Thermo Scientific MSSA Multi-Stream Solution Analyzer Product Specifications The Thermo Scientific Multi-Stream Solution Analyzer (MSSA) provides continuous online analysis of solution streams and is capable of measuring elements as low as sulphur in the periodic table. This provides cost effective, real-time analysis of key elements and other variables from sampled stream lines in hydrometallurgical operations such as SX/EW and refineries. Features Can simultaneously measure 8 elements •  Can receive between 1 and 16 streams per •  analyzer Energy Dispersive XRF technique of analysis •  Stainless steel frame and construction. •  Stainless steel or polypropylene wetted parts. •  Simple, low maintenance design. •  De-multiplexing of streams available if •  required. Metallurgical shift sampling available. •  Flush capabilities for scaling solutions. •  The Thermo Scientific MSSA provides a continuous window into the hydrometallurgical process which allows operators or an automatic control system to increase plant efficiency, quality and recovery. Specific Benefits Monitoring the transfer of metal species through different stages of the plant, viz pregnant liquor solution, organic and electrolytes. Monitoring contaminant species in copper •    SX / EW in aqueous streams. Automation of the rectifiers in the EW •    plant. Optimization of the electrical energy use •    in the EW plant by minimizing peak power loads. Contributes to increased current efficiency by: Optimization of current density based on •    metal concentration in the electrolyte to improve cathode quality. •    Automation of reagent dosing. •    Man power savings due to the decrease •    in control samples analyzed by the laboratory. Operating Principles The MSSA automatically switches each stream past the analysis head in turn. The MSSA sample handling system includes: Sample inlet manifold for connection of •    the incoming sample lines. Multiplexer Unit consisting of inlet valves, •    header tank, level sensor tube, wash system. Flow cell for presentation of the sample to •    the detector. Demultiplexer unit consisting of an array •    of outlet valves (maximum 8 streams). Samplers for calibration and shift average •    sampling.

Upload: others

Post on 27-Jan-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

  • Thermo Scientific MSSAMulti-Stream Solution Analyzer

    Product SpecificationsThe Thermo Scientific Multi-Stream Solution Analyzer (MSSA) provides continuous online analysis of solution streams and is capable of measuring elements as low as sulphur in the periodic table. This provides cost effective, real-time analysis of key elements and other variables from sampled stream lines in hydrometallurgical operations such as SX/EW and refineries.

    FeaturesCan simultaneously measure 8 elements • Can receive between 1 and 16 streams per • analyzerEnergy Dispersive XRF technique of analysis• Stainless steel frame and construction.• Stainless steel or polypropylene wetted parts. • Simple, low maintenance design.• De-multiplexing of streams available if • required.Metallurgical shift sampling available. • Flush capabilities for scaling solutions.• 

    The Thermo Scientific MSSA provides a continuous window into the hydrometallurgical process which allows operators or an automatic control system to increase plant efficiency, quality and recovery.Specific BenefitsMonitoring the transfer of metal species through different stages of the plant, viz pregnant liquor solution, organic and electrolytes.

    Monitoring contaminant species in copper •   SX / EW in aqueous streams. Automation of the rectifiers in the EW •   plant.Optimization of the electrical energy use •   in the EW plant by minimizing peak power loads.

    Contributes to increased current efficiency by:

    Optimization of current density based on •   metal concentration in the electrolyte to improve cathode quality.•   Automation of reagent dosing. •   Man power savings due to the decrease •   in control samples analyzed by the laboratory.

    Operating PrinciplesThe MSSA automatically switches each stream past the analysis head in turn. The MSSA sample handling system includes:

    Sample inlet manifold for connection of •   the incoming sample lines.Multiplexer Unit consisting of inlet valves, •   header tank, level sensor tube, wash system.Flow cell for presentation of the sample to •   the detector. Demultiplexer unit consisting of an array •   of outlet valves (maximum 8 streams). Samplers for calibration and shift average •   sampling.

  • Specifications

    Analysis PerformanceNumber of Streams 1-16, sampled flow rate nominally 10 liters/min.Element Range Up to 8 elements from sulphur to uranium (atomic number 16 to 92)Analysis Method X-Ray Fluorescence (XRF)X-Ray Source Minimal activity encapsulated radioisotope Detection System Energy dispersive solid-state detectorTypical Accuracy 1 – 2% relative or better above 10 g/L. 2 – 4% relative or better in the range 1 – 10 g/L. 2 – 8% relative or better in the range 1mg – 1g/LMeasurement Time Typically 60-100 seconds per stream, including analysis, wash and changeover. Wash sequence programmed to suit.Inlet and Outlet Valves ½”, Normally closed ball valves with pneumatic actuators. Drain Valve ¾” Normally open ball valve with pneumatic actuator.Operating Temperature 0 – 50°CSolution Temperature 0 - 90°C (PP version). 0 - 110°C (SS version).Utilities RequiredPower Requirements Single Phase AC (110V, 220V, 240V or 255V at 50 or 60Hz)Air Instrument quality air 500 – 800 kPa. Air consumption: approx 200 litres per hour at working pressure. Secondary air filter built in.Water Clean water or plant streams at 200 – 800 kPa pressure. Consumption approximately 200 litres per hour. Selection and need for wash fluids will vary with specific application.Liquid Nitrogen 7 – 10 litres of Liquid Nitrogen/week.DimensionsSize Length Width Height 1600mm 1040mm 1735mm (5’3” x 3’5” x 5’9”)Weight Nominally 500 kilograms (1100 lbs)StandardsElectrical rating IP66Quality Assurance Adelaide Manufacturing facility ISO-9001:2000 certified

    © 2008 Thermo Fisher Scientific Inc. All rights reserved. Results may vary under different operating conditions. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details.

    Process Instruments

    Service and SupportRegular service and maintenance of your sampling and analysis equipment will optimize accuracy and availability, providing you with the assurance that you are not left blind to process upsets. We offer a range of service contracts that can be tailored to fit your needs and budget. Possible inclusions are operator training, calibration and a preventative maintenance program.

    Multi-Stream Solution Analyzer for Slurries