TESTING FLAT-PANEL IMAGING SYSTEMS:What the Medical Physicist Needs to Know
JAMES A. TOMLINSON, M.S., D.A.B.R.Diagnostic Radiological Physicist
![Page 2: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/2.jpg)
Topics
• Image Uniformity and Artifacts• Image Quality - Detail• Image Quality - Contrast• Detector Sensitivity
![Page 3: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/3.jpg)
Image Uniformity - Setup
• PMMA – 8” thick x 14”x17”• Copper – 2-3 mm x 14”x17”• Typical radiographic technique or AEC
![Page 4: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/4.jpg)
Image Uniformity - Analysis
• Visual inspection of image to look for shading or signal brightness variation
• Use typical display window, then narrow window
• Provides a quick review
• Comprehensive testing requires analytical tools to measure signal variation across image
![Page 5: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/5.jpg)
Image Artifacts
• Inspect uniformity image for:– signal voids or defects– lines (banding)– residual image burn
• CAUTION: Latent images can result from phosphor burn if detector dose is too high !
![Page 6: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/6.jpg)
Image Quality - Detail
• Primarily controlled by sampling pitch (pixel size) and display pitch
• Use PSF, LSF, or ESF to generate MTF curve
• Line-pair test tool quickly evaluates cut-off frequency in the field
![Page 7: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/7.jpg)
DF Detail
• Past experience using line-pair test tool with digital fluoroshows increasing detail with decreasing image size, because fluoroscopic image has more detail than converted digital matrix , up to 2048x2048.
FOV (cm) lp/mm (1k)20 2.516 3.213 3.9
![Page 8: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/8.jpg)
Flat Panel Detail
• Flat panel detectors often have the same cut-off frequency regardless of FOV.
FOV (cm) lp/mm (1k)20 3.516 3.513 3.5
![Page 9: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/9.jpg)
Digital Zoom
• Because native detector element size is constant, it is limiting factor with detail
• When change FOV, the system does “digital zoom”, by overlaying original data pattern over new display matrix (like digital cameras)
• Since original pattern remains constant from sampling frequency, so does cut-off frequency
![Page 10: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/10.jpg)
detectormatrix
testpattern
![Page 11: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/11.jpg)
detectormatrix
![Page 12: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/12.jpg)
displaymatrix
![Page 13: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/13.jpg)
detectormatrix
zoom area
![Page 14: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/14.jpg)
zoom
same cut-off frequency
![Page 15: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/15.jpg)
Magnification
• If detector matrix (512-1024) ≤ display matrix (1024), image magnification (zoom) results in spread of data across pixels and no increased detail
• If detector matrix (2048) is > display matrix (1024), image magnification (zoom) results in increased detail displayed
![Page 16: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/16.jpg)
Pixel Binning
• Some systems combine a group of detector element data together to form one pixel for display
• Results in larger than expected blur and low cut-off frequencies
• Used in some applications, such as cardiac cath, where data throughput is restricted at high frame rates
![Page 17: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/17.jpg)
Pixel Binning (1k display)
FOV lp/mmxlarge 1.7large 1.7medium 1.7small 3.5
![Page 18: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/18.jpg)
Sampling Pitch
Size Matrix Pixel Pitch3 MP 1500x2000 0.24 mm 2 lp/mm
5 MP 2000x2500 0.17 mm 3 lp/mm
9 MP 2700x3300 0.13 mm 3.8 lp/mm
![Page 19: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/19.jpg)
Aliasing
• To reduce aliasing errors with detector and display matrices , place test object at 450 to the matrix
• Underestimates cut-off frequency
• Divide observed frequency by sin 450 (0.7)
![Page 20: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/20.jpg)
Typical DR Resolution
• 3-5 lp/mm
• Should use magnifying glasses, especially for old senior physicists, due to visual loss with age
![Page 21: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/21.jpg)
Image Contrast
• Contrast sensitivity and resolution is much better for DR systems than screen/film systems
• Image should be viewed at a greater distance than for detail review.
![Page 22: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/22.jpg)
Dynamic Range
Signal
S/F
DR
Log Exposure
![Page 23: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/23.jpg)
Contrast Sensitivity
• Test tools commonly available to determine cutoff contrast level (%) and size
• Typical contrast = < 1%
![Page 24: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/24.jpg)
![Page 25: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/25.jpg)
16% 4.5%
2%
1.5%
0.5%
Sensitivity
![Page 26: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/26.jpg)
10 mm
0.6 mm
Resolution
![Page 27: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/27.jpg)
Detector Radiation Sensitivity
• Measure radiation to the detector, required to produce acceptable image
• Clinical medical physicists need to know what the patiententrance radiation exposures are, for exposure control, organ dose calculations and fetal dose estimates
![Page 28: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/28.jpg)
AEC Calibration
• To assess routine patient exposure, the ESE from AEC should be assessed with 4-10 inches of PMMA
• The kVp should be appropriate for the thickness
• We have found that for patient ESE:CR 2x > Screen/film 2x > DR
![Page 29: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/29.jpg)
![Page 30: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/30.jpg)
My Image Quality Phantom
• With readily available test tools, a combo phantom can be built to provide quick evaluation of detail and contrast.
• Considerable cost savings to other phantoms…
![Page 31: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/31.jpg)
![Page 32: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/32.jpg)
![Page 33: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/33.jpg)
1.0 % contrast
![Page 34: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/34.jpg)
![Page 35: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/35.jpg)
4.0 lp/mm
![Page 36: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/36.jpg)
“Name That Phantom”
• D.R.I.Q.• D.R.I.P.• The TESTOOL• M.I.Q.P.• P.E.D.R.O.
![Page 37: TESTING FLAT-PANEL IMAGING SYSTEMS What the Medical Physicist](https://reader031.vdocument.in/reader031/viewer/2022020704/61fb50212e268c58cd5cb016/html5/thumbnails/37.jpg)