ultimate core performance to maximize your …...to maximize your investment thermo scientific...

16
Ultimate Core Performance to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations

Upload: others

Post on 28-Jul-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

Ultimate Core Performance

to Maximize Your Investment

Thermo Scientific Accucore HPLC Columns for Biomolecule Separations

Page 2: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

Automated Packing ProcessEnhanced automated procedures ensure that all columns are packed with the highest quality

Tight Control of Particle Diameter Enhanced selection process keeps particle size distribution to a minimum and produces high efficiency columns

Solid Core Particles2.6 µm diameter particles with a solid core generate high speed, high resolution separations without excessive backpressure

Advanced Bonding TechnologyOptimized phase bonding creates a series of high coverage, robust phases

The range of Accucore™ HPLC columns packed with 150 Å pore diameter particles allow biomolecule separations to benefit from the superb resolution and high speed enabled by the Core Enhanced Technology™ that Accucore is based on.

The key components of Core Enhanced Technology are:

Core Enhanced Technology – for Biomolecules

150 Å Solid Core

ParticlesTight Control of

Particle DiameterAdvanced Bonding

TechnologyAutomated Packing

ProcessFast Separations • • • •High Peak Capacity • • • •Increased Sensitivity • • • •Lower Pressure • • •Good Loading Capacity • • •Reproducible Chromatography • • •Long Lifetime • • • •Peptide and Protein Analysis • •

The table above shows how the components of Core Enhanced Technology with 150 Å pore diameter particles contribute to the outstanding features of Accucore HPLC columns for biomolecule separations.

Page 3: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

Thermo Scientific Accucore HPLC Columns for Biomolecule Separations 1

Table of Contents

Why Core Enhanced Technology Works 2

150 Å Pore Diameter Particles 3

Column Formats 4

Analytical and Micro Columns. . . . . . . . . . . . . . . . . . . . 4 Defender Guard Cartridges . . . . . . . . . . . . . . . . . . . . . 4 nanoViper Columns . . . . . . . . . . . . . . . . . . . . . . . . . 4Biomolecule Phases 5

Accucore 150-C18 Column 6

Peptide Separations 6

High Peak Capacity . . . . . . . . . . . . . . . . . . . . . . . . . 7 Reproducible Separations . . . . . . . . . . . . . . . . . . . . . . 7 Excellent Loading Capacity . . . . . . . . . . . . . . . . . . . . . 8Accucore 150-C4 Column 9

Protein Separation 9

Excellent Resolution . . . . . . . . . . . . . . . . . . . . . . . . 10 Reproducible Results . . . . . . . . . . . . . . . . . . . . . . . 10 Proven Loading Capacity. . . . . . . . . . . . . . . . . . . . . . 11Thermo Scientific nanoViper Column Separations 12

Ordering Information

Accucore HPLC Columns. . . . . . . . . . . . . . . . . . . . . . 13 Accucore Defender Guard Cartridges . . . . . . . . . . . . . . . 13

Thermo Scientific Accucore HPLC Columns for Biomolecule Separations 1

Protein Separation 9

Excellent Resolution . . . . . . . . . . . . . . . . . . . . . . . . 10Reproducible Results . . . . . . . . . . . . . . . . . . . . . . . 10Proven Loading Capacity. . . . . . . . . . . . . . . . . . . . . . 11

Thermo Scientific nanoViper Column Separations 12

Ordering Information

Accucore HPLC Columns. . . . . . . . . . . . . . . . . . . . . . 13Accucore Defender Guard Cartridges . . . . . . . . . . . . . . . 13

Page 4: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

2 Thermo Scientific Accucore HPLC Columns for Biomolecule Separations

Why Core Enhanced Technology WorksTwo factors that strongly affect chromatographic efficiency are mass transfer and eddy diffusion, the C and A terms respectively from the Van Deemter equation.

Lower BackpressureL Column length (cm)

h Mobile phase viscosity (cP)

F Flow rate (mL/min)

dp2 Particle diameter (μm)

dc2 Column diameter (cm)

The equation above shows how backpressure is related to particle diameter.

The tightly controlled 2.6 µm diameter of Accucore particles also results in lower backpressures than typically seen with sub-2 µm materials.

Resistance to mass transfer is minimized by the solid core design of Accucore particles as the diffusional path of analytes is limited by the depth of the outer porous layer.

The tight control of particle diameter and automated packing process used for Accucore columns result in a tight, highly uniform packed bed that minimizes eddy diffusion.

Material Accucore 2 6 μmParticle Size Distribution (D90/10)

1 15

Bu + CuA +H =

H Height equivalent to theoretical plate (column length/efficiency)

A Eddy diffusion

B Longitudinal diffusion

C Resistance to mass transfer

u Mobile phase linear velocity

250LhF

dp2dc

2DP ~

Page 5: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

Thermo Scientific Accucore HPLC Columns for Biomolecule Separations 3

150 Å Pore Diameter ParticlesThe 150 Å pore diameter solid core particles used in the Accucore 150-C18 and 150-C4 columns are designed specifically to provide the optimum combination of retention and resolution for peptides and proteins.

The chromatograms below show the improved resolution and peak shapes achieved with Accucore 150-C18 compared to a C18 solid core phase with an 80 Å pore diameter.

Accucore 150-C18

Conventional Solid Core C18

1.12

3.76 7.09 8.91 13.468.74

11.20

1.01

3.60 6.98

10.84

13.13

8.52

Peak

Cap

acity

Stationary Phase

607080

90100110120130140

150

Accucore 150-C18 ConventionalSolid Core C18

Peak Capacity Comparison

Sharper peaks result in a significantly higher peak capacity for 150-C18 compared to the conventional solid core C18 column.

Improved resolution Improved peak shape

Page 6: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

4 Thermo Scientific Accucore HPLC Columns for Biomolecule Separations

Column FormatsAccucore HPLC columns for biomolecules are offered in both analytical, micro and nano formats. Optimum conditions and backpressure ratings are shown in the table below.

Analytical and Micro ColumnsAccucore HPLC columns are packed into our high pressure hardware. These stainless steel columns are engineered to the highest quality and have a pressure rating of 1000 bar. Thermo Scientific Defender Guard CartridgesGuard columns are designed to protect your column from particulates introduced from the matrix or instrument and from any strongly retained components in the injected sample.

Defender™ Guard Cartridges have been designed specifically to work with high speed, high efficiency separations. nanoViper ColumnsThe nanoViper fingertight connection system for nanoLC connections eliminates the assembly of PEEK sleeve connections. It is preassembled and fingertight to maximize ease-of-use. The nanoViper fitting is capable of withstanding pressures up to a 1000 bar and is compatible with third party valves and unions.

Accucore nanoViper columns are available in 150 and 500 mm lengths – for ultra-high peak capacity.

Format Column ID Optimum Flow Rate

Optimum Injection Volume

Backpressure Rating

Temperature Rating

Nano 75 μm 300 nL/min 1 μL 800 bar 70 °C

Micro 2.1 mm 400 μL/min 1 μL 1000 bar 70 °C

Analytical 3.0 mm 800 μL/min 3 μL 1000 bar 70 °C

Analytical 4.6 mm 1800 μL/min 5 μL 1000 bar 70 °C

Page 7: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

Thermo Scientific Accucore HPLC Columns for Biomolecule Separations 5

Biomolecule PhasesLike all other Accucore products, the 150 Å pore size particles and the phases based on them are fully characterized using the suite of tests shown below.

Silica Support Properties

Bonded Phase Properties

Hydrophobic Interactions

Secondary Interactions

Acidic Interactions

Surface Area Carbon loadHydrophobic retention Base activity Acid interaction

HR BA AI

Pore SizeHydrophobic selectivity Chelation

Ion exchange capacity (pH 2.7)

HS C IEX(2.7)

Particle Size Steric selectivityIon exchange capacity (pH 7.6)

SS IEX(7.6)

Particle Size Distribution

Hydrogen bonding capacity

HBC

Accucore 150-C18

Phase characteristics are designed for the separation of peptides

Accucore 150-C4

Lower hydrophobic retention, optimal retention of proteins and larger peptides

Details, including the phase characterization, are given below for the Accucore biomolecule phases.

IEX (2.7)

AI

C

BA IEX (7.6)

HR/10

HS

SS

HBC

IEX (2.7)

AI

C

BA IEX (7.6)

HR/10

HS

SS

HBC

Page 8: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

6 Thermo Scientific Accucore HPLC Columns for Biomolecule Separations

• Designed for the separation of peptides

• Outstanding resolution

Accucore 150-C18

IEX (2.7)

AI

C

BA IEX (7.6)

HR/10

HS

SS

HBC

Hydrophobicity

pH Range

Pore Size

Carbon Load (%)

Particle Size 2.6 µm USP L1

0 141 to 11

Low

0 300150 Å

0 257

Peptide SeparationsAccucore 150-C18

Mobile phase A: 0.1 % TFA in 10:90 acetonitrile:water

Mobile phase B: 0.1 % TFA in 70:30 acetonitrile:water

Gradient: 0-50 % B over 15 min; hold for 2 min; drop to 0 % in 0.1 min; hold at 0 % B for 5 min

Flow: 300 µL/min

Temperature: 35 °C

Column: Accucore 150-C18 2.6 µm, 100 x 2.1 mm

Injection: 5 µL

Detection: UV (220 nm)

120000

140000

160000

180000

200000

220000

240000

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

-40000

-20000

0

20000

40000

60000

80000

100000

1

2 34

56

7

Time (min)

mAU

Peak Number tR/min Peptide MW Concentration (µg/mL)

1 1.12 Glycine-Tyrosine 238.24 Da 2.0

2 3.76 Valine-Tyrosine-Valine 379.45 Da 17.0

3 7.09 Met-Enkephalin 573.66 Da 21.0

4 8.74 Angiotensin III 931.09 Da 15.0

5 8.91 Leu-Enkephalin 569.65 Da 21.0

6 11.20 Ribonuclease A ~ 13700 Da 42.5

7 13.46 Insulin 5733.49 Da 30.0

Page 9: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

Thermo Scientific Accucore HPLC Columns for Biomolecule Separations 7

High Peak Capacity

Higher peak capacities facilitate increased peptide identifications. Accucore 150-C18 provides much narrower peak widths, therefore significantly higher peak capacity than a column packed with < 2 µm wide pore fully porous C18.

Reproducible Separations

Precision of retention times is critical for reliable analysis. The Accucore 150-C18 column exhibits excellent retention time reproducibility.

Peak Number Peptide tR (min) % RSD

7 Insulin 13.58 0.11

Accucore 150-C18 Backpressure: 175 bar

Conventional Solid Core C18 Backpressure: 170 bar

< 2 µm Wide Pore Fully Porous C18 Backpressure: 320 bar

uAU

Time (min)

-50000

0

50000

100000

2 4 6 8 10 12 14

Peak

Cap

acity

Stationary Phase

132

134

136

138

140

142

144

146

Accucore 150-C18 < 2 µm Wide PoreFully Porous C18

Peak Capacity Comparison

nc Peak capacity

tg Gradient time

w Average peak width 10% height

tg wnc = 1 + ( )

13.0 13.5 14.0

Page 10: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

8 Thermo Scientific Accucore HPLC Columns for Biomolecule Separations8 Thermo Scientific Accucore HPLC Columns for Biomolecule Separations

Excellent Loading Capacity

The data below demonstrates the impact on peak capacity of increasing the load of peptides on the Accucore 150-C18 column. Only for loads exceeding 2 µg on the column are significant decreases in peak capacity observed.

Total Load on Column (µg) Peak Capacity

0.8 159.24

1.7 152.75

3.4 142.31

6.7 128.87

13.1 107.76

Peak

Cap

acity

Total Load on Column /µg

0

50

100

150

200

0 2 4 6 8 10 12 14

Time (min)

0

-200

400

800

1000

1200

1400

1600

1800

2000

2200

600

200

10 2 3 4 5 6 7 8 9 10 11 12 13 14 15

13.1 µg

6.7 µg

3.4 µg

1.7 µg

0.8 µg

mAU

Loading Capacity

Page 11: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

9Thermo Scientific Accucore HPLC Columns for Biomolecule Separations | Technical Guide

• Significantly lower hydrophobic retention than C18

• Ideal for retention of proteins and larger peptides

Mobile phase A: 0.1 % TFA in 30:70 acetonitrile:water

Mobile phase B: 0.1 % TFA in 98:2 acetonitrile:water

Gradient: 0-30 % B over 8 mins 30-95 % B over 2 mins 95 % B hold for 1 min 0 % B hold for 4 mins

Flow: 400 µL/min

Temperature: 40 °C

Column: Accucore 150-C4 2.6 µm, 100 x 2.1 mm

Injection: 2 µL 10 pmol/µL solution

Detection: UV (214 and 280 nm)

Accucore 150-C4

IEX (2.7)

AI

C

BA IEX (7.6)

HR/10

HS

SS

HBC

Hydrophobicity

pH Range

Pore Size

Carbon Load (%)

Particle Size 2.6 µm USP L26

0 142 to 9

Low

0 300150 Å

0 252

mAU

Time (min)

-10000

0

10000

20000

30000

40000

50000

60000

70000

80000

90000

100000

110000

120000

130000

0 1 2 3 4 5 6 7 8 9 10

1

2

3

*4

5

6

Peak Number tR/min Protein MW Concentration (µg/mL)

1 1.54 Insulin 6 kDa 40

2 2.66 Cytochrome C 12 kDa 80

3 4.42 Lysozyme 14 kDa 100

4 7.38 Myoglobin 18 kDa 120

5 7.88 Carbonic anhydrase 30 kDa 200

6 7.88 Ovalbumin 45 kDa 300

* Carbonic anhydrase impurity

Intact Protein Separation

Page 12: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

10 Thermo Scientific Accucore HPLC Columns for Biomolecule Separations

Excellent Resolution

Accucore 150-C4 provides significantly sharper and higher peaks than a column packed with 5 µm wide pore fully porous C4 – thus offering better resolution and sensitivity. The Accucore 150-C4 also performs better than a column packed with < 2 µm wide pore fully porous C4 and generates only a fraction of the backpressure.

Protein Mean tR / min % RSD Mean peak width Mean at half height / min asymmetry

Ovalbumin (45 kDa) 9.30 0.06 0.05 1.14

Reproducible Results

The Accucore 150-C4 column exhibits excellent peak shape and retention time reproducibility.

Time (min)

0

20000

40000

60000

80000

0

20000

40000

60000

80000

100000

0 2 4

*

6 8 10

*

Time (min)

0

20000

40000

60000

80000

0

20000

40000

60000

80000

100000

0 2 4

*

6 8 10

uAU

uAU

0 2 4 6 8 10 0 2 4 6 8 10

Time (min)

9.2 9.4

Accucore 150-C4 Accucore 150-C4 Backpressure: 185 bar

< 2 µm wide pore fully porous C4 Backpressure: 320 bar

5 µm wide pore fully porous C4

Data From Six Injections

Page 13: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

Thermo Scientific Accucore HPLC Columns for Biomolecule Separations 11

Proven Loading Capacity

As the load on the Accucore 150-C4 column increases up to 1200 pmoles (25 µg) peak shapes remain very sharp as shown by the peak capacity plot.

mAU

Time (min)

0

100000

200000

300000

400000

500000

600000

0 1 2 3 4 5 6 7 8 9 10

30 pmoles

60 pmoles

120 pmoles

300 pmoles

600 pmoles

1200 pmoles

Peak

Cap

acity

Total Load on Column (total pmoles)

0

20

40

60

80

100

120

140

0 600 1200200 400 800 1000

Loading Capacity

Thermo Scientific Accucore HPLC Columns for Biomolecule Separations 11

Page 14: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

12 Thermo Scientific Accucore HPLC Columns for Biomolecule Separations

nanoViper Column Separations

Base peak 50 fmol loading of BSA digest

Mobile phase A: 0.1 % formic acid in water

Mobile phase B: 90 % acetonitrile in water

Gradient: 4-40 % B over 30 mins

40-95 % B over 2 mins

95 % B hold for 2 mins

Flow: 300 nL/min

Temperature: not controlled

Column: Accucore 150-C18, 2.6 µm, 150 x 75 µm

Backpressure: 198 bar (100 % A)

Injection: Direct on-column loading of 1 µL of BSA digest, 50 fmol/µL in water + 0.1% formic acid

Detection: Thermo Scientific LTQ Orbitrap XL Mass Spectrometer coupled with a Proxeon Nano Spray Flex Ion Source

Protein separation using formic acid as an MS compatible mobile phase additive

Mobile phase A: 0.1 % formic acid in water

Mobile phase B: 0.1 % formic acid in acetonitrile

Gradient: 0-30 % B over 1 min

30-60 % B over 10 mins

60-95 % B over 1 min

95 % B hold for 3 mins

Flow: 300 nL/min

Temperature: 40 °C

Column: Accucore 150-C4, 2.6 µm, 150 x 75 µm

Backpressure: 204 bar (100 % A)

Injection: 0.25 µL 2 pmol/µL solution

Detection: UV (214 nm)

Peak Number tR /min Protein

1 6.89 Cytochrome C (12 kDa)

2 7.34 Insulin (6 kDa)

3 8.77 Myoglobin (18 kDa)

4 9.57 Carbonic anhydrase (30 kDa)

5 11.02 Ovalbumin (45 kDa)

* Carbonic anhydrase impurity

mAU

Time (min)

0

5

10

15

20

7

1

2

*

3

4

5

6 8 9 10 11 12 13 14

Rela

tive

Abun

danc

e

Time (min)

0

10

20

30

40

50

60

70

80

90

100

20 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36

2.003.89

5.74

6.426.60 6.69

7.5711.49

11.219.55

11.92

14.36

14.80

16.14

18.02

18.40

19.10

20.16

22.18

20.05

22.43

26.68

27.39

30.07

28.77

31.64 35.33

37.80

22.56 23.81

24.98

32.1036.03

Page 15: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

Thermo Scientific Accucore HPLC Columns for Biomolecule Separations 13

Ordering Information

Accucore HPLC ColumnsDescription Particle Size Length (mm) 2.1 mm ID 3.0 mm ID 4.6 mm ID

Accucore 150-C18 2.6 µm

30 16126-032130 16126-033030 16126-03463050 16126-052130 16126-053030 16126-054630100 16126-102130 16126-103030 16126-104630150 16126-152130 16126-153030 16126-154630

Accucore 150-C4 2.6 µm

30 16526-032130 16526-033030 16526-03463050 16526-052130 16526-053030 16526-054630100 16526-102130 16526-103030 16526-104630150 16526-152130 16526-153030 16526-154630

Accucore Defender Guard Columns (4/pk)Description Particle Size Length (mm) 2.1 mm ID 3.0 mm ID 4.6 mm IDAccucore 150-C18 2.6 µm 10 16126-012105 16126-013005 16126-014005

Accucore 150-C4 2.6 µm 10 16526-012105 16526-013005 16526-014005

Accucore nanoViper ColumnsDescription Particle Size Length (mm) 75 µm ID

Accucore 150-C18 2.6 µm150 16126-157569500 16126-507569

Accucore 150-C4 2.6 µm150 16526-157569500 16526-507569

UNIGUARD Direct-Connection Guard Cartridge HoldersDescription 2.1 mm ID 3.0 mm ID 4.6 mm IDUNIGUARD Drop-In Guard Cartridge Holder 852-00 852-00 850-00

Standard Replacement Tip 850-RT 850-RT 850-RT

• 150 Å pore size solid core particles for fast biomolecule separations

• Superb resolution of peptides and intact proteins at low backpressures

• Exceptionally rugged and easy to use analytical and nano scale columns

Ultimate Core Performance

to Maximize Your Investment

Page 16: Ultimate Core Performance to Maximize Your …...to Maximize Your Investment Thermo Scientific Accucore HPLC Columns for Biomolecule Separations Automated Packing Process Enhanced

For more information visit: www.thermoscientific.com/accucore

© 2012 Thermo Fisher Scientific Inc. All rights reserved. iPad is a registered trademark of Apple Inc., registered in the US and other countries. Agilent is a registered trademark of Agilent Technologies Inc. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

USA and Canada +1 800 332 3331France +33 (0)1 60 92 48 34Germany +49 (0) 2423 9431 20 or 21United Kingdom +44 (0)1928 534110Japan +81 3 5826 1615

China +86 21 68654588 +86 10 84193588+86 20 83145199 800 810 5118India +91 22 6742 9494 +91 27 1766 2352Australia 1 300 735 292 (free call domestic)New Zealand 0800 933 966 (free call domestic)All Other Enquiries +44 (0) 1928 534 050

Technical SupportNorth America +1 800 332 3331Outside North America +44 (0) 1928 534 440

TG20502_E 05/12S

Thermo Scientific Chromatography Columns and Consumables CatalogThis extensive catalog offers 600 pages of proven chromatography tools and product selection guides. Available online, with a robust search tool and optimized for your iPad®. Visit www.thermoscientific.com/catalog

Chromatography Resource CenterOur web-based resource center provides technical support, applications, technical tips and literature to help move your separations forward. Visit www.thermoscientific.com/crc

Resources for ChromatographersResources