Understanding ISO StandardsISO 14644-2:2015 Cleanroom Monitoring

Daniele PandolfiGlobal Product Line Manager, AerosolLife Sciences Division

Page 2: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

Understanding ISO Standard – ISO 14644-2:2015

• Introduction

• Monitoring Plan Alternatives

• Monitoring Set-up

• Risk Assessment

• Alert and Action Limits

• How to chose the right strategy

• Learn from FDA 483 Warning Letters

Page 3: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Introduction

Page 4: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

Introduction

• The recent revision of ISO 14644-1 and -2 has introduced several changes for cleanroom classification and monitoring guidelines.

• The ISO community voted in favor of this revision on October 29th, 2015.

• This presentation discusses those changes and how they affect you.

• Questions will be addressed at the end, but feel free to voice them as they come to mind.

Page 5: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

What Is ISO 14644?

ISO 14644 consists of the following parts, under the general title Cleanrooms and associated controlled environments:

• Part 1: Classification of air cleanliness by particle concentration

• Part 2: Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration

• Part 3: Test methods

• Part 4: Design, construction and start-up

• Part 5: Operations

• Part 7: Separative devices (clean air hoods, gloveboxes, isolators and mini-environments)

• Part 8: Classification of air cleanliness by chemical concentration (ACC)

• Part 9: Classification of surface cleanliness by particle concentration

• Part 10: Classification of surface cleanliness by chemical concentration

Page 6: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

What Is ISO 14644?

ISO 14644 represents the most used standards in Pharma and Electronics controlled environments.

• ISO 14644-2 specifies the requirements of a monitoring plan based on risk assessment of the intended use. The data obtained provides evidence of cleanroom or clean zone performance related to air cleanliness by particle concentration.

• This revision of ISO 14644-2 emphasizes the need to consider a monitoring strategy in addition to the initial or periodic execution of the classification of a cleanroom or clean zone in accordance with ISO 14644-1:2015.

Page 7: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

New 14644-2:2015 Main Goals

Emphasize the needs and advantages of a planned cleanroom contamination monitoring

Provides the method for a correct particles contamination alarm and warning limits setting, based on a careful evaluation of data trends

Define the differences between a simple periodic cleanroom control and a more intensive/complex monitoring strategy

Enhance the installation and process knowledge, as to improve the risk assessment evaluation and a faster reaction to any unexpected cleanroom performance

Over the last five years, the ISO Technical Committee 209

has been working on the revision of the basic airborne

cleanliness classification, 14644-1 and -2.

Page 8: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

New 14644-2:2015 Main Goals

ISO 14644-2:2015 shall not be only considered as a new standard with which to be compliant, but mainly as beneficial tool to use in achieving mature cleanroom environmental control.

Last but not least, the goal dictated by the TC 209 is:

Concretely reduce the operation cost by preventing production loss

Page 9: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Monitoring Plan Alternatives

Page 10: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 - Terminology

MONITORING

• Monitoring is an observation of the process made in accordance with a specific method, able to provide clear evidence of cleanroom performance.

• Monitoring can be “continuous”, “sequential” or “periodic”.

Page 11: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – Monitoring Plan

SEQUENTIAL

• Performed using sequential multiplexing systems.

• Unaccepted methods for Pharmaceutical Industry

• High risk of particle loss in long tubing while measuring particles greater than 1µm

Location1

Location2

Location3Location4

Location5

Location “n”…

Page 12: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – Monitoring Plan

CONTINUOUS

• Uses multiple particle counters, one per individual location

• Continuous flow of data over time

• Immediate evaluation of unexpected contamination events

Location A

Page 13: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – Monitoring Plan

PERIODIC

• Scheduled particle monitoring frequency (i.e. once per week)

• ISO 14644-2:2015 requires the test frequency to be defined and clearly specified

Week 1

Week 2

Week 3

Week «n»

Page 14: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

Monitoring Set-Up

Page 15: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – Sample Point

SAMPLE POINT LOCATION

• Understand the contamination sources and their impact on the activity in the cleanroom

• Locate particle counter probes as close as possible to critical zones

• Maximum allowed distance

• 1 foot = 33cm

Critical Point

Page 16: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – Instrument Selection

INSTRUMENT SELECTION

• Airborne particle collection efficiency, suitability to monitor the selected particle size(s), and accessibility for maintenance, calibration, and repair

• Potential adverse impact of the sampling system on the process

• Air sample flow rate and volume

• 1 cfm / 28.3 lpm commonly used

Page 17: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – Instrument Selection – cont.

ISO REQUIREMENT

• ISO 14644-2:2015 requires that all instruments employed in the cleanroom classification or monitoring activities fully comply with ISO 21501-4.

• ISO 21501-4 states the calibration and performance requirement for LSAPC (Light Scattering Airborne Particle Counters)

Page 18: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – Remote Sampling

HOW LONG CAN MY TUBING BE?

• The use of long sample transport tubes must be avoided, when intending to evaluate particle concentrations at sizes greater than 1 μm.

• ISO requires the adherence to maximum tubing lengths as specified by the particle counter manufacturer, which is typically between 1.5 and 2 meters.

Page 19: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Choosing the Right Monitoring Plan

Page 20: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – Choosing the right monitoring plan

• ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and the recommended monitoring method.

• It is up to the user to choose the most appropriate one based on their specific manufacturing requirements and risk assessment.

Page 21: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – Choosing the right monitoring plan

• Life Science industries may consider other standards to correctly and reasonably set-up their monitoring method and frequency.

• World Health Organization(WHO) issued a document titled Environmental Monitoring of Clean Rooms in Vaccine Manufacturing Facilities in November 2012 which includes instruction on determining the best monitoring frequency approach based on cleanliness class.

Page 22: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – Choosing the right monitoring plan

Page 23: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

The Risk Assessment

Page 24: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – The Risk Assessment

Risk Assessment

• The appropriate creation of risk assessment documentation is an essential requirement for implementing a compliant monitoring plan.

Goals

• Correctly understand:

• the process

• the critical areas/locations

• possible sources of contamination

• element/event that may compromise the cleanroom performance, product quality and, consequently, negatively affect the operation cost

Page 25: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – The Risk Assessment

ICH Q9 – QUALITY RISK ASSESSMENT

• The 2005 International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use is one of the best guidelines available for proper risk assessment development, review and application.

Page 26: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – The Risk Assessment

PDA - PARENTAL DRUG ASSOCIATION

• Fundamentals of an Environmental Monitoring Program, Technical Report No.13 is also a useful document when approaching monitoring plan development, and addresses the needs of a meaningful, manageable and defensible monitoring program.

Page 27: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – The Risk Assessment

RISK ASSESTMENT ALTERNATIVE TOOLS

• Other useful tools to be considered for reliable risk assessment development are:

• HACCP,

• FMEA/FMECA,

• PHA, FTA and HAZOP

Page 28: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 – The Risk Assessment

A responsible understanding of the production process and installation performance aids in the:

• prevention of unexpected out of specification conditions

• Support of power saving activities

Page 29: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Alert and Action LimitsHow to choose the right strategy

Page 30: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 - Alert and Action LimitsHow to choose the right strategy

Action Level

• User set level at which, when exceeded, will require immediate intervention, root cause investigation and corrective actions

Alert Level

• User set level, defined to provide early warning of a drift from normal conditions. This level should be used to prevent action level conditions.

Page 31: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 - Alert and Action LimitsHow to choose the right strategy

• The right determination of action and alert limits is extremely important and must be supported by the risk assessment and a consistent quantity of historical data.

Page 32: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 - Alert and Action LimitsHow to choose the right strategy

• ISO 14644-2:2015 states the importance of a long term evaluation as well as a yearly assessment of limits, method and frequency.

• While not necessarily requiring a change, the assessment is an important exercise in the critical evaluation of a monitoring plan.

• Cleanroom users should frequently question whether their existing monitoring plan is still applicable and consistent with the cleanroom’s actual performance and activities.

Page 33: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 - Alert and Action LimitsHow to choose the right strategy

The standard provides some important recommendations, as well as an applicable strategy to keep in mind when setting alert and action limits. One with high significance is provided in the paragraph B.3.1.2, quoted below:

• B.3.1.3 When setting alert and action levels, it is important to be sensitive to the high variability of airborne particle concentrations with time and at different locations. In particular, special care shall be taken when considering alert and action levels for cleanliness classes ISO Class 5 and cleaner with low concentrations of particles. In these circumstances, the occurrence of “nuisance alarms” due to false counts and/or natural variability of particle concentration is more likely and should

be avoided by careful selection of alert and action levels. Frequent “nuisance” alarms should be avoided as they can lead to alarms being ignored by users.

Page 34: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 - Alert and Action Limits, How to choose the right strategy

STRATEGY 1

• Establish a trigger threshold value based on a series of consecutively high readings.

• For example: 3 consecutive, 1 minute readings all above a specified level

0

2

4

6

8

10

12

14

16

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

PA

RTI

CLE

S

MINUTES

Monitoring Data Particles

Alarm Level

Alarm Event

Page 35: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

0

2

4

6

8

10

12

14

16

18

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30

PA

RTI

CLE

S

MINUTES

Monitoring Data Particles

Alarm Level

ISO 14644-2:2015 - Alert and Action Limits, How to choose the right strategy

STRATEGY 2

• Establish a trigger threshold value based on a high frequency of elevated readings. This method is commonly referred as “x out of y”, where “x” is the number of events and “y” is the number of minutes.

• For example, 3 out of the last 10 readings/minutes are above the specified alarm threshold

Page 36: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Learn from FDA 483 Warning Letters

Page 37: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 - Learn from FDA 483 Warning Letters

• Being compliant with any standard requires experience, knowledge and a critical approach, enabling the harmonization of regulation requirements with specific production environments.

• Jumping on this task can be done well in advance by studying the observations of FDA inspectors made to users in similar situations.

• Several 483 warning letters have been redacted over the last years, many of them strictly linked with monitoring plans' lack of compliance. Some of them are summarized in Scott Sutton's The Environmental Monitoring Program in a GMP Environment.

Page 38: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 - Learn from FDA 483 Warning Letters

483 letter extract

• …Regarding the increased non-routine surveillance monitoring performed to further evaluate the Building 123456 manufacturing facility, there was no plan in place specifying the locations to be tested, method of sampling, and actions to be taken when microbial contamination was noted…

Comments

• This warning letter states the need of having a monitoring plan in place, a description of the location to be tested and a specific sampling method. Monitoring results are considered insufficient if they don't support and link to a clear and approved plan.

Page 39: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 - Learn from FDA 483 Warning Letters

483 letter extract

• …the alert and action limits established for the manufacturing areas are not based on historical data taken from the EM Program…

Comments

• This warning letter, dated 2001, requires the cleanroom user to proactively and critically review the sampling historical data, as it must be referenced to correctly set up the appropriate alert and action levels.

• This requirement was likely difficult to accomplish in 2001, but is now easily achievable using the appropriate software platform (CFR 21 part 11 compliant), capable with a particle counter's data storage.

Page 40: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

ISO 14644-2:2015 - Conclusion

• ISO 14644-2:2015 is not only a new standard to be compliant with, but is also a beneficial tool to use in achieving mature cleanroom environmental control.

• The standard's main goal is to cultivate and promote a strong knowledge of cleanroom performance inside every company.

• ISO 14644-2:2015 was published on December 15th, 2015. All users who want to be compliant with this standard are required to take necessary action immediately.

Page 41: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

Question & Answer session

Page 42: Understanding ISO Standards€¦ · ISO 14644-2:2015 –Choosing the right monitoring plan •ISO 14644-2:2015 does not specifically provide a link between the cleanliness class and

Confidential and proprietary

Thank you for attending!