Understanding Light Curing Improved Clinical Success Howard E. Strassler, DMD, FADM, FAGD Professor, Division of Operative Dentistry University of Maryland School of Dentistry [email protected]

Upload: others

Post on 31-Aug-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Understanding Light Curing Improved Clinical Success

Howard E. Strassler, DMD, FADM, FAGD

Professor, Division of Operative Dentistry

University of Maryland School of Dentistry

[email protected]

Page 2: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Special thank you to

Richard Price, DDS, MSc, PhD

Professor, Prosthodontics

& Biomedical Engineering

Page 3: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Disclosure

Advisory Board BlueLight Analytics

San Francisco Giants fan

Supporter of Baltimore Orioles

Page 4: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

What are we light curing?

Adhesives

Restorative composites

Sealants

Ceramic veneers

Ceramic inlays/onlays

Fiber posts

Bases-liners

Tray materials

Denture base materials

Prosthodontic provisional materials

Desensitizing agents

Bleaching agents

Page 5: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Majority of direct restorations placed Composite Resins

• 130,054 General Dentists in US

• >122 million direct resin restorations

• Plus veneers + orthodontics

ADA 2005-6 Survey of Dental Services

All used light curing!

Page 6: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

“Despite steady progress in learning how to better formulate and cure, or harden, dental composites… Studies have shown that dental resin composites have an average replacement time of 5.7 years due to secondary decay and fracture of the restoration.”

National Institute for Dental and Craniofacial Research, 2009

Page 7: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Understanding Light Curing

Page 8: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Price RB, Fahey J, Felix DM. Knoop hardness mapping used to compare the efficacy of LED, QTH and PAC curing lights. Oper Dent. 2010; 34:58-68. Sharkey S, Ray N, Burke F, Ziada H, Hannigan A. Surface hardness of light activated composites cured by two different visible light sources: an in vitro study. Quintessence Int. 2001; 32:401-5. Rueggeberg FA, Caughman WF, Curtis JW Jr. Effect of light intensity and exposure duration on cure of resin composite. Oper Dent. 1994; 19:26-32. Malhotra N, Mala K. Light curing considerations for resin-based composite materials: a review. Part I. Compend Contin Dent Educ. 2010; 31:498-505. Suh BI, Feng L, Wang YH, et al. The effect of the pulse-delay cure technique on residual strain in composites. Compend Contin Dent Educ (Special Issue); 1999; 20(2):4-12 Duke ES, Light-emitting diodes in composite resin polymerization. Compend Contin Dent Educ. 2001; 22:722-5. Yap AU, Wong NY, Siow KS. Composite cure and shrinkage associated with high intensity curing light. Oper Dent. 2003; 28:357-64. Schattenberg A, Lichtenberg D, Stender E, Willershausen B, et al. Minimal exposure time of different LED-curing devices. Dent Mater. 2008; 24:1043-9. Fan PL, Schumacher RM, Azzolin K, Gerary R, et al. Curing-light intensity and depth of cure of resin based composites tested according to international standards. J Am Dent Assoc. 2002; 133:429-34. Yap AU, Seneviratne C. Influence of light energy density on effectiveness of composite cure. Oper Dent. 2001; 26:460-6. Calheiros FC, Kawano Y, Stansbury JW, Braga RR. Influence of radiant exposure on contraction stress, degree of conversion and mechanical properties of resin composites. Dent Mater. 2006; 22:799-803. Judy RH, Dunn WJ, Patel AB, Swanson T. Effective single-charge end point of cordless light-emitting diode light-curing units. Am J Orthod Dentofacial Orthop. 2006; 130:378-84. Felix D, Price RB. The importance of stating disaggregated irradiance values from curing lights. J Dent Res (IADR abstracts). 2011; 90:abstract no. 260. Caughman WF, Rueggeberg FA, Curtis JW Jr. Clinical guidelines for photocuring restorative resins. J Am Dent Assoc. 1995; 126:1280-6. Strassler HE. Cure depths compared with LED and other curing lights. J Dent Res (Special Issue, AADR abstracts). 2003; Spectral curing lights and evolving technology. ADA Prof Prod Rev. 2009; 4(4):1-12. Xu X, Sandras D, Burgess JO. Shear bond strength with increasing light-guide distance from dentin. J Esthet Restor Den. t 2006; 18:19-28. Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent. 2000; 12:320-7. Felix CA, Price RB. The effect of distance from light source on light intensity from curing lights. J Adhes Dent. 2003; 5:283-91. Hinoura K, Miyazaki M, Onose H. Effect of irradiation time to light-cured resin composite on dentin bond strength. Am J Dent. 1991; 4:273-6. Rueggeberg FA, Jordan D. Light tip distance and cure of resin composite J Dent Res (Special Issue A), 1992; 71:188 abstract. Felix CA, Price RB. Effect of distance on power density from curing lights. JDR 85 (Special Issue B) IADR abstracts. 2006; abstract no. 2486. Pilo R, Oelgresser D, Cardash HS. A survey of output intensity and potential depth of cure among light-curing units in clinical use. J Dent. 1999; 27:235-41. Pires JA Cvitko E, Denehy GE, Swift EJ. Effects of curing tip distance on light intensity and composite resin microhardness. Quintessence Int. 1993; 24:517-21. Price RBT, McLeod ME, Felix CM. Quantifying light energy delivered to a Class I restoration. J Can Dent Assoc. 2010; 76(2):a23. resin proximal contact. Oral Health J. August 2010; 60-73. Rueggeberg F, Mutluay MM, Price RBT, Felix CM. Efficacy of a training device for increasing curing energy delivery. J Dent Res. 2010; 89(Special Issue B, Barcelona IADR, research abstracts): abstract no. 4079 McAndrew R, Lynch CD, Pavli M, Bannon A, et al. Summary of: the effect of disposable infection control barriers and physical damage on the power output of light curing units and light curing tips. Brit Dent J. 2011; 210:E12. Price RB, Rueggeberg FA, Labrie D, Felix CM. Irradiance uniformity and distribution from light curing units. J Esthet Restor Dent. 2010; 22:86-101.

Tested as Vitrebond Powder/Liquid

Science of light curing

Page 9: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Composite Use

• 122,666,950 direct resin restorations a year ADA 2005-6

• One third of dentists have eliminated amalgam. Christensen, Dental Economics Jan 2011

• In the last 12 months in the U.S., the ratio of

composite placed compared to amalgam placed

was two Class II composites for every one Class

II amalgam. Limoli and Associates, Oct 2010

Page 10: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Did you know?

• 122,666,950 direct resin restorations a year: ADA 2005-6

• 50 working weeks a year x 5 days = 250 working days

• 490,667 resins a day

• 490,667 times a curing light was used

• 490,667 times @ $200 each

• $98,133,560 a day in the US ALONE

• $24.5 BILLION a YEAR

Page 11: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Light-curing… so easy a caveman can do it!

Page 12: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

The Problem

Process of “light-curing” is

treated with “little regard” to the exacting science it really is.

“Too easy” and minimal attempts seem to produce an acceptably “hard” restoration.

Page 13: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Understanding the challenge

Teach students-clinicians how to use their light properly and what affects the extent of cure of THEIR RESTORATIONS :

What “spectral output” of the light means

What are the “spectral needs” of the resin

What is the difference between irradiance and energy density

How exposure duration and spectral delivery affect final restoration cure

Page 14: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

The Problem: not curing composite completely

• Lower bond strength

• Increased microleakage

• Increased recurrent caries

• Increased staining-color changes

• Increased wear

Premature restoration failure!

Uncured Resin

Tooth

Page 15: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Not all curing lights are the same

Understanding

The Differences

Page 16: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Power = mWatts

How much energy- what type of energy?

Page 17: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Irradiance (Power/Unit Area) mW/cm2

Page 18: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Irradiance 1735 mW/cm2

0

2

4

6

8

370 390 410 430 450 470 490 510 530

Wavelength (nm)

Abso

lute

Irra

dian

ce (m

W/n

m/c

m2 )

Spectral Radiant Power mWatts/nm

Page 19: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Understanding

Power: mWatts

Irradiance: (Power/Unit Area) mW/cm2

Spectral Radiant Power: mWatts/nm

Energy Density = Irradiance x Time: 8 - 16 J/cm2

Page 20: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

MARC (BlueLight Analytics)

BlueLight makes MARC™− Managing Accurate Resin Curing.

MARC™ was invented by Dr. Richard Price and his research associate, Chris Felix, at Dalhousie University.

MARC™ is the first and only scientifically accurate, clinically relevant and easy-to-use energy measurement system for measuring what resin actually receives.

Page 21: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Consistent calibration enables apples:apples comparisons

Who is MARC™?

Scientifically accurate, clinically relevant

& easy-to-use

MARC™ Patient Simulator

quantifies energy delivery to resin

Page 22: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

LIGHT SPECTRUM

Page 23: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

LIGHT SPECTRUM

Irradiance = 1,200 mW/cm²

Page 24: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Irradiance not uniform

What is the beam profile?

Price, Rueggeberg et al. J Esthetic and Restorative Dentistry 2010

5,834 mW/cm2 388 mW/cm2

1200 mW/cm2

Page 25: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Detects uneven beam distribution

hot spots-cold spots

Page 26: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Operator technique evaluation

Typodont head with detectors in teeth measure

light received

Page 27: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Bottom line…

How do we make the science relevant and create value during a student’s/clinician’s education?

Page 28: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

For a clinical procedure light curing = 15-60 sec

Page 29: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Understanding leads to improved clinical success

Light energy matters!

Page 30: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Safety concerns

Page 31: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Safety concerns

Page 32: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent
Page 33: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent
Page 34: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Survivability Amalgam vs Composite

Page 35: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent
Page 36: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Effect of Distance from the Light

Guide on the Irradiance Received

Felix CA, Price RB. Effect of Distance on Power Density from Curing Lights. Journal of dental research 2006;85:abstract 2486 (www.dentalresearch.org).

Page 37: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

When curing adhesives in deep proximal boxes with quartz halogen light (600 mw/cm2) curing time should be increased to 40-60 seconds to ensure optimal polymerization

Xu X, Sandras D, Burgess JO. Shear bond strength with increasing light-guide distance

from dentin. J Esthet Restor Dent 2006

Page 38: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

To maximize the energy delivered, the operator should wear eye protection, should watch what he or she is doing and should hold the light both close to and perpendicular to the restoration.

Price RB, McLeod ME, Felix C Quantifying light energy delivered to a Class I restoration J Can Dent Assoc 2010

Page 39: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Where is the light relative to what it is curing?

Page 40: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

600 mw/cm2

700 mw/cm2

800 mw/cm2

1000 mw/cm2

1200 mw/cm2

1400 mw/cm2

Page 41: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

370 mw/cm2

450 mw/cm2

600 mw/cm2

800 mw/cm2

900 mw/cm2

1000 mw/cm2

Page 42: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Orientation of the Light

**Right angle **Close to the tooth as is possible

Matrix band moves The light further away!

Page 43: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Mis-Orientation of the Light

Note the height of the matrix

Page 44: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Lights not working optimally

Page 45: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

MARC Scientifically accurate- clinically relevant device. Measures: Irradiance Spectral emission Energy values

Page 46: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Operator technique

MARC for light curing training

Page 47: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Good light curing habits CORE

Curing Light

Operator Technique

Restoration (location, depth, size, opening)

Energy Requirement

Page 48: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Know your Curing light

Page 49: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

37 Operators Same Light, Same Tooth, Same Time

Optimal is 8.0 J/cm² Range from 0.9 – 7.2 J/cm² Average was 4.1 J/cm²

Page 50: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Teaching and Evaluating Light Curing

• NOT looking, NOT stabilizing,

• NOT wearing eye protection!

Wearing eye protection,

looking, stabilizing!

Before Instruction After Instruction

Page 51: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

After Group Instruction

Optimal is 8.0 J/cm² Range from 5.8 – 7.6 J/cm² Average was 6.7 J/cm²

Page 52: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Operator controlled Light tip diameter of 7.5 mm

Sealant on molar- 2 cycles (overlapping tip)

Occlusal Class I- 2 cycles (overlapping tip)

Page 53: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Class I or II premolar- 1 cycle no overlap needed unless the premolar is larger than normal

Class II molar- 2 cycles (overlapping tip)

Page 54: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Facial of a maxillary central incisor two or four cycles with overlap

Facial veneer • Direct composite

• Porcelain veneer

Class IV

Light cure facial and lingual

A maxillary lateral incisor two cycles

Curing porcelain veneers

4 cycles w/ overlap

Page 55: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Negative effects-Restoration Light angulation-Light movement

A 30°angle can reduce energy delivery by 26%

Page 56: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Change the angulation Change the cure

TOP TOP

Page 57: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Energy requirement

Power: mWatts

Irradiance: (Power/Unit Area) mW/cm2

Spectral Radiant Power: mWatts/nm

Energy Density = Irradiance x Time: 8 - 16 J/cm2

Page 58: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

7 Steps to Ensure Better Light Curing

1. Wear orange glasses (blue light blocking) for safety and so you can watch what you are doing

2. Re-position the patient so you can easily see the restoration and access it with the curing light

3. Position yourself comfortably so you can stabilize the curing light directly over the preparation

Page 59: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

7 Steps to Ensure Better Light Curing

4. Adjust the light guide so you can operate the light comfortably. Clean the tip as needed.

5. Stabilize the curing light so the beam is perpendicular to the surface of the resin

6. Begin curing 1mm away from resin and then move as close as possible after 1 second

7. Air-cool or wait between each curing cycle; test the temperature rise from the light on the back of your hand.

Page 60: Understanding Light Curing Improved Clinical Success...Price RB, Derand T, Sedarous M, Andreou P, et al. Effect of distance on the power density from two light guides. J Esthet Dent

Understanding leads to improved clinical success

Light energy matters!