Page 1: Urinalysis Simplified - ARKRAY USA

AUTION HYBRID ™FULLY AUTOMATED INTEGRATED URINE ANALYZER

AU-4050

Urinalysis SimplifiedFully Automated Integrated Urine Chemistry and Sediment Analyzer

• Compact • Accurate • Easy to Use

Page 2: Urinalysis Simplified - ARKRAY USA

• Simply load specimens, walk away, and return to results flagged for review

• Load up to 60 samples at a time

• Process 100-200 samples per hour

Optimize laboratory workflow with the AUTION HYBRID™ AU-4050. A fully automated and integrated urine chemistry and sediment analyzer allows operators to walk away to other activities.

Chemistry Analysis

• Reliable AUTION test strips

• Second generation module

• Easy-to-load strips with no calibration required

Uriflet S 9HA Strips

• Excellent sensitivity and specificity

• Clinically significant reporting ranges

• Color compensation pad corrects results for normal urine colors

• Abnormal color detection alerts users of potential false positive reactions

Competitor A

45"

Competitor B

63" 31.5"

30-50% smaller than the competition

COMPACT DESIGN

CONSOLIDATES PROVEN TECHNOLOGIES

Page 3: Urinalysis Simplified - ARKRAY USA

• Empower clinicians to rapidly screen for UTI on admission to increase reimbursement

• Reduce costs associated with negative or contaminated urine cultures using accurate sediment results

• Capacity to perform sediment analysis on all specimens for improved disease state screening

• Increase the ability to find rare elements automatically analyzing approximately 44 HPFs

Sediment Analysis

• Proven flow cytometry accurately identifies and enumerates formed elements

• Standardize results by reducing subjective sediment interpretations

• Achieve an autoverification rate of up to 92%

• Reduce hands-on time reviewing flagged results

• Wide linearity with no dilutions required

• Accurate bacteria detection using a dedicated reaction chamber, nucleic acid stain, interference removal, and anti-carry over wash

1 HPF 10 HPFs 44 HPFs

Digital Microscopy 500 pictures

Manual Flow Cytometry 65,000 particles

AUTION HYBRID™ AU-4050Proven technology you trust to achieve the quality results you expect.

EASY TO USE

IMPROVED OUTCOMES

• Rapid results in as little as 2 minutes and 15 seconds

• Intuitive Windows® based IPU software

• One easy-to-read consolidated report

• Barcoded consumables management

• Automatically populate QC fields with AUTION controls

• Minimal maintenance, 5 minutes daily/weekly and 10 minutes monthly

Page 4: Urinalysis Simplified - ARKRAY USA

AUTION HYBRID ™FULLY AUTOMATED INTEGRATED URINE ANALYZER

AU-4050

Minneapolis, MN 55439 • 877.538.8872 • [email protected] • www.arkrayusa.com AH016-01 Rev 6/16

Urine Chemistry (CHM)

Measurement Parameters• Leukocyte • Nitrite • Glucose

• Blood • Bilirubin • Urobilinogen

• Specific Gravity• Color• Turbidity

• Ketone • pH• Protein

Measurement Technology

Test strips: Dual wavelength reflectance method Single wavelength reflectance for blood Specific gravity: reflection refractometry Color: light-transmission Turbidity: light-scatter

Onboard Capacity Maximum 400 strips

Urine Sediment (FCM)

Measurement Parameters

Enumerated: • RBC • WBC • Epithelial Cells • Hyaline Casts • Bacteria

Flagged: • Pathological Casts • Small Round Cells • Crystals • Yeast-like Cells • Mucus • Sperm

Measurement Technology• Fluorescent flow cytometry• Two separate chambers (bacteria and sediment) with proprietary polymethine dyes

GeneralDimensions Width: 31.5” Depth: 28.3” Height: 28.3”

Weight 265 lbs (120 kg)

Power Supply 100-240 VAC, 50/60 Hz

Operating EnvironmentTemperature: 10–40°C Humidity: 20–80% R.H.

Data Storage Up to 10,000 analyses

Throughput 100–200 samples/hour (varies depending on CHM and FCM test mix)

Minimum Sample Volume40 mm from bottom of tube (approx. 5 mL) 1 mL manual FCM

Communication RS-232C, Ethernet