TriMed Ulnar Osteotomy System™Surgical Technique
Ulnar Osteotomy Compression PlateTM
Plate Application
• ExposeboneintheintervalbetweentheECUandFCU.
• Applyplateonbone.Ensurethatthetwosmallholesonthesideoftheplatearevisible.
• Positionplatetoavoiddistalimpingementduringshortening. • Insertone3.2mmscrewintheslottedholeattheendfarthest
fromtheosteotomysite,andthree3.2mmscrewsontheoppositesideofplate,startingwiththeendscrewhole.
Guide and Pins
• PlacetheCombinationDrill(CD)Guideontheplate.
• Insertashort1.6mm(0.062”)K-wireflushalongthebackwallofthepinslotintheCDGuideandthroughonecortex.
• RepeatwithlongerK-wireandremoveCDGuide.
Osteotomy
• Selecttheappropriate‘A’guidebasedontheplannedresection(2,3,4,5mm)andinsertintotheplate.Holdguidewithboneclampor1.6mmK-wire(seeTips).
• Makecutwithasawbladeof0.4mmthickness.Irrigateliberally.
• Remove‘A’guide,insert‘B’guide,andmakesecondcut. Removebonewafer.
Application of Compression Clamp
• InserttheCompressionClampintotheholeonthefixededgeoftheplate.
• AdjusttheclampsotheK-wiresleeveisjustbelowthecentralaxisoftheulnaandawayfromtheosteotomysite.
• Inserta1.6mm(0.062”)K-wirethroughthesleeveengagingthefarcortex.
1
2
3
4
UlnarOsteotomyCompressionPlateTM
2
Compression of the Osteotomy
• Loosenthescrewintheslottedholeonly1/4 turn.
• ShortenulnawithCompressionClamp.Useaboneclampforaddedcompressionofosteotomysite.
• Checkbothsidesofosteotomysiteforbonecoaptation.
Prepare Lag Screw Hole
• Re-applyCDGuideanddrilllagscrewholeusing2.3mm(red)drillandguide.
• RemoveCDGuide.
• Measuretothelongsideofthehole.
• Useabonetaptoensurethatlagscrewiseasytoinsert.
Insert Lag Screw
• Insertlagscrewensuringpenetrationoffarcortex.
• Re-tightencorticalscrewinslottedholewatchingthatosteotomyremainscompressed.
Final Fixation
• Inserttheremainingcorticalscrews.
• RemoveK-wiresandclamps.
5
7
6
8
UlnarOsteotomyCompressionPlateTM
3
TIPS
Ulnar Osteotomy Compression PlateUOCP
• A1.6mmK-wireortheboneclampcanbeusedto stabilizethecuttingguidestotheplatewhile osteotomycutsaremade.
• Osteotomieslargerthan5mmmaybemadeby performingsuccessivecuts.
Bone ClampOBC
Compression ClampCOMPC-UOS
Saw GuideUOSG-A-x02mm to 05mm
UOSG-B
K-WireWIRE-1.6/065WIRE-1.6/100
LAG HOLE K-WIRE SLOTS SLOTTEd HOLE
InSTRUMEnT HOLES
CombinationDrill GuideGdUOS-1.6
Cortical ScrewHEX3.2-xx
10mm to 18mm
Cortical Locking ScrewLCBS3.2-xx10mm to 18mm
Cortical Lag ScrewLAG3.2-xx14mm to 26mm
Saw BladeOSB-9x
0.4mm thickness(Use blade lengths of 25–32mm
and widths of 6–10mm)
©2017 TriM
ed, Inc. All rights reserved. TME 840-59 (07-17)
TriMed,Inc./27533AvenueHopkins/Valencia,CA91355USA/800-633-7221/www.trimedortho.com
PatentCoverage:TriMed,Inc.productsarecoveredbypatentsissuedintheU.S.andinforeignjurisdictions.ThepresentlyissuedU.S.patentsare:5,709,682;5,931,839;5,941,878;6,077,266;6,113,603;7,037,308;7,195,633;7,540,874;8,177,822;8,821,508;8,906,070;9,089,376;9,283,010;9,220,546.TheTriMedUlnarOsteotomyCompressionPlatehasU.S.andinternationalpatentspending.TriMedUlnarOsteotomyCompressionPlateisatrademarkofTriMed,Inc.
Thetechniquepresentedisonesuggestedsurgicaltechnique.Thedecisiontouseaspecificimplantandthesurgicaltechniquemustbebasedonsoundmedicaljudgmentbythesurgeonthattakesintoconsiderationfactorssuchasthecircumstancesandconfigurationoftheinjury.