July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of- Field Locations for Radiotherapy Beam Scanning 59 th Annual Meeting of the American Association of Physicists in Medicine Denver, CO Wesley Culberson Department of Medical Physics, University of Wisconsin Madison University of Wisconsin Medical Radiation Research Center (UWMRRC)

Upload: others

Post on 09-Oct-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

July 30, 2017

The Use of Large-Volume Ion Chambers in Out-Of-

Field Locations for Radiotherapy Beam Scanning

59th Annual Meeting of the American Association of Physicists in Medicine

Denver, CO

Wesley Culberson

Department of Medical Physics, University of Wisconsin – Madison

University of Wisconsin Medical Radiation Research Center (UWMRRC)

Page 2: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Disclosures

• None

2 of 15

Page 3: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Background

• Linac output fluctuates

• When acquiring commissioning scans, removing the effects of

output fluctuations helps

• Internal monitor chambers too difficult to utilize

• “Reference detectors” and a second electrometer used to

account for beam fluctuations

3 of 15

Remove linac

fluctuations

Smoother

data

Better TPS

calculations

Page 4: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

PDD

without

reference

detector

Profile

without

reference

detector

Output Fluctuations

4 of 15

PDD with

reference

detector

Profile with

reference

detector

Page 5: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Challenges

• Reference detector location must be continually adjusted

5 of 15

Page 6: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Challenges with small fields

• Reference detector location not available for very small fields

• Some linacs have access for reference detectors in the head,

but this can be tricky

6 of 15

Possible Elekta Infinity

reference detector location

Page 7: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Project Goal

• Test locations around a linac with large-volume chambers

• Quantify ion chamber current for scanning feasibility

7 of 15

Page 8: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Methods

• Three large-volume ion chambers used

1) Exradin A4 – 30 cc

2) Exradin A5 – 100 cc

3) Custom parallel-plate – 30 cc

8 of 15

Image source: www.standardimaging.com

Page 9: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Methods (cont)

• Varian 21EX

• 6 MV, 600 MU/min

• Multiple locations outside of the primary beam path tested

• Field sizes: 1 x 1 cm2 -> 10 x 10 cm2

• Ion recombination checked

• Chambers relocated twice for repeatability testing

• Radiochromic film measurements for hot spots

9 of 15

Page 10: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Results

• Green cells indicate beam currents > 138 pA (adequate for

beam scanning)

• Recombination <2% for all tests

10 of 15

Page 11: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Hot Spots

• Varian 21EX

• Varied little with field size

• Gradient present

11 of 15

Page 12: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

12 of 15

Hot

spot

Page 13: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Hot Spots

• Varian TrueBeam™

13 of 15

Page 14: The Use of Large-Volume Ion Chambers in Out-Of- Field ......July 30, 2017 The Use of Large-Volume Ion Chambers in Out-Of-Field Locations for Radiotherapy Beam Scanning 59th Annual

Summary

• Out of field locations tested for different chambers and locations

around a clinical linear accelerator

• It is possible to use a reference detector in an out-of-field

location

• This means fewer trips into the vault

• Very useful for small fields where reference detector location is

limited