URGENT FIELD SAFETY NOTICE Ortho Clinical Diagnostics Applying VidimSoft Label s to ORTHO BioVue® System Cassettes Date Product l ssue Explanation Impact to R esults Required Actions 4th May 2017 This Urgent Field Safety Notice is in reference to the following products: P roduct Name Product Co de Vi dimSoft OrthoScan'M System 300.0122 ORTHQ'M Workstation for Bi oVue® Cassettes 6904629 The OrthoScan System (OrthoScan), manufactured by VidimSoft, is used to capture electronic images of the aggl uti nation patterns of ORTHO BioVue® System Cassettes (BioVue Cassettes) and assist in interpretation, security and storage of the images. The ORTH O Workstation, manufactured by Ortho Clinical Diagnostics (Ortho), supports the in vitro i mmunohemato logy testing of human blood through the incubation and centrifugation of BioVue Cassettes. As part of a Field Safety Corrective Action, Ortho Clinical Diagnostics (Ortho) init iail:ed this Urgent Field Safety Notice having been made aware that, due to the requirement to manually apply up to six (6) barcode Iabeis to BioVue Cassettes when using OrthoScan, there is the possibil ity that if Iabeis are applied to a cassette incorrectly (with gaps and/or creases with one or multiple Iabeis applied), the cass· ette may not be positioned properly in the ORTHO Workstation Centrifuge Cassette Holder. Centrifugat ion of BioVue Cassettes that ar enot posit ioned properly (i.e. Ievei) into the ORTHO Workstati on Centrifuge Cassette Holder may produce incorrect (false positive) results. Affected sample results occurring prior t o this communication arenot easily identifiable; thus, a review of previous results may be impractical. lf your Iabaratory adds VidimSoft Iabeis to BioVue Cassettes and 1 uses a ORTHO Workstat ion, operators should ensure that: o Label sare applied without a significant gap at the bottom of the cassette and wi thout creases in the Iabei after it is applied to the cassette to help assure the cassette is posit ioned correctly in the Workstation Centrifuge Cassette Holder. Refer to Illustrations on Page 2. o Cassettes are positioned Ievei in the ORTHO Workstation Cenil:rifuge Cassette Holder prior to the init iation of the centrifugation cycle. Refer to Illustrations on Page 2. • Post th is notification by your ORTHO Workstation or wi th your user documentation. • Complete andreturn the Confirmation of Receipt form by 18 May 2017. Contact Piease contact the Ortho Care'M Technical Salutions Center at 0800 895 963 if you have Information further questions or require additional information. Ref. CL2017-096_ EU Page 1 of 2

Upload: others

Post on 04-Mar-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: URGENT FIELD SAFETY NOTICE Ortho Clinical Diagnostics ......As part of a Field Safety Corrective Action, Ortho Clinical Diagnostics (Ortho) initiail:ed this Urgent Field Safety Notice

URGENT FIELD SAFETY NOTICE

Ortho Clinical Diagnostics Applying VidimSoft Labels to ORTHO BioVue® System Cassettes

Date

Product

lssue Explanation

Impact to Results

Required Actions

4th May 2017

This Urgent Field Safety Notice is in reference to the following products:

Product Name Product Code

VidimSoft OrthoSca n'M System 300.0122

ORTHQ'M Workstation for BioVue® Cassettes 6904629

The OrthoScan System (OrthoScan), manufactured by VidimSoft, is used to capture electronic images of the agglutination patterns of ORTHO BioVue® System Cassettes (BioVue Cassettes) and assist in interpretation, security and storage of the images.

The ORTHO Workstation, manufactured by Ortho Clinical Diagnostics (Ortho), supports the in vitro immunohematology testing of human blood through the incubation and centrifugation of BioVue Cassettes.

As part of a Field Safety Corrective Action, Ortho Clinical Diagnostics (Ortho) init iail:ed this

Urgent Field Safety Notice having been made aware that, due to the requirement to manually apply up to six (6) barcode Iabeis to BioVue Cassettes when using OrthoScan, there is the possibility that if Iabeis are applied to a cassette incorrectly (with gaps and/or creases with one or multiple Iabeis applied), the cass·ette may not be positioned properly in the ORTHO

Workstation Centrifuge Cassette Holder.

Centrifugation of BioVue Cassettes that arenot posit ioned properly (i.e. Ievei) into the ORTHO Workstation Centrifuge Cassette Holder may produce incorrect (false positive) results.

Affected sample results occurring prior t o this communication arenot easily identifiable;

thus, a review of previous results may be impractical.

• lf your Iabaratory adds VidimSoft Iabeis to BioVue Cassettes and 1uses a ORTHO Workstat ion, operators should ensure that:

o Labelsare applied without a significant gap at the bottom of the cassette and without

creases in the Iabei after it is appl ied to the cassette to help assure the cassette is posit ioned correctly in the Workstation Centrifuge Cassette Holder. Refer to Illustrations on Page 2.

o Cassettes are positioned Ievei in the ORTHO Workstation Cenil:rifuge Cassette Holder prior to the init iation of the centrifugation cycle. Refer to Illustrations on Page 2.

• Post this notification by your ORTHO Workstation or with your user documentation.

• Complete andreturn the Confirmation of Receipt form by 18 May 2017.

Contact Piease contact the Ortho Care'M Technical Salutions Center at 0800 895 963 if you have Information further questions or require additional information.

Ref. CL2017-096_ EU Page 1 of 2

Page 2: URGENT FIELD SAFETY NOTICE Ortho Clinical Diagnostics ......As part of a Field Safety Corrective Action, Ortho Clinical Diagnostics (Ortho) initiail:ed this Urgent Field Safety Notice

Illustrations of Proper Label Application and Cassette Positioning

Ref. CL2017-096_ EU

Correctly applied Iabei

Cassette correctly positioned in Centrifuge Holder

Cassette incorrectly positioned in Centrifuge Holder