usb vnas - copper mountain tech · equipped a lab with 12 vnas instead of 1. ... manual •download...

16
USB VNAs Extend Your Reach

Upload: trinhthuan

Post on 27-Apr-2018

214 views

Category:

Documents


1 download

TRANSCRIPT

USB VNAs Extend Your Reach

Agenda

• Intro to Performance Testing– Aka Annual Calibration

• Intro to VNAPT

• Getting started with VNAPT

• Wrap up

Engineers Extending Their Reach

Dielectric LLCTraveling to job sites is made easier with S5048

CERNTR5048 is used in the worlds largest particle accelerator

KwiksetR54 deployed in R&D and production of Kwikset KevoBluetooth locks

College of New JerseyEquipped a lab with 12 VNAs instead of 1

USB VNA

Purpose of Performance Test

• Check the VNA’s performance metsspecifications

–Output power, CW frequency, harmonic distortion, etc.

• Recommended performance test interval is annual

Purpose of VNAPT

• Automate VNA performance testing

–Ease the process

–Automatic test report generation

• You can download the VNAPT software for Copper Mountain Technologies website

• Test procedure found in the Performance Test Manual

• Download the PT Manual from the CMT website

Test Content

Intro to VNAPT

• Procedure

1. Connect VNA to PC

2. Launch VNA software, confirm “Ready” status

3. Launch VNAPT

4. Choose the VNA under test, click blue Start button to begin

Required Equipment

• Listed in Performance Test manual

– Mainly: frequency counter, spectrum analyzer, power meter, gages, torque wrenches, calibration kit, and verification kit

– Equipment similar to listed can be used if it satisfies specifications

– NPR-Z51 Thermal Power Sensor is required for automation of output power tests

VNAPT User Interface (1 of 3)

• Metrology

• Accessibility

• Innovation

Click to view

the navigation

Click to

change the

inspection

result

Click to

enter to

test

section

Click to

generate or

print test report

VNAPT User Interface (2 of 3)

• Metrology

• Accessibility

• Innovation

Click to back to

home screen

Enter theoretical

value

First, click the

button to go to

VNA program

and do the

calibration

Then, click to

perform

automatic

measurement

VNAPT User Interface (3 of 3)

• Metrology

• Accessibility

• Innovation

Click to

expand for

detailed

instruction

Click to pop up

a window for

detailed

instruction

Please Note

• Not all VMT VNAs are supported (yet):

–Version 1.4.9.22 of VNAPT supports: Planar 8xx/1, S5048, and “R” family

• VNA COM server must be registered (last step in installer wizard)

– If in doubt, just uninstall/reinstall the VNA program taking care to register the server

Product Families

Reflectometers85 MHz to 14 GHzOperates without the use of a test cable

Compact20 kHz to 4.8 GHz50Ω and 75Ωmodels available

Planar100 kHz to 8 GHz4-port and direct receiver access models available

Cobalt100 kHz to 20 GHzUnmatched combination of size, speed and wide dynamic range

Final Thoughts

• Any suitable lab can execute annual verification/calibration, either manually or with VNAPT

• VNAPT is not a substitute for detailed procedures of the Performance Test manual!

• Please contact us with any questions or for more details on performance test procedures [email protected]