the beam-foil spectra of krypton (2 to 5 mev) · 2020. 4. 2. · i i i i 77-18,667 cardon, bartley...

213
THE BEAM-FOIL SPECTRA OF KRYPTON (2 TO 5 MEV) Item Type text; Dissertation-Reproduction (electronic) Authors Cardon, Bartley Lowell, 1940- Publisher The University of Arizona. Rights Copyright © is held by the author. Digital access to this material is made possible by the University Libraries, University of Arizona. Further transmission, reproduction or presentation (such as public display or performance) of protected items is prohibited except with permission of the author. Download date 07/06/2021 04:55:53 Link to Item http://hdl.handle.net/10150/289579

Upload: others

Post on 27-Jan-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

  • THE BEAM-FOIL SPECTRA OF KRYPTON (2 TO 5 MEV)

    Item Type text; Dissertation-Reproduction (electronic)

    Authors Cardon, Bartley Lowell, 1940-

    Publisher The University of Arizona.

    Rights Copyright © is held by the author. Digital access to this materialis made possible by the University Libraries, University of Arizona.Further transmission, reproduction or presentation (such aspublic display or performance) of protected items is prohibitedexcept with permission of the author.

    Download date 07/06/2021 04:55:53

    Link to Item http://hdl.handle.net/10150/289579

    http://hdl.handle.net/10150/289579

  • INFORMATION TO USERS

    This material was produced from a microfilm copy of the original document. While the most advanced technological means to photograph and reproduce this document

    have been used, the quality is heavily dependent upon the quality of the original

    submitted.

    The following explanation of techniques is provided to help you understand markings or patterns which may appear on this reproduction.

    1. The sign or "target" for pages apparently lacking from the document photographed is "Missing Page(s)". If it was possible to obtain the missing page(s) or section, they are spliced into the film along with adjacent pages. This may have necessitated cutting thru an image and duplicating adjacent pages to insure you complete continuity.

    2. When an image on the film is obliterated with a large round black mark, it is an indication that the photographer suspected that the copy may have moved during exposure and thus cause a blurred image. You will find a

    good image of the page in the adjacent frame.

    3. When a map, drawing or chart, etc., was part of the material being photographed the photographer followed a definite method in "sectioning" the material. It is customary to begin photoing at the upper left hand corner of a large sheet and to continue photoing from left to right in equal sections with a small overlap. If necessary, sectioning is continued again — beginning below the first row and continuing on until

    complete.

    4. The majority of users indicate that the textual content is of greatest value, however, a somewhat higher quality reproduction could be made from "photographs" if essential to the understanding of the dissertation. Silver prints of "photographs" may be ordered at additional charge by writing the Order Department, giving the catalog number, title, author and specific pages you wish reproduced.

    5. PLEASE NOTE: Some pages may have indistinct print. Filmed as received.

    University Microfilms International 300 North Zeeb Road Ann Arbor, Michigan 48106 USA

    St. John's Road, Tyler's Green High Wycombe, Bucks, England HP10 8HR

  • I I

    I I

    77-18,667

    CARDON, Bartley Lowell, 1940-THE BEAM-FOIL SPECTRA OF KRYPTON (2 TO 5 MeV).

    The University of Arizona, Ph.D., 1977 Physics, atomic

    Xerox University Microfilms , Ann Arbor, Michigan 48106

  • THE BEAM-FOIL SPECTRA OF KRYPTON (2 TO 5 MeV)

    by

    Bartley Lowell Cardon

    A Dissertation Submitted to the Faculty of the

    DEPARTMENT OF PHYSICS

    In Partial Fulfillment of the Requirements For the Degree of

    DOCTOR OF PHILOSOPHY

    In the Graduate College

    THE UNIVERSITY OF ARIZONA

    19 7 7

  • THE UNIVERSITY OF ARIZONA

    GRADUATE COLLEGE

    I hereby recommend that this dissertation prepared under my

    direction by Bart lev Lowell Cardon

    entitled The Beam-foil Spectra of Krypton ( 2 to 5 MeVl

    be accepted as fulfilling the dissertation requirement for the

    degree of Doctor of Philosophy

    issertation Director Date

    As members of the Final Examination Committee, we certify

    that we have read this dissertation and agree that it may be

    presented for final defense.

    ^ft,/7? c—w,

    3/29/7;

    %/,. / f t / . y /so)?*

    Final approval and acceptance of this dissertation is contingent on the candidate's adequate performance and defense thereof at the final oral examination.

  • STATEMENT BY AUTHOR

    This dissertation has been submitted in partial fulfillment of requirements for an advanced degree at The University of Arizona and is deposited in the University Library to be made available to borrowers under rules of the Library.

    Brief quotations from this dissertation are allowable without special permission, provided that accurate acknowledgment of source is made. Requests for permission for extended quotation from or reproduction of this manuscript in whole or in part may be granted by the head of the major department or the Dean of the Graduate College when in his judgment the proposed use of the material is in the interests of scholarships. In all other instances, however, permission must be obtained from the author.

    djjy t&A-y

  • ACKNOWLEDGMENTS

    And so, here, the odyssey of ray graduate studies ends. Many

    people over many years, beginning with Sumner Davis at Berkeley, then

    Kenneth Andrew and Robert Stanley at Purdue, Henry Hill at Wesleyan

    and The University of Arizona, and finally John Leavitt at The Univer

    sity of Arizona, have contributed to seeing it all brought to a success

    ful conclusion. It was John Leavitt, above all others, whose calm and

    wise guidance, and example, provided me with a supportive and productive

    environment in which to fruitfully pursue my dissertation. His high

    standards are hopefully reflected in this work. Other members of the

    physics faculty who gave freely of their time and energy and to whom

    I express my appreciation are Carl Tomizuka, John Robson, Leon Blitzer,

    J. D. Garcia, Stanley Bashkin, Larry Mclntyre, John Stoner, Bill Bickel,

    Doug Donahue, Dan Dietrich, Bob Kalbach, John Kessler, Bob Parmenter

    Bob Chambers, Bruce Barrett, Mike Scadron, and John McCullen. Dick Van

    Reeth and the Physics Creative Laboratory were very helpful. Peter

    Stoss saw to the well-being of the Van de Graaff Laboratory and A1

    Sheehan and Richard Lamoreaux saw to my electronics and computer prob

    lems. Jackie Fahey, Evelyn Burros, and Lois Couch provided efficient

    and cheerful secretarial aid. John Howe gave the artist's touch. Hank

    Oona assisted me in several phases of the data taking. Ron Pamachena

    resolved some computer mysteries. Marsha Mapes endured the preparation

    iii

  • iv

    of the final spectral line list and carefully and diligently checked

    ray calculations. Barbara Otke Bickel was my excellent typist and I

    appreciate her care. Delmar Barker and Ed Middlesworth were good

    companions and thoughtful colleagues.

    A special note of appreciation is due my mother and father,

    sisters and their husbands, and Marion Meyer. Their constant support

    and devotion helped buoy me up during the darker moments. I thank them

    for their unselfish love.

  • TABLE OF CONTENTS

    Page-

    LIST OF ILLUSTRATIONS vii

    LIST OF TABLES ix

    ABSTRACT x

    1. INTRODUCTION 1

    2. EXPERIMENTAL APPARATUS AND PROCEDURE 6

    Accelerator and Selector Magnet 6 Excitation Chamber 8 Detector System 11 Spectrometer 14 Spectrometer Calibration 17 Spectrometer Refocusing and Intensity Enhancement .... 25

    Refocusing 29 Intensity Enhancement 32

    Recording and Reduction of Data ' 40

    3. DETERMINATION OF CHARGE STATE 54

    Charge State Distribution of Krypton 55 Multiple Scattering of Krypton in Carbon Foil 68 Charge State Identification Via Electrostatic Deflection . 70

    4. THEORY 81

    Polarization Model of the Atom 83 Spectral Line Intensities 99 Spectral Analysis Miscellany 103

    5. DISCUSSION OF SPECTRA AND CONCLUSIONS 106

    Krypton III 107 Krypton IV 108 Krypton V 108 Krypton VI 109 Krypton VII Ill Krypton VIII 115 Krypton IX-XIII 122 Conclusions 122

    v

  • TABLE OF CONTENTS--Continued

    vi

    Page

    APPENDIX A: GAUS2Z: SPECTRAL LINE FITTING PROGRAM ... 147

    APPENDIX B: PROGRAMS LAMSORT AND ORDSORT 150

    APPENDIX C: THE KRYPTON SPECTRAL LINE LIST f400 - 6000&) . 154

    APPENDIX D: RYDBERG AND LLAMBDA: ENERGY LEVEL AND SPECTRUM GENERATING PROGRAMS 186

    REFERENCES 189

  • LIST OF ILLUSTRATIONS

    Figure • Page

    1. Krypton beam deflected in an electrostatic field 3

    2. Side and top views of the experimental apparatus from the Van de Graaff accelerator to the McPherson monochromator 7

    3. Diagram of the photon counting detection system showing the photomultiplier tube, pre-amp, multiscaler, and digital-analogue output 12

    4. Diagram of the Fabry-Perot filter used to determine the periodic error in the spectrometer calibration 20

    5. Top view of monochromator with incident beam passing through foil and into Faraday cup 22

    6. Spectral line FWHM as a function of wavelength for the several refocused conditions of the spectrometer .... 31

    7. Focus wavelength X versus energy E of K̂r* beam for IE lenses 2 through 8 34

    8. Focus wavelength X versus energy E of 28N2 calibration beam for IE lenses 2 through 8 35

    9. Focus wavelength X versus IE lens diameter D 38

    10. Detector and grating combinations used to record Kr and N2 spectral scans 41

    11. Intensity enhanced spectra of 2 MeV krypton in the region 3375 - 3455& 43

    12. Intensity enhanced spectra of 5 MeV krypton in the region 2959 - 2985X . . . ' 44

    13. Histogramg of the number of observed kryptog wavelengths per 100A interval in the region 400 - 6000A 53

    14. Electrostatic deflection chamber for the determination of charge state distributions in krypton 57

    vii

  • viii

    LIST OF ILLUSTRATIONS—Continued

    Figure Page

    15. Charge state fractions as a function of incident beam energy 60

    16. Equilibrium charge state distribution for 1 and 2 MeV krypton in carbon foil 6 ugm/cm2) 62

    17. Mean charge q versus incident beam energy E 65

    18. Charge state distribution width d versus incident beam energy E 66

    19. The charge state electrostatic deflection apparatus .... 72

    20. Screening parameters versus orbital angular momenta for outer electron n = 4 - 10 in Kr IX 87

    21. Dipole polarizability of krypton ions versus core charge ? 92

    22. Relative intensities for An = 1,2,3 Rydberg supermultiplets 102

  • LIST OF TABLES

    Table Page

    1. Particle densities and numbers for nitrogen (1.67 MeV) and krypton (2,5 MeV) beams 11

    2. Properties of the McPherson 225 gratings 15

    3. Line widths of krypton and nitrogen beam-foil spectral lines 28

    4. Characteristics of the intensity enhancement lenses .... 36

    5. Spectrometer calibration parameters for the McPherson ,225 . 48

    6. Equilibrium charge-state fractions for 1 and 2 MeV krypton in carbon foil 64

    7. Mean charge and distribution width for 2,3,4, and 5 MeV krypton in carbon foil charge-state distributions .... 68

    8. Multiple scattering angles for 1, 2, and 5 MeV krypton in carbon foil 69

    9. Charge state analyzed krypton spectral lines 78

    10. Ground state electron configurations for Kr III - XIII ... 82

    11. Summary of krypton ion electron polarizability results . . 97

    12. Wavelengths and fine structure splittings for 4s4p 3P -4s4d D and 4s4p 3P - 4p2 3P of Kr VII 113

    13. Observed Rydberg transitions in Kr VII 114

    14. Observed and predicted Rydberg transitions in Kr VIII . . . 116

    15. Observed and predicted Rydberg transitions in Kr IX . . . . 123

    16. Observed and predicted Rydberg transitions in Kr X .... 128

    17. Observed and predicted Rydberg transitions in Kr XI . . . . 132

    18. Observed and predicted Rydberg transitions in Kr XII . . . 136

    19. Observed and predicted Rydberg transitions in Kr XIII . . . 141

    ix

  • ABSTRACT

    We evaluated the usefulness of the beam-foil source for pro

    ducing high quality spectral data and extended our knowledge of the

    spectra in the fourth period of the periodic table by recording the

    beam-foil spectra of krypton between 400 - 6000X. The Doppler broadening

    was reduced by refocusing and lens imaging techniques. The spectra were

    recorded with a McPherson 225 one-meter scanning monochromator. Special

    attention was given to determining the calibration of the spectrometer

    with the krypton spectra interposed with calibration scans of known

    nitrogen lines recorded under similar conditions. The incident krypton

    beam energies were 2, 3, 4, and 5 MeV. The photoelectric scans (inten

    sity versus wavelength) were fitted by hand and with a Gaussian curve

    fitting program on the CDC 6400 computer. Approximately 650 nitrogen

    profiles established the spectrometer calibration to within an average

    uncertainty of ± 0.3&. The approximately 2400 krypton line profiles

    were analyzed to produce a list of 965 calibrated vacuum wavelengths.

    We determined the equilibrium charge state distribution of 1 and 2 MeV

    krypton in carbon foil and extrapolated our results to 5 MeV. We mea

    sured the multiple scattering of krypton in carbon foil at 1 and 2 MeV

    and extrapolated our results to 5 MeV. The charge state distributions

    of krypton in carbon foil and the intensity variation in our spectra at

    2, 3, 4, and 5 MeV were used to deduce charge states. We investigated

    a direct method for determining charge states of specific line/

    x

  • xi

    line-complexes by deflecting the post-foil beam in an electrostatic

    field. We determined the charge states of 24 line/line-complexes and

    comment on the limitations of our method.

    A simple core polarization modification of the Bohr-Sommerfeld

    model of the atom is considered. We use this model to obtain values of

    the electronic dipolar polarizability for Kr VIII - XIII from our spec

    tral data. We find a quadrupolar polarizability for Kr XI and XII.

    These polarizabilities are used to predict other An = 1,2,3 (and 4, in

    the case of Kr IX) transitions in these ions. Relative line intensities

    among Rydberg supermultiplets are discussed and a recipe for obtaining

    intensity centers-of-gravity for close lying transitions is presented.

    A method for extrapolating values of the dipolar polarizability along

    an isoelectronic sequence is explained and used to predict the polariza

    bilities for Kr VI and VII. Features of Kr III-VIII, identified in our

    line list from the previous work of other authors, are discussed. Of

    the 956 spectral lines observed, and tabulated in our appendices, 434

    possess some kind of identification, 324 of which were previously

    unreported. Five hundred thirty-one (531) krypton wavelengths have yet

    to be identified. We conclude that the beam-foil source is capable,

    with care, of providing spectroscopic data accurate to about ± O.lA and

    that future effort might reduce this to ± 0.03A.

  • CHAPTER 1

    INTRODUCTION

    The field of beam-foil spectroscopy can properly be said to

    have begun with the pioneering researches of Kay (1963) and Bashkin and

    Meinel (1964). In the short span to 1977 this field has grown and

    developed enormously such that at least some part of it has been the

    basis of investigation at practically all major physics laboratories

    in the world. This has prompted four international conferences to be

    held over a period of nine years devoted to its various aspects. The

    published proceedings of these conferences (Bashkin 1968; Martinson,

    Bromander, and Berry 1970; Bashkin 1973; Sellin and Pegg 1976) chronicle

    the brief history and follow the increasingly sophisticated development

    of beam-foil spectroscopy. Indeed, the scope and character has so

    altered from its original notion of purely spectroscopic study that

    Martinson (Martinson and Gaupp 1974) has suggested that it might be

    more accurately described as atomic physics with ion accelerators.

    To appreciate some of the capabilities of beam-foil spectroscopy

    in the study of atomic processes, let us examine, in simplified terms,

    the beam-foil source. We project, by means of a particle accelerator,

    an incident ion beam on a thin foil target (typically carbon). As a

    result of this encounter, which results in the capture and loss exchange

    of electrons between the beam and the foil material, the beam emerges on

    1

  • 2

    the other side as a charge distribution of ions, the mean ionic charge

    higher, the greater the incident energy. On the back surface of the

    foil, or in the electron rich environment immediately adjacent, the ions

    recombine with electrons and begin radiating and continue to do so as

    they move downstream. In figure 1 we have a photograph of a krypton

    beam emerging from a carbon foil and partially separated into its ion

    components by means of a perpendicular electric field. The entire exci

    tation process occurs in a time period of the order of 10~15 seconds or

    less. We now have a coherent light source (by virtue of the abrupt

    excitation of the atomic states in the ions) which emits radiation from

    a distribution of typically highly ionized atoms, for which the observed

    decrease in light intensity as a function of distance downstream is

    related to the decay time of the atomic states participating (the time

    scale t is t = x/v where x is the distance downstream and v is the beam

    velocity).

    The coherence of the radiation and its anisotropy have been

    exploited to study fine and hyperfine structures, g factors, and Lamb

    shifts. The variation in intensity with distance downstream, which pro

    vides a direct measurement to the lifetime of the atomic states, has

    provided a wealth of lifetime data which had heretofore been very diffi

    cult or impossible to obtain by other spectroscopic sources. Recently,

    Andra (1976) used the beam-foil source in conjunction with laser excita

    tion to obtain cascade free lifetime measurements accurate to ± 0.15%,

    a remarkable achievement. Despite these fruitful applications of the

    beam-foil source, however, the question remains of how useful it can be

    for providing spectral results and be in some way competitive with other

  • 3

    Figure 1. Krypton beam deflected in an electrostatic field.

    A h. vA, 2 MeV krypton beam emerges after colliding with a carbon foil at left and is deflected upwards by a 24 KV/cm electric field. The bottom field plate is positive and is strongly fluorescing due to electron bombardment. The lower edge of the upper plate, 1 cm away, is barely visible. The photograph is a Cibachrome print of a Kodachrome 64 slide taken at f/3.5 for 12 minutes with a 50 mm lens. Scale is 1:1.

  • 4

    spectroscopic sources, e.g., the sliding spark and theta-pinch light

    sources, which also produce high stages of ionization (albeit not as

    readily or easily as the beam-foil source), but with narrower line

    widths. It was early recognized that the price one paid for high stages

    of ionization in the beam-foil source was large Doppler broadening as a

    consequence of the large beam velocities used. This broadening was

    sufficient to obscure all but the grossest features in the visible part

    of the beam-foil produced spectrum. In addition, as the techniques of

    beam-foil spectroscopy have nevertheless yielded valuable lifetime and

    spectral information concerning the lighter elements, where there exists

    ample data from earlier work with other sources to act as a guide, it

    has become desirous to extend spectral analyses to the higher stages of

    ionization of heavier elements where there is much less supportive infor

    mation from other studies. Furthermore, as a result of the efforts now

    being put into controlled thermonuclear fusion, there is a demand for

    spectroscopic information concerning wall materials and diagnostic gases

    which are highly ionized.

    To evaluate the degree to which the beam-foil source would be

    useful in providing such information and to assess the accuracy attain

    able in a beam-foil spectroscopic study, we decided to undertake an

    extensive and thorough analysis of the beam-foil spectra of an element

    incorporating recent advances in the reduction of spectral line width

    from fast-ion sources with special attention paid to obtaining the best

    calibration of the spectrometer consistent with its capabilities. The

    resulting spectra would then be used to test the suitability of a simple

    atomic model for determining electric dipole polarizabilities of the

  • 5

    electron cores and screening parameters. These quantities and their

    extrapolation along isoelectronic sequences have been valuable in system

    atizing the hyrogen-like transitions in the lighter elements and it

    remains to be seen how useful they will be in elucidating the beam-foil

    spectra of the heavier elements.

    The element of study was krypton, isotope A = 84 (56.9% abun

    dant), which comments itself by being monatomic, chemically inert, and

    easy to use in the particle accelerator. Its atomic structure places

    it near the middle of the naturally occurring members of the periodic

    table. Preliminary calculations of the anticipated spectral features,

    using a simple modified Bohr atomic model, indicated fine structures

    which would be resolvable and would therefore reveal something about

    the details of the ions responsible. The well known atomic spectra of

    nitrogen, produced under conditions identical to that of krypton, were

    chosen for internal calibration.

  • CHAPTER 2

    EXPERIMENTAL APPARATUS AND PROCEDURE

    The equipment used to produce and record the krypton spectra and

    attendant nitrogen calibration spectra consisted of four basic elements

    (see figure 2): the particle beam accelerator and selector magnet, the

    excitation chamber wherein the beam-foil interaction occurred and spectra

    were produced, the spectrometer which analyzed the spectra, and the detec

    tor system which photoelectrically recorded the spectrometer output and

    generated analogue and digital records of intensity versus wavelength.

    Each of these elements will be described, with special emphasis on the

    spectrometer, as it was this instrument which determined the final

    quality and reliability of the spectral data obtained.

    Accelerator and Selector Magnet

    The particle accelerator is a model CN Van de Graaff with a

    utilizable accelerator voltage range of 1-5.5 MV. The accelerator is

    mounted vertically. The accelerated beam, upon exit, passes through a

    90° bending magnet which directs the ion beam horizontally to the exci

    tation chamber. The 90° bending magnet can be operated at sustained

    fields of more than the 13,200 gauss required to deflect a 5 MeV 84Kr+

    beam. The 90° deflection angle insures excellent selection of a specific

    q/m, thereby isolating a beam free of contaminants from nearby q/m values.

    6

  • foil wheel viewing (Translatable •«•») port Model CN

    Van de Graaff Accelerator

    Faraday Cup

    p7>—photo-detector ^4 housing

    excitation chamber

    grating

    McPherson 225

    Monochromator 90° analyzing / magnet

    quadrupole

    focusing magnet

    side view top view

    Figure 2. Side and top views of the experimental apparatus from the Van de Graaff accelerator to the McPherson monochromator.

    The side and top view scales are not the same.

  • 8

    This is substantiated by the complete absence of any noticeable impurity

    lines in the krypton spectra.

    Excitation Chamber

    This is a stainless-steel box which fits over the entrance slit

    housing of the spectrometer and is evacuated by both an oil diffusion

    and Ti getter pump, the latter pump possessing a low pumping efficiency

    for the noble gases. These pumps maintain a high vacuum environment

    (1 - 2 x 10"6 torr) and consequent low background gas pressure for the

    incident ion beam. At the downstream end of the chamber the ion beam

    terminates in a large aperture (2" diam.) Faraday cup with electron

    suppressor ring (battery biased at -180 V). The current from the Fara

    day cup permits the amount of the incident ion beam to be continuously

    monitored. A quartz window port-hole opposite the entrance slit enables

    observation of the ion beam as well as the introduction of light into

    the spectrometer from external stationary sources. The foil wheel is

    mounted in the upstream side of the chamber. It is a stainless-steel

    notched wheel mounted on a horizontal shaft and capable of holding 24

    foil holders, each holder containing two foils over 3/16" holes. This

    horizontal shaft may be translated along the beam axis permitting the

    distance from foil to spectrometer optic axis to be varied for lifetime

    measurements. A linkage between the horizontal shaft and the outside of

    the chamber allows any of the 48 possible foil positions to be rotated

    into the beam path. A typical time interval of one hour was required to

    bring the excitation chamber and spectrometer up to atmospheric pressure,

  • 9

    remove the foil wheel, change foils, replace the foil wheel, and pump

    back down.

    The foils used throughout the experiment were made of spectro-

    scopically pure carbon. They were fabricated in this laboratory and

    their thicknesses, selected to be between 4-8 ugm/cm2, were determined

    according to the method described by Stoner (1969). Approximately

    2000 carbon foils were used in the execution of this beam-foil study.

    The lifetimes of the carbon foils dictated the amount of current

    selected for the incident ion beam. For the 1.67 MeV N2 incident beam

    used for producing the calibration spectra, a typical foil lifetime was

    nominally 10 minutes for beam currents of 6 - 8 uamps. For the incident

    krypton ion beams the following average behavior was observed: two min

    ute foil lifetime for a 2 namp, 2 MeV, Kr+ beam and a five minute foil

    lifetime for a 5 pamp, 5 MeV, Kr+ beam. To maintain foil lifetime and

    prevent wandering of the beam position, it was essential that the beam

    be broadly focussed, i.e., that it have a diameter at least the size of

    the foil or 3/16".

    For the above beam parameters we can calculate the approximate

    number of radiating atoms within the acceptance angle of the spectrom

    eter. Let the krypton beam have a velocity v. Then,

    v = gc = c v/2E7931A , (2.1)

    where E is the beam energy in MeV, A the atomic number of the beam ion

    (A = 84 for the krypton isotope used throughout the experiment), and

    c the celerity of light. Let the ion beam current be i. Then,

  • 10

    i = n q e v a , (2.2)

    where n is the beam particle density, q the beam mean charge (determined

    from the charge state distribution of the post-foil beam), e the charge

    of the electron, and a the cross-sectional area of the beam. The number

    of particles which pass through the foil per unit time is

    N/t = nva , (2.3)

    where N is the number of particles contained in the defining cylinder of

    the beam (assuming no divergence) which radiate within that part of the

    beam intercepted by the acceptance angle of the spectrometer. This

    cylinder is ad long and a particle takes time t = ad/v to traverse this

    distance. Hence,

    n = /931A/2E (2.4) qeca

    and

    N = naad . (2.5)

    The acceptance angle a is 1/10 rad., the perpendicular distance of the

    beam from the entrance slit is d = 6.6 cm, and the beam diameter is 4.8

    mm, from which we construct table 1. Note that the background gas at

    a pressure of 5 x 10~6 torr and 300°K has a particle density of

    ̂2 x 1011 particles/cm3.

  • 11

    Table 1. Particle densities and numbers for nitrogen (1.67 MeV) and krypton (2,5 MeV) beams.

    A E (MeV) i (pamp) q n (part./cm3) N (part.)

    28 1.67 7 2.0 3.5 x 105 4.1 x 10̂

    84 2.0 2 5.6 5.6 x 10̂ 6.6 x 103

    84 5.0 5 8.9 5.6 x 10̂ 6.6 x 103

    Detector System

    The photon counting detector system consists of a photomulti-

    plier with power supply, battery operated preamplifier, 4000 channel

    multiscaler, analog strip-chart recorder, and high-speed digital printer.

    A diagram of the detector system is shown in figure 3. To cover the

    O spectral range 400 - 6000A accessible with the spectrometer, three

    photodetectors were employed. These detectors and their operating

    parameters are summarized below:

    (1) Bendix 4219 EIC Spiraltron Electron Multiplier - windowless.

    O Operated at room temperature; range of use 400 - 1600A (1% of peak sensi

    tivity at 1500X); cathode is grounded with collector at +2200V;

    wavelengths observed above 1500A are of second or higher order.

    (2) EMR 541F-08-10 - selected ultraviolet LiF window. Operated

    at room temperature; range of use 1050 - 3000A (abrupt lower cut-off at

    1050A); cathode is at -3800V; wavelengths observed above 2100& can be

    second or higher order.

    (3) RCA 1P28 - Corning #9741 UV transmitting glass window.

    Operated at LN2 cooled temperature; range of use 2500 - 6000A; cathode

  • disc, level

    discriminator ireamp counter memory o-

    Intertechnique Multiscaler

    x>

    Q. CD

    Figure 3. Diagram of the photon counting detection system showing the photomultiplier tube, pre-amp, multiscaler, and digital-analogue output.

    A representative photon pulse is shown, amplified, discriminated, and shaped for the pulse counter. •-•

  • is at -800 V; (1% of peak sensitivity at 1700 - 1800A); wavelengths

    O observed above * 3400A can be second or higher order.

    An important feature of these detectors is the cut-off wave

    lengths for their different regions of response. This greatly facili

    tated culling out second and third order lines in the final reduction

    of the spectral line list. All the detectors performed in a predictable

    and trouble-free manner.

    The preamplifier is an in-house designed and fabricated (Sheehan

    and Lamoreaux 1974) battery-operated circuit employing an MC 1552 mono

    lithic video amplifier chip. This preamp has a bandwidth of about 10

    MHz and gain of 50-100. Its operation was plagued from time to time

    by pickup. This was eventually eliminated by isolation from any nearby

    power supply cables and insuring that both preamp and preamp housing

    were properly grounded.

    The output from the preamp was fed into a DIDAC Intertechnique

    4000 channel multiscaler where it was further amplified and the pulse

    height discriminated to reduce noise, and finally counted. The onset of

    the counting process was activated by a pulse from the spectrometer

    drive, enabling us to begin the photon counting process at a known and

    reproducible setting of the spectrometer. The duration of the photon

    counting for one channel, the channel integration time, was flexible and

    could be set in the multiscaler. When the counting for one channel was

    terminated, the results were stored in the memory of the Intertechnique

    and the counting process was automatically advanced to the next channel.

    The number of channels, up to 4000, was flexible and could be set in the

    multiscaler. Upon completion of the photon counting for the number of

  • 14

    channels chosen, the counting process either terminated or automatically

    restarted. The results of each cycle or scan was displayed on an oscillo

    scope. From the memory, four bins which retained the channel contents

    of four different scans, the number of counts per channel for each scan

    can be retrieved and can be either plotted on a Heath model EU-20B servo-

    recorder (analogue output) or printed out in digital form on adding

    machine paper by a Franklin Printer (Mohawk model 1200 Digital Strip

    Printer). It was typical that both analogue and digital records of the

    spectral scans were generated, the plots providing useful visual evi

    dence of the scans, the digital print-out of photon counts versus chan

    nel number being ultimately transferred entirely to IBM punch cards for

    computer analysis.

    For all the spectral scans in this experiment the integration or

    dwell time per channel was 0.6 seconds, a value chosen so as to make the

    channel duration correspond to convenient intervals of wavelength. For

    O O example, if 50A are to be scanned at a spectrometer scan rate of 5 A/min.,

    it will take 10 minutes to perform the scan. For a 0.6 second dwell

    time per channel we will require 1000 channels. Each channel in turn

    will correspond to 0.05A in wavelength.

    Spectrometer

    The spectrometer used to obtain the spectral data in this

    experiment was a McPherson Model 22S normal incidence one-meter vacuum

    UV scanning monochromator. Two gratings enabled wavelength coverage of

    O 400 - 6000A. The properties of these gratings are listed in table 2.

  • 15

    Table 2. Properties of the McPherson 225 gratings.

    Grating 1 Grating 2

    rulings 600 Jl/mm same

    size 96 mm x 56 mm 90 mm x 50 mm

    radius of curvature 995.4 mm 998.8 mm

    coating A1 with MgF2 overcoat same

    blaze wavelength 1500A 5000&

    construction tripartite monopartite

    manufacturer Bausch § Lomb same

    SSF: ent. slit gap 2.2 mm 2.2 mm

    exit slit gap 2.9 mm 9.0 mm

    micrometer setting 0.360" 0.285"

    The final three rows of table 2 are the optimum settings for

    stationary source focus of the spectrometer. They refer to the gaps

    between the entrance and exit slit housing and the tube extensions of

    the spectrometer body and the respective settings of the grating microm

    eter inside the spectrometer.

    XB The useful range for high efficiency of a blazed grating is ± -y

    about the blaze wavelength (Davis 1970, pp. 9-12). For grating 1 this

    O would mean a range of 750 - 2250A. In practice we found these limits

    could be stretched and the useful range extended from 400 - 3000X. With

    grating 2 the useful range is 2500 - 6000A, the upper wavelength repre

    senting the mechanical limit of the spectrometer drive. All data taken

  • with this grating were within the interval 2500 - 6000X. The plate fac

    tor for both gratings, with the spectrometer at stationary source focus,

    is 16.6 A/mm.

    A six inch oil diffision pump immediately beneath the spectrom

    eter body provided a vacuum of 1 - 5 x 10"6 torr when the spectrometer

    was evacuated. In practice it was essential that the spectrometer be

    evacuated for all the beam-foil studies, ijr calibration work with

    stationary sources and when changing foils in the excitation chamber,

    the spectrometer was filled with dry nitrogen. A high vacuum (< 1 x 10~5

    torr) was mandatory for the proper operation of the windowless Bendix

    spiraltron photodetector.

    As mentioned earlier, the spectrometer provided a starting pulse

    for the multiscaler. This was facilitated by a 3-3/4 inch notched wheel

    attached, by sprocket-chain-idler wheel with a 5:1 reduction, to the

    spectrometer drive shaft near the wavelength odometer (displaying wave-'

    lengths to the nearest Angstrom). These four notches, 90° apart, stroke

    a microswitch in a 4.5 V battery powered circuit which is normally open

    but can be closed momentarily by a pressure switch activated from the

    side of the spectrometer body. Each turn of the master screw corre-

    O sponds to a wavelength change of 50A which, with a 5:1 reduction, means

    O the notched wheel turns through 10A per revolution making available a

    starting pulse every 2.5A. It was crucial to the reproducibility of all

    the spectral scans taken, especially in the case of multiple scans over

    the same spectral region, that the starting pulses be reliably dupli

    cated. The starting pulses were therefore continuously checked during

    the initial (about six months) calibration period of the spectrometer.

  • 17

    These checks were most readily made by scanning many times over the same

    stationary source spectral line or line complex and observing the shift

    in line center and change in line width. In this way, aside from the

    thermal drift in the spectrometer to be discussed shortly, the starting

    pulses were found to be reproducible to ± 0.05A, a value below the un

    certainty in the wavelength calibration of the spectrometer.

    Of the twelve synchronous reversible scanning speeds available

    from the drive mechanism, only four speeds were used to perform actual

    scans (although the screw drive was slewed to starting points at 200 or

    500 A/min). These were 50 A/min, 20 A/min, 10 A/min, and 5 A/min, producing scans of 0.5 A/chan., 0.2 A/chan., 0.1 A/chan., and 0.05 A/chan.

    O The rapid scans of 0.5 A/chan. were called survey scans, the others,

    slower, resolution scans. Most of the krypton spectral scans were taken

    O at 0.1 and 0.2 A/chan.

    Spectrometer Calibration

    The prior use history of our spectrometer was not accurately

    known except that previous studies (Stoner and Leavitt 1971b; the

    difference between the observed and handbook value of the nitrogen

    fine structure separation at X = 3482A reported therein is 0.4A) had

    indicated an anomalous behavior in the wavelength calibration. As it

    was the consuming goal of this spectral study to obtain the highest

    quality data heretofore accessible with a beam-foil source, in coopera

    tion with optimization of the spectral line width possible with refocus-

    sing of the spectrometer, thereby producing spectral lines measurable to

    O ± 0.05 - 0.10A, it was essential to eliminate any peculiarities in the

  • spectrometer calibration and determine accurately and reproducibly the

    wavelength calibration. In a straightforward, although time consuming,

    way the general performance and reproducibility of the wavelength deter

    mination with the spectrometer were systematically checked. This was

    accomplished in four stages. In stage one the entrance and exit slits

    were calibrated and found to be within ± 3p of the slit micrometer

    settings. Relative separations between a large number of stationary

    source wavelengths generated by a Hg penlite source and He, Ar, and Ne

    filled Geissler tubes were measured. It was readily apparent that the

    calibration anomaly had a periodicity of 5oX corresponding to one revo

    lution of the grating drive master screw. A careful inspection was made

    of all mechanical aspects of the drive mechanism seeing to it that all

    set screws in the gear train were properly seated and tight. The master

    and slave screws, as well as the motor and drive bearings, were carefully

    lubricated. No binding or damage to any of the gears could be found.

    It was during this period that a long-period and unpredictable drift in

    the peak location and line width of strong Hg calibration lines lead us

    to suspect that changes in the ambient temperature at the spectrometer

    were also influencing wavelength calibration. Daily records were kept

    of the temperature with a mercury-in-glass thermometer situated atop

    the spectrometer body (2 meters above the floor).

    In stage two a Fabry-Perot etalon was used as a wavelength

    filter (Tolansky 1970) to produce calibration wavelengths at controlled

    and regular spaced wave number intervals for the wavelength interval

    3400 - 5400X. For a Fabry-Perot etalon of spacing t there is a trans

    mission peak for wavelength X at angle 0 with respect to the optic axis

  • 19

    if 2t cos 9. = pX where p is the order of interference. Restricting

    o b s e r v a t i o n o f t h e i n t e r f e r e n c e p a t t e r n t o n e a r t h e o p t i c a x i s , 0 ^ 0 ,

    and 2t = pX. We let a = 1/X so 2ta = p. If two transmission peaks

    of wavenumber a and a + ACT differ by one in order of interference,

    then Aa = l/2t. (Act is called the free-spectral-range of the etalon.)

    Wavelengths which differ from a by multiples of Aa are therefore trans

    mitted. If a continuum light source illuminates the etalon and the

    resulting central region of the interference pattern is projected onto

    the entrance slit of the spectrometer, a series of discrete wavelength

    peaks are produced which are equally spaced by wavenumber ACT = l/2t and

    whose range is limited only by the continuum light source and the reflec

    tivity of the etalon coatings. Figure 4 is a diagram of the Fabry-Perot

    arrangement which produced the calibration fringes used to check for

    periodic error in the spectrometer drive. An air-cooled tungsten halo

    gen lamp and Leitz projector provided sufficient intensity to produce

    O calibration peaks in the range 3400 - 5400 A. The etalon spacer was

    comprised of three tungsten carbide balls, 120° apart, either 1/32" or

    1/64" in diameter producing a free spectral range of 6.2 or 12.5 cm"1

    O ° (1.0 or 2.OA at 4000A). The measured overall finesse (the ratio of the

    free spectral range to the FWHM of the transmission profile) of the

    etalon was 3.8. The transmission peaks of the etalon were scanned by

    the spectrometer and the resulting peak centers determined. A least

    squares fitting program from the IBM Scientific Subroutine Package

    (IBM 1968) was modified to fit the peak number versus wavenumber to a

    first or second order polynomial. If the calibration of the spectrom

    eter was free of error; then the first order fit should have zero

  • ^ 20.5 cm, ^ „ 30.5 cm . ^

    front surface mirror

    f 19 cm

    i fWs ESSS

    port window (quartz)

    _ j ) ^16 dia. stop !

    Fabry-Perot etalon

    ground glass plate

    |ens

    i

    % entrance slit (30/x)

    continuum light source

    laser (to check alignment)

    Figure 4. Diagram of the Fabry-Perot filter used to determine the periodic error in the spectrometer calibration.

  • 21

    residuals between the observed wavenumber and the calculated wavenumber.

    Any periodic error in the calibration or systematic drift would show up

    in wavenumber residuals. Hence, the residuals were determined and dis

    played by plotting the residual wavenumbers, converted to wavelength,

    versus wavelength. These results convincingly demonstrated the presence O

    of a periodic error per screw revolution of magnitude ± 0.5A. Fitting

    of this data to a second order polynomial did not change the conclusions

    except to indicate that there might be long term variations in the spec

    trometer calibration or long term thermal drift in the Fabry-Perot

    etalon.

    In stage three we assumed that the spectrometer calibration had

    two sources of error: a periodic error lodged somewhere in the motor-

    drive,slave screw,and master screw train, and a long term mechanical

    distortion of the spectrometer body due to thermal drift. By observing,

    with a cathetometer, the rate of passage past a cross-hair, the teeth

    in particular bevel gears on the drive shafts leading to the master

    screw, as well as the rate of turning of the drive-box selector dial,

    we determined that the error in the drive up to the master screw was

    accurate to ± 0.1&. One revolution of the master screw corresponded

    to 50&. With the spectrometer drive set at 2 A/min, a periodic error

    O of ± 0.1A would show up in the field of view as the early or late arrival

    by ± 3 seconds of the final tooth in the bevel gear, a time interval

    measurable with a stopwatch. The master screw was further checked by

    placing a depth gauge against the end of the grating table arm resting

    against the master screw nut block (see figure 5). The observed depth

    gauge reading was compared with the expected reading as the screw

  • VDG BEAM

    TOP VIEW:

    FOIL

    CUP GRATING

    SCREW DRIVE^

    GRATING MIC-""^*^ CAM

    PHOTOTUBE

    Figure 5. Top view of monochromator with incident beam passing through foil and into Faraday cup.

    Light passes through the entrance slit, reflects off the grating seated on a base attached to a spring loaded arm which, riding along the adjustable cam, is driven by the master screw, and through the exit slit to the phototube.

    N) N>

  • 23

    turned through one revolution, the fractions of a revolution noted by

    the passage past a cross-hair of a tooth in a bevel gear (25 teeth)

    observed with a telescope. The pitch of the master screw was previously

    O determined to be 0.6983 mils/A. A ± 5 mil feeler gauge rested against

    the top of the nut block to watch for screw and bearing wobble

    (Ingalls 1952a,b,c). Although no noticeable wobble was measured, the

    O master screw was found to have a ± 0.5A error per revolution. We

    decided that the most dependable remedy to this problem was to replace

    the entire master screw assembly with a new one from McPherson. When

    the new master screw assembly was installed, the grating table was

    completely dismantled, cleaned, and reassembled. During reassembly of

    the table, ball bearings upon which the table rides were repositioned

    so as to avoid noticeable wear spots in the kinematic grooves. The

    entire master screw and drive were again thoroughly lubricated, a ritual

    that was followed at regular intervals during the nine months the spec

    tral data were recorded.

    In stage four, the new master screw was checked for periodic

    error and wobble. The improvement was dramatic. The detectable period

    ic error did not exceed ± 0.1&, a value we felt to be consistent with

    the ultimate capability of the instrument. We then determined the spec-

    O trometer wavelength calibration in the region 2500 - 6000A using wave

    lengths from stationary sources of Hg, Ne, and He. The resulting curve,

    a plot of wavelength correction versus McPherson wavelength, showed a

    smooth, monotonic change with the measured points scattered within

    ± O.lA above and below this curve. However, it remained obvious that

    remedying the errors in the master screw did not remove all errors in

  • the calibration. There remained unexplained day to day shifts in the

    location of the peak centers of calibration wavelengths indicating

    potential systematic errors of greater than ± O.lX. for data not recorded

    close together in time. The greatest source of these errors was corre

    lated with changes in the temperature gradient top and bottom of the

    spectrometer body which resulted from changes in the heating of the

    bottom of the spectrometer when the drive motor was turned on and off.

    It, therefore, became a mandatory procedure to turn the drive motor on

    at least one-half hour before taking data and to leave it on for the

    duration of the data taking day. In addition to the thermometer atop

    the spectrometer, a second thermometer was placed in close proximity to

    the motor and spectrometer base. Readings from both thermometers were

    recorded continuously during all data taking. A fan was installed to

    continually flush fresh air under the spectrometer and thereby prevent

    stagnation of heated air under the spectrometer from nearby pumps and

    motors. We attempted to relate the ambient room temperature and temper

    ature differential top and bottom of the spectrometer to observed

    wavelength .shifts and changes, in line width, but our efforts were incon

    clusive and not pursued. Enclosing the entire spectrometer in a temper

    ature controlled environment was considered impractical. The temperature

    sensitivity of the spectrometer was accepted as an inevitable source of

    error in the wavelength calibration and was ultimately the limiting

    factor in the quality of our wavelength determination.

    It might be concluded from our experience above that the

    McPherson 225 is not the instrument to use for these high quality spec

    tral studies; nevertheless, it was the instrument available and our

  • 25

    calibration tests and remedies enabled us to thoroughly and confidently

    understand its characteristics.

    Spectrometer Refocusing and Intensity Enhancement

    The limitation on the quality of the data that can be obtained

    from a beam-foil source is governed by the line width of the spectral

    lines. Poor line width leads to loss of resolution, washing out of

    spectral details, and the poor determination of line centers. This,

    in turn, leads to poor calibration curves and subsequent poor wavelength

    determinations. In the preceding sections the spectrometer calibration

    limits were discussed. In this section attention is given to the mini

    mization of spectral line widths produced by the beam-foil source.

    From the consideration of the relationship of spectrometer reso

    lution to through-put luminosity CJacquinot 1954), it is clear that

    narrowing the line width alone does not improve the spectrometer perform

    ance. Reducing the line width to increase the resolution decreases the

    spectrometer through-put, and increases the detection time. In practice,

    one strives for a balance between narrow line width and high luminosity.

    This is especially important for beam-foil sources where there is low

    inherent luminosity and limited foil lifetime. It is not useful here to

    discuss the theory of the grating spectrometer (Stroke 1967, Davis 1970)

    or the origins of line width in fast ion-beam sources (Stoner and Leavitt

    1973). We merely review here the salient features of grating spectrom

    eters and spectral line width as they pertain to the McPherson 225 and

    this experiment.

  • 26

    The optimum position settings of the entrance and exit slits

    on the Rowland circle for stationary source focus (SSF) of the McPherson

    225 is provided by the manufacturer. These settings are achieved by

    two adjustments: the positioning of the slit housings in their respec

    tive tube extensions in the spectrometer body, a positioning that

    affords a coarse focus setting, and moving the grating table precisely

    by means of a micrometer screw (see figure 5). With the entrance and

    exit slits equal and set at approximately the optimum slit width

    (Sawyer 1963), we observed the change in line width as a function of

    the micrometer setting for a stationary source wavelength (typically

    Hg 5461A, 3650A, or 2537A). The minimum in the plot of line width

    versus micrometer setting established the optimum focus setting. We

    thereby confirmed the SSF settings for the two gratings used in this

    experiment. The values of the body-tube separations for the entrance

    and exit slits and micrometer settings are listed in table 2. The

    cammed arm against which the grating table arm rides assures that once

    the spectrometer is optimally focussed for one wavelength it is focussed

    for all wavelengths. The instrumental line width is governed by the

    dispersion of the spectrometer and, for both gratings used in this

    experiment, was 16.6 A/mm. For 50 p slits this yields a line width of AXj = 0.8A.

    The actual observed line width, however, depends not only on

    the instrumental line width, but the source line width as well. The

    actual line width properly involves the convolution of the natural line

    width (normally neglected as it is typically less than O.OIOA), line

    broadening effects in the source, and the instrumental width. The line

  • broadening effects from the source are the Doppler broadening due to

    the velocity of the ion-beam and finite acceptance angle a of the spec

    trometer, and multiple scattering of the ion-beam in the foil. The

    nonrelativistic (valid for the energies encountered in this exerpiment)

    Doppler broadening due to finite acceptance angle is given by

    AAq = 2XQg sin a/2 . (2.5)

    For small values of a, equation 2.5 becomes

    AXn = X g a (2.6) Do v

    where Xq is the rest frame wavelength, g = v/c with v the beam velocity,

    and a the acceptance angle of the spectrometer. For example, consider

    a 5 MeV 0t*Kr+ beam so that 3 = 0.0113, and let a = 1/10. Then AX̂ =

    1.1A for X = lOOoA and AX_. = 6.8A for X = 6000A. We can reduce the o Do

    Doppler width for a given Xq and g by decreasing a through masking the

    grating with, however, a resultant decrease in through-put for the

    spectrometer.

    The angular spread in the beam due to multiple scattering in

    the foil produces a resulting line width described by

    AX = H JuTTo1 . (2.7) so '

    This result is similar to the Doppler width with the acceptance angle

    replaced by /£n 2 • 0. Here 0 is the r.m.s. multiple scattering angle

    (see discussion of table 8, page 69) Again for a 5 MeV 81fKr+ beam

    0 = 2.3 x 10 2 rads., and AXg = 0.2A at 1000A and 1.3A at 6000A. For

    a 2 MeV K̂r beam these values are multiplied by 1.6.

  • 28

    If we assume that the line profiles due to instrumental effects,

    Doppler broadening, and multiple scattering can all be described by

    Gaussians, the resultant total line width is a simple sum of quadra

    tics, namely,

    AX2 = AX2 + AX2 + AX2 . (2.8) T I D S

    Values of AX̂ , for 50 u slits are listed in table 3. It is evident

    from the above discussion that the ability to do any medium resolution

    (X/AX ̂ 10,000) spectroscopy with a beam-foil source is severely limited

    by the Doppler broadening. Recent investigations, however, by Stoner and

    Leavitt (1971a,b), Leavitt and Stoner (1972), and Leavitt, Robson, and

    Stoner (1973) have shown that the Doppler broadening can be significant

    ly reduced without simultaneously severely reducing the spectrometer

    through-put. Indeed, in the instance of using a lens imaging technique,

    aptly named "intensity enhancement," the available intensity was

    Table 3. Line widths of krypton and nitrogen beam-foil spectral lines.

    Ion-Beam X (X) A^-f(^) O T

    5 MeV 8lfKr+ 1000 1.4

    6000 7.0

    2 MeV 8ItKr+ 1000 1.1

    6000 4.9

    1.67 MeV 28N2 1000 1.4

    6000 6.9

  • o improved while providing line widths of 1 - 2A. We discuss below the

    two techniques, refocusing the spectrometer and intensity enhancement,

    which we employed to reduce the Doppler broadening in our beam-foil

    spectral source.

    Refocusing

    The theory behind refocusing of a concave grating spectrometer

    has been thoroughly discussed by Leavitt, Robson, and Stoner (1973).

    By moving the exit plane back a distance

    D = XQg/K (2.9)

    where K is the plate factor (reciprocal linear dispersion) for the

    spectrometer, the exit slit is placed at the convergence point of the

    Doppler shifted rays coming from the fast-ion beam. For the McPherson

    225, the exit slit is not typically moved. Rather, by means of the

    micrometer adjustment, the cam akd grating table are moved back a

    distance D/2. Refocusing, in this fashion, for a particular Aq insures

    the spectrometer will be refocused for ail Xq, as was also true for the

    SSF, for the particular beam velocity v.

    In practice we refocused the spectrometer beginning at the SSF.

    The position of the entrance slit was left undisturbed. We selected a

    particular XQ and beam energy of interest from which D was then deter

    mined. We next determined the distance along the cam (figure 5) at

    which the grating table arm was located from the zeroeth order position

    for the X under consideration. This enabled us to calculate what o

    fraction f this position is to the position of the arm at 6000A

  • 30

    (the grating arm moves a distance 0.12"/1000X along the cam). When the

    cam micrometer is displaced a distance x, the grating table and grating

    are moved back a distance fx. This in turn is equivalent to moving the

    exit slit a distance 2fx off the Rowland circle. For a refocus distance

    D, we want, therefore, to move the micrometer a distance x = D/2f.

    Often, however, the beam velocity is so large that the micrometer travel

    is not sufficient to refocus for the entire distance D. Under these

    circumstances the exit slit must be moved as well. For the energies

    used in this experiment, all refocusing was accomplished with the cam

    micrometer, the exit slit remaining at its position for SSF. Starting

    with the SSF values in table 2, the calculated micrometer settings used

    for 5 MeV ahKr and 1.67 MeV 28N2 were 0.243" for the 1500& blaze

    grating and 0.168" for the 5000X blaze grating. The actual values

    used in this experiment were 0.248" and 0.200" respectively. Hence,

    for the 1500$. blaze grating the spectrometer was refocused for 5 MeV

    01tKr+. The resulting line width for 2 MeV 8lfKr was found to be essen

    tially the same as for 5 MeV 8ltKr+ permitting us to perform our spectral

    scan at different energies with one refocus setting. The refocus

    O setting of the 5000A blaze grating represents a compromise setting

    between 2 and 5 MeV 8lfKr. The line width as a function of wavelength

    is displayed in figure 6.

    Data which were ultimately to be analyzed were taken with the

    spectrometer refocused and equal slit widths of 50 y or 100 y (except

    for one scan performed with 75 y slits). The 100 y slits were typically

    used in regions with weaker lines. Both choices of slit settings

  • 2.0

    1.8

    °S

    2 X $ li.

    JZ +-

    -g 5 a> c:

    1.6

    i:o

    © - 2 MeV Kr, 1500 A blaze, refocus

    ® - 5 MeV Kr. 1500 A blaze, refocus

    © - average of 2-5 MeV Kr. 5000 A blaze, refocus

    1.4 -

    ©

    1.2 -

    1000 2000 3000 4000 5000 6000

    wavelength (&)

    Figure 6. Spectral line FWHM as a function of wavelength for the several refocused conditions of the spectrometer. w

  • 32

    intentionally exceed ws0pt and ŵ 0pt discussed by Leavitt, Robson, and

    Stoner (1973) because of the low intensity available in the krypton

    beam-foil source. The consequent line widths are therefore slightly

    larger than the optimum line width discussed in the above paper.

    Intensity Enhancement

    For a spectrometer set at SSF, Stoner and Leavitt (1971a) and

    Bergkvist (1976) have shown that a suitably placed cylindrical lens on

    the optic axis between the fast-ion beam and the entrance slit will

    obtain narrow line widths with attendant large entrance slit widths.

    The lens permits an increase in through-put of the spectrometer because

    of increased entrance slit width while maintaining narrow line width

    through partial compensation of the Doppler broadening. In this experi

    ment, the average observed increase in peak intensity was 5.8 fold over

    results taken under the same conditions with the spectrometer refocused.

    This increase in intensity by use of a cylindrical lens is called Inten

    sity Enhancement and the lens, Intensity Enhancement Lens (or IE Lens).

    O The resultant decrease in the total line width at 6000A was a factor of

    three compared to results obtainable with the spectrometer viewing a

    fast-ion beam at SSF.

    The results of Stoner and Leavitt (1971 a) show that the focal

    length f of the cylindrical lens is related to the beam velocity v, the

    rest frame wavelength XQ, the spectrometer plate factor K, the grating

    angles of incidence i and diffraction r, the grating focal length fg,

    and the lens-grating distance p by

  • 33

    W V*WO J. I jj~x __ p UVO X I -1-

    cK cos il f r cos il f V. ./ v. s ./ s (2.10)

    where is the refocus distance of equation 2.9. Effectively this lens

    produces an image of the fast-ion beam, collected over a larger accept

    ance angle than that of the spectrometer, at the entrance slit where the

    Doppler shifted wavelengths across the image are compensated for by the

    linear dispersion of the spectrometer. For the IE lenses used in this

    experiment (six in all), (p-f)/fg = 1 to within 0.5% and cos r/cos i =

    1.01 at 1000A and 1.05 at 6000A. Hence f - D̂ . The cylindrical lens

    is placed approximately a distance f (Berqkvist 1976) from the entrance

    slit. From equation 2.10 it is seen that a particular f is optimum only

    for a particular Xq. The selection of lenses to enable study of an

    extended spectral region, therefore, represents a compromise. These

    choices are best made by recasting equation 2.10 in a different form.

    We replaced the focal length of the cylindrical lens with f = D/2[n(X)-l]

    where D is the diameter of the lens and n is the wavelength dependent

    index of refraction of the lens material (Jenkins and White 1957).

    Expressing the beam velocity in terms of its energy E, we find

    In figures 7 and 8 are plotted X versus E for the IE lenses used in this

    experiment (the IE lens number along with its diameter is tabulated in

    table 4) for a 8l*Kr and a 28N2 beam. The lens material is fused quartz

    and its index of refraction has been tabulated for various wavelengths

    (Handbook of Chemistry and Physics 1971). By drawing a vertical line

    X •O • SCXFT

    (2.11)

  • 34

    6000

    5500

    5000

    4500

    o<

    *< 4000

    DC I— o z UJ 3500

    UI

    1 ̂3000

    2500

    931A

    2000

    1500 0 2 3 7 4 5 6

    ENERGY E (MeV)

    Figure 7. Focus wavelength X versus energy E of 8tfKr+ beam for IE lenses 2 through 8.

  • 35

    6000

    DK (n-l)

    5500 • 931A

    5000-•

    4500- -

    ©<

    ^ 4000- -

    X H* O z UJ _1

    3500- -

    UJ

    p 3000--

    2500- -

    2000 -

    1500

    0 2 7 3 5 4 6

    ENERGY E (MeV)

    Figure 8. Focus wavelength X versus energy E of 28N* calibration beam for IE lenses 2 through 8.

  • Table 4. Characteristics of the intensity enhancement lenses.

    a - Wavelength at which minimum linewidth occurs, b - Minimum linewidth observed. c - Useful wavelength range.

    IE 2 MeV &tKr 5 MeV 84 Kr Lens No. D(mm)

    ent. CM)

    exit CM) Aa(A)

    h ° AX°(A)

    C O 6 X (A)

    ent. CM)

    exit CM)

    Q O AaCA)

    b ° AX CA)

    c o 6X (A)

    2 1.37 ± 0.03 450 50 3010 1.3 2750-3950 350 (some 0 1000)

    50 2800 1.6 1550-2800

    3 2.03 ± 0.03 450 50 4150 1.7 3800-5350 450 50 2400 1.4 2200-3900

    4 2.45 ± 0.03 450 50 6000 1.7 5200-6000 450 50 3700 1.8 3400-4100

    5 3.00 ± 0.03 - - - - - 45051000 50 5300 2.2 4100-4925

    6 3.56 ± 0.03 - - - - - 1000 50 5100 2.2 5800-5600

    8 4.00 ± 0.07 - - - - - 1000 50 5810 2.5 5400-6000

    w ON

  • for the beam energies of interest through this family of curves, we

    selected the appropriate lens to use for a particular wavelength and

    wavelength range. It is important to realize, however, that this

    2 family of curves is sensitive to the value of [(DK/2) (931A/2)] and,

    therefore, to errors in D and K. For example, if the value of the

    above expression is changed by 10%, the resulting curves are shifted

    along the energy axis about 1/3 their respective spacings. The errors

    which arise from neglecting the factor

    rc£sr) fefl COS X I f ̂' S

    are easily incorporated into an error in D. To account for errors in

    the constants of equation 2.11 and for changes in the value of E due to

    energy loss in the foil we found it useful to plot Xq versus D for the

    values of E (figure 9) encountered in this experiment. Hence

    D = 8̂E/951A Xq . (2.13)

    The values of E chosen are for 2 MeV and 5 MeV 81tKr incident on a

    6 Mg/cm2 carbon foil. The energy loss in the foil for 1.67 MeV 28N2

    and 5 MeV 8lfKr+ are negligibly different so that their g's are the same

    and one curve describes both beams. We can now see (figure 9) what

    influence errors in our parameters, all lumped into D, have on our

    choice of optimum wavelength and wavelength interval when using IE

    lenses to produced our observed spectra. In table 4 we summarize the

    relevant IE lens parameters used in this spectral study. We determined

    the optimum slit width by examining with increasing slit widths the

    change in line width to peak intensity.for a particular spectral line.

  • 38

    6000

    5500

    5000

    4500

    o<

    X h O z Ui -J

    3500 •

    > $ 3000 •

    2500 •

    n= — /8E. U K V 93| a

    2000 •

    1500 0 2 3 4

    I E Lens Diameter 0 (mm)

    Figure 9, Focus wavelength, \ versus IE lens diameter D.

    Curve A is for 2 MeV K̂r"*" incident on a 6 ygm/cm2 carbon foil. Curve B is for 1.67 MeV anc* 5 MeV K̂r4" incident on a 6 ugm/cm2 carbon foil.

  • We selected that slit width which, if increased further, would have pro

    duced a larger line width with no significant improvement in intensity.

    Table 4 also lists the minimum observed line width in each wavelength

    interval accessible to the various IE lenses. Because of the increased

    viewing angle of the IE lens, the foil wheel was moved upstream to per

    mit an unocculted view of the beam.

    It was not an expressed objective of this spectral study to per

    form a detailed analysis of the intensity enhancement technique. Never

    theless, several problems were evident, suggesting future investigations.

    The derivation of the focal length relation (equation 2.10) is intui

    tively convincing, but no thorough treatment after the fashion of

    Leavitt, Robson, and Stoner (1973) of the beam-lens-spectrometer system

    has been accomplished. Bergkvist (1976) has treated the case of a beam

    with zero cross-section, but ignores the more realistic finite beam size

    case. In addition, there is no theory of how line width should vary with

    wavelength. The observed behavior is that the line width goes through

    a minimum near a particular wavelength for a given IE lens. This needs

    to be explained, as well as the apparent disagreement between the

    observed wavelength with minimum line width and the calculated wave

    length. A more reasonable agreement between these two wavelengths was

    obtained by decreasing the product DK by 7-8%, part of this decrease

    accountable if one includes the neglected geometric factors (equation

    2.12). Indeed, it is by no means obvious what effect spherical aberra

    tion, surface defects, and diffraction by the cylindrical lens will have

    upon the resultant fast-ion beam image. Finally, we do not know what

    influence variation of the downstream intensity due to decay of atomic

  • states will have upon the resultant spectral profiles,

    observed and computer-fitted line profiles of isolated

    with IE lenses were all symmetrical.

    although the

    lines recorded

    Recording and Reduction of Data

    In figure 10 we show the spectral region 400 - 6000& and the

    particular combinations of photodetectors and gratings chosen to record

    the 2-5 MeV 8£fKr+ and 1.67 MeV 28N* Calibration) beam-foil spectra.

    We oriented ourselves to the strongest lines available in the krypton

    and nitrogen calibration spectra by performing survey scans from 400 -

    6 6000A with the spectrometer refocused, 300 y slits, and at a rate of

    O 50A/min. The incident krypton beam energies were 2, 3.6, and 5 MeV.

    Using these scans as our guide, we made multiple scans of the strongest

    krypton spectral features, spectrometer refocused, using smaller slit

    widths and slower scan speeds. As stated earlier, almost all of these

    scans were recorded at 0.1 or 0.2 A/chan with 50 or 100 p slits. The

    number of multiple scans taken for each spectral feature was arbitrary

    and so chosen as to assure a signal to noise ratio of about 10:1. We

    recorded the entire region from 400 - 1600A using 100 p slits at 0.2

    X/chan. These scans insured that we obtained all the weaker spectral

    features in this region as well. The spectrometer was then adjusted

    for SSF and the IE lenses installed. Krypton spectra for a 5 MeV

    incident beam were then recorded from 1600 - 3000A at 0.2&/chan. No

    spectra were recorded in this region at 2 MeV incident krypton beam

    energy because of a lack of the appropriate IE lenses. From 3000- 6000A

    the entire krypton spectra for both 2 and 5 MeV incident beam energies

  • ©©-Short Scans (541 F)

    X

    -4219 EIC-

    IP28-541 F-

    -4-1000 2000

    •1500 A blaze-

    4- -4- X 3000 4000 5000 6000

    5000 & blaze-

    Photodetectors

    Gratings

    Figure 10. Detector and grating combinations used to record Kr and N2 spectral scans.

    Horizontal scale is wavelength range (X) • •f*.

  • 42

    were recorded at 0.2 A/chan using IE lenses. Strong, individual features Q

    were also multiply scanned at 0.1 A/chan. Selected, complex, spectral

    features were recorded at incident beam energies of 3 and 4 MeV to provide

    additional intensity information as an aid to unravelling their origins.

    Interspersed among these krypton scans were the 1.67 MeV Ng

    calibration scans taken with the spectrometer under the same conditions

    and typically preceding and following a particular sequence of krypton

    scans. With the spectrometer refocused, the nitrogen scans were of the

    O strongest observed lines and recorded at 0.05 A/chan with 50 y slits.

    O In the region 2900 - 6000A, the nitrogen spectra were recorded entirely

    using intensity enhancement at 0.2 A/chan with scans over selected

    strong lines at 0.1 A/chan.

    The record of each multiple scan consisted of intensity versus

    multiscaler channel number, starting point of the spectrometer, scan

    scale in Angstroms per channel, number of repetitions of each scan and

    observed beam current. For scans of such duration that foil changes

    were necessary during the course of a scan, the wavelengths at which the

    changes were made were noted. Temperatures top and bottom of the spec

    trometer were always recorded.

    The results of the krypton and nitrogen scans were printed out

    on strip charts and the digital information on paper tape. The charts

    became an indispensible visual record, greatly aiding the computer

    fitting of all the spectral data. In figures 11 and 12 are sample scans

    of parts of the krypton spectra.

    During the data reduction, these charts were displayed en masse

    on 4' x 8' celotex sheets to permit rapid visual comparison between scans

  • CSA KrSC CSA Kr EL

    3380 3390 3400 3410 3420

    WAVELENGTH (X)

    3430

    2 MeV KRYPTON ON C FOIL (~6fiq /cm4)

    entrance slit > 450fi, exil slit 50u

    2.2/xAMP beam, 4 scans

    Intensity enhancement lens used

    Kr 21 ?

    - - 1.40 A w 100

    3440 3450

    Figure 11. Intensity enhanced spectra of 2 MeV krypton in the region 3375 - 3455X.

    Seen here in second order. 4̂ C/l

  • Kr3Zm (8-7) (2—order)

    400

    5 MeV KRYPTON ON C FOIL (~6/xg/cm2)

    entrance slit: lOOO/i, exit slit: 50fj.

    3.5ft AMP beam , 5 scans

    Intensity enhancement lens used KrX (12—10)(2—order)

    a a> u> Q 300 to s Kr 3ZHT (8-7) (2— order)

    - 1.67 A to 200 z UJ

    100

    5970 5960 5940

    WAVELENGTH (A)

    5920 5930 5950

    O Figure 12. Intensity enhanced spectra of 5 MeV krypton in the region 2959 - 2985A.

    Seen here in second order.

  • over the same spectral region at different energies and other spectral

    regions at the same energy. These displays become in large scale what

    photographic plates are to the classical spectroscopist. The digital

    results were transferred manually to IBM cards for later computer analy

    sis. For quick orientation in locating a particular scan, a "road-map",

    divided according to grating used and incident beam energy, was drawn

    O showing the location of each scan in the region 400 - 6000A.

    All the krypton and nitrogen spectral scans were analyzed either

    by hand or computer fitting of the intensity profiles. If, by visual

    inspection, the signal to noise ratio of weak spectral features was so

    poor that computer fitting was of doubtful advantage, a profile was hand

    drawn over the spectral feature and the peak center determined from the

    profile center at full-width-half-maximum CFWHM). The channel number,

    determined in this way, was decreased by one to strike agreement with

    the channel numbers assigned from the computer fitting program.

    We computer fitted the spectral data with program GAUS2Z (see

    Appendix A). To accomplish this task it was essential that we first

    determine the line width as a function of wavelength for the various

    conditions of the spectrometer under which our data were recorded. To

    this end we fitted selected strong single lines with GAUS2Z allowing the

    program to determine the best line width. The resulting line widths

    versus wavelength were then plotted and a smooth curve was drawn through

    the data. These curves established the line width used as an input

    parameter for fitting the remaining spectral features.

    The results of each computer fitting provided the x2 of the fit,

    peak location (in channel number), uncertainty in the peak location,

  • peak height and its uncertainty, FWHM and its uncertainty, and the

    calculated area under each peak. When the x2 and visual appearance of

    the fit were deemed satisfactory, we stopped the fitting process and

    converted the peak channel number to observed wavelength. In this

    straightforward, albeit tedious, manner the observed wavelengths of

    approximately 2500 krypton and 750 nitrogen lines were determined.

    These wavelengths, to facilitate further treatment, were punched onto

    IBM cards, each card containing a wavelength, its uncertainty, intensity,

    scan identification, and brief comments about the fit.

    From the observed nitrogen calibration wavelengths we proceeded

    to construct calibration correction graphs. As all data had been taken

    in vacuum, we decided to calibrate all our results in terms of vacuum

    O values, including observed wavelengths above 2000A. This involved con-

    verting the tabulated nitrogen calibration wavelengths used above 2000A

    to vacuum values. This was accomplished directly by programming the

    NBS Wavenumber Tables (Coleman, Bozman, and Meggers 1960) into the

    HP-65 calculator (Cardon 1975). The lines in the observed nitrogen

    spectra were identified from Striganov and Sventitskii (1968). Because

    2%2 has a q/m ratio of 28, C0+, Si+, Fe++, A1H+, and MgH* were poten

    tial contaminants in the incident beam and wavelengths from these ele

    ments were considered possible intruders in the nitrogen calibration

    scans. However, in the final analysis, the 652 lines used to establish

    the calibration graphs were mostly nitrogen with the remainder being

    oxygen and carbon lines. The values of AX = X „„ - X , were corr vac HB obs

    plotted versus We then least-squares fitted a straight line to

    the plotted data obtaining the best values for slope and intercept.

  • 47

    In this way we established calibration for the 5 MeV 8lfKr+ data. The

    beam velocity for 2, 3, and 4 MeV 8tfKr+ was different, however, from the

    1.67 MeV Nj> data. For the IE data taken at 2 and 5 MeV, we compared

    prominent lines at both energies and established the calibration of the

    2 MeV results based on the 5 MeV results. For the refocused data taken

    o below 1600A, there was no noticeable shift of the 2 MeV spectra with

    respect to the 5 MeV spectra. The same calibration was used for both

    energies. For the data taken with the spectrometer refocused and the

    5000A blaze grating, a systematic correction to the 5 MeV calibration

    was applied to the 2, 3, and 4 MeV 8I*Kr+ results. This correction was

    necessary to account for the relativistic transverse Doppler shift (Jack

    son 1962, p. 364). This red shift AX of the rest frame wavelength Xq is

    X 2 ax = ~t $ (-2a4)

    for the velocities v encountered in this experiment. The correction

    applied to the wavelengths obtained at 2, 3, and 4 MeV, after determining

    their vacuum wavelengths with reference to the 1.67 MeV calibration,

    was X (5-E)

    AX = 78204- C2-15)

    o where E is the beam energy in MeV. At 5000A, for example, the correc

    tion to the 2 MeV data accountable to the transverse Doppler effect is

    0.19&. We summarize the parameters used to establish the calibration

    of the spectrometer for the diverse conditions encountered in this

    experiment in table 5. A program LAMSORT (see Appendix B) was written

    that read the cards containing the observed krypton wavelengths and

  • 48

    Table 5. Spectrometer calibration parameters for the McPherson 225.

    AX — AX , + B ,. X corr obs vac

    X , + AX . obs corr

    Energy Grating Arrangement (MeV) A B

    1500A Refocus 2,5 0.000038 12.11A

    ISOOK IE lens 2 5 0.00084 11.70&

    1500& IE lens 3 5 0.00084 12.00A

    5000& Refocus (before Sept. 18, 1973)

    2,3,4,5 -0.000109 -3.56A

    5000X Refocus (after Sept. 18, 1973)

    2,3,4,5 -0.000109 2.49A

    5000A IE Lens 2 2 0.000307 4.15&

    5000A IE lens 3 4,5 0.000186 2.2lX

    5000A IE lens 3 2,3 0.00000946 2.40A

    5000& IE lens 4 2 -0.000223 5.17&

    5000A IE. lens 4 5 0.00017 3.79X

    5000A IE lens 5 5 -0.000075 3.05A

    5000X IE lens 6 5 -0.000173 3.59A

    5000A IE lens 8 5 -0.000727 4.62A

  • converted them to calibrated vacuum wavelengths using the parameters A

    and B in table 5. The result of this reduction of the data taken with

    the thirteen different calibrations of the spectrometer, and accounting

    for the transverse Doppler effect, was a listing of 2392 calibrated wave

    length measurements. From this list, multiple measurements of the

    same spectral line were reduced to unweighted averages. For isolated

    single lines this was straightforward. In the case of line complexes

    and blends it was often difficult to decide which measurements belonged

    together, as the spread in the observed wavelengths for a particular

    line often overlapped the wavelengths of nearby lines. For each of

    these cases it was necessary to study the individual scans and make

    judicious guesses. With the averaging process, the line list was re

    duced to 1187 lines. A search was next made by hand, wavelength by

    wavelength, for second and,possibly, third order lines. The wavelength

    o response of the photodetectors and the lower wavelength O 1600A)

    transmission cutoff of the IE lenses greatly facilitated isolating

    those regions where higher orders were likely interlopers. The possi-

    o bility of second order lines appearing in the region 400 - 800A from the

    o first order lines in the region 200 - 400A was carefully examined although

    acknowledged unlikely because of the response of the grating coating in

    this spectral region (Samson 1967) and our extreme displacement from the

    grating blaze. Data were obtained for 5 MeV krypton in the region 100 -

    400A with a grazing incidence VUV spectrometer (Mclntyre and Bernstein

    1976) which showed no strong krypton features that would likely pass to

    the second order with the McPherson 225. We also checked, to no avail,

    for second order wavelengths of transitions seen in a krypton seeded theta

  • o pinch light source observed between 209 - 319A (Englemann, Thomson, and

    Monaghan 1976). We concluded that the 1500X blaze grating in the

    McPherson 225 is not responsive to wavelengths below approximately 400X.

    When a second or third order line was identified (there were

    o only 12 third order lines with the 1500A blaze grating and two third

    o order lines with the 5000A blaze grating), it was averaged with the

    first order results, the wavelengths weighted according to their order

    (1 for first order, 2 for second order, etc.)- If a line could not be

    clearly identified as to its order it was assumed to be a first order

    line. A histogram was plotted of the difference between the first order

    wavelengths and the average wavelength for all orders, where possible.

    The histogram was symmetric about and peaked at the zero wavelength

    difference, indicating no systematic errors in calibration. Confidence

    in the spectrometer was further gleaned from the observed small scatter

    among the multiple measurements of particular wavelengths taken with the

    spectrometer under the diverse calibration conditions. Indeed, it was

    from the standard deviation

    I ZX* - nX2

    ' = V • * C2'16)

    where is a particular measurement, X the wavelength average, and n

    the number of measurements (n > 3) that the wavelength uncertainty for

    three or more measurements was established. The were rounded off

    o to the nearest 0.05A and became the accepted uncertainties. If there

    was only one measurement for a particular wavelength, the assigned un

    certainty was two channels for the scan rate at which the measurement

    was taken and for two measurements it was the deviation of the values

  • 51

    o from the average rounded to the nearest 0.05A. Furthermore, any

    uncertainties greater than 0.6& resulted in the wavelength being

    deleted from the final list. We plotted a histogram of the uncertain

    ties in our final wavelength list (see Appendix C). The plot was

    o symmetric about the peak uncertainty of ± 0.35A; the observed uncer

    tainties extending from ± 0.1-0.6A. Although our observed wavelengths

    are uncertain to order O.lA, they are tabulated to the nearest 0.01A

    so as to preserve and display behavior in this least significant digit.

    o We compared our wavelength list below 1200A with results tabu

    lated in Kelly and Palumbo (1973). These wavelengths are so identified

    in Appendix C. The agreement between our results and those previously

    tabulated and assembled in Kelly and Palumbo is often well within the

    uncertainties assigned our observations. Thus, we further supported

    the validity of our calibration and method of establishing uncertainties.

    No attempt was made to determine the spectral, response of the

    o spectrometer between 400 - 6000A. The intensity or intensities listed

    for each wavelength in Appendix C must therefore be acknowledged as

    approximate count rates for a spectrometer under similar conditions.

    Because the spectral lines were recorded under greatly differing condi

    tions, leading to variations in recorded intensity, all the intensity

    values, realizing that the spectral response was being ignored, were

    normalized to a common beam current, slit settings, and multiscaler

    dwell time. Intensities recorded with the spectrometer refocused were

    all normalized in units

  • 52

    v-umiua + ~ 50 y slits • 0.6 sec dwell time • 10 yamp beam current

    Similarly, intensities recorded using intensity enhancement were norma

    lized in units

    counts ,2 450 y ent - 50 y exit slits • 0.6 dwell time • 10 yamp beam current

    Because the intensities recorded with intensity enhancement are depen

    dent upon transition lifetimes and the consequent post-foil light

    distribution whereas the refocus intensities are taken near the foil and

    from a well-defined portion of the beam, the refocus intensities were

    considered the more reliable. Where both refocus and IE intensity

    values existed for the same spectral line, the ratio of IE intensity

    to refocus intensity was calculated. These ratios were plotted in a

    histogram from which a conversion factor, refocus to IE intensity, was

    determined. This factor, an average of the results for 2 MeV and 5 MeV

    krypton, was 5.8. We thereby reduced all intensities to refocus values

    and expressed the results in the units of equation 2.17. Note that

    this unit can also be written as [counts/C50yslits • 6 yC)] although

    the dwell time is now concealed.

    The krypton spectra line list (Appendix C) extends from 417.44-

    o 5595.96A and consists of 965 lines. Of these, 110 have been previously

    observed and/or identified by other authors. We discuss in Chapter 5

    these previous observations in addition to new identifications which

    follow as a result of this investiation. In figure 13 we plot a histo

    gram of the number of our observed krypton wavelengths in this study.

  • 53

    50-

    CO

    f 40-c _a> a

    -o a> > a> in £X