use of standards in the review of medical devices
TRANSCRIPT
www.elsevier.com/locate/jelectrocard
Journal of Electrocard
Use of standards in the review of medical devices
Charles HoT, Donald Jensen, Frank Lacy, Neal Muni, Sabina Reilly, Elias MallisDepartment of Health and Human Services, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Rockville, MD 20850, USA
Received 10 June 2005; accepted 10 June 2005
Abstract In keeping with its mandate to provide the least burdensome means of product regulation, the US
0022-0736/$ – see fro
doi:10.1016/j.jelectroc
T Corresponding
Food and Drug Administration’s (FDA) Center for Devices and Radiological Health uses many
different standards to facilitate the review of premarket submissions of medical devices. The benefits
of using standards in this manner include providing a set of common requirements and test protocols
to the device manufacturer, thus reducing the manufacturer’s need to breinvent the wheelQ for eachnew bench test to ensure safety and effectiveness of the device. With the current trend toward
international harmonization of standards, tests performed in accordance with an international
standard may be acceptable to several countries. However, there are instances in which the FDA does
not agree with certain provisions in a standard. This article aims to clarify the approaches taken by
the FDA to balance or resolve issues that arise when there is disagreement regarding the acceptability
of a particular standard. One approach begins with the recognition of only some provisions of a
standard, or more commonly, excluding those parts of a standard that are unacceptable to the FDA
for regulatory or scientific validity reasons. Other approaches include working with the standards
development organizations to revise the standard to include language more agreeable to all parties
involved. Specific examples will be presented on medical devices, such as electrocardiogram cables
and connectors, and noninvasive blood pressure monitors.
2005 Published by Elsevier Inc.
Keywords: FDA; Medical devices; Standards; Pressure band method
1. Introduction
The US Food and Drug Administration’s (FDA) Center
for Devices and Radiological Health uses many recognized
standards to facilitate the review of premarket notification
(510(k)) submissions of medical devices. Section 514(c) of
the Food, Drug, and Cosmetic Act, as amended, provides
explicit authority to use FDA-recognized standards in the
premarket review of medical devices. Many benefits are
accrued by using standards in this manner, including the
ability to provide a set of common requirements and test
protocols to the device manufacturer, thus reducing the need
to breinvent the wheelQ for each new bench test to ensure
safety and effectiveness of the device. Furthermore, with the
present trend toward international harmonization of stand-
ards, tests performed in accordance with an international
standard may be acceptable to several countries, facilitating
nt matter 2005 Published by Elsevier Inc.
ard.2005.06.030
author.
simultaneous product approval applications to multiple
international regulatory bodies. In addition, the use of
standards may reduce the need for supporting documents in
the premarket notification submissions, as in the case of
Special 510(k)’s. In this article, unless otherwise stated, the
term standards refers to bvoluntary consensus standards.QBefore the agency can officially recognize a standard, the
standard is reviewed for compatibility with agency regula-
tory and scientific needs. Subsequent to this review, an
associated bsupplementary information sheet Q is developedto accompany the recognized standard. This information
sheet documents the extent of the standard’s recognition,
and whether or not any additional documentation, such as
summary report of test results, is needed. The rationale for
the supplementary information sheet is that when a
manufacturer declares conformity to the standard, the
agency can have a better understanding of the extent of
the conformity.
In most cases, a recognized standard is accepted in its
entirety by the agency. In rare occasions, however, the FDA
iology 38 (2005) 171–174
C. Ho et al. / Journal of Electrocardiology 38 (2005) 171–174172
may not completely agree with the provisions of the
standard, necessitating certain steps to resolve specific
differences. This article aims to clarify the steps taken by
the FDA to balance or resolve such differences, using
illustrative device regulatory submissions as examples for
which differences in FDA and technical standards were
identified and then resolved.
2. Conflict with a Federal regulation
One reason why the FDA might not be able to accept a
voluntary consensus standard is that some provisions of a
standard actually conflict with Federal regulation. In this
case, the FDA must follow its regulatory mandate instead of
the consensus standard.
For example, the Code of Federal Regulations (CFR) spe-
cifies a mandatory bperformance standard for electrode lead
wires and patient cablesQ (21 CFR 898). Many electrical/
electronic devices, such as electrocardiogram monitors,
pulse oximeters, and apnea monitors, include patient cables.
In 21 CFR 898, the Federal performance standard ensures that
the distal connector of a patient cable (the connector located
farthest from the patient) cannot be accidentally plugged into
the AC power outlet.
The regulation, under 21 CFR 898.14 (b), sets forth cer-
tain test requirements and test methods for verifying com-
pliance with this Federal performance standard. However,
the voluntary consensus standard, ANSI/AAMI EC12:2000,1
section 5.3.2, sets forth the verification test as inspection and
does not specify a test finger for verifying compliance.
Because this clause of the EC12 standard disagrees with the
Federal regulation, the FDA supplementary information
sheet indicates that this clause is not recognized. The
manufacturer is required to use the test methods set forth
in 21 CFR 898.14(b). The regulation also specifies that under
special circumstances, the FDAmay grant waivers to devices
used for special purposes.
Another example of a consensus standard being in
conflict with Federal regulations is the use of the English
language in a device’s Instructions for Use document.
Although the regulation under 21 CFR 801.15(c) requires
the use of the English language, except in Puerto Rico, the
NOTE in section 4.1 of EC12:2000 recommends the use of
symbols under AAMI/ISO TIR 15223, without any textual
instruction. As a result, this clause of EC12:2000 is not
recognized by the FDA. The supplementary information
sheet hence indicates that the manufacturer is required to use
the English language in the Instructions for Use. One
notable exception is the use of the symbol bRx Q to indicate
bprescription use only.Q
1 The acronyms ANSI and AAMI represent the American National
Standards Institute and Association for the Advancement of Medical
Instrumentation, respectively.
3. Differing interpretations of standard
Another reason for the FDA to not accept a consensus
standard is a difference of scientific opinion in the technical
validation used to justify the implementation of a standard
and/or its interpretation. An example is the pressure band
method used in clinical tests of noninvasive blood pressure
(NIBP) monitors.
The voluntary consensus standard, ANSI/AAMI SP10
(1992 and 2002 versions),2 has been used by many NIBP
monitor manufacturers to validate the NIBP monitor against
an invasive reference device (such as an indwelling arterial
catheter). This standard can also be interpreted to allow
the use of the pressure band method. However, the FDA
is concerned about potential deficiencies in the scien-
tific validity of the pressure band method. They are dis-
cussed below.
3.1. Pressure band method
An NIBP monitor usually requires at least 10 seconds to
record one set of systolic, diastolic, and mean pressure
readings. In contrast, an invasive reference device (eg, an
indwelling arterial catheter), continuously records several
sets of systolic, diastolic, and mean pressure readings
during this 10-second period. The method for comparing
the NIBP monitor’s single set of readings with the several
sets of reference device readings is not clearly articulated in
the standard.
In the alternative method indicated in SP10:2002, clause
C3, paragraph 2, only the single maximum and minimum
values of the systolic pressure readings of the reference
device recorded during this 10-second period could be used to
define a bpressure bandQ for systolic pressures recorded by thereference device. In similar fashion, a diastolic pressure band
can be defined using only diastolic pressure values.
Although the SP10 standard does not specify the
method of comparing the NIBP monitor and the invasive
reference device, some manufacturers have used the inter-
pretation below.
If a subject device’s systolic pressure reading lies within
the pressure band (ie, if the subject device’s reading is less
than the maximum, but greater than the minimum, recorded
by the invasive reference device), the difference between
subject device and reference device is scored as bzero.Q Ifthe subject device’s systolic pressure reading lies outside the
defined pressure band, only the distance between the subject
device reading and the edge of the pressure band is scored
as the difference between subject device and reference
device. Some manufacturers refer to this as the bpressureband method.Q An illustration of this method is provided
using artificial data.
2 ANSI/AAMI SP10:2002 is the consensus standard entitled,
bManual, electronic, or automated sphygmomanometers.Q The 1992
version is the older version of the standard and covered only automated
sphygmomanometers.
Illustration on Pressure Band Method
0
20
40
60
80
100
120
140
160
0 2 4 6 8 10 12
Time (s)
Sys
tolic
Pre
ssu
re (
mm
Hg
)
Invasive reference data NIBP monitor data
C.Hoet
al./JournalofElectro
cardiology38(2005)171–174
173
C. Ho et al. / Journal of Electrocardiology 38 (2005) 171–174174
The 1987 version of the SP10 standard and some
international draft standards (DIN 58130 and prEN 1060-4),
in fact, provide numerical examples using the pressure
band method to compare blood pressure readings from
the NIBP monitor and the invasive reference device,
although the pressure band method is not stated by name
in these documents.
3.2. Issues with the pressure band method
The FDA has concerns regarding the scientific validity of
the pressure band method because of the following technical
issues:
1. Patients with atrial fibrillation can have a very
significant beat-to-beat variation in blood pressure
depending on the length of time between ventricu-
lar systolic contractions. As a result, these patients
can have very wide pressure bands. In the FDA’s
experience with this methodology, fairly wide
pressure bands have been observed, up to almost
60 mm Hg wide, for patients with (or without)
atrial fibrillation.
2. The upper and the lower edges of the systolic (or
diastolic) pressure band are the 2 readings that define
the maximum and the minimum of the systolic (or
diastolic) pressure readings recorded by the intra-
arterial reference device during the 10-second
recording session. Thus, the pressure band method
uses only 2 readings to represent all the systolic (or
diastolic) values recorded by the invasive reference
device during this period. Hence, the method ignores
information embedded in all the remaining systolic
(or diastolic) data points. By comparison, a mathe-
matical average of all systolic (or diastolic) data
points gathers information from all such points.
Intuitively, methods of evaluation that limit the use
of available data are less optimal than those methods
that incorporate all existing data for analysis.
3. Furthermore, the FDA recognizes that single data
points are susceptible to error that is attributable to
bnoise.Q In this case, any noise interference on the
maximum or minimum data point defining the
pressure band, recorded by the reference device,
will likely increase the width of the pressure band
due to the existence of other data points inside the
pressure band, thus resulting in more zero differ-
ence scores.
The consequence of allowing a wide pressure band is that
the NIBP monitor would appear to be more accurate than
the clinical data currently support and demonstrate.
The FDA believes that there is some merit to the pressure
band method if the width of the pressure band is not
excessive. However, the FDA is not ready to propose a
specification for an acceptable width of the pressure band at
this moment, given the current lack of a broad-based
consensus regarding an appropriate pressure bandwidth.
Hence, we have contacted representatives of the AAMI
Sphygmomanometer Committee to express the agency’s
concerns with this methodology and to engage the com-
mittee to develop appropriate alternative testing strategies. It
is anticipated that many such contacts will need to take
place to develop a broadly supported performance standard
that is agreed to by the industry, the clinical community, and
the FDA.
In the interim, the FDA advises the manufacturer that the
use of the pressure band method in validating NIBP
monitors will be reviewed on a case-by-case basis.
4. Conclusion
The FDA attempts to use accepted consensus standards
whenever appropriate to lessen the burden of device
manufacturers in performing multiple different validation
tests to comply with differing technical standards. However,
as discussed in this article, there are instances in which the
agency cannot accept the use of a consensus standard. For
example, if provisions of a technical standard conflict with
Federal regulations, the agency is mandated to follow the
regulation. Another reason why the agency may not accept a
consensus standard is that there may be specific scientific
concerns regarding provisions of the standard. In all such
cases, the FDA provides a supplementary information sheet
to clearly specify the agency’s requirements when there are
differences with a technical standard.
The FDA understands the many advantages to be gained
in technological innovation and product development by
having the FDA’s regulatory requirements and international
consensus standards as harmonized as possible. To achieve
this goal, the FDA sends representatives to participate in
various standards development committees’ work. Although
this has been helpful in moving toward the goal of har-
monizing standards, the FDA realizes that much of the
work toward the development of standards by committee
must be determined by consensus via balloting, meaning
that there might be rare occasions in which there is dis-
cordance between the FDA’s requirements and the consen-
sus standard that develops. However, the FDA remains
committed to working toward a goal of harmonization of
standards whenever possible within the scope of the
agency’s regulations.