Quick start guide
Proximal pressure line
Exhalationvalve line
1
2
3
4
Overview
Trilogy202
Circuit setup
Setup screen
Prescription setup
Overview1
Remotealarmconnector
Serial connector – to connect to a computer or Alice polysomnograph
Ethernet connector
Externalbatteryconnector
Air inlet and grey foam filter
Universalporting block
Detachablebattery cover
Cancel action or exit menu
Navigate menus or change settings
Select/modify/confirm parameters
Audio pause button
Start/stop button
AC power inlet
SD card to download ventilation data7
6
5
4
3
2
1
Full access mode
SD card inserted in the ventilator
Audio pause active
Ramp active
Status of optional detachable battery
Status of internal battery
Status of external battery
Battery is charging
Symbols and icons
71 2 3
5
6
4
Passive exhalation port circuit Active exhalation valve flow sensor circuit
Circuit setup2
From the Setup screen select "Passive" for circuit type.
Before operating the Trilogy202, you need to configure the circuit type.
Active exhalation valve with proximal airway pressure circuit
From the Setup screen select "Active PAP" for circuit type. From the Setup screen select "Active flow" for circuit type.
High-pressure oxygen inlet fitting
Oxygen blender module
Flow line
Exhalationvalve line
Tracheostomytubing
Exhalation valve
Mask withoutintegrated leak
Bacterial/viral filter(optional)
Flow sensorWhite stripedflow line
Flexible tubing
Tracheostomytube
Non-vented mask
Bacterial/viral filter
Bacterial/viral filter
Tracheostomytube
Flexible tubing
Non-vented mask
Exhalation valveProximalpressure port
Proximalpressure line
Exhalation valve line
Flexible tubing
Tracheostomytube
Non-vented mask
Flow line
Bacterial/viral filter
Flow lineExhalation valve line
White striped flow line
Porting block
Flexible tubing
Exhalation valve line
Exhalation valve
Flex tube
White stripedflow line
Flex tube
Flow sensor
Elbow
Elbow
Passive exhalation port
Adapter(PN 1061398) required for reusable circuits
1068642_20100720.indd 1-51068642_20100720.indd 1-5 8/3/2010 8:45:12 AM8/3/2010 8:45:12 AM
Exit
Settings And Alarms
Options
Alarm log
Event log
Information
Menu 1/10
SelectNavigate
Cancel
Dual Prescription OFCircuit Type Active PAPMode S/T
Menu Settings and Alarms 2/23
OKEdit
Make sure the Trilogy 202 is turned off before taking the following steps
Select Settings And Alarms
Use to choose Active PAP, Active flow, or Passive and then press twice to confirm
B
D
Exit
Dual Prescription OFF
Circuit Type Passive
Mode S/T
Menu Settings and Alarms 2/23
ModifyNavigate
A
C
Press and hold and for a few seconds to access the Setup screen
Navigate to Circuit Type and press
Setup screen3Configuring the circuit type (active or passive)
E
Press to begin therapy or continue navigating through the menu to set up the ventilation parameters
Press to enter Menu AccessNavigate to Options and press
To enable full access to the settings
B
D
Use to choose Full or Limited Access and press to confirm
C
Press andto return to Initial Screen
Avsluta
Inställningar och larmAlternativLarmloggHändelseloggInformation
VäljNavigera Finish
Menu Access Limited
Detailed View OFF
Language English
Menu Options 1/18
ModifyNavigate
Cancel
Menu Access Full
Detailed View OFF
Language English
Menu Options 1/18
OKEdit Finish
Menu Access Full
Detailed View OFF
Language English
Menu Options 2/18
ModifyNavigate
Setting menu access (May be performed while on or off the patient)
Exit
Settings And Alarms
Options
Alarm log
SelectNavigate
Select Settings And Alarms
Prescription setup4
B
Menu
A
From main screen, press Menu
Dual Prescription
If Dual Prescription is enabled, both prescriptions need to be adjusted as previously explained
Dual Prescription ON
Mode S/T
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
Menu Settings And Alarms 1/20
Navigate ModifyFinish
Set both primary and secondary prescriptions
Exit
Switch to Secondary Settings
Safely Remove SD Card
Primary Settings and Alarms
Secondary Settings and Alarms
Options
Menu 3/8
SelectNavigate
RR 20BPM
PIP 5.9cm H2O
Vte 435ml
Leak 21l/min
Peak Flow 67.9l/min
MinVent 8.7l/min
Cancel
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
OKEdit
Press to validate new setting
F
Cancel
AVAPS ON
Tidal Volume 400 ml
IPAP Max Pressure 25.0 cmH20
OKEdit
E
Use to change the value or status of setting
Finish
AVAPS OFFIPAP 30.0 cmH20EPAP 8.0 cmH20
ModifyNavigate Finish
Mode S/TAVAPS OFFIPAP 30.0 cmH20EPAP 8.0 cmH20
ModifyNavigate
Press to change the parameter highlighted in blue
C D
Scroll up and down using to navigate through the settings and alarms
Modus S/TAVAPS EINAtemzugvolumen 400 mlIPAP Max Druck 25.0 cmH2O
Navigieren ÄndernFertig
G
Press andto return to the main screen
A
1068642JH 7/20/10