Temporary Cardiac Pacing
Dr. Md. Mashiul AlamPhase B resident
Chairperson: Prof. F. Rahman
![Page 2: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/2.jpg)
Introduction
Temporary cardiac pacing involves electrical cardiac stimulation to treat a bradyarrhythmia or tachyarrhythmia until it resolves or until long-term therapy can be initiated.
The purpose of temporary pacing is to re-establish circulatory integrity and normal hemodynamics that are acutely compromised by a slow or fast heart rate.
![Page 3: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/3.jpg)
Indication Acute hemodynamically significant bradycardia or
asystole Acute MI: New bifasicular block (RBBB + LAD or RAD) New LBBB with 1st degree AV block Alternating LBBB and RBBB Mobitz Type II and complete heart block ( in case of
hemodynamically significant in Inferior MI) RVI with loss of AV synchrony
![Page 4: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/4.jpg)
Bridge to permanent pacing Termination of tachycardias (Overdrive pacing) Post cardiac surgery which may affect AV node
of His bundle Electrophysiological study Prophylactic during RHC in the setting LBBB Complex PCI in RCA Drug toxicity e.g., digoxin, beta blocker
![Page 5: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/5.jpg)
Relative contraindication
• Poor vascular access• Bleeding disorder or anticoagulaiton therapy: INR should be > 1.8 Platelete count >50,000
![Page 6: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/6.jpg)
Temporary pacing methods
• Transcutaneou ventricular pacing (TCP)• Transesophageal or transgastric pacing• Epicardial pacing• Transvenous pacing
![Page 7: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/7.jpg)
TCP
Emergency treatment of sever bradycardia or asystole and bridge to transvenous pacing
Lead placement:• Anterior posterior position One at the anterior chest wall at V3 positon another one posteriorly below scapula• Anterior lateral position Over sternum and at apex
![Page 8: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/8.jpg)
![Page 9: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/9.jpg)
• Pulse width – 20 to 40 ms• High output – 100 to 200 mA
Common problem:Sever patient discomfortFailure to capture
![Page 10: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/10.jpg)
Transesophageal or gastric pacing
• Flexible electrode on the tip of a catheter down the esophagus for atrial pacing
• Advanced to the fundus of stomach for ventricular pacing
Rarely used
![Page 11: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/11.jpg)
Epicardial Pacing
At the time of aortic valve replacement, AV canal defect repair, tricuspid surgery or Ebstein anomaly correction
Epicardial pacing wires are placed in atrium and ventrical at the time of surgery. Fine small caliber wire are used only for a short period.
![Page 12: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/12.jpg)
Transvenous pacing
• Patient preparaton: Infromed consent if possible ( no need in
emergency condition) Peripheral IV access if the procedure is elective should be performed in monitored setting
equipped with fluoroscopy and crash cart May be perfromed ECG or echo guided in some
situations
![Page 13: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/13.jpg)
• SiteFemoral veinInternal jugular veinSubclavian veinBrachial vein
Ipsilateral subclavian vein should be avoided if permanent pacemaker is planned
Femoral approach is advisable if undergone thrombolysis
![Page 14: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/14.jpg)
![Page 15: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/15.jpg)
Temporary leads should be resited from femoral access sites after 48 hours and from other sites after 5-7 days when possible to prevent infection
• Lead choice- Active fixation leads allows greater pacemaker
stability Passive fixation lead – usually done
![Page 16: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/16.jpg)
![Page 17: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/17.jpg)
• Instruments C-arm image-intensifier/ fluroscope Temporary pacing box 5F temporary pacing wire 5F or 6F introducer sheath External pacemaker / defibrillator Cardiac monitor with non invasive / invasive
monitoring facilities Oxygen and air inlet
![Page 18: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/18.jpg)
![Page 19: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/19.jpg)
![Page 20: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/20.jpg)
• Position of the patient: Supine in bed Tendelenberg postion for internal jugular and
subclavian approach for filling of the vein and prevention of air embolism
![Page 21: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/21.jpg)
• Technique- Venous sheath is inserted by Seldinger
techniqueVentricular lead insertionAtrial lead insertion
![Page 22: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/22.jpg)
![Page 23: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/23.jpg)
![Page 24: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/24.jpg)
Ventricular lead insertion
• Leads are straight or slightly angulated near the tip
• Pacing catheter is advanced through venous sheath under fluoroscopy guidance (20-30 degree LAO projection)
• Catheter advanced through tricuspid valve and turned either clockwise or anticlock wise to direct the tip anteriorly
![Page 25: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/25.jpg)
• Attempt to cross TV directly• If unsuccessful gentle pressure applied with
catheter torqued to allow mid portion of catheter to prolapse across the valve or by looping the tip against the lateral atrial wall and rotating the loop medially
• Once catheter in the ventricle – it is rotated so that the tip points inferiorly to the apex
![Page 26: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/26.jpg)
• The ideal catheter placement site- diaphragmatic surface or floor of the right ventricle anywhere between the midpoint and apex
• Placement at the Apex should be avoided as there is increased chance of perforation
• RVOT is a option but unstable tip location
![Page 27: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/27.jpg)
Ensuring accurate lead placementTip will be located left side of the spine in AP
view and anterior inferior postion in lateral veiw. Some buckling is acceptable.
If tip is in coronary sinus the tip will be posterior superiorly directed in lateral view
• Leads are attached with the pulse generator….
![Page 28: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/28.jpg)
![Page 29: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/29.jpg)
![Page 30: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/30.jpg)
Testing
Once catheter is in a stable position , threshold testing should be performed
![Page 31: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/31.jpg)
![Page 32: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/32.jpg)
Pacing rate: regulates the number of impulses that can be delivered to the heart per minute. The rate setting depends on the physiological needs of the patient.
Threshold: is the lowest energy, which will cause the myocardial cell to contract. The desired threshold is 0.5-1.0mA.
Sensing is the ability of the pacemaker to “see” when a natural (intrinsic) depolarization is occurring.– Pacemakers sense cardiac depolarization by measuring changes in electrical potential of myocardial cells between the anode and cathode.– Expressed in Millivolts (mV).
![Page 33: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/33.jpg)
Setting stimulation threshold
• Set RATE at least 10 bpm above patient’s intrinsic rate.
• Decrease OUTPUT: Slowly turn OUTPUT dial counterclockwise until ECG
shows loss of capture.
• Increase OUTPUT: Slowly turn OUTPUT dial clockwise until ECG shows
consistent capture. This value is the stimulation threshold.
• Set OUTPUT to a value 2 to 3 times greater than the stimulation threshold
value.
• Restore RATE to previous value
.
![Page 34: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/34.jpg)
Setting sensing threshold • Set rate at least 10 bpm below patient’s intrinsic rate.
• Adjust output: Set OUTPUT to 0.1 mA .
• Decrease SENSITIVITY: Slowly turn MENU PARAMETER dial counterclockwise until pace indicator flashes continuously.
• Increase SENSITIVITY: Slowly turn MENU PARAMETER dial clockwise until sense indicator flashes and pace indicator stops flashing. This value is the sensing threshold.
• Set SENSITIVITY to half (or less) the threshold value.
![Page 35: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/35.jpg)
Post placement CXR
• To evaluate pneumothorax and proper position of lead after the procedure
![Page 36: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/36.jpg)
![Page 37: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/37.jpg)
Pacer Care
• Check catheter insertion site daily for signs of infection and apply regular sterile dressing at regular interval
• Obtain daily 12 lead surface ECG with and without pacing
• Check function daily by determining the sensing and pacing threshold and check underlying rhythm daily by decreasing the pacing rate gradually
![Page 38: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/38.jpg)
Complication and troubleshooting
• Related to central venous access (Hematoma, Pneumothorax, hemothorax, air embolism, thrombosis, arterial puncture, AV fistula, infection)
• Cardiac arrythmia• Myocardial perforation echocardiogram should be performed Lead should be withdrawn only when there is all
facilities for percardiocentesis or surgical drainage
![Page 39: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/39.jpg)
• CHB in a patent with prevous LBBB due to irritation of right bundle while lead placement
• Pacemaker relatedGenerator failure – change the generator/
keep an extra generator for emergencyVentricular non capture
![Page 40: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/40.jpg)
Ventricular non capture/ capture failure
![Page 41: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/41.jpg)
• Causes of Failure to Capture: Insufficient output Low pacemaker batteryDisplaced or fractured leadElectrolyte abnormalities: acidosis; hypoxia;
hypokalaemia
![Page 42: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/42.jpg)
• Management: Increase output on Output dial lead may need to be repositioned or replacedCheck electrolytes View rhythm in different leads Change electrodes Check connections Change
battery, cables, pacemaker box
![Page 43: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/43.jpg)
Ventricular undersensing
Potential for ventricula fibrilation ( due to R on T phenomena)
![Page 44: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/44.jpg)
![Page 45: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/45.jpg)
• Management: Increase sensitivity (reduce mV on Sense dial)lead may need to be repositioned or replacedView rhythm in different leads Change electrodes Check connections Change
cables, battery, pacemaker box Check electrolytes
![Page 46: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/46.jpg)
Ventricular oversensing
Pacing does not occur when intrinsic rhythm is inadequate
![Page 47: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/47.jpg)
• Pacemaker too sensitive • Possible lead displacement • Pacemaker failure
Danger of heart block, asystole
![Page 48: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/48.jpg)
• Management: Reduce sensitivity (increase mV on Sense dial)lead may need to be repositioned or replacedView rhythm in different leadsChange electrodes Check connections Change cables, battery, pacemaker box Check
electrolytes
![Page 49: Temporary cardiac pacing](https://reader035.vdocument.in/reader035/viewer/2022062316/588704111a28ab4e3a8b6d5d/html5/thumbnails/49.jpg)
Thank you
There is no alternative to practical experience….