traceability and legal metrology - time and … nist seminar...audio time announcement telephone...
TRANSCRIPT
![Page 1: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/1.jpg)
Traceabilityand
Legal Metrology
Traceabilityand
Legal Metrology
![Page 2: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/2.jpg)
4 Parts of a MeasurementDevice Under TestDevice Under Test
(quartz, rubidium, and cesium)(quartz, rubidium, and cesium)
ReferenceReference(cesium oscillator or transfer standard)(cesium oscillator or transfer standard)
MethodMethod(measurement system and procedure)(measurement system and procedure)
Uncertainty AnalysisUncertainty Analysis(statistics and data reduction)(statistics and data reduction)
![Page 3: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/3.jpg)
ISO definition of traceabilityISO definition of ISO definition of traceabilitytraceability
The property of the result of a measurement or the value of a standard whereby it can be related to stated references, usually national or international standards, through an unbroken chain of comparisons all having stated uncertainties.
![Page 4: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/4.jpg)
The Traceability ChainThe Traceability ChainThe Traceability Chain
The traceability chain is a series of comparisons between the device under test to a reference. The final comparison in the chain is made using the International System (SI) units as a reference.Each comparison is a link in the chain.The uncertainty of each comparison (link) must be known and documented.
![Page 5: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/5.jpg)
The Traceability Chain (cont.)The Traceability Chain (cont.)The Traceability Chain (cont.)National metrology institutes (NMIs) like NIST provide the ultimate measurement references for their countries. The intent of all NMIs is to realize the SI units as closely as possible.Although the goal is to establish traceability to the SI, this is often done by comparing to an NMI that in turn compares its references to the SI.
![Page 6: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/6.jpg)
Time and Frequency SI UnitsTime and Frequency SI UnitsTime and Frequency SI UnitsSecond (s)Second (s)
standard unit for time intervalstandard unit for time intervalintrinsic, defined using cesium 133 atomintrinsic, defined using cesium 133 atomone of 7 base SI unitsone of 7 base SI units
Hertz (Hz)Hertz (Hz)standard unit for frequency (sstandard unit for frequency (s--11))events per secondevents per secondone of 21 SI units derived from base unitsone of 21 SI units derived from base units
![Page 7: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/7.jpg)
The Traceability PyramidThe Traceability Pyramid
![Page 8: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/8.jpg)
The Importance of TraceabilityThe Importance of Traceability
Establishing traceability to national and international standards provides evidence that measurements are being made correctlyTraceable measurements are essential elements of:
quality control systemstrade agreementsallowing compatible products to be manufactured all over the world
![Page 9: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/9.jpg)
The Traceability ProblemThe Traceability ProblemTechnicalTechnical
usually not a difficult technical problem, usually not a difficult technical problem, especially if requirements are modestespecially if requirements are modest
LegalLegalrequires repeatable, reliable proceduresrequires repeatable, reliable proceduresrequires documentation of traceability chain for requires documentation of traceability chain for quality control systemsquality control systemsmay be contractual requirement or a may be contractual requirement or a prerequisite for doing businessprerequisite for doing business
![Page 10: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/10.jpg)
ISO Guide 17025 - 5.6.2.1ISO Guide 17025 - 5.6.2.1For calibration laboratories, the program for calibration of equipment shall be designed and operated so as to ensure that calibrations and measurements made by the laboratory are traceable to the SI (Systeme International) units of measurement.
![Page 11: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/11.jpg)
Laboratory AccreditationLaboratory AccreditationCalibration laboratories that seek accreditation through a program such as NVLAP must comply with ISO 17025 and document the traceability chain for each measurement. The traceability chain must link the measurement back to the SI unit, and the documentation must include the uncertainty of each link of the chain.
![Page 12: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/12.jpg)
Types of Time and Frequency Information
Types of Time and Frequency Types of Time and Frequency InformationInformation
Date and TimeDate and Time--ofof--DayDayrecords when an event happenedrecords when an event happened
Time IntervalTime Intervalduration between two eventsduration between two events
FrequencyFrequencyrate of a repetitive eventrate of a repetitive event
![Page 13: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/13.jpg)
Everything Relates to the Second
Everything Relates to the Everything Relates to the SecondSecond
By counting seconds, we can determine By counting seconds, we can determine date and timedate and time--ofof--daydayBy counting events per second, we can By counting events per second, we can measure frequencymeasure frequency
![Page 14: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/14.jpg)
SI Definition of the SecondSI Definition of the SecondSI Definition of the Second
The duration of 9,192,631,770 periods of the The duration of 9,192,631,770 periods of the radiation corresponding to the transition radiation corresponding to the transition between two hyperfine levels of the ground between two hyperfine levels of the ground state of the cesiumstate of the cesium--133 atom.133 atom.
![Page 15: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/15.jpg)
Uncertainties of physical realizations of the base SI units
Uncertainties of physical Uncertainties of physical realizations of the base SI unitsrealizations of the base SI units
SI Base Unit Physical Quantity Uncertainty candela luminous intensity 1 x 10-4
mole amount of substance 8 x 10-8 kelvin thermodynamic temperature 3 x 10-7
ampere electric current 4 x 10-8 kilogram mass 1 x 10-8
meter length 1 x 10-12 second time interval 1.3 x 10-15
![Page 16: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/16.jpg)
Coordinated Universal Time (UTC)
Coordinated Universal Time Coordinated Universal Time (UTC)(UTC)
About 50 About 50 NMIsNMIs do a continuous comparison of do a continuous comparison of their standards (over 200 atomic oscillators are their standards (over 200 atomic oscillators are involved). The average value of these standards is involved). The average value of these standards is used to create the UTC time scale. The BIPM in used to create the UTC time scale. The BIPM in France handles the data and computes the France handles the data and computes the uncertainties. uncertainties. BIPM Circular T shows difference between UTC BIPM Circular T shows difference between UTC and UTC(k), or the UTC maintained by each NMI.and UTC(k), or the UTC maintained by each NMI.
![Page 17: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/17.jpg)
BIPM Circular T (www.bipm.fr)BIPM Circular T (www.bipm.fr)
![Page 18: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/18.jpg)
Low accuracy measurements that require traceability
Low accuracy measurements that require traceability
![Page 19: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/19.jpg)
High accuracy measurements that require traceability
High accuracy measurements that require traceability
![Page 20: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/20.jpg)
Other Areas of Metrology that require Traceable FrequencyOther Areas of Metrology that require Traceable Frequency
![Page 21: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/21.jpg)
Ways to establish traceability to the SI through NIST
Ways to establish traceability Ways to establish traceability to the SI through NISTto the SI through NIST
NIST controlled broadcastsNIST controlled broadcastsWWV/WWVH, WWVB, ACTS, ITSWWV/WWVH, WWVB, ACTS, ITS
NIST monitored broadcastsNIST monitored broadcastsLORANLORAN--C, GPSC, GPS
NIST controls/monitors measurementsNIST controls/monitors measurementsFrequency Measurement ServiceFrequency Measurement Service
![Page 22: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/22.jpg)
Traceability chain for signals
controlled by a NMI
Traceability chain for signals
controlled by a NMI
![Page 23: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/23.jpg)
The Distribution of UTC(NIST)The Distribution of UTC(NIST)
![Page 24: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/24.jpg)
Traceability chain for signals monitored by a NMI
Traceability chain for signals monitored by a NMI
![Page 25: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/25.jpg)
GPS MonitoringGPS MonitoringMore complex than monitoring ground
based signals because:Each orbiting satellite carries its own frequency standardsSatellites can only be received when they fly over monitoring station
![Page 26: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/26.jpg)
NIST GPS Data Archive
![Page 27: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/27.jpg)
NIST GPS Data Archive
![Page 28: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/28.jpg)
GPS Traceability Chain A
Link Reference Compared To:A SI units UTC(NIST)B UTC(NIST) UTC(USNO)C UTC(USNO) GPS SignalsD GPS Broadcast Signals GPS Received SignalsE GPS Received Signals User’s Device under Test
![Page 29: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/29.jpg)
GPS Traceability Chain B
Link Reference Compared To:A SI units UTC(NIST)B UTC(NIST) GPS Broadcast SignalsC GPS Broadcast Signals GPS Received SignalsD GPS Received Signals User’s Device under Test
![Page 30: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/30.jpg)
NIST traceable signals and linksNIST traceable signals and linksTraceable Signal or Link
Receiving Equipment
TimingOffset
Timing Uncertainty
Frequency Uncertainty
Audio Time Announcement
Telephone (303-499-7111)
< 30 ms
(within USA)
< 1 ms
NA
Automated Computer Time Service (ACTS)
Computer, software, modem, and phone line (303-494-4774)
1 to 150 ms
(< 10 ms with loop test)
< 5 ms
NA
Network Time Service (NTS)
Computer, software, and Internet connection
< 1 s
< 100 ms
NA
WWV and WWVH HF Receiver (2.5, 5, 10, 15, or 20 MHz)
1 to 15 ms
(within USA)
< 500 µs
10-6 to 10-9
WWVB or LORAN-C LF Receiver (60 or 100 kHz)
< 100 µs (calibrated for
path delay)
< 500 ns
10-10 to 10-12
Global Positioning System (GPS)
GPS Receiver (1575.42 MHz)
< 100 ns (calibrated for
equipment delay)
< 20 ns
10-12 to 10-13
GPS common-view, GPS carrier phase, and Two Way Satellite Time Transfer (TWSTT)
Receiving equipment, transmitting equipment (TWSTT only), tracking schedules, and data link with NIST
< 5 ns
< 5 ns
< 1 x 10-13
![Page 31: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/31.jpg)
Measuring 1 x 10-8Measuring 1 x 10-8
Frequency Offset of 1 x 10-8
4000
4100
4200
4300
4400
4500
4600
4700
4800
4900
5000
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
Hours, GPS=1, WWVB=2, GOES=3
Mic
rose
cond
s
Series1Series2Series3
![Page 32: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/32.jpg)
Measuring 1 x 10-9Measuring 1 x 10-9
Frequency Offset of 1 x 10-9
4060
4080
4100
4120
4140
4160
4180
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
Hours, 1=GPS, 2=WWVB, 3=GOES
Mic
rose
cond
s
Series1Series2Series3
![Page 33: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/33.jpg)
Measuring 1 x 10-10Measuring 1 x 10-10
Frequency Offset of 1 x 10-10
3960
3965
3970
3975
3980
3985
3990
3995
4000
4005
4010
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Hours, 1=GPS, 2=WWVB, 3=GOES
Mic
rose
cond
s
Series1Series2Series3
![Page 34: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/34.jpg)
Measuring 1 x 10-11Measuring 1 x 10-11
Frequency Offset of 10-11
3993
3994
3995
3996
3997
3998
3999
4000
4001
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19Hours, 1=GPS, 2=WWVB
Mic
rose
cond
s
Series1Series2
![Page 35: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/35.jpg)
Time and Frequency is well suited for remote calibrationsTime and Frequency is well
suited for remote calibrationsSince frequency standards are sensitive to shipment and environmental changes, calibrations are often made at the customer’s siteRadio reference (often GPS) is used as transfer standardCalibrations can be monitored by an NMI using a modem or Internet link
![Page 36: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/36.jpg)
NIST Frequency Measurement and Analysis Service
NIST Frequency Measurement and Analysis Service
Measurements are made remotely at customer’s Measurements are made remotely at customer’s sitesite
NIST supplies everything neededNIST supplies everything neededMeasurement process is automatedMeasurement process is automated
NIST downloads data by modem, performs NIST downloads data by modem, performs uncertainty analysis, sends calibration reportsuncertainty analysis, sends calibration reports
Provides traceability to NIST with uncertainty of Provides traceability to NIST with uncertainty of 2 x 102 x 10--1313 per dayper day
![Page 37: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/37.jpg)
NIST Frequency Measurement and Analysis Service (cont.)
NIST Frequency Measurement and Analysis Service (cont.)
NIST Service ID Number 76100S5 oscillators can be calibrated at onceCalibrations are automatedRecognized by NVLAPReplacement parts sent when necessary using overnight delivery service
![Page 38: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/38.jpg)
FMAS Customer Map (labs with cesium oscillators in red)
![Page 39: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/39.jpg)
NIST FMAS DisplayNIST FMAS Display
![Page 40: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/40.jpg)
Calibration ReportsCalibration ReportsMonthly calibration reports are sent to each NIST FMAS customer. Report is compliant with ISO Guide 25/17025 and includes a statement of uncertainty.The report defines the calibration period as one day (24 hours). The report annotates all situations where no data were recorded or where the device under test was out of tolerance.
![Page 41: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/41.jpg)
Reporting of Measurement Errors and Outages
Reporting of Measurement Errors and Outages
The monthly calibration report annotates The monthly calibration report annotates the following situations:the following situations:
No data were recordedNo data were recordedGPS reception problemsGPS reception problemsGPS broadcast errorsGPS broadcast errorsMeasurement System Errors including Measurement System Errors including hardware or software failureshardware or software failuresDevice under Test errorsDevice under Test errorsDevice under Test substitutions/changesDevice under Test substitutions/changes
![Page 42: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/42.jpg)
FMAS SpecificationsFMAS Specifications
Number of Measurement Channels 5Input Frequencies Accepted bySystem
1 Hz to 120 MHz in 1 Hzincrements
Single Shot Resolution < 30 ps
Measurement Uncertainty, GPScomparisons(24 hours, 2 sigma)
2 x 10-13
Measurement Uncertainty, oscillatorto oscillator comparisons(24 hours, 2 sigma)
2 x 10-15
![Page 43: Traceability and Legal Metrology - Time and … NIST Seminar...Audio Time Announcement Telephone (303-499-7111) < 30 ms (within USA) < 1 ms NA Automated Computer Time Service (ACTS)](https://reader034.vdocument.in/reader034/viewer/2022050405/5f82aef8e3e3ae3d0772858b/html5/thumbnails/43.jpg)
NIST Policy on TraceabilityNIST Policy on Traceability
For more information about traceability, visit the NIST Traceability Web Site:
http://http://nistnist..govgov/traceability//traceability/