treatment planning of lung cancer: dosimetric considerations · 2012. 4. 30. · brt-based...
TRANSCRIPT
Treatment PlaCancer: Dosimetr
Indrin J ChIndrin J. Ch
Henry Ford Hy
anning of Lung ric Considerations
hetty PhDhetty, PhD
Health Systemy
DiscloDisclo
My department receives r
• NIH/NCI
• Varian Medical Systems• Varian Medical Systems
• Philips HealthCare
osureosure
research support from:
ss
Learning Objeg jTo discuss the physics related to luspecial emphasis on small tumor s
To review dose calculation accuracTo review dose calculation accuraclung planning/dose calculation
To present approaches to improveusing simplistic calculation algorit
To review example volumetric arc h i l itechnical issues
ctives/Outlineung tumor dose coverage, with sizes and location
cy with different algorithms forcy with different algorithms for
e plan quality especially when hms for lung cancer planning
therapy plans and discuss
Pre Test Q
1 For the treatment of small lung tu1. For the treatment of small lung tualgorithm is contra-indicated for S
A. 1-D Pencil Beam B. 3-D Pencil Beam C Superposition/ConvolutionC. Superposition/ConvolutionD. Monte Carlo E. Both A and B
Question
umors located peripherally which dosumors located peripherally, which dosSBRT-based treatment planning:
The Primary Issue: Un
omparison of the 100% IDLs, Pen
nderdosage of the PTV
ncil beam (dashed) and MC (solid
Lateral ScatterM C l i l iMonte Carlo simulati
ring of electrons:10 MV il bon, 10 MV pencil beam
Impact of electron sb fpenumbra: conf
Note the differences in doseNote the differences in dosebroadening
PBPB
scattering on beam f l l lformal lung plane gradient due to penumbrale gradient due to penumbral
MCMC
Loss of Charged PartiCPE exists in a volume if each charCPE exists in a volume if each charvolume is replaced by an identical
broad photon field
volume
In narrow field, CPE is lost and dos
cle Equilibrium (CPE)ged particle (electron) leaving theged particle (electron) leaving the electron entering the volume
narrow photon field
volume
se reduction can be severe
Small field central axis de1.0
0.8
ρ=10.5
ρ=1ρ=0.2 ρ=
0.30 4 8 12
“Build down effect” – severescattering of electrons into tDose builds up in the tumor periphery.
epth dose: slab phantom Ion Chamber
MC (DPM)
6x, 2x2 cm
=1
16 20 24
e dose reduction caused by the lung tissue resulting in underdosage at tumor
Implications for
1.0 Ion Cham
MC (DPM)
0.8
( )
6x, 2x2
0 50.5ρ=1
ρ=0.2 ρ=1
0.30 4 8 12 16 20
Depth (cm)
Ring” of underdosage gets larger
Depth (cm)
Ring of underdosage gets largerpproaches the e’ range) and high
“island” tumors
ρ = 1 0
mber
) ρ = 1.0
“Ring” of
)
cm
ρ = 0.2 ρ = 1
Ring of underdosa”rebuildupof dose
24
r for smaller tumors (as sizer for smaller tumors (as size her energies due to larger e- rang
Pencil Beam
The Energy Effect The Energy Effect
Pencil Beam (6 MV)
95%
90% PTV90% PTV
C-S (6 MV)(6 MV)
95%
90%PTV
Pencil Beam
(low vs. high MV)(low vs. high MV)
(18 MV)
95%
90% PTV90% PTV
C-S (18 MV)
PTV
95%90%
The Energy E
PTV DVHs (PB vs. AAA), 6 PB: mean = 70 2 GyPB: mean = 70.2 GyAAA: mean = 68.9 GyDiff i i PTV dDiff. in min. PTV dose
Effect (6 MV)
AAAPB
MV
11%= 11%
The Energy E
PTV DVHs (PB vs. AAA), 18PB 70 5 GPB: mean = 70.5 GyAAA: mean = 64.7 GyDiff. in min. PTV dose =
Effect (18 MV)
AAAPB
8 MV
= 16%
Accuracy of dose calcuAccuracy of dose calcu
PB80%
PTV80%
95%
PTVITV
ulations for lung SBRTulations for lung SBRT
PTV diam. = 3.2 cm
PTV vol. = 14.6 cc
80%PTV
MCPTV
DVH comparisDVH comparisMC plan recomputed uMC plan recomputed uMC plan recomputed uMC plan recomputed u
58
on for the PTV on for the PTV using MUs from PB planusing MUs from PB planusing MUs from PB planusing MUs from PB plan
MC PB
8% 98%
MC PB
PTV mean dose diff. vs. PTVPTV mean dose diff. vs. PTV
20 40
-5.5
20 40
PTV mean dose diff %dose diff.% MC-PB]
15.5-
25.5-
V diam. (mm): 100 patientsV diam. (mm): 100 patients
PTV diam. (mm)
60 80 10060 80 100
MLD (MC) as a function oMLD (MC) as a function o
10
8
MLD (Gy)6
MLD (Gy) MC
4
2
00 2 40 2 4
MLD (G
of MLD (PB): 50 patientsof MLD (PB): 50 patients
y = 0.9093x - 0.0755R2 0 9948R2 = 0.9948
6 8 10 126 8 10 12Gy) PB
Comparative DoseComparative Dose
Purpose: To investigate dobetween treatment plans
• 1-D Pencil beam (PB) – (il b ( li• 3-D Pencil beam – (Eclip
• Anisotropic Analytic Algp y gconvolution-type, Eclipse)• Pinnacle Collapsed Co• Pinnacle – Collapsed Coconvolution-type, Pinnacl• Monte Carlo (iPlan Brai
e Calculation Study e Calculation Study
osimetric differences computed with:
(iPlan BrainLAB) i )pse, Varian)
gorithm (AAA, g ( ,)ne Convolution (CCCne Convolution (CCC, e, Philips) nLAB)
6X, central tumor: M6X, central tumor: M
100
80
60
40
20
0
20
Dose (Gy)0
20 25 30 35
U for the 1DU for the 1D--PB plan PB plan
AAA
1D-PBMC
3D-PBCCC
40 45 50 55
6X, peripheral 6X, peripheral 100100
80
60
40
20
000 10 20
tumor: 12 Gy x 4tumor: 12 Gy x 4
1D-PBAAAMC
CCC
MC
3D-PB
CCC
30 40 50
% difference in PTV% difference in PTV11 l11 l
% diff 1D PB 3D
11 lung ca11 lung ca
% diff.
Dx-Dmc
1D-PB 3D
Dx Dmc
Ave 29.2 2e 29.2 2
STD D 12 1 1STD Dev 12.1 1
Max Diff. 63.2 4
V min. dose vs. MC, V min. dose vs. MC, ll
D PB AAA CCC
ncer plansncer plans
D-PB AAA CCC
21.6 1.5 -2.021.6 1.5 2.0
12 8 4 6 5 112.8 4.6 5.1
48.5 12.7 8.7
What can be done to
Increase the prescription dose
Increase the PTV margin and hIncrease the PTV margin and hfix the lateral scattering probleincreases unintended dose to hincreases unintended dose to h
Use beams with smallest path i ( h 6X hi henergies (choose 6X over highe
o improve the plan?
– non-uniformity in target
hence the field size – will nothence the field size will not em but will help restore CPE -healthy lung tissuehealthy lung tissue
length through lung and low MV)er MV)
What can be done to
N l b b hNon-coplanar beams may be h– electron lateral scattering is c
ddi l badding more co-planar beams
One size doe
Each case must be judged by lo
o improve the plan?
h l f l d di l tihelpful depending on location cylindrically symmetric;
ill i hwill not improve the coverage
es not fit all!
ocation and field size
VMAT (RapidArcc) Example Case
RapidArc (2 partial arcs) IMRT
DVHs for normal lung g
V20 (RA) = 26.4%; MLDV20 (IMRT) = 27 5%; 14V20 (IMRT) = 27.5%; 14
IMIMIMIM
RARARARA
and R peripheral PTVp p
RARARA
IMRT
RA
IMRT
RA
IMRT
RA
D = 14.2 Gy4 7 Gy4.7 Gy
MRTMRTMRTMRT
VMAT vs. IMRT
Target coverage was somewha
Lungs
MLD (G )
Esophagus
M (G )MLD (Gy) Mean (Gy)
RA 14.7 17.5
IMRT 14.2 20.3
Case: Summary
t better with RA vs. IMRT
Heart
M (G )
Cord
M
MU and (Beam-onMean (Gy) Max
(Gy)
(Beam-onTime)
11.3 35.8 475(0.8 min
8.2 36.5 1563 (2 6 min(2.6 min
VMAT and the IDescribes the interaction betwemotionmotion
From Bortfeld et al Physics in MFrom Bortfeld et al. Physics in M
Interplay effect in IMRT is genefor highly fractionated treatme
Interplay Effecteen organ motion and MLC leaf
Medicine and Biology: 47 2002Medicine and Biology: 47, 2002
erally small (~1%) especially ents
M ti lit d 1 3 i S/
VMAT and the I• Motion amplitude: 1.3 cm in S/• Rapid Arc (RA) plan – two 180 d• Interplay effect incorporated an• Interplay effect incorporated an
both RA and IMRT plans
/I 0 2 i R/L 0 4 i A/P
Interplay Effect/I, 0.2 cm in R/L, 0.4 cm in A/Pdeg. Arcs; IMRT plannd compared to “static case” fornd compared to “static case” for
Tumor Motion: ExhaINH (Axial
INH (Cor)
ale and Inhale StatesEXH (Axial
EXH (Cor)(Co )
DVHS (PTV and normal lung)IMRTIMRT
PTV min. dose diff. =
) with and without interplayT planT plan
with interplay18% with interplaywith interplayp yp ywith interplay
DVHS (PTV and normal lung)R idARapidA
) with and without interplayA lArc plan
with interplay
Overall S
Simulation, planning and deliveconfounded by motion and hetconfounded by motion and het
Lung tissue density impacts tug y psignificantly requiring accurat
umor size is important – pay spwith field sizes close to the elecwith field si es close to the elec
Location! Locat
Convolution/superposition or Mused for lung cancer treatment
Summary
ery of lung cancer is terogeneity in tissue densityterogeneity in tissue density
umor dose deposition pte dose algorithms
pecial attention to small tumoctron rangectron range
tion! Location!
MC-based methods should be t planning
Overall S
High energy photon beams (>
VMAT/R idA ff ffiVMAT/RapidArc offers an efficparticularly in the context of S
Interplay effects tend to be smp yor VMAT, however, further invunderstand these effects in thunderstand these effects in th
Summary
> 6 MV) should be avoided
i t l ti t RT d licient solution to RT delivery, SBRT
mall in the context of IMRT vestigation is needed to fully he VMAT settinghe VMAT setting
AcknowleColleagues
Colleagues at Un
Assistance fr
Colleagues at Un
BrainLab (FeldkirPhilips Radiation OPhilips Radiation O
Varian Medi
NIH/NCI Gra
Dr. Akila Viswanathan (Chair) anCristin Watson and o
dgementsat HFHS
iv of Michigan
rom Industry
iv of Michigan
yrchen, Germany)Oncology SystemsOncology Systemsical Systems
ant Support
d Dr. Ramesh Rengan (Co-Chair)other ASTRO HQ staff
Post Test
1 For the treatment of small lung tu1. For the treatment of small lung tualgorithm is contra-indicated for S
A. 1-D Pencil Beam B. 3-D Pencil Beam C Superposition/ConvolutionC. Superposition/ConvolutionD. Monte Carlo E. Both A and B
Question
umors located peripherally which dosumors located peripherally, which dosSBRT-based treatment planning:
Test Questi
E: Under small field conditions, and lower den,dominant factor influencing dose to the tumor. Peaccount for electron scattering either implicitly inaccurate under such circumstances. Thereforecontext of SBRT planningcontext of SBRT planning. References: S. Benedict, K. Yenice, D. Followil, et al. " Stereo, , ,Task Group 101", Med Phys 37, 4078-4101 (2011
I. Das, G. Ding, A. Ahnesjo, " Small fields: Non206 215 (2008)206-215 (2008).
ion Answer
nsity lung tissue, electron scattering becomes thy g , gencil beam algorithms (1-D or 3-D) do not properor explicitly and have been shown to be quit
e pencil beam algorithms are contra-indicated in th
otactic body radiation therapy: The report of AAPMy py p1).
equilibrium radiation dosimetry", Med Phys 35 (1