupled
groups are not affected by both the refocusing pulses, resulting in a change in the phase of the
measured doublet signal. This results in some signal cancellation at the voxel boundaries,
thereby reducing the overall amplitude of the doublet signal. This approach also elevates the
available signal-to-noise ratio (SNR) by enabling robust lactate data collection at an echo time
of 144 ms (the most common clinical choice of long echo time) rather than 288 ms.
In summary, the present work addresses two aims. The first aim is to develop and
validate a novel MRS pulse sequence for measuring T1 values of metabolites in a time efficient
manner, suitable for use in patient populations and when J-coupled metabolites are present in
low concentration. The second aim is to use the developed pulse sequence to report the T1
value of lactate at 3 T in patients with brain cancer. In particular we measure the of the
lactate doublet averaged over all microenvironments, as it is typically what is of interest in
clinical MRS. These values, taken together with values from the literature as appropriate, are
then used to provide a quantitative estimate of the brain lactate concentration in vivo.
3.2 Theory
A diagram of the prototype pulse sequence for spectroscopic T1 measurement is shown in
Figure 3.1. The main elements consist of an inversion pulse followed some time later (as
selected by the inversion time, ) by point resolved spectroscopy4 (PRESS) localization with
spectroscopic readout. However, rather than performing acquisitions for multiple values in
sequence, as in IR experiments, paired acquisitions are performed at each value. One
acquisition includes the inversion pulse, producing the signal ( ). The other acquisition is
performed without the inversion pulse, producing the signal ( ). When the paired signals are
subtracted to yield a difference signal, it can be shown by solving the Bloch equations that
if TR is chosen to track with (i.e., , where C is a constant) then can be
modeled by a simple two-parameter, monoexponential decay function involving the T1
relaxation time constant. The optimal choice of TI and TR values (given practical constraints,
such as the total time allowed for MRS measurement) can then be determined using
65
computational methods to estimate the T1 value and minimize the uncertainty of the estimate.
This overall approach allows measurements with comparatively small TR values without the
need for a third parameter in the monoexponential model. The approach also provides time
efficiency by removing the need for all measurements to be made with large TR values for full
longitudinal recovery of magnetization, as adopted in conventional IR, and removes the need
for a fourth TI point as in MFIR130. A similar pulse sequence approach for T1 measurement has
been used recently, although without thoroughly validating optimal TI and TR selection as well
as time efficiency in relation to other measurement methods131. These issues are addressed as
part of the experimental methods outlined below.
3.3 Methods
Spectroscopic data were collected using a GE 750MR 3.0T MRI system (General Electric
Healthcare, Waukesha WI) with a standard 8-channel head coil receiver. Considering the
prototype pulse sequence (Figure 3.1) in more detail, the standard PRESS sequence was
modified to include four additional elements: 1) a frequency-selective (not spatially selective)
hyperbolic secant inversion pulse was added prior to the chemical shift selective saturation19
water suppression; 2) BASING RF pulses132 were added after the first and second refocusing
pulses, using linear-phase Shinnar-Le Roux75 (SLR) design with minimal transition width, to
reduce unwanted coherences and reduce co-editing of metabolites other than lactate; 3)
custom SLR 180° refocusing pulses were used to increase SNR and further reduce unwanted
coherences; and 4) the crusher gradient scheme was modified to include bipolar gradients
which selected the proper echo pathway for BASING. BASING is a two-cycle technique which
refocuses the lactate doublet (by inverting the lactate quartet) on the first cycle, and leaves the
lactate doublet unaffected on the second cycle. Subtraction of these two cycles results in the
addition of the lactate doublet, whereas the overlapping lipid signals are theoretically
unaffected by the cycling scheme and thus cancelled. Table 3.1 provides the details of all the RF
pulses within the sequence. The pulse sequence uses a four cycle scheme: two for the
separation of coupled and uncoupled resonances via BASING, and two for the interleaving of
the inversion pulse. For the first and second cycle, the inversion pulse is on, and for the third
and fourth cycle the inversion pulse amplitude is set to zero. For the first and third cycle, the
66
Figure 3.1: Spectroscopic pulse sequence for measuring T1 relaxation. is the inversion pulse
and TI is inversion time. A subsequent PRESS module is characterized by the time between
the isoflip time of the excitation pulse ( ) and the middle of the first refocusing pulse ( );
the time between the middle of and the second refocusing pulse ( ); and , the
time between the middle of and the start of the acquisition time. is equal to
the echo time (TE), 144 ms. The time between the first BASING pulse ( ) and the second
BASING pulse ( ) is TE/2 (72 ms) for this editing scheme.
BASING pulses ( and in Figure 3.1) are centred at water and invert the lactate quartet
at 4.1 ppm. For the second and fourth cycles, the BASING pulses are shifted downfield by 198
Hz to not invert the metabolites. Thus the uncoupled singlet signal, , is obtained from
( ) (3.1)
where the subscript on denotes the cycle number of the measured signal. The coupled
doublet signal, , is obtained from
67
( ) (3.2)
This four cycle scheme is then repeated as necessary for signal averaging. For the work
presented here, 128 excitations were performed for each value. Thus, the four cycle scheme
was averaged over 32 trials in each case.
The signal processing was performed using specially-written scripts in MATLAB (the
Mathworks, Inc., Natick, MA). Prior to reconstruction, the first point in the raw FID of the
unsuppressed water acquisition was used to estimate the coil sensitivity, which was then used
to rephase and scale the signals from each of the individual coils. By scaling and phasing prior to
summation over the eight coils, SNR is substantially improved compared to direct summation.
The net signals were then combined over the four-cycle scheme according to Equation 3.1 for
singlets and Equation 3.2 for lactate, averaged over all repetitions as appropriate, and input to
the freeware known as Totally Automatic Robust quantitation in NMR (TARQUIN94). Algorithms
within TARQUIN enabled fitting of the data to the four metabolites of interest: NAA, lactate,
creatine and choline. Because the BASING pulses have some effect on some downfield smaller
peaks, a more physically realistic basis set was used by taking the default basis set in TARQUIN
and removing the smaller resonances.
The concentration values obtained from TARQUIN were then used to estimate values. For
computational simplicity, the monoexponential equation for was linearized by taking the
natural logarithm and weighting the noise contribution at potential TI values appropriately.
This enabled use of linear least squares fitting to estimate T1, and optimization of TI and TR
values using a closed-form expression to minimize the uncertainty of the estimate. White
Gaussian noise was assumed and standard error propagation was used to derive the expression
for estimating the standard deviation of , . To estimate reliably while accounting for
imperfect model fitting or spurious artefacts in the experimental data which were not well
represented by the standard deviation of the spectral components, the calculation was
performed using both the Cramer-Rao lower bounds from TARQUIN, and from the standard
deviation of the residuals of the least squares fit. The larger of the two estimates was reported.
68
The unsuppressed water acquisition was then used to provide an absolute reference, setting
the concentration of water at 42.3 M, as is common when attempting to quantify metabolite
Table 3.1: Summary of radiofrequency (RF) pulses in the prototype pulse sequence (Figure 3.1).
Pulses and are centred on water for the first and second cycles of the four-cycle
scheme (and invert the lactate quartet) and for the third and fourth cycles they are shifted
downfield from water to leave the metabolites unaffected. Pulse inverts all the metabolites
for cycles 1 and 2 but is set to zero for cycles 3 and 4. Shinnar Le-Roux75 (SLR) pulses are used
for all but the inversion pulse, and BASING128 pulses are the minimum-phase SLR pulses used
for the lactate-editing sequence. B1 is the amplitude of the applied RF pulses.
Name Pulse type Pulse duration (ms)
Bandwidth (Hz)
Offset frequency (Hz)
Max B1 ( )
Custom adiabatic hyperbolic secant
10.0 1200 -256 16.01/0
Stock excitation SLR
3.6 2367 -256 14.22
Custom spin-echo SLR
9.5 842 -256 17.66
Custom minimum-phase BASING SLR
30.0 230 0/198 5.74
Custom spin-echo SLR
9.5 842 -256 17.66
Custom minimum-phase BASING SLR
30.0 230 0/198 5.74
concentrations from MRS data64. All concentrations were obtained using water as an internal
reference, as implemented in TARQUIN, similar to the procedure by Madan et al.35 The water
unsuppressed signal had its concentration set to be 42.3 M, with a water attenuation factor of
exp(-TE/T2water), where the T2 of water was 56 ms for healthy white matter or 156 ms for glioma
tissue35. For each metabolite the straight line intercept of the concentration vs TI plot was used
to correct for inversion effects. This value was then corrected for non-equilibrium steady-state
69
T1 effects and T2 relaxation using T2 values obtained from the literature for lactate35 and the
other metabolites122. Statistical comparison between the T1 of lactate and the other three
metabolites was subsequently performed using a Wilcoxon signed-rank test, with the threshold
for statistical significance set according to a Type 1 error of .
The additional pulse sequence parameters for this initial work included an echo time
(TE) of 144 ms (the TE value providing the maximum SNR achievable with this editing
scheme128) and a voxel size of 20 mm 20 mm 20 mm. Water suppression was
implemented using the standard CHESS19 method immediately preceding the excitation pulse
for all inversion times.
The uncertainty on T1 derived from the least squares fit was numerically minimized to
yield the optimal sets of TIs and TRs. The optimization was subject to three physical constraints:
(the minimum allowable inversion time to perform CHESS prior to the excitation
pulse); TR exceeds the time from pulse sequence onset to the end of readout;
(necessary for the monoexponential model to hold); and a fixed total experiment time (chosen
for in vivo work to be 30 minutes, ignoring calibration time). The number of individual
measurements (i.e., TI points within the 30 minute measurement time) was allowed to vary
from 3 to 8. Numerical minimization of the expected uncertainty on T1 was achieved using the
iterative sequential quadratic programming algorithm in MATLAB that required a priori
knowledge of the value of interest. The only inputs required to optimize the choice of TI and
TR values were the estimated T1 and the total experiment time. The algorithm commenced with
a starting value of , the reported of rat glioma measured at 4.7 T60, to
approximate the expected human analogue.
Briefly summarizing the subsequent experiments, tests were first undertaken in
phantoms to confirm that the prototype pulse sequence produced results in agreement with
a standard interleaved inversion recovery sequence with equilibrium achieved in-between
successive excitations. Next, validation experiments were undertaken with two healthy
volunteers (23 year old male and 36 year old female for volunteer 1 and 2, respectively) to
ensure that the sequence provided physically realistic measurements of relaxation in vivo.
70
Finally, the sequence was used to estimate lactate in a group of patients with brain cancer.
All volunteers participated with free and informed consent and with the approval of the
research ethics board at Sunnybrook Health Sciences Centre.
The values of lactate, NAA, creatine and choline were initially estimated from the
“Braino” phantom (GE Healthcare). As a proof of principle the measurement was done using
four linearly spaced values of 160, 260, 360, 460 ms using both the prototype pulse
sequence, and for comparison an inversion recovery (IR) PRESS sequence with an interleaved
adiabatic hyperbolic secant inversion pulse76,77 added prior to water suppression with a TR of 4
seconds. The TI, TR as well as the total experiment time for all experiments are shown in Table
3.2. For these initial measurements, identical values were used in both sequences to control
for possible systematic variation in the precision on associated with the sampling of
longitudinal relaxation. The TR values for the prototype pulse sequence were 1.5, 1.6, 1.7 and
1.8 s, respectively, with a total experiment time of 14 minutes. For the standard IR PRESS
sequence, an appropriate TR of 4 s was chosen to allow close to full recovery between each
successive inversion for the metabolites of interest, with a total experiment time of 34 minutes.
For practical application of the prototype pulse sequence, the optimal values were
determined for a given total acquisition time according to the optimization procedure
described above. Notable exceptions were that a) the optimization algorithm was run with an
initial value of T1 = 720 ms (the variance-weighted mean of the previous two measurements of
the T1 of lactate in this phantom) and b) the duration of the experiment constrained to 14
minutes, as for application of the prototype pulse sequence above. The TIs for the optimized
measurements in the “Braino” phantom were 160, 942 and 942 ms, and the associated TR
values were 1680, 2460 and 2460 ms for the three points, respectively. A similar numerical
optimization algorithm was written for the standard IR sequence which constrained the time
between the 2nd refocusing pulse and the start of the next TR interval to equal 5T1. For this
sequence the number of TIs and the number of averages were allowed to vary while keeping
the total experiment time equal to the experiment time of the prototype sequence. The
optimal TIs were determined as 160, 897 and 897 ms and the associated TR values were 3909,
4646 and 4646 ms, respectively, with 64 averages performed at each TI.
71
For in vivo work, the prototype pulse sequence was used to measure T1 with the
optimization algorithm prescribing TI values of 160, 2011 and 2011 ms with TR values of 3454,
5304 and 5304 ms, respectively. (The larger prescribed in vivo TIs and TRs are due to the larger
in vivo T1 values that are expected in comparison to those of the Gadolinium-doped “Braino”
phantom).
Table 3.2: The TI, TR and total scan time values for all experiments.
TI1
(ms) TI2
(ms) TI3
(ms) TI4
(ms) TR1
(ms) TR2
(ms) TR3
(ms) TR4
(ms) Total scan time (minutes)
“Braino” IR
160 260 360 460 4000 4000 4000 4000 34.13
“Braino” prototype
160 260 360 460 1500 1600 1700 1800 14.08
“Braino” IR optimized
160 897 897 n/a 3909 4646 4646 n/a 14.08
“Braino” prototype optimized
160 942 942 n/a 1680 2460 2460 n/a 14.08
In vivo prototype optimized
160 2011 2011 n/a 3454 5304 5304 n/a 30.00
When performing MRS of patients, care was taken to maximize the amount of tumor
tissue within the voxel by comparing the voxel placement with -weighted anatomical images
acquired after administration of Gd-DTPA contrast agent from previous exams, which were
available in 3 of the 6 patients. Care was also taken to avoid placing the voxel too close to the
skull or ventricles. A representative anatomical image with overlaid voxel placement is shown in
Figure 3.2.
72
3.4 Results
Figure 3.3 shows the results of the four relaxation measurements (prototype sequence and
IR-PRESS sequence with the same linearly-spaced values, the prototype sequence with
optimized values, and the optimized IR sequence) for the four main metabolites in the
“Braino” phantom. The measurements are in excellent agreement in all cases, as evident by
observing the similar slopes of the plots (proportional to
) for each metabolite, and the
Figure 3.2: T1 – weighted anatomical image with voxel placement (white square) overlaid for
patient 5. The singlet and doublet spectra at both long and short TI are displayed in Figure 3.4
for this patient.
differences are comparable to the measurement error. For display purposes, the amplitudes
have been arbitrarily scaled to distinguish each line easily. However, the steady-state values for
the prototype sequence versus the standard IR sequence are lower by 9%, 3%, 18% and 15%,
for creatine, lactate, NAA and choline, respectively. The decrease observed for lactate is
73
substantially less than the other three metabolites despite its relatively large value due to
the refocusing effect of BASING129, as explained further in the discussion. The metabolite
concentrations within the phantom were also estimated as described in the Methods section,
using the values of Figure 3.3 as well as previously measured T2 values. Values of
mM, mM, mM, and mM were estimated for the concentrations
of creatine, NAA, choline and lactate, respectively. The singlets are within approximately 10 %
of the known concentrations133 of 10 mM, 12.5 mM and 3 mM for creatine, NAA and choline,
respectively, whereas the lactate doublet is within approximately 25 % of the known
concentration of 5.0 mM. The lower estimate for lactate is likely due to residual anomalous J-
modulation24.
Figure 3.3 also shows that the estimation error is largest for lactate, due to its lower
concentration (and thus SNR) within the “Braino” phantom. The error on the estimated T1 of
lactate is reduced by a factor of 0.41 for the optimized prototype pulse sequence, compared to
the error obtained when using the arbitrary linearly spaced values. The reduction factor
predicted by the optimization algorithm is 0.47, in good agreement with this result. The error
on the estimated T1 of lactate is reduced by a factor of 0.74 for the optimized prototype
sequence compared to that obtained with the optimized IR sequence. The numerical
optimization predicts a reduction factor of 0.82, again in good agreement. Overall, these results
indicate a) that substantially improved precision of the estimate can be obtained if the
value is known reasonably well a priori; and b) that by removing the constraint of large TR
values, an improvement in precision can be gained for a constant experiment time.
In addition, the technique developed was applied to probe the white matter of two
healthy volunteers, producing the estimated values for choline, NAA and creatine shown in
Table 3.3. Adequate agreement is obtained with previous estimates122,131,134, with the present
estimate slightly elevated for creatine. Using these values, the mean metabolite concentrations
74
Figure 3.3: Inversion recovery results for “Braino” phantom for the four major metabolites
observed at TE = 144 ms (creatine, lactate, NAA, choline). The top line for each metabolite
corresponds to the prototype pulse sequence applied with optimized choice of TI and TR. The
second from the top line is the optimized interleaved IR technique. The second from the
bottom line is the prototype pulse sequence with linearly spaced inversion times, and the
bottom line is the standard interleaved IR technique applied with TR = 4000 ms. Amplitude
values were scaled arbitrarily for display purposes. The acquisition time for both applications of
the prototype pulse sequence and the optimized IR sequence was the same (approximately 14
minutes), and the acquisition for the linearly spaced IR sequence was approximately 34
minutes. The slope of the line is proportional to and the parameter estimates are also
shown including error estimates. Excellent agreement is observed between measurements for
all four metabolites. Substantially reduced error in estimating was obtained for both lactate
and NAA when using the optimized TI and TRs for both the prototype and the standard IR
sequence, as predicted by numerical simulation.
75
for the two healthy volunteers were estimated as mM, mM and
mM for choline, NAA and creatine, respectively. The concentration estimates are within the
range previously reported in the literature of 1-5 mM for choline, 10-25 mM for NAA and 6-14
mM for creatine135 and in excellent agreement with those reported using a very similar
absolute quantitation scheme35. No lactate was detected within the doublet spectra from the
healthy volunteers, as expected.
Table 3.3: T1 values measured within the white matter in two healthy volunteers as well as the
mean values obtained from healthy volunteers within the literature at 3 T.
Subject Choline T1 (ms) NAA T1 (ms) Creatine T1 (ms)
1
2
Chen, et al.131
Li, et al.122
Träber, et al.134 (inversion recovery)
Träber, et al.134 (saturation recovery)
Figure 3.4 shows the spectra obtained from a grade 3 glioma according to the voxel
prescribed in Figure 3.2, for the optimized values of 160, 2011 and 2011 ms. There is good
singlet suppression in the doublet spectra, consistent with previous work132 (as shown by the
significantly reduced amplitude of NAA, Cr and Cho), indicating negligible contamination from
lipids in the vicinity of lactate. Table 3.4 lists the estimated T1 values for lactate, choline, NAA
and creatine across the group of patients. The T1 estimates for choline and NAA agree well with
those obtained previously from glioma patients122, although the value presented here for
creatine is slightly larger. Potential reasons for this discrepancy are given in the discussion.
There is a significant difference between the T1 of lactate and choline (P = 0.004) as well as
lactate and NAA (P = 0.009). No significant difference is observed between the T1 of lactate and
76
creatine (P = 0.537). Given the estimated T1 of lactate, calculations indicate that √ is
maximized for a standard PRESS sequence with TE = 144 ms or TE = 288 ms when TR = 2830 ms.
Figure 3.4: Spectra obtained from patient 5 (from the voxel prescribed as shown in Figure 3.2).
The top row shows singlet spectra (choline, creatine, NAA and lipids) whereas the bottom row
shows doublet spectra (primarily lactate). Spectra in the left column were acquired with TI =
160 ms and those in the right column were acquired with TI = 2011 ms. Only two TI values are
displayed, since the second and third acquisitions of the optimized prototype pulse sequence
used the same TI value and the respective spectra differed only by noise. Note that there is
some non-lactate signal in the doublet spectrum in the 2-3 ppm range, likely due to the co-
editing of NAA and glutamate.
Table 3.5 displays the absolute concentrations for the four measured metabolites for all
six glioma patients. Good agreement is observed between the present results and those
previously estimate35,136 by a very similar absolute quantification procedure. Most importantly,
the lactate concentration estimate of ( ) mM is within experimental variation of the
previously reported values of ( ) mM35 and ( ) mM136.
77
Table 3.4: Estimated T1 values from six patients with high grade glioma. The measured spectra
from patient 1 indicated very large lipid, lactate and water peaks with minimal other
metabolites indicating that the voxel was placed primarily within a necrotic core; T1 could only
be measured for lactate in this patient. All patients were diagnosed with glioblastoma (grade 4),
except patient 5 who had a grade 3 glioma. The third last row is the mean standard deviation
across all patients. The second last row is a previous measurement of the T1 values obtained
from a population of glioma patients at 3 T. The last row is a previous measurement of the T1
values obtained from a population of brain tumor patients at 1.5 T.
Patient number Lactate T1 (ms) Choline T1 (ms) NAA T1 (ms) Creatine T1 (ms)
1* n/a n/a n/a
2
3
4
5
6
Mean
Li et al.122 n/a
Sijens et al.123 (1.5 T)
78
Table 3.5: Estimated absolute concentration of metabolites from the same patients with high
grade glioma as reported in Table 3.3. Spectra from patient 1 indicated very large lipid, lactate
and water peaks with minimal other metabolites indicating that the voxel was placed primarily
within a necrotic core. All patients were diagnosed with glioblastoma (grade 4), except patient
5 who had a grade 3 glioma. The second last row is the mean standard deviation across all
patients. The bottom row is a previous measurement of the absolute concentration values
obtained from a population of high grade glioma patients at 3 T.
Patient number [Lac] (mM) [Choline] (mM) [NAA] (mM) [Creatine] (mM)
1* n/a n/a n/a
2
3
4
5
6
Mean
Madan, et al.35
3.5 Discussion
In the present work, a novel pulse sequence was developed for measuring of low-
concentration J-coupled species in a time-efficient manner. This sequence was then applied to a
population of high grade glioma patients to measure the of the methyl group of lactate in
vivo.
A number of issues are worth discussion in relation to the pulse sequence that was
developed. First, the numerical procedure for minimizing in a specific experiment duration
(30 minutes in humans) prescribed that data should be acquired at a single low TI value, and
79
two identical high TI values. This prescription is not intuitive, however, it is understandable by
thinking about simple linear least squares procedures for estimating the slope of a line. In such
cases, the error on the slope has a dominator term equal to ∑ ( ), where x is the
independent variable, is the mean, and the subscript i refers to each instance of the
independent variable. The error on the slope is thus minimized if the values are spread
widely apart in relation to . This accounts for the wide separation in TI values that was
prescribed. In addition, the latter TI value has low SNR (as almost full recovery has occured)
and so repeated measurements are prescribed to obtain a more precise estimate of the signal
at this time point.
A related issue concerns linearization of the monoexponential equation for which
was performed to simplify computational aspects of optimizing TI and TR values for the
prototype pulse sequence. For the TI values with low SNR, no correction was made for how the
noise distribution was skewed by taking the natural logarithm. To check for potential
systematic error due to this approach, T1 estimates were re-calculated using nonlinear least
squares fitting of the monoexponential function. The nonlinear fit results deviated by <1 % from
the values listed in Table 3.3 and Table 3.4. Thus, linearization was useful in the present case,
but it is not essential to the success of the method and may not be advisable for data acquired
at lower SNR levels.
The prototype pulse sequence was also found to be robust to experimental
imperfections. Numerical simulation of the Bloch equations revealed that variation in
the prescribed flip angle of the excitation and refocusing pulses perturbed a T1 value of 2000 ms
by only ms, well within the biological variation of the experimental results. No obvious
signs of unwanted coherences were evident in the acquired spectra, possibly because
decreased spoiling power is required for the coarse voxel size used here26. More gradient
spoiling could be added to the sequence in the future, if required. Slightly increased noise was
consistently observed for the 2-3 ppm range in both the “Braino” phantom and volunteers,
however (eg. See Figure 3.4). This noise did not affect quantification of lactate, and likely
results from co-editing of J-coupled resonances of NAA and glutamate which are known to lie
within the pass band of the BASING pulse27. The effect of voxel shift due to chemical shift mis-
80
registration is also unlikely to be important because it can be reasonably assumed that all
lactate originates from the tumor, which was typically larger than the prescribed MRS voxel.
Lastly, water suppression could be achieved solely through BASING9 by modifying the frequency
scheme of the BASING pulses. This was not implemented here, as CHESS has been shown to be
robust across a range of static magnetic field strengths, whereas simultaneous spectral editing
and water suppression with BASING is challenging at 1.5 T due to spectral overlap effects with
the pulses used here. However, sole use of BASING could reduce the minimum allowable TI to
7.6 ms (down from 160 ms), which could help to improve measurement precision. Overall, the
quality of the spectra was judged to be similar to that obtained in previous work using the
identical BASING scheme132.
The optimal choice of TI values has previously been investigated for MFIR130. Similar
results were obtained in comparison to the present work, with optimal results obtained when
one low TI value and one high TI value were selected. The effect of low SNR motivating a
repeated measurement at the high TI value was not considered. In contrast, Ogg and
Kingsley130 have suggested taking measurements with two additional intermediate TI values.
Differences between the TI prescriptions are likely due to the differences in underlying
mathematical model. The present work uses a two-parameter mono-exponential model which
is appropriate under MRS conditions, whereas Ogg and Kingsley adopted a three-parameter
model more appropriate for proton MRI with higher SNR130.
Although a priori knowledge of the T1 is required for numerical minimization of the
uncertainty on T1, this is not a major limitation and is not a requirement for the
monoexponential model of to hold. In practice, the initial input of T1 for optimization did
not have a strong impact on the prescription of optimized TI and TRs. If the actual value of
varies from the input initial value of = 1725 ms by plus or minus 500 ms the result is an
increase in the predicted uncertainty by a factor of 1.11 and 1.02, respectively. In any case, this
is not a limiting issue. Some implicit knowledge of is always assumed when choosing TI
values to sample recovery curves appropriately, as part of obtaining high quality results in
inversion or saturation recovery experiments.
81
The MRS measurement precision in patients likely was also reduced compared to the
data quality that is achievable in healthy humans119. This is expected because of the lower
signal from tumor tissue and the difficulty in maximizing the amount of tumor tissue within the
voxel. These effects are general to the use of single voxel MRS and not specific to the pulse
sequence. However, TI values were chosen to optimize T1 measurements for lactate, which
recovers in the longitudinal direction more slowly than some of the other metabolites, slightly
compromising their measurement. Other MRS studies134 measuring the longitudinal relaxation
within tumors have achieved coefficients of variation (ratio of the standard deviation to the
mean) for choline, creatine and NAA of approximately 20%, similar to the values observed here.
The T1 estimate for creatine was slightly larger than previously measured122,123, which could be
due to differences in the patient populations studied, voxel positioning or other sources of
statistical variation. The discrepancy likely did not result from an imperfection in the model
used here. Such a systematic bias likely would impact NAA and choline results as well, whereas
these showed excellent agreement with the literature.
With appropriate modification, the technique presented here could also be used in a
variety of other measurement applications, such as to estimate the T1 value of 2-
Hydroxyglutarate64. Furthermore, the technique is not constrained to single voxel applications.
It could be incorporated in a typical MR spectroscopic imaging technique, or a multivoxel
technique that avoids or minimizes k-space encoding, such as Hadamard encoding137 or
constrained source space MRS138,139. This could be advantageous for measuring the value of
metabolites in pathologies with large heterogeneous regions, or to expedite control
measurements from regions of healthy tissue. The lactate-editing scheme chosen here is not
the only available technique for disentangling lactate and overlapping lipid signals. The
inversion sequence here could use the LASER (localization by adiabatic selective refocusing)
method instead, which substantially reduces the chemical shift mis-registration, and the
separation of lactate and lipid signals can be done using a multiple-quantum filter140. Other
alternatives are MEGA-SPECIAL141, which adopts a similar editing technique referred to as
MEGA for spectral separation and SPECIAL for spatial localization, or similarly MEGA-sLASER142.
These options could potentially improve SNR and, since the multiple-quantum filter is a single
82
shot technique, reduce sensitivity to motion at the cost of signals from the other metabolites.
For BASING to provide perfect lipid suppression, the spectra in the two-cycle subtraction
scheme must have identical phase and amplitude and thus there is the potential for motion to
introduce errors. However, the present results suggest that such errors are small in practical
circumstances.
The lactate T1 values reported for brain cancer patients in the present work agree well
with those for rat gliomas at 4.7 T60 but are substantially higher than those measured
previously at 1.5 T in a separate population by another research group who did not include
spectral editing123. It is possible to add the signals from the lipid and lactate resonances in the
present work, representing the scenario whereby measurements are made without use of
BASING. This provides a estimate of ( ) ms that is within experimental
uncertainty of the previous estimate of ( ) ms at 1.5 T123, strongly suggesting
that lipid contamination produces a substantial erroneous reduction in lactate . There is a
significant difference between the of the lipid plus lactate compared to the lactate alone
(P = 0.004). This discrepancy could also be partially due to the differences in field strengths
used for measuring (3 T here vs 1.5 T previously123), as in many biological compounds the
value exhibits a field strength dependency. It should be noted that there exists a correlation
between the T1 relaxation times measured from Patients 2-6 between lactate and the other
metabolites, as evident by the R2 values of 0.45, 0.67 and 0.26 for NAA, Cr and Cho,
respectively.
It should also be mentioned that the estimates provided here are probably slightly
different from the "true" values that would be obtained for each resonance in the absence of
magnetization transfer and chemical exchange effects. Such effects likely occur at some level in
the present experiments, for example due to use of CHESS pulses. This water suppression
scheme is very common, and the vast majority of in vivo MRS work using this and other
schemes does not account for exchange effects. Investigating the extent of perturbation
due to magnetization transfer or chemical exchange is an interesting topic for future research
which would likely benefit from the use of a time-efficient procedure for precisely measuring
longitudinal relaxation, such as developed here. From a practical standpoint, the present work
83
makes the appropriate estimates that are required to correct for partial longitudinal
recovery, as well as for optimizing √ in MRS experiments.
The absolute concentration values that were estimated in the present work are also in
good agreement with previous reports35,136. Overall, the general trend was confirmed of
increased lactate and choline, decreased NAA and roughly equal creatine concentration for
glioma tissue as compared to healthy tissue35. These estimates are based on the assumption
that the water resonance has a constant concentration of 42.3 M35,64, however, whereas the
actual concentration within glioma tissue could vary from this value, as well as from patient to
patient. This limitation can be removed by including an external reference sample in the
measurement protocol and data analysis. For now, clinical MRS will continue to rely on
interpreting MRS data using ratios of neurometabolite signals ratios with respect to creatine
(which has been shown to vary substantially within a patient population35). Irrespective of this,
the use of reference samples, literature values for and for each spectral component
(including the revised data for lactate reported here), enable absolute concentrations to be
estimated for use in basic and clinical MRS research. In the future, such work may facilitate the
use of lactate as a biomarker for improved cancer treatment and survival.
3.6 Conclusions
A novel MRS technique was developed and validated at 3 T for time-efficient measurements
of the methyl group of lactate without contamination from lipids. The resulting T1 value of
(2000 280) ms was obtained for a group of six glioma patients. After correcting for T1 (and T2
from literature values) the absolute lactate concentration was estimated as ( ) mM.
Lactate T1 exhibits similar variations as other major metabolites observable by MRS in high
grade gliomas. The T1 estimate provided here will be useful for future MRS studies that wish to
optimize pulse sequence parameters, or to report relaxation-corrected estimates of lactate
concentration as an objective tumor biomarker.
84
Chapter 4
Diffusion-Weighted J-Resolved Spectroscopy
A paper published in Magnetic Resonance in Medicine [Epub ahead of print] by Karl Landheer,
Rolf Schulte, Ben Geraghty, Christopher Hanstock, Christian Beaulieu, Albert P. Chen, Charles H.
Cunningham, and Simon J. Graham
4.1 Introduction
Proton magnetic resonance spectroscopy (MRS) is a powerful non-invasive technique used to
measure biomarker activity within the brain and body. The diffusion characteristics of the
spectral components can also be investigated by appending diffusion-sensitizing gradients to
standard protocols such as Point Resolved Spectroscopy4 (PRESS) and Stimulated Echo
Acquisition Mode113 (STEAM), as implemented by Posse et al.101 These techniques, referred to
as diffusion-weighted magnetic resonance spectroscopy (DW-MRS), provide novel information
and allow metabolites and their microstructural environment to be probed noninvasively within
the intracellular and extracellular space in vivo. The DW-MRS data features depend on factors
such as active transport, cytosol viscosity and compartmentalization. With the exception of
glucose and lactate, the metabolites that can be probed in this manner are predominantly
intracellular143.
In principle, each individual metabolite offers specific information about the
microstructural compartments where it resides. Previous DW-MRS results show that
metabolites diffuse freely along cell fibers, suggesting that the metabolites are not confined
inside cell bodies103. The pathophysiological changes in DW-MRS signals have been measured
for a range of disorders such as cancer, ischemia and excitotoxicity of the brain143, in addition to
glial reactivity in response to inflammation in systemic lupus erythematosus106. Functional MR
techniques have also been combined with DW-MRS to observe increased apparent diffusion of
85
metabolites during visual stimulation,107 providing a unique tool for investigating physiological
aspects of brain activity.
To date, DW-MRS of the human brain has primarily focused on the N-aceytlaspartate
(NAA), creatine (Cr) and choline (Cho) resonances due to their relatively high signal-to-noise
ratio (SNR) and strong non-overlapping singlets. Other metabolites are of substantial interest,
however. For example, glutamate (Glu) and gamma-aminobutyric acid (GABA) have important
roles as the main excitatory and inhibitory neurotransmitters, respectively46. Glutamate and
GABA typically reside in the synaptic vesicles29,46 within neurons, but during neurotransmission
they are released into the synaptic cleft where they bind to postsynaptic receptors. To the
authors’ knowledge, however, only three proton DW-MRS studies have investigated
metabolites in the human brain beyond NAA, Cr and Cho: glutamate, glutamine (Gln) and N-
acetyl aspartyl glutamate (NAAG) in healthy volunteers at 7 T108, lactate in edema and tumors
at 3 T144 and myo-inositol (mI) in healthy volunteers at 3 T145. The major difficulties in
measuring such other metabolites are: 1) low SNR, necessitating long scan time; and 2) spectral
overlap with larger resonances. These problems are particularly troublesome at field strengths
of 3 T and below. Alternatively, using higher field strength systems for small animal MRI, the
diffusion characteristics of up to 10 metabolites have been investigated in ischemic rat brains at
4.7 T146 and 12 metabolites have been investigated similarly in healthy rat brain at 9.4 T147.
Several MRS techniques have been developed for accurately measuring smaller
resonances in the absence of diffusion weighting. Spectral editing J-difference techniques have
been developed on clinical-grade MRI systems for detection of Glu and Gln8, GABA148, lactate129
and 2-Hydroxyglutarate64, among others. These techniques have good efficacy but usually
require the phase of two subsequent acquisitions to be consistent. It is well known that the
phase varies greatly from one excitation to the next in DW-MRS due to involuntary subject
motion, however, necessitating re-phasing of the spectra prior to summation101. Re-phasing is
a challenge under conditions of low SNR and this may negatively impact the ability to edit and
quantify the smaller resonances of interest. A small deviation from the 180° phase difference
requirement in spectral editing can lead to substantial error. In addition, spectral editing
techniques often improve detection of specific resonances while disrupting other portions of
86
the spectrum. Two-cycle spectral editing is probably not ideal for measuring the diffusion
properties of small resonances, therefore, in many cases.
An alternative spectral editing scheme based on double-quantum coherence-transfer
has been used to measure the diffusion of lactate in the presence of contaminating lipids149–151,
but discards information about other metabolites. Another approach uses biexponential fitting
to estimate the apparent diffusion coefficient (ADC) of lactate within a tumor in the presence of
lipids152. This method is well suited to obtaining lactate ADC values, but would be challenging to
apply to other metabolites such as glutamate due to spectral overlap, and the need for very
high SNR to discern each metabolite properly.
Small resonances can also be quantified at 3 T using localized two-dimensional (2D) J-
resolved spectroscopy (JPRESS)13,153, originally implemented in a half-echo acquisition mode. In
JPRESS, a second spectral dimension is added by sequentially increasing the echo time for each
spectroscopic data acquisition. As the initial phase of the signal collected from a J-coupled
metabolite varies depending on the echo time, collecting data at a series of different echo
times enables careful sampling of the dispersion in this second spectral dimension. Subsequent
2D Fourier transformation of the time domain data into the frequency domain enables
quantification of some of the smaller resonances which are overlapped in traditional one-
dimensional (1D) spectroscopy. The efficacy of the JPRESS sequence has been demonstrated on
clinical-grade 3 T MRI systems including acquisition of in vivo data, with a maximum echo
sampling scheme to improve sensitivity16. Using JPRESS and 2D fitting software referred to as
ProFit14,15, signals from 17 metabolites have been reliably measured in vivo at 3 T. Based on
these promising initial developments, a novel technique called diffusion-weighted JPRESS (DW-
JPRESS) is proposed here. The purpose of this study is to describe, characterize and validate use
of DW-JPRESS to measure the ADCs of metabolites beyond NAA, Cr and Cho at 3 T.
4.2 Methods
A prototype pulse sequence for DW-JPRESS was implemented using a MR750 3 T MRI system
(General Electric Healthcare, Waukesha WI) with a standard 8-channel head coil receiver. The
associated pulse sequence diagram is shown in (Figure 4.1), and includes three major
87
modifications from the basic JPRESS sequence. First, maximum-echo sampling was
implemented, whereby sampling consistently commenced 1 ms after the last gradient pulse,
instead of at the spin echo maximum as in typical MRS readouts. Maximum-echo sampling was
used because it has been shown to increase SNR and decrease overlap of resonance tails16. The
1 ms time delay following the last gradient pulse was included as a conservative measure to
allow for dissipation of any fast decaying eddy currents that were potentially present, prior to
commencing data collection. Second, diffusion-sensitizing gradients (DSGs) were added prior to
the first crusher and after the last crusher in the three orthogonal directions simultaneously.
Third, instead of automatically averaging the data from all identical excitations, all individual
spectral traces for identical excitations were saved for rephasing prior to averaging, as
necessary in DW-MRS101. The time between the start times of the DSGs, , was set to be 60.636
ms for the first TE value (74 ms), and incremented by = 1 ms for each TE step. The DSG width,
, was 8.396 ms with a rise time, , of 1 ms for all echo times (a slew rate of 50 T/s/m, ie. one
quarter of the maximum value available on the MRI system, 200 T/s/m). For in vivo
experiments, the amplitude of the DSG for the first TE value, , equaled 50 mT/m (the
maximum allowable gradient amplitude) and 5 mT/m for diffusion-weighted spectra acquired
with high and low b-values, respectively (see below for further details). The diffusion-weighting
parameters were chosen to provide a sufficiently large b-value for adequate diffusion weighting
while maintaining a short echo time. A non-zero b-value for the low b acquisition was used to
assist in suppressing unwanted echo pathways. Outer volume suppression154 and global water
suppression via CHESS19 were implemented using the standard pulse sequence components
available. Outer volume suppression was used both to reduce lipid contamination from the
scalp and for inner volume saturation25 to reduce anomalous J-modulation effects.
In each in vivo DW-JPRESS experiment, two data sets were collected with CHESS water
suppression on: one with the DSG amplitude set to the high b-value, and the other with DSG set
to the low b-value. In both cases, data were acquired with the number of excitations (NEX) set
at 8 for each TE step, for signal averaging purposes. Two additional calibration data sets were
collected with identical parameters except a NEX of 1 and water suppression off: one at high b-
value and one at low b-value. All 4 data sets were acquired with a minimum TE value of 74 ms,
88
Figure 4.1: DW-JPRESS pulse sequence for the a) initial echo time and b) intermediate kth echo time.
Shaded gradients are crushers. The large gradients played out in the three orthogonal directions are the
diffusion-sensitizing gradients (DSGs). Acquisition begins approximately 1 ms after the end of the second
DSG, for maximum echo-sampling. Note that the amplitude of the DSGs are reduced in b) compared to
a) according to Equation 4.3 to keep the b-value constant across all echo steps. The reduction in
amplitude is exaggerated for display purposes. The peak of the spin echo occurs at the echo time (TE)
and is equal to and the second refocusing pulse in b) is played out at a time later
than in a). Outer-volume suppression and global water suppression (CHESS) is executed prior to the 90°
excitation pulse. Gradient amplitudes are not drawn to scale (crushers are substantially smaller than the
DSGs for all TE values).
89
5000 Hz receiver bandwidth (corresponding to the bandwidth of the F2 dimension in the JPRESS
spectra), the aforementioned of 2 ms (corresponding to a bandwidth of 500 Hz in the F1
dimension) and 100 incremental TE steps. The water unsuppressed JPRESS spectra were used
for eddy current correction20 and estimation of coil sensitivity. Cardiac gating was also included
to reduce motion-related errors in vivo due to cerebrospinal fluid pulsatility, with the TR value
set to two R-R cycles (~2 sec) and a trigger delay of 300 ms101. For the phantom experiments
the TR was set to 1.5 s. For the phantom experiment performed on the “BRAINO” phantom,
two additional intermediate b-values were also collected (both the water suppressed and water
unsuppressed data). These data were subsequently used to compare ADCs estimated with four
b-values to those estimated with only two b-values, as part of validating the latter approach.
The signal processing of DW-JPRESS data was performed using specially-written scripts
in MATLAB (the Mathworks, Inc., Natick, MA) using a pipeline as shown in Figure 4.2. The data
from each coil were eddy-current corrected using the water-unsuppressed JPRESS spectrum20,
weighted by the coil sensitivity and then averaged over all eight coils. The coil-averaged signal
was then multiplied by a novel “streak correction” factor to maintain the water resonance at
constant amplitude across all NEX to correct any residual cardiac pulsatility artefacts not
eliminated by cardiac gating. The streak correction algorithm is similar to the algorithm
originally proposed by Posse et al.101 to reject individual data traces that exhibit excessive signal
losses. The algorithm used here first determined the excitation which produced the highest
water resonance amplitude (ie. the least non-linear motion) for each of the 100 individual echo
times. Results for the other seven excitations were then scaled to match this water resonance
amplitude. In the absence of motion, the water resonance amplitudes for all 8 NEX had a
comparatively small noise envelope and thus the streak correction had negligible impact. When
non-linear motion was present, however, the algorithm substantially reduced the spurious
fluctuations and subsequent streak artifacts in the 2D FID. The water resonance was then
removed using Hankel singular value decomposition (HSVD). The signal was then Fourier
transformed and automatically phase corrected for zero and first order phase terms118 which
were obtained from the water resonance prior to removal, inverse Fourier transformed and
90
Figure 4.2: Flow chart of the processing steps used to estimate ADCs from the raw DW-JPRESS data.
The 2D free induction decay (FID) results are processed through two separate pipelines referred to as
the 2D pipeline (left) and the 1D pipeline (right). The arrow connecting the 1D pipeline and the 2D
pipeline indicates that the 1D Cr303 data are multiplied with the results from ProFit (because Profit
outputs spectra as a ratio to Cr303), ensuring that the diffusion characteristics of Cr303 do not bias ADC
estimates for all metabolites. The preprocessing steps are done for each TE value (ie. 100 times) with
two repetitions: once for high diffusion and once for low diffusion. This produces two 2D FIDs which are
then combined to calculate the ADCs in the bottom portion of the flow chart.
91
averaged over all NEX. These preprocessing steps were repeated for each echo step (100 times)
for the data obtained at both high b-value and low b-value.
The preprocessed FIDs were then submitted to two spectral analysis pipelines,
subsequently referred to as the 2D and 1D pipelines (left and right columns in Figure 4.2,
respectively). The 2D pipeline consisted of analysing the 2D spectra at both diffusion weightings
with ProFit, using a basis set simulated from the complete density matrix with the General
Approach to Magnetic resonance Mathematical Analysis (GAMMA) library155 using the same
metabolites as in the original ProFit implementation14 minus glucose (due to its very small
concentration). The basis set consisted of 19 metabolites: alanine (Ala), ascorbic acid (Asc),
aspartate (Asp), Cr, phophorylcholine (PCh), GABA, Gln, Glu, glycine (Gly), glutathione (GSH),
glycerophosphorylcholine (GPC), lactate (Lac), myo-insitol (mI), NAAG, phosphorylethanolamine
(PE), scyllo-inositol (Scy) and taurine (Tau). For the work presented here, it was found that
there was insufficient spectral resolution to separate NAAG from NAA and PCh from GPC. For
this reason, the results from NAAG and NAA were summed and subsequently referred to as
total NAA (tNAA), and the results from PCh and GPC were summed and referred to as total
choline (tCho). Additionally, Cr was split into two separate metabolites, referred to as Cr303
and Cr391 for the two resonances at 3.03 ppm and 3.91 ppm, respectively, as is typical with
ProFit14.
Because ProFit results are output as a ratio to the Cr303 peak, the 2D data were then
multiplied by the Cr303 value obtained from the 1D pipeline (represented by the horizontal
arrow connecting the two pipelines in Figure 4.2) to estimate ADCs using unscaled
concentration values rather than ratios to Cr303. This processing step was important to remove
bias, as otherwise the 2D spectral results at each b-value would be weighted by the specific
diffusion characteristics of Cr303. The estimated diffusion characteristics for Cr303 were thus
identical for both the 1D and 2D pipelines. The 1D pipeline consisted of separately analysing the
data for each of all 100 TE steps with the freeware known as Totally Automatic Robust
quantitation in NMR (TARQUIN94). Because TARQUIN requires the onset of data acquisition to
92
occur at the echo maximum, all data acquired prior to the peak of each spin echo were
discarded for this analysis. This corresponded to truncating the first ( ) samples for the
kth line in the raw 2D FID. The average of the metabolite values obtained over the 100 individual
lines was then used to estimate the 1D pipeline ADCs. The basis set was automatically
generated in TARQUIN and contained identical in vivo metabolites to the basis set used in the
2D pipeline, as described above.
In preliminary work, DW-JPRESS experiments were also conducted on the GE Healthcare
"BRAINO" phantom that contains a restricted set of metabolites with known concentrations133.
In these cases, the basis set for the 1D and 2D pipelines was restricted according to a priori
knowledge of the metabolite set (Cho, Cr, NAA, mI, Lac and Glu) to avoid overfitting. The 1D
pipeline was also used for estimating the in vivo ADC values of NAA, Cr, Cho, mI and Glu. These
results provided a useful comparison with the analogous estimates obtained from the 2D
spectral analysis.
The ADCs were estimated for both pipelines according to the standard equation
[ ( ) ( )]
(4.1)
where ( ) is the peak area value obtained from the fitting software for each particular
metabolite, m, from the spectra with b-value , and ( ) is the analogous metabolite peak
area value from the signal with b-value . The uncertainties of ADC parameters estimated by
the 1D pipeline were obtained by propagating the uncertainties in the concentration estimates
from the 100 individual TE values. For the 2D pipeline, the analogous uncertainties were
obtained from the uncertainty of the Cr303 values estimated by the 1D pipeline and the
Cramer-Rao lower bounds outputted by ProFit, according to standard error propagation. The b-
values were calculated according to the equation for a trapezoidal gradient:
[ (
) ]
(4.2)
93
where is the gyromagnetic ratio for protons, and is the amplitude of the DSG at the first TE
value. Because each TE step resulted in a different duration between the two DSGs, it was
necessary to adjust the amplitude of the DSGs at each TE value to keep the b-value constant.
This was achieved by the following equation:
√ (
)
( )
(4.3)
where is the amplitude of the DSG at the kth TE step. The b-values were 2188 and 22 s/mm2
(G1 amplitude of 50 mT/m and 5 mT/m) for the high b and low b spectra for in vivo
experiments, respectively. For the phantom experiments, the four b-values were 1012 s/mm2,
1264 s/mm2, 1544 s/mm2 and 1852 s/mm2 (G1 amplitude of 34 mT/m, 38 mT/m, 42 mT/m and
46 mT/m), respectively, reflecting the change in material properties. A smaller maximum b-
value was used in the latter case because the ADC values were known to be substantially higher
in the phantom than in vivo. For the same reason, a larger minimum b-value was used to assist
in water suppression in the phantom. To investigate the efficacy of using Equation 4.3 to
modify the amplitude of the DSGs for each TE step, ADCs were calculated and compared for the
first 15 and last 15 TE values in the phantom experiment.
To investigate the potential for bias between the two data processing pipelines, the
ADCs from both pipelines for tNAA, tCho, Glx and mI were compared using a Mann-Whitney U
Test, with the threshold for statistical significance set according to a Type 1 error of .
Additionally, to quantify the deviation between the two pipeline ADC estimates for tNAA, tCho,
Glx and mI, the root mean square relative difference, for each metabolite, m, was
calculated:
√
∑(
)
(4.4)
94
where and are the estimates from the 2D and 1D pipeline for metabolite
m and the ith subject, respectively (N subjects in total). This value was not calculated for Cr303
because the values from both pipelines were identical, as described above.
Once the DW-JPRESS pulse sequence was debugged and validated using the “BRAINO”
phantom, experiments were subsequently conducted in 8 young healthy adult volunteers free
from previous or existing neurological or psychological deficits. All volunteers participated with
free and informed consent and with the approval of the Research Ethics Board at Sunnybrook
Health Sciences Centre. One dataset was discarded due to poor shim, and another due to
motion that resulted in negative ADCs. Thus the data for 6 volunteers are presented (one
female) with an age range of 23 to 35 years. For the phantom experiment, the voxel was placed
within the middle of the “BRAINO” phantom and was 2.53 cm by 2.67 cm by 2.44 cm in size
(16.5 cm3). In vivo, voxels were consistently placed in predominantly parietal white matter with
sizes that ranged from 4.76 cm by 1.86 cm by 2.29 cm (20.3 cm3) to 4.54 cm by 2.78 cm by 2.54
cm (32.1 cm3) in the anterior/posterior, right/left and superior/inferior directions, respectively.
Figure 4.3 shows a representative voxel overlaid on top of axial and coronal anatomical images.
The total experiment time for each subject was approximately 75 minutes, which included a
localizer and T1-weighted anatomical imaging (FSPGR IR, 256 by 256 pixels, pixel size = 0.86 mm
by 0.86 mm, TR/TE = 8.2/3.2 ms, flip angle = 8°). Each individual water-suppressed JPRESS
spectra was acquired in approximately 26 minutes, although due to the cardiac gating
procedure, this depended on the heart rate of the subject.
4.3 Results
Table 4.1 lists the ADCs estimated from the “BRAINO” phantom using both the 1D and 2D
pipelines. The measured temperature within the bore of the magnet was 19°C and, for
comparison with literature values typically given at 20°C, the data from the phantom were thus
scaled assuming equal activation energy between water and the metabolites156 using cubic
interpolation of the water ADC values from Sacco et al.157 The values listed in Table 4.1 include
an increase of approximately 3 % above the unscaled values due to this temperature correction
95
Figure 4.3: a) Axial prescription and b) coronal prescription of the DW-JPRESS voxel for Subject 3. DW-
JPRESS spectra for this subject are displayed in Figure 4.4.
factor. The ADC of water, as estimated from the residual water signal after water suppression,
yielded a value corrected to 20°C of x10-3 mm2/s using four b-values. The
analogous corrected value was x10-3 mm2/s using the lowest and highest b-values
only, as subsequently used in vivo. These estimates agree well with each other and the ADC
values reported in the literature for water at room temperature, which range between 1.95
x10-3 mm2/s and 2.1 x10-3 mm2/s101,102,108,157. Table 4.1 also shows excellent agreement
between ADC estimates for both pipelines for the three metabolites typically measured by DW-
MRS: Cr (both components at 3.03 and 3.91 ppm), choline and NAA. The measured percent
difference of the ADC estimated from the first 15 TE values compared to the last 15 TE values of
the 2D data set, using the b-values of 1012 s/mm2 and 1852 s/mm2, was ( ) %,
( ) %, ( ) %, ( ) %, ( ) %, ( ) % and
96
( ) % for Cr303, Cr391, tNAA, Cho, Lac, Glu, and mI, respectively. Negligible
differences were
Table 4.1: ADC estimates obtained from the “BRAINO” phantom for the 2D and 1D pipelines scaled to
20°C. For each pipeline, results are listed for fitting data acquired at 4 b-values (1012 s/mm2, 1264
s/mm2, 1544 s/mm2, 1852 s/mm2) to Equation 1 using linear least squares, as well as for fitting with 2 b-
values (1012 s/mm2, and 1852 s/mm2).
Metabolite 2D-pipeline ADC (x 10-3 mm2/s) 4 b-values
2D-pipeline ADC (x 10-3 mm2/s) 2 b-values
1D-pipeline ADC (x 10-3 mm2/s) 4 b-values
1D-pipeline ADC (x 10-3 mm2/s) 2 b-values
Cr303
Cr391
NAA
Cho
Glu
Lac
mI*
*Because of rapid T2 relaxation, only the first 10 TE steps were used in this case.
observed between ADC pairs in all cases except Cr391, verifying that it was acceptable to use
Equation 4.3 to maintain a constant b-value for all TE values. It can also be seen Table 4.1 that
the ADCs estimated using two b-values are in excellent agreement to those obtained using 4 b-
values for both pipelines, validating the use of two b-values for the in vivo experiments.
Figure 4.4 shows typical in vivo 2D JPRESS spectra obtained with both high b-values and
low b-values, including the spectral fitting and residual results obtained from ProFit. Similar fit
quality is obtained in both conditions, indicating that the large DSGs have negligible impact on
the quality of the spectral data (eg. linewidth).
Table 4.2 lists the ADCs estimated from both pipelines for 6 healthy volunteers. The mean was
calculated using only values which were physically realistic (ADC value 0.80 x 10-3 mm2/s, the
mean ADC of water in brain measured here, which the metabolites should not surpass, and
0.09 x 10-3 mm2/s, which is the lowest measured ADC in healthy tissue of any metabolite
previously reported109). Some of the weakest resonances were not reliably quantified by ProFit
(as observed by extremely large Cramer-Rao bounds or by non-physical
97
Figure 4.4: Water-suppressed JPRESS spectra obtained from Subject 3 with voxel prescription as shown
in Figure 4.3. The real spectra (top), fits (middle) and residuals (bottom) are plotted on a logarithmic
color scale. High b value weighting is shown in the left column of plots, with low b value weighting
shown in the right column. Note that due to the scaling performed by ProFit the amplitude appears to
be the same; however, the high b spectrum has its signal amplitude reduced by ~35 %. As can be seen
from the comparable residuals, similar quality fits are obtained for both low and high b-values. The
subfigures are logarithmically scaled and window and levelled consistently throughout, with negative
values depicted in blue.
negative ADC values) and were therefore excluded. The metabolites Ala, Asp, Asc, Pe, GABA,
Gln, Gly, and Tau were found to have physically unrealistic ADCs (either negative or several
times larger than water) for at least one subject. As expected, a wide range in ADC estimates is
observed across the remaining resonances, with the highest mean ADC observed for Tau
(although this result may not be reliable), and the lowest for mI and tCho. These results suggest
that knowledge of the ADC values for some of these other metabolites beyond Cr, tNAA and
tCho may provide additional valuable information.
98
Table 4.2: ADCs estimated from 6 subjects for 2D and 1D pipelines.
Metabolite
Subject 1 Subject 2 Subject 3 Subject 4 Subject 5 Subject 6 Mean
2D Pipeline ADC (x 10-3 mm2 /s)
Cr303
Cr391
tNAA
tCho
Ala -
- - - -
Asp
Asc
GABA -
-
Glu
Gln
Glx
Gsh
Gly
-
Lac -
-
mI
Pe
- - - -
Scy
Tau -
-
1D Pipeline ADC (x 10-3 mm2 /s)
Cr303
tNAA
tCho
Glx
mI
Water
99
No significant difference is observed between the median ADC values estimated from
the two pipelines for tNAA (P = 0.82), tCho (P = 0.39), Glx (P = 0.82) and mI (P = 0.70). Figure 4.5
demonstrates the correlation between the 1D and 2D pipeline estimates for tNAA, tCho, Glx, mI
across the six subjects. The measured Drms,m values were ( ) %, ( ) %,
( ) %, and ( ) %, for tNAA, tCho, Glx and mI, respectively. The larger
variation for Glx and mI, as well as reduced precision, is likely due to the difficulties in
estimating these resonances using traditional 1D spectroscopy.
Figure 4.5: Plot of ADCs estimated from the 2D pipeline versus those estimated from the 1D
pipeline. Strongest agreement is observed for tNAA, followed by tCho; poorest agreement is
observed for Glx and mI.
100
4.4 Discussion
The quality of the spectra shown in Figure 4.4 was obtained by carefully controlling for the
effects of eddy currents. A previous study found that a single water unsuppressed spectrum
was sufficient for eddy current correction in typical JPRESS spectra16 because the majority of
the eddy currents arose from the last set of gradients applied. However, preliminary
experiments related to the present work (not shown here) found that water-unsuppressed
spectra acquired at a single TE value were insufficient to remove eddy current effects in DW-
JPRESS spectra for all TE steps, as apparent by distortions in the lineshapes. This is likely due to
the fact that substantial eddy currents result from the very large DSGs, which have modified
timing for each TE step. Differences in gradient coil hardware characteristics between the
research MRI system used here and that used in the previous work16 could also be an important
factor. Irrespective of the cause, one full water-unsuppressed JPRESS spectra was acquired for
each b-value to correct for eddy currents with good efficacy across all TE steps. The eddy
current correction was done for each individual coil element prior to averaging, as shown in
Figure 4.2, because in principle each may be sensitive to slightly different eddy currents.
Due to its simplicity, Equation 4.3 was used at each TE step to calculate the amplitude of
the DSGs. This equation neglects the small effect of crushers, however, which at the larger TE
values produce a slightly larger b-value than the nominal b-value listed. The b-value for each
echo step was also calculated using the alternate approach of integrating the square of the k-
space trajectory mapped out by the gradient waveforms, which included all DSGs and crushers.
It was found that the estimate of the b-value through this technique was approximately 0.5 %
greater for the last echo than the first echo, indicating a negligible change for the precision of
the experiments presented here. This was corroborated by the negligible changes observed in
the phantom experiment when comparing ADC values calculated from the first 15 TE values to
those calculated from the last 15 TE values. As part of these observations, the very large
uncertainties in the ADC difference values for Glu and mI are due to the relatively low
concentrations as well as the low T2 values of these metabolites in the phantom (resulting in
large variations in the ADCs estimated at larger TE values), whereas the large uncertainty
observed for Cr391 is due to contamination from the water peak, as explained further below.
101
It is also important to emphasize that the present work does not provide rotationally
invariant estimates of ADCs, as DSGs were applied in all three orthogonal directions
simultaneously. The rotationally invariant estimate is usually achieved by acquiring diffusion-
weighted spectra from gradients successively applied in three orthogonal directions, but this
was impractical in the present work due to experiment time constraints and the need to
generate high b-values while keeping the TE values relatively low.
The initial DW-JPRESS data from the BRAINO phantom (Table 4.1) confirmed that both
the 1D and 2D pipelines produced ADC estimates in good agreement with each other for all
metabolites except mI. The discrepancy observed for mI was likely due to quantification
difficulties using 1D spectroscopy, and the small T2 value of this metabolite in the phantom. The
results also agree with ADC values previously estimated from an identical phantom as
x10-3 mm2/s, x10-3 mm2/s , and x10-3 mm2/s for tNAA,
creatine and choline, respectively, using a DW-MRS PRESS sequence at 7 T100, whereas the
values obtained from the 2D pipeline were x10-3 mm2/s, x10-3 mm2/s
and x10-3 mm2/s. The previously measured ADC of glutamate within a glutamate-
only phantom was x10-3 mm2/s at 7 T108, which is about 20 % larger than the value
presented here of x10-3 mm2/s for the two b-value estimate. This discrepancy
could be due to differences in the temperature at which these values were measured,
differences in eddy current correction schemes or other differences in the data processing
pipelines between studies, such as the confounding overlap of glutamate and NAA at 2.03 ppm.
For the other metabolites (NAA, Cho and Lac) the difference between the ADCs values
estimated by both pipelines is comparable to the experimental uncertainty. The 2D pipeline
provides substantially improved precision, however, likely because of the improved fitting
capability that arises from introducing a second spectral dimension. Overall, these results
indicate that the 2D pipeline can be used to extract ADC estimates in agreement with those
obtained at higher field strengths in phantom experiments. Due to the difficulty in quantifying
smaller resonances using 1D MRS at 3 T, no attempt was made to quantify ADCs beyond tNAA,
Cr, Cho, Glx and mI using the 1D pipeline for in vivo data (which exhibit much broader
linewidths and lower SNR than the phantom data).
102
Previously reported ADC values for tCho, tNAA and Cr in white matter108 are
x10-3 m2/s , x10-3 mm2/s and x10-3 mm2/s, respectively. The
mean values for tCho, tNAA and Cr303 from the 2D pipeline used in the present work are
mm2/s, mm2/s and mm2/s, respectively. The present
mean ADC value estimated for tCho is slightly elevated compared to the previous finding.
However, the results between 2D and 1D pipelines are in excellent agreement for all subjects
except Subject 1, for which the tCho estimate for the 2D pipeline was slightly elevated,
although this was still within experimental uncertainty. This observation notwithstanding, no
statistically significant differences were observed between the two pipelines for tNAA, tCho, Glx
or mI ADC estimates across all subjects (Cr was not tested because it was identical for the 2
pipelines, as described in the Methods). There is, however, an increase in the standard
deviation across all 6 patients from the 2D pipeline as compared to the 1D pipeline. This is due
to the uncertainty on the 2D ADC estimates being derived from the uncertainty from both the
1D pipeline Cr303 and the uncertainty on the 2D individual metabolites from ProFit. It should
also be mentioned that due to the inherent variation in diffusion time as well as TE value within
this DW-JPRESS, possible correlations between relaxation and diffusion-weighted signals may
impact the estimated ADC values. It has previously been suggested that no correlation between
relaxation and diffusion properties exist for mouse brain158, however a significant correlation
was found for NAA and Cr within human brains159 (although the effect was small). The present
results are also consistent with a small effect, given that the estimated ADC values agree well
with previous estimates obtained with standard 1D DW-MRS. Nevertheless, mitigation
strategies should be considered for future studies (such as use of a smaller value for ).
Future work will investigate removing the Cr303 scaling by ProFit, which in turn will
make it unnecessary to perform the processing step of multiplying DW-JPRESS results by Cr303
values obtained from the 1D pipeline. The net result should be improved precision on the ADC
values estimated by DW-JPRESS. The estimated ADC of water obtained here across all six
subjects of x10-3 mm2/s agrees well with the previous invariant trace/3 ADC
estimate in healthy white matter at 3 T102 of x10-3 mm2/s. The values obtained
within the present work for Cr, NAA, Cho lie in the range previously reported for rat
103
brain146,147,160–163. However, The estimates obtained here for mI, Gln and GSH (which reside in
glial cells) and Glu (which resides in axons) are elevated compared to rat studies146,147,
suggesting a possible differences in the compartments within the respective cells, physiological
differences between species, or differences in acquisition and analysis schemes. Furthermore,
the effects of restricted diffusion have been completely neglected in the present work due to
the minimum number of b-values chosen, as well as the relatively small maximum b-value used.
Future studies may separate the fast and slow-diffusing components using more b-values and a
biexponential fit, which can be used to estimate the intra-extracellular distribution of
metabolites, as has previously been done in rat brains147.
Possible causes of the slightly elevated tCho ADC estimates in the present work,
compared to previous reports, include residual errors from cardiac pulsatility and T1 recovery.
The effects of cardiac pulsatility were elevated by use of a somewhat large MRS voxel to
improve SNR in DW-JPRESS. Although the "streak-removal" procedure was effective at
suppressing pulsatility effects that remained present even in the presence of cardiac gating,
close inspection of the data indicated that a small amount of pulsatility artifact still remained.
In addition, cardiac gating introduced a variability in the TR interval that created small signal
fluctuations related to differences in the extent of T1 recovery for each of the resonances. The
former issue can be addressed by moving to a smaller voxel (see below) or, for example, by
implementing navigator-based reacquisition of corrupted data100. The T1 recovery effect can be
addressed by increasing the diffusion-weighting, which would likely improve the ADC precision,
or by increasing the TR so that small variations in TRs have less impact, at the expense of
increased experiment time. It is possible that tCho was most affected by small variations in TR
values because it exhibits a shorter T1 value than Cr or tNAA119. It is also important to mention
that the present work is particularly sensitive to bulk tissue motion due to the relatively long
experiment time, and the use of large diffusion sensitizing gradients to detect molecular
displacements. It is likely that this sensitivity can be reduced by future technical development,
for example by implementing a new version of the DW-JPRESS pulse sequence optimized for
newer MRI systems with enhanced SNR and gradient amplitude. Additionally the voxel used
was predominantly white matter, as it allowed for the largest voxel possible, although due to
104
the relatively higher concentration of metabolites in grey matter than white matter in future
implementations it may be possible to increase the SNR by moving to a smaller voxel placed in
grey matter.
Previously reported ADC values for Glu, Gln and Glx (Glu+Gln) in parietal white matter of
healthy volunteers are x10-3 mm2/s, x10-3 mm2/s and x10-
3 mm2/s, respectively, using a DW-MRS PRESS sequence at 7.0 T108. For comparison, the
present work obtained values of x10-3 mm2/s, x10-3 mm2/s and
x10-3 mm2/s, respectively, in good agreement. The one notable exception involved
Subject 5, for which the ADC estimated for Gln was uncharacteristically low. This anomalous
result was likely due to the inability of ProFit to distinguish between the two glutamate-
containing moieties. Thus, the reliability of DW-JPRESS to separate Glu from Gln reliably at 3 T
remains an open question and likely depends on the available SNR as well as shim quality.
Additionally, the value for Glu obtained here agrees well with a previous value of
x10-3 mm2/s which was obtained in a monkey brain using Carbon-13 labelled glutamate164.
The 2D pipeline ADC estimate for mI was x10-3 mm2/s, in good agreement
with previous results obtained in an anaesthetized monkey165 ( x10-3 mm2/s), and in
healthy volunteers using diffusion-tensor spectroscopy145 ( x10-3 mm2/s).
Interestingly, the ADC value for mI was among the lowest of all metabolites that were studied
(along with tCho). It was also observed that the ADC of mI (which primarily resides within glial
cells) was smaller than Scy, despite an equal molecular weight. This suggests that there is a
potential difference in either cellular localization or compartmentalization of these molecules.
This difference was close to statistically significant (P = 0.06). Although the estimate for the ADC
of Scy was anomalously high for Subject 1, even when excluding this subject a difference trend
was observed between the ADC estimate for Scy and mI (P = 0.08). There was a weak
correlation (R2 = 0.13) between the ADC and the square root of the molecular weight. This
indicates that although some of the variation in the ADCs can be attributed to size of the
molecules, most is due to various compartmentalization factors.
105
The remaining metabolites of interest were probed less reliably by the prototype
implementation of DW-JPRESS. Only Subjects 3 and 4 had a coefficient of variation below 25 %
for GABA. The T2 of GABA is approximately 88 ms166, similar to the initial TE value of 74 ms
which ensures appreciable T2-weighting of the DW-JPRESS results. Furthermore, GABA has a
relatively low concentration and strong overlap with larger resonances. The ADC values for Ala,
Pe and Tau were not reliable in any subjects, likely for similar reasons. Future work to reduce
the minimum TE value could help to alleviate this problem (using an MRI system with higher
gradient strength) and allow ADCs to be estimated reliably for a larger range of biomolecules
than was achieved in the present work. It would also be interesting to investigate the
implementation of DW-JPRESS at 7 T where there is improved SNR and decreased overlap
between resonances.
In principle, identical ADC values should be estimated for Cr303 and Cr391 using the 2D
pipeline. The results listed in
Table 4.2 show consistently higher in vivo values for Cr391 than for Cr303, however (except for
Subjects 5 and 6 where the estimate is markedly low). This effect likely arises from use of the
HSVD method to remove water signal, which affects the Cr391 resonance because it is near the
HSVD cut-off frequency. Conversely, the phantom experiments provided very similar ADC
estimates for Cr303 and Cr391 for both pipelines because of substantially reduced linewidths in
comparison to in vivo conditions, and thus reduced effect from water contaminating the Cr391
peak.
A relatively large voxel size was used in this proof-of-principle work, even in the context
of single voxel experiments. In general, JPRESS requires a large voxel size for accurate detection
of the smaller coupled resonances, and the diffusion-weighting introduces an additional
reduction in SNR of ~35%. The b-value used for the in vivo measurements was 2188 s/mm2,
whereas the b-value that minimizes the standard deviation of Equation 4.1 is approximately
80% of 1/ADC. It is therefore possible that the optimal b-value which minimizes the expected
uncertainty on the ADCs is somewhat higher, which is corroborated by Ellegood et al.167
although care must be taken due to the potential of increased weighting from the restricted
106
diffusion regime at higher b-values. With the gradients used here this would necessitate
increasing the gradient duration, , which would increase TE values, increase the effects of
eddy currents and bring the sequence further away from the q-space condition of
instantaneous diffusion, resulting in artificially inflated ADCs. Improvements can likely be
obtained by optimizing diffusion sensitization, choice of TR as well as the number of TE steps.
Typical 1D DW-MRS has previously been optimized at 3 T, where a recommendation of 13
minutes duration was prescribed105. A similar reproducibility study should be done to
determine optimal DW-JPRESS measurement parameters to obtain reliable ADC estimates with
low variability. This will be investigated in the future, and it is likely that optimized
implementations of DW-JPRESS can significantly reduce the total scan time or voxel size while
keeping a similar precision on the ADCs obtained here. In addition, bipolar gradient schemes109
have been shown to reduce eddy currents108 and their implementation should be investigated
for DW-JPRESS.
In principle DW-JPRESS could be combined with parallel imaging to measure the change
in ADCs in two or more voxels simultaneously138, which would be critical in reducing experiment
times for future DW-MRS applications such as those that compare results between healthy and
diseased tissues. However, care must be taken to avoid the possibility of introducing artefacts
from a multivoxel reconstruction.
4.5 Conclusions
A novel technique which combines JPRESS with DW-MRS was developed to measure the ADCs
of metabolites beyond NAA, Cr and Cho at 3 T. The proposed technique was found to provide
consistent estimates for the ADCs of tNAA, Cr and tCho when compared to a typical DW-MRS
pipeline. Additionally the new technique provided realistic estimates for the ADCs of glutamate
+ glutamine, and myo-inositol in all subjects and additionally glutathione and scyllo-inositol in
all but one subject. With further technical development to address the main limitation of long
acquisition times, DW-JPRESS will become more practical and may provide useful information
about the diffusion characteristics of metabolites beyond NAA, Cr and Cho at 3 T.
107
Chapter 5
Conclusion
The most common in vivo MRS sequences are PRESS4 and STEAM5 which were developed
almost 30 years ago and, in principle, remain identical to their initial inception. Despite its
mature status, however, MRS continues to find new applications due to its underlying ability to
non-invasively measure physiological information to supplement anatomical information from
MRI. What follows is a discussion of the outcomes of the thesis, as well as recommended future
work. It is evident that there remains considerable scope for further development of in vivo
MRS technology. It will be interesting to see how the field of in vivo MRS evolves as the
translation of in vitro techniques becomes increasingly possible, due to ongoing improvements
in hardware such as increased field strength, improved high order shimming, and increased
gradient amplitude and slew rate.
5.1 Summary
In Chapter 1, a brief overview was presented of the necessary physics to understand the work
developed in the thesis. In particular, the quantum mechanical basis for proton MRS was
explained and the product operator formalism was introduced. The basics of in vivo MRS were
explained including the metabolites observable in the brain, how the signal is spatially localized,
as well as topics such as parallel imaging, absolute MRS and diffusion-weighted MRS, that were
relevant for the work that was undertaken. In addition, the present status of MRS as applied to
brain cancer and brain cancer treatment was also discussed.
Within Chapter 2 a study was presented which extended SVS to include multiple voxels
through amplitude modulation of the excitation RF pulses and use of localized multi-channel
coil sensitivity to reconstruct the individual voxels. By this approach, referred to as “constrained
source space MRS” (CSSMRS), it was shown that high quality data from two voxels, with
relatively few artefacts and very little “bleed” from one voxel to another. In addition, the
increase in noise due to the geometry of the head coil was investigated as a function of
108
distance between the two voxels, which was necessary to determine whether there is a time
benefit in using CSSMRS. Based on the positive results of this experiment, the CSSMRS
technique was applied to a cohort of brain cancer patients. This was undertaken because SVS is
often used in that particular clinical setting to differentiate between neoplastic and non-
neoplastic lesions, typically with additional data acquisition from a contralateral brain region as
a control. Positive results were again obtained and it was concluded that CSSMRS could
potentially halve the duration of clinical MRS protocols where spectra are measured from both
lesion and control locations.
Within Chapter 3 a novel pulse sequence was developed which was a hybrid of fast IR
and saturation recovery used it to estimate the longitudinal relaxation time of lactate, filling a
gap in the in vivo MRS literature. The sequence was similar to a standard IR experiment except
that the inversion pulse was interleaved on and off successively, and the total repetition time
(TR) was linked to the inversion time (TI). By making these two changes: 1) the difference signal
at each TI value followed a monoexponential function of T1, enabling simple fitting procedures
to estimate the relaxation time; and 2) the consistent need for long TR values (as required in
standard IR) was removed, providing increased time efficiency. The predicted uncertainty on
the estimated T1 of lactate was then numerically minimized by finding the optimal TI and TR
pairs to use within fixed measurement time. Spectral editing pulses were used to separate
lactate from the contaminating lipid signals and the first lipid-free estimates of lactate T1 were
thus reported in vivo. From specific validation experiments in phantoms, the technique was
found to provide approximately 25 % improvement in measurement precision over standard IR.
The mean lactate T1 value was subsequently determined to be ( ) ms over a group
of 6 patients with high grade glioma.
In Chapter 4 a novel technique was developed which combines a diffusion-weighting
module with the 2D JPRESS technique. Typical DW-MRS experiments at 3 Tesla only investigate
the diffusion of NAA, choline and creatine due to their relative ease of measurement. JPRESS,
however, allows for resolving smaller J-coupled resonances because of the introduction of a
second spectral dimension. By combining these two techniques a more accurate measurement
of the diffusion characteristics of glutamate + glutamine, myo-inositol, glutathione and scyllo-
109
inositol was obtained. Good agreement was observed between the two techniques estimates
for the typical metabolites diffusion characteristics, namely NAA, creatine and choline in both
phantom and healthy volunteer experiments. Optimization of this technique may allow for the
estimation of additional metabolites beyond those mentioned above.
From these collective research milestones, multiple avenues of future research are
possible. Several examples are provided in the following sections.
5.2 Future Directions for CSSMRS
The CSSMRS work developed in this thesis focused exclusively on acquired spectral information
from two voxels simultaneously, as it was the simplest extension of SVS. However, extension to
a larger number of voxels is desirable in certain applications, such as when multiple disease foci
are present (eg. brain metastases). There are two main issues to address when extending
CSSMRS beyond two voxels: 1) how to perform appropriate spatial localization; and 2) whether
√ advantages are maintained due to the increase in noise from the condition number
of the reconstruction matrix. Considering the first issue, the simple localization scheme
presented in Chapter 2 does not enable additional voxels to be positioned arbitrarily. One
alternative is to make a small pulse sequence modification to cosine modulate one of the
refocusing RF pulses, which then enables four voxels to be localized. Three of these four voxels
can then be arbitrarily positioned (subject to the condition that the three voxels cannot lie
along a single line), although the position of the fourth voxel becomes very constrained. This
was implemented in preliminary experiments, as shown in Figure 5.1. This approach could also
be extended further by modulating both of the refocusing pulses, which would allow four out of
a total of eight voxels to be arbitrarily localized (subject to the condition that the four voxels
cannot lie in one plane).
In principle this implementation is straightforward, but the cosine modulation has the
effect of reducing the bandwidth of the pulses by a factor of two (assuming the refocusing
pulses are played out at the physical maximum allowed B1 amplitude). The bandwidth
110
Figure 5.1: Four voxel profile overlaid on an anatomical axial MR image. Three of the four
voxels were positioned and the fourth voxel was localized based on an algorithm that made the
voxels as square as possible.
reduction results in greater chemical shift displacement and, for J-coupled chemical species, a
reduction in the measured signal amplitude due to partial cancellation at the borders of the
voxel. A more robust solution is the arbitrary excitation of multiple voxels through the
excitation k-space approach72. Mapping out excitation k-space is typically quite slow, however,
so it is likely that a parallel excitation approach utilizing an array of transmit coils168 such as
Transmit SENSE169 (also known as “parallel RF transmission”) is needed to reduce the duration
of the excitation pulse.
Considering now the second issue, acquiring more voxels with CSSMRS while using the
same multi-channel receiver coil is expected to increase the g-factor, and because CSSMRS is
only beneficial when the g-factor is less than the square root of the number of voxels, future
multi-voxel implementations will need to investigate the g-factor dependence in detail. The g-
factor increase relates to the increase in the condition number of the matrix ( ) in
111
Equation 2.5 as the number of voxels increases, due to the increase in overlap of the signal
measured by each coil. In principle, the g-factor can be improved by increasing the number of
coil elements and reducing the sensitive volume of each element. Coil channel count can be
increased in a reasonably straightforward manner from the 8-channel configuration
investigated in Chapter 2, as head coils with up to 64 channels are now commercially available.
Additionally the sensitivity profile of coils at higher static magnetic field strengths (e.g., 7 T)
typically varies more steeply which will result in a reduction in g-factor.
There are also potential advantages to combining the CSSMRS technique with 2D
JPRESS, as the alternative of combining 2D JPRESS with MRSI is very challenging. For example,
2D JPRESS usually acquires between 1 and 8 repetitions (TR intervals) for signal averaging at
each TE step, whereas an MRSI acquisition with 8 by 8 voxels (as is often used) requires at least
64 TR intervals for spatial encoding. For a typical 1D MRS experiment at a single TE value, 64 TR
intervals is acceptable and leaves some experiment time for signal averaging. For a 2D JPRESS
experiment, however, 64 TR intervals per TE step leads to an experiment time of ~3 hours.
Instead, CSSMRS could easily be applied to JPRESS to permit acquisition of 2D spectra from two
(or possibly more) separate voxels simultaneously. There are some artefacts introduced from
the CSSMRS approach138,139, and so a careful investigation of how CSSMRS impacts the ability to
extract signals from brain metabolites present at small concentrations is needed.
5.3 Future Directions for DW-JPRESS
Before DW-JPRESS is applied to investigate changes in the diffusion of metabolites associated
with neuropathology, there is substantial technical optimization that could be performed on
the pulse sequence. In particular, it will be important to reduce the echo time, as well as a
combination of (1) the total scan duration; 2) the voxel size; and most importantly 3) the
variability of the estimated ADCs. There is an inherent tradeoff between these latter three
factors, which must be considered carefully. Given the long experiment times associated with
the current implementation of DW-JPRESS in Chapter 4 (75 minutes), these pulse sequence
optimizations are preferable before progressing to clinical research applications.
112
Reducing the minimum TE value has the effect of increasing the SNR in DW-JPRESS
(substantially for some of the interesting metabolites such as GABA, which has a T2 of 80 ms166).
Eddy currents, which are dependent on the gradient plateau duration, , will also be reduced.
The minimum TE value used in Chapter 4 was 74 ms, which can be greatly reduced by moving to
an MRI system with higher maximum gradient amplitude, either in a whole-body or head insert
gradient configuration. Additional reduction of the minimum TE value is also potentially
possible by increasing the gradient slew rate, which in the present work was kept to a quarter
of its maximum value to reduce vibromechanical motions that can corrupt diffusion-weighted
measurements. Preliminary experiments found that there was substantial table motion when
the MRI system used in this thesis was run with a higher slew rate. Assuming vibromechanical
effects can be overcome, using the maximum slew rate available on whole-body MRI systems
(200 T/s/m) and a maximum gradient amplitude of 80 mT/m (a factor of 2 increase compared
to the hardware used in Chapter 4) theoretically it should be possible to reduce the minimum
TE value to ~60 ms while keeping the same b-value. This is a larger minimum echo time than is
typically used for DW-MRI due to the reduced ADCs of metabolites as compared to water.
Nevertheless this reduction in echo time could improve SNR by ~15% for the fast-decaying
metabolites such as GABA. The minimum TE value could be reduced even further by including a
bipolar pair of gradients between the two refocusing pulses.
Furthermore, the choice of TE step (chosen in Chapter 4 to be 2 ms, consistent with
previous JPRESS implementations), and TR could be optimized for a particular metabolite of
interest (provided both the T1 and T2 are known). Increasing the TE step has the effect of
greater spectral resolution in the J-resolved direction, which will increase separation of
metabolites. However the increase in spectral resolution will reduce the SNR significantly due
to increased T2-weighting. Thus there is some optimal TE step that provides the minimum
measurement uncertainty for a given scan time. In principle a numerical optimization
procedure should be undertaken analogous to that of Chapter 3, that includes all of interwoven
experimental factors to minimize the expected uncertainty on the estimated ADC of one (or the
average of multiple) metabolites of interest with the uncertainties on individual metabolites
estimated using Cramer-Rao Lower Bounds (CRLBs)170,171. It has previously been shown that the
113
optimization of CRLBs in a typical JPRESS experiment had a reduction in the coefficients of
variation for Glutamate of approximately 30 %171. Thus, combining optimal pulse sequence
parameters with increased maximum gradient amplitude and slew rate should allow reduction
in the total experiment time by a factor of two or more while maintaining reliable
measurements of metabolite ADCs. Additionally, by moving to a magnetic field strength of 7 T,
which provides increased SNR119 as well as the greater spectral separation, the experiment time
or voxel size could be further reduced.
The original motivation for development of DW-JPRESS in this thesis was to measure the
change in the diffusion properties of neurometabolites including lactate, associated with early
radiation treatment response. Upon developing a proof-of-principle DW-JPRESS sequence it
became apparent that this research application was premature given the suboptimal nature of
the prototype and the long experiment time. Alternatively, a separation of lactate and lipid
could be achieved through a biexponential model due to their substantially different diffusion
rates152. In addition, as lactate is one of the few metabolites that resides primarily in the
extracellular space it is unclear whether investigating the nature of its diffusion is beneficial.
DW-JPRESS does, however, provide the ability to measure other J-coupled resonances not
normally available at 3 T, which could be useful for a variety of other applications.
With the implementation of technical improvements, several interesting applications
could be explored with DW-JPRESS. It was shown in Chapter 4 that even with the proof-of-
principle version, the ADC of glutamate/glutamine was reliably quantified. This suggests that
DW-JPRESS could be combined with functional MRS (which combines functional magnetic
resonance-like tasks with MRS acquisition to investigate how the metabolite profiles change
with brain activation), to determine if glutamate changes with activation. It has previously been
observed that the ADCs of NAA, Cr and Cho increase by varying amounts with activation in the
visual cortex107. It would be interesting to test the hypothesis that glutamate, due to its role as
the main excitatory neurotransmitter46, exhibits increased diffusion during neuronal activation,
when it primarily resides in the synaptic cleft. Furthermore DW-JPRESS may be used to provide
information about microstructural changes in pathologies which are known to have a large
114
disruption of normal metabolites, such as schizophrenia45, bi-polar, post-traumatic stress
disorder, obsessive-compulsive disorder and unipolar major depression172, as well as stroke173.
5.4 Early Radiation Treatment Response
It has previously been shown that brain cancer patients who had an increase in the ratio of
Cho/NAA at the 3rd week of treatment as compared to the start of treatment had a significantly
greater chance of early progression69. This ratio of choline and NAA is dependent on the
underlying concentration of the two metabolites, as well as the respective T1 and T2 values. It is
therefore hypothesized that by using spectroscopic methods to quantify T1, T2 and
concentration of choline and NAA it will provide improved ability to detect response to
radiation therapy compared to the use of metabolite ratios. A similar experiment to the one
which showed a correlation between Cho/NAA and treatment response69 should be undertaken
which measures the quantitative values using the procedure described in Chapter 3. In
addition, an optimized measurement of T2 values, as well as the ADCs of the metabolites, since
these will be susceptible to changes in tissue microstructure and could be a useful additional
biomarker of therapeutic response. The CSSMRS method (as developed in Chapter 2) with
BASING lactate editing (as used in Chapter 3) and the recommended TR of 2830 ms (based on
the T1 of lactate from Chapter 3) should be performed to measure the lipid-free lactate levels,
as well as to provide a control by measuring the contralateral normal-appearing brain tissue. In
addition to investigate the diffusion characteristics of metabolites in response to radiation an
optimized DW-JPRESS sequence should be used, as developed in Chapter 4 (provided the scan
duration can be reduced sufficiently). All the above measurements, including the clinical MRI
protocol need to be completed in less than one hour and ideally repeated weekly throughout
the radiation treatment.
Previously an increase in the Cho/NAA ratio after the 3rd week of radiation therapy
resulted in a hazard ratio of 2.7269. Using this as a lower bound for the hazard ratio (as we will
be at least as sensitive to metabolite changes as this previous work), a sample size of 31
patients should be used for this study using the proportional-hazards regression model174, with
a type I error rate of 0.05 and a type II error rate of 0.20. By making these quantitative
115
measurements it will be possible to investigate which values (or combination of values) provide
the best predictor for early radiation treatment response and the earliest time point at which
significant changes can be measured. Ultimately, this protocol could provide information useful
for clinicians in determining the optimal radiation treatment course for patients with brain
cancer. This could save non-responding patients unnecessary treatment, thereby improving
their quality of life in the short term and reducing the financial burden on our healthcare
system. Alternatively non-responders could also be selected for other treatment regimens such
as chemotherapy agents, with the hope of extending survival.
5.5 Final Remarks
Although the phenomenon of NMR has been known about for over 70 years (and much of the
underlying theoretical physics was developed by the early 1950s) the field continues to grow
and make profound impact in applications involving solid state physics, organic and physical
chemistry and, as discussed in this thesis, medicine. Over the past 30 years, in vivo MRS has
been developed and applied to a wide variety of pathologies in both the brain (as discussed
here) and other parts of the body. Continual improvements in MRS capabilities in terms of
hardware and technical developments as well as the demonstration of its benefits are allowing
MRS to become more applicable in a clinical setting. It will be interesting to see how MRS grows
in the coming decades.
116
Bibliography 1. Malom H. Levit Spin Dynamics. (2008).
2. Sørensen, O., Eich, G., Levitt, M., Bodenhausen, G. & Ernst, R. Product operator
formalism for the description of NMR pulse experiments. Progress in NMR Spectroscopy
16, 163–192 (1984).
3. Hahn, E. Spin echoes. Physical Review 80, 580–594 (1950).
4. Bottomley, P. A. Spatial Localization in NMR Spectroscopy in Vivo. Ann N Y Acad Sci 508,
333–348 (1987).
5. Haase, A., Frahm, J., Matthaei, D., Hanicke, W., Bomsdorf, H., Dietmar, K., Tischler, R. MR
imaging using stimulated echoes (STEAM). Radiology 160, 787–790 (1986).
6. McDonald, G. & Leigh, J. A new method for measuring longitudinal relaxation times.
Journal of Magnetic Resonance (1969) 362, 358–362 (1973).
7. Carr, H. & Purcell, E. Effects of diffusion on free precession in nuclear magnetic
resonance experiments. Physical review 94, 630–638 (1954).
8. Lee, H. K., Yarnan, A. & Nalcioglu, O. Homonuclear J-Refocused Spectral Editing
Technique for Quantification of Glutamine and Glutamate bv 1H NMR Spectroscopy.
Magn Reson Med 34, 253–259 (1995).
9. Star-Lack, J., Nelson, S. J., Kurhanewicz, J., Huang, L. R. & Vigneron, D. B. Improved water
and lipid suppression for 3D PRESS CSI using RF band selective inversion with gradient
dephasing (BASING). Magn Reson Med 38, 311–321 (1997).
10. Bax, A. & Lerner, L. Two-dimensional nuclear magnetic resonance spectroscopy. Science
232, 960–967 (1986).
117
11. Aue, W., Bartholdi, E. & Ernst, R. Two‐dimensional spectroscopy. Application to nuclear
magnetic resonance. The Journal of Chemical Physics 64, 2229–2246 (1976).
12. Macura, S. & Ernst, R. R. Elucidation of cross relaxation in liquids by two-dimensional
N.M.R. spectroscopy. Molecular Physics 41, 95–117 (1980).
13. Thomas, M. A., Hattori, N., Umeda, M., Sawada, T. & Naruse, S. Evaluation of two-
dimensional L-COSY and JPRESS using a 3 T MRI scanner: from phantoms to human brain
in vivo. NMR Biomed 16, 245–51 (2003).
14. Schulte, R. F. & Boesiger, P. ProFit: two-dimensional prior-knowledge fitting of J-resolved
spectra. NMR Biomed 19, 255–63 (2006).
15. Fuchs, A., Boesiger, P., Schulte, R. F. & Henning, A. ProFit revisited. Magn Reson Med 71,
458–68 (2014).
16. Schulte, R. F., Lange, T., Beck, J., Meier, D. & Boesiger, P. Improved two-dimensional J-
resolved spectroscopy. NMR Biomed 19, 264–70 (2006).
17. Prost, R. W. Magnetic resonance spectroscopy. Medical Physics 35, 4530 (2008).
18. Arnold, J. T., Dharmatti, S. S. & Packard, M. E. Chemical Effects on Nuclear Induction
Signals from Organic Compounds. The Journal of Chemical Physics 19, 507 (1951).
19. Haase, A., Frahm, J., Hänicke, W. & Matthaei, D. 1H NMR chemical shift selective (CHESS)
imaging. Phys Med Biol 30, 341–344 (1985).
20. Klose, U. In Vivo Proton Spectroscopy in Presence of Eddy Currents. Magn Reson Med 30,
26–30 (1990).
21. Jehenson, P., Westphal, M. & Schuff, N. Analytical Method for the Compensation of
Eddy-Current Effects Induced by Pulsed Magnetic Field Gradients in NMR Systems. Jour
Magn Reson 90, 264–278 (1990).
118
22. Kreis, R. Issues of spectral quality in clinical 1H-magnetic resonance spectroscopy and a
gallery of artifacts. NMR Biomed 17, 361–81 (2004).
23. Felblinger, J., Kreis, R. & Boesch, C. Effects of physiologic motion of the human brain
upon quantitative 1 H-MRS : analysis and correction by retro- gating†. NMR Biomed 11,
107–114 (1998).
24. Lange, T., Dydak, U., Roberts, T. P. L., Bjeljac, M. & Boesiger, P. Pitfalls in Lactate
Measurements at 3T. Am J Neuroradiol 27, 895–901 (2006).
25. Edden, R. a E., Schär, M., Hillis, A. E. & Barker, P. B. Optimized Detection of Lactate at
High Fields Using Inner Volume Saturation. Magn Reson Med 56, 912–917 (2006).
26. Starck, G., Carlsson, A., Ljungberg, M. & Forssell-Aronsson, E. k-space analysis of point-
resolved spectroscopy (PRESS) with regard to spurious echoes in in vivo (1)H MRS. NMR
Biomed 22, 137–47 (2009).
27. Govindaraju, V., Young, K. & Maudsley, A. A. Proton NMR chemical shifts and coupling
constants for brain metabolites. NMR Biomed 13, 129–153 (2000).
28. Coyle, J. T. The Nagging Question of the Function of N-Acetylaspartylglutamate.
Neurobiology of Disease 4, 231–238 (1997).
29. Brand, A., Richter-Landsberg, C. & Leibfritz, D. Multinuclear NMR Studies on the Energy
Metabolism of Glial and Neuronal Cells. Dev Neurosci 15, 289–298 (1993).
30. Haris, M., Cai, K., Singh, A., Hariharan, H. & Reddy, R. In vivo mapping of brain myo-
inositol. Neuroimage 54, 2079–2085 (2011).
31. Seaquist, E. & Gruetter, R. Identification of a High Concentration of Scyllo-‐inositol in the
Brain of a Healthy Human Subject Using 1H- and 13C-NMR. Magn Reson Med 39, 313–
316 (1998).
119
32. Ross, B. Biochemical Considerations in 1H Spectroscopy. Glutamate and Glutamine; Myo-
inositol and Related Metabolites. NMR Biomed 4, 59–63 (1991).
33. Kalra, S., Hanstock, C. C., Martin, W. R. W., Allen, P. S. & Johnston, W. S. Detection of
Cerebral Degeneration in Amyotrophic Lateral Sclerosis Using High-Field Magnetic
Resonance Spectroscopy. Arch Neurol 63, 1144–1148 (2006).
34. Saunders, D. E., Howe, F. A., Boogaart, A. v d, Griffiths, J. R. & Brown, M. M. Aging of the
Adult Human Brain: In Vivo Quantitation of Metabolite Content With Proton Magnetic
Resonance Spectroscopy. Jour Magn Reson Imag 9, 711–716 (1999).
35. Madan, A. et al. Proton T2 Measurement and Quantification of Lactate in Brain Tumors
by MRS at 3 Tesla In Vivo. Magn Reson Med 73, 2094–9 (2015).
36. Li, B. S. Y., Wang, H. & Gonen, O. Metabolite ratios to assumed stable creatine level may
confound the quantification of proton brain MR spectroscopy. Magn Reson Imaging 21,
923–928 (2003).
37. Dragon, F. et al. Behavioral Effects of L-a-Glycerylphosphorylcholine: Influence on
Cognitive Mechanisms in the Rat. Pharmacol Biochem Be 41, 445–448 (1992).
38. Millington, W. R. & Wurtman, R. J. Choline Administration Elevates Brain
Phosphorylcholine Concentrations. Jour Neurochem 6, 1748–1752 (1982).
39. Hasselmo, M. The Role of Acetylcholine in Learning and Memory. Curr Opin Neurobiol 16,
710–715 (2006).
40. Herholz, K. et al. In Vivo Imaging of Glucose Consumption and Lactate Concentration in
Human Gliomas. Ann Neurol 31, 319–327 (1991).
41. Lin, A.-Q., Shou, J.-X., Li, X.-Y., Ma, L. & Zhu, X.-H. Metabolic changes in acute cerebral
infarction: Findings from proton magnetic resonance spectroscopic imaging. Exp Ther
Med 7, 451–455 (2014).
120
42. Chu, W. et al. Psychiatry Research : Neuroimaging Magnetic resonance spectroscopy
imaging of lactate in patients with bipolar disorder. Psychiatry Res Neuroimaging 213,
230–234 (2013).
43. Horska A, Barker, P. B. Imaging of Brain Tumors: MR Spectroscopy and Metabolic
Imaging. Neuroimag Clin N Am 20, 293–310 (2010).
44. Graff-Radford, J. & Kantarci, K. Magnetic resonance spectroscopy in Alzheimer’s disease.
Neuropsychiatr Dis Treat 9, 687–696 (2013).
45. Port, J. D. & Agarwal, N. MR spectroscopy in schizophrenia. Jour Magn Reson Imag 34,
1251–61 (2011).
46. Petroff, O. A. C. GABA and Glutamate in the Human Brain. Neuroscientist 8, 562–573
(2002).
47. Bröer, S. et al. Alanine metabolism, transport, and cycling in the brain. J Neurochem 102,
1758–70 (2007).
48. Harrison, F. & May, J. Vitamin C Function in the Train: Vital Role of the Ascorbate
Transporter (SVCT2). Free Radic Biol Med 46, 719–730 (2009).
49. Storck, T. & Schulte, S. Structure, expression, and functional analysis of a Na(+)-
dependent glutamate/aspartate transporter from rat brain. Proc Natl Acad Sci USA 89,
10955–10959 (1992).
50. Albrecht, J., Sidoryk-Węgrzynowicz, M., Zielińska, M. & Aschner, M. Roles of glutamine in
neurotransmission. Neuron Glia Biology 6, 263–76 (2011).
51. Campistron, G., Buijs, R. & Geffard, M. Glycine neurons in the brain and spinal cord.
Antibody production and immunocytochemical localization. Brain Res 376, 400–405
(1986).
121
52. Dringen, R. Metabolism and functions of glutathione in brain. Prog Neurobiol 62, 649–
671 (2000).
53. Bakovic, M., Fullerton, M. D. & Michel, V. Metabolic and molecular aspects of
ethanolamine phospholipid biosynthesis: the role of CTP: phosphoethanolamine
cytidylyltransferase (Pcyt2). Biochem Cell Biol 300, 283–300 (2007).
54. Jia, F. et al. Taurine is a potent activator of extrasynaptic GABA(A) receptors in the
thalamus. J Neurosci 28, 106–15 (2008).
55. Porter, K. R., McCarthy, B. J., Freels, S., Kim, Y. & Davis, F. G. Prevalence estimates for
primary brain tumors in the United States by age, gender, behavior, and histology.
Neuro-Oncology 12, 520–7 (2010).
56. DeAngelis, L. Brain Tumors. N Engl J Med 344, 114–123 (2001).
57. Ricard, D. et al. Primary brain tumours in adults. Lancet 379, 1984–1996 (2012).
58. Gropler, R. et al. Comparison of Carbon-11-Acetate With Fluorine-18-Fluorodeoxyglucose
for Delineating Viable Myocardium by Positron Emission Tomography. JACC 22, 1587–
1597 (1993).
59. Smith, J. S. et al. Role of extent of resection in the long-term outcome of low-grade
hemispheric gliomas. Journal of clinical oncology : official journal of the American Society
of Clinical Oncology 26, 1338–45 (2008).
60. Terpstra, M. et al. Relationships Among Lactate Concentration, Blood Flow and
Histopathologic Profiles in Rat C6 Glioma. NMR Biomed 9, 185–194 (1996).
61. Kousi, E. et al. Spectroscopic Evaluation of Glioma Grading at 3T: The Combined Role of
Short and Long TE. Sci World J 2012, 1–11 (2012).
122
62. Wang, W. et al. Evaluation of the Diagnostic Performance of Magnetic Resonance
Spectroscopy in Brain Tumors: A Systematic Review and Meta-Analysis. PloS one 9,
e112577 (2014).
63. Wang, X. et al. Comparison of magnetic resonance spectroscopy and positron emission
tomography in detection of tumor recurrence in posttreatment of glioma: A diagnostic
meta-analysis. Asia Pac J Clin Oncol 11, 97–105 (2015).
64. Choi, C. et al. 2-hydroxyglutarate detection by magnetic resonance spectroscopy in IDH-
mutated patients with gliomas. Nat Med 18, 624–629 (2012).
65. Yan, H., Parsons, D. & Jin, G. IDH1 and IDH2 mutations in gliomas. Curr Neurol Neurosci
Rep 13, 345–357 (2009).
66. Theony, H. C. & Ross, B. D. Predicting and Monitoring Cancer Treatment Response with
DW-MRI. J Magn Reson Imaging 32, 2–16 (2010).
67. Al-Mefty, O., Kersh, J., Routh, A. & Smith, R. The long-term side effects of radiation
therapy for benign brain tumors in adults. Journal of neurosurgery 73, 502–512 (1990).
68. Sagiyama, K. et al. In vivo chemical exchange saturation transfer imaging allows early
detection of a therapeutic response in glioblastoma. PNAS 111, 4542–7 (2014).
69. Muruganandham, M. et al. 3D-MR Spectroscopic Imaging at 3Tesla for Early Response
Assessment of Glioblastoma Patients during External Beam Radiation Therapy. Int J
Radiat Oncol Biol Phys 90, 181–189 (2014).
70. Verma, N., Cowperthwaite, M. C., Burnett, M. G. & Markey, M. K. Differentiating tumor
recurrence from treatment necrosis: a review of neuro-oncologic imaging strategies.
Neuro-oncology 15, 515–34 (2013).
71. Fink, K. R. & Fink, J. R. Imaging of brain metastases. Surg Neurol Int 4, S209–19 (2013).
123
72. John Pauly, Dwight Nishimura, A. M. A k-Space Analysis of Small-Tip-Angle Excitation.
Jour Magn Reson 81, 43–56 (1989).
73. Nishimura, D. G. Principles of Magnetic Resonance Imaging. (2010).
74. Shinnar, M., Bolinger, L. & Leigh, J. S. The Use of Finite Impulse Response Filters in Pulse
Design. Magnetic Resonance in Medicine 12, 81–87 (1989).
75. Pauly, J., Le Roux, P., Nishimura, D. & Macovski, A. Parameter Relations for the Shinnar-
Le Roux Selective Excitation Pulse Design Algorithm. IEEE Trans Med Imag 10, 53–65
(1991).
76. Silver, M. S., Joseph, R. I., Chen, C. N., Sank, V. J. & Hoult, D. I. Selective population
inversion in NMR. Nature 310, 681–683 (1984).
77. Silver, M. S., Joseph, R. I. & Hoult, D. I. Highly Selective pi/2 and pi Pulse Generation. Jour
Magn Reson 59, 347–351 (1984).
78. Hennig, J. Echoes—how to generate, recognize, use or avoid them in MR-imaging
sequences. Part I: Fundamental and not so fundamental properties of spin echoes.
COncept Magnetic Res 3, 125–143 (1991).
79. Academy, N. & States, U. NMR chemical shift imaging in three dimensions. Proc Natl
Acad Sci USA 79, 3523–3526 (1982).
80. Maudsley, A. A., Hilal, S. K., Perman, W. H. & Simon, H. E. Spatially Resolved High
Resolution Spectroscopy by “Four-Dimensional” NMR. Jour Magn Reson 51, 147–152
(1983).
81. Frahm, J. et al. Localized Proton NMR Spectroscopy in Different Regions of the Human
Brain in Viva Relaxation Times and Concentrations of Cerebral Metabolites. Magn Reson
Med 63, 47–63 (1989).
124
82. Spielman, D., Pauly, J., Macovski, A. & Enzmann, D. Spectroscopic Imaging with
Multidimensional Pulses for Excitation: SIMPLE. Magn Reson Med 19, 67–84 (1991).
83. Segebarth, C. M., Balériaux, D. F., Luyten, P. R. & Den Hollander, J. A. Detection of
Metabolic Heterogeneity of Human Intracranial Tumors in Vivo by 1H NMR Spectroscopic
Imaging. Magn Reson Med 13, 62–76 (1990).
84. Roemer, P. B., Edelstein, W. a, Hayes, C. E., Souza, S. P. & Mueller, O. M. The NMR
Phased Array. Magn Reson Med 16, 192–225 (1990).
85. Griswold, M. a et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA).
Magn Reson Med 47, 1202–10 (2002).
86. Pruessmann, K. P., Weiger, M., Scheidegger, M. B. & Boesiger, P. SENSE: Sensitivity
Encoding for Fast MRI. Magn Reson Med 42, 952–62 (1999).
87. Lin, F.-H. et al. Sensitivity-encoded (SENSE) proton echo-planar spectroscopic imaging
(PEPSI) in the human brain. Magnetic Resonance in Medicine 57, 249–57 (2007).
88. Bonekamp, D., Smith, M. a, Zhu, H. & Barker, P. B. Quantitative SENSE-MRSI of the
human brain. Magn Reson Imaging 28, 305–13 (2010).
89. Kuo, A. Y.-C., Chiew, M., Tam, F., Cunningham, C. & Graham, S. J. Multiecho Coarse Voxel
Acquisition for Neurofeedback fMRI. Magn Reson Med 65, 715–24 (2011).
90. Chiew, M. & Graham, S. J. Constrained Source Space Imaging: Application to Fast, Region-
Based Functional MRI. Magn Reson Med 70, 1058–1069 (2012).
91. Provencher, S. Estimation of Metabolite Concentrations from Localized in Vivo Proton
NMR spectra. Magn Reson Med 30, 672–679 (1993).
92. Naressi, A., Couturier, C., Castang, I., De Beer, R. & Graveron-Demilly, D. Java-based
graphical user interface for MRUI, a software package for quantitation of in vivo/medical
magnetic resonance spectroscopy signals. Comput Biol Med 31, 269–86 (2001).
125
93. Poullet, J. & Sima, D. An automated quantitation of short echo time MRS spectra in an
open source software environment: AQSES. NMR Biomed 20, 493–504 (2007).
94. Reynolds, G., Wilson, M., Peet, A. & Arvanitis, T. N. An Algorithm for the Automated
Quantitation of Metabolites in in Vitro NMR Signals. Magn Reson Med 56, 1211–1219
(2006).
95. Longo, R., Pollesello, P. & Ricci, C. Proton MR Spectroscopy in Quantitative in Vivo
Determination of Fat Content in Human liver Steatosis. Jour Magn Reson Imag 5, 281–
285 (1995).
96. Li, C., Kuo, Y., Chen, CY Kuo, YT, Chiu, YY She, F. & Liu, G. Quantification of choline
compounds in human hepatic tumors by proton MR spectroscopy at 3 T. Magn Reson
Med 776, 770–776 (2005).
97. Woessner, D. E. Effects of Diffusion in Nuclear Magnetic Resonance Spin-Echo
Experiments. J Chem Phys 34, 2057 (1961).
98. Stejskal, E. O. & Tanner, J. E. Spin Diffusion Measurements: Spin Echoes in the Presence
of a Time-Dependent Field Gradient. J Chem Phys 42, 288 (1965).
99. Bihan, D. Molecular Diffusion, Tissue Microdynamics and Microstructure. NMR Biomed 8,
375–386 (1995).
100. Ercan, A. E., Techawiboonwong, A., Versluis, M. J., Webb, A. G. & Ronen, I. Diffusion-
weighted chemical shift imaging of human brain metabolites at 7T. Magn Reson Med 73,
2053–2061 (2015).
101. Posse, S., Cuenod, C. & Bihan, D. Le Human Brain: Proton Diffusion MR Spectroscopy.
Radiology 188, 719–725 (1993).
102. Ellegood, J., Hanstock, C. C. & Beaulieu, C. Trace apparent diffusion coefficients of
metabolites in human brain using diffusion weighted magnetic resonance spectroscopy.
Magn Reson Med 53, 1025–32 (2005).
126
103. Najac, C., Branzoli, F., Ronen, I. & Valette, J. Brain intracellular metabolites are freely
diffusing along cell fibers in grey and white matter, as measured by diffusion-weighted
MR spectroscopy in the human brain at 7 T. Brain Struct Funct 221, 1245–1254 (2014).
104. Ronen, I., Ercan, E. & Webb, A. Axonal and glial microstructural information obtained
with diffusion-weighted magnetic resonance spectroscopy at 7T. Front Integr Neurosci 7,
1–9 (2013).
105. Wood, E. T. et al. Reproducibility and optimization of in vivo human diffusion-weighted
MRS of the corpus callosum at 3 T and 7 T. NMR Biomed 28, 976–87 (2015).
106. Ercan, E. et al. Glial and axonal changes in systemic lupus erythematosus measured with
diffusion of intracellular metabolites. Brain 1–11 (2016).
107. Branzoli, F., Techawiboonwong, A., Kan, H., Webb, A. & Ronen, I. Functional diffusion-
weighted magnetic resonance spectroscopy of the human primary visual cortex at 7 T.
Magn Reson Med 69, 303–309 (2013).
108. Kan, H. E. et al. Differences in Apparent Diffusion Coefficients of Brain Metabolites
Between Grey and White Matter in the Human Brain Measured at 7 T. Magn Reson Med
67, 1203–1209 (2012).
109. Cao, P. & Wu, E. X. In vivo diffusion MRS investigation of non-water molecules in
biological tissues. NMR Biomed [epub] (2016).
110. Gu, M., Liu, C. & Spielman, D. M. Parallel Spectroscopic Imaging Reconstruction with
Arbitrary Trajectories Using k-space Sparse Matrices. Magn Reson Med 61, 267–72
(2009).
111. Posse, S., Otazo, R., Tsai, S.-Y., Yoshimoto, A. E. & Lin, F.-H. Single-shot magnetic
resonance spectroscopic imaging with partial parallel imaging. Magn Reson Med 61,
541–547 (2009).
127
112. Oshio, K., Kyriakos, W. & Mulkern, R. V Line Scan Echo Planar Spectroscopic Imaging.
Magn Reson Med 44, 521–524 (2000).
113. Frahm, J., Merboldt, K.-D. & Hänicke, W. Localized Proton Spectroscopy using Stimulated
Echoes. Jour Magn Reson 72, 502–508 (1987).
114. Inglese, M. et al. Field, Coil, and Echo-Time Influence on Sensitivity and Reproducibility of
Brain Proton MR Spectroscopy. Am J Neuroradiol 27, 684–8 (2006).
115. Jin, J., Liu, F., Weber, E., Li, Y. & Crozier, S. An electromagnetic reverse method of coil
sensitivity mapping for parallel MRI–Theoretical framework. Jour Magn Reson 207, 59–
68 (2010).
116. M. J. Allison, S. Ramani, J. A. F. Regularized MR Coil Sensitivity Estimation Using
Augmented Lagrangian Methods. International Symposium on Biomedical Imaging (ISBI)
32, 394–397 (2012).
117. Allison, M. J., Ramani, S. & Fessler, J. a Accelerated Regularized Estimation of MR coil
Sensitivities Using Augmented Lagrangian Methods. IEEE Trans Med Imaging 32, 556–64
(2013).
118. Chen, L., Weng, Z., Goh, L. & Garland, M. An efficient algorithm for automatic phase
correction of NMR spectra based on entropy minimization. Jour Magn Reson 158, 164–
168 (2002).
119. Li, Y. et al. T1 and T2 Metabolite Relaxation Times in Normal Brain at 3T and 7T. J Mol
Imag Dynamic 1–5 (2013).
120. Ryner, L., Sorenson, J. & Thomas, M. Localized 2D J-resolved 1 H MR spectroscopy:
Strong coupling effects in vitro and in vivo. Magn Reson Imaging 13, 853–869 (1995).
121. Gill SS, Thomas DG, Van Bruggen N, Gadian DG, Peden CJ, Bell JD, Cox IJ, Menon DK, Iles
RA, Bryant DJ, et al. Proton MR Spectroscopy of Intracranial Tumours: In Vivo and In Vitro
Studies. J Comput Assist Tomogr 14, 497–504 (1990).
128
122. Li, Y. et al. Comparison of T1 and T2 metabolite relaxation times in glioma and normal
brain at 3T. J Magn Reson Imaging 28, 342–350 (2008).
123. Sijens, P. E. & Oudkerk, M. 1H chemical shift imaging characterization of human brain
tumor and edema. Eur Radiol 12, 2056–61 (2002).
124. Ethofer, T. et al. Comparison of Longitudinal Metabolite Relaxation Times in Different
Regions of the Human Brain at 1.5 and 3 Tesla. Magn Reson Imaging 50, 1296–1301
(2003).
125. Mylnarik, V., Gruber, S. & Moser, E. Proton T1 and T2 relaxation times of human brain
metabolites at 3 Tesla. NMR Biomed 14, 325–331 (2001).
126. Look, D. C. & Locker, D. R. Time Saving in Measurement of NMR and EPR Relaxation
Times. Rev Sci Instrum 41, 250–251 (1970).
127. Gupta, R. K., Ferretti, J. A., Becker, E. D. & Weiss, G. H. A Modified Fast Inversion-
Recovery Technique for Spin-Lattice Relaxation Measurements *. Jour Magn Reson 38,
447–452 (1980).
128. Star-Lack, J. et al. In Vivo Lactate Editing with Simultaneous Detection of Choline,
Creatine, NAA, and Lipid Singlets at 1.5 T Using PRESS Excitation with Applications to the
Study of Brain and Head and Neck Tumors. J Magn Reson 133, 243–254 (1998).
129. Kelley, D. A. C., Wald, L. L. & Star-lack, J. M. Lactate Detection at 3T : Compensating J
Coupling Effects With BASING. J Magn Reson Imaging 9, 732–737 (1999).
130. Ogg, R. J. & Kingsley, P. B. Optimized Precision of Inversion-Recovery T1 Measurements
for Constrained Scan Time. Magn Reson Med 51, 625–30 (2004).
131. Chen, A. P. et al. T1 measurement of brain metabolites at 3T with a saturated-inversion
recovery method. Proc Intl Soc Mag Reson Med 15, 1376 (2007).
129
132. Park, I. et al. Implementation of 3 T Lactate-Edited 3D 1H MR Spectroscopic Imaging with
Flyback Echo-Planar Readout for Gliomas Patients. Ann Biomed Eng 39, 193–204 (2011).
133. Schirmer, T. & Auer, D. P. On the reliability of quantitative clinical magnetic resonance
spectroscopy of the human brain. NMR Biomed 13, 28–36 (2000).
134. Träber, F., Block, W., Lamerichs, R., Gieseke, J. & Schild, H. H. 1H Metabolite Relaxation
Times at 3.0 Tesla: Measurements of T1 and T2 Values in Normal Brain and
Determination of Regional Differences in Transverse Relaxation. J Magn Reson Imaging
19, 537–45 (2004).
135. Minati, L., Aquino, D., Bruzzone, M. & Erbetta, A. Quantitation of normal metabolite
concentrations in six brain regions by in-vivo 1 H-MR spectroscopy. J Med Phys 35, 154–
163 (2010).
136. Howe, F. A. et al. Metabolic Profiles of Human Brain Tumors Using Quantitative In Vivo
1H Magnetic Resonance Spectroscopy. Magn Reson Med 49, 223–232 (2003).
137. Bolinger, L. & Leigh, J. S. Hadamard Spectroscopic Imaging (HSI) for Multivolume
Localization. J Magn Reson 80, 162–167 (1988).
138. Landheer, K., Sahgal, A., Das, S. & Graham, S. J. Constrained Source Space MR
Spectroscopy: Multiple Voxels, No Gradient Readout. Am J Neuroradiol 37, 1–8 (2015).
139. Boer, V. O., Klomp, D. W. J., Laterra, J. & Barker, P. B. Parallel Reconstruction in
Accelerated Multivoxel MR Spectroscopy. Magn Reson Med 74, 599–606 (2015).
140. Payne, G. S., DeSouza, N. M., Messiou, C. & Leach, M. O. Single-shot single-voxel lactate
measurements using FOCI-LASER and a multiple-quantum filter. NMR Biomed 28, 496–
504 (2015).
141. Near, J., Simpson, R., Cowen, P. & Jezzard, P. Efficient γ-aminobutyric acid editing at 3T
without macromolecule contamination: MEGA-SPECIAL. NMR Biomed 24, 1277–1285
(2011).
130
142. Andreychenko, A., Boer, V. O., Arteaga de Castro, C. S., Luijten, P. R. & Klomp, D. W. J.
Efficient Spectral Editing at 7 T: GABA Detection with MEGA-sLASER. Magn Reson Med
68, 1018–1025 (2012).
143. Nicolay, K., Braun, K. P. J., Graaf, R. A. De, Dijkhuizen, R. M. & Kruiskamp, M. J. Diffusion
NMR spectroscopy. NMR Biomed 14, 94–111 (2001).
144. Harada, M. et al. Diffusion-weighted in vivo localized proton MR spectroscopy of human
cerebral ischemia and tumor. NMR Biomed 15, 69–74 (2002).
145. Ellegood, J., Hanstock, C. C. & Beaulieu, C. Diffusion Tensor Spectroscopy of Myo-Inositol
in Human Brain. Proc Intl Soc Mag Reson Med 16, 573 (2008).
146. Dreher, W., Busch, E. & Leibfritz, D. Changes in Apparent Diffusion Coefficients of
Metabolites in Rat Brain After Middle Cerebral Artery Occlusion Measured by Proton
Magnetic Resonance Spectroscopy. Magn Reson Med 389, 383–389 (2001).
147. Pfeuffer, J., Tkáč, I. & Gruetter, R. Extracellular–intracellular distribution of glucose and
lactate in the rat brain assessed noninvasively by diffusion-weighted 1H nuclear magnetic
resonance. J Cereb Blood Flow Metab 20, 736–746 (2000).
148. Mescher, M., Merkle, H., Kirsch, J., Garwood, M. & Gruetter, R. Simultaneous in vivo
spectral editing and water suppression. NMR Biomed 272, 266–272 (1998).
149. Sotak, C. H. A Method for Measuring the Apparent Self-Diffusion Coefficient of in Vivo
Lactic Acid Using Double-Quantum Coherence-Transfer Spectroscopy. Jour Magn Reson
90, 198–204 (1990).
150. Sotak, C. & Moore, S. A volume-localized, two-dimensional NMR method for lactate-
edited diffusion coefficient measurements using zero-quantum coherence-transfer. Jour
Magn Reson 92, 581–589 (1991).
131
151. Trimble, L., Shen, J., Wilman, A. & Allen, P. Lactate editing by means of selective-pulse
filtering of both zero-and double-quantum coherence signals. Jour Magn Reson 86, 191–
198 (1990).
152. Wang, A. M. et al. Separation and quantification of lactate and lipid at 1.3 ppm by
diffusion-weighted magnetic resonance spectroscopy. Magn Reson Med 77, 480–489
(2017).
153. Yue, K., Marumoto, A., Binesh, N. & Thomas, M. A. 2D JPRESS of human prostates using
an endorectal receiver coil. Magn Reson Med 47, 1059–64 (2002).
154. Posse, S., DeCarli, C. & Bihan, D. Le Three-dimensional echo-planar MR spectroscopic
imaging at short echo times in the human brain. Radiology 192, 733–738 (1994).
155. Smith, S., Levante, T., Meier, B. & Ernst, R. Computer simulations in magnetic resonance.
An object-oriented programming approach. J Magn Reson Ser A 75–105 (1994).
156. Cuenod, C., Posses, S., Mattiello, J. & Le Bihan, D. Relationship between Diffusion and
Temperature for N-Acetyl Aspartate and Choline : A Localized 1H NMR Spectroscopy
study. Proc. SMRM 11th Annual Meeting, Berlin 1215 (1992).
157. Holz, M., Heila, S. R. & Saccob, A. Temperature-dependent self-diffusion coefficients of
water and six selected molecular liquids for calibration in accurate 1H NMR PFG
measurements. Phys Chem Chem Phys 2, 4740–4742 (2000).
158. Ligneul, C., Palombo, M. & Valette, J. Metabolite diffusion up to very high b in the mouse
brain in vivo: Revisiting the potential correlation between relaxation and diffusion
properties. Magn Reson Med [Epub ahead of print] (2016).
159. Branzoli, F., Ercan, E., Webb, A. & Ronen, I. The interaction between apparent diffusion
coefficients and transverse relaxation rates of human brain metabolites and water
studied by diffusion-weighted spectroscopy at 7 T. NMR in biomedicine 27, 495–506
(2014).
132
160. Merboldt, K., Horstermann, D., Hanicke, W., Bruhn, H. & Frahm, J. Molecular self‐
diffusion of intracellular metabolites in rat brain in vivo Investigated by Localized Proton
NMR Diffusion Spectroscopy. Magn Reson Med 125–129 (1993).
161. Wick, M., Nagatomo, Y., Prielmeier, F. & Frahm, J. Alteration of Intracellular Metabolite
Diffusion in Rat Brain In Vivo During Ischemia and Reperfusion. Stroke 26, 1930–1934
(1995).
162. Toorn, A. Van Der, Dijkhuizen, R. M., Tulleken, C. A. F. & Nicolay, K. Diffusion of
Metabolites in Normal and Ischemic Rat Brain Measured by Localized 1H MRS. Magn
Reson Med 36, 914–922 (1996).
163. Abe, O. et al. Temporal Changes of the Apparent Diffusion Coefficients of Water and
Metabolites in Rats With Hemispheric Infarction: Experimental Study of
Transhemispheric Diaschisis in the Contralateral Hemisphere at 7 Tesla. J Cereb Blood
Flow Metab 20, 726–735 (2000).
164. Valette, J. et al. Diffusion-weighted NMR Spectroscopy Allows Probing of 13C Labeling of
Glutamate Inside Distinct Metabolic Compartments in the Brain. Magn Reson Med 60,
306–11 (2008).
165. Valette, J. et al. Isoflurane strongly affects the diffusion of intracellular metabolites, as
shown by 1H nuclear magnetic resonance spectroscopy of the monkey brain. J Cereb
Blood Flow Metab 27, 588–596 (2007).
166. Edden, R. A. E., Intrapiromkul, J., Zhu, H., Cheng, Y. & Barker, P. B. Measuring T2 In Vivo
With J-Difference Editing: Application to GABA at 3 Tesla. Jour Magn Reson Imag 35,
229–234 (2012).
167. Ellegood, J., Hanstock, C. C. & Beaulieu, C. Considerations for measuring the fractional
anisotropy of metabolites with diffusion tensor spectroscopy. NMR Biomed 24, 270–80
(2011).
133
168. Zhu, Y. Parallel Excitation With an Array of Transmit Coils. Magn Reson Med 51, 775–84
(2004).
169. Katscher, U., Börnert, P., Leussler, C. & Van den Brink, J. S. Transmit SENSE. Magn Reson
Med 49, 144–50 (2003).
170. Cavassila, S., Deval, S., Huegen, C., Ormondt, D. v & Graveron-Demilly, D. Cramér–Rao
bounds: an evaluation tool for quantitation. NMR Biomed 14, 278–283 (2001).
171. Bolliger, C. S., Boesch, C. & Kreis, R. On the use of Cramér-Rao minimum variance bounds
for the design of magnetic resonance spectroscopy experiments. NeuroImage 83, 1031–
40 (2013).
172. Maddock, R. & Buonocore, M. MR Spectroscopic Studies of the Brain in Psychiatric
Disorders. Curr Topics Behav Neurosci 11, 199–251 (2012).
173. Dani, K., An, L., Henning, E., Shen, J. & Warach, S. Multivoxel MR spectroscopy in acute
ischemic stroke comparison to the stroke protocol MRI. Stroke 43, 2962–2967 (2012).
174. Schoenfeld, D. Sample-size formula for the proportional-hazards regression model.
Biometrics 39, 499–503 (1983).