the franklin institute power input. following this, rf ... 3.3 output indicator and power supply...
TRANSCRIPT
F Fnai Report F-B1SOS
C
(Evaluation* of MARK 1 MOD 0 Squih and
L MARK 2 MOD 0 Ignition Element)
by
C/) Paul F. MohrbachC-5 Robert F. Wood
Norman P. Faunce
16ftftJanuary 23, 1961 to August 15, 1962
C%4mE.~ Prepared for
U.S. NAVAL WEAPONS LABORATORYITDahigren, Virginia
Code WHR
Contract No. N178-8730
THE FRANKLIN INSTITUTELABORATORIES FOR RESEARCH AND DEVELOPMENT
P HI L A D E L PHI A P E NNS YL VA N IA
THE FRANKLIN INSTITUTE Lborw for Raumh &a Dmsu wW
Final Report F-B1805
PRECISION RF SENSITIVITY STUDIES
(Evaluations of MARK 1 MOD 0 Squib and MARK 2 MOD 0 Ignition Element)
by
Paul F. MohrbachRobert F. WoodNorman P. Faunce
January 23, 1961 to August 15, 1962
Prepared for
U.S. NAVAL WEAPONS LABORATORYDahlgren, Virginia
-Code WHR
* , Contract No. N178-8730
THE FRANKLIN INSTITUTE • Lbonumin for Rmami aM Dewlog4 t
F-B1805
ABSTRACT
Developments leading to the establishment of precision RF firingevaluation techniques are set forth in this final summary report. TheRF sensitivity of the MARK 1 MOD 0 squib and MARK 2 MOD 0 ignition elementhas been determined by these newly developed precision firing techniques.,Data from 5 to 1000 Mc CW are obtained, differing little from the dcpower sensitivity determined from 10-second constant current pulse data.
Many developments preceded the actual tests. Studies of systemlosses indicated that special matching devices would be desirable, forlimiting to reasonable values the power lost in the RF plumbing. As aconsequence, a novel shorted stub was developed for use at the higherfrequencies. Low-loss lumped-parameter tuning networks were fabricatedfor the lower frequencies. Toward this same end, the design andconstruction of the fixture for mounting the initiator under test wasgiven special attent n.
Results with 'recision never before duplicated were obtainedby use of a special cali'ation procedure that was developed. A photo-conductive cell is focused onto the exposed bridgewire of the deviceto be tested and its response is calibrated against a steady-state dcpower input. Following this, RF is fed to the same device and the ratioof the RF to dc level for the same bridgewire illumination is taken asthe system calibration factor. In so doing, all losses between the pointat which we measure input RF power and the bridge element are lumped.Because of this it is important first to evaluate the losses in the plugor base of the initiator, and, then with this to correct the systemcalibration factor. For the two itenms evaluated no detectable base losswas evident up to 1000 Mc.
THE FKANKlIN INSTITUE Labomsur 1ev URmi d Dsumv ut
F-EL805
TAKlE OF CONTENTS
ABSTRACT . i
1. INTRODUCTION . 1
2. INVESTIGATIONS OF FIRING SYSTEM LOSSES . . . . . . . . . 2
2.1 Losses Determined by Use of Thermocouple Loads . 32.*2 Infrared Detector Above a Bridgewire for Loss
Determination .. .. .. .... ..... . ...... 62.3 Photo-Conductive Cell for Loss Determination . . . 72.4 Evaluations of Losses Attributable to the Squib or
Ignition Element .. .. .. .. .. .. .. ..
3. FIRING SYS=D AND INSTIIJNENTAT ION DEVELOPKM T. .. ..... 10
3.1 Matching Systems. .. .. .... ..... . ... 11
3.1.1 A Novel Adjustable Shorted Stub .. .. . .11
3.1.2 An Expedient Shorted Stub Adjustable Line1;Matching System. .. .. .. .... ...... 123.1.3 Low Frequency Matching Systems. .. .. ..... 15
3.3 Output Indicator and Power Supply for*Clairex 1
3.4 Calorimeter owr Reference'Souc.........2
4. DETERMINATION OF SYSTEM LOSS FACTOR; SYSTEM CALIBRATION 27
5. IRF SENSITIVITY TESTS OF MARiK1MODO0SQUIB AND MARK 2MODO0IGNITION EMT . .. .. .... . ..... ..... 30
6. CONCLUSICNSAND REC E NDATIONS. .. .. .... ...... 33ACN4EGMNS. .. .. ..... . .... ........ 36APPENDIX A - Systems Calibration Data .. .. ........ A
APPENDIX B - Bruceton Data and Analysis Forms. .. ..... BIAPPENDIX C - Denver Research Institute Simlaktor Tests. C
THE FRANKLIN INSTITUTE *Labomsn fi Away&i ad Dwomu~mu
F-refl05
LIST OF FIGURES
2-1 Percent of. Incident Power into Thermocouple VersusHeater Resistance (load) .... .......... 5
2-2 Comparison of Standard and "Loseless" MARK 2 MOD0 Ignition Element. . . . . . . . . . . . 10
3-1 Adjustable Shorted Stub - Disassembled. . . . . 13
3-2 Close-up of Collets for Adjustable Short. . . . 13
3-3 Matching System Employing the Shorted Stub andSection of Adjustable Air Line .... ......... 14I3-4 Theoretical Circuit for Matching MARK 2 IgnitionElement at Low Frequencies ........ ... 15
3-5 A Typical Lumped Element Matching Network . . 17
3-6 Cross Section of MARK 2 MOD 0 Firing Mount.. 19
3-7 Firing Mount and Photocell Housing .......... 193-8 Adapter Parts for Accommodating MARK 1 MOD 0
Squib ....... .. ..................... 203-9 Firing Mount and Photocell Housing (Adapted for
MARK 1 MOD 0 Squib) .............. ... 20
3-10 Clairex Cell Output Indicator and Power Supply. 21
3-11 Schematic Diagram - Clairex Cell Output Indicatorand Power Supply ...... ................ 22
3-12 Simplified Circuit of HP Model 434A CalorimetricPower Meter ............ . . . . 24
3-13 Diagram of Calorimeter Calibrator Circuit . . . 26
3-14 Change in Load Power for Change in LoadResistance. .... ................. . . . 27
4-1 Basic Arrangement for Firing System Calibration 28
LIST OF TABLES
4-1 System Loss Factors ................ .... 29
5-1 RF Sensitivity of MARK 1 MOD 0 Squib and MARK2 MOD 0 Ignition Element ..... .......... 31
iii
:1 THE FRANKLIN INSTITUTE *LabWumvW for Rmdi aidd Deuh9..
F-B1805
1. INTRODUCTION
The purpose of this program was to determine the RF sensitivity
over a wide range of frequencies of the MARK 1 Squib and the NARK 2 MOD 0Ignition Element with an error of less than 5%. The plan required thedevelopment and careful calibration of suitable systems of instrumentation.
The test procedure was to use the Bruceton routine wherein power is thevariable and each initiator is tested at one of several predetermined
power levels; the particular level for each being one step higher orlower depending upon whether the preceding result was a non-fire or fire
respectively. By following this procedure, testing randomly sampled
groups of 40-100 initiators at each of several frequencies we can derive
a curve of 37 power versus frequency.
In the early days of the program, exhaustive tests were performed
to evaluate losses in various matching devices, over a frequency range of250 to 500 megacycles: Experimental tests were supplemented with
theoretical studies of impedance matching, which led to the conclusion
that large system losses are inevitable when matching into loads having
a smll resistive component or large reactive components. Acknowledgingthat system losses of 50 per cent or more were unavoidable, we proceeded to
assemble and test practical systems for firing the two items being studied.
Calibrated RF thermocouples were used in the first attemptsto evaluate system losses. These thermocouples were useful for comparing
the relative losses of different matching systems. It was soon realized,
however, that a better way to determine absolute system loss would involve
determination of the RF power actually reaching the bridgewire of the
ignition element or squib under test. This was done by means of photo-
conductive cells mounted above the bridgewire of an inert device.
---
THE "A MU INS7ITUTE . wm for Rmw m Dwa
F-B185
The system lose evaluation technique above has one inherentsource of error, in that the losses in the base of the ignitionl element
are included in the lose factor obtained. Base losses, if present,
should not be included in the loss factor.
Several experiments were made to determine the loss in the
base of the MAR 2 MOD 0 ignition element at 250 megacycles. Nonumerical values were obtained since the losses were too small to measure.For frequencies below 1000 Mc, therefore, it is sufficiently accurateto ignore the base loss. A Bruceton test is conducted and a figure forsystem input power for 50% functioning obtained. Multiplication of the
system input power by the system efficiency factor gives a calculatedvalue for the power at the bridgewire. The value for calculated powerto the bridgewire should remain essentially the same for all frequencies
provided no factor other than bridgewire temperature causes initiation.
A valuable consequence of these studies was the evidence ac-
cumulated to show that stock radio frequency components (shorted stubsin particular) were not ideally suited to precision firing systems. This
discovery resulted in extensive development of components, not originallycontemplated in the scope of this project.
2. INVESTIGATIONS OF FIRING SYSTEK LOSSES
The accuracy of EED RF functioning sensitivity evaluation
depends upon the degree to which it is possible to resolve the powerdelivered to the items being tested. Firing systems which feed knownamounts of power into a matched termination will yield accurate resultsonly if the hardware between the load (EED) and the point at which the
power is introduced can be considered lossless. As these elementsbecome lossy - that is, as their dissipation of power becomes sig-
nificant in relation to the power dissipated by the load - the accuracyof resulting data is correspondingly reduced. However, if the amount of
-2-
THE FRANKLIN INSTITUTE Labouusiu for Rmu.ra tam Dowbom
F-B1805
power lost in these parts can be determined with certainty equal to that
associated with the measurement of input power, we can deduce the powerdelivered to the EED with comparable accuracy.
A two-prong program was initiated early in the study to selectthe most effective matching system and to specify its loss with reasonableaccuracy. Since the loss would probably be related to the magnitude ofthe terminating impedance, we planned tests using UHF vacuum thermocouples
of varying resistance as loads to investigate low frequency losses. Fortests at higher frequency, above the limit for the thermocouples, weemployed thermal radiation detectors whose response was calibrated to
be proportional to the power dissipated in the bridgewire of the ED tobe evaluated.
2.1 Losses Determined by Use of Thermocouple Loads
Lose measurements on impedance matching devices were initially
restricted to the frequency range of 250 to 500 megacycles. The results
were to serve as a guide for an intelligent approach to the more difficultrange extending up to 1000 Mc, and beyond.
Stub stretchers, double stub tuners and triple stub tuners aremade up from various combinations of adjustable stubs and line sections,
and are almost universally employed for impedance matching at frequenciesabove 200 megacycles. A line stretcher, which is an adjustable length
of coaxial line, is ordinarily inserted as a series component. An
adjustable stub is a variable length of coaxial line which is shortedat one end and must of necessity be employed as a shunt. Open circuitstubs are seldom used because of radiation problems at the open end.These devices utilize the principle of combining series and shunt coaxialimpedance elements to produce a match between the load and source impedance.
THE FRANKLIN INSTITUTE . LAbwd for Rmawch & DaWoomm
F-BIS05
An experiment was designed to determine the relative losses
of several types of coaxial matching devices. A double stub tuner,
a stub stretcher and an assortment of line stretchers and shorted stubs
were available for tests. Six coaxially mounted UHF vacuum thermocouples
were employed as load terminations. The nominal values of dc heater
resistance for these thermocouples are 50, 9, 2, 0.6, 0.4 and 0.2 ohms.
An additional experiment was designed to determine the amount
of power delivered to each of the thermocouples without matching. Data
from this experiment were used to calculate an unmatched voltage
standing wave ratio (VSWR) for each thermocouple. The aim was to
determine the correlation between the unmatched VSWR and the powerdelivered to the thermocouples when matched.
Details of this evaluation may be found in report Q-B1805-2.
Figure 2-1 is included here to show the nature of the results. Of the
several matching systems examined, the stub stretcher showed the least
loss in these tests. The figure suggests that we may expect about 35%
system losses for loads comparable to the MARK 1 MOD 0 (1.0 ohm) and
70 to 80% for the MARK 2 MOD 0 (0.1 ohm), in the 300 to 500 Mc frequency
range.
The dotted curves included in Figure 2-1 indicate the nature
of losses that would result if we failed to match these loads. From
these data we determined an estimate of VSWR which was to be compared
to that value computed on the basis of impedance measurements. However,
when the impedance measurements were made it became apparent that the
thermocouples' impedances at these higher frequencies were in no way
similar to the impedances expected of the EED terminations. Accordingly
it was presumed that the system loss data acquired in this manner could
not be relied upon to give the needed accuracy. Emphasis was therefore
shifted to examination of photodetector techniques.
-4-
THE FRANKLIN INSTITUTE *Lubouries for Ruwwh and DemLoPnmn
{ F-B1803
110
loo
x 0
WITH STUS STETHE
o /o
~20
ALOAD NOT MATCH4ED/ (STUB STRETCHER
REMOVED)I0.1 3 3 4 aSS?Sol it 3 4 5S6?SS* a 4THERMOCOUPLE HEATER RESISTANCE (ohm)
m~ z-. PERCENTr or RI/CNT POwER iNT THEROCOUPLE VERSS
HEATER ARSISTANCE (LOAN)
THE FRANKLIN INSTITUTE - LAbmuWYW for Ramh and Dw
F-Br.805
2.2 Infrared Detector Above a Bridgewire for Loss Determination
It appeared reasonable that an infrared detector mounted above
and close to a heated bridgewire should have a response related to heat
energy from the wire. It should make no difference whether the wire is
energized by dc or RF power, if the wire has a small diameter and a
length which is short compared with the wavelength of the applied RF
energy. These assumptions were experimentally validated before this
technique was accepted for use.
We know from the experience gained with thermocouple loads
that matching losses could be made negligible at 250 megacycles for a
load of approximately 50 ohms. On the other hand, we had been unable
to prevent excessive losses when matching a 0.2-ohm load. Therefore,
we could be reasonably certain that the method was sound if the detector
above a 50-ohm wire provided equal indications for equal powers of dc
and RF excitation.
The instrumentation was designed around a pair of Kodak Ektron
infrared detectors, type N-1, mounted on a common heat sink to provide
temperature compensation. One cell is exposed directly to a radiating
source while the other is shielded.
The photo-conductive cells were connected as opposing arms of
an ac resistance bridge with a l000-cps excitation voltage. Relative
output of the infrared detector is indicated by balancing the bridge
and reading a calibrated dial.
The infrared detector mount provides a means for placing the
cell in the best position relative to the bridgewire. One end of the
device under test is installed in an adapter fitting for RF input. The
other end, with the bridgewire exposed, is put into an opening in the
detector mount.
-6-
THE FRANKLIN INSTITUTE • Laborau for Ruuimk and DAWOmW
F-B1805
As was previously mentioned, this measuring technique would
be valid if equal amount of RF and dc power in a 50-ohm bridgewire would
give equal output from the detector. A value of exactly 50-ohms could
not be obtained; instead, a wire measuring 79 + J 16.5 ohms at 250 Mc was
employed.
Several tests with the 79-ohm wire indicated an efficiency of95 per cent at 250 megacycles. This efficiency was derived from the
ratio of the dc power to the RF power required for unity output of the
detector. Since some loss was expected when matching 79-ohms instead
of 50-ohms, the figure of 95 per cent efficiency was considered to
indicate a confirmation of the infrared method.
Additional tests were made with bridgewires having dc resistances
of 0.28 and 0.40 ohms. The efficiency, using a stub stretcher, varied
between 60 and 65 per cent. These tests were conducted with RF power
of 300 milliwatts.
The system used in these preliminary experiments of photo-
conductor techniques was not ideally adaptable to RF firing system
requirements. However, indications were that the technique was feasible.
While considering the needs for modifying the IR system to this new use
it was decided to investigate other photoconductors such as the Clairex
photoconductive cell. This cell was found to be more suited to the task
of systems loss evaluation than were the cells tried first.
2.3 A Photo-Conductive Cell For Loss Determination
From a number of photoconductors a Clairex Type 404 photo-
conductive cell whose spectral response peaks at 6900 angstroms, and is
down 30 per cent at 6300 and 9000 angstroms, was selected. The cell wasmounted in a light-tight aluminum cover, which was designed to be easilyslipped on and off the firing mount.
-7
THE FRANKLIN INSTITUTE Laboraw for Ruserd ad D'1m W
I F-BI805
We planned to operate the heated brigewire at a temperature
just below the visible glow point. Because of the high output obtained
from the cell at this temperature, an ordinary ohneter, such as a
Triplett Model 630 multimeter, was adequate for initial output indication.
The resistance of this Clairex cell decreases from more than 20 megohms
when dark down to about 500 kilohms for the bridgewire temperatures
involved in our tests.
We exposed the detector above a 50-ohm heater in a matched RF
system to prove again that the method was valid. A length of high
resistance wire was installed on a MARK 2 ignition element plug assembly
(not on the 50-ohm line as was done in the previous experiment) in place
of the low resistance bridgewire. The effective series RF impedance of
this wire was 31-J 10.5 ohms at 250 megacycles.
The ignition element, with the 31-ohm bridgewire, was installed
in the firing mount and matched with a stub stretcher at 250 megacycles.
The ratio of the dc power to the RF power required for equivalent output
of the photo-detector was 96.5 per cent. This was assumed to be near
enough to 100 per cent to prove the technique was valid. We should expect
a loss of at least 3.5 per cent with a load of 31-J 10.5 ohms.
This Clairex photo-cell and housing was used for all of the
system calibrations performed in this study.
2.4 Evaluations of Losses Attributable
to the Squib or Ignition Element
The calibration procedure to be described in the following
section assumes that the items to be tested have no, inherent loss at
the test frequencies. The total power entering the base of the items
is, therefore, propagated to the bridgewire.
-8
-I
THE FRANKLIN INSTITUTE • IAov1 for Rownh and Dmpmw
F-BIB05
This assumption appeared to be valid since attempts to determine
the loss associated with both the MARK 1 squib plug or the MARK 2 ignition
element base gave indication of small power loss.
Attenuation measurements through the base of the MARK 1 MOD 0
squib were made at 250, 500 and 1000 megacycles and values between zero
and 0.2 db were obtained. In some cases the 0.2 db value was for 250
megacycles, and the zero db figure was at 1000 megacycles. This indicatesthat the attenuation is too low to be determined by the measuring
equipment.
Similar results were obtained for MARK 2 MOD 0 ignition elements.These were measured at 500 and 1000 megacycles only. No values higher
than 0.1 db were recorded for the attenuation through the base of the
ignition element.
In section 2.3, an efficiency of 96.5% at 250 megacycles was
cited for an element with a 31-ohm bridgewire; this indicates that
attenuation through the plug cannot be very great at that frequency.However, we wished to be certain-that-tbe greater part of the 3.5% powvr.
loss was not expended in the plug.
Reference to Figure 2-2 (a) will give some idea of the complexityof the various thin sections of dielectric materials in the base of the
MARK 2 ignition element. A simulated element was constructed to have
the same outside physical dimensions and bridgewire resistance as a
standard MARK 2. Its losses were expected to be less than those ,of the
standard because of the use of a polystyrene insulator in place of the
thin nylon insulators and the mica washer. The dimensions were such
that the characteristic impedance of the base is 50-ohms up to the point
of bridgewire attachment. A drawing of this simulant is given in
Figure 2-2 (b).
i _9
THE FRANKLIN INSTITUTE Lboratvies for Research and D opm
F-B1805
(NYLON INSULATOR OUTER ELECTRODE
MICRE
SOLDER *OLDR
ELECTRIJBRASS
MICA NYLON INSULATORINSULATOR
(a) Sltadard Mark 2 Mot 0 Element with ExploSivU (b) "Loseleg" Simulato MarkZ ModO EIrlmntM Part of OutW Cas. Rem.
FIG 2-2. COMPARSON of STANDARD AND "LDSSLESS" MARK! M2 0 16NI "V ELEMENT
The purpose of the unit was to serve as a reference for
comparison with a group of standard MARK 2 ignition elements. Losses
exhibited by the standard item could then be attributed to the base.
At 250 Mc there was no significant difference between the two systems.
3. FIRING SYSTEMS AND INSTRUMENTATION DEVELOPMENTS
Results of our investigation into system losses led to
consideration of means for their reduction, if not their total
elimination. Accordingly a number of development efforts were initiated.
Though we had performed firing tests prior to the inception
of this program, using stock RF components, we did so without the full
realization of the magnitude of the losses thereby introduced. We had
suspected that our systems were far from ideal, but not so far as the
evidence accumulated in this program indicated to us. To effect an
improvement, we designed a special adjustable shorted stub, and took
every precaution to keep the length as short as possible, so that losses
would be minimized.
-10-
THE FRANKLIN INSTITUTE . Laboratories for Researh and Development
F-BI05
Similarly we took especial care in designing mounts, to
eliminate probable causes for loss. Selection and calibration of
instrumentation, likewise, was made with an awareness of the rigorous
demands of this totally new concept of precision firing. As a consequence,
valuable knowledge was acquired, but more importantly, reliable systems
became available and were used to perform RF firing evaluations with
precision not previously obtainable.
3.1 Matching Systems
Our initial system loss evaluations (see Section 2) indicated
that commercially available matching sections were responsible for unduly
high system losses. Their chief drawback may be attributed to one or
both of two shortcomings in design: (1) The use of sliding contacts
which either become dirty or lose their tension and thus give "high
resistance" contact, or (2) because of the need for spring or finger
contacts, the minimum stub length is unreasonably long. To offset these
effects, we designed a unique shorted stub,which was the outgrowth of an
arrangement assembled to fire one test at 250 Mc.
For the lower frequencies stub tuners become unwieldy, and it
is necessary to resort to lumped parameter networks to perform the
matching task.
3.1.1 A Novel Adjustable Shorted Stub
The novelty of our shorted stub lies first in the fact that
it may be adjusted from 60 cm down to a length approaching zero and
second in the unusual arrangement of collets which provide the actual
short circuiting condvctance across the line. A third feature is the
provision of an external knob which allows the collets to be positively
locked in any desired position. A micrometer adjustment for fine
positioning is also included.
For our purpose they are shortcomings; for their ordinary intendeduse, the consequences may be not serious, and this may be an~unjustaccusation. - 11
t THE FRANKLIN INSTITUTE Lubuumg for Rmmdi and DmIApsnut
F-B1805
Figure 3-1 is a view of the adjustable shorted stub, partially
disassembled to expose some of the inner constructional details. It
consists basically of a 70-cm length of coaxial air line (outer and inner
conductor) and an assembly containing the shorting slug which is inserted
within the coaxial line. The 70-cm line is so constructed that it may
be attached to the Junction port of a coaxial tee fitting in such a way
as to form a smooth continuation of the outer and inner conductors of
the tee.
The shorting slug consists of two concentric silver platedcollets. The outer collet has an internal taper which is a mating fit
with the external taper of the inner collet. When the inner collet is
drawn up into and flush with the outer collet, the locking action forces
the contacting fingers against both the inner and outer conductor. The
intimate contact between the tapered surfaces of the collets combines
with finger-spreading collet action to create a plane of high conductance
across the line at the Junction of the coaxial tee. Figure 3-2 is a
close-up view of the collets in the condition just described.
Figure 3-3 is a view of the completely assembled shorted stub
with a section of adjustable air line attached for matching the MARK 1MOD 0 squib at 250 megacycles. A longer adjustable line is shown just
below the main assembly, and a short adjustable line at the bottom of
the figure. These components can be interchanged as required.
3.1.2 An Expedient Shorted-Stub Adjustable Line Matching System
One attempt to solve an immediate matching problem was to
assemble an adjustable line and a type "N" tee Junction with a shorting
cap screwed on one of the ports. This formed.a simple stub stretcher
with a fixed stub and an adjustable line. A trial showed that this"tuner" could almost match a MARK 2 ignition element inserted in the
firing mount. A slight modification, to bring the shorting cap a
little closer to the line Junction, produced a match with a ratio of
incident to reflected power of better than 200 to 1.
1 -12
( ~ ~~THE FRANKLIN INSTITUTE *LAbosmdu for Rmmch cad DmiissomKB1
rfi .5-/. AWISTAE SAVRWE SliD -UMASEMED
-13.
THE FRANKUN INSTITUTE *Laboiuwvia for Romhd and Dewelopmmer
0 g1 a a 4 £ 0 0 1 0 I itM
M* 3-3. AMM'd SYFnW fMWPY ME 5AN?7t 57 AND#CTIG' OF AON*4ft AOF LA
THE FRANKLIN INSTITUTE - Iorawoa for Rmfth and DewloPmnwt
Tests with a number of ignition elements suggested that a
fixed shorted stub and an adjustable line stretcher would produce an
acceptable match for any of the MARK 2 ignition elements at 250 megacycles.
Furthermore, the system appeared to be exceptionally stable compared
with previous systems. Consequently this arrangement was used to fire
the 250 Mc test of the MARK 2 MOD 0 ignition elements while the adjustable
stub was being developed.
3.1.3 Low Frequency Matching Systems
At frequencies df 5 and 30 megacycles, impedance measurements
show that we may assume that the real part of the impedance of the MARK 2
ignition element corresponds to the dc resistance of the bridgewire. The
bridgewire, normally 0.1 ohms when cold, is about 0.14 ohms when heated
to a dull red.
Figure 3-4 is a theoretical circuit for matching a MARK 2 ignition
element to a 50-ohm line. The values of reactance required for matching
were obtained by substitution in equations (3-1) and (3-2)
TRANSMISSION LINE JRZo= 50n
I R. +jojX,2.65n (0.14 n)
tF 3-4. TrVORE7ICAL CIRUr FOR MAN/wo MARKS lNITov ELEAMET AT LOW F$
2 . o a a (3-1)
X2 Z (3-2)-'2
-15 -
THE FRANKLIN INSTITUTE •Labmwovk r Ra mc dand [email protected]*W
F-B1805
These equations yield almost equal values of 2.65 and 2.64 ohms fork and X . At 5 megacycles, a capacitor of 0.012 microfarads and an
inductor of 0.082 microhenries are required for the calculated reactancesof 2.65 ohms. At 30 megacycles, the component values are 0.002
microfarads and 0.0136 microhenries.
The equations from which we calculated X and X assume thatthe load has no reactance. The load in Figure 2-2 is therefore denotedas R + Jo which in the case of the MARK 2 element is approximately 0.14ohms. The net reactance of the load will modify the calculated value ofX2 , however, this will not be considered here since the exact value of
X2 will be obtained empirically in actual practice.
Similar calculations were made to determine the theoreticalvalues of components required for matching the MARK 1 MOD 0 squib at 5and 30 megacycles. The nominal dc resistance of the squib bridgewire
is 1 ohm (cold) and about 1.4 ohms when heated to a dull red. The cal-culated values for Xl and X2 are 8.49 and 8.24 ohms respectively. At5 megacycles, these reactances may be obtained with a capacitor of0.00375 microfarads and an inductor of 0.262 microhenries. At 30
megacycles, these values become 0.000625 microfarads and 0.0436 microhenries.
The matching networks used in the firing of the squib and
ignition elements were simple combinations of inductive and capacitive
elements mounted in an aluminum boc as shown in Figure 3-5. Here canbe seen a variable 1000-picofarad tuning capacitor, a fixed shunt
capacitor, and a series tuning inductor. The fixed capacitor and the
series inductor are removable so that other values may be substitutedaccording to the frequency and the impedance of the item under test.
It would ordinarily be assumed that both the series inductor
and the shunt capacitor would have to be variable in order to allow forthe normal variation in resistance and reactance of the EED being tested.However, it was found that the MAK 1 MOD 0 squibs and the MARK 2 MOD 0
- 16 -
THE FRANKLIN INSTITUTE • Laboawos for ResCdh and D£WIoPmeM
m.3-5. A rYPCAL LUWPED Lw.EMENr AA7 wMU PETwoRK
ignition elements were so uniform in their electrical characteristics
that a fixed inductance and capacitance in combination with a single
variable capacitor was enough to produce an acceptable match for any
individual in a group under test.
3.2 Firing Mounts
The mount which adapts the EED to the RF line is not the
least important part of the system. Mount designs must insure repeatability
of mounting EED's so that no variation in results can be attributed to
deficiendes at this point. Reliable contact between leads and line
conductors must also be assured. It is for this latter reason as well
as to make provision for insertion of a voltage probe, that a redesign
of the mount initially constructed for this test schedule was required.
- 17 -
THE FRANKLIN INSTITUTE • Ok for Ruww& aid Dm Au
F-B1805
Figure 3-6 shows a cross section of the mount finally used to
perform the tests reported herein. Unique features of this mount include
a collet arrangement to insure a positive electrical contact between
the outer case of the ignition element and the body of the mount. Figure
3-7 shows this feature more clearly. The MARK 2 ignition element with
the bridgewire exposed can be seen held by the collet.
A voltage probe can be inserted to make contact with the center
conductor, at a point close to the base of the element. The probe is
removable, and a plug closes the opening during actual firing. The probe
was used for trials to measure power at the input to the EED, by a voltage-
impedance method.
A series of adaptor parts permit this mount to accept the MARK
1 MOD 0 squib. These are shown in Figure 3-8. The leads of the squib
must be cut to exact length end shaped before installation in the mount.
This is done with a special holding and stripping device, not shown. A
cut and dressed squib is in the foreground of the picture, which also
shows another squib mounted in the center conductor. A socket-head screw
is turned to fasten one of the leads in the center conductor. A slot
is provided in the collet to force the ground lead against the bottom
edge of the outer case of the squib. The collet-closer brings the collet
into intimate contact with the bottom circumference of the squib case.
The screw that tightens the center conductor is reached by a special
wrench inserted through a hole in the main body of the mount. The
complete assembly with squib in place is shown in Figure 3-9.
3.3 Output Indicator And Power Supply for Clairex Cell
A Triplett multimeter, Model 630, may be used for indication
of the resistance of the Clairex 404 photocell. When set on the O
ohms scale, this meter indicates 500,000 ohms mid-scale. With the
Clairex cell about 1/16-inch above a squib or ignition element bridgewire
~- 18 -
THE FRANKLIN INSTITUTE *Laborawkis for Research and Dewdopmen
F-B1805
*BODY BRASS VOLTAGE PROSE
A STANDARDGENERAL RADIO -POLYSTYRENECONNECTOR Is INSULATO
INSTALLED HERE / / /
c OUTPUTLEADS
CCLATRE PHOTO-CONDCTORCONDUCTIVE CELL
POLYSYRENELIGHT TIGHT HOUSINGSUPPORT (REMOVABLE)
MARK 2 MOD 0IGNITION ELEMENT
(EXPLOSIVE REMOVED)
FI. 5-6. MRSS-SECTa MV F ~'W2 MOO 0 FIRW AMA'(Photocou Housing Attacod)
PLW
AM .3-? 7iA MO r AMD R'07 LL #LA
-19-
THE FRANKLIN INSTITUTE *Laborutaries for Reseand and D"LOM~'d
LEAD i GLOA D LA
LE.3IRAD ROUAND LEA7~LtCJ/
F/c% 5-& A (Adrpte forr FORi ACOMOr MqAM/MM02)I
AIN -OD-
THE FRANKLIN INSTITUTE • Lboames for Reeach and Demlopment
F-Bl805
I
F/a3-/. C AREx caL ourpur INID/CArOR AND POWER SUPPL
this would correspond to a dull red glow of the bridgewire. Anticipating
the need for greater sensitivity, we developed an improved meter system.
Initial firings indicated that the 50% fire level could be
considerably below a dull red glow of the bridgewire. To facilitate tests
at these lower power levels, we designed a combined photocell power supply
and output indicator, specifically for use with the Clairex cell. Figure
3-10 is a view of this instrument. A photocell in a housing attached to
the firing mount can be seen connected to the instrument by a shielded
cable. A seven-inch 50-microampere meter is used as an indicator.
Although the instrument functions as a high resistance 6hmmeter, there
is no necessity for calibrating the scale in ohms. The original
calibration of the scale combined with observation of the range dial
setting (1 to 10) is all that we need. Description of the electronic
circuit follows.
-a1
THE FRANKLIN INSTITUTE *Libuwwu for Raw& ad Djwmm
F-BM85
TI YIr ~ ___
04? M"LA
III Its.
LN
la () 04 THOU Nit D0* atWINE 30430 PRECISION o -Mi.
500*IRWTTIOO + 036-L404
ANES ITOR A 1%ROTO. all mSOO912 WPE0ATET SOW
W0 N D / T ATT nitn i-u00Ko 0 01NSS
Fle. 5-11 SCHEMAMI DIAN-CLAIREX CELL OMflWW INVAATOR AAV POWER softy7
When the multimeter is used and set to the X10.0 ohm resistance
scale, a 30 volt battery, an indicating mter, and a series resistor to
limit current to 60 microamperes constitute the essential elements of
the circuit. Such a circuit can be made to measure higher values of
unknown resistance if the battery voltage is increased to say, 90 volts,
and the series resistance increased accordingly.
The instrument we designed to replace the multimeter is
essentially a regulated power supply and an ohmmeter circuit.* Details
of the circuit are shown in the schematic diagram Figure 3-11. The
regulated power supply is a conventional supply' with an 0B2 gas-tube
voltage regulator. The output across this tube is approximately 107 volts.
Ris a seri~es rheostat, which is adjusted to give a voltage drop of
exactly 100 volts across the precision voltage divider Rto R,
-22 -
THE FRANKLIN INSTITUTE • Labmw for Rommdi d Dms mft
F-B S05
The voltage divider, consisting of ten precision wirewound
resistors of 500 ohms each, draws a constant 20 milliampere current from
the 100 volt supply. Voltages from 10 to 100 volts, in steps of 10 volts,
are selected by the range switch S2. Series resistors R,4 to R23 are so
chosen as to cause the "calibration" current to be 50 microamperes at
each range. When the switch S is thrown to "Calibration Check" position,3the Clairex cell is shorted out and the meter will read 50 microamperes
+ 1% on all ranges. The first range position, S set to R3, is considered
to be the reference position. R3 is adjusted so that the meter reads
exactly 50 microamperes or full scale when S3 is on CAL. check and the
range switch is on range 1. All other ranges will then be automatically
adjusted to within 1%, which is the tolerance of the resistors used.
The value of 20 milliamperes for the current through the voltage
divider was chosen to provide a maximum loading effect of 0.4% when the
meter is indicating over the range of zero to 50 microamperes.
Voltage ranges of 10 to 100 volts in steps of 10 were chosen
in place of normal scales 1X, lOX, lOOX, etc., because steps of 10 volts
are more practical for our special application. When-used on the highest
range, the meter is capable of indicating bridgewiie heat aout 40% below
a dull red glow. The meter will then indicate at about mid-scale or 25
microamperes. This unit is a considerable improvement over an ordinary
multimeter for indication of photocell resistance.
3.4 Calorimetric Power Reference Source
A Hewlett-Packard Model 434A calorimetric power meter was
obtained as an RF reference standard for this program. This meter was
selected for several reasons. First, it has several ranges, from 10
w full scale to 10 watts full scale. The power levels at which theMARK 1 MOD 0 squib and the MRK 2 MOD 0 ignition element are expected
to fire should fall within these ranges.
- 23 -
THE FRANKLIN INSTITUTE • L4bmor es for Rmawh and DftvloPnan
F-B1O5
RF InputInput Head
- Oil Flow
Comparison Head
Fl, 3"-2. SIMPLIFIED CIRCUIT Of Np mAOEL 4344 CALORIMETRIC POWER SUPPLY
Second, the calorimetric power meter can be calibrated simply
by substituting an equivalent dc power for the RF power. This dc power
can be accurately measured with instruments whose calibration is
certified by the Bureau of Standards. The overall accuracy of the
calorimeter can be held within ± 2% of the full scale reading when special
calibration techniques are employed.
Basically, the 434A power meter consists of a balanced bridge
circuit which becomes unbalanced when applied HF power heats one branch,
An error voltage produced by this unbalance is amplified and fed back
to the bridge, bringing it back into balance. Figure 3-12 shows a
simflified diagram of the power meter's circuit. The input head consists
of a 50-ohm resistive termination for the RF input and a thermistor, which
is a temperature sensitive resistor in close proximity. A flow of oil
over these elements maintains them at the same temperature. The compari-
son head is constructed in the same way. The temperature sensitive
resistors are in adjacent legs of an audio frequency bridge circuit. RP
power, When dissipated in the terminating resistor, will raise the
-24 -
THE FRANKLIN INSTITUTE •Ldbwdg for Rmdt andmDavuspm
F-B1805
temperature of the thermistor, causing the bridge to become unbalanced.
An error voltage, produced by this unbalance, is amplified and fed backinto the comparison head load resistor. The power dissipated in this
resistor causes the bridge to become balanced. A meter movement isconnected to this feedback circuit and is deflected by an amount pro-
portional to the power required to rebalance the bridge. To eliminatezero drift and maintain stability, the oil flowing in the input andcomparison heads is passed through a heat exchanger prior to enteringthe heads. This maintains both heads at the same temperature initially.
A series flow system is used so that the flow rates in the heads areidentical.
A 100 mv + 1% dc power supply is provided with the calorimeter.When the instrument is calibrated with this supply the overall accuracyof the instrument is within ± 5% of full scale. This error includes
BF and calibration losses but does not include mismatch losses whichmight occur between the system being evaluated and the RF input head.
The accuracy of the power meter can be greatly improved by calibrating
at the same level as the BF power to be measured. In this way thepower meter serves only as a transfer instrument and errors due to dif-ferences in range settings and meter tracking errors are eliminated.To do this a variable dc power supply of high stability is necessary.
The power produced by this supply may be accurately measured by the
following technique.
The HF input termination resistor, with a nominal value of50 ohms, may differ from this by as much as ± 5 ohms. A calculationof power based on the input voltage and the load resistor would, therefore,
be in considerable error unless the value of this resistor were accuratelyjmeasured at each input power level. The error produced by changes in
this resistor may be reduced greatly by the following method. A precision50 ohm resistor (R.) is placed in series with the load resistor (RL) as
- 25 -
THE FRANKLIN INSTITUTE • LAboratorks for Reseadh and DeeloPmoe
F-B1805
shown in Figure 3-13. The voltage
is now measured across the series
combination of the two resistors.
Power dissipated in the load (RL) 0-5
may now be computed as follows: Power Supply E EL
E2 (3-1)=L
EL PGF1. 3-13. D14AM OF CALARAVETCALIRATOR CIRCUIT
and= RL
EL E Rs+---(3-2
therefore
E RL (3-3)
The change in load power (PL) as EL changes from 40 to 60 ohms is plotted
in Figure 3-14. As indicated in the plot, a change in RL of ± 5 ohms
from the nominal value of 50 ohms changes the power dissipated in RL by
less than 0.3%. Therefore, to measure PL' accurately, all we need do is
measure the voltage across the series combination and compute the power
from equation 3-3, using the nominal value of R, since changes in this
resistance produce negligible error. However, since the power is a func-
tion of the voltage squared, this voltage must be measured to an accuracytwice that of the accuracy required for PL" By employing a precision
potentiometer calibrated against a standard cell we should be able to
compute the dc calibration power to within + 1%.
- 26 -
THE FRANKLIN INSTITUTE 1 aborawvie for Resawch and Demelopment
100.0
0
0 99.02
zW
CL
96.5
35 40 46 50 55 60LOAD RESISTOR (ohms)
FIG. 3-14. CHANGE IN LOAD POWER FOR CHANGE N WAD RESISrNCE
4. DETERMINATION OF SYSTEM LOSS FACTOR; SYSTEM CALIBRATION
The total power entering a completely matched RF system can be
accurately measured by conventional means. The RF power arriving at the
exposed bridgewire of an EED can be detected as explaired in Section 2.3.
If there is no significant attenuation caused by the base of the element,
we may reasonably assume that the system loss is equal to the total power
entering the system minus the power absorbed in the bridgewire.
System calibration is, therefore, performed on the equipment
after it is assembled in preparation to fire the test. Figure 4-1 shows the
general form of our firing systems; although the physical character ofthe separate components may vary with frequency, their function remains
fixed. The procedure for calibratinn is as follows:
-27-
THE FRANKLIN INSTITUTE *Labmsmriu for Ruwmad DmI.omme
F-B1805
£w
Wh
I~II Iwid~wI
-28-x
THE FRANKLIN INSrITUTE . Lbodon for Rawrc aid D uvlu
F-BI05
A number of EED's from the lot to be evaluated are strippedof their explosive charge to expose the bridgewire. One of these is
mounted as a termination, with the photocell fixed to respond to the
bridgewire illumination. The matching section is adjusted to give the
best possible match, indicated by a minimum deflection of the galvanometer
which responds to the reflected power out of the dual directional coupler.
RF power is then adjusted to give a reasonable photocell indication
(500 Kilohm, read by the multimeter, was standard for all calibrations
made in this series). The magnitude of this RF power is determined by
switching the generator output into the calorimeter power meter. Thefiring mount and photocell fixtures are then removed from the RF system
and connected to a dc power source. DC power, measured by use of a
Leeds and Northrop K-2 potentiometer, is adjusted to give the samephotocell response as was noted for the RF input (500 Kilohms). The
ratio of the HF to dc power magnitudes averaged for the number of EED's
prepared for this purpose is taken as the system calibration (efficiency)
factor.
Table 4-1 gives the results of system loss factors determined
for ignition elements intended for use in this series of tests. The
data from which these values are obtained are included in Appendix A.
Table 4-1
SYSTEM LOSS FACTORS
Frequency System Loss Factor(Mc) MARK 1 Souib NM 2 MOD 0
5 0.948 o.69630 0.718 0.449
250 0.810 0.4121000 0.514 0.560
- 29 -
THE FRANKLIN INSTITUTE W L orie for Rmah and Dmbpmu
F-B1805
As stated earlier, the same basic system was used for each calibration.
Why then are not the loss factors identical? First, because the matching
system for 5 and 30 Mc is a lumped parameter network, whereas that for
200 and 1000 Mc is the stub and line-stretcher tuner. Second, for any
system we would expect the loss to increase as frequency is increased,
and as the load impedance departs from 50 ohms. Except for the MARK 2MOD 0 loss factor at 1000 Mc, the data bear out these expectations.
5. RF SENSITIVITY TESTS OF MARK 1 SQUIB ANDMARK 2 IGNITION ELEMNET
Both the squib and the ignition element have been evaluated
for sensitivity at frequencies of 5, 30, 250, and 1000 Mc. Tests were(1)run in accordance with the Bruceton Plan, and 100 of the deviceswere expended in each test to insure a reasonable confidence level. The
sample group of each item was taken from a single production lot. Two
large lots one of each item, were selected, so as to avoid the chance
of a wide lot-to-lot deviation.
The firing systems were carefully calibrated for system
efficiency as discussed in the previous section. Accordingly, data are
given as input power to the bridgewire for various probabilities of
functioning. On the assumption that power loss in the base of the
element is negligible within the frequency span of the tests, we may
construe the data to be equally representative of power input to the
base of the element. Recognizing that the loss of the base plug willbecome more significant as frequency increases, we are led to be
moderately suspicious of the data given for 1000 Mc.
Results of the firing evaluations are given in Table 5-1;
raw data and calculations are appended. The appearance of two sets of
data for MARK 2 MOD 0 elements evaluated at 30 Mc warrants some explaft04on.
Those results tabulated at the bottom of the table were obtained first.
(1) For a description of the Bruceton Plan and the manner in which data arereduced therefrom see: AMP Report Na. 101.12 "Statistical Analysis for
a Now Procedure in Sensitivity Experiment," July 1944.
- 30 -
f THE FRANKLIN INSTITUTE - Laborazovies for Research and Dewiopment
F-31805
4~~1 (D44 4 .
0 4 .4 4 -
0 1 01 0
'--P
I31
F-B1805
A comparison of the 50% level to other data which was available at the
time indicated a difference much greater than had been anticipated.
A study led to the discovery that the items used in this test were
inadvertently taken from a second lot of elements in storage. Repeating
the test with devices from the lot reserved for this program produced
the data tabulated in its proper order. These two sets of data indicate
the order of magnitude of the lot-to-lot variations that were eliminated
from our results by working with a single lot.
Calculations made with dc firing data indicate that the meanfiring (50%) level for the MARK 1 squib and the MARK 2 ignition element,expressed as power, should be 0.09 and 0.9 watts respectively. The
results of calculated power to the bridgewire given in Table 5-1 confirms
our long-held belief that the inherent (bridgewire mode of power application)
sensitivity of a hot-wire bridged device is adequately defined by its
dc sensitivity for frequencies below 1000 Mc. As frequency is increased
beyond this point we can expect a significant difference between RF and
dc power sensitivity. The data at 1000 Mc determined in this program,particularly that for the MARK 2 MOD 0 ignition element, indicate that
differences between RF and dc sensitivity may, for some devices, show
up at a lower frequency.
Evaluations at frequencies up to and including 10,000 Mc were
initially planned to be a part of this program. However, technical
problems had to be resolved to allow power at the input of the device to
be measured with a precision comparable to that of the data already
presented. It became apparent that these solutions could not be obtained
under the limitations of the present program. Consequently no firings
above 1000 Mc were performed. The data for the sensitivity of the MARK
2 MCI 0 ignition element must therefore be accepted at face value only.No plausible explanation can be given for its deviation until firings
performed at higher frequencies are available to establish a trend.
- 32 -
-- - -- -- -
THE FRANU INSTITUTE L Abomr r Ruuih d 1 w
F-B1805
6. CONCLUSIONS AND RECOMMNDATIONS
With the completion of Bruceton tests at 5, 30, 250 and 1000
megacycles, we are in a position to draw limited conclusions regarding
the sensitivity of the MARK 2 ignition element. The loss in the base
of this item is considered to be negligible at frequencies of 1000
megacycles and below. Therefore, the figure obtained for sensitivity
at the bridgewire is the sensitivity at the base of the item. The average
sensitivity of the MARK 2 element over the range of 5, 30, 250 and 1000
megacycles is, then, 0.893 watts. This figure corresponds very closely
with the dc firing level, which for a 10 second pulse is 0.9 watts.4Bruceton tests have also been completed for the MARK 1 MOD 0
squib at these same frequencies. The loss in the base of the squib using
negligible, we may give a tentative figure of 0.0910 watts for the
sensitivity of this item. The equivalent dc sensitivity of this item,
0.09 watts, is as we would expect very close to the RF value.
While a considerable part of the original aims of the program
were accomplished, several unforeseen complications arose during the study.
It should be understood, that the attention given these complications
has in nearly every case further extended the present state of knowledge
in this field. These complications can be divided into two major
categories; RF power detection at the base of the EED, and impedance
matching.
With regard to power detection, it was indicated from the
preliminary studies that vacuum thermocouples of varying resistances
could be used to obtain a calibration of the system losses as a function
of terminating resistance. However, while this still remains a pos-
sibility, it was found that the mounted thermocouples represented
impedances that were so different from the actual EED's that it was
decided to go to a different system. We now use for the lower frequencies
-3
THE FRANKLIN INSTITUTE • /osuOr ROOm& dnd Dw
F-BBO5
a photo detector which detects the energy dissipated in the bridgewire.
This system works very well at the lower frequencies, but is valid only
as long as the RF energy is dissipated entirely in the bridgewire and there
is no direct RF interaction with the detector. Both of these points are
questionable at high frequencies.
The more direct and most useful approach to this problem is to
determine the power directly at the base of the plug. However, this is
more difficult to do, but additional study should lead to practical
solutions.
With regard to matching techniques, great strides have been
made. It was determined during this study that system losses are almost
entirely functions of the degree of mismatch between the generating
* system and the termination, and the physical length of the system from
the matching device to the EED. For the tremendous mismatches which
normally occur between a typical 50 ohm system and an EED these losses
can become quite large; so large in fact, that situations can occur which
make it impossible or extremely difficult to obtain a match. A limited
theoretical development indicated that present comnercially available
matching systems were not adequate and pointed out the necessary direction
in which to go to develop more effective matching devices. These devices
have subsequently been developed and proven in use. This then brings usto the final status of the contract. At this time the following items have
been accomplished.
a). Development of precision firing instrumentation forall four frequencies (5, 30, 250, 1,000 Mc) applicableto hot-wire bridged EED's having negligible loss intheir base.
* b). Development of system loss calibration techniques forthis same frequencies, and actual calibration of thesefiring systems.
-34 ---
THE FRANKLIN INSTITUTE .LAbsi for RaurA wd D mlopwo
F-B1B05
o). Development of refined rtching systems for thefrequencies stated.
d). Successful precision firing tests for both the MARK1 MOD 0 squib and the MARK 2 MOD 0 ignition elementat the frequencies stated.
Included in the scope of this contract, but not completed,
were tests at 3 and 10 Gc. Because of the great number and complexity of
problems that arose during the performance of the work documented herein,
it was not possible to perform these tests within the limitations of thisprogram. Additional problems requiring more intensive study than could
be supported by this program were envisioned. It was concluded, therefore,
that these would be deferred to a subsequent investigation.
On the other side of the balance, the solutions to the many
unpredicted problems encountered in this investigation have yielded
knowledge of considerable future value.
Though much has been learned during the performance of this work
still more is to be learned. Means to provide this same sort of reliabletest data at 3 and 10 Gc (or higher) must be developed. Likewise, in
view of the evaluations of "RF protected" squibs and fixes, some thought
must be given to the manner in which these will necessitate altering thetechnique proposed in this document.
Finally, many areas of potential hazard have been seeminglyignored in this study. Some thought should be given to "precision
firings," to determine the sensitivity of EED's to: (1) pulsed RF signals,
(2) lead-to-case functioning, and (3) plug heating leading to cook-off.
More important than the mere evaluation of these functioning modes is thethought of investigating their interrelation one to another. Some evidence
gained in years of RF testing suggests that pulsed signals of average powercontent much less than that required of a dc signal can induce a
functioning of some EED's. To know why and how this is accomplished (if
it is indeed so, and not merely an unwarranted conclusion) is to have
knowledge of significant application to the HERO effort.
- 35 -
!o- W _ I I
THE FRANKLIN INSTUT Ldbors for Ruwch .nd Duum u
F-B1805
ACKNOWIILEDGWNTS
The successful completion of this study is primarily due tothe unceasing efforts of John P. Warren who was charged with the major
portion of the technical support. Aid and consultation was freelygiven by the total staff of the Applied Physics Laboratory, but inparticular, George H. McKay, Ramie H. Thompson.
Major portions of this report were prepared by these men citedin proportion to their contribution to the productive effort.
Paul F. Mohrbach Norman P. FaunceGroup Leader Project Engineer
~ Approved by:
E. E. Hannum, Manager F. L. Jackson
Applied Physics Laboratory Director of Laboratories
-36-
*mp-
THE FRANKUiN iNsTITUT * Lbouas for Rnwwmrd d DiwhiMa
APPENDIX A
SYSTEM CALIBRATION DATA
MARK 1 MOD 0 SQUIB .. .. .. ... .......... . .... Al - A4
MAK 2 MDO0IGNITION ELEMENT. .. .. .. ..... ...... A5 -A9
'.1.
THE FRANKLIN INSTITUTE *Laborawdoe for Resmd anid Dewiopmeri
F-180
b w000000000~anMU~'1
00000000In~w~g
0, R
414 O V 4 0. in r l4
VKWJo~ eror aig
* p***
Al0000
j THEl FRANKLIN INSTITUTE *Laboumiu for Rmda aNd 1D.bWgw
F-D1805
0000000000
fa 088000 0009
r4) r- r4 H r r
N CV.1 * * * * *
;ad i% r a a~awto* al 48848488443
j0 00 o0 00
~A
. ......
THE FRANKLIN INSTITUTE *Laborlas fo Rusawf arod DemloPnm
Nil-81805
~O H 0000000000
Ac- c- l- %0 lH r4 Hr4 H i H r *1 r
~ I J iC;
JUOJU0 00000800004
d 6(4
I.-ID
THE FRANKLIN INSTITUTE L AWoMfWW for Robc aid D&P.
§ F-B1805
,~ 0000000000;
t;~~~ '0 '4 4
0fl
04 r.~
0 000nnnnn~0 nnnn
000000000
a0 N 0ncmO
JOo dm 000000
09 ~ * .(n4 -*&A NO.9 tLe
A4'4 '
THE FRANKLIN INSTITUTE *Laboramvfsfo Rmmeor and DewvoPmmn
7-31805
.00N 0004
0
:1 A. .j
-4~~ pI NN
43o000000
Q 000000
4, N44H
TT., 1551515
A5
THE FRANKLIN INSTITUTE *Labommviu for Rmivrdt dnd DSuIOM.W
F-E1805
tb I E C
0 %04I 000000000
IC)t 0C' H40WI O00 00
0000000000 0 0
m2 0000000000
0(U40 h4 w
soug 0000000000
tO4 W, NO 8 OOwdu 4RC' ll"N.
jozun 0 t q0 0 wHc
(ulm~) HHH HHHHH
THE FRANKUN INSTITUTE - Laboratries for Reseivc and Dewiopm
JI
~ 4r4.-4r
0.
00000 0 00
4
000000 0 0;
8-, 3 z
vH r 4 F1
4 C4 C4e 44 4 C
A7
THE FRANILIN INSTITUTE - LdbmWiu for PANDd mi Dhwgstmf
&4I F-B1805
000 00000000j
& UN %0
* 14 OOOOOOOO~O 0
43C ..... .
0 0 0 0 0 0 0 0 0
CLI C'0
cv cH HHHHHHHoH H
14at R"oc4S-Ccc,0 U -r C - C
4 4 4 4 4 4 N
F98
L: 1U l L Ll U
THE FRANKLIN INSTITUTE *Ldbaboi orir Ruunh and D&whfMii*
F-B31905
000000H0000
o0 00I 0 0 II I0 It 00
1^ U
WN ~ ~ 0N 0 a
0000000000 ?Hm%
clH t Of c n 0 cr \ m
*t4 4OOr-40000000cv 'D %0aD %0
(eurqo) wwwb wdw
o Hw NH H -H toJ Hl
-4~ 4 I 0 H4 0
A9
THE FRANKLIN INSTITUTE Lbmuuu for Ru..&A dw DAhmm,
F-B1805
APPENDIX B
BRUCETON DATA
MARK i OD O SQUIB ............................ B1 - B8
MARK 2 MOD O IGNITION ELENT .... ............... .B9 - B18
B
------
THE FRANKLIN INSTITUTE *Laborazovics for Research and Devdlopnent
F-B1805
B-1 MRUCLYI TEST MIK 1 ROD 0 SWUIB 5 MC ____
u ncik~ m Levels Q~) Fu ctonina Levels (
0-ft~~ I
- _a_ I I I
0 ~ A~ 5- [Ai 7
0 I __ 1 0, = L-A ,1-, 6
I I
k-9 Is 63
24 z _0-g 69
2 2012-7
1 1.~~ ~ 1 26 -
a-.1- 27ff 42 Iim
28 1 - 7-
464
- -- 7 _--------11711Le4a
-1 10 -g I - fl -' A a! i / .BO7
- 1L ,.& 39 &f___fG-..j.
1 09
THE FRANKLIN INSTITUTE •LdbmtW for Rmumh and DuiInm
F-B1805
i j, ;- Proalilit7 Oe*Uau resIlly~ Oarene,_ _ LwLee Lvls Level0 1 o 1A 0 ,. rS fo p% fjo1.!
5 25 .GO..,. GO- /--. -... __6 G 2L /. 3 .-- q/, d. .rnjg, u - d.. .... o@ ,9s
- "4.1-d od- _s . c . , .
Total: N":J- _____ N - .. A NO. j M. H
special Pornmto GO a*~3c - (lot C) . _. .
d -(ogC)(lag)m V is Inter Cn I i N I Intgr
A-Zia2 2
it n Ft= 11r0, P. 19. at givenPm or uD0tB, p. 19, at gin Por X
M - NB-A)/t$11 - c *d.Am/i*% GFrom W for 0 H ywasa S. Ol'Flro Ut for 0 & vers S.
#. 0d( .0"). . Ue oraphs, II& ii. u.. Grphs III & iT.h3 ?s .5, mid Onmp When a? .5, ad 0ph v
*Uh + f Io's; - r " O WhenS < .5 Wh < .5
n ""Vulid fw N ;0.1 sIT o.thsWi . ow-s fut bu t l i' C pst lb.10.1, 3. atikti..l Awlywis S A Nw L (Y) 12 (T) 1/
- is S"itiwity b imat" 2 + A2 2 12 /.2.1
uY.1P.. "b ,-,i t c< , kx t ,For we! s Far ox!'B
"A Z. -, /I /,r.9 ,
. +. " /, Crop ) ( ." t Co ) .- +.pc T (..) ( Cai n) - * + TsemffdariY tatistic. w.ZSW21o6 units vna1.Lo ite
I i Jf L(1,1_) (fX9 Cut) - ~r (_Z_) (foCoei) - mo-kq- Y- A&?Lis units .g&fj* unmits,
B2
THE FRANKLIN INSTITUTE *LAborawries for Research and DeteloPnm
F-BJ.85
B42 BRUCET4 TEST HK 1 MOD 0 SQUIB 30 XC __________
*F .k~n Leel (0 Pi"I1"j Levls (a)
3 '4/ 63
- III - 7
x ISI
O&Pi40.4 S6
17____ 1 1V 11
-~ ~ 6:::.nA
e~33
THE FRANKLIN INSTITUTE - Labortoe for Rescalch and DeM1lopmei
F-3,805
- - ProkebMlty Oot1ins PI MAMYit og1demoo0 a. L.A eavel. Lewd, i..i.
0o VO - % o '
2 'L 0.Z~~Q . . 2.,7~ a.?. 26 ~____2--31
6 16 ____ s. oa/.5 k?. SS 03 a.049C-2 d tl,3 (-0?7 d0.3 G
Totas: N. ~Nj .o ff_ Komza... .- ff~ w-2... 040.L.KOO
(16Ca~ (log V, - Lj o_________ ________ -
whe XlsC~ InI j=l JIMNU.UftSr a hnNii nee
imiy StatisticsnAh N in am ue integer.. n when Nt isM Intager
B - I im FroW. ,p. 19, at givot P or 10Frm*I, p. 19, at givert P or~ I
M - ( ?4B-A')M Oo forG & H ewm. S. *From OP for 0& verusS3.o - c 4- d (AAN t* 0 s rpsII T s )oh I Ta -1.62 d (M # 0.029)j r.Gal !&X.U.OeinII&X.
WhenS 2 .5, maMGraph V We 3 .51 mid Gaph V*Ue#fra'-f OWhn S< .5 Whn S <.5
**%lid f.~ m ;I 3 IV. ethelveotem-________________"Ilt lbketm Poewtl CA Uqet 1b. )(Y
j 101.13~~~~~~~~~~. Sette Aelyms fee A lb 3Fi1wwL() .aia.~m 1PFmee~ is 8"esitivity bpetimmts h2 2b 2 1/2 2.212
July, 1344) Pilo N,. lb-i
For 1'sO Freeza,'kxVef_ Tk
If O,4-3 fti- L.0 9A i
Aix. -D Mo 2_2145 Oig dit, =2 .~jo itsS3eaoder StAtiftics
%D +9Ok %A,-
THE FRANKLIN INSTITUTE * LAbovatuvrs for Remvach and Deiopmeu
F-B1805
B-A BMMETMr T~EST MK 1 MOD 0 SMIUB 250 C _ ___________-
Wes
W I 7
1 e- -1 1 9- - -
1 16 1eu 9-silIN IW464
-- - -- - - 6.915 - JAM
66
X. ~ 1- 6. 6. IIO
27
- - - -- - - - - - _ _ _ ____
-4-9 26 X.76
-e -- - -09_"A - , 2930 , -77
- 1.- Ab - - -- -- i~ ~ I '- IIII /* 32 A -z : ___ 62
FA- &3 _k __
I 1-- - - .Z t '0, 3 6
I - 396
-- -- -- - ~ 42 -~ - 2J6, IA AG 43 loallyz 93
- - - - f l -1-164 94.
._1 1 1. - 45 1 1 Itv 1 46 A
: - 97
D5
THE FRANKLIN INSTITUTE *Laboratories for Reseawh and Delopmenc
F-31S05
jT vaU1ty Common.. PnfbbltY oommom...%bes U Lauel Leve Lows". Lowel
...±L .~ ILL P ~ .Ia _ -4 o C v # k * L a-~ a - p ~
4 ..L.. 6
6~ 34~@ ~ .- 4f1
fpecial Pmrsomrs a n (1) 2
a * (log C) L/AfL __4Z_ ___
d - lg0* (lo Vi - '61 C)D na when x is M Integer %ba N in j Integer
n 2A N is MMnteger L21*,,X I gf Inege
0 i.sm' Pot Wn, P. 19, at givnPw or IFrea, P. 19, at SIYM Por
1. 62 a (3. 0. gooJ:~ Gueapho III A IT. ue Graph In A IT.Pz1 WhnS k.59 and Gaph V WbS 2.5, ani Graph V
Nb # fo Ys o a* When S < .5 WhenS < .5**%lid for H i $-.3 only. otbari.. s-
Soft Imus. RoperoI UM FPP .101.1114 Isuttitil Analysis for A Now£~4n~I~ (T) fgafdMUMCM1 MPrmodave is amwiis bimenwts, 2 & 1
July, 1946) file. b-1 w t! aFor 8ot Far, zal
Seondry Statiatica
W M -dc/,00d (a) (to Gotm) - lmko- y ()(hcow) - a-py- y
[N e. *; k - Z4r41~zL.. units Z.flJtZ&los wdto
TIAL9
B6
THE FRANKUN INSTITUTE U Labmodav fr Rmomh mW Dmwloim
MtN M HK 1. )DD 0 SQUI 1000 No
- ; ':i_I 0 92 a I I
0 -_ _ _ ! -r,I 1.3_7865
0 4:
t.xL .91o xi xA3 W 67 1 1 2- . It--
x I1.0222D cM --. W
" ;...Z . - ' - - ."0
--6 Is :.05 is I
-0 !- 0 10
- 55
1 0 - -: 06
0.- Ml -0y R 1 t : :
0 - oTo
B? 10.56778I IL 0.943 W I
0~~1 -
7 ,.'
0 . Ax0 0. .99 orio~
0 1- . .0
Io -1 0 1 o. n
Z~ ~ ~ ~~~X I.6 ___" ": ..L~4 0 1 L t942, oO
0-1W
THE FRANKLIN INSTITUTE * Lbom Rnv mid De wl*omm
- - - 1,.Wlty 05An. Psbam1ty common"i l I % % It-wl I'wl LoaLWL a
i- I W' -. a -=- a--- 90 p -
1 i ,6 6 IV
l II 2DO 1L. I!? e O( 2- .,b# !.L zA-
3*Z * 2 P3m0 1 p. 19, Go. .1 rm i, p. 1 , oti. P.
u • ( NI.A),tV
's a a.SW)~! Insp 1-f a* aue9235 3 a w xxxl 0x.01
N ma ..5, af h IV hiS .5, N.phV
fps ho ms 50O. X3. .~~ 2 Mso N 4-
Smi Iee bhim (amibu b.A t S' b Ti-l l
,.. .713 * .OiO6i. 07. 0 IP., i 6m 9 at ie I , . 1 a
d (AM.... MW 00 use 2 01)eII a a.y* use 1 O ust)e II N
* k .,ad , When3 .a a --,,
202 ') .U 0r). - (I) b(=it.). a-y -
.a n s ,, s .- , .2rn ./02..&11/2,
a !LI
38
THE FRANKLIN INSTITUTE * Labommatois for RescaAd and DovioPrnen
F-B1805
B-4 MBMT TEST MX 2 MOD 0 T 5 XC -
a -xI__I x
4 54
- - f~ 1 0 -0 1 1 ~ ~ h 161
Zi I I (1o
I- I& _ _ _ C -4 I x- 4s]--
-- I 22 &. 0- .J 9
24 - 74A.AW 6
27 I-!---
1 73
-7 3 AW261 1____
~~iiiI 32 ------ _
1 ID77
- 30 1- -- so-
-r Zw 40
1 6
I:: x 1*OV 391AlI 0 1 - - so . 000-X
[ I I x I I1 0B9
THE FRANKLIN INSTITUTE * LabmarW for Rom* and DNuOWm
F-BI805
i i ProbitY Oetitdow Poaifty , fido,,L.a.l Level Levels Level
4J 16, 6 - - - o. 2O3. 4,-.i1 - , ,,L , k1®.L .t
. 5 d- GD. 1,. ._ d- 0.03 .3L a i
36 - -- S If o G /, S ) - o CL -/Toal:N,67N.49 N HO N-La /O H
*msisl Paramters R * ~ 2*L.2~~'' ~O N2-- (lo 4)1.. -____A
d (og ) (16 i :D - whn XIniminteger (Dn - I enN is integer2 2~itee
PrimA Statisticsn - = when N in W integer n " 12 when N is Winteger
B I it" From B1*, p. 19, at given P or 10From BR*, p. 19, at given P orI
m NB-A')M'c (d mAs)* )Frc .BR* for" & H vesus S. O)Fo BR* for O & H Yarns S.
• 1.62 d (N . 0.)29 F- Use Graphs III & IV. Use Graphs III & 1V.- 5 ( 0 N-I.) When S 2 .5, and Graph V When S 2 .5, m Graph V
t.+ f r "ae,;" - tfr ',** When S < .5 When S < .5
~Val d for N )10. samly. otbsroifs ass-
101. IA.s sta t ' ( bml sis fwAr Confidence Interval CT Confidence Interval (T)
in.3 hs itivl AssIwris Lw A2 ,.2 21/22 & 21Paseitd, 1 ) Il N. M il rt If.
JsI~~~. k. Finb a- A~aa T - k.For 'os' For 'Oxs'
A_ . - :,r /,/ o.. o . II // 4 -_~B.- 71 L ~ ..57
hi -A. SILAO6 0. ~ J.ga~u. Fl Cloulaticlis" .IV14(.LCaif) - a.,k* T Lo) - Y Ccnf) -. 1.. ° ./. ,. 4 lot, t .__._,l___z .,
,,.14 ' *. .. ,. .,'
" .. , "+ (,) .Cf) - a Coe) -k,-T
B0.6 0
Aotiloo a 14-2t
1ITIALI S 6
B10
THE FRANKLIN INSTITUTE • Laboraovls for Reseach and DemloPnent
ERCETON TEST MK 2 MOD 0 IGNITION ELDNEM F-BI05
I | ,------------__ eIt Testen
___ W WIN___S
- - - - - - - - - - - ,,,,L... .. - J Si. o_ 921%%
Ix I 1 . 2 1 0 1 _0 1 1 .095 a 1I 16332 .1 0 so
x I oL____ oo__ L-- - I-------25-7--02-
- -- missed .2-- -.~ - - - - - 4 jai_ x_ I_152L 0;$
SI -I
.- -I? *942L...Q.- ~ ~ ~ ~ ~ ~ -- -- 00 3 .094.3
U&S56 . !- - - - .358085 .100"o-- missed .170
.3792101107
- -- 1003 0__01_?2
33-------------------------------------- J*Q37___ 34 74
S.0,49036 1S*451 -- 32 1--
z 3550 .10 ?9608 " - -- - --~.077 2 * 4 -- - - - -~.10 2 61
.02 .199 ia .10,67xx L 7k .:L0 04'
.2 28661 .---.10 6 I
3* ----- - U
- .09
, L .10 6?- .12
2 .... a 49
, 5 5 I I 26 _90
iit l- .09 94x- .6 . 95
M _ -- .099 i2j _ _ _ _ _ Bl
THE FRANKLIN INSTITUTE • LAbmumviu for Rumn& ad DwuommF-BIB05
- - - Probability Cauidwe Probability ConfidenceLelas Level Lewls Level0 o 0 1. 67 "5 .9 z%. 90 1 4 .
1 1 14 4 1.79 1004-. 0.1 l - -.
L4 L1 1.91
9 ~ 28205 a- .28708309 o4 is 0 4 2 .20 a. *0282 1*282 _ ..o 29 k?®-S, IS o. o. ._ o. 2 U G.O,-
.., .. 1 - .~_ . O 0. -1.012 d _ - /36 2 102 4144::~ 0 q
TA lnt n N- 51e4 N iN 100 Hntger
3B- 1t3 ®Prou l34, p. 19, at givet p or I?wa i 3, p. 19, at ivLenP or I
tOc (AA IN-A )" . C1 Frm DR* tor 0 & H versus S. ( Fro. 3* for" 0 & H versus S.
cr -1. d (M* 0 gW~ Ue Gte, ba Il &IV.U.. Graphu III & IV.
n1.1.Ud0.O m91.8496enrapi. iH2 AZ41
lflen S >.5, aM~dOaph V When S .5, e,,i Gaph Va (le '.s(e -m '0 When < .i When < .5
eels y b st ws' +U p~
1 .bad eis Esie lie ing ( tg Conf" en i nte gerA-isni 2. 2et s b .
a -.," I* its From Me p 9,atgve PoF o m B- % p./ 19 tgvnPor
l. 1H) ... . 7 When S k , h.5V
**" f . m al.e6 aM,--WnS. _,5ohen_ ,<_ ,
9 * .542899 .53Ao M f MA* .884 1.873
(2 B Conf) mkla Y (ky) (1.k Coo+)
- 139 0 m wite C.7I 7 1gmt
W. N 2.4946 watts t- 17%.28708 (03)(JCoe) - a-y( - T i.)(r Cue) - 3olkpO - Y
*0S _ :LI6_o visits ~ * 209iqits
a. 1:9360 wins
~2-1-62 Im__ _ _ _
j _____B12
THE FRANKLIN INSTITUTE a LAomiodw for Rmawk and DmAjimF-B1805
Bruceton Firing Data For MARK 2 MOD 0 Ignition Element - 250 Me
itmiaLeve (V Fuwtioninq Levels W
U ],-- lf 'l .I . /I s
'4 w Wlli~~ \Aa 4z 16 6--/
N- II
Q---------- - a 'ZJ -- iZ ,/ a
-Y3
-L Y
-- - -- - -- - ., .J22
- - - - - - - - * ,/ I 'iF, * .l 62'a
t X , ItU ,q
Is_____ I R5
{ X,1 f ~ , ,/ '1 3. . 6 3 t
_ __,4 ,04
.1 _ . Go__ 66
X #a/ # IN -" LO 0?
- - - - . o#3i;'t"o *4
-- i&2LLZ,' lo #a 70
- 1 ,#- -- _ _ _ 21l * 7/, 5'- *,a.Lt,1 ,#Li r°fiT.~ i
I- x 1~~~~~z 25 75
- X og O.l7 • ,,i sii X,.L I - t~ i , i Il
___ 23 73
- - -- ,- il - , ic
130 1 , i S, o l
x I.~ l#l 3, V3 I , l 3 I^t 1 1
3- - -I #
/a./34 - X.-1 I I .1.7/14 /aA 4I
i/ AA .- - I , - .
tl - ,, '7 -- - - - - - - - .6
WO as
* *0 ,!- .,.~ l t ,, . i lli Iof
! .!0 to0
I I I - M .1- - '7
--- A- -- ____UA!?- 1' :)tai0 ~
I_ A'idE! -r RAI
B13
THE FRANKLIN INSTITUTE - Labomawks for Researh and Deivwopmet
F-B.805
I I 7 . % - robsUity fa.dmoe Pftberm1t7 oo1dmseP % - ,_ _ L_.a_ Ll L-wa_7..
S.17. o ia06 .O0i ,,,-.4 , L .Z l ®. ,®
/ . t~k3 1 9 1 , a."Y " o,: &C .. , .9$a '.€"
4.-~ 16 "goo 1- ?~f' 0.,/2 2 .644~fl ~ i__d- .a65 0()- ,O_ d- 6,0' G01- /620
- .'s o- _ 110.1 £a Z.. o3.1 40
Spe.ial P.' ers CD. x . - i.. ).
d (lo C)-. (lagC). nijuhefXis intr n Ijm N is integer
Prim"y statistics .1 1
WWI
A-Zin 2 ~ l*2 nijegB -Ir its FromW*, p. 19, at give P or IFroEn, p.1 9, at gven P or I
ii - ( NB-A)/Wc #i (A44 tB-s)*I O)Fr BR* for G & versus S. V ram S for 0 & N vems S..2 d (1 + 0.0 Oraph@ I & . Ue Grphs III & iv.
)1 When S ? .5, ead Craph V When S 2 .5, i Graph V"IJe Vfar Wen e he < .5 Wh nS < .5VelAd f.. m ) 0.3 8 ly, tbeei. w-
mit 'baeta RoomI (Ai r a 1b.M. I, Uta.tlml ieiei, f- A N, W (y) gUdfILjlwia (Y)F eeiee ej+ t L\2 (2 . W2 L/2Je ly, 1944) i Nle .l b - I IY
For 0e' Fm Is,@
i * *
m ...A9 4 .0 ) ((A cof Elsa - (0%_) (52 cm') .a-S O +-
N %!i
B14
THE FRANKLIN INSTITUTE • L bonnohies for Reseamdh and Developem
BRUCETON TEST MIK 2 MDD 0 IGNITION ElMVT 1000 lk F-BI805
6 -7 1 1 I
n- 0 -4ROW0 9
I 0 26222 |0- .10216SV+ 1 01 0,57752 4 X072 . .1 4
ma fl _ - .i00. IC
11 -- a FI I .20--892 7 X.2231.
1 01- _ 0 .0 9
1 0 _ x
0 .m7166 0 .
x 0.30865 ,4 0 .09063.3
0.49039 14 0 0
<n - Ix 1 0.33761 .91o
- _ 0-* r o0-16572 ,;x o.4.0852 .~s*
0 _to 0 .-100?, ,6 097 TO
X 0.0510M t I O. I02 .1sIx 0 1 13 1 - Lmq ?3
01 0 1 1- - 00 7
Ix 1 0.30358 -s X. I.95 .0 75I -_ IW29 790 - 0- .100 79
-1 1 , 1 0.06355 .100 1,- 0- 1 0.1264 .095
I xI 0.162 I I - .05 3
0 1 . US------------------___ W4 100.183
0 1 --
1 X 1_ 04 92 T 0 _ 1 l-00 as
0 ____,0
101 O. 58 1 0. 3 0 . n
o0 .'3:7, o7 - -- .i0
r- I _ _ _-7 _-JLL
0 -1 1C 4 2~3. 1' - 0 - - _
• .'~sq 441 1 0. M 9 MZ 1140 1- 40 0 .10.2197 4 0.- 1
10 1 - 41 1 i 1m.0.37996 9__ n4.
-59536 40 8 o 1 - .00- 4 1 .40252 0-0.89 X - -_
0 11 8L5 1 1 , 25X O O4D Ima 2II151 I IiM a25 0 u 10 - 25
B15
THE FRANKLIN INSTITUTE *Ldboraeve for Resoarh and Developmen
F-B1805
£ 1
~ - - - lwbablUty 0uida4ii P1,01ability uU sa. % w~tt ___ Levl Levels Lord
7 __ .. L 0 1.19 P% a .9...... Za 902 1% u.... at1 1 22 1 1.27 100.1% .1 0%
243 r22 1.392pugj2 2ot5Th 1.49 3&
4 16 0 2 1.60 .091 g W -128 a. L i2
4 36 14 - -69 02 .17 3 1 .0& 0 2~ LL
0A -(06C! -03 t.~m x 1 m teo flumD . h N. M Integer
V *.. .diUs Af~eOEem te -Mi s m.ee
0.1.U ism W~ O himSWP 9 tSv fhimS W, .19 at X~vmP or
*?svo W* fo 0 A. mlp wem 3. __________for___&______ _3.
uss, iauI in a. iv se.aeII T
-%wfwab . my aied o. vlm 0.5 1&Ah < o .5//*i leum 0 11 w or U4...1 ft a.
, M Ro~aZle6 rwvie -kx V._k T k
No_ _ _ _ _ _ _ 1.63 watts * / O
.373 (U)Q Cff am u- ( .S oe316
a j2__)j dopgwt
ITHE FRANKLIN INSTITUTE *Laboratovie for Research and Dvom
B-5 BRIJCECON TEST______2_______0_T 30___
'. ~ a
je I I 'A A 4 1 4
o-C - - 1-09 6
- 591
7-'
111 1~ 14 x ___
S175 67
77
- __ ci , ?7o--:?o
23 -- ?37
26 dol -
- -0 3 4,
4(, AX:7 -X - --
- -3 0 so ~6
443
-is & _
36 -,z& ±0
92
054 9
THE FRANKLIN INSTITUTE * Laeovws for RemawJh and DeveIopnwU
F-B1805
f is - -ma Probability Comfdemee ProbabilitY 0atumee' ~ L*Vel LOW11 Levels ee
1 ~ ~ A 1x /f 10 lo-%AL %.
I-- I--L AIL G..2bJ s1ddJ13 0.
Totl9 afO~ .,.13 k . 3,01 .o,3S - 23L......
*IParytamtrsc (D . 2ben 1.m41ne n.I~ rhn L
3 - 7 ism FProm 3*, p. 19, at given P or X 0 From 3R*, p. 19, at given P orM *(NB-AS)7E
d (A/M t%)* Frm * for 0&H versusS. GFrom Btefor 0&Hersus S.Wrhen S .,etapVWhen S 2 .5, and Graph V
*Um t" *a; - fg VOnS. When S <.5**valid for m al 0. 3 "ITy. Oh c, am . Whn <.
a.l mrbums lbpt' (A %bdt lb. w3xjmmQ()CniestIera(Y101.13, bSatistical Anlyisa for A Paw fli*Zi 1 J (1/) 1/2.~n C'Prsd", AuUmitiviey albeut.'e H%2) ~ .
For *'os Fr It's" -
m - 4vI 10 6'gl_________
SwnayStatistics 0- l ats2.u nt
M. -14. Af0b)(10j Conf) - a-kpaY - 1 J (10 Cue) .- M~-O-y
B18
THE FRANKLIN INSTITUTE *Lbabouoin for R~UiaWh drd DWA"Wmu
F-BIS05
APFENDfl C
- I DRI SIMULATOR TESTS
c
THE FRANKLIN INSTITUTE • Labomu INr R mmh dnd D wA
F-BS05
I C. RF TESTS ON SIMLATED NRK 2 MOD 0 IGNITION EMENTSFROM DENVER RESEARCH INSTITUTE
During the performance of this program the Denver Research
Institute delivered to us 28 simulated MARK 2 MOD 0 ignition elements.
These simulators consist of the standard MARK 2 MOD 0 assembly complete
through the bridgewires, with a thermocouple substituted for the explosive.The elements were divided into four groups according to the spacing of
the thermocouples above the bridgewires. The MARK 2 MOD 0 element has
two paralleled bridgewires. The simulated elements have two thermocouplesdesignated A and B - one for each of the parallel bridgewires. The
spacings previously referred to are: very far (VF), far (F), medium (M),
and close (C).
JTests were conducted at 5, 30, 250 and 1000 megacycles. These
tests have shown that the simulated elements have an increase in
thermocouple output as frequency is raised from 5 to 1000 megacycles.
In some cases the output is considerably more than doubled, over this
frequency range. At 5 megacycles, however, the output tends to level
Data for millivolt output versus frequency are given in
Table C-1. The simulators were all tested in the same RF systems used
for the Bruceton tests of the complete MARK 2 MOD 0 ignition elements.
Input power was set at the 50% mean fire level.
Curves showing output as a function of frequency for a limited
number of simulators from the group of 28 are plotted in Figures C-1 andC-2.
A few of the DRI simulators were also tested with dc power
applied to the bridgewires. The output was a linear function of dc input
Figure C-3 displays this relationship.
C-l
IiTHE FRANKLIN INSTITUTE e LdbomwoW for R.oi and Dmwagnm
F-B1805
Table C-1
RF TESTS AT FRANKLIN INSTITUTE ON DRI SIMULATEDMARK 2 MOD 0 IGNITION ELEMENTS
Thsrmocounle - Millivolt Outnut5X 30 me 25Q0 . iOOO Me Between
Element Thermocouple Thermocouple Thermocouple Thermocouple ThermocoupleA B A B A B A B a
761 5.18 1.88 5.3 1.6 7.3 5.6 11.0 5.2 very far763 6.55 3.90 7.1 1.4 9.7 4.4 16.2 10.0 very far764 5.9 5.05 6.0 5.3 5.8 5.2 8.65 9.0 very far766 6.0 8.9 6.4 9.4 8.0 8.9 8.2 10.8 very far767 cut open 7.8 6.6 9.5 9.2 12.0 13.1 very far
715 6.10 5.40 6.05 5.7 6.0 5.6 3.5 3.28 Far724 0 9.6 0 9.6 7.9 10.0 0 13.7 Far728 2.65 2.4 2.55 2.5 2.7 2.7 3.35 3.1 Far729 3.6 3.7 3.4 3.45 4.0 4.0 5.5 5.9 Far730 3.44 1.55 3.22 1.44 3.6 1.8 4.3 2.78 Far733 6.00 3.2 5.81 3.1 11.0 4.5 11.4 9.0 Far736 3.2 4.0 3.30 4.05 4.3 4.2 6.21 5.0 Far740 5.58 3.3 5.70 3.35 6.4 4.1 7.66 8.1 Far716 6.6 8.1 6.3 7.6 6.7 7.7 7.41 10.8 Medium717 12.1 10.8 12.0 11.8 18.0 15.0 16.0 0 Medium719 7.55 0 7.8 0 8.5 15.8 10.2 0 Medium732 11.1 10.7 10.8 10.8 10.5 13.7 10.3 16.0 Medium734 7.2 6.43 6.8 6.3 9.4 7.8 8.0 13.0 Medium735 0 4.25 0 4.7 13.5 5.2 0 9.8 Medium742 8.6 2.9 8.6 2.9 11.3 5.2 14.4 10.9 Medium714 7.7 3.6 7.6 3.75 10.8 7.0 13.8 7.4 Close718 4.7 5.45 4.8 5.10 5.0 5.2 5.9 6.0 Close721 13.4 10.5 13.2 3.55 16.6 9.2 18.5 10.0 Close727 Cut open 0 0 0 4.0 0 11.0 Close731 7.05 0 6.8 0 6.9 12.6 9.0 0 Close737 3.15 6.2 3.15 5.9 6.0 8.5 Not Measured Close738 8.6 7.8 8.3 7.8 11.5 9.4 Not Measured Close741 9.1 7.9 8.8 8.2 11.5 11.5 Not Measured Close
C-2
It
THE FRANKLIN INSTITUTE *Laboratories for Research and Detwlopment
F-3805
Fig. Cl. outut (mIllivote) Cr! Simulated Amr 2 moo0 lfomme o~meit16, As A Functlon Of' Frequency - Groupl
1412-
910
7669
Fig. c-3. output nillvoi pr! Slwubated IlK 2 JEw o Ifoltin iemntE1- As A FuncoG Of owem -nu Trop BZdeWr
766A
76A46------------- -
As Fnc~mOfDC POWR INPut roordg W
103
THE FRANKLIN INSTITUTE ° Laboratovs foy Rewah and Dodorwnt
F-B1805
DISTRIBUTION LIST
Commander Officer-in-ChargeU.S. Naval Weapons Laboratory U.S. Naval Explosive Ordnance DisposalDahlgren, Virginia Technical CenterCode: WHR (2) U.S. Naval Propellant Plant
Indian Head, Maryland
Chief, Naval OperationsDepartment of the Navy Commanding OfficerWashington 25, D.C. U.S. Naval Underwater Ordnance StationCode: OP-411H Newport, Rhode Island
Chief, Bureau of Naval Weapons DirectorDepartment of the Navy U.S. Naval Research LaboraioryWashington 25, D.C. Washington 25 D.C.Code: C-132 Code: 5410 12)
RAAV-3421RM-15 Commandant of the Marine CorpsRKMO-224 Washington 25, D.C.RHMO-235 Code: AO4CRM40-32RWO-33 CommanderRMMO-4 Pacific Missile RangeRNMO-43 P.O. Box 8RHMO-44 Point Mugu, CaliforniaRMMP-343 Code: 3260RREN-32DIS-313 (4) Commanding Officer and Director
U.S. Navy Electronics LaboratoryChief, Bureau of Yards and Docks San Diego 52, CaliforniaDepartment of the Navy Attn: LibraryWashington 25, D.C.Code: D-200 Commanding Officer
U.S. Naval Ammunition DepotCommander Crane, IndianaU.S. Naval Ordnance aboratory Code: 34White Oak, MarylandCode: ED Commanding Officer
NO U.S. Naval Ordnance PlantLV Macon, GeorgiaTechnical Library Code: PD 270
Commanding Officer Commander Naval Air ForceU.S. Naval Ordnance Laboratory U.S. Atlantic FleetCorona, California U.S. Naval Air StationCode: 561 Norfolk 11, Virginia
Code: CNAL-74'+BCommanderU.S. Naval Ordnance Test Station CommanderChina Lake, California Training CommandCode: 556 U.S. Pacific Fleet
4572 c/o U.S. Fleet Anti-Submarine WarfareSchool
Commanding Officer San Diego 47, CaliforniaU.S. Naval Air Development CenterJohnsville, PennsylvaniaCode: EL 94
a
THE FRANKLIN INSTITUTE * Labowonu f Ramch aQ Dw m
F-B1805
DISTRIBUTION LIST (Cont.)
Ccmanding General Commanding Officerheadquarters, Fleet Marine Force, Pacific Harry Diamond Fuse Laboratoriesc/o Fleet Post Office Washington 25, D.C.San Francisco, California Attn: Harry F. GodfreyAttn: Force Communications Electronic
Officer Commanding OfficerU.S. Army Nuclear Weapon
Commander-in-Chief Coordination GroupU.S. Pacific Fleet Fort Belvoir, Virginiac/o Fleet Post OfficeSan Francisco, California DirectorAttn: Code 4 U.S. Army Engineer Research and
Development LaboratoriesCommander Fort Belvoir, VirginiaSeventh Fleet Attn: Chief, Basic Research Groupc/o Fleet Post OfficeSan Francisco, California Commanding Officer
Picatinny ArsenalCommander Dover, New JerseyNew York Naval Shipyad Attn: Artillery Ammunition andWeapons Division Rocket Development LaboratoryNaval Base Mr. S. M. AdelmanBrooklyn 1, New York Attn: Evaluation Unit, InstrumentationAttn: Code 290 Section, Building 352 -
Mr. Abraham GrinochCommanderPearl Harbor Naval Shipyard Commanding GeneralNavy No. 128, Fleet Post Office Headquarters 2DRAADCOMSan Francisco, California Oklahoma City, A.F.S.Attn: Code 280 Oklahoma City, Oklahoma
Commander Commanding OfficerPortsmouth Naval Shipyard U.S. Army Signal Research andPortsmouth, New Hampshire Development Laboratory
Fort Monmouth, New JerseyCommander Attn: SIG0/EL-GFNaval Air ForceU.S. Pacific Fleet Commanding OfficerU.S. Naval Air Station U.S. Army OrdnanceNorth Island, San Diego, California Frankford Arsenal
Philadelphia 37, PamsylvaniaDepartment of the ArmyOffice Chief of Ordnance CommanderWashington 25, D.C. U.S. Army Rocket and GuidedCode: ORDGU-SA Missile Agency
ORDTB Redstone Arsenal, AlabamaORDTW Attn: ORDXR-R (Plans)
Office Chief Signal Officer Commanding OfficerResearch and Development Division Office of Ordnance Research,Washington 25, D.C. U.S. ArmyCode: SIGRD-8 Box CM, Duke Station
Durham, North CarolinaAttn: Internal Research Division
b
THE FRANKLIN INSTITUTE Labomovi for Rmach ad Dro/owt
F-B1805
DISTRIBUTION LIST (Cont.)
Coaanding General CommanderWhite Sands Missile Range Griffiss Air Force BaseNew Mexico RADC, Rome, New YorkCode: ORDBS-G3 Attn: RCLS/Philip L. Sandler
Commanding Officer CommanderWhite Sands Missile Range, Air Force Special Weapons CenterNew Mexico Kirtland Air Force BaseU.S.A. SMSA Albuquerque, New MexicoCode: SIGWS-AJ Attn: SWVSA
Commanding General Director of MissileU.S. Army Electronic Proving Safety ResearchGround Deputy IG for Safety
Fort Huachuca, Arizona Norton Air Force Base, CaliforniaAttn: Technical Library
Commanding GeneralDirector of Office of Special Air Fleet Marine Force, Pacific
Weapons Developments MCAS, El ToroU.S. Continental Army Command Santa Ana, CaliforniaFort Bliss, TexasAttn: Capt. Chester I. Peterson Commanding General
T S Control Officer Strategic Air CommandOffutt Air Force Base, Nebraska
Headquarters Code: DOSEKAir Research and Development CommandAndrews Air Force Base CommanderWashington 25, D.C. Headquarters Ground Electronics EngineeringCode: RDMMS-3 Installation Agency
Griffiss Air Force BaseCommander Rome, New YorkAir Force Missile Test Center Code: ROZMWTPatrick Air Force Base, FloridaCode: MTRCF Armed Services Explosives Safety Board
Department of DefenseDirector Nuclear Safety Rm. 2075, Gravelly Point, Building T-7Kirtland Air Force Base, Washington 25, D.C.
N.M. ResearchCode: AFCNS Commander
Hanscom Air Force Base, MassachusettsHeadquarters Codes: AFCCDD/CCEI-IOgden Air Materiel Ares Attn: Capt. LongHill Air Force BaseOgden, Utah HeadquartersCode: OOYSS Armed Services Technical Information
Agency (TIPCR)Commander Arlington Hall StationAir Force Missile Development Arlington 12, Virginia
Center VIA: U.S. Naval Weapons LaboratoryHolloman Air Force Base Dahlgren VirginiaAlamiagordo, New Mexico Code: WHR (1O)Attn: NDBG
c
VTHE FRANKLIN INSTITUTE e Labotia for Rmanh mn Dwlogm.at
F-B1805
DISTR IRTION LIST (Cont.)
Commander Aerojet-General CorporationField Command P.O. Box 296Defense Atomic Support Agency Azusa, CaliforniaAlbuquerque, New Mexico Attn: Myra Z. Grenier, LibrarianAttn: FCDR3
Grumman Aircraft Engineering CorporationU.S. Atomic Energy Commission Weapons Systems DepartmentDivision of Military Application Bethpage, Long Island, New YorkWashington 25, D.C. Attn: Mr. E. J. Bonan
Defense Research Staff Jansky and 3ailey, Inc.British Embassy 1339 Wisconsin Avenue. N.W.3100 Massachusetts Ave., N.W. Washington, D.C.Washington 8, D.C. Attn: Mr. F. T. Mitchell, Jr.Attn: Mr. G. R. NiceVIA: U.S. Naval Weapons Lab. Lockheed Aircraft Corporation
Dahlgren, Virginia P.O. Box 504Code: WHR Sunnyvale, California
Attn: Missile Systems Division, Dept. 62-20Naval Member Mr. I. B. GluckmanCanadian Joint Staff Attn: Missiles and Space Division, Dept. 81-623450 Massachisetts Ave., N.W. Mr. E. W. TiceWashington 25, D.C. Attn: Missiles and Space Division, Dept. 81-71Attn: Staff Officer (Weapons) Mr. R. A. FuhrmanVIA: U.S. Naval Weapons Laboratory
Dahlgren, Virginia McCormick-Selph AssociatesCode: WHR Hollister, California
Attn: Technical LibraryAmerican Machine and Foundry
Compa n :
Alexandria Division Midwest Research Institute1025 North Royal Street 425 Volker BoulevardAlexandria, Virginia Kansas City, Missouri
Attn: Security OfficerAtlas Chemical Industries, Inc.Reynolds Ordnance Section RCA Service CompanyP.O. Box 271 System Engineering FacilityTamaqua, Pennsylvania Government Service DepartmntAttn: Mr. R. McGirr 838 N. Henry Street
Alexandria, VirginiaThe Bendix Corp. Attn: C. M. FisherScintilla DivisionSidney, New York Sandia CorporationAttr: R. M. Purdy Division 1262
Albuqaerque, New MexicoPermits Powder Company VIA: Commander, Field Comandk2116 West Soledad Canyon Road Defense Atomic Support AgencySaugus, California Albuquerque, Now MexicoAttn: Mr. L. LoFiego
University of DenverChance Vought Corporation Denver Research InstituteP.O. Box 5907 Denver 10, ColoradoDallas 22, Texas Attn: Mr. R. B. FeaginAttn: R. D. Henry
d
r THE FRANKLIN INSTITUTE for Research and lXena
F-B1805
DISTRIBk0 LIST (Cont.)
Flare DivisionAtlantic Reeearch Corporation Welex Electronic. Corporal: ion19701 W. Goodvale Road Solar Building, Suite 201Saugus, California 16th and K Street@, N.W.Attn: Mr. N. E. Eckert, Head, Waahington 5, D.C.
R&D GroupNorth American Aviation, Inc.
Vitro Laboratories Cocnioation Services1.4000 Georgia Avenue 4300 E. 5th Ave.Silver Spring, Maryland Columbus 16, Ohio
Attn: Miss C. Jaques, Librari
t.Le