Operating Instructions
TypeMS01
pH sensor cube
-
We reserve the right to make technical changes without notice.Technische Änderungen vorbehalten.Sous réserve de modifications techniques.
© Bürkert SAS, 2014
Operating Instructions 1409/0_EU-ML 00566496 Original EN
-
1
Contents
1 AbouttheoperAtinginstructions...........................................................................................................................4
1.1 symbolsused..........................................................................................................................................................................4
1.2 Definitionoftheword"product"....................................................................................................................................4
1.3 Definitionoftheword"system"......................................................................................................................................4
1.4 Definitionoftheword"büs".............................................................................................................................................5
2 intenDeDuse....................................................................................................................................................................................5
3 bAsicsAfetyinformAtion....................................................................................................................................................6
4 generAlinformAtion................................................................................................................................................................7
4.1 contact........................................................................................................................................................................................7
4.2 Warrantyconditions..............................................................................................................................................................7
4.3 informationsontheinternet............................................................................................................................................7
5 Description.......................................................................................................................................................................................8
6 technicAlDAtA................................................................................................................................................................................9
6.1 conditionsofuse...................................................................................................................................................................9
6.2 conformitytostandardsanddirectives....................................................................................................................9
6.3 materialstheproductismadeof..................................................................................................................................9
6.4 fluiddata....................................................................................................................................................................................9
6.5 measurementdata.............................................................................................................................................................10
6.6 electricaldata.......................................................................................................................................................................10
6.7 communication....................................................................................................................................................................10
7 instAllAtion...................................................................................................................................................................................11
7.1 safetyinstructions.............................................................................................................................................................11
7.2 mountingtheproductonthebackplane................................................................................................................11
8 ADjustmentAnDoperAtion..............................................................................................................................................12
8.1 safetyinstructions.............................................................................................................................................................12
8.2 howtoadjusttheproduct.............................................................................................................................................12
8.3 Adjustmentsthatmustbedone.................................................................................................................................13
8.4 generalinformationonthedisplaysoftwaretypeme21..............................................................................13
Type MS01
English
-
2
Contents
8.5 "Device"viewoftheproduct.........................................................................................................................................14
8.6 "function"viewoftheproduct.....................................................................................................................................15
8.7 Detailedviewsofthebüsfunction...........................................................................................................................16
8.8 "parameter"viewofthesensor...................................................................................................................................17
8.8.1 Configuringthetriggeringofthewarnings...................................................................................17
8.8.2 Configuringthetriggeringoftheerrors........................................................................................18
8.9 "Diagnosis"viewofthesensor....................................................................................................................................18
8.10 "maintenance"viewofthesensor..............................................................................................................................19
8.10.1 CalibratingtheoffsetvalueoftheISFETsensor.........................................................................19
8.10.2 Calibratingtheoffsetand/orslopevaluesofthepHsensor....................................................21
8.10.3 Doinga1pointcalibrationprocedureofthepHsensor...........................................................22
8.10.4 Doinga2pointcalibrationprocedureofthepHsensor...........................................................23
8.10.5 Adjustingtheoffsetvalueofthetemperature..............................................................................27
8.10.6 Simulatingsomedata........................................................................................................................27
8.10.7 Readingthedateofthelastcalibration........................................................................................27
8.10.8 Readingthedateofthenextduecalibration...............................................................................27
8.10.9 Settingthetimeintervalbetweentwocalibrations.....................................................................27
9 mAintenAnceAnDtroubleshooting.......................................................................................................................28
9.1 safetyinstructions.............................................................................................................................................................28
9.2 cleaningoftheproduct...................................................................................................................................................28
9.3 replacingtheexternalreferenceelectrode.........................................................................................................28
9.4 troubleshoutingifnomessageisdisplayed.......................................................................................................29
9.5 troubleshootingifthestatusleDoftheproductisredororange........................................................30
9.5.1 Message"Toohighleakagecurrent".............................................................................................30
9.5.2 Message"Toolowleakagecurrent"...............................................................................................31
9.5.3 Message"ToohighpHvalue".........................................................................................................31
9.5.4 Message"ToolowpHvalue"...........................................................................................................32
9.5.5 Message"Toohightemperaturevalue".........................................................................................33
9.5.6 Message"Toolowtemperaturevalue"..........................................................................................33
9.5.7 Message"Toohighreferenceimpedance"...................................................................................34
9.5.8 Message"Toolowreferenceimpedance"....................................................................................35
10 spArepArtsAnDAccessories......................................................................................................................................36
English
-
3
11 pAckAging,trAnsport..........................................................................................................................................................36
12 storAge..............................................................................................................................................................................................36
12.1 tostoretheproductforamaximumof4days..................................................................................................36
12.2 tostoretheproductformorethan4daysandlessthan1month.........................................................37
12.3 tostoretheproductforamorethan1month...................................................................................................37
12.4 Afterstorage..........................................................................................................................................................................37
13 DisposAloftheproDuct..................................................................................................................................................37
English
Type MS01
-
4
AbouttheOperatingInstructions
1 AbouT The operATing inSTrucTionSTheOperatingInstructionsdescribetheentirelifecycleoftheproduct.PleasekeeptheOperatingInstructionsinasafeplace,accessibletoallusersandanynewowners.
theoperatinginstructionscontainimportantsafetyinformation.
Failuretocomplywiththeseinstructionscanleadtohazardoussituations.
▶ TheOperatingInstructionsmustbereadandunderstood.
1.1 Symbols used
danger
Warnsagainstanimminentdanger.
▶ Failuretoobservethiswarningcanresultindeathorinseriousinjury.
Warning
Warnsagainstapotentiallydangeroussituation.
▶ Failuretoobservethiswarningcanresultinseriousinjuryorevendeath.
attention
Warnsagainstapossiblerisk.
▶ Failuretoobservethiswarningcanresultinsubstantialorminorinjuries.
note
Warnsagainstmaterialdamage.
▶ Failuretoobservethiswarningmayresultindamagetotheproductorsystem.
Indicatesadditionalinformation,adviceorimportantrecommendations.
ReferstoinformationcontainedintheOperatingInstructionsorinotherdocuments.
▶ Indicatesaninstructiontobecarriedouttoavoidadanger,awarningorapossiblerisk.
→ Indicatesaproceduretobecarriedout.
» Indicatestheresultofaspecificinstruction.
1.2 Definition of the word "product"
Theword"product"usedwithintheseOperatingInstructionsalwaysreferstothepHsensorcubetypeMS01.
1.3 Definition of the word "system"
Theword"system"usedwithintheseOperatingInstructionsalwaysreferstotheOnlineAnalysisSystemtype8905.
English
Type MS01
-
5
AbouttheOperatingInstructions
1.4 Definition of the word "büS"
Theword"büS"usedwithintheseOperatingInstructionsalwaysreferstothefieldbusdeveloppedbyBürkert.
2 inTenDeD uSe
useofthisproductthatdoesnotcomplywiththeinstructionscouldpresentriskstopeople,nearbyinstallationsandtheenvironment.
▶ TheproductisintendedsolelyforthemeasurementofthepHinwaterwithina8905system.▶ Thisproductmustbeprotectedagainstelectromagneticinterference,ultravioletraysand,wheninstalledoutdoors,theeffectsofclimaticconditions.
▶ ThisproductmustbeusedincompliancewiththecharacteristicsandcommissioninganduseconditionsspecifiedinthecontractualdocumentsandintheOperatingInstructions.
▶ Requirementsforthesafeandproperoperationoftheproductarepropertransport,storageandinstallation,aswellascarefuloperationandmaintenance.
▶ Onlyusetheproductasintended.
▶ Observeanyexistingrestraintswhentheproductisexported.
English
Type MS01
-
6
Basicsafetyinformation
3 bASic SAfeTy inforMATionThissafetyinformationdoesnottakeintoaccount:
• anycontingenciesoroccurrencesthatmayariseduringassembly,useandmaintenanceoftheproduct.
• thelocalsafetyregulationsthattheoperatormustensurethestaffinchargeofinstallationandmaintenanceobserve.
Variousdangeroussituations.
Toavoidinjurytakecare:
▶ topreventanyunintentionalpowersupplyswitch-on.
▶ tocarryouttheinstallationandmaintenanceworkbyqualifiedandskilledstaffwiththeappropriatetools.
▶ tousetheproductonlyifinperfectworkingorderandincompliancewiththeinstructionsprovidedintheseOperatingInstructions.
▶ toobservethegeneraltechnicalrulesduringtheplanninganduseoftheproduct.
▶ nottousethisproductinexplosiveatmospheres.
▶ nottousethisproductinanenvironmentincompatiblewiththematerialsfromwhichitismade.
▶ nottomakeanyexternalorinternalmodificationstotheproduct.
note
elements/componentssensitivetoelectrostaticdischarges
• Thisproductcontainselectroniccomponentssensitivetoelectrostaticdischarges.Theymaybedamagediftheyaretouchedbyanelectrostaticallychargedpersonorobject.Intheworstcasescenario,thesecompo-nentsareinstantlydestroyedorgooutoforderassoonastheyareactivated.
• Tominimiseorevenavoidalldamageduetoanelectrostaticdischarge,takealltheprecautionsdescribedintheEN61340-5-1norm.
• Alsoensurethatyoudonottouchanyoftheliveelectricalcomponents.
English
Type MS01
-
7
Generalinformation
4 generAl inforMATion
4.1 contact
Tocontactthemanufactureroftheproductusefollowingaddress:
BürkertSAS
RueduGiessen
BP21
F-67220TRIEMBACH-AU-VAL
TheaddressesofourinternationalbranchescanbefoundontheInternetat:www.burkert.com
4.2 Warranty conditions
TheconditiongoverningthelegalwarrantyistheconforminguseoftheproductinobservanceoftheoperatingconditionsspecifiedintheseOperatingInstructions.
4.3 informations on the internet
YoucanfindtheOperatingInstructionsandtechnicaldatasheetsregardingthetypeMS01at:www.burkert.com
English
Type MS01
http://burkert.comhttp://burkert.com
-
8
Description
5 DeScripTionThepHsensorcubeisusedinthesystemtype8905.
Theelectricalandfluidconnectionsaremadeviathebackplaneofthesystemtype8905.
Leverto:
- lock/unlocktheproduct.
- carryoutmaintenanceoperations.
Unlockedposition
Lockedposition
Maintenanceposition
Push-buttonwithmea-suredprocessvariable
Fluidconnections
Electricalinterface
Pin
Adaptationpins
Housingoftheexternalreferenceelectrode
SensorcubestatusLED
Slotforthemicro-SIMcard
Fig. 1: Description of the product
English
Type MS01
-
9
Technicaldata
6 TechnicAl DATA
6.1 conditions of use
Ambienttemperature 0to+40°C
Airhumidity <90%,withoutcondensation
Protectionratingacc.toEN60529
• IP65,whenpluggedinthebackplane
• IP20,asstandaloneproduct
Max.heightabovesealevel 2000m
6.2 conformity to standards and directives
TheproductconformstotheCEdirectivesofthesystemtype8905,onlywhentheproducttypeMS01ispluggedinthesystemtype8905.
6.3 Materials the product is made of
part material
Housing PPE+PS
Seal EPDM
Lever PC
6.4 fluid data
• Typeoffluid
• pHvalue
• Water,withoutparticles:drinkingwater,industrialwater
• pH4to9Minimalflowrate 3l/h;recommended6l/h
Watersamplepressure PN6
Watersampletemperature 0to+40°C,notfreezing
English
Type MS01
-
10
Technicaldata
6.5 Measurement data
pHmeasurement
• Measuringrange
• Sensorresolution
• Measurementdeviation("measurementbias",asdefinedinthestandardJCGM200:2012)
• Linearity
• Repeatability
• Responsetime(t90)
• Measurementsensor
• Electrolyteoftheexternalreferenceelectrode
• pH4to9
• pH0.02
• pH±0.1
• pH±0.05
• pH±0.05
• <10s
• ISFET
• 3molKClTemperaturemeasurement
• Measuringrange
• Measurementsensor
• 0to50°C
• Pt1000ClassB,nocontactwiththewatersampleMaintenanceintervaloftheexternalreferenceelectrode 12months,nominal,dependingonthewaterquality
6.6 electrical data
Operatingvoltage 24VDCthroughthebackplaneofthesystemtype8905
Powerconsumption 0.8VA
6.7 communication
Internalcommunication throughbüS
ExternalcommunicationbystatusLED accordingtoNAMURNE107
English
Type MS01
-
11
Installation
7 inSTAllATion
7.1 Safety instructions
note
riskofdamagetotheproductduetonon-conforminginstallation.
• Theelectricalandfluidicinstallationscanonlybecarriedoutbyqualifiedandskilledstaffwiththeappropriatetools.
• Respecttheinstallationinstructionsforthesystem.
note
riskofdamagetotheproductduetothepowersupply
• Shutdownandisolatetheelectricalpowersourcebeforecarryingoutworkonthesystem.
note
riskofdamagetotheproductduetotheenvironment
• Protecttheproductagainstelectromagneticinterference,ultravioletraysand,wheninstalledoutdoors,theeffectsoftheclimaticconditions.
7.2 Mounting the product on the backplane
TheproductispluggedinthebackplaneofthesystemType8905.
4
3
3
1. Drythesurfaceofthebackplanethatwillbeincontactwiththeproduct.
2. Drythesurfaceoftheproduct.
3. Whilethepush-buttonispushed,turnthebayonetlevertotheright,
ontheunlockedposition,.Donotpushthelevertothemaintenanceposition.
4. Insertthetwoadaptationpinsintheirholesandthenplugtheproductinthebackplane.
5. Whilethepush-buttonispushed,turnthebayonetlevertotheleft,
onthelockedposition, .
Fig. 2: Mounting a product on the backplane of the system
English
Type MS01
-
12
Adjustmentandoperation
8 ADjuSTMenT AnD operATion
8.1 Safety instructions
note
riskofdamagetotheproductduetonon-conformingadjustment.
• TheoperatorsinchargeofadjustmentmusthavereadandunderstoodthecontentsoftheseOperatingInstructions.
• TheoperatorsinchargeofadjustmentmusthavereadandunderstoodthecontentsoftheOperatingInstruc-tionsofthedisplaysoftwaretypeME21and/orthecontentsoftheOperatingInstructionsoftheBürkertCommunicatorsoftwaretype8920and/ortheOperatingInstructionsofthecontrollermoduletypeME25.
• Inparticular,observethesafetyrecommendationsandintendeduse.
• Theproduct/installationmustonlybeadjustedbysuitablytrainedstaff.
note
riskofdamagetotheproductduetonon-conformingcommissioning.
• Beforecommissioning,makesurethatthestaffinchargehavereadandfullyunderstoodthecontentsoftheseOperatingInstructions.
• Inparticular,observethesafetyrecommendationsandintendeduse.
• Theproduct/theinstallationmustonlybecommissionedbysuitablytrainedstaff.
note
riskofdamagetotheproductduetonon-conformingoperation.
• TheoperatorsinchargeofoperationmusthavereadandunderstoodthecontentsoftheseOperatingInstructions.
• Inparticular,observethesafetyrecommendationsandintendeduse.
• Theproduct/installationmustonlybeoperatedbysuitablytrainedstaff.
8.2 how to adjust the product
Theadjustmentoftheproductcanbemade:
• eitherwiththedisplayofthesystemtype8905.ThedisplayofthesystemismanagedbythesoftwaretypeME21.Seechap.8.4.
• orwithaPCandtheBürkertCommunicatorsoftwaretype8920.Togetgeneralinformationaboutthesoftwaretype8920,refertotheOperatingInstructionsofthetype8920.
English
Type MS01
-
13
Adjustmentandoperation
8.3 Adjustments that must be done
Dotheadjustments:
• aftertheproducthasbeeninstalledinthesystem.
• afterthefluidicandelectricalinstallationshavebeenmadeonthesystem.
• afterthetightnessofthesystemhasbeenchecked.
• afterthesystemhasoperatedforapolarizationtimeof12hours.
• Beforecommissioningtheproduct/systemforthefirsttimeandforthecorrectoperationoftheproductandofthesystem,dothefollowingadjustmentsfortheproduct:
- calibratetheoffsetvalueofthepHsensor(refertochap.8.10)or,
- calibratetheoffsetvalueandtheslopevalueofthepHsensor(refertochap.8.10).
• Beforecommissioningtheproduct/systemafteramaintenanceoperation(i.e.replacementoftheexternalref-erenceelectrodeorreplacementoftheISFETsensor),letthesystemoperateforapolarizationtimeof12hoursanddothefollowingadjustmentsfortheproduct:
- ifthepHsensorhasbeenreplaced,calibratetheoffsetvalueandtheslopevalueofthepHsensor(refertochap.8.10).
- iftheISFETsensorhasbeenreplaced,calibratetheoffsetvalueoftheISFETsensorandcalibratetheoffsetvalueandtheslopevalueofthepHsensor(refertochap.8.10).
• Calibratetheproducteverymonth.
8.4 general information on the display software type Me21
TheseOperatingInstructionsexplaintheadjustmentsthatarespecifictotheproducttypeMS01.
→ TogetgeneralinformationaboutthedisplaysoftwaretypeME21,refertotheOperatingInstructionsofthetypeME21thatisontheCDdeliveredwiththesystemandthatisalsoavailableatwww.burkert.com.
TheOperatingInstructionsofthedisplaysoftwaretypeME21:
• givegeneralinformationonthesoftware,suchas:descriptionoftheuserinterface,structureofthemenus,descriptionofthepossibleviews("Device"viewforexample),descriptionofthenavigationbuttons...
• explainhowtomakethegeneraladjustmentssuchas:thedisplaylanguage,thelocatingoftheproduct...
• explainhowtoconfigureandcustomizethe"Desktop"viewswithvaluesorgraphs.
• givegeneralinformationontheerrormessagesandtheoperatingofthesystemstatuslight.
English
Type MS01
http://burkert.com
-
14
Adjustmentandoperation
8.5 "Device" view of the product
The"Device"viewshowssomeofthemeasurementdatarelatedtotheproduct.
→ Ifthedisplayshowsa"Desktop"view,tap toaccessthe"Device"view.
→ Todisplaythe"Device"viewoftheproduct,selecttheproductinthelistofdevicesontheleftofthedisplay.
Table 1: "Device" view of the product
Controller
5648300001
Device 5648300001
Location TreatmentunitNr1 Status N/A
pH pH7.00 pHvoltage 0.00V
Temperature 24.10°CBoardtemperature 26.20°C
Deviceview
HMIU
12.08.201416:55
pH
ListofdevicesthatareconnectedtothebüS
Thefollowingdatacanbereadfromthe"Device"viewoftheproduct:
• thedevicesthatareconnectedonthebüSwiththeir"Uniquedevicename".Bydefault,the"Uniquedevicename"isbuiltupwithtwonumbers:theordercodeoftheproduct(forexample564830,fortheproduct)andtheseriesnumber(0001,forexample).
• wheretheproductisgeographicallylocated.
• themeasuredvalueofthepHofthewatersample.
• themeasuredvalueofthepotentialdifference.
• themeasuredvalueofthetemperatureofthewatersample.
• themeasuredvalueofthetemperatureoftheinternalmeasurementboard.
→ Todisplaythe"Function"viewoftheproduct,tap .
English
Type MS01
-
15
Adjustmentandoperation
8.6 "function" view of the product
The"Function"viewshowsthefunctionsavailableforaproductand,foreachfunction,themaindatarelatedtoeachfunction.
Todisplaythe"Function"viewoftheproduct:
1. selecttheproductinthelistofdevices,ontheleftofthedisplay,
2. tap .
Table 2: "Function" view of the product
büS
function büS
Uniquedevicename 5648300001 Location TreatmentUnitNr1
Description Springwater
IsfetSensor5648300001 12.08.201416:55
pH Sensor
"function"viewofthebüs
→ Todisplaythe"Function"viewofthebüStheproductisconnectedto,select"büS"inthelistofdevicesontheleftofthedisplay.
→ Tap toaccessthe"Parameter"view.Seechap.8.7.
Sensor
function Sensor
Measurement pH7.00 pHvoltage 0.00V
Slope -55.00mV/pHOffset pH7.00
Temperature 24.10°COffset 0.00°C
IsfetSensor5648300001 12.08.201416:55
büS
pH
"function"viewofthesensor
→ Todisplaythe"Function"viewofthesensor,select"Sensor"inthelistofdevicesontheleftofthedisplay.
→ Tap toaccessthe"Parameter"view.Seechap.8.8.
English
Type MS01
-
16
Adjustmentandoperation
8.7 Detailed views of the büS function
Table 3: "Parameter" view of the "büS" function
parameter büS
BasicSettings UniqueDeviceName 5648300001
TreatmentUnitNr1
0
0
Location
Description
Physicalgroup
Logicalgroup
ISFETsensor5648300001 04.04.201412:21
Groupingsettings
"parameter"detailedviewofthebüs
Tosettheuser-specificdatafortheidentificationofthedeviceonthebüS:
• UniqueDeviceNameoftheproduct.
→ MakesureyouchooseauniquenamefortheproductbecausethebüSmustrecognizetheproduct.
→ Makesureyouchooseauniquenamethatisselfexplanatorytoidentifytheproductbecause,iftheuniquenameischanged,allthesettingsmadeonthebüSmustbechanged.
• Description
• Location
• Groupingsetting
Table 4: "Diagnosis" view of the "büS" function
Diagnosis büS
büSinformation Address 126
500kbit/s
ISFETsensor
564830
683328
A.00.02.42
A.01.00.00
Baudrate
Devicename
Identitynumber
Softwareidentitynumber
Softwareversion
Hardwareversion
ISFETsensor5648300001 04.04.201412:21
Deviceinformation
"Diagnosis"detailedviewofthebüs
ToreadsomebüSdataanddevicedata:
• Addressandbaudrate
• Devicenameanddeviceordercode(Identitynumber)
• Softwareandhardwareinformation
• InformationrelatingtotheTFT
• Devicedriverinformation
Nochangesbyuserarepossible.
English
Type MS01
-
17
Adjustmentandoperation
Table 5: "Maintenance" view of the "büS" function
Maintenance büS
Resettypes Device Reset
ISFETsensor5648300001 12:21:16"maintenance"detailedviewofthebüs
Restarttheproductforareset.
8.8 "parameter" view of the sensor
→ Inthe"Function"view,tap toaccessthe"Parameter"view.
parameter Sensor
Temperature Desiredunit °C
ON
2s
Measurement
Responsetime
Warnings
Errors
Messagestriggers
IsfetSensor5648300001 12:21:16
Referenceimpedance
Filter
The"Parameter"viewofthesensormakesitpossible:
• tochangetheunitsofthedisplayedtemperature.
• toselectwhethertheimpedanceofthereferenceelectrodeismeasuredornot.
• toenterthevalueoftheT10-90responsetimeforthemeasuredsignal.
• toconfigurethetriggeringofthewarnings,seechap.8.8.1.
• toconfigurethetriggeringoftheerrors,seechap.8.8.2.
Table 6: "Parameter" view of the sensor
8.8.1 configuring the triggering of the warnings
Thevaluesofsomeprocessvariablesandthevaluesofsomesensorcubevariablescanbemonitoredtodetectaproblemintheprocessoronthesensor.
ForthepHsensorcube,thefollowingvaluescanbemonitored:
• thepHofthewatersample,
• thetemperatureofthewatersample,
• theimpedanceofthereferenceelectrode,
• theleakagecurrentoftheISFETsensor.
English
Type MS01
-
18
Adjustmentandoperation
Procedure:
1. TapWarnings inthe"Parameter"viewofthesensor.
2. TapActivationflags inthe"SensorParameter"view.
3. Selecttheprocessvariablesandthesensorcubevaluestobemonitoredandvalidate.
4. Settheminimumand/ormaximumthresholdvalues.
Whentheparameteredthresholdvaluesarereachedandtheactivationflagoftherelatedprocessvariableor
sensorcubevalueisselected,awarningmessageisgeneratedinthe"MessagesList":tap toaccessthe"MessagesList".Seechap.9.4forthetroubleshootingdetails.
8.8.2 configuring the triggering of the errors
Thevaluesofsomeprocessvariablesandthevaluesofsomesensorcubevariablescanbemonitoredtodetectaproblemintheprocessorontheproduct.
ForthepHsensorcube,thefollowingvaluescanbemonitored:
• thepHofthewatersample,
• thetemperatureofthewatersample,
• theimpedanceofthereferenceelectrode,
• theleakagecurrentoftheISFETsensor.
Procedure:
1. Taperrors inthe"Parameter"viewofthesensor.
2. TapActivationflags inthe"Parameter"viewofthesensor.
3. Selecttheprocessvariablesandthesensorcubevaluestobemonitoredandvalidate.
4. Settheminimumand/ormaximumthresholdvalues.
Whentheparameteredthresholdvaluesarereachedandtheactivationflagoftherelatedprocessvariableor
sensorcubevalueisselected,anerrormessageisgeneratedinthe"MessagesList":tap toaccessthe"MessagesList".Seechap.9.4forthetroubleshootingdetails.
8.9 "Diagnosis" view of the sensor
The"Diagnosis"viewmakesitpossibletoreadthefollowingvalues:
• themeasuredvalueofthepHofthewatersample.
• themeasuredvalueofthepotentialdifference.
• thevalueoftheoffsetoftheISFETsensor,calculatedthroughcalibrationorenteredinthe"Maintenance"view.
• thevalueoftheslopeofthepHsensor,calculatedthroughcalibrationorenteredinthe"Maintenance"view.
• thevalueoftheoffsetofthepHsensor,calculatedthroughcalibrationorenteredinthe"Maintenance"view.
• themeasuredvalueofthewatersampletemperature.
• themeasuredimpedanceofthereferenceelectrode.
English
Type MS01
-
19
Adjustmentandoperation
• thetemperatureoffsetthathasbeenenteredinthe"Maintenance"viewofthesensor.
• themeasuredvalueofthetemperatureofthemeasurementboard.
• themeasuredvalueoftheISFETleakagecurrent.
• thetimetheproducthasalreadyoperated.
• thecalibrationlimitsoftheoffsetvalueoftheISFETsensor.
• thecalibrationlimitsoftheoffsetvalueofthepHsensor.
• thecalibrationlimitsoftheslopevalueofthepHsensor.
→ Fromthe"Parameter"viewofthesensor,tap toaccessthe"Diagnosis"view.
8.10 "Maintenance" view of the sensor
The"Maintenance"viewmakesitpossible:
• tomodifytheoffsetvalueoftheISFETsensor,
• tomodifytheoffsetvalueofthepHsensor,
• tomodifytheslopevalueofthepHsensor,
• toautomaticallycalibratethesensor,withthecalibrationwizard,
• tomodifytheoffsetvalueofthewatersampletemperature,
• tochecktheoperatingoftheproductbysimulatingsomedata,
• toreadthedateofthelastcalibration,
• toreadthedateofthenextduecalibration,
• tosetthenumberofdaysbetweentwocalibrations.
→ Fromthe"Parameter"viewofthesensor,tap toaccessthe"Maintenance"view.
8.10.1 calibrating the offset value of the iSfeT sensor
TheoffsetoftheISFETsensoriscalibratedinthefactory.
IftheISFETsensormustbereplacedbyanewone,thenewISFETsensormustbecalibrated.
→ IfyouknowtheoffsetvalueofthenewISFETsensor,gotothe"Maintenance"viewofthesensor,tapphcali-brationwizardisfetoffset andentertheoffsetvalueoftheISFETsensor.
→ IfyoudonotknowtheoffsetvalueofthenewISFETsensor,dothefollowingcalibrationprocedure.
1. Connectaperistalticpumpinthewatersampleinletcircuitofthesystem,asshowninFig.3.Ifthefluidicconnectionsarenotdonecorrectly,theproductandthesystemcanbedamagedbecausethepressureintheproductandinthesystemistoohigh.
English
Type MS01
-
20
Adjustmentandoperation
24 VDC
+-
OUT IN
Peristalticpump
Fig. 3: Installation for the calibration of the sensor in a system type 8905
2. Makesurethedirectionoftheflowiscorrect.
3. Makesurethefluidicinstallationforthecalibrationistight.
4. Letcleanwaterflowthroughthesystemtorinsetheproduct.
5. PrepareabuffersolutionwithapH7value.
6. Inthe"Maintenance"viewofthesensor,tapphcalibrationwizardZeroadjustment .
7. Step1/5:letthebuffersolutionflowthroughtheproduct.
8. Tap .
9. Step2/5:tapinputbuffervalue andenterthepHvalue7ofthebuffersolution.
10. Validate.
11. Tap .
12. Step3/5:whenthepHmeasurementisstable,tap .
English
Type MS01
-
21
Adjustmentandoperation
13. Step4/5:
• ifthecalibrationhassucceeded,thecalculatedoffsetvalueisdisplayed.Tap togotothestep5/5.
• ifanerrormessageisdisplayed,refertoTable7.
• ifawarningmessageisdisplayed,refertoTable8.
14. Step5/5:validateorcancelthecalibration.
Table 7: Error message at the end of the calibration of the ISFET sensor
Displayedmessage ComputedcalibrationvalueoutoferrorlimitssymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thesensorcalibrationhasfailedbecausethecalculatedoffsetvalueis
outoftheerrorrange.Whattodo? → SendtheproductbacktoBürkert.
Table 8: Warning message at the end of the calibration of the ISFET sensor
Displayedmessage ComputedcalibrationvalueoutofwarninglimitssymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thecalculatedoffsetvalueisoutofthewarningrangebecauseawrong
buffersolutionhasbeenusedforthecalibration.Whattodo? → Makesurethebuffersolutionusedisthecorrectone.
→ Ifitisnotthecorrectone,abortthecalibration,useacorrectbuffersolutionanddothecompletecalibrationprocedureagain.
→ Youcanchoosetoeithervalidateorcancelthecalibration.
• Ifyouvalidatethecalibration,thenewcalculatedoffsetvalueisusedtodeterminethepHvalue,andthelastcalibrationdateisupdated.
• Ifyoucancelthecalibration,thecurrentoffsetvalueisusedtodeterminethepHvalue,andthelastcalibrationdateisnotupdated.
8.10.2 calibrating the offset and/or slope values of the ph sensor
TomeasurepHvalueswithaslessdeviationaspossibleyoumustcalibratethepHsensor.
Thiscanbedone:
• eitherbyadjustingbyhandtheslopevalueand/ortheoffsetvalueofthepHsensor:inthe"Maintenance"viewofthesensor,tapphslope andenteraslopevalueand/ortapphoffset andenteranoffsetvalue.
• orbydoinga1pointcalibrationproceduretoautomaticallyadjusttheoffsetvalueofthepHsensor.Seechap.8.10.3.
• orbydoinga2pointcalibrationproceduretoautomaticallyadjusttheoffsetvalueandtheslopevalueofthepHsensor.Seechap.8.10.4.
English
Type MS01
-
22
Adjustmentandoperation
8.10.3 Doing a 1 point calibration procedure of the ph sensor
Dangerduetothenatureofthefluid
▶ Respecttheregulationsonaccidentpreventionandsafetyrelatingtotheuseofaggressivefluids.
Doa1pointcalibrationproceduretoautomaticallyadjusttheoffsetvalueofthepHsensor.
1. Connectaperistalticpumpinthewatersampleinletcircuitofthesystem,asshowninFig.3,chap.8.10.1.Ifthefluidicconnectionsarenotdonecorrectly,theproductandthesystemcanbedamagedbecausethepressureintheproductandinthesystemistoohigh.
2. Makesurethedirectionoftheflowiscorrect.
3. Makesurethefluidicinstallationforthecalibrationistight.
4. Letcleanwaterflowthroughthesystemtorinsetheproduct.
5. PrepareabuffersolutionwithapHvalueascloseaspossibletothewatersampleandwithapHvaluethatiscompatiblewithallthesensorcubesofthesystem.
6. Inthe"Maintenance"viewofthesensor,tapphcalibrationwizard1point .
7. Step1/5:letthebuffersolutionflowthroughtheproduct.
8. Tap .
9. Step2/5:tapinputbuffervalue andenterthepHvalueofthebuffersolution.
10. Validate.
11. Tap .
12. Step3/5:whenthepHmeasurementisstable,tap .
13. Step4/5:
• ifthecalibrationhassucceeded,thecalculatedoffsetvalueisdisplayedandthedateofthelastcalibrationis
updated(seechap.8.10.7).Tap togotothestep5/5.
• ifanerrormessageisdisplayed,refertoTable9.
• ifawarningmessageisdisplayed,refertoTable10.
14. Step5/5:validateorcancelthecalibration.
English
Type MS01
-
23
Adjustmentandoperation
Table 9: Error message at the end of the 1 point calibration of the pH sensor
Displayedmessage ComputedcalibrationvalueoutoferrorlimitssymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thesensorcalibrationhasfailedbecausethecalculatedoffsetvalueis
outoftheerrorrange.Whattodo? 1. Comparethecalculatedoffsetvaluewiththecalibrationlimitsthat
canbereadinthe"Diagnosis"view.
2. Replacetheexternalreferenceelectrodebyanewone.
3. Dothecalibrationagain.
4. Ifthecalibrationfailsagain,replacetheproductbyanewone.
Table 10: Warning message at the end of the 1 point calibration of the pH sensor
Displayedmessage ComputedcalibrationvalueoutofwarninglimitssymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thecalculatedoffsetvalueisoutofthewarningrangebecause:
• eitherawrongbuffersolutionhasbeenusedforthecalibration.
• ortheexternalreferenceelec-trodeisaging.
Whattodo? 1. Makesurethebuffersolutionusedisthecorrectone.
2. Ifitisnotthecorrectone,abortthecalibration.
3. Useacorrectbuffersolutionanddothecompletecali-brationprocedureagain.
→ Youcanchoosetoeitherval-idateorcancelthecalibration.
• Ifyouvalidatethecalibration,thenewcalculatedoffsetvalueisusedtodeterminethepHvalue,andthelastcalibrationdateisupdated.
• Ifyoucancelthecalibration,thecurrentoffsetvalueisusedtodeterminethepHvalue,andthelastcalibrationdateisnotupdated.
→ Plantoreplacetheexternalreferenceelectrode.
8.10.4 Doing a 2 point calibration procedure of the ph sensor
Dangerduetothenatureofthefluid
▶ Respecttheregulationsonaccidentpreventionandsafetyrelatingtotheuseofaggressivefluids.
Doa2pointcalibrationproceduretoautomaticallyadjusttheoffsetvalueandtheslopevalueofthepHsensor.
1. Connectaperistalticpumpinthewatersampleinletcircuitofthesystem,asshowninFig.3,chap.8.10.1.Ifthefluidicconnectionsarenotdonecorrectly,theproductandthesystemcanbedamagedbecausethepressureintheproductandinthesystemistoohigh.
2. Makesurethedirectionoftheflowiscorrect.
English
Type MS01
-
24
Adjustmentandoperation
3. Makesurethefluidicinstallationforthecalibrationistight.
4. Letcleanwaterflowthroughthesystemtorinsetheproduct.
5. PreparetwobuffersolutionswithpHvaluesthatarecompatiblewithallthesensorcubesofthesystem.
6. Inthe"Maintenance"viewofthesensor,tapphcalibrationwizard2point .
7. Step1/8:letthefirstbuffersolutionflowthroughtheproduct.
8. Tap .
9. Step2/8:tapinputvalueofbuffer1 andenterthepHvalueofthebuffersolution.
10. Tap tovalidate.
11. Tap .
12. Step3/8:whenthepHmeasurementisstable,tap .
13. Letcleanwaterflowthroughthesystemtorinsetheproduct.
14. Step4/8::letthesecondbuffersolutionflowthroughtheproduct.
15. Tap .
16. Step5/8:tapinputvalueofbuffer2 andenterthepHvalueofthebuffersolution.
17. Tap tovalidate.
18. Tap .
19. Step6/8:whenthepHmeasurementisstable,tap .
20. Step7/8:
• ifthecalibrationhassucceded,thecalculatedslopevalueandthecalculatedoffsetvaluearedisplayedandthe
dateofthelastcalibrationisupdated(seechap.8.10.7).Tap togotothestep8/8.
• ifanerrormessageisdisplayed,refertoTable11.
• ifawarningmessageisdisplayed,refertoTable12.
English
Type MS01
-
25
Adjustmentandoperation
21. Step8/8:validateorcancelthecalibration.
Table 11: Error message at the end of the 2 point calibration of the pH sensor
Displayedmessage ComputedcalibrationvalueoutoferrorlimitssymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thesensorcalibrationhasfailedbecausethecalculatedoffsetvalue
and/orthecalculatedslopevalueareoutoftheerrorrange.
→ Comparethecalculatedoffsetvalueandthecalculatedslopevaluewiththecalibrationlimitsthatcanbereadinthe"Diagnosis"view.
• Iftheoffsetvalueisoutoftheerrorlimits:
• Iftheslopevalueisoutoftheerrorlimits:
Whattodo? 1. Replacetheexternalreferenceelectrodebyanewone.
2. Dothecalibrationagain.
3. Ifthecalibrationfailsagain,replacetheproductbyanewone.
→ SendtheproductbacktoBürkerttohavetheISFETsensorreplaced.
English
Type MS01
-
26
Adjustmentandoperation
Table 12: Warning message at the end of the 2 point calibration of the pH sensor
Dis
play
edm
essa
geC
ompu
ted
calib
ratio
nva
lue
outo
fwar
ning
lim
itss
ymbo
ldis
play
edo
nth
eu
niqu
eD
evic
e n
ame
oft
hep
rodu
ctp
ossi
ble
caus
eTh
eca
lcul
ated
offs
etv
alue
and
/ort
hec
alcu
late
dsl
ope
valu
ear
eou
toft
hew
arni
ngra
nge:
•If
the
offs
etv
alue
iso
uto
fthe
war
ning
lim
its:
•If
the
slop
eva
lue
iso
uto
fth
ew
arni
ngli
mits
:•
eith
era
wro
ngb
uffe
rsol
utio
nha
sbe
enu
sed
fort
hec
alib
ratio
n.•
orth
eex
tern
alre
fere
nce
elec
-tr
ode
isa
ging
.W
hat
tod
o?1.
M
ake
sure
the
buffe
rsol
utio
nus
edis
the
corr
ecto
ne.
2.
Ifit
isn
otth
eco
rrec
tone
,abo
rt
the
calib
ratio
n.
3.
Use
ac
orre
ctb
uffe
rsol
utio
nan
ddo
the
com
plet
eca
li-br
atio
npr
oced
ure
agai
n.
→
You
can
choo
seto
eith
erv
al-
idat
eor
can
celt
hec
alib
ratio
n.
•If
you
valid
ate
the
calib
ratio
n,th
ene
wc
alcu
late
dof
fset
val
ueis
us
edto
det
erm
ine
the
pHv
alue
,an
dth
ela
stc
alib
ratio
nda
teis
up
date
d.
•If
you
canc
elth
eca
libra
tion,
th
ecu
rren
toffs
etv
alue
isu
sed
tod
eter
min
eth
epH
val
ue,a
nd
the
last
cal
ibra
tion
date
isn
ot
upda
ted.
→
Pla
nto
repl
ace
the
exte
rnal
ref-
eren
cee
lect
rode
.
→
Sen
dth
epr
oduc
tbac
kto
B
ürke
rtto
hav
eth
eIS
FET
sens
orre
plac
ed.
English
Type MS01
-
27
Adjustmentandoperation
8.10.5 Adjusting the offset value of the temperature
Thetemperaturemeasuredbythesensorcanbecorrectedwithanoffsetvalue.
1. Inthe"Maintenance"viewofthesensor,taptemperatureoffset andenterthevalueofthetem-peratureoffset,inthedisplayedtemperatureunits.
2. Tochangetheunitsofthetemperature,gotothe"Parameter"viewofthesensor,taptemperatureDesiredunit andchoosethetemperatureunits:theoffsetvalueofthetemperatureisautomaticallyconverted.
8.10.6 Simulating some data
Checktheoperating(forexample,makesurethewarningand/orerrorlimitsarecorrectlyset)oftheproductand/orprocessbysimulatingsomedata.
1. Inthe"Maintenance"viewofthesensor,tapsimulationstatus andchoose"ON"toactivatethesimulation.
2. Validate.
3. TosimulateapHvalue,tapsimulationph andenterthepHvaluetobesimulated.
4. Tosimulateatemperaturevalue,tapsimulationtemperature andenterthetemperaturevaluetobesimulated.
8.10.7 reading the date of the last calibration
Toreadthedateofthelastcalibrationthathassucceeded.
→ Inthe"Maintenance"viewofthesensor,readthedateinthefieldcalibrationschedulelastcalibration.
8.10.8 reading the date of the next due calibration
Toreadthedateofthenextcalibrationthatmustdone.
→ Inthe"Maintenance"viewofthesensor,readthedateinthefieldcalibrationschedulenextcalibration.
8.10.9 Setting the time interval between two calibrations
Tosetthetimeinterval,indays,betweentwocalibrations:
1. Inthe"Maintenance"viewofthesensor,tapcalibrationscheduleintervalindays .
1. Enterthenumberofdaysbetweentwocalibrations.Werecommendtocalibratetheproducteverymonth.
2. Validate.
Whentheduecalibrationdateisreached,awarningmessageisdisplayedintheMessagesList.
English
Type MS01
-
28
Maintenanceandtroubleshooting
9 MAinTenAnce AnD TroubleShooTing
9.1 Safety instructions
Warning
riskofinjuryduetonon-conformingmaintenance.
▶ Maintenancemustonlybecarriedoutbyqualifiedandskilledstaffwiththeappropriatetools.
note
riskofdamagetotheproductduetothepowersupply
• Shutdownandisolatetheelectricalpowersourcebeforecarryingoutworkonthesystem.
9.2 cleaning of the product
danger
riskofinjuryduetothenatureofthedetergent.
▶ Respecttheregulationsonaccidentpreventionandsafetyrelatingtotheuseofaggressivefluids.
Theproductcanbecleanedwithaclothdampenedwithwater(max.40°C)orwithanaciddetergent(withmax.5%ofhydrochloricacid).
9.3 replacing the external reference electrode
TheexternalreferenceelectrodemustbereplacediftheerrorcalibrationlimitsfortheoffsetvalueofthepHsensorareexceeded.
Proceduretoreplacetheexternalreferenceelectrode:
1. RemovethepHsensorcubefromthesystemasdescribedintheOperatingIsntructionsofthesystemtype8905.
2. Liftthedropleverofthereferenceelectrodeandturnitbyaquarterturntotheleft.
Droplever
3. Pullthelevertoremovetheexternalreferenceelectrode.
4. Disposeoftheusedexternalreferenceelectrode.Keeptotheexistingprovisionsonthesubjectofwastedisposalandenvironmentalprotection.
English
Type MS01
-
29
Maintenanceandtroubleshooting
5. Ifnecessary,cleantheemptyseatingoftheelec-trodereferencewithadryclothorcottonswabs.
6. Preparethenewreferenceelectrode.
7. CarefullyremovetheprotectivecapbecauseitcontainssomeKCl.
8. Makesurethesealisinitsgrooveontheexternalreferenceelectrode.
9. Ifnecessary,removethesoliddepositsfromthecapillary.
10. Moistenthesealwithwater.Ifnot,thesealcanbedamaged.
Seal
Capillary
11. Liftthedropleverofthereferenceelectrodeandturnitbyaquarterturntotheleft.
12. Insertthenewexternalreferenceelectrodefully.
13. Turnthedropleverbyaquarterturntotherightandloweritfullyinitsseating.
14. MountthepHsensorcubebackinthesystemasdescribedintheOperatingIsntructionsofthesystemtype8905.
9.4 Troubleshouting if no message is displayed
colouroftheproductstatusleD OFF
possiblecause Theproduct/thesystemisnotenergized.
Whattodo? 1. Checkthewiring.
2. Makesurethevoltagesupplyis24VDC.
3. Checkthatthepowersupplysourceisworkingproperly.
English
Type MS01
-
30
Maintenanceandtroubleshooting
9.5 Troubleshooting if the status leD of the product is red or orange
Ifanerrororawarningmessagehasbeengeneratedbythesystem:
• thestatuslightofthesystemisredororange,
• thestatusLEDoftheproductisredororange,
• thesymbol or appearsontheUniqueDeviceNameoftheproduct,inthelistofdevices,
• thesymbol appearsinthetopleftcornerofthedisplay.
→ Tap toaccessthe"MessagesList".
9.5.1 Message "Too high leakage current"
Displayedmessage ToohighleakagecurrentsymboldisplayedontheuniqueDevicenameoftheproductpossiblecause TheleakagecurrentoftheISFETsensorisoutofrange.
Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):
• theerror"activationflag"forthehighthresholdoftheleakagecurrentisselected,
• andthehighthresholdsetfortheleakagecurrentoftheISFETsensorhasbeenreached.
Whattodo? → SendtheproductbacktoBürkerttohavetheISFETsensorreplaced.
Displayedmessage ToohighleakagecurrentsymboldisplayedontheuniqueDevicenameoftheproductpossiblecause TheleakagecurrentoftheISFETsensorisoutofrange.
ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):
• thewarning"activationflag"forthehighthresholdoftheleakagecurrentisselected,
• andthehighthresholdsetfortheleakagecurrentoftheISFETsensorhasbeenreached.
Whattodo? → Acknoledgethewarningmessage:theproductcontinuestooperate.
→ or,sendtheproductbacktoBürkerttohavetheISFETsensorreplaced.
→ orreplacetheproductbyanewone.
English
Type MS01
-
31
Maintenanceandtroubleshooting
9.5.2 Message "Too low leakage current"
Displayedmessage ToolowleakagecurrentsymboldisplayedontheuniqueDevicenameoftheproductpossiblecause TheleakagecurrentoftheISFETsensorisoutofrange.
Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):
• theerror"activationflag"forthelowthresholdoftheleakagecurrentisselected,
• andthelowthresholdsetfortheleakagecurrentoftheISFETsensorhasbeenreached.
Whattodo? → Deselecttheerror"activationflag"forthelowthresholdoftheleakagecurrent.
Displayedmessage ToolowleakagecurrentsymboldisplayedontheuniqueDevicenameoftheproductpossiblecause TheleakagecurrentoftheISFETsensorisoutofrange.
ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):
• thewarning"activationflag"forthelowthresholdoftheleakagecurrentisselected,
• andthelowthresholdsetfortheleakagecurrentoftheISFETsensorhasbeenreached.
Whattodo? → Deselectthewarning"activationflag"forthelowthresholdoftheleakagecurrent.
9.5.3 Message "Too high ph value"
Displayedmessage ToohighpHvaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause ThepHvalueofthewatersampleisoutofrange.
Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):
• theerror"activationflag"forthehighthresholdofthepHvalueisselected,
• andthehighthresholdsetforthepHvaluehasbeenreached.Whattodo? → Checktheprocess.
English
Type MS01
-
32
Maintenanceandtroubleshooting
Displayedmessage ToohighpHvaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause ThepHvalueofthewatersampleisoutofrange.
ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):
• thewarning"activationflag"forthehighthresholdofthepHvalueisselected,
• andthehighthresholdsetforthepHvaluehasbeenreached.Whattodo? → Checktheprocess.
9.5.4 Message "Too low ph value"
Displayedmessage ToolowpHvaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause ThepHvalueofthewatersampleisoutofrange.
Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):
• theerror"activationflag"forthelowthresholdofthepHvalueisselected,
• andthelowthresholdsetforthepHvaluehasbeenreached.Whattodo? → Checktheprocess.
Displayedmessage ToolowpHvaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause ThepHvalueofthewatersampleisoutofrange.
ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):
• thewarning"activationflag"forthelowthresholdofthepHvalueisselected,
• andthelowthresholdsetforthepHvaluehasbeenreached.Whattodo? → Checktheprocess.
English
Type MS01
-
33
Maintenanceandtroubleshooting
9.5.5 Message "Too high temperature value"
Displayedmessage ToohightemperaturevaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thetemperaturevalueofthewatersampleisoutofrange.
Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):
• theerror"activationflag"forthehighthresholdofthewatersampletemperaturevalueisselected,
• andthehighthresholdsetforthewatersampletemperaturevaluehasbeenreached.
Whattodo? → Checktheprocess.
Displayedmessage ToohightemperaturevaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thetemperaturevalueofthewatersampleisoutofrange.
ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):
• thewarning"activationflag"forthehighthresholdofthewatersampletemperaturevalueisselected,
• andthehighthresholdsetforthewatersampletemperaturevaluehasbeenreached.
Whattodo? → Checktheprocess.
9.5.6 Message "Too low temperature value"
Displayedmessage ToolowtemperaturevaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thetemperaturevalueofthewatersampleisoutofrange.
Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):
• theerror"activationflag"forthelowthresholdofthewatersampletemperaturevalueisselected,
• andthelowthresholdsetforthewatersampletemperaturevaluehasbeenreached.
Whattodo? → Checktheprocess.
English
Type MS01
-
34
Maintenanceandtroubleshooting
Displayedmessage ToolowtemperaturevaluesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Thetemperaturevalueofthewatersampleisoutofrange.
ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):
• thewarning"activationflag"forthelowthresholdofthewatersampletemperaturevalueisselected,
• andthelowthresholdsetforthewatersampletemperaturevaluehasbeenreached.
Whattodo? → Checktheprocess.
9.5.7 Message "Too high reference impedance"
Displayedmessage ToohighreferenceimpedancesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Theimpedanceofthereferenceelectrodeisoutofrange.
Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):
• theerror"activationflag"forthehighthresholdoftheimpedanceofthereferenceelectrodeisselected,
• andthehighthresholdsetfortheimpedanceofthereferenceelec-trodehasbeenreached.
Whattodo? 1. Gointothe"Diagnostic"viewofthesensortoreadtheimpedancevalueofthereferenceelectrode.
2. Makesurethehighthresholdisconsistentwiththevalueofthecon-ductivityofthewatersample.
3. Ifnecessary,replacetheexternalreferenceelectrode.
4. Iftheexternalreferenceelectrodehasbeenreplaced,calibratethepHsensor.
English
Type MS01
-
35
Maintenanceandtroubleshooting
Displayedmessage ToohighreferenceimpedancesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Theimpedanceofthereferenceelectrodeisoutofrange.
ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):
• thewarning"activationflag"forthehighthresholdoftheimpedanceofthereferenceelectrodeisselected,
• andthehighthresholdsetfortheimpedanceofthereferenceelec-trodehasbeenreached.
Whattodo? 1. Gointothe"Diagnostic"viewofthesensortoreadtheimpedancevalueofthereferenceelectrode.
2. Makesurethehighthresholdisconsistentwiththevalueofthecon-ductivityofthewatersample.
3. Plantoreplacetheexternalreferenceelectrode.
9.5.8 Message "Too low reference impedance"
Displayedmessage ToolowreferenceimpedancesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Theimpedanceofthereferenceelectrodeisoutofrange.
Themessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggerserrors ("Parameter"viewofthesensor):
• theerror"activationflag"forthelowthresholdoftheimpedanceofthereferenceelectrodeisselected,
• andthelowthresholdsetfortheimpedanceofthereferenceelec-trodehasbeenreached.
Whattodo? → Deselecttheerror"activationflag"forthelowthresholdoftheimpedanceofthereferenceelectrode.
Displayedmessage ToolowreferenceimpedancesymboldisplayedontheuniqueDevicenameoftheproductpossiblecause Theimpedanceofthereferenceelectrodeisoutofrange.
ThemessageisdisplayedwhenthefollowingsettingshavebeenmadeinmessagestriggersWarnings ("Parameter"viewofthesensor):
• thewarning"activationflag"forthelowthresholdoftheimpedanceofthereferenceelectrodeisselected,
• andthelowthresholdsetfortheimpedanceofthereferenceelec-trodehasbeenreached.
Whattodo? → Deselectthewarning"activationflag"forthelowthresholdoftheimpedanceofthereferenceelectrode.
English
Type MS01
-
36
Sparepartsandaccessories
10 SpAre pArTS AnD AcceSSorieS
caution
riskofinjuryand/ordamagecausedbytheuseofunsuitableparts.
Incorrectaccessoriesandunsuitablereplacementpartsmaycauseinjuriesanddamagetheproductandthesurroundingarea.
▶ UseonlyoriginalaccessoriesandoriginalreplacementpartsfromBürkert.
Accessory ordercodeBuffersolution,pH=5 566031Buffersolution,pH=7 418541Buffersolution,pH=8 onrequestExternalreferenceelectrode 566084
11 pAckAging, TrAnSporT
note
Damageduetotransport
Transportmaydamageaninsufficientlyprotectedproduct.
• Transporttheproductinshock-resistantpackagingandawayfromhumidityanddirt.
• Donotexposetheproducttotemperaturesthatmayexceedtheadmissiblestoragetemperaturerange.
• Protecttheelectricalinterfacesusingprotectiveplugs.
12 STorAge
note
poorstoragecandamagetheproduct.
▶ Dependingonthedurationofthestoragetime,obeythestorageinstructionsinthechapter12.1to12.3.
▶ Afteranystorageperiod,obeytheinstructionsinchap.12.4.
12.1 To store the product for a maximum of 4 days
1. Rinsetheproductwithtapwater.
2. Purgetheproductwithairatamax.pressureof2bar.
3. Storetheproductatroomtemperature(about23°C±10°C).
4. Storetheproductinadryplaceawayfromdust.
English
Type MS01
-
37
Sparepartsandaccessories
12.2 To store the product for more than 4 days and less than 1 month
1. Rinsetheproductwithtapwater.
2. DONOTPURGETHEPRODUCTbecauseitwilldryout.
3. Storetheproductatroomtemperature(about23°C±10°C).
4. Storetheproductinadryplaceawayfromdust.
12.3 To store the product for a more than 1 month
1. Removetheexternalreferenceelectrode.
2. Rinsetheproductwithtapwater.
3. Purgetheproductwithairatamax.pressureof2bar.
4. Storetheproductatroomtemperature(about23°C±10°C).
5. Storetheproductinadryplaceawayfromdust.
12.4 After storage
Beforecommisioningaproductthathasbeenstored:
1. Mountanewexternalreferenceelectrode.Seechap.9.3.
2. Plugtheproductinthesystem.
3. WhilethepowersupplyofthesystemisOFF,letthewatersampleflowthroughtheproductforatleast2hours.
4. Calibratetheproduct.Seechap.8.10.
13 DiSpoSAl of The proDucT→ Disposeoftheproductanditspackaginginanenvironmentally-friendlyway.
noteDamagetotheenvironmentcausedbyproductscontaminatedbyfluids.
• Keeptotheexistingprovisionsonthesubjectofwastedisposalandenvironmentalprotection.