Upgrade the RIXS Beamline to Ultrahigh Resolution at NSRRC:
Mechanical Stability Issues D.J. Wang, C.C. Chiu, T.C. Tseng, L.J. Huang, H.S. Fung, H.W. Fu, S.W. Lin, C.S. Lin, C.H. Lai, W.B. Wu, S.C. Chung, D.J. Huang, and C.T. Chen
Reporter: C.S. Lin
National Synchrotron Radiation Research Center Hsinchu, Taiwan
outline
• 1. Introduction (optical layout, stability issues)
• 2. Original design and setup
• 3. Measurement instruments
• 4. Upgraded design and engineering
• 5. Performance of the upgrade
• 6. Summary
C. H. Lai, H. S. Fung, W. B.Wu, H. Y. Huang, H.W. Fu, S.W. Lin,C. C. Chiu, D. J. Wang, L. J. Huang, T. C. Tseng, S. C. Chung, C.T. Chen and D. J. Huang Through this new system, the commissioning results successfully showed spin-flip, dd, and charge-transfer excitations of NiO. These results are consistent with previous published results, while maintaining a better spectral resolution and a higher measurement efficiency. The best energy resolution of our setup in terms of full width at half maximum is 77 meV at the incident photon energy tuned about the Ni L3-edge.
Journal of Synchrotron Radiation 2014, Mar 21, p325-332
High-Resolution, High-Efficiency Beamline and Spectrometer for Reasonace Inelastic Soft X-ray Scattering (RIXS)
Layout of RIXS beamline
Base on the energy compensation principle, we build the AGM –AGS beamline for RIXS.
The first optical element is a horizontal focusing mirror (HFM) located at distance 12.05m from the center of an elliptically polarized undulator (EPU), which is the source of X ray.
A vertical focusing mirror (VFM) is located at distance 1.4 m from the HFM to focus the X-ray beam onto an entrance slit.
The light is further focused and dispersed onto the sample by the active-grating monochromator (AGM).
An aperture is placed just before the sample to define the energy spread of the incident photons.
After the sample, a horizontal refocusing mirror (HRFM) enhances the efficiency of collection of scattered X-rays in the horizontal direction.
An active-grating spectrometer (AGS) is used also to disperse and to focus the scattered light onto a two-dimensional charge-coupled device (CCD).
Bendable grating
Entrance slit opening
don’t scarify resolution
Anticipation of Beamline User : High resolution and High Stability
In this innovative optical design,
theoretical resolution is very high but limited by the
realistic optics quality. (so far 0.5urad rms slope
error, A)
High stability is related from source through each
optical component and CCD. (below 0.1urad p-p, B)
Grating
2 PZT actuator
Real resolution〜 〔(A2 +B2+…)〕1/2
NiO RIXS (energy loss) Spectra
3 minutes/spectrum
Spectrometer @ 90 degrees
545 counts/pixel
For the energy of point A, CCD takes the energy loss spectrum 3 min to get FWHM, and further to determine the resolution.
Due to Low stability of AGM, AGS and CCD will then deteriorate the resolution
(weak photon signal expt)
Why is the stability sensitive for this experiment
Monochromator
AGS
CCD
Entrance slit
Sample chamber
Typical RIXS original Setup
To collect RIXS signal at different scattering angle, AGS could be moved to different location. Blue steel frame moves along the circle trace and carries the AGS and CCD to desired location.
How to stably support the movable AGS and CCD?
Before 2013 Feb
0 2 4 6 8 10 12 14 16
-0.2
0.0
0.2
0.4
0.6
0.8
1.0
1.2
Time(sec)
m
vertical
background RMS:0.010m
steel framework RMS:0.122m
Vertical and horizontal vibration of original
steel frame
Instrument: Lion precision capacitive sensor relative to ground
Vertical vibration is 10 times higher than background
0 2 4 6 8 10 12 14 16
-0.08
-0.04
0.00
0.04
0.08
0.12
0.16
0.20
m
Time(sec)
Horizontal
background RMS:0.004m
steel framework RMS:0.018m
Measurement instruments
• High-resolution autocollimator and stable support for measuring grating angle (20Hz, 0.1urad) on grating surface
• Capacitive sensor for measuring displacement of support (1Kz, 0.001um)
• Linear encoder of grating actuator (20Hz, 2nm)
• PCB accelerometer for measuring chamber (4-100HZ)
Autocollimator
Mirror Granite support
Upgraded engineering
• AGS, CCD granite engineering and air floating.
• AGM and AGS chamber will be rebuilt.
• AGS & CCD are located on granite base with flatness = 5um. The granite base will be lifted height 30um @3Kg/cm^2 when operating 4 air purge holes.
• Each granite platform segment was polished at a factory in advance. The installation flatness is about 50um/10m, and pitch direction is below 20urad.
• Slow curing epoxy was used to fill with the gap between granite and ground.
• Repeatability of sliding granite in pitch direction is about 1 urad by lifting it up and down.
Granite engineering
Granite platform
Sliding granite base
Wire pulling mechanism of the sliding granite
AGS & CCD sliding granite base
moving mechanism
Using 8 steel wires with spring force to pull the granite base, and then to prevent it from distortion by steel frame
TLS RIXS BL upgrade
2013 Feb
New AGM chamber system
New system
MEDSI2012 C.C. Chiu D. J. Wang
Using Stable granite base to support chamber
Introducing Vertical big chamber Decoupling Grating adjusting
mechanism with Vertical big chamber
AGM design
Grating rotation center
Flexible pivot
Motor Micos Connection rod, 30cm vertical
Weight balance
The grating support is mounted to granite and decouple with chamber via bellows, and therefore prevents the vibration from both sides of the chamber
Bellows
Grating support
Granite
Resolution of AGM by Micos motor
MEDSI 2014 Resolution is good enough for the grating
Refine chamber mounting
Because the weight and vacuum force are different on both sides of the chamber, we will introduce bellows support to reduce extra moment on the chamber.
beam
Bellows support
bellows
Chamber mounting from view of beam direction
Ion pump
bellows
Due to Weight of IP and GV is more than AGM chamber, we use bellows to decouple with the chamber to further prevent the indeterminate supporting points of this system.
0 50 100 150 200
0.00008
0.00010
RMS= 2.30 nm
Po
sitio
n(m
m)
Time (sec)
Best before Aug. shutdown
20140930 after upstream frame
RMS= 2.33 nm
AGM chamber with bellows support
Data obtained from Micos motor encoder
User response: Bellows support help shorten the pendulum-phenomenon duration of grating holder in the chamber of AGM subjected to external excitation
Bellows support
Without bellows support
367mm
10nm p-p
10nm p-p/367mm
≈ 27.2nrad
AGS design Insert Bellows support locked by screw rod
• AGS is mounted on the bottom flange due to height limitation
• The adjusting mechanism for angle driving of AGS is the same as the one of AGM
Add wt on bottom flange
2000 2500 3000
1.5
2.0
2.5
3.0
Time(sec)
ra
d
AK background on 05A granite
RMS 0.13 rad
60cm
Stability test of Autocollimator
Autocollimator
Mirror
Wind shield
No bellows support No wt
No bellows support Wt 80Kg
1 bellows support Wt 80Kg
2 bellows support Wt 80Kg
Grating encoder (nm rms)
3.6 2.3 2.2 2.18
Autocollimator (urad)
0.38 0.178 0.187 0.167
AGS chamber mounting stability
The bellows support are tentative, and just shows the trend
24
[100] 𝜎-pol
[001]
857
856
855
854
853
852
Incid
en
t P
ho
ton
En
erg
y (
eV
)
-8 -6 -4 -2 0
Energy Loss (eV)
700
600
500
400
300
200
100
0
Spin d-d CT
1.0
0.8
0.6
0.4
XA
S (
arb
. u
nits)
865860855850
Photon Energy (eV)
NiO XAS
300
200
100
0Inte
nsity (
arb
. u
nits)
-0.4 -0.2 0.0 0.2Energy Loss (eV)
853.0 855.0 855.5 857.0
FWHM = 108 meV for several hours
Commissioning results: RIXS spectra on NiO
Excitation of local spin-flip
The surface normal of the NiO sample was along the [001] direction; the incident angle was 30 from the sample surface.
RIXS spectra excited with incident photons of energies 853, 855 and 857 eV.
RIXS spectra in an energy range near the elastic scattering excited with of selected incident energies labeled in units of eV.
Color mapping of RIXS intensity across the L3 absorption edge, obtained by combining RIXS spectra excited every 500 meV.
We chose NiO as a test sample for high resolution RIXS measurements using a side branch of the EPU beamline at the TLS.
NiO is a prototypical system of strongly correlated electrons that exhibit many-body characteristic properties.
Summary
• After upgrading the stability of RIXS beamline, users are
satisfied with the resolution and stability of RIXS system.
• From grating encoder reading, it seems that we can
attain below 30nrad p-p mechanical stability
(corresponding to 10nm obtained from Micos motor encoder) p-p mechanical stability for RIXS system, and
this mechanical stability can last for several hours.
• Beamline signals obtained from CCD are more sensitive
than most commercial measurement tools.