Quick start guide Proximal pressure line Exhalation valve line 1 2 3 4 Overview Trilogy202 Circuit setup Setup screen Prescription setup Overview 1 Remote alarm connector Serial connector – to connect to a computer or Alice polysomnograph Ethernet connector External battery connector Air inlet and grey foam filter Universal porting block Detachable battery cover Cancel action or exit menu Navigate menus or change settings Select/modify/confirm parameters Audio pause button Start/stop button AC power inlet SD card to download ventilation data 7 6 5 4 3 2 1 Full access mode SD card inserted in the ventilator Audio pause active Ramp active Status of optional detachable battery Status of internal battery Status of external battery Battery is charging Symbols and icons 7 1 2 3 5 6 4 Passive exhalation port circuit Active exhalation valve flow sensor circuit Circuit setup 2 From the Setup screen select "Passive" for circuit type. Before operating the Trilogy202, you need to configure the circuit type. Active exhalation valve with proximal airway pressure circuit From the Setup screen select "Active PAP" for circuit type. From the Setup screen select "Active flow" for circuit type. High-pressure oxygen inlet fitting Oxygen blender module Tracheostomy tube Non-vented mask Bacterial/viral filter Bacterial/viral filter Tracheostomy tube Flexible tubing Non-vented mask Exhalation valve Proximal pressure port Proximal pressure line Exhalation valve line Flexible tubing Tracheostomy tube Non-vented mask Flow line Bacterial/viral filter Flow line Exhalation valve line White striped flow line Porting block Flexible tubing Exhalation valve line Exhalation valve Flex tube White striped flow line Flex tube Flow sensor Elbow Elbow Passive exhalation port Adapter (PN 1061398) required for reusable circuits 1068642_20100720.indd 1-5 1068642_20100720.indd 1-5 8/3/2010 8:45:12 AM 8/3/2010 8:45:12 AM

Upload: others

Post on 26-May-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Quick start guide

Proximal pressure line

Exhalationvalve line

1

2

3

4

Overview

Trilogy202

Circuit setup

Setup screen

Prescription setup

Overview1

Remotealarmconnector

Serial connector – to connect to a computer or Alice polysomnograph

Ethernet connector

Externalbatteryconnector

Air inlet and grey foam filter

Universalporting block

Detachablebattery cover

Cancel action or exit menu

Navigate menus or change settings

Select/modify/confirm parameters

Audio pause button

Start/stop button

AC power inlet

SD card to download ventilation data7

6

5

4

3

2

1

Full access mode

SD card inserted in the ventilator

Audio pause active

Ramp active

Status of optional detachable battery

Status of internal battery

Status of external battery

Battery is charging

Symbols and icons

71 2 3

5

6

4

Passive exhalation port circuit Active exhalation valve flow sensor circuit

Circuit setup2

From the Setup screen select "Passive" for circuit type.

Before operating the Trilogy202, you need to configure the circuit type.

Active exhalation valve with proximal airway pressure circuit

From the Setup screen select "Active PAP" for circuit type. From the Setup screen select "Active flow" for circuit type.

High-pressure oxygen inlet fitting

Oxygen blender module

Flow line

Exhalationvalve line

Tracheostomytubing

Exhalation valve

Mask withoutintegrated leak

Bacterial/viral filter(optional)

Flow sensorWhite stripedflow line

Flexible tubing

Tracheostomytube

Non-vented mask

Bacterial/viral filter

Bacterial/viral filter

Tracheostomytube

Flexible tubing

Non-vented mask

Exhalation valveProximalpressure port

Proximalpressure line

Exhalation valve line

Flexible tubing

Tracheostomytube

Non-vented mask

Flow line

Bacterial/viral filter

Flow lineExhalation valve line

White striped flow line

Porting block

Flexible tubing

Exhalation valve line

Exhalation valve

Flex tube

White stripedflow line

Flex tube

Flow sensor

Elbow

Elbow

Passive exhalation port

Adapter(PN 1061398) required for reusable circuits

1068642_20100720.indd 1-51068642_20100720.indd 1-5 8/3/2010 8:45:12 AM8/3/2010 8:45:12 AM

Exit

Settings And Alarms

Options

Alarm log

Event log

Information

Menu 1/10

SelectNavigate

Cancel

Dual Prescription OFCircuit Type Active PAPMode S/T

Menu Settings and Alarms 2/23

OKEdit

Make sure the Trilogy 202 is turned off before taking the following steps

Select Settings And Alarms

Use to choose Active PAP, Active flow, or Passive and then press twice to confirm

B

D

Exit

Dual Prescription OFF

Circuit Type Passive

Mode S/T

Menu Settings and Alarms 2/23

ModifyNavigate

A

C

Press and hold and for a few seconds to access the Setup screen

Navigate to Circuit Type and press

Setup screen3Configuring the circuit type (active or passive)

E

Press to begin therapy or continue navigating through the menu to set up the ventilation parameters

Press to enter Menu AccessNavigate to Options and press

To enable full access to the settings

B

D

Use to choose Full or Limited Access and press to confirm

C

Press andto return to Initial Screen

Avsluta

Inställningar och larmAlternativLarmloggHändelseloggInformation

VäljNavigera Finish

Menu Access Limited

Detailed View OFF

Language English

Menu Options 1/18

ModifyNavigate

Cancel

Menu Access Full

Detailed View OFF

Language English

Menu Options 1/18

OKEdit Finish

Menu Access Full

Detailed View OFF

Language English

Menu Options 2/18

ModifyNavigate

Setting menu access (May be performed while on or off the patient)

Exit

Settings And Alarms

Options

Alarm log

SelectNavigate

Select Settings And Alarms

Prescription setup4

B

Menu

A

From main screen, press Menu

Dual Prescription

If Dual Prescription is enabled, both prescriptions need to be adjusted as previously explained

Dual Prescription ON

Mode S/T

AVAPS ON

Tidal Volume 400 ml

IPAP Max Pressure 25.0 cmH20

Menu Settings And Alarms 1/20

Navigate ModifyFinish

Set both primary and secondary prescriptions

Exit

Switch to Secondary Settings

Safely Remove SD Card

Primary Settings and Alarms

Secondary Settings and Alarms

Options

Menu 3/8

SelectNavigate

RR 20BPM

PIP 5.9cm H2O

Vte 435ml

Leak 21l/min

Peak Flow 67.9l/min

MinVent 8.7l/min

Cancel

AVAPS ON

Tidal Volume 400 ml

IPAP Max Pressure 25.0 cmH20

OKEdit

Press to validate new setting

F

Cancel

AVAPS ON

Tidal Volume 400 ml

IPAP Max Pressure 25.0 cmH20

OKEdit

E

Use to change the value or status of setting

Finish

AVAPS OFFIPAP 30.0 cmH20EPAP 8.0 cmH20

ModifyNavigate Finish

Mode S/TAVAPS OFFIPAP 30.0 cmH20EPAP 8.0 cmH20

ModifyNavigate

Press to change the parameter highlighted in blue

C D

Scroll up and down using to navigate through the settings and alarms

Modus S/TAVAPS EINAtemzugvolumen 400 mlIPAP Max Druck 25.0 cmH2O

Navigieren ÄndernFertig

G

Press andto return to the main screen

A