thermal mass flowmeter and temperature transmitter mti10

126
F1 F2 F3 F4 INSTALLATION AND MAINTENANCE INSTRUCTIONS IM-8-632-US May 2015 Thermal Mass Flowmeter and Temperature Transmitter MTI10 Insertion and MTL10 Inline

Upload: trinhdieu

Post on 11-Jan-2017

230 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

INSTALLATION AND MAINTENANCE INSTRUCTIONS

IM-8-632-US May 2015

Thermal Mass Flowmeter andTemperature Transmitter

MTI10 Insertion and MTL10 Inline

Page 2: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Notice

2

This publication must be read in its entirety before performing any operation. Failureto understand and follow these instructions could result in serious personal injury

and/or damage to the equipment. Should this equipment require repair or adjustmentbeyond the procudures given herein, contact the factory at:

Spirax Sarco Inc.2150 Miller Drive

Longmont, CO 80501Phone: 800.356.9362

Fax: 303.682.7069www.spiraxsarco.com/us

Spirax Sarco believes that the information provided herein is accurate;

however, be advised that the information contained herein is NOT a guaranteefor satisfactory results. Specifically, this information is neither a warranty nor

guarantee, expressed or implied, regarding performance, merchantability, fitness,or any other matter with respect to the products; nor recommendation for the use

of the product/process information in conflict with any patent. Please note that Spirax Sarco. reserves the right to change and/or improve the product

design and specification without notice.

All Spirax Manuals and software for the MTI10/MTL10 are available in English only.

Page 3: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Table Of Contents1. Introduction ..................................................................................................................Page 42. Installation (Mechanical) ..............................................................................................Page 8 a. Insertion Type .................................................................................................Page 9 b. Inline Type ....................................................................................................Page 133. Wiring (Electrical) ......................................................................................................Page 15 a. Input Power ..................................................................................................Page 17 b. Signal Wiring ................................................................................................Page 19 c. Frequency/Alarm Wiring ...............................................................................Page 21 d. Remote Switch .............................................................................................Page 22 e. Remote Wiring ..............................................................................................Page 234. Operation (Standard Operation) ................................................................................Page 25 a. Start Up ........................................................................................................Page 25 b. Display Screens ............................................................................................Page 26 c. Engineering Screens ....................................................................................Page 27 d. Programming ................................................................................................Page 28 e. Menu Trees ...................................................................................................Page 555. Calibration .................................................................................................................Page 62 a. Introduction ...................................................................................................Page 62 e. Calibration Validation ....................................................................................Page 62 b. Start Up ........................................................................................................Page 66 c. Operation - CAL-V ........................................................................................Page 74 d. Operation - Zero CAL-CHECK .....................................................................Page 76 e. Troubleshooting Calibration Validation .........................................................Page 816. Modbus .....................................................................................................................Page 82 a. Introduction ...................................................................................................Page 82 b. Modbus Protocol ...........................................................................................Page 82 c. RS-485 Wiring ..............................................................................................Page 89 d. Programming Using Local Display ...............................................................Page 907. Maintenance ..............................................................................................................Page 948. Troubleshooting .........................................................................................................Page 979. Appendices ..............................................................................................................Page 104 a.Specifications .............................................................................................Page 104 b. Agency Approvals .......................................................................................Page 107 c. MTI10/MTL10 with 2 Gas Curves ...............................................................Page 108 d. Dimensions ................................................................................................Page 112 e. Installation Variations .................................................................................Page 120 f. Warranty ......................................................................................................Page 122 g. Returning your meter ..................................................................................Page 124

3

Page 4: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Thank you for purchasing the Spirax Sarco MTI10 insertion or MTL10 inline Thermal Mass Gas Flowmeter and Temperature Transmitter. The Model MTI10/MTL10 is one of the most technically advanced flowmetersintheworld.TheMTI10/MTL10utilizethesamesensorandelectronics.TheMTL10includesacompanysuppliedflowtube,whilethe MTI10 is inserted into an existing pipe. Extensive engineering effort hasbeeninvestedtodeliveradvancedfeatures,accuracymeasurementperformance and outstanding reliability.

This Instruction Manual contains the electrical and mechanical installationinstructionsaswellasdetailsforprogramming,maintainingand troubleshooting the meter.

Thismanualisdividedintothefollowingsections:Introduction,Installation,Wiring,Operation,CalibrationValidation,Modbus,Maintenance,Troubleshooting,andAppendices. Theory of OperationThe MTI10/MTL10 is an innovative Thermal Mass Gas Flowmeter andTemperatureTransmitter.Itismicroprocessor-basedandfieldprogrammable. The MTI10/MTL10 thermal sensor operates on the law that gases absorb heat. A heated sensor placed in an air or gas stream transfers heat in proportion to the stream’s mass velocity. There are two sensor elements connected to a balanced bridge circuit. One sensor element detects the gas temperature and a second element is maintained at a constant temperature above the gas temperature. The energy applied to the heated sensor to maintain a constant temperature differential(constant∆T)isdirectlyproportionaltothemassflowvelocity.TheMTI10/MTL10flowmetermaintainsaccurateflowmeasurement over a large temperature and pressure range.

Mass FlowTheModelMTI10/MTL10measuresmassflow;anadvantageoverotherflowmeterswhichmeasurevolumetricflowrate.Volumetricflowis incomplete because temperature and pressure are unknown and mustbemeasuredseparately.Forexample,themassflowofagasdepends on its temperature and pressure. As temperature and pressure changes,thegasvolumechangesbutnotitsmass.Thereforeadevicemeasuringmassflowisindependentoftemperatureandpressure.TheModelMTI10/MTL10providesadirectmeasurementofgasflowinMassunits(kg/hr,lb/hr),standardunits(SCFM,SLPM)ornormalunits(NM3/hr,NLPM)withnoadditionaltemperatureorpressuremeasurementsrequired.

4

Introduction

Calibration Validation

I/O Description

Product Description

Mass Flow

Welcome

Flow Calibration

Page 5: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Calibration ValidationSpirax Sarco has developed a method to validate the calibration of the flowmeterinthefield.ThismethodiscalledCalibrationValidationandit is made up of two distinct tests: CAL-V™ and Zero CAL-CHECK™. The goal of Calibration Validation is to provide operators with the ability to verify that the meter is capturing accurate data at scheduled recalibration times - or at any time - instead of sending the meter back to the factory for recalibration.

ByperformingCAL-V™inthefield,operatorscanverifythatthemeteris running accurately by testing the functionality of the sensor and its associated signal processing circuitry. This test can be done in the pipeandinnormalprocessingconditions.Thesecondtest,ZeroCAL-CHECK™,ensurestheeffectivenessandsensibilityofthesensorata"noflow"condition.

Flow CalibrationSpirax Sarco maintains instrument calibration records on every flowmeter.ThisdatacanalsobeaccessedviaacomputerusingMTI10/MTL10 View software within the instrument. Computer-generated calibrationdocumentsdescribespecificinstrumentdetailsthatcanbesortedbyserialnumber,tagnumberorcustomerpurchaseorder.

Calibrationfilesincludedetailsonprocessconditions,calibrationfluid,linesizeandotherinformation.AllNIST-traceableequipmentutilizedforthecalibrationprocedureisidentified,asisthecalibrationhistoryofallreference equipment.

InadditiontotheCalibrationCertificate,acertifiedflowtablethatcorrelatescurrentoutputswithscaledunitsofflowisproducedforeachcalibrated device.

I/O DescriptionThe MTI10/MTL10 feature two galvanically isolated 4 to 20mA analog outputs: one isolated digital output that can be used for frequency or alarm,oneprogrammablediscreteinputandaUSBconnectionforcommunication with a computer. Thefirst4to20mAoutputisforflowrate.Thesecond4to20mAoutputcanbeconfiguredeitherforflowrateorprocessgastemperature.Both4 to 20mA outputs can be scaled by the user. The frequency output is programmabletorepresentflowrateandcanbescaledformaximum

5

Introduction

Calibration Validation

I/O Description

Flow Calibration

Page 6: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

IR (infrared) Buttons

Display Window

Push Buttons

flow/maximumfrequency,units-per-pulseorpulse-per-units.Themaximumfrequencyis100Hz.Anisolated24VDCoutputpoweroptionis provided for use with these outputs. It can supply a 42mA maximum total load (do not use for other external devices).

MTI10/MTL10 View interfaces to the USB port and is a free PC-basedsoftwareprogramthatdisplaysflowmeterreadingsandpermitsflowmeterconfiguration. RS485Modbus,andindustrystandardcommunicationoption,isavailableontheMTI10/MTL10flowmeter.

Optional Display Panel & Configuration PanelTheconfigurationpanelallowstheusertochangeavarietyofsettings.The display is 2 lines x 16 characters with 4 mechanical and 4 IR (infrared) buttons. The IR and mechanical buttons perform the same function but the IR buttons can be used without opening the cover. The IRbuttonscanbecalibrated(p.54)forbetteroperationinthefieldor disabled (p. 54) when the meter is used in snow or ice in order to avoid false key detection.

Fig.1.1:OptionalDisplayandConfigurationPanel

6

Introduction

Display

Page 7: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

Standard I/O

4 to 20 mA Flow

Contact Input

4 to 20 mA Flow or Temperature

Frequency or Alarm

USB(Free meter View Software)

Outputs and Communications are Galvanically Isolated

Optional Digital Communications

RS485 Modbus

24VDC Input Power100-240VAC Optional

Optional Display andCon�guration Panel

Standard Digital Communications

Functional DiagramAnoptionalon-boarddisplayisavailabletoviewflowrate,totalflow,elapsedtime,processgastemperatureandalarms.ThedisplayisalsousedinconjunctionwiththeConfigurationPanelforfieldconfigurationofflowmetersettingssuchas4to20mAscaling,pulseoutputfrequencyscaling,pipearea,zeroflowcutoff,flowfilteringordamping,displayconfigurations,diagnosticsandalarmlimits.

Fig. 1.2: Functional Diagram

Functional Diagram

7

Introduction

Page 8: Thermal Mass Flowmeter and Temperature Transmitter MTI10

ScopeThis section describes how to install the Spirax Sarco MTI10/MTL10 Flowmeter and how to get started. Installation methods will vary accordingtotheflowmetertype(insertionorinline).

For Insertion Types (MTI10):1. Determine lateral position on the pipe 2. Sensor installation depth 3. Sensororientationinrelationtosensorlengthanddirectionofflow4. Propertighteningofcompressionfittingformountingmeter

For Inline Types (MTL10):1. Determine lateral position on the pipe2. Flowbodyorientationinrelationtodirectionofflowinpipe3. Propertighteningofcompressionfitting

Installation procedures must be performed using a combination of the enduser’sbestengineeringpractices,incompliancewithlocalcodes,and manufacturer’s recommendations.

General PrecautionsThe following general precautions should be observed: 1. Exercisecarewhenhandlingtheflowmetertoavoiddamagingthe

probe,sensororenclosure.2. The enclosure covers must be closed except during installation.3. Mounting MTI10 or MTL10 in direct sunlight can cause the

temperatureinsidetheenclosuretoincreasebeyonddesignlimits,resulting in failure of LCD display and reduced component life. It is recommended that a sunshade be installed to avoid direct sunlight.

4. Ensuretheflowdirectionarrowpointsinthedirectionofflow.5. DonotinstalltheMTI10orMTL10enclosurenearanigniter,igniter-

controller or switching equipment.6. Do not install an external power supply in a cabinet containing an

igniter controller or switching equipment.7. Ensure that good engineering practices and applicable industry

codes are followed throughout the installation process.8. Foraccurateflowmeasurement:reviewflowmeterplacement

instructionsbeforeinstallationtoensureaproperflowprofileinthepipe.

Installation -MTI10/MTL10 Flowmeter

8

Installation: General

Insertion Flowmeter Lateral Placement

Page 9: Thermal Mass Flowmeter and Temperature Transmitter MTI10

FLOW

10X Pipe ID

ProperProfile

15X Pipe ID

IrregularProfile

F1 F2 F3 F4

F1 F2 F3 F4

Instructions for Insertion Flowmeter Lateral PlacementInstalltheModelMTI10Insertionstyleflowmetersothatitisfarenoughawayfrombendsinthepipe,obstructions,orchangesinlinesizestoensureaconsistentflowprofile.Fifteendiametersofstraightpipeupstream and ten downstream are recommended.

Forexample,a2"pipewouldrequire30"upstreamand20"downstream,but a 4" pipe would require 60" upstream and 40" downstream.

Fig. 2.1: Upstream and Downstream Pipe IDs for Insertion Meters

Special Conditions of Use:• Consult the manufacturer if dimensional information on the

flameproofjointsisnecessary.• Follow the manufacturer's instructions to reduce the potential of

anelectrostaticcharginghazard.

9

Installation: Insertion Type

Insertion Flowmeter Lateral Placement

Page 10: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Half Coupling,3/4" NPT Female(Supplied byCustomer)

Compression Fitting(Supplied by Spirax Sarco)

Customer’s Pipe

22mm (0.87”)LC

F1 F2 F3 F4

Installation DepthThe installation depth of the sensor in the pipe is dependent on the pipe size.Togetthemostaccuratereading,properplacementofthesensorwindowwithinthepipeisnecessary.AsshowninFig2.2,theendofthe sensor window should be 22mm (0.87") past the center line of the pipe.

Fig. 2.2: Cross Section of Insertion Sensor Depth in Pipe

10

Installation: Insertion Type

Installation Depth Sensor Orientation - Direction of Flow

Page 11: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

FLOW

Pipe

Flow Direction Arrow

+10°

-10°

FLOW

FLOW

Fig. 2.3: Orientation of MTI10 Insertion Type Flowmeter

Note:Someflowmetersareshippedwiththesensorelementsthatareoffset(seefigure2.4).Othersareshippedwithsensorsthathaveequallengthelements(seefigure2.5).Thesensortypesuppliedwasselectedat the factory to be the best suited for your application. Follow the appropriate sensor orientation instructions.

Unequal Length Sensor ElementsInstall the shorter sensor element upstream from the longer one.

Fig. 2.4: Unequal Length Sensor Elements

11

Installation: Insertion Type

Sensor Orientation - Direction of Flow

Sensor Orientation - Unequal Length Sensors

Page 12: Thermal Mass Flowmeter and Temperature Transmitter MTI10

+10°

-10°

FLOW

FLOW

Equal Length Sensor ElementsInstallflowmeterwithboth sensor elementsfacingtheflowstreamwithin ±3mm (0.1”). Fig. 2.5: Equal Length Sensor Elements

Insertion Mounting Instructions - Compression FittingsThe MTI10 is mounted through a ¾" hole and a ¾" female NPT half couplingprovidedinthecustomer'spipe.Insertionstyleflowmetersarenot designed for use in pipes smaller than 1½".• Installthecompressionfittingintothe¾-inchfemaleNPThalf

coupling.• Wheninstallingina50mm(2")pipeorlarger,installtheendofthe

probe 22mm (0.87") past the center line of the pipe and tighten the compressionfittingnut(refertoFigure2.2onpage10).

• When installing into a 40mm (1.5") pipe carefully install the probe into the pipe until it touches the opposite wall and pull back 3mm (0.1”).Tightenthecompressionfittingnut.

Caution:Oncethecompressionfittingislockedontotheprobe,theprobecanberemovedorrotated,buttheinsertiondepthislockedinplace.Note: Donotovertightencompressionfitting.

Fig. 2.6: Proper Tightening of the Compression Fitting Nut

Whileholdingthefittingbodysteady,tightenthenutoneandone-quarterturntothe9o'clock position.

Installation: Inline Type

Flowmeter Placement -Inline Type

1212

Installation: Insertion Type

Mounting -Insertion Type

Sensor Orientation - Equal Length Sensors

Page 13: Thermal Mass Flowmeter and Temperature Transmitter MTI10

FLOW

4X Pipe ID

ProperProfile

8X Pipe ID

IrregularProfile

F1 F2 F3 F4

F1 F2 F3 F4

Instructions for Inline Flowmeter PlacementInstalltheMTL10Inlinestyleflowmetersothatitisfarenoughawayfrombendsinthepipe,obstructions,orchangesinlinesizestoensureaconsistentflowprofile.Eightdiametersofstraightpipeupstreamandfour downstream are recommended (for ¼" meters: 6” (152 mm) of straight,unobstructedpipeupstreamanddownstreamarerequired).

Forexample,a2"pipewouldrequire16"upstreamfromtheedgeoftheflowbodyand8"downstreamfromtheotherendoftheflowbody,whereas a 4" pipe would require 32" upstream and 16" downstream.

TheMTL10iswelded,threadedorflangedtothecustomer’spipe.Careshould be taken to ensure that the diameter of the mating pipe is the samediameterastheMTL10flowbodyorerrorsinflowreadingscanoccur. The installation procedure should be a combination of the end user’sbestengineeringpractices,incompliancewithlocalcodes,andthe manufacturer’s recommendations.

See Figure 2.7 for a detailed look at upstream and downstream pipe diameters for inline meters.

Fig. 2.7: Upstream and Downstream Pipe IDs for Inline Meters

13

Installation: Inline Type

Flowmeter Placement -Inline Type

Page 14: Thermal Mass Flowmeter and Temperature Transmitter MTI10

FLOW

Flow Body

F1 F2 F3 F4

Inline OrientationInstalltheflowbodysothattheengravedarrowonthefittingandthearrowontheflowbodyarepointingwiththedirectionofflow.

Fig. 2.8: Orientation of an Inline Meter - Directional Arrows

Tightening Compression FittingsThecompressionfittinghasbeenplacedaccordingtotheproperdepthintheflowbodybySpiraxSarcofactorytechnicians.Aftertheflowbodyhasbeencorrectlyfittedtotheprocesspipe,thecompressionfittingmay

Note:• Refer to the Spirax Sarco Calibration Validation Chapter

forinformationonsettingthefieldbaselineforZeroCAL-CHECK™ tests if you plan to perform these tests in the pipe.

• Please save the PVC sensor cover that was shipped with your meter. It will be needed to perform Zero CAL-CHECK™ tests out of pipe.

Wiring: General

Scope

Precautions

Power Wiring

14

Installation: Inline Type

Mounting - Inline Type

Flow Body Orientation -Inline Type

Page 15: Thermal Mass Flowmeter and Temperature Transmitter MTI10

needtobetightenedcorrectly(seefigure2.6onp.12).Wiring InstructionsWiretheMTI10/MTL10byopeningtherearenclosurecover,bringingcustomer-supplied wires in through the conduit openings and connecting to the terminal blocks. The MTI10/MTL10 has two conduit openings to maintain separation between AC input power and output signal wiring. To eliminate thepossibilityofnoiseinterference,useaseparateconduitforACpowerandcut all wires short for a minimum service loop.

Wiring Precautions• WARNING - DO NOT OPEN WHEN ENERGIZED OR AN EXPLOSIVE

ATMOSPHERE IS PRESENT.• Allplumbingandelectricalinstallationsofflowmetersmustbein

compliancewithlocalcodes,theenduser’sbestengineeringpractices,and manufacturer’s recommendations.

• An external power disconnect and 16A over-current protection are required for the AC and DC powered MTI10/MTL10

• DonotinstalltheMTI10/MTL10enclosurenearanigniter,igniter-controller or switching equipment.

• Do not install an external power supply in a cabinet containing an igniter controller or switching equipment.

• Thisflowmetercontainscomponentsthatcanbedamagedbystaticelectricity. You must discharge yourself by touching a grounded steel pipe orothergroundedsteelmaterialpriortoworkinginsidethisflowmeter.

• Fortheremotesensoroption,theserialnumberoftheelectronicsenclosure must match the remote sensor probe.

• Closeanyunusedentriesusingsuitablycertifiedplugs Power Wiring Forpowerwiring,usestrandedcopperwire,nolargerthan16-gauge.Ifanexternal24VDCpowersourceisused,twistedpairshieldedcableisrecommended. Supply connection wiring must be rated for at least 90°C (194°F).

GroundingThe enclosure must be properly grounded with a quality earth ground. 16 gauge,strandedwireisrecommended.

Signal WiringForsignalwiring,therecommendedwiregaugeis18to22AWG.Alwaysusetwisted pair shielded cable. The cable shield should not be connected at the flowmeter,itshouldbeconnectedatthepowersupplyACgroundterminalorinstrumentation AC ground. Do not route the power and signal wires in the same conduit. Power wires must enter left-hand conduit entry. Signal and remote sensor (where applicable) must enter right-hand conduit entry.

15

Wiring: General

Scope

Precautions

Power Wiring

Grounding

Signal Wiring

Page 16: Thermal Mass Flowmeter and Temperature Transmitter MTI10

2x 3/4 inch NPT Female F1 F2 F3 F4

Signal Wiring,Serial Communication,Remote Sensor Wiring

InputPower

Serial Communication WiringIfyouhavepurchasedtheRS485communicationoption,pleaserefer to the Modbus section of this manual on p. 89.

Remote Sensor WiringNote: Remote wiring is only required when the Remote Electronics optionsisprovided.Fivewireshieldedcablerequired,therecommended wire gauge is 18AWG. Make sure that the cable length does not exceed 100 feet and the wire resistance does not exceed one ohm. Do not connect the cable shield at the electronics enclosure end.

Fig. 3.1: MTI10/MTL10 Wiring

Note:Serial numbers: Ifyouhavemorethanonemeter,youmustensurethattheserialnumbersoftheprobe/J-Box,remoteelectronics,housing,and/orflowbodymatchoneanother.Theseitems have been manufactured and calibrated to operate as a unit and cannot be mismatched.

Installation wiring: Obtain the correct length for the MTI10/MTL10 power and signal wires using one of these methods:• Trim the wires to extend 2.5 inches out of the enclosure

after the conduit and wires are routed to the MTI10/MTL10 (preferred method).

• Trim the wires to extend 6 inches from the end of the conduit before it is attached to the MTI10/MTL10.

16

Wiring: General

Serial Communication Wiring

Remote Sensor Wiring

MTI10/MTL10 Wiring

Power Input Wiring

Page 17: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Power Input Requirements: 24VDC SupplyExternalDCpowersupplymustprovide24VDC±10%,at0.7Ampsminimum.

The enclosure must be properly grounded with a quality earth ground. Sixteen(16)gauge,strandedwire,isrecommendedforpowerandearthground.

Fig. 3.2: Connections for 24VDC Supply

4-20FLOW

TS424VOUT

4-20#2

Red 1

Red2

Yel 3

Wh 4

Wh 5

PULSE/ALM

TS2

TS1

F1

J2

1

24V, 0.75A

2 3

IN

+1

-2 +1

-2

+3

-4+5

-6+7

-8

TS3REMOTESENSOR

Earth Ground

24V Return

+24VDC

Caution:• Supply connection wiring must be rated for at least

90°C (194° F).

17

Wiring: Input Power

Power Input Wiring

Page 18: Thermal Mass Flowmeter and Temperature Transmitter MTI10

4-20FLOW

TS424VOUT

4-20#2

Red 1

Red2

Yel 3

Wh 4

Wh 5

PULSE

TS2

TS1

F1

J2

1

24V, 0.75A

2 3

IN

+1

-2 +1

-2

+3

-4+5

-6+7

-8

TS3REMOTESENSOR

Earth Ground

AC (L)

AC (N)

Power Input Requirements: 100 to 240VAC SupplyIftheMTI10/MTL10hastheACpoweroption,theACpowermustprovide100 to 240VAC -15% / +10% (85 to 264VAC) at 0.2 Amps minimum.

The enclosure must be properly grounded with a quality earth ground. Sixteen(16)gauge,strandedwire,isrecommended.

Fig. 3.3: Connections for optional AC Power

Caution:• Supply connection wiring must be rated for at least

90°C (194° F).

18

Wiring: Input Power

Power Input Wiring 4 to 20mA Loop Power Provided by Customer(Recommended)

Page 19: Thermal Mass Flowmeter and Temperature Transmitter MTI10

4-20FLOW

TS424VOUT

4-20#2

Red 1

Red2

Yel 3

Wh 4

Wh 5

PULSE/ALM

TS2

TS1

F1

J2

1

24V, 0.75A

2 3

IN

+1

+1 -2

-2

+3

-4

+5

-6+7

-8

TS3REMOTESENSOR

+24VDC

+24VDC

4 to 20mA FLOW RATE

4 to 20mA TEMPERATURE OR FLOW RATE

24VDC Return

Customer PLC or DCSMTI10MTL10

24VDC Return

* (see important note below)

* (see important note below)

4 to 20mA Output Wiring: Customer-Supplied Power SourceBring the 4 to 20mA wiring in through the right-hand conduit hub. ConnectFLOWRATE4to20mAwiringtoTS2,1(+)&2(-).Connect4to20mAoutput#2wiringtoTS2,3(+)&4(-).

Fig. 3.4: 4 to 20mA Output Wiring for Customer-Supplied Power Source

Important NoteThe load resistor on the MTI10/MTL10 Flowmeter 4 to 20mA signal is typically 250 ohms and is located in or at the customers PLC or DCS. A 250 ohm resistor in the 4 to 20mA line will result in a 1 to 5VDC signal to the PLC or DCS. Some PLC/DCS equipment has theloadresistorbuiltintotheunit;pleaserefertothePLC/DCStechnical manual. Do not exceed a 600 ohm load on the MTI10/MTL10 Flowmeter 4 to 20mA signal.

19

Wiring: Signal Wiring

4 to 20mA Loop Power Provided by Customer(Recommended)

Page 20: Thermal Mass Flowmeter and Temperature Transmitter MTI10

4-20FLOW

TS424VOUT

4-20#2

Red 1

Red2

Yel 3

Wh 4

Wh 5

PULSE/ALM

TS2

TS1

F1

J1

1

24V, 0.75A

2 3

IN

+1

+1 -2

-2

+3

-4+5

-6+7

-8

TS3REMOTESENSOR 4 to 20mA FLOW RATE

4 to 20mA TEMPERATURE OR FLOW RATE

Customer PLC or DCSMTI10MTL10

+

(+)1( )2

+

-

-

TS424VOUT

- * (see important note below)

* (see important note below)

4 to 20mA Output Wiring: Loop Power Provided by MTI10/MTL10Bring the 4 to 20mA wiring in through the right-hand conduit hub. Connect the 4 to 20mA wiring to terminal blocks TS2 and TS4 as shown in the diagram below.

Fig. 3.5: 4 to 20mA Output Wiring for Loop Power Provided by MTI10/MTL10

Note: This wiring option is only available with the isolated 24V ouput power option.

Important NoteThe load resistor on the MTI10/MTL10 Flowmeter 4 to 20mA signal is typically 250 ohms and is located in or at the customers PLC or DCS. A 250 ohm resistor in the 4 to 20mA line will result in a 1 to 5VDC signal to the PLC or DCS. Some PLC/DCS equipment has theloadresistorbuiltintotheunit;pleaserefertothePLC/DCStechnical manual. Do not exceed a 600 ohm load on the MTI10/MTL10 Flowmeter 4 to 20mA signal.

Caution:

Do not exceed 42mA total load (ie including 4-20mA outputs).

20

Wiring: Signal Wiring

4 to 20mA Loop Power Provided by MTI10/MTL10

Frequency/Alarm Output Wiring

Page 21: Thermal Mass Flowmeter and Temperature Transmitter MTI10

4-20FLOW

TS424VOUT

4-20#2

Red 1

Red2

Yel 3

Wh 4

Wh 5

PULSE/ALM

TS2

TS1

F1

J1

1

24V, 0.75A

2 3

IN

+1

+1 -2

-2

+3

-4+5

-6+7

-8

TS3REMOTESENSOR

2.4K to 10K OHM

Frequency or Alarm Output

Customer PLC or DCSMTI10MTL10

+

(-)2 (+)1

TS424VOUT

4-20FLOW

TS424VOUT

4-20#2

Red 1

Red2

Yel 3

Wh 4

Wh 5

PULSE/ALM

TS2

TS1

F1

J1

1

24V, 0.75A

2 3

IN

+1

+1 -2

-2

+3

-4+5

-6+7

-8

TS3REMOTESENSOR +24VDC

24VDC Return

2.4K to 10K OHM

Frequency or Alarm Output

Customer PLC or DCSMTI10MTL10

Frequency/Alarm Output WiringBring frequency/alarm wiring in through the right-hand conduit hub. ConnecttoTS2,5(+)&6(-).Thefrequency/alarmoutputisanopencollector circuit capable of sinking a maximum of 10mA of current. Frequency or Alarm selection is programmed using the display. Only oneoption,frequencyoralarm,canbeactiveatatime.

Fig. 3.6: Frequency/Alarm Output Isolated (Recommended)

Fig. 3.7: Frequency/Alarm Output Local +24V Power Option

Note:

The MTI10/MTL10 Frequency/Alarm output is typically used to drive digital circuitry or solid-state relays. The output of a solid staterelaymay,inturn,operateloads such as electromechanical relays or alarm indicators.

Caution:

Do not exceed 42mA total load (ie including 4-20mA outputs).

21

Wiring: Frequency/Alarm Wiring

Frequency/Alarm Output Wiring

Page 22: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Remote Switch WiringAremoteswitchcanbeusedtoresettheTotalizerandelapsedtime,ifenabled in the programming settings. There is no polarity requirement ontheseconnections.UseTS2,7(+)&8(-).

Whenthe2gascurveoptionisordered,theswitchcanbeusedtoswitch between curves.

Fig. 3.8: Remote Switch Wiring

IfyouhavepurchasedRS485Modbuscommunications,pleaserefer to the Modbus chapter in the IMI.

22

4-20FLOW

4-20#2

PULSE/ALM

TS2

TS1

F1

J1

1

24V, 0.75A

2 3

IN

+1

-2

+3

-4+5

-6+7

-8

EXTERNAL SWITCH

FT3 Customer PLC or DCS

Wiring: Remote Switch

Remote Switch Wiring

MTI10MTL10

Page 23: Thermal Mass Flowmeter and Temperature Transmitter MTI10

2X 3/4 " NPT,Female

Remote Cable, 5 Conductor,Shielded, 100ft (30.48m) Max.

F1 F2 F3 F4

InputPower

Signal Wiring

Remote WiringRemote wiring will be the same for both MTI10 insertion and MTL10 inline flowmeters.

Fig. 3.9: Remote Wiring

Signal Wiring includes:4to20mA,pulse,alarmoutput,contactinput,remoteswitch,USB,and communications options. Power input is 24VDC or optional 100-240VAC (+10% / -15%).

Note: Remote wiring is only required when the Remote Electronics option is provided.

Five wire shielded cable required. The shielded cable should be run through a separate grounded steel conduit (no other cables or wires in theconduit).Ifyouareusingyourowncable,makesurethatthecablelength does not exceed 100 feet and has a wire resistance that does not exceed one ohm (18 AWG recommended).

Do not connect the cable shield at the electronics enclosure end.

Use an extension cable to connect the terminals of the remote probe enclosure to connector TS3 of the electronics enclosure as shown in Figure 3.10 and Table 3.1 (p. 24).

23

Wiring: Remote Sensor Option

Remote Wiring

Page 24: Thermal Mass Flowmeter and Temperature Transmitter MTI10

TS424VOUT

TS 5

TS1

F1

J2

1

24V,0.75A

2 3

-2 +1

TS3

#4

#6

#3

#5

#2

#1

Black Dot Denotes Pin #1

Cable Shield

Electronics Enclosure

RED

RED

WHTWHT

GRNWHT

Shield

GRN

BRN

YEL

BLK

RED

Probe Sensor Wired By FoxSENSOR

RED*

BLK*

BRN*

WHT*

GRN*

1

2

3

4

5

RED

RED

YEL

WHT

WHT

Remote Enclosure

Sensor Wires

Fig. 3.10: Remote Sensor Wiring

*Wire colors listed here represent the wire colors of cables supplied by Spirax Sarco. Colors may vary if customer is supplying their own cable.

Table 3.1: Remote Sensor Cable Wiring

24

Wiring: Remote Sensor

Electronics Enclosure

Terminal Numbers

Extension CableWire Color

Remote EnclosureTerminal Numbers

Sensor WireColor

1 Red 1 Red2 Black 2 Red3 Brown 3 Yellow

No Connection Shield 4 Green4 White 5 White5 Green 6 White

Remote Sensor Cable Wiring

Input WiringRemote Sensor

Operation: Start Up

Start UpSequence

Page 25: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Start Up SequenceThe program automatically enters the Run/Measure mode after power up.IftheLocaldisplayisinstalled,thescreenwillshowthesoftwareversions for the MTI10/MTL10 and the display module during power up.ProgrammingoftheflowmetercanalsobeaccomplishedusingaWindows-based PC program called MTI10/MTL10 View.

USB InterfaceThe USB interface is a standard feature which allows communication to aPCinordertomonitorreadingsandconfiguresettings.MTI10/MTL10View,isafreeapplicationprogramfromSpiraxSarcothatconnectstotheUSBinterfaceandallowsdatamonitoring,configurationsetting,dataloggingtoExcel,andanoptiontosaveandrecallMTI10/MTL10configurationdata.Aserialcommunicationmanualisavailableforuserswho want to create their own PC application.

MTI10/MTL10 Optional Display Panel & Configuration PanelThe MTI10/MTL10 display is a 2 line x 16 character display with 4 mechanical and 4 IR (infrared) buttons. The IR and mechanical buttons perform the same function but the IR buttons can be used without opening the cover. The IR buttons can be calibrated (p. 54) for better operationinthefieldordisabled(p.54)whenthemeterisusedinsnow or ice in order to avoid false key detection.

Fig.4.1:MTI10/MTL10OptionalDisplayandConfigurationPanel

F1 F2 F3 F4

IR (infrared) Buttons

Display Window

Push Buttons

25

Operation: Start Up

Start UpSequence

Page 26: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Measurement ModeInthemeasurementmode,therearefourdifferentdisplayscreens(display1,2,3andapromptscreentoentertheprogrammingmode),two display screens are user programmable (refer to Display Setup p. 36). Scrolling through the display is accomplished by pressing the F1 or F2 key to view the next or previous screen. Pressing the F1 and F2 keys at the same time enters the Engineering Menu screens (display 10 through 26). Keys F1 and F2 are used to scroll through the different screens and key F4 in order to exit to Display #1. Pressing the F3 and F4 keys at the same time brings up the Reset Total screen (see p. 46) prompt.

Fig. 4.2: MTI10/MTL10 Display Screen Navigation

Operation: Display Screens

Local Display Screens

26

F1 key: Moves up one screenF2 key: Moves down one screen

(User programable screen)

(User programable screen)

Enter “totalizer reset screen” when F3 & F4 are pressed at the same time

(Fixed screen)

Enter programming screenRequires password.Default is 1234.

Display #1

Display #2

Display #3

Display #4(Fixed screen)

F1

F1

F1

F1 F2

F2 F4

F2

F2 F3 F4

1456.5 SCFM123456 SCF

Elp = 14.6 HR88.5˚ F

Alarm = NoneElp = 14.6 HR

Set Parameter?No Yes

Page 27: Thermal Mass Flowmeter and Temperature Transmitter MTI10

3124.6 SCFMcsv=0.3432 Volt

Enter: Press F1 & F2 at the same timePress F4 to return to normal mode

Display 10

CsvAv=366809Vel=112345.7 FT/M Display 11

FloFlt=3666805.3Vel=2356.45 M/H Display 12

TsiAvr=512.5 cntTsvAvr=323.7 cnt Display 13

Tsi= 2.1345 VoltTsv=0.9856 Volt Display 14

Tsi = 0.0435 AmpTsi = 221.5 Ohm Display 15

RTD9= 345.5 cntGas Temp=123.7 °C Display 16

CH1_420=2167 cntCH2_420=1234 cnt Display 17

Feq=1234.5 cntAlarm=33,35 Display 18

FloHi= 1234 SCFMFloLo=0 SCFM Display 19

TmpHi=300 °CTmpLo=10 °C Display 20

Elp=12.5 HRStat(hex)=2800 Display 21

FT3 V3.02dDisplay V2.03b Display 22

Pwr_Cycl=24Err_tot=0 Display 23

Tsi=221.5 OhmRTD9=10.3 Ohm Display 24

CAL-V=23.51CAL-V Chk=0.2% Display 25

F2 KeyF1 Key

BrShtDnCnt=0 cnt Display 26

F3 & F4 pressed at the same time will initiate a "Total" reset

Flow in selected unitSensor voltage in volts

Sensor average voltsVelocity in selected unit

Sensor filtered average in voltsVelocity in meters/hour

TSI average countTSV average count

TSI in voltsTSV in volts

TSI current in ampsTSI resistance in ohms

RTD9 countGas Temperature in degrees C

CH1 4-20ma current loop countCH2 4-20 ma current loop count

Frequency output countAlarm codes

High flow limit alarmLow flow limit alarm

High temperature limit alarmLow temperature limit alarm

Elapsed time in hoursStatus in hexadecimal

MTI10/MTL10 main board firmware revisionMTI10/MTL10 display board firmware revision

Power cycle countError with totalizer count

TSI resistance in ohmsRTD9 resistance in ohms

CAL-V ValueCAL-V last verify value

Bridge shutdown detection count

ZRO_Pref=xx.xxxxZRO_diff=x.xx%

ZRO_Bref=xx.xxxxZRO_diff=x.xx%

Display 27

Display 28

Zero CAL-CHECK Pipe RefZero CAL-CHECK % difference

Zero CAL-CHECK Bottle RefZero CAL-CHECK % difference

MTI10/MTL10 Engineering DisplaysPressingtheF1&F2keysatthesametimeinthenormalmode,bringsup the engineering displays. These displays show internal parameters of the MTI10/MTL10 which are used by Spirax Sarco service technicians.

PressF4toexit.UsetheF1&F2keystonavigate.

Fig. 4.3: MTI10/MTL10 Engineering Displays

27

Operation: Engineering Displays

EngineeringDisplay

Page 28: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Data Entry using the local display moduleThere are 2 basic types of menu entries: one for changing value or string and one for selecting from a selection list.

To Change a Value or String :

Press CHG (F1)keytochangethevalue,OK (F4) to accept the value.

Press the UP (F1) or DN (F2)keytoselectanewdigitorcharacter,thecursor points to the selected digit. Press NXT (F3) to select the next digit and OK (F4) to accept the entry.

Note: If the UP (F1) or DN (F2)keyishelddownformorethan1second,the program will progressively select new digits at increasing speed as time increases.

To Select from a List:

Press NXT (F1) key repeatedly until the correct selection is made and OK (F4) key to accept the entry.

Entering the Programming Mode

Toentertheprogrammingmode,presstheF1 or F2 key repeatedly in the normal running mode until the following screen is shown:

VALUE = 0.91234CHG OK

VALUE = 0.91234UP DN NXT OK

FLO UNT = SCFMNXT OK

28

Operation: Programming

Programming Usingthe Local Display

Page 29: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Press YES (F4) and the following screen will prompt the user to enter the password if it is active:

Enterthecorrectpassword,thenfollowtheinstructionsforchangingavalueasspecifiedabove.ThedefaultLevel1passwordis“1234”.

Ifthewrongpasswordisentered,themessage“WrongPassword”willbe displayed for a few seconds and then returns to the programming entry screen. Ifthepasswordisaccepted,thebaseprogrammingscreenwillbeshown:

This is the base screen for the programming mode.

Press EXIT (F4)repeatedlyuntil“NormalMode”isseenbrieflytoexitthe programming mode.

Analog 4 to 20mA OutputThe following menu allows the scaling of the analog 4 to 20mA outputs.

Fromthebasescreen,pressI/O (F1) and then in the next screen press 420 (F3).

PASWD:_UP DN NXT OK

SET PARAMETERSI/O FLO DSP EXIT

SET I/OCOM I/O 420 EXIT

SET PARAMETERS ?No Yes

29

Programming Usingthe Local Display

Page 30: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Select CH1 (F1) to program channel 1.

Enter the value for the 20mA and press OK (F4) key to accept the setting.

Then the following screen will display:

Enter the value for the 4mA and press OK (F4).

Note: 4mA is normally set to 0.

This menu allows the user to select an alarm level on the 4 to 20mA when a serious issue is detected that is preventing the calculation of a correct flowvalue.

The options are:Force the 4 to 20mA signal to 3.6 mAForce the 4 to 20mA signal to 21 mADo not force 4 to 20mA signal (not used)

Press (F4)repeatedlyuntil“NormalMode”isseenbrieflytoexittheprogramming mode.

Note:Whentheflowrateexceedstheprogrammedvalueforthe20mAsetpoint,theanalogoutputwillstayat20mAandanalarm code will be generated.

SET 4-20 mACH1 CH2 EXIT

4 mA = 0 SCFMCHG OK

20 mA = 3751 SCFMCHG OK

mA Fault = 3.6 mANXT OK

30

Programming Usingthe Local Display

Operation: Programming

Page 31: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Select CH2 (F2) to program channel 2.Channel2isprogrammableforflow(CH2=Flow)ortemperature(CH2=Temp).

Press NXT (F1) to select Flow or Temperature and then press OK (F4).

Enter the value for the 20mA and press OK (F4) key to accept setting.

Then the following screen will show:

Enter the value for the 4mA and press OK (F4). Press EXIT (F4)repeatedlyuntil“NormalMode”isseenbrieflytoexittheprogramming mode.

Note:Whentheflowrateexceedstheprogrammedvalueforthe20mAsetpoint,theanalogoutputwillstayat20mAandanalarmcodewillbegenerated.

Frequency OutputFromthemainmenu,pressI/O (F1), I/O (F2) and then OUT (F2)

CH2= TempNXT OK

20 mA = 300 ° FCHG OK

4 mA = 10 ° FCHG OK

SET I/OINP OUT EXIT

31

Operation: Programming

Programming Usingthe Local Display

Page 32: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Press OUT (F2) to select output and the following screen may show:

Press NEXT (F1) to cycle through output options until you have the selection for "OUT=Frequency" and press OK (F4).

Thefrequencyoutputcanbeconfiguredinoneofthreeways:(1)specifyingamaximumfrequencytoadefinedmaximumvalueofflowrate,(2)specifyinghowmanyflowunitstotalperpulse,U/P(i.e.,0.1SCFperpulse)or(3)specifyinghowmanypulsesperunit,P/U(i.e.,10pulsesper SCF). All of these approaches are equivalent.

Use P/U (F1)toenterpulseperunit,U/P (F2) for Unit per pulse or FEQ (F3)toentertheflowandmaximumfrequencytoscalethefrequencyoutput.

Note:Whendataisenteredwithanyofthethreedescribedmethods,theother values will be re-calculated according to the settings.

Entering data in Pulse per Unit:Press P/U (F1) and the following screen will show:

Press CHG (F1) to change the setting and then OK (F4) to accept entry.

Thevalueenteredisinpulseperselectedflowunittotal(i.e.,10pulsesper SCF)

OUT = FrequencyNXT OK

PLS/UNT = 1.2CHG OK

FREQUENCY OUTPUT P/U U/P FEQ EXIT

32

Operation: Programming

Programming Usingthe Local Display

Pulse per Unit

Page 33: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Entering data in Unit per Pulse:Press U/P (F2) and the following screen will show:

Press CHG (F1) to change the setting and then OK (F4) to accept entry.Thevalueenteredisinunitperpulse(i.e.0.01flowunittotalperpulse)

Entering data with flow and maximum frequency:Press FEQ (F3) and the following screen will show:

Enter the maximum frequency and press OK (F4)(Maximumfrequencyshouldnotexceed100Hz)

The next screen will show:

Note:Whentheflowrateexceedsthemaximumfrequencysetpoint,theoutput will stay at that maximum frequency but the MTI10/MTL10 will issue an alarm code.

Caution:Equationtoensurepulseratemustnotexceed100Hz.

Alarm OutputToprogramtheAlarmoutput,pressI/O (F1) key from the "SET PARAMETERS"menuscreen,thenselectI/O (F2) and the screen will show:.

Then press OUT (F2) and the screen may show:

UNT/PLS = 0.01CHG OK

MaxFreq=98.5HzCHG OK

MaxFlo=4999.8 SCFMCHG OK

SET I/OINP OUT EXIT

33

Operation: Programming

Programming Usingthe Local Display

Max Flow and Frequency

Page 34: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Then press NXT (F1) to select the correct alarm and press OK (F4).

Selections are:HiFloAlm = High Flow AlarmLoFloAlm = Low Flow AlarmHiTempAlm = High Temperature AlarmLoTempAlm = Low Temperature AlarmNot usedFrequency

Enter the value for the limit by pressing CHG (F1) and then OK (F4).

Note: There is only one output to operate as a frequency output or an alarm output. Both cannot operate at the same time.

For Discrete Input Settings:Fromthemainmenu,pressI/O (F1) and then I/O (F2) and then INP (F1) key to select input. The following menu will display:

Press NXT (F1) repeatedly until the correct selection is shown and then press OK (F4) to accept the setting.

Selections are: Not usedTotReset ResetthetotalizerSwitch Crv Switch between calibration curve 1 and curve 2 (only if 2 gas curve ordered)

Press EXIT (F4) repeatedly until you exit programming mode.

OUT = HiFloAlmNXT OK

HiFloAlm=500 SCFMCHG OK

INP = Not UsedNXT OK

34

Operation: Programming

Programming Usingthe Local Display

Discrete Input

Page 35: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Serial Communication SettingsToprogramtheSerialcommunicationsettings,pressI/O (F1) key from the base menu.

Press COM (F1) to select Serial communication:

Options for serial communication are:Not UsedModbus

Anyselectionotherthan“NotUsed”requiresinstallationofanoptionboard for the selected communication type.

IfRS485Modbusisselected,youwillbeaskedtoenterBaudrate,parity,stopbitandaddress(seebelow):

Baud rate (for Modbus selection only):

Press NXT (F1) to select the baud rate and press OK (F4) Selections are:

192009600480024001200

SET I/OCOM I/O 420 EXIT

Comm=ModbusNXT OK

Baud=9600NXT OK

35

Operation: Programming

Programming Usingthe Local Display

Page 36: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Parity (for Modbus selection only):

Press NXT (F1) to select the parity and press OK (F4) Selections are:

NONEODDEVEN

Address: For Modbus selection only

Press CHG (F1) to change the Modbus address and then press OK (F4). ToavoidconflictsontheModbusthismustbeauniqueaddress.

Enabling/Disabling USB Serial communication checksum

The USB serial communication checksum may be enabled or disabled using the display keypad.

Disabling the communication checksum is sometimes useful in order to use a simple serial communication program such as HyperTerminal to communicate with the meter.

Press NXT (F1) to enable or disable the USB checksum and press OK (F4).

Display SetupRemember,therearefourdisplayscreensthatyoucancyclethroughin normal operating mode (see Figure 4.2 on p. 26). Two of the four displayscreensarefixedandcannotbechanged(displays#3&4).The

Parity=NONENXT OK

Address = 01CHG OK

USB CKSM=EnabledNXT OK

36

Operation: Programming

Programming Usingthe Local Display

Page 37: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

other two screens are programmable to show the information that you prefer and is discussed in this section.

Display #1 Display #2

Selections are:DSP1L1 Display1,Line1DSP1L2 Display1,Line2DSP2L1 Display2,Line1DSP2L2 Display2,Line2

To Program Display Screens #1 & 2:

From the base programming menu press DSP (F3) to select the display menu:

Press DSP (F1) key. The display will show:

These are the selections for the display #1 line #1. Selections are:

Flo rate Flow rateTotal Total massElps Elapsed timeTemp TemperatureAlarm Error codes

DISPLAY/PASSWORDDSP IR PSW EXIT

DSP1L1 = TotalNXT OK

DSP2L1DSP2L2

DSP1L1DSP1L2

SET PARAMETERSI/O FLO DSP EXIT

37

Operation: Programming

Programming Usingthe Local Display

Page 38: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Whentheselectioniscorrect,pressOK (F4) to accept. The display will then go through the same process for all 4 lines of the 2 programmable displays(DSP1L1,DSP1L2,DSP2L1andDSP2L2).

Afterthelastlineofdisplay2isaccepted,thedisplaywillshowthefollowing menu:

This menu allows you to alternate between menu display 1 and 2 every few seconds.

Selections are: On or Off

Press OK (F4) to accept selection.

Press EXIT (F4)repeatedlyuntil“NormalMode”isseenbrieflytoexittheprogramming mode.

PasswordTherearetwouserlevelpasswords,onlyLevel 1 is programmable and gives access to all the normal settings. The second password is used to allow access to calibration factors and should normally never be changed unlessadvisedbytheSpiraxSarcoTechnialSupport,ortosetanewpassword in the event that the user forgets the Level 1 password.

Default Level 1passwordis“1234”,andLevel 2passwordis“9111”.

The Level 1programmablepasswordcanbedisabledbysettingitto“0”.

From the base programming menu press DSP (F3) to select the display menu:

ALTERNATE = OffNXT OK

SET PARAMETERSI/O FLO DSP EXIT

DISPLAY/PASSWORDDSP IR PSW EXIT

38

Operation: Programming

Programming Usingthe Local Display

Page 39: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

To Program the Password:Press PSW (F2) key to select password.

This screen displays the current Level 1 password.Press CHG (F1) key to change the password and enter new value (see

p. 28 for further details).

Press OK (F4) to accept new data and exit programming by pressing EXIT (F4) key repeatedly until out of the programming mode.

Note: Password can be number or letter characters up to 4 digits.

Units SettingsThismenuisusedtosettheunitsformassflow,temperature,pressurereferenceandthesettingsofreferencetemperature,referencepressureand density of gas when using Lbs/time or Kg/time.

ThesevalueswillbesetatSpiraxSarco,usingtheApplicationDataSheetvalues.Ifthecustomerchangestheapplication,thesevaluescan be changed to match the new application. Check with Spirax Sarco Technical Support before changing the application gas.

The unit setting is accessed from the base programming menu by pressing FLO (F2):

PASSWD = 1234CHG OK

FLOW PARAMETERS 1DGN UNT PRM EXIT

SET PARAMETERS I/O FLO DSP EXIT

39

Operation: Programming

Programming Usingthe Local Display

Page 40: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

Press UNT (F2) for Unit selection:

Press NXT (F1) to change selection and OK (F4) to accept.

Selections for Flow unit are: SCFM LBS/S NLPSSCFH NLPH MSCFD (MCFD)NM3/H NLPM SM3/HNM3/M SMPS MT/HKG/H NMPS NM3/DKG/M SFPM MMSCFM

(MMCFM)KG/S MMSCFD

(MMCFD)SCFD

LBS/H LBS/D MCFD (MSCFD)LBS/M SLPM SM3/M

SM3/D

Note:Thetotalizerwillrolloverwhenreachingacertainvalue.Themaximum value is dependent on the unit selected.

Maximum Total Rollover Value:Mostflowunits: 99,999,999,999MSCFD: 999,999,999MMSCFM: 9,999,999MMSCFD: 999,999

WARNING:TheMTI10/MTL10re-calculatesarea,4and20mAvalues,maximumflowforthefrequencyoutputandzeroflowcutoffwhenchangingflowunitsexceptforvelocityunits.Whengoingtoorfromvelocityunits,theMTI10/MTL10 will not recalculate these values and these values must be re-entered manually.

FLO = SCFMNXT OK

40

Operation: Programming

Programming Usingthe Local Display

Page 41: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

After pressing OK (F4) to accept the Flow unit the display will prompt for the temperature unit setting:

Press NXT (F1) to change selection and OK (F4) to accept.Selections for Temperature unit are:

Deg CDeg F

After pressing OK (F4)toacceptthetemperatureunitsetting,thedisplay will prompt for temperature reference in selected unit.

Press CHG (F1) to change the reference and OK (F4) to accept.

After pressing OK (F4)toacceptthereferencetemperature,thedisplaywill prompt for the pressure unit selection:

Press NXT (F1) to select next entry and OK (F4) to accept.Selections are:

mmHG Millimeters of mercuryPsia Pounds per square inch atmospherebara Bar atmosphere

Afterthepressureunitselectionismade,thedisplaywillshowamenuto enter the pressure reference:

TMP UNT= Deg FNXT OK

TmpRef = 60 °FCHG OK

PRES UNT= PsiaNXT OK

PresRef= 14.7CHG OK

41

Operation: Programming

Programming Usingthe Local Display

Page 42: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Press CHG (F1) to change it and OK (F4) to accept.Afterthepressurereferenceisaccepted,thedisplaywillpromptforthegasdensityifLBSorKGwasselectedforflowunit:

Press CHG (F1) to change and OK (F4) to accept.

Note: The density entry is only used when KG/time or LBS/time is selectedforflowrateunits.

Density conditions are referenced to 0 C° at 760 mmHg.

Flow Parameters

Thisisthemenuusedtosetvariousflowparametervalues.Theyare:Flowcutoff,pipearea,filter,highandlowalarmforflowandtemperature.

The menu is accessed from the base programming menu by pressing FLO (F2):

Then press PRM (F3):

Note: The CAL and SPC function key will only appear and be accessible from a Level 2 password.

DNS = 0.988876 KG/m3CHG OK

SET PARAMETERS I/O FLO DSP EXIT

FLOW PARAMETER 1DGN UNT PRM EXIT

FLOW PARAMETER 2CAL SPC PRM EXIT

42

Operation: Programming

Programming Usingthe Local Display

Page 43: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Then press PRM (F3):

EnterthevalueforthepercentlowflowcutoffandthenpressOK (F4).

Whentheflowratefallsbelowthezeroflowcutoff,theflowmeterwilldisplayaflowvalueofzero.

Enter the pipe area in square meters or square feet and then press OK (F4).

UsesquaremeterformetricflowunitselectionandsquarefeetforEnglishflowunitselection.

Thefiltervalue is also referred to as a dampening factor and is used to quietthereadings.Thefiltervalueisanexponentialfilterthatdampensthe noise and is used as follows:

Flow Value = (FA * new value) + (FB * average) WhereFA=filtervalue,FA+FBisequalto1.0.

Alowerfiltervaluewillincreasedampeningoftheflowrateandsmooththereading.Alowerfiltervaluewillalsoslowthemeter’sresponse.Forexample,ifweenterafilterof0.8,theweightratiofornewaverageis:

New average = (80% new sample) + (20% last average) Filterrangeis0.01to1.0,0.01beingahighfiltervalueand1.0=nofilter.

EnterthefiltervalueandthenpressOK (F4).

CUTOFF = 2.0 SCFMCHG OK

FILTER = 0.8CHG OK

A^2 = 0.05672 Ft^2CHG OK

43

Operation: Programming

Programming Usingthe Local Display

Page 44: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Thisistheupperflowlimitalarmvaluethatcanbeassociatedwithadiscreteoutput.Analarmcodeisgeneratedwhentheflowvalueexceedsthislimit.Ifnocheckingisneeded,thisvalueshouldbesettozero.

Press OK (F4) to accept the value.

Thisisthelowerflowlimitalarmvaluethatcanbeassociatedwithadiscreteoutput.Analarmcodeisgeneratedwhentheflowvalueisbelowthislimit.Ifnocheckingisneeded,thisvalueshouldbesettozero.

Press OK (F4) to accept the value.

This is the upper temperature limit alarm value that can be associated with a discrete output. An alarm code is generated when the temperature valueexceedsthislimit.Ifnocheckingisneeded,thisvalueshouldbesettozero.

HiFloAlm = 1234 SCFMCHG OK

LoFloAlm = 100 SCFMCHG OK

HiTmpAlm = 230 CCHG OK

44

Operation: Programming

Programming Usingthe Local Display

High Flow Rate Alarm

Low Flow Rate Alarm

High Temp Alarm

Filter Response (Sec.) 65% of Target

0.09 0.100.8 0.150.7 0.200.6 0.250.5 0.300.4 0.350.3 0.400.2 0.600.1 1.00

0.05 2.000.03 3.000.01 10.3

Page 45: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Press OK (F4) to accept the value.

This is the lower temperature limit alarm value that can be associated with a discrete output. An alarm code is generated when the temperaturevalueisbelowthislimit.Ifnocheckingisneeded,thisvalueshouldbesettozero.

Press OK (F4) to accept the value.

Note: If the programming menu was entered with a Level 2password,then more menus will be shown that deal with factory set parameters that should not be changed. Calibration Parameters

This menu allows changing the factory calibrated setting of the flowmeterandisaccessiblewithaLevel 2 password. Calibration parameter values are set for temperature and pressure at 0 degree C and 760 mmHg.

These settings should normally never be changed except by Spirax Sarco personnel at the factory.

This menu is entered from the base menu and pressing FLO,PRM and CAL.

Press CAL (F1) then the display will show:

LoTmpAlm = 50 CCHG OK

FLOW PARAMETER 2CAL SPC PRM EXIT

CAL TABLETB1 - - - - - - EXIT

45

Operation: Programming

Programming Usingthe Local Display

Page 46: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

Press TB1 (F1) then the display will show:

Press NXT (F3) then the display will show:

Use the CHG (F1)keytochangetheentry,PRV (F2) to move to the previousentry,NXT (F3) to move to the next entry and EXIT (F4) to return.

Pressing the NXT (F3) key will show the data point voltage and then mass velocityandthengotothenextdatapoint.ThenumberafterVolt(i.e.,Volt1)orFlo(i.e.,Flo1)indicatedthedatapointnumber.

The calibration table can hold up to 20 data pair points. Each data point has a voltage and mass velocity associated with it.

Reset Total and Elapsed Time

Theresettingofthetotalizerandelapsedtimeisaccomplishedbypressing the F3 and F4 keys at the same time in the normal running mode.

F1 F2 F3 F4

Press YES (F4) to reset total and NO (F1) to cancel.

Note: This feature is not available on non-resettable units.

Totalizer Rollover: The MTI10 and MTL10 has an automatic roll-over function. The unit will begin roll-over as follows:

Volt1 = 0.92367CHG PRV NXT EXIT

Flo1 = 0CHG PRV NXT EXIT

RESET TOTAL ?NO YES

46

Operation: Programming

Programming Usingthe Local Display

Page 47: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Mostflowunits: 99,999,999,999MSCFD: 999,999,999MMSCFM: 9,999,999MMSCFD: 999,999

Restore Database

Restoringtheoriginalfactorysettingsisaccomplishedfromthe“FlowParameter 2” menu by entering a Level 2password“9111”andpressingthe SPC key (F2).

F1 F2 F3 F4

Upon pressing OK (F4),anoptiontorestorethedatabasewillfollow:

F1 F2 F3 F4

Press YES (F1) ONLY if you want to restore your database to the initial factory setting that the meter was shipped with. All current user-entered settings will be overwritten.

ThegreenLP1LEDwillflashatafasterpaceuntiltherecallisperformed.The "RESET CRC" screen will follow "RESTORE DATABASE".

Reset CRCIftheNVRAMCRCcheckfails(ErrorCode36),theprogrammedsettingsvalueswillneedtobeverifiedandcorrectedbeforeclearingtheerror.CallSpirax Sarco Technial Support if you need assistance.

F1 F2 F3 F4

RESET CRC?YES NO

47

Operation: Programming

Programming Usingthe Local Display

K fact = 0 %CHG OK

RESTORE DATABASE ?YES NO

Page 48: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Press YES (F1) ONLY if you want to reset the CRC and generate a new CRC value.

Simulation

Thismenuallowsforthesimulationofflowrate,temperatureandflowinput voltage. It should only be used for testing and demonstration purposes.

Make sure to return all of these simulation values to zero, before returning to the normal mode of operation.

Note:Simulatedvaluesareonlyenabledwhennotsettozero.

Caution: If the 4 to 20mA and/or the pulse outputs are connected to controllers,setthecontrollersto“manual”.Thiswillensurethatthesimulated signals do not cause false controller action.

The menu is accessible from the main programming menu by pressing FLO,andDGN (F1):

F1 F2 F3 F4

Pressing DGN (F1) will show:

F1 F2 F3 F4

Pressing SIM (F1) will show:

F1 F2 F3 F4

Enter the value and then press OK (F4).

FLOW PARAMETER 1DGN UNT PRM EXIT

DIAGNOSTICSIM TST EXIT

FloSim = 0 SCFMCHG OK

48

Operation: Programming

Programming Usingthe Local Display

Page 49: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Note:Enterzerotodisablethisfeature.

F1 F2 F3 F4

Enter the value and then press OK (F4).

Note:Enterzerotodisablethisfeature.

F1 F2 F3 F4

Enter the value and then press OK (F4).

Note: This value is used to simulate the Current Sense Voltage (CSV) andshouldbesettozerofornormalmode.

F1 F2 F3 F4

Press YES (F1)tostartthesimulationmode,otherwisepressNO (F4).Uponpressingeitherkey,theprogramwillreturntotheFLOWPARAMETER 1 menu.

Note: Simulation Mode will be cleared if the power is cycled.

Note:• Refer to the Calibration Validation Chapter for information on

settingthefieldbaselineforZeroCAL-CHECK™testsifyouplan to perform these tests in the pipe.

• Please save the PVC sensor cover that was shipped with your meter. It will be needed to perform Zero CAL-CHECK™ tests out of pipe.

TmpSim = 0 CCHG OK

CsvSim = 0 VCHG OK

ENABLE SIM?YES NO

49

Operation: Programming

Programming Usingthe Local Display

Page 50: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

CAL-V™ - Calibration Validation Test 1ThismenuallowstheusertoconfirmthecalibrationoftheMTI10/MTL10 by verifying the functionality of the sensor and sensor signal processingcircuitry.DuringtheCAL-V™calibrationvalidationtest,themicroprocessor adjusts current to the sensor elements and determines the resulting electrical characteristics. These site characteristics are compared with the data that was collected at the factory during the original meter calibration. Matching data within established tolerances confirmsthemeterisaccurate.Thistestcanbeperformedundernoflowornormalflowconditions.Thetesttakesuptofourminutestocomplete.Attheconclusionofthetest,aPassorFailmessagewillbedisplayed.Press F4 at the conclusion of the test to return to normal measuring mode or to terminate the test. Press FLO (F2) from the main menu. The display will show:

Press DGN (F1). The display will show:

Press TST (F2).

FLOW PARAMETER 1DGN UNT PRM EXIT

DIAGNOSTICSIM TST EXIT

Caution:• The CAL-V™ test is valid for checking the calibration accuracy

offlowmetersinstalledintheapplicationsforwhichitwascalibratedincludingthegas/gasmixture,calibrationrangeandpipesizeshownonthecalibrationcertificate.

• Forapplicationswithtemperatureexceeding250°F(121°C),CAL-V™ test results may vary.

• Periodic inspection for damage and cleaning of the sensor elements is required.

50

Operation: Programming

Programming Usingthe Local Display

Page 51: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4 F1 F2 F3 F4

The display will show:

Press CALV (F1)toperformtheCAL-V™verificationtest.

Note:TheMTI10/MTL10willstopmeasuringflowwhenperformingthistest. Press EXIT (F4) to exit if you do not wish to continue.

ToselectwhattheflowoutputwilldoduringaCAL-V,choosefromthese options:

GoToZero:Flowoutputwillbezeroduringthetest(i.e.4mA)Hold Value: Flow will hold last value during the test

Select the option and press OK (F4).

WARNING:Ifyouareusingaclosedloopcontrol,thesystemneedstobe taken off-line during the test.

Press OK (F4) to start CAL-V™. CAL-V™ test screen:

Thistestwilltakeupto4minutes(lesstimeifthereisflow)andwillshow the Cal value changing as the power to the sensor is adjusted. The T=xxxisacountdowntimerindicatinghowmuchtimeislefttofinishthetest.A“PleaseWait”messagewillbeflashingonandoffonline2duringthis test.

DIAGNOSTICCALV ZRO EXIT

Verifying CAL-VCal = 12.7 T=123

Take Controloff-line EXIT OK

Flo: Go to ZeroNXT OK

Verifying CAL-V Please Wait

51

Operation: Programming

Programming Usingthe Local Display

Page 52: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Upontestcompletion,thefinalCAL-V™valuewillbedisplayedalongwith a Pass/Fail message.

Zero CAL-CHECK™ - Calibration Validation Test 2The Zero CAL-CHECK™ test is a companion test to CAL-V™.UnlikeCAL-V™,whichmaybeperformedinthepipeandatprocessconditions,ZeroCAL-CHECK™mustbeperformedatzeroflowtoensureavalidtestresult.ThistestisusedtoconfirmthattheflowmeterstillretainsitsoriginalNIST-traceablecalibrationatzeroflowandthatthesensorisfreeoffilmorresiduethatmayaffectreadings.Thetesttakeslessthan5minutestocomplete.Attheconclusionofthetest,aPass or Fail message will be displayed. Press F4 at the conclusion of the test to return to normal measuring mode or to terminate the test. See Calibration Validation User's Guide for more details. Press FLO (F2) from the main menu. The display will show:

Press DGN (F1). The display will show:

Press TST (F2). The display will show:

Press ZRO (F2) to enter the Zero CAL-CHECK™ menu.

Ifperformingthetestinthepipe,a"noflow"conditionmustbemet.Ifperformingoutofthepipe,themetermustberemovedandthesensor

FLOW PARAMETER 1DGN UNT PRM EXIT

DIAGNOSTICSIM TST EXIT

CAL-V = 0.51Passed OK

DIAGNOSTICCALV ZRO EXIT

52

Operation: Programming

Programming Usingthe Local Display

Page 53: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

protected by the PVC sensor cover originally shipped with the meter from the factory.

Press PIP (F1) to choose to perform the test in the pipe. Press BTL (F2) to choose to perform the test out of the pipe.

The display will show:

F1 F2 F3 F4

OR

F1 F2 F3 F4

Press VER (F1) key to verify the Zero CAL-CHECK™.

Press YES (F1) key to verify the Zero CAL-CHECK™.WARNING:Ifyouaredoinga"Pipe"test,youmustverifythatthereisanoflowconditionbeforeproceeding.Ifyouareperformingthetestina

ZERO PIPE TESTVER SET PRM EXIT

ZERO BTL TESTVER EXIT

VERIFY ZRO? YES NO

Process Zero and Stable YES

ZERO CHK MENU PIP BTL EXIT

Note:• ThefieldbaselineforaZeroCAL-CHECK™testperformedin

the pipe ("ZERO PIPE TEST") must be set before performing the test.

• See the Calibration Validation Chapter for details on performing all diagnostic tests.

• Please use the PVC sensor cover that was shipped with your meter perform Zero CAL-CHECK™ tests out of pipe ("ZERO BTL TEST").

53

Operation: Programming

Programming Usingthe Local Display

Page 54: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

Verifying ZROSd = x.xe-xx T=xx

bottle,besuretoisolatethesensorinabottle-anyairmovement(evenfrom a fan) can result in a false "fail" result.

Onceprocessisstable,pressYES (F4) key to begin the Zero CAL-CHECK™.

This test will take less than 5 minutes. The T=xx is a count down timer indicatinghowmuchtimeislefttofinishthetest.

Upontestcompletion,thefinalpercentagevaluewillbedisplayedalongwith a Pass/Fail message.

Enabling/Disabling the Infrared Keypad (IR Buttons)

The IR buttons may be disabled from the menu to avoid being triggered by frost or snow on the window. This menu is accessed by pressing DSP (F3) from the main menu then IR (F2):

Press NXT (F1) key to enable or disable the IR buttons.

Note:Afterselecting“Disable”andpressingOK (F4),theIRbuttonswillno longer operate. It will be necessary from now on to open the cover and operatetheconfigurationpanelusingthemechanicalpushbuttons.Toreturntothenormaldisplaymode,usemechanicalbuttonsorwaitfortheprogramming mode timeout.

Calibrating the Infrared Keypad (IR Buttons)The IR buttons are calibrated in the factory before shipment, but conditions in the field may alter the way the buttons read. To allow the IR buttons to perform

Verifying ZRODiff = xx.xx T=xx

Diff=x.xxx % Passed OK

IR KEY= EnabledNXT OK

54

Operation: Programming

Programming Using the Local Display

Page 55: Thermal Mass Flowmeter and Temperature Transmitter MTI10

better, it may be necessary to calibrate the keys in the field. Use your finger to activate the IR buttons using this process:

Note: Your finger must activate each button approximately 0.1” to .5” from the surface of the glass.

Press the (F1) button until the display shows ‘SET PARAMETER?’ Then use YES (F4).

Use buttons (F1) (F2) and (F3) to enter "1111" (up/down and next).

Use buttons (F1) (F2) and (F3) to enter "0000" then press OK (F4).

Turn OFF power to the MTI10/MTL10.

Turn ON power to the MTI10/MTL10 while placing your finger on the (F1) IR key.

55

Operation: Programming

Page 56: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Fig. 4.4: MTI10/MTL10 Menu Tree - Main Menu

SET PARAMETERSI/O FLO DSP EXIT

SET I/OCOM I/O 420 EXIT

Comm=MODBUSNXT OK

NoneModbus

SET I/OINP OUT EXIT

INP= Not usedNXT OK

Not usedTot Reset

SET 4-20 mACH1 CH2 EXIT

20 mACnt=3265CHG OK

4 mACnt=367 CHG OK

20 mA=2345.6 SCFMCHG OK

4 mA=0 SCFMCHG OK

CH2=TempNXT OK

TempFlow

20mACnt=3748CHG OK

4mACnt=274CHG OK

20 mA=300 °FCHG OK

4 mA=0 °FCHG OK

Level 2

Level 2

Analog Outputs

Select either flow or temperature for channel 2

Baud=9600NXT OK

Parity=EVENNXT OK

Address=02CHG OK

Modbus only

1200240048009600

19200

NONEODDEVEN

01-247

Enter menu by scolling to display 4 and entering the password

mA Fault=Not useNXT OK

Not use3.6 mA21 mA

Display Menu, p. 60

Flow Parameters 1 Menu, p. 58

Digital Output Menu, p. 5756

Operation: Menu Tree

Digital Output Main Menu

Page 57: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Fig. 4.5: MTI10/MTL10 Menu Tree - Digital Output

OUT= Not usedNXT OK

Not usedFrequencyHiFloAlmLoFloAlm

HiTempAlmLoTempAlm

FREQUENCY OUTPUTP/U U/P FEQ EXIT

PLS/UNT=2CHG OK

UNT/PLS=0.5CHG OK

MaxFreq=100HzCHG OK

MaxFlo=5678.5 SCFMCHG OK

HiFloAlm=500 SCFMCHG OK

LoFloAlm=100 SCFMCHG OK

HiTmpAlm=300° FCHG OK

LoTmpAlm=10 FCHG OK

High Limit Alarm

Low Limit Alarm

High Temp Alarm

Low Temp Alarm

Select 1 of 3 methods to scale the frequency output

OUT= FREQUENCYNXT OK

Frequency

OUT= HiFloAlmNXT OK

OUT= LoFloAlmNXT OK

OUT= HiTmpAlmNXT OK

OUT= LoTmpAlmNXT OK

57

Digital Output

Operation: Menu Tree

Page 58: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Fig. 4.6: MTI10/MTL10 Menu Tree - Parameter Menu 1

FLOW PARAMETER 1DGN UNT PRM EXIT

DIAGNOSTICSIM TST EXIT

FloSim=0 SCFMCHG OK

TmpSim=0 ° FCHG OK

CsvSim=0 VCHG OK

ENABLE SIM?YES NO

Note: simulation value needs to be greater than zero

to be taken(i.e 0.0001 for a value close to

zero)

FLO UNT=SCFMNXT OK

SCFMSCFH

NM3/HNM3/M

KG/HKG/MKG/SLBS/HLBS/MLBS/SNLPHNLPMSMPSNMPSSFPM

MMSCFDLBS/DSLPMNLPS

MSCFDSM3/HMT/H

NM3/DMMSCFM

MCFDSCFD

SM3/MSM3/D

TMP UNT=° FNXT OK

Deg FDeg C

TmpRef= 0 °FCHG OK

PRES UNT=mmHGNXT OK

PresRef=760CHG OK

mmHGPsiabara

STP

FlowParameter2Menu,p.59

DiagnosticTestMenu,p.61

5858

Parameter Menu 2

Operation: Menu Tree

Parameter Menu 1

Page 59: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Fig. 4.7: MTI10/MTL10 Menu Tree - Parameter Menu 2

FLOW PARAMETER 2CAL SPC PRM EXIT

Volt1 = 0.2479CHG PRV NXT EXIT

Flo1 = 0.0CHG PRV NXT EXIT

Volt20 = 1.2479CHG PRV NXT EXIT

Flo20 = 7046.5CHG PRV NXT EXIT

K fact = 0%CHG OK

RESTORE DATABASE?YES NO

RESET CRC?YES NO

Cutoff=12.5 SCFMCHG OK

A2=0.0233CHG OK

Filter=0.8CHG OK

HiFloAlm=0 SCFMCHG OK

LoFloAlm=0 SCFMCHG OK

HiTempAlm=0 °FCHG OK

LoTempAlm=0 °FCHG OK

Level 2

Level 2Calibration Table Parameters

Flow cutoff in selected unit

Pipe Area im M2 or FT2

Flow Exponential Filter

High Flow Alarm in selected unit

Low Flow Alarm in selected unit

High Temperature Alarm in selected unit

Low Temperature Alarm in selected unit

velocity in Meter/hour for point #1

sensor voltage for point #1

CAL TABLETB1 (TB2) (TB3) EXIT

Cycle throughup to 20 settings

Level 1

5959

Parameter Menu 2

Operation: Menu Tree

Page 60: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Fig. 4.8: MTI10/MTL10 Menu Tree - Display Menu

Note:All readings updated every second• Flo Rate = Flow rate of process gas• Total=Totalflowofprocessgas• Elps=Elapsedtimesinceresetofflowtotal• Temp = Temperature of process gas• Alarm = High/Low Flow Rate or Temperature Alarm

DISPLAY/PASSWORDDSP IR PSW EXIT

DSP1L1=FLo rateNXT OK

FLo rateTotalElps

TempAlarm

DSP1L2=TempNXT OK

DSP2L1=AlarmNXT OK

DSP2L2=ElpsNXT OK

ALTERNATE=OffNXT OK

OnOff

PASSWD=1234CHG OK

FLo rateTotalElps

TempAlarm

FLo rateTotalElps

TempAlarm

FLo rateTotalElps

TempAlarm

When alternate "ON", flashes between the 2 displays

Display 1 Line 1

Display 1 Line 2

Display 2 Line 1

Display 2 Line 2

IR KEY=EnabledNXT OK

EnabledDisabled

60

OperationSection 1

Operation: Menu Tree

Display Menu

Page 61: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Fig. 4.9: MTI10/MTL10 Menu Tree - Diagnostic Tests Menu

ZERO PIPE TESTVER SET PRM EXIT

SET ZRO Ref?YES NO

VERIFY ZRO?YES NO

Level 1

Process Zero andStable? EXIT YES

Verifying ZRODiff=x.xx T=xx

Diff=x.xxx %Pass OK

Process Zero andStable? EXIT YES

Setting ZRO refZR=x.xxxxx T=xx

Ref=x.xxxxOK

Level 2

Display alternates between

the 2 screens

Displays count down timer

(T=xx)

Displays“Pass” or

“Fail”

Test=300 secCHG OK

ZERO CHK MENUPIP BTL EXIT

DIAGNOSTICCALV ZRO EXIT

Verifying CAL-VCal=xxx T=yyy

Verifying CAL-VPlease Wait

CAL-V=0.51Pass OK

Display alternates between the 2

screens until the test is completed

in about 4 minutes. Pressing

the F4 key will abort the test

“Pass” or ”Fail” result is displayed at end of test.

CAL VER EXIT

Flo:Go to zeroNXT EXIT OK

Hold ValueGo to zero

Take Controloff-line EXIT OK

Level 2

Level 1

Verifying ZROSd=x.xe-xx T=xx

ZERO BTL TESTVER EXIT

VERIFY ZRO?YES NO

Level 1

Process Zero andStable? EXIT YES

Verifying ZRODiff=x.xx T=xx

Diff=x.xxx %Pass OK

Display alternates between

the 2 screens

Displays “Pass” or “Fail”

Verifying ZROSd=x.xe-xx T=xx

FlowParameter1Menu,p.58

61

OperationSection 1

Operation: Menu Tree

Diagnostic Tests Menu

Page 62: Thermal Mass Flowmeter and Temperature Transmitter MTI10

62

Calibration

CalibrationValidation

The Spirax Sarco MTI10/MTL10 View software allows users to easily displaydataandconfiguretheMTI10/MTL10totheirspecificapplicationparameters.Then,logflow/temperaturedatatoanExcelfile.Thesoftware can also activate the Spirax Sarco CAL-V™ and Zero CAL-CHECK™ diagnostic functions.

This section contains the operation instructions for the Calibration Validation diagnostic tests: CAL-V™ and Zero CAL-CHECK™.

Calibration of the Spirax Sarco MTI10/MTL10 flowmeterToensurethatallflowmetersmeetspecifiedperformanceparametersandprovideaccurate,repeatablemeasurementsinthefield,allcalibrationsareperformedwithNIST-traceableflowstandards.Eachmeter is shipped from the factory with a Spirax Sarco Calibration Certificate.

Calibration ValidationValidatingthecalibrationofyourMTI10/MTL10flowmeterissimpleandeasy.Byperformingtwosimpletestsinthefield,operatorscanverifythat the meter is running accurately by testing the functionality of the sensor and its associated signal processing circuitry.

Thefirsttest,CAL-V™,teststhefunctionalityofthesensoranditsassociated signal processing circuitry and can be done in the pipe and in normal processing conditions (see "CAL-V™ Calibration Validation Test" onpage64).Thesecondtest,ZeroCAL-CHECK™,ensurestherepeatability and cleanliness of the sensor (see "Zero CAL-CHECK™ Calibration Validation Test" on page 65).

See Figure 1.1 on the following page to understand Calibration Validation with these two tests.

Page 63: Thermal Mass Flowmeter and Temperature Transmitter MTI10

63

Calibration

Fig. 5.1: Total Calibration Validation

Figure 5.1 above illustrates the Calibration Validation feature and a summary of the details for each of the two tests.

SpiraxSarcohasdevelopedCalibrationValidation,usingtheCAL-V™and Zero CAL-CHECK™ tests to help our customers avoid sending the meter back for annual or biennial re-calibrations.

Using Calibration Validation allows our customers to validate the accuracyandfunctionalityofthemeterinthefieldwithapushofabutton.

CalibrationValidation

Calibration ValidationSpirax Sarco MTI10/MTL10 Gas Flowmeter

CAL-VTM Zero - CAL-CHECKTM

Sensor and Electronics Test

Sensor Test at Zero Flowvs. Field Baseline

Sensor Test at Zero Flowvs. Factory Baseline

• Complete test of sensor elements and electronics.

• Inyourpipe,undernormalprocess conditions.

• Operator initiated via front panel,MTI10/MTL10,orMODBUS.

• Hold outputs at last value or go tozero;operator-selectable.

• Testtakeslessthan5minutes;3 minutes typical.

• Test results in pass/fail message.

• Data saved in meter for look-up anytime.

• Calibration Validation Certificatecanbegeneratedif test is initiated using MTI10/MTL10 ViewTM software.

• Customer-setzeroflowbaselineestablishedundernormalzeroflowconditions.

• Test compares sensor characteristics atzeroflowwithcustomer-setzeroflowbaseline.

• Operator-initiatedfromfrontpanel,MTI10/MTL10View™,orMODBUS.

• Fail condition indicates possible dirty sensor.

• Use of Spirax Sarco Packing Gland Assembly to retract probe is a convenientwaytoestablishazeroflowcondition.

• Test takes less than 5 minutes after zeroflowcondition.established

• CalibrationValidationCertificatecanbe generated if test is initiated using MTI10/MTL10 View™software.

• Test compares sensor characteristicsatzeroflowat ambient temperature and atmospheric pressure with factory characteristics.

• UsedtoconfirmZeroCAL-CHECK™wheninsituzeroflowconditioncannotbeestablished.

• Operator-initiated from front panel or MTI10/MTL10 View™.

• Test takes less than 5 minutes after out of pipe set-up complete.

• CalibrationValidationCertificatecan be generated if test is initiated using MTI10/MTL10 View™ software.

Page 64: Thermal Mass Flowmeter and Temperature Transmitter MTI10

64

Calibration

CAL-V™ Calibration Validation Test ThismenuallowstheusertoconfirmthecalibrationaccuracyoftheMTI10/MTL10 Flowmeter by verifying the functionality of the sensor and sensor signal processing circuitry. During the CAL-V™ calibration validationtest,themicroprocessoradjuststhecurrenttothesensorelements and determines the resulting electrical characteristics. These site-determined characteristics are compared with the data that was collected at the factory during the original meter calibration. Matching datawithinestablishedtolerancesconfirmsthemeterisaccuratewithinpublishedaccuracyspecifications.

Thistestcanbeperformedundernoflowornormalflowconditions.Thetesttakesabout4minutestocomplete.Attheconclusionofthetest,aPass or Fail message will be displayed.

Note: If the CAL-V™ test is performed using the MTI10/MTL10 View™ Software,atthecompletionofthetest,aCAL-V™Certificatemaybeprintedforarecordofthetest.ThiscertificatewilldisplaytheCAL-V™values and a pass/fail result.

Fig. 5.2: Normal Mode vs. CAL-V™ Mode

CAL-VTM

Page 65: Thermal Mass Flowmeter and Temperature Transmitter MTI10

65

Calibration

Zero CAL-CHECK™ Calibration Validation Test TheZeroCAL-CHECK™testisusedtoensurethattheflowmeterstillretainsitsoriginalNIST-traceablecalibrationatzeroflow.Ifzeroflowcanbeestablished,thesensordoesnotneedtoberemovedandtheprocedurecanbedoneinthepipe.Alternatively,aPackingGlandAssembly can be used to remove the sensor from the gas stream to createa“noflow”condition.

Note: If the Zero CAL-CHECK™ test is performed using the MTI10/MTL10 View™Software,atthecompletionofthetest,aZeroCAL-CHECK™Certificatemaybeprintedforarecordofthetest.Thiscertificatewilldisplay a pass/fail result.

Fig. 5.3: Normal Mode vs. Zero CAL-CHECK™ Mode

Page 66: Thermal Mass Flowmeter and Temperature Transmitter MTI10

66

Calibration

Starting Up CAL-V™CAL-V™ may be performed at any time after installation and at any flowrate,includingzeroflow.Youdonotneedanyextraequipmentorpaperwork to perform a CAL-V™ test. Refer to "Performing a CAL-V™ Calibration Validation Test" on p. 74 of this User's Guide to start CAL-V™.

Starting Up Zero CAL-CHECK™There are a few ways that Zero CAL-CHECK™ is different than CAL-V™:• The Zero CAL-CHECK™ test must be performed at a "no flow"

condition(iezeroflow)• IftheZeroCAL-CHECK™testistobeperformedinthepipe,thefield

baseline must be set before an actual test can be performed (see "Setting Field Baseline for In-Situ Zero CAL-CHECK™ Tests" on p. 67).

• IftheZeroCAL-CHECK™testistobeperformedoutofthepipe,the meter must be set upside-down (probe pointing up) and the PVC sensor protector that the meter was shipped with must be placed back over the sensor to achieve the factory baseline that the meter has been set with.

Techniques for Achieving Zero Flow - In the PipeIn-situ (in the pipe) Zero CAL-CHECK™ testing can be achieved in one of two common ways:1. Packing Gland Assembly

Thefirstin-situoptionisachievedthroughtheuseofthePackingGland Assembly that may be ordered as an option for most inline- or insertion- type meters. The assembly allows the operator to retract and isolate the sensor from the process in order to conduct the Zero CAL-CHECK™test.Thisisparticularlybeneficialforapplicationsinwhich the process is not easily stopped.

2. Pipe Bypass / Valving-OffThesecondin-situoption,ifspaceallows,istoredirecttheflowthrough a bypass pipe section or valve off the meter in order to isolate the meter's sensor in the place where it has been installed. While the flowisredirected,theZeroCAL-CHECK™testcanbeperformed.Oncethetestiscomplete,thevalvestothebypassmaybeclosedandflowmaybedirectedbacktothemeter'ssensorwhereflowmonitoring can continue as normal.

Achieving Zero Flow - Out of PipeIfspacelimitationspreventin-situtestingatzeroflowaslistedabove,then Out of Pipe testing must be performed.

CAL-V™

Zero CAL-CHECK™

Achieving Zero Flow - In-Situ (In the Pipe)

Achieving Zero Flow - Out of Pipe

Page 67: Thermal Mass Flowmeter and Temperature Transmitter MTI10

67

Calibration

Withthisconfiguration,themetermustberemovedfromtheprocess,thetestperformed,andthenthemeterreturnedtotheprocessaftertestinghas been completed.

DuetothehighsensitivityofthePowerPro™sensor,itisnecessaryto isolate the sensor once the meter has been removed from the pipe. Therefore,SpiraxSarcoprovidesasensorcoverwhenthemeterisshipped to the customer. An alternative to the sensor cover is to use a bottle or other closed container in order to isolate the sensor and achieve the"noflow"conditionnecessarytoperformtheZeroCAL-CHECK™test.

Fig. 5.4: Factory-Supplied PVC Sensor Protector vs. Plastic Bottle

NOTE: For best results:1. Use the factory-supplied PVC sensor protector shipped with the meter

(see Fig. 5.3 above).2. Placethemeterupside-downonaflat,solidsurfacebeforestartingthe

test.3. Do not allow the meter to get jostled - make sure the meter is stable

and the sensor completely isolated.

Setting Field Baseline for In-Situ Zero CAL-CHECK™ testsAftercalibrationofeveryMTI10/MTL10flowmeter,alabtechniciansetsthe factory baseline for Zero CAL-CHECK™ tests. If you are planning onremovingthemeterfromtheprocesstoperformthetest,youdonothavetosetthebaseline;however,ifyouplantoperformthetestin-situ(inthepipe)orusingthePackingGlandAssembly,youmustsetthefieldbaseline before performing the test.

Setting FieldBaseline

Page 68: Thermal Mass Flowmeter and Temperature Transmitter MTI10

68

Calibration

Packing GlandAssembly

Using the Packing Gland AssemblyThe following are the instructions for using the Packing Gland Assembly tosetthefieldbaselinefor-orperforming-anin-situZeroCAL-CHECK™test.IfyouarenotusingthePackingGlandAssembly,moveahead to "Starting the Field Baseline Set" on p. 71.

Note: If you need information on the installation of the Packing Gland Assembly,refertotheDocument105440.

Fig. 5.5: MTI10/MTL10 and Parts of the Packing Gland Assembly

F1 F2 F3 F4

Packing Gland Assembly

Pipe

Exposed probe section

Valve Shut-o� Handle

Probe in pipe

Sensor

Fittings

MTI10/MTL10 Flowmeter

Page 69: Thermal Mass Flowmeter and Temperature Transmitter MTI10

69

Calibration

Fig. 5.6: Close-Up: Fittings of the Packing Gland Assembly

Inordertoisolatethesensor,itmustberetractedfromthepipe.Loosenthesmallnutofthecompressionfittingusing7/8"and13/16"wrenches.Onceloosened,theprobewillbeabletoslideupwards(seeFigure2.4below).

Fig. 5.7: Loosening the Compression Fitting

Large Nut

Small Nut

End Fitting

Adjacent Hex

Fittings

7/8” wrench

13/16” wrench

Small Nut

Adjacent Hex

Page 70: Thermal Mass Flowmeter and Temperature Transmitter MTI10

70

Calibration

Withonehandsupportingthemeter,slidetheprobeupthroughthepacking Gland Assembly until the internal stop makes contact. Close theValveShut-offhandlebyturningthelever90˚clockwisetoisolatethesensor within the chamber of the Packing Gland Assembly. Then slightly tightenthecompressionfittingtokeepasealinthechamber.

Fig. 5.8: Isolating the Sensor in Packing Gland Assembly Chamber

Packing GlandAssembly

Pull handle down CW90° to shut off valve

Lift to unlock handle

F1 F2 F3 F4

Sensor removedfrom Process

Slide meter up

Pull handle down CW90° to shut o� valve

Sensor in Chamber

Small NutAdjacent Hex

Page 71: Thermal Mass Flowmeter and Temperature Transmitter MTI10

71

Calibration

Packing Gland Assembly

Setting Field Baseline

Now that the sensor is isolated in the chamber of the Packing Gland Assembly,theMTI10/MTL10flowmeterisreadytoeitherperformtheZeroCAL-CHECK™ test or set the Field baseline for future tests.

Whenthemeterhascompletedthetest,besuretoopentheshut-offvalve,slidethemeterdownwarduntiltheferrulerestsinthefittings,andtightenthesmallnutofthefittingsinordertosealtheprobeinthestreamagain.

Check to be sure the meter has returned to normal operation.

Starting the Field Baseline SetAfter the meter has been installed and wired correctly (see the Wiring Instructionsectionofthismanual),poweronthemeter.The meter should default to display #1. Press "F1" to program the meter. Press Yes (F1) to set Parameters and you will be prompted to enter the password. Use a Level 2 password (9111). The display will show:

F1 F2 F3 F4

Press FLO (F2) to go to the Flow Parameter 1 Menu.

F1 F2 F3 F4

Press DGN (F1) to go to the Diagnostic Test Menu.

F1 F2 F3 F4

Press TST (F2) to get to the Diagnostic Test Sub-Menu.

SET PARAMETERSI/O FLO DSP EXIT

FLOW PARAMETER 1DGN UNT PRM EXIT

DIAGNOSTIC SIM TEST EXIT

Page 72: Thermal Mass Flowmeter and Temperature Transmitter MTI10

72

Calibration

F1 F2 F3 F4

Choose ZRO (F2) for Zero CAL-CHECK™.

F1 F2 F3 F4

Choose PIP (F1) for in-situ Zero CAL-CHECK™.

F1 F2 F3 F4

Tosetthefieldbaseline,chooseSET (F2).

F1 F2 F3 F4

Choose YES (F1)tosetthefieldbaseline.

F1 F2 F3 F4

Note:Atthispoint,youmustmakesurethatthereisanoflowconditionin the pipe (see "Techniques for Achieving Zero Flow - In the Pipe" on p. 66). Also be sure that the meter will be stable for the duration of the setting process.

Setting Field Baseline

Process Zero and Stable? YES

SET ZRO Ref? YES NO

ZERO PIPE TEST VER SET PRM EXIT

ZERO CHK MENU PIP BTL EXIT

DIAGNOSTIC CALV ZRO EXIT

Page 73: Thermal Mass Flowmeter and Temperature Transmitter MTI10

73

Calibration

Setting Field Baseline

Ifthereisnoflowandthemeterisstable,pressYES (F4). The display will show:

F1 F2 F3 F4

Asthemeterissettingthefieldbaseline,themeterwilldisplaythereference value for the Zero baseline (ZR=xx.xxxx) and a countdown timer (T=xx) to approximate the time until the completion of the set.

Dependingonthemeterconfiguration,thesetmaytakebetween5-15minutes to complete.

Note: Do not interrupt the set by touching the meter or changing conditions in the pipe.

Attheconclusionoftheset,thedisplaywillshow:

F1 F2 F3 F4

Press OK (F4) to return to the Zero CAL-CHECK™ menu.

The meter may now perform Zero CAL-CHECK™ tests at the Field Baseline at any time.

Setting ZRO ref ZR=x.xxxxx T=xx

Ref=x.xxxx OK

Page 74: Thermal Mass Flowmeter and Temperature Transmitter MTI10

74

Calibration

Performing the CAL-V™ Calibration Validation TestThismenuallowstheusertoconfirmthecalibrationoftheMTI10/MTL10 by verifying the functionality of the sensor and sensor signal processingcircuitry.DuringtheCAL-V™calibrationvalidationtest,themicroprocessor adjusts current to the sensor elements and determines the resulting electrical characteristics. These site characteristics are compared with the data that was collected at the factory during the original meter calibration. Matching data within established tolerances confirmsthemeterisaccurate.Thistestcanbeperformedundernoflowornormalflowconditions.Thetesttakesuptofourminutestocomplete.Attheconclusionofthetest,aPassorFailmessagewillbedisplayed.Press F4 at the conclusion of the test to return to normal measuring mode or to terminate the test. Press FLO (F2) from the main menu. The display will show:

F1 F2 F3 F4

Press DGN (F1). The display will show:

F1 F2 F3 F4

Press TST (F2). The display will show:

F1 F2 F3 F4

Press CALV (F1)toperformtheCAL-V™verificationtest.

Note:TheMTI10/MTL10willstopmeasuringflowwhenperforming this test. Press EXIT (F4) to exit if you do not wish to continue.

CAL-VTM

FLOW PARAMETER 1DGN UNT PRM EXIT

DIAGNOSTIC SIM TST EXIT

DIAGNOSTICCALV ZRO EXIT

Page 75: Thermal Mass Flowmeter and Temperature Transmitter MTI10

75

Calibration

ToselectwhattheflowoutputwilldoduringCAL-V™,choosefromtheseoptions:

GoToZero:Flowoutputwillbezeroduringthetest(ie4mA)Hold Value: Flow will hold last value during the test

F1 F2 F3 F4

Select the option using NXT (F1) and then press OK (F4).

F1 F2 F3 F4

WARNING:Ifyouareusingclosedloopcontrol,thesystemneedstobetaken off-line during the test.

Press OK (F4) to start CAL-V™. CAL-V™ test screen:

F1 F2 F3 F4

Thistestwilltakeupto4minutes(lesstimeifthereisflow)andwillshowthe Cal value changing as the power to the sensor is adjusted.

Caution:• The CAL-V™ test is valid for checking the calibration accuracy of

flowmetersinstalledintheapplicationsforwhichitwascalibratedincludingthegas/gasmixture,calibrationrangeandpipesizeshownonthecalibrationcertificate.

• Forapplicationswithtemperatureexceeding250°F(121°C),CAL-V™ test results may vary.

• Periodic inspection for damage and cleaning of the sensor elements is required.

CAL-VTM

Take Controloff-line EXIT OK

Verifying CAL-VCal = 12.7 T=123

Flo:GotozeroNXT EXIT OK

Page 76: Thermal Mass Flowmeter and Temperature Transmitter MTI10

76

Calibration

CAL-VTM TheT=xxxisacountdowntimerindicatinghowmuchtimeislefttofinishthetest.A“PleaseWait”messagewillbeflashingonandoffonline2during this test.

F1 F2 F3 F4

Upontestcompletion,thefinalCAL-V™valuewillbedisplayedalongwitha Pass/Fail message.

Performing the Zero CAL-CHECK™ Calibration Validation Test The Zero CAL-CHECK™ test is a companion test to CAL-V™. Unlike CAL-V™,whichmaybeperformedinthepipeandatprocessconditions,ZeroCAL-CHECK™mustbeperformedatzeroflowtoensureavalidtestresult.ThistestisusedtoconfirmthattheflowmeterstillretainsitsoriginalNIST-traceablecalibrationatzeroflowandthatthesensorisfreeoffilmorresiduetatmayaffectreadings.Thetesttakeslessthan5minutestocomplete.Attheconclusionofthetest,aPassorFailmessagewill be displayed. Press F4 at the conclusion of the test to return to normal measuring mode or to terminate the test. Press FLO (F2) from the main menu. The display will show:

F1 F2 F3 F4

Press DGN (F1). The display will show:

F1 F2 F3 F4

Zero CAL-CHECK™

CAL-V = 0.51Passed OK

FLOW PARAMETER 1 DGN UNT PRM EXIT

DIAGNOSTICSIM TST EXIT

Page 77: Thermal Mass Flowmeter and Temperature Transmitter MTI10

77

Calibration

Press TST (F2). The display will show:

F1 F2 F3 F4

Press ZRO (F2) to choose the type of Zero CAL-CHECK™ test. If performingthetestinthepipe,a"noflow"conditionmustbecreated.Ifperformingoutofthepipe,themetermustberemovedandthesensorprotected by a bottle.

F1 F2 F3 F4

Performing the Zero CAL-CHECK™ - In the PipePress PIP (F1) to choose to perform the test in the pipe. The display will show:

F1 F2 F3 F4

Press VER (F1) key to verify the Zero CAL-CHECK™.

F1 F2 F3 F4

Press YES (F1) key to verify the Zero CAL-CHECK™.

F1 F2 F3 F4

Zero CAL-CHECK™

Zero CAL-CHECK™ - In-situ (In the Pipe)

ZERO CHK MENU PIP BTL EXIT

DIAGNOSTICCALV ZRO EXIT

ZERO PIPE TEST VER SET PRM EXIT

VERIFY ZRO? YES NO

Process Zero and Stable YES

Page 78: Thermal Mass Flowmeter and Temperature Transmitter MTI10

78

Calibration

WARNING:Ifyouareusing"Pipe"test,youmustverifythatthereisanoflowconditionbeforeproceeding.Ifyouareperformingthetestinabottle,besuretoisolatethesensorinabottle-anyairmovement(evenfrom a fan) can result in a false "fail" result.

Onceprocessisstable,pressYES (F4) key to begin the Zero CAL-CHECK™.

F1 F2 F3 F4

This test will take less than 5 minutes. The T=xx is a count down timer indicatinghowmuchtimeislefttofinishthetest.

F1 F2 F3 F4

Upontestcompletion,thefinalpercentagevaluewillbedisplayedalongwith a Pass/Fail message.

Performing the Zero CAL-CHECK™ - Out of PipeRemove the meter from the pipe and isolate in an area that will allow the test to go undisturbed (see Note below).

Press FLO (F2) from the main menu. The display will show:

NOTE: For best results:1. Use the factory-supplied PVC sensor protector shipped with the meter

(see Fig. 2.1 on p. 67).2. Placethemeterupside-downonaflat,solidsurfacebeforestarting

the test.3. Do not allow the meter to get jostled - make sure the meter is stable

and the sensor completely isolated.

Verifying ZRODiff = xx.xx T=xx

Diff=x.xxx % Passed OK

Zero CAL-CHECK™ - In Pipe

Zero CAL-CHECK™ - Out of Pipe

Page 79: Thermal Mass Flowmeter and Temperature Transmitter MTI10

79

Calibration

Zero CAL-CHECK™ - In Pipe

F1 F2 F3 F4

Press DGN (F1).

The display will show:

F1 F2 F3 F4

Press TST (F2). The display will show:

F1 F2 F3 F4

Press ZRO (F2) to choose the type of Zero CAL-CHECK™ test.

F1 F2 F3 F4

Press BTL (F2) to choose to perform the test out of the pipe. The display will show:

F1 F2 F3 F4

Press VER (F1) key to verify the Zero CAL-CHECK™.

DIAGNOSTIC SIM TST EXIT

DIAGNOSTICCALV ZRO EXIT

FLOW PARAMETER 1 DGN UNT PRM EXIT

ZERO CHK MENU PIP BTL EXIT

ZERO BTL TESTVER EXIT

Page 80: Thermal Mass Flowmeter and Temperature Transmitter MTI10

80

Calibration

F1 F2 F3 F4

Press YES (F1) key to verify the Zero CAL-CHECK™.

F1 F2 F3 F4

WARNING:Youmustverifythatthereisanoflowconditionbeforeproceeding. Be sure to isolate the sensor completely - any air movement (even from a fan) can result in a false "Fail" result.

Onceprocessisstable,pressYES (F4) key to begin the Zero CAL-CHECK™.

F1 F2 F3 F4

This test will take less than 5 minutes. The T=xx is a count down timer indicatinghowmuchtimeislefttofinishthetest.

F1 F2 F3 F4

Upontestcompletion,thefinalvaluewillbedisplayedalongwithaPass/Fail message.

Diff=x.xxx % Passed OK

Zero CAL-CHECK™ - In Pipe

Process Zero and Stable? YES

Verifying ZRODiff = xx.xx T=xx

VERIFY ZRO? YES NO

Page 81: Thermal Mass Flowmeter and Temperature Transmitter MTI10

81

Calibration

Troubleshooting CAL-V™If the MTI10/MTL10 Flowmeter fails a CAL-V™ Calibration Validation test,themetermustbereturnedtothefactoryforevaluation.Seethe"Returning Your Meter" section in the Appedix of this Installation and Instruction Guide or contact Spirax Sarco Applications at 1-800-356-9362 for information on how to return the meter.

Troubleshooting Zero CAL-CHECK™If the MTI10/MTL10 Flowmeter fails a Zero CAL-CHECK™ Calibration Validationtest,thereareafewreasonsthatcouldbethecause:

1. The sensor may be dirty• Try cleaning the sensor and try the test again• Ifthemeterfailsagain,moveto#2

2. The sensor may not be properly covered/isolated• Out of Pipe:

• Wind currents (fans in room included) could be affecting the sensor

• Forbestresults,besuretousethefactory-suppliedPVCsensor cover (see Fig. 5.4 on Page 67)

• Ifthefactory-suppliedPVCsensorcoverisunavailable,usea clean dry plastic beverage bottle

• In Pipe:• IfusingthePackingGlandAssembly,besurethattheShut-

off valve has been closed• Makesurethatthereisa"noflow"orzeroflowconditionon

the meter's sensor• Try the test again• Ifthemeterfailsagain,moveto#3

3. Themetermaynothavestabilizedproperly• Make sure the meter is not being affected by vibrations or other

movement• Allowthemetertostabilizewithoutbeingmovedortouchedfor

15 minutes• Try the test again• Ifthemeterfailsagain,contactSpiraxSarcoApplicationsat

1-800-356-9362

Troubleshooting CAL-V™

Troubleshooting Zero CAL-CHECK™

Page 82: Thermal Mass Flowmeter and Temperature Transmitter MTI10

82

MODBUS

Introduction ScopeThis section describes the MODBUS implementation for the Spirax Sarco MTI10/MTL10 Mass Flow Meter based on the Modicon Modbus Protocol (PI-MBUS-300 Rev. J).

MODBUS is an application layer messaging protocol that provides client/server communications between devices. MODBUS is a request/reply protocolandoffersservicesspecifiedbyfunctioncodes.

ThesizeoftheMODBUSProtocolDataUnitislimitedbythesizeconstraintinheritedfromthefirstMODBUSimplementationonSerialLinenetwork(max.RS485ApplicationDataUnit=256bytes).Therefore,MODBUS PDU for serial line communication = 256 – Server address (1 byte) – CRC (2 bytes) = 253 bytes.

RS-485 ADU = 253 + Server address (1 byte) + CRC (2 bytes) = 256 bytes.

For more information on MODBUS go to the web site http://www.modbus.org/

Command Request:<Meter Address> <Function code> <Register start address high> <Register start address low> <Register count high> <Register count low> <CRC high> <CRC low>

Command Response:<Meter Address> <Function code> <Data byte count> <Data register high> <Data register low>… <Data register high> <Data register low> <CRC high> <CRC low>Note: The data shown in brackets < > represent one byte of data.

MTI10/MTL10 Commands supportedThe MTI10/MTL10 supports the following commands:1. Function 03: Read holding registers.2. Function 04: Read input register.3. Function 06: Preset single register.

Read Holding Registers (command 03)This command reads the basic variables from the MTI10/MTL10 and has the following format:

Protocol

Page 83: Thermal Mass Flowmeter and Temperature Transmitter MTI10

83

MODBUS

Protocol Command Request:<Meter Address> <Function code=03> <Register start address high> <Register start address low><Register count high> <Register count high> <CRC low> <CRC low>

Command Response:<Meter Address> <Function code=03> <Byte count> <Data high> <Data low>…<Data high> <Data low> <CRC high> <CRC low>

Table 6.1: MTI10/MTL10 Modbus register assignments for command 0x03Register Address

Modbus Address

Data type Scaling Comment

0x00 40001 Flow in Eng unit (low) No Massflowinselectedunit0x01 40002 Flow in Eng unit (high) No0x02 40003 Total (low) No Total in selected unit0x03 40004 Total (High) No0x04 40005 Temperature (low) *10 Temperature in selected

unit * 100x05 40006 Temperature (high) *100x06 40007 Elapsed time (low) *10 Elapsed time in hours * 100x07 40008 Elapsed time (high) *100x08 40009 Velocity (Low) No Velocity in Nm/hr0x09 40010 Velocity (high) No0x0A 40011 Flow in Eng unit * 10 *10 Massflowinselectedunit

* 100x0B 40012 Flow in Eng unit *100 *100 Massflowinselectedunit

* 1000x0C 40013 Total *100 *100 Total in selected unit * 1000x0D 40014 Total2(low,2gascurvesonly) No Total #2 for 2 gas curves0x0E 40015 Total2(high,2gascurvesonly) No Total #2 for 2 gas curves0x0F 40016 Status No Status0x10 40017 Status2 No Status 20x11 40018 Spare/ Not used0x12 40019 Spare/ Not used0x13 40020 FlowinEngUnit(float,upper16bits) No Massflowinselectedunit0x14 40021 FlowinEngUnit(float,lower16bits) No Massflowinselectedunit0x15 40022 TotalinEngUnit(float,upper16bits) No Total in selected unit0x16 40023 TotalinEngUnit(float,lower16bits) No Total in selected unit0x17 40024 Total#2for2gascurve(float,upper

16 bits)No Total in selected unit

0x18 40025 Total#2for2gascurve(float,lower16 bits)

No Total in selected unit

0x19 40026 Temperatureinselectedunit(float,upper 16 bits)

No Temperature in selected unit

0x1A 40027 Temperatureinselectedunit(float,lower 16 bits)

No Temperature in selected unit

Page 84: Thermal Mass Flowmeter and Temperature Transmitter MTI10

84

MODBUS

Protocol 0x1B 40028 Elapsedtimeinhours(float,upper16bits)

No Elapsed time in hours

0x1C 40029 Elapsedtimeinhours(float,lower16bits)

No Elapsed time in hours

0x1D 40030 Velocityinselectedunit(float,upper16 bits)

No Velocity in selected unit

0x1E 40031 Velocityinselectedunit(float,lower16 bits)

No Velocity in selected unit

0x1F 40032 CAL-VDiff(floatupper16bits) No CAL-V Diff0x20 40033 CAL-VDiff(floatlower16bits) No CAL-V Diff0x21 40034 CAL-VSet(floatupper16bits) No CAL-V Set0x22 40035 CAL-VSet(floatupper16bits) No CAL-V Set0x23 40036 Spare/ Not used0x24 40037 Total24hrs,Lasttotalrecord,low

registerNo Tot24hrs: Last total record

0x25 40038 Total24hrs,Lasttotalrecord,highregister

No Tot24hrs :Last total record

0x26 40039 Total24hrs,CurrentDay(0-6) No Tot24hrs :Current Day0x27 40040 Total24hrs,CurrentHour(0-23) No Tot24hrs: Current Hour0x28 40041 Total24hrs,Recordday1,low

registerNo Tot24hrs: Record day 1

0x29 40042 Total24hrs,Recordday1,highregister

No Tot24hrs: Record day 1

0x2A 40043 Total24hrs,Recordday2,lowregister

No Tot24hrs: Record day 2

0x2B 40044 Total24hrs,Recordday2,highregister

No Tot24hrs :Record day 2

0x2C 40045 Total24hrs,Recordday3,lowregister

No Tot24hrs: Record day 3

0x2D 40046 Total24hrs,Recordday3,highregister

No Tot24hrs :Record day 3

0x2E 40047 Total24hrs,Recordday4,lowregister

No Tot24hrs :Record day 4

0x2F 40048 Total24hrs,Recordday4,highregister

No Tot24hrs :Record day 4

0x30 40049 Total24hrs,Recordday5,lowregister

No Tot24hrs: Record day 5

0x31 40050 Total24hrs,Recordday5,highregister

No Tot24hrs: Record day 5

0x32 40051 Total24hrs,Recordday6,lowregister

No Tot24hrs :Record day 6

0x33 40052 Total24hrs,Recordday6,highregister

No Tot24hrs :Record day 6

0x34 40053 Total24hrs,Recordday7,lowregister

No Tot24hrs :Record day 7

0x35 40054 Total24hrs,Recordday7,highregister

No Tot24hrs: Record day 7

0x36 40055 Total24hrs,LastTotal,lowregister No Tot24hrs :Last Total0x37 40056 Total24hrs,LastTotal,highregister No Tot24hrs :Last Total0x38 40057 ZeroCheckMean(float,floatupper

16 bits)No Zero Check Mean value

0x39 40058 ZeroCheckMean(float,floatlower16 bits)

No Zero Check Mean value

Page 85: Thermal Mass Flowmeter and Temperature Transmitter MTI10

85

MODBUS

Protocol 0x3A 40059 ZeroCheckStdev(float,floatupper16 bits)

No Zero Check Standard deviation

0x3B 40060 ZeroCheckStdev(float,floatlower16 bits)

No Zero Check Standard deviation

0x3C 40061 ZeroCheckPipeRef(float,floatupper16 bits)

No Zero Check Pipe Reference

0x3D 40062 ZeroCheckPipeRef(float,floatlower16 bits)

No Zero Check Pipe Reference

0x3E 40063 ZeroCheckPipeDiff(float,floatupper16 bits)

No Zero Check Pipe Difference %

0x3F 40064 ZeroCheckPipeDiff(float,floatlower16 bits)

No Zero Check Pipe Difference %

0x40 40065 ZeroCheckBottleRef(float,floatupper 16 bits)

No Zero Check Bottle Reference

0x41 40066 ZeroCheckBottleRef(float,floatlower 16 bits)

No Zero Check Bottle Reference

0x42 40067 ZeroCheckBottleDiff(float,floatupper 16 bits)

No Zero Check Bottle Difference %

0x43 40068 ZeroCheckBottleDiff(float,floatlower 16 bits)

No Zero Check Bottle Difference %

0x44 40069 ZeroCheckTestTime(integer,lower16 bits)

No Zero Check Test Time (sec)

* The data in registers with scaling must be multiplied by 10 or 100 as indicated to be scaled properly. Note:RegistersA,B&Careprovidedtogetmoreresolutionforlowflowandtotal.Whenthevalueexceedsthemaximumvalueofthe16bitregisters,theywillbefrozenwithall16bitsset.Itisalsopossibletousethevelocitytocalculatetheflowinengineeringunitsbyusingthepipearea and conversion factor for the selected unit.

Example:Request data register at starting address 0x0000 and specifying only 1 register:<0x01> <0x03> <0x00> <0x00> <0x00> <0x01> <0x0a> <0x84>

Command Response:<0x01> <0x03> <0x02> <xx> <xx> <CRC high> <CRC low> Where xx xx is the data register value.

Read Input Register (MTI10/MTL10 status and status 2 command 04)This command is used to report the MTI10/MTL10 status information. It is a READ ONLY command.

Page 86: Thermal Mass Flowmeter and Temperature Transmitter MTI10

86

MODBUS

Command Request:<Meter Address> <Function code=04> <Register address =0> <Register address =0><Register count =0> <Register count =1> <CRC high> <CRC low>

Command Response:<Meter Address> <Function code=04> <Byte count-=2> <Status High><Status Low><CRC high> <CRC low>

The MTI10/MTL10 supports only reading of the MTI10/MTL10 status. The registeraddressmustbesettozero(ModbusAddress30001)andtheregister count must be set to 1.

Table6.2:Statusbitsdefinitions(ModbusAddress30001)Bit Definition Comment0 Power up indication Reset when out of the power up sequence1 Flow rate reached high limit threshold Setlimittozerotodisable2 Flow rate reached low limit threshold Setlimittozerotodisable3 Temperature reached high limit threshold Setlimittozerotodisable4 Temperature reached low limit threshold Setlimittozerotodisable5 Sensor reading is out of range Check sensor wiring6 Velocityflowrateoutsideofcalibrationtable Check sensor wiring7 Incorrect Settings Check settings8 In simulation mode Set simulation value to 0 to disable9 Frequency output is out of range Check frequency output settings10 Analog4-20mAforflowisoutofrange Check analog output settings11 Analog 4-20 mA for temperature is out of range Check analog output settings12 Busy Check wiring from RS485 to Anybus IC13 Bridge Shutdown Check RTC14 CRC error Check parameters and reset CRC15 Tot Error Reset total

Table6.3:Status2bitsdefinitions(ModbusAddress30002)Bit Definition Comment0 CAL-V in progress CAL-V in progress1 ADC12<>ADC24 too far apart Internal ADC calibration out of range2 CAL-V Diff out of range CAL-V Diff out of range3 Curve #2 Selected (for 2 gas curve only) Curve #2 Selected (2 gas curve only)4 Zero Check Failed Zero Check Failed5 CAL-V/Zero Check Aborted CAL-V/Zero Check Aborted

Protocol

Page 87: Thermal Mass Flowmeter and Temperature Transmitter MTI10

87

MODBUS

Protocol Preset Single Register (command 06)This command is used to perform miscellaneous functions such as clearingthetotalizerandinitiatingdiagnosticoperations.Theregisteraddressis0x0a(10decimal,Modbus=40011)andthedatatowriteisdescribed on the following page.

Command Request: <Meter Address> <Function code=06> <Register address high=0x00> <Register address low=0x0a><Register data high=0x00> <Register data low =0x02> <CRC high> <CRC low>

Command Response:<Meter Address> <Function code=06> <Register address =0x00> <Register address =0x0a><Register data=0x00> <Register data =0x02> <CRC high> <CRC low>

Reset Total:Address=40011,data=0x02ThiscommandisusedtocleartheTotalizerandelapsedtimeregisters

Reset 24 hours Total:Address=40011,data=180(0xB4)This command reset the 24 hours 7 days record including the day and hours counters

Reset 24 hour time:Address=40011,data=181(0xB5)This command reset the 24 hours day and hours counters

24 hours Event:Address=40011,data=182(0xB6)Thiscommandgeneratesa24hoursevent,thesamewayaswhenthe24 hours counter rolls over.This may be useful to record total over a shorter period.

Page 88: Thermal Mass Flowmeter and Temperature Transmitter MTI10

88

MODBUS

Protocol CAL-V Verify:Address=40011,data=161(0xA1)Thiscommandinitiatesa“CAL_VVerify.Thisoperationmaytake4minutestocompleteandwillstopthemeterfromcalculationflow.TheStatus2 bit D0 may be monitored to check for completion.

Zero Check In-Pipe Verify:Address=40011,data=173(0xAD)Thiscommandinitiatesa“ZeroCheckIn-PipeVerify”.Thisoperationdoesnotaffectflowcalculations.Theregister40069maybemonitoredtocheck for completion.

Zero Check In-Bottle Verify:Address=40011,data=176(0xB0)Thiscommandinitiatesa“ZeroCheckIn-BottleVerify”.Thisoperationdoesnotaffectflowcalculations.Theregister40069maybemonitoredtocheck for completion.

Switch to Curve #1:Address=40011,data=170(0xAA)This command initiates a command to switch to gas curve 1 when configuredfor2gascurves.Makesurethattheinputcontactisnotprogrammed for curve switching

Switch to Curve #2:Address=40011,data=171(0xAB)This command initiates a command to switch to gas curve 2 when configuredfor2gascurves.Makesurethattheinputcontactisnotprogrammed for curve switching

Page 89: Thermal Mass Flowmeter and Temperature Transmitter MTI10

89

MODBUS

RS-485 Wiring The RS-485 Modbus communication wiring connections are made to terminal block TS5 of the MTI10/MTL10 RS-485 board. The Tx/Rx+signalmustconnecttopin1,Tx/Rx-mustconnecttopin2,communicationcommontopin3,andthecableshieldtopin4asshowinFigure 4.1.

RS-485 Termination ResistorConnect a termination resistor across the receive/transmit signals of the last device on the RS-485 communication line. To connect the 121ohm terminationresistorontheMTI10/MTL10,setjumperJP1totheTERMposition. Disconnect the termination resistor on all other external RS-485 devices. The termination resistor of the MTI10/MTL10 is disconnected by setting jumperJP1totheNC(NotConnected)position.Seefigure4.1.

Figure6.4:RS-485WiringandTerminationResistorConfiguration

Tx/Rx(+)Tx/Rx(-)

Communication Common

Cable Shield

Termination Resistor Jumper JPI

Page 90: Thermal Mass Flowmeter and Temperature Transmitter MTI10

90

MODBUS

ProgrammingUsing The LocalDisplay

Entering the programming modeToentertheprogrammingmode,presstheF1orF2keysrepeatedlyinthenormal running mode until the following screen is displayed.

SET PARAMETERS? NO YES

F1 F2 F3 F4

Press YES (F4) and the following screen will prompt the user to enter the password if enabled:

PASWD: UP DN NXT OK

F1 F2 F3 F4

Enter the correct password (Note: the default password for Level 1 is 1234).

PASWD: UP DN NXT OK

F1 F2 F3 F4

Press the UP (F1) or DN (F2) keys to scroll to a new digit or character. The cursor indicates the selected digit. Press NXT (F3) to select the next digit and OK (F4) to accept the entry. Note: If the UP (F1) or DN (F2) keys are held down for more than 1 second,newdigitswillbeselectedatanincreasingrate.

Ifanincorrectpasswordisentered,themessage“WrongPassword”will be displayed for a few seconds and then return to the programming entryscreen.Ifthepasswordisaccepted,thefollowingscreenwillbedisplayed:

SET PARAMETERS I/O FLO DSP EXIT

F1 F2 F3 F4

This is the base screen of the programming mode.

Page 91: Thermal Mass Flowmeter and Temperature Transmitter MTI10

91

MODBUS

CommunicationProtocol and Parameters

Note:Iftheprogrammingmodemustbeexited,pressEXIT(F4)repeatedlyuntilthe“NormalMode”screenisdisplayed.Toprogramthecommunicationparameters,pressI/O(F1)keyfromthebase screen of the programming mode.

SET I/O I/O FEQ 420 EXIT

F1 F2 F3 F4

Then press I/O (F1) again:

SET I/O COM CTC EXIT

F1 F2 F3 F4

Then press COM (F1) to select communication parameters

Set Bus protocol:

Bus=Modbus NXT OK

F1 F2 F3 F4

Press NXT (F1) repeatedly until the correct selection is shown and then press OK (F4) to accept the setting.Selections are: “Modbus” “None”Set Baud Rate communication parameter:

Baud=9600 NXT OK

F1 F2 F3 F4

Page 92: Thermal Mass Flowmeter and Temperature Transmitter MTI10

92

MODBUS

CommunicationProtocol and Parameters

Press NXT (F1) repeatedly until the correct Baud Rate is shown and then press OK (F4) to accept the setting.Selectionsare:“19200”“9600”“4800”“2400”“1200”

Parity=EVEN NXT OK

F1 F2 F3 F4

Press NXT (F1) repeatedly until the correct Parity is displayed and press OK (F4) to accept the setting.Selectionsare:“NONE”“ODD”“EVEN”Set Unit Address:

Address=01 CHG OK

F1 F2 F3 F4

Press CHG (F1) key to change the Modbus communication Address.

Address=01 UP DN NXT OK

F1 F2 F3 F4

Press the UP (F1) or DN (F2) keys to select the value for the Modbus Address. The cursor points to the selected digit. Press NXT (F3) to select the next digit and OK (F4) to accept the entry.

Page 93: Thermal Mass Flowmeter and Temperature Transmitter MTI10

93

MODBUS

CommunicationProtocol and Parameters

Note: If the UP (F1) or DN (F2) keys are held down for more than 1 second,newdigitswillbeselectedatanincreasingrate.

It is very important that there are not two devices with the same Modbus address.Toavoidconflicts,eachModbusslavemusthaveauniqueaddress. Range is from 1 to 247.

Note: Power to the MTI10/MTL10 must be cycled off and on for new Modbus settings to take effect.

Page 94: Thermal Mass Flowmeter and Temperature Transmitter MTI10

GB

ES

WARNING!BEFOREATTEMPTINGANYMAINTENANCE,TAKETHENECESSARYSAFETY PRECAUTIONS BEFORE REMOVING THE PROBE FROM THE DUCT (EXAMPLE:PURGELINESOFTOXICAND/OREXPLOSIVEGAS,DEPRESSURIZE,ETC...).

WARNING! EXPLOSION HAZARD. DO NOT REMOVE OR REPLACE COMPONENTS OR FUSES UNLESS POWER HAS BEEN DISCONNECTED WHEN A FLAMMABLE OR COMBUSTIBLE ATMOSPHERE IS PRESENT.

WARNING! EXPLOSION HAZARD. DO NOT DISCONNECT EQUIPMENT WHEN A FLAMMABLE OR COMBUSTIBLE ATMOSPHERE IS PRESENT.

WARNING! TURN OFF INPUT POWER BEFORE REMOVING OR INSTALLING A CIRCUIT BOARD ASSEMBLY FROM THE ENCLOSURE.

Access to ElectronicsAccessing electronics is not normally required for maintenance purposes. If a loose connectionissuspected,unscrewtherearend-capofthemeterenclosuretoaccesstheterminations

CAUTION: BE SURE POWER TO METER IS SWITCHED OFF BEFORE ATTEMPTING TO ACCESS ELECTRONICS. Ifthereisaproblemandalooseconnectionisnotfound,please contact Spirax Customer Service for technical assistance at (800) 356-9362.

¡ADVERTENCIA!ANTESDEINTENTARCUALQUIERMANTENIMIENTO,TOMELAS PRECAUCIONES DE SEGURIDAD NECESARIAS ANTES QUE RETIRAR LA SONDA DEL DUCTO (EJEMPLO: PURGUE LAS LÍNEAS DE GASES TÓXICOS Y/O EXPLOSIVOS,DESPRESURICE,ETC...).

¡ADVERTENCIA! PELIGRO DE EXPLOSIÓN. NO RETIRE O REEMPLACE COMPONENTES O FUSIBLES A MENOS QUE LA ENERGÍA HAYA SIDO DESCONECTADA SIEMPRE QUE ESTÉ PRESENTE UNA ATMÓSFERA INFLAMMABLE O COMBUSTIBLE.

¡ADVERTENCIA! PELIGRO DE EXPLOSIÓN. NO DESCONECTE NINGÚN EQUIPO CUANDO UNA ATMÓSFERA INFLAMABLE O COMBUSTIBLE ESTÉ PRESENTE PRONTO.

¡ADVERTENCIA! DESCONECTE LA ENERGÍA DE ALIMENTACIÓN ANTES DE REMOVER O INSTALAR UN ENSAMBLE DE TARJETA DE CIRCUITO DEL GABINETE.

Acceso a la ElectrónicaNormalmente no se requiere tener acceso a la electrónica para propósitos de mantenimiento.Sisesospechadeunaconexiónsuelta,desatornillelatapaposteriordela caja del medidor para tener acceso a las terminales

PRECAUCIÓN: ASEGÚRESE QUE LA ALIMENTACIÓN DEL MEDIDOR ESTE DESCONECTADA ANTES DE INTENTAR EL ACCESO A LA ELECTRÓNICA. Si existe algúnproblemaynoseencuentraningunaconexiónsuelta,porfavorpóngaseencontacto con el Servicio al Cliente de Spirax para asistencia técnica al número (800) 356-9362.

Precautions

Precauciones

94

Maintenance: Precautions

Page 95: Thermal Mass Flowmeter and Temperature Transmitter MTI10

AVERTISSEMENT!AVANTTOUTETENTATIVEDEMAINTENANCE,OBSERVERLES CONSIGNES DE SECURITE NECESSAIRES AVANT DE RETIRER LA SONDE DELACONDUITE(PAREXEMPLE,PURGERLESLIGNESDESGAZEXPLOSIFS/TOXIQUESQU’ELLESPOURRAIENTCONTENIR,DEPRESSURISERLECONTENEUR,ETC.).

AVERTISSEMENT ! RISQUE D’EXPLOSION. NE PAS RETIRER NI REMPLACER DES COMPOSANTS OU DES FUSIBLES SI LA SOURCE D’ALIMENTATION N’A PAS ETE DEBRANCHEE DANS UNE ATMOSPHERE INFLAMMABLE OU COMBUSTIBLE.

AVERTISSEMENT ! RISQUE D’EXPLOSION. NE PAS DEBRANCHER UN EQUIPEMENT DANS UNE AMBIANCE COMBUSTIBLE OU INFLAMMABLE.

AVERTISSEMENT ! COUPER L’ALIMENTATION AVANT DE RETIRER OU D’INSTALLER UN ENSEMBLE DE CARTE DE CIRCUITS IMPRIMES DU BOITIER.

Accès aux composants électroniquesL’accès aux composants électroniques n’est généralement pas nécessaire dans le cadredelamaintenance.Siuneconnexionlâcheestsuspectée,dévisserlecapuchond’extrémité arrière du boîtier du compteur pour accéder aux terminaisons

ATTENTION : S’ASSURER QUE L’ALIMENTATION DU COMPTEUR EST COUPEE AVANT D’ACCER AUX COMPOSANTS ELECTRONIQUES. En cas de problème et qu’aucuneconnexionlâchen’estdétectée,veuillezcontacterleserviceclientdeSpiraxpour obtenir une assistance technique au (800) 356-9362.

ACHTUNG! BITTE ERGREIFEN SIE DIE ERFORDERLICHEN SICHERHEITSMAßNAHMEN,BEVORSIEIRGENDWELCHEWARTUNGSARBEITENDURCHFÜHREN UND DIE MESSSONDE AUS DEM ROHR ENTFERNEN (BEISPIEL: LEITUNGEN ZUR ENTFERNUNG VON GIFTIGEN UND/ODER EXPLOSIVEN GASENREINIGEN,DRUCKSENKEN,USW.).

ACHTUNG! EXPLOSIONSGEFAHR. KOMPONENTEN ODER SICHERUNGEN BITTE ERSTDANNENTFERNENODERAUSTAUSCHEN,WENNDERSTROMGETRENNTWURDE,FALLSEINEENTZÜNDLICHEODERBRENNBAREATMOSPHÄREVORHANDEN IST.

ACHTUNG!EXPLOSIONSGEFAHR.GERÄTNICHTVOMSTROMTRENNEN,WENNEINEENTZÜNDLICHEODERBRENNBAREATMOSPHÄREVORHANDENIST.

ACHTUNG!EINGANGSLEISTUNGAUSSCHALTEN,BEVORLEITERPLATTENBAUGRUPPENAUSDEMGEHÄUSEAUSGEBAUTODERINDIESES EINGEBAUT WERDEN.

Zugriff auf die ElektronikDerZugriffaufdieElektronikistzuWartungszweckennormalerweisenichterforderlich.FallseineloseVerbindungvermutetwird,schraubenSiediehintereEndkappedesMessgerätgehäusesab,umaufdieAnschlüssezugreifenzukönnen.

VORSICHT:STELLENSIESICHER,DASSDERSTROMANDASMESSGERÄTAUSGESCHALTETIST,BEVORSIEVERSUCHEN,AUFDIEELEKTRONIKZUZUGREIFEN. Falls Probleme auftreten und keine lose Verbindung gefunden werdenkann,wendenSiesichbittezweckstechnischerUnterstützungandenSpirax-Kundendienst unter der Nummer (800) 356-9362.

FRPrécautions

DE

Vorsichtsmaßnahmen

95

Maintenance: Precautions

Page 96: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Broken or Damaged ProbeIfthesensorisbrokenordamaged,theprobeandelectronicsmustbereturnedtothefactory.Anew sensor will be installed and calibrated. Refer to "Returning Your Meter" on p. 124.

Flow Calibration and Calibration ValidationToensurecontinuedhighaccuracyofyourModelMTI10/MTL10flowmeter,SpiraxSarcoprovidesa full NIST traceable calibration. It is recommended that the meter be returned to Spirax Sarco for a calibration check in our NIST traceable labs after three years of operation. The CAL-V™ and ZeroCAL-CHECK™testsshouldbeperformedannuallyorasrequiredbyfederal,stateandlocalagencies,ifapplicable.

Fuse ReplacementVerify the fuse is defective by measuring it with an Ohm Meter (Two replacement fuses are provided with each unit). The fuse F1 is located near the power terminal block and can be removed by using tweezersorneedle-nosepliers.ReplacementfuseisLittelfusepartnumber0454.750MR.

Warning:• Turn input power OFF before removing or installing a fuse. Use only recommended fuse

replacements.• Itistheuser’sresponsibilitytoinstalltheflowmeterinanappropriatelydesignedsystem

with adequate safety protections. • DONOTremovetheflowinstrumentfromtheflowbodywhilethesystemisunderflow

conditions. • This product may experience temperatures from as low as -40F (-40C) and as high as

649F (343C). It is the user’s responsibility to take safety precautions regarding operating temperatureoftheflowmeter.

• Iflargeflowbodyhasbeenpurchased,userisadvisedtouseadoubleslingtopreventrotationofflowbodyduringinstallationorothermethodtopreventdamageofflowinstrument.

Sensor Wiring Note: Sensor terminations are performed at the factory except when the Remote Electronics option is used or ordered.

Sensor CleaningThesensorisinsensitivetosmallamountsofresidue,butcontinueduseindirtyenvironmentswillnecessitateperiodiccleaning.Toinspectthesensor,removepowerfromelectronicsandremovetheunitfromthepipeorduct,exposingthesensorelements.Iftheyarevisiblydirty,cleanthemwith water or alcohol (ethanol) using an appropriate brush until they appear clean again. Even thoughthesensorelementsareruggedandbreakageresistant,avoidtouchingthemwithanysolidobject and use a light touch while cleaning them.

96

Maintenance: General

Page 97: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Problem Possible Cause Action

Display-Main BdComm. Error

Display and main board not communicating

Check status of LP1 on the Main Board and LP6 on the display board. Are green LEDs blinking once per second? IfLEDsarenotblinking,cycle power to reset me-ter. Call Spirax Sarco Tech Support.

Meter does not read up to full scale

Calibration table may be corrupted

•Check the calibration table for a corrupted location. •Enter the password 9111. •Start on p. 42 of the MTI10/MTL10 manual and follow the steps to get to Flow Parameters 2 menu screen. •Select CAL (F1) •Select NXT (F1) to cycle through calibration table to verify entries match calibrationcertificate.•Check for CRC error code

Velocity measurement seems low

1.Probe not oriented properly

2.Sensor dirty

1.Orient probe per installation sections: Insertion(p.9),Inline(p. 13).

2.Clean sensor (p. 96)

TroubleshootingCaution!Theelectronics,sensorandsensorinterconnectwiressuppliedbySpiraxSarcoarecalibratedasasingleprecisionmassflowmeter.Interchanging sensors or sensor wiring will impair the accuracy of the flowmeter.IfyouexperienceanyproblemwithyourMTI10/MTL10,callSpirax Sarco Technical Support at 800-356-9362.

97

Troubleshooting: GeneralPrecautionsTroubleshooting

Page 98: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Problem Possible Cause Action

Unit will not power-up a)No power inputb)Bad fusec)Bad Power supply

•Check fuse (F1) located next to TS1 on main board.•Check for correct power supply voltage at TS1 on main board.

If fuse is OK and unit stillwon’tpowerup,callSpirax Sarco for additional assistance

Meterdoesnotinitialize Electromechanical interference

•Check meter power cycles.•Press and release F1 andF2atthesametime;the display will enter Engineering screens.•Press F1 to get to screen #23;recordpowercyclevalue. •Press F4 to return to normaloperation;monitormeter until problem returns. •Return to screen #23 to see if power cycles have increased;microprocessoris resetting due to EMI electrical noise entering the meter. •Check Power input and output cables grounding and routing.

9898

Troubleshooting: General

Page 99: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Problem Possible Cause Action

Velocity measurement is erraticorfluctuating

1.Veryturbulentflow

2.Sensor dirty

3.Sensor broken

4.Probe not mounted securely

5.Malfunctioninflowmeter

6.Meter installed incorrectly

1.Increase dampening (see filtersettingsin"FlowPa-rameters" on p. 42)

2.Clean sensor (Refer to Maintenancesection,p.96)

3.ReturnflowmetertoSpi-rax Sarco for repair (Refer to p. 124 for shipping instructions)

4.Remount probe (see Installationsection,p.9andp.13);mustbemounted securely with-out vibration. If vibration persists,chooseanewmounting location without location.

5.ReturnflowmetertoSpi-rax Sarco for repair (Refer to p. 124 for shipping instructions)

6.Re-install meter accord-ing to instructions (Refer to installationsection,p.9and p. 13)

99

Troubleshooting

99

Troubleshooting: General

Page 100: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Installation ProblemsThe following is a summary listing of problems that may be encountered with the installation of the MTI10/MTL10 Thermal Mass Flowmeter.1. Improper wiring connections for power and/or 4 to 20mA output

signal. The MTI10/MTL10 require a separate power source for the main board and the two 4 to 20mA output signals. Two wires supply 24VDC power to the main board. Two wires are used for each of the 4 to 20mA output signals. Refer to Figure 3.4 (p. 19) and Figure 3.5 (p. 20). Also refer to "Wiring Precautions" and "Helpful Hints" in Wiring section (p. 15) for further guidance.

2. Inadequate power source. Forthosemodelsthatarepoweredby24VDC,a24VDC±10%,0.75 amp or greater power supply is recommended. If the voltage supplied is not within this range or if the power supply is not rated for25wattsminimum,avarietyofproblemscanoccurincludinginaccurateflowreadings,dimdisplayandfaultyprogrammingaction.The input voltage must be within the range of 23 to 25VDC as measuredatthepowerinputterminalsoftheflowmeterelectronics.

3. Flow measurement seems inaccurate.• ChecktoensurethattheflowmeterisinstalledsothattheFlow

DirectionArrowengravedontheflatsurfaceofthefittingbelowthe electronics housing is properly pointing in the direction of flow.RefertoFigure2.7(p.14).Ifnot,changeorientationofmeter.

• IfyouhaveaMTI10insertiontypeflowmeter,checkthattheinsertion depth of the sensor/probe is correct. The end of the probe should be adjusted as per Figure 2.2 (p. 10).

• Forinlinemetertypes,ensurethatthereareaminimumoftendiametersofstraightpipeupstreamofthesensorandfivediametersdownstream(¼"meters:6”[152mm]ofstraight,unobstructed pipe upstream and downstream required). For insertionmetertypes,ensurethatthereareaminimumoffifteendiametersofstraightpipeupstreamofthesensorandtendiametersdownstream.Ifcomplexflowdisturbancesareupstreamofthesensor,extensionofthestraightpipemayberequiredtoensureaccurateflowmeasurement.ContactSpiraxSarco for assistance.

• Ensure that pipe area data in the meter matches data on the SpiraxSarcoCalibrationCertificate.Thepipeinternalcrosssectionalareaisprogrammedintotheflowmeterthroughthefrontpanel(seeFlowParameters,p.42).Thisareaisprogrammed in square feet or square meters. The Calibration Certificatedeliveredwiththeflowmetercontainstheareathat

100100100

Troubleshooting: Installation Problems

Problem Summary

Troubleshooting: Installation Problems

Page 101: Thermal Mass Flowmeter and Temperature Transmitter MTI10

wasprogrammedintotheflowmeterattheSpiraxSarcofactory.Check to ensure that this area is correct.

4. Erraticflowreadingespeciallyaflowreadingspikinghigh. Thismaybeasymptomofmoistureintheflowstream.MTI10/MTL10flowmetersaredesignedtoworkinrelativelydrygasapplications only. Contact Spirax Sarco to discuss resolutions to this problem.

5. Flowmeterisnotrespondingtoflow. This problem could be caused by a number of reasons: • Checktoensureadequatepowerissuppliedtotheflowmeter

asdescribedabove.Ifthingsappeartobecorrect,aneasyfunctional test can be performed. Carefully remove the probe andsensorfromthepipeorflowbody.Caution:thesensorisHOT.Forthoseflowmeterswithadisplay-andifthedisplayisreadingzero-blowonthesensortoseeifaresponseoccurs.Ifnothinghappens,takeadampragorspongeandplaceitincontact with the sensor. A reading should occur. Contact Spirax Sarco Technical Support with this information.

• Acorruptedcalibrationtablemayleadtoazeroflowreading.Verify that all Cal Flow Parameter settings are correct by accessing the "Calibration Parameters" information on meter (see p. 45). Check meter data for any non-whole numbers and call customer service for assistance.

6. Displayand/or4to20mAsignalreadingabovezeroflowwhennoflowisoccurringinthepipe.Ifthereadingislessthan5%offullscale,itislikelythisisanormalconditioncausedbyconvectionflowcreatedbytheheatedsensor.Itdoesnotmeanthatthezerooftheinstrumentisimproperlyset.TheSpiraxSarco,sensorisextremelysensitivetogasflowandcanevenreadthesmallflowcausedbyconvection.Ifthisisanunacceptablecondition,pleasecontactSpiraxSarcoTechnicalSupportforalternatives.

7. Mismatched serial numbersIfyouhavemorethanonemeter,youmustensurethattheserialnumbersofmeter,remote,and/orflowbodymatchoneanother.These items have been manufactured and calibrated to operate as a unit and cannot be mismatched.

101101

Installation Problems

Troubleshooting: Installation Problems

Page 102: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Alarm Code

Reason Action

13 Flow rate above high limits

Refer to the PARAMETER MENU 2 section on p. 42 of this Manual to verify limit is within range. Check ALM = HiFloAlm under PRM.

14 Flow rate below low limits

Refer to the PARAMETER MENU 2 section on p. 42 of this Manual to verify limit is within range. Check ALM = LoFloAlm under PRM.

15 Temperature above high limits

Refer to the PARAMETER MENU 2 section on p. 42 of this Manual to verify limit is within range. Check ALM=HiTempAlm under PRM.

16 Temperature below low limits

Refer to the PARAMETER MENU 2 section on p. 42 of this Manual to verify limit is within range. Check ALM = LoTempAlm

22 Sensor out of range Refer to the ENGINEERING DISPLAY MENU on p. 61 of this Manual and the SpiraxSarcofactoryCalibrationCertificateto check CSV voltage. Compare Display 10 valuetoCalibrationCertificateCSVvoltageand verify it's within range.

23 Velocity out of calibration table range

Refer to the ENGINEERING DISPLAY MENU on p. 61 of this Manual and the SpiraxSarcofactoryCalibrationCertificateto check CSV voltage. Compare Display 10 valuetoCalibrationCertificateCSVvoltageand verify it's within range.

24 Check settings One or more internal settings are corrupted or out of spec. Contact Spirax Sarco Service for instructions to verify settings.

25 Simulation mode Meter is in Simulation Mode. Refer to the PARAMETER MENU 1 section on p. 48 of this Manual. Use the SIM Section under Diagnostics to return to normal operation.

26 Frequency output over range

Refer to the DIGITAL OUTPUT MENU on p. 57 of this Manual. Verify the Frequency Output settings are within limits.

Alarm CodesInformation to diagnose and clear alarm codes is on p. 59 under the Menu Tree section. Enter password (9111) and follow the block diagram to get to the section affected by the error code.

102

Troubleshooting: Alarm Codes

Alarm Codes

Page 103: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Alarm Code

Reason Action

32 4to20mAforflowrateis out of range

Refer to the MAIN MENU on p. 56 of this Manual. Use the Set I/O section to verify range limits under FLO Set 4 to 20mA.

33 4 to 20mA for temperature is out of range

Refer to p. 44 of this Manual. Use the Set I/O section to verify range limits under FLO Set 4 to 20mA. Channel #2 can be set forflowortemperature.

34 Busy Meter is recalculating new parameters.

35 Sensor Bridge Shutdown

The MTI10/MTL10 probe is getting too hot. Open sensor circuit and check sensor wiring.

36 Database CRC Error Refer to the PARAMETER MENU 2 section on p. 42 of this Manual to reset CRC. Use SPC section of menu to reset CRC. Contact Spirax Sarco Technical Support for possible causes.

37 TotalizerErrorDetected

See "Reset the Total and Elapsed Time" on p. 46 for steps to clear Error Code. Contact Spirax Sarco for possible causes.

38 CAL-V in progress WaituntiltheCAL-VorCalSetisfinished.

39 ADC12 versus ADC24 too far

The tolerance between the 2 ADC is out of specification.Recallingmanufacturedefaultmay correct the problem.

40 CAL-V Diff Fail The CAL-V Diff Failed. Check sensor wiring and verify that the sensor's resistance is correct. Call Tech Support.

41 Zero CAL-CHECK Fail Allowmetertostabilizefor15minutesandperform the test again. If another "Fail" test results,callTechSupport.

103

Troubleshooting: Alarm Codes

Alarm Codes

Page 104: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Performance SpecsFlow Accuracy:

MTL10 Inline Flowmeter: ± 1% of reading ± 0.2 % of full scale. ¼"size:6”(152mm)ofstraight,unobstructedpipeupstreamanddownstreamrequired.Allothersizes:8diametersofstraight,unobstructedpipeupstreamand4downstreamrequired.

MTI10 Insertion Flowmeter: ± 1% of reading ± 0.2 % of full scale. 15diametersofstraight,unobstructedpipeupstreamand10downstreamrequired.

Flow Repeatability: ± 0.2% of full scale

Flow Response Time: 0.9 seconds (one time constant)

Temperature Accuracy:±1.8°F(±1.0°C)-40to250°F(-40to121°C);±3.6°F(±2.0°C),250to650°F(121to343°C);60SFPMminimum.

Calibration:Factory Calibration to NIST traceable standardsCAL-V™&ZeroCAL-CHECK™:Insitu,operator-initiatedcalibrationvalidation

Operating SpecsUnits of Measurement (field selectable): SCFM,SCFH,NMPS,NM3/M,NM3/H,NM3/D,NLPS,NLPM,NLPH,SCFD,MSCFD,MMSCFD,MMSCFM,SMPS,SM3/H,SM3/D,SM3/M,LB/S,LB/M,LB/H,LB/D,KG/S,KG/M,KG/H,SLPM,SFPM,MT/H,MCFD

Flow Rates for MTI10 Insertion Flowmeter: 15to60,000SFPM(0.07to280NMPS)-Airat70°F(20°C)&1ATMTurndown:upto1000:1;100:1typical

Typical Flow Ranges for Insertion FlowmetersPipe size SCFM NM3/hr

1.5" (40mm) 0-840 0-1,320

2" (50mm) 0-1,400 0-2,200

3" (80mm) 0-3,080 0-4,860

4" (100mm) 0-5,300 0-8,360

6" (150mm) 0-12,000 0-18,900

8" (200mm) 0-20,800 0-32,800

12" (300mm) 0-46,600 0-73,500

104

Appendices:Specifications

Page 105: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Full Scale Flow Ranges for Inline Flowmeters:

Size SCFM NM3/hr0.25" 0-20 0-320.5" 0-90 0-140

0.75" 0-180 0-2801" 0-320 0-500

1.25" 0-580 0-9101.5" 0-840 0-1,3202" 0-1,400 0-2,200

2.5" 0-2,000 0-3,1503" 0-3,080 0-4,8604" 0-5,300 0-8,3606" 0-12,000 0-18,900

Note: Standard conditions of air at 70°F and one atmosphere. Consult factory for other gases and for flow rangesabovethoselisted.Inlinemetersabove5,000SCFM(7,900NM3/H)airmayrequirethirdpartyCalibration. Contact Spirax Sarco.

Gas Pressure (maximum): Insertion: 500 psig (34.5 barg) Inline(1/4"through6"):NPT500psig(34.5barg);150#flange230psig(16barg) Check with factory for higher pressure options.

Note: Pressure ratings stated for temperature of 100°F (38°C).

RelativeHumidity:90%RHmaximum;non-condensing

MaximumAltitude:6,562ft(2,000m)max.

Temperature: ST sensor: -40 to 250°F (-40 to 121°C)

HT Sensor: -40 to 650°F (-40 to 343°C) Enclosure:

Without display or AC power supply: -40 to 158°F (-40 to 70°C)Withdisplayand/orACpowersupply:-4to158°F,(-20to70°C)Remote sensor junction box: -40 to 212°F (-40 to 100°C).

Input Power:24 VDC (± 10%), 0.7 Amps (standard DC power) 100 to 240VAC~(+10%/-15%), 50-60Hz, 0.2 Amps (with AC power option)

Note: Fluctuations of AC and DC power supply are not to exceed ± 10% of rating.

— – – –

105

Appendices:Specifications

Page 106: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Class I Equipment (Electrical Grounding Required for Safety). Installation (Over-voltage) Category II for transient over-voltages.

Outputs:Twoisolated4to20mAoutputs(outputoneisforflowrateandoutputtwoisprogrammableforflowrateortemperature);faultindicationperNAMURNE43.Isolatedpulseoutput0to100Hz,5to24voltsp/pforflow(thepulseoutputcanbeusedasanisolatedsolidstateoutputforalarms);10mAmax.

Serial Communication:USB communication port is standard. The free PC-based software tool - MTI10/MTL10 View™ - providescompleteconfiguration,remoteprocessmonitoring,anddataloggingfunctions.Optional serial communication: RS485 Modbus.

4 to 20mA Loop Verification: Simulation mode used to align 4 to 20mA output with the input to customer’s PLC/DCS.

Physical SpecsSensor material:

316stainlesssteelstandard;HastelloyC276optional

Enclosure:NEMA4X(IP68),aluminum,dualconduitentrieswith¾”NPToroptionalM20x1.5mm.Cabling toremoteenclosure:5-conductor,18AWG,twisted,shielded,100feetmaximum.

Retractor Assemblies: Packing gland assembly: 125 psig (8.6 barg ) max. Highpressure(crank)retractor:NPT600psig(41.4barg),ANSI150flange&ANSI300flange,no valve supplied.

Insertion Flowmeter Installation: SpiraxSarco-suppliedcompressionfittingconnectstocustomer-supplied¾”femalecouplingwelded to pipe.

106

Appendices:Specifications

Page 107: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Agency ApprovalsCE: Approved EMCDirective;2004/108/ECEmissions and Immunity Testing: EN61326-1:2008Low Voltage Directive (LVD): 2006/95/ECProduct Safety Testing: EN 61010-1: 2010Pressure Equipment Directive: 97/23/ECWeldTesting:ENISO15614-1andENISO9606-1,ASMEB31.3

FM and FMc: ApprovedClassI,Div.1,GpsB,C,D;ClassII,Div.1,GpsE,F,G;ClassIII,Div.1;T3C,Ta=-40°Cto70°C;ClassI,Zone1,AEx/ExdIIB+H2(T6,T4,orT1*);Ta=-20°Cto70°C;Type4X(IP67)

ATEX (FM12ATEX0034X): ApprovedII2GExdIIB+H2(T6,T4,orT1*)GbTa=-20°Cto70°C;IP67II2DExtbIIIC(T85°C,T135°C,orT450°C*)DbTa=-20°Cto70°C;IP67

IECEx (IECEx FMG 12.0010X): ApprovedExdIIB+H2(T6,T4,orT1*)GbTa=-20°Cto70°C;IP67ExtbIIIC(T85°CorT135°C*)DbTa=-20°Cto70°C;IP67**

ATEX and IECEx Standards:EN 60079-0: 2009 IEC 60079-0: 2011EN 60079-1: 2007 IEC 60079-1: 2007EN 60079-31: 2009 IEC 60079-31: 2008EN60529:1991,+A1:2000 IEC60529:2001

*Temperature code ratings for Zones are dependent on external process temperature factors and equipment enclosureconfiguration.Seethetableaboveforspecifictemperaturecoderatings.**The IECEx dust rating does not apply to the Remote Enclosure.

Note:TheEUPressureEquipmentDirective(PED)requiresthattheminimumambientandfluidtemperatureratingforcarbonsteelflowbodiesnotbebelow-29C. 107

Appendices: Agency Approvals

Model Code Temperature Code Marking (Gas) Temperature Code Marking (Dust)

Enclosure Sensor Type Main Enclosure Remote Enclosure

Main Enclosure Remote Enclosure

E1 ST T4 N/A 135°C N/A

E2 ST T4 N/A 135°C N/A

E3 ST T6 T4 85°C 135°C

E4 ST T6 T4 85°C 135°C

E3 HT T6 T1 85°C 450°C

E4 HT T6 T1 85°C 450°C

Page 108: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

SWITCH TO CRV 1? NO YES

PASSWORD:_ UP DN NXT OK

SWITCH TO CRV 1

108

Appendices

ScopeThis section describes added features to the standard MTI10/MTL10 flowmeter when using the 2 gas curves firmware option.

MTI10/MTL10 2 Gas CurvesThe 2 Gas Curves firmware allows the use of two different calibration tables when running with different gases. One of two methods can be used to switch between the two calibration curves:

1) Use of Contact Input:Whenthecontactinputisprogrammedforcurveswitching,anopencontact will select curve #1 and a contact closure will select curve #2.

2) Use of the Keypad:Ifthecontactinputisnotprogrammedforcurveswitching,pressingF2 and F3 simultaneously will prompt an operator to manually switch curve upon entering a password and confirming the action by pressing the appropriate key.

PressingF2&F3simultaneously:

Password needs to be entered if active (default: 1234):

Afterenteringavalidpassword,abriefconfirmationmessagewillbedisplayed for 1 second:

MTI10/MTL10 2 Gas Curves

Appendices: MTI10/MTL10 with 2 Gas Curves

Page 109: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

INP=SWITCH CRV NXT OK

109

MTI10/MTL10 2 Gas Curves

3) Programming Contact Input for Curve Switching:Enter the menu using steps outlined in "Discrete Input Settings" section (p. 34) and select "Switch CRV". Please note that the flowmeter needs to be programmed for 2 gas curves at the Spirax Sarco factory before you can select this function. Flowmeters are shipped with pre-programmed user requested settings.

Selections are: "Not used" "Tot Reset" "Switch Crv"

Helpful Hint:Fromnormaldisplaymode,pressF4toviewthecurrentgascurveselection.

Appendices: MTI10/MTL10 with 2 Gas Curves

Page 110: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

DNS1 = 0.9876 CHG OK

DNS2 = 1.2876 CHG OK

20mA = 500SCFM CHG OK

4mA = 0 CHG OK

110

Appendices: MTI10/MTL10 with 2 Gas Curves

4) Programming Densities for Curve 1 and Curve 2:Whentheselectedflowunitismass/time,twodifferentdensitieswillbeused for each curve if the meter has been programmed for 2 gas curves. To change the densities:

Go to the unit menu following "Unit Settings" section.

DNS1 is the density associated with curve 1. Change it as needed and press OK.

DNS2 is the density associated with curve 1. Change it as needed and press OK.

5) Programming 4 to 20mA settings for Curve 1 and Curve 2:Whenthemeterhasbeenprogrammedfor2gascurves,2setsof4to20mA settings for flow rate are used. To program these settings:

Go to the 4 to 20mA setting following the "Analog 4 to 20mA Settings" section.

20mA is the upper limit associated with curve 1. Change it as needed and press OK.

4mA is the lower limit associated with curve 1. Change it as needed and press OK.

MTI10/MTL10 2 Gas Curves

Page 111: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

20 maCv2=450 SCFM CHG OK

4 maCv2=0 CHG OK

111

Appendices: MTI10/MTL10 with 2 Gas Curves

20 maCv2 is the upper limit associated with curve 2. Change it as needed and press OK.

4 maCv2 is the lower limit associated with curve 2. Change it as needed and press OK.

6) Operation:• Toavoidconfusion,onlyoneoftwotechniquesisenabled.Ifthe

contactinputisassignedtoswitchgascurves,thentheabilityto switch using the F2 and F3 function keys on the front panel is disabled.

• Twototalizers(Total1andTotal2)andtwoelapsedtimecountersareavailable on the display and through the USB serial communication. Theresetfunctionwillresetalltotalizersandelapsedtimecountertozero.

• Intheeventofapowerfailure,thesoftwarewillrememberthelastcurveinuse.Uponpoweringupagain,theMTI10/MTL10unitwillcontinue to use that curve.

• Switching between gas curves will require a password unless the passwordissetto"0",whichdisablesit.

• The calibration certificates for order with 2 gas curves will identify which gas is Gas 1 and Gas 2.

• Whenmeasuringinmassunits,adensityvaluemustbeenteredforeach gas curve.

MTI10/MTL10 2 Gas Curves

Page 112: Thermal Mass Flowmeter and Temperature Transmitter MTI10

4.6(11.7)

3.9(9.9)

8.1(20.6)

2X 3/4” NPT, FEMALE

“LL”

10.3±0.3(26.2±0.8)

6.5(16.5)

0.87(2.2)

112

AppendicesSection 1

Appendices: Dimensions

Local withRetractor

Fig. 7.1 Local Insertion Meter with Retractor Dimensions

Table 7.1 Local Insertion Meter with Retractor

ProbeSize ProbeSize Dimension“LL”±.01Model Code mm (inches) mm (inches)

375R 375 (15) 15 (375)450R 450 (18) 18 (450)600R 600 (24) 24 (600)750R 750 (30) 30 (750)900R 900 (36) 36 (900)

Page 113: Thermal Mass Flowmeter and Temperature Transmitter MTI10

2X 3/4” NPT,FEMALE

0.87(2.2)

10.3±0.3(26.2±0.8)

ø3.6(ø9.1)

4.4(11.2)

2.0(5.1)

5.9(15.0)

“LL”

5.2(13.2)

3.9(9.9)

CONDUIT, 3/4”, METAL

REMOTE CABLE, 5 CONDUCTOR,SHIELDED, 100ft (30.48m) MAX8.1

(20.6) 4.6(11.7)

113

AppendicesSection 1

Appendices: Dimensions

Remote withRetractor

Fig. 7.2 Remote Insertion Meter with Retractor Dimensions

Table 7.2 Remote Insertion Meter with Retractor

ProbeSize ProbeSize Dimension“LL”±.01

Model Code mm (inches) mm (inches)375R 375 (15) 15 (375)450R 450 (18) 18 (450)600R 600 (24) 24 (600)750R 750 (30) 30 (750)900R 900 (36) 36 (900)

ProbeSize ProbeSize Dimension“LL”±.01Model Code mm (inches) mm (inches)

375R 375 (15) 15 (375)450R 450 (18) 18 (450)600R 600 (24) 24 (600)750R 750 (30) 30 (750)900R 900 (36) 36 (900)

Page 114: Thermal Mass Flowmeter and Temperature Transmitter MTI10

“LL”

“HH”

2x 3/4 inch NPT Female

4.6(11.7)

5.2(13.2)

F1 F2 F3 F4

8.1(20.6)

4.3(10.9)

FLOW LC

SignalWiring

InputPower

Appendices: Dimensions

114

Local Insertion Meter

Fig. 7.3 Insertion Meter Dimensions

Table 7.3 Insertion Meter with 316 stainless steel probe

ProbeSize ProbeSize Dimension “LL”±.01

Dimension “HH”±.01

Model Code mm (inches) mm (inches) mm (inches)100I 150 (6) 150 (6) 315 (12.5)225I 225 (9) 225 (9) 390 (15.5)300I 300 (12) 300 (12) 470 (18.5)375I 375 (15) 375 (15) 550 (21.5)450I 450 (18) 450 (18) 620 (24.5)600I 600 (24) 600 (24) 770 (30.5)750I 750 (30) 750 (30) 930 (36.5)900I 900 (36) 900 (36) 1080 (42.5)

Page 115: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

FLOW LC

3.6(9.1)

4.4(11.2)

2.0(5.1)

4.5(11.4)

“HH”

Remote Cable, 5 Conductor,Shielded, 100ft (30.48m) Max.

“LL”

5.2(13.2)

4.6(11.7)

InputPower

2x 3/4 inch NPT Female

ProbeSize ProbeSize Dimension “LL”±.01

Dimension “HH”±.01

Model Code mm (inches) mm (inches) mm (inches)100l 150 (6) 150 (6) 300 (11.9)225l 225 (9) 225 (9) 380 (14.9)300I 300 (12) 300 (12) 455 (17.9)375I 375 (15) 375 (15) 530 (20.9)450I 450 (18) 450 (18) 600 (23.9)600I 600 (24) 600 (24) 760 (29.9)750I 750 (30) 750 (30) 900 (35.9)900I 900 (36) 900 (36) 1060 (41.9)

Appendices: Dimensions

115

Remote Insertion Meter

Fig 7.4: Insertion Remote Meter Dimensions

Table 7.4 Insertion Remote Meter with 316 stainless steel probe

Page 116: Thermal Mass Flowmeter and Temperature Transmitter MTI10

2x 3/4 inch NPT Female

“H”

“L”

LC

2x NPT Male Thread

4.6(11.7)

5.2(13.2)

F1 F2 F3 F4

8.1(20.6)

4.3(10.9)

SignalWiring

InputPower

116

Local Inline NPT Meter

Appendices: Dimensions

Table7.5InlineMeterwith316stainlesssteelflowbodyandNPTEndConnections

Remote Inline NPTFig. 7.5: Inline Meter with 316 Stainless Steel Flow Body and NPT End Connections Dimensions

*AlsoavailableinA106GradeBcarbonsteelpipe(20PC,25PC,30PC,and40PCmodelcodes)

BodySize BodySize Dimension“L” Dimension“H”Model Code mm (inches) mm (inches) mm (inches)

8 8 (0.25) 145 (5.8) 265 (10.5)15 15 (0.50) 300 (12) 265 (10.5)20 20 (0.75) 300 (12) 265 (10.5)25 25 (1.00) 300 (12) 265 (10.5)32 32 (1.25) 300 (12) 265 (10.5)40 40 (1.50) 300 (12) 265 (10.5)50 50 (2.00) 300 (12) 265 (10.5)65 65 (2.25) 450 (18) 270 (10.6)80 80 (3.00) 450 (18) 270 (10.6)

100 100 (4.00) 450 (18) 280 (11.1)

Page 117: Thermal Mass Flowmeter and Temperature Transmitter MTI10

3.6(9.1)

4.4(11.2)

2.0(5.1)

2.0(5.1)

“HH”

Remote Cable, 5 Conductor,Shielded, 100FT Max.

“L”

LC

2x NPT Male Thread

4.6(11.7)

5.2(13.2)

F1 F2 F3 F4

InputPower

117

Appendices: Dimensions

Remote Inline NPT

Table7.6InlineRemoteMeterwith316stainlesssteelflowbodyandNPTEndConnections

Fig. 7.6: Inline Remote Meter with 316 Stainless Steel Flow Body and NPT End Connections Dimensions

*AlsoavailableinA106GradeBcarbonsteelpipe(20PC,25PC,30PC,and40PCmodelcodes)

BodySize BodySize Dimension“L” Dimension“HH”Model Code mm (inches) mm (inches) mm (inches)

15 15 (0.50) 300 (12) 265 (10.5)20 20 (0.75) 300 (12) 265 (10.5)25 25 (1.00) 300 (12) 265 (10.5)32 32 (1.25) 300 (12) 265 (10.5)40 40 (1.50) 300 (12) 265 (10.5)50 50 (2.00) 300 (12) 265 (10.5)65 65 (2.25) 450 (18) 270 (10.6)80 80 (3.00) 450 (18) 270 (10.6)

100 100 (4.00) 450 (18) 280 (11.1)

Page 118: Thermal Mass Flowmeter and Temperature Transmitter MTI10

2x 3/4 inch NPT Female

“H”

“L”

LC

4.6(11.7)

5.2(13.2)

2X Flange, Raised Face, ANSI B16.5, 316 SST

8.1(20.6)

4.3(10.9)

F1 F2 F3 F4

118

Local Inline Flange Meter

Appendices: Dimensions

Table7.7InlineMeterwith316stainlesssteelflowbodyand150#RFFlangeEndConnections

Remote InlineFlange Meter

Fig. 7.7: Inline Meter with 316 Stainless Steel Flow Body and 150# RF Flange End Connections Dimensions

*AlsoavailableinA106GradeBcarbonsteelpipewithA105flanges(20FC,25FC,30FC,and40FCmodelcodes)

BodySize BodySize Dimension“L” Dimension“H”Model Code mm (inches) mm (inches) mm (inches)

15 15 (0.50) 300 (12) 265 (10.5)20 20 (0.75) 300 (12) 265 (10.5)25 25 (1.00) 300 (12) 265 (10.5)32 32 (1.25) 300 (12) 265 (10.5)40 40 (1.50) 300 (12) 265 (10.5)50 50 (2.00) 300 (12) 265 (10.5)65 65 (2.25) 450 (18) 270 (10.6)80 80 (3.00) 450 (18) 270 (10.6)

100 100 (4.00) 450 (18) 280 (11.1)150 150 (6.00) 600 (24) 310 (12.2)

Page 119: Thermal Mass Flowmeter and Temperature Transmitter MTI10

2X 3/4 " NPT,Female

3.6(9.1)

4.4(11.2)

2.0(5.1)

4.5(11.4)

“HH”

“L”

CL

Remote Cable, 5 Conductor,Shielded, 100FT Max.

4.6(11.7)

F1 F2 F3 F45.2(13.2)

119

Appendices: Dimensions

Remote InlineFlange Meter

Table7.8InlineRemoteMeterwithstainlesssteelflowbodyand150#RFFlangeEndConnections

Fig. 7.8: Inline Remote Meter with Stainless Steel Flow Body and 150# RF Flange End Connections Dimensions

*AlsoavailableinA106GradeBcarbonsteelpipewithA105flanges(20FC,25FC,30FC,and40FCmodel codes)

BodySize BodySize Dimension“L” Dimension“HH”Model Code mm (inches) mm (inches) mm (inches)

15 15 (0.50) 300 (12) 265 (10.5)20 20 (0.75) 300 (12) 265 (10.5)25 25 (1.00) 300 (12) 265 (10.5)32 32 (1.25) 300 (12) 265 (10.5)40 40 (1.50) 300 (12) 265 (10.5)50 50 (2.00) 300 (12) 265 (10.5)65 60 (2.25) 450 (18) 270 (10.6)80 80 (3.00) 450 (18) 270 (10.6)

100 100 (4.00) 450 (18) 280 (11.1)150 150 (6.00) 600 (24) 310 (12.2)

Page 120: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1 F2 F3 F4

F1 F2 F3 F4

F1 F2 F3 F4

: Installation Variations (moisture)

120

AppendicesSection 1Tilt Installations Tilt Installation

at45°,CWTilt InstallationsThese variations on installations help prevent moisture and condensation from forming on the sensor and disrupting accurate flow measurement. SpiraxSarcorecommends180°installation,ifpossible.

Tilt Installation at 180°

Tilt Installation at90°,CCW

Tilt Installation at90°,CW

Page 121: Thermal Mass Flowmeter and Temperature Transmitter MTI10

F1F2

F3

F4

: Installation Variations (limited space)

121

AppendicesSection 1Tilt Installation at45°,CW

Tilt Installation at 45°Whenrestrictedphysicalinstallationspaceexists,theMTI10/MTL10can also be installed at a 45° angle. Please note that the display's orientation will remain aligned with the top of the meter.

Note: Displays are rotatable only in 90° angle increments.

Formoreinformationaboutdisplayconfigurations,visit www.spiraxsarco.com/us/products-services/products/flowmeters.asp to view other display configurations.

Page 122: Thermal Mass Flowmeter and Temperature Transmitter MTI10

WarrantyLIMITED WARRANTY AND REMEDIES. Spirax Sarco warrants to Purchaser that the products of Spirax Sarco's own manufacture supplied hereunderwill,foraperiodof12monthsfromthedateofshipment,unlessotherwisespecified,toPurchaser,befreefromdefectsinmaterialand workmanship under normal and proper operating conditions and service. The obligation of Spirax Sarco and Purchaser's sole and exclusiveremedypursuanttothiswarrantyshallbe,atSpiraxSarco'soption,torepairorreplaceanyproductorpartthereofwhichisreturnedtoSpiraxSarco,Inc.,2150MillerDriveSuiteH,Longmont,CO80501,with transportation charges prepaid that is determined by Spirax Sarco tobedefective.Notwithstandingtheforegoing,SpiraxSarcoshallhaveno obligation hereunder if all payments due from Purchaser have not beenmade,ortheproductorpartbecomesdefectiveinwholeorinpartastheresultofrepairsnotmadebySpiraxSarcoorPurchaser,orastheresultofremoval,improperuse,operationaboveratedcapacities,ormisapplication thereof after it has been delivered to the Purchaser.

Products,partscomponentsandaccessories,irrespectiveorattachmentorassembly,madebyothermanufacturer'sarewarrantedonlytotheextent of the original manufacturer's warranty to Spirax Sarco. If the productsoldisdescribedas"used",itissold"asis,whereis"withoutany guarantee or warranty whatsoever. EXCEPT AS SET FORT H HEREIN,SPIRAXSARCOMAKESNOOTHERWARRANTIES,EXPRESS,IMPLIEDORSTATUTORY,INCLUDINGWARRANTIESOFMERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. PURCHASER ACKNOWLEDGES THAT THIS IS NOT RELYIN G UPON SPIRAX SARCO’S SKILL OR JUDGMENT TO SELECT OR FURNISH GOODS SUITABLE FOR ANY PARTI CULAR PURPOSE OR UPON ANY AFFIRMATIONS OR FACT OR PROMISES OF SPIRAX SARCO WHICH EXTEND BEYOND SPECIFICATIONS MUTUALLY AGREED UPON IN WRITING BY SPIRAX SARCO AND PURCHASER. IN NO EVENT SHALL SPIRAX SARCO BE LIABLE TO PURCHASER FOR DAMAGES beyond the cost of repair or replacement of defective products,punitivedamages,ORFORSPECIAL,INDIRECT,ORCONSEQUENTIALDAMAGES,INCLUDINGBUTNOTLIMITEDTO,INTERRUPTIONOFOPERATIONS,LOSSOFANTICIPATEDPROFITS,ORDAMAGETOPURCHASER’SBUSINESSREPUTATION.

122

Appendices: Warranty

Warranty Warranty

Page 123: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Warranty Claims and Return Procedure. Products returned to Spirax Sarco by Purchaser for a credit or under a warranty claim will not be accepted for exchange or credit without Spirax Sarco’s prior written authorizationinaccordancewiththeReturnofProduct/MaterialProcedure in effect at the time of the claim.

Out of Warranty ConditionsThis policy has been established to cover repair or replacement of equipment sold by seller which as exceed the warranty expiration date. Seller agrees to repair or replace such equipment at Buyer’s expense under the following terms and conditions:

a) All equipment must be returned to Seller’s plant at Buyer’s expense and risk.b) Invoices on equipment which is repaired and returned to Buyer will bebasedonthedirectandindirectlabor,materialsandoverheadcostsassociatedwithrepair,tests,inspections,etc.c) The warranty set forth in Seller’s warranty policy for its equipment appliestonewequipmentonly.Nowarranty,expressorimplied,appliesto equipment repaired and returned under provisions of this policy.d)BuyerwillbenotifiedimmediatelyifSellerdeterminestherepaircost per item will exceed 50% of the current selling price of a new replacement. Buyer’s advice to repair the defective item must be received before any further action is taken.e)Allexporttaxes,feesanddutiesconnectedwithout-of-warrantyshipments outside the continental United States will be the responsibilityof the Buyer.f) A minimum charge of $150 will be charged to Buyer for all out-of-warranty repairs.

123

Appendices: Warranty

Warranty

Page 124: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Returning Your MeterTheSpiraxSarcoApplicationsGroup(phone800-356-9362,fax303-682-7069) can help you through the process of returning a meter for service.

IfitbecomesnecessarytoreturnaSpiraxSarcoflowmeterforserviceorrecalibration,pleasefollowthesesteps:

1. Pleasehaveyourmeter’sserialnumber(s)readysothatwecanfindthe records for your meter(s) quickly.

2. Read the Spirax Sarco RMA Request Form carefully for detailed instructions on the RMA process.

3. Fill out the Spirax Sarco RMA Customer Information Form.4. ObtainaReturnMaterialAuthorization(RMA)Numberfromthe

Spirax Sarco Applications Group. 5. Unlessspecificallyinstructedtodootherwise,theentireflowmeter

mustbereturned,includingallelectronics.(Forremoteunitsorflowbodies,ALLserialnumbersmustmatchtheircorrespondingmeters.)

6. Clean and decontaminate all wetted parts before returning to Spirax Sarco.

What to expect while your meter is being servicedDependingonthetypeofmeter,therearevaryingturnovertimesforservicing a meter. The average time needed to service the meter is 10 - 14 days (not including shipping or peak production times).

If you have already shipped your meter to Spirax Sarco for servicing andwouldliketocheckthestatusofyourmeter,pleasecontacttheApplications Department.

Rush recalibration service is available for a fee. Restrictions apply.

124

DisclaimerSection 1Appendices: Returning Your Meter

What to Expect During Servicing

Returning Your Meter

Page 125: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Caution - (refer to accompanying documents): Please follow the specified instructions and general safety practices.

Indicates compliance with the WEEE Directive. Please dispose of the product in accordance with local regulations and conventions.

Indicates compliance with the applicable European Union Directives for Safety LVD (Low Voltage Directive 2006/95/EC), EMC (Electromagnetic Compatibility Directive 2004/108/EC), and PED (Pressure Equipment Directive 97/23/EC).

Enclosure Protection Classification per IEC 60529: Protected against the ingress of dust and Immersion.

IP67

125

DisclaimerSection 1

Page 126: Thermal Mass Flowmeter and Temperature Transmitter MTI10

Spirax Sarco Technical Support Department Toll Free at: 1-800-356-9362

SPIRAXSARCO,INC.•1150NORTHPOINTBLVD.•BLYTHEWOOD,SC29016PHONE 803-714-2000 • Fax 803-714-2222

www.spiraxsarco.com/us