ultrasonic flow meter dulcoflow dfmaprominentfluid.com/uploads/files 2016/manuals 2...ultrasonic...

57
Ultrasonic Flow Meter DulcoFlow ® DFMa Operating instructions Part no. 986007 BA MAZ 014 01/11 EN Please carefully read these operating instructions before use! · Do not discard! The operator shall be liable for any damage caused by installation or operating errors! Technical changes reserved.

Upload: others

Post on 29-Jan-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

  • Ultrasonic Flow MeterDulcoFlow® DFMa

    Operating instructions

    Part no. 986007 BA MAZ 014 01/11 EN

    Please carefully read these operating instructions before use! · Do not discard!The operator shall be liable for any damage caused by installation or operating errors!

    Technical changes reserved.

  • 986007, 1, en_GB

    © 2010

    ProMinent Dosiertechnik GmbH

    Im Schuhmachergewann 5-11

    69123 Heidelberg

    Germany

    Telephone: +49 6221 842-0

    Fax: +49 6221 842-617

    email: [email protected]

    Internet: www.prominent.com)

    2

  • Supplementary information

    Read the following supplementary informationin its entirety! Should you already know thisinformation, you have an even greater need ofthe Operating Instructions.

    The following are highlighted separately in thedocument:

    n Enumerated lists

    Instructions

    ð Outcome of the instructions

    - see (reference)

    Information

    This provides important informationrelating to the correct operation of thesystem or is intended to make your workeasier.

    Safety information

    Safety information is identified by pictograms -see "Safety Chapter".

    General non-discriminatory approach

    In order to make it easier to read, this docu‐ment uses the male form in grammatical struc‐tures but with an implied neutral sense. It isaimed equally at both men and women. Wekindly ask female readers for their under‐standing in this simplification of the text.

    Supplemental instructions

    3

  • Table of contents1 Identity code............................................................................................................................ 6

    2 About this product................................................................................................................... 8

    3 Safety chapter......................................................................................................................... 9

    4 Storage and transport........................................................................................................... 12

    5 Device overview.................................................................................................................... 13

    6 Functional description........................................................................................................... 14

    7 Assembly and installation...................................................................................................... 157.1 Installation.................................................................................................................... 157.2 Hydraulic Installation.................................................................................................... 157.3 Electrical Installation..................................................................................................... 17

    8 Settings ................................................................................................................................ 208.1 Operating unit............................................................................................................... 208.1.1 LCD screen............................................................................................................... 208.1.2 LEDs.......................................................................................................................... 208.1.3 Control keys.............................................................................................................. 208.2 Checking adjustable variables...................................................................................... 238.3 Operating menu overview............................................................................................ 248.4 Changing to adjustment mode..................................................................................... 258.4.1 "Mode" main menu.................................................................................................... 258.4.2 "Set-up" main menu.................................................................................................. 258.4.3 "Calibration" main menu............................................................................................ 298.4.4 "Zero set" main menu................................................................................................ 408.4.5 "Info" main menu....................................................................................................... 418.4.6 "Language" main menu............................................................................................. 41

    9 Start up.................................................................................................................................. 42

    10 Maintenance, repair and disposal......................................................................................... 4410.1 Maintenance............................................................................................................... 4510.2 Repairs....................................................................................................................... 4510.3 Disposal...................................................................................................................... 45

    11 Troubleshooting.................................................................................................................... 4611.1 DulcoFlow® error........................................................................................................ 4711.2 Pump error................................................................................................................. 4811.3 All Other Faults........................................................................................................... 49

    Table of contents

    4

  • 12 Technical data....................................................................................................................... 50

    13 Dimensions sheet.................................................................................................................. 53

    14 EC Declaration of Conformity................................................................................................ 54

    15 Decontamination declaration................................................................................................. 55

    16 Index..................................................................................................................................... 56

    Table of contents

    5

  • 1 Identity code

    Product range DulcoFlow® flow meter

    DFMa Type For pump product ranges

    05 Beta, gamma/ L: 1000 - 0413/0713; delta: 1608 - 1612

    08 Beta, gamma/ L: 0220/0420 - 0232; delta: 1020 - 0280

    Seal material

    E EPDM

    T PTFE

    V FPM

    Hydraulic connection

    1 6/4 mm

    2 8/5 mm

    3 12/9 mm

    Electrical connection

    A 2 m European

    B 2 m Swiss

    C 2 m Australian

    D 2 m USA

    Signal output

    0 No output

    1 Current output

    2 Counter output

    Version

    0 With ProMinent® Logo

    2 Without ProMinent® Logo

    Accessories

    0 No accessories

    Identity code

    6

  • Identity code

    7

  • 2 About this productThe flow meter DulcoFlow® is intended for usein measuring pulsing volume flows in the rangefrom 0.1 to 80 l/h. All parts coming into contactwith flow media are made from PVDF. Thisensures that aggressive media can also bemeasured without problem. The device isinstalled approximately 30 cm after the pump inthe metering line. Interfering influences, suchas air bubbles, are identified and forwarded tothe analysis unit as an error message. With thedelta, it can only be used with the fast meteringstroke version.

    The DulcoFlow® flow meter can not only beused for recording and measurement of volumeflows, but also for monitoring individualmetering strokes. In this case the pump is cali‐brated to the lifting volume set at the pump. Alower and upper limit can be entered, which ifexceeded or undershot, results in no feedbackto the pump. This creates an error message.The connection to the pump takes place via theinput for the "Flow Control" dosing monitor

    The device is designed for wall mounting.

    About this product

    8

  • 3 Safety chapter

    Explanation of the safety information

    The following signal words are used in theseoperating instructions to identify differentseverities of a hazard:

    Signalword

    Meaning

    WARNING

    Denotes a possibly hazardous sit‐uation. If this is disregarded, youare in a life-threatening situationand this can result in serious inju‐ries.

    CAU‐TION

    Denotes a possibly hazardous sit‐uation. If this is disregarded, itcould result in slight or minor inju‐ries or material damage.

    Warning signs denoting different types ofdanger

    The following warning signs are used in theseoperating instructions to denote different typesof danger:

    Warning signs Type of danger

    Warning – high-voltage.

    Warning – dangerzone.

    Correct and proper use

    n The device may only be used with liquidmetering chemicals.

    n The device can only be used with pulsingliquid flows with a clear zero flow.

    n The device may only be used after it hasbeen correctly installed and commissionedin accordance with the technical data andspecifications contained in the operatinginstructions.

    n Observe the general limitations with regardto viscosity limits, chemical resistance anddensity - see also ProMinent resistance list(In the product equipment catalogue or atwww.prominent.com)!

    n Any other uses or modifications are pro‐hibited.

    n The device is not suitable for measuringcontinuous liquid flows.

    n The device may not be used to measuregaseous media or solids.

    n The device may not be used with combus‐tible media without appropriate protectivemeasures.

    n The device may not be used with explo‐sive media.

    n The device may not be used with radioac‐tive media.

    n The device is not intended for exteriorapplications without use of suitable protec‐tive equipment.

    n The device should only be operated bytrained and authorised personnel, see thefollowing "Qualifications" table.

    n You are obliged to observe the informationcontained in the operating instructions atthe different phases of the system'sservice life.

    Safety chapter

    9

  • Safety information

    WARNING!

    Danger of electric shockA mains voltage may exist inside thehousing.

    – If the housing has been damaged,you must disconnect the device fromthe mains immediately. It may only bereturned to service after an authorisedrepair.

    WARNING!

    Warning of dangerous or unknown feedchemicalShould a dangerous or unknown feedchemical be used: It may escape from thehydraulic components during maintenancework.

    – Take appropriate protective measuresbefore working on the device (safetyglasses, safety gloves, ...). Observethe safety data sheet for the feedchemical.

    – Drain and flush the hydraulic partsbefore working on the device.

    CAUTION!

    Warning of feed chemical spraying aroundFeed chemical can spray out of thehydraulic components if they are manipu‐lated or opened due to pressure in thehydraulic and adjacent parts of the system.

    – Disconnect the pump of the devicehydraulic system from the mainspower supply and ensure that itcannot be switched on again by unau‐thorised persons.

    – Depressurise the system before com‐mencing any work on hydraulic parts.

    CAUTION!

    Danger of personnel injury and materialdamageThe use of untested third party parts canresult in personnel injuries and materialdamage.

    – Only fit parts to the device, whichhave been tested and recommendedby ProMinent.

    NOTICE!

    Warning of illegal operationObserve the regulations that apply wherethe device is installed.

    Information in the event of an emergency

    In an emergency, disconnect the device fromthe mains!

    If feed chemical escapes, also depressurise thedevice hydraulic system. Adhere to the safetydata sheet for the feed chemical.

    Safety chapter

    10

  • Qualification of personnel

    Activity Qualification level

    Storage, transport,unpacking

    Instructed personnel

    Installation, installa‐tion of hydraulicsystem

    Technical personnel

    Electrical installation Electrician

    Operation Instructed personnel

    Maintenance Technical personnel

    Repairs Customer service -authorised by ProMi‐nent

    Decommissioning,disposal

    Technical personnel

    Troubleshooting Technical personnel,electrician, instructedpersonnel

    Technical personnel

    A qualified employee is deemed to be a personwho is able to assess the tasks assigned to himand recognise possible hazards based on his/her technical training, knowledge and experi‐ence, as well as knowledge of pertinent regula‐tions.

    Electrician

    Electricians are deemed to be people, who areable to complete work on electrical systemsand recognize and avoid possible hazardsindependently based on their technical trainingand experience, as well as knowledge of perti‐nent standards and regulations. Electriciansshould be specifically trained for the workingenvironment in which the are employed andknow the relevant standards and regulations.Electricians must comply with the provisions ofthe applicable statutory directives on accidentprevention.

    Instructed personnel

    An instructed person is deemed to be a personwho has been instructed and, if required,trained in the tasks assigned to him/her andpossible dangers that could result fromimproper behaviour, as well as having beeninstructed in the required protective equipmentand protective measures.

    Customer Service department

    Customer Service refers to service technicians,who have received proven training and havebeen authorised by ProMinent® to work on thesystem.

    Safety chapter

    11

  • 4 Storage and transportSafety information

    WARNING!

    Only return the device for repair in acleaned state and with flushed hydraulicparts - refer to the chapter "Decommis‐sioning"!

    Only send the unit complete with a filled inDecontamination Declaration form. TheDecontamination Declaration constitutesan integral part of an inspection / repairorder. A unit can only be inspected orrepaired when a Decontamination Declara‐tion Form is submitted that has been com‐pleted correctly and in full by an authorisedand qualified person on behalf of the oper‐ator.

    The "Decontamination Declaration Form"can be found in the Appendix or underwww.prominent.com.

    NOTICE!

    Danger of material damageThe device can be damaged by incorrector improper storage or transportation!

    – The device should only be stored ortransported in a well packaged state -preferably in its original packaging.

    – The packaged unit should also onlybe stored or transported in accord‐ance with the stipulated storage con‐ditions.

    – The packaged unit should be pro‐tected from moisture and the ingressof chemicals.

    Ambient conditions

    Data Value Unit

    Minimum storage and transport temperature -10 °C

    Maximum storage and transport temperature +50 °C

    Air humidity < 95 % rel.humidity*

    * non-condensing

    Storage and transport

    12

  • 5 Device overview

    1

    24

    5

    3

    Fig. 1: Overview of equipment DulcoFlow® with arrows showing flow1 Control elements - see chapter "Settings" -

    "Operating unit"2 Mains connection3 Signal output (option)

    4 Feed chemical inlet5 Feed chemical outlet

    Device overview

    13

  • 6 Functional descriptionThe DulcoFlow ® flow meter measures thevolume flow of pulsing flows. The ultrasonic,time of flight measurement method is used. Forthe time of flight measurement, a sound signalis alternately transmitted in and against thedirection of flow. The time difference is then ameasure of the mean flow velocity. Use of theultrasound measurement method automaticallycompensates any temperature inducedchanges in the medium. Operation withoutmoving parts guarantees a long service life andwear-free operation.

    The DulcoFlow® calculates the mass flow fromthe volume flow and the density of the feedchemical.

    Additionally the DulcoFlow® can measure thepressure surges of the metering pump andhence replace a dosing monitor such as theFlow Control. A metering pump such as thegamma/ L or delta® can use these signals asacknowledge pulses for its individual strokes. Ifthe acknowledge pulses are missing or if thecapacity exceeds the specified limit values, themetering pump stops after an adjustablenumber of missing pulses and goes into faultmode - see metering pump operating instruc‐tions.

    The DulcoFlow® gives the flow reading, strokefeedback, or error messages via the variousoutput types.

    Functional description

    14

  • 7 Assembly and installationSafety information

    WARNING!

    Carry out the assembly work before theelectrical installation is undertaken.

    WARNING!

    Observe the information in the "Technicaldata" chapter.

    WARNING!

    Danger of an electric shockIf the device is used outdoors without acover or weatherproof roof, water may beable to collect on the seals and penetratethe housing or direct sunlight may causethe housing to be corroded.

    – Always use a cover or weatherproofroof when using the unit outdoors.

    CAUTION!

    Warning of illegal operationObserve the regulations that apply wherethe unit is to be installed.

    The device is resistant to normal atmos‐pheres in plant rooms.

    7.1 Installation

    Install the device so that the hose betweenit and the pump is approximately 30 cmlong.This ensures it measures accurately.

    Mount the device vertically.This ensures it measures accurately.

    Mount the device vertically on the wallusing both eyes on the housing.

    Do not forget the washers.

    7.2 Hydraulic Installation

    CAUTION!

    Warning of escaping feed chemicalFeed chemicals can escape in the eventthat the hose lines are incorrectly installed.

    – Only use original hoses with thespecified hose dimensions.

    – Avoid reducing the hose sizes.

    Assembly and installation

    15

  • Flow direction

    The flow direction through the device goesfrom the bottom to the top.

    Injection valves, back pressurevalves, relief valvesInjection valves, back pressure valves orrelief valves must not have any effect onthe measurement.

    Relief valves

    It is best to install a relief valve before theflow meter, so that the displayed flow orflow volume actually corresponds to thevalue which is being metered into thesystem if the relief valve triggers.

    Hydraulic Dampers

    Hydraulic dampers, like accumulators,inline dampers or bladder accumulators /diaphragm accumulators must not haveany effect on the measurement.A zero flow must be available.The damping must not push the flow meanvalue beneath the measurement threshold.

    Installing the hosing:

    1. Cut off the ends of the hoses (6) so thatthey are straight.

    2. Unscrew the union nut (5) and pushover the hose together with the clampring (4).

    3. Push the hose end (6) up to the stopover the nozzle (3).

    4. Tighten the union nut (5).

    5. Pull on the hose (6) and tighten up theunion nut (5).

    1

    3

    2

    4

    5

    6

    P_MAZ_0041_SW

    Fig. 2: Installing the hose line1 Connector2 Seal3 Nozzle4 Clamp ring5 Union nut6 Hose

    Assembly and installation

    16

  • P_MAZ_0040_SW

    2

    PD

    1

    3

    4

    5

    6

    7

    Fig. 3: DulcoFlow® hydraulic system1 Target use2 Pressure maintenance device3 Hydraulic Damper4 DulcoFlow®5 Overflow device6 Metering pump7 Storage tank

    7.3 Electrical Installation

    WARNING!

    Danger of electric shockThe use of a residual current circuitbreaker drastically increases survivalchances should persons come into contactwith the mains voltage due to an electricalaccident.

    – Always fit a residual current circuitbreaker on-site.

    A metering pump, PLC or other devices can beconnected to the flow meter.

    The following alternative electrical outputs areavailable via cable:

    n Current output (identity code characteristic"signal output "1"):

    n Counter output (identity code characteristic"signal output "2"):

    Assembly and installation

    17

  • Current output

    The following can be signalled via the current output (standard signal output (mA)):

    n Instantaneous flown Fault

    Technical data:

    Data Value Unit

    Current* 0/4 .. 20 mA

    Maximum load 400 Ω

    * zero volt connection

    Counter output as stroke feedback output

    The counter output as stroke feedback outputcan be used to signal stroke feedback, pro‐vided it is set accordingly - see chapter "Set‐tings".

    1. To report stroke feedback at themetering pump such as gamma/ L anddelta® , plug the cable from the flowmeter to the pump in the "dosing mon‐itor" terminal.

    ð The identifier for the dosing monitormust appear on the LCD screen ofthe pump.

    2. If it does not appear, make the neces‐sary settings at the metering pump.

    Assembly and installation

    18

  • Counter output as frequency output

    The counter output as frequency output can be used to signal flow or an error, provided it is setaccordingly - see chapter "Settings".

    1. Remove the connected socket.

    2. Remove the insulation of the leads to match the terminals of the other device.

    3. Crimp on a suitable cable end sleeve.

    ð In this format, the cable can be connected to another device.

    4. Connect the cable to both devices in accordance with the wiring diagram. DulcoFlow® wiringdiagram - see appendix.

    Technical data:

    Data Value Unit

    Voltage with open contacts < 30 V

    Current with closed contacts < 10 mA

    Residual voltage with closed contacts < 0.7 V

    Switching frequency < 10 kHz

    Lead Function

    brown Supply Voltage

    white Error signal

    blue Frequency signal

    black Mass

    Assembly and installation

    19

  • 8 Settings8.1 Operating unit

    P_MAZ_0039_SW

    V: 1145.69 mLQ: 502.13 mL/h

    i

    DulcoFlow ®1

    2

    3

    4

    Fig. 4: Operating unit DulcoFlow®

    1 LCD display2 Status LED3 Control keys4 Stroke feedback LED

    8.1.1 LCD screenThe LCD screen comprises a two-line display.

    8.1.2 LEDsThe status LEDs (left) shows the followinginformation:

    LED Information

    green Status OK

    orange Warning

    red Error

    The stroke feedback LED (right) shows the fol‐lowing information:

    LED Information

    off No pressure surge identified

    green, illumi‐nated

    Pressure surge detected -within tolerance

    red, illumi‐nated

    Pressure surge detected -outside tolerance

    8.1.3 Control keysThe control keys are designated as follows:

    Key Designation

    i [ i ]

    [DOWN]

    [UP]

    [P]

    Settings

    20

  • P

    P 3 s

    P

    P 3 scont.display

    main menu set-up

    set-up display

    = set-up option

    Fig. 5: Navigation within the operating menuThe control keys have different functions in the continuous display, in the operating menu and in themenu branches.

    In the continuous display

    Key Functions

    i Changes to continuous dis‐playChanges into the operatingmenu

    In the operating menu

    Key Functions

    Change to the previous menubranch

    Change to the next menubranch

    Open menu branch (switch tothe first item of the selectedmenu branch).

    Settings

    21

  • In a menu item of a menu branch

    Key Functions

    i Switches between the continuous changing of a number and digitby digit changingIncrease/change the set value

    Reduce/change the set value

    Accept the configured value and change to the next menu item ofthe menu branch.

    In digit by digit changing: Within a number change to the nextfigure.

    B0210

    density1.3425 g/mL

    density1.3425 g/mL

    density1.3425 g/mL

    a) b)

    Fig. 6: a) Change between continuous changing of a number and digit by digit changing; b) Changesthe position within the number

    Key [P] - additional generally-applicable func‐tions:

    Press dura‐tion

    Function

    approx. 2 s Open the operating menu

    approx. 3 s Quit the operating menuwithout saving the configuredvalues and return to the con‐tinuous display

    A modified set value is only adopted, if ithas previously been confirmed by briefpressing of the [P] key.

    Settings

    22

  • 8.2 Checking adjustable varia‐bles

    Before you adjust the flow meter, you cancheck the actual settings of the adjustable vari‐ables:

    Press the key [ i ] ("i" for "Info"), if theLCD screen is displaying a continuousdisplay - i.e. no display of the operatingmenu.

    ð After every press on the [ i ] key,you see a different continuous dis‐play.

    The number and appearance of the continuousdisplays depends on the selected measuredvariable and whether a setpoint has beenentered - see below.

    The key [ i ] is used to change betweencontinuous displays.

    Continuous display for the "Volume" measuredvariable

    Continuous display Description

    V: 243,32 mLQ: 171,05 mL/h

    Total quantity* V andinstantaneous volumeflow Q

    V: 243,32 mLN: 637

    Total quantity* V andnumber of strokes* N

    V/H: 0,382 mL%SH: 102,3 %

    Volume per stroke V/H and deviation fromthe set point of thelifting volume** %SH

    * since the last reset

    ** if a „volume (stroke)“ has been entered under„Calibration è Stroke feedback è Set-point“

    Continuous displays for the "Mass" measuredvariable

    (If „mass“ was set under „Modeè Measured variable“ .)

    Continuous display Description

    m: 326,05 gQ: 229,21 g/h

    Total quantity* m andinstantaneous massflow Q

    m: 326,05 gN: 527

    Total quantity* m andnumber of strokes* N

    m/H: 0,619 g%SH: 97,8 %

    Mass per stroke m/Hand deviation from theset point of the strokemass** %SH

    * since the last reset

    ** if a„mass (stroke)“ has been entered under„Calibration è Stroke feedback è Set-point“

    Reset values

    – To reset the total quantity andstrokes, press both [arrow keys]simultaneously.

    – In the main menu "Zero set", the totalquantity and the strokes can be set to"zero" independently of each other.

    Settings

    23

  • 8.3 Operating menu overview

    B0209

    P

    P

    P

    cont.display

    2 s

    set to zero quantity and strokes

    acknowledge error

    main menumode

    menumeasurandmeasurand

    Pmain menucalibrationcalibrationmenu

    stroke feedb.stroke feedb.

    menuquantityquantity

    Pmain menuzero setzero setmenu

    quantityquantity

    menustrokesstrokes

    Pmain menuset-upset-upmenu

    displaydisplay

    menuserviceservice

    menucounter outp.*counter outp.*

    menucurrent output*current output*

    * Dependent on identity code

    Settings

    24

  • main menuinfoinfo

    main menulanguagelanguage

    Fig. 7

    8.4 Changing to adjustmentmode

    If the [P] key is pressed for 2 seconds in a con‐tinuous display, the device changes to adjust‐ment mode.

    The following main menus can be selected inadjustment mode:

    1 - Mode

    2 - Set-up

    3 - Calibration

    4 - Zero set

    5 - Info

    6 - Language

    8.4.1 "Mode" main menumain menu mode

    The measured variables can be selected fromthe "Operation" main menu:

    n „Volume“ (-flow)n „Mass“ (-flow)The appearance of a few menus is dependenton this.

    8.4.2 "Set-up" main menumain menu set-up

    The following menus can be selected from the"set-up" main menu:

    1 - „Display“3 - „Current output“ (identity code character‐

    istic "signal output" "1")

    2 - „Counter output“ (identity code character‐istic "signal output" "2")

    4 - „Service“ (for customer service only)

    Settings

    25

  • 8.4.2.1 "Display" menu

    P P P PB0228

    set-up display

    flow unit mL/h

    quantity unit mL

    damping flow 30.00 s

    In the „display“ menu, the units can be selected for the quantities and flow.Moreover, the damping of the displayed flow values can be changed (not for quantities), if theychange too quickly / slowly in the display.

    The greater the set integration constant in the menu item „damping flow“, the greater the damping ofthe displayed flow values.

    Non-metric units and their conversion

    Unit Meaning Conversion

    1 gal 1 US liquid gallon = 3.785421 l

    1 lb(s) 1 Pound = 453.59237 g

    8.4.2.2 "Counter output" menu(identity code characteristic "signal output" "2")

    P

    P

    P

    P P

    P

    P P

    P P

    B0235

    set-up counter output

    puls/freq.open-collector

    counter output puls/freq.

    counter output stroke feedback

    open-collector NPN

    K factor 5000 1/mL

    puls/freq.signal output

    K factor unit 1/mL

    puls/freq.error output

    errorbubbles

    In the „counter output“ menu the counter output can be set either as a pulse/frequency output tooutput the instantaneous flow, as an error signal or as a feedback message.

    Settings

    26

  • puls/freq. (frequency output)

    Firstly the type of the output („Open-Collector“)can be selected so that it is suitable for the con‐nected device (NPN / PNP).

    The K factor can be set via the „signal output“menu, and is then used by the DulcoFlow® toconvert the instantaneous flow prior to outputvia the counter output to an external device.

    In the „error output“ menu a setting can bemade to determine whether the DulcoFlow®outputs an error signal via the counter output toan external device. It works like a relay instatus NO.

    stroke feedback

    For stroke feedback, e.g. to a Prominentmetering pump, as with a Flow Control® dosingmonitor, the cable must be fed from the counteroutput to the "dosing monitor" terminal of themetering pump.

    Then the „stroke feedback“ must be calibratedunder „calibration“, see chapter "Calibration".

    Settings

    27

  • 8.4.2.3 "Current output" menu(identity code characteristic "signal output" "1")

    P P P P

    P P

    P PB0234

    set-up current output

    current range 4...20 mA

    current output signal current

    flow unit mL/h

    0/4 mA value 0.68 mL/h

    20 mA value 897.65 mL/h

    current outputerror current

    error currentbubbles

    In the „current output“ menu the standard signal output (mA) can be set either to output the instanta‐neous flow or as a feedback message.

    In the „signal current“ menu a setting can bemade to determine how the DulcoFlow® outputsthe instantaneous flow via the current output toan external device.

    In the „error output“ menu a setting can bemade to determine whether the DulcoFlow®outputs an error signal via the current output toan external device (23 / 3.6 mA).

    Settings

    28

  • 8.4.2.4 "service" menu

    P PB0229

    set-up service

    service code 0000

    The "service" menu is password protected and only for customer service.

    8.4.3 "Calibration" main menumain menu calibration

    From the "calibration" menu either the flowmeasurement can be calibrated or the strokefeedback set up.

    8.4.3.1 Calibrate "stroke feedback"8.4.3.1.1 For "volume"The permitted range for the lifting volume V/Hcan be specified in this menu. If the liftingvolume moves outside of this range, e.g. due toa changed back pressure, the DulcoFlow® nolonger gives any stroke feedback to themetering pump and the right LED now illumi‐nates as a steady red instead of green. ProMi‐nent metering pumps such as the gamma/ L ordelta® go into fault mode after a series of defec‐tive strokes (pump set up).

    Settings

    29

  • About setpoint and tolerances

    P P P P

    P P PB0230

    calibration stroke feedback

    volume (stroke) 0.347 mL

    stroke feedback set-point

    volume (stroke) 0.347 mL

    lower tolerance -010.00 %

    tolerance tolerance (%)

    upper tolerance +020.00 %

    Specify the allowed range for the lifting volume V/H via the setpoint of the lifting volume V/H andtolerances in %:

    1. Follow the menu path „calibration è stroke feedback è set-point“ and press the key [P] .

    ð The menu item „volume (stroke)“ displays the currently stored setpoint.

    2. Start the metering pump.

    ð The actual measured value is displayed.

    3. Turn the stroke length adjustment knob until the desired setpoint is displayed.

    4. Press key [P] .

    ð The displayed measured value is saved as a setpoint and the menu item „tolerancetolerances“ appears.

    The setpoint is valid as 100%.

    5. Stop the metering pump.

    6. Press key [P] .7. Set the „lower tolerance“ with the [arrow keys] and press the [P] .8. Set the „upper tolerance“ with the [arrow keys] and press the [P] .

    ð The continuous display appears again.

    The desired setpoint can also be entered, without having to use the stroke adjustment dial orthe pump having to be running, directly under „volume (stroke)“ using the [arrow keys] .

    Settings

    30

  • Upper limit values

    B0231

    calibration stroke feedback

    volume (stroke) 0.347 mL

    stroke feedback set-pointP P P P

    volume (stroke) 0.347 mL

    tolerance tolerance (%) P

    lower limit 0.312 mL

    tolerance limits P P P

    upper limit 0.416 mL

    Alternatively, the desired limits for the allowed range of the lifting volume can be entered under„lower limit“ and „upper limit“:1. Follow the menu path „calibration è stroke feedback è set-point“ and press the key [P] .

    ð The menu item „volume (stroke)“ displays the currently stored setpoint.

    2. Start the metering pump.

    ð The actual measured value is displayed.

    3. Turn the stroke length adjustment knob until the desired setpoint is displayed.

    4. Press key [P] .

    ð The displayed measured value is saved as a setpoint and the menu item „tolerancetolerances“ appears.

    The setpoint is valid as 100%.

    5. Using the [arrow keys] change to the menu item „limits“ and press key [P] .6. Turn the stroke length adjustment dial down until the desired lower limit is reached and press

    key [P] .7. Proceed analogously for the upper limit.

    ð The continuous display appears again.

    8. Stop the metering pump.

    Settings

    31

  • The desired setpoint can also be entered, without having to use the stroke adjustment dial orthe pump having to be running, directly under „volume (stroke)“ using the [arrow keys] .

    8.4.3.1.2 For "mass"The permitted range for the mass per stroke m/H can be specified in this menu. If the mass perstroke moves outside of this range, e.g. due toa changed back pressure, the DulcoFlow® nolonger gives any stroke feedback to themetering pump and the right LED now illumi‐nates as a steady red instead of green. ProMi‐nent metering pumps such as the gamma/ L ordelta® go into fault mode after a series of defec‐tive strokes (pump set up).

    Settings

    32

  • About setpoint and tolerances

    B0232

    P P P P

    P P P

    calibration stroke feedback

    mass (stroke) 0.347 g

    stroke feedback set-point

    mass (stroke) 0.347 g

    lower tolerance -010.00 %

    tolerance tolerance (%)

    upper tolerance +020.00 %

    Specify the allowed range for the mass per stroke m/H via the setpoint of the lifting volume m/H andtolerances in %:

    1. Follow the menu path „calibration è stroke feedback è set-point“ and press the key [P] .

    ð The menu item „volume (stroke)“ displays the currently stored setpoint.

    2. Start the metering pump.

    ð The actual measured value is displayed.

    3. Turn the stroke length adjustment knob until the desired setpoint is displayed.

    4. Press key [P] .

    ð The displayed measured value is saved as a setpoint and the menu item „tolerancetolerances“ appears.

    The setpoint is valid as 100%.

    5. Press key [P] .6. Set the „lower tolerance“ with the [arrow keys] and press the [P] .7. Set the „upper tolerance“ with the [arrow keys] and press the [P] .

    ð The continuous display appears again.

    8. Stop the metering pump.

    The desired setpoint can also be entered, without having to use the stroke adjustment dial orthe pump having to be running, directly under „mass (stroke)“ using the [arrow keys] .

    Settings

    33

  • Upper limit values

    B0233

    P P P P

    P

    P P P

    calibration stroke feedback

    mass (stroke) 0.347 g

    stroke feedback set-point

    mass (stroke) 0.347 g

    tolerance tolerance (%)

    lower limit 0.312 g

    tolerance limits

    upper limit 0.416 g

    Alternatively, the desired limits for the allowed range of the stroke mass can be entered under „lowerlimit“ and „upper limit“:1. Follow the menu path „calibration è stroke feedback è set-point“ and press the key [P] .

    ð The menu item „volume (stroke)“ displays the currently stored setpoint.

    2. Start the metering pump.

    ð The actual measured value is displayed.

    3. Turn the stroke length adjustment knob until the desired setpoint is displayed.

    4. Press key [P] .

    ð The displayed measured value is saved as a setpoint and the menu item „tolerancetolerances“ appears.

    The setpoint is valid as 100%.

    5. Using the [arrow keys] change to the menu item „limits“ and press key [P] .6. Turn the stroke length adjustment dial down until the desired lower limit is reached and press

    key [P] .7. Analogously, proceed in the same way for the upper limit.

    ð The continuous display appears again.

    8. Stop the metering pump.

    Settings

    34

  • The desired setpoint can also be entered, without having to use the stroke adjustment dial orthe pump having to be running, directly under „mass (stroke)“ using the [arrow keys] .

    8.4.3.2 Calibrate "quantity"

    Only calibrate the quantity flow if the dis‐played values do not attain the expectedaccuracy.

    Settings

    35

  • 8.4.3.2.1 By corr. factor

    B0244

    P P Pcalibration quantitycorr. factor 101.23 %

    quantity corr. factor

    calibration stroke feedback

    If the new correction factor is know in %, it can be entered directly here.

    It is obtained by dividing a value you have measured yourself by the displayed value and multiplyingthe result by 100.

    1. Follow the menu path „calibration è stroke feedback“.2. Using the [arrow keys] change to the menu item „calibration quantity“ and press key [P] 2x.3. Using the [arrow keys] Enter the „corr. factor“ and press the key [P] .

    ð The continuous display appears again.

    8.4.3.2.2 By meas. Value

    CAUTION!

    Danger with dangerous feed chemicalsProvided the following handling instruc‐tions are followed, contact with the feedchemical is possible.

    – If the feed chemical is dangerous,take appropriate safety precautionswhen carrying out the following han‐dling instructions.

    – Observe the feed chemical safetydata sheet.

    Depending on the set measured variable, a cal‐ibration menu appears for:

    n Volumen Mass

    Settings

    36

  • Volume

    B0226

    P P

    P P P

    P

    calibration quantity

    quantity corr. factor

    actual volume 54.72 mL

    quantity meas. Value

    corr. factor 95.27 %

    nominal volume 52.13 mL

    calibration stroke feedback

    Settings

    37

  • Requirements:

    n 1 measuring cylinder which can be read sufficiently accuratelyn The metering pump suction line is fed, bubble-free into the measuring cylinder.

    1. Record the fluid level in the measuring cylinder.

    2. Change from the menu item „calibration stroke feedback“ using the [arrow keys] to the menuitem „calibration quantity“ and press key [P].

    ð The menu item „quantity corr. factor“ appears.

    3. Change from the menu item „quantity corr. factor“ using the [arrow keys] to the menu item„quantity meas. Value“ and press key [P].

    ð The menu item „actual volume“ appears.

    4. Start the metering pump.

    Select the number of strokes so that the reading error at the measuring cylinder (half ofthe smallest readable volume graduation divided by the metered total volume) is smallerthan the Dulcoflow measurement error.

    5. Stop the metering pump.

    6. Record the fluid level in the measuring cylinder and calculate the difference.

    7. Press key [P] .

    ð The menu item „nominal volume“ appears.

    8. Adjust the value in the menu item „nominal volume“ using the [arrow keys] based on this dif‐ference and press the [P] key.

    ð The menu item „corr. factor“ appears. It shows the calculated correction factor.

    9. Press key [P] , to return to the continuous display.

    It is also possible to manually calculate the correction factor and enter it directly under „entry“ - „corr.factor“.

    Settings

    38

  • Mass

    B0227

    P P

    P P P

    P

    calibration quantity

    quantity corr. factor

    actual mass 54.72 g

    quantity meas. Value

    corr. factor 95.27 %

    nominal mass 52.13 g

    calibration stroke feedback

    Settings

    39

  • Requirements:

    n 1 weighing instrument which can be read sufficiently accuratelyn 1 vessel with feed chemicaln the suction line is fed, bubble-free into the measuring vessel.

    1. Zero the weighing instrument.

    2. Change from the menu item „calibration stroke feedback“ using the [arrow keys] to the menuitem „calibration quantity“ and press key [P].

    ð The menu item „quantity corr. factor“ appears.

    3. Change from the menu item „quantity corr. factor“ using the [arrow keys] to the menu item„quantity meas. Value“ and press key [P].

    ð The menu item „actual mass“ appears.

    4. Start the metering pump.

    Select the number of strokes so that the reading error at the weighing instrument (half ofthe smallest readable mass graduation divided by the metered total mass) is smallerthan the Dulcoflow measurement error.

    5. Stop the metering pump.

    6. Read off the weight from the weighing instrument.

    7. Press key [P] .

    ð The menu item „nominal mass“ appears.

    8. Adjust the value in the menu item „nominal mass“ using the [arrow keys] to the read-offweight and press the key [P] .

    ð The menu item „corr. factor“ appears. It shows the calculated correction factor.

    9. Press key [P] , to return to the continuous display.

    It is also possible to manually calculate the correction factor and enter it directly under „entry“ -„correction factor“.

    8.4.4 "Zero set" main menumain menu zero set

    In the main menu "Zero set", the total quantityand the strokes, which are displayed in the con‐tinuous display, can be set to "zero" independ‐ently of each other.

    Settings

    40

  • 8.4.5 "Info" main menumain menu info

    This information can be read-off in the "info"main menu:

    Code Meaning

    ID Identity code

    SN Serial number

    HW Hardware version

    SW Software version

    BL Bootloader

    8.4.6 "Language" main menumain menu language

    The operating language can be selected fromthe "language" main menu.

    Settings

    41

  • 9 Start up1. Connect the device hydraulically to the

    installation.

    2. Connect the device to the supplyvoltage.

    3. If necessary set:

    n the language - see Ä Chapter 8.4.6„"Language" mainmenu“ on page 41

    n the measured variable to "mass"and density of the feed chemical -see Ä on page 25

    n the units - see Ä Chapter 8.4.2.1„"Display" menu“ on page 26

    n the current output - see Ä Chapter8.4.2.3 „"Current output"menu“ on page 28

    n the counter output - see Ä Chapter8.4.2.2 „"Counter output"menu“ on page 26

    4. Allow the metering pump to prime -press both [arrow keys] simultaneouslyat the pump control unit.

    5. At the device acknowledge the error„bubbles“ by pressing the key [P] .

    6. Allow the metering pump to run.

    7. Check whether the displayed values areplausible.

    8. Check whether the stroke feedback tothe metering pump is plausible.

    9. Check whether the frequency signalsand the mA signals of the device havethe expected effect if these signals arebeing used.

    If the metering pump goes into fault modeduring start up, press the [P] key toacknowledge.

    Use with the metering pump delta® is onlypossible if the pump is set to "dosing" -"set-up" (discharge stroke)" - "fast".

    Operation as a flow meter

    1. In the menu „calibration“ calibrate the„quantity“ - see Ä Chapter 8.4.3.2 „Cali‐brate "quantity"“ on page 35

    2. Check whether the displayed values areplausible.

    Operation as a dosing monitor

    1. Under „set-up è set counter output“ to„stroke feedback“.

    2. Connect the device cable with the"dosing monitor" terminal of the pump.

    3. In the menu „calibration“ set the „strokefeedback“ - see Ä Chapter 8.4.3.1 „Cali‐brate "stroke feedback"“ on page 29

    Function "transmit error signal"

    (for identity code characteristic "signal output""1")

    1. Under „set-up è current outputè error current“ set the desired errorresult and set the „error current“.

    2. Cause the error to occur and checkwhether everything functions as desired.

    Start up

    42

  • (for identity code characteristic "signal output""2")

    1. Under „set-up è counter outputè puls/freq. è error output“ set thedesired error result.

    2. Cause the error to occur and checkwhether everything functions as desired.

    Start up

    43

  • 10 Maintenance, repair and disposal

    WARNING!

    Danger from chemical residuesThere is normally chemical residue in themeasurement pipe and housing after ope‐ration. This chemical residue could be haz‐ardous to people.

    – It is mandatory that the safety infor‐mation relating to the "Storage, trans‐port and unpacking" chapter is readbefore shipping or transporting theunit.

    – Thoroughly clean the measurementpipe and the housing to removechemicals and dirt. Adhere to thesafety data sheet for the feed chem‐ical.

    Maintenance, repair and disposal

    44

  • 10.1 Maintenance

    Interval Maintenance work

    Regularly Check whether the correct flow value is displayed.

    If this is not the case, recalibrate the flow - see chapter "calibrate".

    If stroke feedback is used: Adjust the stroke length with the metering pump run‐ning and set just above the programmed upper limit - the flow identifier at thepump should no longer flash.

    Adjust the stroke length with the metering pump running and set just below theprogrammed lower limit - the flow identifier at the pump should no longer flash.

    If this is not the case, check for the cause and as necessary readjust the Dulco‐Flow® - see Ä Chapter 8.4.3.1 „Calibrate "stroke feedback"“ on page 29

    If stroke feedback is used: Check whether the stroke feedback LED (right-handside of the device) illuminates in time with the strokes.

    If this is not the case, check for the cause and rectify as necessary.

    Check whether feed chemical is coming out.

    If necessary, carefully wipe the device with a soft cloth and soapy water.

    10.2 RepairsOnly ProMinent of customer service authorisedby ProMinent may repair the DulcoFlow® flowmeter.

    10.3 Disposal

    CAUTION!

    Environmental hazard due to electronicwasteThere are electronic components in thedevice, which can have a toxic effect onthe environment.

    – Separate the electronic componentsfrom the remaining parts.

    – Observe the current applicable regu‐lations in your country.

    Maintenance, repair and disposal

    45

  • 11 Troubleshooting

    WARNING!

    Warning of dangerous or unknown feedchemicalShould a dangerous or unknown feedchemical be used: It may escape from thehydraulic components during maintenancework.

    – Before maintenance work, takeappropriate protective measures(safety glasses, safety gloves, ...).Observe the safety data sheet for thefeed chemical.

    – Drain and flush the liquid end of themetering pump before working on it.

    Troubleshooting

    46

  • 11.1 DulcoFlow® errorFaults with error messages

    The left device LED lights up red if an error exists.

    Fault description Cause Remedy

    Bubbles detected There are too many bubbles or particlesin the feed chemical.

    Avoid bubbles or particles in thefeed chemical.

    Warning with error message

    The left device LED lights up orange if awarning exists.

    LED signals

    For further information about LED signals,see the "Settings" chapter.

    Troubleshooting

    47

  • 11.2 Pump error

    To return the pump to operating status after one of the following errors, press the key [P] .

    In the event of an error an LED lights up red and the flow identifier flashes.

    Fault description Cause Remedy

    The pump stopsduring priming.

    Due to air in the liquid end,the DulcoFlow® has notoutput an acknowledgepulse.

    During priming pull the cable to the DulcoFlow®out - the function "Flow" is disabled which thecable is out.

    The pump stopduring Dulco‐Flow® set-up.

    The DulcoFlow® has givenout too few sequentialacknowledge pulses.

    Press key [P] .

    The pump stopswhile running.

    There is air in the liquid end,gaseous feed chemical.

    n Pull out the cable to the DulcoFlow® fromthe pump.

    n Bleed the liquid end.n Plug the cable to the DulcoFlow® into the

    socket on the pump.n Increase the number of acknowledge

    pulses in the pump's menu.

    Troubleshooting

    48

  • Fault description Cause Remedy

    There is air in the liquid end,the chemical feed containeris empty.

    n Fill the metering tank.n Pull out the cable to the DulcoFlow® from

    the pump.n Bleed the liquid end.n Plug the cable to the DulcoFlow® into the

    socket on the pump.

    There is gas in the dosinghead - leaks in the pathbetween the chemical feedcontainer and the Dulco‐Flow®.

    n Repair the leak.n Pull out the cable to the DulcoFlow® from

    the pump.n Bleed the liquid end.n Plug the cable to the DulcoFlow® into the

    socket on the pump.

    Blockage between Dulco‐Flow® and metering tank

    n Clear the blockage.n Pull out the cable to the DulcoFlow® from

    the pump.n Bleed the liquid end.n Plug the cable to the DulcoFlow® into the

    socket on the pump.

    The stroke adjustment dialis mis-set

    - see metering pump operating instructions

    The feed chemical is tooviscous

    - see metering pump operating instructions

    11.3 All Other FaultsAll other faults:

    Inform your customer service department oryour ProMinent branch.

    Troubleshooting

    49

  • 12 Technical dataPerformance data

    Data Value Unit

    Measuring range, pulsing, type 05: 0.1... 13 L/h

    Measuring range, pulsing, type 08: 0.6... 80 L/h

    Smallest measurable lifting volume, pulsing approx. 0.03 mL/stroke

    Accuracy over at least 100 strokes: ± 2 %*

    * relative to the measured value

    Electrical data

    Data Value Unit

    Stroke feedback, output**: 1 Contact/stroke

    Frequency output**: < 10 kHz

    Current output, max. load: 400 Ω

    Protective system: IP 65

    Supply voltage: 100...230 V AC

    Mains supply frequency: 50/60 Hz

    ** open collector

    Media requirements

    Material compatibility with: PVDF, sealing material

    Data Value Unit

    Medium pressure: 0...16 bar

    Medium temperature: -10...+45 °C

    Dynamic viscosity (η): 0.5...2000 mPas

    Sound pressure level: 1000 ... 2500 m/s

    Technical data

    50

  • Ambient conditions

    Data Value Unit

    Minimum storage and transport temperature -10 °C

    Maximum storage and transport temperature +50 °C

    Ambient temperature in operation, min. -10 °C

    Ambient temperature in operation, max. +50 °C

    Maximum air humidity * 95 % rel. humidity

    * non-condensing

    Material

    Component Material

    Measurement pipe PVDF

    Seals, hydraulic - see identity code

    Housing PPE+GF20

    Screws, etc. A2

    Electronics Electronic components

    Dimensions

    Data Value Unit

    Hose connection - nominal widths: 6/4, 8/5, 12/9 mm

    Dimensions (HxBxD): 183.6 x 121 x 122.7 mm

    Technical data

    51

  • Compatibility

    Type Pumps

    05 Beta, gala: 1000 - 0413/0713

    delta: 1608 - 1612

    08 Beta, gala: 0220/0420 - 0232

    delta: 1020 - 0280

    Technical data

    52

  • 13 Dimensions sheet

    107

    183.6

    35.5121.6

    160.5

    80.4

    18

    5.4

    10

    121

    91.9

    M20x1,5

    Fig. 8: Dimensional drawing DulcoFlow® - dimensions sheet in mm

    Dimensions sheet

    53

  • 14 EC Declaration of Conformity

    - Original -

    EC Declaration of Conformity

    We, ProMinent Dosiertechnik GmbH

    Im Schuhmachergewann 5 - 11 DE - 69123 Heidelberg

    hereby declare that, the product specified in the following complies with the relevant basic health and safety

    rules of the EC Directive, on the basis of its functional concept and design and in the version marketed by us.

    This declaration loses its validity in the event of a modification to the product not agreed with us.

    Description of the product: DulcoFlow ultrasonic flow meter

    Product type: DFMa...

    Serial no.: Please refer to type plate on the device

    Relevant EC

    Directives:

    EC Low Voltage Directive (2006/95/EC)

    EC EMC Directive (2004/108/EC)

    Harmonised standards applied, in

    particular:

    EN 60335-1, EN 61010-1 EN 55011, EN 61000-6-3, EN 61326-1

    Other applicable national

    standards and specifications:

    EN 60529, EN 61000-4-2/3/4/5/6/11

    Technical documents have been

    compiled by documentation

    specialists:

    Norbert Berger

    Im Schuhmachergewann 5-11

    DE-69123 Heidelberg

    Date / manufacturer's signature:

    Details of the signatory:

    09.12.2010

    Dr. Johannes Hartfiel, Assistance Development Director

    EC Declaration of Conformity

    54

  • 15 Decontamination declaration

    Decontamination declaration

    55

  • 16 IndexAAbout this product.......................................... 8BBL................................................................. 41bubbles......................................................... 28CCalibration.................................................... 29Checking...................................................... 23Checking adjustable variables..................... 23Control keys................................................. 20corr. factor.................................................... 36Correct and proper use.................................. 9Correction factor............................... 36, 37, 39Counter output....................................... 18, 26Current output........................................ 18, 28Ddamping....................................................... 26Display......................................................... 26Disposal....................................................... 44dosing monitor.............................................. 27Dosing monitor....................................... 18, 42EEmergency................................................... 10error output............................................. 27, 28Error signal................................................... 42Explanation of the safety information ............ 9FFault............................................................. 18Flow.............................................................. 18Flow meter................................................... 42frequency output.......................................... 27Frequency output......................................... 19Functional description.................................. 14

    HHW............................................................... 41Hydraulic system.......................................... 17IID.................................................................. 41Identity code................................................... 6Info............................................................... 41Information in the event of an emergency.... 10Installation.................................................... 15LLanguage..................................................... 41LCD screen.................................................. 20Leads........................................................... 19LEDs............................................................ 20Limit values............................................ 31, 34MMain menu................................................... 25Maintenance................................................. 44Mass................................................. 25, 32, 39Mass flow..................................................... 25meas. Value................................................. 36Mode............................................................ 25OOperating unit............................................... 20Ppuls/freq....................................................... 27QQualification of personnel............................. 11RRepair........................................................... 44SSafety chapter................................................ 9Service......................................................... 29Setpoint.................................................. 30, 33Set-up........................................................... 25

    Index

    56

  • SN................................................................ 41Standard signal output (mA)........................ 18Start up......................................................... 42Storage......................................................... 12stroke feedback...................................... 27, 29Stroke feedback........................................... 18Supplementary information............................ 3SW............................................................... 41TTechnical data.............................................. 50Tolerances............................................. 30, 33

    Transport...................................................... 12Troubleshooting........................................... 46UUnits............................................................. 26VVolume....................................... 25, 29, 35, 37Volume flow.................................................. 25WWarning sign.................................................. 9ZZero set........................................................ 40

    Index

    57

    Table of contents1 Identity code2 About this product3 Safety chapter4 Storage and transport5 Device overview6 Functional description7 Assembly and installation7.1 Installation7.2 Hydraulic Installation7.3 Electrical Installation

    8 Settings8.1 Operating unit8.1.1 LCD screen8.1.2 LEDs8.1.3 Control keys

    8.2 Checking adjustable variables8.3 Operating menu overview8.4 Changing to adjustment mode8.4.1 "Mode" main menu8.4.2 "Set-up" main menu8.4.2.1 "Display" menu8.4.2.2 "Counter output" menu8.4.2.3 "Current output" menu8.4.2.4 "service" menu

    8.4.3 "Calibration" main menu8.4.3.1 Calibrate "stroke feedback"8.4.3.1.1 For "volume"8.4.3.1.2 For "mass"

    8.4.3.2 Calibrate "quantity"8.4.3.2.1 By corr. factor8.4.3.2.2 By meas. ValueVolumeMass

    8.4.4 "Zero set" main menu8.4.5 "Info" main menu8.4.6 "Language" main menu

    9 Start up10 Maintenance, repair and disposal10.1 Maintenance10.2 Repairs10.3 Disposal

    11 Troubleshooting11.1 DulcoFlow® error11.2 Pump error11.3 All Other Faults

    12 Technical data13 Dimensions sheet14 EC Declaration of Conformity15 Decontamination declaration16 Index