the effect of the size of the apical foramen and coronal flaring on the accuracy of three electronic...
DESCRIPTION
electronic tooth length measuring devicesTRANSCRIPT
THE EFFECT OF THE SIZE OF THE APICAL
FORAMEN AND CORONAL FLARING ON THE
ACCURACY OF THREE ELECTRONIC TOOTH
LENGTH MEASURING DEVICES.
(AN INVITRO STUDY)
Hossam M. Tewfik. Assoc. Prof. Of Endodontics. Ain shams University. Ahmed A. Hashem. Lecturer of Endodontics. Ain Shams University.
Abstract
The purpose of the present study was to evaluate the effect of the size of
the apical foramen and the coronal flaring of the root canals on the
accuracy of three apex locators. The actual lengths of forty mesio-buccal
canals were measured under surgical microscope. Canals were then
divided into two equal groups each representing size of the apical
foramen (equivalent to #15 and #35 K-files). Each group was subdivided
into two equal subgroups according to the degree of canal coronal flaring.
The electronic tooth measurements were done using three different apex
locators (Tri auto ZX, Neo Sono and AFA apex finder). The results
showed no difference in accuracy among the three tested devices. Canals
with open apices decreased the accuracy of the electronic measurements
as compared to canals with closed apices. On the other hand, coronal
flaring of the root canals increased the accuracy of the three tested
electronic devices.
Introduction
Cleaning and shaping of the root canal system cannot be initiated
without accurate determination of tooth length. Several methods have
been used for proper establishment of working length. Perhaps the Ingle’s
radiographic technique (1) received the greatest popularity being the most
widely used method for length determination. Limitations and
shortcomings of the radiographic technique led to the introduction of the
electronic devices for tooth length measurement. Electronic devices for
tooth length measurement were first introduced by Sunada(2) and since
then, different devices based on several electrical terms were presented
and tested.
O’Neill et al(3) tested the accuracy of the Sono explorer apex
locator invivo as compared to the actual tooth length after extraction. The
results showed that both measurements were equal in 83% of the times
while in the other 17% the electronic measurement was shorter than the
actual length. This high accuracy in the electronic measurement was not
attained in the study by Seidberg et al(4) where the Sono Explorer was
accurate in only 48% of the times. The authors thus suggested that this
device is not a reliable tool for locating the working length.
Fouad et al(5) evaluated five brands of electronic tooth length
measuring devices in invivo conditions and compared them to the
radiographic estimates of canal length. The accuracy of the electronic
devices ranged from 55-75% with no statistical difference between
different brands.
Keller et al(6) clinically compared the Endocater apex locator with
the traditional radiographic length determination performed by an
experienced endodontist. Results showed that the evaluator was 95.8%
accurate in positioning the file within +/- 1 mm from the cemento-
dentinal junction. On the other hand, the Endocater was only 67.7%
accurate. Close results were recorded by Wu et al(7) when testing the Sono
Explorer Type III. The actual root canal lengths were measured after
tooth extraction and compared to the electronic measurement. The rate of
agreement of the two measurements was 77.5% within a range of +/-
0.5mm.
Again, Hembrough et al(8) evaluating the Sono explorer Mark III
presented similar data where they found the device to be accurate in
73.1% as compared to 88.5% achieved by the radiographic method.
Shabahang et al(9) evaluated the newly introduced Root ZX apex locator
in locating the apical foramen in unprepared root canals of vital teeth and
found it to be efficient in locating the apical foramen within the range of
+/- 0.5 mm with an accuracy of 96.2%.
Different root canal conditions have direct effect on the accuracy
of different root canal electronic measurement devices. These conditions
include the type of tissue inside the canal (vital or non vital), type of
fluids inside the canal, size of apical foramen and degree of canal
enlargement.
Fouad and Krell(10)
compared five different electronic devices
under different root canal conditions. The tested variables were the
presence of different fluids, presence of conductive gutta percha and the
size of the root canal. The results showed that no difference was seen
among different devices while most fluids caused short measurements
with the greatest effect with blood and sodium hypochlorite. Further
more, canal enlargement increased the accuracy of all electronic devices.
Stein et al(11)
studied whether width of the apical foramen
influenced the accuracy of the Neosono-D apex locator and found that the
increase in width of the foramen altered the accuracy of the device.
The efficiency of different brands of apex locators invivo in both
vital and necrotic cases were tested by Mayeda et al(12)
and showed no
statistical difference in measurements between vital and necrotic cases.
Similar results were reported by Dunlap et al(13)
where the measurements
recorded in vital and necrotic cases by the Root ZX apex locator were
82.3% accurate to within 0.5 mm from apical constriction.
The effect of root canal instrumentation in recapitulated and non-
recapitulated canals on the accuracy of apex locators was examined by
Rivera and Seraji(14)
. They found that recapitulation of non-patent root
canals was necessary to obtain accurate length determination when using
apex locators.
Fouad et al(15)
tested four apex locators under different root canal
fluids with the variation of apical foramen size. The results showed that
no significant difference was noted among devices in dry canals
regardless of the foramen size. Only the Endex apex locator was superior
to the other devices in the moist canals. Similarly Frank and
Torabinejad(16)
reported an accuracy of 89.64% when Endex apex locator
was used in moist root canals.
In 1994, Kobayashi and Suda(17)
developed a new method for
electronic canal measurement to overcome the effect of presence of
electrolytes in the canal. The method depend on measuring two
impedances of the canal using current sources with two different
frequencies. The accuracy in electronic tooth measurement devices was
shown not to be affected by the condition of the root canal either dry or
wet regardless the type of device used(18)
. Recently, Jenkins et al(19)
and
Kaufman et al(20)
in two different studies evaluated the efficiency of Root
ZX with different root canal irrigants and showed high reliability of the
device.
Stabholz et al(21)
showed in a clinical study that the ability to
determine the apical constriction by tactile sensation was significantly
increased when the root canals were preflared. Should this alter the
accuracy of electronic devices for tooth measurement ? A question which
was addressed by several studies.
Nguyen et al(22)
evaluated the root ZX as a reliable method for
tooth length measurement. Furthermore, they tested the accuracy in
measurement when using small or large files in enlarged canals. Results
showed that the file size is not a contributing factor in the accuracy of
measurement.
Moor et al(23)
compared four apex locators in unflared dry, flared
dry or flared wet root canals and showed that early coronal flaring did
not ensure better or more precise electronic tooth length measurement.
However, opposite to the later study, Ibarrola et al(24)
showed that root
canal preflaring increased the efficacy of the Root ZX apex locator
suggesting that the preflaring allow the working length files to reach the
apical foramen more consistently thus improve their accuracy.
The effect of some of the previously reviewed canal conditions on
the accuracy of apex locators needs further testing. Thus, it was the
purpose of the present study to evaluate the effect of the size of the apical
foramen and coronal flaring of the root canal on the accuracy of three
apex locators.
Materials and methods
1- Selection and classification of samples
Forty human freshly extracted mandibular molars were selected. Teeth
were intact with no carious lesions or restorations. Absence of internal
calcifications was verified radiographically. Any molar with external
resorption, fracture or extensive curvature was discarded.
Samples were classified into two equal groups according to the size of
the apical foramen. Samples of group I had an apical foramen which
permitted size 15 K-file to project 1-2 mm beyond the apical foramen
while samples of group II had larger apical foramen which permitted the
passage of size 35 K-file 1-2 mm beyond the apical foramen.
Each group was subdivided into two equal subgroups where samples
of subgroup A (groups I and II) had their lengths measured without any
manipulation or enlargement of their root canals. Lengths of samples of
subgroup B (groups I and II) were measured after enlargement of the
coronal 2/3 of their root canals.
2- Preparation of samples
All molars were scaled and curetted to remove external debris and
calculus deposits. The occlusal surface of all samples was slightly
grounded to produce a flat surface for better adjustment of the rubber
stopper during length measurement. An endodontic coronal access
preparation was done for all samples using carbide round burs and
tapered diamond stones mounted on high speed contra angle with water
as a coolant. Pulp chambers were thoroughly irrigated with sodium
hypochlorite 2.6%.
All length measurements for all molars were done for the mesio-
buccal canal only. The forty mesio-buccal canals were checked for root
canal patency using # 15 K-files. Using a tapered diamond stone with
water coolant, one millimeter root-end resection was done for the mesial
root of samples of group II. This horizontal cut resulted in larger apical
foramen which permitted size 35 K-file to pass beyond the apical
foramen. All canals were irrigated with sodium hypochlorite 2.6%.
Samples of subgroup A (groups I and II) were stored with no further
manipulation until length measurements. Samples of subgroup B (groups
I and II) were subjected to flaring of the coronal 2/3 of the canal. Flaring
was done using gates glidden drills # 4, 3, 2 mounted on slow speed
contra angle in a crown-down fashion. Root canals were thoroughly
irrigated and patency was rechecked.
3- Actual tooth length Measurement:
All forty samples were subjected to tooth length measurement by
introducing a # 15 K file (group I ) or # 35 Kfile (group II) until it
appeared just projecting from the apical foramen and flushing with the
root end. This position was verified under the surgical microscope at X8
magnification. The rubber stopper on each file was adjusted on the
flattened cusp, file was withdrawn from the canal and length was
measured to the nearest 0.25mm.
4- Invitro model for electronic tooth measurement:
The in vitro model used was that introduced by Aurelio et al(25)
and
modified by Nahmias et al(26)
. The model consists of 12x72 mm
polystyrene culture tube containing agar in phosphate buffered saline.
The crown of each tooth was tightly fitted in the tube cap such that the
roots would be totally emersed in the medium. A ¾ inch nail extending
from the medium through the sides of the tube at a 3 cm distance from the
base of the tube acted as a negative electrode.
5- Electronic tooth length measurements:
Three different electronic tooth length measuring devices (Fig. 1) were
evaluated which were: the Tri auto ZX (J Morita, Japan), Neo Sono
(Satellic, France) and the Apex Finder (AFA)(Analytic Technology,
USA). The manufacturer instructions for operation of each device were
accurately followed. The file holder for each device was mounted with K
file either #15 (group I) or #35 (group II) while the lip clip for each
device was hanged on the metal nail at the base of the culture tube. Each
file was slowly introduced inside the root canal until the apical foramen
was reached as indicated by each device (Fig. 2). The instrument stopper
was adjusted on the flattened cusp, file was withdrawn from the canal,
measurements were verified three times and recorded to the nearest 0.25
mm.
6- Data collection
The mean difference between the electronic measurement and the
actual tooth length measurement was used for data evaluation. A positive
value indicated long measurement while negative value indicated short
measurement. Measurements that fell within +/- 0.5 mm were determined
as acceptable. Data were subjected to multiple range analysis of variance
(ANOVA) and student t-test at p<0.05.
Results
The difference in length measurement by the three tested devices
among the two groups and subgroups as compared to the actual tooth
length is presented in tables 1-4.
The total number of measurements (120 measurements)
representing the readings of the forty samples measured by each of the
three electronic devices is presented in table 5. This overall data showed
that the electronic measurement was exactly the same as the actual tooth
length (zero value) in 44 measurements (36.6%). A difference of +/- 0.5
mm or less in the electronic measurement as compared to the actual tooth
length was seen in 62 measurements (51.6 %). A difference more than +/-
0.5 mm was recorded in only 14 measurements (11.6%). Considering the
data that fell within +/- 0.5 mm to be clinically acceptable showed that
the overall accuracy of the electronic measurements regardless of the
device used was 88.2%.
As regard the accuracy of the three tested devices, no significant
difference was recorded among the data recorded at p<0.05. However, the
readings of the Tri auto ZX and the Neo Sono were identical and showed
higher degree of accuracy reaching 90% whereas, the AFA recorded
lower degree of accuracy (85%).
The degree of accuracy of electronic length measurement among
the two groups with both subgroups is presented in table 6. The highest
degree of accuracy was attained when the root canals were flared before
measurement. This was seen in samples with closed apices (group I
subgroup B) as well as in samples with open apices (group II subgroup
B). This high degree of accuracy reached 100% when adding the
measurements with zero value and the measurements that fell within the
+/-0.5 mm. This was followed by the samples that had closed apices and
their coronal 2/3 was not flared (group I subgroup A). These samples had
their electronic measurement exact as the actual measurement in 33.3 %
of the times. While in 60 % of the times the electronic measurement was
considered clinically acceptable. The lowest degree of accuracy was seen
in samples which had open apices and were not coronaly flared (group II
subgroup A). These samples had the highest percent of unacceptable
measurements (40%). While the acceptable measurements were 59.9%.
The data recorded for such subgroup (Group II subgroup A) was
statistically significant than the other subgroups at p<0.05.
The effect of the size of the apical foramen on the accuracy of the
electronic length measurement is presented in table 7. Adding the zero
value to clinically acceptable values (+/- 0.5 mm) showed that the degree
of accuracy of the three tested devices reached 96.65% when the apical
foramen was not enlarged. On the other hand, canals with large apical
foramina decreased the accuracy of electronic measurement reaching a
value of 79.95%, however, this decrease was not statistically significant.
The effect of coronal flaring on the accuracy of electronic length
measurement is presented in table 8. Again, adding the zero value to the
+/- 0.5 mm values showed that coronal flaring resulted in a 100%
accuracy of the three tested devices. On the other hand, unflared canals
showed significant decrease in the accuracy of electronic length
measurement reaching 76.6%.
Table 1: The difference in length measurements of samples with closed apices and
no coronal flaring (Group I Subgroup A)
Sample Tri ZX Neo Sono AFA
1 0 0 - 0.25
2 - 0.5 - 0.5 - 0.25
3 - 0.5 - 0.5 - 1.0
4 0 0 0
5 - 0.25 - 0.25 + 0.5
6 0 - 0.5 0
7 - 0.5 0 + 0.5
8 - 0.25 0 - 0.25
9 -0.5 - 0.25 - 1.0
10 0 - 0.5 - 0.25
Mean - 0.25 - 0.25 - 0.2
Table 2: The difference in length measurements of samples with closed apices and
coronal flaring (Group I Subgroup B)
Sample Tri ZX Neo Sono AFA
1 + 0.5 + 0.5 0
2 + 0.5 0 0
3 - 0.5 0 - 0.5
4 0 - 0.5 - 0.5
5 - 0.5 - 0.5 0
6 0 - 0.5 0
7 - 0.5 0 0
8 - 0.5 0 - 0.5
9 + 0.5 - 0.5 0
10 + 0.5 + 0.5 - 0.5
Mean 0 - 0.1 - 0.2
Table 3: The difference in length measurements of samples with open apices and
no coronal flaring (Group II Subgroup A)
Sample Tri ZX Neo Sono AFA
1 + 1.0 + 1.0 + 1.0
2 0 0 + 0.25
3 - 1.0 - 1.0 0
4 0 0 + 0.75
5 + 0.25 + 0.25 + 0.25
6 0 + 0.25 0
7 + 0.25 0 + 0.25
8 0 0 + 0.25
9 + 1.0 + 1.0 + 1.0
10 - 1.0 - 1.0 + 0.75
Mean + 0.05 + 0.05 + 0.45
Table 4: The difference in length measurements of samples with open apices and
coronal flaring (Group II Subgroup B)
Sample Tri ZX Neo Sono AFA
1 - 0.5 0 - 0.5
2 0 + 0.25 0
3 0 0 + 0.5
4 0 - 0.5 - 0.25
5 + 0.25 + 0.5 + 0.5
6 0 - 0.5 0
7 - 0.5 0 + 0.5
8 + 0.25 0 - 0.25
9 0 + 0.5 - 0.5
10 0 + 0.25 + 0.5
Mean - 0.05 + 0.05 + 0.05
Table 5: The total number of measurements for all samples with the representative
percentages in between brackets.
Tri ZX
(n=40)
Neo Sono
(n=40)
AFA
(n=40)
Total
(n=120)
0 16 (40%) 16 (40%) 12 (30%) 44 (36.6%)
+/- 0.5 mm 20 (50%) 20 (50%) 22 (55%) 62 (51.6%)
> 0.5 mm 4 (10%) 4 (10%) 6 (15%) 14 (11.6%)
Table 6: Comparison among groups with representative percentages in between
brackets.
Group Group I /A
(n=30)
Group I /B
(n=30)
Group II /A
(n=30)
Group II /B
(n=30)
0 10 (33.3%) 12 (40%) 10 (33.3%) 12 (40%)
+/- 0.5 mm 18 (60%) 18 (60%) 8 (26.6%) 18 (60%)
> 0.5 mm 2 (6.6%) 0 12 (40%) 0
Table 7: Effect of the size of apical foramen on the accuracy of measurement
represented as percentages.
Closed apex Open apex
0 36.65% 36.65%
+/- 0.5 mm 60% 43.3%
> 0.5 mm 3.3% 20%
Table 8: Effect of the coronal flaring on the accuracy of measurement represented
as percentages.
Unflared canals Flared canals
0 33.3% 40%
+/- 0.5 mm 43.3% 60%
> 0.5 mm 23.3% 0%
Discussion
The accuracy of electronic apex locators have been
evaluated by many studies with conflicting results ranging from 48% up
to 96.2%. In the present study the overall data showed an accuracy of
88.2% with no difference between the three tested devices. This was not
much different than most of the recent studies where the new generations
of apex locators proved to be equally accurate(10,15,18,26).
The purpose of the present investigation was to evaluate the effect
of certain canal conditions on the accuracy of three different electronic
devices for tooth length measurements. Since the introduction of the first
generation of apex locators by Sunada(2), it have been always recognized
that wet contaminants in the root canal affects the accuracy of measuring
devices. Therefore, all recently introduced newer generations of apex
locators were designed to operate very accurately in different root canal
fluids. The three tested devices in the present study are recommended to
be used in wet environment so sodium hypochlorite was selected as the
root canal irrigant in all measurements to mimic clinical conditions.
Evaluating the accuracy of electronic devices invivo presents some
difficulties in testing certain conditions. Therefore, the use of an invitro
model is considered a needed alternative. Different invitro models for
examining the accuracy of electronic devices for tooth length
measurement have been suggested and tested(7,17,18,27)
. In the present
study, the agar model was used which was developed by Aurelio et al(25)
,
modified by Nahmias et al(26)
and tested for its validity by several
studies(10,15,26)
. This model provides the advantage of being easy to
assemble, highly accurate and its data are reproducible.
All electronic measurements recorded in the present investigation
were compared to the actual tooth length and categorized into three
levels: Zero value (representing the actual tooth length), +/- 0.5 mm and
more than +/- 0.5 mm. Measurements falling within the range of zero to
+/- 0.5 mm were considered clinically acceptable. This assumption was
followed by most of the studies testing these electronic devices based on
the fact that deviation of the apical foramen from the exact tooth apex is
considered a normality(5,8,9,28)
.
Two variables were evaluated in the present study the first was the
effect of the size of the apical foramen on the accuracy of the electronic
measurement. Wide apical foramen can be seen in certain stages of apical
development, presence of periapical lesions or in cases of iatrogenic over
instrumentation. Should this affect the accuracy of apex locators?
Huang(29)
noted that the larger the size of the apical foramen, the less
accurate was the measuring device. Other studies(11,30,31)
also showed that
the electronic devices are inaccurate in presence of wide or immature
apical foramen. Wu et al(7) reported a direct relation ship between the
size of the apical foramen and the accuracy of electronic apex locators.
The results of the present study were not much different than those
previously reviewed studies where a decrease in the degree of accuracy of
the electronic measurements was noted among teeth with open apices as
compared to those with closed apices (table 7). However, under the
conditions of the present investigation this difference was not significant.
Perhaps the only disagreement was reported by Fouad et al(15)
where they
showed in their study superiority of the measurements done by the Endex
apex locator when the apical foramen was widened.
The second variable included in the present study was to evaluate
the effect of coronal preflaring on the accuracy of the electronic length
measurement. Our results showed that coronal flaring did significantly
increased the accuracy of the tested devices reaching 100 % when adding
the zero values to the +/- 0.5 mm values. This was similar to the study by
Ibarrola et al(24)
who found that preflaring increase the accuracy of apex
locators. It is suggested that coronal flaring aids in removing any
interference caused by cervical dentin which provides unobstructed
access to the apical foramen(21)
. Opposite to these results, Moor et al(23)
suggested that early coronal flaring did not ensure better root length
measurement.
Although the invitro model utilized in the present study proved to
be efficient, and enabled us to examine variables that are not practical to
be tested clinically yet, the obtained results should not be directly
extrapolated to clinical situation. Future invivo investigations including
such tested variables are strongly recommended.
References
1- Ingle JI, Taintor J: Endodontics 3 rd. ed. Lea and Fabiger.
Philadelphia. 1985.
2- Sunada I: New method for measuring the length of the root canal.
J. Dent. Res. 2:375-387, 1962.
3- O’ Neil LJ: A clinical evaluation of electronic root canal
measurement. Oral Surg. 38: 479-483, 1974.
4- Seidberg BH, Alibrandi BV, Fine H, Logue B: Clinical
investigation of measuring working lengths of root canals with an
electronic device and with digital tactile sense. JADA. 90:379-
387, 1975.
5- Fouad AF, Krell KV, McKendry DJ, Koorbusch GF, Olson RS: A
clinical evaluation of five electronic root canal length measuring
instruments. J Endodon 16:446-449, 1990.
6- Keller ME, Brown CE, Newton CW: A clinical evaluation of the
endocater. An electronic apex locator. J Endodon 17:271-274,
1991.
7- Wu YN, Shi JN, Haung LZ, Xu YY: Variables affecting electronic
root canal measurement. Int Endo J. 25:88-92, 1992.
8- Hembrough JH, Weine FS, Pisano JV: Accuracy of an electronic
apex locator: A clinical evaluation in maxillary molars. J Endodon
19:242-246, 1993.
9- Shabahang S, Goon WY, Gluskin AH: An invivo evaluation of
root ZX electronic apex locator. J Endodon 22:616-618, 1996.
10- Fouad AF, Krell KV: An invitro comparison of five root canal
length measuring instruments. J Endodon 15:573-577, 1989.
11- Stein TJ, Corcoran JF, Zillich RM: The influence of the major and
the minor foramen diameters on apical electronic probe
measurements. J Endodon 16:520-522, 1990.
12- Mayeda DL, Simon JH, Aimar DF, Finley k: In vivo measurement
accuracy in vital and necrotic canals with the endex apex locator. J
Endodon 19:545-548, 1993.
13- Dunlap CA, Remeikis NA, Degole EA, Rauschenberger CR: An
invivo evaluation of an electronic apex locator that uses the ratio
method in vital and necrotic canals. J Endodon 24:48-50, 1998.
14- Rivera EM, Seraji MK: Effect of recapitulation on the accuracy of
electronically determined canal length. Oral Surg. 76:225-230,
1993.
15- Fouad AF, Rivera EM, Krell KV: Accuracy of the Endex with
variations in canal irrigants and foramen size. J Endodon 19:63-
67, 1993.
16- Frank Al, Torabinejad M: An invivo evaluation of Endex apex
locator. J Endodon 19:177-179, 1993.
17- Kobayashi C, Suda H: New electronic canal measuring device
based on the ratio method. J Endodon 20:111-114, 1994.
18- Czerw RJ, Fulkerson MS, Donnelly JC, Walmann JO: In vitro
evaluation of the accuracy of several electronic apex locators. J
Endodon 21:572-575, 1995.
19- Jenkins JA, Walker WA, Schindler WG, Flores CM: An invitro
evaluation of the accuracy of the root ZX in the presence of
various irrigants. J Endodon 27:209-211, 2001.
20- Kaufman AY, Kelia S, Yoshpe M: Accuracy of a new apex
locator: An invitro study. Int. Endo. J. 35:186-192, 2002.
21- Stabholz A, Rotstein I, Torabinejad M: Effect of preflaring on
tactile detection of the apical constriction. J Endodon 21:92-94,
1995.
22- Nguyen HQ, Kaufman AY, Komorowski RC, Friedman S:
Electronic length measurement using small and large files in
enlarged canals. Int Endo J. 29:359-364, 1996.
23- Moor RJ, Hommez GM, Martens LC, De Boever: Accuracy of
four electronic apex locators. An invitro evaluation. Endod. Dent.
Traumatol. 15:77-82, 1999.
24- Ibarrola JL, Chapman BL, Howand JH, Knowles KI, Ludlow MO:
Effect of preflaring on root ZX apex locators. J Endodon 25:625-
626, 1999.
25- Aurelio JA, Nahmias Y, Gerstein H. A model for demonstrating
an electronic canal length measuring device. J Endodon 9:568-
569, 1983.
26- Nahmias Y, Aurelio JA, Gerstein H. An invitro model for
evaluation of electronic root canal length measuring devices. J
Endodon 13:209-214, 1987.
27- Tinaz AC, Alacam T, Topuz O: A simple model to demonstrate
the electronic apex locator. Int Endo. J. 35:940-945, 2002.
28- Ricard O, Roux D, Bourdeau L, Woda A: Clinical evaluation of
the accuracy of the Evident RCM Mark II apex locator. J Endodon
17:567-569, 1991.
29- Huang L: An experimental study of the principle of electronic root
canal measurement. J Endodon 13:60-64, 1987.
30- Berman LH, Fleishman SB: Evaluation of the accuracy of the
NeoSono D. electronic apex locator. J Endodon 10:164-167, 1984.
31- Hulsmann JH, Pieper K: Use of an electronic apex locator in the
treatment of teeth with incomplete root formation. Endod Dent
Traumatol. 5:238-241, 1989.
Fig. 1: The three tested electronic devices for tooth length measurement.
Fig 2: The invitro model during tooth length measurement