THE SOLUTION FOR ACCURATE GAS FLOW M E A S U R E M E N T ISO 9001:2008 Certified Quality System T HERMO S YNC ® Revolutionary Finned Thermowells Reduces Thermal Transfer from the Pipeline More Accurate than a Standard Thermowell Fast Thermal Response Reduces Unaccounted Error Accepts Standard Probes Mounts in Existing 3/4 NPT Thermowell Port Flow Tested and Rated to 100 FPS in 1000 PSI Natural Gas

Upload: hoangkhuong

Post on 15-Sep-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

T H E S O L U T I O N F O R A C C U R A T E G A S F L O W M E A S U R E M E N T

ISO 9001:2008 Certified Quality System

THERMOSYNC ®

Revolutionary Finned

Thermowells

Reduces Thermal Transfer

from the Pipeline

More Accurate than a Standard

Thermowell

Fast Thermal Response

Reduces Unaccounted Error

Accepts Standard Probes

Mounts in Existing 3/4 NPT

Thermowell Port

Flow Tested and Rated to 100

FPS in 1000 PSI Natural Gas

Page 2: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

INTRODUCTION

Thermowell – Technical Information Reducing Thermowell Conduction Errors in Gas Pipeline Temperature Measurement 2 Temperature Error - Can a few degrees really make a difference? 6 ThermoSync High Velocity/Pressure Flow Tests, CEESI Iowa 7 ThermoSync Test Results: Temp Error vs. Flowrate and Prove Response Time 10

Additional Test Results: ThermoSync vs. Conventional Thermowells and Performance Evaluations can be obtained by calling Parker PGI's Customer Service at 1-800-231-0233 or 1-713-744-9850 (USA). TECHNICAL AND ORDERING INFORMATION ThermoSync Thermowell Dimensions 11 ThermoSync Thermowell Ordering Information 12 ATP-1000 and ATP-1001 Probe/Sensor 13 ATP-1000-R6 and ATP-1001-R6 Dual Probe/Sensor 14 ATP-1000 RTP Probe/Sensor Calibration 15 ATP High Accuracy RTD Probe/Sensor Ordering Information 16 ATP High Accuracy RTD Probe/Sensor Options (photos) 17 ATA-Series Ordering Information 18 ATA-Series Options (photos) 19

USER’S LIST

Domestic End Users and OEM Users List 20

Delrin® and Teflon® are registered trademarks of the E.I. DuPont de Nemours Company.

INDEX

Page 3: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

Abstract. It is well known that conventional thermowells will thermally “couple” with the vessel in which they are mounted resulting in an error in the temperature measurement. Conduction error, commonly called immersion error, is present whenever a temperature gradient exist between the vessel or pipe the thermowell is installed into, and the substance being measured. Sources of conduction error in gas temperature measurement and methods of reducing it, specifically the use of finned thermowells, are discussed.

In the gas pipeline industry, gas temperature, along with pressure and flow, is used to calculate volumetric flow. Any error in the temperature measurement results in an error in the flow calculation. This error directly shifts the bottom line, resulting in accounting for too much gas or not enough. This unaccounted error in the gas volume measurement can have significant cost associated with it. Any time there is a difference between the pipe temperature and the flowing gas temperature there will be a conduction error. Even the most accurate temperature measurement equipment will have errors. Understanding the sources of conduction error and how to minimize it will increase the accuracy of gas temperature measurement.

For the sensor to approach the temperature of the flowing gas there must be a thermal coupling or heat transfer between the two. For a net heat transfer from hot to cold to take place there must be a temperature difference between the sensor and the gas. When there is no heat transfer, the measurement system is in thermal equilibrium and the sensor is as close to the gas temperature as it will get. In reality, the sensor temperature will never be the same as the gas temperature due mainly to conduction errors.

Reducing Thermowell Conduction Measurement Errors in Gas Pipeline Temperature Measurement

INTRODUCTION

To accurately measure the flowing gas temperature, the sensor must be in thermal contact with the gas, but not disturb the gas temperature. Figure 1 shows a cut-away view of a typical thermowell installation in a gas pipeline. A threaded fitting, welded to the top of the pipe, provides the mount for the thermowell. A thermowell fitting connects the thermowell to a protection head or temperature transmitter to allow field wiring to the sensor. The thermowell fitting should include a spring to keep the sensor protection tube pressed firmly into the thermowell providing a good thermal contact between the sensor tip and the bottom of the thermowell. Thermally conductive grease at the sensor tip will also improve performance as any air gap between the sensor tip and the thermowell will add to the measurement error and reduce the response time.

THE MEASUREMENT PROCESS

FIGURE 1. Typical sensor installation including thermowell and protection head for gas temperature measurement in a pipeline.

2 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

REDUCING THERMOWELL ERRORS

Page 4: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

SOURCES OF CONDUCTION ERROR

Heat transfer can be subdivided into three categories, conduction, convection, and radiation. Conduction takes place when there is a temperature gradient in a solid, liquid, or gas. The molecules in the hot area increase the violence of their vibrations as they heat up. Then, as they collide with their slower moving neighbors, some of their energy of motion is shared, and they in turn pass it on from one molecule to the next. The equation for thermal conductivity in one dimension is [1]:

where qcond is the quantity of heat flow per unit time, A is the cross sectional area, T2-T1 is the temperature difference, and L is the length. When a metal rod is heated on one end, heat will be conducted through the metal to heat the other end.

Convection is the transfer of heat from one place to the other by the actual motion or flow of the material. The equation for convection heat transfer, also called Newton’s Law of cooling is [2]:

(2) TThAq sconv

where qconv is the heat power being transferred through the surface, A is the surface area, Ts is the surface temperature, T¥ is the temperature of the fluid distant from the surface, and h is the convection heat transfer coefficient (25 to 250 for forced convection gases [3]). An example of convection heat transfer would be a fan blowing on a hot object. The heat from the hot object is transferred to the cooler flowing air. In gas temperature measurement, the heat from the thermowell sensing-section and sensor is transferred to the flowing gas mainly by convection.

Heat can also be transferred by radiation, which is the continuous emission of energy from the surface of all objects. The equation for radiation heat transfer is [2]:

where qrad is the radiated power, A is the surface area, σ is the Stefan-Boltzmann constant (5.67x10-8Watt/m-2 x K-4), TS is the surface temperature of the object, TSUR is the surface temperature of the ambient surroundings. At the relatively low temperatures seen in natural gas pipelines, typically less than 70°C, the rate of radiation heat transfer is small and is at long wavelengths producing little influence on the measurement accuracy. Radiation heat transfer can be minimize by manufacturing the thermowell of a low emissivity material, such as polished stainless steel with emissivity = 0.17 [4].

44SURSrad TTAq (3)

HEAT TRANSFER

L

TTkAqcond

12 (1)

FIGURE 2. Cutaway pipe section showing (a) sensor in flowing gas without conduction path, (b) conduction path through the sensor wires, and (c) large conduction path through thermowell, protection tube, and sensor wires.

Conduction errors are present whenever there is a difference in temperature between the flowing gas and the pipeline walls. The pipeline running underground is kept at a fairly constant temperature but at the metering station, a section of the pipeline is brought above ground to allow access for the measurement equipment. The metering section of pipe and the equipment installed on the pipe, are heated and cooled by the changing ambient environment. As the temperature difference increases, the conduction error for any installation also increases.

To understand the sources of conduction errors, first consider a temperature sensor inserted into the gas stream as shown in Figure 2a.

3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

REDUCING THERMOWELL ERRORS

Reducing Thermowell Conduction Measurement Errors in Gas Pipeline Temperature Measurement

Page 5: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

The sensor is only influenced by convection and a small influence from radiation heat transfer and will eventually be equal to or very near the gas temperature. There is no conduction error because there is no conduction path to the sensor. In Figure 2b, sensor wires are added, which creates a conduction path between the warm pipe and the sensor. Heat is conducted from the pipe walls through the wires to the sensor, and actually shifts the sensor temperature, resulting in a conduction error.

As long as there is a temperature difference between the pipe and the gas, the sensor will never equal the gas temperature due to the thermal energy conducted through the sensor wires. In Figure 2c, the thermowell, sensor protection tube, and protection head are added to the sensor. This significantly increases the cross sectional area of the conduction path to the sensor. Referring to Equation 1, the increased cross sectional area increases the conduction heat transfer to the sensor, increasing the conduction error. If the pipe was cooler than the gas, the same error exists for the same differential temperature but in the opposite direction. The following relation may be used to determine the extent of the conduction error [5]:

The most common method for minimizing conduction errors is to increase the insertion length. Typically the insertion length of the thermowell should equal a minimum of 10 times the diameter of the thermowell [6]. Increasing the insertion length will reduce conduction errors as shown in Equation 4 but in small diameter pipes this may be impossible. Reducing the thermowell diameter, to reduce the conduction path, is not practical in high pressure, high flow velocity pipelines because the strength of the well will be compromised. Flow rates can be as high as 30 meters per second with pressures up to 7,000 kPa.

Reducing the thermal conductivity of the thermowell material (k in equation 4) will reduce the conducted thermal energy between the pipe walls and the sensor, resulting in a reduced influence to the measurement. The drawback is that it will also reduce the thermal conduction through the sensing section of the thermowell to the sensor, reducing the influence of the coefficient of heat transfer (h in equation 4) and reducing the sensor response time. The thermowell must be manufactured of materials compatible with the application. To minimize radiation heating, the material should have a low emissivity surface. Typically thermowells used in gas pipeline temperature measurement are made of polished stainless steel, a tradeoff between accuracy and strength.

Further examination of equation 4 indicates that if the surface area of the thermowell at the sensor was increased (p in equation 4), without increasing the conduction cross sectional area (a in equation 4) or the thermal conductivity (k in equation 4), conduction errors will be reduced. This can be accomplished by adding an array of fins at the sensing section of the thermowell. Figure 3a shows a conventional thermowell for comparison to a thermowell with the added fins to increase the surface area of the sensing section, Figure 3b. Adding the 13 fins increases the surface area in the sensing section over 7.5 times that of a conventional thermowell.

(4)

)(mLCosh

TTTT

sgwisg

where Tsg is the static temperature of the gas, Ti is the indicated sensor temperature, Tw is the pipe wall temperature, L is the immersion length, m = (hp/ka)1/2 , h is the convection coefficient of heat transfer, p is the surface area of the thermowell (at the sensor), k is the thermal conductivity of the thermowell, and a is the conduction cross-sectional area of the thermowell. The larger mL becomes, the closer the indicated temperature, Ti, approaches the true gas temperature, Tsd. Any means of increasing the mL product will result in a reduced conduction error. The steady state temperature of the sensor is a result of a balance between convection heat transfer from the gas to the thermowell, and conduction and radiation heat transfer between the sensor and its surroundings.

REDUCING CONDUCTION ERROR

4 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

REDUCING THERMOWELL ERRORS

Reducing Thermowell Conduction Measurement Errors in Gas Pipeline Temperature Measurement

Page 6: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

FIGURE 3. (a) Conventional thermowell (b) Finned thermowell

The increased surface area at the sensing section of the finned thermowell significantly increases the convection heat transfer between the gas and the sensor, shifting the sensor temperature closer to the actual gas temperature, reducing the conduction error. Equation 2 shows that an increase in the surface area, A, will increase the heat power transferred, qconv.

FINNED THERMOWELL PERFORMANCE

FIGURE 4. Flow comparison of a conventional thermowell, top curve, and a finned thermowell, bottom curve. [7]

Actual flow testing clearly shows the performance improvements the increased surface area provides. A conventional thermowell and a finned thermowell were run in a 7.5cm inside diameter schedule 80 pipe. The pipe was heated to 38˚C and held at this temperature while various flow rates were observed.

Figure 4 shows the difference between the actual gas temperature and the indicated temperature. The finned thermowell, indicated by the bottom trace, was very close to the actual gas temperature. The conventional thermowell had over 1.6˚C (3˚F) error over the entire test. Slower flow velocities produced larger errors as expected. Sensor response time was four times faster using the finned thermowell [7].

REFERENCES 1. Sears F.W., Zemansky M.W., Young H.D., College

Physics Fifth Edition 1980.

2. Bently R.E.. Temperature and Hummidity Measurement, Volume 1, 1998.

3. Incropera F.P., DeWitt D.P., editors. Fundamentals of Heat and Mass Transfer. John Wiley & Sons, Inc. New York 1990.

4. OS550 Industrial Infrared Thermometer Users Guide, Emissivity Table (Omega Engineering).

5. ASME Temperature Measurement, (American Society of Mechanical Engineers), PTC 19.3 – 1974, reaffirmed 1998, Part 3.

6. Liptak B.G., editor-in-chief, Process Measurement and Analysis, chapter 4.

7. TAN-34CO-L4 Thermowell Performance Evaluation, ThermoSync Temperature Measurement System, PGI International, October 2000.

5 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

REDUCING THERMOWELL ERRORS

Reducing Thermowell Conduction Measurement Errors in Gas Pipeline Temperature Measurement

Page 7: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

CAN A FEW DEGREES MAKE A DIFFERENCE?

Using the flow lab test results, two flow rates were used to compare the conventional stainless steel thermowell to the ThermoSync temperature measurement system. At 60 CFM the temperature error for the conventional system is 7.5°F and at 300 CFM 3.8°F. For the ThermoSync system, 0.2°F at 60 CFM and 0.0°F at 300 CFM. These are conservative errors taken with only a 30°F difference between the pipe temperature and actual gas temperature. In extreme ambient conditions, where the difference is larger, expect larger temperature errors. The two flow rates were converted to SCFH using 200 PSI at 70°F.

60 ACFM = 48 MCF/Hour

300 ACFM = 240 MCF/Hour

Using AGA3 equations, the following parameters were used to determine the effect of the temperature error on a calculated flow volume for an orifice plate flow meter.

Flange Taps Upstream Pressure Base 14PSIA Temp Base 60°F Barometric Pressure 14.7 Static Press. 200 PSIG Pipe I.D. 2.97" Orifice 1.50" Specific Gravity 1.00 CO2 Mole Percent 0.00 N2 Mole Percent 0.00

Flow Rate Thermowell Temp MCF/Hour

MCF Error/Hour

Dollar Error/Hour

Dollar Error/24

Hours

Dollar Error/31

days

60 CFM

Actual 70.0 48.003 0 $ 0.00 $ 0.00 $ 0.00

ThermoSync 70.2 47.991 0.012 $ 0.05 $ 1.14 $ 35.35

Conventional 77.5 47.584 0.419 $ 1.66 $ 39.82 $ 1,234.47

Actual 70.0 240.000 0 $ 0.00 $ 0.00 $ 0.00

ThermoSync 70.0 240.000 0.000 $ 0.00 $ 0.00 $ 0.00

Conventional 73.8 238.931 1.069 $ 4.23 $ 101.60 $ 3,149.53

300 CFM

Dollar error per hour for the two temperature measurement methods (based on $3.96 per MCF)

Even under these conservative conditions, a conventional thermowell may be costing you $101.60 per day...or $37,084 per year!

$4.23 x 24 hours per day = $101.60

$101.60 x 365 days per year = $37,084!

Temperature Induced Error in Turbine or PD Meters

Flow Rate Thermowell Temp MCF/Hour

MCF Error/Hour

Dollar Error/Hour

Dollar Error/24

Hours

Dollar Error/31

days

60 CFM

Actual 70.0 48.144 0 $ 0.00 $ 0.00 $ 0.00

ThermoSync 70.2 48.124 0.020 $ 0.08 $ 1.90 $ 58.92

Conventional 77.5 47.404 0.740 $ 2.93 $ 70.33 $ 2,180.22

Actual 70.0 239.981 0 $ 0.00 $ 0.00 $ 0.00

ThermoSync 70.0 239.981 0.000 $ 0.00 $ 0.00 $ 0.00

Conventional 73.8 238.095 1.886 $ 7.47 $ 179.25 $ 5,556.61

300 CFM

To determine the measurement error in ultrasonic, turbine and PD meters, the SCFM is calculated using the actual gas temperature. This rate is compared to the SCFM calculated using the measured temperatures from the ThermoSync system and the conventional system.

Even under these conservative conditions, a conventional thermowell may be costing you $179.25 per day...or $65,426.25 per year!

$7.47 x 24 hours per day = $179.25

$179.25 x 365 days per year = $65,426.25

6 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

Page 8: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

THERMOSYNC FLOW STABILITY TEST REPORT

High Velocity—High Pressure Flow Test, CEESI Iowa

TAN-34CO-L2, TAN-34CO-L4, TAN-34CO-L8, TAN-34CO-L12

January 15, 2001

Design equations for the strength of a thermowell, natural frequency, and the wake or Strouhal frequency of thermowells have been available for some time, and are well- documented (ASME PTC 19.3 Temperature Measurement). These equations however, are developed around a conventional thermowell profile and do not apply to the finned ThermoSync design. To determine the application limitations for the ThermoSync thermowells, testing at the maximum operating conditions must be done. It was determined that the maximum pressure normally seen in a natural gas pipeline is 1000 PSI. Similarly, the maximum flow velocity was determined to be 100 feet per second. Working with CEESI Iowa natural gas flow measurement facility in Garner Iowa, we were able to define the test requirements.

Background

The Colorado Engineering Experiment Station, Inc. (CEESI) is an independent commercial calibration and flow research facility. CEESI also participates in flow measurement standards committees organized by the ASME, AGA, API, and GRI. Many years of experience in meter proving have gained CEESI the international reputation as a leader in flow measurement. The CEESI Iowa high flow test facility has the unique ability to flow natural gas up to 20,000 ACFM at 1050 PSI in a controlled test environment.

CEESI

To simplify the test, a spool was designed to allow six thermowells to be installed and tested together. The spool is a 10” schedule 40 pipe with ANSI 600 lb. flanges as shown in figure 1.

Test Set-Up

FIGURE 1. Test spool design.

FIGURE 2. Test spool and thermowells installed in CEESI flow test pipe.

7 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

Page 9: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

The thermowells in locations C, D, E, and F were fitted with Entran miniature EGA accelerometers to measure vibration. Each accelerometer shared the same amplifier/power supply with a gain of 41.6 making the output of each sensor 50mv/g (1.2mv/g x 41.6 = 50mv/g).

Accelerometer Specifications

Model EGA-250-/R Range +/-250g Limit +/-1250g Temp. Range -40 to 250°F Non-Linearity +/-1% Output Approx. 1.2mv/g

The conventional thermowell in location A (with a conventional RTD probe) and the ThermoSync thermowell in location B (with the ATP-1000) were used for a temperature accuracy comparison. Both probes were connected to a Rosemount 3244MV transmitter. The transmitter was setup to provide an output proportional to the difference of the two temperatures. If both sensors are at the same temperature, no thermowell error, the temperature output will be 12ma (1/2 scale). Any temperature error will shift the transmitter’s output. While running the test, the pipe temperature was a fairly constant 60°F and the gas temperature 69.23°F (measured by CEESI). With a 9°F difference between pipe wall and gas, the measured temperature difference was between 0.1°F and 0.2°F at the 50ft/sec. flow rate.

Flow velocity = 100 feet per second Static Pressure .............................................................................................. 1080 PSI TAN-34CO-L2 ...................................................................................................... 8.8g TAN-34CO-L4 ...................................................................................................... 9.2g TAN-34CO-L8 ..................................................................................................... 10.4g TAN-34C0-L12 .................................................................................................... 11.2g

All of the thermowells produced only low level background vibration at 100 ft/sec. Flow Velocity = 112 feet per second Static Pressure .............................................................................................. 1080 PSI TAN-34CO-L2 ....................................................................................................... 22g TAN-34CO-L4 ....................................................................................................... 20g TAN-34CO-L8 ...................................................................................................... 112g TAN-34C0-L12 ................................................................................................. > 250g

Increasing the flow velocity from 100 to 112 feet per second resulted in L12 breaking into oscillation at 540 Hz. Most of the vibration in the other wells was produced by L12 interference. This is clearly shown by the data taken with L12 removed. When the wake or Strouhal frequency of a well approaches the natural frequency, the thermowell breaks into oscillations. An audible tone at the vibration frequency clearly indicated when this happened. The two waveforms below show the L12 thermowell at 100 ft/sec and 112 ft/sec.

TEST RESULTS

8 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

THERMOSYNC FLOW STABILITY TEST REPORT

Page 10: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

Flow Velocity = 112 feet per second, L12 Removed Static Pressure .............................................................................................. 1080 PSI TAN-34CO-L2 ......................................................................................................... 8g TAN-34CO-L4 .................................................................................................... 11.2g TAN-34CO-L8 .................................................................................................... 14.4g

L12 was removed from the spool to allow the other probes to be measured at higher flow velocities. Removing L12 reduced the vibration levels measured on the other wells. Flow Velocity = 135 feet per second, L12 Removed Static Pressure .............................................................................................. 1080 PSI TAN-34CO-L2 .................................................................................................... 10.4g TAN-34CO-L4 .................................................................................................... 17.6g TAN-34CO-L8 .................................................................................................... 25.6g Increasing the flow velocity slightly increased measured vibration levels.

Flow Velocity = 150 feet per second, L12 Removed Static Pressure .............................................................................................. 1080 PSI TAN-34CO-L2 .................................................................................................... 15.2g TAN-34CO-L4 .................................................................................................... 31.2g TAN-34CO-L8 ....................................................................................................... 28g

At 150 feet per second, there is no excessive vibration. The L12 thermowell is unacceptable for use above 100 feet per second. The L2, L4 and L8 thermowells performed without excessive vibration up to the 150 ft/sec and should provide long-term performance in flow velocities up to 100 feet per second. The L12 should be tested again without the other probes installed in the spool to determine if there is interference from the other thermowells. Notice in Figure 1 that the other probes may interfere with the L12 flow profile. The measured temperature error between the ThermoSync L4 system and a conventional thermowell / probe (3” insertion length) at 50 feet per second was approximately 0.2°F. Calculating the volume using 69.23°F and 69.03°F, there could be a 5.7 MCF per hour error in the calculated flow volume due to pipe wall thermal coupling in the thermowell. At a gas price of $9.00 per MCF, that’s a $51 per hour, $1,231 per day or $449,388 per year unaccountable error. Using the ThermoSync system can significantly reduce pipe wall coupling errors.

SUMMARY

9 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

THERMOSYNC FLOW STABILITY TEST REPORT

Page 11: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

Conventional Thermowellwith Stainless Steel Probe

TAN-34C0-L4 Thermowellwith ATP-1000 Probe

TEMPERATURE ERROR vs. FLOW RATE 3” schedule 80 pipe at 100°F with a nominal supply of air temp of 70°F at 0 PSIG

The zero error readings at zero CFM show the test pipe and temperature sensors have stabilized at 100° F. As the flow rate increases the temperature of the air at the air t empera ture sensor quickly drops to the nominal air temp of 70° F showing a large error. With continued increase in flow rate, the error drops as the cooling effect of the flowing gas overcomes the heating influence of the pipe. The error continues dropping with increasing flow rate, but levels out substantially beyond 200 CFM.

At time zero with the temperature sensors and the test pipe stabilized at 100° F there is no error. The flow rate was suddenly changed to 100 CFM causing a sudden maximum error. The flow rate is held at 100 CFM through the duration of the test. The response time for the ThermoSync thermowell is about 120 seconds. For the conventional thermowell, response time is about 480 seconds. The probe type had no effect on response time of either thermowell.

PROBE RESPONSE TIME 3” schedule 80 pipe at 100°F with a nominal supply of air temp of 70°F at 0 PSIG At time zero, test pipe and temperature sensors have stabilized at 100°F. At time zero, flow is switched from 0 to 100 CFM.

Response Time 3-inch Schedule 80 Pipe at 100°F with a 70°F

nominal supply air at 0 PSIG—100CFM

Err

or,

Deg

rees

Fah

ren

hei

t

Flow, Cubic Feet / Minute

Seconds

Err

or,

Deg

rees

Fah

ren

hei

t

10 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

THERMOSYNC TEST RESULTS

Page 12: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

Process Connection

PART NUMBER A B C D E F G H I J K L M Stem

TAN-12C0-L2 1/2” FNPT 1.25” 1/2” MNPT 1.69” 0.633” .495” 2.22” 1.20” .260” .37” .645” 3.88” .90” Stepped

TAN-12C0-L4 1/2” FNPT 1.25” 1/2” MNPT 1.69” 0.633” .495” 2.96” 1.20” .260” .37” .645” 4.75” .90” Stepped

TAN-12C0-L8 1/2” FNPT 1.25” 1/2” MNPT 1.69” 0.633” .495” 4.59” 1.20” .260” .37” .645” 6.37” .90” Stepped

TAN-12C0-L12 1/2” FNPT 1.25” 1/2” MNPT 1.69” 0.633” N/A 6.66” 1.20” .260” .37” .645” 8.45” .90” Straight

TAN-12C0-L24 1/2” FNPT 1.25” 1/2” MNPT 1.69” 0.633” N/A 9.89” 1.20” .260” .37” .645” 11.67” .90” Straight

TAN-34C0-L2 1/2” FNPT 1.25” 3/4” MNPT 1.69” 0.808” .495” 2.22” 1.20” .260” .37” .85” 3.82” .90” Stepped

TAN-34C0-L4 1/2” FNPT 1.25” 3/4” MNPT 1.69” 0.808” .495” 2.96” 1.20” .260” .37” .85” 4.56” .90” Stepped

TAN-34C0-L8 1/2” FNPT 1.25” 3/4” MNPT 1.69” 0.808” .495” 4.59” 1.20” .260” .37” .85” 6.20” .90” Straight

TAN-34C0-L12 1/2” FNPT 1.25” 3/4” MNPT 1.69” 0.808” N/A 6.66” 1.20” .260” .37” .85” 8.26” .90” Straight

TAN-34C0-L24 1/2” FNPT 1.25” 3/4” MNPT 1.69” 0.808” N/A 9.89” 1.20” .260” .37” .85” 11.49’ .90” Straight

TAN-10C0-L4 1/2” FNPT 1.375” 1” MNPT 1.69” 0.808” .495” 2.96” 1.20” .260” .37” .85” 4.75” .90” Stepped

TAN-10C0-L8 1/2” FNPT 1.375” 1” MNPT 1.69” 0.808” .495” 4.59” 1.20” .260” .37” .85” 6.37” .90” Stepped

TAN-10C0-L12 1/2” FNPT 1.375” 1” MNPT 1.69” 0.808” N/A 6.66” 1.20” .260” .37” .85” 8.45” .90” Straight

TAN-10C0-L24 1/2” FNPT 1.375” 1” MNPT 1.69” 0.808” N/A 9.89” 1.20” .260” .37” .85” 11.67” .90” Straight

All Thermowells: Material: Press/Temp: Flow:

316L SS 4900 PSI Max @ 330° F 100 FPS (L2, L4, L8, L12) or 50 FPS (L24) max in 1000 PSI Natural Gas

NOTE: Use a thermal coupling paste or fluid to couple the probe to the well ONLY in the lower .5 inches of the well. DO NOT fill the well with thermal coupling fluid. Spring load the probe to contact the bottom of the well.

CRN: OH7489.5123467890YTN

11 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

THERMOSYNC DIMENSIONS

Page 13: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

PROCESS CONNECTION SIZE MATERIAL FITS PIPELINE SIZE OPTIONS

TAN - XX CO - - XXXXXXX XXX

PROCESS CONNECTION SIZE

CODE DESCRIPTION

12 1/2” MNPT Process Connection

34 3/4” MNPT Process Connection

10 1” MNPT Process Connection (not available for the 2” Pipeline Size)

MATERIAL

CODE DESCRIPTION

CO 316L Stainless Steel

FITS PIPELINE SIZE

CODE DESCRIPTION

L2 2” Pipeline

L4 3” - 4” Pipeline

L8 6” - 8” Pipeline

L24 16” - 24” Pipeline

L12 10” - 12” Pipeline

OPTIONS

CODE DESCRIPTION

P Cap & Chain Assembly (Can be ordered separately as Spare Part # ATA-1002)

V Vent/Bleed Hole in Thermowell (1/8”)

W1 316 SS Identification Tag

12 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

THERMOWELL ORDERING INFORMATION

Process Connection

300 SS Clip Ring

300 SS Chain

OPTIONS

Aluminum Cap “P” Option

Viton O-Ring

1/8” Vent Hole “V” Option

Page 14: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

The ATP-1000/1001 RTD probes reduce thermal coupling errors from thermowells by using a unique two-piece protection tube. The body of the protection tube is made of insulating PVC plastic, thermally insulating the RTD sensor from the thermowell walls. This helps block the pipe thermal transfer down the protection tube to the sensor. The RTD sensor is mounted in a thermal conductive body for best accuracy and response.

For maximum accuracy, the ATP-1000 and ATP-1001 probes should be used with a TAN ThermoSync thermowell.

SPECIFICATIONS 4-wire platinum wire-wound RTD Element

100 Ohms at 0° C (IEC 751)

PT 100 typical calibration coefficients (ITS-90) A (°C) 3.90830 x 10-3 B (°C) -5.77500 x 10-7 C (°C) -4.18301 x 10-12 Alpha (°C) 3.850 x 10-3 Delta (°C) 1.49990 Beta (°C) 0.10863

Band 5 sensor, interchangeability accuracy ±0.025°C at 0°C, ±0.07°C at 50°C1

Calibration and accuracy certification service available

-40°C (-40°F) to +60°C (+140°F) temperature range

Sensor has high thermal conductivity hard-anodized aluminum tip for maximum thermal transfer between the thermowell and sensor

PVC protection tube thermally insulates the thermowell neck from the sensor

26 AWG Teflon® leads

Probe length user-cut to fit application

Probe body is laser marked with serial number and ratings for calibration traceability

SENSOR LEAD CONNECTIONS The 4-wire probe can be used with 4-wire, 3-wire and 2-wire RTD transmitters. For 3-wire systems, use two red leads and one white lead. Insulate the unused white lead (do not connect the unused white lead). For 2-wire systems, connect both red leads to one terminal and both white leads to the other terminal.

PROTECTION TUBE CUTTING PROCEDURE 1. Determine the probe length required for your

installation. Be sure to leave room for the spring-loaded fitting.

2. Using a small tubing cutter, score the plastic tubing about half-way through. To prevent damage to the sensor leads, never cut through the plastic protection tube with the tubing cutter. Minimum probe length is 2 inches.

3. Remove the tubing cutter and apply bending pressure at the score mark. The protection tube should break at the score mark. If excessive pressure is required, use the tubing cutter to make the score deeper.

PATENTED

13 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

ATP-1000 & ATP-1001 PROBE/SENSOR

Page 15: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

The ATP-1000/1001-R6 RTD Dual RTD probes reduce thermal coupling errors from thermowells by using a unique two-piece protection tube. The body of the protection tube is made of insulating PVC plastic, thermally insulating the RTD sensors from the thermowell walls. This helps block the pipe thermal transfer down the protection tube to the sensors. The RTD sensors are mounted in a thermal conductive body for best accuracy and response.

For maximum accuracy, the ATP-1000-R6 and ATP-1001-R6 Dual RTD probes should be used with a TAN ThermoSync thermowell.

SPECIFICATIONS Wire-wound RTD Elements

100 Ohms at 0° C (IEC 751)

PT 100 typical calibration coefficients (ITS-90) A (°C) 3.90830 x 10-3 B (°C) -5.77500 x 10-7 C (°C) -4.18301 x 10-12 Alpha (°C) 3.850 x 10-3 Delta (°C) 1.49990 Beta (°C) 0.10863

Band 5 sensor, interchangeability accuracy ±0.025°C at 0°C, ±0.07°C at 50°C1

Calibration and accuracy certification service available

-40°C (-40°F) to +60°C (+140°F) temperature range

Sensor has high thermal conductivity hard-anodized aluminum tip for maximum thermal transfer between the thermowell and sensor

PVC protection tube thermally insulates the thermowell neck from the sensor

26 AWG Teflon® leads

Probe length user-cut to fit application

Probe body is laser marked with serial number and ratings for calibration traceability

SENSOR LEAD CONNECTIONS The Dual RTD probe can be used with 3-wire and 2-wire RTD transmitters. For 3-wire systems, use the two red leads and one white lead for the first sensor, and the two orange leads and one yellow lead for the second sensor. For 2-wire systems, connect both red leads to one terminal and the white lead to the other terminal for the first sensor, and connect both orange leads to one terminal and the yellow lead to the other terminal for the second sensor.

PROTECTION TUBE CUTTING PROCEDURE 1. Determine the probe length required for your

installation. Be sure to leave room for the spring-loaded fitting.

2. Using a small tubing cutter, score the plastic tubing about half-way through. To prevent damage to the sensor leads, never cut through the plastic protection tube with the tubing cutter. Minimum probe length is 2 inches.

3. Remove the tubing cutter and apply bending pressure at the score mark. The protection tube should break at the score mark. If excessive pressure is required, use the tubing cutter to make the score deeper.

PATENTED

14 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

ATP-1000-R6 & ATP-1001-R6 PROBE/SENSOR

Page 16: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

Serial # 1634

TEST EQUIPMENT: Model Description Serial Number Calibration Date Calibration Due Hart 1521 6-1/2 Digit Thermometer A13579 5/16/2001 Hart 7102 -5 to 125 C Micro-Bath A13424 2/12/2001 Agilent 34401A 6-1/2 Digit Multimeter US36130957 11/19/2003 11/19/2004

TEST MEASUREMENTS: Temperature °F Temperature °C Resistance 32.957 0.532 100.185 75.043 23.913 109.292 89.984 32.213 112.516 119.958 48.866 118.949

Temperature 22°C / 72°F Date 8/13/2004

CALLENDAR—VAN DUSEN COEFFICIENTS: Coefficient IPRT Typical ** Calibrations Coefficient’s R @°C 100.000 99.969 Alpha 0.00385 0.00385698 Delta 1.4999 1.47929 Beta 0.10863 * A 3.908300 E-03 3.91404 E-3 B -5.77500 E-07 -5.70561 E-7 C -4.18301 E-12 -4.18984 E-12

Calibration accuracy: +/- 0.025°C or +/- 0.05°F

*The beta coefficient is only used for temperatures below °C. The value for the beta coefficient is not solved because there are no samples below 0°C. Use the Typical beta coefficient.

**Coefficient’s IAW ITS90.

15 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

ATP-1000 RTP PROBE/SENSOR CALIBRATION

Page 17: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

PROBE/SENSOR LENGTH OPTIONS

ATP - XXXX - XXXXXXX

PROBE/SENSOR LENGTH

CODE DESCRIPTION

1000 12.60” Probe/Sensor Length

1001 18.60” Probe/Sensor Length

All probe bodies are laser marked with a serial number and ratings of traceability. User cut to length with tubing cutter.

OPTIONS

CODE DESCRIPTION

A Armored Cable (Valid only with LXX Option)

C RTD Calibration (4-Point or 6-Point, depending on RTD Selected)

L12 12 foot Shielded Cable Leads with Weather-Tite Connector Seal

L24 24 foot Shielded Cable Leads with Weather-Tite Connector Seal

L50 50 foot Shielded Cable Leads with Weather-Tite Connector Seal

NXX Non-Standard RTD Length (Valid only with LXX Option; 1” Increments; Minimum Length 2”)

R6 6-Wire Dual RTD Probe

S Spring Loaded (Includes Spring, Locking Collar and 1/2” MNPT Fitting)

W1 316 SS Identification Tag

W1C 316 SS Tag with Calibration data (Valid only with C option)

PATENTED

Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com 16

ATP PROBE/SENSOR ORDERING INFORMATION

Page 18: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

Standard 12’ Shielded Cable

316 SS Identification

Tag

No Cable

Spring-Loaded Only

ATP-1000-L12SW1

ATP-1000-S

17 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

RTD PROBE/SENSOR OPTIONS

Page 19: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

FITS PIPELINE SIZE OPTIONS

ATA - XXXX XXX - XXXXXXX

PROCESS CONNECTION

CODE DESCRIPTION 1000 3/4” MNPT Process Connection 2000 1/2” MNPT Process Connection

PROCESS CONNECTION

3000 1” MNPT Process Connection (not available for the 2” Pipeline Size)

FITS PIPELINE SIZE

CODE DESCRIPTION

L2 2” Pipeline

L4 3” - 4” Pipeline

L8 6” - 8” Pipeline

L12 10” - 12” Pipeline

L24 16” - 24” Pipeline

OPTIONS

CODE DESCRIPTION

B Explosion-Proof Conduit Box (includes Connector for RTD—1/2” NPT Connections)

C RTD Calibration (4-Point or 6-Point, depending on RTD Selected)

F 316 SS RTD Fitting (Delrin Standard)

P Cap & Chain Assembly (Can be ordered separately as Spare Part # ATA-1002)

R4 4-Wire Spring-Loaded RTD Probe

R6 6-Wire Spring-Loaded Dual RTD Probe

U 316 SS Union (PVC Standard)

V Vent/Bleed Hole in Thermowell (1/8”)

W1 316 SS Identification Tag

W1C 316 SS Tag with Calibration data (Valid only with C option)

ATA Base Models Consist of: 316 SS ThermoSync Thermowell, 316 SS Close Nipple, PVC Union and Delrin Coupler

18 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

ATA-SERIES ORDERING INFORMATION

Page 20: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

ATA-1000L4-R4

ATA-1000L4-R4U

ATA-1000L4-BR4U

Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com 19

ATA-SERIES OPTIONS

Page 21: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

Thermosync Thermowell RTD Probe/Sensor Gas Temperature Measurement

End Users (USA)

ATMOS Energy Corporation .......................................... Mississippi

Coastal Oil & Gas ......................................................... Texas

Coastal Flow Measurement ........................................... Texas

Devon Energy/Devon Gas Service ................................. Texas, Oklahoma

Blue Water Constructors, Inc. ....................................... Florida

El Paso Natural Gas Company ....................................... Texas, New Mexico, Louisiana

Eott Energy ................................................................. California

Florida Gas Transmission .............................................. Florida, Texas

Gemma Power Systems, LLC ........................................ Florida

Kinder Morgan Inc. ...................................................... Colorado, Wyoming, Texas, New Mexico, Illinois

Northern Border Pipeline Company ................................ Minnesota, Indiana, Illinois

Northern Natural Gas Company..................................... Alabama, Kansas, Minnesota, Louisiana, Nebraska, New Mexico, Florida, Iowa, Texas

PanHandle Eastern Pipeline ......................................... Kansas, Texas, Oklahoma

Peoples Gas ................................................................ Florida

Reliant Energy Field Services, Inc. ................................. Louisiana

Sierra Pacific Power ..................................................... Nevada

Transwestern Pipeline .................................................. New Mexico

Questar Gas Pipeline / Transmission .............................. Wyoming, Utah

Western Gas Resources ................................................ Wyoming

XTO Energy ................................................................ Texas, Oklahoma, Alaska, Colorado, New Mexico, Wyoming, Louisiana, Utah, Kansas

OEM Accounts (USA)

Automation X .............................................................. New Mexico

Instromet, Inc. ............................................................ Texas

J-W Measurement Company ......................................... Colorado, Texas

J.W. Williams-Flint ....................................................... Wyoming

Red Man Pipe & Supply ................................................ Oklahoma

Thermo Electron .......................................................... Texas

Wedge Measurement & Control .................................... Texas

20 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 www.pgiint.com

DOMESTIC END USERS & OEM USERS LIST

Page 22: ThermoSync Brochure 2013 - Team Technical · 3 Parker PGI 16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 REDUCING THERMOWELL ERRORS Reducing Thermowell Conduction

16101 Vallen Drive Houston, TX 77041 USA 713-466-0056 1-800-231-0233 Fax: 1-800-568-9228 [email protected] www.pgiint.com © Parker PGI, 2013 [3-22-13 Rev. 6]

Parker PGI provides this information in good faith, and it is intended only as an informative guide to Parker PGI products and services. Individuals using this literature must exercise their independent judgment in evaluating product selection and determining product appropriateness for their particular purpose, system requirements and certifications. Parker PGI reserves the right to change product designs and specifications without notice.