I R;WTheChildren \v
FerozHospital Si llte Institute of Child Healthzepur Road, Lahore Phone # (92) (42) 99230901-23 Fax # 99231560
D
3
No. P.C./ >V $-77 /CII&ICH Dated ,2019To
Mr. Abdul VVahabI/C Programmer (l.T.)Health DLahore.
eparlment, Civil Secretariat.
Subject: - REQUEST
REPORTS_
EQUIPMENT FOR r
YEAR 201S-2019WEBSITE
FOR UPLOADING
__HIE PROCUREMENT OF ELECTRO MEDICAL
THE CHILDREN’S HOSPITAL. LAHORE FOR THEON PPRA WEBSITE / HEALTH DEPARTMENT
THE TECHNICAL EVALUATIONFOR
Please find enclosed herewith bids technical evaluation for the procurement of
of Institute of Paediatric
in The Children's Hospital,
Paediatric Surgery
Cardiology & Cardiac Surgery Scheme for the year 2018-2019 iLahore need to be uploaded on the PPRA website / Health Deps
y & Allied Specialties & Establishment
artment.
Your positive response will be highly appreciated.
Pifof. Dr. IV ammad Saleem} (FCPS. MME-HPE)
rsdiatric SurgeryPrivfessdr ofMedical Directors
Sr. # Evaluation Parameters Medequips Pvt Ltd Vertex Medical Shirazi Trading Al-Kareem
01 Complete Package / Tender (Mandatory) Yes Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Responsive Responsive Responsive Responsive
Sr. # Evaluation Parameters Medequips Pvt Ltd Vertex Medical Shirazi Trading Al-Kareem
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes Yes
02 Technical & Engineering capability (As defined for the specific Bid in specifications) Yes Yes Yes Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes Yes Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals No Yes Yes No
05 Availability of relevant Tools and Testing / Calibration Equipment Yes Yes Yes Yes
06 Warranty Yes Yes Yes Yes
Non Responsive Responsive Responsive Non-Responsive
Evaluated by Purchase Committee
Remarks:
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
The Children’s Hospital & The Institute of Child Health Lahore
Remarks:
PART - II
Evaluated by Technical Committee
Required Quantity: 2
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Name of the Equipment: Phototherapy Unit
Technical EvaluationPackage No. / Bid No. : 15426/ 01 / CH & ICH
Required Item Name
Phototherapy Item # 1
Brand Atom Drager GE Healthcare Veiele
Sr. # Quoted Model Bili Therapy BiliLux Lullaby LED LED 1000
01 Country of Manufacturer Japan Germany USA Sweden
02 Manufacturing site Japan Germany India Sweden
03 Compliance with defined Quality Standards Compliance Compliace Compliance Compliance
04 Specifications Compliance Feature Wise Non-Compliance Compliace Compliance Non-Compliance
05 Technical Eligibility of Product Not Accepted Accepted Accepted Not Accepted
06 Technical Eligibility of Firm Not Accepted Accepted Accepted Not Accepted
Non Responsive Responsive Responsive Non Responsive
Evaluated by Technical Committee
Technical Status
Medequips Pvt Ltd Shirazi Trading Al-Kareem Vertex Medical
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
PART - III
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Imran Hashim
Assistant Prof. Paediatric Surgery
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman)
Dr. Aslam Khan
Assistant Prof. of Plastic
& Reconstructive Surgery
(Member)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Shirazi Trading Vertex Medical Al-Kareem Medical
01 Complete Package / Tender (Mandatory) Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes
Responsive Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Shirazi Trading Vertex Medical Al-Kareem Medical
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes Yes Yes
03Up-to-Date After Sales Service Certificate from Manufacturer /
Principal Yes Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes Yes Yes
06 Warranty Yes Yes Yes
Responsive Responsive Responsive
The Children’s Hospital & The Institute of Child Health Lahore
Remarks:
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
Technical EvaluationPackage No. / Bid No. : 15426/ 02 / CH & ICH
Required Quantity: 3
Name of the Equipment: Over Head Electric Baby Warmer
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Evaluated by Technical Committee
Required Item Name
Over Head Electric Baby Warmer Item # 2
Brand GE USA Drager Local
Sr. # Quoted Model Lullaby warmer Baby therm 8004 AK-174
01 Country of Manufacturer USA Germany PAK
02 Manufacturing site India USA/Germany PAK
03 Compliance with defined Quality Standards Compliance Compliance Compliance
04 Specifications Compliance Feature Wise Compliance Compliance Not Accepted
05 Technical Eligibility of Product Accepted Accepted Not Accepted
06 Technical Eligibility of Firm Accepted Accepted Accepted
Responsive Responsive Non-Responsive
Vertex Medical
Technical Status
Shirazi Trading Al-Kareem Medical
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
PART - III
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Imran Hashim
Assistant Prof. Paediatric Surgery
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman)
Dr. Aslam Khan
Assistant Prof. of Plastic &
Reconstructive Surgery
(Member)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Orient Medical Metro C E
01 Complete Package / Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Orient Medical Metro C E
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment N/A N/A
06 Warranty Yes Yes
Responsive ResponsiveRemarks:
Name of the Equipment: BP Apparatus with Neonatal Cuff (Desk Type)
Technical EvaluationPackage No. / Bid No. : 15264/ 03/ CH & ICH
The Children’s Hospital & The Institute of Child Health Lahore
Required Quantity: 8
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Remarks:
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Technical Committee
Required Item Name
BP Apparatus with Neonatal Cuff (Desk Type) Item # 3
Brand MDF Riester
Sr. # Quoted Model 840 Big Ben CAT 1453 round
01 Country of Manufacturer USA Germany
02 Country of Origin of Product USA Germany
03 Compliance with defined Quality Standards Compliance Compliance
04 Specifications Compliance Feature Wise Compliance Compliance
05 Technical Eligibility of Product Accepted Accepted
06 Technical Eligibility of Firm Accepted Accepted
Responsive ResponsiveTechnical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Orient Medical Metro CE
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Imran Hashim
Assistant Prof. Paediatric Surgery
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman)
Dr. Aslam Khan
Assistant Prof. of Plastic &
Reconstructive Surgery
(Member)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Orient Medical Metro C E
01 Complete Package / Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Orient Medical Metro C E
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment N/A N/A
06 Warranty Yes Yes
Responsive ResponsiveRemarks:
Name of the Equipment: BP Apparatus (Floor Type)
Technical Evaluation
Package No. / Bid No. : 15462/ 04/ CH & ICH
The Children’s Hospital & The Institute of Child Health Lahore
Required Quantity: 08
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Remarks:
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Technical Committee
Required Item Name
BP Apparatus (Floor Type) Item # 4
Brand MDF Riester
Sr. # Quoted Model 830 Big Ben Cat 1468 round
01 Country of Manufacturer USA Germany
02 Country of Origin of Product USA Germany
03 Compliance with defined Quality Standards Compliance Compliance
04 Specifications Compliance Feature Wise Compliance Compliance
05 Technical Eligibility of Product Accepted Accepted
06 Technical Eligibility of Firm Accepted Accepted
Responsive ResponsiveTechnical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Orient Medical Metro C E
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Imran Hashim
Assistant Prof. Paediatric Surgery
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman)
Dr. Aslam Khan
Assistant Prof. of Plastic &
Reconstructive Surgery
(Member)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Al Kareem Medi Serve Vertex Medical Total Technology
01 Complete Package / Tender (Mandatory) Yes No Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Responsive Non-Responsive Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Al Kareem Medi Serve Vertex Medical Total Technology
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes Yes Yes Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes Yes Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals No No Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment N/A N/A N/A N/A
06 Warranty Yes Yes Yes Yes
Non-Responsive Non-Responsive Responsive Responsive
Remarks:
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
Required Quantity: 30
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Name of the Equipment: Suction Venturi
The Children’s Hospital & The Institute of Child Health Lahore
Technical Evaluation
Package No. / Bid No. : 15426/ 05 / CH & ICH
Evaluated by Technical Committee
Required Item Name
Suction Venturi Item # 5
Brand Oxylitre GCE TM Flow Meter
Sr. # Quoted Model S723R Not Mentioned TM2 Ejector Easy Air
01 Country of Manufacturer UK Czech Republic France Italy
02 Country of Origin of Product UK Czech Republic France Italy
03 Compliance with defined Quality Standards Compliance Compliance Compliance Compliance
04 Specifications Compliance Feature Wise Compliance Compliance Compliance Compliance
05 Technical Eligibility of Product Accepted Accepted Accepted Accepted
06 Technical Eligibility of Firm Not Accepted Not Accepted Accepted Accepted
Non-Responsive Non-Responsive Responsive ResponsiveTechnical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Al Kareem Medi Serve Vertex Medical Total Technology
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Imran Hashim
Assistant Prof. Paediatric Surgery
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman)
Dr. Aslam Khan
Assistant Prof. of Plastic
& Reconstructive Surgery
(Member)
Nebulizer (Imported)
7
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Medi Serve
01 Complete Package / Tender (Mandatory) Yes
02 Original Receipt of Tender (Mandatory) Yes
03 Affidavit from Bidder Yes
04 Bid Security / CDR (Mandatory) Yes
05 Bid Validity (Mandatory) Yes
06 Delivery Period (Mandatory) Yes
Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Medi Serve
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal No
04 Bidder Past Performance for Quoted Product in Teaching Hospitals No
05 Availability of relevant Tools and Testing / Calibration Equipment N/A
06 Warranty Yes
Non- Responsive
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
Required Quantity:
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Remarks:
The Children’s Hospital & The Institute of Child Health Lahore
Technical EvaluationPackage No. / Bid No. : 15426/ 06 / CH & ICH
Name of the Equipment:
Evaluated by Technical Committee
Name
Nebulizer (Imported) Item # 6
Brand Elmasler
Sr. # Quoted Model UN300A
01 Country of Manufacturer Turkey
02 Country of Origin of Product Turkey
03 Compliance with defined Quality Standards Not Accepted
04 Specifications Compliance Feature Wise Not Accepted
05 Technical Eligibility of Product Not Accepted
06 Technical Eligibility of Firm Not Accepted
Non-Responsive
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Medi Serve
Technical Status
PART - III
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Imran Hashim
Assistant Prof. Paediatric Surgery
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman)
Dr. Aslam Khan
Assistant Prof. of Plastic &
Reconstructive Surgery
(Member)
Defibrillator
1
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Medequips
01 Complete Package / Tender (Mandatory) Yes
02 Original Receipt of Tender (Mandatory) Yes
03 Affidavit from Bidder Yes
04 Bid Security / CDR (Mandatory) Yes
05 Bid Validity (Mandatory) Yes
06 Delivery Period (Mandatory) Yes
Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Medequips
01Exclusive Authorization / Sole Agent Certificate by the
ManufacturerYes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes
03Up-to-Date After Sales Service Certificate from Manufacturer /
Principal Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes
06 Warranty Yes
Responsive
The Children’s Hospital & The Institute of Child Health Lahore
Package No. / Bid No. : 15426/ 08/ CH & ICH
Name of the Equipment:
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
Required Quantity:
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Remarks:
Evaluated by Technical Committee
Name
Defibrillator
Brand NIHON Khoden
Sr. # Quoted Model TEC5631K
01 Country of Manufacturer Japan
02 Country of Origin of Product Japan
03 Compliance with defined Quality Standards Compliance
04 Specifications Compliance Feature Wise Compliance
05 Technical Eligibility of Product Accepted
06 Technical Eligibility of Firm Accepted
ResponsiveTechnical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Medequips
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Imran Hashim
Assistant Prof. Paediatric Surgery
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman)
Dr. Aslam Khan
Assistant Prof. of Plastic &
Reconstructive Surgery
(Member)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Claris Medical Medi Urge Coral Medical Vertex Medical
01 Complete Package / Tender (Mandatory) Yes Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Responsive Responsive Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Claris Medical Medi Urge Coral Medical Vertex Medical
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)N/A N/A N/A N/A
03Up-to-Date After Sales Service Certificate from Manufacturer /
Principal Yes Yes Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes Yes Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment N/A N/A N/A N/A
06 Warranty Yes Yes Yes Yes
Responsive Responsive Responsive Responsive
The Children’s Hospital & The Institute of Child Health Lahore
As per ListRequired Quantity:
Name of the Equipment: General Surgery Instrument (Imported)
Technical EvaluationPackage No. / Bid No. : 15426/ 9/ CH & ICH
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Remarks:
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
Evaluated by Technical Committee
Required Item Name
General Surgery Instrument (Imported)
Item # 9
Brand
Sr. # Quoted Model
01 Country of Manufacturer Germany USA Germany Germany
02 Country of Origin of Product Germany USA Germany Germany
03 Compliance with defined Quality Standards Compliance Compliance Compliance Compliance
04 Specifications Compliance Feature Wise
Not User friendly
,not furnished well
few are fine TNA
Accepted finishing
with adequate
Finishing TA
Not user friendly
,Finishing is crude
TNA
Good Finishing User
friendly TA
05 Technical Eligibility of Product Not Accpeted Accepted Not Accpeted Accepted
06 Technical Eligibility of Firm Accepted Accepted Accepted Accepted
Non Responsive Responsive Non Responsive ResponsiveTechnical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Claris Medical Medi Urge Vertex Medical
ASANUSIntegra (Brand-
Jarit)
B. Braun Aesculap
AG
Coral Medical
Medicon GmbH
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Muhammad Sharif
Prof. of Peads Surgery, Mayo
Hospital, Lahore
(Member)
Dr. Saeed Ahmad
Assistant Prof. of Maxillofacial Surgery
(Member)
Dr. Abdul Latif Sami
Prof. of orthopedic Surgery
(Member)
Dr. Seema Qayyum
Prof. of Peads Ophthalmology
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman)
Dr. M. Nadeem Malik
Associate Prof. Neurosurgery
(Member)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Vertex Medical FAB
01 Complete Package / Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes No
04 Bid Security / CDR (Mandatory) Yes No
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes No
Responsive Non Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Vertex Medical FAB
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes No
02Technical & Engineering capability
(As defined for the specific Bid in specifications)N/A N/A
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes No
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes N/A
05 Availability of relevant Tools and Testing / Calibration Equipment N/A N/A
06 Warranty Yes No
Responsive Non Responsive
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Remarks:
The Children’s Hospital & The Institute of Child Health Lahore
Technical Evaluation
Package No. / Bid No. : 15462/ 11/ CH & ICH
Name of the Equipment: Orthopeadic Instrument Sets (Imported)
Required Quantity: As per list
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Evaluated by Technical Committee
Required Item Name
Orthopeadic Instrument set imported Item # 11
Brand B Braun Medicon
Sr. # Quoted Model Acsculape
01 Country of Manufacturer Germany Germany
02 Country of Origin of Product Germany Germany
03 Compliance with defined Quality Standards Compliance Compliance
04 Specifications Compliance Feature Wise Techincally Accepted Compliance
05 Technical Eligibility of Product Accepted Not Accepted
06 Technical Eligibility of Firm Accepted Not Accepted
Responsive Non Responsive
Vertex Medical FAB
Technical Status
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Saeed Ahmad
Assistant Prof. of Maxillofacial
Surgery
(Member)
Dr. Abdul Latif Sami
Prof. of orthopedic Surgery
(Member)
Dr. Seema Qayyum
Prof. of Peads Ophthalmology
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman
Dr. M. Nadeem Malik
Associate Prof. Neurosurgery
(Member)
Dr. Mian Muhammad Hanif
Prof. of Orthopedic Surgery, LGH
(For Orthopedic Surgery
Instruments Imported)
(Member)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Vertex Medical Medi Urge
01 Complete Package / Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder on Stamp paper Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Vertex Medical Medi Urge
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes Yes
06 Warranty Yes Yes
Responsive ResponsiveRemarks:
Remarks:
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
The Children’s Hospital & The Institute of Child Health Lahore
Technical EvaluationPackage No. / Bid No. : 15426/12/ CH & ICH
Required Quantity: 6
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Name of the Equipment: Humby's Skin Graft Knife
Evaluated by Technical Committee
Required Item Name
Humby's Skin Graft Knife Item # 12
Brand B Bruan Miltex
Sr. # Quoted Model Asculape Watson
01 Country of Manufacturer Germany USA
02 Country of Origin of Product Germany USA/Europe
03 Compliance with defined Quality Standards Compliance Compliance
04 Specifications Compliance Feature Wise Compliance Compliance
05 Technical Eligibility of Product Accepted Accepted
06 Technical Eligibility of Firm Accepted Accepted
Responsive ResponsiveTechnical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Medi Urge Vertex Medical
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Imran Hashim
Assistant Prof. Paediatric Surgery
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman)
Dr. Aslam Khan
Assistant Prof. of Plastic &
Reconstructive Surgery
(Member)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters FDS Ture Dynamic Surgiquips G MED
01 Complete Package / Tender (Mandatory) Yes Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes Yes
03 Affidavit from Bidder on Stamp paper Yes Yes Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes Yes
Responsive Responsive Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters FDS Ture Dynamic Surgiquips G MED
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes Yes Yes Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal No No Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals No No No Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes Yes Yes Yes
06 Warranty Yes Yes No Yes
Non Responsive Non Responsive Non Responsive Responsive
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Remarks:
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
PART - II
Technical Evaluation
The Children’s Hospital & The Institute of Child Health Lahore
Required Quantity:
Remarks:
Part - I
01
Package No. / Bid No. : 15426/13/ CH & ICH
Name of the Equipment: POP Cutter
Evaluated by Technical Committee
Required Item Name
POP Cutter Item # 13
Brand Treu Lawton Dimeda Desutter
Sr. # Quoted Model Not Mentioned CC5
01 Country of Manufacturer Germany Germany Germany Uk
02 Country of Origin of Product Germany Germany Germany Uk
03 Compliance with defined Quality Standards Compliance Compliance Compliance Compliance
04 Specifications Compliance Feature Wise - - - Compliance
05 Technical Eligibility of Product - - - Accepted
06 Technical Eligibility of Firm Not Accepted Not Accepted Not Accepted Accepted
Non ResponsiveNon
Responsive
Non
Responsive ResponsiveTechnical Status
Surgiquips G MED
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
FDS Ture Dynamic
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Imran Hashim
Assistant Prof. Paediatric Surgery
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman)
Dr. Aslam Khan
Assistant Prof. of Plastic & Reconstructive
Surgery
(Member)
High Speed Autoclave
(Flash Autoclave)
1
Sr. # Evaluation Parameters Glow Pak
01 Complete Package / Tender (Mandatory) Yes
02 Original Receipt of Tender (Mandatory) Yes
03 Affidavit from Bidder Yes
04 Bid Security / CDR (Mandatory) Yes
05 Bid Validity (Mandatory) Yes
06 Delivery Period (Mandatory) Yes
Responsive
Sr. # Evaluation Parameters Glow Pak
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes
02 Technical & Engineering capability (As defined for the specific Bid in specifications) Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes
06 Warranty Yes
Responsive
Evaluated by Technical Committee
The Children’s Hospital & The Institute of Child Health Lahore
Technical EvaluationPackage No. / Bid No. : 15426/ 17/ CH & ICH
Name of the Equipment:
Required Quantity:
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Remarks:
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
Name
High Speed Autoclave (Flash Autoclave) Item # 17
Brand W&H
Sr. # Quoted Model Lara 22
01 Country of Manufacturer Austria
02 Country of Origin of Product Austria
03 Compliance with defined Quality Standards Compliance
04 Specifications Compliance Feature Wise Compliance
05 Technical Eligibility of Product Accepted
06 Technical Eligibility of Firm Accepted
ResponsiveTechnical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Evaluated by Technical Committee
Glow Pak
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Asma Mushtaq
Senior Registrar Pead
Neorosurgery
(Member)
Dr. Nadeem Malik
Associate Prof. Peads Neuro
Surgery
(Member)
Dr. Seema Qayyum
Prof. of Peads Ophthalmology
(Chairman)
Dr. Laeeq Ur Rehman
Assist. Prof. Neuro Surgery
(Member)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Vertex Medical Amtronech
01 Complete Package / Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder on Stamp paper Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Vertex Medical Amtronech
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes Yes
06 Warranty Yes Yes
Responsive Responsive
Remarks:
The Children’s Hospital & The Institute of Child Health Lahore
Technical EvaluationPackage No. / Bid No. : 15426/22/ CH & ICH
Name of the Equipment: Neurophysiological Monitoring System
Required Quantity: 1
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
Evaluated by Technical Committee
Required Item Name
Neurophysiological Monitoring System Item # 22
Brand Nihon Khoden Natus
Sr. # Quoted Model NeuroMaster Xltek Protektor32 IOM
01 Country of Manufacturer Japan Canada
02 Country of Origin of Product Japan Canada
03 Compliance with defined Quality Standards Compliance Compliance
04 Specifications Compliance Feature Wise Compliance Compliance
05 Technical Eligibility of Product Accepted Accepted
06 Technical Eligibility of Firm Accepted Accepted
Responsive Responsive
Vertex Medical Amtronech
Technical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Asma Mushtaq
Senior Registrar Pead
Neorosurgery
(Member)
Dr. Nadeem Malik
Associate Prof. Peads Neuro
Surgery
(Member)
Dr. Seema Qayyum
Prof. of Peads Ophthalmology
(Chairman)
Dr. Laeeq Ur Rehman
Assist. Prof. Neuro Surgery
(Member)
1
Evaluated by Purchase Committee
Sr. # Evaluation Parameters National Enterprises BIOS Medi Urge
01 Complete Package / Tender (Mandatory) Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes
Responsive Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters National Enterprises BIOS Medi Urge
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes Yes Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals No Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes Yes Yes
06 Warranty Yes Yes Yes
Non Responsive Responsive Responsive
Part - I
The Children’s Hospital & The Institute of Child Health Lahore
Technical EvaluationPackage No. / Bid No. : 15426/ 23 / CH & ICH
Name of the Equipment:
Required Quantity:
Ultrasonic Surgical Aspirator with
Straight and Curved Hand Pieces
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Remarks:
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
Evaluated by Technical Committee
Ultrasonic Surgical Aspirator with Straight and curved hand pieces
Item # 23
Brand Misonix Soring Integra
Sr. # Quoted Model System n360-0 SONOCA 300 CUSA EXCEL+
01 Country of Manufacturer USA Germany USA
02 Country of Origin of Product SouthAfrica Germany USA
03 Compliance with defined Quality Standards Compliane Compliane Compliance
04 Specifications Compliance Feature Wise - Compliane Compliance
05 Technical Eligibility of Product - Accepted Accepted
06 Technical Eligibility of Firm Not Accepted Accepted Accepted
Non Responsive Responsive Responsive Technical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Medi UrgeBIOSNational Enterprises
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Asma Mushtaq
Senior Registrar Pead
Neorosurgery
(Member)
Dr. Nadeem Malik
Associate Prof. Peads Neuro
Surgery
(Member)
Dr. Seema Qayyum
Prof. of Peads Ophthalmology
(Chairman)
Dr. Laeeq Ur Rehman
Assist. Prof. Neuro Surgery
(Member)
Maxillofacial Surgery Set
(Part-B)
As per list
Sr. # Evaluation Parameters Glow Pak
01 Complete Package / Tender (Mandatory) Yes
02 Original Receipt of Tender (Mandatory) Yes
03 Affidavit from Bidder Yes
04 Bid Security / CDR (Mandatory) Yes
05 Bid Validity (Mandatory) Yes
06 Delivery Period (Mandatory) Yes
Responsive
Sr. # Evaluation Parameters Glow Pak
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes
02 Technical & Engineering capability (As defined for the specific Bid in specifications) Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes
06 Warranty Yes
Responsive
Remarks:
The Children’s Hospital & The Institute of Child Health Lahore
Technical EvaluationPackage No. / Bid No. : 15426/ 25/ CH & ICH
Name of the Equipment:
Required Quantity:
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Technical Committee
Remarks:
Name
Maxillofacial Surgery Set (Part-B) As per list Item # 25
Brand W&H
Sr. # Quoted Model
01 Country of Manufacturer Austria
02 Country of Origin of Product Austria
03 Compliance with defined Quality Standards Compliance
04 Specifications Compliance Feature Wise Compliance
05 Technical Eligibility of Product Accepted
06 Technical Eligibility of Firm Accepted
ResponsiveTechnical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Evaluated by Technical Committee
Glow Pak
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Saeed Ahmad
Assistant Prof. of
Maxillofacial Surgery
(Member)
Dr. Abdul Latif Sami
Prof. of orthopedic Surgery
(Member)
Dr. Seema Qayyum
Prof. of Peads Ophthalmology
(Member)
Dr. Farrukh Mahmood
Prof. of Plastic Surgery
(Member)
Dr. Nabila Talat
Prof. of Paediatric Surgery
(Chairman
Dr. M. Nadeem Malik
Associate Prof. Neurosurgery
(Member)
Evaluated by Purchase CommitteeSr. # Evaluation Parameters M/s Commtel Systems M/s Vital International M/s Taha Enterprises
01 Complete Package / Tender (Mandatory) Yes Yes Yes02 Original Receipt of Tender (Mandatory) Yes Yes Yes03 Affidavit from Bidder Yes Yes Not on Stamp Paper04 Bid Security / CDR (Mandatory) Yes Yes No05 Bid Validity (Mandatory) Yes Yes No06 Delivery Period (Mandatory) Yes Yes No
Responsive Responsive Non-Responsive
Evaluated by Technical CommitteeSr. # Evaluation Parameters M/s Commtel Systems M/s Vital International M/s Taha Enterprises
01 Exclusive Authorization / Sole Agent Certificate by the ManufacturerYes Yes No
02 Technical & Engineering capability
(As defined for the specific Bid in specifications) Yes Yes No
03 Up-to-Date After Sales Service Certificate from Manufacturer /
Principal Yes Yes No
04 Bidder Past Performance for Quoted item Yes Yes No05 Availability of relevant Tools and Testing / Calibration Equipment
Yes Yes No
06 Warranty Yes Yes No
Responsive Responsive Non-Responsive
Name of the Equipment: Video Conference Equipment
The Children’s Hospital & The Institute of Child HealthFerozpur Road, Lahore Phone No. +92 42 99230901 – 23
Technical EvaluationPackage No. / Bid No. : P.C./15426/26/CH&ICH
Required Quantity: 04
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Remarks:
Package No. / Bid No. : P.C./15426/26/CH&ICH
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
Evaluated by Technical Committee M/s Commtel Systems M/s Vital International M/s Taha Enterprises
Required Item Name
Video Conference Equipment # 26
Brand Huawei (VC)
Crown (UPS)
TCL(LED)
Dahua (VC)
APC (UPS)
Sansung (LED)
Yealink (VC)
APC (UPS)
Samsung (LED)Sr. # Quoted Model TE 40 (VC)
CMUS 1200 (UPS)
55P65 (LED)
PH-VCS-TS51A0 (VC)
SRT 1000I (UPS)
UDHK Smart TV NU8000 (LED)
VC-800 (VC)
SRV (UPS)
55KU5350 LED
01 Country of Manufacturer China China China
02 Country of Origin of Product China China China
03 Compliance with defined Quality Standards Yes Yes No04 Specifications Compliance Feature Wise Yes No No
05 Technical Eligibility of Product Yes No No06 Technical Eligibility of Firm Yes Yes No
Responsive Non-Responsive Non-Responsive
Package No. / Bid No. : P.C./15426/26/CH&ICH
Technical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Mr. Mazhar Iqbal Noor
Computer Programmer (Member)
Mr. Muhammad Riaz Shahid
I.T. Incharge (Member / Convener)
Ali Raza
Technical Representative of PITB, Lahore
Dr. Usaid Qureshi
Assistant Prof. of Pead Cardiology
(Member)
Dr. Syed Najam Haider
Prof. of Peads Cardiology
(Chairman)
Muhammad Hafeez
Purchase Officer / Coordinator (Member)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters Orient Medical Metro C E
01 Complete Package / Tender (Mandatory) Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes
03 Affidavit from Bidder Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes
05 Bid Validity (Mandatory) Yes Yes
06 Delivery Period (Mandatory) Yes Yes
Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters Orient Medical Metro C E
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes
02 Technical & Engineering capability
(As defined for the specific Bid in specifications)
Yes Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment N/A N/A
06 Warranty Yes Yes
Responsive Responsive
Remarks:
The Children’s Hospital & The Institute of Child Health Lahore
Technical Evaluation
Package No. / Bid No. : 15462/ 27/ CH & ICH
Name of the Equipment:BP Apparatus with all Cuff sizes (Wall
Mounted)
Required Quantity: 31
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Remarks:
Evaluated by Technical Committee
Required Item Name
BP Apparatus with all Cuff sizes (Wall Mounted)
Brand MDF Riester
Sr. # Quoted Model 840 Big Ben Cat 1459 Round
01 Country of Manufacturer USA Germany
02 Country of Origin of Product USA Germany
03 Compliance with defined Quality Standards Compliance Compliance
04 Specifications Compliance Feature Wise Compliance Compliance
05 Technical Eligibility of Product Accepted Accepted
06 Technical Eligibility of Firm Accepted Accepted
Responsive Responsive
Orient Medical Metro C E
Technical Status
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Usaid Qureshi
Assistant Prof. Paediatric Cardiology
(Member)
Dr. Syed Najam Haider
Prof. of Paediatric Cardiology
(Chairman)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters A.M Sales Roche Hoora International
01 Complete Package / Tender (Mandatory) Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes
Responsive Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters A.M Sales Roche Hoora International
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes Yes Yes
03Up-to-Date After Sales Service Certificate from Manufacturer /
Principal Yes Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals Yes Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes Yes Yes
06 Warranty Yes Yes Yes
Remarks:
The Children’s Hospital & The Institute of Child Health Lahore
Technical Evaluation
Package No. / Bid No. : 15462/ 29/ CH & ICH
Name of the Equipment: ABG Machine
Required Quantity: 01 No.
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Responsive Responsive Responsive
Evaluated by Technical Committee
Required Item Name
ABG Machine Item # 29
Brand Nova Biomedical cobas Seimens Healthnineers
Sr. # Quoted Model Start Profile Phox
OximetertB 221 POC Rapid Lab 348
01 Country of Manufacturer USA Switzerland USA
02 Country of Origin of Product USA Switzerland USA
03 Compliance with defined Quality Standards Yes Yes Yes
04 Specifications Compliance Feature Wise Compliance Compliance Compliance
05 Technical Eligibility of Product Accepted Accepted Accepted
06 Technical Eligibility of Firm Accepted Accepted Accepted
Responsive Responsive Responsive
A.M Sales Hoora International
Technical Status
Remarks:
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Roche
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Usaid Qureshi
Assistant Prof. Paediatric
Cardiology
(Member)
Dr. Syed Najam Haider
Prof. of Paediatric Cardiology
(Chairman)
Evaluated by Purchase Committee
Sr. # Evaluation Parameters FDS KASBAN Total Technologies
01 Complete Package / Tender (Mandatory) Yes Yes Yes
02 Original Receipt of Tender (Mandatory) Yes Yes Yes
03 Affidavit from Bidder Yes Yes Yes
04 Bid Security / CDR (Mandatory) Yes Yes Yes
05 Bid Validity (Mandatory) Yes Yes Yes
06 Delivery Period (Mandatory) Yes Yes Yes
Responsive Responsive Responsive
Evaluated by Technical Committee
Sr. # Evaluation Parameters FDS KASBAN Total Technologies
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes Yes Yes
02Technical & Engineering capability
(As defined for the specific Bid in specifications)Yes Yes Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes Yes Yes
04 Bidder Past Performance for Quoted Product in Teaching Hospitals No Yes Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes Yes Yes
06 Warranty Yes Yes Yes
Non Responsive Responsive Responsive Remarks:
The Children’s Hospital & The Institute of Child Health Lahore
Technical EvaluationPackage No. / Bid No. : 15426/ 31 / CH & ICH
Name of the Equipment:
Required Quantity:
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Remarks:
PART - II
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
2
Air Filtration Sterilizer System for ICUs
Evaluated by Technical Committee
Air Filtration Sterilizer System for ICUs (Item # 31) FDS KASBAN Total Technologies
Brand Misonix Kover NateoSante
Sr. # Quoted Model System n360-0 Genius 485J CTI EOLIS 600
01 Country of Manufacturer USA Italy France
02 Country of Origin of Product SouthAfrica Italy France
03 Compliance with defined Quality Standards Compliance Compliance Compliance
04 Specifications Compliance Feature Wise - Compliance Compliance
05 Technical Eligibility of Product - Accepted Accepted
06 Technical Eligibility of Firm Not Accepted Accepted Accepted
Non Responsive Responsive Responsive
PART - III
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
Technical Status
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Muhammad Hafeez
Purchase Officer/Coordinator
(Member)
Dr. Usaid Qureshi
Assistant Prof. Paediatric Cardiology
(Member)
Dr. Syed Najam Haider
Prof. of Paediatric Cardiology
(Chairman)
Cochlear Implant
(Hearing Devices)
20
Sr. # Evaluation Parameters Med el Pakistan
01 Complete Package / Tender (Mandatory) Yes
02 Original Receipt of Tender (Mandatory) Yes
03 Affidavit from Bidder Yes
04 Bid Security / CDR (Mandatory) Yes
05 Bid Validity (Mandatory) Yes
06 Delivery Period (Mandatory) Yes
Responsive
Sr. # Evaluation Parameters Med el Pakistan
01 Exclusive Authorization / Sole Agent Certificate by the Manufacturer Yes
02 Technical & Engineering capability (As defined for the specific Bid in specifications) Yes
03 Up-to-Date After Sales Service Certificate from Manufacturer / Principal Yes
04 Bidder Past Performance for Quoted Product Yes
05 Availability of relevant Tools and Testing / Calibration Equipment Yes
06 Warranty Yes
Responsive
KNOCKED DOWN CRITERIA - (VENDOR EVALUATION)
Evaluated by Technical Committee
Remarks:
Required Quantity:
Part - I
KNOCKED DOWN CRITERIA - (BID EVALUATION)
Evaluated by Purchase Committee
Remarks:
The Children’s Hospital & The Institute of Child Health Lahore
Technical EvaluationPackage No. / Bid No. : 56918/ 33/ CH & ICH
Name of the Equipment:
PART - II
Name
Choclear implant Item # 33
Brand Med-el
Sr. # Quoted Model Mi1210 & Opous 2
01 Country of Manufacturer Austria
02 Country of Origin of Product Austria
03 Compliance with defined Quality Standards Compliance
04 Specifications Compliance Feature Wise Compliance
05 Technical Eligibility of Product Accepted
06 Technical Eligibility of Firm Accepted
Responsive Technical Status
Evaluated by Technical Committee
Med el Pakistan
KNOCKED DOWN CRITERIA - (PRODCUT EVALUATION)
PART - III
Engr. Irfan Haider
Bio-Medical Engineer
(Member)
Engr. Shujaat Ali
Bio-Medical Engineer
(Member)
Dr Imran Saeed
Associate Professor of Pead ENT
(Member)
Dr. Tahir Rashid
Prof. of ENT, Lahore General
Hospital, Lahore
(Chairman)
Dr. Samra
Senior Medical Officer
Audiology Department
(Member)