Page 1: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Operating InstructionsBedienungsanleitung Manuel utilisateur

Type 8202

pH or redox transmitter

pH- oder Redox-Transmitter

Transmetteur de pH ou de potentiel d‘oxydo-réduction

Page 2: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

We reserve the right to make technical changes without notice.Technische Änderungen vorbehalten.Sous réserve de modifications techniques.

© 2008 Bürkert SAS

Operating Instructions july2008/0_EU-ML_560329

Page 3: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:
Page 4: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:
Page 5: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

pH or redox transmitter Type 8202

Contents:

About this mAnuAl1. .................................................................................6

symbols used1.1. ..........................................................................................6

intended use2. ................................................................................................7

Restraints2.1. ...................................................................................................7

Foreseeable misuse2.2. .............................................................................7

bAsic sAFety inFoRmAtion3. ...............................................................8

GeneRAl inFoRmAtion4. ...........................................................................9

contents of the delivery4.1. .....................................................................9

Warranty conditions4.2. ..............................................................................9

information on the internet4.3. ............................................................ 10

descRiption5. ................................................................................................10

Area of application5.1. ............................................................................. 10

General description5.2. ........................................................................... 10

Design5.2.1. .............................................................................10

pHorRedoxsensor5.2.2. .....................................................11

description of the label5.3. ................................................................... 12

Versions available5.4. ............................................................................... 12

technicAl dAtA6. ........................................................................................13

conditions of use6.1. ................................................................................ 13

conformity to standards and directives6.2. .................................. 13

General technical data6.3. ..................................................................... 14

Mechanicaldata6.3.1. ............................................................14

Generaldata6.3.2. ..................................................................16

Electricaldata6.3.3. ................................................................17

Electricalconnections6.3.4. .................................................17

pH/Redoxsensor6.3.5. .......................................................18

Assembly7. .......................................................................................................21

safety instructions7.1. .............................................................................. 21

unscrew the cover7.2. ............................................................................. 22

Fitting the cover7.3. ...................................................................................22

Fitting the display module7.4. .............................................................. 23

Removing the display module7.5. ..................................................... 23

Fitting the sensor into the armature (without fluid)7.6. ......... 24

Fitting the electronic module to the armature7.7. .................... 24

3

Type 8202

english

Page 6: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

instAllAtion And WiRinG8. ................................................................. 25

safety instructions8.1. .............................................................................. 25

installation onto the pipe8.2. ................................................................ 26

electrical wiring8.3. ....................................................................................27

Assemblingthemaleorfemaleconnector8.3.1. (accessories:seeChap.12)........................................ 27

WiringaversionwithoneM12fixedconnector8.3.2. ...28

WiringaversionwithtwoM12fixedconnectors8.3.3. .30

commissioninG9. ........................................................................................33

safety instructions9.1. .............................................................................. 33

Adjustment And Functions10. ....................................................................................................................................................................34

safety instructions10.1. ........................................................................... 34

Functions10.2. ...............................................................................................34

using the browse button10.3. .............................................................. 36

description of the display10.4. ............................................................ 38

Descriptionoftheiconsandleds10.4.1. .............................38

Whenswitchingonthedevice10.4.2. ..................................39

Whenbrowsinginthemenus10.4.3. ....................................39

Read mode and access to the menus mode10.5. ..................... 40

Functions available per menu10.6. .................................................... 41

parameters menu ("param")10.7. ....................................................... 42

Setthedataparametersindependentlyofthe10.7.1. measurements("System"menu)...............................42

Setthedisplayparameters("Display"menu)10.7.2. ........43

Outputconnectionmode10.7.3. (menu"Outputs"-Function"HWMode")...............44

Settheparametersforthecurrentoutputs10.7.4. (menu"Outputs"-Functions"AC1"and"AC2")..45

Settheparametersforthetransistoroutputs10.7.5. (menu"Outputs"-functions"TR1"and"TR2")....47

Setthesensorparameters("Sensor"menu)10.7.6. .........49

calibration menu ("calib")10.8. ........................................................... 50

Activate/deactivatetheHoldfunctionandmodify10.8.1. theaccesscodetotheCalibrationmenu ("System"menu)...........................................................50

Adjustthecurrentoutputs("Outputs"menu)10.8.2. ........51

Calibratethesensor10.8.3. (menuSensor-function"Probe")............................51

Inputanoffsetforthetemperaturemeasurement10.8.4. (menu"Sensor"-function"Temperature")............56

diagnostic menu ("diagnostic")10.9. ................................................ 56

ModifytheDiagnosticmenuaccesscode10.9.1. ("System"menu)...........................................................56

Monitorthesensor(menu"Sensor"-Functions10.9.2. "Glasselectrode"and"Refelectrode")...................57

4

english

Type 8202

Page 7: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Monitorthefluidtemperature10.9.3. (menu"Sensor"-Function"Temperature")............59

test menu ("test")10.10. ....................................................................... 61

ModifytheTestmenuaccesscode10.10.1. ("System"menu)...........................................................61

Checkthattheoutputsareworkingcorrectly10.10.2. ("Outputs"menu)..........................................................61

Checkthattheoutputsarebehavingcorrectly10.10.3. ("Sensor"menu)...........................................................62

information menu ("info")10.11. ........................................................ 62

mAintenAnce And tRoubleshootinG11. ................................ 63

safety instructions11.1. ........................................................................... 63

cleaning of the transmitter11.2. ......................................................... 63

MaintenanceofthepH/redoxsensor11.2.1. ......................63

Replacing the sensor11.3. ...................................................................... 65

if you encounter problems11.4. .......................................................... 66

spARe pARts And AccessoRies12. ............................................... 72

pAckAGinG, tRAnspoRt13. ................................................................... 73

stoRAGe14. .......................................................................................................73

disposAl oF the pRoduct15. ........................................................... 74

5

english

Type 8202

Page 8: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

AbouT This mAnuAl1. Thismanualdescribestheentirelifecycleofthedevice.Pleasekeepthismanualinasafeplace,accessibletoallusersandanynewown-ers.

this manual contains important safety information.

Failure to complywith these instructions can lead to hazardoussituations.Thismanualmustbereadandunderstood.•

symbols used1.1.

danger

Warns you against an imminent danger.Failuretoobservethiswarningcanresultindeathorinserious•injury.

Warning

Warns you against a potentially dangerous situation.Failuretoobservethiswarningcanresultinseriousinjuryoreven•death.

CauTion

Warns you against a possible risk.Failuretoobservethiswarningcanresultinsubstantialorminor•injuries.

noTe

Warns you against material damage.Failuretoobservethiswarningcanresultindamagetothedevice•orsystem.

indicatesadditionalinformation,adviceorimportantrecom-mendationsforyoursafetyandforthecorrectoperationofthedevice.

referstoinformationcontainedinthismanualorinotherdocuments.

indicatesaproceduretobecarriedout.→

6

Aboutthismanual

english

Type 8202

Page 9: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

inTended use2.

use of the ph/Redox transmitter that does not comply with the instructions could present risks to people, nearby installations and the environment.The8202transmitterisintendedsolelyforthemeasurementof:•

thepHincleanliquidsorliquidscontainingsolids,sulphides-orproteins.

ortheoxidationreductionpotentialincleanliquidsorliquids-containingsolids,sulphidesorproteinswhichmaypresentlowconductivity.

Thisdevicemustbeprotectedagainstelectromagneticinterfer-•ence,ultravioletraysand,wheninstalledoutdoors,theeffectsoftheclimaticconditions.Thisdevicemustbeusedincompliancewiththecharacteristics•andcommissioninganduseconditionsspecifiedinthecontractualdocumentsandintheusermanual.Requirementsforsafeandproperoperationarepropertransport,•storageandinstallationaswellascarefuloperationandmainte-nance.Onlyusethedeviceasintended.•

Restraints2.1.

Observeanyexistingrestraintswhenthedeviceisexported.

Foreseeable misuse2.2.

The8202 transmittermustnotbeused inapotentiallyexplosive•atmosphere.

Donotusefluidthatisincompatiblewiththematerialsofwhichthe•transmitterismade.

Donotsubjectthedevicetomechanicalloads(e.g.byplacingobjects•ontopofitorbyusingitasastep).

Donotmakeanyexternalmodificationstothedevice.Donotpaint•orvarnishanypartofthedevice.

7

Intendeduse

english

Type 8202

Page 10: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

bAsic sAFeTy inFoRmATion3. Thissafetyinformationdoesnottakeintoaccount:

anycontingenciesoroccurencesthatmayariseduringassembly,•useandmaintenanceofthedevices.

thelocalsafetyregulationsthattheoperatormustensurethestaff•inchargeofassemblyobserve.

danger due to high pressureCutoffthepressureanddepressurizethepipesbeforeloosening•thepipesandfittings.

danger due to electrical voltageBeforestartingwork,makesure thatyouswitchoff thesupply•voltageandsecureittopreventrestarting.Observeallapplicableaccidentprotectionandsafetyguidelines•forelectricalequipment.

danger due to high temperatures of the fluidUsesafetyglovestohandlethedevice.•Stopthecirculationoffluidandpurgethepipebeforeloosening•thepipesandconnections.Keepalleasilyflammablefabricsandfluidsawayfromthedevice.•Donotinsulatethedevice.

danger due to the nature of the fluid.Respect theprevailing ruleson accidentprevention and safety•relatingtotheuseofaggressivefluids.

Various dangerous situations

Toavoidinjurytakecareto:preventanypowersupplyswitchon.•carryouttheinstallationandmaintenanceworkbyqualifiedand•skilledstaffwiththeappropriatetools.guaranteeasetorcontrolled restartingof theprocess,aftera•powersupplyinterruption.usethedeviceonlyifinperfectworkingorderandincompliance•withtheinstructionsprovidedintheusermanual.observethegeneraltechnicalrulesduringtheplanninganduse•ofthedevice.

noTe

chemical compatibility of materials in contact with the fluid.Systematicallycheckthechemicalcompatibilityofthecomponent•materialsofthetransmitterandtheproductslikelytocomeintocontactwithit(forexample:alcohols,strongorconcentratedac-ids,aldehydes,alkalinecompounds,esters,aliphaticcompounds,ketones,halogenatedaromaticsorhydrocarbons,oxidantsandchlorinatedagents).

8

Basicsafetyinformation

english

Type 8202

Page 11: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

noTe

elements / components sensitive to electrostatic dischargesThisdevicecontainselectroniccomponentssensitivetoelectro-•staticdischarges.Theycangetdamagediftheyaretouchedbyanelectrostaticallychargedpersonorobject.Intheworstcasescenario,thesecomponentsareinstantlydestroyedorgooutoforderassoonastheyareactivated.Tominimiseorevenavoidalldamageduetoanelectrostaticdis-•charge,takealltheprecautionsdescribedintheEN100015-1norm.Alsoensurethatyoudonottouchanyoftheliveelectricalcom-•ponents.

Thetransmittertype8202wasdevelopedwithdueconsid-erationgiventoacceptedsafetyrulesandisstate-of-the-art.However,risksmayarise.

Failure toobserve these instructionsaswellasanyunauthorisedworkonthedeviceexcludesusfromanyliabilityandalsonullifiesthewarrantywhichcoversthedeviceanditsaccessories.

GeneRAl inFoRmATion4.

contents of the delivery4.1.

Whenyoureceivethemerchandise,makesurethatthecontentsofthedeliveryhavenotbeendamagedinanywayandensurethattheycorrespondexactlywiththedeliverynoteorpackinglist.

Ifthisisnotthecase,contactusimmediately.

Theaddressesofourinternationalbranchescanbefoundonthelastpagesofthismanual.

Alsoontheinternet,at:

www.burkert.com Bürkert Company Locations

Warranty conditions4.2.

Thisdocumentdoesnotcontainanywarrantyconditions.Tofindoutthisinformation,pleaserefertoourgeneralsalesterms.Thecondi-tiontobenefitfromthewarrantyistheconforminguseofthe8202inobservanceofthespecifiedconditionsofoperating.

Thewarrantyonlyextendstodefectsinthetransmittertype8202anditscomponents.

Weacceptnoliabilityforanykindofconsequentialdamagewhichmayoccurduetothefailureorthemalfunctioningofthedevice..

9

Generalinformation

english

Type 8202

Page 12: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

information on the internet4.3.

Youcanfindtheusermanualsandtechnicaldatasheetsregardingthetype8202at:

www.burkert.com Documentation DatasheetsorManuals/Ap-provals Type8202

Thecompletemanualcanbeorderedundercode:560329

descRipTion5.

Area of application5.1.

The8202transmitterisintendedsolelyforthemeasurementof:

thepHincleanliquidsorliquidscontainingsolids,sulphidesor•proteins.

ortheoxidationreductionpotentialincleanliquidsorliquids•containingsolids,sulphidesorproteinswhichmaypresentlowconductivity.

Thanksto2fullyprogrammabletransistoroutputs,thetransmittercanbeusedtoswitchasolenoidvalve,activateanalarmand,thanksto1or24-20mAcurrentoutputs,establishoneortwocontrolloops.

General description5.2.

design5.2.1.

The8202transmittercomprises:

amoduleformeasuringphysicalparameters,comprising:•

apHorRedoxsensormeasuringapotentialdifference(PD)in-mV

aPt1000temperaturesensorintegratedinthearmatureofthe-pHorRedoxsensormeasuringaresistance.

10

Description

english

Type 8202

Page 13: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

anacquisition/conversionmoduleforthephysicalparameters•measured:

PDacquisitionmeasuredinmV-

conversionofthemeasuredPDintopHunits(foratransmitter-withpHsensoronly)

acquisitionoftheresistancemeasuredandconversioninto-temperature

adisplaymodulewithbrowsebuttonusedtoreadand/orconfig-•uretheparametersofthedevice.Thedisplaymoduleisavailableasanaccessory.

Oneversionofthe8202transmitterwith2transistoroutputsanda4-20mAoutletoperatesona2-wiresystemandrequiresapowersupplyof14-36VDC.Forsuchaversion,electricalconnectionisdoneviaanM12,5-point,malefixedconnector.

Oneversionofthe8202transmitterwith2transistoroutputsand24-20mAoutletsoperatesona3-wiresystemandrequiresapowersupplyof12-36VDC.Forsuchaversion,electricalconnec-tionisdoneviaanM12,5-point,malefixedconnectorandanM12,5-point,femalefixedconnector.

ph or Redox sensor5.2.2.

The8202transmittercanbefittedwithastandardsensor120mmlong,measuringthepHortheoxidationreductionpotential.

Itisscrewedintoanarmaturewiththebuilt-inPt1000temperaturesensor.

ThepHorRedoxsensorisaglassmembranewithvariableselectiv-ityaccordingtothepHortheRedox:

WhenthepHsensorisimmersedinasolution,adifferenceinpo-•tentialisformed,duetothehydrogenions(H+),betweentheglassmembraneandthesolution.Thisdifferenceinpotential,measuredinrelationtoareferenceelectrode,isdirectlyproportionaltothepHvalue(59.16mVperpHunitat25°C).

WhenaRedoxsensorisimmersedinasolution,anexchangeof•electronsoccursbetweentheoxidisedformandthereducedformofanelectrolyte.Theresultingvoltageistheoxidationreductionpotential.

11

Description

english

Type 8202

Page 14: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

description of the label5.3.

Fig. 5.1 Label on the 8202 transmitter

Typeoftransmitterandparametermeasured1.Electricalpowersupply2.Outputspecifications3.Sensorspecifications4.Temperaturerangeofthefluid,withoutsensor5.Nominalpressureofthefluid,withoutsensor6.AllocationofthepinsontheM12fixedconnectors7.Manufacturercode8.Orderreference9.Serialnumber10.Protectionrating11.

Versions available5.4.

Thefollowingversionsofthe8202pHtransmitterareavailable.ThesereferencesincludetheelectronicmoduleandthesensorarmatureincludingthePt1000.

ThepH/Redoxsensor,thedisplaymoduleandthefittingforcon-nectiontotheprocessshouldbeorderedseparately.ForthepH/Redoxsensorandthedisplaymodule,seethelistofaccessoriesinChapter12.

sup

ply

volt-

age

out

puts

sen

sor

ele

ctri

cal c

on-

nect

ion

uni

on n

ut

mat

eria

l

ord

er c

ode

14-36VDC 2transistors+1x4-20mA

with-out

M12,5-pinmalefixedconnector

PVC 559630PVDF 559632

12-36VDC 2transistors+2x4-20mA

with-out

M12,5-pinmalefixedconnectorandM12,5-pinfemalefixedcon-nector

PVC 559631PVDF 559633

12

Description

english

Type 8202

Page 15: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Pressure:complyingwitharticle3of§3from97/23/CEdirective.•Acc.tothe97/23/CEpressuredirective,thedevicecanonlybeusedinthefollowingcases(dependingonmax.pressure,pipediameterandfluid):

type of fluid conditionsFluidgroup1,§1.3.a onlyDN25Fluidgroup2,§1.3.a DN ≤ 32

orDN>32andPNxDN ≤ 1000Fluidgroup1,§1.3.b DN ≤ 25

orDN>25andPNxDN ≤ 2000Fluidgroup2,§1.3.b DN ≤ 125

TechnicAl dATA6.

conditions of use6.1.

Ambienttemperature: -10to+60°C(operatingandstoragewithoutthepH/Redoxsensor)

Airhumidity: <85%,noncondensated

Protectionrating: IP67withconnectorspluggedinandtightenedand electronicmodule cover fully screweddown

conformity to standards and 6.2. directives

EMC:EN61000-6-2,EN61000-6-3•

Vibration:EN60068-2-6•

Shock:EN60068-2-27•

13

Technicaldata

english

Type 8202

Page 16: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

General technical data6.3.

mechanical data6.3.1.

component materialBox/seals stainlesssteel,PPS/EPDM

Cover/seal PC/EPDM

Displaymodule PC/PBT

M12fixedcon-nector

nickel-platedbrass

Fixedconnectorbase

stainlesssteel1.4404(316L)

Screw stainlesssteel

Tighteningnut PVCorPVDF

Sensorarmature/seal

PVDF,stainlesssteel1.4571(316Ti)/EPDM

pHorRedoxsen-sor

refertothecorrespondingmanual

Materialsincontactwiththefluid:•PVDF,stainlesssteel1.4571(316Ti).Alsorefertothesensorusermanual,particularlyforsensorsotherthanthosemanufac-turedbyBürkert.

PC

EPDM

PPS

EPDM

PPS

PVDF

PVCorPVDF

EPDM

PVDF

Stainlesssteel316Ti

Stainlesssteel

Nickel-platedbrass

pH/Redoxsensor

Fig. 6.1 Materials used in the 8202 transmitter

14

Technicaldata

english

Type 8202

Page 17: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Fig. 6.2 Dimensions of the 8202 transmitter [mm]

A:applicationrangeofa8202withaPVDFnut•

B:applicationrangeofa8202withaPVCnut•

Themeasureshavebeenmadeatanambienttemperatureof60°C

Fig. 6.3 Fluid temperature and pressure dependency (8202, without sensor)

15

Technicaldata

english

Type 8202

Page 18: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

General data6.3.2.

Diameterofthepipes DN25toDN125(DN15toDN20dependingonconditions)

Typeoffitting PVCorPVDF;

32equal90°Tee(pipeDN25)•

d-32-dreducing90°Tee(pipe• ≥ 32)

25-32-20-25reducingbushes•(pipeDN15)or32-25reducingbushes(pipeDN20),mountedona32equal90°Teefitting

Unionnutforconnectiontothefitting

G11/2''internalthread

Fluidpressure PN16(seefig.6.3);fluidpressuremaybelimitedbythesensorused:seecorrespondinginstructionmanual

Fluidtemperature dependsonusedsensor(seecorre-spondinginstructionmanualandfig.6.3)

ph measurement

Measurementrange•

Resolution•

Measurementerror•

Recommendedmin.•divergenceofthepHrangecorrespondingtothe4-20mAsignal

-2to16pHor-580to+580mV•

0.001pHor0.1mV•

±0.02pHor0.5mV•

0.5pHunitor30mV•(eg:range6.7to7.2pHor-20mVto+10mVcorrespondstothe4-20mAoutputcurrent)

Redox potential meas-urement

Measurementrange•

Resolution•

Measurementerror•

Recommendedmin.•divergenceoftheredoxrangecorrespondingtothe4-20mAsignal

-2000mVto+2000mV•

1mV•

±3mV•

50mV•(eg:range1550to1600mVcor-respondstothe4-20mAoutputcurrent)

Temperaturesensor Pt1000integratedinthesensorar-mature

temperature measure-ment

Measurementrange•

Resolution•

Measurementerror•

Recommendedmin.di-•vergenceofthetempera-turerangecorrespondingtothe4-20mAsignal

-40°Cto+130°C•

0.1°C•

±1°C•

10°C•(eg:range10to20°Ccorre-spondstothe4-20mAoutputcurrent)

Temperaturecompensation automatic(integratedPt1000)Referencetemperature=25°C

16

Technicaldata

english

Type 8202

Page 19: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Currentoutput

Responsetime(10%•-90%)

versionwith1current•output

versionwith2current•outputs

4-20mA,sinkorsource(dependingonprogramming)

150ms(bydefault)•

max.loopimpedance:1100• Wat36VDC,610 Wat24VDC,180 Wat14VDC

max.loopimpedance:1100• Wat36VDC,610 Wat24VDC,100 Wat12VDC

electrical connections6.3.4.

number of fixed con-nectors

type of cable plug

1M12malefixedcon-nector

femaleM12,5-pin(notsupplied)

1M12malefixedcon-nector+1M12femalefixedconnector

femaleM12,5-pin(notsupplied)+maleM12,5-pin(notsupplied)

Typeofcablerecom-mended

shielded

electrical data6.3.3.

Powersupply

versionwith3outputs•

versionwith4outputs•

14-36VDC,filteredandregulated•

12-36VDC,filteredandregulated•Consumption,withchargesontransistors

Powerconsumption(withoutcharge):

versionwith3outputs•

versionwith4outputs•

≤2A

≤ • 25mA(at14VDC)

≤ • 5mA(at12VDC)

Protectionagainstpolar-ityreversal

yes

Protectionagainstspikevoltages

yes

Protectionagainstshortcircuits

yes,forthetransistoroutputs

Transistoroutput NPN(/sink)orPNP(/source)(depend-ingonprogramming),collectoropen,700mAmax.NPNoutput:0.2-36VDCPNPoutlet:supplyvoltage

17

Technicaldata

english

Type 8202

Page 20: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

ph / Redox sensor6.3.5.

Sensor pHorRedoxsensor,combined,120mmlongwithheadPG13.5,withouttemperaturesensor

loGotRode ph

Typeoffluid•

Measurementrange•

Fluidpressure•

Fluidtemperature•

Ambienttemperature•

Minimumconductivity•

Max.pressureatmax.•temperature

Numberandtypeofdia-•phragms

Referenceelectrolyte•

clean•

2to14pH•

0to6bar•

-10to+60°C(seefig.6.3)•

0to+60°C(whenrunning)•+4to+30°C(storage)

2µS/cm•

6bar•

1"singlepore• TM"

polymer•

unitRode plus ph

Typeoffluid•

Measurementrange•

Fluidpressure•

Fluidtemperature•

Ambienttemperature•

Minimumconductivity•

Max.pressureatmax.•temperature

Numberandtypeofdia-•phragms

Referenceelectrolyte•

contaminatedorcontainingsul-•phidesorproteins

0to14pH•

0to6bar•

0to+130°C(seefig.6.3)•

0to+60°C(whenrunning)•+4to+30°C(storage)

2µS/cm•

6bar•

2"singlepore• TM"

polymer•

18

Technicaldata

english

Type 8202

Page 21: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

plAstRode ph

Typeoffluid•

Measurementrange•

Fluidpressure•

Fluidtemperature•

Ambienttemperature•

Minimumconductivity•

Max.pressureatmax.•temperature

Numberandtypeofdia-•phragms

Referenceelectrolyte•

drinking,aquariumorswimming•poolwater

0to14pH•

0to6bar•

-10to+40°C(seefig.6.3)•

0to+60°C(whenrunning)•+4to+30°C(storage)

2µS/cm•

6bar•

1"singlepore• TM"

polymer•

ceRAtRode ph

Typeoffluid•

Measurementrange•

Fluidpressure•

Fluidtemperature•

Ambienttemperature•

Minimumconductivity•

Max.pressureatmax.•temperature

Numberandtypeofdia-•phragms

Referenceelectrolyte•

athighpressure,ahighflowrate•

0to14pH•

0to16bar•

0to+130°C(seefig.6.3)•

0to+60°C(whenrunning)•+4to+30°C(storage)

50µS/cm•

6bar•

3inceramicHP•

gel•

19

Technicaldata

english

Type 8202

Page 22: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

loGotRode redox

Typeoffluid•

Measurementrange•

Fluidpressure•

Fluidtemperature•

Ambienttemperature•

Minimumconductivity•

Max.pressureatmax.•temperature

Numberandtypeofdia-•phragms

Referenceelectrolyte•

clean•

-2000to+2000mV•

0to6bar•

-10to+50°C(seefig.6.3)•

0to+60°C(whenrunning)•+4to+30°C(storage)

2µS/cm•

6bar•

1"singlepore• TM"

polymer•

unitRode plus redox

Typeoffluid•

Measurementrange•

Fluidpressure•

Fluidtemperature•

Ambienttemperature•

Minimumconductivity•

Max.pressureatmax.•temperature

Numberandtypeofdia-•phragms

Referenceelectrolyte•

clean,contaminated,withlow•conductivity,containingsulphidesorproteins

-2000to+2000mV•

0to6bar•

0to+130°C(seefig.6.3)•

0to+60°C(whenrunning)•+4to+30°C(storage)

2µS/cm•

6bar•

2"singlepore• TM"

polymer•

20

Technicaldata

english

Type 8202

Page 23: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Warning

Risk of injury due to non-conforming assembly.Thedevicemustonlybeassembledbyqualifiedandskilledstaff•withtheappropriatetools.

Risk of injury due to unintentional switch on of power supply or uncontrolled restarting of the installation.Take appropriatemeasures to avoid unintentional activation of•theinstallation.Guaranteeasetorcontrolledrestartingoftheprocesssubsequent•totheassemblyofthedevice.

Assembly7.

safety instructions7.1.

danger

Risk of injury due to high pressure in the installation.Cutoffthepressureanddepressurizethepipesbeforeloosening•thepipesandfittings.

Risk of injury due to electrical discharge.Beforestartingwork,switchoffthepowersupplyandsecureit•topreventrestarting.Observeallapplicableaccidentprotectionandsafetyguidelines•forelectricalequipment.

Risk of injury due to the nature of the fluid.Respect theprevailing ruleson accidentprevention and safety•relatingtotheuseofaggressivefluids.

21

Assembly

english

Type 8202

Page 24: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

unscrew the cover7.2.

noTe

the tightness of the transmitter is not guaranteed when the cover is removed.Takeanyprecautionsnecessarytopreventtheprojectionoffluid•insidethebox.

the transmitter may be damaged if a metal component comes into contact with the electronics.Takeanyprecautionsnecessarytopreventcontactoftheelectron-•icswithametalcomponent(screwdriver,forexample).

Tounscrewthecover,→useyourhandoratoolwhichcanbeusedasalever,takingcarenottoscratchtheglass

Turnthecoveruntilfully→unscrewed

Fig. 7.1 Unscrewing the cover

Fitting the cover7.3.

Check that there is a seal→on the cover and that it isnot damaged. Replace it ifnecessary.

Greasethesealifnecessary,→usingacomponentcompat-ible with thematerial fromwhichitismade.

Fullytightenbyhandtoguar-→anteetightness.

Fig. 7.2 Fitting the cover

22

Assembly

english

Type 8202

Page 25: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Fitting the display module7.4.

Unscrewthecover(see→§7.2).

Setthedisplaymodule→atanangleofca.20°inrelation to the desiredposition.

Themodulecanbefit-→ted in4differentposi-tions,at90°intervals.

Fullypushinthemodule→and,usingthepalmofyour hand, turn to therighttolockit.

Fig. 7.3 Fitting the display module

Removing the display module7.5.

Unscrewthecoverifneces-→sary(see§7.2).

Placethepalmofyourhand→flatonthemoduleandturnbyca.20°totheleft.Onceunlocked, the module israisedslightlybythespringaction.

Remove themodule from→itshousing.

Fig. 7.4 Removing the display module

23

Assembly

english

Type 8202

Page 26: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Fitting the sensor into the 7.6. armature (without fluid)

SensorheadCompressionwasher

Sealring

CheckthatdimensionHon→the sensor is between34and 46mm. If necessary,adjust the height of thecompressionwasher.

Remove the protective→plug

At first use, apply water→orsoapywateron the"A"gasket.

Insert the tightness seal→suppliedintothegrooveonthearmature.

Insertthesensorfittedwith→its seal into the armaturefromabove.Tightenthesensorheadus-→ingasuitablewrench.

Fig. 7.5 Fitting the sensor into the armature (without fluid)

Fitting the electronic module to 7.7. the armature

Check that the sensor is→fitted into the armature(see§7.6).

Checkthatseal"A"onthe→armatureisingoodcondi-tion. Replace it if neces-sary(seeChap.12,SpareParts)

Inserttheelectronicmodule→into the armature,makingsurethepolarisingslotsarecorrectlypositioned.

Apply slight vertical pres-→sure to engage the tight-nessseal.

Fastentheelectronicmod-→ule and the armature to-gether by tightening thenut. Tighten the nutG2''byhandonly,until itstopsturning, to ensure goodelectricalcontact.

Fig. 7.6 Fitting the electronic module to the armature

24

Assembly

english

Type 8202

Page 27: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

insTAllATion And WiRinG8.

safety instructions8.1.

danger

Risk of injury due to high pressure in the installation.Cutoffthepressureanddepressurizethepipesbeforeloosening•thepipesandfittings.

Risk of injury due to electrical discharge.Beforestartingwork,switchoffthepowersupplyandsecureit•topreventrestarting.Observeallapplicableaccidentprotectionandsafetyguidelines•forelectricalequipment.

Risk of injury due to the nature of the fluid.Respect theprevailing ruleson accidentprevention and safety•relatingtotheuseofaggressivefluids.

Warning

Risk of injury due to non-conforming installation.Theelectricalandfluidinstallationcanonlybecarriedoutbyquali-•fiedandskilledstaffwiththeappropriatetools.Install appropriate safety devices (correctly rated fuse and/or•circuit-breaker).Respecttheassemblyinstructionsforthefittingused.•

Risk of injury due to unintentional switch on of power supply or uncontrolled restarting of the installation.Take appropriatemeasures to avoid unintentional activation of•theinstallation.Guaranteeasetorcontrolledrestartingoftheprocesssubsequent•totheassemblyofthedevice.

25

Installationandwiring

english

Type 8202

Page 28: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

installation onto the pipe8.2.

danger

Risk of injury due to high pressure in the installation.Cutoffthepressureanddepressurizethepipesbeforeloosening•thepipesandfittings.

Risk of injury due to the nature of the fluid.Respect theprevailing ruleson accidentprevention and safety•relatingtotheuseofaggressivefluids.

The8202transmitterisinsertedintoafittingfittedtothepipe.

Selectanappropriatepositiononthepiping→

Fig. 8.1 Assembly positions in the piping

Fitthefittingatanangleof±75°max.tothevertical→

Fig. 8.2 Angle to the vertical

Fitthedisplaymodule(see§7.4)tobeabletocalibrateand→programmethetransmitter

Fitthetransmitter(see§7.6and7.7)→Calibratethetransmitter(see10.8.3)→

Thesensormustalwaysbeimmersedinthefluidtopreventitdryingout.

Fitthetransmitterintothefitting,asshowninfigure8.3:→

26

Installationandwiring

english

Type 8202

Page 29: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Checkthatseal"A"is→presentonthefittingandthatitisnotdamaged.Re-placethesealnecessary.

Carefullyinsertthecomplete→transmitterintothefitting.

Adjustthetransmittertothe→desiredposition.

Tightentheholdingnuton→thefitting.

Fig. 8.3 Installation into a fitting

electrical wiring8.3.

danger

Risk of injury due to electrical discharge.Beforestartingwork,switchoffthepowersupplyandsecureit•topreventrestarting.Observeallapplicableaccidentprotectionandsafetyguidelines•forelectricalequipment.

use a high quality electrical power supply (filtered and regulated).

Assembling the male or female 8.3.1. connector (accessories: see chap. 12)

Completelyunscrewthe→nut[1]

Removetherearsectionof→theconnector[2].

Maketheconnections→(see8.3.2or8.3.3)

Fig. 8.4 M12 multi-pin connector (not supplied)

27

Installationandwiring

english

Type 8202

Page 30: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Wiring a version with one m12 8.3.2. fixed connector

V+(14-36VDC)

Transistoroutput1

Transistoroutput2

0V

Fig. 8.5 Wiring of the male fixed connector on a version with 1 M12 fixed connector

pin for the m12 female cable available as op-tional equipment (order ref. 438680)

colour of the conductor

1 brown2 white3 blue

4 black5 grey

Charge1(solenoidvalve,forexample)

Charge2(solenoidvalve,forexample)

Powersupply

white

blue

blackbrown

grey

Fig. 8.6 Connection of the 2 transistor outputs, in NPN mode (software pro-gramming "NPN/sink", see § 10.7.3), on a version with 1 fixed connector

Charge1(solenoidvalve,forexample)

Charge2(solenoidvalve,forexample)

Powersupply

white

blue

blackbrown

grey

Fig. 8.7 Connection of the 2 transistor outputs, in PNP mode (software pro-gramming "PNP/sink", see § 10.7.3), on a version with 1 fixed connector

28

Installationandwiring

english

Type 8202

Page 31: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

powersupply

4-20mAinput

brown

blue

grey

Fig. 8.8 Connection of the current output, in sink mode (software program-ming "NPN/sink", see § 10.7.3), on a version with 1 fixed connector

powersupply

4-20mAinput

brownblue

grey

Fig. 8.9 Connection of the current output, in source mode (software pro-gramming "PNP/source", see § 10.7.3), on a version with 1 fixed connector

Charge2

Charge1 4-20mAinput

brown

greyblue

white

black

powersupply

Fig. 8.10 Connection of the 2 transistor outputs and the current output, in NPN/sink mode (software programming "NPN/sink", see § 10.7.3), on a version with 1 fixed connector

Charge2

Charge1 4-20mAinput

brown

grey

bluewhite

black

powersupply

Fig. 8.11 Connection of the 2 transistor outputs and the current output, in PNP/source mode (software programming "PNP/source", see § 10.7.3), on a version with 1 fixed connector

29

Installationandwiring

english

Type 8202

Page 32: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Wiring a version with two m12 fixed 8.3.3. connectors

Transistoroutput1

Currentoutput1

0V V+(12-36VDC)

male fixed connector

V+(12-36VDC)

Transistoroutput2

Currentoutput2

0V

Female fixed connectorFig. 8.12 Wiring on the M12 male fixed connector and the M12 female fixed connector

pin for the m12 female cable available as op-tional equipment (order ref. 438680)

colour of the conductor

1 brown2 white3 blue4 black5 grey

pin for the m12 male cable available as op-tional equipment (order ref. 559177)

colour of the conductor

1 brown2 white3 blue4 black5 grey

Charge1(solenoidvalve,forexample)

Charge2(solenoidvalve,forexample)

Powersupply

white

blue

brown

grey

white

brown

Fig. 8.13 Connection of the 2 transistor outputs, in NPN mode, on a version with 2 fixed connectors (software programming "NPN/sink", see § 10.7.3)

Charge1(solenoidvalve,forexample)

Charge2(solenoidvalve,forexample)

Powersupply

white

blue

brown

grey

white

blue

Fig. 8.14 Connection of the 2 transistor outputs, in PNP mode, on a version with 2 fixed connectors (software programming "PNP/source", see § 10.7.3)

30

Installationandwiring

english

Type 8202

Page 33: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Powersupply

4-20mA1stinput

brown

blue

grey

4-20mA2ndinput

black

brown

black

Fig. 8.15 Connection of the 2 current outputs, in sink mode, on a version with 2 fixed connectors (software programming "NPN/sink", see § 10.7.3)

Powersupply

4-20mA1stinput

brown

blue

grey

4-20mA2ndinput

black

blue

black

Fig. 8.16 Connection of the 2 current outputs, in source mode, on a version with 2 fixed connectors (software programming "PNP/source", see § 10.7.3)

Charge1

4-20mA1stinput

brown

grey

blue

white

black

Powersupply

4-20mA2ndinput

Charge2

brown

black

white

Fig. 8.17 Connection of the 2 transistor outputs and the 2 current outputs, in NPN/sink mode, on a version with 2 fixed connectors (software program-ming "NPN/sink", see § 10.7.3)

31

Installationandwiring

english

Type 8202

Page 34: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Charge1

4-20mA1stinput

brown

grey

blue

white

black

powersupply

4-20mA2ndinput

Charge2

blue

black

white

Fig. 8.18 Connection of the 2 transistor outputs and the 2 current outputs, in PNP/source mode, on a version with 2 fixed connectors (software program-ming "PNP/source", see § 10.7.3)

32

Installationandwiring

english

Type 8202

Page 35: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

commissioninG9.

safety instructions9.1.

Warning

danger due to non-conforming commissioning.

Non-conformingcommissioningcouldleadtoinjuriesanddamagethedeviceanditssurroundings.Beforecommissioning,makesure that thestaff inchargehave•readandfullyunderstoodthecontentsofthemanual.Inparticular,observethesafetyrecommendationsandintended•use.Thedevice/installationmust only be commissionedby suitably•trainedstaff.Protectthisdeviceagainstelectromagneticinterference,ultravio-•letraysand,wheninstalledoutdoors,theeffectsoftheclimaticconditions.

33

Commissioning

english

Type 8202

Page 36: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

AdjusTmenT And FuncTions10.

safety instructions10.1.

Warning

Risk of injury due to non-conforming adjustment.

Non-conformingadjustmentcouldleadtoinjuriesanddamagethedeviceanditssurroundings.Theoperatorsinchargeofadjustmentmusthavereadandunder-•stoodthecontentsofthismanualInparticular,observethesafetyrecommendationsandintended•use.Thedevice/installationmustonlybeadjustedbysuitablytrained•staff.

Functions10.2.

Thedevicehas2modesofuse:

Read mode

Thismodeisusedtoread:

themeasuredvaluesof2physicalparametersselectedinthe•Parametersmenu

thecurrentvaluesemittedonthe4-20mAoutputs•

menus mode

Thismodecomprises5menus:

the parameters menu (§ 10.7) to:•

settheparametersforthetransmitterdata,independentofthe-measurements(date,time,Parametersmenuaccesscode...)

restorethedefaultdataoftheReadmodeandtheoutputs-

programmetheinformationtobedisplayedinReadmode,-independentlyoftheparametersprogrammedforthecurrentandtransistoroutputs

definethedisplaycontrastand,onaversionwith2fixedcon--nectorsonly,thebrightnessofthebacklighting

settheparametersforeachcurrentandtransistoroutput,inde--pendentlyofeachotherandindependentlyoftheparametersselectedfortheReadmode

definethebehaviourofeach4-20mAoutputintheeventof-diagnosticerror

definetheelectricalconnectionmodefortheoutputs(sink/-NPNorsource/PNP)

definethemeasurementparametersforthepHoroxidation-reductionpotentialsensor

the calibration menu (§ 10.8) to:•

activatetheHoldmode-

programmetheCalibrationmenuaccesscode-

adjusteach4-20mAoutput-

calibratethesensor-

inputanoffsettomeasurethetemperature-

34

Adjustmentandfunctions

english

Type 8202

Page 37: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

the diagnostic menu (§ 10.9) to:•

programmetheDiagnosticmenuaccesscode-

configurethedevice'sbehaviourifproblemsaredetectedon-thepH/redoxsensorelectrodes

activateordeactivatetheuseof"error"and"warning"messag--esforthepH/redoxsensorifthedefinedimpedancethresholdsareexceeded

configurethe2impedancerangesoutsideofwhich"error"and-"warning"messagesareused

readthemeasuredvaluesoftheimpedancesonthepH/redox-sensor

activateordeactivatetheuseof"error"and"warning"mes--sagesifthedefinedfluidtemperaturethresholdsareexceeded

configurethe2temperaturerangesoutsideofwhich"error"-and"warning"messagesareused

readthemeasuredvalueofthefluidtemperature-

the test menu (§ 10.10) to:•

programmetheTestmenuaccesscode-

checkthattheoutputsareworkingandbehavingcorrectly-

the information menu (§ 10.11) to:•

readthemeaningofcertainicons,whentheyaredisplayed-

readthesoftwareversionoftheacquisition/conversionmod--uleforthephysicalparameters

35

Adjustmentandfunctions

english

Type 8202

Page 38: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

using the browse button10.3.

Symbolisedby

inthismanual

Symbolisedby

inthismanual

Symbolisedby inthismanual

Symbolisedbyinthismanual

Symbolised

by inthisma-nual

Fig. 10.1 Using the browse button

you want to... press...

...browseinReadmode

Nextscreen:•

Previousscreen:•...accesstheProgramming•mode

...displaytheParameters•menu

foratleast2sec.fromany

screeninReadmode

...browseintheProgrammingmodemenus Nextmenu:•

Previousmenu:•...selectthemenudisplayed

...browseinthemenufunctions

Nextfunction:•

Previousfunction:•...selectthehighlightedfunction

36

Adjustmentandfunctions

english

Type 8202

Page 39: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

you want to... press...

...browseinthedynamicfunc-tionsbar(MEAS,BACK,ABORT,OK,YES,NO)

Nextfunction:•

Previousfunction:•...confirmthehighlighteddy-namicfunction

...modifyanumericalvalueincrementthefigurese--lected

-decrementthefigurese--lected -selectthepreviousfigure-

-selectthenextfigure-

-allocatethe"+"or"-"sign-tothenumericalvalue - totheextremeleft

ofthenumericalvaluethen

untilthedesiredsignisdisplayed

movethedecimalpoint-- totheextremerightofthenumericalvaluethen

untilthedecimalpointisinthedesiredplace

37

Adjustmentandfunctions

english

Type 8202

Page 40: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

description of the display10.4.

description of the icons and leds10.4.1.

OPEN LOCK

mV_pH1423 mV

tempc55 °cERR

Redled:indi-catesanerror;see§11.14

Notused

Yellowled:indicatesthattran-sistor1hasbeenswitched

Yellowled:indicatesthattran-sistor2hasbeenswitched

Fig. 10.2 Position of the icons and description of the leds

Theledsofthedisplaymoduleareduplicatedontheelec-tronicboardthatislocatedunderthedisplaymodule:theseledsbecomevisiblewhenthetransmitterisnotequippedwiththedisplaymodule.

icon meaning and alternatives

Sensoringoodorderandfluidtemperaturewithinthepro-grammedrange.

Ifthemonitoringoftheimpedanceontheelectrodesand/orthefluidtemperaturehasbeenactivated,thealternativesinthispositionare:

• ,associatedwith :see§10.9.2,10.9.3and11.4

• ,associatedwith ERR :see§10.9.2,10.9.3and11.4Devicecurrentlymeasuring.

Thealternativesinthispositionare:

HOLD!

• flashing:HOLDmodeactivated(see§10.8.1)

T• :runningcheckthattheoutputsareworkingandbe-havingcorrectly(see§10.10.2and10.10.3)

"Maintenance"event;see§10.8.3and11.4

"Warning"event;see§10.8.3,10.9.2,10.9.3and11.4

ERR "Error"event;see§10.9.2,10.9.3and11.4

38

Adjustmentandfunctions

english

Type 8202

Page 41: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

When switching on the device10.4.2.

Whenthedeviceisswitchedon,thedisplayindicatesthesoftwareversionofthedisplay.

ThedisplaythenshowsthefirstscreeninREADmode:

OPEN LOCK

mV_ph1423 mV

tempc55 °c

Fig. 10.3 Display when the device is switched on

Seeparagraph10.7.2tochoosethedatadisplayedinReadmode

When browsing in the menus10.4.3.

Fonction2:Fonction3:Fonction4:MEAS

System

BACK OK

This iswhen thedevice is be-ing parame-tered................................

Thearrowindicatesthepresenceofthepreviousfunc-tions,accessible

via

Thearrowindicatesthepresenceofthenextfunctions,accessiblevia

Thisiconidentifiesthemenuinwhichyouwillfindit.

Titleofthemenu,thesub-menuorthefunctionwhichyouarecurrentlybrowsing.

Dynamicfunctions(accessible

via and ):

MEAS:togobacktotheREADmode•

BACK:togobacktotheparentmenu•

ABORT(appearsinplaceof"BACK"):togo•backtotheparentmenuwithoutvalidatingtheinput

OK:tovalidatethevalueorinputselected•andreturnautomaticallytotheparentmenu

YESandNO:toanswerthequestionasked•

39

Adjustmentandfunctions

english

Type 8202

Page 42: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Read mode and access to the menus mode10.5.

This iswhen thedevice is be-ing parame-tered................................

This iswhen thedevice is be-ing parame-tered................................

param >2s

Readmode

mV_ph1243 mV

tempc55 °c

mV_ph1243 mV

tempc55 °c

18.3 mA

Ac118.3 mA

Ac27.5 mA

Ac1

7.5 mAAc2

calib

This iswhen thedevice is be-ing parame-tered................................

infoThis iswhen thedevice is be-ing parame-tered................................

diagnostic

See§10.6fordetailsofthefunctionspermenu→

Menusmode

A

A

B

B

test

40

Adjustmentandfunctions

english

Type 8202

Page 43: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Functions available per menu10.6.

This iswhen thedevice is be-ing parame-tered................................

This iswhen thedevice is be-ing parame-tered................................

This iswhen thedevice is be-ing parame-tered................................

param

calib

test

This iswhen thedevice is be-ing parame-tered................................

SystemDisplayOutputsMEAS BACK

Param

SystemOutputsSensorMEAS BACK

Calibration

SystemOutputsSensorMEAS BACK

TestSystem menu: see § 10.7.1

Display menu: see § 10.7.2

Outputs menu: see § 10.7.3, 10.7.4 and 10.7.5

Sensor menu: see § 10.7.6

Code"Param"correct

Codeincorrect1)

diagnostic

SystemSensor

MEAS BACK

Diagnostic

System menu: See § 10.8.1

Outputs menu: see § 10.8.2

Sensor menu: see § 10.8.3 and 10.8.4

System menu: see § 10.10.1

Outputs menu: see § 10.10.2

Sensor menu: see § 10.10.3

System menu: see § 10.9.1

Sensor menu: see § 10.9.2 and 10.9.3

This iswhen thedevice is be-ing parame-tered................................

SensorMEAS BACK

ParamDisplayOutputs

This iswhen thedevice is be-ing parame-tered................................

Code"Diagnostic"

correct

Code"Calib"correct

Code"Test"correct

info

ErrorWarningMaintenanceMEAS BACK

Info

Error, Warning, Maintenance, Smiley and Software menus: see § 10.11

SoftwareMEAS BACK

InfoMaintenanceSmiley

Codeincorrect1)

Codeincorrect1)

Codeincorrect1)

1)Ifthecodehasbeenlost,usethe"9100"universalcode.

41

Adjustmentandfunctions

english

Type 8202

Page 44: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

parameters menu ("param")10.7.

set the data parameters independently of the measurements ("system" menu)10.7.1.

Refertoparagraphs10.5and10.6toaccesstheParammenu.

This iswhen thedevice is be-ing parame-tered................................

DateTime

MEAS BACK

System

This iswhen thedevice is be-ing parame-tered................................

Code

FactorySetMEAS BACK

System

DATE:Setthedate(inputformat:year/month/dayintheformYYYY/MM/DD)

TIME:Setthetime(inputformat:hours:minutesseconds)

CODE:ModifythePARAMmenuaccesscode(4figures;0000bydefault).

FACTORY SET:RestorethedefaultdataoftheReadmodeandtheoutputs.TimeCode

42

Adjustmentandfunctions

english

Type 8202

Page 45: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

set the display parameters ("display" menu)10.7.2.

Warning

Risk of injury due to non-conforming adjustment.Choose the type of sensor (see10.7.6) fitting the transmitter•beforesettingthedisplayparameters.

Refertoparagraphs10.5and10.6toaccesstheParammenu.

1)or"Line2"

This iswhen thedevice is be-ing parame-tered................................

Line1Line2

MEAS BACK

Display

This iswhen thedevice is be-ing parame-tered................................

Contrast

BacklightMEAS BACK

Display

This iswhen thedevice is be-ing parame-tered................................

PVar:Unit:Filter:MEAS

Line 1 1)

BACK

set the parameters for the data displayed in Read mode:

PVAR:Choosethephysicalparametertobedisplayedonline1orline2ofthedisplayrespectively.Thepossiblechoicesdependontheselectedsensortype,whetherpHorORP.

UNIT:ChoosetheunitinwhichthephysicalparameterselectedinthePVARfunctionaboveisdisplayed

FILTER:Choosetheattenuationlevelforthemeasurementvariationsofthephysicalparameterdisplayedonline1orline2.Seedetailsnextpage

set the parameters for the reading comfort level of the data displayed:

CONTRAST:Choosethedisplaycontrastlevel(asa%)

BACKLIGHT:Onaversionwith2fixedconnectorsonly,choosethelightintensityofthedisplay(asa%)

Line2Contrast

43

Adjustmentandfunctions

english

Type 8202

Page 46: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Attenuate the measurement variations on the display ("Filter" function in "line1" or "line2" menu)

Threelevelsofattenuationareproposed:"slow","fast"or"none"

"Slow"filtration

"Fast"filtration

"None"filtration

Fig. 10.4 Filtration graphs

output connection mode (menu 10.7.3. "outputs" - Function "hWmode")

Refertoparagraphs10.5and10.6toaccesstheParammenu.

This iswhen thedevice is be-ing parame-tered................................

HWModeAC1AC2MEAS BACK

Outputs

This iswhen thedevice is be-ing parame-tered................................

TR1TR2MEAS BACK

Outputs

Outputs menu, functions AC1 and AC2: see 10.7.4

Outputs menu, functions TR1 and TR2: see 10.7.5

HWMODE:Choosetheconnectionmodeforalloutputs(currentandtransistor)

AC2

Theconnectionmodeisthesameforalloutputs.

Ifyouchoose"sinkNPN",thecurrentoutputsmustbeconnectedinsinkmodeandthetransistoroutputsinNPNmode.

Ifyouchoose"sourcePNP",thecurrentoutputsmustbeconnectedinsourcemodeandthetransistoroutputsinPNPmode.

Seethewiringfortheoutputsinparagraph8.3.

44

Adjustmentandfunctions

english

Type 8202

Page 47: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

set the parameters for the current outputs (menu "outputs" - Functions "Ac1" and "Ac2")10.7.4.

Warning

Risk of injury due to non-conforming adjustment.Choose the type of sensor (see10.7.6) fitting the transmitter•beforesettingtheoutputparameters.

Refertoparagraphs10.5and10.6toaccesstheParammenu.The2ndcurrentoutput"AC2"isonlyavailableonaversionwith2fixedcon-nectors.

This iswhen thedevice is be-ing parame-tered................................

HWModeAC1AC2MEAS BACK

Outputs

This iswhen thedevice is be-ing parame-tered................................

TR1TR2MEAS BACK

Outputs

This iswhen thedevice is be-ing parame-tered................................

PVar:4mA:20mA:MEAS BACK

Output.AC11)

This iswhen thedevice is be-ing parame-tered................................

Filter:

MEAS BACK

Output.AC11)

PVAR:Chooseaphysicalparameterassociatedwithcurrentoutput1orcurrentoutput2respectively.Thepossiblechoicesdependontheselectedsensortype,whetherpHorORP.

4mA:Choosethevalueofthephysicalparameter(previouslyselected),associatedwithacurrentof4mA,foreachcurrentoutput.See"Currentoutputparameters",nextpage.

20mA:Choosethevalueofthephysicalparameter(previouslyselected),associatedwithacurrentof20mA,foreachcurrentoutput.See"Currentoutputparameters",nextpage.

FILTER:Choosethelevelofattenuationforthecurrentvariationsforeachcurrentout-put.Seenextpageandfig.10.4.

DIAG. MODE:Choosetoemitacurrentof22mAoncurrentoutput1orcurrentoutput2respectivelywhenan"error"eventrelatedtothediagnostic(see§10.9.2and10.9.3)isgeneratedbythetransmitterorleavecurrentoutput1orcurrentoutput2respectivelytooperatenormally(select"none").

Seealso"Ifyouencounterproblems"at§11.4.

Outputs menu - Functions TR1 and TR2: see 10.7.5

Diag.mode:

Outputs menu - "HWMode" function: see 10.7.3

AC2

1)or"Output.AC2"

20mA:

45

Adjustmentandfunctions

english

Type 8202

Page 48: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Attenuate the measurement variations on the current output ("Filter" function in "output.Ac1" or "output.Ac2" menu)

Threelevelsofattenuationareproposed:slow,fastornone

Thebehaviourofthefiltersforthecurrentoutputsisidenticaltothatofthefiltersfordisplay.Refertofigure10.4.

set the parameters of a current output (functions "4mA" and "20mA" in the "outputs" menu)

Functions"4mA"and"20mA"areusedtodefinethemeasurementrangeforthephysicalparameterassociatedwiththecurrentonthe4-20mAoutput.

WeshouldnoteP1respectivelyP2thevaluesassociatedwitha4mAor20mAcurrentrespectively

Physicalparameterchosen

Fig. 10.5 Current 4-20 mA according to the chosen physical parameter

please note:

P1maybegreaterthanP2:ifthisisthecase,thesignalisinvertedandrangeP1-P2correspondstotherangeforthe20-4mAcurrent.

46

Adjustmentandfunctions

english

Type 8202

Page 49: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

set the parameters for the transistor outputs (menu "outputs" - functions "TR1" and "TR2")10.7.5.

Warning

Risk of injury due to non-conforming adjustment.Choose the type of sensor (see10.7.6) fitting the transmitter•beforesettingtheoutputparameters.

Refertoparagraphs10.5and10.6toaccesstheParammenu.

This iswhen thedevice is be-ing parame-tered................................

HWModeAC1AC2MEAS BACK

Outputs

This iswhen thedevice is be-ing parame-tered................................

TR1TR2MEAS BACK

Outputs

This iswhen thedevice is be-ing parame-tered................................

PVar:Mode:Low:MEAS BACK

Output.TR11)

This iswhen thedevice is be-ing parame-tered................................

High:Contact:

MEAS BACK

Output.TR11)

Delay:

PVAR:Chooseaphysicalparameterassociatedwithtransistoroutput1ortransistoroutput2respectivelyorlinkthegenerationofa"warning"event(see§10.8.3,10.9.2and10.9.3)totransistoroutput1ortransistoroutput2re-spectively.Thepossiblechoicesdependontheselectedsensortype,whetherpHorORP.Iftheselectedtransistoroutputislinkedtothe"warning"event,thetransistorswitchesassoonassuchaneventisgeneratedbythetransmitter.Seealso"Ifyouencounterproblems",at§11.4.

MODE 2):Choosetheswitchingmodefortransistoroutput1ortransistoroutput2.Seedetailsnextpage.

LOW 2):Choosethelowswitchingthresholdvaluefortransistoroutput1ortransistoroutput2.Seedetailsnextpage.

HIGH 2):Choosethehighswitchingthresholdvaluefortransistoroutput1ortransistoroutput2.Seedetailsnextpage.

CONTACT:Choosethetypeofcontactonstandby(normallyopen,NO,ornormallyclosed,NC)fortransistoroutput1ortransistoroutput2.Seedetailsnextpage.

DELAY:Choosethevalueofthedelaypriortoswitchingforeachtransistoroutput.Seedetailsnextpage.

Outputs menu - HWMode function: see 10.7.3

Outputs menu - Functions AC1 and AC2: see 10.7.4

AC2

1)or"Output.TR2"

2)Arenotavailableifthegenerationofa"warning"eventhasbeenlinkedtotheselectedtransistoroutput.

47

Adjustmentandfunctions

english

Type 8202

Page 50: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

possible switching modes and switching thresholds on a tran-sistor output (functions "mode", "low", "high" and "delay" in the menu "output.tR1" or "output.tR2")

hysteresis mode

Thechangeofstatusisdonewhenathresholdisdetected(increas-ingflowrate:thresholdhigh(functionHigh)tobedetected;de-creasingflowrate:thresholdlow(functionLow)tobedetected).

ON

OFF

contact

Low High

physicalparameter

NOON

OFF

contact

Low High

physicalparameter

NC

Fig. 10.6 Hysteresis mode

Window mode

Thechangeofstatusisdonewheneveroneofthethresholdsisdetected.

ON

OFF

contact

Low High

physicalparameter

NOON

OFF

contact

Low High

physicalparameter

NC

Fig. 10.7 Window mode

NO=Normallyopen;NC=Normallyclosed

the "delay" functionisusedtoprogrammethetimingpriortoswitchingandisvalidforbothoutputthresholds.

Switchingisonlydoneifoneofthethresholds,highorlow(func-tions"High"or"Low"),isexceededforadurationlongerthanthistiming.

48

Adjustmentandfunctions

english

Type 8202

Page 51: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

set the sensor parameters 10.7.6. ("sensor" menu)

Warning

Risk of injury due to non-conforming adjustment.Choose the typeof sensormountedon the transmitter before•settingthedisplayandoutputparameters

Themonitoring(see§10.9.2)oftheredox("ORP")sensorisimpossibleifthemeasurementmodeissetto"asymmetri-cal".

Refertoparagraphs10.5and10.6toaccesstheParammenu.

This iswhen thedevice is be-ing parame-tered................................

Type:Mode:MainsFcy:MEAS BACK

Sensor TYPE:Choosethetypeofsensorused,pHorRedox("ORP"choice).

MODE:Choosethetypeofmeasurement,symmetrical(differential)orasymmetrical.

MAINS FCY:Choosethefrequencyofyourmainselectricity,50or60Hz.Thisfrequen-cyisfilteredbythetransmittertoensurestablemeasurements.

symmetrical or asymmetrical measurement ("mode" function in the "sensor" menu)

symmetrical measurement

Thesymmetricalmeasurementisadifferentialmeasurement:inthistypeofmeasurement,thestainlesssteelringonthesensorarmatureisusedasareference.

(1):Measurementelectrode

(2):Referenceelectrode

(3):Stainlesssteelring

Fig. 10.8 Schematic diagram of symmetrical measurement

Asymmetrical measurement

Intheasymmetricalmeasurement,themeasurementisdoneinrela-tiontothereferenceelectrode.

(1):Measurementelectrode

(2):Referenceelectrode

Fig. 10.9 Schematic diagram of asymmetrical measurement

49

Adjustmentandfunctions

english

Type 8202

Page 52: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

calibration menu ("calib")10.8.

Activate/deactivate the hold 10.8.1. function and modify the access code to the calibration menu ("system" menu)

Refertoparagraphs10.5and10.6toaccesstheCalibmenu.

HoldCode

MEAS BACK

System HOLD:ChoosewhethertheHOLDfunc-tionisenabledordisabled.

CODE:ModifytheCalibrationmenuaccesscode(4figures;0000bydefault)

"hold" function in the "system" menu

TheHoldmodesiautomaticallydeactivatedwhenthetrans-mitterrestartsafterapowerinterruption,iftheHoldmodewasactivatedatthemomentofthepowercut-off.

TheHoldmodeisusedtocarryoutmaintenanceworkwithoutinter-ruptingtheprocess.

ToactivatetheHOLDmode:

enterthe"HOLD"function→choose"enabled"→validateby"OK"→

Inpractice,whenthedeviceisinHoldmode:

the• HOLD!

iconisdisplayedinplaceofthe icon

thecurrentemittedoneach4-20mAoutputisfixedatthevalue•ofthelastmeasurementofthephysicalparameterassociatedwitheachoutput

eachtransistoroutputisfixedatthestatusacquiredatthemo-•menttheHoldfunctionisactivated.

thedeviceisinHoldmodeuntiltheHOLDfunctionisdeactivated.•

TodeactivatetheHOLDmode:

enterthe"HOLD"function→choose"disabled"→validateby"OK"→

50

Adjustmentandfunctions

english

Type 8202

Page 53: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Adjust the current outputs ("outputs" menu)10.8.2.

Warning

Risk of injury due to non-conforming adjustment.MakesuretheHoldmodeisdisabled(see10.8.1).•

Refertoparagraphs10.5and10.6toaccesstheCalibmenu.

AC1AC2

MEAS BACK

Outputs

4mA:Adjusttheoffsetoncurrentoutput1orcurrentoutput2.

Whenthe"4mA"functionisselected,thetransmittergeneratesacurrentof4mA:measurethecurrentemittedbythe4-20mAoutputusingamultimeterandinputthevaluegivenbythemultimeterinthefunction"AC1.4mA"or"AC2.4mA".

20mA:Adjustthespanoncurrentoutput1orcurrentoutput2.

Whenthe"20mA"functionisselected,thetransmittergeneratesacurrentof20mA:measurethecurrentemittedbythe4-20mAoutputusingamultimeterandinputthevaluegivenbythemultimeterinthefunction"AC1.20mA"or"AC2.20mA".

4mA20mA

MEAS BACK

AC1 1)

1)or"AC2"

calibrate the sensor (menu sensor - function "probe")10.8.3.

Choosethetypeofsensor(§10.7.6)thetransmitterisfit-tedwithbeforecalibratingthesensor.

Refertoparagraphs10.5and10.6toaccesstheCalibmenu.

51

Adjustmentandfunctions

english

Type 8202

Page 54: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

ProbeTemperature

MEAS BACK

Sensor

Calibtemperat.Calibration

MEAS BACK

Probe

ProbeconstantsCalibintervalMEAS BACK

Probe

CALIB TEMPERAT.:Choosethetypeoftemperaturecompensationforthecalibrationprocess:eitherthetemperaturemeasured(select"Auto")orafixedvalue(select"constant"theninputreferencetemperature).

CALIBRATION:CalibratethepHsensorin1or2pointsortheRedoxsensorin1point.Seedetailsfollowingpages.Acalibrationprocessupdatesthelastcalibrationdate("Lastcal.date"functionoftheCALIBINTERVALsub-menuhereafter).

Thiscalibrationcanbereplacedbytheinputinthe"Probeconstants"functionbelow,oftheoffsetand/orspanvaluesindicatedonthesensorcertificate,ifsupplied.Enteringtheprobeconstantsdoesnotupdatethelastcalibrationdate("Lastcal.date"functionoftheCALIBINTERVALsub-menuhereafter).

PROBE CONSTANTS:Inputtheoffsetand/orspanvaluesindicatedonthepH/Redoxsensorcertificate,ifsupplied.Thisinputreplacesacalibrationmadebythe"Calibration"functionabovebutdoesnotupdatethelastcalibrationdate("Lastcal.date"functionoftheCALIBINTERVALsub-menuhereafter).

CALIB INTERVAL:Readthedateofthelastcalibration(function"Lastcal.date")andprogrammetheperiodicityofcalibrations,indays(function"Interval"):thetransmittergen-

eratesa"maintenance"eventbydisplayingthe iconanda"warning"event,eachtimeacalibrationisdue.Setfunction"Interval"to"0000days"toignorethefunction.

The"warning"eventmaybeassociatedwithoneorother•orbothtransistoroutputs(see§10.7.5).

Seealso"Ifyouencounterproblems"at§11.4.•

Probeconstants

Calibration

52

Adjustmentandfunctions

english

Type 8202

Page 55: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

calibrate the ph / Redox sensor ("calibration" function in the "probe" menu)

ThepHsensorcanbecalibratedaccordingtoa1-pointora•2-pointprocedure.

TheRedoxsensorcanbecalibratedaccordingtoa1-pointproce-•dure.

Inordernottointerrupttheprocess,activatetheHOLD•function(see§10.8.1).

Beforeeachcalibration,correctlycleantheelectrodewith•asuitableproduct.

Ina2-pointcalibration,thebuffersolutionsusedmustbe•atthesametemperature.

Programmetheperiodicityofcalibrationsinthe"Interval"•functioninthesub-menu"Calibinterval"(seepreviouspage):eachtimeacalibrationisdue,thetransmittergen-eratesa"maintenance"eventanda"warning"event.

1-point calibration

The1-pointcalibrationprocedureisusedforrapidcalibrationbyadjustingtheoffsetofthemeasurementgraphwithabuffersolutionwithaknownpH(tocalibrateapHsensor:seep.55)oraknownoxidationreductionpotential(tocalibrateaRedoxsensor:seep.54).

2-point calibration

The2-pointcalibrationprocedureforapHsensorisusedfortheprecisecalibrationoftheoffsetandthegradient("span")ofthesen-sormeasurementgraph.

Thisoperationrequires2buffersolutions:ingeneralafirstsolutionwithapHof7andasecondsolutionwithapHveryclosetothatoftheprocessvaluetobemeasured.Seep.55.

errors in ph sensor calibration

AttheendofcalibrationofthepHsensor,twotypesofmessagesmaybedisplayed:

the"Warning:span/offsetoutofrange"messagesignalseither•anerrorinthebuffersolutionortheageingofthesensor(ithasreachedhalfwayinitslifespan).Ageingofthesensorisconfirmedbya"span"valueofbetween50mV/pHand53mV/pHorbe-tween63mV/pHand65mV/pHandanoffsetvalueofbetween-60mVand-35mVorbetween35mVand60mV.Thesevaluescanneverthelessbesaved.

the"Error:span/offsetoutofrange"messageindicatesthatthe•sensormustbereplaced.Thisisconfirmedbya"span"value<50mV/pHor>65mV/pHoranoffsetvalue<-60mVor>60mV.Inthiscase,thecalibrationisnottakenintoaccount.Recalibratethetransmitterwhenthenewsensorisinplace.

53

Adjustmentandfunctions

english

Type 8202

Page 56: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

detailed procedure for the 1-point calibration of the oxidation reduction potential sensor

465.0mV1st point

475.0mV

Offset:

Cal.Result

Immersethecleansensorinthebuffersolution;iftheHoldmodeisdeactivated,thetransmitteralternatelydisplays:

thesolutionDDPmeas-•ured

thesolutiontemperature•measured

InputtheDDPofthebuffersolution(indi-catedonthebottle).

OK

Thetransmitterdisplaysthecalibrationresult.

Youarerequestedtosavethecalculatedvalueornot.

Seebelowforerrorsinthecalibrationofanoxida-tionreductionpotentialsensor.

MEAS BACK

MEAS ABORT OK

BACK

errors in the calibration of an oxidation reduction potential sensor

Attheendofcalibrationoftheoxidationreductionpotentialsensor,twotypesofmessagemaybedisplayed:

the"Warning:span/offsetoutofrange"messagesignalseitheranerrorinthebuffersolutionortheageingofthesensor(ithasreached•halfwayinitslifespan).Ageingofthesensorisconfirmedbyanoffsetvalueofbetween-60mVand-35mVorbetween35mVand60mV.Thesevaluescanneverthelessbesaved.

the"Error:span/offsetoutofrange"messageindicatesthatthesensormustbereplaced.Thisisconfirmedbyanoffsetvalue<-60mVor>•60mV.Inthiscase,thecalibrationisnottakenintoaccount.Recalibratethetransmitterwhenthenewsensorisinplace.

54

Adjustmentandfunctions

english

Type 8202

Page 57: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

detailed procedure for the 1- or 2-point calibration of the ph sensor

7.100pH1st point

Doyouwanttocali-bratea2ndpoint?

YES NO

1st point

7.000pH

Offset:Span:

Cal.Result

Immersethecleansensorinthefirstbuffersolution;iftheHoldmodeisde-activated,thetransmitteralternatelydisplays:

thesolutionpHmeas-•ured

thesolutiontempera-•turemeasured

InputthepHofthebuffersolution(indi-catedonthebottle).

Rinse

Rinsethesensor.WhenthepHreachesthedesiredvalue,validaterinsingwithOK

23.6°C2nd point

10.00pH

Immersethecleansensorinthesecondbuffersolution;iftheHoldmodeisdeactivated,thetransmitteralternatelydisplays:

thesolutionpHmeasured•

thesolutiontemperaturemeasured•

InputthepHofthesecondbuffersolu-tion(indicatedonthebottle).

OK

Thetransmitterdisplaysthecalibrationresult.

Youarerequestedtosavethecalcu-latedvaluesornot.

Seep.53forerrorsinthecalibrationofapHsensor.

BACK

6.153pH25.00°C

OK

MEAS ABORT OK

MEAS BACK

MEAS ABORT OK

MEAS ABORT OKOK

YES NO

55

Adjustmentandfunctions

english

Type 8202

Page 58: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

input an offset for the temperature 10.8.4. measurement (menu "sensor" - function "Temperature")

Refertoparagraphs10.5and10.6toaccesstheCalibmenu.

ThetemperaturetransmittedbythePt1000sensormaybecor-rected.

Thiscorrectionfactoristhetemperatureoffset.

BACKProbeTemperature

MEAS BACK

SensorMEAS

Calib.temp

OK

Inputthetempera-tureoffset(±5°C)

-3,500°C

diagnostic menu ("diagnostic")10.9.

modify the diagnostic menu access 10.9.1. code ("system" menu)

Refertoparagraphs10.5and10.6toaccesstheDagnosticmenu.

SystemSensor

MEAS BACK

DiagnosticCode

MEAS BACK

System

CODE:ModifytheDiagnosticmenuaccesscode(4figures;0000bydefault).

56

Adjustmentandfunctions

english

Type 8202

Page 59: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

monitor the sensor (menu "sensor" - Functions "Glass electrode" and "Ref electrode")10.9.2.

Refertoparagraphs10.5and10.6toaccesstheDagnosticmenu.

configure the device's behaviour in case of electrode consumption or if there are problems on the ph sensor (glass electrode, function "Glass elec" and/or reference electrode, function "Ref.elec.") or on the Redox sensor (reference electrode, function "Ref.elec.")

ACTIVATE:Choosewhetherornottoactivatemonitoringoftheimpedanceonthese-lectedelectrode.

Thismonitoringisdonebythegenerationofa"warning"eventiftheimpedancerangede-finedinthe"Warnhi/lo"functionsbelowisexceededandan"error"eventiftheimpedancerangedefinedinthe"Errhi/lo"functionsbelowisexceeded.

Seealso"Ifyouencounterproblems"at§11.4.

IMPEDANCE:ReadtheimpedancemeasuredinrealtimeonthepHelectrodeandthereferenceelectroderespectivelyofthepH/Redoxsensor.

WARN HI:Inputtheimpedancevalueabovewhicha"warning"eventisgenerated.Seedetailsnextpage.

WARN LO:Inputtheimpedancevaluebelowwhicha"warning"eventisgenerated.Seedetailsnextpage.

ERR HI:Inputtheimpedancevalueabovewhichan"error"eventisgenerated.Seedetailsnextpage.

ERR LO:Inputtheimpedancevaluebelowwhichan"error"eventisgenerated.Seedetailsnextpage.

GlasselectrodeRefelectrodeTemperatureMEAS BACK

Sensor

MEAS BACK

Glass elec.1)

Activate:Impedance:Warnhi:

MEAS BACK

Glass elec.1)

Warnlo:Errhi:Errlo:

1)or"Ref.elec."

Sensor menu - Tem-perature function: see 10.9.3

57

Adjustmentandfunctions

english

Type 8202

Page 60: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

"Warning" and "error" events (functions "Activate", "Warn hi", "Warn lo", "err hi" and "err lo" in the menus "Glass elec." and "Ref.elec.")

Amalfunctioninyourprocessorsimplyinthemeasurementsensormayberevealedbytoolowortoohighanimpedanceononeand/orotherelectrodeonthemeasurementsensor.

Tobewarnedwhenanelectrodehastoolowortoohighanimped-ance:

activatemonitoringonthiselectrodeinthefunction"activate",→then

programmeanimpedancerange(inM→ W forthepHelectrodeandinkWforthereferenceelectrodeonthepH/Redoxsensor)outsideofwhichthetransmittergeneratesa"warning"eventand

displaysthe and icons.

programmeanimpedancerange(inM→ W forthepHelectrodeandinkWforthereferenceelectrodeonthepH/Redoxsensor)outsideofwhichthetransmittergeneratesan"error"eventand

displaysthe andERR icons.

Whena"warning"or"error"eventisgeneratedbythetransmitter:

gointothe"Info"menutoreadthecauseoftheeventgeneration→and/orgointothe"Sensor"functionoftheDiagnosticmenuto→readtheimpedancevaluesforeachelectrodeinordertoidentifythecauseofanoutofrangeimpedance.

ifnecessary,cleanthenrecalibratethemeasurementsensoror→replaceit

The"warning"eventmayalsobeassociatedwithoneor•otherorbothtransistoroutputs.See§10.7.5,function"Output.TR1"or"Output.TR2".

The"error"eventmayalsobeassociatedwithoneor•otherorbothcurrentoutputs.See§10.7.4,function"Output.AC1"or"Output.AC2".

Seealso"Ifyouencounterproblems"at§11.4.•

58

Adjustmentandfunctions

english

Type 8202

Page 61: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

monitor the fluid temperature (menu "sensor" - Function "Temperature")10.9.3.

Refertoparagraphs10.5and10.6toaccesstheDagnosticmenu.

monitor the fluid temperature and configure the device's behaviour if the ranges defined are exceeded ("temperat.")

ACTIVATE:Choosewhetherornottoactivatemonitoringofthefluidtemperature.

Thismonitoringisdonebythegenerationofa"warning"eventifthefluidtemperaturerangedefinedinthe"Warnhi/lo"functionsbelowisexceededandan"error"eventifthefluidtemperaturerangedefinedinthe"Errhi/lo"functionsbelowisexceeded.

Seealso"Ifyouencounterproblems"at§11.4.

TEMPERATURE:Readthefluidtemperaturemeasuredinrealtimebythebuilt-inPt1000sensor.

WARN HI:Inputthefluidtemperaturevalueabovewhicha"warning"eventisgenerated.Seedetailsnextpage.

WARN LO:Inputthefluidtemperaturevaluebelowwhicha"warning"eventisgenerated.Seedetailsnextpage.

ERR HI:Inputthefluidtemperaturevalueabovewhichan"error"eventisgenerated.Seedetailsnextpage.

ERR LO:Inputthefluidtemperaturevaluebelowwhichan"error"eventisgenerated.Seedetailsnextpage.

GlasselectrodeRefelectrodeTemperatureMEAS BACK

Sensor

MEAS BACK

Temperat.

Activate:TemperatureWarnhi:

MEAS BACK

Temperat.Warnlo:Errhi:Errlo:

Sensor menu - Functions Glass electrode and Ref electrode: see 10.9.2

59

Adjustmentandfunctions

english

Type 8202

Page 62: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

"Warning" and "error" events (functions "Activate", "Warn hi", "Warn lo", "err hi" and "err lo" in the menu "temperature")

Amalfunctioninyourprocessorthebuilt-intemperaturesensormayberevealedeitherbytoolowortoohighafluidtemperatureorbyanincorrecttemperaturemeasurement.

Tobewarnedwhenthetemperaturemeasurementisoutofrange:

activatemonitoringonthefluidtemperatureinthefunction"acti-→vate",then

programmeatemperaturerange(in°C)outsideofwhichthe→transmittergeneratesa"warning"eventanddisplaysthe and

icons.

programmeatemperaturerange(in°C)outsideofwhichthe→transmittergeneratesan"error"eventanddisplaysthe andERR icons.

Whena"warning"or"error"eventisgeneratedbythetransmitter:

gointothe"Info"menutoreadthecauseoftheeventgeneration→and/orgointothe"Sensor"functionoftheDiagnosticmenuto→readthemeasuredfluidtemperaturevalue.

ifnecessary,checkwhetherthebuilt-inPt1000isworkingcor-→rectlybymeasuringafluidwithaknowntemperature.IfthePt1000isfaulty,returnthedevicetoBürkert.

ifthePt1000isnotthecauseoftheproblem,checktheproc-→ess.

The"warning"eventmayalsobeassociatedwithoneor•otherorbothtransistoroutputs.See§10.7.5,function"Output.TR1"or"Output.TR2".

The"error"eventmayalsobeassociatedwithoneor•otherorbothcurrentoutputs.See§10.7.4,function"Output.AC1"or"Output.AC2".

Seealso"Ifyouencounterproblems"at§11.4.•

60

Adjustmentandfunctions

english

Type 8202

Page 63: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Test menu ("Test")10.10.

modify the Test menu access code 10.10.1. ("system" menu)

Refertoparagraphs10.5and10.6toaccesstheTestmenu.

SystemOutputsSensorMEAS BACK

TestCode

MEAS BACK

System

CODE:ModifytheTestmenuaccesscode(4figures;0000bydefault).

check that the outputs are working 10.10.2. correctly ("outputs" menu)

Refertoparagraphs10.5and10.6toaccesstheTestmenu.

SystemOutputsSensorMEAS BACK

Test

MEAS BACK

Outputs

AC1AC2Tr1

Outputs

Tr2MEAS BACK

AC2Tr1

AC1:Checkthatcurrentoutput1isworkingcorrectlybyinputtingacurrentvalueandthenselecting"OK".

AC2:Checkthatcurrentoutput2isworkingcorrectlybyinputtingacurrentvalueandthenselecting"OK".

TR1:Checkthattransistoroutput1isworkingcorrectlybyselectingthestatusofthetransistor("ON"or"OFF")then"OK".

TR2:Checkthattransistoroutput2isworkingcorrectlybyselectingthestatusofthetransistor("ON"or"OFF")then"OK".

61

Adjustmentandfunctions

english

Type 8202

Page 64: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

check that the outputs are 10.10.3. behaving correctly ("sensor" menu)

Refertoparagraphs10.5and10.6toaccesstheTestmenu.

Thefeaturemakesitpossibletosimulatethemeasurementofthephysicalparametertocheckiftheoutputsarecorrectlypro-grammed.

SystemOutputsSensorMEAS BACK

Test

MEAS BACK

Sensor

PVar:Value:

PVAR:Choosethephysicalparametertobetested.

VALUE:Inputaphysicalparameterselectedfromthe"PVAR"func-tionabovetocheckoutputbehaviour.

information menu ("info")10.11.

Refertoparagraphs10.5and10.6toaccesstheInfomenu.

ErrorWarningMaintenanceMEAS BACK

Info

SoftwareMEAS BACK

InfoMaintenanceSmiley

Thismenumakesitpossibletoread:

ashortdescriptionofthereasonwhythefollowingiconsaredis-•playedbythetransmitter,ononehand:

ERROR:- ERR

WARNING:-

MAINTENANCE:-

SMILEY:- or

Seealso"Ifyouencounterproblems"at§11.4.

thesoftwareversion(function"Software")oftheacquisition/•conversionmoduleforthephysicalparametersmeasured,ontheotherhand.

62

Adjustmentandfunctions

english

Type 8202

Page 65: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

mAinTenAnce And 11. TRoubleshooTinG

safety instructions11.1.

danger

Risk of injury due to high pressure in the installation.Cutoffthepressureanddepressurizethepipesbeforeloosening•thepipes.

Risk of injury due to electrical discharge.Beforestartingwork,switchoffthepowersupplyandsecureit•topreventrestarting.Observeallapplicableaccidentprotectionandsafetyguidelines•forelectricalequipment.

Risk of injury due to the nature of the fluid.Respect theprevailing ruleson accidentprevention and safety•relatingtotheuseofaggressivefluids.

Warning

Risk of injury due to non-conforming maintenance.Maintenancemustonlybecarriedoutbyqualifiedandskilledstaff•withtheappropriatetools.Guaranteeasetorcontrolledrestartingof theprocess,aftera•powersupplyinterruption.

cleaning of the transmitter11.2.

The8202transmittercanbecleanedwithwaterorusingadeter-gentcompatiblewiththematerialsitismadeof.

PleasefeelfreetocontactyourBürkertsupplierforanyadditionalinformation.

maintenance of the ph/redox 11.2.1. sensor

danger

Risk of injury due to the nature of the cleaning or regeneration solution.Respect theprevailing ruleson accidentprevention and safety•relatingtotheuseofaggressivefluids.

noTe

the process may be polluted by the cleaning solution.After each cleaning operation, rinse the sensorwith distilled•waterandimmerseit inaKCl3Msolutionorrunningwaterfor10minutes.

lifespan of the sensor

ThelifespanofapH/redoxsensordependsonthefluidtobemeas-uredandtheconditionsinwhichmeasurementsaretaken.

Whenthefluidtobemeasuredisaggressive,forexample,and/orthefluidtemperaturereacheshighvalues,thelifespanofthesensormaybeshortened.

63

Maintenanceandtroubleshooting

english

Type 8202

Page 66: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Whenthemeasuringconditionsarefavourableandthefluidtemper-atureremainsintheregionof25°C,thesensorhasalifespanof1to3years;itdropsto3monthswhenthefluidtemperaturereaches90°C.

storing the sensor

Tostorethesensor:•

putKCl3Msolution,availableasanaccessory,intheprotective→cap

placetheprotectivecaponthesensor→Ifthesensorhasdriedoutduringstorage:•

letitsoakinaKCIsolutionforonefullnightifdehydrationis→majortoobviateincorrectmeasurementsanddrift

cleaning the sensor

Whenthesensorisdirty,cleanitaccordingtothetypeofdirt:•

cloggingbygreasesoroils:useatensioactive-basedcleaner-

limescaleormetalhydroxidedesposit:usedilutedhydrochloric-acid(10%)

sulphurousprecipitate(e.g.inpurificationstations):useamix--tureofdilutedhydrochloricacid(10%)andpepsin(saturated)

Particularcaseofasensorwithaceramicdiaphragm:•

eggwhitecolourpollution:soakthesensorinasolutioncom--posedof0.4%HCland5g/lofpepsinforseveralhours

blackcolourationofthediaphragm(silverprecipitate):immerse-thesensorinasolutioncomposedof0.4%HCland76g/lofthiourea

whentheredoxsensorispollutedbyions:•

useaquaregia,asolutioncomposedof25%nitricacidand→75%hydrochloricacid

immersethesensorinthesolutionobtainedforamaximumof→1minutetopreventthisaquaregiasolutiondissolvingtheplate

rinsethoroughly→Regenerating the sensor

ToregenerateapHsensor:•

soakthesensorinanNaOHsolution(0.1-1M)for10minutes→soakthesensorinanHCIsolution(0.1-1M)for10minutes→rinsethesensorbysoakingitinaKCl3Msolution(seeChap.→12)foratleast15minutes

Toregeneratearedoxsensor:•

cleanthemetalsurfacewithaslightlyabrasivematerialsuchas→toothpasteorveryfinescouringpowder.

64

Maintenanceandtroubleshooting

english

Type 8202

Page 67: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Replacing the sensor11.3.

noTe

the transmitter loses the ip67 protection rating when dis-mounted.Takeanyprecautionsnecessarytoprotecttheinsideofthedis-•mountedtransmitter.

Removethesensorasshownbelow.→Unscrewtheholdingnuton→thesensorarmatureandtheelectronicmodule.

Remove the electronic→modulebypullingitstraightout.Theremaybeaslightresistanceduetothetight-nessseal.

Unscrewthesensorusing→asuitablewrench.

Removeitcarefullyfromthe→armature.

Fig. 11.1 Removing the sensor from the armature

Fitanewsensorintothearmatureasshownin§7.6.→Refittheelectronicmoduletothearmatureasshownin§7.7.→

65

Maintenanceandtroubleshooting

english

Type 8202

Page 68: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

if you encounter problems11.4.

Red led

current output

transistor output

icon message displayed within the info menu

meaning What to do

ON 22mA dependingonthresholds

ERR +

"E:Sat.ORPStage"

"E:Sat.pHStage"

ThepHorredoxvalueisnotbeingcorrectlymea-suredduetothesatura-tionoftheinputstageofthemeasuringboard

Checkthecorrectwiringoftheearthpoints→checktheequipotentialsurfacesoftheinstal-→lation

ON 22mA dependingonthresholds

ERR +

"Nosensorboard" Thelinktothemeasure-mentmoduleisinter-rupted

Switchthepowersupplyoffthenonagain→iftheerrorpersists,returnthedevicetoBürkert→

ON 22mA dependingonthresholds

ERR +

"SEEFactIRead" Factorydataislost.

Theprocesscontinuesbuttheaccuracyofthedeviceismodified

Switchthepowersupplyoffthenonagain→iftheerrorpersists,returnthedevicetoBürkert→

"SEEFactRead"

ON 22mA dependingonthresholds

ERR +

"SEEUserRead" Userdataforthesensorislost(eg.typeofsensor)

Switchthepowersupplyoffthenonagain→checkthesensorparametersinallthe"Sensor"→menusthensavethemagain

iftheerrorpersists,returnthedevicetoBürkert→"SEEUserWrite"

ON 22mA dependingonthresholds

ERR +

"SPTMissing" TheconnectiontothePt1000sensorislost

CheckthatthefasteningnutbetweenthepH→sensorandtheelectronicmoduleiscorrectlyscrewed

iftheerrorpersists,returnthedevicetoBürkert→

66

Maintenanceandtroubleshooting

english

Type 8202

Page 69: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Red led

current output

transistor output

icon message displayed within the info menu

meaning What to do

ON 22mA dependingonthresholds

ERR +

"SPTRegulation" Thefluidtemperatureisnotbeingcorrectlymea-sured.

Theprocessisstopped.

Switchthepowersupplyoffthenonagain→iftheerrorpersists,returnthedevicetoBürkert→

ON 22mA dependingonthresholds

ERR +

"SRTCClock" Theclockisfaulty.

Theprocesscontinues.

ReturnthedevicetoBürkertiftheclockises-→sential.

ON 22mA dependingonthresholds

ERR +

"TRCOMMeasure" Theacquisition/conver-sionmoduleofthephysi-calparametersisfaulty.

Theprocessisstopped.

Switchthepowersupplyoffthenonagain→iftheerrorpersists,returnthedevicetoBürkert→

ON 22mA dependingonthresholds

ERR +

"TREEFactRead" Parameterreadingerror Switchthepowersupplyoffthenonagain→iftheerrorpersists,setthedevicebacktothe→defaultsettings(§10.7.1)

iftheerrorpersists,returnthedevicetoBürkert→"TREEUserRead"

ON 22mA dependingonthresholds

ERR +

"TREEUserWrite" Parametersavingerror Switchthepowersupplyoffthenonagain,→savetheparametersagain→iftheerrorpersists,setthedevicebacktothe→defaultsettings(§10.7.1)

iftheerrorpersists,returnthedevicetoBürkert→

67

Maintenanceandtroubleshooting

english

Type 8202

Page 70: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Red led

current output

transistor output

icon message displayed within the info menu

meaning What to do

OFF 4-20mA dependingonthresholds

+

"SBuffDiffORP"

"SBuffDiffpH"

"SBuffpHORP"

ThepHorredoxvalueisnotbeingcorrectlymea-suredduetothesatura-tionoftheinputstageofthemeasuringboard(commonmode).

Checkthewiring→checktheequipotentialsurfacesoftheinstal-→lation

"SRTCReinit" Thedateandtimearelostbecausethedevicehasnotbeenpoweredupforatleast5days.

Themessageisonlydis-playedatthefirstpower-ingup.

Programthedateandtime(see§10.7.1).→powerupthetransmitteratleast4hoursso→thatthedateandtimearebatterypoweredfor5days.

ON 22mA1) dependingonthresholds

ERR +

"E:ImpedanceRef" Theimpedanceofthereferenceelectrodeisoutofrange.

Themessageisdisplayedifthemonitoringoftheimpedanceoftherefer-enceelectrodehasbeenactivated,dependingontheprogrammedthresh-oldsERRLOandERRHI(see§10.9.2).

Gointothe"Sensor"functionoftheDiagnostic→menutoreadtheimpedancevalueoftherefer-enceelectrode(§10.9.2)

ifnecessary,cleanthenrecalibratethemeas-→urementsensororreplaceit

1) if the DIAG MODE function of the "Output.AC1" or "Output.AC2" menu is set to "22 mA" (see § 10.7.4); else, the current output delivers a standard current between 4 and 20 mA

68

Maintenanceandtroubleshooting

english

Type 8202

Page 71: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Red led

current output

transistor output

icon message displayed within the info menu

meaning What to do

ON 22mA1) dependingonthresholds

ERR +

"E:Imped.Glass" TheimpedanceofthepHelectrodeisoutofrange.

ThemessageisdisplayedifthemonitoringoftheimpedanceofthepHelectrodehasbeenacti-vated,dependingontheprogrammedthresholdsERRLOandERRHI(see§10.9.2).

Gointothe"Sensor"functionoftheDiagnostic→menutoreadtheimpedancevalueofthepHelectrode(§10.9.2)

ifnecessary,cleanthenrecalibratethemeas-→urementsensororreplaceit

ON 22mA1) dependingonthresholds

ERR +

"E:Temperature" Thefluidtemperatureisoutofrange.

Themessageisdisplayedifthemonitoringofthefluidtemperaturehasbeenac-tivated,dependingontheprogrammedthresholdsERRLOandERRHI(see§10.9.3).

Gointothe"Sensor"functionoftheDiagnostic→menutoreadthemeasuredfluidtemperature(§10.9.3).

ifnecessary,checkwhetherthebuilt-inPt1000→isworkingcorrectlybymeasuringafluidwithaknowntemperature.

IfthePt1000isfaulty,returnthedeviceto→Bürkert.

ifthePt1000isnotthecauseoftheproblem,→checktheprocess.

1) if the DIAG MODE function of the "Output.AC1" or "Output.AC2" menu is set to "22 mA" (see § 10.7.4); else, the current output delivers a standard current between 4 and 20 mA

69

Maintenanceandtroubleshooting

english

Type 8202

Page 72: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Red led

current output

transistor output

icon message displayed within the info menu

meaning What to do

OFF 4-20mA Switched2) +

"W:ImpedanceRef" Theimpedanceofthereferenceelectrodeisoutofrange.

Themessageisdisplayedifthemonitoringoftheim-pedanceofthereferenceelectrodehasbeenacti-vated,dependingontheprogrammedthresholdsWARNLOandWARNHI(see§10.9.2).

Gointothe"Sensor"functionoftheDiagnostic→menutoreadtheimpedancevalueoftherefer-enceelectrode(§10.9.2)

ifnecessary,cleanthenrecalibratethemeas-→urementsensororreplaceit

OFF 4-20mA Switched2) +

"W:Imped.Glass" TheimpedanceofthepHelectrodeisoutofrange.

ThemessageisdisplayedifthemonitoringoftheimpedanceofthepHelectrodehasbeenacti-vated,dependingontheprogrammedthresholdsWARNLOandWARNHI(see§10.9.2).

Gointothe"Sensor"functionoftheDiagnostic→menutoreadtheimpedancevalueofthepHelectrode(§10.9.2)

ifnecessary,cleanthenrecalibratethemeas-→urementsensororreplaceit

2) If the "PVAR" function of the "Output.TR1" and/or "Output.TR2" menus is set to "warning" (see § 10.7.5); else, the transistor outputs are operating depending on the programmed thresholds.

70

Maintenanceandtroubleshooting

english

Type 8202

Page 73: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

Red led

current output

transistor output

icon message displayed within the info menu

meaning What to do

OFF 4-20mA Switched2) +

"W:Temperature" Thefluidtemperatureisoutofrange.

Themessageisdisplayedifthemonitoringofthefluidtemperaturehasbeenac-tivated,dependingontheprogrammedthresholdsWARNLOandWARNHI(see§10.9.3).

Gointothe"Sensor"functionoftheDiagnostic→menutoreadthemeasuredfluidtemperature(§10.9.3).

ifnecessary,checkwhetherthebuilt-inPt1000→isworkingcorrectlybymeasuringafluidwithaknowntemperature.

IfthePt1000isfaulty,returnthedeviceto→Bürkert.

ifthePt1000isnotthecauseoftheproblem,→checktheprocess.

OFF 4-20mA Switched2) "M:Calib.Date" Acalibrationisdue.

Theperiodicityofthecali-brationsissetwithinthe"INTERVAL"functionofthe"CALIBINTERVAL"menu(see§10.8.3)

Calibratethesensor(§10.8.3)→

2) If the "PVAR" function of the "Output.TR1" and/or "Output.TR2" menus is set to "warning" (see § 10.7.5); else, the transistor outputs are operating depending on the programmed thresholds.

71

Maintenanceandtroubleshooting

english

Type 8202

Page 74: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

spARe pARTs And 12. AccessoRies

CauTion

Risk of injury and/or damage by the use of incorrect parts.

Incorrectaccessoriesandunsuitablereplacementpartsmaycauseinjuriesanddamagethedeviceandthesurroundingarea.Useonlyoriginalaccessoriesandoriginalreplacementpartsfrom•Bürkert.

spare part order code

SealinEPDM,Ø 46x2mm,forthesensorarmature

559169

Accessory order codeDisplaymodule 559168pHsensor,LOGOTRODEpH120mm 427114pHsensor,CERATRODEpH120mm 418319pHsensor,UNITRODEPLUSpH120mm 560376pHsensor,PLASTRODEpH120mm 560377Redoxsensor,UNITRODEPLUSRedox120mm

560378

Redoxsensor,LOGOTRODERedox120mm 560379StoragesolutionforpH/Redoxsensor(KCl3M),500ml

418557

Accessory order codeBuffersolution,500ml,pH=4 418540Buffersolution,500ml,pH=7 418541Buffersolution,500ml,pH=10 418543Buffersolution,500ml,redoxpotential=475mV

418555

M12,5-pinfemaleconnector,tobewired 917116M12,5-pinfemaleconnector,mouldedonshieldedcable(2m)

438680

M12,5-pinmaleconnector,mouldedonshieldedcable(2m)

559177

72

Sparepartsandaccessories

english

Type 8202

Page 75: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

pAckAGinG, TRAnspoRT13.

noTe

damage due to transport

Transportmaydamageaninsufficientlyprotecteddevice.Transportthedeviceinshock-resistantpackagingandawayfrom•humidityanddirt.Avoidtheeffectsofheatandcoldwhichcouldcausethestorage•temperaturerangetobeexceeded.Protecttheelectricalinterfacesbyusingprotectioncaps.•

sToRAGe14.

noTe

poor storage can damage the device.Storethedeviceinadryplaceawayfromdust.•Storagetemperatureofthe8202:-10to+60°C.•Storagetemperatureofthesensor:see§6.3.5orcorresponding•instructionmanual.

73

Packaging,Transport

english

Type 8202

Page 76: Type 8202 - Kele Gas_Specialty_Sensors/PDFs/DS8202... · 2011. 11. 14. · Documentation Datasheets or Manuals/Ap-provals Type 8202 The complete manual can be ordered under code:

disposAl oF The pRoducT15. Disposeof thedeviceand itspackaging inanenvironmentally-→friendlyway.

noTe

damage to the environment caused by products contaminated by fluids.Keeptotheexistingprovisionsonthesubjectofwastedisposal•andenvironmentalprotection.

please note

Complywiththenationaland/orlocalregulationswhichconcerntheareaofwastedisposal.

74

Disposaloftheproduct

english

Type 8202