    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 1© Pfeiffer Vacuum 2007

    Leak Detector

    SmartTestdesign and function

  • 8/16/2019 Training SmartTest HLT560

    2/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 2© Pfeiffer Vacuum 2007

    The participant will be able to:

    explain the design and the function

    operate the unit

    perform maintenance work

    do troubleshooting

    Course objectives

  • 8/16/2019 Training SmartTest HLT560

    3/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 3© Pfeiffer Vacuum 2007

    Definition “ LEAK"

     A LEAK is an untight spot in a container or line,through which gases, vapors, or even liquids flow

    from an area where the pressure is higher to an area

    where the pressure is lower.

  • 8/16/2019 Training SmartTest HLT560

    4/90EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 4© Pfeiffer Vacuum 2007

    SI - Standard = mol/s (1mol = 6.025E23 particles)

    Fluid technology = Nm/s, sccm, sccs

    Vacuum technology = mbar l/s, Pa m3/s, Torr l/s,

    atm cc/s

     Air conditioning = g/a, oz/y

    Concentration = ppm (no leak rate unit)

    Leak rate units

  • 8/16/2019 Training SmartTest HLT560

    5/90EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 5© Pfeiffer Vacuum 2007

    The leak rate is 1 mbar l/s if in an evacuated container

    with a volume of 1 Liter, the pressure rises by 1 mbarper second , or if in a pressurized container with the

    same volume, the pressure drops by 1 mbar per

    second.

    Definition of the unit "mbar l/s"

  • 8/16/2019 Training SmartTest HLT560

    6/90EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 6© Pfeiffer Vacuum 2007

     A bicycle tube may have a maximum pressure

    loss of 1000 mbar in 175 days.Volume = 1.5 liters

    175 days = 1.5x107 seconds

    Which is the maximumadmissible leak rate?

    How do you test a bicycle

    tube?

    Example 1: Leak rate in mbar l/s

  • 8/16/2019 Training SmartTest HLT560

    7/90EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 7© Pfeiffer Vacuum 2007

    Example 1

    Result:

    QL=[(p

    2- p

    1) * V] / t

    s / lmbar 101,0101,5

    1,51000

    s

    lmbar Q

      4

    7  L  ∗∗=

    ∗=

    ∗=

      −

    Bubble test

    The leakrate should be not higher than 1x10-04 mbarl/s.

  • 8/16/2019 Training SmartTest HLT560

    8/90EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 8© Pfeiffer Vacuum 2007

    Example 2: Leak rate in mbar l/s

     A bicycle tube is pumped up to 3.0 bar.

    The default leak rate is taken from example

    Volume = 1.5 liters

    1year = ca. 3.0 x 107 seconds

    Leak rate = 1,0 E-04 mbar l/s

    What is the pressure in the bicycle tube after 1

    year?

  • 8/16/2019 Training SmartTest HLT560

    9/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 9© Pfeiffer Vacuum 2007

    Example 2

    bar mbar  x p

    mbar  x p

    lsslmbar  x x p

    tube

    loss

    loss

    0.1100.1

    100.2

    ***

    5.1100.3*100.1

    3

    3

    74

    ==

    =

    ===

    Volumen

    Timerate Leak 

     ploss*

    =

    Result

    The pressure in the bicycle tube is 1 bar after 1

    year?

  • 8/16/2019 Training SmartTest HLT560

    10/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 10© Pfeiffer Vacuum 2007

    Test object pressurized:

    Pressure decay

    Bubble test

    He sniffing with SmartTest

    Leak detection methods

  • 8/16/2019 Training SmartTest HLT560

    11/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 11© Pfeiffer Vacuum 2007

    E1 E0 E-8 E-9 E-10 E-11 E-12E-3 E-4 E-5 E-6 E-7E-1 E-2

    He sniffing with SmartTest

    mbar l/s

    Detection limits pressurized

    Bubble test

    Pressure decay

  • 8/16/2019 Training SmartTest HLT560

    12/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 12© Pfeiffer Vacuum 2007

    Test object under vacuum:

    Pressure increase

    Gas type dependence ( Pirani gauge)

    He leak test with SmartTest

    Leak detection methods

  • 8/16/2019 Training SmartTest HLT560

    13/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 13© Pfeiffer Vacuum 2007

    Detection limits vacuum

    E1 E0 E-8 E-9 E-10 E-11 E-12E-3 E-4 E-5 E-6 E-7E-1 E-2

    Pressure increase

    Gas type dependence PiraniGas type dependence Pirani

    He leak test with SmartTest

    mbar l/s

  • 8/16/2019 Training SmartTest HLT560

    14/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 14© Pfeiffer Vacuum 2007

    Sniffing methods

       T  e  s   t  g  a  s

    Integral

       T  e  s   t  g  a  s

    Local

  • 8/16/2019 Training SmartTest HLT560

    15/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 15© Pfeiffer Vacuum 2007

    Vacuum methods integral

       T  e  s   t  g  a  s

    Test object evacuated

    Test chamber pressurized

       T  e  s   t  g  a  s

    Test object pressurized

    Test chamber evacuated

  • 8/16/2019 Training SmartTest HLT560

    16/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 16© Pfeiffer Vacuum 2007

       T

      e  s   t  g  a  s

    Test object evacuated

    Vacuum method local

  • 8/16/2019 Training SmartTest HLT560

    17/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 17© Pfeiffer Vacuum 2007

    SmartTest HLT 5xx

    1 Unit HLT 5xx

    2 Cart to SmartTest

    3 Backing pump

    1

    2

    3

    Modular concept

  • 8/16/2019 Training SmartTest HLT560

    18/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 18© Pfeiffer Vacuum 2007

    Type overview

    SmartTest

    SmartTest HLT 550

    HLT 560

    HLT 570

    Rotary vane pump

    without Backing pump

    Diaphragm pump

    SmartTest

    Dry pump TS 620

    SmartTest

    SmartTest

    HLT 565

    HLT 572

    HLT 575SmartTest

    Rotary vane pump

    UNO 5

    UNO 30 M

    Dry pump XtraDry 150-2

  • 8/16/2019 Training SmartTest HLT560

    19/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 19© Pfeiffer Vacuum 2007

    The leak detector SmartTest

    1 Test port

    2 Instrument operation

    3 Inlet air filter

    4 Connection for

    external backing pump

    1 2

    3

    4

  • 8/16/2019 Training SmartTest HLT560

    20/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 20© Pfeiffer Vacuum 2007

    Instrument operation

    1 START / STOP – button

    2 ZERO – button

    3 Softkeys with

    changeable meaning

    4 Display

    12

    3

    4

  • 8/16/2019 Training SmartTest HLT560

    21/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 21© Pfeiffer Vacuum 2007

    V4

    V1

    V3

    V6

    V5

    V2

    P2

    P1

    Disign

    1 Test port

    2 Pressure at Test port

    3 Clibrated leak

    4 Vent of sniff connection

    5 Backing pump

    6 Fore vacuum pressure

    7 Turbo drag pump

    8 Mass spectrometer 

    1

    2

    3

    7

    4

    6

    8

    5

  • 8/16/2019 Training SmartTest HLT560

    22/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 22© Pfeiffer Vacuum 2007

    Mass spectrometer 

    M > 4

    M < 4

    M = 4

    3 12

    5

    9

    11

    8

    7

    6

    410

    1 Cathode 1

    2 Anode

    3 Cathode 2

    4 Shielding

    5 Extractor 

    6 Shielding

    7 Magnetical field

    8 Suppressor 

    9 Shielding10 Faraday cup

    11 Pre amplifier 

  • 8/16/2019 Training SmartTest HLT560

    23/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 23© Pfeiffer Vacuum 2007

    Ionsource

    Gas inlet

    Pre amplifier 

    Magnet

    Magnetic field 

     +

     +

     +

     +

     

    +

     +

    Principle of measurement

  • 8/16/2019 Training SmartTest HLT560

    24/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 24© Pfeiffer Vacuum 2007

    V4

    V1

    V3

    V6

    V5

    V2

    P2

    P1

    START/STOP pressedPressure value P2

    1000

    15

    0,5

    5

    mbar 

    0

  • 8/16/2019 Training SmartTest HLT560

    25/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 25© Pfeiffer Vacuum 2007

    V4

    V1

    V3

    V6

    V5

    V2

    P2

    P1

    Pump fore vacuum

    P2 > 15 mbar 

  • 8/16/2019 Training SmartTest HLT560

    26/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 26© Pfeiffer Vacuum 2007

    V4

    V1

    V3

    V6

    V5

    V2

    P2

    P1

    Measure „ Counter flow“

    P2 < 15 mbar 

    P2 > 5 mbar 

  • 8/16/2019 Training SmartTest HLT560

    27/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 27© Pfeiffer Vacuum 2007

    V4

    V1

    V3

    V6

    V5

    V2

    P2

    P1

    Measure "Twin-Flow low"

    P2 < 5 mbar 

    P2 > 0.5 mbar 

  • 8/16/2019 Training SmartTest HLT560

    28/90

  • 8/16/2019 Training SmartTest HLT560

    29/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 29© Pfeiffer Vacuum 2007

    V4

    V1

    V3

    V6

    V5

    V2

    P2

    P1

    Sniffing „ Twin-Flow low“

  • 8/16/2019 Training SmartTest HLT560

    30/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 30© Pfeiffer Vacuum 2007

    V4

    V1

    V3

    V6

    V5

    V2

    P2

    P1

    Calibration

  • 8/16/2019 Training SmartTest HLT560

    31/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 31© Pfeiffer Vacuum 2007

    Running Up

    The run-up of the turbo pump is visualized in percentage via

    bar graph.

    Details Menu with selected system data

    Sprache, Language Direct link to the menu Language

    Idioma

    Setup Settings of the operating parameters

  • 8/16/2019 Training SmartTest HLT560

    32/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 32© Pfeiffer Vacuum 2007

    Details

    For l ine pressure p1 Present fore-vacuum pressure

    Rotation speed Present rotation of the turbo pump

    TMP current Present consumption of the turbo pump

    Emission „ On“ if the rotation >= 1450 Hz , p1 < 10 mbar  

    and the filament test was okay

    Filament Active filament 1 or 2

    Vent TMP The filament and the turbo pump

    will switch „ off“ automatically

    The menu “ Vent TMP” will displayed only if the menu

    „ Maintenace enable at „ Yes“ .

  • 8/16/2019 Training SmartTest HLT560

    33/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 33© Pfeiffer Vacuum 2007

    Venting TMP (1)

    The turbo pump is switch „ off“ and the leak detector is waiting

    unti l the rotation speed of the turbo pump < 150 Hz.

    Restart Stops the automatic venting procedure of the turbopump and leads back to the run up mode of the unit.

  • 8/16/2019 Training SmartTest HLT560

    34/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 34© Pfeiffer Vacuum 2007

    Venting TMP (2)

    The rotation speed of the turbo pump

  • 8/16/2019 Training SmartTest HLT560

    35/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 35© Pfeiffer Vacuum 2007

    Venting TMP (3)

    The turbo pump is vented and the leak detector can be switched off.

    Vent TMP Vents the turbo pump again

    Restart Stops the automatic venting procedure of the turbo

    pump and leads back to the run up mode of the unit.

  • 8/16/2019 Training SmartTest HLT560

    36/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 36© Pfeiffer Vacuum 2007

    Ready to start

    Display

    internal test leak

    Start the calibration

    SettingsMode, Mass,

    Last calibrationFilter 

    Background signal

    Setup main menu

    Start the regeration

    Direct link to the

    menu Mode & Mass

  • 8/16/2019 Training SmartTest HLT560

    37/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 37© Pfeiffer Vacuum 2007

    Measured value display

     Analog

    anddigital

    Graphic

    Changedisplay mode

  • 8/16/2019 Training SmartTest HLT560

    38/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 38© Pfeiffer Vacuum 2007

     Activate Zero

    0

    10

    20

    30

    40

    50

    t1

    t

    ZERO

    Leak rate

    Displayed leak rate

    Measured leak rate

  • 8/16/2019 Training SmartTest HLT560

    39/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 39© Pfeiffer Vacuum 2007

    Zero at falling back ground

    0

    10

    20

    30

    40

    50

    t2

    t

    t1

    Leak rate

    Measured leak rate

    Displayed leak rate

    Subrtraction value

  • 8/16/2019 Training SmartTest HLT560

    40/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 40© Pfeiffer Vacuum 2007

    Deactivate Zero

    0

    10

    20

    30

    40

    50

    t3

    t

    ZERO

    > 3 s

    Leak rate

    Measured leak rate

    Displayed leak rate

  • 8/16/2019 Training SmartTest HLT560

    41/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 41© Pfeiffer Vacuum 2007

    Setup main menu

    View Display configuration

     Access control Set access control

    Language Selection the language

    Info Information

    Calibration settings Setting for the calibration

    User settings Configuration of the user parameters

    Maintenance & Service Menus Maintenance & Service

  • 8/16/2019 Training SmartTest HLT560

    42/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 42© Pfeiffer Vacuum 2007

    View

    Contrast Settings for display contrast

    Units Settings for leak rate and pressure

    Time & Date Setting for time and date

    Display range Configuration of the display range

    Lower display limit Setting for lower display limit

    Background Setting for Display background signal

    at ready to start

  • 8/16/2019 Training SmartTest HLT560

    43/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 43© Pfeiffer Vacuum 2007

    Contrast

    Setting contrast range 0 – 99

    Invert display normal / invert display

  • 8/16/2019 Training SmartTest HLT560

    44/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 44© Pfeiffer Vacuum 2007

    Units

    Unit Leak rate mbar*l/s, Pa*m3/s, Torr*l/s

    sccm, sccs, atm*cc/s,

    ppm, g/a, oz/yr (only selectable in „ Sniffing“ mode)

    Unit Pressure mbar, Pa, atm, Torr  

  • 8/16/2019 Training SmartTest HLT560

    45/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 45© Pfeiffer Vacuum 2007

    Time & Date

    Date Setting of the actual date

    Time Setting of the actual time (24 hours format)

  • 8/16/2019 Training SmartTest HLT560

    46/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 46© Pfeiffer Vacuum 2007

    Display range

    Scale linear Display linear  

    logarithmic Display logarithmic

    2…9 Number of decades in

    log. Display

    Range automatic Automatic range selection

    manual Manual range selection

    Time axis 16…960 Time axis, time in seconds

  • 8/16/2019 Training SmartTest HLT560

    47/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 47© Pfeiffer Vacuum 2007

    Lower display limit

    Lower display limit for unit mbar*l/s

    1E-12 mbar*l/s

    1E-11 mbar*l/s

    1E-10 mbar*l/s1E-09 mbar*l/s

    This setting limits the display of the leak rate downwards in

    the measuring mode. It is only effective for the vacuum mode.

  • 8/16/2019 Training SmartTest HLT560

    48/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 48© Pfeiffer Vacuum 2007

    Background at „ Ready to start“

    yes Background displayed

    at „ Ready to Start“

    no Background not displayed

    at „ Ready to Start“

  • 8/16/2019 Training SmartTest HLT560

    49/90

  • 8/16/2019 Training SmartTest HLT560

    50/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 50© Pfeiffer Vacuum 2007

    Maintenance enable

    Yes Menu „ Maintenance & Service“

    is enabled

    No Menu „ Maintenance & Service“is disabled

  • 8/16/2019 Training SmartTest HLT560

    51/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 51© Pfeiffer Vacuum 2007

    Calibration enabled

    Yes Calibration at „ Ready to Start“ enabled

    No Calibration at „ Ready to Start“ disabled

  • 8/16/2019 Training SmartTest HLT560

    52/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 52© Pfeiffer Vacuum 2007

    Change device PIN

    New PIN 0000 – 9999 New device PIN

    New PIN (verify) 0000 – 9999 New device PIN

    (repeat for confirmation)

  • 8/16/2019 Training SmartTest HLT560

    53/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 53© Pfeiffer Vacuum 2007

    Change menu PIN

    New PIN 0000 – 9999 New menu PIN

    New PIN (verify) 0000 – 9999 New menu PIN

    (repeat for confirmation)

    L

  • 8/16/2019 Training SmartTest HLT560

    54/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 54© Pfeiffer Vacuum 2007

    Language

    The fol lowing languages are available for operation:

    Language

    Francais French

    Deutsch German Operating language

    Espanol Spanish

    English English

    I f ti

  • 8/16/2019 Training SmartTest HLT560

    55/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 55© Pfeiffer Vacuum 2007

    Information

    Settings Display menu parameters

    System data Display system data

    Vacuum diagram Display vacuum diagram

    Error list Display the last 12 errors

    Calibration history Display the last 12 calibrations

    S tti

  • 8/16/2019 Training SmartTest HLT560

    56/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 56© Pfeiffer Vacuum 2007

    Settings

    Display of the current parameter settings in 4 pages.

    S t d t

  • 8/16/2019 Training SmartTest HLT560

    57/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 57© Pfeiffer Vacuum 2007

    System data

    Display of the system data/components in 4 pages.

    Vacuum diagram

  • 8/16/2019 Training SmartTest HLT560

    58/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 58© Pfeiffer Vacuum 2007

    Vacuum diagram

    Vacuum diagram with the current states of valves, turbo pump

    and emission.

    Leak rate and pressure values P1, P2 are displayed.

    Error list

  • 8/16/2019 Training SmartTest HLT560

    59/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 59© Pfeiffer Vacuum 2007

    Error list

    List of the last 12 errors in following format:

    Date / Time / Error no. / Remark

    Calibration history

  • 8/16/2019 Training SmartTest HLT560

    60/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 60© Pfeiffer Vacuum 2007

    Calibration history

    List of the last 12 Calibration in following format:

    Date / Time / Location / Test gas / Calibration factors

  • 8/16/2019 Training SmartTest HLT560

    61/90

    User Settings

  • 8/16/2019 Training SmartTest HLT560

    62/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 62© Pfeiffer Vacuum 2007

    User Settings

    Mode & Mass Operation mode and detected mass

    Filter & Zero Filter and zero settings

     Alarm Alarm sett ings

    Interfaces Interface settings

    Parameter save / load Sets of parameter, factory settings

    Monitor ing funct ions Diverse funct ions can be monitored

  • 8/16/2019 Training SmartTest HLT560

    63/90

    Filter & Zero

  • 8/16/2019 Training SmartTest HLT560

    64/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 64© Pfeiffer Vacuum 2007

    Filter & Zero

    Filter without No leak rate filter  static Filter with fixed time constant

    dynamic Filter with dynamic adaptation of time constant

    Zero with Start Zero is executed immediately after the

    2s…5 min specif ied time.

    enabled Manual Zero releaseddisabled Manual Zero locked

    constant Manual Zero released and the zero value is retained

    after pressing STOP. Automatic BG -tracing is locked.

    MS-BG

    sub- On Internal background is subtracted at START

    traction Off Internal background is not subtracted at START

    Alarm

  • 8/16/2019 Training SmartTest HLT560

    65/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 65© Pfeiffer Vacuum 2007

     Alarm

    Mode prop. leak rate Acoustic signal proportional to the bar displaytrigger alarm A mult i-frequency signal is generated as soon

    0… 4.5 min as the leak rate exceeds the setpoint value.

     An alarm delay t ime can be entered.

    set point A tone with the frequency proport ional to the

    0… 4,5 min leak rate is emitted as soon as the leak rate

    exceeds the setpoint value. An alarm delay t ime can be entered.

    pinpoint Acoustic signal proportional to the leak rate

    between 0,1 and 10*setpoint value

    Setpoint 1E-12…9,9E+2 Alarm setpoint value

    Warn limit 1…100 % Warning limit as percentage of setpoint value

    Interfaces

  • 8/16/2019 Training SmartTest HLT560

    66/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 66© Pfeiffer Vacuum 2007

    Interfaces

     Analog output Output of the analog signal

    Compact gauge Compact gauge settings

    Control location Selection of the control location

    Relay Relay settings

    Serial port Selection of port and protocol

    Bypass option Bypass option settings

     Analog output

  • 8/16/2019 Training SmartTest HLT560

    67/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 67© Pfeiffer Vacuum 2007

    g p

    Channel 1 off Channel is off(0V)Pressure p1 Output pressure p1

    Pressure p2 Output pressure p2

    LR mantissa Output leak rate mantissa, l inear from 0…10 V

    LR exponent Output leak rate exponent as a step function

    LR linear Output leak rate linear, linear from 0…10 V

    LR log. Output leak rate logarithmic, upper limit 10VPressure p(ext) Output external pressure

    Channel 2 see channel 1

    Scale upper limit upper limi t(=10V) for setting „ LR log“

    1E-11…1E+06 and “LR linear”

    V / decade 0.5, Volt per decade for setting „ LR log“

    1, 2, 2.5, 5, 10

    Compact gauge

  • 8/16/2019 Training SmartTest HLT560

    68/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 68© Pfeiffer Vacuum 2007

    p g g

    Full scale 0.1, 1, 10,100, 1000, 2000, 5000,(only for linear gauges) 10000, 50000 mbar 

    Type APR, CMR, PCR, PKR, TPR

    Setpoint value Setpoint value for relay output

    1E-10…9,9E+02

    Pressure p(ext) Current pressure value p(ext)

    Press. internal Inlet pressure determined with

    P2 source internal gauge

    external Inlet pressure determined with

    external gauge

    Control location

  • 8/16/2019 Training SmartTest HLT560

    69/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 69© Pfeiffer Vacuum 2007

    Local Controlled by START, STOP and ZERO keys

    Local and Control led by START, STOP and ZERO keys

    RS232/RS485 and via RS232/RS485 interface

    RS232/RS485 Controlled via RS232/RS485 interface

    PLC Controlled via digital input

     ALL Controlled by START, STOP and ZERO keys,

    via digi tal input and via RS232/RS485 interface

    Relay

  • 8/16/2019 Training SmartTest HLT560

    70/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 70© Pfeiffer Vacuum 2007

    y

    Relay 1,2Off Relay is always dropped-out.

    START Relay is active when V2 is open.

    STOP Relay is active when V6 is open.

    START/STOP Relay is active when „ START“ is pressed and

    is dropped-out when „ STOP“ is pressed.

    Ready Relay is active in measure.Setpoint Relay is active when leak rate exceeds setpoint.

    On Relay is always active.

    Warning limit Relay is active when leak rate exceeds warn limit.

    Pressure setpoint Relay is active when pressure exceeds setpoint.

    Serial ports

  • 8/16/2019 Training SmartTest HLT560

    71/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 71© Pfeiffer Vacuum 2007

    Port RS 232 Control via RS 232RS 485 Control via RS 485

    Protocol Pfeiffer Pfeiffer interface protocol

    HLT 2xx HLT 2xx interface protocol

    (selected commands)Binary Diagnosis interface protocol

    (only for internal use)

     Address 1…255 Bus-address of SmartTest

    in Pfeifer protocol

    Bypass option

  • 8/16/2019 Training SmartTest HLT560

    72/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 72© Pfeiffer Vacuum 2007

    Mode No bypass No bypass option selected

    Quick pump The optional Backing pump is ready

    for the pumping down process without

    being active in the measuring mode.

    Partial flow The optional Backing pump is readyfor the pumping down process and the

    measuring mode.

    Internal on / off Delay time for the act ivat ion of the internal

    evacuation Backing pump is enabled / disabled.

    delay

  • 8/16/2019 Training SmartTest HLT560

    73/90

    Load parameters

  • 8/16/2019 Training SmartTest HLT560

    74/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 74© Pfeiffer Vacuum 2007

    View Displays values of parameter set

    Parameter set „PARA SET“ 1,2 .(4 pages)

    Load Loads the parameter set „ PARA SET“ 1 or 2.

    Save parameters

  • 8/16/2019 Training SmartTest HLT560

    75/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 75© Pfeiffer Vacuum 2007

    The current parameter setting will be saved after pressing

    the button „ Save“ in „ PARA SET 1“ or „ PARA SET 2“ .

    Load factory settings

  • 8/16/2019 Training SmartTest HLT560

    76/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 76© Pfeiffer Vacuum 2007

    The factory settings will be loaded after pressing the button „ OK“ .

    Monitoring functions

  • 8/16/2019 Training SmartTest HLT560

    77/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 77© Pfeiffer Vacuum 2007

    Flow Settings for the flow control

    (sniffing mode only)

    Contamination Settings of the protection against

    protection helium contamination

    Min. Volume & Beep Settings of minimal volume and beep

    Valves Settings valve V1, V3 and V4

    Evacuation time Settings for evacuation time and vent mode

    & Vent

    Calibration request Enables and disables request for a cal ibration

    Flow

  • 8/16/2019 Training SmartTest HLT560

    78/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 78© Pfeiffer Vacuum 2007

    Flow min „ Flow too low“ appears when the current

    1…40 sccm value falls below this threshold

    Flow max „ Flow too high“ appears when the current

    10…50 sccm value exceeds t this threshold

    Current flow Current flow of the sniffer line (measure mode)

    These settings only take place in “ sniff ing mode” .

  • 8/16/2019 Training SmartTest HLT560

    79/90

  • 8/16/2019 Training SmartTest HLT560

    80/90

    Valves

  • 8/16/2019 Training SmartTest HLT560

    81/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 81© Pfeiffer Vacuum 2007

    Twin-Flow high enable Valve V4 opensdisable Valve V4 always closed

    0,01… 0,5 Threshold value when

    mbar V4 opens

    Twin-Flow low enable Valve V3 opens

    disable Valve V3 always closed

    0,1… 5 mbar Threshold value whenV3 opens

    Counterflow enable Valve V1 opens

    disable Valve V1 always closed

    0,1… 25 mbar Threshold value when

    V1 opens

    Evacuation time & vent

  • 8/16/2019 Training SmartTest HLT560

    82/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 82© Pfeiffer Vacuum 2007

    Maximum 1 s…30min This is the threshold for maximumevacuation t ime inf ini tely pump down t ime before this process

    will stopped.

    Vent with Stop Inlet is vented automatically after Stop.

    manual Inlet can be vented in „Ready to Start“

    by pressing the „ Vent“ softkey.

    no Venting of the inlet is locked.

    Regeneration Start-Stop-cycles with short range intended for

    the reduction of a raised helium background.

    Regeneration

  • 8/16/2019 Training SmartTest HLT560

    83/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 83© Pfeiffer Vacuum 2007

     Automated Start-Stop-cycle with short range intended

    for the reduction of a raised helium background.

     An active “ Regenerat ion” will be announced in the display

    and stops after 60 minutes automatically.

    This funct ion can be deactivate in general with the STOP

    key or with STOP in the setting “ Venting with STOP” menu.

    The “ Regeneration” can only be successful activated

    in the setting “ Venting with STOP” menu!

    Calibration request

  • 8/16/2019 Training SmartTest HLT560

    84/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 84© Pfeiffer Vacuum 2007

    Yes The request for calibration comes up:

    - 30 min after switching HLT „ On“

    - temperature of HLT has changed

    more than 5°C

    No The request for calibration does not come up.

    Maintenance & Service

  • 8/16/2019 Training SmartTest HLT560

    85/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 85© Pfeiffer Vacuum 2007

    Burn- in Automatic start-stop cycle

    Maintenance intervals Displays the current operating hours

    components of the components fore pump,

    turbo pump and ion source

    View Maintenance List of maintenance intervals

    intervals list

    Service Access to menu Service

    The maintenance menu can only be chosen when the maintenance

    is enabled under Setup / Access control / Maintenance enabled.

    Burn-In

  • 8/16/2019 Training SmartTest HLT560

    86/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 86© Pfeiffer Vacuum 2007

    Start without Automatic start-stop cycle

    Calibration without calibration

    Start with Automatic start-stop cycle

    Calibration with calibration

    You can see in the display when the Burn–In is activated.

    The Burn-In can be deactivated with the STOP key.

    Burn-In activ

  • 8/16/2019 Training SmartTest HLT560

    87/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 87© Pfeiffer Vacuum 2007

    Run time Time in minutes since activating

    the Burn-In function

    Cal factor min. Calibration factor low

    Cal factor max. Calibration factor high

     Anode pot . min. Anode potent ial low Anode pot . max. Anode potent ial h igh

    Stop Deactivate the Burn-In

    This menu is available when the Burn-In is active.

    (Setup / Maintenance & Service / Burn-In)

  • 8/16/2019 Training SmartTest HLT560

    88/90

    Maintenance interval list

  • 8/16/2019 Training SmartTest HLT560

    89/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 89© Pfeiffer Vacuum 2007

    List of the last 12 Maintenance in following format:

    Date / Time / Component / Operating hours

    Service

  • 8/16/2019 Training SmartTest HLT560

    90/90

    EPM, Wolfgang Flatau, 2007-08, Training SmartTest page 90© Pfeiffer Vacuum 2007

    The service menu is available after entering the Service-PIN.

    This menu offers more information an settings for the service

    engineers.