the gc-apci interface for the agilent q-tof lc/ms … · the 6550 q-tof lc/ms system provides the...
TRANSCRIPT
IntroductionThe Agilent 3212 GC-APCI interface for use with the Agilent 6530, Agilent 6540, and Agilent 6550 Q-TOF LC/MS systems provides the highest GC chromatographic peak resolution and best mass accuracy in a robust and easy to use platform. The unique design of the GC-APCI ion source enables rapid switching between GC and LC/MS operation and the fl exibility to handle a wide range of challenging applications. All of the Agilent 6500 series Q-TOF products offer fast acquisition speeds up to 50 spectra/sec to provide high quality data for very narrow GC peak widths.
The GC-APCI technology delivers:• Very narrow chromatographic peak widths, especially for high boiling
compounds, providing greater sensitivity and compound separation
• Improved fl exibility and productivity for analysis of a broad variety of sample types
• Up to 3-fold faster analysis times due to higher gas fl ow rates
• Enhanced compound ionization, sensitivity and structural characterization for a wide range of analytes, due to easy access to a selection of makeup gases
This technical overview illustrates the advantages of the GC-APCI Interface for a wide range of applications, including environmental methods such as nitrosamine analysis and Method 8270 for semivolatile organic compounds. It can also provide superior performance for analysis of foods and natural products such as cooked meats and plant extracts. Superior sensitivity and linearity of quantitation are also demonstrated.
The GC-APCI Interface for the Agilent Q-TOF LC/MS System Improves Sensitivity, Mass Accuracy, and Speed for a Wide Range of GC Applications
Technical Overview
AuthorsPaul GoodleyAgilent Technologies, Inc.Santa Clara, CA USA
Sheher Bano MohsinAgielnt Technologies, Inc.Santa Clara, CA USA
2
Simple and secure coupling device provides fast switching between GC and LC/MS operationThe unique design of the GC-APCI interface enables switching between GC and LC operation within a few minutes. A short, heated transfer tube with an outer tube for extra protection connects the GC capillary to the APCI source, which provides the corona discharge for ion generation. This design helps ensure uniform transfer line heating and good resolution of compounds with high boiling points
The quick lock connection between the APCI interface housing and APCI source registers alignment with the ion sources, eliminating the need for adjustments to either column positioning or corona needle positioning, and providing very robust switching to GC-operation.
Combining both GC and LC capability with the TOF or Q-TOF mass analyzer provides the fl exibility to handle a wide variety of sample types, from neutral to highly polar, and signifi cantly increases the productivity of existing MS instrumentation.
The mass accuracy of the Q-TOF generates highly confi dent compound identifi cation from the spectra of each compound. This mass accuracy advantage is also useful for providing identifi cation of unknowns, when screening samples for compounds such as pesticides.
Quick throw locking mechanism that registers alignment with the ion source
Transfer line coupling to GC and oven wall pass-through
Heated transfer line with outer tube integral column effluent tip
APCI-removable needle
GC column inlet
Figure 1. Agilent 3212 GC-APCI interface for the Agilent 6500 Series Q-TOF LC/MS systems.
Figure 2. Illustration of the use of the Agilent 3212 GC-APCI interface to couple the Agilent 7890A GC to the Agilent 6540 Q-TOF LC/MS.
Quick throw loading mechanism
Coupling of GC transfer line to Q-TOF APCI ion source
3
Excellent Sensitivity Some compounds do not easily ionize using ESI, and APCI provides an excellent alternative for their detection and quantitation. Low picogram detection (on the column) can be achieved for many compounds, with excellent linearity of quantitation.
For example, a one picogram injection of octafl uoronaphthalene (OFN) onto the GC column of the combined GC-APCI/6530 Q-TOF mass spectrometer (Figure 3) shows an approximate 10:1 or better signal-to-noise (S/N) ratio for the extracted ion chromatogram (EIC). Note that the lower panel shows the mass spectrum having primarily the radical ion [M]*+ at 271.9870 accurate mass.
×102
00.20.40.60.81.01.21.4
×103
00.20.40.60.81.01.21.41.6
8.145
271.9870
Acquisition time (min)
Mass-to-charge (m/z)
Coun
tsCo
unts
2
256 258 260 262 264 266 268 270 272 274 276 278 280 282 284 286 288 290 292
2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11
EIC m/z 272 of 1 pgOFN on column
Primary ion of OFN at 1 pg
Octafluoronaphthalene
F
F
FF F
F FF
Figure 3. EIC (upper panel) and mass spectrum (lower panel) of 1 pg of octafl uoronaphthalene (OFN) onto the GC column of the Agilent 3212 GC-APCI/Agilent Q-TOF mass spectrometer.
4
Reliable QuantitationLinearity of quantitation across two orders of magnitude is good for several compounds that do not easily ionize by ESI, such as 2,6-dimethyl phenol and cholesterol (Figures 4 and 5), and for compounds that do not ionize at all in ESI, such as methyldecanoate (Figure 6). All of the coeffi cients of calibration (R2 values) for these compounds were > 0.995. The rapid scan speed of the 6550 Q-TOF LC/MS helps ensure accurate quantitation for very narrow peaks.
Figure 4. Calibration curve for 2,6-dimethylphenol, 1 to 500 pg on column.
DimethylphenolR2 = 0.99877.0
×105
6.56.05.55.04.54.03.53.02.52.01.51.00.5
0
0 50 100 150 200 250Concentration (ng/mL)
Resp
onse
s
300 350 400 450 500-0.5
Figure 5. Calibration curve for cholesterol, 1 to 500 pg on column.
CholesterolR2 = 0.9963
7
8
9
×106
6
5
4
3
2
1
0
0 50 100 150 200 250Concentration (ng/mL)
Resp
onse
s
300 350 400 450 500
Methyl decanoateR2 = 0.9959
2.4
2.8
×106
2.0
1.6
1.2
0.8
0.4
0
0 50 100 150 200 250Concentration (ng/mL)
Resp
onse
s
300 350 400 450 500
Figure 6. Calibration curve for methyl decanoate, 0.5 to 500 pg on column. This compound does not ionize using ESI.
5
VersatilityThe Grob test mix is often used to characterize the performance of a GC/MS system, including signal response and mass accuracy, inertness of the fl ow path, and adequate transfer line heating. It is also a good test of the versatility of a system, since it contains a wide variety of small polar and nonpolar compounds, from alkanes to organic acids to aromatics.
The GC-APCI/6550 Q-TOF LC/MS system provides very narrow peaks for the Grob mix compounds, with widths < 2 seconds (Figure 7). The 6550 Q-TOF LC/MS system provides the fast scanning speed needed to generate such high quality data.
The system generates good separations for the wide variety of polarities represented in the mix. For example, butanediol is a highly polar nonvolatile compound, yet it exhibits minimal peak tailing (Figure 8).
The GC-APCI interface also enables analysis of compounds, such as methyldodecanoate, in the mix that do not ionize easily by ESI. The Q-TOF can provide high mass accuracy MS/MS spectra as well. (Figure 9).
Mass accuracy is also very good with the Grob test mix, ranging from –1.06 to +0.53 parts per million (ppm), across several of the compounds.
Cholesterol
Butanediol
1-Octanol
Nonanal
DimethylamineMethyldecanoate
Methylundecanoate
Dicyclohexylamine
Methyldodecanoate
Peak widths are < 2 seconds
2.4
2.8
×106
2.0
1.6
6.8661.2
0.8
0.4
00 2 3 54 6 7
Acquisition time (min)Co
unts
8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
Dimethylphenol
Figure 7. EIC Grob test mix analyzed on the Agilent 3212 GC-APCI/Agilent 6540 Q-TOF LC/MS.
m/z = 71.064973.064996(M+H)+[-H2O]
72.055450M*+[-H2O]
ButanediolHO OH
1.21.4
×104
1.0
0.60.8
0.40.2
03.8 4.2
*4.601
4.6 5.0Acquisition time (min)
Coun
ts
5.4 5.8 6.2 6.6 7.0 7.4 7.8 8.2 8.6
2.4
3.2
4.0×103
1.6
0.8
069 70 72 71 7671 73 72
Mass-to-charge (m/z)
Coun
ts
77 78 79 81 83 85 87 89 92 9380 82 84 86 78 90 91
Figure 8. EIC and mass spectrum of highly polar and nonvolatile butanediol, illustrating minimal peak tailing.
6
Quantitation of PAHs in Food MatricesPolycyclic aromatic hydrocarbons (PAHs) make up a class of more than 100 chemicals composed of up to six benzene rings fused together. PAHs are a human health concern because a number of studies have shown increased incidence of cancer (lung, skin, and urinary cancers) in humans exposed to PAH mixtures.
PAHs can occur in food either by uptake from the environment or as a result of food processing. While the list of priority PAHs varies in different countries, the United States Environmental Protection Agency (USEPA) and the European Union (EU) have identifi ed 16 priority PAHs that require monitoring.
The GC-APCI/6540 Q-TOF LC/MS system was used to develop a method for detecting PAHs in cooked hamburgers, as the cooking process can generate these compounds. A sample of the cooked hamburger (10 g) was spiked with 100 pg of a mix of PAH standards. The sample was then extracted with organic solvents and cleaned up by passing it over a silica column.
MS/MS fragmentation of methyldodecanoate
77.03850
57.06985142.00296
Precursor ion of methyldodecanoate
2.42.8
×105
2.0
1.21.6
0.80.4
0201 203
215.20045(M+H)+
214.19248(M+H)+
216.20356(M+H)+ 218.21040
(M+H)+
205 207Acquisition time (min)
Coun
ts
209 211 213 215 217 219 221 223 225 227
0.8
1.0×105
0.6
0.4
0.2
045 55 75 85 10565 95
Mass-to-charge (m/z)
Coun
ts
115 125 145 155 175 205 225 235135 165 185 195 215
H3CCH3O
O
Figure 9. MS and MS/MS spectra of methyldodecanoate, which does not ionize easily by ESI.
7
Quantitation of PAHs in Food MatricesAn extracted ion chromatogram illustrates very narrow peaks, complete baseline resolution of the 16 compounds, identifi cation of all of the compounds
3.0
3.5
4.0
4.5
5.0
5.5
6.0
6.5
7.0
×105
2.5
1.5
2.0
1.0
0.5
05 7
10.5Cpd 3: Acenaphthene
10.2Cpd 2: Acenaphthylene
8.1Cpd 1: Naphthalene
11.7Cpd 4: Fluorene
14.9Cpd 5: Anthracene
15.1Cpd 6: Anthracene
20.3Cpd 8: Pyrene
21.5Cpd 9: Pyrene
20.3Cpd 7:
Dibenzo(def,mno)chrysene
22.8Cpd 10:
Dibenzo(def,mno)chrysene
26.7Cpd 12:
Benzo(c)phenanthrene
26.5Cpd 11:
Benzo(c)phenanthrene
35.7Cpd 17:
1,2,5,6-Dibenzanthracene
35.7Cpd 16:
Dibenzo(def,mno)chrysene
36.6Cpd 18:
Dibenzo(def,mno)chrysene
30.4Cpd 14:
Benzo(a)pyrene30.3
Cpd 13: Benzo(a)pyrene 31.5
Cpd 15: Benzo(a)pyrene
9 11Acquisition time (min)
Coun
ts
13 15 17 19 21 23 25 27 29 31 33 35 37 39
Figure 10. Extracted ion chromatogram of a standard mix of 16 PAH compounds identifi ed using their accurate mass. The system was calibrated using the background ions from the GC column bleed.
using accurate mass, and partial resolution of some of the isomers , (Figure 10). Expanding the region from 28.6 to 33.2 minutes illustrates the partial resolution and identifi cation of the four isomers, with mass accuracy ≤ 0.8 ppm (Figure 11).
4th isomer
30.4Cpd 14:
Benzo(a)pyrene 31.5Cpd 15:
Benzo(a)pyrene
30.3Cpd 13:
Benzo(a)pyrene
Benzo(a)pyrene C20 H12 30.330.431.5
253.10101253.10111253.10125
252.09372252.09381252.09396
-0.72-0.370.23
322
334,291375,124268,214
C20 H12C20 H12
Benzo(a)pyreneBenzo(a)pyrene
4.8×105
4.44.03.63.22.82.42.01.61.20.80.4
28.6 29.0 29.4 29.8 30.2 30.6 31.0 31.4 31.8 32.2 32.6 33.00
Acquisition time (min)
Coun
ts
Figure 11. Expanded section of the extracted ion chromatogram shown in Figure 10 illustrating the partial separation and identifi cation of the four isomers of benzo(a)pyrene.
8
Figure 12. The mass accuracies obtained for the 16 PAHs spiked into the hamburgers.
Identifi cation of the PAHS by mass only requires very accurate mass determinations. In this case, the mass accuracies ranged from as low as 0.02 ppm to no higher than 2 ppm, enabling confi dent compound identifi cations (Figure 12).
Compounds with very low proton affi nity can be identifi ed using the mass of the free radical cation (M)*+, enabling confi rmation (Figure 13).
This data was graciously provided by Agnieszka Olas and Karl Pettit of Marchwood Scientifi c.
Phenanthrene
Cpd 6: Phenanthrene: +FBF spectrum (14.6–14.6 minutes)2.8
×104
2.2
2.4
2.6
2.0
1.8
1.6
1.4
1.2
1.0
0.8
0.6
0.4
0.2
8 9 10 12 14 16 1811 13 15 17 19 20 21 22 25 27 29 31 32 3323 24 26 28 300
Acquisition time (min)
Mass-to-charge (m/z)
Coun
ts
Coun
ts
×103
7
6
5
4
3
2
1
177.5 178.0 178.5 179.0 179.5 180.0 180.50
178.07763M*+
179.08506(M+H)+
180.08844(M+H)+
Figure 13. Phenanthrene is a low proton affi nity compound, and it can be identifi ed using both the M*+ and (M+H)+ions.
9
Analysis of Semivolatile Compounds (EPA Method 8270)EPA Method 8270 is widely used for the analysis of semivolatile compounds in environmental matrices, including water and soil. The list of compounds contains many pesticides, as well as PAHs.
The GC-APCI/6540 Q-TOF LC/MS system provides excellent results with Method 8270, using a 50 pg injection of an 8270 standard mix of 50 compounds (Figure 14). Many of the PAHs have similar molecular structures as well as similar and common molecular weights. Figure 14 demonstrates the benefi ts of having an isothermal transfer line, which results in very narrow peak widths and excellent resolution, particularly for the high boiling PAH compounds, such as those at the 1,120 second retention time (Figure 15). In addition, there is very little drift in the baseline.
Method 8270 standard mix
460 500 580 660 740 820540
450
458
471
480
487
499
510
518
534
540
548
559
554
577
603
615
586
632
621
637
642
670
688
677
665
648
699
714
733
738
746
754
781
791
786
774
811
824
835
848
857 87
3
817
880
887
897
911
922
934
940
950
964
974
982 98
610
0010
0810
2310
4310
5810
6910
8010
8910
9911
1911
4011
6011
6811
7811
97
620 700 780 860 900 940 980 11001020 1060 1140 1180
×104
0.7
0.91.01.11.21.31.41.51.61.71.81.92.0
0.5
0.30.4
0.6
0.8
0.20.1
Coun
ts
Acquisition time (sec)
Figure 14. Base peak chromatogram of a 50 pg injection of an Method 8270 standard mix.
720 760 800 920 960
725 733
738
746
744
754
763
786
781
791
811
824
835
848
670
688
887
857
817
873
911
897
922
931 93
495
094
0
964
974
982
986
996
1000
1008
1015
1023
1043
1058
1069
1076
1080
1084
1099
1107
1123
1132
1140
1148
1160
1119
.171
1168
1178
1197 1211
1226
840 880 1000 1040 1080 11601120 1200
×104
0.7
0.91.01.11.21.31.41.5
0.5
0.30.4
0.6
0.8
0.20.1
Coun
ts
Acquisition time (sec)
Figure 15. Expanded section of the base peak chromatogram shown in Figure 14 with an inset illustrating the excellent resolution of the late-eluting high boiling compounds, including two shown in the inset at 1,110 and 1,120 seconds, respectively.
10
In addition to narrow column widths, using the Q-TOF mass spectrometer provides the ability to identify compounds with very similar retention times based on very small mass differences, due to its excellent mass accuracy (Figure 16).
Figure 16. A base peak chromatogram and mass spectra illustrating the successful identifi cation of two PAH compounds with small differences, due to the benefi t of excellent chromatography and the accurate mass capability of the system.
Measured ion mass277.0983 m/z
Measured ion mass279.1143 m/z
Dibenz[a,h]anthraceneActual ion mass[M+H]+ C22H15 = 279.1170
×105
×105
1.2
1.6
0.8
0.40.6
1.0
1.4
0.20
1.21.00.80.60.40.2
110 120 140 160 180 100130 150 170 190 210 220 230 240 270 290 310 330250 260 280 300 3200
Indeno[1,2,3-cd]pyreneActual ion mass[M+H]+ C22H13 = 277.1014
Mass-to-charge (m/z)
710 720 740 760 780 800730 750 770 790 810 820 830 840 870 890 910 930 940 950 960850 860 880 900 920Acquisition time (min)
Coun
ts
×105
2.6
3.43.8
1.8
1.01.4
2.2
3.0
0.60.2
Coun
tsCo
unts
279.1143
277.0983
11
Nitrosamines in Environmental SamplesNitrosamines are classifi ed as carcinogens by the USEPA and the Food and Drug Administration (FDA). Research studies have suggested that certain nitrosamines could be mutagens and could cause birth defects at very low concentrations.
Nitrosamines are found in in cured meat products, tobacco, rubber, cosmetics, and other consumer products. Concern has also grown regarding the health effects of naturally formed nitrosamines caused by added nitrites and amines in food.
Effective analytical techniques are needed by many industries to measure these toxic compounds. Gas chromatography combined with nitrogen and chemiluminescent detection offers reliable separation and sensitive detection, but not absolute molecular confi rmation and identifi cation. Gas chromatography combined with accurate mass measurements can provide a higher level of molecular structure confi rmation.
Figure 17 illustrates the separation and detection of nitrosamine standards using the GC-APCI/6540 Q-TOF LC/MS system. The peak at 7.7 minutes appears to be a single chromatographic peak, but when ion extraction is performed using accurate masses, more peaks appear (Figure 18). The coelution of three nitrosamines is readily observed when using accurate mass to determine the compound.
Figure 17. Extracted ion chromatogram of a standard mix of nitrosamines.
×105
00.20.40.60.8
11.21.41.61.8
22.22.42.62.8
33.23.43.63.8
44.2
Acquisition time (min)
Coun
ts
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
N-Nitrosodi-n-butyl-amine
1-Nitrosopiperidine
N-Nitroso-N-methylethylamine
Diethylnitrosoamine
A group of three coeluting nitrosamines
Figure 18. Expanded section of the extracted ion chromatogram shown in Figure 17 illustrating the ability of accurate mass to determine the identities of three coeluting compounds.
00.20.40.60.8
11.21.41.61.8
22.22.42.62.8
33.23.43.63.8
44.2
5
5.8 6 6.2 6.4 6.6 6.8 7 7.2 7.4 7.6 7.8 8 8.2 8.4 8.6 8.8 9 9.2 9.4 9.6 9.8 10 10.2 10.4
1-Nitrosopyrrolidine
Nitrosomorpholine
N-Nitroso-dipropylamine
×105
Acquisition time (min)
Coun
ts
12
Natural Products AnalysisCurry leaf essential oil is used in Indian cooking as well as traditional medicine native to the Indian subcontinent. It is said to strengthen the gums and teeth, prevent nausea and cure stomach upsets, treat skin irritations and poisonous bites, and act as an antibacterial, antifungal, and astringent.
Accurate mass measurements can be used to identify two compounds with the same nominal mass, that differ by as little as 36 millimass units in accurate mass (Figure 19).
00.10.20.30.40.50.60.70.80.9
1
+APCI EIC(131.0815) Scan Frag=130.0V SAmple5-006.d
*3
00.10.20.30.40.50.60.70.80.9
1
+APCI EIC(131.1179) Scan Frag=130.0V SAmple5-006.d
*1
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
HEEDAm/z 131.0815
C5H10N2O2131.0817C5H11N2O2
131.1175C6H15N2O
149.0229
C6H14N2O
N-Nitroso-dipropylaminem/z 131.1179
×102
×102
Acquisition time (min)
Coun
ts (%
)
0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
130 131 132 133 134
×103
Mass-to-charge (m/z)
Coun
ts
0
1
2
3
4
5
6
7
8
120 130 140 150 160
×104
Mass-to-charge (m/z)
Coun
ts
Coun
ts (%
)
Figure 19. The advantages of accurate mass measurements for identifying two compounds with isobaric masses at nominal mass 131. The difference in accurate mass for the two identifi ed coompounds is only 36 milimass units.
This commercially valuable oil has been shown to exhibit chemical diversity depending on the geographic location. This diversity impacts the aroma and fl avor of the oil, as well as its potential health benefi ts.
13
Figure 20. Extracted ion chromatogram and MS spectrum of components of curry oil.
00.20.40.60.8
11.21.41.61.8
2 205.1951C15 H25
148.1246
Mass-to-charge (m/z)140 160 180 200 220 240 260 280 300 320 340 360 380 400 420 440 460 480
0
1
2
7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13 13.5 14 14.5 15 15.5 16 16.5 17 17.5
Caryophyllene/cubenene
MS of caryophyllene from curry leaves
×105
×105
Acquisition time (min)
Coun
tsCo
unts
Figure 21. A graph in MPP of mass versus retention time, also showing the identities of several of the compounds determined from RT and accurate mass.
Mass vesus retention time
SabinenePhellandrenePinene
CaryophylleneCubenene
700
600
500
400
300
200
100
00 5 10
Retention time (min)
Mas
s (Da
)
15 20
The GC-APCI/6550 Q-TOF LC/MS system and Mass Profi ler Professional (MPP) software are excellent tools to study this chemical diversity. The retention time and accurate mass can be used to identify large numbers of compounds in the oil, and the analysis tools in MPP can be used to elicit differences in the compounds contained in oil sourced from different locations.
Once accurate masses have been determined for compounds, their retention times can be compared to those in the METLIN database using MPP, in order to confi rm their identity. In this case, caryophyllene and cubenene have the same mass and retention time (Figure 20). Several other compounds identifi ed using MPP and the METLIN database are shown in Figure 21.
www.agilent.com/chem/LCMS
This information is subject to change without notice.
© Agilent Technologies, Inc., 2013Published in the USA, April 22, 20135991-2210EN
ConclusionThe GC-APCI Interface for Agilent Q-TOF LC/MS systems improves productivity and reduces cost by adding GC capability to your laboratory. This cross-platform approach has attracted a lot of attention in several areas of analytical chemistry, especially metabolomics. For metabolic profi ling, the traditional approach would use at least two different instruments: the GC/MS for the volatile components and LC/MS for the nonvolatile more polar species. This requires a lab to have two or more instruments to cover the range of polarity/volatility of compounds. With GC-APCI, the same mass spectrometer can be used for both GC and LC analyses.
Atmospheric pressure ionization enables as much as fi ve times higher gas fl ow than dedicated GC systems. The end result is faster analysis times and larger sample loads. A wide range of makeup gases that can be readily attached to the interface assures ionization for a wide variety of samples. You gain the fl exibility to handle a wide range of challenging applications, including those highlighted here. Very narrow GC peaks and the accurate mass capability of the Q-TOF LC/MS provide greater sensitivity, higher resolution and confi dent compound identifi cations.