the babylog 8000 plus with pressure support and volume guarantee the key to breathing harmony
TRANSCRIPT
The Babylog 8000 plus with Pressure Support and Volume Guarantee
The Key to Breathing Harmony
Babylog 8000 plus
• A ventilator designed specifically for
critically ill newborn infants
• Comprehensive modes: CPAP, IMV,
SIMV, A/C, PSV
• Volutrauma protection strategy: Volume
Guarantee option in SIMV, A/C, and PSV
modes
• Leak- adapted flow synchronization
PSV with DifferencesFirst the Similarities:
• Pressure targeted breath (both spontaneous and set)
• Flow terminated at 15% of peak inspiratory flow
• Inspiratory time will vary and is patient dependent
• Preset Insp. Time still functions as a back-up, breath will not exceed clinician setting
Now the big differences:•A “back-up” rate is set - so yes, in theory a paralyzed patient can be placed on PSV, Babylog style
•All breaths, whether patient assisted or not will be pressure targeted and flow terminated
•PSV may be combined with the volume guarantee (VG) option – allowing clinicians to set tidal volume
PSV with Differences
Paw
V•
insp
expPatient or vent ilator initiated
inspiration PSV cycledexpiration
Peak flow
Drop to 15% of peak flow
Pressure Support Ventilation (PSV)
Set Pinsp
Flow terminationautomatically compensates
for leak
Onset of inspiration
Onset ofexpiration
Leak flow
Pressure Support Ventilation with leak compensation
Patient- Set Inspiratory Time in Neonatal Pressure Support Ventilation (PSV)
Examples courtesy of Prof. JC Rozé, Nantes, France
The average Inspiratory Time of the four patients on PSV above The average Inspiratory Time of the four patients on PSV above was ~ .25seconds – expect to see shorter I times, and was ~ .25seconds – expect to see shorter I times, and
consequently a drop in the MAPconsequently a drop in the MAP
Why let the infant end inspiration (set Tinsp ) ?
CBF variability increases IVH riskPerlman , et al. N Engl J Med 1983
CBF variability and IVH increases with infant- ventilator asynchrony
Perlman , et al. N Engl J Med 1985
Rennie, et al. Arch Dis Child, 1987
So, harmonizing ventilator- infant interaction should reduce IVH.
Why give mechanical support for all breaths?
• Decreases Work of Breathing Jarreau, Moriette, Mussat, et al; Am J Resp CCM, Mar 1996
• Decreases Oxygen ConsumptionRoze, Liet, Gournay, Debillon, Gaultier; Eur Resp J, Nov 1997
• Reduced Stress Hormone LevelsQuinn, de Boer, Ansari, Baumer; Arch Dis Child, May 1998
• Decreased Effort and Respiratory RateBendel- Stenzel, Bing, Meyers, Connett, Mammel; Ped Res, May 1998
• Less Variation in Vt After SurfactantMrozek, Bendel- Stenzel, Meyers, Bing,, Mammel; Ped Res, May 1998
Problems with a Preset Ti: Patient Asynchrony
The infant attempts to breathe during a set ventilator inspiratory cycle, resulting in lower lung pressure and
excessive volume.
•The VG option allows for delivery of a set volume during mandatory pressure breaths in A/C, SIMV, and PSV
•Similar to Autoflow on the Dura, the PIP will adjust automatically up to a set maximum, compensating for changes in resistance and/or compliance, to ensure the set tidal volume is delivered
•The inspiratory Pressure knob now functions as the maximum pressure allowed, NOT the set PIP. PIP is not set, and will vary.
Volume Guarantee: How does it work?
•Similar to autoflow, high respiratory demand from the patient will result in a lower PIP delivered to the airway, i.e. less support for the patient.
•PATIENTS MUST BE CLOSELY MONITORED FOR SIGNS OF INCREASED WOB WHEN UTILIZING THE VOLUME GUARENTEE MODE!
Problems with Volume Guarantee
Why volume - oriented ventilation ( VOV ) in infants?
VOV vs SIMV: Equivalent ventilation at lower MAP
Herrera, at al. Ped Res 1999
VOV reduces IVH, acute lung injuryRosen WC, et al. Ped Pulm 1993 Sinha SK, et al. Arch Dis Child Fetal Neo Ed
1997
Consistent Vt Stable PaCO2 Stable CBF Less IVH
Postulated mechanism of action:
Why volume - oriented ventilation ( VOV ) in infants?
Pressure ventilation may cause hypoventilation or overdistention when CL changesDavis, et al. N Engl J Med 1988
Gibson, et al. Eur J Pediatr 1994
Bjorklund, et al. Am J Respir Crit Care Med 1996
Dimitriou, et al. J Perinat Med 1997
Lung overdistention creates acute / chronic lung injuryHerandez, et al. J Appl Physiol 1989
Bjorkland, et al. Ped Res 1997
Variable PaCO2 in pressure ventilation increases IVHStewart, et al. Pediatrics 1981
Pressure Support Mode with Volume Guarantee :
Concept of “Autoweaning”
PIP
C lung
Vt
Extubate!Time
Tidal volume delivery in mechanically ventilated preterm infants
Appropriate tidal volume for mechanical ventilation of preterm infants with
surfactant deficiency is
4 - 6 mL/ kg body weight.
PSV+VGA Happy Child
Two Extra Pictures - The Neonatal flow sensor
• Hot wire anemometer
• Sensitive to 0.17 mL
• 0 - 30 lpm range
• Weighs 10 grams
• 0.5 mL added deadspace
• Inexpensive
• 6 month use
Flow Sensor Measurement Principle
Hot wire anemometer:• Two tiny platinum wires are heated to 400°C
• One wire is shaded to determine direction of gas flow
• Wire cooling is proportional to gas flow
• Flow is integrated with time for volume measurement