the exposure index for digital radiography (iec 62494-1...
TRANSCRIPT
3/11/2015
1
MDACC Imaging Physics1
AAPM Spring Clinical Meeting
2015
The Exposure Index for Digital Radiography
(IEC 62494-1 and AAPM Report 116)
MDACC Imaging Physics
S. Jeff Shepard, MS, DABR, FAAPMImaging Physics DepartmentDiagnostic Imaging Division
The University of Texas M. D. Anderson Cancer CenterHouston, Texas
Acknowledgement:Michael Flynn, PhDHenry Ford Hospital System
MDACC Imaging Physics2
Learning Objectives• Recognizing good technique in DR - Noise• Understand causes and solutions to “exposure
creep”• Understand IEC/TG116 Exposure Index• Verifying EI calibration in the clinic• Be aware of additional recommendations in TG116
MDACC Imaging Physics3
Identifying Correct Clinical TechniqueRecognizing Bad Images
• In the film-screen world, under- and over- exposures were easily recognized by the appearance of the recorded image – Repeat at a higher or lower technique
based on optical density
3/11/2015
2
MDACC Imaging Physics4
Recognizing Bad Images• With Digital Radiography, under- and
over-exposures are not so easily recognized
– Adequate images over a much wider range of exposure
– Post-processing can hide mistakes– Excellent dynamic range may have down-sides
Identifying Correct Clinical Technique
MDACC Imaging Physics5
Exposure Indices
MDACC Imaging Physics6
Exposure Indices
3/11/2015
3
MDACC Imaging Physics7
Exposure “Creep”• Under-Exposure
– Higher noise– Detail visibility suffers
• Over-Exposure– Lower noise (“Pretty” – improved SNR)– High patient dose
• Radiologists may complain about noise, but not usually about over-exposure
– Technologists learn quickly how to avoid criticism– If no one is paying attention, exposures will
“creep” up.
MDACC Imaging Physics8
Exposure “Creep”Widely known problem that’s been around
for a long time• Freedman M, Pe E, Mun SK, Lo SCB, Nelson M,
“The potential for unnecessary patient exposure from the use of storage phosphor imaging systems,” SPIE 1897:472-479 (1993).
• Gur D, Fuhman CR, Feist JH, Slifko R, Peace B, “Natural migration to a higher dose in CR imaging,” Proc Eighth European Congress of Radiology, Vienna Sep 12-17, 154 (1993).
MDACC Imaging Physics9
Vendor-Specific IndicesManuf. Index Lin or
LogExposure relation
Std kV/Filter Cal
Agfa lgM Log Direct 75/1.5 mm Cu 1.96 bel @ 2.5 uGyAlara EIV Log Direct 70/7.1 mm HVL 2000 mbel @ 10 µGyCanon EXP Linear Direct 80/8.2 mm Al 2000 @ 10 µGyCanon REX Linear Direct -- 106 @ 10 µGy*Carestream EI Log Direct 80/0.5 Cu + 1 Al 3000 @ 1 mRFuji S Linear Inverse 80/3 mm Al
“Total”200@1 mR
GE UDExp Linear Direct 80/21 mm Al 2.85 @ 0.5 mAsImaging Dynamics
SE Linear Inverse 80/1 mm Cu 200 @ 10 uGy
Konika S Linear Inverse 80/3 mm Al “Total”
200@1 mR
Philips EI_s Linear Inverse 70/7.1 mm HVL 400 @ 2.53 µGySiemens EXI Linear Direct 70/0.6 mm Cu 100 @ 1 uGy
3/11/2015
4
MDACC Imaging Physics10
Exposure IndicesIEC 62494-1 & AAPM TG116
• Recommendation for a standard detector exposure index for all digital radiography
• Published 2008• Mike Flynn and Jeff Shepard represented the
USA (USNC TAG members) and AAPM on the IEC working group
MDACC Imaging Physics11
Exposure Index (EI):• Index is proportional to the air Kerma that the
detector would have received under standard beam conditions for the same raw pixel value in the relevant image region.
• Calibrated for the imaging system over the specified operating range of image receptor air kerma such that:
EI = (100 uGy-1) * KCALwhere KCAL is the image receptor air kerma in μGyunder the calibration conditions.
• For Krel = 10 uGy, EI = 1000• For Krel = 2.5 uGy, EI = 250
Exposure Index – IEC
MDACC Imaging Physics12
“Relevant” Image Region• Gray histogram is for the
entire image.• Black histogram is for the
relevant anatomic region or Values of Interest (VOI)
3/11/2015
5
MDACC Imaging Physics13
“Relevant” Image Region• Gray histogram is for the
entire image.• Black histogram is for the
relevant anatomic region or Values of Interest (VOI)
MDACC Imaging Physics14
“Relevant” Image Region
• AAPM TG116: “The median is recommended rather than the mean or mode because it is less affected by data extremes and outliers.”
• IEC 62494-1: “The [indicator] shall be calculated using the mean, median, mode, trimmed mean, trimean, or other recognized statistical method for the description of central tendency of the [values] in the relevant image region.”
MDACC Imaging Physics15
Region to assess signal indicator
Systems vary in the region used to assess the signal for an image.
• Full Image
• Regular regions
• Anatomic regions (segmentation)
3/11/2015
6
MDACC Imaging Physics16
Deviation IndexBoth standards also call for a “Deviation Index”
DI = 10 × Log10{EI/EIT(b,v)}
• EIT (b,v) is a table of target values storedby body part (b) and view (v)
DI = 0 is a perfect exposureDI = +1 means exposure was high/low by about 28%
(one density or mAs step)
• EIT tables to be customized for each site• If not customized, default value of DI is 0.
MDACC Imaging Physics17
Deviation Index• Exposure indices are saved in the DICOM header
– 0018,1411 (EI)
– 0018,1412 (EIT)
– 0018,1413 (DI)
– 0018,1405 (EI) and 0018,6000 (Sensitivity) are old and no longer recommended
• DI Format:– AAPM: Decimal string with one decimal place (tracking
and trend management)
– IEC: Integer
• Both indices change with VOI modification by the tech
MDACC Imaging Physics18
Values of interest – VOIPixel values to be filtered/re-scaled for
presentation
Num
ber
of P
ixels
Pixel Value
Values of Interest
EI and DI calculated from
median pixel value
EI = EITDI = 0.0
3/11/2015
7
MDACC Imaging Physics19
VOI Recognition Failure• Gonadal shields, prosthetics, surgical mods• False EI & DI reported
Num
ber
of P
ixels
Pixel Value
CorrectValues of Interest
EI and DI calculated from incorrect median
pixel value
Incorrect Values of Interest
EI ≠ EITDI = -1.3
EI = EITDI = 0.0
MDACC Imaging Physics20
EI = EITDI = 0.0
Manual Correction of VOI FailureTech manually returns the VOI to the proper position
by reprocessing
Num
ber
of P
ixels
Pixel Value
Correct EI and DI calculated from new median pixel value
CorrectValues of Interest
MDACC Imaging Physics21
RQA5 is the standard beam condition• RQR5 pre-filtered beam (70 kVp, 2.6 mm Al
HVL)• 70 kVp• 21 mm added pure Al filtration• 6.8 mm Al HVL
But …• Pure Aluminum is impractical for field
measurements (expensive, heavy, hard to find).• An alternative is to use copper (cheap, portable
and widely available).• Valid substitution?
Standard Beam
3/11/2015
8
MDACC Imaging Physics22
Beam Quality – spectral shapeReport 116 reported equivalent spectral shape with RQA5 conditions and Cu/Al filtration
72.59 kVp3.02 mm Al pre-filtered HVL
• 21 mm type 1090 Al (99.9%)• Filtered HVL : 6.80 mm Al
+ + + + +• 0.5 mm Cu + 2.8 mm Al• Filtered HVL : 6.80 mm Al
HVL6.8 mm Al
Normalized Spectra,Cu/Al spectrum is about 2X that of RQA5
MDACC Imaging Physics23
Standard BeamReport 116 illustrated how the addition of different Al filters to a Cu filter could compensate for differences in the unfiltered beam quality.
Approximate Al to add to 0.5 mm Cu to achieve 6.8 mm HVL at 70 kVp. Averaged from about 25 systems tested.
MDACC Imaging Physics24
AAPM Report 116• HVL = 6.8 mm Al• Adjust Al and (if necessary) kVP to get HVL• 70 ± 4 kVP• 0.5 mm Cu + (0 – 4) mm Al (type 1100)
– Brass acceptable as a Cu substitute– 21 mm pure Al acceptable.
IEC Beam condition• 70 kVP• 0.5 mm Cu + 2 mm Al• HVL = 6.8 +/- 0.3 mm Al
Beam Quality – IEC vs Report 116
• Assumes 2.9 mm HVL unfiltered beam
3/11/2015
9
MDACC Imaging Physics25
AAPM Report 116• HVL = 6.8 mm Al• Adjust Al and (if necessary) kVP to get HVL• 70 ± 4 kVP• 0.5 mm Cu + (0 – 4) mm Al (type 1100)
– Brass acceptable as a Cu substitute– 21 mm pure Al acceptable.
IEC Beam condition• 70 kVP• 0.5 mm Cu + 2 mm Al• HVL = 6.8 +/- 0.3 mm Al
Beam Quality – IEC vs Report 116
• Assumes 2.9 mm HVL unfiltered beam
Verifying EI calibration
MDACC Imaging Physics27
Verifying EI Calibration• Check generator kV calibration/reproducibility • Set up the Standard Beam HVL (add copper and
adjust Al filtration)• Establish a reference dosimeter away from the CR• Obtain grid and table-top attenuation factors from
the manufacturer• Set up for a 10 µGy exposure to the detector (EI =
1000)• Compare to EI that the system reports
3/11/2015
10
MDACC Imaging Physics28
Beam setup – step 1Prior to any measurements verify that the
x-ray source has acceptable exposure reproducibility (coefficient of variation < 0.03) and kV accuracy (within ± 3%) at the standardized condition.
MDACC Imaging Physics29
Beam setup – step 2Collimate the x-ray beam to only
cover the ion chamber with no more than 1 inch margins.
Add 0.5 mm copper filtration at the face of the collimator.
For DR systems, the detector should be covered with a lead apron or similar barrier when making the exposures for HVL determination and adjustment.
MDACC Imaging Physics30
Aluminum for HVL
Beam setup – step 3Measure the HVL of the filtered
beam
Adjust the kVP and/or aluminium filtration within the limits specified to obtain a HVL as close as possible to 6.8 mm Al.
70 kVP + 4
0 – 4 mm Al
3/11/2015
11
MDACC Imaging Physics31
Beam setup – step 4 (DR)The detector should be placed as far from the
x-ray source as is practical, at least 100 cm. Collimate to the edges of the detector.
If present, remove the grid and any other components between the ion chamber and the image detector. If any components cannot be removed, obtain the attenuation factors from the DR system or component vendor.
MDACC Imaging Physics32
Source
> 25 cm(CR only)
Source to Detector Distance
(maximum possible)
Lab Jack(CR only)
Collimator
Added Filtration
Lead (CR only)
Beam setup – step 4 (CR)For CR, the cassette should be
separated from any surface that may increase backscatter from that surface, as recommended in AAPM Report 93 (TG 10). Use lead behind the plate to further reduce backscatter.
If the detector is not square, the long axis of the detector should be perpendicular to the x-ray tube A-C axis.
MDACC Imaging Physics33
Detector housing
Beam setup – step 5Place a calibrated ion
chamber at the center of the beam approximately midway between the source and detector (Position A).
All distances should be measured from the focal spot as indicated on the x-ray tube housing.
3/11/2015
12
MDACC Imaging Physics34
Source
Source to Detector Distance
(maximum possible)
Ion chamber(position A)
Source to Chamber Distance
Chamber to Detector Distance
Collimator
Added Filtration
Beam setup – step 6Use lead to protect a DR detector.
Make an exposure. Using an inverse square correction and applying the grid and table attenuation factors (if present) determine the air kerma at the detector (KCal).
Change the mAs setting to obtain a Kcal value that is in the middle of the detector’s response range (suggest 10 uGy).
MDACC Imaging Physics35
Source
Source to Detector Distance
(maximum possible)
Source to Chamber Distance
Chamber to Detector Distance
Collimator
Added Filtration
Ion chamber(position B)
Beam setup – step 7Move the ion chamber perpendicular to
the tube axis such that it is at the edge of the field of view (Position B).
Ensure that the entire ion chamber is in the radiation beam and is not shadowed by a collimator blade.
Make an exposure using the mAs found earlier and determine the ratio of the air kerma at Position A to that at Position B.
MDACC Imaging Physics36
Beam setup – step 8Remove the protective lead and
expose the detector again.
Verify the resulting KCal by monitoring the exposure with the chamber at position B.
Compare the corrected reading to the Exposure Index reported by the system/detector. EI should equal KCal *100 if properly calibrated,
3/11/2015
13
MDACC Imaging Physics37
EI CalibrationEI calibration should be verified
• At acceptance• Annually thereafter• After service events and software
upgrades
MDACC Imaging Physics38
Rules for repeatsLittle clinical information on reject
thresholds in literature• Van Metter and Yorkston
–Chest and abdominal imaging –Wide range of patient thicknesses –Most AEC controlled images fell within
the range of DI = ± 1.2.
MDACC Imaging Physics39
Rules for RepeatsEmulating film/screen limits
ΔOD = γ*Log10 {E2/E1}, and DI = 10*Log10 {E2/E1}
So: Δ DI = 10*ΔOD / γγ (film gamma): slope of H&D at 1.0 above B+F
For an OD range of +0.3 (0.6 OD total) and γ = 2.5:
Δ DI = 10 * {0.6/2.5} = 2.4Acceptable DI range is + 1.2
3/11/2015
14
MDACC Imaging Physics40
DI action levels (Table 2)DI EI Action
Between -1.0 and +1.0 79% < exp < 126% Check for noise and
clipping (always)
< -1.0 < 79% of targetCheck for noise and
consult with radiologist/management
on need for repeat, investigate cause.
Between +1.0 and +3.0
126% – 200% of target
Repeat only if relevant anatomy is clipped or
“burned out”
> 3.0 > 200% of target
Repeat only if relevant anatomy is clipped or “burned out”, require
immediate management follow-up
MDACC Imaging Physics41
DI Action LevelsSites reporting excessive follow-up.
Action levels based on screen/film emulation is admittedly arbitrary.
Little or no published literature.
MDACC Imaging Physics42
DI Action LevelsAAPM Report 232
To investigate the current state of practice for CR/DR Exposure and Deviation Indices based on AAPM Report 116 and IEC 62494, for the purpose of establishing achievable goals (reference levels) and action levels in digital radiography. The products of this task group will be a brief report and an updated version of Table 2 from AAPM Report #116.Jaydev Dave and Kyle Jones, Co-Chairs
3/11/2015
15
MDACC Imaging Physics43
Other TG116 Recommendations• VOI Histograms and graphical pixel
overlays• Repeat/Reject logs• Import/Export of EIT tables• Export of for-processing image data• Dependence of EI on kV
MDACC Imaging Physics44
AEC and Technique Charts Still important in controlling patient
exposure• AEC may need to be adjusted for CR and
add-on DR– Energy dependence may not be the same as
GdO2S• Technique charts are important for
portables and extremities
MDACC Imaging Physics45
AEC Calibration Using EIEI should remain consistent with:
• Varying energy• Varying phantom thickness• Varying AEC cell• Varying dose rate (mA)
• EI should be reproducible over multiple exposures
3/11/2015
16
MDACC Imaging Physics46
What the Exposure Index is NOT for
Patient dose estimation– Need beam HVL, pt. thickness, SSD and
SID, grid atten, AEC pickup atten, detector assy. input atten …
– If you have all these, you don’t need EI!
System intercomparisons– Index says nothing about detector energy
dependence, efficiency. - VOI recognition strategy will dramatically
affect indices.
MDACC Imaging Physics47
Exposure Index MonitoringQC programs based on exposure indices
are successful– Seibert JA, Shelton DK, and Moore EH,
“Computed Radiography X-ray ExposureTrends,” Academic Radiology 3, 313-318(1996).
Katie Hulme will cover this next.
MDACC Imaging Physics48
SummaryExposure Indices
• Noise, not density• Exposure creep• Exposure indices
– AAPM Report 116– IEC 62494-1
• Calibration• Rules for repeats
3/11/2015
17
MDACC Imaging Physics49
THANK [email protected]
MDACC Imaging Physics50
References:Fredman M, Pe E, Mun SK, Lo SCB, Nelson M, The
potential for unnecessary patient exposure from the use of storage phosphor imaging systems, SPIE 1897:472-479 (1993).
Gur D, Fuhman CR, Feist JH, Slifko R, Peace B, Natural migration to a higher dose in CR imaging, Proc. Eighth European Congress of Radiology, Vienna,Sep 12-17, 154 (1993).
Yorkston J. Flat-panel DR detectors for radiography and fluoroscopy. In: Specifications, Performance Evaluations, and Quality Assurance of Radiographic and Fluoroscopic Systems in the Digital Era,Goldman LW and Yester MV eds. Madison, WI: Medical Physics Publishing (2004)177-228.
MDACC Imaging Physics51
References:Willis CE, Thompson SK and Shepard SJ. Artifacts
and Misadventures in Digital Radiography. Applied Radiology 33(1):11-20, January 2004.
R.E. Alvarez, J.A. Seibert, and S. K. Thompson,Comparison of dual energy detector system performance, Medical Physics31(3), 556-565 (2004).
JA Seibert, DK Shelton, and EH Moore, Computed Radiography X-ray Exposure Trends, Academic Radiology 3, 313-318 (1996).
J A Seibert, et al, AAPM Report #93, “Acceptance Testing and Quality Control of PhotostimulableStorage Phosphor Imaging Systems: Report ofAAPM Task Group 10.” AAPM (2006)
3/11/2015
18
MDACC Imaging Physics52
References:Richard S., Siewerdsen J. H., Jaffray D. A., Moseley
D. J. and Bakhtiar B., Generalized DQE analysis of radiographic and dual-energy imaging using flat-panel detectors, Med. Phys. 32 (5), May 2005, 1397 – 1415
Lehman L. A., Alvarez R. A., Macovski A., Brody W. R., Pelc N. J., Reiderer S. J., and Hall A., Generalized image combinations in dual KVP digital radiography, Med. Phys. 8 (5), Sept/Oct 1981, 659 – 667
Shepard S.J., et al, AAPM Report 116, An Exposure Index for Digital Radiography (Executive Summary), Med. Phys., 2009
MDACC Imaging Physics53
References:IEC 62494-1, Medical electrical equipment- Exposure
index of digital X-ray imaging systems Part 1: Definitions and requirements for general radiography, International ElectrotechnicalCommission, 2008