technical factors or prime factors bushong ch 15 1
TRANSCRIPT
Technical Factors or Prime Factors
Bushong Ch 15
1
PRIME FACTORS
What is ldquotechniquerdquo
How does it affect the ldquoimagerdquo
2
Exposure Factors ndash 3 or 4
The four prime exposure factors are Voltage = kVp Current = mA Exposure time = seconds or fractions of a sec Source-to-image distance = SID
3
PRIME FACTORS
bull KVP
bull MAS
bull DISTANCE
4
kVp
Kilovolts controls how fast the electrons are sent across the tube
Controls quality penetrability amp contrast
Increasing kVp also increases scattered photons reducing image quality
Does kVp influence OD
5
kVp
bull Low kVp (50 ndash 60)
bull Short scale
bull High contrast
bull ldquoBone workrdquo
6
kVp
bull High kVp (90 ndash 120)
bull Long scale
bull Low contrast
bull ldquoChest imagesrdquo
7
mA
Determines the number of photons radiation quantity OD amp patient dose
Changing mA does not change the kinetic energy of e-
Available mA stations are usually 50 100 200 300 400 amp 600
8
Exposure Time
Should be kept as short as possible for most examinations To minimize the risk of patient motion
mA X s = mAs mAs controls OD
mAs determines the number of photons in the primary beam
9
Distance
Affects exposure of the IR because of the Inverse Square Law
SID largely determines the intensity of photons at the IR
Distance has no effect on radiation quality
10
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
PRIME FACTORS
What is ldquotechniquerdquo
How does it affect the ldquoimagerdquo
2
Exposure Factors ndash 3 or 4
The four prime exposure factors are Voltage = kVp Current = mA Exposure time = seconds or fractions of a sec Source-to-image distance = SID
3
PRIME FACTORS
bull KVP
bull MAS
bull DISTANCE
4
kVp
Kilovolts controls how fast the electrons are sent across the tube
Controls quality penetrability amp contrast
Increasing kVp also increases scattered photons reducing image quality
Does kVp influence OD
5
kVp
bull Low kVp (50 ndash 60)
bull Short scale
bull High contrast
bull ldquoBone workrdquo
6
kVp
bull High kVp (90 ndash 120)
bull Long scale
bull Low contrast
bull ldquoChest imagesrdquo
7
mA
Determines the number of photons radiation quantity OD amp patient dose
Changing mA does not change the kinetic energy of e-
Available mA stations are usually 50 100 200 300 400 amp 600
8
Exposure Time
Should be kept as short as possible for most examinations To minimize the risk of patient motion
mA X s = mAs mAs controls OD
mAs determines the number of photons in the primary beam
9
Distance
Affects exposure of the IR because of the Inverse Square Law
SID largely determines the intensity of photons at the IR
Distance has no effect on radiation quality
10
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Exposure Factors ndash 3 or 4
The four prime exposure factors are Voltage = kVp Current = mA Exposure time = seconds or fractions of a sec Source-to-image distance = SID
3
PRIME FACTORS
bull KVP
bull MAS
bull DISTANCE
4
kVp
Kilovolts controls how fast the electrons are sent across the tube
Controls quality penetrability amp contrast
Increasing kVp also increases scattered photons reducing image quality
Does kVp influence OD
5
kVp
bull Low kVp (50 ndash 60)
bull Short scale
bull High contrast
bull ldquoBone workrdquo
6
kVp
bull High kVp (90 ndash 120)
bull Long scale
bull Low contrast
bull ldquoChest imagesrdquo
7
mA
Determines the number of photons radiation quantity OD amp patient dose
Changing mA does not change the kinetic energy of e-
Available mA stations are usually 50 100 200 300 400 amp 600
8
Exposure Time
Should be kept as short as possible for most examinations To minimize the risk of patient motion
mA X s = mAs mAs controls OD
mAs determines the number of photons in the primary beam
9
Distance
Affects exposure of the IR because of the Inverse Square Law
SID largely determines the intensity of photons at the IR
Distance has no effect on radiation quality
10
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
PRIME FACTORS
bull KVP
bull MAS
bull DISTANCE
4
kVp
Kilovolts controls how fast the electrons are sent across the tube
Controls quality penetrability amp contrast
Increasing kVp also increases scattered photons reducing image quality
Does kVp influence OD
5
kVp
bull Low kVp (50 ndash 60)
bull Short scale
bull High contrast
bull ldquoBone workrdquo
6
kVp
bull High kVp (90 ndash 120)
bull Long scale
bull Low contrast
bull ldquoChest imagesrdquo
7
mA
Determines the number of photons radiation quantity OD amp patient dose
Changing mA does not change the kinetic energy of e-
Available mA stations are usually 50 100 200 300 400 amp 600
8
Exposure Time
Should be kept as short as possible for most examinations To minimize the risk of patient motion
mA X s = mAs mAs controls OD
mAs determines the number of photons in the primary beam
9
Distance
Affects exposure of the IR because of the Inverse Square Law
SID largely determines the intensity of photons at the IR
Distance has no effect on radiation quality
10
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
kVp
Kilovolts controls how fast the electrons are sent across the tube
Controls quality penetrability amp contrast
Increasing kVp also increases scattered photons reducing image quality
Does kVp influence OD
5
kVp
bull Low kVp (50 ndash 60)
bull Short scale
bull High contrast
bull ldquoBone workrdquo
6
kVp
bull High kVp (90 ndash 120)
bull Long scale
bull Low contrast
bull ldquoChest imagesrdquo
7
mA
Determines the number of photons radiation quantity OD amp patient dose
Changing mA does not change the kinetic energy of e-
Available mA stations are usually 50 100 200 300 400 amp 600
8
Exposure Time
Should be kept as short as possible for most examinations To minimize the risk of patient motion
mA X s = mAs mAs controls OD
mAs determines the number of photons in the primary beam
9
Distance
Affects exposure of the IR because of the Inverse Square Law
SID largely determines the intensity of photons at the IR
Distance has no effect on radiation quality
10
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
kVp
bull Low kVp (50 ndash 60)
bull Short scale
bull High contrast
bull ldquoBone workrdquo
6
kVp
bull High kVp (90 ndash 120)
bull Long scale
bull Low contrast
bull ldquoChest imagesrdquo
7
mA
Determines the number of photons radiation quantity OD amp patient dose
Changing mA does not change the kinetic energy of e-
Available mA stations are usually 50 100 200 300 400 amp 600
8
Exposure Time
Should be kept as short as possible for most examinations To minimize the risk of patient motion
mA X s = mAs mAs controls OD
mAs determines the number of photons in the primary beam
9
Distance
Affects exposure of the IR because of the Inverse Square Law
SID largely determines the intensity of photons at the IR
Distance has no effect on radiation quality
10
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
kVp
bull High kVp (90 ndash 120)
bull Long scale
bull Low contrast
bull ldquoChest imagesrdquo
7
mA
Determines the number of photons radiation quantity OD amp patient dose
Changing mA does not change the kinetic energy of e-
Available mA stations are usually 50 100 200 300 400 amp 600
8
Exposure Time
Should be kept as short as possible for most examinations To minimize the risk of patient motion
mA X s = mAs mAs controls OD
mAs determines the number of photons in the primary beam
9
Distance
Affects exposure of the IR because of the Inverse Square Law
SID largely determines the intensity of photons at the IR
Distance has no effect on radiation quality
10
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
mA
Determines the number of photons radiation quantity OD amp patient dose
Changing mA does not change the kinetic energy of e-
Available mA stations are usually 50 100 200 300 400 amp 600
8
Exposure Time
Should be kept as short as possible for most examinations To minimize the risk of patient motion
mA X s = mAs mAs controls OD
mAs determines the number of photons in the primary beam
9
Distance
Affects exposure of the IR because of the Inverse Square Law
SID largely determines the intensity of photons at the IR
Distance has no effect on radiation quality
10
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Exposure Time
Should be kept as short as possible for most examinations To minimize the risk of patient motion
mA X s = mAs mAs controls OD
mAs determines the number of photons in the primary beam
9
Distance
Affects exposure of the IR because of the Inverse Square Law
SID largely determines the intensity of photons at the IR
Distance has no effect on radiation quality
10
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Distance
Affects exposure of the IR because of the Inverse Square Law
SID largely determines the intensity of photons at the IR
Distance has no effect on radiation quality
10
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
INTENSITY IS SPREAD OUThellip
11
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Inverse Square Law Formula
Intensity 1
Intensity 2
Distance 2 - Squared
Distance 1 - Squared
12
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Direct Square Law
bull New mAs = New distance 2
Old mAs Old distance 2
13
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Focal-Spot Changes
14
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Tube voltage (kVp)
bull Determines the maximum energy in the beam
bull spectrum and affects the quality of the output spectrum
bull Efficiency of x-ray production is directly related to tube voltage
15
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Influencing factors kVp
15 rule 15 kVp = doubling of exposure to the film
15 kVp = halving of exposure to the film
15 rule will always change the contrast of the image because kV is the primary method of changing image contrast
Remember 15 change ( ) KVP has the same effect as
doubling or frac12 the MAS on density
16
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
kVp Changes
bull The kVp setting must be changed by at least 4 to produce visual changes an image
17
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
4 kVp Changes
18
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Radiographic Technique
Technique charts are based on the ldquoaverage patientrdquo
The thicker the part the more x-radiation is required to penetrate Calipers should be used
Keep in mind not only the measurement but the type of tissue you need to penetrate (fat vs muscle)
19
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Technique
In general Soft tissue = low kVp and high mAs
Extremity (soft tissue amp bone) = low kVp
Chest (high subject contrast) = high kVp Abdomen (low subject contrast) = middle kVp
20
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Pathology
Can appear with increased radiolucency or radiopacity
Some pathology is destructive causing tissue to be radiolucent
Others can be additive causing tissue to be radiopaque
21
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Technique selection ndash Fixed kVp For each anatomic part there is an optimum
kVp
mAs is varied based on part thickness or pathological condition
22
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Image Quality
Bushong Ch 16
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Objectives
bull Image Quality ndash Factors
bull Geometric Factors
bull Subject Factors
bull Artifacts
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Image Quality
bull Is the exactness of the representation of the patientrsquos anatomy
bull 3 major factors affecting image quality that is under the control of the technologist Image Receptor selectionuse Geometric factors amp Subject factors
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Judging Image Quality
bull The most important characteristic of radiographic quality are Spatial Resolution Contrast Resolution Noise amp Artifacts
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Main Factors Affecting Recorded Detail
bull kVp amp mAsbull Motionbull Object
Unsharpnessbull SID (Source to
Image Distance)bull OID (Object to
Image Distance)
bull Material Unsharpness Film Screen Combo
bull Focal Spot Size bull MTF (modulation
transfer function)
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Recorded Detail
bull Other names- detail-sharpness of detail-definition-resolution-degree of noise- visibility of detail
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Resolution
bull Is the ability to image two separate objects and visually distinguish one from the other
bull Spatial resolution is the ability to image small objects that have high subject contrast Ex bone-soft tissue interface breast calcifications calcified lung nodule
bull Conventional radiography has excellent spatial resolution
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
RESOLUTION TEST
TOOLS
LINE PAIRS MM
Depicts how well you can see the differences in structures
More lines=more detail
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Measuring Resolution for an x-ray imaging system
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
SMPTE Test Pattern
bull In 1985 the Society of Motion Picture and Television Engineers (SMPTE) published a recommended practice (RP-122)
bull Specifications for Medical Diagnostic Imaging Test Patterns for Television Monitors and Hard-copy Cameras
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
SMPTE Test Pattern
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Focal spot size of the cathode
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Line-focus principle
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Modulation Transfer Function
bull The ability of a system to record available spatial frequencies
bull The sum of the components in a recording system cannot be greater than the system as a whole
bull When any componentrsquos function is compromised because of some type of interference the overall quality of the system is affected
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Contrast Resolution
bull Is the ability to distinguish anatomic structures of similar subject contrast Ex liver-spleen gray matter-white matter
bull Magnetic Resonance Imaging has the highest contrast resolution
bull Computed Tomography is excellent as well
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
MRI CT
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
The contrast of an object is expressed relative to its surrounding background
That is what determines its visibility
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Radiographic Contrast
bull Is the product of image receptor contrast and subject contrast
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
ldquoNoiserdquo
bull Borrowed from electrical engineering
bull Audio noise = hum or flutter heard from a stereo
bull Video noise = ldquosnowrdquo on a TV
bull Radiographic noise = random fluctuation on the OD of the image
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy
image - MORE COMMON IN CR SYSTEMS
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Radiographic noise
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Radiographic Noise
bull Four components
bull Film graininess structure mottle quantum mottle amp scatter radiation
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Radiographic Noise
bull Film graininess ndash distribution amp size of the silver halide grains in the emulsion
bull Structure mottle ndash speed of the intensifying screen Phosphor size amp DQECE
bull Not under the control of the technologist
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Image Noise
bull Speckled background on the imagebull Caused when fast screens and high kVp
techniques are used Noise reduces image contrast
bull The percentage of x-rays absorbed by the screen is the detective quantum efficiency (DQE)
bull The amount of light emitted for each x-ray absorbed is the conversion efficiency (CE)
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Quantum Mottle
bull An image produced with just a few x-rays will have more quantum mottle
bull The use of very fast intensifying screens or not enough mAs or kVp will increase quantum mottle
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Quantum mottle
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Screen Speed
bull Efficiency of a screen in converting x-rays to light is Screen Speed
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Speed
bull Fast image receptorsndash Noise Spatial resolutionndash Contrast resolution
bull Slow image receptorsndash Noise Spatial resolutionndash Contrast resolution
ndash See pg 274 for answers
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Speed
bull Low noise = fast or slow speed
bull High contrast resolution = fast or slow speed
bull Reduced spatial resolution = fast or slow speed
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
PARALLAX ndasheach emulsion has an image
single image overlaped edges edge sharp less sharp
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Other Film FactorsCharacteristic Curve
bull Is used to describe the relationship between OD and radiation exposure
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
What is the useful OD range
Characteristic curveof radiographic film
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
The latitude of an image receptor is the exposure range over which it responds with diagnostically
useful OD
bull Depending on the manufacturing characteristics
radiographic film will respond differently to
radiation exposure
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
FS vs Digital Dynamic Range
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Unexposed film
bull Appears like a frosted glass window
bull ODs of unexposed film are due to base density and fog density
bull Base density ndash tint added to the base to reduce eye strain and crossover Has a densitometer value of approximately 01
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
CROSSOVER
bull Reducing crossover by adding a dye to the base
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Unexposed film
bull Fog Density ndash inadvertent exposure of film during storage chemical contamination improper processing radiation exposure etc
bull Fog density contributes to reduction of radiographic contrast
bull Fog density should not exceed 01
bull Base + fog OD = 01 to 03
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
2 principal characteristics of any image are Spatial amp Contrast Resolution
bull Spatial resolution ndash Resolution is the ability to image two separate
objects and visually distinguish one from the otherndash Spatial resolution is the ability to image small
objects that have high subject contrast (eg bone-soft tissue interface calcified lung nodules)
ndash Determined by focal-spot size and other factors that contribute to blur
ndash Diagnostic x-ray has excellent spatial resolution It is measured in line pairs per mm
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Other factors affecting the finished radiograph
bull The concentration of processing chemicals
bull The degree of chemistry agitation during development
bull Development time
bull Development temperature
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Image Quality in Digital
Matrix size is determined by
bull Receptor size (Field of View FOV)
bull Pixel size
bull CR - Sampling frequency
bull DR - DEL size
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Spatial Resolution determined by
1048697 Pixel sizebull CR- sampling frequencybull DR ndash DEL sizebull 1048697 There are relationships betweenbull Pixel sizebull Receptor sizebull Matrix sizebull 1048697 pixel size = larger matrixbull 1048697 receptor size = larger matrixbull Spatial resolution is not related the amount of exposure
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Sampling Frequency
bull The sampling frequency is the rate at
bull which the laser extracts the image data
bull from the plate
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Signal Sampling Frequency Good sampling under sampling
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Nyquist Frequency
bull The Nyquist Frequency will be frac12 of thebull sampling frequencybull A plate that is scanned using a sampling
frequency of 10 pixels per millimeter would not be able to demonstrate more than 5 line pairs per millimeter based upon the Nyquist Frequency
bull The Nyquist Frequency allows the determination of the spatial resolution for a given sampling frequency
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Geometric Factors
bull Producing high quality radiographs Technologists must maximize geometric conditions
bull Three principal geometric conditions affect radiographic quality Magnification Distortion amp Focal-spot blur
bull Review table 16-4 pg 295 for summary
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Object Unsharpness
bull Main problem is trying to image a 3-D object on a 2-D film
bull Human body is not straight edges and sharp angles
bull We must compensate for object unsharpness with factors we can control focal spot size SID amp OID
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Magnification
bull All image on the radiograph are larger than the object they represent
bull For most exams minimizing magnification is desired There are a few exams where some magnification can be helpful
bull TUBE CLOSE TO THE PART (SID)bull PART FAR FROM THE CASSETTE (OID)
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Minimizing Magnification
bull Large SID use as large a source-to-image receptor distance as possible
bull Small OID place the object as close to the Image receptor as possible
bull In terms of recorded detail and magnification the best image is produced with a small OID and a large SID
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Size Distortion amp SID
bull Major influences SID amp OID
bull As SID magnification bull Standardized SIDrsquos allow radiologist to
assume certain amt of magnification factors are present
bull Must note deviations from standard SID
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
The position of the tube (SID) to IR Will influence how the structures appear on the image The farther away ndash the less magnified uarrSID darr MAGNIFICATION
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
SID
bull Shine a flashlight on a 3-D object shadow borders will appear ldquofuzzyrdquo
-On a radiograph called Penumbrabull Penumbra (fuzziness) obscures true
border ndash umbrabull Farther the flashlight from object =
sharper borders Same with radiography
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
OIDObject to Image Distance
bull The closer the object to the film the sharper the detail
bull OID penumbra sharpness bull OID penumbra sharpness bull Structures located deep in the body
radiographer must know how to position to get the object closest to the film
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Size Distortion amp OID
bull If source is kept constant OID will affect magnification
bull As OID magnification bull The farther the object is from the film the
more magnification
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
The position of the structure in the body will influence how magnified it will be seen on the image
The farther away ndash the more magnified
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Minimal magnification small OID
Magnification - large OID
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
40rdquo SID VS 72rdquo SID
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Magnification Factor
source-to-image receptor distance
bull MF = source-to-object distancendash SOD difficult to measure accurately
usually an estimated value
SID
bull MF = SOD
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Finding SOD
bull SID ndash OID = SOD
bull SID = 100 cm
bull OID = 7 cm
bull What is the SOD
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Finding magnification of the heart on a lateral CXR
bull SID ndash OID = SOD
bull SID = 72 inches
bull OID = 8 inches (estimated)
bull What is the SOD
bull What is the Mag Factor
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Distortion
bull Misrepresentation of the true size or shape of an object
-MAGNIFICATION (size distortion)
-TRUE DISTORTION (shape distortion)
bull Shape distortion unequal magnification of different portions of the same object
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Shape Distortion
bull Depends on
bull Object thickness
bull Object position
bull Object shape
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Object Thickness
bull Thick objects have more OID and are more distorted than thinner structures
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Object Position
bull If the object plane and the image plane are parallel the image is not distorted
bull CR perpendicular
to the part
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Position Distortion
bull Foreshortened = anatomy at an incline to the CR displays smaller than true size
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
D amp E = shape distortion
(foreshortening of part)
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Position Distortion
bull Elongation anatomy at an incline and lateral to the central axisndash Could be
foreshortened as well
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Elongation Foreshortened Normal
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Position Distortion
bull Spatial distortion = anatomy positioned at various OIDs but superimposed only one can be seen
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Position Distortion ndash Irregular Anatomy
bull Anatomy or objects can cause considerable distortion when imaged off the central axis
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Focal-Spot Blur
bull Dependent on the size of the effective focal spot
bull Smaller the effective focal spot = less blur and better spatial resolution
bull Focal-spot blur is the most important factor in determining spatial resolution
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Focal Spot Size
bull Smaller x-ray beam width will produce a sharper image
bull Fine detail = small focal spot (ie small bones)
bull General radiography uses large focal spot
bull Beam from penlight size flashlight vs flood light beam
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
ANODE
ANODE
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Focal spot size ndash determined by filament in cathode amp surface area used at anode
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
THE SMALLER THE BEAM TOWARDS THE PATIENT - THE BETTER THE DETAIL OF THE IMAGE PRODUCED
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Focal-spot blur is caused by the effective size of the focal spot which is larger to the
cathode side of the image
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Focal-spot blur is small when the object-to-image receptor distance (OID) is small
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Image QualitySubject Factors
bull Patient thickness
bull Effective atomic number
bull Object shape
bull Subject contrast
bull Tissue mass density
bull kVp
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Object shape
Objects with structurehaving a form that coincides with thex-ray beam has maximum detail
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Patient Motion
bull Can be voluntary or involuntary
bull Best controlled by short exposure times
bull Use of careful instructions to the pt
bull Suspension of pt respiration
bull Immobilization devices
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Blurring of image due to patient movement during exposure
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-
Questions
- PRIME FACTORS
- PRIME FACTORS
- Slide 6
- Slide 7
- Inverse Square Law Formula
- Direct Square Law
- Focal-Spot Changes
- Tube voltage (kVp)
- Influencing factors kVp
- kVp Changes
- 4 kVp Changes
- Image Quality
- Objectives
- Image Quality
- Judging Image Quality
- PowerPoint Presentation
- Main Factors Affecting Recorded Detail
- Slide 29
- Recorded Detail
- Slide 31
- Slide 32
- Slide 33
- Resolution
- Slide 35
- Measuring Resolution for an x-ray imaging system
- Slide 37
- Slide 38
- SMPTE Test Pattern
- SMPTE Test Pattern
- Focal Spot Size
- Focal spot size of the cathode
- Line-focus principle
- Modulation Transfer Function
- Contrast Resolution
- MRI CT
- Slide 47
- The contrast of an object is expressed relative to its surrounding background That is what determines its visibility
- Radiographic Contrast
- Slide 50
- Slide 51
- Slide 52
- Slide 53
- ldquoNoiserdquo
- QUANTUM MOTTLE Not enough PHOTONS ndash can create a mottled or grainy image - MORE COMMON IN CR SYSTEMS
- Radiographic noise
- Radiographic Noise
- Slide 58
- Image Noise
- Quantum Mottle
- Quantum mottle
- Slide 62
- Slide 63
- Screen Speed
- Speed
- Slide 66
- PARALLAX ndash each emulsion has an image single image overlaped edges edge sharp less sharp
- Other Film Factors Characteristic Curve
- What is the useful OD range
- Slide 70
- The latitude of an image receptor is the exposure range over which it responds with diagnostically useful OD
- Slide 72
- FS vs Digital Dynamic Range
- Unexposed film
- CROSSOVER
- Slide 76
- Slide 77
- 2 principal characteristics of any image are Spatial amp Contrast Resolution
- Other factors affecting the finished radiograph
- Slide 80
- Image Quality in Digital
- Slide 82
- Slide 83
- Slide 84
- Spatial Resolution determined by
- Slide 86
- Sampling Frequency
- Slide 88
- Signal Sampling Frequency Good sampling under sampling
- Nyquist Frequency
- Slide 91
- Slide 92
- Slide 93
- Slide 94
- Slide 95
- Slide 96
- Geometric Factors
- Object Unsharpness
- Magnification
- Minimizing Magnification
- Size Distortion amp SID
- Slide 102
- SID
- Slide 104
- Slide 105
- OID Object to Image Distance
- Size Distortion amp OID
- Slide 108
- Slide 109
- Slide 110
- Slide 111
- Slide 112
- Slide 113
- 40rdquo SID VS 72rdquo SID
- Slide 115
- Magnification Factor
- Finding SOD
- Finding magnification of the heart on a lateral CXR
- Distortion
- Slide 120
- Shape Distortion
- Object Thickness
- Slide 123
- Object Position
- Position Distortion
- D amp E = shape distortion (foreshortening of part)
- Slide 127
- Elongation Foreshortened Normal
- Slide 129
- Position Distortion ndash Irregular Anatomy
- Focal-Spot Blur
- Slide 132
- Slide 133
- Slide 134
- Slide 135
- Focal-spot blur is caused by the effective size of the focal spot which is larger to the cathode side of the image
- Focal-spot blur is small when the object-to-image receptor distance (OID) is small
- Image Quality Subject Factors
- Object shape
- Patient Motion
- Slide 141
- Slide 142
- Slide 143
- Slide 144
- Questions
-