the nuts and bolts of icd therapy - cardiolandcardioland.org/ecg/sjm/the nuts and bolts of icd...
TRANSCRIPT
1
For Cardiology Fellows
Electrical Management of Cardiac Rhythm Disorders
December 5-8Austin, Texas
The Nuts and Bolts of ICD Therapy
2
3
Action Potential – Localized Differences in Conduction
• 0.2 meters/second in the AV node• 0.4 meters/second in ventricular muscle• 1 meter/second in atrial muscle• 4 meters/second in the Purkinje fibers
Conduction velocity within cardiac tissue varies from slowest to fastest as follows:
TachycardiaMechanism
Reentry80 -90%
Non-reentryTorsade de pointes
Acute ischemia
Mechanisms
5
Normal Heart Scarring Due to Myocardial Infarction
Ischemic Re-entry
Mechanisms
6
Mechanisms
Gray areas represent scar
7
Mechanisms
8
THE BIG FOUR
• Sensing• Detection• Diagnosis• Therapy
SenseAbility™
10
Ideal Sensing Circuit
• Ideal sensing circuit:
– Senses R-waves– Senses varying amplitudes of arrhythmias from sinus rhythm to
low amplitude ventricular fibrillation– Does not over sense P-waves, T-waves, myopotentials or EMI– Filters the signal to achieve a clean, noise-free waveform
suitable for detection
11
Sensing Challenges
• Sensing challenges
– Combine the functions of an ICD with a dual chamber rate responsive pacemaker
– Add the ability to use atrial sensing as an SVT discriminator– Accurate sensing in both the atrium and ventricle is critical for
appropriate therapy delivery
12
Sensing Options Pacemakers
• Fixed sensitivity
NSR
VT/VF
13
Programmable Sensing Features
• Programmable sensing features – Threshold start– Decay delay– Maximum sensitivity– Refractory periods
• All parameters can be adjusted or “fine tuned” in the presence of sensing problems
14
Sensed Refractory
• Sensed Refractory :• Ventricular: 125 (nominal) or 157 ms • Atrial: 93 (nominal), 125 or 157 ms
Maximum Amplitude
R-Wave or P-Wave Sensed
Sensed Refractory
15
Threshold Start
• Definition: percentage of maximum peak amplitude sensed during the sensed refractory period used to begin the linear decay– Ventricular threshold start
• 50%, 62.5% (nominal), 75% or 100%– Atrial threshold start
• 50% (nominal) , 62.5%, 75% or 100%
16
Threshold Start
• Nominally 62.5% of measured R– (Max start value 3.75 mV if R waves are >6mV))
• Nominally 50% of measured P– (Max start value 1.5 mV if P-waves are >3mV))
• Fixed rate of decay– Atrium 0.5 mV per 312 ms– Ventricle 1.0 mV per 312 ms
50%
62.5%
R-Wave T-Wave
17
Rules to SenseAbilityTM Threshold Start
• Ventricular: Threshold start will not start higher than 6.0 mV or lower than 1.0 mV
• Atrium: Threshold start will not start higher than 3.0 mV or lower than 0.3 mV
• I.E. If R-waves are 10 mV and threshold start is 62.5%, it will begin at 3.75 mV
18
Decay Delay
• Holds threshold at starting value for a period of time• May be increased if oversensing
– T-wave oversensing on RV channel– Far field R-waves on RA channel
0 ms
60 ms
19
Decay Delay Programming
• Ventricular – Post sensed: 0, 30, 60 ms (nominal), 95, 125, 160, 195 and 220
ms– Post paced: Auto, and the options above
• Atrium– Post sensed: 0 (nominal), 30, 60, 95, 125, 160, 195 & 220 ms– Post paced: Same as post sensed
20
Maximum Sensitivity (The “Floor”)
• Definition: The most sensitive level to which the device can decay – Anything below the maximum sensitivity is not seen by the
device– Separately programmable (ICD and pacemaker)
• Nominals:– Ventricular 0.3 mV – Atrial 0.2 mV
21
Maximum Sensitivity
• Decays until reaches max sensitivity or another signal is sensed
V- Max Sensitivity = 0.3 mVA- Max Sensitivity= 0.2 mV
0.3 mV
22
Post Paced
• There is a separate algorithm used when the device is pacing
• There is no P- or R-wave to measure• The threshold start and decay delay are based upon the
pacing rate
23
Post Paced
• As the rate increases, the threshold start will decrease• This allows the device to become more sensitive at fast
rates
VRPVRP VRP VRP
Automatic Post Pace Threshold Start
24
Atrial SensingAutomatic or Fixed
25
Pacemaker and defib max sensitivity can be different
26
Can Change for Double Counting Due to Wide QRS
DETECTION
28
Detection – Ideal Design Considerations and Challenges
• High sensitivity for detecting VT and VF
• Diagnose tachyarrhythmias in a timely manner
• Make appropriate diagnosis decisions– Avoid diagnosing SVT as VT and treating inappropriately
• Classify hemodynamically stable vs. unstable VT with rate cut offs
29
Detection
• Proper rhythm detection is dependent upon:
– Accurate sensing: • Undersensing (signal dropout) • Oversensing (double-counting or T-wave sensing)
• This will result in incorrect interval measurement and classification
• The ICD makes detection decisions based upon the programmed rate cut-off only– It is not a "clinical" diagnosis
30
Detection
• The ICD uses the bipole IEGM• Detection decisions are based upon the programmed
ventricular rate cut-off, but rate is not enough
31
Detection
• The recognition by the device of a minimum number of intervals that satisfy the programmed rate criteria for
– Fibrillation– Tachycardia– Sinus rhythm
32
Detection - Terminology
• Interval– Current interval– Interval average
• Binning– Fibrillation detection– Tachycardia detection– Sinus redetection
33
Detection
• Current interval– Measures the time in milliseconds between any two sensed
events, paced events or combination
• Interval average– The current interval, in milliseconds, plus the 3 prior intervals
divided by 4– This helps to provide a “smoothing effect” on detection– Thus, 1 or 2 beat variations don’t have a large effect on
detection
34
Detection Zones
• 1 Zone– VF only
• 2 Zone – VF and VT
• 3 Zone– VF zone and VT-1 and VT-2
• OFF– Defibrillator OFF
• Bradycardia pacing is available in any Detection Zone configuration
35
36
Detection
• 3 zone– VF with VT-1 and VT-2– Tiered therapy with 2 tach zones and VF – Arrhythmias in tach zones receive:
• ATP and low energy cardioversion– Used in patients with 2 different VTs
• Fast VT• Slow VT
37
3 Zone
• Must have 30 ms between pacing upper rate, VT-1, VT-2 and VF Zone
Normal Bradycardia Rhythm VT 1 Zone VT 2 Zone VF Zone
140 bpm 160 200 bpm
Rhythm:VVI(R) or DDD(R)
Pacing
38
Detection
• OFF– Provides bradycardia pacing only and needs to be programmed
ON – Used if patient is experiencing inappropriate shocks– When patient is undergoing surgery involving electrocautery
• Shocks during delicate surgery could be disastrous
39
Disable/ Enable Tachy Therapy
• Disable/ Enable Tachy Therapy– A red line will cross over the VT/ VF parameters on the FastPath
Summary page
40
Monitor Zone
• Will monitor, store EGM and not treat tachy arrhythmias• Multiple uses: palpitations, additional tachy arrhythmias
41
Detection – Binning Rules
• When the interval and interval average are the same, the interval is binned in that zone
• When the interval and interval average are not the same classification, but are tach/ fib the interval is binned in the faster zone
• Always errors on the safe side!• The interval is not binned if the interval or interval
average is sinus and the other is VT/VF• Prevents 1 beat variations from impacting detection!
42
Detection – Binning Rules
Interval Interval Average Binned
Sinus Sinus SinusVT/VF Sinus Not BinnedSinus VT/VF Not BinnedVT-1 VF VFVT-2 VT-1 VT-2VF VT-2 VFVF VF VF
43
Interval Average Calculation
• Programmed Parameters– VT Detection: 140 bpm (430 msec)
• Current Interval = 418 – Interval Average = 418 + 422 + 422 + 430 = 1692/ 4 = 423
Episode Detection at 142 bpm (420 ms)
44
Diagnosis
• Diagnosis– The point at which a rhythm meets all the programmed criteria
once diagnosis is made therapy will be given– But in order for a diagnosis, the device needs to know how many
intervals must be binned to qualify
45
Diagnosis
• Tach Detection: T-1/140 bpm,12 Intervals
• T-2/ 160 bpm, 12 Intervals• F- 182, 12 Intervals
CI IA422 + 422 + 430 + 434 = 427 ms418 + 422 + 422 + 430 = 423 ms422 + 422 + 418 + 422 = 421 ms
46
Sinus Redetection
• For the device to consider the episode over after a tachyarrhythmia has been detected:– The appropriate number of sinus intervals must be binned– Brady pacing counts– Intervals must be consistent (non-tach/ fib) but not consecutive
• Programmable • Nominal = 5• Slow = 7• Fast = 3
47
Sinus Redetection
• Brady pacing counts as sinus
48
Redetection After Therapy
• After therapy has been delivered a certain number of intervals must be binned before the next therapy can be initiated
• This insures the arrhythmia is still present
49
Redetection After Therapy
• Fib redetection – Fixed at 6 intervals
• Tach 1 & 2 redetections are separately programmable – 6 to 20– Nominal = 6
50
Tach - Redetection
• Tach detection = 400 ms/ 12 Intervals• Fib detection = 300 ms/ 12 Intervals• Tach redetect – 6 Intervals
51
Redetection After Therapy
• An arrhythmia may continue at a slower rate after therapy
• If the rate is slower than the programmed initial detection rate, the device must be able to recognize and continue to treat the arrhythmia
52
What happens if my patient has PVCs in a bigeminal pattern?
• Tach / sinus ratio counterAKA bigeminal avoidance
53
Tach/ Sinus Ratio Counter
• AKA bigeminal avoidance– Tach/sinus ratio counter protects against inadvertently treating
ventricular bigeminy• This is NOT a lethal arrhythmia and should not be treated• Patients are usually not symptomatic
– Must protect against detecting a bigeminal rhythm as a tachycardia (do not want to treat)!
– Also could happen with T-wave oversensing
54
Tach/ Sinus Ratio CounterVT = 400Number of Intervals = 12
T/S counter = 0
2.5 2.5 2.5 2.5200 200 200 200 200 200 200 200 200 200 200200
500 500 500 500 500 500 500 500 500 500 500200
500200
500
Tach/Sinus CounterFor Each Sinus – 1For Each 2.0 Seconds – 1For Each non Sinus + 1If T/S Counter ≥ 3 = TachIf T/S Counter < 3 = Bigeminy
Will Tell You if Bigeminal Avoidance in Diagnostics
Reason Store = “Bigeminal Avoidance”
55
Discrimination/Diagnosis
• Covered by Tom Kenny
High Voltage Therapy
57
High Voltage Therapy
• Purpose of high voltage therapy– To terminate:
• Ventricular tachycardia• Ventricular fibrillation
58
Shock Therapy
• Shock therapy can be successful no matter what phase the cell is in as long as we capture the cell
• The trick is capturing the cell
59
The Challenge of Stopping VF
• The more muscle mass you can place between the shocking electrodes, the more likely you are to be successful with the shock
Shocking Coils
Can Electrode
Left Ventricle
60
Battery
• Lithium silver vanadium oxide• 3.2 volt battery• Gradual decline in voltage
gives good end of service indicator
61
Capacitors
• The battery alone can’t deliver energy rapidly enough and doesn’t have enough voltage– Capacitors use to be the largest component
in the defibrillator– High-voltage capacitors are used to store
the charge and then deliver the shock
62
High Voltage Therapy Delivery – Detection During Charging
Charge
**
11
**
22
**
33 44 55 66
Interval and Interval Average must be Non Sinus
SHOCK
After beginning to charge, the device must reconfirm the presence of the tachyarrhythmia by binning 6 non-
sinus intervals before delivering the shock
TT TT TT
63
High Voltage Therapy – Aborted Therapy
* ******* * *
1
*
2
*
3
*
4
*
5
Sinus Redetect
Stop Charging
If device redetects “sinus rhythm” during charging, the shock is aborted and not delivered to the patient
F FF FF FF F F VS VS VS VS VS*
65
What Is a High DFT?
• When the amount of energy being delivered is not enough to safely defibrillate the patient on a regular basis, the patient has a high DFT
• Many physicians believe that there must be at least a 10 J safety margin between your patient’s DFT and the maximum programmable energy of the device
66
Who Will Have Hight DFTs?
• Almost impossible to predict• Common indicators
– Low LV ejection fraction– Higher NYHA class– Previous history of bypass surgery– Amiodarone use within the past six weeks– History of prior VF
67
How Were Your Physicians Trained to Deal with High DFTs?
• Reverse polarity (non-invasive)• Reposition the lead (invasive)• Start adding hardware (invasive)
– Manually add an SVC coil if you started with a single-coil lead– Manually remove the SVC coil from the header if you started
with a dual-coil lead– Add an SVC coil in the coronary sinus– Add an array
68
Each of the Traditional Options Require
• Additional shocks• More hardware
– More incisions– If an array is implanted, the patient can be left with chronic back
pain
• More time
69
An Overview of St. Jude Medical’s Portfolio of Advanced Options
• Programmable polarity
• Programmable tilt
• Programmable fixed pulse widths
• Programmable shocking vectors
• Unsurpassed delivered energy (36 J)
70
Programmable Shocking Vectors
• In a 2 coil lead system, program SVC Coil ON or OFF• Allows change in the location of current flow• Eliminates the need to invasively change lead-header
configuration, preventing additional surgical procedures
Nominal: RV to SVC/Can
Nominal: RV to Can
71
Waveform Programmability
• What do we mean by waveform programmability?
– We mean that the waveforms have programmable “fixed tilts”and “fixed pulse widths”
– St. Jude branded ICD waveforms (both monophasic and biphasic) can be programmed to either fixed tilt or fixed pulse width
72
Waveform Programmability – Fixed Tilt
• Definition: Tilt represents the percentage fall in voltage on the capacitor from the beginning to the end of each phase over the course of the entire pulse
• For example, a 200 V shock with 65% tilt would deliver 130 V in the first phase then shift its polarity– (200V X .65 = 130); (200 – 130 = 70 V left over)
73
Cellular Time Constant• The example below shows a shock being delivered and the cell's response. The cell
ends its response at about 4 ms, but notice the shock continues. The energy is wasted, but worse than that, it is counter-productive to the cell. It reduces the cell’s final response.
-20
0
20
40
60
80
100
0 2 4 6 8 10milliseconds
shock voltage
membraneresponse
Wastedenergy
Optimal PW
74
Waveform Programmability – Fixed Tilt (Biphasic)
• Pulse width will adjust based on programmed shock impedance
• Fixed tilt - shock therapy will be in joules (a voltage reference is provided)
Tilt = 50%Charge to 400 V (aprox 10 J)
Biphasic
InitialVoltage
Phase 1Width
Phase 2Width
TILT
TILT
400 V
200 V
75
Waveform Programmability – Fixed Tilt
• One might think that a higher tilt which would deliver more energy would be best (more is better)– Studies have shown that tilts between 40% and 65% are better
than other tilts
• A fixed tilt device (or a device programmed to a fixed tilt) delivers a constant energy by varying the pulse width as a function of the patient’s defibrillation lead impedance
76
Waveform Programmability – Pulse Width (Model Specifics)
• Each pulse width phase is programmable in volts (joules provided as reference)– St. Jude branded devices
• 1st phase: 3.0 to 10.0 ms in 0.5 ms increments – Nominal: 5.5 ms
• 2nd phase: 1.2 or 1.5 - 10 ms in 0.5 ms increments– Nominal: 5.5 ms
77
Waveform Programmability – Burping Theory Cont.
• Recommended pulse widths are derived• Optimal pulse widths are a function of
– High voltage lead impedance– Device capacitance– Cell membrane time constant
• Shortening of the second phase pulse width lowers DFTs in patients on class III anti-arrhythmics
78
Pulse Width Optimization Table
Device Family
Resistance
Block Number
ATP Therapy
80
Antitachycardia Pacing (ATP)
• Advantages– When this therapy is utilized the patient usually has minimal
symptoms– No charge time, so it is a quick therapy– Pain free
• Disadvantages– Requires time to deliver multiple bursts– ATP may accelerate the VT– Prolongs time to shock therapy
81
Antitachycardia Pacing
• Works best with reentrant ventricular tachycardias
82
EP “Talk”
• Burst pacing– Delivery of multiple fast stimuli
• Burst or train– A single series of paced stimuli
• Ramp pacing– The paced rate is progressively
increased within a burst
• Scanning– An automatic change in cycle length
from burst to burst attempt
83
Burst Cycle Length (BCL)
• Fixed or adaptive– Nominal: adaptive 85%
• Adaptive - % of average tach interval at the time of diagnosis
• Example:– Adaptive ON @ 75%– Tach interval 400 ms– BCL = 300 ms
Burst Cycle Length Fixed
X Milliseconds
X
Burst Cycle Length Adaptive
X % of Tach Detect Rate
X
84
ATP Burst Therapy
• Simple burst therapy– BCL = 80% (adaptive)– Tach cycle length = 400 ms– 400 ms X 80% = 320 ms
Tach CL 400 ms
Burst CL 320 ms 320 ms 320 ms 320 ms
BCL: 80% No. of Stimuli = 4 No. of Bursts: 4
85
ATP Parameters
• Ramp – Successively decreases the intervals between pulses within a
burst which will pace faster• ON / OFF (nominal)
– Ramp step• Intra burst step size• Nominal 8 ms when ON
86
ATP Parameters
• Ramp pacing • Ramp pacing can be combined with readaptive to adapt
the first cycle of each burst to the VT cycle length can be combined with readaptive to adapt the first cycle of each burst to the VT cycle length
Ramp: On Ramp Step: 20 ms BCL: 90% No. of Stimuli = 4 No. of Bursts: 4
Tachy CL 400 ms
87
Scanning
• A scanning burst sequence consists of multiple bursts where the rate is faster (cycle length shorter) between bursts– ON (decremental) or OFF, nominal ON (dec)– Scan step (nominal 12 ms)
Scanning: On Scan Step: 10ms BCL: 90% No. of Stimuli = 4 No. of Bursts: 4
VT CL 400 ms
Burst Cycle Length 360 ms 350 ms 340 ms 330 ms
88
ATP Parameters
• Ramp with scanning
BCL = 90 % Scanning Step 10 msRamp Step = 25 ms400
360 335 310
400
350 325 300400
340 315 290
285
275
265
89
ATP Parameters
• Minimum BCL – Shortest possible cycle length delivered during any ATP scheme– Will not pace faster than the min BCL– Nominal 200 ms
• ATP therapy is based on the initial detected tachycardia rate and will not change if the tachycardia cycle length is altered, unless readaptive is ON
90
ATP Parameters
• ATP may need more energy than traditional brady pacing– Amplitude choices of 7.5 V (nominal) or 10.0 V (max)– Pulse width choices of 1.0 ms (nominal) or 1.9 ms (max)
91
ATP Delivery
• When tach bin is full, the first ATP pulse is delivered synchronous with the next tach event (QRS)– The surface ECG typically shows the stimulus 40 to 80 ms after
the onset of the QRS– The remaining stimuli will be delivered as VOO
92
Questions?