  • Operating Instructions

    TypeMS01

    pH sensor cube

  • We reserve the right to make technical changes without notice.Technische Änderungen vorbehalten.Sous réserve de modifications techniques.

    © Bürkert SAS, 2014

    Operating Instructions 1409/0_EU-ML 00566496 Original EN

  • 1

    Contents

    1 AbouttheoperAtinginstructions...........................................................................................................................4

    1.1 symbolsused..........................................................................................................................................................................4

    1.2 Definitionoftheword"product"....................................................................................................................................4

    1.3 Definitionoftheword"system"......................................................................................................................................4

    1.4 Definitionoftheword"büs".............................................................................................................................................5

    2 intenDeDuse....................................................................................................................................................................................5

    3 bAsicsAfetyinformAtion....................................................................................................................................................6

    4 generAlinformAtion................................................................................................................................................................7

    4.1 contact........................................................................................................................................................................................7

    4.2 Warrantyconditions..............................................................................................................................................................7

    4.3 informationsontheinternet............................................................................................................................................7

    5 Description.......................................................................................................................................................................................8

    6 technicAlDAtA................................................................................................................................................................................9

    6.1 conditionsofuse...................................................................................................................................................................9

    6.2 conformitytostandardsanddirectives....................................................................................................................9

    6.3 materialstheproductismadeof..................................................................................................................................9

    6.4 fluiddata....................................................................................................................................................................................9

    6.5 measurementdata.............................................................................................................................................................10

    6.6 electricaldata.......................................................................................................................................................................10

    6.7 communication....................................................................................................................................................................10

    7 instAllAtion...................................................................................................................................................................................11

    7.1 safetyinstructions.............................................................................................................................................................11

    7.2 mountingtheproductonthebackplane................................................................................................................11

    8 ADjustmentAnDoperAtion..............................................................................................................................................12

    8.1 safetyinstructions.............................................................................................................................................................12

    8.2 howtoadjusttheproduct.............................................................................................................................................12

    8.3 Adjustmentsthatmustbedone.................................................................................................................................13

    8.4 generalinformationonthedisplaysoftwaretypeme21..............................................................................13

    Type MS01

    English

  • 2

    Contents

    8.5 "Device"viewoftheproduct.........................................................................................................................................14

    8.6 "function"viewoftheproduct.....................................................................................................................................15

    8.7 Detailedviewsofthebüsfunction...........................................................................................................................16

    8.8 "parameter"viewofthesensor...................................................................................................................................17

    8.8.1 Configuringthetriggeringofthewarnings...................................................................................17

    8.8.2 Configuringthetriggeringoftheerrors........................................................................................18

    8.9 "Diagnosis"viewofthesensor....................................................................................................................................18

    8.10 "maintenance"viewofthesensor..............................................................................................................................19

    8.10.1 CalibratingtheoffsetvalueoftheISFETsensor.........................................................................19

    8.10.2 Calibratingtheoffsetand/orslopevaluesofthepHsensor....................................................21

    8.10.3 Doinga1pointcalibrationprocedureofthepHsensor...........................................................22

    8.10.4 Doinga2pointcalibrationprocedureofthepHsensor...........................................................23

    8.10.5 Adjustingtheoffsetvalueofthetemperature..............................................................................27

    8.10.6 Simulatingsomedata........................................................................................................................27

    8.10.7 Readingthedateofthelastcalibration........................................................................................27

    8.10.8 Readingthedateofthenextduecalibration...............................................................................27

    8.10.9 Settingthetimeintervalbetweentwocalibrations.....................................................................27

    9 mAintenAnceAnDtroubleshooting.......................................................................................................................28

    9.1 safetyinstructions.............................................................................................................................................................28

    9.2 cleaningoftheproduct...................................................................................................................................................28

    9.3 replacingtheexternalreferenceelectrode.........................................................................................................28

    9.4 troubleshoutingifnomessageisdisplayed.......................................................................................................29

    9.5 troubleshootingifthestatusleDoftheproductisredororange........................................................30

    9.5.1 Message"Toohighleakagecurrent".............................................................................................30

    9.5.2 Message"Toolowleakagecurrent"...............................................................................................31

    9.5.3 Message"ToohighpHvalue".........................................................................................................31

    9.5.4 Message"ToolowpHvalue"...........................................................................................................32

    9.5.5 Message"Toohightemperaturevalue".........................................................................................33

    9.5.6 Message"Toolowtemperaturevalue"..........................................................................................33

    9.5.7 Message"Toohighreferenceimpedance"...................................................................................34

    9.5.8 Message"Toolowreferenceimpedance"....................................................................................35

    10 spArepArtsAnDAccessories......................................................................................................................................36

    English

  • 3

    11 pAckAging,trAnsport..........................................................................................................................................................36

    12 storAge..............................................................................................................................................................................................36

    12.1 tostoretheproductforamaximumof4days..................................................................................................36

    12.2 tostoretheproductformorethan4daysandlessthan1month.........................................................37

    12.3 tostoretheproductforamorethan1month...................................................................................................37

    12.4 Afterstorage..........................................................................................................................................................................37

    13 DisposAloftheproDuct..................................................................................................................................................37

    English

    Type MS01

  • 4

    AbouttheOperatingInstructions

    1 AbouT The operATing inSTrucTionSTheOperatingInstructionsdescribetheentirelifecycleoftheproduct.PleasekeeptheOperatingInstructionsinasafeplace,accessibletoallusersandanynewowners.

    theoperatinginstructionscontainimportantsafetyinformation.

    Failuretocomplywiththeseinstructionscanleadtohazardoussituations.

    ▶ TheOperatingInstructionsmustbereadandunderstood.

    1.1 Symbols used

    danger

    Warnsagainstanimminentdanger.

    ▶ Failuretoobservethiswarningcanresultindeathorinseriousinjury.

    Warning

    Warnsagainstapotentiallydangeroussituation.

    ▶ Failuretoobservethiswarningcanresultinseriousinjuryorevendeath.

    attention

    Warnsagainstapossiblerisk.

    ▶ Failuretoobservethiswarningcanresultinsubstantialorminorinjuries.

    note

    Warnsagainstmaterialdamage.

    ▶ Failuretoobservethiswarningmayresultindamagetotheproductorsystem.

    Indicatesadditionalinformation,adviceorimportantrecommendations.

    ReferstoinformationcontainedintheOperatingInstructionsorinotherdocuments.

    ▶ Indicatesaninstructiontobecarriedouttoavoidadanger,awarningorapossiblerisk.

    → Indicatesaproceduretobecarriedout.

    » Indicatestheresultofaspecificinstruction.

    1.2 Definition of the word "product"

    Theword"product"usedwithintheseOperatingInstructionsalwaysreferstothepHsensorcubetypeMS01.

    1.3 Definition of the word "system"

    Theword"system"usedwithintheseOperatingInstructionsalwaysreferstotheOnlineAnalysisSystemtype8905.

    English

    Type MS01

  • 5

    AbouttheOperatingInstructions

    1.4 Definition of the word "büS"

    Theword"büS"usedwithintheseOperatingInstructionsalwaysreferstothefieldbusdeveloppedbyBürkert.

    2 inTenDeD uSe

    useofthisproductthatdoesnotcomplywiththeinstructionscouldpresentriskstopeople,nearbyinstallationsandtheenvironment.

    ▶ TheproductisintendedsolelyforthemeasurementofthepHinwaterwithina8905system.▶ Thisproductmustbeprotectedagainstelectromagneticinterference,ultravioletraysand,wheninstalledoutdoors,theeffectsofclimaticconditions.

    ▶ ThisproductmustbeusedincompliancewiththecharacteristicsandcommissioninganduseconditionsspecifiedinthecontractualdocumentsandintheOperatingInstructions.

    ▶ Requirementsforthesafeandproperoperationoftheproductarepropertransport,storageandinstallation,aswellascarefuloperationandmaintenance.

    ▶ Onlyusetheproductasintended.

    ▶ Observeanyexistingrestraintswhentheproductisexported.

    English

    Type MS01

  • 6

    Basicsafetyinformation

    3 bASic SAfeTy inforMATionThissafetyinformationdoesnottakeintoaccount:

    • anycontingenciesoroccurrencesthatmayariseduringassembly,useandmaintenanceoftheproduct.

    • thelocalsafetyregulationsthattheoperatormustensurethestaffinchargeofinstallationandmaintenanceobserve.

    Variousdangeroussituations.

    Toavoidinjurytakecare:

    ▶ topreventanyunintentionalpowersupplyswitch-on.

    ▶ tocarryouttheinstallationandmaintenanceworkbyqualifiedandskilledstaffwiththeappropriatetools.

    ▶ tousetheproductonlyifinperfectworkingorderandincompliancewiththeinstructionsprovidedintheseOperatingInstructions.

    ▶ toobservethegeneraltechnicalrulesduringtheplanninganduseoftheproduct.

    ▶ nottousethisproductinexplosiveatmospheres.

    ▶ nottousethisproductinanenvironmentincompatiblewiththematerialsfromwhichitismade.

    ▶ nottomakeanyexternalorinternalmodificationstotheproduct.

    note

    elements/componentssensitivetoelectrostaticdischarges

    • Thisproductcontainselectroniccomponentssensitivetoelectrostaticdischarges.Theymaybedamagediftheyaretouchedbyanelectrostaticallychargedpersonorobject.Intheworstcasescenario,thesecompo-nentsareinstantlydestroyedorgooutoforderassoonastheyareactivated.

    • Tominimiseorevenavoidalldamageduetoanelectrostaticdischarge,takealltheprecautionsdescribedintheEN61340-5-1norm.

    • Alsoensurethatyoudonottouchanyoftheliveelectricalcomponents.

    English

    Type MS01

  • 7

    Generalinformation

    4 generAl inforMATion

    4.1 contact

    Tocontactthemanufactureroftheproductusefollowingaddress:

    BürkertSAS

    RueduGiessen

    BP21

    F-67220TRIEMBACH-AU-VAL

    TheaddressesofourinternationalbranchescanbefoundontheInternetat:www.burkert.com

    4.2 Warranty conditions

    TheconditiongoverningthelegalwarrantyistheconforminguseoftheproductinobservanceoftheoperatingconditionsspecifiedintheseOperatingInstructions.

    4.3 informations on the internet

    YoucanfindtheOperatingInstructionsandtechnicaldatasheetsregardingthetypeMS01at:www.burkert.com

    English

    Type MS01

    http://burkert.comhttp://burkert.com

  • 8

    Description

    5 DeScripTionThepHsensorcubeisusedinthesystemtype8905.

    Theelectricalandfluidconnectionsaremadeviathebackplaneofthesystemtype8905.

    Leverto:

    - lock/unlocktheproduct.

    - carryoutmaintenanceoperations.

    Unlockedposition

    Lockedposition

    Maintenanceposition

    Push-buttonwithmea-suredprocessvariable

    Fluidconnections

    Electricalinterface

    Pin

    Adaptationpins

    Housingoftheexternalreferenceelectrode

    SensorcubestatusLED

    Slotforthemicro-SIMcard

    Fig. 1: Description of the product

    English

    Type MS01

  • 9

    Technicaldata

    6 TechnicAl DATA

    6.1 conditions of use

    Ambienttemperature 0to+40°C

    Airhumidity <90%,withoutcondensation

    Protectionratingacc.toEN60529

    • IP65,whenpluggedinthebackplane

    • IP20,asstandaloneproduct

    Max.heightabovesealevel 2000m

    6.2 conformity to standards and directives

    TheproductconformstotheCEdirectivesofthesystemtype8905,onlywhentheproducttypeMS01ispluggedinthesystemtype8905.

    6.3 Materials the product is made of

    part material

    Housing PPE+PS

    Seal EPDM

    Lever PC

    6.4 fluid data

    • Typeoffluid

    • pHvalue

    • Water,withoutparticles:drinkingwater,industrialwater

    • pH4to9Minimalflowrate 3l/h;recommended6l/h

    Watersamplepressure PN6

    Watersampletemperature 0to+40°C,notfreezing

    English

    Type MS01

  • 10

    Technicaldata

    6.5 Measurement data

    pHmeasurement

    • Measuringrange

    • Sensorresolution

    • Measurementdeviation("measurementbias",asdefinedinthestandardJCGM200:2012)

    • Linearity

    • Repeatability

    • Responsetime(t90)

    • Measurementsensor

    • Electrolyteoftheexternalreferenceelectrode

    • pH4to9

    • pH0.02

    • pH±0.1

    • pH±0.05

    • pH±0.05

    • <10s

    • ISFET

    • 3molKClTemperaturemeasurement

    • Measuringrange

    • Measurementsensor

    • 0to50°C

    • Pt1000ClassB,nocontactwiththewatersampleMaintenanceintervaloftheexternalreferenceelectrode 12months,nominal,dependingonthewaterquality

    6.6 electrical data

    Operatingvoltage 24VDCthroughthebackplaneofthesystemtype8905

    Powerconsumption 0.8VA

    6.7 communication

    Internalcommunication throughbüS

    ExternalcommunicationbystatusLED accordingtoNAMURNE107

    English

    Type MS01

  • 11

    Installation

    7 inSTAllATion

    7.1 Safety instructions

    note

    riskofdamagetotheproductduetonon-conforminginstallation.

    • Theelectricalandfluidicinstallationscanonlybecarriedoutbyqualifiedandskilledstaffwiththeappropriatetools.

    • Respecttheinstallationinstructionsforthesystem.

    note

    riskofdamagetotheproductduetothepowersupply

    • Shutdownandisolatetheelectricalpowersourcebeforecarryingoutworkonthesystem.

    note

    riskofdamagetotheproductduetotheenvironment

    • Protecttheproductagainstelectromagneticinterference,ultravioletraysand,wheninstalledoutdoors,theeffectsoftheclimaticconditions.

    7.2 Mounting the product on the backplane

    TheproductispluggedinthebackplaneofthesystemType8905.

    4

    3

    3

    1. Drythesurfaceofthebackplanethatwillbeincontactwiththeproduct.

    2. Drythesurfaceoftheproduct.

    3. Whilethepush-buttonispushed,turnthebayonetlevertotheright,

    ontheunlockedposition,.Donotpushthelevertothemaintenanceposition.

    4. Insertthetwoadaptationpinsintheirholesandthenplugtheproductinthebackplane.

    5. Whilethepush-buttonispushed,turnthebayonetlevertotheleft,

    onthelockedposition, .

    Fig. 2: Mounting a product on the backplane of the system

    English

    Type MS01

  • 12

    Adjustmentandoperation

    8 ADjuSTMenT AnD operATion

    8.1 Safety instructions

    note

    riskofdamagetotheproductduetonon-conformingadjustment.

    • TheoperatorsinchargeofadjustmentmusthavereadandunderstoodthecontentsoftheseOperatingInstructions.

    • TheoperatorsinchargeofadjustmentmusthavereadandunderstoodthecontentsoftheOperatingInstruc-tionsofthedisplaysoftwaretypeME21and/orthecontentsoftheOperatingInstructionsoftheBürkertCommunicatorsoftwaretype8920and/ortheOperatingInstructionsofthecontrollermoduletypeME25.

    • Inparticular,observethesafetyrecommendationsandintendeduse.

    • Theproduct/installationmustonlybeadjustedbysuitablytrainedstaff.

    note

    riskofdamagetotheproductduetonon-conformingcommissioning.

    • Beforecommissioning,makesurethatthestaffinchargehavereadandfullyunderstoodthecontentsoftheseOperatingInstructions.

    • Inparticular,observethesafetyrecommendationsandintendeduse.

    • Theproduct/theinstallationmustonlybecommissionedbysuitablytrainedstaff.

    note

    riskofdamagetotheproductduetonon-conformingoperation.

    • TheoperatorsinchargeofoperationmusthavereadandunderstoodthecontentsoftheseOperatingInstructions.

    • Inparticular,observethesafetyrecommendationsandintendeduse.

    • Theproduct/installationmustonlybeoperatedbysuitablytrainedstaff.

    8.2 how to adjust the product

    Theadjustmentoftheproductcanbemade:

    • eitherwiththedisplayofthesystemtype8905.ThedisplayofthesystemismanagedbythesoftwaretypeME21.Seechap.8.4.

    • orwithaPCandtheBürkertCommunicatorsoftwaretype8920.Togetgeneralinformationaboutthesoftwaretype8920,refertotheOperatingInstructionsofthetype8920.

    English

    Type MS01

  • 13

    Adjustmentandoperation

    8.3 Adjustments that must be done

    Dotheadjustments:

    • aftertheproducthasbeeninstalledinthesystem.

    • afterthefluidicandelectricalinstallationshavebeenmadeonthesystem.

    • afterthetightnessofthesystemhasbeenchecked.

    • afterthesystemhasoperatedforapolarizationtimeof12hours.

    • Beforecommissioningtheproduct/systemforthefirsttimeandforthecorrectoperationoftheproductandofthesystem,dothefollowingadjustmentsfortheproduct:

    - calibratetheoffsetvalueofthepHsensor(refertochap.8.10)or,

    - calibratetheoffsetvalueandtheslopevalueofthepHsensor(refertochap.8.10).

    • Beforecommissioningtheproduct/systemafteramaintenanceoperation(i.e.replacementoftheexternalref-erenceelectrodeorreplacementoftheISFETsensor),letthesystemoperateforapolarizationtimeof12hoursanddothefollowingadjustmentsfortheproduct:

    - ifthepHsensorhasbeenreplaced,calibratetheoffsetvalueandtheslopevalueofthepHsensor(refertochap.8.10).

    - iftheISFETsensorhasbeenreplaced,calibratetheoffsetvalueoftheISFETsensorandcalibratetheoffsetvalueandtheslopevalueofthepHsensor(refertochap.8.10).

    • Calibratetheproducteverymonth.

    8.4 general information on the display software type Me21

    TheseOperatingInstructionsexplaintheadjustmentsthatarespecifictotheproducttypeMS01.

    → TogetgeneralinformationaboutthedisplaysoftwaretypeME21,refertotheOperatingInstructionsofthetypeME21thatisontheCDdeliveredwiththesystemandthatisalsoavailableatwww.burkert.com.

    TheOperatingInstructionsofthedisplaysoftwaretypeME21:

    • givegeneralinformationonthesoftware,suchas:descriptionoftheuserinterface,structureofthemenus,descriptionofthepossibleviews("Device"viewforexample),descriptionofthenavigationbuttons...

    • explainhowtomakethegeneraladjustmentssuchas:thedisplaylanguage,thelocatingoftheproduct...

    • explainhowtoconfigureandcustomizethe"Desktop"viewswithvaluesorgraphs.

    • givegeneralinformationontheerrormessagesandtheoperatingofthesystemstatuslight.

    English

    Type MS01

    http://burkert.com

  • 14

    Adjustmentandoperation

    8.5 "Device" view of the product

    The"Device"viewshowssomeofthemeasurementdatarelatedtotheproduct.

    → Ifthedisplayshowsa"Desktop"view,tap toaccessthe"Device"view.

    → Todisplaythe"Device"viewoftheproduct,selecttheproductinthelistofdevicesontheleftofthedisplay.

    Table 1: "Device" view of the product

    Controller

    5648300001

    Device 5648300001

    Location TreatmentunitNr1 Status N/A

    pH pH7.00 pHvoltage 0.00V

    Temperature 24.10°CBoardtemperature 26.20°C

    Deviceview

    HMIU

    12.08.201416:55

    pH

    ListofdevicesthatareconnectedtothebüS

    Thefollowingdatacanbereadfromthe"Device"viewoftheproduct:

    • thedevicesthatareconnectedonthebüSwiththeir"Uniquedevicename".Bydefault,the"Uniquedevicename"isbuiltupwithtwonumbers:theordercodeoftheproduct(forexample564830,fortheproduct)andtheseriesnumber(0001,forexample).

    • wheretheproductisgeographicallylocated.

    • themeasuredvalueofthepHofthewatersample.

    • themeasuredvalueofthepotentialdifference.

    • themeasuredvalueofthetemperatureofthewatersample.

    • themeasuredvalueofthetemperatureoftheinternalmeasurementboard.

    → Todisplaythe"Function"viewoftheproduct,tap .

    English

    Type MS01

  • 15

    Adjustmentandoperation

    8.6 "function" view of the product

    The"Function"viewshowsthefunctionsavailableforaproductand,foreachfunction,themaindatarelatedtoeachfunction.

    Todisplaythe"Function"viewoftheproduct:

    1. selecttheproductinthelistofdevices,ontheleftofthedisplay,

    2. tap .

    Table 2: "Function" view of the product

    büS

    function büS

    Uniquedevicename 5648300001 Location TreatmentUnitNr1

    Description Springwater

    IsfetSensor5648300001 12.08.201416:55

    pH Sensor

    "function"viewofthebüs

    → Todisplaythe"Function"viewofthebüStheproductisconnectedto,select"büS"inthelistofdevicesontheleftofthedisplay.

    → Tap toaccessthe"Parameter"view.Seechap.8.7.

    Sensor

    function Sensor

    Measurement pH7.00 pHvoltage 0.00V

    Slope -55.00mV/pHOffset pH7.00

    Temperature 24.10°COffset 0.00°C

    IsfetSensor5648300001 12.08.201416:55

    büS

    pH

    "function"viewofthesensor

    → Todisplaythe"Function"viewofthesensor,select"Sensor"inthelistofdevicesontheleftofthedisplay.

    → Tap toaccessthe"Parameter"view.Seechap.8.8.

    English

    Type MS01

  • 16

    Adjustmentandoperation

    8.7 Detailed views of the büS function

    Table 3: "Parameter" view of the "büS" function

    parameter büS

    BasicSettings UniqueDeviceName 5648300001

    TreatmentUnitNr1

    0

    0

    Location

    Description

    Physicalgroup

    Logicalgroup

    ISFETsensor5648300001 04.04.201412:21

    Groupingsettings

    "parameter"detailedviewofthebüs

    Tosettheuser-specificdatafortheidentificationofthedeviceonthebüS:

    • UniqueDeviceNameoftheproduct.

    → MakesureyouchooseauniquenamefortheproductbecausethebüSmustrecognizetheproduct.

    → Makesureyouchooseauniquenamethatisselfexplanatorytoidentifytheproductbecause,iftheuniquenameischanged,allthesettingsmadeonthebüSmustbechanged.

    • Description

    • Location

    • Groupingsetting

    Table 4: "Diagnosis" view of the "büS" function

    Diagnosis büS

    büSinformation Address 126

    500kbit/s

    ISFETsensor

    564830

    683328

    A.00.02.42

    A.01.00.00

    Baudrate

    Devicename

    Identitynumber

    Softwareidentitynumber

    Softwareversion

    Hardwareversion

    ISFETsensor5648300001 04.04.201412:21

    Deviceinformation

    "Diagnosis"detailedviewofthebüs

    ToreadsomebüSdataanddevicedata:

    • Addressandbaudrate

    • Devicenameanddeviceordercode(Identitynumber)

    • Softwareandhardwareinformation

    • InformationrelatingtotheTFT

    • Devicedriverinformation

    Nochangesbyuserarepossible.

    English

    Type MS01

  • 17

    Adjustmentandoperation

    Table 5: "Maintenance" view of the "büS" function

    Maintenance büS

    Resettypes Device Reset

    ISFETsensor5648300001 12:21:16"maintenance"detailedviewofthebüs

    Restarttheproductforareset.

    8.8 "parameter" view of the sensor

    → Inthe"Function"view,tap toaccessthe"Parameter"view.

    parameter Sensor

    Temperature Desiredunit °C

    ON

    2s

    Measurement

    Responsetime

    Warnings

    Errors

    Messagestriggers

    IsfetSensor5648300001 12:21:16

    Referenceimpedance

    Filter

    The"Parameter"viewofthesensormakesitpossible:

    • tochangetheunitsofthedisplayedtemperature.

    • toselectwhethertheimpedanceofthereferenceelectrodeismeasuredornot.

    • toenterthevalueoftheT10-90responsetimeforthemeasuredsignal.

    • toconfigurethetriggeringofthewarnings,seechap.8.8.1.

    • toconfigurethetriggeringoftheerrors,seechap.8.8.2.

    Table 6: "Parameter" view of the sensor

    8.8.1 configuring the triggering of the warnings

    Thevaluesofsomeprocessvariablesandthevaluesofsomesensorcubevariablescanbemonitoredtodetectaproblemintheprocessoronthesensor.

    ForthepHsensorcube,thefollowingvaluescanbemonitored:

    • thepHofthewatersample,

    • thetemperatureofthewatersample,

    • theimpedanceofthereferenceelectrode,

    • theleakagecurrentoftheISFETsensor.

    English

    Type MS01

  • 18

    Adjustmentandoperation

    Procedure:

    1. TapWarnings inthe"Parameter"viewofthesensor.

    2. TapActivationflags inthe"SensorParameter"view.

    3. Selecttheprocessvariablesandthesensorcubevaluestobemonitoredandvalidate.

    4. Settheminimumand/ormaximumthresholdvalues.

    Whentheparameteredthresholdvaluesarereachedandtheactivationflagoftherelatedprocessvariableor

    sensorcubevalueisselected,awarningmessageisgeneratedinthe"MessagesList":tap toaccessthe"MessagesList".Seechap.9.4forthetroubleshootingdetails.

    8.8.2 configuring the triggering of the errors

    Thevaluesofsomeprocessvariablesandthevaluesofsomesensorcubevariablescanbemonitoredtodetectaproblemintheprocessorontheproduct.

    ForthepHsensorcube,thefollowingvaluescanbemonitored:

    • thepHofthewatersample,

    • thetemperatureofthewatersample,

    • theimpedanceofthereferenceelectrode,

    • theleakagecurrentoftheISFETsensor.

    Procedure:

    1. Taperrors inthe"Parameter"viewofthesensor.

    2. TapActivationflags inthe"Parameter"viewofthesensor.

    3. Selecttheprocessvariablesandthesensorcubevaluestobemonitoredandvalidate.

    4. Settheminimumand/ormaximumthresholdvalues.

    Whentheparameteredthresholdvaluesarereachedandtheactivationflagoftherelatedprocessvariableor

    sensorcubevalueisselected,anerrormessageisgeneratedinthe"MessagesList":tap toaccessthe"MessagesList".Seechap.9.4forthetroubleshootingdetails.

    8.9 "Diagnosis" view of the sensor

    The"Diagnosis"viewmakesitpossibletoreadthefollowingvalues:

    • themeasuredvalueofthepHofthewatersample.

    • themeasuredvalueofthepotentialdifference.

    • thevalueoftheoffsetoftheISFETsensor,calculatedthroughcalibrationorenteredinthe"Maintenance"view.

    • thevalueoftheslopeofthepHsensor,calculatedthroughcalibrationorenteredinthe"Maintenance"view.

    • thevalueoftheoffsetofthepHsensor,calculatedthroughcalibrationorenteredinthe"Maintenance"view.

    • themeasuredvalueofthewatersampletemperature.

    • themeasuredimpedanceofthereferenceelectrode.

    English

    Type MS01

  • 19

    Adjustmentandoperation

    • thetemperatureoffsetthathasbeenenteredinthe"Maintenance"viewofthesensor.

    • themeasuredvalueofthetemperatureofthemeasurementboard.

    • themeasuredvalueoftheISFETleakagecurrent.

    • thetimetheproducthasalreadyoperated.

    • thecalibrationlimitsoftheoffsetvalueoftheISFETsensor.

    • thecalibrationlimitsoftheoffsetvalueofthepHsensor.

    • thecalibrationlimitsoftheslopevalueofthepHsensor.

    → Fromthe"Parameter"viewofthesensor,tap toaccessthe"Diagnosis"view.

    8.10 "Maintenance" view of the sensor

    The"Maintenance"viewmakesitpossible:

    • tomodifytheoffsetvalueoftheISFETsensor,

    • tomodifytheoffsetvalueofthepHsensor,

    • tomodifytheslopevalueofthepHsensor,

    • toautomaticallycalibratethesensor,withthecalibrationwizard,

    • tomodifytheoffsetvalueofthewatersampletemperature,

    • tochecktheoperatingoftheproductbysimulatingsomedata,

    • toreadthedateofthelastcalibration,

    • toreadthedateofthenextduecalibration,

    • tosetthenumberofdaysbetweentwocalibrations.

    → Fromthe"Parameter"viewofthesensor,tap toaccessthe"Maintenance"view.

    8.10.1 calibrating the offset value of the iSfeT sensor

    TheoffsetoftheISFETsensoriscalibratedinthefactory.

    IftheISFETsensormustbereplacedbyanewone,thenewISFETsensormustbecalibrated.

    → IfyouknowtheoffsetvalueofthenewISFETsensor,gotothe"Maintenance"viewofthesensor,tapphcali-brationwizardisfetoffset andentertheoffsetvalueoftheISFETsensor.

    → IfyoudonotknowtheoffsetvalueofthenewISFETsensor,dothefollowingcalibrationprocedure.

    1. Connectaperistalticpumpinthewatersampleinletcircuitofthesystem,asshowninFig.3.Ifthefluidicconnectionsarenotdonecorrectly,theproductandthesystemcanbedamagedbecausethepressureintheproductandinthesystemistoohigh.

    English

    Type MS01

  • 20

    Adjustmentandoperation

    24 VDC

    +-

    OUT IN

    Peristalticpump

    Fig. 3: Installation for the calibration of the sensor in a system type 8905

    2. Makesurethedirectionoftheflowiscorrect.

    3. Makesurethefluidicinstallationforthecalibrationistight.

    4. Letcleanwaterflowthroughthesystemtorinsetheproduct.

    5. PrepareabuffersolutionwithapH7value.

    6. Inthe"Maintenance"viewofthesensor,tapphcalibrationwizardZeroadjustment .

    7. Step1/5:letthebuffersolutionflowthroughtheproduct.

    8. Tap .

    9. Step2/5:tapinputbuffervalue andenterthepHvalue7ofthebuffersolution.

    10. Validate.

    11. Tap .

    12. Step3/5:whenthepHmeasurementisstable,tap .

    English

    Type MS01

  • 21

    Adjustmentandoperation

    13. Step4/5:

    • ifthecalibrationhassucceeded,thecalculatedoffsetvalueisdisplayed.Tap togotothestep5/5.

    • ifanerrormessageisdisplayed,refertoTable7.

    • ifawarningmessageisdisplayed,refertoTable8.

    14. Step5/5:validateorcancelthecalibration.

    Table 7: Error message at the end of the calibration of the ISFET sensor

    Displayedmessage ComputedcalibrationvalueoutoferrorlimitssymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thesensorcalibrationhasfailedbecausethecalculatedoffsetvalueis

    outoftheerrorrange.Whattodo? → SendtheproductbacktoBürkert.

    Table 8: Warning message at the end of the calibration of the ISFET sensor

    Displayedmessage ComputedcalibrationvalueoutofwarninglimitssymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thecalculatedoffsetvalueisoutofthewarningrangebecauseawrong

    buffersolutionhasbeenusedforthecalibration.Whattodo? → Makesurethebuffersolutionusedisthecorrectone.

    → Ifitisnotthecorrectone,abortthecalibration,useacorrectbuffersolutionanddothecompletecalibrationprocedureagain.

    → Youcanchoosetoeithervalidateorcancelthecalibration.

    • Ifyouvalidatethecalibration,thenewcalculatedoffsetvalueisusedtodeterminethepHvalue,andthelastcalibrationdateisupdated.

    • Ifyoucancelthecalibration,thecurrentoffsetvalueisusedtodeterminethepHvalue,andthelastcalibrationdateisnotupdated.

    8.10.2 calibrating the offset and/or slope values of the ph sensor

    TomeasurepHvalueswithaslessdeviationaspossibleyoumustcalibratethepHsensor.

    Thiscanbedone:

    • eitherbyadjustingbyhandtheslopevalueand/ortheoffsetvalueofthepHsensor:inthe"Maintenance"viewofthesensor,tapphslope andenteraslopevalueand/ortapphoffset andenteranoffsetvalue.

    • orbydoinga1pointcalibrationproceduretoautomaticallyadjusttheoffsetvalueofthepHsensor.Seechap.8.10.3.

    • orbydoinga2pointcalibrationproceduretoautomaticallyadjusttheoffsetvalueandtheslopevalueofthepHsensor.Seechap.8.10.4.

    English

    Type MS01

  • 22

    Adjustmentandoperation

    8.10.3 Doing a 1 point calibration procedure of the ph sensor

    Dangerduetothenatureofthefluid

    ▶ Respecttheregulationsonaccidentpreventionandsafetyrelatingtotheuseofaggressivefluids.

    Doa1pointcalibrationproceduretoautomaticallyadjusttheoffsetvalueofthepHsensor.

    1. Connectaperistalticpumpinthewatersampleinletcircuitofthesystem,asshowninFig.3,chap.8.10.1.Ifthefluidicconnectionsarenotdonecorrectly,theproductandthesystemcanbedamagedbecausethepressureintheproductandinthesystemistoohigh.

    2. Makesurethedirectionoftheflowiscorrect.

    3. Makesurethefluidicinstallationforthecalibrationistight.

    4. Letcleanwaterflowthroughthesystemtorinsetheproduct.

    5. PrepareabuffersolutionwithapHvalueascloseaspossibletothewatersampleandwithapHvaluethatiscompatiblewithallthesensorcubesofthesystem.

    6. Inthe"Maintenance"viewofthesensor,tapphcalibrationwizard1point .

    7. Step1/5:letthebuffersolutionflowthroughtheproduct.

    8. Tap .

    9. Step2/5:tapinputbuffervalue andenterthepHvalueofthebuffersolution.

    10. Validate.

    11. Tap .

    12. Step3/5:whenthepHmeasurementisstable,tap .

    13. Step4/5:

    • ifthecalibrationhassucceeded,thecalculatedoffsetvalueisdisplayedandthedateofthelastcalibrationis

    updated(seechap.8.10.7).Tap togotothestep5/5.

    • ifanerrormessageisdisplayed,refertoTable9.

    • ifawarningmessageisdisplayed,refertoTable10.

    14. Step5/5:validateorcancelthecalibration.

    English

    Type MS01

  • 23

    Adjustmentandoperation

    Table 9: Error message at the end of the 1 point calibration of the pH sensor

    Displayedmessage ComputedcalibrationvalueoutoferrorlimitssymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thesensorcalibrationhasfailedbecausethecalculatedoffsetvalueis

    outoftheerrorrange.Whattodo? 1. Comparethecalculatedoffsetvaluewiththecalibrationlimitsthat

    canbereadinthe"Diagnosis"view.

    2. Replacetheexternalreferenceelectrodebyanewone.

    3. Dothecalibrationagain.

    4. Ifthecalibrationfailsagain,replacetheproductbyanewone.

    Table 10: Warning message at the end of the 1 point calibration of the pH sensor

    Displayedmessage ComputedcalibrationvalueoutofwarninglimitssymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thecalculatedoffsetvalueisoutofthewarningrangebecause:

    • eitherawrongbuffersolutionhasbeenusedforthecalibration.

    • ortheexternalreferenceelec-trodeisaging.

    Whattodo? 1. Makesurethebuffersolutionusedisthecorrectone.

    2. Ifitisnotthecorrectone,abortthecalibration.

    3. Useacorrectbuffersolutionanddothecompletecali-brationprocedureagain.

    → Youcanchoosetoeitherval-idateorcancelthecalibration.

    • Ifyouvalidatethecalibration,thenewcalculatedoffsetvalueisusedtodeterminethepHvalue,andthelastcalibrationdateisupdated.

    • Ifyoucancelthecalibration,thecurrentoffsetvalueisusedtodeterminethepHvalue,andthelastcalibrationdateisnotupdated.

    → Plantoreplacetheexternalreferenceelectrode.

    8.10.4 Doing a 2 point calibration procedure of the ph sensor

    Dangerduetothenatureofthefluid

    ▶ Respecttheregulationsonaccidentpreventionandsafetyrelatingtotheuseofaggressivefluids.

    Doa2pointcalibrationproceduretoautomaticallyadjusttheoffsetvalueandtheslopevalueofthepHsensor.

    1. Connectaperistalticpumpinthewatersampleinletcircuitofthesystem,asshowninFig.3,chap.8.10.1.Ifthefluidicconnectionsarenotdonecorrectly,theproductandthesystemcanbedamagedbecausethepressureintheproductandinthesystemistoohigh.

    2. Makesurethedirectionoftheflowiscorrect.

    English

    Type MS01

  • 24

    Adjustmentandoperation

    3. Makesurethefluidicinstallationforthecalibrationistight.

    4. Letcleanwaterflowthroughthesystemtorinsetheproduct.

    5. PreparetwobuffersolutionswithpHvaluesthatarecompatiblewithallthesensorcubesofthesystem.

    6. Inthe"Maintenance"viewofthesensor,tapphcalibrationwizard2point .

    7. Step1/8:letthefirstbuffersolutionflowthroughtheproduct.

    8. Tap .

    9. Step2/8:tapinputvalueofbuffer1 andenterthepHvalueofthebuffersolution.

    10. Tap tovalidate.

    11. Tap .

    12. Step3/8:whenthepHmeasurementisstable,tap .

    13. Letcleanwaterflowthroughthesystemtorinsetheproduct.

    14. Step4/8::letthesecondbuffersolutionflowthroughtheproduct.

    15. Tap .

    16. Step5/8:tapinputvalueofbuffer2 andenterthepHvalueofthebuffersolution.

    17. Tap tovalidate.

    18. Tap .

    19. Step6/8:whenthepHmeasurementisstable,tap .

    20. Step7/8:

    • ifthecalibrationhassucceded,thecalculatedslopevalueandthecalculatedoffsetvaluearedisplayedandthe

    dateofthelastcalibrationisupdated(seechap.8.10.7).Tap togotothestep8/8.

    • ifanerrormessageisdisplayed,refertoTable11.

    • ifawarningmessageisdisplayed,refertoTable12.

    English

    Type MS01

  • 25

    Adjustmentandoperation

    21. Step8/8:validateorcancelthecalibration.

    Table 11: Error message at the end of the 2 point calibration of the pH sensor

    Displayedmessage ComputedcalibrationvalueoutoferrorlimitssymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thesensorcalibrationhasfailedbecausethecalculatedoffsetvalue

    and/orthecalculatedslopevalueareoutoftheerrorrange.

    → Comparethecalculatedoffsetvalueandthecalculatedslopevaluewiththecalibrationlimitsthatcanbereadinthe"Diagnosis"view.

    • Iftheoffsetvalueisoutoftheerrorlimits:

    • Iftheslopevalueisoutoftheerrorlimits:

    Whattodo? 1. Replacetheexternalreferenceelectrodebyanewone.

    2. Dothecalibrationagain.

    3. Ifthecalibrationfailsagain,replacetheproductbyanewone.

    → SendtheproductbacktoBürkerttohavetheISFETsensorreplaced.

    English

    Type MS01

  • 26

    Adjustmentandoperation

    Table 12: Warning message at the end of the 2 point calibration of the pH sensor

    Dis

    play

    edm

    essa

    geC

    ompu

    ted

    calib

    ratio

    nva

    lue

    outo

    fwar

    ning

    lim

    itss

    ymbo

    ldis

    play

    edo

    nth

    eu

    niqu

    eD

    evic

    e n

    ame

    oft

    hep

    rodu

    ctp

    ossi

    ble

    caus

    eTh

    eca

    lcul

    ated

    offs

    etv

    alue

    and

    /ort

    hec

    alcu

    late

    dsl

    ope

    valu

    ear

    eou

    toft

    hew

    arni

    ngra

    nge:

    •If

    the

    offs

    etv

    alue

    iso

    uto

    fthe

    war

    ning

    lim

    its:

    •If

    the

    slop

    eva

    lue

    iso

    uto

    fth

    ew

    arni

    ngli

    mits

    :•

    eith

    era

    wro

    ngb

    uffe

    rsol

    utio

    nha

    sbe

    enu

    sed

    fort

    hec

    alib

    ratio

    n.•

    orth

    eex

    tern

    alre

    fere

    nce

    elec

    -tr

    ode

    isa

    ging

    .W

    hat

    tod

    o?1.

    M

    ake

    sure

    the

    buffe

    rsol

    utio

    nus

    edis

    the

    corr

    ecto

    ne.

    2.

    Ifit

    isn

    otth

    eco

    rrec

    tone

    ,abo

    rt

    the

    calib

    ratio

    n.

    3.

    Use

    ac

    orre

    ctb

    uffe

    rsol

    utio

    nan

    ddo

    the

    com

    plet

    eca

    li-br

    atio

    npr

    oced

    ure

    agai

    n.

    You

    can

    choo

    seto

    eith

    erv

    al-

    idat

    eor

    can

    celt

    hec

    alib

    ratio

    n.

    •If

    you

    valid

    ate

    the

    calib

    ratio

    n,th

    ene

    wc

    alcu

    late

    dof

    fset

    val

    ueis

    us

    edto

    det

    erm

    ine

    the

    pHv

    alue

    ,an

    dth

    ela

    stc

    alib

    ratio

    nda

    teis

    up

    date

    d.

    •If

    you

    canc

    elth

    eca

    libra

    tion,

    th

    ecu

    rren

    toffs

    etv

    alue

    isu

    sed

    tod

    eter

    min

    eth

    epH

    val

    ue,a

    nd

    the

    last

    cal

    ibra

    tion

    date

    isn

    ot

    upda

    ted.

    Pla

    nto

    repl

    ace

    the

    exte

    rnal

    ref-

    eren

    cee

    lect

    rode

    .

    Sen

    dth

    epr

    oduc

    tbac

    kto

    B

    ürke

    rtto

    hav

    eth

    eIS

    FET

    sens

    orre

    plac

    ed.

    English

    Type MS01

  • 27

    Adjustmentandoperation

    8.10.5 Adjusting the offset value of the temperature

    Thetemperaturemeasuredbythesensorcanbecorrectedwithanoffsetvalue.

    1. Inthe"Maintenance"viewofthesensor,taptemperatureoffset andenterthevalueofthetem-peratureoffset,inthedisplayedtemperatureunits.

    2. Tochangetheunitsofthetemperature,gotothe"Parameter"viewofthesensor,taptemperatureDesiredunit andchoosethetemperatureunits:theoffsetvalueofthetemperatureisautomaticallyconverted.

    8.10.6 Simulating some data

    Checktheoperating(forexample,makesurethewarningand/orerrorlimitsarecorrectlyset)oftheproductand/orprocessbysimulatingsomedata.

    1. Inthe"Maintenance"viewofthesensor,tapsimulationstatus andchoose"ON"toactivatethesimulation.

    2. Validate.

    3. TosimulateapHvalue,tapsimulationph andenterthepHvaluetobesimulated.

    4. Tosimulateatemperaturevalue,tapsimulationtemperature andenterthetemperaturevaluetobesimulated.

    8.10.7 reading the date of the last calibration

    Toreadthedateofthelastcalibrationthathassucceeded.

    → Inthe"Maintenance"viewofthesensor,readthedateinthefieldcalibrationschedulelastcalibration.

    8.10.8 reading the date of the next due calibration

    Toreadthedateofthenextcalibrationthatmustdone.

    → Inthe"Maintenance"viewofthesensor,readthedateinthefieldcalibrationschedulenextcalibration.

    8.10.9 Setting the time interval between two calibrations

    Tosetthetimeinterval,indays,betweentwocalibrations:

    1. Inthe"Maintenance"viewofthesensor,tapcalibrationscheduleintervalindays .

    1. Enterthenumberofdaysbetweentwocalibrations.Werecommendtocalibratetheproducteverymonth.

    2. Validate.

    Whentheduecalibrationdateisreached,awarningmessageisdisplayedintheMessagesList.

    English

    Type MS01

  • 28

    Maintenanceandtroubleshooting

    9 MAinTenAnce AnD TroubleShooTing

    9.1 Safety instructions

    Warning

    riskofinjuryduetonon-conformingmaintenance.

    ▶ Maintenancemustonlybecarriedoutbyqualifiedandskilledstaffwiththeappropriatetools.

    note

    riskofdamagetotheproductduetothepowersupply

    • Shutdownandisolatetheelectricalpowersourcebeforecarryingoutworkonthesystem.

    9.2 cleaning of the product

    danger

    riskofinjuryduetothenatureofthedetergent.

    ▶ Respecttheregulationsonaccidentpreventionandsafetyrelatingtotheuseofaggressivefluids.

    Theproductcanbecleanedwithaclothdampenedwithwater(max.40°C)orwithanaciddetergent(withmax.5%ofhydrochloricacid).

    9.3 replacing the external reference electrode

    TheexternalreferenceelectrodemustbereplacediftheerrorcalibrationlimitsfortheoffsetvalueofthepHsensorareexceeded.

    Proceduretoreplacetheexternalreferenceelectrode:

    1. RemovethepHsensorcubefromthesystemasdescribedintheOperatingIsntructionsofthesystemtype8905.

    2. Liftthedropleverofthereferenceelectrodeandturnitbyaquarterturntotheleft.

    Droplever

    3. Pullthelevertoremovetheexternalreferenceelectrode.

    4. Disposeoftheusedexternalreferenceelectrode.Keeptotheexistingprovisionsonthesubjectofwastedisposalandenvironmentalprotection.

    English

    Type MS01

  • 29

    Maintenanceandtroubleshooting

    5. Ifnecessary,cleantheemptyseatingoftheelec-trodereferencewithadryclothorcottonswabs.

    6. Preparethenewreferenceelectrode.

    7. CarefullyremovetheprotectivecapbecauseitcontainssomeKCl.

    8. Makesurethesealisinitsgrooveontheexternalreferenceelectrode.

    9. Ifnecessary,removethesoliddepositsfromthecapillary.

    10. Moistenthesealwithwater.Ifnot,thesealcanbedamaged.

    Seal

    Capillary

    11. Liftthedropleverofthereferenceelectrodeandturnitbyaquarterturntotheleft.

    12. Insertthenewexternalreferenceelectrodefully.

    13. Turnthedropleverbyaquarterturntotherightandloweritfullyinitsseating.

    14. MountthepHsensorcubebackinthesystemasdescribedintheOperatingIsntructionsofthesystemtype8905.

    9.4 Troubleshouting if no message is displayed

    colouroftheproductstatusleD OFF

    possiblecause Theproduct/thesystemisnotenergized.

    Whattodo? 1. Checkthewiring.

    2. Makesurethevoltagesupplyis24VDC.

    3. Checkthatthepowersupplysourceisworkingproperly.

    English

    Type MS01

  • 30

    Maintenanceandtroubleshooting

    9.5 Troubleshooting if the status leD of the product is red or orange

    Ifanerrororawarningmessagehasbeengeneratedbythesystem:

    • thestatuslightofthesystemisredororange,

    • thestatusLEDoftheproductisredororange,

    • thesymbol or appearsontheUniqueDeviceNameoftheproduct,inthelistofdevices,

    • thesymbol appearsinthetopleftcornerofthedisplay.

    → Tap toaccessthe"MessagesList".

    9.5.1 Message "Too high leakage current"

    Displayedmessage ToohighleakagecurrentsymboldisplayedontheuniqueDevicenameoftheproductpossiblecause TheleakagecurrentoftheISFETsensorisoutofrange.

    Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):

    • theerror"activationflag"forthehighthresholdoftheleakagecurrentisselected,

    • andthehighthresholdsetfortheleakagecurrentoftheISFETsensorhasbeenreached.

    Whattodo? → SendtheproductbacktoBürkerttohavetheISFETsensorreplaced.

    Displayedmessage ToohighleakagecurrentsymboldisplayedontheuniqueDevicenameoftheproductpossiblecause TheleakagecurrentoftheISFETsensorisoutofrange.

    ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):

    • thewarning"activationflag"forthehighthresholdoftheleakagecurrentisselected,

    • andthehighthresholdsetfortheleakagecurrentoftheISFETsensorhasbeenreached.

    Whattodo? → Acknoledgethewarningmessage:theproductcontinuestooperate.

    → or,sendtheproductbacktoBürkerttohavetheISFETsensorreplaced.

    → orreplacetheproductbyanewone.

    English

    Type MS01

  • 31

    Maintenanceandtroubleshooting

    9.5.2 Message "Too low leakage current"

    Displayedmessage ToolowleakagecurrentsymboldisplayedontheuniqueDevicenameoftheproductpossiblecause TheleakagecurrentoftheISFETsensorisoutofrange.

    Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):

    • theerror"activationflag"forthelowthresholdoftheleakagecurrentisselected,

    • andthelowthresholdsetfortheleakagecurrentoftheISFETsensorhasbeenreached.

    Whattodo? → Deselecttheerror"activationflag"forthelowthresholdoftheleakagecurrent.

    Displayedmessage ToolowleakagecurrentsymboldisplayedontheuniqueDevicenameoftheproductpossiblecause TheleakagecurrentoftheISFETsensorisoutofrange.

    ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):

    • thewarning"activationflag"forthelowthresholdoftheleakagecurrentisselected,

    • andthelowthresholdsetfortheleakagecurrentoftheISFETsensorhasbeenreached.

    Whattodo? → Deselectthewarning"activationflag"forthelowthresholdoftheleakagecurrent.

    9.5.3 Message "Too high ph value"

    Displayedmessage ToohighpHvaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause ThepHvalueofthewatersampleisoutofrange.

    Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):

    • theerror"activationflag"forthehighthresholdofthepHvalueisselected,

    • andthehighthresholdsetforthepHvaluehasbeenreached.Whattodo? → Checktheprocess.

    English

    Type MS01

  • 32

    Maintenanceandtroubleshooting

    Displayedmessage ToohighpHvaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause ThepHvalueofthewatersampleisoutofrange.

    ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):

    • thewarning"activationflag"forthehighthresholdofthepHvalueisselected,

    • andthehighthresholdsetforthepHvaluehasbeenreached.Whattodo? → Checktheprocess.

    9.5.4 Message "Too low ph value"

    Displayedmessage ToolowpHvaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause ThepHvalueofthewatersampleisoutofrange.

    Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):

    • theerror"activationflag"forthelowthresholdofthepHvalueisselected,

    • andthelowthresholdsetforthepHvaluehasbeenreached.Whattodo? → Checktheprocess.

    Displayedmessage ToolowpHvaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause ThepHvalueofthewatersampleisoutofrange.

    ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):

    • thewarning"activationflag"forthelowthresholdofthepHvalueisselected,

    • andthelowthresholdsetforthepHvaluehasbeenreached.Whattodo? → Checktheprocess.

    English

    Type MS01

  • 33

    Maintenanceandtroubleshooting

    9.5.5 Message "Too high temperature value"

    Displayedmessage ToohightemperaturevaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thetemperaturevalueofthewatersampleisoutofrange.

    Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):

    • theerror"activationflag"forthehighthresholdofthewatersampletemperaturevalueisselected,

    • andthehighthresholdsetforthewatersampletemperaturevaluehasbeenreached.

    Whattodo? → Checktheprocess.

    Displayedmessage ToohightemperaturevaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thetemperaturevalueofthewatersampleisoutofrange.

    ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):

    • thewarning"activationflag"forthehighthresholdofthewatersampletemperaturevalueisselected,

    • andthehighthresholdsetforthewatersampletemperaturevaluehasbeenreached.

    Whattodo? → Checktheprocess.

    9.5.6 Message "Too low temperature value"

    Displayedmessage ToolowtemperaturevaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thetemperaturevalueofthewatersampleisoutofrange.

    Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):

    • theerror"activationflag"forthelowthresholdofthewatersampletemperaturevalueisselected,

    • andthelowthresholdsetforthewatersampletemperaturevaluehasbeenreached.

    Whattodo? → Checktheprocess.

    English

    Type MS01

  • 34

    Maintenanceandtroubleshooting

    Displayedmessage ToolowtemperaturevaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thetemperaturevalueofthewatersampleisoutofrange.

    ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):

    • thewarning"activationflag"forthelowthresholdofthewatersampletemperaturevalueisselected,

    • andthelowthresholdsetforthewatersampletemperaturevaluehasbeenreached.

    Whattodo? → Checktheprocess.

    9.5.7 Message "Too high reference impedance"

    Displayedmessage ToohighreferenceimpedancesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Theimpedanceofthereferenceelectrodeisoutofrange.

    Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):

    • theerror"activationflag"forthehighthresholdoftheimpedanceofthereferenceelectrodeisselected,

    • andthehighthresholdsetfortheimpedanceofthereferenceelec-trodehasbeenreached.

    Whattodo? 1. Gointothe"Diagnostic"viewofthesensortoreadtheimpedancevalueofthereferenceelectrode.

    2. Makesurethehighthresholdisconsistentwiththevalueofthecon-ductivityofthewatersample.

    3. Ifnecessary,replacetheexternalreferenceelectrode.

    4. Iftheexternalreferenceelectrodehasbeenreplaced,calibratethepHsensor.

    English

    Type MS01

  • 35

    Maintenanceandtroubleshooting

    Displayedmessage ToohighreferenceimpedancesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Theimpedanceofthereferenceelectrodeisoutofrange.

    ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):

    • thewarning"activationflag"forthehighthresholdoftheimpedanceofthereferenceelectrodeisselected,

    • andthehighthresholdsetfortheimpedanceofthereferenceelec-trodehasbeenreached.

    Whattodo? 1. Gointothe"Diagnostic"viewofthesensortoreadtheimpedancevalueofthereferenceelectrode.

    2. Makesurethehighthresholdisconsistentwiththevalueofthecon-ductivityofthewatersample.

    3. Plantoreplacetheexternalreferenceelectrode.

    9.5.8 Message "Too low reference impedance"

    Displayedmessage ToolowreferenceimpedancesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Theimpedanceofthereferenceelectrodeisoutofrange.

    Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):

    • theerror"activationflag"forthelowthresholdoftheimpedanceofthereferenceelectrodeisselected,

    • andthelowthresholdsetfortheimpedanceofthereferenceelec-trodehasbeenreached.

    Whattodo? → Deselecttheerror"activationflag"forthelowthresholdoftheimpedanceofthereferenceelectrode.

    Displayedmessage ToolowreferenceimpedancesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Theimpedanceofthereferenceelectrodeisoutofrange.

    ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):

    • thewarning"activationflag"forthelowthresholdoftheimpedanceofthereferenceelectrodeisselected,

    • andthelowthresholdsetfortheimpedanceofthereferenceelec-trodehasbeenreached.

    Whattodo? → Deselectthewarning"activationflag"forthelowthresholdoftheimpedanceofthereferenceelectrode.

    English

    Type MS01

  • 36

    Sparepartsandaccessories

    10 SpAre pArTS AnD AcceSSorieS

    caution

    riskofinjuryand/ordamagecausedbytheuseofunsuitableparts.

    Incorrectaccessoriesandunsuitablereplacementpartsmaycauseinjuriesanddamagetheproductandthesurroundingarea.

    ▶ UseonlyoriginalaccessoriesandoriginalreplacementpartsfromBürkert.

    Accessory ordercodeBuffersolution,pH=5 566031Buffersolution,pH=7 418541Buffersolution,pH=8 onrequestExternalreferenceelectrode 566084

    11 pAckAging, TrAnSporT

    note

    Damageduetotransport

    Transportmaydamageaninsufficientlyprotectedproduct.

    • Transporttheproductinshock-resistantpackagingandawayfromhumidityanddirt.

    • Donotexposetheproducttotemperaturesthatmayexceedtheadmissiblestoragetemperaturerange.

    • Protecttheelectricalinterfacesusingprotectiveplugs.

    12 STorAge

    note

    poorstoragecandamagetheproduct.

    ▶ Dependingonthedurationofthestoragetime,obeythestorageinstructionsinthechapter12.1to12.3.

    ▶ Afteranystorageperiod,obeytheinstructionsinchap.12.4.

    12.1 To store the product for a maximum of 4 days

    1. Rinsetheproductwithtapwater.

    2. Purgetheproductwithairatamax.pressureof2bar.

    3. Storetheproductatroomtemperature(about23°C±10°C).

    4. Storetheproductinadryplaceawayfromdust.

    English

    Type MS01

  • 37

    Sparepartsandaccessories

    12.2 To store the product for more than 4 days and less than 1 month

    1. Rinsetheproductwithtapwater.

    2. DONOTPURGETHEPRODUCTbecauseitwilldryout.

    3. Storetheproductatroomtemperature(about23°C±10°C).

    4. Storetheproductinadryplaceawayfromdust.

    12.3 To store the product for a more than 1 month

    1. Removetheexternalreferenceelectrode.

    2. Rinsetheproductwithtapwater.

    3. Purgetheproductwithairatamax.pressureof2bar.

    4. Storetheproductatroomtemperature(about23°C±10°C).

    5. Storetheproductinadryplaceawayfromdust.

    12.4 After storage

    Beforecommisioningaproductthathasbeenstored:

    1. Mountanewexternalreferenceelectrode.Seechap.9.3.

    2. Plugtheproductinthesystem.

    3. WhilethepowersupplyofthesystemisOFF,letthewatersampleflowthroughtheproductforatleast2hours.

    4. Calibratetheproduct.Seechap.8.10.

    13 DiSpoSAl of The proDucT→ Disposeoftheproductanditspackaginginanenvironmentally-friendlyway.

    noteDamagetotheenvironmentcausedbyproductscontaminatedbyfluids.

    • Keeptotheexistingprovisionsonthesubjectofwastedisposalandenvironmentalprotection.